internal quality control in clinical laboratories.pptx
DESCRIPTION
Internal Quality control in Clinical Laboratories.TRANSCRIPT
04/17/2023Nazar Ahmed Mohamed Abd-
Alla(Sangoor)1
سبحان خلق منشئ كل
بقدر
Internal Quality control in Clinical Laboratoreis
04/17/2023Nazar Ahmed Mohamed Abd-
Alla(Sangoor)2
NAZAR AHMED MOHAMED ABD-ALLA
Topics
04/17/2023Nazar Ahmed Mohamed Abd-
Alla(Sangoor)3
Quality definition.Quality Activity.Type of error in clinical laboratory.Specimens and Requests .Method selection, and Reagents
storage condition .Laboratory instruments.instruments and equipment calibration.SOPs.Use of calibrators and control material .
Topics
04/17/2023Nazar Ahmed Mohamed Abd-
Alla(Sangoor)4
Implementation of quality programsInterpretation of internal quality
results.Corrective action.Turn around time.Documentation.
Whats Quality
quality is defined as the totality of features and characteristics of a product or service that bear on its ability to satisfy stated or implied needs.
Medical laboratories must provide a high quality service by producing accurate, precise, relevant and comprehensive data that can be applied to the medical management of patients.
tests requested must be appropriate to the medical problem, must be analytically correctly performed and their results interpreted correctly.
04/17/2023Nazar Ahmed Mohamed Abd-
Alla(Sangoor)5
Quality assurance: Part of quality management focused on providing confidence that quality requirements will be fulfilled . witch contain many activity.
Quality control: Part of quality management focused on fulfilling quality requirements.
Quality management system: Management system to direct and control an organization with regard to quality.
04/17/2023Nazar Ahmed Mohamed Abd-
Alla(Sangoor)6
Appropriateness of tests can only be obtained by a dialogue between the clinician and the medical pathologist.
Correct analytical results are based on: (i) quality management within the
laboratory. (ii) the quality of industrially prepared
reagents (kits) and instruments .(iii) quality management of the pre-
analytical phase outside the laboratory along with analytical & post-analytical phase.
A bad system, a wrong sampling or a kit with poor performance can never produce a reliable result, even in a laboratory with the best quality management system.
04/17/2023Nazar Ahmed Mohamed Abd-
Alla(Sangoor)7
principles aspects of making reliable analytical measurements:
1. Analytical measurements should be made to satisfy an agreed requirement.
2. Analytical measurements should be made using methods and equipment that have been tested to ensure they are fit for purpose.
3. Staff making analytical measurements should be both qualified and competent to undertake the task.
04/17/2023Nazar Ahmed Mohamed Abd-
Alla(Sangoor)8
principles aspects of making reliable analytical measurements:
4. There should be a regular independent assessment of the technical performance of a laboratory.
5. Analytical measurements made in one location should be consistent with those elsewhere.
6. Organizations making analytical measurements should have well defined quality control and quality assurance procedures.
04/17/2023Nazar Ahmed Mohamed Abd-
Alla(Sangoor)9
Quality Assurance
Quality assurance is a comprehensive and systematic process that strives to ensure reliable patient results.
This process includes: every level of laboratory operation. Phlebotomy services.competency testing, error analysis, standard protocols, PPE, quality control, and turnaround time .
04/17/2023Nazar Ahmed Mohamed Abd-
Alla(Sangoor)10
*QA activities encompass all of the non-analytic activities, those activities that are not part of the clinical testing process. The laboratory organizes it activities to provide the best possible health care to the patient.
Nazar Ahmed Mohamed Abd-Alla(Sangoor)
QA activities
04/17/202311
* A. management and monitoring personnel, B. using quality materials (reagents instruments, supplies, etc.), C. using established procedures and established statistics (a procedure manual),D. specimen collection, identification, transport, accession, and handling prior to testing,E. reporting results,F. fee charges for tests performed, G. using corrective actions to obtain desired results,H. monitoring patient satisfaction.Nazar Ahmed Mohamed Abd-
Alla(Sangoor)
Examples of QA Activities
04/17/202312
*The Clinical laboratory is concerned about quality and accuracy of the tests that are reported to primary care givers. The laboratory monitors where these errors can appear that will affect the accuracy of test results.
Nazar Ahmed Mohamed Abd-Alla(Sangoor) 04/17/202313
*These errors can occur prior to the test analysis and if they manifest, they are called preanalytical errors or variables.
*If the error occurs during the testing process, then it become an analytical error.
*If the error appears after the test is performed and reported, then it is known as a post-analytical error.
*The preanalytical error occurs before the test is performed. This error source can occur at the beginning of test ordering and filling out the requisition.
Nazar Ahmed Mohamed Abd-Alla(Sangoor)
Type Of Error
04/17/202314
*type of error *Pre-Analytical Error: 01- duplicate or missing requisitions 02- tests omitted from the requisition
03- incorrect ordering of tests04- patient identification error05- incorrect blood collection 06- specimen transport error07- specimen handling/processing in the lab
*Analytical errors: occur during the testing process. 01- deteriorated or wrong reagents
02- any instrument malfunction03- laboratorial error04- incorrect recording of test results
Nazar Ahmed Mohamed Abd-Alla(Sangoor) 04/17/202315
* Post Analytical Error:* Examples of these are:
01- failure to notify the physician of critical values: (Critical values may imply a life-threatening situation for the patient and are brought to the immediate attention of the physician and/or the patient care area responsible for the patient ).02- failure to report test results in a timely manner.03- placement of report in the chart of the wrong patient 04- miscommunications that are detrimental to the patient regarding the tests performed.
Nazar Ahmed Mohamed Abd-Alla(Sangoor) 04/17/202316
Specimens and Requests
04/17/2023Nazar Ahmed Mohamed Abd-
Alla(Sangoor)17
1- Request forms. Request forms should include: Patient details.(age,sex,race,resedance) Specimen details .(type,preservatie,time of collection) Clinical Remarks .(spleenomegaly,hepatomegly,fever,join
pain)
2- Primary sample collection.• Patient Preparation• Containers selection• Blood Collection• Plasma Preparation• Specimen storage and Transportation
Specimens and Requests
04/17/2023Nazar Ahmed Mohamed Abd-
Alla(Sangoor)18
3- Samples acceptance criteria.Complete request form.Patient identification.Suitable container selected.Sufficient amount of blood collected.Sample labeled properly.Sample prepared properly.Sample separated and store properly.
Specimens and Requests
04/17/2023Nazar Ahmed Mohamed Abd-
Alla(Sangoor)19
4- Samples Rejection CriteriaIncomplete request form.Rung Patient identification.Unsuitable container selected.Unlabelled or incorrectly labeled sample.Haemolysed sample.Sample changes due to : concentration changes compostion changes. bacterail changes. and enzymatic changes.
METHOD VALIDATION
04/17/2023Nazar Ahmed Mohamed Abd-Alla(Sangoor)20
“Method validation” is a term used for the suite of procedures to which an analytical method is subjected.
to provide objective evidence that the method, if used in the manner specified, will produce results that conform to the statement of the
method validation parameters.
ISO Definition
1. The process of establishing the performance characteristics and limitations of a method and the identification of the influences which may change these characteristics and to what extent. Which analytes can it determine in which matrices in the presence of which interferences? Within these conditions what levels of precision and accuracy can be achieved?
2. The process of verifying that a method is fit for purpose, i.e. for use for solving a particular analytical problem. 04/17/2023
Nazar Ahmed Mohamed Abd-Alla(Sangoor)21
Continue: Method validation
Method validation involves the evaluation of the fitness of analytical methods for their purpose.
The process of proving that an analytical method is acceptable for its intended purpose.
the concept of fitness for purpose, a method is validated for a particular use under particular circumstances. If those circumstances vary, then the method would need to be re-validated at least for the differences.
04/17/2023Nazar Ahmed Mohamed Abd-
Alla(Sangoor)22
METHOD VALIDATION
04/17/2023Nazar Ahmed Mohamed Abd-Alla(Sangoor)23
Error assessment is what method validation is about.
However, before getting to the assessment of errors, you have to first select the method to be validated.
Method selection is a different process that needs to be understood in relation to the validation process that will follow.
Importance of Method Validation Practices
04/17/2023Nazar Ahmed Mohamed Abd-Alla(Sangoor)24
Laboratory regulations require that method performance for any new method be "verified" prior to reporting patient test results.
Precision and accuracy are specifically identified, along with analytical sensitivity, analytical specificity, reportable range, reference values, and any other applicable characteristic.
Types of Errors to be assessed by method validation
04/17/2023Nazar Ahmed Mohamed Abd-Alla(Sangoor)25
imprecision or random errors, inaccuracy, bias, or systematic errors,
which can be of two types constant systematic error or proportional systematic error.
All these errors can be recognized when a group of measurements are compared to the correct or true values.
Verification
Much of the work of method validation is done by international organizations that publish standard methods. The reason such methods appear to be written in a kind of legalese is that there must be no doubt as to what the method is and how it should be implemented. When accuracy and precision data are published from interlaboratory trials, there is some confidence that the method has undergone extreme scrutiny and testing.
04/17/2023Nazar Ahmed Mohamed Abd-
Alla(Sangoor)26
Verification
A laboratory that uses a method for the first time should spend some time in going through the analysis with standard materials so that when used with field samples, the method will yield satisfactory results.
This is verification and must be done to an appropriate level before any method is used. By its nature, (verification comes under the heading of Single Laboratory Validation).
04/17/2023Nazar Ahmed Mohamed Abd-
Alla(Sangoor)27
Method performance parameters assessed in a method validation
studyIdentity: Measurement correctly applies to the
stated measurand.Selectivity Specificity: Determination of the
extent of the effects of interfering substances and the ability of the method to measure the measurand; analysis in different matrices covered by the scope of the validation.
Limit of detection: Minimum value of the measurand at which the presence of the analyte can be determined with a given probability of a false negative, and in the absence of the analyte, at a given probability of a false positive.
[Limit of determination] Minimum value that can be obtained with a specified measurement uncertainty
04/17/2023Nazar Ahmed Mohamed Abd-
Alla(Sangoor)28
Method performance parameters assessed in a method validation
study
Calibration [linearity] Model parameters [sensitivity] :Adequacy of the calibration model; parameters with uncertainties.
Calibration range [linear range] Range of values of the measurand in which the validation holds
Bias and recovery [accuracy] :Demonstration of the absence of significant systematic error after corrections have been made for bias and/or recovery
04/17/2023Nazar Ahmed Mohamed Abd-
Alla(Sangoor)29
Method performance parameters assessed in a method validation
study
Robustness or ruggedness: Ability of method to remain unaffected by small variations in method parameters (some authors make the distinction bebetween the property robustness and a ruggedness test in which deliberate changes are made in a method to assess the robustness)
04/17/2023Nazar Ahmed Mohamed Abd-
Alla(Sangoor)30
Method Selection
04/17/2023Nazar Ahmed Mohamed Abd-
Alla(Sangoor)31
1. General Characteristics: Made in ,Expire date, Package Content,
Accessories, Package size ,Stability after open, and Reagent storage considerations.
2. Application Characteristics: Specimen type, Sample volume, Turnaround
Time, Stability of reaction product, Cost-per-test, Filter used, and Safety considerations
3. Methodology Characteristics: Type of Reaction, Reaction Principle,
Measurement reaction, Temperature, and Time period of measurement.
Experiments for estimating analytical errors
04/17/2023Nazar Ahmed Mohamed Abd-Alla(Sangoor)32
METHOD COMPARISSION FOR HBG ESTIMATION BETWEEN TOW DIFFERENT
INSTRUMENT04/17/2023
Nazar Ahmed Mohamed Abd-Alla(Sangoor)33
6 8 10 12 14 16 180
2
4
6
8
10
12
14
16
18
HBG
HBG
Clinical laboratory instruments example
04/17/2023Nazar Ahmed Mohamed Abd-
Alla(Sangoor)34
Spectrophotometers .Water bath.Vortex.Hematology analyzer.Microplate reader & washer.Centrifuge.Flame photometers or ISE.Coagulometer.Hot air oven.
Maintenance and calibration of Photometers
04/17/2023Nazar Ahmed Mohamed Abd-
Alla(Sangoor)35
Includes checking for:Wavelength accuracyLinearity of instrumentStray lightBase line stabilityPhotometric accuracyShort circuitsCheck 0% TTemperature
04/17/2023Nazar Ahmed Mohamed Abd-
Alla(Sangoor)36
04/17/2023Nazar Ahmed Mohamed Abd-
Alla(Sangoor)37
04/17/2023Nazar Ahmed Mohamed Abd-
Alla(Sangoor)38
Micropipettes Calibration
04/17/2023Nazar Ahmed Mohamed Abd-
Alla(Sangoor)39
Gravimetric method (Distilled water
weighing method) (Recommended method)
Work Requirement:1. Calibration tools2. Double distilled water. 3. New compatible tip4. Analytical electronic balance(3-5 digits)5. Temperature controlled atmosphere6. Small plastic beaker.
Checking Micropipettes Calibration
04/17/2023Nazar Ahmed Mohamed Abd-
Alla(Sangoor)40
Weighing Should takes place at 20-25ċ constant to ± 0.5ċ.Set the desire testing volume of your pipette.Carefully fit the tip onto the tip cone.Aspirate the distilled water 5 times( Humidity equilibriumCarefully aspirate the fluid, keeping the pipette vertical. Pipette distilled water into a tred container and read the
weight in mgs.Repeat at least five times (ten times )and record each
result in mgs.
Checking Micropipettes Calibration
04/17/2023Nazar Ahmed Mohamed Abd-
Alla(Sangoor)41
Convert the recoded weight to volume (V1) either by divide the weight of the water by its density ( at 20ċ : 0.9982 ) or by multiply the weight by the Z correction factor (= 1.002899 µl/mg at 20Ċ ).
Calculate the mean volume (V1) Calculate the standard d deviation SD.Calculate the Coefficient of Variation. Calculate the Inaccuracy and Imprecision .Calculate the F max value.
04/17/2023Nazar Ahmed Mohamed Abd-
Alla(Sangoor)42
Systematic Error and Random Error
04/17/2023Nazar Ahmed Mohamed Abd-
Alla(Sangoor)43
Compare the Systematic Error
( Inaccuracy), Random Error
( Imprecision ) and F max with the values in the performance specification .
If the results fall within the specifications, the pipette is ready for use .
Otherwise check both Systematic Error and Random Error and when necessary recalibrate the pipette.
Table Shows The Maximum Permitted Errors( F Max )
PipettesType
Nominal volume
M.P.E (F Max )
Nominal volume
M.P.E (F Max )
Single channel pipettes
5 µl ±0.3 µl 200 µl ±2 µl
10 µl ±0.3 µl 250 µl ±2.5 µl
20 µl ±0.4 µl 500 µl ±5 µl
25 µl ±0.5 µl 1000 µl ±10 µl
50 µl ±0.8 µl 2000 µl ±20 µl
100 µl ±1.5 µl 5000 µl ±50 µl
Multi- channel pipettes
10 µl ±0.6 µl 250 µl ±5.0 µl
50 µl ±1.6 µl 300 µl ±6.0 µl04/17/2023
Nazar Ahmed Mohamed Abd-Alla(Sangoor)44
Table Shows The Recommended Specification
Nominal volume
Inaccuracy Imprecision Nominal
volume
Inaccuracy Imprecision
±≤%
(A) µl % ≤ µl %±≤ (A) µl % ≤ µl
5 µl 1.0 O.05 0.8 0.04 200 µl 0.5 1 0.2 0.410 µl 1.0 0.1 0.8 0.08 250 µl 0.5 1.25 0.2 0.5
20 µl 0.7 0.14 0.4 0.08 500 µl 0.5 2.5 0.2 1.0
25 µl 0.7 0.175 0.4 0.1 1000 µl 0.5 5 0.2 2.0
50 µl 0.7 0.35 0.4 0.2 10-100 µl 0.5 0.5 0.2 0.2
100 µl 0.5 0.5 0.2 0.2 20-200 µl 0.5 1 0.2 0.4
0.5-10 µl 1.0 0.1 0.8 0.08 25-250
µl 0.5 1.25 0.2 0.5
5-50 µl0.7 0.35 0.4 0.2 100-
1000 µl 0.5 5 0.2 2.004/17/2023
Nazar Ahmed Mohamed Abd-Alla(Sangoor)45
Standard Operating Procedures
04/17/2023Nazar Ahmed Mohamed Abd-
Alla(Sangoor)46
SOPs are written , up to date instructions and information which cover in details how to perform individual tests to insure the correct use , availability , reliability , timeliness, and reporting of blood tests and correct interpretation of the test result
*Quality control (QC) encompasses quality assurance as it focuses on analytical activities that are associated with the testing process. QC consists of:A. running control samples with patient samples,B. using established statistical methods to determine reliability of test procedures and test results,C. monitoring instrument and laboratorial performance.
Nazar Ahmed Mohamed Abd-Alla(Sangoor)
Quality control
04/17/202347
Quantitative QC Materials
04/17/2023Nazar Ahmed Mohamed Abd-
Alla(Sangoor)48
Calibrator: a solution which has a known amount of analyte weighed in or has a value determined by repetitive testing using a reference or definitive test method
Control: material or preparation used to monitor the stability of the test system within predetermined limits
Sources for Control Materials
04/17/2023Nazar Ahmed Mohamed Abd-
Alla(Sangoor)49
Commercial product.Diagnostic samples (qualitative
QC).“Homemade” or “In-house”.Obtained from:
Another laboratory .EQA provider.
04/17/2023Nazar Ahmed Mohamed Abd-
Alla(Sangoor)50
Parallel Testing.-Validation of new controls
Whenever possible, new lots of control material must be assayed in parallel alongside the current
in use lot. This is to enable the calculation of
laboratory QC ranges and to demonstrate that the
QC material is performing as expected .
04/17/2023Nazar Ahmed Mohamed Abd-
Alla(Sangoor)51
Controls for Quantitative Assays :
a. In order to validate new controls, the new lot of controls will be run in parallel with the old lot of controls 2-3 times a day for 5-10 days, to give a minimum of 20 values to enable the calculation of laboratory specific QC ranges. The mean and QC ranges for the new lot of controls will be reviewed and signed off by the laboratory supervisor or director before being put into use .
04/17/2023Nazar Ahmed Mohamed Abd-
Alla(Sangoor)52
b. For hematology the new lot of controls should be
run in parallel with the old lot of controls to give a
minimum of 10 values over a period 5 days if possible.
The mean and ranges for the new lot of controls will
be reviewed and signed off by the laboratory
supervisor or director before being put into use .
04/17/2023Nazar Ahmed Mohamed Abd-
Alla(Sangoor)53
2 .Controls for Qualitative Assays :
Each new lot of QC for qualitative assays must be run
and give an expected response. The lot of controls will
be reviewed and signed off by the laboratory
supervisor or director before being put into use .
04/17/2023Nazar Ahmed Mohamed Abd-
Alla(Sangoor)54
Types of Control Materials
Assayedmean calculated by the
manufacturermust verify in the laboratory
Unassayedless expensivemust perform data analysis
“Homemade” or “In-house” pooled sera collected in the
laboratorycharacterizedpreserved in small quantities for
daily use
04/17/2023Nazar Ahmed Mohamed Abd-
Alla(Sangoor)55
Levey-Jennings Chart
04/17/2023Nazar Ahmed Mohamed Abd-
Alla(Sangoor) 57q211q`
o A graphical method for displaying control results and evaluating whether a procedure is in-control or out-of-control
o Control values are plotted versus time o Lines are drawn from point to point to accent
any trends, shifts, or random excursions
Levey-Jennings Chartcontrol sample included in each assay
rundata plotted graphically (assay value
verses timeof day)control (or confidence) limits mean standard deviation (usually + 2 SD) if control limits are not met, then no
patient samples run in that batch can be reported.
if control limits are met, then patient samples run in that batch can be reported
04/17/2023Nazar Ahmed Mohamed Abd-
Alla(Sangoor)58
QC changes detectable with Levey-
Jennings plots
Drift: control value moves progressively in one direction from the mean for at least 3 days e.g deterioration of reagent or control
Dispersion :increase in random errors e.g inconsistency in technique
Shift: sudden problem develops e.g instrument malfunction or technique change
04/17/2023Nazar Ahmed Mohamed Abd-
Alla(Sangoor)59
Findings Over Time
04/17/2023Nazar Ahmed Mohamed Abd-
Alla(Sangoor)60
Ideally should have control values clustered about the mean (+/-2 SD) with little variation in the upward or downward direction
Imprecision = large amount of scatter about the mean. Usually caused by errors in technique
Inaccuracy = may see as a trend or a shift, usually caused by change in the testing process
Random error = no pattern. Usually poor technique, malfunctioning equipment
Levey-Jennings Chart -Record and Evaluate the Control Values
04/17/2023Nazar Ahmed Mohamed Abd-
Alla(Sangoor)61
80
85
90
95
100
105
110
115
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24
Mean
Day
+1SD
+2SD
+3SD
-1SD
-2SD
-3SD
What are the "Westgard rules”
04/17/2023Nazar Ahmed Mohamed Abd-
Alla(Sangoor)62
They are different combination of rules depending on the number of control being used, the total allowable error, and your work environment.
rules That are used as conjunction with each other to provide a high level of errors detection, while reducing the incidence of false rejection.
Typical rules combination
04/17/2023Nazar Ahmed Mohamed Abd-
Alla(Sangoor)63
For control run in multiples of 2 “ typical chemistry”
13s / 22s / R4s / 41s / 10 x
For control run in multiples of 3“ typical haematology,coagulation, and immunoassays”.
13s / 2 of 32s / R4s / 31s / 12 x
Westgard Rules (Generally used where 2 levels of control material are analyzed per run)
04/17/2023Nazar Ahmed Mohamed Abd-
Alla(Sangoor)64
12S rule 13S rule 22S rule
R4S rule 41S rule 10X rule
Westgard – 12S Rule
04/17/2023Nazar Ahmed Mohamed Abd-
Alla(Sangoor)65
“warning rule”One of two control results falls outside ±2SD
Alerts tech to possible problems
Not cause for rejecting a runMust then evaluate the 13S
rule
12S Rule = A warning to trigger careful inspection of the
control data
04/17/2023Nazar Ahmed Mohamed Abd-
Alla(Sangoor)66
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24
Mean
Day
+1SD
+2SD
+3SD
-1SD
-2SD
-3SD
12S rule violation
Westgard – 13S Rule
04/17/2023Nazar Ahmed Mohamed Abd-
Alla(Sangoor)67
If either of the two control results falls outside of ±3SD, rule is violated
Run must be rejectedIf 13S not violated, check 22S
13S Rule = Reject the run when a single control measurement
exceeds the +3SD or -3SD control limit
04/17/2023Nazar Ahmed Mohamed Abd-
Alla(Sangoor)68
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24
Mean
Day
+1SD
+2SD
+3SD
-1SD
-2SD
-3SD
13S rule violation
Westgard – 22S Rule
04/17/2023Nazar Ahmed Mohamed Abd-
Alla(Sangoor)69
2 consecutive control values for the same level fall outside of ±2SD in the same direction, or
Both controls in the same run exceed ±2SD
Patient results cannot be reportedRequires corrective action
22S Rule = Reject the run when 2 consecutive control measurements exceed the same
+2SD or -2SD control limit
04/17/2023Nazar Ahmed Mohamed Abd-
Alla(Sangoor)70
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24
Mean
Day
+1SD
+2SD
+3SD
-1SD
-2SD
-3SD
22S rule violation
Westgard – R4S Rule
04/17/2023Nazar Ahmed Mohamed Abd-
Alla(Sangoor)71
One control exceeds the mean by –2SD, and the other control exceeds the mean by +2SD
The range between the two results will therefore exceed 4 SD
Random error has occurred, test run must be rejected
R4S Rule = Reject the run when 1 control measurement
exceed the +2SD and the other exceeds the -2SD control limit
04/17/2023Nazar Ahmed Mohamed Abd-
Alla(Sangoor)72
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24
Mean
Day
+1SD
+2SD
+3SD
-1SD
-2SD
-3SD
R4S rule violation
Westgard – 41S Rule
04/17/2023Nazar Ahmed Mohamed Abd-
Alla(Sangoor)73
o Requires control data from previous runso Four consecutive QC results for one level of
control are outside ±1SD, oro Both levels of control have consecutive
results that are outside ±1SD
41S - reject when 4 consecutive control measurements exceed the same mean ± 1s limit.
04/17/2023Nazar Ahmed Mohamed Abd-
Alla(Sangoor)74
Westgard – 10X Rule
04/17/2023Nazar Ahmed Mohamed Abd-
Alla(Sangoor)75
Requires control data from previous runsTen consecutive QC results for one level of
control are on one side of the mean, orBoth levels of control have five consecutive
results that are on the same side of the mean
10x Rule = Reject the run when 10 consecutive control
measurements fall on one side of the mean
04/17/2023Nazar Ahmed Mohamed Abd-
Alla(Sangoor)76
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24
Mean
Day
+1SD
+2SD
+3SD
-1SD
-2SD
-3SD
10x rule violation
Moving Averages (Bull’s method)
large laboratories (built into blood analysers)
assumes the population sampled each day remains constant
therefore the calculated indices (MCV, MCH and MCHC) remain stable
determine mean indice values for each batch of 20 patients, plot on control chart
any change: instrument or technical fault
as accurate as 4C preparations 04/17/2023Nazar Ahmed Mohamed Abd-
Alla(Sangoor)77
Moving Averages
MCV = Hct RCC MCH = Hb RCC MCHC = Hb Hct
Changes in the moving
averages graphs indicate where the problem might be in the system. eg. If the light source for Hb is becoming weak, then the calculated MCH and MCHC values will fall
04/17/2023Nazar Ahmed Mohamed Abd-
Alla(Sangoor)78
Moving averages
CAUSE MCV MCH MCHC
LowHb no change low low
High Hb no chang high high
Low RCC high high no change
High RCC low low no change
Low Hct low no change high
High Hct high no change low
04/17/2023Nazar Ahmed Mohamed Abd-
Alla(Sangoor)79
Duplicate tests on patient samples
tests precisionwill not detect incorrect calibration (accuracy) smaller laboratories METHOD test 10 samples repeat the tests calculate the difference between pairs
of results and derive a standard deviation SD should always be < 2SD 04/17/2023
Nazar Ahmed Mohamed Abd-Alla(Sangoor)80
Check tests on patient samples
monitors day-to-day precision (small lab)
detects deterioration within apparatus and reagents
METHOD» select 3-5 normals in the afternoon,
record and average values (WCC, RCC, Hb). Store at 4oC.
» re-assay same samples next morning» tests should agree within 2SD Must ensure there has been no
change in samples 04/17/2023Nazar Ahmed Mohamed Abd-
Alla(Sangoor)81
Corrective action
04/17/2023Nazar Ahmed Mohamed
Abd-Alla(Sangoor)82
Errors assessment and management:Change Old Habits - Recognize Problems.Bad habit Repeat the control .Inspect the control charts or rules violated
to determine type of error.Relate the type of error to potential
causes.Consider factors in common on multitest
systems Relate causes to recent changes Verify the solution and document the
remedy
Turn around Times (TATs)
04/17/2023Nazar Ahmed Mohamed Abd-
Alla(Sangoor)83
Laboratory test turnaround times (TATs) In contrast to laboratorians, the majority of clinicians defined a TAT start time as test ordering, and a TAT ending time as result reporting.
Timely reporting of patient tests can increase efficiency of care and improve customer satisfaction.
In study done 2008 found that postanalytical phase accounted for 64-88% of total tumround time, the pre-analytical phase for 7-17%, and the analytical phase for 2-29%.
04/17/2023Nazar Ahmed Mohamed Abd-
Alla(Sangoor)84
Documentations
Definitions
04/17/2023Nazar Ahmed Mohamed Abd-
Alla(Sangoor)86
Forms: Blank form design to special work to be fill by data at time of use.
Records: worksheets, forms, charts, labels, Used to capture information, activities, or results when performing a procedure
Documents: written policies, process descriptions, procedures, and forms Used to communicate information
May be paper or electronic
Hierarchy of Documents
04/17/2023Nazar Ahmed Mohamed Abd-Alla(Sangoor)87
Policies
Processes
Procedures / Work Instructions
“what to do”
“how it happens”
“how to do it” - (SOPs)
Document Control
04/17/2023Nazar Ahmed Mohamed Abd-
Alla(Sangoor)88
Advantages:Assures that the most current version is used
Ensures availability when neededOrganizational tool
Summary: Document Preparation and Control Process
04/17/2023Nazar Ahmed Mohamed Abd-
Alla(Sangoor)89
Approval
Issue / Distribution
Review
Revision
Preparation
ANY QUESTION?
SUMMARY
04/17/2023Nazar Ahmed Mohamed Abd-
Alla(Sangoor)91
Pateint preparation, and good specimen collection, preparation , handling, and storage of specimen.
Clinical Laboratory instruments daily, weekly, and monthly maintenance and calibration done regular.
Micropipettes recalibration takes place monthly .
Good clinical method selected, and Reagents storage condition verified by monitoring of refrigerators temperature.
SUMMARY
04/17/2023Nazar Ahmed Mohamed Abd-
Alla(Sangoor)92
SOPs written, approved ,and followed carefully, then Sops critical point check list used daily.
MORE than one levels of control sera at least should used in all batches with patient samples.
Results of control sera register in quality book and blotted in levey Jennings chart.
Westgard rules used as guidance for acceptability or rejection of patient results applied.
Errors assessment and management takes place, and corrective action documented.
Turn a round time to all investigation verify monthly.
Documentation (to all lab activities)