introduction to magnetic resonance imaging
DESCRIPTION
Introduction to Magnetic Resonance Imaging. Howard Halpern. Basic Interaction. Magnetic Moment m ( let bold indicate vectors) Magnetic Field B Energy of interaction: E = - µ∙B = -µB cos( θ ). Magnetic Moment . μ ~ Classical Orbital Dipole Moment - PowerPoint PPT PresentationTRANSCRIPT
![Page 1: Introduction to Magnetic Resonance Imaging](https://reader035.vdocuments.site/reader035/viewer/2022062305/568164ad550346895dd6b485/html5/thumbnails/1.jpg)
Introduction to Magnetic Resonance Imaging
Howard Halpern
![Page 2: Introduction to Magnetic Resonance Imaging](https://reader035.vdocuments.site/reader035/viewer/2022062305/568164ad550346895dd6b485/html5/thumbnails/2.jpg)
Basic Interaction
• Magnetic Moment m (let bold indicate vectors)
• Magnetic Field B• Energy of interaction:
– E = -µ∙B = -µB cos(θ)
![Page 3: Introduction to Magnetic Resonance Imaging](https://reader035.vdocuments.site/reader035/viewer/2022062305/568164ad550346895dd6b485/html5/thumbnails/3.jpg)
Magnetic Moment
• μ ~ Classical Orbital Dipole Moment– Charge q in orbit with diameter r, area A = πr2
– Charge moves with velocity v; Current is qv/2πr– Moment µ = A∙I = πr2∙qv/2πr = qvr/2 = qmvr/2m
• mv×r ~ angular momentum: let mvr “=“ S• β = q/2m; for electron, βe = -e/2me (negative for e)• µB = βe S ; S is in units of ħ
–Key 1/m relationship between µ and m• µB (the electron Bohr Magneton) = -9.27 10-24 J/T• µP (the proton Bohr Magneton) = 5.05 10-27 J/T****
![Page 4: Introduction to Magnetic Resonance Imaging](https://reader035.vdocuments.site/reader035/viewer/2022062305/568164ad550346895dd6b485/html5/thumbnails/4.jpg)
Magnitude of the Dipole Moments
• Key relationship: |µ| ~ 1/m• Source of principle difference between
– EPR experiment– NMR experiment
• This is why EPR can be done with cheap electromagnets and magnetic fields ~ 10mT not requiring superconducting magnets
![Page 5: Introduction to Magnetic Resonance Imaging](https://reader035.vdocuments.site/reader035/viewer/2022062305/568164ad550346895dd6b485/html5/thumbnails/5.jpg)
Torque on the Dipole in the Magnetic Field B0
• Torque, τ, angular force• τ = -∂E/∂θ = µB0 sin θ = -| µ × B |• For moment of inertia I• I ∂2θ/∂2t = -|µ|B0 sin θ ~ -|µ|B0θ for small θ• This is an harmonic oscillator equation
(m∂2x/∂2t+kx=0) • Classical resonant frequency (ω0=√(k/m)) ω0=√(|µ|B0/I)
![Page 6: Introduction to Magnetic Resonance Imaging](https://reader035.vdocuments.site/reader035/viewer/2022062305/568164ad550346895dd6b485/html5/thumbnails/6.jpg)
Damped Oscillating Moment: Pulsed Experiment
• Solution to above is ψ = A exp(iωt) + B exp(-iωt): Undamped, Infinite in time, Unphysical
• To I ∂2θ/∂2t + µB0θ =0 add a friction term Γ∂θ/∂t to get: I ∂2θ/∂2t + Γ∂θ/∂t + µB0θ =0
ω = iΓ/2I ± (4IµB0-Γ2)1/2/2I so that (t) = A exp(-t Γ/2I - i ((4IµB0-Γ2)1/2/2I)t ) + B exp(-t Γ/2I + i ((4IµB0-Γ2)1/2/2I)t ) Lifetime: T ~ 2I/ ΓRedfield Theory: Γ~µ2 so T1s &T2s α 1/μ2
Interaction with environment
“Reality Term”
![Page 7: Introduction to Magnetic Resonance Imaging](https://reader035.vdocuments.site/reader035/viewer/2022062305/568164ad550346895dd6b485/html5/thumbnails/7.jpg)
Damped Driven Oscillating Moment: Continuous Wave Measurement
• If we add a driving term to the equation so that I ∂2θ/∂2t + Γ∂θ/∂t + µB0θ =B1exp(iω1t),
• In the steady state θ(t) = B1exp(iω1t)/[ω02-ω1
2+2i ω1 Γ/2I] ~B1exp(iω1t)/[2ω0(ω0-ω1)+2i ω1 Γ/2I]a resonant like profile with ω0
2=µB/I, proportional to the equilibrium energy.
Re [θ(t)]~B1/ ω12 [(Δω)2+ (Γ/2I)2] for
ω0~ω1
Lorentzian shape, Linewidth Γ/2I
![Page 8: Introduction to Magnetic Resonance Imaging](https://reader035.vdocuments.site/reader035/viewer/2022062305/568164ad550346895dd6b485/html5/thumbnails/8.jpg)
Plot
0 50 1000.002
0.001
01.601 10 6
2 10 3
Im w(
1001 w50 1000.002
0
0.002
9.081 10 4
Re w(
1001 w
![Page 9: Introduction to Magnetic Resonance Imaging](https://reader035.vdocuments.site/reader035/viewer/2022062305/568164ad550346895dd6b485/html5/thumbnails/9.jpg)
Profile of MRI Resonator• Traditionally resonator thought to amplify the
resonant signal• The resonant frequency ω0 and the linewidth term
Γ/2I are expressed as a ratio characterizing the sharpness of the resonance line: Q= ω0/(Γ/2I)
• This is also the resonator amplification term• ω0/Q also characterizes the width of frequencies
passed by the resonator: ω0/Q = Δω
![Page 10: Introduction to Magnetic Resonance Imaging](https://reader035.vdocuments.site/reader035/viewer/2022062305/568164ad550346895dd6b485/html5/thumbnails/10.jpg)
Q.M. Time Evolution of the Magnetization
• S is the spin, an operator• H is the Hamiltonian or Energy Operator• In Heisenberg Representation:
∂S/ ∂t = 1/iћ [H,S], Recall, H= -µB∙B =- βe S ∙B. [H,Si]=HSi-SiH=-βe(SjSi-SiSj)Bj= βeiћS x B
∂S/ ∂t = βe S x B; follows also classically from the torque on a magnetic moment ~ S x B as seen above. Bloch Equations follow with (let M = S, M averaging over
S):∂S/ ∂t = βe S x B -1/T2(Sxî +Syĵ)-1/T1(Sz-S0)ķ
![Page 11: Introduction to Magnetic Resonance Imaging](https://reader035.vdocuments.site/reader035/viewer/2022062305/568164ad550346895dd6b485/html5/thumbnails/11.jpg)
Density Matrix • Define an Operator ρ associating another operator S with
its average value:• <S> = Tr(ρS) ; ρ=Σ|ai><ai|; basically is an average over
quantum states or wavefunctions of the system• ρ : Density Matrix: Characterizes the System• ∂ρ/∂t = 1/iћ [H, ρ], 1/iћ [H0+H1, ρ],
• ρ*= exp(iH0t/ћ) ρ exp(-iH0t/ћ)=e+ ρ e-; H1*=e+H1e-
• ∂ρ*/∂t = 1/iћ [H*1(t), ρ*(t)], • ρ*~ ρ*(0) +1/iћ ∫dt’ [H*(t’)1, ρ*(0)]+ 0
(1/iћ)2 ∫dt’dt”[H*1(t’),[H*1(t”), ρ*(0)]
Here we break the Hamiltonian into two terms, the basic energy term, H0 = -µ∙B0, and a random driving term H1=µ∙Br(t), characterizing the friction of the damping
![Page 12: Introduction to Magnetic Resonance Imaging](https://reader035.vdocuments.site/reader035/viewer/2022062305/568164ad550346895dd6b485/html5/thumbnails/12.jpg)
The Damping Term (Redfield)• ∂ρ*/∂t=1/iћ [H*(t’)1, ρ*(0)]+ 0(1/iћ)2 ∫dt’[H*1(t),[H*1(t-t’), ρ*(t’)]• Each of the H*1 has a term in it with H= -µ∙Br and µ ~
q/m• ∂ρ*/∂t~ (-) µ2 ρ* This is the damping term:• ρ*~exp(- µ2t with other terms)• Thus, state lifetimes, e.g. T2, are inversely
proportional to the square of the coupling constant, µ2
• The state lifetimes, e.g. T2 are proportional to the square of the mass, m2
![Page 13: Introduction to Magnetic Resonance Imaging](https://reader035.vdocuments.site/reader035/viewer/2022062305/568164ad550346895dd6b485/html5/thumbnails/13.jpg)
Consequences of the Damping Term
• The coupling of the electron to the magnetic field is 103 times larger than that of a water proton so that the states relax 106 times faster
• No time for Fourier Imaging techniques• For CW we must use
1. Fixed stepped gradients 1. Vary both gradient direction & magnitude (3 angles)
2. Back projection reconstruction in 4-D
![Page 14: Introduction to Magnetic Resonance Imaging](https://reader035.vdocuments.site/reader035/viewer/2022062305/568164ad550346895dd6b485/html5/thumbnails/14.jpg)
Major consequence on image technique
• µelectron=658 µproton (its not quite 1/m);• Proton has anomalous magnetic moment
due to “non point like” charge distribution (strong interaction effects)
• mproton= 1836 melectron
• Anomalous effect multiplies the magnetic moment of the proton by 2.79 so the moment ratio is 658
![Page 15: Introduction to Magnetic Resonance Imaging](https://reader035.vdocuments.site/reader035/viewer/2022062305/568164ad550346895dd6b485/html5/thumbnails/15.jpg)
Relaxtion times:
• Water protons: Longitudinal relaxation times, T1 also referred to as spin lattice relaxation time ~ 1 sec
T2 transverse relaxtion time or phase coherence time ~ 100s of milliseconds• Electrons T1e 100s of ns to µs T2e 10s of ns to µs
![Page 16: Introduction to Magnetic Resonance Imaging](https://reader035.vdocuments.site/reader035/viewer/2022062305/568164ad550346895dd6b485/html5/thumbnails/16.jpg)
Magnetization: Magnetic Resonance Experiment
• Above touches on the important concept ofmagnetization M as opposed to spin S
• From the above, M=Tr(ρS), • Thus magnetization M is an average of the spin
operator over the states of the spin system• These can be
– coherently prepared or,– as is most often the case, incoherent states – Or mixed coherent and incoherent
![Page 17: Introduction to Magnetic Resonance Imaging](https://reader035.vdocuments.site/reader035/viewer/2022062305/568164ad550346895dd6b485/html5/thumbnails/17.jpg)
Magnetization
• As above, we define M=<S>, the state average of spin
• If 1/T1,1/T2=0 equations: ∂M/ ∂t = βeM x B; B=Bz
• Bx=cos(ω0t)• By=sin(ω0t)• ω0= βeBz/ℏ • i.e., the magnetization precesses in the magnetic field
![Page 18: Introduction to Magnetic Resonance Imaging](https://reader035.vdocuments.site/reader035/viewer/2022062305/568164ad550346895dd6b485/html5/thumbnails/18.jpg)
L-C Resonator
C
L
Zeeman (B0)Magnetic Field Directions:
Bz
Bx
By
![Page 19: Introduction to Magnetic Resonance Imaging](https://reader035.vdocuments.site/reader035/viewer/2022062305/568164ad550346895dd6b485/html5/thumbnails/19.jpg)
Resonator Functions
1. Generate Radiofrequency/Microwave fields to stimulate resonant absorption and dispersion at resonance condition: hn=mB0
2. Sense the precessing induction from the spin sample
1. Spin sample magnetization M couples through resonator inductance
2. Creates an oscillating voltage at the resonator output
![Page 20: Introduction to Magnetic Resonance Imaging](https://reader035.vdocuments.site/reader035/viewer/2022062305/568164ad550346895dd6b485/html5/thumbnails/20.jpg)
Two modes
• Continuous wave detection• Pulse
![Page 21: Introduction to Magnetic Resonance Imaging](https://reader035.vdocuments.site/reader035/viewer/2022062305/568164ad550346895dd6b485/html5/thumbnails/21.jpg)
Continuous Wave (CW)• Elaborate• Narrow band with high Q (ω0/Q = Δω) highly tuned
resonator. For the time being ω0 is frequency, not angular frequency
• Narrow window in ω is swept to produce a spectrum• Because resonator is highly tuned sweep of B0,
Zeeman field• h ω0/μ=B= B0 +BSW : Narrow window swept through
resonance
![Page 22: Introduction to Magnetic Resonance Imaging](https://reader035.vdocuments.site/reader035/viewer/2022062305/568164ad550346895dd6b485/html5/thumbnails/22.jpg)
Bridge
• Generally the resonator is part of a “bridge” analogous to the Wheatstone bridge to measure a resistance by balancing voltage drops across resistance and zeroing the current
![Page 23: Introduction to Magnetic Resonance Imaging](https://reader035.vdocuments.site/reader035/viewer/2022062305/568164ad550346895dd6b485/html5/thumbnails/23.jpg)
Homodyne Bridge
Resonator tuned to impedance of 50ΩThus, when no resonance all signal from Sig Gen absorbed in resonatorWhen resonance occurs energy is absorbed from RF circuit: RESISTANCE resonator impedance changes => imbalance of “bridge” giving a signal
![Page 24: Introduction to Magnetic Resonance Imaging](https://reader035.vdocuments.site/reader035/viewer/2022062305/568164ad550346895dd6b485/html5/thumbnails/24.jpg)
For this “Bridge” balance with system impedance (resistance): 50 Ω• Resonance involves absorption of energy from the RF
circuit, appearing as a proportional resistance • Bridge loses balance: resonator impedance no longer
50 Ω.• Signal: 250 MHz voltage• Mixers combine reference from SIG GEN• Multiply Acoswt*coswt= A[cos(w+wt+cos(wwt]=A + high frequency which we filter
• Demodulating the Carrier Frequency.
![Page 25: Introduction to Magnetic Resonance Imaging](https://reader035.vdocuments.site/reader035/viewer/2022062305/568164ad550346895dd6b485/html5/thumbnails/25.jpg)
More• Field modulation• Very low frequency “baseline” drift: 1/f noise• Solution: Zeeman field modulation• Add to B0 a Bmod=coswmodt term where wmod is ~audio
frequency.• Detect with a Lock-in amplifier whose output is the
input amplitude at a multiple (harmonic) of wmod • First harmonic ~ first derivative like spectrum
![Page 26: Introduction to Magnetic Resonance Imaging](https://reader035.vdocuments.site/reader035/viewer/2022062305/568164ad550346895dd6b485/html5/thumbnails/26.jpg)
First hamonic spectrum
![Page 27: Introduction to Magnetic Resonance Imaging](https://reader035.vdocuments.site/reader035/viewer/2022062305/568164ad550346895dd6b485/html5/thumbnails/27.jpg)
Pulse• Offensive lineman’s approach to magnetic resonance• Depositing a broad-band pulse into the spin system
and then detect its precession and its decay times• CW: driving a harmonic oscillator with frequency w
and measuring amplitude response• Pulse: striking mass on spring and measuring
oscillation response including many modes each with– A frequency– An amplitude– Fourier transform give spectral frequency response
![Page 28: Introduction to Magnetic Resonance Imaging](https://reader035.vdocuments.site/reader035/viewer/2022062305/568164ad550346895dd6b485/html5/thumbnails/28.jpg)
Broadband Pulse
• Real Uncertainty principle in signal theory:• ΔνΔT=1• Exercise: For a signal with temporal
distribution F(t)=exp(-t^2) what is the product of the FWHH of the temporal distribution and that of its Fourier transform?
• What is that of F(t)=exp(-(at)^2)?
![Page 29: Introduction to Magnetic Resonance Imaging](https://reader035.vdocuments.site/reader035/viewer/2022062305/568164ad550346895dd6b485/html5/thumbnails/29.jpg)
MRI broadband
• ΔT=1μs• Δν=1 MHz• Water proton gyromagnetic ratio: 42
MH/T• => 1μs pulse give 0.024 T excitation
(.24 KGauss)
![Page 30: Introduction to Magnetic Resonance Imaging](https://reader035.vdocuments.site/reader035/viewer/2022062305/568164ad550346895dd6b485/html5/thumbnails/30.jpg)
EPR Broadband
• 1 ns pulse => 1 GHz excitation• Electron gyromagnetic ratio: 28 GHz/T• 1 ns pulse gives 0.035 T wide excitation• NMR lines typically much narrower than EPR• EPR is a hard way to live
![Page 31: Introduction to Magnetic Resonance Imaging](https://reader035.vdocuments.site/reader035/viewer/2022062305/568164ad550346895dd6b485/html5/thumbnails/31.jpg)
Pulse bridge
• Simpler particularly for MRI• Requires same carrier demodulation• Requires lower Q for broad pass band• Excites all spins in the passband window so, in
principle, more efficient acqusition
![Page 32: Introduction to Magnetic Resonance Imaging](https://reader035.vdocuments.site/reader035/viewer/2022062305/568164ad550346895dd6b485/html5/thumbnails/32.jpg)
Magnetic field gradients encode position or location
• Postulated by Lauterbur in landmark 1973 Nature paper
• Overcomes diffraction limit on 40 MHz RF– λ = 7.5 m
• Gradient: Gi= dB/dxi where x1=x; x2=y; x3=z• B is likewise a vector but generally take to be
Bz.
![Page 33: Introduction to Magnetic Resonance Imaging](https://reader035.vdocuments.site/reader035/viewer/2022062305/568164ad550346895dd6b485/html5/thumbnails/33.jpg)
Standard MRI z slice selection
• z dimension distinguished by pulsing a large gradient during the acqusition
• GzΔz = ∂Bz/∂zΔz = ΔB = hΔν/βe
• Δν = GzΔz/hβe
• If Δν > ω0/Q the frequency pass band of the resonator, then the sensitive slice thickness is defined by Δz= hβeω0/QGz
![Page 34: Introduction to Magnetic Resonance Imaging](https://reader035.vdocuments.site/reader035/viewer/2022062305/568164ad550346895dd6b485/html5/thumbnails/34.jpg)
X and Y location encoding
• For standard water proton MRI, within the z selected plane
• Apply Gradient Gx-Gy plane• Frequency and phase encoding
![Page 35: Introduction to Magnetic Resonance Imaging](https://reader035.vdocuments.site/reader035/viewer/2022062305/568164ad550346895dd6b485/html5/thumbnails/35.jpg)
Phase encoding
• Generally one of the Gx or Gy direction generating gradients selected. Say Gy
• The gradient is pulsed for a fixed time Δt before signal from the precessing magnetization is measured.
• Magnetization develops a phase proportional to the distance along the y coordinate Δφ=2πyGy/βp where βp is the proton gyromagnetic ratio.
![Page 36: Introduction to Magnetic Resonance Imaging](https://reader035.vdocuments.site/reader035/viewer/2022062305/568164ad550346895dd6b485/html5/thumbnails/36.jpg)
Frequency Encoding
• The orthogonal direction say x has the gradient imposed in that direction during the acquisition of the magnetization precession signal
• This shifts the frequency of the precession proportional to the x coordinate magnitude Δω=xGx
![Page 37: Introduction to Magnetic Resonance Imaging](https://reader035.vdocuments.site/reader035/viewer/2022062305/568164ad550346895dd6b485/html5/thumbnails/37.jpg)
Signal Amplitude vs location
• The complex Fourier transform of the voltage induced in the resonator by the precession of the magnetization gives the signal amplitude as a function of frequency and phase
• The signal amplitude as a function of these parameters is the amplitude of the magnetization as a function of location in the sample
![Page 38: Introduction to Magnetic Resonance Imaging](https://reader035.vdocuments.site/reader035/viewer/2022062305/568164ad550346895dd6b485/html5/thumbnails/38.jpg)
EPR: No time for Phase Encoding
• Generally use fixed stepped gradients• Both for CW and Pulse• Tomographic or FBP based reconstruction
![Page 39: Introduction to Magnetic Resonance Imaging](https://reader035.vdocuments.site/reader035/viewer/2022062305/568164ad550346895dd6b485/html5/thumbnails/39.jpg)
Imaging: Basic StrategyConstraint: Electrons relax 106 times faster than water protons
Image Acquisition: Projection Reconstruction: Backprojection
![Page 40: Introduction to Magnetic Resonance Imaging](https://reader035.vdocuments.site/reader035/viewer/2022062305/568164ad550346895dd6b485/html5/thumbnails/40.jpg)
Key for EPR Images
• Spectroscopic imaging• Obtain a spectrum from each voxel• EPRI usually uses an injected reporter
molecule• Spectral information from the reporter
molecule from each voxel quantitatively reports condition of the fluids of the distribution volume of the reporter
![Page 41: Introduction to Magnetic Resonance Imaging](https://reader035.vdocuments.site/reader035/viewer/2022062305/568164ad550346895dd6b485/html5/thumbnails/41.jpg)
Projection Acquisition in EPR
( ,f B x0 0
,2 2 xB BB Bn n
+
( 0app swB x B B G x + +
• Spectral Spatial Object Support ~
TOTh g Bn
![Page 42: Introduction to Magnetic Resonance Imaging](https://reader035.vdocuments.site/reader035/viewer/2022062305/568164ad550346895dd6b485/html5/thumbnails/42.jpg)
=α
Imposition of a Gradient on aspectral spatialobject (a) here 3 locations each with spectra acts to shear the spectral spatial object (b) giving the resolved spectrum shown. This is equivalent to observing (a) at an angle a (c). This is a Spectral-Spatial Projection.
![Page 43: Introduction to Magnetic Resonance Imaging](https://reader035.vdocuments.site/reader035/viewer/2022062305/568164ad550346895dd6b485/html5/thumbnails/43.jpg)
Projection Description
• With s(Bsw, Ĝ) defined as the spectrum we get with gradients imposed
( ( ( 2
2
, ,x
B
sw sw swB
s B G f B x B B G x dxdB
+
![Page 44: Introduction to Magnetic Resonance Imaging](https://reader035.vdocuments.site/reader035/viewer/2022062305/568164ad550346895dd6b485/html5/thumbnails/44.jpg)
More Projection
• The integration of fsw is carried out over the hyperplane in 4-space by
swB B G x +
DefiningˆcG x
BcL
with
The hyperplane becomes1ˆ( )sw swB B G G x B c G + +
with 1tan c Ga
![Page 45: Introduction to Magnetic Resonance Imaging](https://reader035.vdocuments.site/reader035/viewer/2022062305/568164ad550346895dd6b485/html5/thumbnails/45.jpg)
And a Little More
( ( ( 2
2
ˆ, cos , cos cos sinx
B
sw sw swB
s B G f B x B B cG x dxdBa a a a
+
+
So we can write B = Bsw + c tanα Ĝ∙x
![Page 46: Introduction to Magnetic Resonance Imaging](https://reader035.vdocuments.site/reader035/viewer/2022062305/568164ad550346895dd6b485/html5/thumbnails/46.jpg)
So finally it’s a Projection
( ( ( ˆ, cosr
sw r Gs B G f r r dra a
with
( (
cos
,
ˆˆ cos ,sin
, .2 2
sw
G
r x
B
r B c x
G
B B c
a
a a a
![Page 47: Introduction to Magnetic Resonance Imaging](https://reader035.vdocuments.site/reader035/viewer/2022062305/568164ad550346895dd6b485/html5/thumbnails/47.jpg)
Projection acquisition and image reconstructionImage reconstruction: backprojections in spectral-spatial space
Resolution: δx=δB/Gmax THUS THE RESOLUTION OF THE IMAGE PROPORTIONAL TO δB
each projection filtered and subsampled;
Interpolation of Projections: number of projections x 4 with sinc(?) interpolation,Enabling for fitting
Spectral Spatial ObjectAcquisition: Magnetic field sweep w stepped gradients (G)Projections: Angle a: tan(a=G*L/H
![Page 48: Introduction to Magnetic Resonance Imaging](https://reader035.vdocuments.site/reader035/viewer/2022062305/568164ad550346895dd6b485/html5/thumbnails/48.jpg)
The Ultimate Object:Spectral spatial imaging
![Page 49: Introduction to Magnetic Resonance Imaging](https://reader035.vdocuments.site/reader035/viewer/2022062305/568164ad550346895dd6b485/html5/thumbnails/49.jpg)
Preview: Response of Symmetric Trityl (deuterated) to Oxygen
• So measuring the spectrummeasures the oxygen. Imaging the spectrum: Oxygen Image