introduction to lc-ms to lc-ms dr. michael baynham senior specialist, mass spectrometry applications...
TRANSCRIPT
Introduction to LC-MS
Dr. Michael Baynham Senior Specialist, Mass Spectrometry ApplicationsClinical Research Network Coordinator
2 24th April 2006
Applied Biosystems — Advancing Quality Science
© 2006 Applied Biosystems
Overview
● Comparison of LC-MS with more traditional techniques such as LC with DAD detection, GC-MS & immunoassay analysis
● What is the difference between LC-MS and LC-MS/MS?● LC-MS/MS system technology● LC-MS/MS source technology● What effects the way a compound forms an ion?● Summary
Comparison of LC-MS with Traditional Techniques
4 24th April 2006
Applied Biosystems — Advancing Quality Science
© 2006 Applied Biosystems
Diode Array Detection- not specific enough for differentiation
5 24th April 2006
Applied Biosystems — Advancing Quality Science
© 2006 Applied Biosystems
Mass Spectrometry Specificity
324308
6 24th April 2006
Applied Biosystems — Advancing Quality Science
© 2006 Applied Biosystems
Comparison of LC-MS with GC-MS
● GC-MS requires compounds to be volatile to be ionised– traditionally electron impact source is used.
● LC-MS can be used to detect compounds from poly-aromatic (non-polar) to peptide and proteins.
● GC-MS is still able to detect long chain aliphatic compounds (petroleum based anlaytes) and very low mass volatile material better than LC-MS.
7 24th April 2006
Applied Biosystems — Advancing Quality Science
© 2006 Applied Biosystems
Comparison of LC-MS/MS with Immunoassay
● LC-MS/MS is more specific than immunoassay – LC-MS/MS detection is based on column retention time,
parent mass and structure. Therefore it is compound specific.
● Furthermore, derivitisation is not normally needed to detect the target compound, saving analysis time and money.– Immunoassay is normally based on compound class and
will not be able to determine which compound in a group is detected.
8 24th April 2006
Applied Biosystems — Advancing Quality Science
© 2006 Applied Biosystems
Comparison of LC-MS/MS with Immunoassay
● LC-MS/MS is generally more sensitive than immunoassay● LC-MS/MS can screen for more than one compound class
with one injection ● Immunoassay measures one compound class per assay
and therefore needs more sample when compound class screening
● Immunoassay is generally more rapid than LC-MS/MS
9 24th April 2006
Applied Biosystems — Advancing Quality Science
© 2006 Applied Biosystems
General reasons to look at LC-MS(/MS)
● Multi-component analysis– LC-MS(/MS) can detect over 300 compounds of different
classes in just one run using often a low volume of sample– Reduces sample analysis time and allows for multiple
analyses to be done on low volumes of samples with minimal sample pre-treatment
What is the difference between LC-MS and LC-MS/MS?
11 24th April 2006
Applied Biosystems — Advancing Quality Science
© 2006 Applied Biosystems
LC-MS (first stage of MS/MS)
+ P r e c ( 1 2 2 .1 0 ) : E x p 1 , 6 .8 6 1 m in f r o m S a m p le 1 ( B u s p i r o n e M e ta b o l i t e s ) o f B u s p ir o n e _ P r e 1 2 2 _ 0 1 . . . M a x . 1 .1 e 4 c p s .
2 6 0 2 8 0 3 0 0 3 2 0 3 4 0 3 6 0 3 8 0 4 0 0 4 2 0 4 4 0 4 6 0 4 8 0 5 0 0 5 2 0 5 4 0 5 6 0 5 8 0 6 0 0 6 2 0 6 4 0m /z , a m u
5 %
1 0 %
1 5 %
2 0 %
2 5 %
3 0 %
3 5 %
4 0 %
4 5 %
5 0 %
5 5 %
6 0 %
6 5 %
7 0 %
7 5 %
8 0 %
8 5 %
9 0 %
9 5 %
1 0 0 %
402
Many compounds
enter the MS as ions
Compounds filtered based on
their mass/charge ratio and detected.
Single masses selected when performing
quantitation.
Compounds Ionised
12 24th April 2006
Applied Biosystems — Advancing Quality Science
© 2006 Applied Biosystems
Many analytesenter the MS
One analyteselected and
broken up by collisions with gas
All the fragment ions from the one analyte detected
+EPI (402.25) Charge (+1): Exp 3, 6.891 min from Sample 1 (Buspirone Metabolites) of Buspirone... Max. 1.6e6 cps.
60 80 100 120 140 160 180 200 220 240 260 280 300 320 340 360 380 400 420 440 460m/z, amu
5%
10%
15%
20%
25%
30%
35%
40%
45%
50%
55%
60%
65%
70%
75%
80%
85%
90%
95%
100% 402.3
122.1
219.2150.1
238.2
178.2
98.1
196.2 281.2
N
NN N
N
O
O122
281238
196
OH
LC-MS/MS - Product Ion Scan
13 24th April 2006
Applied Biosystems — Advancing Quality Science
© 2006 Applied Biosystems
LC-MS/MS - Multiple Reaction Monitoring (MRM)
Many analytesenter the MS
Only one of the fragment ions from the one analytefiltered off and detected
One analyteselected and
broken up by collisions with gas
XIC for Prazepam (325/271)500 fg on column (in urine)
14 24th April 2006
Applied Biosystems — Advancing Quality Science
© 2006 Applied Biosystems
Why use LC-MS/MS rather than LC-MS
● Specificity and Sensitivity– LC-MS/MS is often 20 -100 times more sensitive compared
to LC-MS in a multi-targeted screening approach.– LC-MS/MS is much more specific than LC-MS as there is a
second filtering process. ○ Can use shorter columns allowing quicker run times as there is
less background interference from the sample.
– Allows the generation of library searchable spectra
15 24th April 2006
Applied Biosystems — Advancing Quality Science
© 2006 Applied Biosystems
FROM MS TO MS/MS
1.12e9 cpsTIC of +Q1: from Full Scan 100 - 800
% In
tens
ity
0.5 1.0 1.5 2.0 2.5 3.0
Time, min
0.120.41
0.88 1.111.34
1.521.81
2.10 2.28
2.45 2.74
5101520253035404550556065707580859095
10 ng of drug injected on-column
16 24th April 2006
Applied Biosystems — Advancing Quality Science
© 2006 Applied Biosystems
FROM MS TO MS/MS
5.20e5 cpsXIC of +Q1: from 482.0 amu from Full Scan
0.06 0.35 0.53 0.700.93 1.11
1.28
1.64
1.81 2.10
2.57
2.803.04
0.5 1.0 1.5 2.0 2.5 3.0
Time, min
5101520253035404550556065707580859095
% In
tens
ity
17 24th April 2006
Applied Biosystems — Advancing Quality Science
© 2006 Applied Biosystems
FROM MS TO MS/MS
6.14e5 cpsTIC of +Q1 SIM (482): from SIM
0.09 0.240.36 0.610.74 0.921.03 1.28 1.45 1.701.87 2.02 2.242.37
2.57
2.762.883.01
0.5 1.0 1.5 2.0 2.5 3.0
Time, min
5101520253035404550556065707580859095
% In
tens
ity
18 24th April 2006
Applied Biosystems — Advancing Quality Science
© 2006 Applied Biosystems
FROM MS TO MS/MS
4.20e4 cpsMRM from 482 - 258
0.18 0.54 0.77 1.00 1.241.36 1.57 1.76 2.00 2.20 2.37
2.64
2.94
0.5 1.0 1.5 2.0 2.5 3.0
Time, min
5101520253035404550556065707580859095
% In
tens
ity
19 24th April 2006
Applied Biosystems — Advancing Quality Science
© 2006 Applied Biosystems
LC-MS/MS Systems for Small Molecule Analysis
● Iontrap (traditional spherical traps)● Quadrupole Time of Flight (QqTOF) systems● Triple quadrupole (QQQ) systems.● Q TRAP® systems.
20 24th April 2006
Applied Biosystems — Advancing Quality Science
© 2006 Applied Biosystems
Similarities between these systems?
● They all separate ions based on their mass/charge (m/z) ratio
● All are capable of LC-MS/MS analysis● In clinical samples the charge (z) is normally 1
Triple Quadrupole (QQQ) Systems
22 24th April 2006
Applied Biosystems — Advancing Quality Science
© 2006 Applied Biosystems
QQQ systems
● Qualitative analysis (spectral data acquisition)– Relative low sensitivity in scan mode due to lower duty
cycle.– Low mass accuracy and resolution versus QqTOF systems.
● Quantitative analysis (multi-targeted applications)– MRM possible – High sensitivity and linearity– Simultaneous multi component analysis possible.
Spherical (3-dimensional) Ion Traps
ions
Heated transfer capillary
Ion transfer 3D Trap
24 24th April 2006
Applied Biosystems — Advancing Quality Science
© 2006 Applied Biosystems
Spherical Ion Traps Product Ion Operation
● MS/MS in time, not in space– Ion processing (trapping, isolating, excitation
(fragmentation) and scanning take place at the same location (inside the trap cavity).
● Can do MSn
25 24th April 2006
Applied Biosystems — Advancing Quality Science
© 2006 Applied Biosystems
Many analytesenter the MS
One mass isolated and then
fragmented by excitation.
All the fragment ions trapped in the spherical trap are then scanned out based on their mass/charge
Spherical Ion Trap Product Ion Product Ion SScancan
26 24th April 2006
Applied Biosystems — Advancing Quality Science
© 2006 Applied Biosystems
Spherical Ion Traps● Qualitative analysis (spectral data acquisition)
– High sensitivity in full scan and product ion scan mode – Low mass accuracy and resolution versus QqTOF systems.– Space charge effects possible (Resolution loss, mass shifts,
isotope ratio shift, loss of dynamic range and accuracy).– Low mass cut off.
● Quantitative analysis (multi-targeted applications)– Lower sensitivity and linearity.– No real MRM possible.– No simultaneous multi component analysis (relative low scan
speed compared to MRM on triple).
QqTOF Technology
28 24th April 2006
Applied Biosystems — Advancing Quality Science
© 2006 Applied Biosystems
QqTOF Product Ion MS/MS Operation
Scan precursors Detect all productsCAD
Q1 LINAC® Q2Collision Cell
product ionspectrum
Q1 transmits a given mass, sending a single precursor into Q2 Ions are fragmented in Q2 to produce product ions TOF scans selected mass range to detect product ions TOF
29 24th April 2006
Applied Biosystems — Advancing Quality Science
© 2006 Applied Biosystems
QqTOF systems
● Qualitative analysis (spectral data acquisition)– High sensitivity in full scan and product ion scan mode – High mass accuracy and resolution versus iontraps and
QQQ systems (but mass accuracy does not help distinguish samples where they have have the same molecular formulae but different structures).
● Quantitative analysis (multi-targeted applications)– Lower sensitivity and linearity– No real MRM possible– No simultaneous multi component analysis (relative low
scan speed compared to MRM on triple)
Hybrid Linear Ion Trap Technology (Q TRAP® System)
31 24th April 2006
Applied Biosystems — Advancing Quality Science
© 2006 Applied Biosystems
Location of Linear Ion Trap
Q0
N2 CAD Gas
linear ion trap3x10-5 Torr
Exit lens
Aux AC
LINACTM
Collision Cell
Q1 Q2 Q3
The ion trap is in the Q3 region of a triple quadrupole.
Because the trap is now a quadrupole it can hold more ions than a conventional 3D trap which reduces space charge effects. This leads to better spectra, giving you more confidence in the results.
32 24th April 2006
Applied Biosystems — Advancing Quality Science
© 2006 Applied Biosystems
Many analytesenter the MS and are separated by a quadrupolemass filter
One analyteselected and
broken up by collisions with gas
All the fragment ions trapped in the Linear trap and then scanned out based on their mass/charge
Q TRAP® System - Enhanced Product Ion Scan
33 24th April 2006
Applied Biosystems — Advancing Quality Science
© 2006 Applied Biosystems
Q TRAP® Systems● Qualitative analysis (spectral data acquisition)
– High sensitivity in full scan and product ion scan mode – Higher mass accuracy and resolution versus QQQ systems
but less than QqTOF systems.– Reduced potential of space charge effects.– No low mass cut off.
● Quantitative analysis (multi-targeted applications)– High sensitivity and linearity.– MRM possible.– Simultaneous multi component analysis possible.
34 24th April 2006
Applied Biosystems — Advancing Quality Science
© 2006 Applied Biosystems
LC-MS/MS Source Technology
35 24th April 2006
Applied Biosystems — Advancing Quality Science
© 2006 Applied Biosystems
Ionization techniques for small molecule applications
● Electrospray Ionisation● Atmospheric Pressure Chemical Ionisation
Electrospray Ionisation
37 24th April 2006
Applied Biosystems — Advancing Quality Science
© 2006 Applied Biosystems
Electrospray Source Principle
Ion Spray: Electrospray Ionisation (ESI)softest ionisation techniquepolar to ionic substances
38 24th April 2006
Applied Biosystems — Advancing Quality Science
© 2006 Applied Biosystems
Electrospray Source Principle
Ion Spray: Electrospray Ionisation (ESI)softest ionisation techniquepolar to ionic substances
39 24th April 2006
Applied Biosystems — Advancing Quality Science
© 2006 Applied Biosystems
Electrospray Source Principle
Ion Spray: Electrospray Ionisation (ESI)softest ionisation techniquepolar to ionic substances
40 24th April 2006
Applied Biosystems — Advancing Quality Science
© 2006 Applied Biosystems
Electrospray Source Principle
Ion Spray: Electrospray Ionisation (ESI)softest ionisation techniquepolar to ionic substances
41 24th April 2006
Applied Biosystems — Advancing Quality Science
© 2006 Applied Biosystems
Electrospray Source Principle
Ion Spray: Electrospray Ionisation (ESI)softest ionisation techniquepolar to ionic substances
Atmospheric Pressure Chemical Ionization
43 24th April 2006
Applied Biosystems — Advancing Quality Science
© 2006 Applied Biosystems
Heated Nebulizer (APCI)
Heated Nebulizer:Atmospheric Pressure Chemical Ionisation (APCI)corona discharge needlepolar to unpolar and thermally stable compounds
44 24th April 2006
Applied Biosystems — Advancing Quality Science
© 2006 Applied Biosystems
Heated Nebulizer (APCI)
Heated Nebulizer:Atmospheric Pressure Chemical Ionisation (APCI)corona discharge needlepolar to unpolar and thermally stable compounds
45 24th April 2006
Applied Biosystems — Advancing Quality Science
© 2006 Applied Biosystems
Heated Nebulizer (APCI)
Heated Nebulizer:Atmospheric Pressure Chemical Ionisation (APCI)corona discharge needlepolar to unpolar and thermally stable compounds
46 24th April 2006
Applied Biosystems — Advancing Quality Science
© 2006 Applied Biosystems
Heated Nebulizer (APCI)
Heated Nebulizer:Atmospheric Pressure Chemical Ionisation (APCI)corona discharge needlepolar to unpolar and thermally stable compounds
47 24th April 2006
Applied Biosystems — Advancing Quality Science
© 2006 Applied Biosystems
Heated Nebulizer (APCI)
Heated Nebulizer:Atmospheric Pressure Chemical Ionisation (APCI)corona discharge needlepolar to unpolar and thermally stable compounds
48 24th April 2006
Applied Biosystems — Advancing Quality Science
© 2006 Applied Biosystems
Antiretroviral agents AntidepressantsAmino acids
ImmunosuppressantsMMA
Vitamins
Electrospray
AnalyteAnalyte
Ionisation Methods of Choice
What effects the way compounds form an ions?
50 24th April 2006
Applied Biosystems — Advancing Quality Science
© 2006 Applied Biosystems
Acid or base (pH)?
● For compounds containing Nitrogen, a low pH or high concentration of H+ ions will increase the amount of compound forming the ions (typically add formic or acetic acid to the LC system).
● For compounds containing an acid group, a higher pH or lower concentration of H+ ions will increase the amount of compound forming the ions (typically add ammonia to the LC system).
51 24th April 2006
Applied Biosystems — Advancing Quality Science
© 2006 Applied Biosystems
Can Standard LC conditions be Used?
● Yes, but some LC Buffers may cause problems.● Typically methanol and acetonitrile systems buffered with
ammonia acetate or formate are used in LC-MS.● Flow rates for nL/min to mL/min● One example of a troublesome LC-MS buffer is
Phosphate buffer and the use of a non-volatile ion pairing agent (e.g. SDS) which can can cause severe suppression and complex spectra.
52 24th April 2006
Applied Biosystems — Advancing Quality Science
© 2006 Applied Biosystems
LC/MS/MS Summary● LC-MS/MS offers several advantages over more traditional
analytical techniques including higher sensitivity and selectivity.
● There are many different types of LC-MS/MS systems which vary functionality. It is important to understand the type of analyses intended the technology.
● There are different types of source available. Having an idea of the polarity of the compound will guide you to the most appropriate source, but generally Electrospray is the most flexible.
● Take care when transferring methods to LC-MS from LC with UV detection as the original buffer may not be ideal.
53 24th April 2006
Applied Biosystems — Advancing Quality Science
© 2006 Applied Biosystems
Tandem Mass Spectrometry SystemsFor Clinical Research
API 2000™system
Entry level system
API 4000™ system
High performance, high sensitivity
systemfor applications such as targeted steroid analysis
API 5000™ System
Highestperformance,
ultimate sensitivity
system
API 3200™ System
Economichigh performance
systemIdeal for
TDM / IEM / Steroids / etc.
Increased Sensitivity
54 24th April 2006
Applied Biosystems — Advancing Quality Science
© 2006 Applied Biosystems
Use of TMS in Clinical Research ApplicationsTDM
ImmunosuppressantAntiretroviral
AntipsychoticsAnticonvulsive
AntiviralBenzodiazepine
3200
NBS
30-40 DiseasesFatty Acid
Oxidation DisorderAmino AcidDisorders
Organic AcidDisorders
Other Diseases
3200
Biomarker
SteroidsT3/T4
Vitamin DHomocysteine
Orotic AcidVLCFA
Bile AcidCancer
Etc.3200 to 5000
Clin Tox
General unknown Screening (GUS)
Drugs Of Abuse (DOA)
Triple / QTRAP
55 24th April 2006
Applied Biosystems — Advancing Quality Science
© 2006 Applied Biosystems
LC/MS/MS for Research on
Therapeutic Drug Monitoring
56 24th April 2006
Applied Biosystems — Advancing Quality Science
© 2006 Applied Biosystems
Why adopt Tandem Mass Spectrometry for TDM?● Cost advantage vs Immunoassays
● Time To Result – Increased speed and throughput
● Flexibility adding new drugs to the panel
● Simultaneous detection of all drugs
● More accurate quantitation due to less cross reactivity with metabolites – Increased selectivity and specificity
● LC/MS/MS shows excellent reproducibility and dynamic range
● Very robust assay
57 24th April 2006
Applied Biosystems — Advancing Quality Science
© 2006 Applied Biosystems
● Immunosuppressants–Cyclosporin A–Tacrolimus–Sirolimus–Everolimus–Mycophenolic Acid (MPA)
● LC/MS/MS, the ideal alternative for immunoassays● Simultaneous detection and quantitation● Excellent sensitivity and very good signal-to-noise ratio
Sirolimus
Clinical Research: Therapeutic Drug Monitoring
58 24th April 2006
Applied Biosystems — Advancing Quality Science
© 2006 Applied Biosystems
Instrument configuration: modified setting to include MPA
API 3200 system
AB POROS® R1/20 Perfusion Column
API 3200 built-in switching valve.
Autosampler
Pump B
Pump A
Phenomenex Column Luna Phenyl-Hexyl
Column oven at 60 d.C
Methanol 10% +0.1 % Acetic Ac
Methanol 97% + 0.1 % Acetic Ac. + 10 mM CH3COONH4
59 24th April 2006
Applied Biosystems — Advancing Quality Science
© 2006 Applied Biosystems
XIC of +MRM (14 pairs): 931.6/864.8 amu from Sample 11 (E 4) of 030806 Patienten Merc Ph... Max. 2700.0 cps.
0.2 0.4 0.6 0.8 1.0 1.2 1.4 1.6 1.8 2.0 2.2 2.4 2.6 2.8 3.019 38 57 76 95 114 133 152 171 190 209 228 247 266 286
Time, min
0.0
2000.0
4000.0
6000.0
8000.0
1.0e4
1.2e4
1.4e4
1.6e4
1.8e4
2.0e4
2.2e4
2.4e4
2.6e4
2.8e4
3.0e4
3.2e4
1.20
0.19 0.28 0.660.48 2.201.570.78 2.51
XIC of +MRM (14 pairs): 931.6/864.8 amu from Sample 10 (E4) of 030807 Patienten Bruno Ca... Max. 6800.0 cps.
0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8 0.9 1.0 1.1 1.2 1.3 1.4 1.5 1.6 1.7 1.8 1.9 2.0 2.1 2.2 2.3 2.4 2.510 19 29 38 48 57 67 76 86 95 105 114 124 133 143 152 162 171 181 190 200 209 219 228 238
Time, min
0.00
1.00e4
2.00e4
3.00e4
4.00e4
5.00e4
6.00e4
7.00e4
8.00e4
9.00e4
1.00e5
1.10e5
1.17e5
2.04
New 2 D LC-MS/MS methodology >
< Previous LC-MS/MSmethodology
60 24th April 2006
Applied Biosystems — Advancing Quality Science
© 2006 Applied Biosystems
61 24th April 2006
Applied Biosystems — Advancing Quality Science
© 2006 Applied Biosystems
Fast Separation and Detection of Multiple Immunosuppressants and Internal Standards
62 24th April 2006
Applied Biosystems — Advancing Quality Science
© 2006 Applied Biosystems
Calibration Curves with Blood Calibrators
63 24th April 2006
Applied Biosystems — Advancing Quality Science
© 2006 Applied Biosystems
Dynamic Range and Linearity for EVEROLIMUS
64 24th April 2006
Applied Biosystems — Advancing Quality Science
© 2006 Applied Biosystems
0 250 500 750 1000
0
250
500
750
1000
1250
Y = 1.06X + 8.404
CsA Co nc ent r at io n b y L C/ MS/ MS ( ng / m L)
CsA Co nc .b y EMIT( ng / mL)
CyclosporinA: EMIT Immunoassay vs. LC/MS/MS
65 24th April 2006
Applied Biosystems — Advancing Quality Science
© 2006 Applied Biosystems
Cyclosporin A Tacrolimus Sirolimus Everolimus (CyA) (TRL) (SRL) (RAD)
LOD 1.3 0.1 0.1 0.1 [ng ml-1]
Linearity 0.9999 0.9997 0.9998 0.9998 [R2]
Recovery [%] concentration 1 88.3 114.4 116.9 118.8
concentration 2 103.7 101.9 114.5 102.8
RSD [%] Intra-day (n=100) 4.6 10.2 9.3 10.3
RSD [%] Inter-day (n=40) 12.3 13.5 12.3 10.9
Accuracy [%] 87.7 – 115.8 90.3 – 110.8 91.0 - 107.5 91.2 – 112.1 (n=40) (Mean: 99.2) (Mean: 96.9) (Mean: 98.9) (Mean: 103.1)
Koal et al.: J.Chromatogr. B 805 (2004) 215-222
Method performance parameters determined for 25 µl spiked blood sample
66 24th April 2006
Applied Biosystems — Advancing Quality Science
© 2006 Applied Biosystems
MPA features
● MPA has a free acidic moiety which should make it
more sensitive in Negative Ion Mode.
● MPA is quite labile in the ionization process.
● Reported to be metabolized at a significant extent
(mainly to the Glucuronide conjugate).
● MPA prone to stick on surfaces.
67 24th April 2006
Applied Biosystems — Advancing Quality Science
© 2006 Applied Biosystems
Fast 2D-LC MS/MS Method including MPA
68 24th April 2006
Applied Biosystems — Advancing Quality Science
© 2006 Applied Biosystems
Conclusions
● LC/MS/MS technology and principles
● LC/MS/MS is superior to conventional methods with regard to specificity, accuracy, dynamic range and cost per assay
● LC/MS/MS is easy to use
● Wide range of potential uses for Therapeutic Drug Monitoring, Inborn Errors of Metabolism, Clinical Biomarkers – growing research into these areas focused on mass spectrometry.
Thank You For Your Attention!
69 24th April 2006
Applied Biosystems — Advancing Quality Science
© 2006 Applied Biosystems
● Some of the publications cited in this presentation present off-label uses of Applied Biosystems products. Applied Biosystems’ citing of these publications does not constitute an endorsement of, nor warranty of, the data or conclusions contained in these publications.
● Applied Biosystems products have not been cleared or approved by the U.S. FDA or any other agency, or under the European IVD directive, for any purpose. Applied Biosystems does not promote or sell its instruments or other products for diagnostic or other clinical use.
● No claim or representation is made or intended by Applied Biosystems (i) as to any clinical use of any product (whether diagnostic, prognostic, therapeutic, blood banking or any other clinical use), or (ii) that any product has been cleared, approved, registered or otherwise qualified by Applied Biosystems with the united states Food and Drug Administration or any other regulatory agency for use in any diagnostic or other clinical procedure. Applied Biosystems’ products are labeled “For Research Use Only. Not For Use in Diagnostic Procedures,” or are General Laboratory Instruments.
70 24th April 2006
Applied Biosystems — Advancing Quality Science
© 2006 Applied Biosystems
For Research Use Only. Not for use in diagnostic procedures.
Applera is a trademark and Applied Biosystems, POROS and AB (design) are registered trademarks of Applera Corporation or its subsidiaries in the U.S. and/or certain other countries.
API 2000, API 3200, API 4000, API 5000, Turbo V, Curtain Gas and Ionspray are trademarks. Photospray and LINAC are registered trademarks of Applied Biosystems/MDS SCIEX a joint venture between Applera Corporation, and MDS SCIEX Inc.
All other trademarks are the sole property of their respective owners
©2006 Applied Biosystems. All rights reserved.
Trademark Acknowledgements