instructions for horiba raman confocal imaging microscope · raman confocal imaging microscope...

17
Dianwen Zhang (Email: [email protected]) Raman confocal imaging microscope Instructions for Horiba Raman confocal imaging microscope Imaging Technology Group Beckman Institute for Advanced Science and Technology University of Illinois at Urbana‐Champaign

Upload: phamngoc

Post on 09-May-2018

238 views

Category:

Documents


2 download

TRANSCRIPT

Page 1: Instructions for Horiba Raman confocal imaging microscope · Raman confocal imaging microscope Instructions for Horiba Raman confocal imaging microscope ... Close the Labspec software

DianwenZhang(Email:[email protected])

Raman confocal imaging microscope

Instructions for Horiba Raman confocal imaging microscope

ImagingTechnologyGroupBeckmanInstituteforAdvancedScienceandTechnology

UniversityofIllinoisatUrbana‐Champaign

Page 2: Instructions for Horiba Raman confocal imaging microscope · Raman confocal imaging microscope Instructions for Horiba Raman confocal imaging microscope ... Close the Labspec software

MicroscopySuite,ImagingTechnologyGroupBeckmanInstituteforAdvancedScienceandTechnologyUniversityofIllinoisatUrbana‐Champaign

OperationmanualofRamanconfocal

405NorthMathews,UrbanaIL61801Tel:217‐244‐0170;FAX:217‐244‐6219

http://itg.beckman.illinois.edu/

RevisionHistoryAuthorsandEditors RevisionDateDianwenZhang 4/20/2012DianwenZhang 12/22/2014DianwenZhang 06/28/2016DianwenZhang 05/10/2017

Page 3: Instructions for Horiba Raman confocal imaging microscope · Raman confocal imaging microscope Instructions for Horiba Raman confocal imaging microscope ... Close the Labspec software

MicroscopySuite,ImagingTechnologyGroupBeckmanInstituteforAdvancedScienceandTechnologyUniversityofIllinoisatUrbana‐Champaign

OperationmanualofRamanconfocal

405NorthMathews,UrbanaIL61801Tel:217‐244‐0170;FAX:217‐244‐6219

http://itg.beckman.illinois.edu/

TableofContents

TheRamanConfocalMicroscopeSystem:Overview..................................................................1 

StandardOperatingProcedure(SOP)forLasers.........................................................................2 

Laserspecifications.............................................................................................................................................2 

DuringnormaloperationsoftheLaser......................................................................................................3 

Laserstart‐upprocedure..................................................................................................................................4 

Lasershutdownprocedure..............................................................................................................................5 

Emergencyresponse..........................................................................................................................................5 

StartUpTheMicroscopeSystem.......................................................................................................6 

ShutDowntheMicroscopeSystem...................................................................................................8 

BasicLabspecSoftwareSettings........................................................................................................9 

GeneralSpectralAcquisitionProcedure......................................................................................10 

DefaultSystemConfigurationforGeneralUsers......................................................................12 

Troubleshooting...................................................................................................................................13 

Page 4: Instructions for Horiba Raman confocal imaging microscope · Raman confocal imaging microscope Instructions for Horiba Raman confocal imaging microscope ... Close the Labspec software

MicroscopySuite,ImagingTechnologyGroupBeckmanInstituteforAdvancedScienceandTechnologyUniversityofIllinoisatUrbana‐Champaign

OperationmanualofRamanconfocal

405NorthMathews,UrbanaIL61801Tel:217‐244‐0170;FAX:217‐244‐6219

http://itg.beckman.illinois.edu/

TheRamanConfocalMicroscopeSystem:Overview

Theleftrearview

Thefrontview

Synapsecamera

AndorNewtonEMCCDcamera

Halogenlampforepi‐illumination

Coarsefocusing

Finefocusing

Laserbutton

Objectiveturret

633nmlaser

830nmlaser 785 nmlaser

405 nmlaser

532 nmlaser

Page 5: Instructions for Horiba Raman confocal imaging microscope · Raman confocal imaging microscope Instructions for Horiba Raman confocal imaging microscope ... Close the Labspec software

MicroscopySuite,ImagingTechnologyGroupBeckmanInstituteforAdvancedScienceandTechnologyUniversityofIllinoisatUrbana‐Champaign

OperationmanualofRamanconfocal

405NorthMathews,UrbanaIL61801Tel:217‐244‐0170;FAX:217‐244‐6219

http://itg.beckman.illinois.edu/ ‐2‐

StandardOperatingProcedure(SOP)forLasers

Laserspecifications

Wavelength  405 nm (blue)  Max power  40 mW 

Laser Class  Class 3B  Laser Type  Continuous wave 

Manufacturer  ONDAX  Serial #  X6000 

Model  LM‐405‐PLR‐40‐2 

Wavelength  532 nm (green)  Max power  50 mW 

Laser Class  Class 3B  Laser Type  Continuous wave 

Manufacturer  Laser Quantum  Serial #  0420506‐82664 

Model  Ventus 532 50mW 

Wavelength  633 nm (red)  Max power 35 mW 

Laser Class  Class 3B  Laser Type  Continuous wave 

Manufacturer  Melles Griot  Serial #   

Model  05‐LHP‐928 

Wavelength  785 nm (invisible)  Max power  575 mW (set to 330 mW) 

Laser Class  Class 4  Laser Type  Continuous wave 

Manufacturer  Toptica Photonics AG  Serial #  XTRA_3V0_01113 

Model  XTRA II 

Wavelength  830 nm (invisible)  Max power  100 mW 

Laser Class  Class 3B  Laser Type  Continuous wave 

Manufacturer  Sacher Lasertechnik group  Serial #  RL‐830‐0711‐0322 

Model  N/A 

Page 6: Instructions for Horiba Raman confocal imaging microscope · Raman confocal imaging microscope Instructions for Horiba Raman confocal imaging microscope ... Close the Labspec software

MicroscopySuite,ImagingTechnologyGroupBeckmanInstituteforAdvancedScienceandTechnologyUniversityofIllinoisatUrbana‐Champaign

OperationmanualofRamanconfocal

405NorthMathews,UrbanaIL61801Tel:217‐244‐0170;FAX:217‐244‐6219

http://itg.beckman.illinois.edu/ ‐3‐

DuringnormaloperationsoftheLaserTheRamanconfocalisaHoribaLABRAMproductandaclass3Blaserproduct,whichmeans

Laserbeamsandreflectedbeamscouldbedangerous.Diffusedreflectedbeamsinmostcasesdonotpresentanyproblem.

Lasersarehighintensitylightsourcesproducingvisibleorinvisiblelight(forexample,785nmand830nmlasersinthesystem)atspecificwavelengths.Thisconcentratedenergyinanarrowlaserbeammaycausedamagetobiologicaltissues,especiallytoeyes.Allbelowlabelsonthemicroscopeindicateahazardoussituationwherethereisariskofseriousinjurytobecausedbylaserradiation.Specialattentionmustbepaid.

Assuredoorremainsclosedandlockedfromtheoutsidewhenlaserisinuse.Theexterior“InUse”lightmustbeturnedonwheneverthelaserisinoperation

Beforeremovingorchangingobjectives,thelaserbuttononthemicroscopefrontpanelmustbein“off”position.Incaseofremovingobjectives,theemptyobjectivepositionsontheobjectiveturretmustbecoveredwithobjectivecaps.

Page 7: Instructions for Horiba Raman confocal imaging microscope · Raman confocal imaging microscope Instructions for Horiba Raman confocal imaging microscope ... Close the Labspec software

MicroscopySuite,ImagingTechnologyGroupBeckmanInstituteforAdvancedScienceandTechnologyUniversityofIllinoisatUrbana‐Champaign

OperationmanualofRamanconfocal

405NorthMathews,UrbanaIL61801Tel:217‐244‐0170;FAX:217‐244‐6219

http://itg.beckman.illinois.edu/ ‐4‐

Whenusing785nmand830nmlasers(invisible),allusersarerequiredtowearlasergogglesthroughoutexperiments.

Usershouldavoidwearingreflectiveaccessories/jewelry(e.g.watches).

Laserstart‐upprocedureThelasercontrollersareshowninFigure1.Onlyonelaserissupposedtobeusedatatime;

DoNOTadjustthesettingsonthelasercontrollers.

Forthe405nmlaser(Figure1(a)),plugthepowersupplytothenearbypoweroutlet.Forthe532nmlaser(Figure1(b)),turnonthekeyswitch,andpressthe“Enable”buttonwhenitstartsflashing.Forthe633nmlaser(Figure1(b)),turnonthekeyswitchonthelasercontroller.Forthe785nmlaser(Figure1(c)),turnonthekeyswitchlocatedontheopticaltableinfrontofthe830nmlasercontroller.Forthe830nmlaser(Figure1(b)),pressonthepowerbutton,waitfor30secondsandpresstheshutter“on/off”buttonon.

Figure1 (a)405nmlaserpowersupply;(b)532,633and830nmlasercontrollers;(c)785laser

Shutter

633nmlasercontroller

532nmlasercontroller

830nmlasercontrollerPower

(b)(a) 405nmlaserpowersupply

785nmlaser“on/off”switch

(c)

Page 8: Instructions for Horiba Raman confocal imaging microscope · Raman confocal imaging microscope Instructions for Horiba Raman confocal imaging microscope ... Close the Labspec software

MicroscopySuite,ImagingTechnologyGroupBeckmanInstituteforAdvancedScienceandTechnologyUniversityofIllinoisatUrbana‐Champaign

OperationmanualofRamanconfocal

405NorthMathews,UrbanaIL61801Tel:217‐244‐0170;FAX:217‐244‐6219

http://itg.beckman.illinois.edu/ ‐5‐

LasershutdownprocedureForthe405nmlaser(Figure1(a)),unplugthepowersupply.Forthe532nmlaser(Figure1(b)),turnoffthekeyswitchonthelasercontroller.Forthe633nmlaser(Figure1(b)),turnoffthekeyswitchonthelasercontroller.Forthe785nmlaser(Figure1(c)),turnoffthekeyswitchlocatedontheopticaltableinfrontofthe830nmlasercontroller.Forthe830nmlaser(Figure1(b)),pressoffthepowerbuttononthelasercontroller.

EmergencyresponseINEVENTOFEMERGENCY:

‐ Turnoffalllasersimmediately.‐ Dial911.‐ Ifalasereyeinjuryissuspected,havetheinjuredpersonkeeptheirheaduprightand

motionlesstorestrictbleedingintheeye.Aphysicianshouldevaluatelasereyeinjuriesassoonaspossible[1].

‐ Ifthereisafire,leavethearea,pullthefirealarm,andcontactthefiredepartmentbycalling911.Donotfightthefireunlessitisverysmallandyouhavebeentrainedinfire‐fightingtechniques[1].

‐ InformthePI,thegroupsafetyofficer,andthebuildingmanagerassoonaspossible.Ifthereisaninjury,thePIneedstosubmitareportofinjurytoRiskManagement.Provideasmanydetailsaspossible[1].

‐ InformtheDRSat217‐333‐2755assoonaspractical.Provideasmanydetailsaspossible[1].

References

[1]UIUCDRSLaserSafetyManual.http://www.drs.illinois.edu/site‐documents/LaserSafetyManual2014.pdf

Page 9: Instructions for Horiba Raman confocal imaging microscope · Raman confocal imaging microscope Instructions for Horiba Raman confocal imaging microscope ... Close the Labspec software

MicroscopySuite,ImagingTechnologyGroupBeckmanInstituteforAdvancedScienceandTechnologyUniversityofIllinoisatUrbana‐Champaign

OperationmanualofRamanconfocal

405NorthMathews,UrbanaIL61801Tel:217‐244‐0170;FAX:217‐244‐6219

http://itg.beckman.illinois.edu/ ‐6‐

StartUpTheMicroscopeSystem

1. Turnonthelaser.

2. SetupthelaserpathtomicroscopeFollowingthenoteontheyellowlabelbesideeachplunger(#1‐4inFigure2),setupthefourfrontplungerstoguidethelasertothelaserentranceportofthemicroscope.

Fortheuserswhouse633nmlaseranddon’tuseSuperhead,pleasealsocheckthefifthplunger(Figure3)inthesystem:

3. Checkthetwoopticalfiltersinsidetheslidingchamber(Figure4)ontopofthemicroscope.Ifthelabelsoftheinstalledfiltersdoesn’tmatchthelaserwavelength,changethem.

Plunger1

Laserentrance(ahole)tothemicroscope

Plunger3

Plunger2

Plunger4

Plunger5needstobepulledupfortheuseof633nmlaserwhenusingthemicroscope.

Figure2 Thefrontplungers

Figure3 Therearplunger

Page 10: Instructions for Horiba Raman confocal imaging microscope · Raman confocal imaging microscope Instructions for Horiba Raman confocal imaging microscope ... Close the Labspec software

MicroscopySuite,ImagingTechnologyGroupBeckmanInstituteforAdvancedScienceandTechnologyUniversityofIllinoisatUrbana‐Champaign

OperationmanualofRamanconfocal

405NorthMathews,UrbanaIL61801Tel:217‐244‐0170;FAX:217‐244‐6219

http://itg.beckman.illinois.edu/ ‐7‐

4. TurnontheLEDlampforepi‐illumination.

5. Openthecontrolsoftware–Labspecandloadsampleformeasurements.IfLabspecfreezesorgivesanerrormessage“ConnectionwithS3000fails,”followtheinstructioninthesectionofTroubleshootingtorestartthemicroscopesystem.TheLabspecsoftwaremanualcanbefoundinthefolderof“Manuals”onthewindowsdesktop.

Openthechambercarefully Carefullychangethesetwofiltersifnecessary.

Figure4Filterschamber

Figure5LEDlightforepi‐illuminationwhitelight

Page 11: Instructions for Horiba Raman confocal imaging microscope · Raman confocal imaging microscope Instructions for Horiba Raman confocal imaging microscope ... Close the Labspec software

MicroscopySuite,ImagingTechnologyGroupBeckmanInstituteforAdvancedScienceandTechnologyUniversityofIllinoisatUrbana‐Champaign

OperationmanualofRamanconfocal

405NorthMathews,UrbanaIL61801Tel:217‐244‐0170;FAX:217‐244‐6219

http://itg.beckman.illinois.edu/ ‐8‐

ShutDowntheMicroscopeSystem1. ClosetheLabspecsoftware,andlogoffthecomputer.2. DecreasetheLEDlightintensitytominimumandthenturnofftheLEDlamp(The

powerbuttonattherearofthelamp).3. Turnthe“Laser”buttononthemicroscopefrontpaneltoOFFposition.4. Turnoffthelasersthatyouhaveused.5. Cleanuptheworkingplace(Takeawayallyourstuffs;usedglassslidesorcoverslips

shouldbethrownintosharpswastebin;etc.).

Page 12: Instructions for Horiba Raman confocal imaging microscope · Raman confocal imaging microscope Instructions for Horiba Raman confocal imaging microscope ... Close the Labspec software

MicroscopySuite,ImagingTechnologyGroupBeckmanInstituteforAdvancedScienceandTechnologyUniversityofIllinoisatUrbana‐Champaign

OperationmanualofRamanconfocal

405NorthMathews,UrbanaIL61801Tel:217‐244‐0170;FAX:217‐244‐6219

http://itg.beckman.illinois.edu/ ‐9‐

BasicLabspecSoftwareSettings

1. Selecttherightlaserline.2. Selecttherightdensityfilter.ThenumbersareopticaldensityOD=Log(Power

transmissionfactor).Thedashlinesgivenodensityfilterinuse.3. Theconfocalpinhole(Hole)controlsmicroscopespatialresolution.Forbestoptical

resolution,setitto200‐300μm,orotherwiseusethemaximumvalue950μm.Fillinadesiredvalue,click“enter”onkeyboard.

4. Clicktheredthunderbuttontoinitializethe“Hole”.5. Theslitsize(Slit)determinesspectralresolution.Fillinadesiredvalue,click“enter”on

keyboard.6. Clicktheredthunderbuttontoinitializethe“Slit”.7. Selectthedesiredgrating.Bydefault,only300g/mmand1800g/mmareavailable.300

g/mmforbroadspectralrange,butpoorspectralresolution,while1800fornarrowspectralrange,butgoodspectralresolution.

8. Payattentiontotheunit.Followthebelowpicturetochangeitifnecessary.

9. Setthecentralvalueofthex‐scaleofthespectrum.Changeitforvaryingspectralrange

under“singlewindow”modespectralacquisition.Fillinadesiredvalue,click“enter”onkeyboard.

10. Clicktheredthunderbuttontoinitializethe“Spectrometer”.11. Selecttheobjectivelenstobeused.Itmustbesetrightbeforeyourmappingacquisition.

Alldefaultobjectivesarelongworkingdistance(LWD)lens.12. Theprefixofyourdatafilename(optional).13. Makesurethisis“XYZMotorized”.14. Thesethreeredthunderbuttonsallowyoutodefinethecurrentsamplepositiontobe

theoriginofcoordinatesinthesystem.Usuallyitisneededtoclickthebuttonforz‐axisbeforeyoudoz‐axisscanningmapping.

15. Theexposuretimeforrealtimespectralacquisition.Theunitisseconds.

1 2

3

74 6 10 1315

11 14

5 89171216

Page 13: Instructions for Horiba Raman confocal imaging microscope · Raman confocal imaging microscope Instructions for Horiba Raman confocal imaging microscope ... Close the Labspec software

MicroscopySuite,ImagingTechnologyGroupBeckmanInstituteforAdvancedScienceandTechnologyUniversityofIllinoisatUrbana‐Champaign

OperationmanualofRamanconfocal

405NorthMathews,UrbanaIL61801Tel:217‐244‐0170;FAX:217‐244‐6219

http://itg.beckman.illinois.edu/ ‐10‐

16. Theexposuretimeforsinglespectrumacquisitionormappingacquisition.Theunitisseconds.

17. Setthenumberofaccumulationsperspectralacquisition.

GeneralSpectralAcquisitionProcedure

1. Checktheabovementionedsettingsatthebottompanelofthesoftwareandmakesuretheyareallcorrect.

2. Checkthe“Laser”buttononthefrontpanelofthemicroscopeandmakesureitisonthe"OFF"position.

3. Startvideobyclicking .4. YoushouldbeabletoseetheLEDwhitelightshiningonyoursamplethroughthe

objecivelens.Adjustthemicroscopefocustogetyourinterestedsampleareaintothe

objectivefocalplaneandthenstopthevideobyclicking .Youcansavetheimageinthevideowindowtoa*.ngvorotherimageformatfile.

5. Turnoffthefluorescenceroomlight,butyoumaykeeptheHalogenceilinglighton.Nowyouarereadyforspectralacquisition.

a. Ifyouaregoingtodoreal‐timespectralacquisition ,turnonthe“Laser”buttononthemicroscopefrontpanel,click tostartitandclick tostopit.Youcansavetheacquiredspectrumtoa*.ngsfileandconvertittoa*.txtfile.

b. Ifyouaregoingtodosinglespectrumacquisition ,followthebelowsteps:1) Ifyouhavecheckedthe“extendedrangeacquisition ”andknownthere

isnothingneededtochange,skipthisstep.Orotherwiseclick ,selecteither“Singlewindow”or“Multiplewindows”mode.Ifyouaregoingtousethe“Multiplewindows”mode,youalsoneedtosetforaspecificspectralrange.

2) Turnonthe“Laser”buttononthemicroscopefrontpanel,click togetasinglespectrum.Everytimeyouclick ,yougetonemorespectrum.

3) Usingthemenu“File→SaveAll”,youcansaveallacquiredspectratoasingle*.ngs ile,oryoucanalsousethemenu“File→SaveAs”tosaveasingleselectedspectrumtoa*.ngsfileorconvertittoa*.txtfile.

c. Ifyouaregoingtodomappingacquisition ,followthebelowsteps:1) Settouse“SingleWindow”modeinthe“extendedrangeacquisition ”.2) Setthemappingproperty .3) Turnonthe“Laser”buttononthemicroscopefrontpanel,click tostart

mappingacquisition.

Page 14: Instructions for Horiba Raman confocal imaging microscope · Raman confocal imaging microscope Instructions for Horiba Raman confocal imaging microscope ... Close the Labspec software

MicroscopySuite,ImagingTechnologyGroupBeckmanInstituteforAdvancedScienceandTechnologyUniversityofIllinoisatUrbana‐Champaign

OperationmanualofRamanconfocal

405NorthMathews,UrbanaIL61801Tel:217‐244‐0170;FAX:217‐244‐6219

http://itg.beckman.illinois.edu/ ‐11‐

4) Yougetthreenewwindowsforthemappingacquisition.Youonlyneedtosavetheallspectradatawindow“SpIm”–clicktoactivethewindow,andsaveittoa*.ngcfile.

Page 15: Instructions for Horiba Raman confocal imaging microscope · Raman confocal imaging microscope Instructions for Horiba Raman confocal imaging microscope ... Close the Labspec software

MicroscopySuite,ImagingTechnologyGroupBeckmanInstituteforAdvancedScienceandTechnologyUniversityofIllinoisatUrbana‐Champaign

OperationmanualofRamanconfocal

405NorthMathews,UrbanaIL61801Tel:217‐244‐0170;FAX:217‐244‐6219

http://itg.beckman.illinois.edu/ ‐12‐

DefaultSystemConfigurationforGeneralUsers

PushintheDualScan/Standardplunger

Pullout–onlySynapseforgeneralusers

Pushinformicroscopemode

Usemicroscopemode

Optional

Turnitononlywhenlaserisneeded UseonlySynapse

forgeneralusers

UseonlyXYstageforgeneralusers

Optional

Page 16: Instructions for Horiba Raman confocal imaging microscope · Raman confocal imaging microscope Instructions for Horiba Raman confocal imaging microscope ... Close the Labspec software

MicroscopySuite,ImagingTechnologyGroupBeckmanInstituteforAdvancedScienceandTechnologyUniversityofIllinoisatUrbana‐Champaign

OperationmanualofRamanconfocal

405NorthMathews,UrbanaIL61801Tel:217‐244‐0170;FAX:217‐244‐6219

http://itg.beckman.illinois.edu/ ‐13‐

Troubleshooting1. Problem:WhenstartingLabspec,itbecomesfrozenorshowsthebelowerrormessage

beforethefinalmainwindowcanappear.

Solution:Followthebelowstepstorestartandresetthemicroscope.

1) CloseLabspec.SometimesyoualsoneedchecktheWindowsTaskManagertomakesurethatthereisnoaprocesscalled“NGSLabspec.exe”runninginbackground.

2) Restartthemicroscope.Findthebelowsocketandthemicroscopepowerboxbesidethelasercontrollersontheopticaltable.Turnoffthesocketswitchpointedbytheyellowarrow.Waitmorethan1minuteandthenturnitbackon.Nowthemicroscopeisrestarted.

3) StartLabspec.Labspecshouldrunnormally,andafterafewcommondevice

initialization,itwillshowthebelowstagecalibrationmessage.Beforeyouclick“OK”,makesurethatnosampleisloadedortheloadedsamplewon’thittheobjectiveswhenthestagemovinginitsfullrange.Click“OK”tostartthestagecalibration.

4) Initializethethreekeymicroscopecomponents‐“Hole,”“Slit”and

“Spectrometer”.Seethebelowimage.Clicktheredthunderbuttonatthetop‐rightonthe“Hole”controlpanel,waituntiltheinitializationprocessiscomplete,andthenseparatelyrepeatthisprocedurefor“Slit”and“Spectrometer”.

Page 17: Instructions for Horiba Raman confocal imaging microscope · Raman confocal imaging microscope Instructions for Horiba Raman confocal imaging microscope ... Close the Labspec software

MicroscopySuite,ImagingTechnologyGroupBeckmanInstituteforAdvancedScienceandTechnologyUniversityofIllinoisatUrbana‐Champaign

OperationmanualofRamanconfocal

405NorthMathews,UrbanaIL61801Tel:217‐244‐0170;FAX:217‐244‐6219

http://itg.beckman.illinois.edu/ ‐14‐