introduction to tomosynthesis - amos...

132
Introduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical Physics Laboratory Department of Radiology and Imaging Sciences & Winship Cancer Institute Emory University Atlanta, Georgia

Upload: lenhi

Post on 15-Apr-2018

232 views

Category:

Documents


1 download

TRANSCRIPT

Page 1: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

Introduction to Tomosynthesis

Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical Physics Laboratory

Department of Radiology and Imaging Sciences & Winship Cancer Institute

Emory University Atlanta, Georgia

Page 2: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

Disclosure

• Institutional Research Collaborations:

– Barco

– Hologic

• Consultant:

– Fuji Medical Systems USA

2

Page 3: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

Learning Objectives

• To understand the fundamental principles behind tomosynthesis

• To explain the possible different system designs

• To explain the determinants of image quality

• To list the factors that affect radiation dose

• To understand the common artifacts in tomosynthesis

3

Page 4: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

Outline

Motivation

Introduction

System Design and Design Considerations

Image Reconstruction

Radiation Dose

X-Ray Scatter

Artifacts

Synthetic Mammograms

4

Page 5: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

MOTIVATION

5

Page 6: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

6

Page 7: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

7

Page 8: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

0.16 lesion

localization

fraction

8 Vikgren et al, Radiology 249(3), 1034-1041 (2008).

Page 9: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

9 Vikgren et al, Radiology 249(3), 1034-1041 (2008).

Page 10: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

10

Page 11: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

11

Page 12: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

12

Page 13: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

29% of missed cancers were missed due to being “obscured by overlying tissue”

13 Birdwell et al, Radiology 219, 192-202 (2001).

Page 14: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

14

Page 15: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

15

Page 16: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

16

Page 17: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

17 Courtesy GE Medical Systems

Page 18: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

18

Page 19: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

Computed Tomography

More expensive Higher radiation dose 100x chest CT over chest radiograph 2-5x breast CT over mammography Metal problematic Slower to read (?) ….otherwise, fantastic!

19

Page 20: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

Is there a halfway??

(can we get the best of both worlds?)

20

Page 21: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

21

Page 22: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

Linear Tomography

22 Bushberg et al, The Essential Physics of Medical Imaging, 2nd edition.

Page 23: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

Towards Tomographic Imaging

23

Standard

Transmission

Imaging

Stereoscopic

Imaging

Digital

Tomosynthesis

Computed Tomography

2+ D 2.2 D 2 D 3 D

(If your optical system can handle it!)

(Is more always

better?)

Linear

Tomography

2.1 D

(If you plan ahead!)

Page 24: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

DIGITAL TOMOSYNTHESIS

24

Page 25: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

25

Detector

Translated X-ray source

X-ray beam

Lesions of Interest

This information is used to reconstruct the volume

Page 26: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

Shift correlates with vertical location

26

Page 27: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

Recall

Courtesy of Hologic Inc. 27

Page 28: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

CC view .IDC

Courtesy of Hologic Inc. 28

Page 29: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

Courtesy of Hologic Inc. 29

Page 30: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

Benefits

Similar to Radiography/Mammography System

Workflow

Interpretation

Dose

…but with some discrimination of vertical position!

30

Page 31: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

SYSTEM DESIGN AND DESIGN CONSIDERATIONS

31

Page 32: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

FFDM System Breast Tomo System

32

Page 33: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

33 Courtesy Joseph Lo (via youtube)

Page 34: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

34 https://www.youtube.com/watch?v=g9AjqhQJwAs

Page 35: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

35 Courtesy of Philips Digital Mammography AB

Page 36: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

System

Fuji

AMULET

Innovality

GE Essential

Hologic

Selenia

Dimensions

IMS Giotto

TOMO

Philips

MicroDose

Planmed

Nuance Excel

DBT

Siemens

MAMMOMAT

Inspiration

Detector Type

Full field -

Direct (a-Se)

(Hexagonal

pixels)

Full field -

Indirect

Full field -

Direct (a-Se)

Full field -

Direct (a-Se)

Linear Slit Scan

– Spectral

Photon

Counting (Si)

Full field -

Direct (a-Se)

Full field -

Direct (a-Se)

Detector

Motion Static Static Rotating Static

Continuous Slit

Scan

Rotating during

exposure Static

X-Ray Tube

Motion Continuous Step-and-Shoot Continuous Step-and-Shoot Continuous Continuous Continuous

Detector to

Center of

Rotation

Distance (cm)

4 4 0 2 -40 4.37 4.7

Angular Range 15 25 15 40 11 30 50

Number of

Projections 15 9 15 13 21 15 25

Scan Time (sec) 4 7 3.7 12 3 – 10 20 25

Reconstruction

Method

Modified

FBP Iterative FBP

Iterative with

Total Variation

Regularization

Iterative Iterative FBP

Development

Stage

Commercial

System**

Commercial

System

Commercial

System

Commercial

System** Prototype Prototype

Commercial

System

36 **Currently not approved for clinical use in the U.S. by the Food and Drug Administration (FDA)

Page 37: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

37 http://2014.bhpa.eu/wp-content/uploads/formidable/Marshall_Nicolas.pdf

Page 38: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

38 Courtesy of Otto Zhou, Applied Nanotechnology Laboratory, University of North Carolina at Chapel Hill

Page 39: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

CNT s-DBT

DBT

System MTF

~30% increase in system resolution for standard 15 degree, 15 view scan

A. Tucker, et al, Med Phys 2012

Page 40: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

S-DBT reconstructions above

Continuous motion DBT reconstructions

MC # 1 MC # 2 MC # 3 MC # 4 MC # 5 MC # 6

Micro-calcification visibility

Page 41: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

41 Shan et al, Phys. Med. Biol. 60, 81-101, 2015

Page 42: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

System

Fuji

AMULET

Innovality

GE Essential

Hologic

Selenia

Dimensions

IMS Giotto

TOMO

Philips

MicroDose

Planmed

Nuance Excel

DBT

Siemens

MAMMOMAT

Inspiration

Detector Type

Full field -

Direct (a-Se)

(Hexagonal

pixels)

Full field -

Indirect

Full field -

Direct (a-Se)

Full field -

Direct (a-Se)

Linear Slit Scan

– Spectral

Photon

Counting (Si)

Full field -

Direct (a-Se)

Full field -

Direct (a-Se)

Detector

Motion Static Static Rotating Static

Continuous Slit

Scan

Rotating during

exposure Static

X-Ray Tube

Motion Continuous Step-and-Shoot Continuous Step-and-Shoot Continuous Continuous Continuous

Detector to

Center of

Rotation

Distance (cm)

4 4 0 2 -40 4.37 4.7

Angular Range 15 25 15 40 11 30 50

Number of

Projections 15 9 15 13 21 15 25

Scan Time (sec) 4 7 3.7 12 3 – 10 20 25

Reconstruction

Method

Modified

FBP Iterative FBP

Iterative with

Total Variation

Regularization

Iterative Iterative FBP

Development

Stage

Commercial

System**

Commercial

System

Commercial

System

Commercial

System** Prototype Prototype

Commercial

System

42 **Currently not approved for clinical use in the U.S. by the Food and Drug Administration (FDA)

Page 43: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

Total Angular Range Oblique Incidence

43 Acciavatti and Maidment, Medical Physics, 38(11), 2011

(20 deg)

Page 44: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

Oblique Incidence – Direct Detectors

44 Zhao and Zhao, Medical Physics, 35(5), 2008

Page 45: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

Oblique Incidence – Indirect Detectors

45 Mainprize et al, Medical Physics, 33(9), 2006

Page 46: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

Acquisition Geometry

Radiography:

1 position, 1 shot

CT:

full revolution, 1000 shots

Tomosynthesis:

???

46

Page 47: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

Acquisition Geometry Optimization

Acquisition parameters:

Angular range

Number of projection angles

47

Page 48: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

48 Maidment et al, Proceedings of SPIE, 5745, 2005

Page 49: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

Artifact Spread Function

49

s BG

s 0 BG 0

I z -I zASF z =

I z -I z

Wu et al, Medical Physics, 31(9), 2004

Page 50: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

Angular Range

50 Hu et al, Medical Physics, 35(12), 2008

Page 51: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

Image Acquisition Optimization

Computer simulated breast volume and lesions

63 different acquisition geometries

In-plane quality and vertical resolution

51

Page 52: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

52 Sechopoulos and Ghetti, Medical Physics 2009, 36, 1199-1207.

(a) Mass

Masses: Increased in-plane quality with increased angular range, fewer projections

Page 53: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

53 Sechopoulos and Ghetti, Medical Physics 2009, 36, 1199-1207.

µCa: Increased in-plane quality with decreased angular range ( mammo), fewer projections (small effect)

Page 54: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

54 Sechopoulos and Ghetti, Medical Physics 2009, 36, 1199-1207.

Vertical resolution increases with angular range

Page 55: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

55 Sechopoulos and Ghetti, Medical Physics 2009, 36, 1199-1207.

Threshold number of projections to improve vertical resolution

Page 56: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

Acquisition Geometry and Vertical Resolution

56 Sechopoulos and Ghetti, Medical Physics 2009, 36, 1199-1207.

aSubstantial artifacts due to narrow angular range

Page 57: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

Subsequent studies:

Threshold number of projections for each angular range was confirmed by others

57

Tucker et al, Proc. SPIE 8313, 831307-831310 (2012) A. S. Chawla et al, Med. Phys. 36, 4859-4869 (2009) I. Reiser and R. M. Nishikawa, Med. Phys. 37, 1591-1600 (2010) Goodsitt et al, Phys. Med. Biol. 59 (2014) 5883

Page 58: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

Chest Tomosynthesis

58

No gain in increase in projections beyond a certain number

Söderman et al, Medical Physics, Vol. 42, No. 3, 2015

Page 59: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

Acquisition Geometry

↑ angular range ↑ vertical resolution

↑ # of projections ↑ vertical resolution up to a point

Have to consider:

scan time

anatomy

detector size

59

Page 60: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

ACQUISITION TECHNIQUE

60

Page 61: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

Tube Voltage Selection

Multiple studies reported higher kV than mammo optimal for tomo

One study reported lower energies beneficial

61

Ren et al, Proceedings of SPIE 5745, 550–561 (2005). Zhao et al, Proceedings of SPIE 5745, 1272–1281 (2005). Wu et al, Proceedings of SPIE 6142, 61425E (2006) Glick and Gong, Proceedings of SPIE 6142, 61421L–61429L (2006).

Page 62: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

Technique and Dosimetric Characterization of a Clinical System

62 Feng and Sechopoulos, Radiology, 2012; 263(3): 35-42

Page 63: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

Breast

Thickness

(cm)

2

3

4

5

6

7

8

Filter

Tube

Voltage

(kVp)

1st HVL

(mm Al)

Rh 25 0.453

Rh 26 0.494

Rh 28 0.517

Rh 29 0.551

Rh 31 0.567

Ag 30 0.586

Ag 32 0.611

63

Filter

Tube

Voltage

(kVp)

1st HVL

(mm Al)

Al 26 0.441

Al 28 0.476

Al 29 0.490

Al 31 0.541

Al 33 0.572

Al 35 0.600

Al 38 0.660

Tomosynthesis Mammography

Feng and Sechopoulos, Radiology, 2012; 263(3): 35-42

Page 64: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

TOMOSYNTHESIS RECONSTRUCTION

64

Page 65: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

System

Fuji

AMULET

Innovality

GE Essential

Hologic

Selenia

Dimensions

IMS Giotto

TOMO

Philips

MicroDose

Planmed

Nuance Excel

DBT

Siemens

MAMMOMAT

Inspiration

Detector Type

Full field -

Direct (a-Se)

(Hexagonal

pixels)

Full field -

Indirect

Full field -

Direct (a-Se)

Full field -

Direct (a-Se)

Linear Slit Scan

– Spectral

Photon

Counting (Si)

Full field -

Direct (a-Se)

Full field -

Direct (a-Se)

Detector

Motion Static Static Rotating Static

Continuous Slit

Scan

Rotating during

exposure Static

X-Ray Tube

Motion Continuous Step-and-Shoot Continuous Step-and-Shoot Continuous Continuous Continuous

Detector to

Center of

Rotation

Distance (cm)

4 4 0 2 -40 4.37 4.7

Angular Range 15 25 15 40 11 30 50

Number of

Projections 15 9 15 13 21 15 25

Scan Time (sec) 4 7 3.7 12 3 – 10 20 25

Reconstruction

Method

Modified

FBP Iterative FBP

Iterative with

Total Variation

Regularization

Iterative Iterative FBP

Development

Stage

Commercial

System**

Commercial

System

Commercial

System

Commercial

System** Prototype Prototype

Commercial

System

65 **Currently not approved for clinical use in the U.S. by the Food and Drug Administration (FDA)

Page 66: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

(really) Filtered Back Projection

66

Hfilter (ωy, ωz) = Hspectrum(ωy) ⋅ Hprofile(ωz) ⋅ Hinverse(ωy, ωz)

Hspectrum(ωy): Hanning filter to control noise

Hinverse(ωy, ωz): Ramp-type filter

Hprofile(ωz): Slice profile filter for constant depth resolution

Mertelmeier et al, SPIE 6142, 61420F (2006)

Page 67: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

67 Mertelmeier et al, SPIE 6142, 61420F (2006)

Page 68: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

68 Mertelmeier et al, SPIE 6142, 61420F (2006)

Page 69: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

69 Zhou et al, Medical Physics, Vol. 34, No. 3, March 2007

Page 70: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

70 Zhou et al, Medical Physics, Vol. 34, No. 3, March 2007

FBP w/ramp only + Hanning & thickness filter + Hanning & thickness filter 2

+ modified ramp SBP Iterative

Page 71: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

Iterative Reconstruction Methods

71

Guess the reconstructed

volume

Simulate the projections that would result from the guessed volume

N=N0e-ΣµT

Simulated projections of

guessed volume

Acquired projections

Compare

Adjust guess

1

i

v vp p

pi i

v v

vp

p

l E Y

l R E

Page 72: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

Other Reconstruction Methods

SIRT

SART

ART

MLEM

TVR

72

Page 73: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

73 Van de Sompel et al, Medical Image Analysis 2011, 15, 53–70

Page 74: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

Comparison of Reconstructions

Optimal acquisition might differ for different recons

Challenging for a single group to implement and optimize all recons

Most appropriate metric(s)?

74

Page 75: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

Tomosynthesis Reconstruction

Mono-energetic Assumption Or constant spectral beam

No explicit definition

Same case with CT reconstruction algorithms

75

Page 76: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

Standard Tomosynthesis Spectrum

0.0

0.5

1.0

1.5

2.0

2.5

3.0

3.5

0.00

0.01

0.02

0.03

0.04

0.05

0.06

0 5 10 15 20 25 30 35 40

Lin

ear

Att

en

uat

ion

Co

eff

icie

nt

(cm

-1)

No

rmal

ize

d S

pe

ctru

m

X-Ray Energy (keV)

W/Al 32 kVp

W/Al 32 kVp + 6 cm of breast tissue

µ Breast Tissue (cm^-1)

µ = 21.7 keV

µ = 25.2 keV

76

Page 77: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

Breast Tomosynthesis Acquisition Model

biθ acquired signal at pixel i for projection θ

ψ(ε) incident energy fluence at energy ε

µ(x,ε) linear attenuation coefficient of voxel x at energy ε

L θ,i line from source to pixel i for projection θ

,

,

L i

x dl

ib e d

77

Page 78: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

Breast Tomosynthesis Acquisition Model

Minimize Poisson likelihood :

Using iterative gradient descent optimization method

arg min

log

MLE

i i i

L

L b b b

X X

X

78 Sechopoulos et al, European Congress of Radiology, 2013

Page 79: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

Homogeneous Phantom + Masses

79

0% 80%

60% 40% 20%

100%

0% 80%

60% 40% 20%

100%

0% 80%

60% 40% 20%

100%

FBP MLEM Spectral

Page 80: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

Homogeneous Phantom + Masses

80

-2

0

2

4

6

8

10

12

14

0 20 40 60 80 100

SDN

R

Lesion Glandular Density (%)

Spectral

MLEM

FBP

Page 81: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

Homogeneous Phantom + Microcalcifications

81

FBP Spectral

Page 82: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

Homogeneous Phantom + Microcalcifications

82

FBP Spectral

Page 83: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

RADIATION DOSE

83

Page 84: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

Breast Tomosynthesis Dosimetry Model

• Mammography:

• Tomosynthesis:

84

MAMMO g MAMMOAGD D N AK

MAX

MIN

TOMO g MAMMO

RGD

AGD D N AKN

AAPM Task Group Report 234

Page 85: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

Projection Angle (deg)

-30 -20 -10 0 10 20 30

RG

D(

)

0.0

0.2

0.4

0.6

0.8

1.0

1.2

1%

25%

50%

75%

100%

Projection Angle (deg)

-30 -20 -10 0 10 20 30

RG

D(

)

0.0

0.2

0.4

0.6

0.8

1.0

1.2

Mo/Mo 25 kVp

Mo/Mo 27 kVp

Mo/Rh 29 kVp

Rh/Rh 31 kVp

Rh/Rh 35 kVp(a) (b)

Projection Angle (deg)

-30 -20 -10 0 10 20 30

RG

D(

)

0.0

0.2

0.4

0.6

0.8

1.0

1.2

1.4

7 cm

10 cm

13 cm

16 cm

19 cm

Projection Angle (deg)

-30 -20 -10 0 10 20 30

RG

D(

)

0.0

0.2

0.4

0.6

0.8

1.0

1.2

2 cm

3 cm

4 cm

5 cm

6 cm

7 cm

8 cm(c) (d)

(a) Glandular fraction (d) X-ray spectrum

(c) Chest wall to nipple distance

(b) Thickness

Relative Glandular Dose

85 Sechopoulos et al, Med Phys, 2007; 3(1): 221-232

Page 86: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

Mammography and Tomosynthesis Dose

86 Feng and Sechopoulos, Radiology, 2012, 263(1): 35-42

Page 87: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

AGD Ratio of Tomo / Mammo

Breast Thickness

(cm)

Glandular Density (%)

1 14.3 25 50 2 2.47 2.34 1.87 1.78 3 2.40 1.94 1.49 1.39 4 2.66 2.11 1.84 1.28 5 2.37 1.90 1.53 1.08 6 1.91 1.83 1.94 1.25 7 2.26 1.75 1.38 1.12 8 2.13 1.85 1.46 1.16

87 Feng and Sechopoulos, Radiology, 2012, 263(1): 35-42

Page 88: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

Mean AGD [mGy] GE SenoClaire Essential

Breast Thickness N Mammo Tomo

All 236 1.62 1.49

<40 mm 28 1.13 1.14

41-50 mm 46 1.34 1.33

51-60 mm 74 1.48 1.41

61-70 mm 55 1.82 1.62

>70 mm 33 2.39 1.98

88 Paulis et al, Investigative Radiology, online ahead of print

Page 89: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

Chest Imaging Effective Dose [mSv]

2-View CXR Chest Tomo Chest CT

0.056 0.124 7

89 Sabol, Medical Physics, Vol. 36, No. 12, 2009

Page 90: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

90 Detector

X-ray beam

x mAs

x mAs x mAs x mAs

x mAs

Does the exposure distribution have to be uniform?

Page 91: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

91 Detector

X-ray beam

x mAs

x mAs y mAs x mAs

x mAs

How about:

y < x ? y > x ?

Page 92: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

92 Detector

X-ray beam

x mAs x mAs

y mAs x mAs

x mAs

Or even:

y < x ? y > x ?

Page 93: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

93 Nishikawa, Reiser et al, Proceedings of SPIE 6510, 65103C–65108C (2007).

Mammo Tomo Proposal

Page 94: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

μCa detectability: center projection < single center slice of reconstruction

Mass detectability: no statistically significant difference

94 Das et al, Medical Physics 2009, 36(6), 2009

Page 95: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

95

Standard Tomo Variable 7 central / 18 total proj

Variable 5 central / 20 total proj

Hu and Zhao, Med. Phys. 38(5), 2455–2466 (2011).

Page 96: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

Uneven distribution of exposure and non-uniform angular sampling used by one commercial manufacturer in systems outside the US.

96

Page 97: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

What if??

Improved image quality?

Dose reduction?

Single-pass contrast enhanced imaging?

97

x kVp

y kVp x kVp y kVp

x kVp

Page 98: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

Dual Spectrum Single Pass Tomo

98

AEC kVp

49 kVp + Cu AEC kVp 49 kVp + Cu

AEC kVp

Sechopoulos et al, European Congress of Radiology, 2015

Page 99: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

8 cm Homogeneous Phantom + Masses

AEC (38 kVp, 84 mAs) AEC + 49 kVp/0.254 mm Cu

99 Sechopoulos et al, European Congress of Radiology, 2015

Page 100: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

100

p = 0.412826

Sechopoulos et al, European Congress of Radiology, 2015

Page 101: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

101

p = 0.232631

Sechopoulos et al, European Congress of Radiology, 2015

Page 102: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

Results

Thickness SDNR Difference Dose

5 cm -16.0 ± 9.25% (p>0.08)

-48%

8 cm -3.2 ± 19.9% (p>0.41)

-28%

102 Sechopoulos et al, European Congress of Radiology, 2015

Page 103: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

X-RAY SCATTER

103

Page 104: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

Effect of Scatter in Tomosynthesis

104 Wu et al, Proc SPIE, 2007

Page 105: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

Linear Grid

Cellular Grid

Radio-transparent

Material

Radio-opaque Material

Radio-opaque Material

Air

Incident x-rays

Imaged object

Scattered x-rays

Scatter grid

Primary-only x-rays

Detector

How is Scatter Normally Dealt With?

105

A clinical grid transmits ~80% of primary and ~20% of scatter x-rays

Page 106: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

Grids in Tomosynthesis

Cut-off at higher projection angles

Primary photon absorption

Prone to image artifacts

106

Page 107: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

GE SenoClaire Essential

Uses anti-scatter grid for DBT acquisition

Septa perpendicular to standard position

High number of lines per unit length

107

Page 108: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

Alternatives

Post-acquisition processing

Correction during reconstruction

108

Page 109: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

TOMOSYNTHESIS ARTIFACTS

109

Page 110: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

110 Wu et al, Medical Physics. 33(7), 2461–2471 (2006).

High contrast off-plane objects introduce artifacts “Voting” strategy to identify projections in which appropriate information is included, others ignored Especially important for acquisitions with low number of projections

Page 111: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

111 Zhang et al, Med. Phys. 34(9), 3603–3613 (2007)

“Mask” to reconstruct only inside the breast Faster reconstruction Avoids artifacts outside breast

Page 112: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

112

Breast tissue outside wide projection FOV

Breast tissue outside reconstructed volume

but that contributes to attenuation

Reconstructed volume

Sechopoulos, Medical Physics, Vol. 40, No. 1, 2013

Page 113: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

113 Zhang et al, J. Comput. Assist. Tomogr. 33(3), 426–435 (2009).

Varying number of projections contributes to the volume update, introducing discontinuities: introduced equalization using neighbor values updated by previous projection Bright artifact due to tissue attenuation outside volume: assume extension of “average” breast tissue outside of field of view to avoid bright artifact

Page 114: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

114

Uncorrected Previous Improved

Lu et al, Proceedings of the 11th IWDM 2012, pp. 745–752.

Improved estimation of x-ray path length in tissue outside field of view

Page 115: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

SYNTHETIC MAMMOGRAMS

115

Page 116: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

116

Mammogram Orig. Synthetic Tomo Slice

Gur et al, Academic Radiology, Vol 19, No 2, 2012

Page 117: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

Recall Rates

117

DBT + FFDM DBT + Synthetic

False Positive

Rate

% Detected Cancers

False Positive

Rate

% Detected Cancers

1st Generation 53.1 83.5 46.1 77.7

2nd Generation 45.6 87.3 45.2 85.5

Skaane et al, Radiology, Vol 271(3), 2014

Page 118: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

Synthetic Mammograms

Included in various commercial systems

118

Page 119: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

Current Research

CADe and CADx for tomosynthesis

Need to lower reading time

Contrast enhanced tomosynthesis

Phase contrast tomosynthesis

Tomosynthesis elastography

119

Page 120: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

Multimodality Imaging

Tomosynthesis

+ US

+ SPECT

+ Electrical Impedance

+ Optical

120

Page 121: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

Summary

Fast digital detectors Advanced imaging

Need to lower anatomic noise

Tomosynthesis similar to planar radiography

System footprint

Workflow

Image interpretation

121

Page 122: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

Summary

Acquisition geometry large impact on image quality

Dosimetry > but similar to planar radiography

Ongoing research in:

Reconstruction algorithms

Other techniques (enhanced, phase, etc.)

Multimodality

122

Page 123: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

QUESTIONS

123

Page 124: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

In terms of image acquisition, what is the main difference between linear tomography and digital tomosynthesis?

1. Linear tomography acquisition takes substantially longer than digital tomosynthesis.

2. Linear tomography results in a single plane being in focus per acquisition while in digital tomosynthesis any number of planes can be reconstructed to be in focus.

3. Linear tomography results in circular images while digital tomosynthesis results in rectangular images.

4. In linear tomography the x-ray tube moves in a straight line while in digital tomosynthesis it moves in a circle.

1. 2. 3. 4.

0%

13%13%

75%

Page 125: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

In terms of image acquisition, what is the main difference between linear tomography and

digital tomosynthesis?

(2) In linear tomography, the in-focus plane has to be selected before acquisition, while in digital tomosynthesis, the reconstruction of the acquired projections results in many planes being in focus.

125

Bushberg et al, “The Essential Physics of Medical Imaging”, 3rd Edition

Page 126: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

How does increasing the angular range of the tomosynthesis scan affect the vertical resolution of the

reconstructed image?

1. Consistently decreases

2. Decreases up to a point, then remains constant

3. Vertical resolution does not increase with increasing angular range

4. Increases up to a point, then remains constant

5. Consistently increases 1. 2. 3. 4. 5.

50% 50%

0%0%0%

Page 127: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

How does increasing the angular range of the tomosynthesis scan affect the vertical resolution

of the reconstructed image?

(5) Increasing the angular range covered by the swing of the x-ray source consistently increases the vertical resolution of the reconstructed tomosynthesis image.

127 Sechopoulos and Ghetti, Medical Physics 2009, 36, 1199-1207.

Page 128: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

How does increasing the number of projections of the tomosynthesis scan affect the vertical resolution of the

reconstructed image?

1. Consistently decreases

2. Decreases up to a point, then remains constant

3. Vertical resolution does not increase with increasing number of projections

4. Increases up to a point, then remains constant

5. Consistently increases 10

1. 2. 3. 4. 5.

0% 0% 0%0%0%

Page 129: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

How does increasing the number of projections of the tomosynthesis scan affect the vertical

resolution of the reconstructed image?

(4) Increasing the number of projections during a tomosynthesis acquisition increases the vertical resolution up to a certain threshold, beyond which the resolution remains constant unless the angular range is increased.

129 Sechopoulos and Ghetti, Medical Physics 2009, 36, 1199-1207.

Page 130: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

Why is there a high interest in chest tomosynthesis?

10

0%

0%

0%

0%

0% 1. Much lower dose than chest radiography with out-of-plane blurring

2. Much lower dose than chest CT with out-of-plane blurring

3. Same vertical resolution as chest CT but lower dose

4. Improved vertical resolution as chest CT at same dose

5. Same vertical resolution and dose as chest CT, but considerably cheaper

Page 131: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

Why is there a high interest in chest tomosynthesis?

(2) Chest tomosynthesis involves, in general, higher dose than chest radiography, but considerably lower than chest CT. Although it doesn’t result in true tomographic images as chest CT, it provides enough vertical resolution to be superior to chest radiography and sufficient for some clinical applications.

131

Sabol, Medical Physics, Vol. 36, No. 12, 2009 Dobbins and McAdams, European Journal of Radiology 72 (2009) 244–251

Page 132: Introduction to Tomosynthesis - AMOS Onlineamos3.aapm.org/abstracts/pdf/99-28889-359478-110669.pdfIntroduction to Tomosynthesis Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical

Introduction to Tomosynthesis

Ioannis Sechopoulos, Ph.D., DABR Diagnostic Medical Physics Laboratory

Department of Radiology and Imaging Sciences & Winship Cancer Institute

Emory University Atlanta, Georgia