New product - The device is not sterile. - Cleaning and sterilization are required prior to first use. Cleaning and Disinfection: - Clean instrument with running water to remove any blood or tissue immediately after use. - Immerse instrument in an approved cleaning/disinfecting solution. - Use of an ultrasonic cleaner is highly recommended. - DO NOT USE agents containing high concentration of chlorine or agents contain- ing oxalic acid. - Use distilled water to prevent water stains. Sterilization - All dental instruments must be be sterilized prior to each use. - Recommended autoclave sterilization protocol: temperature of 134ºC (273ºF), pressure of ~315 Kpa for 6 minutes. - Do not exceed 134ºC. Maintenance - Perform a visual and functional inspection of the instrument prior to sterilization. - Dispose of damaged instruments. Key to codes used Attention, see instructions for use Catalog Number Batch Code Manufacturer MIS Implants Technologies Ltd. P.O.Box 7, Bar Lev Industrial Park, 20156, ISRAEL Website: www.mis-implants.com MIS Implants Technologies GmbH Postfach 1651, 32376 Minden, Germany Tel: +49 571-972-7620; Email: [email protected] EC REP Instrument Maintenance Clean thoroughly immediately after use. Store the ratchet on its own, not attached to any tools. MP-UI054 Rev. 1, Oct. 2013

Upload: others

Post on 01-Feb-2021

1 views

Category:

Documents


0 download

TRANSCRIPT

  • New product

    - The device is not sterile. - Cleaning and sterilization are required prior to first use.

    Cleaning and Disinfection:

    - Clean instrument with running water to remove any blood or tissue immediately after use.

    - Immerse instrument in an approved cleaning/disinfecting solution.

    - Use of an ultrasonic cleaner is highly recommended.

    - DO NOT USE agents containing high concentration of chlorine or agents contain-ing oxalic acid.

    - Use distilled water to prevent water stains.

    Sterilization

    - All dental instruments must be be sterilized prior to each use.

    - Recommended autoclave sterilization protocol: temperature of 134ºC (273ºF), pressure of ~315 Kpa for 6 minutes.

    - Do not exceed 134ºC.

    Maintenance

    - Perform a visual and functional inspection of the instrument prior to sterilization.

    - Dispose of damaged instruments.

    Key to codes used

    Attention, see instructions for use

    Catalog Number

    Batch Code

    Manufacturer

    MIS Implants Technologies Ltd.P.O.Box 7, Bar Lev Industrial Park, 20156, ISRAELWebsite: www.mis-implants.com

    MIS Implants Technologies GmbHPostfach 1651, 32376 Minden, GermanyTel: +49 571-972-7620; Email: [email protected]

    EC REP

    Instrument Maintenance

    Clean thoroughly immediately after use.

    Store the ratchet on its own, not attached to any tools.

    MP-UI054 Rev. 1, Oct. 2013

  • User ManualRatchet Wrench MT-RI030

    General Information The new ratchet is a mono-block instrument with a unique mechanism that simplifies use and cleaning. To prevent damage to the mechanism, it is critical that the ratchet is used only with keys and adapters that are specifically designed for it.

    The ratchet wrench can be used for implant placement and tightening or loosening screws.

    It should be emphasised that MIS recommends the use of a torque controlled driver whenever possible.

    The ratchet wrench MT-RI030 may transfer torque levels that do not correlate to the recommendations specified for implant placement or screw fastening.

    Excess loads may result in damage to implants, components, screws, the wrench itself and even to the bone-to-implant interface.

    ! Warnings