insertos completos de elitech

50
 SEPPIM S.A.S. ZONE INDUSTRIELLE 61500 SEES - FRANCE : 02 33 81 21 00 : 02 33 28 77 51 DIAGNOSTIC REAGENTS MANUAL METHODS Version : MMAN-09 – April 2010

Upload: eduardo-canul

Post on 19-Oct-2015

1.719 views

Category:

Documents


136 download

TRANSCRIPT

  • SEPPIM S.A.S.

    ZONE INDUSTRIELLE 61500 SEES - FRANCE

    : 02 33 81 21 00 : 02 33 28 77 51

    DIAGNOSTIC REAGENTS

    MANUAL METHODS

    Version : MMAN-09 April 2010

  • SEPPIM S.A.S.

    ZONE INDUSTRIELLE 61500 SEES - FRANCE

    : 02 33 81 21 00 : 02 33 28 77 51

    REAGENTS

    A Version

    ACID PHOSPHATASE .................................MM-PACI-1 ALBUMIN .................................................. MM-ALBU-1

    ALP (DEA) SL .............................................. MM-PASL-1 ALP (DEA) ................................................... MM-PALC-1

    ALT/GPT 4+1 SL .................................... MM-ALSL4/1-1 ALT/GPT ..................................................... MM-ALAT-2

    AMYLASE SL ............................................. MM-AMSL-3 AST/GOT 4+1 SL .................................... MM-ASSL4/1-1 AST/GOT ..................................................... MM-ASAT-2

    B

    BILIRUBIN TOTAL & DIRECT 4+1 .......... MM-BITD-2

    C

    CALCIUM ARSENAZO ............................. MM-CALA-1 CALCIUM OCPC ........................................ MM-CALO-1

    Update CHLORIDE ................................................. MM-CHLO-2

    CHOLESTEROL SL ..................................... MM-CHSL-2 Update CHOLESTEROL ......................................... MM-CHOL-3 CHOLESTEROL HDL SL 2G .................... MM-HDLL-1 CHOLESTEROL LDL SL 2G ...................... MM-LDLL-1 HDL CHOLESTEROL ................................ MM-HDLC-2 CHOLINESTERASE .................................... MM-CHES-2

    CK MB SL ................................................... MM-CMSL-2 CK MB ........................................................ MM-CKMB-2

    CK NAC SL .................................................. MM-CKSL-2 CK NAC ....................................................... MM-CKNA-2 COPPER........................................................ MM-CUIV-1

    CREATININE JAFFE .................................. MM-CRCO-1 CREATININE PAP SL ................................. MM-CRSL-3

    G Version GAMMA GT SL ........................................... MM-GASL-1 GAMMA GT ............................................... MM-GAGT-1

    GLUCOSE HK SL ........................................ MM-GHSL-1 GLUCOSE PAP SL ...................................... MM-GPSL-2 GLUCOSE PAP ............................................ MM-GLUP-1

    I IRON CHROMAZUROL ............................. MM-FECA-1 IRON FERROZINE ................................... MM-FEFR-2

    L

    LACTATE .................................................... MM-LACT-2

    New LDH-L SL ......................................................MM-LLSL-1 LDH-P .......................................................... MM-LDHP-2

    M

    MAGNESIUM CALMAGITE ................... MM-MAGN-1 MICROPROTEIN ........................................ MM-PRTP-2

    P

    PHOSPHORUS ............................................ MM-PHOS-1

    T TOTAL PROTEIN PLUS ............................. MM-PROB-1 TRIGLYCERIDES MONO SL NEW ...... MM-TGMLN-2 TRIGLYCERIDES ........................................ MM-TRIG-2

    U

    UREA UV SL ............................................... MM-URSL-2 UREA UV .................................................... MM-URUV-1

    URIC ACID ................................................. MM-ACUR-2 URIC ACID MONO SL .............................. MM-AUML-2 Update URIC ACID SL ............................................ MM-AUSL-2

  • ELITECH ELITECH ELITECH

    SEPPIM S.A.S.- ZONE INDUSTRIELLE - 61500 SEES - FRANCE

    ACID PHOSPHATASE

    Ref. : PACI- 0030 18 x 3 mL

    MANUAL METHOD

    PROCEDURE Wavelength : 405 nm Temperature : 37C Cuvette : 1 cm light path Read against reagent blank.

    Mix and after a 5 minute incubation measure the change of optical density per minute (OD/min.) during 3 minutes. CALCULATION Total ACP : Activity (U/L) = OD/min. x 853 NP ACP : Activity (U/L) = OD/min. x 860 Prostatic fraction (U/L) = Total ACP (U/L) - NP ACP (U/L). LINEARITY If OD/min. exceeds 0.070 at 405 nm, repeat test using serum diluted 1/10 with sodium chloride solution (9 g/L). Multiply result by 10. NOTE: For more details, see the Instructions for use

    (07/2005) MM-PACI-1

    Total acid phosphatase Non-prostatic acid phosphataseWorking Reagent 1 1 mL 1 mL Working Reagent 2 10 L Sample 100 L 100 L

  • ELITECH ELITECH ELITECH

    SEPPIM S.A.S.- ZONE INDUSTRIELLE - 61500 SEES - FRANCE

    ALBUMIN

    Ref.: ALBU - 0600 2 x 125 mL ALBU - 0700 4 x 250 mL

    MANUAL METHOD

    PROCEDURE Wavelength : 628 nm Temperature : 37C Cuvette : 1 cm light path Read against reagent blank.

    Mix and read the optical density (OD) after a 5 minute incubation. The final colour is stable for at least 15 minutes. CALCULATION OD Sample x n OD Standard n = standard concentration

    NOTE: For more details, see the Instructions for use

    (07/2005) MM-ALBU-1

    BLANK STANDARD SAMPLE Reagent 1 mL 1 mL 1 mL Distilled Water 10 L Standard 10 L Sample 10 L

  • ELITECH ELITECH ELITECH

    SEPPIM S.A.S.- ZONE INDUSTRIELLE - 61500 SEES - FRANCE

    ALP (DEA) SL

    Rf. : PASL- 0400 2 x 62.5 mL PASL- 0420 4 x 62.5 mL PASL- 0500 5 x 125 mL

    MANUAL METHOD

    PROCEDURE Wavelength : 405 nm (410) Temperature : 37C, 30C, 25C Cuvette : 1 cm light path Read against distilled water. One-reagent procedure

    Mix and after a 1 minute incubation, measure the change of optical density per minute (OD/min.) during 3 minutes. Two-reagent procedure

    Mix, wait 1 minute and add:

    Mix and after a 1 minute incubation, measure the change of optical density per minute (OD/min.) during 3 minutes. CALCULATION One reagent procedure Two reagents procedure At 405 nm: Activity (U/L) = OD/min. x 2 750 3424 At 410 nm: Activity (U/L) = OD/min. x 2 910 3623 LINEARITY If OD/min. exceeds 0.250 at 405 nm, repeat the test using serum diluted 1/10 with sodium chloride solution (9 g/L). Multiply result by 10. NOTE: For more details, see the Instructions for use

    (07/2005) MM-PASL-1

    Working reagent : 1 mL Sample : 20 L

    R1 : 1 mL Sample : 20 L

    R2 : 250 L

  • ELITECH ELITECH ELITECH

    SEPPIM S.A.S.- ZONE INDUSTRIELLE - 61500 SEES - FRANCE

    ALP (DEA)

    Rf. : PALC- 0030 20 x 3 mL PALC- 0200 12 x 20 mL

    MANUAL METHOD

    PROCEDURE Wavelength : 405 nm (410) Temperature : 25C, 30C, 37C Cuvette : 1 cm light path Read against distilled water.

    Mix and after a 1 minute incubation, measure the change of optical density per minute (OD/min.) during 3 minutes. CALCULATION At 405 nm: Activity (U/L) = OD/min. x 2 750 At 410 nm: Activity (U/L) = OD/min. x 2 910 LINEARITY If OD/min. exceeds 0.250 at 405 nm, repeat the test using serum diluted 1/10 with sodium chloride solution (9 g/L). Multiply result by 10. NOTE: For more details, see the Instructions for use

    (07/2005) MM-PALC-1

    Working reagent : 1 mL Sample : 20 L

  • ELITECH ELITECH ELITECH

    SEPPIM S.A.S.- ZONE INDUSTRIELLE - 61500 SEES - FRANCE

    ALT/GPT 4+1 SL

    Ref.: ALSL 0410 2 x 62.5 mL ALSL 0430 4 x 62.5 mL ALSL 0510 5 x 125 mL

    MANUAL METHOD

    PROCEDURE Wavelength : 340 nm Temperature : 37C Cuvette : 1 cm light path Read against distilled water. One-reagent procedure

    Mix and after a 1 minute incubation, measure the change of optical density per minute (OD/min.) during 3 minutes. Two-reagent procedure

    Mix, wait 1 minute and add:

    Mix and after a 1 minute incubation, measure the change of optical density per minute (OD/min.) during 3 minutes. CALCULATION For the two procedures, at 340 nm: Activity (U/L) = OD/min. x 1 746 LINEARITY If OD/min. exceeds 0.150 at 340 nm, repeat test using serum diluted 1/10 with sodium chloride solution (9 g/L). Multiply result by 10. NOTE: For more details, see the Instructions for use

    (07/2005) MM-ALSL4/1-1

    Working reagent : 1 mL Sample : 100 L

    R1 : 1 mL Sample : 125 L

    R2 : 250 L

  • ELITECH ELITECH ELITECH

    SEPPIM S.A.S.- ZONE INDUSTRIELLE - 61500 SEES - FRANCE

    ALT/GPT

    Ref.: ALAT - 0200 12 x 20 mL ALAT - 0400 9 x 50 mL

    MANUAL METHOD

    PROCEDURE Wavelength : 340 nm (334 365) Temperature : 30C, 37C Cuvette : 1 cm light path Read against distilled water.

    Mix and after a 1 minute incubation, measure the change of optical density per minute (OD/min.) during 3 minutes. CALCULATION At 340 nm: Activity (U/L) = OD/min. x 1 746 At 334 nm: Activity (U/L) = OD/min. x 1 780 At 365 nm: Activity (U/L) = OD/min. x 3235 LINEARITY If OD/min. exceeds 0.150 at 340 nm, repeat test using serum diluted 1/10 with sodium chloride solution (9 g/L). Multiply result by 10. NOTE: For more details, see the Instructions for use

    (03/2008) MM-ALAT-2

    Working reagent : 1 mL Sample : 100 L

  • ELITECH ELITECH ELITECH

    SEPPIM S.A.S.- ZONE INDUSTRIELLE - 61500 SEES - FRANCE

    AMYLASE SL

    Rf. : AMSL - 0390 1 x 50 mL AMSL - 0400 6 x 50 mL

    MANUAL METHOD

    PROCEDURE Wavelength : 405 nm Temperature : 30C, 37C Cuvette : 1 cm light path Read against distilled water.

    Mix and after a 1 minute incubation, measure the change of optical density per minute (OD/min.) during 3 minutes. CALCULATION 37 C 30 C At 405 nm: Activity (U/L) = OD/min. x 4640 2812 Take dilution factor into account for the calculation of the activity in urine. LINEARITY Up to 900 U/L If OD/min. exceeds 0.190 at 405 nm, repeat test using sample diluted 1/10 with sodium chloride solution (9 g/L). Multiply result by 10. NOTE: For more details, see the Instructions for use

    (11/2006) MM-AMSL-3

    37 C 30 C Reagent 1 mL 1 mL Sample 15 L 25 L

  • ELITECH ELITECH ELITECH

    SEPPIM S.A.S.- ZONE INDUSTRIELLE - 61500 SEES - FRANCE

    AST/GOT 4+1 SL

    Rf. : ASSL- 0410 2 x 62.5 mL ASSL- 0430 4 x 62.5 mL ASSL- 0510 5 x 125 mL

    MANUAL METHOD

    PROCEDURE Wavelength : 340 nm Temperature : 37C Cuvette : 1 cm light path Read against distilled water. One-reagent procedure

    Mix and after a 1 minute incubation, measure the change of optical density per minute (OD/min.) during 3 minutes. Two-reagent procedure

    Mix, wait 1 minute and add:

    Mix and after a 1 minute incubation, measure the change of optical density per minute (OD/min.) during 3 minutes. CALCULATION For the two procedures, at 340 nm: Activity (U/L) = OD/min. x 1 746 LINEARITY If OD/min. exceeds 0.150 at 340 nm, repeat test using serum diluted 1/10 with sodium chloride solution (9 g/L). Multiply result by 10. NOTE: For more details, see the Instructions for use

    (07/2005) MM-ASSL4/1-1

    Working reagent : 1 mL Sample : 100 L

    R1 : 1 mL Sample : 125 L

    R2 : 250 L

  • ELITECH ELITECH ELITECH

    SEPPIM S.A.S.- ZONE INDUSTRIELLE - 61500 SEES - FRANCE

    AST/GOT

    Ref.: ASAT - 0200 12 x 20 mL ASAT - 0400 9 x 50 mL

    MANUAL METHOD

    PROCEDURE Wavelength : 340 nm (334 365) Temperature : 30C, 37C Cuvette : 1 cm light path Read against distilled water.

    Mix and after a 1 minute incubation, measure the change of optical density per minute (OD/min.) during 3 minutes. CALCULATION At 340 nm: Activity (U/L) = OD/min. x 1 746 At 334 nm: Activity (U/L) = OD/min. x 1 780 At 365 nm: Activity (U/L) = OD/min. x 3235 LINEARITY If OD/min. exceeds 0.150 at 340 nm, repeat test using serum diluted 1/10 with sodium chloride solution (9 g/L). Multiply result by 10. NOTE: For more details, see the Instructions for use

    (07/2008) MM-ASAT-2

    Working reagent : 1 mL Sample : 100 L

  • ELITECH ELITECH ELITECH

    SEPPIM S.A.S.- ZONE INDUSTRIELLE - 61500 SEES - FRANCE

    BILIRUBIN

    TOTAL & DIRECT 4+1

    Ref.: BITD 0600 Total 125 mL Direct 125 mL

    BITO - 0600 Total 2 x 125 mL

    BIDI - 0600 Direct 2 x 125 mL

    MANUAL METHOD

    PROCEDURE for TOTAL BILIRUBIN

    Wavelength : 550 nm Temperature : 37C Cuvette : 1 cm light path

    Read against distilled water.

    CALIBRATION TEST

    R1 BILIRUBIN TOTAL 4+1 800 L 800 L

    Calibrator 100 L

    Sample 100 L

    Mix and read the optical density (OD) after 5 minutes of incubation at 37C

    OD1 OD2

    Then add:

    R2 BILIRUBIN TOTAL & DIRECT 4+1 200 L 200 L

    Mix and read the optical density (OD) after 5 minutes of incubation at 37C

    OD3 OD4

    PROCEDURE for DIRECT BILIRUBIN

    Wavelength : 550 nm Temperature : 37C Cuvette : 1 cm light path

    Read against distilled water.

    CALIBRATION TEST

    R1 BILIRUBIN DIRECT 4+1 800 L 800 L

    Calibrator 100 L

    Sample 100 L

    Mix and read the optical density (OD) after 5 minutes of incubation at 37C

    OD1 OD2

    Then add:

    R2 BILIRUBIN TOTAL & DIRECT 4+1 200 L 200 L

    Mix and read the optical density (OD) after 5 minutes of incubation at 37C

    OD3 OD4

    CALCULATION FOR BOTH PROCEDURES (Total and Direct Bilirubin)

    With Factor: (OD4 - OD2) x F

    mg/dL: F = 14.0

    mg/L: F = 140

    mol/L: F = 240

    With calibrator: OD4 - OD2 x n

    OD3 - OD1

    n = Calibrator concentration

    NOTE: For more details, see the Instructions for use (08/2009)

    MM-BITD-2

  • ELITECH ELITECH ELITECH

    SEPPIM S.A.S.- ZONE INDUSTRIELLE - 61500 SEES - FRANCE

    CALCIUM ARSENAZO

    Ref.: CALA - 0600 2 x 125 mL

    MANUAL METHOD

    PROCEDURE Wavelength : 650 nm (600) Temperature : 37C Cuvette : 1 cm light path Read against distilled water.

    Mix and read the optical density (OD) after a 1 minute incubation. The final colour is stable for at least 1 hour. CALCULATION OD Sample x n OD Standard n = standard concentration Take dilution factor into account for the calculation of the concentration in urine.

    NOTE: For more details, see the Instructions for use

    (07/2005) MM-CALA-1

    BLANK STANDARD SAMPLE Reagent 1 mL 1 mL 1 mL Distilled Water 10 L Standard 10 L Sample 10 L

  • ELITECH ELITECH ELITECH

    SEPPIM S.A.S.- ZONE INDUSTRIELLE - 61500 SEES - FRANCE

    CALCIUM OCPC

    Ref.: CALO - 0600 2 x 125 mL

    MANUAL METHOD

    PROCEDURE Wavelength : 570 nm Temperature : 37C Cuvette : 1 cm light path Read against distilled water.

    Mix and read the optical density (OD) after a 5 minute incubation. The final colour is stable for at least 1 hour. CALCULATION OD Sample x n OD Standard n = standard concentration Take dilution factor into account for the calculation of the concentration in urine.

    NOTE: For more details, see the Instructions for use

    (07/2005) MM-CALO-1

    BLANK STANDARD SAMPLE Working Reagent 1 mL 1 mL 1 mL Distilled Water 10 L Standard 10 L Sample 10 L

  • ELITECH ELITECH ELITECH

    (09/2009)

    MM-CHLO-2

    SEPPIM S.A.S.- ZONE INDUSTRIELLE - 61500 SEES - FRANCE

    CHLORIDE

    Rf. : CHLO 0600 2 x 125 mL 250 mL

    MANUAL METHOD

    PROCEDURE

    Wavelength : 480 nm

    Temperature : 25C, 30C, 37C

    Cuvette : 1 cm light path

    Read against reagent blank.

    Mix and read the optical density (OD) after a 5 minute incubation.

    The final colour is stable for at least 1 hour.

    Information for urine sample: The urine sample must be diluted with distilled water before analysis.

    CALCULATION

    Serum/Plasma: Urine:

    n StandardOD

    SampleOD 2n

    StandardOD

    SampleOD

    n= Standard concentration

    NOTE: For more details, see the Instructions for use

    BLANK STANDARD SAMPLE

    Reagent 1 mL 1 mL 1 mL

    Distilled Water 10 L

    Standard 10 L

    Sample 10 L

  • ELITECH ELITECH ELITECH

    SEPPIM S.A.S. ZONE INDUSTRIELLE - 61500 SEES - FRANCE

    CHOLESTEROL SL

    Rf. : CHSL- 0490 1 x 100 mL CHSL- 0500 6 x 100 mL CHSL- 0700 4 x 250 mL CHSL- 0507 6 x 100 mL* CHSL- 0707 4 x 250 mL*

    MANUAL METHOD

    PROCEDURE Wavelength : 500 nm (492-550) Temperature : 37C Cuvette : 1 cm light path Read against reagent blank.

    Mix and read the optical density (OD) after a 5 minute incubation. The final colour is stable for at least 1 hour. CALCULATION OD Sample x n OD Standard

    n= Standard concentration NOTE: For more details, see the Instructions for use

    (07/2007) * these kits contain a standard MM-CHSL-2

    BLANK STANDARD SAMPLE Reagent 1 mL 1 mL 1 mL Distilled Water 10 L Standard 10 L Sample 10 L

  • ELITECH ELITECH ELITECH

    SEPPIM S.A.S.- ZONE INDUSTRIELLE - 61500 SEES - FRANCE

    CHOLESTEROL

    Ref.: CHOL - 0220 12 x 20 mL CHOL - 0420 9 x 50 mL

    MANUAL METHOD

    PROCEDURE Wavelength : 500 nm (492 550) Temperature : 37C Cuvette : 1 cm light path Read against reagent blank.

    Mix and read the optical density (OD) after a 5 minute incubation. The final colour is stable for at least 1 hour. CALCULATION OD Sample x n OD Standard

    n= Standard concentration

    NOTE: For more details, see the Instructions for use

    (03/2010) MM-CHOL-3

    BLANK STANDARD SAMPLE Working Reagent 1 mL 1 mL 1 mL Distilled Water 10 L Standard 10 L Sample 10 L

  • ELITECH ELITECH ELITECH

    SEPPIM S.A.S.- ZONE INDUSTRIELLE - 61500 SEES - FRANCE

    CHOLESTEROL HDL SL 2G

    Ref.: HDLL 0380 1 x 80 mL HDLL 0390 3 x 80 mL

    MANUAL METHOD

    PROCEDURE Wavelength : 600 nm Temperature : 37C Cuvette : 1 cm light path Read against reagent blank. BLANK STANDARD SAMPLE R1 900 L 900 L 900 L Distilled water 9 L Standard 9 L Sample 9 L Mix and read absorbance (A1) after 5 minutes of incubation at 37C. Then add: R2 300 L 300 L 300 L Mix and read absorbance (A2) after 5 minutes of incubation at 37C CALCULATION

    nCalibratorFdilAA

    SampleFdilAA

    )()(

    12

    12 n = Calibrator concentration

    Fdil = Dilution factor = (R1 volume + sample volume)/(R1 volume+R2 volume + sample volume) NOTE: For more details, see the Instructions for use

    (04/2009) MM-HDLL-1

  • ELITECH ELITECH ELITECH

    SEPPIM S.A.S.- ZONE INDUSTRIELLE - 61500 SEES - FRANCE

    CHOLESTEROL LDL SL 2G

    Ref.: LDLL 0380 1 x 80 mL

    MANUAL METHOD

    PROCEDURE Wavelength : 550 nm Temperature : 37C Cuvette : 1 cm light path Read against reagent blank. BLANK STANDARD SAMPLE R1 900 L 900 L 900 L Distilled water 9 L Standard 9 L Sample 9 L Mix and read absorbance (A1) after 5 minutes of incubation at 37C. Then add: R2 300 L 300 L 300 L Mix and read absorbance (A2) after 5 minutes of incubation at 37C CALCULATION

    nCalibratorFdilAA

    SampleFdilAA

    )()(

    12

    12 n = Calibrator concentration

    Fdil = Dilution factor = (R1 volume + sample volume)/(R1 volume+R2 volume + sample volume) NOTE: For more details, see the Instructions for use

    (04/2009) MM-LDLL-1

  • ELITECH ELITECH ELITECH

    SEPPIM S.A.S.- ZONE INDUSTRIELLE - 61500 SEES - FRANCE

    HDL

    CHOLESTEROL

    Ref.: HDLC - 0060 3 x 10 mL x 100 mL

    MANUAL METHOD

    PROCEDURE

    1) Sample preparation: Add to 500 L of sample, 50 L of precipitating reagent. Mix, wait for 10 minutes and centrifuge at 5000 rpm

    for 15 minutes. The supernatant is collected for HDL determination.

    2) HDL determination:

    Wavelength : 500 nm (492-550)

    Temperature : 37C

    Cuvette : 1 cm light path

    Read against reagent blank.

    Mix and read the optical density (OD) after a 5 minute incubation.

    The final colour is stable for at least 1 hour.

    CALCULATION

    OD Sample x n x 1.1

    OD Standard

    n = Standard concentration

    1.1 = Sample dilution factor

    NOTE: For more details, see the Instructions for use

    (09/2009)

    MM-HDLC-2

    BLANK STANDARD SAMPLE

    Working Reagent 1 mL 1 mL 1 mL

    Distilled Water 50 L

    Standard 50 mg/dL 50 L

    Supernatant 50 L

  • ELITECH ELITECH ELITECH

    SEPPIM S.A.S.- ZONE INDUSTRIELLE - 61500 SEES - FRANCE

    CHOLINESTERASE

    Ref.: CHES - 0053 16 x 3 mL

    MANUAL METHOD

    PROCEDURE Wavelength : 405 nm (400 440) Temperature : 25C, 30C, 37C Cuvette : 1 cm light path Read against distilled water.

    Mix and measure the change of optical density every 30 seconds (OD/30 s.) during 90 seconds. CALCULATION Activity (U/L) = OD/30 s. x 23 460 LINEARITY If OD/30 sec. exceeds 0.200 at 405 nm, repeat test using serum diluted 1/5 with sodium chloride solution (9 g/L). Multiply result by 10. NOTE: For more details, see the Instructions for use

    (03/2008) MM-CHES-2

    Working reagent 1 : 3 mL Sample : 20 L Working reagent 2 : 100 L

  • ELITECH ELITECH ELITECH

    SEPPIM S.A.S.- ZONE INDUSTRIELLE - 61500 SEES - FRANCE

    CK MB SL

    Rf. : CMSL- 0410 2 x 62.5 mL CMSL- 0430 4 x 62.5 mL

    MANUAL METHOD

    PROCEDURE Wavelength : 340 nm Temperature : 30C, 37C Cuvette : 1 cm light path Read against distilled water. One-reagent procedure

    Mix and after a 3 minute incubation, measure the change of optical density per minute (OD/min.) during 3 minutes. ) Two-reagent procedure

    Mix and wait 30 seconds

    Mix and after a 5 minute incubation, measure the change of optical density per minute (OD/min.) during 3 minutes. CALCULATION For the two procedures, at 340 nm: Activity (U/L) = OD/min. x 8254 LINEARITY Inhibition : up to 2 000 U/L of CK-MM at 37 C. If OD/min. exceeds 0.075 dilute the sample (one part with nine parts saline solution), repeat the test and multiply the result by 10. Linear up at least 600 U/L. NOTE: For more details, see the Instructions for use

    (11/2008) ): Modification from the previous version MM-CMSL-2

    Working reagent : 1 mL Sample : 40 L

    R1 : 1 mL R2 : 250 L

    Sample : 50 L

  • ELITECH ELITECH ELITECH

    SEPPIM S.A.S.- ZONE INDUSTRIELLE - 61500 SEES - FRANCE

    CK MB

    Rf. : CKMB- 0030 20 x 3 mL

    MANUAL METHOD

    PROCEDURE Wavelength : 340 nm Temperature : 30C, 37C Cuvette : 1 cm light path Read against distilled water.

    Mix and after a 5 minute incubation, measure the change of optical density per minute (OD/min.) during 3 minutes. CALCULATION

    a) Total CK activity : Determine total CK activity in serum using the CK-NAC reagent. b) CK-B activity : CK-B (U/L) = OD/min x 3 333 c) CK-MB activity : CK-MB (U/L) = CK-B (U/L) x 2 d) Percentage of CK-MB activity in sample : % CK-MB = CK-MB x 100 Total-CK

    LINEARITY Inhibition : up to 2000 U/L of CK-MM at 37 C. If the assay of the total CK exceeds 1200 U/L at 37C, dilute the sample appropriately with physiological saline before assay of CK-MB. Multiply the result by the dilution factor to obtain the correct value of the isoenzyme. NOTE: For more details, see the Instructions for use

    (05/2006) MM-CKMB-2

    Working reagent : 1 mL Sample : 50 L

  • ELITECH ELITECH ELITECH

    SEPPIM S.A.S.- ZONE INDUSTRIELLE - 61500 SEES - FRANCE

    CK NAC SL

    Rf. : CKSL- 0410 2 x 62.5 mL CKSL- 0430 4 x 62.5 mL

    MANUAL METHOD

    PROCEDURE Wavelength : 340 nm Temperature : 30C, 37C Cuvette : 1 cm light path Read against distilled water. One-reagent procedure

    Mix and after a 2 minute incubation, measure the change of optical density per minute (OD/min.) during 3 minutes. ) Two-reagent procedure

    Mix and wait 30 seconds

    Mix and after a 2 minute incubation, measure the change of optical density per minute (OD/min.) during 3 minutes. CALCULATION For the two procedures, at 340 nm: Activity (U/L) = OD/min. x 4127 LINEARITY If OD/min. exceeds 0.250 at 340 nm, repeat test using serum diluted 1/10 with sodium chloride solution (9 g/L). Multiply result by 10. NOTE: For more details, see the Instructions for use

    (11/2008) ): Modification from the previous version MM-CKSL-2

    Working reagent : 1 mL Sample : 40 L

    R1 : 1 mL R2 : 250 L

    Sample : 50 L

  • ELITECH ELITECH ELITECH

    SEPPIM S.A.S.- ZONE INDUSTRIELLE - 61500 SEES - FRANCE

    CK NAC

    Rf. : CKNA- 0030 20 x 3 mL

    MANUAL METHOD

    PROCEDURE Wavelength : 340 nm (334-365) Temperature : 25C, 30C, 37C Cuvette : 1 cm light path Read against distilled water.

    Mix and after a 2 minute incubation, measure the change of optical density per minute (OD/min.) during 3 minutes. CALCULATION At 340 nm : Activity (U/L) = OD/min. x 8 095 At 334 nm : Activity (U/L) = OD/min. x 8 252 At 365 nm : Activity (U/L) = OD/min. x 14 571

    LINEARITY If OD/min. exceeds 0.150 at 340 nm, repeat test using serum diluted 1/10 with sodium chloride solution (9 g/L). Multiply result by 10. NOTE: For more details, see the Instructions for use

    (03/2008) MM-CKNA-2

    Working reagent : 1 mL Sample : 20 L

  • ELITECH ELITECH ELITECH

    SEPPIM S.A.S.- ZONE INDUSTRIELLE - 61500 SEES - FRANCE

    COPPER

    Rf. : CUIV- 0050 5 x 10 mL

    MANUAL METHOD

    PROCEDURE Wavelength : 580 nm (570-590) Temperature : 37C Cuvette : 1 cm light path Read against reagent blank. One-reagent procedure

    Mix and read the optical density (OD) after a 5 minute incubation. Two-reagent procedure

    Mix and read the optical density (OD1).

    Mix and read the optical density (OD2) after a 5 minute incubation. The final colour is stable for at least 1 hour. CALCULATION One-reagent procedure : OD Sample x n OD Standard

    Two-reagent procedure : OD2 - OD1 Sample x n OD2 - OD1 Standard

    n = standard concentration NOTE: For more details, see the Instructions for use

    (07/2005) MM-CUIV-1

    BLANK STANDARD SAMPLE Working Reagent 2 1 mL 1 mL 1 mL Distilled Water 50 L Standard 50 L Sample 50 L

    BLANK STANDARD SAMPLE Working Reagent 1 1 mL 1 mL 1 mL Distilled Water 50 L Standard 50 L Sample 50 L

    Reagent 2 50 L 50 L 50 L

  • ELITECH ELITECH ELITECH

    SEPPIM S.A.S.- ZONE INDUSTRIELLE - 61500 SEES - FRANCE

    CREATININE JAFFE

    Ref.: CRCO - 0600 2 x 125 mL CRCO - 0700 4 x 250 mL

    MANUAL METHOD

    PROCEDURE Wavelength : 492 nm (480 520) Temperature : 37C Cuvette : 1 cm light path Read against distilled water.

    Mix and read the optical density (OD1) 10 seconds after the sample or standard addition. Exactly 2 minutes after the first reading, take second reading (OD2). CALCULATION OD Sample x n OD Standard Take dilution factor into account for calculation of concentration in urine. NOTE: For more details, see the Instructions for use

    (07/2005) MM-CRCO-1

    Working reagent : 1 mL Sample : 100 L

  • ELITECH ELITECH ELITECH

    SEPPIM S.A.S.- ZONE INDUSTRIELLE - 61500 SEES - FRANCE

    CREATININE PAP SL

    Ref.: CRSL - 0630 2 x 133 mL

    MANUAL METHOD

    PROCEDURE Wavelength : 550 nm Temperature : 37C Cuvette : 1 cm light path Read against reagent blank.

    Mix, and wait 5 minutes at 37C. Then, add:

    Mix and read the change of optical density (OD) between 1 and 4 minutes.. CALCULATION OD Sample x n OD Standard n = standard concentration Take dilution factor into account for the calculation of the concentration in urine.

    NOTE: For more details, see the Instructions for use

    (03/2008) MM-CRSL-3

    BLANK STANDARD SAMPLE R1 1.5 mL 1.5 mL 1.5 mL Distilled Water 35 L Standard 35 L Sample 35 L

    R2 0.5 mL 0.5 mL 0.5 mL

  • ELITECH ELITECH ELITECH

    SEPPIM S.A.S.- ZONE INDUSTRIELLE - 61500 SEES - FRANCE

    GAMMA GT SL

    Ref.: GASL - 0400 2 x 62.5 mL GASL - 0420 4 x 62.5 mL GASL - 0500 5 x 125 mL

    MANUAL METHOD

    PROCEDURE Wavelength : 405 nm Temperature : 30C, 37C Cuvette : 1 cm light path Read against distilled water. One-reagent procedure

    Mix and after a 1 minute incubation, measure the change of optical density per minute (OD/min.) during 3 minutes. Two-reagent procedure

    Mix, wait 1 minute and add:

    Mix and after a 1 minute incubation, measure the change of optical density per minute (OD/min.) during 3 minutes. CALCULATION

    One reagent procedure Two reagents procedure At 405 nm: Activity (U/L) = OD/min. x 1 158 1421 LINEARITY If OD/min. exceeds 0.200 at 405 nm, repeat test using serum diluted 1/10 with sodium chloride solution (9 g/L). Multiply result by 10. NOTE: For more details, see the Instructions for use

    (07/2005) MM-GASL-1

    Working reagent : 1 mL Sample : 100 L

    R1 : 1 mL Sample : 100 L

    R2 : 250 L

  • ELITECH ELITECH ELITECH

    SEPPIM S.A.S.- ZONE INDUSTRIELLE - 61500 SEES - FRANCE

    GAMMA GT

    Ref.: GAGT 0030 20 x 3 mL GAGT 0200 12 x 20 mL

    MANUAL METHOD

    PROCEDURE Wavelength : 405 nm Temperature : 30C, 37C Cuvette : 1 cm light path Read against distilled water.

    Mix and after a 1 minute incubation, measure the change of optical density per minute (OD/min.) during 3 minutes. CALCULATION At 405 nm: Activity (U/L) = OD/min. x 1 111 LINEARITY If OD/min. exceeds 0.200 at 405 nm, repeat test using serum diluted 1/10 with sodium chloride solution (9 g/L). Multiply result by 10. NOTE: For more details, see the Instructions for use

    (07/2005) MM-GAGT-1

    Working reagent : 1 mL Sample : 100 L

  • ELITECH ELITECH ELITECH

    SEPPIM S.A.S.- ZONE INDUSTRIELLE - 61500 SEES - FRANCE

    GLUCOSE HK SL

    Ref.: GHSL - 0600 5 x 125 mL

    MANUAL METHOD

    PROCEDURE Wavelength : 340 nm (334-365) Temperature : 37C Cuvette : 1 cm light path Read against reagent blank. One-reagent procedure

    Mix and read the optical density (OD) after a 5 minute incubation. Two-reagents procedure

    Mix, wait 1 minute and add:

    Mix and read the optical density (OD) after a 5 minute incubation. CALCULATION OD Sample x n OD Standard

    n= Standard concentration

    NOTE: For more details, see the Instructions for use

    (07/2005) MM-GHSL-1

    BLANK STANDARD SAMPLE Working Reagent 1 mL 1 mL 1 mL Distilled Water 10 L Standard 10 L Sample 10 L

    BLANK STANDARD SAMPLE R1 1 mL 1 mL 1 mL Distilled Water 10 L Standard 10 L Sample 10 L

    R2 250 L 250 L 250 L

  • ELITECH ELITECH ELITECH

    SEPPIM S.A.S. ZONE INDUSTRIELLE - 61500 SEES - FRANCE

    GLUCOSE PAP SL

    Ref.: GPSL - 0490 1 x 100 mL GPSL - 0500 6 x 100 mL GPSL - 0700 4 x 250 mL GPSL - 0507 6 x 100 mL* GPSL - 0707 4 x 250 mL*

    MANUAL METHOD

    PROCEDURE Wavelength : 500 nm (492-550) Temperature : 37C Cuvette : 1 cm light path Read against reagent blank.

    Mix and measure the optical density (OD) after a 10 minute incubation. The reading can be done during 30 minutes. CALCULATION OD Sample x n OD Standard

    n= Standard concentration

    NOTE: For more details, see the Instructions for use

    (07/2007) * these kits contain a standard MM-GPSL-2

    BLANK STANDARD SAMPLE Working Reagent 1 mL 1 mL 1 mL Distilled Water 10 L Standard 10 L Sample 10 L

  • ELITECH ELITECH ELITECH

    SEPPIM S.A.S.- ZONE INDUSTRIELLE - 61500 SEES - FRANCE

    GLUCOSE PAP

    Ref.: GLUP - 0700 4 x 250 mL GLUP - 0800 5 x 500 mL

    MANUAL METHOD

    PROCEDURE Wavelength : 500 nm (492-550) Temperature : 37C Cuvette : 1 cm light path Read against reagent blank.

    Mix and read the optical density (OD) after a 10 minute incubation. The final colour is stable for at least 1 hour. CALCULATION OD Sample x n OD Standard

    n= Standard concentration

    NOTE: For more details, see the Instructions for use

    (07/2005) MM-GLUP-1

    BLANK STANDARD SAMPLE Working Reagent 1 mL 1 mL 1 mL Distilled Water 10 L Standard 10 L Sample 10 L

  • ELITECH ELITECH ELITECH

    SEPPIM S.A.S.- ZONE INDUSTRIELLE - 61500 SEES - FRANCE

    IRON CHROMAZUROL

    Ref.: FECA - 0600 2 x 125 mL

    MANUAL METHOD

    PROCEDURE Wavelength : 637 nm Temperature : 37C Cuvette : 1 cm light path Read against reagent blank.

    Mix and read the optical density (OD) after a 15 minute incubation. The final colour is stable for at least 1 hour. CALCULATION OD Sample x n OD Standard

    n= Standard concentration

    NOTE: For more details, see the Instructions for use

    (07/2005) MM-FECA-1

    BLANK STANDARD SAMPLE Reagent 1 mL 1 mL 1 mL Distilled Water 50 L Standard 50 L Sample 50 L

  • ELITECH ELITECH ELITECH

    SEPPIM S.A.S.- ZONE INDUSTRIELLE - 61500 SEES - FRANCE

    IRON FERROZINE

    Ref.: FEFR 0600 2 x 125 mL

    MANUAL METHOD

    PROCEDURE Wavelength : 560 nm (550-570) Temperature : 37C Cuvette : 1 cm light path Read against distilled water.

    Mix and read the optical density (OD).

    Then add:

    Mix and read the optical density (OD) within one hour.

    CALCULATION OD4 - OD2 x n OD3 - OD1 n = standard concentration

    NOTE: For more details, see the Instructions for use

    (09/2009) MM-FEFR-2

    BLANK STANDARD SAMPLE Working Reagent 1 mL 1 mL 1 mL Distilled Water 200 L Standard 200 L Sample 200 L

    - OD1 OD2

    R2 50 L 50 L 50 L

    - OD3 OD4

  • ELITECH ELITECH ELITECH

    SEPPIM S.A.S.- ZONE INDUSTRIELLE - 61500 SEES - FRANCE

    LACTATE

    Rf. : LACT- 0100 10 x 10 mL

    MANUAL METHOD

    PROCEDURE Wavelength : 540 nm (500-550) Temperature : 37C Cuvette : 1 cm light path Read against reagent blank.

    Mix and read the optical density (OD) after a 5 minute incubation. CALCULATION OD Sample x n OD Standard

    n= Standard concentration NOTE: For more details, see the Instructions for use

    (05/2006) MM-LACT-2

    BLANK STANDARD SAMPLE Working Reagent 1 mL 1 mL 1 mL Distilled Water 10 L Standard 10 L Sample 10 L

  • ELITECH ELITECH ELITECH

    SEPPIM S.A.S.- ZONE INDUSTRIELLE - 61500 SEES - FRANCE

    LDH-L SL

    Ref.: LLSL - 0400 2 x 62.5 mL LLSL - 0420 4 x 62.5 mL

    MANUAL METHOD

    PROCEDURE Wavelength : 340 nm Temperature : 37C Cuvette : 1 cm light path Read against reagent blank. Two-reagent procedure BLANK SAMPLE R1 1 mL 1 mL Distilled Water 35 L Sample 35 L Mix, wait 3 minute and add: R2 250 L 250 L Mix and after a 100 second incubation, measure the change of optical density per minute (OD/min.) during 3 minutes. CALCULATION At 340 nm: Activity (U/L) = OD/min. x 5828 NOTE: For more details, see the Instructions for use

    (10/2007) MM-LLSL-1

  • ELITECH ELITECH ELITECH

    SEPPIM S.A.S.- ZONE INDUSTRIELLE - 61500 SEES - FRANCE

    LDH-P

    Ref. : LDHP - 0030 20 x 3 mL

    MANUAL METHOD

    PROCEDURE Wavelength : 340 nm (334-365) Temperature : 30C, 37C Cuvette : 1 cm light path Read against distilled water.

    Mix and after a 1 minute incubation, measure the change of optical density per minute (OD/min.) during 3 minutes. CALCULATION At 340 nm: Activity (U/L) = OD/min. x 8095 At 334 nm: Activity (U/L) = OD/min. x 8252 At 365 nm: Activity (U/L) = OD/min. x 15000 LINEARITY If OD/min. exceeds 0.150 at 340 nm, repeat test using serum diluted 1/5 or 1/10 with sodium chloride solution (9 g/L). Multiply result by 5 or 10. NOTE: For more details, see the Instructions for use

    (03/2008) MM-LDHP-2

    Working reagent : 1 mL Sample : 20 L

  • ELITECH ELITECH ELITECH

    SEPPIM S.A.S.- ZONE INDUSTRIELLE - 61500 SEES - FRANCE

    MAGNESIUM CALMAGITE

    Ref.: MAGN - 0600 2 x 125 mL

    MANUAL METHOD

    PROCEDURE Wavelength : 520 nm (500-550) Temperature : 37C Cuvette : 1 cm light path Read against reagent blank.

    Mix and read the optical density (OD) after a 5 minute incubation. The final colour is stable for at least 1 hour. CALCULATION OD Sample x n OD Standard

    n= Standard concentration Take dilution factor into account for the calculation of magnesium concentration in urine.

    NOTE: For more details, see the Instructions for use

    (07/2005) MM-MAGN-1

    BLANK STANDARD SAMPLE Working Reagent 1 mL 1 mL 1 mL Distilled Water 10 L Standard 10 L Sample 10 L

  • ELITECH ELITECH ELITECH

    SEPPIM S.A.S.- ZONE INDUSTRIELLE - 61500 SEES - FRANCE

    MICROPROTEIN

    Ref.: PRTP - 0600 2 x 125 mL

    MANUAL METHOD

    PROCEDURE Wavelength : 598 nm Temperature : 37C Cuvette : 1 cm light path Read against reagent blank.

    Mix and read the optical density (OD) after a 5 minute incubation. ) The final colour is stable for 10 minutes. CALCULATION OD Sample x n OD Standard

    n= Standard concentration

    NOTE: For more details, see the Instructions for use

    (11/2008) ): Modification from the previous version MM-PRTP-2

    BLANK STANDARD SAMPLE Reagent 1 mL 1 mL 1 mL Distilled Water 20 L Standard 20 L Sample 20 L

  • ELITECH ELITECH ELITECH

    SEPPIM S.A.S.- ZONE INDUSTRIELLE - 61500 SEES - FRANCE

    PHOSPHORUS

    Ref.: PHOS - 0600 2 x 125 mL

    MANUAL METHOD

    PROCEDURE Wavelength : 340 nm Temperature : 37C Cuvette : 1 cm light path Read against reagent blank.

    Mix and read the optical density (OD) after a 5 minute incubation. The final colour is stable for at least 1 hour. CALCULATION OD Sample x n OD Standard

    n= Standard concentration Take dilution factor into account for the calculation of phosphorus concentration in urine.

    NOTE: For more details, see the Instructions for use

    (07/2005) MM-PHOS-1

    BLANK STANDARD SAMPLE Reagent 1 mL 1 mL 1 mL Distilled Water 10 L Standard 10 L Sample 10 L

  • ELITECH ELITECH ELITECH

    (08/2009)

    MM-PROB-1

    SEPPIM S.A.S.- ZONE INDUSTRIELLE - 61500 SEES - FRANCE

    TOTAL PROTEIN PLUS

    Rf.: PROB - 0600 2 x 125 mL

    PROB - 0700 4 x 250 mL

    MANUAL METHOD

    PROCEDURE

    Wavelength : 546 nm

    Temperature : 37C

    Cuvette : 1 cm light path

    Read against reagent blank.

    Mix and read the optical density (OD) after 10 minutes incubation.

    The final colour is stable for 20 minutes.

    CALCULATION

    nStandardOD

    SampleOD

    n= Standard concentration

    NOTE: For more details, see the Instructions for use

    BLANK STANDARD SAMPLE

    Reagent 1 mL 1 mL 1 mL

    Distilled Water 10 L

    Standard 10 L

    Sample 10 L

  • ELITECH ELITECH ELITECH

    SEPPIM S.A.S. ZONE INDUSTRIELLE - 61500 SEES - FRANCE

    TRIGLYCERIDES MONO SL NEW

    Rf. : TGML-0425 6 x 50 mL TGML-0515 6 x 100 mL TGML-0700 4 x 250 mL TGML-0427 6 x 50 mL* TGML-0517 6 x 100 mL* TGML-0707 4 x 250 mL*

    MANUAL METHOD

    PROCEDURE Wavelength : 500 nm Temperature : 37C Cuvette : 1 cm light path Read against reagent blank.

    Mix and read the optical density (OD) after a 10 minute incubation. CALCULATION OD Sample x n OD Standard

    n= Standard concentration NOTE: For more details, see the Instructions for use

    (07/2007) * these kits contain a standard MM-TGMLN-2

    BLANK STANDARD SAMPLE Reagent 1 mL 1 mL 1 mL Distilled Water 10 L Standard 10 L Sample 10 L

  • ELITECH ELITECH ELITECH

    SEPPIM S.A.S.- ZONE INDUSTRIELLE - 61500 SEES - FRANCE

    TRIGLYCERIDES

    Ref.: TRIG - 0200 12 x 20 mL TRIG - 0400 9 x 50 mL

    MANUAL METHOD

    PROCEDURE Wavelength : 546 nm (520 570) Temperature : 37C Cuvette : 1 cm light path Read against reagent blank.

    Mix and read the optical density (OD) after a 5 minute incubation. The final colour is stable for at least 30 minutes. CALCULATION OD Sample x n OD Standard

    n= Standard concentration

    NOTE: For more details, see the Instructions for use

    (07/2007) MM-TRIG-2

    BLANK STANDARD SAMPLE Working Reagent 1 mL 1 mL 1 mL Distilled Water 10 L Standard 10 L Sample 10 L

  • ELITECH ELITECH ELITECH

    SEPPIM S.A.S.- ZONE INDUSTRIELLE - 61500 SEES - FRANCE

    UREA UV SL

    Rf. : URSL- 0400 2 x 62.5 mL URSL- 0420 4 x 62.5 mL URSL- 0500 5 x 125 mL URSL- 0407 2 x 62.5 mL* URSL- 0427 4 x 62.5 mL* URSL- 0507 5 x 125 mL*

    MANUAL METHOD

    PROCEDURE Wavelength : 340 nm Temperature : 37C Cuvette : 1 cm light path Read against distilled water. One-reagent procedure

    Mix and read the variation of optical density (OD) between 30 seconds and 90 seconds. Two-reagent procedure

    Mix, wait 1 minute and add:

    Mix and read the variation of optical density (OD) between 30 seconds and 90 seconds. CALCULATION OD Sample x n OD Standard n = standard concentration NOTE: For more details, see the Instructions for use

    (07/2007) * these kits contain a standard MM-URSL-2

    Working reagent : 1 mL Sample : 10 L

    R1 : 1 mL Sample : 10 L

    R2 : 250 L

  • ELITECH ELITECH ELITECH

    SEPPIM S.A.S.- ZONE INDUSTRIELLE - 61500 SEES - FRANCE

    UREA UV

    Rf. : URUV-0400 9 x 50 mL URUV-0500 6 x 100 mL

    MANUAL METHOD

    PROCEDURE Wavelength : 340 nm Temperature : 37C, 30C Cuvette : 1 cm light path Read against reagent blank.

    Mix and read the variation of optical density (OD) between 30 seconds and 90 seconds. CALCULATION OD Sample x n OD Standard

    n= Standard concentration

    Take the dilution factor into account for the calculation of urea concentration in urine.

    NOTE: For more details, see the Instructions for use

    (07/2005) MM-URUV-1

    BLANK STANDARD SAMPLE Working Reagent 1 mL 1 mL 1 mL Distilled Water 10 L Standard 10 L Sample 10 L

  • ELITECH ELITECH ELITECH

    SEPPIM S.A..S- ZONE INDUSTRIELLE - 61500 SEES - FRANCE

    URIC ACID

    Rf. : ACUR - 0200 12 x 20 mL

    ACUR - 0400 9 x 50 mL

    MANUAL METHOD

    PROCEDURE

    Wavelength : 510 nm (492-550)

    Temperature : 37C

    Cuvette : 1 cm light path

    Read against reagent blank.

    Mix and read the optical density (OD) after a 5 minute incubation.

    The final colour is stable for at least 15 minutes.

    CALCULATION

    OD Sample x n

    OD Standard

    n= Standard concentration

    Take the dilution factor into account for the calculation of uric acid concentration in urine.

    NOTE: For more details, see the Instructions for use

    (09/2009)

    MM-ACUR-2

    BLANK STANDARD SAMPLE

    Working Reagent 1 mL 1 mL 1 mL

    Distilled Water 25 L

    Standard 25 L

    Sample 25 L

  • ELITECH ELITECH ELITECH

    SEPPIM S.A.S.- ZONE INDUSTRIELLE - 61500 SEES - FRANCE

    URIC ACID MONO SL

    Rf. : AUML - 0420 6 x 50 mL AUML - 0500 6 x 100 mL AUML - 0700 4 x 250 mL AUML - 0427 6 x 50 mL* AUML - 0507 6 x 100 mL* AUML - 0707 4 x 250 mL*

    MANUAL METHOD

    PROCEDURE Wavelength : 555 nm Temperature : 37C Cuvette : 1 cm light path Read against reagent blank.

    Mix and read the optical density (OD) after a 5 minute incubation. CALCULATION OD Sample x n OD Standard

    n= Standard concentration

    Take the dilution factor into account for the calculation of uric acid concentration in urine.

    NOTE: For more details, see the Instructions for use

    (07/2007) * these kits contain a standard MM-AUML-2

    BLANK STANDARD SAMPLE Reagent 1 mL 1 mL 1 mL Distilled Water 25 L Standard 25 L Sample 25 L

  • ELITECH ELITECH ELITECH

    SEPPIM S.A.S.- ZONE INDUSTRIELLE - 61500 SEES - FRANCE

    URIC ACID SL

    Rf. : AUSL 0400 2 x 62.5 mL AUSL 0600 5 x 125 mL

    MANUAL METHOD

    PROCEDURE Wavelength : 555 nm Temperature : 37C Cuvette : 1 cm light path Read against reagent blank. One-reagent procedure

    Mix and read the optical density (OD) after a 5 minute incubation. Two-reagent procedure

    Mix, wait 1 minute and add:

    Mix and read the optical density (OD) after a 5 minute incubation. CALCULATION OD Sample x n OD Standard

    n= Standard concentration Take the dilution factor into account for the calculation of uric acid concentration in urine.

    NOTE: For more details, see the Instructions for use

    (03/2010) MM-AUSL-2

    BLANK STANDARD SAMPLE Working Reagent 1 mL 1 mL 1 mL Distilled Water 25 L Standard 25 L Sample 25 L

    BLANK STANDARD SAMPLE R1 1 mL 1 mL 1 mL Distilled Water 25 L Standard 25 L Sample 25 L

    R2 250 L 250 L 250 L

    MAGNESIUM CALMAGITE ................... MM-MAGN-1