introduction to the new agilent 7900 - alfresco.ubm...
TRANSCRIPT
Introduction to the New Agilent 7900 Redefining ICP-MS Performance
Spectroscopy Webinar
February 2014
Presenter: Ed McCurdy – ICP-MS Product
Marketing, Agilent Technologies
Agilent 7900 ICP-MS Introduction - Agenda
• ICP-MS Market and Agilent Technologies ICP-MS
• Introduction to the new Agilent 7900
- Development
- Key Performance Illustrations
- Unique new Hardware Features
- New MassHunter 4.1 software
- Support, Maintenance and User Training
• Q&A
Agilent 7900 ICP-MS Introduction - Agenda
Or:
• What it is
• Where it comes from
• What it does
• How it works
• What it’s like to use
Agilent’s History of Innovation in ICP-MS The first 25 years - 1987 to 2012
1987 2000 2009 1994
First computer-controlled ICP-MS
PMS series
4500
7500
Enabling high
sensitivity
metal analysis
Enabling routine
robust ICP-MS
analysis
Enabling a new level
of interference
management
First benchtop ICP-MS
Enabling a new
level of ease of
use in ICP-MS
First effective He mode collision/reaction cell
First high-matrix (HMI) ICP-MS
Enabling a new era
in ICP-MS analysis
8800 ICP-QQQ
2012
World’s First ICP-QQQ
February 2014
4
7700
ICP-MS Market Summary ICP-MS now a mainstream analytical technique in many regions
Worldwide ICP-MS market was estimated at ~$275M in 2013
(~1700 instruments)
ICP-MS has moved beyond the research lab and been adopted
for routine trace metals analysis in high throughput labs
For many of these labs, the key analytical requirements are:
- Low detection limits (requires high sensitivity and low background)
- Robustness (tolerance of complex or difficult samples)
- Accuracy (freedom from interferences)
- Dynamic range (ability to measure high & low concentrations in 1 run)
- Productivity (sample run time; fewer sample reruns)
- Ease of use (quick training of new/occasional users)
- Flexibility (ability to handle a wide range of sample types)
- Low Detection Limits – (requires high sensitivity and low background)
- Robustness (tolerance of complex or difficult samples)
- Accuracy (freedom from interferences)
- Dynamic range (ability to measure high & low concentrations in 1 run)
- Productivity (sample run time; fewer sample reruns)
- Ease of use (quick training of new/occasional users)
- Flexibility (ability to handle a wide range of sample types)
So How are ICP-MS Manufacturers Responding? Current Technology to Address Routine Labs’ Performance Needs
- Low Detection Limits – (requires high sensitivity and low background)
- Robustness (tolerance of complex or difficult samples)
- Accuracy (freedom from interferences)
- Dynamic range (ability to measure high & low concentrations in 1 run)
- Productivity (sample run time; fewer sample reruns)
- Ease of use (quick training of new/occasional users)
- Flexibility (ability to handle a wide range of sample types)
Low Detection Limits – Sufficient for ppt level analysis
Accuracy – He mode provides simple removal of most
polyatomic overlaps in most sample types
Flexibility – interfaces for organic solvents, aggressive
acids, HF, small samples volumes; easy coupling to
alternative sample intro devices (LC, GC, FFF, Laser
Ablation, etc.)
Existing instruments (Agilent 7700) already addressed many of these requirements:
Agilent’s ICP-MS Product Development Focus Recent ICP-MS Launches
7700 Series quadrupole ICP-MS – launched 2009 • Unrivalled He mode performance for multi-element interference removal • Unmatched matrix tolerance with HMI (up to 2% TDS) • 9 Orders dynamic range at the detector
8800 triple quadrupole ICP-MS – launched 2012 • Unique QQQ configuration allows operation in MS/MS mode • MS/MS mode provides the only reliable way to remove interferences in complex
or variable samples using reactive cell gases
8800 ICP-QQQ Won Four Industry Awards
Within a Year of Launch Divergence between:
• High-throughput, more routine analysis
(quadrupole ICP-MS with He mode)
• Ultra-flexible, ultra-high performance for
advanced applications, semicon, materials,
ultra-low DLs (ICP-QQQ in reaction mode)
Let’s see what else we’ve been working on
since the 7700 was launched…
7700 had:
Better matrix tolerance than any other ICP-MS
• More robust plasma (lower CeO/Ce ratio) than any other system under standard tuning conditions , plus HMI for routine analysis of % level dissolved solids
Best interference removal with Helium cell gas – eliminates need for reaction gases in all common applications
• 7700 ORS3 improvements - removes all polyatomics in He mode, giving accurate results in complex or variable sample types – impossible on ICP-MS systems that use reactive cell gases or mixtures
Wider dynamic range than any other quadrupole ICP-MS
• Full 9 orders dynamic range at the detector – linear to 500ppm without changing conditions or hardware
Why Introduce another new ICP-MS? Building on the success of the Agilent 7700 Series:
Low Detection Limits – (requires high sensitivity and low background)
Robustness (tolerance of complex or difficult samples)
Accuracy (freedom from interferences)
Dynamic range (ability to measure high & low concentrations in 1 run)
Productivity (sample run time; fewer sample reruns)
Ease of use (quick training of new/occasional users)
Flexibility (ability to handle a wide range of sample types)
Low Detection Limits – (requires high sensitivity and low background)
Robustness (tolerance of complex or difficult samples)
Accuracy (freedom from interferences)
Dynamic range (ability to measure high & low concentrations in 1 run)
Productivity (sample run time; fewer sample reruns)
Ease of use (quick training of new/occasional users)
Flexibility (ability to handle a wide range of sample types)
Areas Where Even Better Performance was Needed Focus for next generation ICP-MS after 7700
Robustness – Nominal limit of 0.2% (2000ppm) total dissolved solids (2% with
HMI) – not sufficient for routine direct analysis of some sample types
Dynamic range – sub-ppt DLs adequate at the low end, but over-range at a few
100ppm – not sufficient for majors and traces in same run
Productivity – Faster analysis needed: 60 seconds per sample or less, with
optimum cell gas mode for all elements
Ease of use – Simpler user interface, intelligent method setup and user training
Requirements to improve on existing performance and extend ICP-MS scope into
new applications & sample types:
Presenting Agilent’s Game-Changing 7900 ICP-MS
We took the world’s best-
selling, highest performing
quadrupole ICP-MS, and
made it 10x better!
Agilent 7900 ICP-MS Key Performance Gains “10x better performance” than the Agilent 7700 ICP-MS
10x higher matrix tolerance – handles even tougher samples than the 7700/HMI - Patented HMI is still unique to Agilent. On the Agilent 7900, the optional ultra-HMI (UHMI)
extends capability to matrix levels of up to 25% TDS
10x wider dynamic range – increases upper measurement limit - 7700 had 9 orders (class-leading). 7900 extends this by at least an order of magnitude (up
to 11 orders measurement range), allowing % levels to be quantified - a first for ICP-MS
10x better signal to noise – lower Detection Limits - Higher ion transmission lens and interface, with orthogonal detector for low background
Improved productivity – faster analysis even when switching cell gases - New ultra fast ORS4 with less than 3 seconds switching time between modes
- New ISIS-3 for fast unattended start-up, autotune and sample delivery; full EPA-6020 analysis with 2 gas modes for optimum measurement in <1 minute per sample
30x faster detector –faster transient signal measurement (TRA) - 0.1ms integration time means more flexibility in single nanoparticle analysis
Easier to use – new software user interface, plus advanced usability tools - Method Wizard and remote monitor/control from tablet or Smartphone
New 7900 ICP-MS; New Technology
Components retained from 7700 Series:
• RF generator
• Sampling cone and retaining ring
• Quadrupole (still the only hyperbolic profile quad in ICP-MS)
• Agilent Mass Flow Controller (AMFC) gas control module
• Turbo pump
… That’s it!!
Everything else is new or re-engineered for the 7900!
You may think (or be told!) that Agilent 7900 as “just a facelift”
In fact it’s almost completely new!
How Much is Really New?? Easier to say “what did we keep from the 7700?”
New 7900 ICP-MS Performance Highlights
February 2014
14
Illustrations of Performance Improvements
• Robustness (matrix tolerance)
• Dynamic Range
• Productivity
- Faster discrete sampling
- Faster cell gas switching
• Ease of Use
- New software UI and features
- User training; routine maintenance
Introducing the Ground-Breaking Agilent UHMI
UHMI – much more than
just a simple T-piece
UHMI uses optimized gas mixing
geometry and sophisticated
plasma/gas-flow tuning algorithm
to set reproducible conditions for
predictable aerosol dilution rate
UHMI gas port
NEW 7900 Ultra High Matrix Introduction (UHMI) What’s different compared to HMI on 7700 and 8800?
Increased dilution range to x100 – even higher matrix capability
Less matrix loading to interface, so better long-term stability
Maintain high carrier gas flow through spray chamber, so faster
gas replacement and washout
7700 HMI New UHMI
HMI-4 (HMI-L) 0.6 L/min 0.8 L/min
HMI-8 (HMI-M) 0.35 0.68
HMI-25 (HMI-H) 0.23 0.5
HMI-50 N.A. 0.4
HMI-100 N.A. 0.33
UHMI
Dilution
Gas Port
Test of Real-World Matrix Tolerance with UHMI
• 7900 with UHMI autotuned as normal
• Argon gas humidifier used – normal for high salt matrices
• Multi-element calibration in simple aqueous standards
• Variable NaCl matrices were then run, each spiked with multi-
element QC spike
- Check recovery of spike level in each NaCl matrix
0
5
10
15
20
25
30
35
40
45
50
spike 0%
Spike Recovery at 0% NaCl (first point is true spike amount)
75 As [ 25 ppb ] 114 Cd [ 50 ppb ] 208 Pb [ 50 ppb ] 201 Hg [ 1 ppb ]
0g / 100ml = 0%
Demonstration of UHMI Performance NaCl matrix analysis with calibration against simple aqueous standards
Data supplied by Wim Proper, Eurofins Analytico, NL
0
5
10
15
20
25
30
35
40
45
50
spike 0% 0.5% 1% 1.5% 2% 5% 10% 25%
Spike Recovery at 0, 0.5, 1, 1.5, 2, 5, 10 and 25% NaCl
75 As [ 25 ppb ] 114 Cd [ 50 ppb ] 208 Pb [ 50 ppb ] 201 Hg [ 1 ppb ]
25g / 100ml = 25%
“Big Four” Toxic Elements in Variable NaCl Matrices – 25% is 125 times the recommended maximum for typical (non-HMI) ICP-MS
Data supplied by Wim Proper, Eurofins Analytico, NL
NaCl Amount 75 As [ 25 ppb ] 114 Cd [ 50 ppb ] 208 Pb [ 50 ppb ] 201 Hg [ 1 ppb ]
0% 26.9 49.2 49.7 0.85
0.5% 24.2 49.0 50.1 0.99
1% 24.8 51.5 50.2 0.93
1.5% 25.5 50.0 50.5 0.88
2% 24.6 50.0 49.7 1.03
5% 25.4 48.7 50.7 0.89
10% 22.8 46.1 49.8 0.91
25% 26.2 45.4 49.0 0.96
Average 25.1 48.7 50.0 0.93
% Recovery 100% 97% 100% 93%
% RSD 5% 4% 1% 6%
0.5g / 100ml = 0.5% 1g / 100ml = 1% 1.5g / 100ml = 1.5% 2g / 100ml = 2% 5g / 100ml = 5% 10g / 100ml = 10% 25g / 100ml = 25% 0g / 100ml = 0%
“Big Four” Spiked into Different Salt Matrices
Data supplied by Wim Proper, Eurofins Analytico, NL
Interfered Elements in Variable NaCl Matrices V-51 (ClO), Cr-52 (ClOH), Ni-60 (NaCl) and Cu-63 (ArNa)
0
10
20
30
40
50
60
70
spike 0%
Spike Recovery at 0% NaCl
51 V [ 50 ppb] 52 Cr [ 50 ppb ] 60 Ni [ 50 ppb ] 63 Cu [ 50 ppb ]
0g / 100ml = 0%
Data supplied by Wim Proper, Eurofins Analytico, NL
0
10
20
30
40
50
60
70
spike 0% 0.5% 1% 1.5% 2% 5% 10% 25%
Spike Recovery at 0, 0.5, 1, 1.5, 2, 5, 10 and 25% NaCl
51 V [ 50 ppb] 52 Cr [ 50 ppb ] 60 Ni [ 50 ppb ] 63 Cu [ 50 ppb ]
25g / 100ml = 25%
Interfered Elements in Variable NaCl Matrices – 25% is 125 times the recommended maximum for typical (non-HMI) ICP-MS
Data supplied by Wim Proper, Eurofins Analytico, NL
NaCl Amount 51 V [ 50 ppb ] 52 Cr [ 50 ppb ] 60 Ni [ 50 ppb ] 63 Cu [ 50 ppb ]
0% 49.2 49.1 49.9 49.6
0.5% 47.3 50.3 48.3 48.6
1% 49.5 49.3 48.8 48.8
1.5% 50.5 50.3 49.9 49.6
2% 49.7 49.1 49.4 48.7
5% 48.9 50.3 47.2 47.7
10% 47.8 50.3 46.3 47.7
25% 48.0 48.7 50.9 50.5
Average 48.9 49.7 48.8 48.9
% Recovery 98% 99% 98% 98%
% RSD 2% 1% 3% 2%
0.5g / 100ml = 0.5% 1g / 100ml = 1% 1.5g / 100ml = 1.5% 2g / 100ml = 2% 5g / 100ml = 5% 10g / 100ml = 10% 25g / 100ml = 25% 0g / 100ml = 0%
Interfered Elements Spiked into Different Salt Matrices
Data supplied by Wim Proper, Eurofins Analytico, NL
Performance Highlights
• Robustness (matrix tolerance)
• Dynamic Range
• Productivity
- Faster discrete sampling
- Faster cell gas switching
• Ease of Use
- New software UI and features
- User training; routine maintenance
Far Wider Measurement Range Than Any Other ICP-MS 11 orders - low and high level calibrations in a single run
Cd (1ppt - 1ppb) and Na (100ppb - 10,000ppm (1%)) in the same run
Concentration range (11 orders)
and upper measurement limit
(>1%) are at least 10x better
than any other ICP-MS
Both calibrations are linear.
Total concentration range
covered from Cd blank (BEC
of <0.1ppt) to Na top
standard (1%) is 11 orders
NEW 7900 Orthogonal Detector System Improved signal to noise
Higher sensitivity
• High sensitivity EM (increase secondary electron generation by higher
voltage at the 1st dynode)
Lower background
• Off-axis from Q-pole to Detector
Improved S/N (average 10x better than 7700 ICP-MS)
• Reduced noise on pulse signal
• New advanced discriminator system to identify and separate noise
Agilent 7900 Capable of Very High Signal/Noise High sensitivity combined with low background (and noise)
Tuned like “typical” ICP-MS –
CeO/Ce <2.5%
Uranium calibration in No Gas
mode:
Ultra-high sensitivity
1.38 GHz/ppm
Ultra-low background (1cps):
DL: 1.3ppq; BEC: 0.48ppq
Signal (Mcps/ppm) Background (cps) SBR
Example 1 300 1 300
Example 2 500 2 250
Example 3 1000 5 200
Importance of background in
ultra-trace level measurements
Note in most “normal” samples
the background is limited by
contamination
0.1 msec This enables the
measurement of the
single NP peak signal
arriving on the detector.
30 nm Au Nanoparticle
February 2014
29
Minimum dwell time for TRA acquisition is shortened to 0.1 msec on the 7900
to allow faster sampling of transient signals.
NEW 7900 Orthogonal Detector System Faster TRA measurement of fast transient signals
Performance Highlights
• Robustness (matrix tolerance)
• Dynamic Range
• Productivity
- Faster discrete sampling
- Faster cell gas switching
• Ease of Use
- New software UI and features
- User training; routine maintenance
New Integrated Sample Introduction System (ISIS 3) Fully compliant multi-mode EPA 6020 analysis* in <1 minute
7 port valve (incl.
online
ISTD port)
Piston
pump
3-way
valve
New features in ISIS 3
• Close-coupled valve – very short tube length so minimal stabilization/rinse delay
• Piston pump for faster sample uptake
• 3-way valve to switch between on-line ISTD or tune solution
• ISIS is now compatible with Startup auto-optimization functions and full autotune
* EPA 6020 includes 23 analytes, plus 8 recommended ISTDs, so up to ~40 analytes (2 gas modes gives optimum data in terms of sensitivity and interference removal)
New Integrated Sample Introduction System (ISIS 3)
Increase sample throughput: 30% faster
26+ elements analysis in soils, etc. (e.g. EPA 6020):
• 7700+ISIS 2: ~75 sec.
• 7900+ISIS 3: <60 sec.
February 2014
32
7700 H2: 6.0
0 mL/min
10 sec.
7900 H2: 6.0
0 mL/min
2 sec.
Fast Cell Gas Switching – ORS4
Users confirm they can set
0 seconds stabilization
time for switch between no-
gas, He and HE He modes
Performance Highlights
• Robustness (matrix tolerance)
• Dynamic Range
• Productivity
- Faster discrete sampling
- Faster cell gas switching
• Ease of Use
- New software UI and features
- User training; routine maintenance
MassHunter 4.1: Simpler and More Powerful Software Dashboard with Gadgets replaces old mixed UI concept
Instrument status monitor (right) can be
displayed on top of DA window, to show status
and access top level functions without the need
to open entire Top Level Application.
• Gadget icons are “live”; change appearance
depending on current status
• Also have short pull-down menus for quick
access to most common functions
Software so Powerful it Can Write Your Methods! A new era in simple method setup and ease of use
Method Wizard – Develops a complete method in three steps!
1. Select pre-set method template and choose matrix level
2. Confirm analytes and internal standards…
Software so Powerful it Can Write Your Methods! A new era in simple method setup and ease of use
Method Wizard – Develops a complete method in three steps!
3. Choose whether to optimize method for speed or DLs
…and click “Optimize”
Select “Speed” or “Low DL”
The new batch is ready to run It can be edited or save as a template for future use
February 2014
38
unit DL BEC
Speed Low DL Speed Low DL
Be ppt 2.7 0.3 0.5 0.2
V ppt 1.4 0.0 0.3 0.0
Cr ppt 4.2 1.7 27.8 22.8
Co ppt 1.4 0.4 5.3 5.5
Cu ppt 3.7 2.2 9.3 8.8
As ppt 7.2 0.8 1.4 0.6
Mo ppt 0.8 0.3 0.7 0.4
Ag ppt 1.3 0.7 6.8 5.8
Cd ppt 0.0 0.0 0.0 0.1
Sb ppt 0.2 0.1 0.3 0.2
Ba ppt 7.4 1.8 21.7 17.6
Pb ppt 1.9 0.5 3.6 3.0
U ppt 0.0 0.0 0.0 0.0
Acquisition time
Speed: 2.4 min
Low DL: 5.7 min
Remote Monitor App (iOS and Android devices) View instrument status and perform basic system control
An Industry First!
• View instrument, queue and error status
• Ignite/extinguish plasma
• Pause/resume Queue
Performance Highlights
• Robustness (matrix tolerance)
• Dynamic Range
• Productivity
- Faster discrete sampling
- Faster cell gas switching
• Ease of Use
- New software UI and features
- User training; routine maintenance
Improving Installation & Familiarization Installation
• Intuitive Installation Checkout
• Site Prep tool for Software Installation
Familiarization tools
• Familiarization Tutorials
- Familiarization Guide
- Familiarization Video
- Familiarization Slide Set
Plus:
Remote Advisor Now available
for ICP-MS!
Familiarization Tutorial – Over 20 Video Clips Videos show key theory and detailed operation
New Maintenance Intervals Redefined based on actual 7700 user maintenance periods
Item Current (7700) New (7900)
Ar gas filter Replace – 6 months Replace – as needed
Foreline Pump Oil Replace – 3 months Replace – 6 months
Shield Plate Clean – 1 month Replace – as needed
Extraction Lens, Omega Lens Clean – 3 ~ 6 months Clean – as needed
Cell Entrance Lens Clean – 6 months Clean – as needed
Plate Bias Lens Clean – 6 months Clean – as needed
Octopole Replace – 12 months Replace – as needed
Deflect Lens Clean – 6 months Clean – as needed
EM Replace – as needed Replace – as needed
43
Reviewed and agreed by R&D,
Support, and Marketing
• Simpler maintenance
• Less downtime and cost
Summary – New Agilent 7900 ICP-MS Better customer experience
Better Analytical performance experience
Ultra high matrix tolerance
Superior sensitivity and lower background noise
Wider dynamic range
New Productivity Option (ISIS 3)
Ultra fast scan speed for Single Nanoparticle analysis
Better Software experience
ICP-MS MassHunter 4.1
Method Wizard
Mobile device support
Better Support experience
Familiarization Tutorials/Videos
Remote Advisor support
March 26, 2014 Page 45
Q&A Questions Unanswered at Live Event
Question Response Why is HMI any better than an online auto-dilution system? HMI is in fact a dilution technique. The main difference is that it dilutes the sample aerosol, not the bulk liquid. This has several advantages, but the main
ones are that there are no sample handling steps prior to analysis, so no contamination or dilution errors. And compared to an automatic on-line dilution
system, the main benefits of HMI are the simplicity (no tubing connectors to leak or need maintenance), and the flexibility. With HMI you can run the
sample once and using different tune steps you can measure it at several different dilutions, all from a single visit to the sample. it make method
development really quick and easy. You said that triple quad was the only reliable way to measure
interfered elements when using reactive cell gases. Why? QQQ allows operation in MS/MS mode, which is where the first quadrupole (the one before the cell) operates as a unit mass filter. This means that only
the analyte mass and any on-mass interferences enter the cell, which means the reaction chemistry is really well controlled. In practice this means that
you don't get any cell-formed reaction product ions giving you new interferences on other analytes, and any analyte that you measure as a product ion
itself (such as As-O at m/z 91) is not going to suffer any interferences from another analyte already at the product ion mass (such as Zr91) How often do you calibrate the detectors ? There are 2 sides to this question. The first is how often do you need to calibrate the detector (to ensure linearity between the pulse-count and analog
ranges, so-called P/A Factor calibration). The answer to that would be about once a week if you are routinely working across both detector ranges (ie
calibrating and measuring across the two ranges). But the second part is how often do users typically do the calibration in practice, and that might be
much more frequently because P/A factor calibration can be set to run automatically as part of the system Startup process every time the plasma is lit.
That way you can be sure that it is always calibrated and you don't need to remember to run it as a regular tuning action. Li 6 as internal std is giving us some problems what else can we
use for the low mass calibration? The choice of a really low mass ISTD element can be quite limited because most of the elements in that mass region are either matrix elements or
required analytes. However, if you optimize the instrument to give really robust operating conditions, the mass bias and ionization effects should be quite
limited, and so a higher mass ISTD such as Co or Ge will still correct reasonably well for low mass analytes. The available ISTD elements will depend on
your samples and your method. How long does it take for the vacuum pressure to be back up
after a power interruption? On the Agilent ICP-MS systems the vacuum automatically starts up again after the power is restored, so you just have to wait the normal warmup period
of about 10 minutes. Even if the power is off for a long time, the vacuum in the high vacuum region only takes a few minutes to pump down. It's only a
little longer (maybe 20 minutes) if the vacuum system has been opened for maintenance. Do you have any data that indicates how fast the system rinses
out between two different sample types ? Do you need to clean
the sample interface between different sample types ?
The required wash out time will depend on two things 1) how different the two sample types are and 2) how low you need to measure the analytes in the
second sample type. There are some obvious "worst-case" scenarios such as a lab that measures pure Co and pure Ni and needs to measure each of
the matrix elements as a trace contaminant in the other. In those cases, it would take a very long time to wash out from 1000's ppm level to ppt level,
and it's probably more practical to keep two sets of sample introduction and interface parts, and reserve each set for its own matrix. For other sample
types, such as wastewater and clean drinking water, or clinical labs that run whole blood, urine and plasma, an extended rinse (maybe 10 minutes)
followed by a few repeats of the blank for the new matrix will work fine. by any chance, do you have further data illustrating performance
at lower spike levels? Maybe 1ppb or lower? Hg should be a good indicator for that. hard to analyze for and the recovery at 25% NaCl is phenomenal
What IS was used for the Na study? The ISTD elements used for the NaCl tests were Li6, Sc, Ge, Rh, In and Ir. ISTD assignment was just based on mass. What solid sampling accessories are available and how do they
compare to solution operation ? By far the most common accessory used for solid sampling into an ICP-MS is laser ablation. This is well-established and routine (used for geochemical
dating in prospecting studies, for example), and there are several well-established suppliers of laser ablation systems. It's difficult to compare solution
and laser analysis because there are pros and cons for each. Sample preparation is often much simpler for direct solid samples, but sample
homogeneity can be an issue. Laser gives elemental distribution information on a micron scale, but it can be much more difficult to find or make suitable
standards for the analysis. Other approaches can be used for solid sampling , including ETV introduction of powders, or even slurry nebulization, but
these are much less widely applicable. Will the new hardware features (ISIS 3, UHMI, new detector,
new collision/reaction cell, ion transmission) also be available on
the 8800?
There are no upgrade paths for those hardware parts for the 8800 at the moment, but of course we are always working on future product development
across all our ICP-MS platforms.
How the polyatomic ions are removed in ORS in no gas mode? In no gas mode, there is not a significant reduction in polyatomic ions in the cell, although a little bit of energy discrimination can still take place even
when the cell is unpressurized. But on the Agilent ICP-MS systems, we always focus a lot on making sure the plasma is tuned for very robust conditions,
which means the molecular ions are being decomposed effectively in the plasma. This is monitored using the strongly-bound Ce-O molecule, so reducing
the CeO level indicates that the plasma is working well to dissociate other interfering species (such as CaO, SO, SiO, etc)
Q&A Questions Unanswered at Live Event
Question Response Do you have a method wizard for drinking water, can't use the
ORS. We have pre-set methods for drinking water with He cell mode and without (so the EPA 200.8 pre-set method uses only no gas mode, for example). But
the Method Wizard can build a method even without an appropriate pre-set method. All you need to do is define the analyte and internal standard
masses and the method optimization will still work fine. The pre-set method just saves some time by giving you a template with many parameters already
predefined, but you can still edit the method to suit your specific requirements. And then you can save the modified method as a new template to be used
as the basis for future methods. Is this new software Mas Hunter 4.1. compatible with ICP-MS
7700 yes it is
r s d meaning relative standard deviation With all the capabilities of the 7900, it sounds like you don't need
ICP-AES any more! There's certainly more overlap in capabilities now, especially for high matrix samples and high concentration analyte measurements. But it's wrong to say
that there's no need for ICP-OES anymore. Two of the most important criteria in selecting an analytical instrument are often budget and fitness for
purpose. Many organizations and individuals are aiming to purchase the minimum compliant solution; in other words the lowest cost instrument that is fit
for purpose,. In cases where your analytical method doesn't need particularly low limits of detection, the lower cost of ICP-OES means it is often the
system of choice, even if ICP-MS could also do the analysis. Many standard methods still reference ICP-OES as well, so ICP-MS cannot be used for
those methods. In many laboratories, the switch from ICP-OES to ICP-MS is actually driven by the fact that the ICP-MS can do the measurements that
ICP-OES can't, such as trace analytes that are currently run using GFAAS, or hydride/Atomic Fluorescence for As, Se, Hg, etc). Since ICP-MS can do all
those trace elements in addition to the traditional ICP-OES workload, it often makes sense to consolidate all the analyses onto a single instrument.
Finally, ICP-OES precision is generally better than ICP-MS, so for high-precision major element assays, OES is still superior. Any experience (e.g. application notes) about the direct
detection of metal contaminants in pharmaceutical low molecular
weight compounds directly out of DMSO without further sample
preparation?
Yes, DMSO can be measured directly on the 7900 and this (and other) solvents are commonly used for sample prep for some pharmaceutical materials,
APIs etc. We have a White Paper on Pharmaceutical Analysis by ICP-MS which references some of these methods.
can the new version of Mass Hunter be applied backward to the
7700X Yes it can. There is an upgrade available which can also include the new Win 7 64bit PC, if your 7700 is currently running an earlier 32bit version of
software. For single element, single particle nanoparticle studies, can you
run with no quadrupole settling time? When you measure single nanoparticles, you have to monitor only a single mass, otherwise you could miss the signal for a particle being monitored at
mass a while you were measuring at mass b. In this single ion monitoring time resolved analysis mode, there is no wait time or settling time between
measurements. What are the possible matrix levels with out the ultra HMI
option? On the 7900, the plasma is very robust anyway (<1% CeO), and it can easily tolerate routine analysis of 0.2% total dissolved solids (TDS), even for
materials that are particularly prone to deposit on the interface (so oxides of Al, Si, Ca, etc.). For relatively simple matrices, you can run level sup to
0.5%, possible a little higher if you tune for extra robustness (so low sample flow rate, lower carrier gas flow, etc.). does the new UHMI still require the argon humidifier? The Ar humidifier is more related to the sample nebulization than the UHMI dilution step. At really high salt levels (and 25% NaCl is a saturated solution),
when the sample reaches the nebulizer tip, the pressure drop causes the salts to crystallize out of solution. Using the Ar humidifier reduces this effect, so
you don't get salt crystals building up on the nebulizer tip or devitrification of the glass. The humidifier for the 7900/UHMI is a different design from the
older one, though. It is now two channel, and it uses gas permeation rather than bubbling, so the internal volume is lower and gas changes don't require
such a long stabilization time. High energy mode - is it equal to current Normal plasma? What
about Cool plasma? Thanks The mode we refer to as "High Energy He mode" or HE He) is a cell gas mode rather than a plasma setting. HE mode uses a higher flow of He cell gas
combined with a higher cell voltage to increase the ion energies in the cell. This helps to reduce some interferences by collisional dissociation. For the
plasma modes, we still have normal plasma (at various levels of robustness or CeO ratios) and cool plasma. Cool plasma requires specific sample
introduction hardware and uses a different ion lens design, which can be added to the 7900 as an optional kit. Does the UHMI come standard when ordering the 7900 or is it
an add on? UHMI is optional for the 7900. That's really only because many labs run relatively clean samples and wouldn't appreciate having to purchase the
hardware that they will never need to use. How is sensitivity affected by use of the HMI system? HMI is a dilution system, so broadly speaking the sensitivity decreases by the UHMI dilution factor you set (100x lower sensitivity at UHMI 100x dilution,
for example). Actually it's not quite as simple as this because, in a matrix, the use of UHMI reduces signal suppression, so the net signal loss is not as
great as the nominal dilution factor. Also, because UHMI increases plasma robustness, it reduces suppression of the poorly ionized elements by an even
greater degree, so when measured in a matrix, the relative signal drop for elements like As, Se, Cd, Hg, etc. is much less than the nominal UHMI dilution
factor.
Q&A Questions Unanswered at Live Event
Question Response
What is the typical effect on LOD/BEC when running in UHMI
mode? For clean samples, the BEC and method DL (i.e. calculated back to the original undiluted sample) will degrade by approximately the same factor as the
UHMI dilution factor (so 100x poorer for UHMI 100). However, this doesn’t apply to high matrix samples, where the use of UHMI makes the plasma
much more robust so signals are not suppressed as much as they are without UHMI. With UHMI, the decomposition of polyatomics is also more
effective (lower CeO/Ce ratio), so some matrix interferences are lower even before the cell processes have their effect. This means that the net effect on
BEC and DL is not as great as the nominal UHMI dilution factor..
How much does it cost to run the instrument per sample if at full
capacity for an 8 hour day? (gas cost, maintenance, etc.) The main running cost is the Argon gas, and an ICP-MS uses approximately one standard (K/L size) cylinder of Ar (10,000 liters of gas) per 8 hours of
operation. Depending on the sample type and the sample introduction you are using, you should expect to get one solution analyzed (uptake,
stabilization, measurement and washout) every 1 to 4 minutes, so in an 8 hour day you’d get between about 100 and 400 samples measured, assuming
around a 15% QC overhead. If you use the instrument a lot (or have multiple ICP instruments) the argon costs less when purchased as liquid Ar, so
many labs use cryo containers or bulk liquid argon tanks. There is the cost of electricity to run the instrument (it draws around 3.5KW when the plasma is
on, plus about 1.3KW for the water recirculator), but maintenance costs on a daily basis are negligible as all the main consumables last many weeks or
months or more (much more in the case of costly items like the EM detector).
How does the 7900 compare with the 8800 from the standpoint
of sensitivity alone. The sensitivity of the 7900 is almost the same as the 8800, but the 8800 probably has slightly lower background.
Our most difficult matrix is CaSO4, rather than NaCl. We
experience severe salt deposition on the nebulizer, with no
notable improvement with the use of an argon humidifier. Has
any work been done on this or other non-NaCl matrices?
I’ve only come across CaSO4 as a matrix component following its use as an extractant for soil sample analysis, but at those concentrations (I think
0.01M) it could be run routinely. Of course handling high matrix samples requires careful method development to ensure the method is compatible with
the instrument hardware and operating conditions being used. Often there’s a balance between modifying the instrument conditions to allow it to handle
an extreme matrix (by using HMI, for example) and modifying the sample to allow it to be measured under more typical operating conditions. In the case
of HMI and UHMI, we have run several other high matrix sample types, including complex salt mixes and digested metals (1% Cu solution, for example),
but I don't remember seeing data on CaSO4. If you make an enquiry through your local Agilent applications person, we can investigate it.
One question: Your talked about 7900, what is 8800 set for?
what is the expectations for 8800? Yes, this webinar was focused on our new 7900 quadrupole ICP-MS system only, but we did several equivalent presentations when we launched the
8800 in 2012. The 8800 works well for all standard applications. However, it sets itself apart when it comes to really problematic analytes in really
difficult sample types. The unique capability of the 8800 is with its MS/MS capabilities, which control the ions that enter the reaction cell, so reactive cell
gases can be used selectively to remove interferences. This is different from normal quadrupole ICP-MS in reaction mode, where all the ions and matrix
elements enter the cell, so the reaction processes can change completely from one matrix (or combination of analytes) to another. The 8800 can
therefore use reaction chemistry to remove interferences much more effectively and reliably, which allows it to measure elements such as S and P at
much lower levels than quadrupole ICP-MS, and measure interfered elements accurately at the ultra-trace levels required in semiconductor or high
performance materials analysis. And unlike quadrupole ICP-MS, the 8800 can also give reliable and accurate data in reaction mode, even when the
sample matrix is complex or variable. There are many examples, but if you search on the Agilent website you should be able to find a link to the 8800
Applications Handbook (publication number 5991-2802EN).
Could you describe the unattended startup with the ISIS-3 in
more detail? During the Startup process for Agilent ICP-MS systems, a series of user-selectable optimization steps can be performed automatically. So things like
torch alignment, EM detector cross-calibration, lens tuning, and generating a standard system performance report can all be run automatically every time
the plasma is ignited. Previously these Startup functions couldn't be used with ISIS in discrete sampling (DS) mode, because DS gives a relatively short-
lived transient signal pulse from the loop injection, so the signal wasn't stable for long enough for the auto-optimization processes to be completed. The
new ISIS includes a T-connector so that the tune solution can be added to the carrier continuously, in place of the ISTD solution that is normally added
online. This means that Startup has a steady-state tune signal to work with so the Startup tasks can be completed just like for normal (non-DS) sample
intro. Can the ECM server handle a 7700 and a 7900 simultaneously? Yes, An ECM server can handle an almost unlimited number of instruments and can even manage electronic records from non-Agilent instruments.
What's the long-term analyte stability (4 hours) running 25%
dissolved solids using the new aerosol dilution system? We didn't have time to show that data, but it will be included in an upcoming application note, For most elements, the stability at the same spike levels
we showed (50ppb for most trace elements) was between 2 and 5% RSD over the 4 hour sequence (alternating 25% NaCl and 25% NaCl plus spike).
Q&A Questions Unanswered at Live Event
Question Response
Does the method wizard include TRA for chromatographic
analyses? Not at the moment, mainly because the instrument doesn’t have a database of the information about all the possible sample introduction types and signal
characteristics (which might be anything from monitoring a steady-state signal to a very rapidly changing transient signal). TRA acquisitions are much
more difficult to define in a way that a software algorithm can work with, because of the additional variable of the signal change with time. We don’t have
pre-set methods for speciation analysis as yet either, although they are on the software development plan.
what sample introduction and hardware has to be separated
going from high matrix method to low matrix method ie. 300g/LZn
matrix to low water matrix
none. there is no need to switch replace or change any HW items from Sample intro.
is there any changes / improvements to Semi Quant methods Semiquant generally works very well on the 7700 already, especially in He mode, but we didn’t make any specific changes to semiquant calibration in the
latest revision., How that compares to what you have now will depend on which revision you are working with at the moment.
Do standards need to be matrix-matched when using HMI
system? No, they don't; that is one of the main benefits of the improved robustness that UHMI provides. The NaCl matrix spike recoveries (up to 25% NaCl matrix)
that we showed were all measured against a calibration in simple aqueous standards (no NaCl matrix matching).
does the ISIS for the 7700 involve 7port valves and piston pump
as well No, ISIS 3 (the close-coupled-valve version with the piston pump) is currently only available for the 7900.
How much was the IS('s) suppressed during the Na study? Around 50% signal loss in 25% NaCl, and reasonably uniform suppression across the mass range. There is an App Note coming that looks into this data
in a bit more detail.