intraocular pressure measurement accuracy and...

9
Intraocular Pressure Measurement Accuracy and Repeatability of a Modified Goldmann Prism: Multicenter Randomized Clinical Trial SEAN MCCAFFERTY, KYLE TETRAULT, ANN MCCOLGIN, WARREN CHUE, JASON LEVINE, AND MELISSA MULLER PURPOSE: To clinically evaluate a modified surface Goldmann applanation tomometer (GAT) prism for intraocular pressure (IOP) accuracy, repeatability, and safety. DESIGN: Prospective, open-label, randomized, controlled, multicenter reference device reliability and validity analysis. METHODS: A GAT and a modified surface GAT prism measured IOP on 173 unique eyes. The study design and analysis complied with FDA 510(k) and ANSIZ80.10- 2014 guidelines. All eyes were randomized to IOP measure- ment by 1 of 5 standard prisms or 5 modified prisms, each from a different manufacturing lot. Pressures were measured by 6 investigators, 2 times with each prism, for a total of 1384 IOP measurements. Analysis included Bland- Altman difference accuracy, intraoperator and interoperator IOP measurement, and manufactured lot repeatability. RESULTS: Bland-Altman indicated no IOP measure- ments pairs outside the ±5 mm Hg guidelines. Operator and manufactured lot repeatability F tests and 1-way ANOVAs indicated no statistical difference between the standard and modified prisms (all P > .10). The dif- ference in IOP measurements of the standard and modi- fied prisms correlated well to Dresdner central corneal thickness (CCT) correction (P [ .01). CONCLUSION: A modified surface replacement prism is statistically equivalent to a flat-surfaced prism. The modi- fied surface prism indicated statistically significant correction for CCT requiring further testing outside the ANSI standard limits (0.500 mm < CCT < 0.600 mm) to examine its full potential. (Am J Ophthalmol 2018;196:145–153. Ó 2018 The Author(s). Published by Elsevier Inc. This is an open access article under the CC BY license (http:// creativecommons.org/licenses/by/4.0/).) F OR 60 YEARS, GOLDMANN APPLANATION TONOM- etry has been established as the standard for measure- ment of intraocular pressure (IOP). However, biomechanical variability in corneal thickness, rigidity, curvature, or corneal tear film among patients can lead to numerous errors in the Goldmann applanation tonometer (GAT) IOP measurement. These additive errors may result in potentially sight-threatening conditions in a significant population of patients such as those with glaucoma or undiagnosed ocular hypertension from other causes. 1 Data from the Ocular Hypertension Treatment Study (OHTS) first noted that IOP readings tend to be overesti- mated in thick corneas, whereas they are underestimated in thinner corneas. This error may lead to a misdiagnosis of glaucoma. 2 Based on these findings, the preferred practice has since changed to include a measurement of central corneal thickness (CCT) with a nomogram to adjust the pressure based on CCT. 3–5 Additionally, it is recognized that the effects of LASIK surgery and the corneal biomechanical differences found in children render accurate IOP measurement by the GAT problematic. 6–9 Attempts have been made to correct the various error components in GAT measurement to yield a standard IOP reading that is comparable between patients. 10,11 However, the process in practice is error prone and cumbersome, leading to almost no clinical adoption by clinicians, with the exception of CCT. The correcting applanation tonometry surface (CATS) tonometer prism is a modification of the standard GAT prism. The CATS prism is cleared by the U.S. Food and Drug Administration (FDA) for the measurement of IOP in applanation tonometers. The only alterations to the original GAT design are the modification of the flat appla- nating surface to a sinusoidal curved surface and a compen- satory lengthening of the prism. 10 The compensatory lengthening of the prism was necessary to maintain a zero average bias between the GAT and CATS prisms over a large standard patient population. The zero bias maintains long-established GAT IOP benchmarks (ie, 16 mm Hg as average normal and 21 mm Hg as borderline high). The prism was designed to significantly decrease patient IOP dependence on corneal biomechanical and tear-film prop- erties, which is the primary source of individual error. No new materials were introduced in the design process (Figure 1, Supplemental Video; Supplemental Material Supplemental Material available at AJO.com. Accepted for publication Aug 30, 2018. From Arizona Eye Consultants, Intuor Technologies, Tucson, Arizona, USA (S.M.); Denver Eye Surgeons, Lakewood, Colorado, USA (K.T.); Cornea Associates, Tucson, Arizona, USA (A.M.); and Arizona Eye Consultants, Tucson, Arizona, USA (W.C., J.L., M.M.). Inquiries to Sean McCafferty, Arizona Eye Consultants, Intuor Technologies, 6422 E. Speedway Blvd, Tucson, AZ 85710, USA; e-mail: [email protected] 0002-9394 https://doi.org/10.1016/j.ajo.2018.08.051 145 © 2018 THE AUTHOR(S). PUBLISHED BY ELSEVIER INC.

Upload: others

Post on 31-May-2020

5 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Intraocular Pressure Measurement Accuracy and ...doclibrary.com/MSC167/PRM/cats_multicenter_trial_AJO4250.pdfbiomechanical variability in corneal thickness, rigidity, curvature, or

SAccepted fo

From ArizUSA (S.M.)Cornea AssConsultants,

InquiriesTechnologiee-mail: SJMc

0002-9394https://doi.or

Intraocular PressureMeasurement Accuracy andRepeatability of a Modified Goldmann Prism:

Multicenter Randomized Clinical Trial

SEAN MCCAFFERTY, KYLE TETRAULT, ANN MCCOLGIN, WARREN CHUE, JASON LEVINE, ANDMELISSA MULLER

� PURPOSE: To clinically evaluate a modified surfaceGoldmann applanation tomometer (GAT) prism forintraocular pressure (IOP) accuracy, repeatability, andsafety.� DESIGN: Prospective, open-label, randomized,controlled, multicenter reference device reliability andvalidity analysis.� METHODS: A GAT and a modified surface GAT prismmeasured IOP on 173 unique eyes. The study design andanalysis complied with FDA 510(k) and ANSIZ80.10-2014 guidelines.All eyes were randomized to IOPmeasure-ment by 1 of 5 standard prisms or 5 modified prisms, eachfromadifferentmanufacturing lot.Pressuresweremeasuredby 6 investigators, 2 times with each prism, for a total of1384 IOP measurements. Analysis included Bland-Altmandifferenceaccuracy, intraoperator and interoperatorIOP measurement, and manufactured lot repeatability.� RESULTS: Bland-Altman indicated no IOP measure-ments pairs outside the ±5 mm Hg guidelines. Operatorand manufactured lot repeatability F tests and 1-wayANOVAs indicated no statistical difference betweenthe standard and modified prisms (all P > .10). The dif-ference in IOP measurements of the standard and modi-fied prisms correlated well to Dresdner central cornealthickness (CCT) correction (P [ .01).� CONCLUSION: A modified surface replacement prism isstatistically equivalent to a flat-surfaced prism. The modi-fied surface prism indicated statistically significantcorrection for CCT requiring further testing outside theANSI standard limits (0.500 mm < CCT <0.600 mm) to examine its full potential. (Am JOphthalmol 2018;196:145–153. � 2018 TheAuthor(s). Published by Elsevier Inc. This is an openaccess article under the CC BY license (http://creativecommons.org/licenses/by/4.0/).)

upplemental Material available at AJO.com.r publication Aug 30, 2018.ona Eye Consultants, Intuor Technologies, Tucson, Arizona,; Denver Eye Surgeons, Lakewood, Colorado, USA (K.T.);ociates, Tucson, Arizona, USA (A.M.); and Arizona EyeTucson, Arizona, USA (W.C., J.L., M.M.).to Sean McCafferty, Arizona Eye Consultants, Intuors, 6422 E. Speedway Blvd, Tucson, AZ 85710, USA;[email protected]

g/10.1016/j.ajo.2018.08.051© 2018 THE AUTHOR(S). PU

FOR 60 YEARS, GOLDMANN APPLANATION TONOM-

etry has been established as the standard for measure-ment of intraocular pressure (IOP). However,

biomechanical variability in corneal thickness, rigidity,curvature, or corneal tear film among patients can lead tonumerous errors in the Goldmann applanation tonometer(GAT) IOPmeasurement. These additive errors may resultin potentially sight-threatening conditions in a significantpopulation of patients such as those with glaucoma orundiagnosed ocular hypertension from other causes.1

Data from the Ocular Hypertension Treatment Study(OHTS) first noted that IOP readings tend to be overesti-mated in thick corneas, whereas they are underestimated inthinner corneas. This error may lead to a misdiagnosis ofglaucoma.2 Based on these findings, the preferred practicehas since changed to include a measurement of centralcorneal thickness (CCT) with a nomogram to adjust thepressure based on CCT.3–5 Additionally, it is recognizedthat the effects of LASIK surgery and the cornealbiomechanical differences found in children renderaccurate IOP measurement by the GAT problematic.6–9

Attempts have been made to correct the various errorcomponents in GAT measurement to yield a standardIOP reading that is comparable between patients.10,11

However, the process in practice is error prone andcumbersome, leading to almost no clinical adoption byclinicians, with the exception of CCT.The correcting applanation tonometry surface (CATS)

tonometer prism is a modification of the standard GATprism. The CATS prism is cleared by the U.S. Food andDrug Administration (FDA) for the measurement of IOPin applanation tonometers. The only alterations to theoriginal GAT design are the modification of the flat appla-nating surface to a sinusoidal curved surface and a compen-satory lengthening of the prism.10 The compensatorylengthening of the prism was necessary to maintain a zeroaverage bias between the GAT and CATS prisms over alarge standard patient population. The zero bias maintainslong-established GAT IOP benchmarks (ie, 16 mm Hg asaverage normal and 21 mm Hg as borderline high). Theprism was designed to significantly decrease patient IOPdependence on corneal biomechanical and tear-film prop-erties, which is the primary source of individual error. Nonew materials were introduced in the design process(Figure 1, Supplemental Video; Supplemental Material

145BLISHED BY ELSEVIER INC.

Page 2: Intraocular Pressure Measurement Accuracy and ...doclibrary.com/MSC167/PRM/cats_multicenter_trial_AJO4250.pdfbiomechanical variability in corneal thickness, rigidity, curvature, or

FIGURE 1. Rendering of the applanating surfaces of the Gold-mann applanation tonometer (GAT) and correcting applanationtonometry surface (CATS) prisms demonstrating the flat(GAT) vs centrally concave and annularly convex (CATS)prisms.

available at AJO.com). The CATS prism has demon-strated improved IOP accuracy mathematically and bydirect GAT comparison, as well as by comparison to trans-ducer intracameral pressure.12–16

The CATS prism uses the existing Goldmann measure-ment armature without recalibration and the clinicianmeasures IOP using the same measurement protocol andtechniques as the current GAT prism. Sensitivity to thecorneal biomechanical error parameters is reduced bypartially matching the negative curvature of the tonometerto positive curvature of the cornea.12–16 The scientificpremise of the CATS tonometer prism is a minimizationof the intracorneal stress during applanation. Thereduced intracorneal stress minimizes the contribution ofthe corneal deformation to the total force on the prismface, thus measuring predominantly the force due toIOP.12 Simultaneously, the annular outer curvature(away from the cornea) minimizes the tear-film adhesionforce and subsequent tear-film error.12,16

The present study was designed to evaluate the perfor-mance accuracy of the CATS prism as compared to thereference GAT tonometer prism in IOP measurement inhealthy subjects. The study was designed to comply withISO 8612:2009, ANSI Z80.10-2014, and Guidance forIndustry and FDA Staff Tonometers – Premarket Notifica-tion [510(k)] Submissions standards.

146 AMERICAN JOURNAL OF

METHODS

HEALTHY ADULT SUBJECTS WERE ENROLLED TO A SERIES OF

IOP measurements with the CATS tonometer prism andthe GAT prism at a single visit. The study was designedas a prospective, open-labeled, randomized, controlled,multicenter reference device comparison. It was prospec-tively registered with clinicaltrials.gov underNCT02989909 and prospectively institutional reviewboard (IRB) approved with informed consent by Chesa-peake IRB. Eligible subjects were screened, enrolled, andevaluated according to the study protocol on the sameday (day 1). Subjects were recruited from 2 U.S. sites inArizona. Subjects meeting the inclusion and exclusioncriteria, 18 years or older, who provided written informedconsent participated in the study.This clinical study was conducted in accordance with

the ethical principles contained within the Declarationof Helsinki, Protection of Human Volunteers (21 CFR50), Institutional Review Boards (21 CFR 56), and Obliga-tions of Clinical Investigators (21 CFR 812).

� DESCRIPTION OF STUDY POPULATION: Healthy adultsubjects, including those with refractive errors and highastigmatism, as well as high-IOP subjects, were enrolledinto the study. The selection of the sample size was basedon ISO 8612:2009, ANSI Z80.10-2014, and Guidance forIndustry and FDA Staff Tonometers – Premarket Notifica-tion [510(k)] Submissions.The subjects with the following conditions were

excluded from participation in the study: subjects withcorneal scarring; subjects with lid, corneal, or ocular condi-tions, disease, disorders, or infection that may haveconfounded the study results; and subjects with uncon-trolled systemic disease that in the opinion of the investi-gator would put the subject’s health at risk. Additionally,pregnant or nursing women and contact lens wearerswere excluded. Ocular surgery within 3 months of enroll-ment or corneal surgery at any time was prohibited.Furthermore, eyes displaying an oval contact image orCCT > 0.600 mm or < 0.500 mm were excluded.

� PROTOCOL: Each subject underwent a standardophthalmic examination by 1 of 6 trained and licensed in-vestigators. A Zeiss HD-OCT-5000 spectral-domain opti-cal coherence tomographer (Zeiss, Jena, Germany) wasused by the assistant to measure CCT. Finally, the assistantinvestigator completed a corneal topography with a ZeissAtlas model 9000 (Jena, Germany) and a corneal curvaturewas used for analysis over the central 3-mm diameter of thecornea in accordance with ANSI Z80.23. Each investigatorconducting IOP measurements was masked to the results ofthe assistant investigator’s tests. Investigators were alsomasked to the randomized and alternated use of theCATS and GAT prism. Use of the test or reference devicewas chosen by random number generator and each

DECEMBER 2018OPHTHALMOLOGY

Page 3: Intraocular Pressure Measurement Accuracy and ...doclibrary.com/MSC167/PRM/cats_multicenter_trial_AJO4250.pdfbiomechanical variability in corneal thickness, rigidity, curvature, or

TABLE 1. Intraocular Pressure Measurement CategoryTolerances for Demonstrating Paired Difference Substantial

Equivalence

IOP Range (mm Hg) Tolerance of Paired Differences (mm Hg)

7 to 16 6 5

>16 to < 23 6 5

>_ 23 6 5

IOP ¼ intraocular pressure.

TABLE 2. Intraocular Pressure Measurements in EachPressure and Astigmatism Category

Astigmatism

IOP

Total

Low

(<_16 mm Hg)

Medium

(>16 to <23 mm Hg)

High

(>_23 mm Hg)

Low (<3 D) 42 57 40 139

High (>_3 D) 13 10 11 34

Total 55 67 51 173

D ¼ diopter; IOP ¼ intraocular pressure.

Results are number of eyes.

subsequent IOP measurement was alternated. Topicalanesthetic drops with fluorescein (fluorescein sodium andbenoxinate hydrochloride ophthalmic solution 0.25%/0.4%; Bausch& Lomb, Tampa, Florida, USA) were appliedprior to each measurement so that examination conditionswere equivalent. Measurements were conducted on a dailycalibrated Haag-Streit model 900 applanation tonometerarmature (Haag-Streit, Mason, Ohio, USA) using 1 of 5cleaned and disinfected Haag-Streit standard GAT prismsand alternatively 1 of 5 cleaned and disinfected CATSprisms. Measurements of IOP were made 2 times with theCATS prism (1 measurement was considered by averagingmeasurements at 180 and 90 degrees to correct for astigma-tism) and 2 times with the Goldmann prism (again axisaveraged). If the sequential measurements with 1 prismwere more than 2 mm Hg different, then a third measure-ment was obtained. All 3 measurements were then aver-aged. The third measurements was included in the studyif it was within the range of the first 2; otherwise all mea-surements were discarded. Four measurements were taken(2 with each prism, 4 total) with at least 5 minutes, butno more than 10 minutes, between each measurement.

Test product was 5 CATS tonometer prisms fromdifferent lot numbers. Control product was 5 Haag-StreitGAT prisms from different lot numbers.

� OBJECTIVES: The primary performance objective of theclinical study was to demonstrate equivalence of theCATS tonometer prism (test) to the GAT prism (control)in the measurement of IOP.

The secondary performance objective was to assess theoperator repeatability of IOPmeasurement. An exploratoryobjective was to examine improved IOP accuracy in accor-dance with expected standard Dresdner CCT correction forIOP.

� ENDPOINTS: The primary performance endpoint was ac-curacy of IOPmeasurement readings, as measured by appla-nation tonometry. The success criterion for the primaryendpoint was that no more than 5% of the paired differ-ences between the reference tonometer (GAT prism) andthe test tonometer (CATS prism) readings would begreater than ANSI Z80.10-2014 prescribed tolerance levels(Table 1).

VOL. 196 ACCURACY AND REPEATABILITY OF A

IOP measurements were taken from participating sub-jects, using the CATS tonometer prism and the GATprism. Enrollment continued until a minimum of 40eligible eyes were identified in each of the prespecifiedIOP range categories as defined by ISO 8612:2009 andANSI Z80.10-2014. At least 10 highly astigmatic eyes(>3 diopter [D] of corneal astigmatism) were required ineach of the low (7–16 mm Hg), medium (>16 to<23 mm Hg), and high (>_23 mm Hg) IOP ranges in thestudy.

� STATISTICAL METHODS: The Full Analysis Set (FAS)was the primary analysis set for the analysis of the effective-ness endpoint and the safety data. All eligible eyes wereincluded in the FAS.Descriptive statistics were provided on continuous vari-

ables includingmean, standard deviation,median, and range.An assessment of change from baseline was provided for IOP,using 2-sided confidence intervals (Cis) and a ¼ 0.05 or1-sided CIs and a ¼ 0.025. The primary endpoint wasanalyzed per the ANSI Z80.10-2014 using a total leastsquares regression (TLSR) as well as a Bland-Altman-typepaired differences analysis according to the above-cited table.For the secondary performance endpoint, a homoscedas-

tic, 2-tailed t test (a ¼ 0.05) was performed, as well as ananalysis of variance (ANOVA) to determine the variancebetween the individual operator readings and collectivereadings. Furthermore, a Student t test and ANOVAwere completed for the analysis of difference in means be-tween the CATS and GAT measurements within individ-ual operators. Finally, a linear regression analysis of thedifference in paired GAT and CATS IOP measurementswas correlated to CCT.

RESULTS

ONE HUNDRED SEVENTY-THREE EYES WERE MEASURED.

There were 139 eyes in the low (<3 D) astigmatism groupand 34 eyes in the high (>3 D) astigmatism group. The

147MODIFIED GOLDMANN PRISM

Page 4: Intraocular Pressure Measurement Accuracy and ...doclibrary.com/MSC167/PRM/cats_multicenter_trial_AJO4250.pdfbiomechanical variability in corneal thickness, rigidity, curvature, or

TABLE 3. Summary of Prism Effectiveness Objectives, Endpoints, and Statistical Methods

Objective Endpoint Methods and Summary Statistics Used for Analysis

Primary:

To demonstrate equivalence of the CATS

tonometer prism to the GAT prism in the

measurement of IOP.

Primary:

Accuracy of IOP measurement: no more

than 5% of the paired differences

between the reference GAT tonometer

(Goldmann prism) and the test tonometer

(CATS prism) readings for each pressure

range are greater than the tolerance level

of 65 mm Hg.

Primary:

Descriptive statistics including mean, SD,

median, min, andmax, using 95%CI, and

a ¼ 0.05.

Scatterplots of the test and reference

measurements as well as Bland-Altman

paired difference analysis.

Secondary:

To assess the inter-operator reliability of

IOP measurement interpretation.

Secondary:

Repeatability of IOP measurement

compared to GAT found statistically

equivalent.

Secondary:

Summary statistics of the individual and

overall measurement variations, t test, F

test, 1-way ANOVA, and a GLME

analysis.

Exploratory:

CATS-GAT Correction Correlation to CCT

Exploratory:

CATS CCT corrected IOP measurement

similar to standard GAT correction.

Exploratory:

Descriptive statistics of test minus

reference measurements.

CATS ¼ correcting applanation tonometry surface; CCT ¼ central corneal thickness; GAT ¼ Goldmann applanation tomography; GLME ¼Generalized Linear Mixed Effects; IOP ¼ intraocular pressure.

FIGURE 2. Scatterplot of the average intraocular pressure(IOP) values from each of the 173 measurements for Goldmannapplanation tonometer (GAT) and correcting applanationtonometry surface (CATS) prisms along with a reference y [x line.

analysis included 170 unique eyes (Table 2). The only sub-set that included repeat measurements was the group thatrequired both high astigmatism (>3 D) and high IOP(>_23 mmHg). This group included 8 unique eyes for a totalof 11 separate measurements to meet the required 10 mea-surements in this category per ANSI Z80.10-2014. Mea-surements within all groups were completed inaccordance with the ANSI Z80.10-2014 guidelines. Therewere 60 (36%) male and 110 (64%) female subjects whocompleted the study. The average age was 58.0 6 18.6years. There was no outlier data in the paired difference pri-mary effectiveness described in Table 3 and as defined byANSI A80.10-2014.

� ACCURACY OF INTRAOCULAR PRESSURE (PRIMARYEFFECTIVENESS ENDPOINT): Figure 2 shows the CATSvs GAT prism scatterplot of all data according to FDATonometer Guidelines section 5a: Scatter Plot of MeasuredIOP Values. It includes the regression line slope, intercept,and correlation coefficient. The scatterplot indicates excel-lent correlation between the CATS and GAT IOP mea-surements throughout the useful range of measurementIOPs with a correlation coefficient of R2 ¼ 0.95,slope ¼ þ0.97, and y-intercept of þ1.04.

The subgroups of low (<3 D) astigmatism and high (>3D) astigmatism are broken out in Figure 3. The (<3 D)astigmatism group indicates a correlation coefficient ofR2 ¼ 0.95, with a slope of þ0.99 and a y-interceptof þ0.68. The>3 D astigmatism subgroup shows a correla-tion coefficient of R2¼ 0.96, with a slope ofþ0.88 and a y-intercept of þ2.47. Both the high and low astigmatismgroups indicated an excellent correlation between CATS

148 AMERICAN JOURNAL OF

and GAT IOP measurements meeting the primary studyendpoint.The Bland-Altman-type plot in Figure 4 shows the

paired differences of the average IOP between the CATSand GAT according to ANSI Z80.10-2014 Section 4.2.1Table 1. The <_3 D astigmatism group indicates a mean dif-ference ofþ0.416 1.49 mmHg (61 SD). The>3 D astig-matism subgroup shows a mean difference of þ0.18 61.43 mm Hg (61 SD). All paired measurement differencesreside within the þ5 mm Hg tolerance given in Table 1 inboth groups and satisfy the ANSI requirement on the

DECEMBER 2018OPHTHALMOLOGY

Page 5: Intraocular Pressure Measurement Accuracy and ...doclibrary.com/MSC167/PRM/cats_multicenter_trial_AJO4250.pdfbiomechanical variability in corneal thickness, rigidity, curvature, or

FIGURE 3. Scatterplots for the (Left) low astigmatism (£3 di-opters) and (Right) high astigmatism (>3 diopters) measure-ments.

FIGURE 4. Bland-Altman-type plot, low astigmatism (£3 diop-ters) and high astigmatism (>3 diopters).

paired testing of reference tonometer with test tonometers.A separate analysis of paired differences in the IOP rangebetween 25 and 35 was completed and shown to be statis-tically equivalent by paired t test (P ¼ .28).

The TLSR fits of the CATS average values against theGAT average values are shown separately in both thehigh (>3 D) and low (<_3 D) astigmatism categories inFigure 5.

The slope of the low (<3 D) astigmatism TLSR isþ1.02,the offset of the regression is þ0.11 mm Hg, and the stan-dard error of the regression is 1.03 mm Hg (Figure 6). Theslope of the high (>3 D) astigmatism TLSR is þ0.90, theoffset of the regression is þ1.96 mm Hg, and the standarderror of the regression is 0.93 mm Hg. The results supportthe primary endpoint of the study and indicate statisticalequivalence between the CATS andGAT prisms satisfyingANSI Z80.10-2014 4.2.2: Total Least Squares Regression,in both high and low astigmatism groups.

The slope of the combined astigmatism group regressionis þ0.99, the offset of the regression is þ0.55 mm Hg, andthe standard error of the regression is 1.04 mm Hg. Thecombined astigmatism groups’ TLSR confirms the statisti-cal equivalence between the CATS and GAT prisms.

VOL. 196 ACCURACY AND REPEATABILITY OF A

� REPEATABILITY OF INTRAOCULAR PRESSURE (SECOND-ARY EFFECTIVENESS ENDPOINT): Figure 7 shows the scat-terplot with the markers colored by the operator (personperforming the measurement). The spread of operatorsacross the GAT average IOP value shows little visual indi-cation of tonometer operator bias between the CATS andGAT prisms.

� INTRAOPERATOR MEASUREMENT REPEATABILITY:

Each IOP measurement consisted of 2 separate measure-ments with the CATS and the GAT prisms. A 2-samplet test on the mean IOP examining intraoperator IOP mea-surement repeatability produced a P value of .524. There-fore, at the 5% significance level, there is no significantdifference in repeatability between the CATS and GATprisms.A 2-sample F test on the standard deviation examining

intraoperator IOP measurement repeatability produced aP value of .132. Therefore, at the 5% significance level,there is no significant difference in repeatability betweenthe CATS and GAT prisms.Each of the 173 lines of measurement consisted of 2

GAT measurements and 2 CATS measurements. A statis-tical analysis examined the difference (delta) between thefirst and second measurements with both the CATS and

149MODIFIED GOLDMANN PRISM

Page 6: Intraocular Pressure Measurement Accuracy and ...doclibrary.com/MSC167/PRM/cats_multicenter_trial_AJO4250.pdfbiomechanical variability in corneal thickness, rigidity, curvature, or

FIGURE 5. Total least squares regression of correcting appla-nation tonometry surface (CATS) average values against Gold-mann applanation tonometer (GAT) average values, low (£3diopters) and high (>3 diopters) astigmatism.

FIGURE 6. Total least squares regression regression of correct-ing applanation tonometry surface (CATS) average valuesagainst Goldmann applanation tonometer (GAT) averagevalues, all tested eyes, combined astigmatism groups.

FIGURE 7. Scatter plot of average intraocular pressure mea-surements, colored by operator performing measurement (notedby initials).

GAT prisms. The variances were found to be almost iden-tical, as shown in the box plot in Figure 8.

� INTEROPERATOR INTRAOCULAR PRESSURE MEASURE-MENT REPEATABILITY: Each of the 173 lines of measure-ments were performed by 1 of 6 operators where each lineconsists of 2 GAT measurements and 2 CATS measure-ments.

The means for variability in paired IOP measurement forall individual operators were compared, using a 1-wayANOVA (a¼ 0.05 significance level). The observed differ-ence in interoperator variability was not statistically signifi-cant (P ¼ .1487), indicating no significant interoperatorvariability in the measurement of IOP with the CATS prism(Figure 9).

The 1-way ANOVA was performed on the differences ofthe CATS IOP readings from the GAT IOP readings. Inother words, the ANOVA assumes the GAT reading to bethe true IOP of the eye and measures only the variance forthe CATS IOP readings. It should be noted, however, thatthis is a very conservative statistical comparison. TheGAT devices themselves have some error in measurement.

150 AMERICAN JOURNAL OF

From the perspective of the 1-way ANOVA, the effects ofthe GAT interoperator variability are observed as solelythe effects of the CATS interoperator variability. Evenwith this conservative approach the CATS prism is not sta-tistically different than the GAT prism. Considering theinfeasibility of a third true intracameral IOP measurementcomparison, this approachwas determined to bemost appro-priate and robust in analyzing interoperator variability.

� LOTNUMBERVARIABILITY: The means for variability inIOP measurement for all CATS prism manufactured lotnumbers were compared, using a 1-way ANOVA (a ¼0.05 significance level). The observed difference(Figure 10) in interoperator variability was not statisticallysignificant (P¼ .090), indicating no significant lot number

DECEMBER 2018OPHTHALMOLOGY

Page 7: Intraocular Pressure Measurement Accuracy and ...doclibrary.com/MSC167/PRM/cats_multicenter_trial_AJO4250.pdfbiomechanical variability in corneal thickness, rigidity, curvature, or

FIGURE 8. Difference in first measurement from second mea-surement using Goldmann applanation tonometer (GAT) andcorrecting applanation tonometry surface (CATS) tonometerprisms.

FIGURE 9. One-way analysis of variance (ANOVA) for inter-operator variability ANOVA and box plot outputs.

FIGURE 10. Box plot of delta measurements vs correctingapplanation tonometry surface (CATS) lot number.

TABLE 4. Summary of Prism Use Safety

Objective

Endpoint/Criteria

for Success Outcome

To demonstrate

equivalence of the

CATS tonometer

prism to the GAT

prism in the

measurement of IOP.

Incidence of

AEs

The rate of any AE

type was 0.0%

AE ¼ adverse event; CATS ¼ correcting applanation tonom-

etry surface; GAT ¼ Goldmann applanation tomography; IOP ¼intraocular pressure.

variability in the measurement of IOP with the CATSprism.

In a likewise fashion to the interoperator variabilityabove, a 1-way ANOVA was performed on the differencesin the CATS and GAT IOP readings for lot number vari-ability. Again, this assumes the GAT reading to be the trueIOP and is a very conservative statistical comparison. Evenwith this conservative approach the CATS prism is not sta-tistically different from the GAT prism in regard to lotnumber repeatability.

� ANALYSIS OF SAFETY: There were no unanticipatedadverse events (AEs) in the study (Table 4).

� SUPPLEMENTAL ANALYSIS (EXPLORATORY ENDPOINT):

The paired IOP measurement difference in CATS and

VOL. 196 ACCURACY AND REPEATABILITY OF A

GAT prisms was calculated and correlated to CCT. Thesubject’s average CCT was 551 6 24 mm, which wasrestricted in the protocol to between 500 and 600 mm.The results shown in Figure 11 indicate a negative GATcorrection slope of �0.023 mm Hg/mm for CCT. Themean IOP with the CATS prism was 19.8 mm Hg,compared to 19.5 mm Hg with the Goldmann. TheCATS prism reduced the IOP error due to CCT by61.7 mm Hg over the GAT prism, which compares wellto the published Dresdner GAT error over this same rangeof CCT values.17 The correlation coefficient associatedwith CCT was statistically significant (P¼ .01), indicatinggood correlation between the difference in IOP betweenthe CATS and GAT prisms over the range of correspond-ing CCT values (R2 ¼ 0.14).To conclude, The CATS prism had no statistically sig-

nificant difference in IOP measurement during assessmentfrom the reference GAT prism according to ANSI Z80.10-

151MODIFIED GOLDMANN PRISM

Page 8: Intraocular Pressure Measurement Accuracy and ...doclibrary.com/MSC167/PRM/cats_multicenter_trial_AJO4250.pdfbiomechanical variability in corneal thickness, rigidity, curvature, or

FIGURE 11. Measured intraocular pressure difference betweencorrecting applanation tonometry surface (CATS) minus Gold-mann applanation tonometer (GAT) correlated to study-limitedcentral corneal thickness (CCT) (500 mm < CCT <600 mm), 95% confidence interval on average mean differenceand slope indicated by dashed lines.

2014 and the FDA tonometer guidelines for 510(k) submis-sion standards. There were no AEs or device failures.

The IOP measurement assessment included a robustanalysis of interoperator and intraoperator repeatability,as well as manufacture lot repeatability. The study exam-ined the clinical accuracy, repeatability, and safety ofCATS tonometer prism, compared to the referenceGAT tonometer prism, and provided statistical evidencethat the CATS prism is as accurate, repeatable, and safeas the GAT prism. Although the CATS and GAT tonom-eter prisms were statistically equivalent in the low, me-dium, and high ranges of IOP, the sample size wasinsufficient to substantiate equivalence in those subjectswith pressures less than or equal to 10 mm Hg and greaterthan 35 mm Hg.

Measurement variability assessment was limited owingto the number of possible repeat measurements on agiven subject because of anesthetic corneal toxicity. Apreviously completed cadaver eye study demonstratingstatistical equivalence was better suited to assess intraop-erator and same eye interoperator measurementvariability.15

152 AMERICAN JOURNAL OF

The clinical study indicates a significant reduction inCATS prism sensitivity to CCT, which is a recognizedcorneal biomechanical error in GAT IOP measure-ment.18 The results verify the previously publishedmathematical modeling, direct CATS/GAT compari-sons, and intracameral IOP comparisons examining thedifference between CATS and GAT measurementswhen correlated to corneal thickness.12–16 There areseveral other tonometers that have been compared toGAT to demonstrate substantial equivalence for FDAclearance. Since the scope of this study was a directcomparison to the GAT reference tonometer and theCATS prism is simply a replacement part for theGAT, no comparison to other tonometers wascompleted. However, previous studies on the CATSprism did show statistically significant CATS minusGAT IOP correlation to corneal hysteresis and cornealresistance factor measured by an Ocular ResponseAnalyzer (ORA; Reichert, Depew, New York,USA).13,15

The present study was limited by the protocol to healthyeyes with nominal CCTs between 500 and 600 mm. Anadditional study examining those patients with CCTsgreater than 600 mm and less than 500 mm is underway.Future studies in pediatric patients will examine CATSprism and GAT prism IOP measurements compared tointracameral pressure. Another planned study will be con-ducted on patients before and after a LASIK refractive pro-cedure. These 2 populations, which together comprisenearly 30% in the United States, could benefit greatlyfrom improved IOP accuracy.19

Lastly, a study is underway comparing the difference inCATS and GAT IOP measurements in keratoconus pa-tients both before and after a corneal cross-linking proced-ure. The difference in CATS and GAT IOP measurementis in effect a population-standardized measurement of thecorneal contribution to applanation IOP. The CATSminus GAT difference is predominantly a measurementof the biomechanical ‘‘bendability’’ of the cornea, whichis a combined effect of corneal thickness and corneal rigid-ity or modulus of elasticity. Therefore, we can use this dif-ference in CATS/GAT IOP to measure procedurallyinduced corneal biomechanical changes in live humaneyes.

FUNDING/SUPPORT: THIS STUDY WAS SUPPORTED IN PART BY NIH SBIR GRANT R43 EY026821-01 AND ARIZONA EYE CONSUL-tants, Tucson, Arizona, USA. Financial Disclosures: Sean McCafferty has an interest in Intuor Technologies (Tucson, Arizona, USA), which ownsthe technology being examined in this clinical trial. Additional grant support unrelated to this study has been provided by Abbott Medical Optics (SantaAna, California, USA) andAlcon, Inc (Ft.Worth, Texas, USA). Jason Levine has unrelated study grant support from Innfocus, Inc. The following authorshave no financial disclosures: Kyle Tetrault, Ann McColgin, Warren Chue, and Melissa Muller. All authors attest that they meet the current ICMJEcriteria for authorship.

Other Acknowledgments: The authors thank Arizona Eye Consultants, Tucson, Arizona, USA, for extensive facilities use.

DECEMBER 2018OPHTHALMOLOGY

Page 9: Intraocular Pressure Measurement Accuracy and ...doclibrary.com/MSC167/PRM/cats_multicenter_trial_AJO4250.pdfbiomechanical variability in corneal thickness, rigidity, curvature, or

REFERENCES

1. Susanna JR,DeMoraesCG,CioffiGA,RitchR.Whydopeople(still) go blind from glaucoma? Trans Vis Sci Tech 2015;4(2):1.

2. Kass M, Heuer D, Higginbotham E, et al. The Ocular Hyper-tension Treatment Study: a randomized trial determines thattopical ocular hypotensive medication delays or prevents theonset of primary open-angle glaucoma.Arch Ophthalmol 2002;120:701–713.

3. Lester M, Mete M, Figus M, Frezzotti P. Incorporating cornealpachymetry into the management of glaucoma. J CataractRefract Surg 2009;35(9):1623–1628.

4. Kotecha A, Elsheikh A, Roberts C, Haogang Z, Garway-Heath D. Corneal thickness- and age related biomechanicalproperties of the cornea measured with the Ocular ResponseAnalyzer. Invest Ophthalmol Vis Sci 2006;47:5337–5347.

5. Whitacre M, Stein R. Sources of error with use of Goldmann-type tonometers. Surv Ophthalmol 2002;38(1):1–30.

6. Schallhorn J, Schallhorn S, Ou Y. Factors that influenceintraocular pressure changes after myopic and hyperopic LA-SIK and photorefractive keratectomy: a large populationstudy. Ophthalmology 2015;122(3):471–479.

7. Tsai A, Loon S. Intraocular pressure assessment after laserin situ keratomileusis: a review. Clin Exp Ophthalmol 2012;40:295–304.

8. Beck A, Giangiacomo A. Pediatric glaucoma: review ofrecent literature. Curr Opin Ophthalmol 2017;28(2):199–203.

9. Feng C, Jin K, Yi K, Choi D. Comparison of intraocular pres-sure measurements obtained by rebound, noncontact, andGoldmann applanation tonometry in children.Am J Ophthal-mol 2015;160:937–943.

10. Damji K, Muni R, Munger R. Influence of corneal variableson accuracy of intraocular pressure measurement. J Glaucoma

2003;12(1):69–80.

VOL. 196 ACCURACY AND REPEATABILITY OF A

11. Liu J, Roberts C. Influence of cornea biomechanicalproperties on intraocular pressure measurement: Quanti-tative analysis. J Cataract Refract Surg 2005;31(1):146–155.

12. McCafferty S, Lim G, Duncan W, Enikov E,Schwiegerling J. Goldmann tonometer prism with an opti-mized error correcting applanation surface. Transl Vis SciTechnol 2016;5:1–5.

13. McCafferty S, Lim G, DuncanW, Enikov E, Schwiegerling J.Goldmann tonometer error correcting prism: clinical evalua-tion. Clin Ophthalmol 2017;11:835–840.

14. McCafferty S, Levine J, Schwiegerling J, Enikov E. Gold-mann applanation tonometry error relative to true intracam-eral intraocular pressure in vitro and in vivo. BMCOphthalmol 2017;17:215.

15. McCafferty S, Levine J, Schwiegerling J, Enikov ET. Gold-mann and error correcting tonometry prisms compared tointracameral pressure. BMC Ophthalmol 2018;18:2.

16. McCafferty S, Enikov E, Schwiegerling J, Ashley S. Gold-mann tonometry tear-film error and partial correction witha shaped applanation surface. Clin Ophthalmol 2018;1:71–78.

17. Kohlhaas M, Boehm AG, Spoerl E, Pursten A, Grein HJ,Pillunat LE. Effect of central corneal thickness, corneal cur-vature, and axial length on applanation tonometry. ArchOphthalmol 2006;124(6):471–476.

18. Prum B, Rosenberg L, Gedde S, et al. American Academy ofOphthalmology. Preferred Practice Pattern; Primary Open-Angle Glaucoma. Available at: 2015. https://doi.org/10.1016/j.ophtha.2015.10.053. Accessed August 7, 2018.

19. Eisenberg D, Sherman B, Mckeown C, Schuman J. Tonom-etry in adults and children: a manometric evaluation of pneu-motonometry, applanation, and tonopen in vitro and in vivo.Ophthalmology 1998;105:1173–1181.

153MODIFIED GOLDMANN PRISM