The Daisi MT FP combines the Daisi MT-V3 with the new Floating Pastille fixture technology from DATA-PIXEL (patent pending). This major technological breakthrough for multi-fiber connector fixtures is perfectly suited for highly accurate X/Y endface angle measurements. With the Floating Pastille fixtures, measurement results are completely independent from the influence of the fiber optic cables and the operator. Floating Pastille is perfectly suitable for the inspection of connectors mounted with heavy cables. With the help of the Floating Pastille, the Daisi MT-V3 is now easier to operate, without compromising any of its existing features : accuracy of measurements, reproducibility, speed, robustness. No calibration is now required, as it is factory set, per fixture, by DATA-PIXEL. DAISI MT FP Automated interferometer with Floating Pastille Technology KEY FEATURES: X/Y ANGLES MEASUREMENTS: Without Floating Pastille fixture With Floating Pastille fixture SPECIFICATIONS: Measurement Method : Connector Type (PC/APC) : Illumination : Measurement speed (sec.) : Lateral resolution (µm) : Field of view (mm) : Power source : USB outputs : Weight (kg) : Dimensions H x W x L (mm) : Phase shifting + Z scanning MT12/24 & MT16/32 White light + Red light 6 (MT12) 2.5 5.6 x 5.6 12V USB 3.0 + USB 2.0 6.3 133 x 171 x 244 INTERFEROMETRY DAISI MT FP 27 rue Saturne, 74 650 Chavanod, France www.data-pixel.com . e-mail : [email protected] Tél. : +33 (0) 450 673 980 Combined White-Light and Red-Light phase-shifting interferometer for precise tip radius and core-dip measurement Fast Autofocus that simplifies operator’s handling Very high accuracy and reproducibility for small X/Y angles measurements Suitable for field heavy cables, standard small cables MTP/MPO connectors and pigtails with MT ferrules No calibration required Floating Pastille Flange is compatible with existing DAISI MT units* Closed-loop high precision 30 microns Z-scan Vibration insensitive High resolution optics and camera for a better resolution Large field of view able to scan up to 96 fibers * existing units may need to be upgraded, depending on their configuration/version

Upload: others

Post on 20-Jul-2020

2 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: INTERFEROMETRY DAISI MT FP - Data Pixel · 27 rue Saturne, 74 650 Chavanod, France . e-mail : info@data-pixel.com Tél. : +33 (0) 450 673 980 Combined White-Light and Red-Light phase-shifting

The Daisi MT FP combines the Daisi MT-V3 with the new Floating Pastille fixture technology from DATA-PIXEL (patent pending).

This major technological breakthrough for multi-fiber connector fixtures is perfectly suited for highly accurate X/Y endface angle measurements.

With the Floating Pastille fixtures, measurement results are completely independent from the influence of the fiber optic cables and the operator. Floating Pastille is perfectly suitable for the inspection of connectors mounted with heavy cables. With the help of the Floating Pastille, the Daisi MT-V3 is now easier to operate, without compromising any of its existing features : accuracy of measurements, reproducibility, speed, robustness. No calibration is now required, as it is factory set, per fixture, by DATA-PIXEL.

D A I S I M T F PAutomated interferometer with Floating Pasti l le Technology

K E Y F E A T U R E S :

X / Y A N G L E S M E A S U R E M E N T S :

Without Floating Pastille fixture With Floating Pastille fixture

S P E C I F I C A T I O N S :

Measurement Method :

Connector Type (PC/APC) :

Illumination :

Measurement speed (sec.) :

Lateral resolution (µm) :

Field of view (mm) :

Power source :

USB outputs :

Weight (kg) :

Dimensions H x W x L (mm) :

Phase shifting + Z scanning

MT12/24 & MT16/32

White light + Red light

6 (MT12)

2.5

5.6 x 5.6

12V

USB 3.0 + USB 2.0

6.3

133 x 171 x 244

I N T E R F E R O M E T R YDA I S I M T F P

27 rue Saturne, 74 650 Chavanod, Francewww.data-pixel.com . e-mail : [email protected] Tél. : +33 (0) 450 673 980

Combined White-Light and Red-Light phase-shifting interferometer for precise tip radius and core-dip measurement

Fast Autofocus that simplifies operator’s handling

Very high accuracy and reproducibility for small X/Y angles measurements

Suitable for field heavy cables, standard small cables MTP/MPO connectors and pigtails with MT ferrules

No calibration required

Floating Pastille Flange is compatible with existing DAISI MT units*

Closed-loop high precision 30 microns Z-scan

Vibration insensitive

High resolution optics and camera for a better resolution

Large field of view able to scan up to 96 fibers

* existing units may need to be upgraded, depending on their configuration/version

Page 2: INTERFEROMETRY DAISI MT FP - Data Pixel · 27 rue Saturne, 74 650 Chavanod, France . e-mail : info@data-pixel.com Tél. : +33 (0) 450 673 980 Combined White-Light and Red-Light phase-shifting

O R D E R O P T I O N S :

DAISI MT FP unit with Blink Interferometry

Flange Floating Pastille Female MT 12/24

Flange Floating Pastille Female MT 16/32

Adapter MPO for MT12/24 ferrules

Adapter MPO for MT16/32 ferrules

Flange Male MT 12/24

Flange Male MT 16/32

Flange Female MT 12/24

Flange Female MT 16/32

70-DAMT-V3-A02-01

30-FL-002323

30-FL-002325

10-AD-001017

30-AD-001776

10-FL-001354

10-FL-001729

30-FL-001355

30-FL-001357

I N T E R F E R O M E T R YDA I S I M T F P

2019-02-07

Fiber #Fiber # Fiber heightFiber height AdjacentAdjacentheightheight

CoreCoredipdip

Fiber tipFiber tipradiusradius

11 2113.42113.4 nmnm 73.973.9 nmnm -- nmnm 4.14.1 mmmm22 2114.22114.2 nmnm 33.733.7 nmnm -- nmnm 4.34.3 mmmm33 2125.62125.6 nmnm 16.816.8 nmnm -- nmnm 44 mmmm44 2131.72131.7 nmnm 23.823.8 nmnm -- nmnm 4.24.2 mmmm55 2147.12147.1 nmnm 56.456.4 nmnm -- nmnm 4.14.1 mmmm66 2132.82132.8 nmnm 23.723.7 nmnm -- nmnm 3.83.8 mmmm77 2156.42156.4 nmnm 56.356.3 nmnm -- nmnm 3.73.7 mmmm88 2199.72199.7 nmnm 8282 nmnm -- nmnm 3.83.8 mmmm99 2197.22197.2 nmnm 80.280.2 nmnm -- nmnm 3.93.9 mmmm1010 2191.22191.2 nmnm 75.275.2 nmnm -- nmnm 3.53.5 mmmm1111 2187.82187.8 nmnm 7171 nmnm -- nmnm 3.53.5 mmmm1212 21452145 nmnm 57.357.3 nmnm -- nmnm 3.33.3 mmmm

Fiber #Fiber # Fiber heightFiber height AdjacentAdjacentheightheight

CoreCoredipdip

Fiber tipFiber tipradiusradius

1313 2039.52039.5 nmnm 73.973.9 nmnm -- nmnm 44 mmmm1414 2080.52080.5 nmnm 4141 nmnm -- nmnm 4.14.1 mmmm1515 2108.82108.8 nmnm 28.428.4 nmnm -- nmnm 4.44.4 mmmm1616 2107.92107.9 nmnm 23.823.8 nmnm -- nmnm 4.54.5 mmmm1717 2090.62090.6 nmnm 56.456.4 nmnm -- nmnm 4.24.2 mmmm1818 2123.32123.3 nmnm 32.732.7 nmnm -- nmnm 44 mmmm1919 2100.22100.2 nmnm 56.356.3 nmnm -- nmnm 3.93.9 mmmm2020 2117.72117.7 nmnm 8282 nmnm -- nmnm 3.93.9 mmmm2121 2117.12117.1 nmnm 80.280.2 nmnm -- nmnm 3.83.8 mmmm2222 21162116 nmnm 75.275.2 nmnm -- nmnm 3.63.6 mmmm2323 2116.82116.8 nmnm 7171 nmnm -- nmnm 3.63.6 mmmm2424 2087.82087.8 nmnm 57.357.3 nmnm -- nmnm 3.63.6 mmmm

Interferometry report Sample name:

ParameterParameter min allowedmin allowed max allowedmax allowed Measurement resultMeasurement result PASS/FAILPASS/FAILFerrule X angle [SX]Ferrule X angle [SX] -0.15-0.15 °° 0.150.15 °° -0.11-0.11 °° PASSPASSFerrule Y angle [SY]Ferrule Y angle [SY] 7.87.8 °° 8.28.2 °° 8.0418.041 °° PASSPASSFerrule X radius [RX]Ferrule X radius [RX] 20002000 mmmm |-10000||-10000| mmmm 4620646206 mmmm PASSPASSFerrule Y radius [RY]Ferrule Y radius [RY] 55 mmmm |-9999999||-9999999| mmmm 544544 mmmm PASSPASSMax adjacent height diff.Max adjacent height diff. -- 300300 nmnm 8282 nmnm PASSPASSMax core dip [CD]Max core dip [CD] -- 100100 nmnm -- nmnm IGNOREDIGNOREDMinus coplanarity [CF]Minus coplanarity [CF] -- 400400 nmnm 43.943.9 nmnm PASSPASSFibers X angle [GX]Fibers X angle [GX] -0.15-0.15 °° 0.150.15 °° -0.111-0.111 °° PASSPASSFibers Y angle [GY]Fibers Y angle [GY] 7.87.8 °° 8.28.2 °° 8.0468.046 °° PASSPASSGeometry limit [GL]Geometry limit [GL] -- 3535 NN -- NN IGNOREDIGNOREDFlatness deviationFlatness deviation -5000-5000 nmnm 50005000 nmnm 262.3262.3 nmnm PASSPASSValid pixelsValid pixels 2323 %% -- 8484 %% PASSPASS

Generated by Blink v2018.12 build 1001 (32bit) | Copyright Data-Pixel

Device: serial:

Daisi MT 4103350802-2005498027

Profile: Type:

24F APC sub1 MT-24_APC_HQ-pastille Full Report.efd

Custom fields Sample name:

Operator:Station s/n:

Mathieu

4103350802-2005498027

PASSPASS

P E R F O R M A N C E :

Parameters

X/Y angles (°) :

Fiber height (µm) :

Repeatability *

0.001

0.003

Reproducibility **

< 0.01

0.01

Range

± 1 (PC or APC)

Up to 20

* values calculated from 50 measurements without interaction on connector between measurements** values calculated from 50 measurements while removing and inserting connector in ferrule holder between measurements

27 rue Saturne, 74 650 Chavanod, Francewww.data-pixel.com . e-mail : [email protected] Tél. : +33 (0) 450 673 980