international federation for consulting gmbh ifc
DESCRIPTION
The Commission Decision 2002/657/EC Performance criteria by Dr. Michael Scheutwinkel. International Federation for Consulting GmbH IFC. Content of Annex of CD 2002/657/EC. 1. 1 Definitions 2 Performance criteria and other requirements for analytical methods - PowerPoint PPT PresentationTRANSCRIPT
![Page 1: International Federation for Consulting GmbH IFC](https://reader036.vdocuments.site/reader036/viewer/2022062301/56815c12550346895dc9ee57/html5/thumbnails/1.jpg)
1
International Federation forConsulting GmbH IFC
TheTheCommission Decision Commission Decision
2002/657/EC2002/657/ECPerformance criteriaPerformance criteria
by
Dr. Michael Scheutwinkel
![Page 2: International Federation for Consulting GmbH IFC](https://reader036.vdocuments.site/reader036/viewer/2022062301/56815c12550346895dc9ee57/html5/thumbnails/2.jpg)
2
Content of Annex of CD 2002/657/EC
1 Definitions
2 Performance criteria and other requirements for analytical methods
2.1 General criteria to be fulfilled by any technique, procedure, etc.
2.2 Criteria for screening methods 2.3 Criteria for confirmatory methods (organics)
2.4 Criteria for confirmatory methods (elements)
3 Recommended valdidation procedures - classical validation - alternative validations, e.g. InterVal
11
22
33
![Page 3: International Federation for Consulting GmbH IFC](https://reader036.vdocuments.site/reader036/viewer/2022062301/56815c12550346895dc9ee57/html5/thumbnails/3.jpg)
3
Performance criteria according to CD 2002/657/EC
2.1 General requirements2.1 General requirements
Sample handlingSample handling Test performanceTest performance
SpecifitySpecifity
Recovery with each batch of samples
Recovery with each batch of samples
Maximum chance for the detection
of the analyte
Maximum chance for the detection
of the analyte
1122
![Page 4: International Federation for Consulting GmbH IFC](https://reader036.vdocuments.site/reader036/viewer/2022062301/56815c12550346895dc9ee57/html5/thumbnails/4.jpg)
4
Commission Decision 2002/657/EC
![Page 5: International Federation for Consulting GmbH IFC](https://reader036.vdocuments.site/reader036/viewer/2022062301/56815c12550346895dc9ee57/html5/thumbnails/5.jpg)
5
Performance criteria according to CD 2002/657/EC
2.2 Screening methods2.2 Screening methods
False compliant rate < 5 % (β-error)
False compliant rate < 5 % (β-error)
In case of a suspected non-compliant result,
a confirmation is obligatory.
In case of a suspected non-compliant result,
a confirmation is obligatory.
![Page 6: International Federation for Consulting GmbH IFC](https://reader036.vdocuments.site/reader036/viewer/2022062301/56815c12550346895dc9ee57/html5/thumbnails/6.jpg)
6
Measuring technique
Substances Annex I
96/23/EEC
Limitations
LC or GC with MS
Groups A and B
- Only if following either an on-line or off-
line chromatographic separation- Only if full scan technique are used or
using at least 3 (group B) or 4 (group A)
identification points for techniques that
do not record the full mass spectra
LC-full-scan DAD
Group B Specific requirements for absorption in UV spectrometry have to be met
Others -----
Suitable confirmatory methods according to
CD 2002/657/EC, here only extract
Table 1
![Page 7: International Federation for Consulting GmbH IFC](https://reader036.vdocuments.site/reader036/viewer/2022062301/56815c12550346895dc9ee57/html5/thumbnails/7.jpg)
7
Commission Decision 2002/657/EC Table 9
![Page 8: International Federation for Consulting GmbH IFC](https://reader036.vdocuments.site/reader036/viewer/2022062301/56815c12550346895dc9ee57/html5/thumbnails/8.jpg)
8
Commission Decision 2002/657/EC Table 10
![Page 9: International Federation for Consulting GmbH IFC](https://reader036.vdocuments.site/reader036/viewer/2022062301/56815c12550346895dc9ee57/html5/thumbnails/9.jpg)
9
Performance criteria according to CD 2002/657/EC
2.3 Confirmatory methods for analysis of organics, common requirements
2.3 Confirmatory methods for analysis of organics, common requirements
No discrimination in cases that two compounds give same response,
only chromatographic analysis without spectrometric
detection is not appropriate.
No discrimination in cases that two compounds give same response,
only chromatographic analysis without spectrometric
detection is not appropriate.
Use of internal standard
or confirmation by co-chromatogaphy
Use of internal standard
or confirmation by co-chromatogaphy
Use of CRMor fortification of
material
Use of CRMor fortification of
material
1133
33
![Page 10: International Federation for Consulting GmbH IFC](https://reader036.vdocuments.site/reader036/viewer/2022062301/56815c12550346895dc9ee57/html5/thumbnails/10.jpg)
10
Performance criteria according to CD 2002/657/EC
2.3 Confirmatory methods for analysis of organics, specific requirements
2.3 Confirmatory methods for analysis of organics, specific requirements
Minimum trueness(Table 2)
Minimum trueness(Table 2)
Reproducibility CV(%)(Table 3)
Reproducibility CV(%)(Table 3)
CV = interlaboratory coefficient
of variation in %
CV = interlaboratory coefficient
of variation in %MRL existing?
CV = 50% of limitMRL existing?
CV = 50% of limit
1122
![Page 11: International Federation for Consulting GmbH IFC](https://reader036.vdocuments.site/reader036/viewer/2022062301/56815c12550346895dc9ee57/html5/thumbnails/11.jpg)
11
Commission Decision 2002/657/ECTable 3
![Page 12: International Federation for Consulting GmbH IFC](https://reader036.vdocuments.site/reader036/viewer/2022062301/56815c12550346895dc9ee57/html5/thumbnails/12.jpg)
12
Commission Decision 2002/657/EC
Table 2
![Page 13: International Federation for Consulting GmbH IFC](https://reader036.vdocuments.site/reader036/viewer/2022062301/56815c12550346895dc9ee57/html5/thumbnails/13.jpg)
13
Commission Decision 2002/657/EC
![Page 14: International Federation for Consulting GmbH IFC](https://reader036.vdocuments.site/reader036/viewer/2022062301/56815c12550346895dc9ee57/html5/thumbnails/14.jpg)
14
Performance criteria according to CD 2002/657/EC
Confirmatory methods for analysis of organics,performance criteria for MS
Confirmatory methods for analysis of organics,performance criteria for MS
Chromatographicseparation
Chromatographicseparation
MS detectionMS detection
Usually full scan,< 10% deviation
from reference MS
Usually full scan,< 10% deviation
from reference MS
Full scan + SIM see table 4
Full scan + SIM see table 4
SIM S/N ratio > 3:1
SIM S/N ratio > 3:1
![Page 15: International Federation for Consulting GmbH IFC](https://reader036.vdocuments.site/reader036/viewer/2022062301/56815c12550346895dc9ee57/html5/thumbnails/15.jpg)
15
Commission Decision 2002/657/EC
![Page 16: International Federation for Consulting GmbH IFC](https://reader036.vdocuments.site/reader036/viewer/2022062301/56815c12550346895dc9ee57/html5/thumbnails/16.jpg)
16
Commission Decision 2002/657/EC
![Page 17: International Federation for Consulting GmbH IFC](https://reader036.vdocuments.site/reader036/viewer/2022062301/56815c12550346895dc9ee57/html5/thumbnails/17.jpg)
17
LC with differentdetection techniques
LC with differentdetection techniques
2-D TLC coupled to full-scan UV/VIS
spectrometricdetection
2-D TLC coupled to full-scan UV/VIS
spectrometricdetection
Chromatography coupled to
infrared detection
Chromatography coupled to
infrared detection
2.3 Confirmatory methods for organics, performance criteria2.3 Confirmatory methods for organics, performance criteria
GC-ECDGC-ECD
![Page 18: International Federation for Consulting GmbH IFC](https://reader036.vdocuments.site/reader036/viewer/2022062301/56815c12550346895dc9ee57/html5/thumbnails/18.jpg)
18
Commission Decision 2002/657/EC
![Page 19: International Federation for Consulting GmbH IFC](https://reader036.vdocuments.site/reader036/viewer/2022062301/56815c12550346895dc9ee57/html5/thumbnails/19.jpg)
19
Commission Decision 2002/657/EC
![Page 20: International Federation for Consulting GmbH IFC](https://reader036.vdocuments.site/reader036/viewer/2022062301/56815c12550346895dc9ee57/html5/thumbnails/20.jpg)
20
Commission Decision 2002/657/EC
Table 5
![Page 21: International Federation for Consulting GmbH IFC](https://reader036.vdocuments.site/reader036/viewer/2022062301/56815c12550346895dc9ee57/html5/thumbnails/21.jpg)
21
Commission Decision 2002/657/EC Table 6
![Page 22: International Federation for Consulting GmbH IFC](https://reader036.vdocuments.site/reader036/viewer/2022062301/56815c12550346895dc9ee57/html5/thumbnails/22.jpg)
22
Table 4
Performance criteria according to CD 2002/657/EC
![Page 23: International Federation for Consulting GmbH IFC](https://reader036.vdocuments.site/reader036/viewer/2022062301/56815c12550346895dc9ee57/html5/thumbnails/23.jpg)
23
Commission Decision 2002/657/EC Table 6
![Page 24: International Federation for Consulting GmbH IFC](https://reader036.vdocuments.site/reader036/viewer/2022062301/56815c12550346895dc9ee57/html5/thumbnails/24.jpg)
24
Commission Decision 2002/657/EC
![Page 25: International Federation for Consulting GmbH IFC](https://reader036.vdocuments.site/reader036/viewer/2022062301/56815c12550346895dc9ee57/html5/thumbnails/25.jpg)
25
Commission Decision 2002/657/EC
![Page 26: International Federation for Consulting GmbH IFC](https://reader036.vdocuments.site/reader036/viewer/2022062301/56815c12550346895dc9ee57/html5/thumbnails/26.jpg)
26
Differential pulse anodicstrippingvoltametry
Differential pulse anodicstrippingvoltametry
ICP-AESICP-AES
Trueness of quantitative tests
Trueness of quantitative tests
2.4 Confirmatory methods for elements, see also table 7
2.4 Confirmatory methods for elements, see also table 7
AASAAS
ICP-MSICP-MS
![Page 27: International Federation for Consulting GmbH IFC](https://reader036.vdocuments.site/reader036/viewer/2022062301/56815c12550346895dc9ee57/html5/thumbnails/27.jpg)
27
Commission Decision 2002/657/ECTable 7
![Page 28: International Federation for Consulting GmbH IFC](https://reader036.vdocuments.site/reader036/viewer/2022062301/56815c12550346895dc9ee57/html5/thumbnails/28.jpg)
28
Commission Decision 2002/657/EC
Table 8