immuno-monitoring using the scepter™ 2.0 cell counter and software pro

Upload: emd-millipore-bioscience

Post on 07-Apr-2018

225 views

Category:

Documents


2 download

TRANSCRIPT

  • 8/6/2019 Immuno-monitoring using the Scepter 2.0 Cell Counter and Software Pro

    1/12

    Immuno-monitoring usingthe Scepter 2.0 Cell Counterand Software Pro

    Application Note

    AbstractBiological samples such as primary isolates or cultured

    cells are oten heterogeneous mixtures o cells that

    dier by type and/or unction. Such dierences in

    cellular attributes are most commonly determined by

    multicolor uorescent antibody detection o cell type-

    specifc surace marker(s) using ow cytometry. Notably,

    in addition to variations in protein expression, many

    cell types and physiological states are also uniquely

    distinguishable on the basis o size alone. The ability to

    identiy population subsets on the basis o phenotypic

    dierences and urther determine their relative

    requencies (and concentrations) is critical to many

    aspects o research.

    The Scepter cell counter combines the ease o automated

    instrumentation and the accuracy o the Coulter

    principle o impedance-based particle detection in an

    aordable, handheld ormat. The instrument uses a

    combination o analog and digital hardware or sensing,

    signal processing, data storage, and graphical display. The

    precision-made, consumable polymer sensor has a laser-

    drilled aperture in its cell sensing zone that enables the

    EMD Millipore is a division of Merck KGaA, Darmstadt, Germany.

    Amedeo Cappione, Ph.D.1; Emily Crossley2; Nagaraja Thirumalapura, DVM, Ph.D.3; and Debra Hoover, Ph.D.11EMD Millipore; 2Dept. o Pathology, Dept. Microbiology & Immunology, University o Texas Medical Branch;3Dept. o Pathology, University o Texas Medical Branch

    instrument to use the Coulter principle to discriminate

    cell diameter and volume at submicron and subpicoliter

    resolution, respectively.

    Using a sensor with a 40 m aperture, the Scepter cell

    counter can accurately and precisely count a broad range

    o cell types, including small cells (> 3 m in diameter)

    and peripheral blood mononuclear cells (PBMC)1. This

    article outlines three examples o experiments using

    the Scepter cell counters sensitive size-discriminating

    capability to demonstrate rapid, qualitative assessment

    o individual cell population requencies in complex cell

    mixtures.

    http://www.millipore.com/life_sciences/flx/scepter&open&cid=BIOS-S-EPDF-1043-1106-DShttp://www.millipore.com/life_sciences/flx/scepter&open&cid=BIOS-S-EPDF-1043-1106-DShttp://www.millipore.com/life_sciences/flx/scepter&open&cid=BIOS-S-EPDF-1043-1106-DShttp://www.millipore.com/life_sciences/flx/scepter&open&cid=BIOS-S-EPDF-1043-1106-DShttp://www.millipore.com/life_sciences/flx/scepter&open&cid=BIOS-S-EPDF-1043-1106-DShttp://www.millipore.com/life_sciences/flx/scepter&open&cid=BIOS-S-EPDF-1043-1106-DShttp://www.millipore.com/life_sciences/flx/scepter&open&cid=BIOS-S-EPDF-1043-1106-DShttp://www.millipore.com/?cid=BIOS-S-EPDF-1043-1106-DS
  • 8/6/2019 Immuno-monitoring using the Scepter 2.0 Cell Counter and Software Pro

    2/12

    2

    ExamplE 1:Lymphocyte vs. monocyte subset

    discrimination in reshly isolated PBMCs

    IntroductionThe human immune system is comprised o cell

    subsets, with distinct unctional profles, that fght

    pathogens. The assessment o immune profles o the

    various immune cell subsets, such as lymphocytes andmonocytes, can help identiy molecular signatures

    that may acilitate research, ranging rom translational

    medicine, prognostic advancements, and early decision-

    making tools in vaccine development. The Scepter

    cell counter, when used in combination with Scepter

    Sotware Pro, provides a tool or rapid determination o

    lymphocyte and monocyte concentrations as well as the

    relative requency o these cell types in PBMC isolates.

    Materials and Methods

    Human blood sample prepHuman PBMCs were isolated rom whole heparinized

    blood o healthy donors by Ficoll-Paque density

    centriugation (GE Healthcare). Briey, 9 mL o blood

    was diluted to 25 mL with phosphate-buered saline

    (1X EmbryoMax PBS, EMD Millipore). and layered over

    15 mL o Ficoll. Samples were centriuged at 400 x g or

    30 minutes with no brake, and the resulting PBMC layer

    was recovered. The PBMC ractions were washed twice

    by centriugation using PBS. Ater fnal spin, cell pellets

    were resuspended in PBS or sufcient conductivity

    during Scepter cell counting.

    Scepter cell counting

    The Scepter cell counter was used to count samples

    ollowing the simple on-screen instructions or each step

    o the counting process. Briey, the user attaches a

    40 m sensor, depresses the plunger, submerges the

    sensor into the sample, then releases the plunger

    drawing 50 L o cell suspension through the cell-

    sensing aperture. The Scepter cell counter detects each

    cell passing through the sensors aperture, calculates cell

    concentration, and displays a histogram as a unction o

    cell diameter or volume on its screen.

    Scepter 2.0 sotware was then used to upload test fles

    rom the device and perorm subsequent data analysis

    to determine the concentrations and relative cellrequencies or the lymphocyte and monocyte ractions.

    Guava easyCyte cell counting

    10 L o each PBMC sample was diluted in 190 L PBS.

    Samples were then analyzed on a guava easyCyte HT

    system to determine the concentrations and relative cell

    requencies or the lymphocyte and monocyte ractions.

    Cell viability determination using

    guava ViaCount assay

    10 L o each PBMC sample was mixed with 190 L

    ViaCount assay reagent, incubated or 5 minutes at room

    temperature (RT). Viability data were acquired on a guava

    easyCyte instrument using guava ExpressPro sotware.

    Cell surface staining and subset determination

    For each sample, 100,000 PBMCs were resuspended

    in 100 L PBS+0.1% bovine serum albumin (BSA).

    To distinguish the discrete cell subsets present in

    PBMC samples, they were stained with the ollowing

    combination o uorescently labeled antibodies: CD3-PE

    (T cells), CD19-Alexa Fluor 488 (B cells), CD16/CD56-

    APC (NK cells), and CD14-PECy7 (monocytes) (antibodies

    all rom eBioscience). Singly stained samples and isotype

    controls were included with each staining set to ensure

    proper instrument setup. Samples were incubated

    at RT or 20 minutes, washed twice with PBS, then

    resuspended in 200 L PBS prior to acquisition. Samples

    were analyzed (3000 cells/sample well) on a guava

    easyCyte HT system using guava ExpressPro sotware.

    01

    http://www.millipore.com/search.do?q=embryomax&cid=BIOS-S-EPDF-1043-1106-DShttp://www.millipore.com/search.do?q=embryomax&cid=BIOS-S-EPDF-1043-1106-DShttp://www.millipore.com/search.do?q=embryomax&cid=BIOS-S-EPDF-1043-1106-DShttp://www.millipore.com/search.do?q=embryomax&cid=BIOS-S-EPDF-1043-1106-DShttp://www.millipore.com/search.do?q=embryomax&cid=BIOS-S-EPDF-1043-1106-DS
  • 8/6/2019 Immuno-monitoring using the Scepter 2.0 Cell Counter and Software Pro

    3/12

    3

    ResultsViability assessment of PBMC

    Freshly prepared or rozen PBMC samples consisted

    o live cells, dead cells, and a considerable amount

    o cellular debris. The debris component varied with

    respect to properties inherent to the blood sample as

    well as the methods o storage and PBMC preparation.

    To understand the relative proportions o live cells, deadcells and debris, reshly isolated PBMC were stained with

    ViaCount reagent.

    ViaCount reagent distinguished viable and non-viable

    cells based on dierential permeabilities o two DNA-

    binding dyes. The cell-permeant nuclear dye stained

    all nucleated cells, while the cell-impermeant viability

    dye brightly stained dying cells. Cellular debris was not

    stained by either dye.

    Results rom a representative assay are presented

    in Figure 1A. In this example, more than 95% o the

    cells were viable. The three components were also

    distinguishable on the basis o particle size. The

    histograms in Figure 1B show the distribution o particles

    in the dierent ractions as unction o orward scatter,

    a ow cytometry-based correlate o particle size. While

    there was some overlap between size distributions o

    dead cells and debris, the live cells constituted a distinct

    raction made up o two dierently-sized cell types.

    Quantifying subsets of PBMCs

    (T cells, B cells, NK cells, monocytes)

    PBMC can be subdivided into many distinct cell types

    based on the varying expression o specifc surace

    markers and unctional capabilities (or example, cytokine

    secretion). The abundance and relative distribution o

    these subsets are unctions o developmental state as

    well as overall health. Two main populations o PBMC

    are lymphocytes and monocytes. The lymphocyte subset

    can be urther subdivided into T cells (CD3+), B cells

    (CD19+), and NK cells (CD16/56+). Monocytes can be

    distinguished rom all lymphocytes on the basis o CD14

    expression.

    Figure 1. Representative ow

    cytometry data demonstrating

    size-based discrimination o live

    vs dead cells. (A) live cells (blue),

    dead cells (red), and debris (green)

    ractions were defned using

    ViaCount reagent. (B) The our

    histograms show the relativedistribution o each raction as a

    unction o particle size (orward

    scatter).

    Figure 2. Representative data

    rom ow cytometry analysis o

    multicolor staining perormed on

    PBMC ractions. (A) Live cells (red)

    are distinguished rom debris/

    dead cells (green) on the basis o

    orward vs. side scatter. (B) Live

    cells rom (A) are ractionated in

    the 2 plots based on variations in

    expression o CD14, CD3, CD16/56,

    and CD19. (C) The our histograms

    show the localization o each

    cell raction with regards to the

    two peaks defned by orward

    scatter. As shown, the three major

    lymphocyte subsets are ound

    predominantly in the smaller-

    cell peak (blue) while the CD14+

    monocytes are restricted to the

    large cell raction (pink).

    10e4

    10e3

    10e2

    Red

    Fluorescence

    10e1

    10e0

    10e0 10e1

    Yellow Fluorescence

    10e2 10e3

    M1

    M2

    M3

    10e4

    M3

    M1 M1

    M1 M1

    100

    otal EventsA B Live Cells

    Dead Cells Debris

    75

    50

    25

    0

    0 2500 5000

    Forward Scatter

    Count

    7 50 0 1 00 00

    100

    75

    50

    25

    0

    0 2500 5000

    Forward Scatter

    Count

    7 50 0 1 00 00

    100

    75

    50

    25

    0

    0 2500 5000

    Forward Scatter

    Count

    7 50 0 1 00 00

    50

    38

    25

    13

    0

    0 2500 5000

    Forward Scatter

    Count

    7 50 0 1 00 00

    Live Cells

    Debris/

    RBCM1

    10000

    7500

    5000

    Side

    Scatter

    2500

    0

    0 2500

    Forward Scatter

    5000 7500 10000

    M1 Monocytes

    T cellsM2

    M1

    M2

    NK cells

    B cells

    A B C

    CD3+10e4

    10e3

    10e2

    10e1

    10e0

    75

    100

    50

    25

    0

    0

    1 0e 0 1 0e 1 1 0e 2 1 0e 3 1 0e 4

    2500 5000

    Forward Scatter

    Count

    CD3-PE (T-cells)

    CD14-PECy7(Monos)

    10e4

    10e3

    10e2

    10e1

    10e0

    1 0e 0 1 0e 1 1 0e 2 1 0e 3 1 0e 4

    CD19-A488 (B-cells)

    CD56-APC(NKCells)

    7500 10000

    M1

    M2

    M3

    CD14+

    75

    100

    50

    25

    0

    0 25 00 5 00 0

    Forward Scatter

    Count

    7500 10000

    M1

    M2

    M3

    M1

    M2

    M3

    CD19+

    75

    100

    50

    25

    0

    0 2 50 0 5 000

    Forward Scatter

    Count

    7500 10000

    D16/56+

    75

    100

    50

    25

    0

    0 25 00 5 00 0

    Forward Scatter

    Count

    7500 10000

    M1

    M2

    M3

    M1

    M2

    M3

  • 8/6/2019 Immuno-monitoring using the Scepter 2.0 Cell Counter and Software Pro

    4/12

    4

    To investigate the distribution o these our subsets with

    regards to size, PBMCs were stained with uorescent

    antibodies specifc or each surace marker (Figure 2). The

    dot plots confrmed defnition o each subset by staining

    with each cell-type specifc marker. Figure 2C indicated

    that B, T, and NK cells were generally smaller in size (low

    orward scatter) while CD14+ monocytes were larger in

    size (higher orward scatter).

    To examine the ability o the 40 m Scepter sensor

    to discriminate lymphocytes rom monocytes, PBMC

    ractions were isolated rom resh human blood by

    Ficoll gradient centriugation. Diluted samples were

    analyzed using both a guava easyCyte ow cytometer

    and the Scepter cell counter. Representative histogram

    plots are presented in Figure 3. In each case, three main

    peaks were distinguishable by each analysis method,

    with greater peak resolution in the ow cytometry

    data, particularly in the separation o lymphocytes rom

    debris. The peaks correspond to lymphocytes (small cells),

    monocytes (large cells), and a debris/dead cell raction.

    Across the nine PBMC samples analyzed (Table 1),

    the average mean cell diameters were 7.230.30 m

    and 10.020.20 m or lymphocytes and monocytes,

    respectively. Resulting values were consistent with

    previously reported size ranges2

    . In addition, cellrequencies were determined by three methods:

    Scepter diameter plot, guava easyCyte orward

    scatter, and antibody staining. The Scepter values

    slightly underestimated the lymphocyte raction while

    overestimating the monocyte subset. This is likely the

    result o the Scepter cell counters comparatively lower

    resolution as well as subjectivity and user bias in the

    placement o gates defning each subset. Overall, there

    was good agreement between the dierent analytical

    techniques with values varying by

  • 8/6/2019 Immuno-monitoring using the Scepter 2.0 Cell Counter and Software Pro

    5/12

    5

    ExAMPE 2:

    Human T-cell activation

    IntroductionIn vivo, T lymphocytes are activated and induced

    to prolierate upon binding o the T cell receptor to

    antigen-presenting cells. In response to stimulation,

    T cells undergo physical, biological, and phenotypicchanges, including increased cell size, secretion o

    cytokines, and up-regulation o CD25 surace expression,

    ultimately culminating in the production o specifcally-

    tuned immune responses. Elevated CD25 expression

    levels are a late-stage indicator o T-cell activation4.

    Assays measuring changes in T- and B-cell activation

    are commonly used to identiy patterns o immune

    response in clinical diagnostics as well as therapeutic

    development.

    The in vivo immune response by T cells can be mimickedex vivo by binding T cells to co-immobilized anti-CD28

    and anti-CD3 monoclonal antibodies3. PBMCs can

    also be activated through exposure to more generic

    inducers o cell prolieration, such as the plant lectin

    phytohemagglutinin (PHA) or pokeweed mitogen. Ex

    vivo activation enables detailed studies o the molecular

    mechanisms regulating T-cell activation and response.

    Materials and MethodsCell culture/activation

    Human PBMC ractions were isolated as previously

    described. Prior to culture, initial cell concentrations were

    determined using the Scepter cell counter. All culturing

    experiments were perormed in RPMI 1640 supplemented

    with 10% etal bovine serum (FBS) (R10) in the absence

    o antibiotics. PBMC (400,000 cells/mL, 24-well plates)

    were stimulated or two days in the presence o soluble

    anti-CD28 mAb (clone 28.2, BD Pharmingen, 2 g/mL)

    on plates pre-coated with anti-CD3 mAb (clone HIT3a;

    BD Pharmingen, 10 g/mL). Mitogenic stimulation was

    carried out in the presence o 2 g/mL PHA (Sigma).

    Unstimulated control cultures were also analyzed.

    Scepter cell countingSample acquisition and data analysis was perormed

    as previously described. Briey, ollowing stimulation, a

    small aliquot o each culture was harvested, diluted in

    PBS, then analyzed on the Scepter cell counter. Test fles

    were uploaded and analyzed using Scepter Sotware Pro

    to determine the degree o cell activation (% activated

    cells) as well as concentrations or both unstimulated

    and activated ractions.

    CD25 staining for activation

    Following two-day stimulation, cultures were harvested,

    washed twice with PBS, and then counted using the

    Scepter cell counter. For each sample, 100,000 cells were

    resuspended in 100 L PBS+0.1% BSA. To distinguish

    the activated T-cell raction, samples were stained with

    anti-CD3-PE (T cells) and anti-CD25-APCeFluor780

    (antibodies rom eBioscience). Singly stained samples and

    Isotype controls were included with each staining set to

    ensure proper instrument setup. Samples were incubated

    at RT or 20 minutes, washed twice with PBS, then

    resuspended in 200 L PBS prior to acquisition. Samples

    were analyzed (3000 cells/sample well) on a guava

    easyCyte HT system using guava ExpressPro sotware.

    02

  • 8/6/2019 Immuno-monitoring using the Scepter 2.0 Cell Counter and Software Pro

    6/12

    6

    The dot plots in Figure 4B showed three main

    populations o cells: resting T cell (CD3+/CD25-),

    activated T cells (CD3+/CD25+), and non-T cells (CD3-/

    CD25-). By comparison, cultures stimulated with CD3/

    CD28 Ab demonstrated signifcantly greater levels o

    T-cell activation than those exposed to PHA. Control

    cultures showed very low requencies o activated cells.

    Under all conditions, CD25 expression also correlatedwith an increase in overall cell size (Figure 4C).

    ResultsTo address the potential use o the Scepter cell counter

    or rapid qualitative monitoring o immune cell

    activation as a unction o the cell size shit, reshly

    isolated PBMCs were stimulated in culture using two

    dierent mechanisms: (1) CD3/CD28 antibody-mediated

    co-stimulation o the T cell receptor (TCR) and (2)

    mitogenic stimulation by PHA.

    Figure 4. Expression o the cell activation marker CD25 correlates with an increase in cell size. (A) An elliptical gate was used to identiy live cells. (B) Dot plots depict the ractionation o live

    cells on the basis o CD3 and CD25 expression. (C) The orward scatter histogram plots demonstrate that expression o CD25 is restricted to the larger-cell raction or all culture conditions.

    CD25 eFluor780

    CD3-PE

    Side

    Scatter

    Side

    Scatter

    Forward Scatter

    Side

    Scatter

    Forward Scatter

    CD3-PE

    CD3-PE

    CD25 eFluor780

    CD25 eFluor780

    Forward Scatter

    Side

    Scatter

    Forward Scatter

    Side

    Scatter

    Forward Scatter

    Side

    Scatter

    Forward Scatter

    Side

    Scatter

    Forward Scatter

    Side

    Scatter

    Forward Scatter

    Side

    Scatter

    Forward Scatter

    A B C

  • 8/6/2019 Immuno-monitoring using the Scepter 2.0 Cell Counter and Software Pro

    7/12

    7

    the live, unstimulated raction. This overlap was clearly

    demonstrated in the untreated sample, in which two

    peaks were seen in the ow cytometry histogram plot o

    total cells (Marker R5). By comparison, the live cell-only

    histogram showed a single peak within R5. The resulting

    data clearly demonstrated the subset o activated T cells

    but could not be used discriminate the live, unstimulated

    T cell population.

    In Figure 5, the Scepter cell counters ability to detect cell

    activation was compared to that o the guava easyCyte

    ow cytometry platorm. While the Scepter cell counter

    was able to detect the presence o the larger, activated

    cell raction in both stimulated cultures, the device

    was unable to simultaneously discriminate the smaller

    unstimulated population in these samples. This was most

    likely due to the presence o a large number o deadcells (red cells in Figure 5A) that overlapped in size with

    Figure 5. Representative data demonstrating size-based discrimination o resting vs. activated cells in control and stimulated PBMC cultures. (A) Live cell (blue), dead cell (red), and de bris

    (green) ractions were defned using ViaCount reagent. (B) The two histograms in each row show the relative distribution o each raction as a unction o particle size (orward scatter). Plots

    are based on total events and live cells, respectively. (C) Histogram data or each sample ollowing analysis using the Scepter cell counter with 40 m sensor.

    Side

    Scatter

    Side

    Scatter

    Forward Scatter

    Side

    Scatter

    Forward Scatter

    Forward Scatter

    Count

    Forward Scatter

    Count

    Forward Scatter

    Count

    Diameter (m)

    Count

    Diameter (m)

    Count

    Diameter (m)

    Count

    Forward Scatter

    Count

    Forward Scatter

    Count

    Forward Scatter

    Count

    Forward Scatter

    A B C

  • 8/6/2019 Immuno-monitoring using the Scepter 2.0 Cell Counter and Software Pro

    8/12

    8

    ExAMPE 3:Splenic cell shift in murine

    model of Ehrlichiosis

    IntroductionHuman monocytotropic ehrlichiosis (HME) is an

    emerging tick-borne disease caused by the obligately

    intracellular pathogen Ehrlichia chaffeensisthat resides

    in mononuclear phagocytes. HME initially maniestsas nonspecifc u-like symptoms but can progress to

    lie-threatening toxic shock-like syndrome with anemia,

    thrombocytopenia, and multi-organ ailure. The presence

    o extensive inammation in the absence o bacterial

    burden suggests that mortality is the consequence o

    unregulated immunopathology5,6. In this study, a murine

    model o HME based on inection o C57BL/6 mice

    with Ehrlichia muriswas used to investigate aspects

    o the immune response, in particular, changes to the

    population dynamics o splenocytes.

    Figure 6. Cell surace staining defnes cell subsets within the mouse

    spleen. Forward vs. side scatter plots reveal the presence o two main

    splenocyte populations. (A) The raction containing larger cells is

    comprised primarily o leukocytes (B, NK, and T cells), while the raction

    o small cells (B) is predominantly er ythrocytes (Ter119+).

    Materials and methodsInfection of mice with Ehrlichia muris

    Six- to eight-week-old C57BL/6 mice were inected

    with Ehrlichia muris (~1 x 104 bacterial genomes) by

    the intraperitoneal route. Mice were sacrifced on day

    30 post-inection and spleens were harvested. Single-

    cell suspensions o splenocytes were prepared using

    the GentleMACS Tissue Dissociator ollowing themanuacturers instructions (Miltenyi Biotec Inc., CA).

    Determination of splenocyte cell distribution using

    the Scepter cell counter

    Cells were diluted to appropriate concentrations

    (~1 2 x 105 cells/mL) and loaded into the Scepter cell

    counter ftted with the 40 M sensor. The gates were

    set to exclude red blood cells using a splenocyte sample

    rom an uninected, nave mouse. Splenocytes rom three

    uninected mice and three mice inected with Ehrlichia

    muriswere used. Data were analyzed using Scepter

    Sotware Pro.

    Determination of splenocyte cell distribution

    by ow cytometry

    Splenocytes were incubated with a uorescent viability

    dye (Near IR LIVE/DEAD Fixable Dead Cell Stain Kit,

    Invitrogen, CA) or 10 minutes in PBS at RT and protected

    rom light. Cells were washed and blocked with Fc Block

    (BD Biosciences, CA) and purifed rabbit, rat and mouse

    IgGs (Jackson ImmunoResearch, PA) in ow cytometry

    buer (PBS + 2% FBS + 0.09% sodium azide). Cells were

    then incubated with uorescently labeled antibodies

    to the surace markers B220 (B cells) and Ter119

    (Erythrocytes) (BD Biosciences, CA and eBioscience, CA,

    respectively) and analyzed by ow cytometry.

    ResultsSingle-cell suspensions o mouse splenocytes were

    analyzed by multicolor ow cytometry (Figure 6). Size-

    based discrimination using orward scatter confrmed the

    presence o two distinct cell populations in all samples.Staining with antibodies specifc or B-cells (B220) and

    erythrocytes (Ter-119) demonstrated that erythrocytes

    were restricted to the subset containing smaller cells,

    while the raction o large cells was composed primarily

    o lymphocytes.

    03

    A

    B

    SSC

    FSC

    SSC

    FSC

    B200-

    PerCP-

    Cy5.5

    Ter119-APC

    B200-

    PerCP-

    Cy5.5

    Ter119-APC

    B cells

    T/NK cells

    1

    2

    70

    27

    Erythroid

    194

    32

  • 8/6/2019 Immuno-monitoring using the Scepter 2.0 Cell Counter and Software Pro

    9/12

    9

    Flow cytometry-derived histograms o total cells showed

    the presence o three distinct populations corresponding

    to debris, lymphocytes (large cells) and erythrocytes

    (small cells). Comparative analysis o splenic cell isolates

    rom Ehrlichia-inected and control mice revealed an

    overall shit in the relative cell distribution. Specifcally,

    inected samples showed an increase in the percentage

    o erythrocytes when compared to the lymphocyte

    subset (Figure 7B and Table 2). In contrast, the Scepter

    histogram was able to detect only two peaks; the third

    peak corresponding to the debris raction had particle

    sizes below the lower limit o detection. In addition, the

    smaller erythrocytes (red blood cells) were at the limits

    o detection or the Scepter cell counter and were not

    Table 2. Total lymphocyte and erythrocyte subset requencies rom six individual splenocyte isolates (three control, three inected). Aliquots rom each

    sample were analyzed using a ow cytometer and Scepter cell counter.

    Count

    Dead/debris

    Erythroid

    Lymphocyte

    Count

    Diameter (m)

    Count

    A B C

    SSC

    FSC FSC

    FSC

    SSC

    FSC

    Diameter (m)

    Figure 7. Size-based ractionation o mouse splenocytes. Untreated andEhrlichia-inected samples were analyzed using a ow cytometer (A, B) and

    using the Scepter cell counter (C). Three distinct peaks corresponding to lymphocytes, ery throcytes, and debris can be visualized using ow cytometry.

    For the Scepter cell counter, the particle size o the debris peak is below the level o detection or the 40 m sensor.

    Scepter Flow Cytometry

    Test Sample Erythrocyte ymphocyte Erythrocyte ymphocyte

    Control-1 27 73 16 84

    Control-2 29 71 21 79

    Control-3 22 78 14 86

    Inected - 1 40 60 42 58

    Inected - 2 30 70 29 71

    Inected - 3 33 67 38 62

  • 8/6/2019 Immuno-monitoring using the Scepter 2.0 Cell Counter and Software Pro

    10/12

    10

    counted, skewing the erythrocyte counts. However, the

    Scepter cell counters quantitation o splenocyte counts

    and percentages were in very good agreement with

    the values determined by ow cytometry. An overall

    assessment o cell concentration values indicated the

    shit was due to an expansion o the erythrocyte subset

    in the spleen while lymphocyte values remained constant(Table 3). These fndings were consistent with previously

    published results7.

    Overall ConclusionsEvaluation o immune responsiveness to viral, bacterial,

    cellular and environmental exposure is an important

    tool in assessment o pathogenicity and toxicity .

    The availability o simplifed methods or identiying

    activation status o cells and or measuring dierences

    in the relative requencies (and concentrations) o more

    than one cell type in samples or co-cultures is very

    useul or accelerating research in these felds.

    We have presented data indicating that both the guava

    easyCyte benchtop ow cytometry platorm as well as

    the Scepter cell counter can determine such dierences

    in resh primary isolates and cultured samples. While

    the ow cytometry system is the more sophisticated

    and quantitative tool, the Scepter is a rapid tool that can

    be used at the hood to provide very good insight into

    cellular responses in concert with existing methods.

    The unctionality o Scepter counting is a result o the

    precision-engineered sensor and the unique handheld

    instrumentation based upon the Coulter principle. Theability o the Scepter device to ensure reproducible cell

    counts improves data quality during experimental setup

    and downstream cell-based analyses.

    Table 3. Total lymphocyte and erythrocyte cell concentration or all six samples.

    References

    1. Smith, J et al. The New Scepter 2.0 Cell Counter Enables the

    Analysis o a Wider Range o Cell Sizes and Types With High

    Precision. EMD Millipore Cellutions 2011 Vol. 1: p 19-22.

    2. Daniels, V. G., Wheater, P. R., & Burkitt, H. G. (1979).

    Functional Histology: A Text and Colour Atlas. Edinburgh:

    Churchill Livingstone. ISBN 0-443-01657-7.3. Prager, E. et al. (2001) Induction o Hyporesponsiveness and

    Impaired T Lymphocyte Activation by the CD31 Receptor:

    Ligand Pathway in T-Cells. J. Immunol. 166: 2364-2371.

    4. Levine, B. L., et al. (1997). Eects o CD28 Costimulation on

    Long-term Prolieration o CD4+ T-cells in the Absence o

    Exogenous Feeder Cells. J. Immunol. 159:5921 5930.

    5. Olano, J. P., G. Wen, H. M. Feng, J. W. McBride, and D. H.

    Walker. 2004. Histologic, Serologic, and Molecular Analysis

    o Persistent Ehrlichiosis in a Murine Model. Am. J. Pathol.

    165:9971006.

    6. Paddock, C. D., and J. E. Childs. 2003. Ehrlichia chaeensis:

    a Prototypical Emerging Pathogen. Clin. Microbiol. Rev.

    16:3764.

    7. MacNamara, K.C., Racine, R. Chatterjee, M., Borjesson, D, and

    Winslow, G.M. (2009) Diminished Hematopoietic Activity

    Associated with Alterations in Innate and Adaptive Immunity

    in a Mouse Model o Human Monocytic Ehrlichiosis. Inect.

    And Immunity. 77:4061-69.

    Cell Concentration (105)

    Test Sample Erythrocyte ymphocyte

    Control-1 1.86 4.65

    Control-2 2.61 5.54

    Control-3 2.17 5.78

    Inected 1 4.3 5.22

    Inected 2 2.82 4.93

    Inected - 3 3.5 5.84

  • 8/6/2019 Immuno-monitoring using the Scepter 2.0 Cell Counter and Software Pro

    11/12

    11

    Description Quantity Catalog No.

    Scepter 2.0 Handheld Automated Cell Counter

    with 40 m Scepter Sensors (50 Pack) 1 PHCC20040

    with 60 m Scepter Sensors (50 Pack) 1 PHCC20060

    Includes:

    Scepter Cell Counter 1

    Downloadable Scepter Sotware 1

    O-Rings 2

    Scepter Test Beads 1 PHCCBEADS

    Scepter USB Cable 1 PHCCCABLE

    Scepter Sensors, 60 m 50 PHCC60050

    500 PHCC60500

    Scepter Sensors, 40 m 50 PHCC40050

    500 PHCC40500

    Universal Power Adapter 1 PHCCP0WER

    Scepter O-Ring Kit, includes 2 O-rings and 1 flter cover 1 PHCC0CLIP

    Description Catalog No.

    Guava ViaCount Reagent 4000-0041

    FlowCellect Human Lymphocyte ZAP-70 Characterization Kit FCIM025122

    FlowCellect Mouse FoxP3 Treg Identifcation Kit FCIM025126

    guava Cell Toxicity Reagents Kit 4500-0230

    FlowCellect Human FOXP3 Treg Characterization Kit FCIM025118

    FlowCellect Mouse Th1 Intracellular Cytokine Kit FCIM025123

    FlowCellect Mouse Th2 Intracellular Cytokine Kit FCIM025124

    FlowCellect Mouse Th17 Intracellular Cytokine Kit FCIM025125

    FlowCellect Mouse Th1/Th2 Intracellular Cytokine Kit FCIM025137

    FlowCellect Mouse Th1/TH17 Intracellular Cytokine Kit FCIM025138

    Ordering Information

    Scepter 2.0 Cell Counter

    Description Quantity Catalog No.

    High-Throughput Sampling Instruments

    guava easyCyte 8HT Base System 1 0500-4008

    guava easyCyte 6HT/2L Base System 1 0500-4007

    guava easyCyte 5HT Base System 1 0500-4005

    PCA-96 Base System 1 0100-8710

    Single Sampling Instruments

    easyCyte 5 Base System 1 0500-5005

    easyCyte 6-2L Base System 1 0500-5007

    easyCyte 8 Base System 1 0500-5008

    Guava easyCyte Benchtop Flow Cytometers

    FlowCellect Flow Cytometry Assay Kits and Milli-MarkFluorescent Conjugated Antibodies for Immunology

    Discover the power o ow cytometry or multiparametric analysis o the immune system:

    visit www.millipore.com/CellularAnalysisor complete listings o our ully validated conjugates

    and assay kits optimized or guava easyCyte systems.

    http://www.millipore.com/catalogue/item/PHCC20040&cid=BIOS-S-EPDF-1043-1106-DShttp://www.millipore.com/catalogue/item/PHCC20060&cid=BIOS-S-EPDF-1043-1106-DShttp://www.millipore.com/catalogue/item/PHCCBEADS&cid=BIOS-S-EPDF-1043-1106-DShttp://www.millipore.com/catalogue/item/PHCCCABLE&cid=BIOS-S-EPDF-1043-1106-DShttp://www.millipore.com/catalogue/item/PHCC60050&cid=BIOS-S-EPDF-1043-1106-DShttp://www.millipore.com/catalogue/item/PHCC60500&cid=BIOS-S-EPDF-1043-1106-DShttp://www.millipore.com/catalogue/item/PHCC40050&cid=BIOS-S-EPDF-1043-1106-DShttp://www.millipore.com/catalogue/item/PHCC40500&cid=BIOS-S-EPDF-1043-1106-DShttp://www.millipore.com/catalogue/item/PHCCP0WER&cid=BIOS-S-EPDF-1043-1106-DShttp://www.millipore.com/catalogue/item/PHCC0CLIP&cid=BIOS-S-EPDF-1043-1106-DShttp://www.millipore.com/catalogue/item/4000-0041&cid=BIOS-S-EPDF-1043-1106-DShttp://www.millipore.com/catalogue/item/FCIM025122&cid=BIOS-S-EPDF-1043-1106-DShttp://www.millipore.com/catalogue/item/FCIM025126&cid=BIOS-S-EPDF-1043-1106-DShttp://www.millipore.com/catalogue/item/4500-0230&cid=BIOS-S-EPDF-1043-1106-DShttp://www.millipore.com/catalogue/item/FCIM025118&cid=BIOS-S-EPDF-1043-1106-DShttp://www.millipore.com/catalogue/item/FCIM025123&cid=BIOS-S-EPDF-1043-1106-DShttp://www.millipore.com/catalogue/item/FCIM025124&cid=BIOS-S-EPDF-1043-1106-DShttp://www.millipore.com/catalogue/item/FCIM025125&cid=BIOS-S-EPDF-1043-1106-DShttp://www.millipore.com/catalogue/item/FCIM025137&cid=BIOS-S-EPDF-1043-1106-DShttp://www.millipore.com/catalogue/item/FCIM025138&cid=BIOS-S-EPDF-1043-1106-DShttp://www.millipore.com/catalogue/item/0500-4008&cid=BIOS-S-EPDF-1043-1106-DShttp://www.millipore.com/catalogue/item/0500-4007&cid=BIOS-S-EPDF-1043-1106-DShttp://www.millipore.com/catalogue/item/0500-4005&cid=BIOS-S-EPDF-1043-1106-DShttp://www.millipore.com/catalogue/item/0100-8710&cid=BIOS-S-EPDF-1043-1106-DShttp://www.millipore.com/catalogue/item/0500-5005&cid=BIOS-S-EPDF-1043-1106-DShttp://www.millipore.com/catalogue/item/0500-5007&cid=BIOS-S-EPDF-1043-1106-DShttp://www.millipore.com/catalogue/item/0500-5008&cid=BIOS-S-EPDF-1043-1106-DShttp://www.millipore.com/cellbiology/flx1/insightful_cell_analysis&open&cid=BIOS-S-EPDF-1043-1106-DShttp://www.millipore.com/cellbiology/flx1/insightful_cell_analysis&open&cid=BIOS-S-EPDF-1043-1106-DShttp://www.millipore.com/catalogue/item/FCIM025138&cid=BIOS-S-EPDF-1043-1106-DShttp://www.millipore.com/catalogue/item/FCIM025137&cid=BIOS-S-EPDF-1043-1106-DShttp://www.millipore.com/catalogue/item/FCIM025125&cid=BIOS-S-EPDF-1043-1106-DShttp://www.millipore.com/catalogue/item/FCIM025124&cid=BIOS-S-EPDF-1043-1106-DShttp://www.millipore.com/catalogue/item/FCIM025123&cid=BIOS-S-EPDF-1043-1106-DShttp://www.millipore.com/catalogue/item/FCIM025118&cid=BIOS-S-EPDF-1043-1106-DShttp://www.millipore.com/catalogue/item/4500-0230&cid=BIOS-S-EPDF-1043-1106-DShttp://www.millipore.com/catalogue/item/FCIM025126&cid=BIOS-S-EPDF-1043-1106-DShttp://www.millipore.com/catalogue/item/FCIM025122&cid=BIOS-S-EPDF-1043-1106-DShttp://www.millipore.com/catalogue/item/4000-0041&cid=BIOS-S-EPDF-1043-1106-DShttp://www.millipore.com/cellbiology/flx1/insightful_cell_analysis&open&cid=BIOS-S-EPDF-1043-1106-DShttp://www.millipore.com/catalogue/item/0500-5008&cid=BIOS-S-EPDF-1043-1106-DShttp://www.millipore.com/catalogue/item/0500-5007&cid=BIOS-S-EPDF-1043-1106-DShttp://www.millipore.com/catalogue/item/0500-5005&cid=BIOS-S-EPDF-1043-1106-DShttp://www.millipore.com/catalogue/item/0100-8710&cid=BIOS-S-EPDF-1043-1106-DShttp://www.millipore.com/catalogue/item/0500-4005&cid=BIOS-S-EPDF-1043-1106-DShttp://www.millipore.com/catalogue/item/0500-4007&cid=BIOS-S-EPDF-1043-1106-DShttp://www.millipore.com/catalogue/item/0500-4008&cid=BIOS-S-EPDF-1043-1106-DShttp://www.millipore.com/catalogue/item/PHCC0CLIP&cid=BIOS-S-EPDF-1043-1106-DShttp://www.millipore.com/catalogue/item/PHCCP0WER&cid=BIOS-S-EPDF-1043-1106-DShttp://www.millipore.com/catalogue/item/PHCC40500&cid=BIOS-S-EPDF-1043-1106-DShttp://www.millipore.com/catalogue/item/PHCC40050&cid=BIOS-S-EPDF-1043-1106-DShttp://www.millipore.com/catalogue/item/PHCC60500&cid=BIOS-S-EPDF-1043-1106-DShttp://www.millipore.com/catalogue/item/PHCC60050&cid=BIOS-S-EPDF-1043-1106-DShttp://www.millipore.com/catalogue/item/PHCCCABLE&cid=BIOS-S-EPDF-1043-1106-DShttp://www.millipore.com/catalogue/item/PHCCBEADS&cid=BIOS-S-EPDF-1043-1106-DShttp://www.millipore.com/catalogue/item/PHCC20060&cid=BIOS-S-EPDF-1043-1106-DShttp://www.millipore.com/catalogue/item/PHCC20040&cid=BIOS-S-EPDF-1043-1106-DS
  • 8/6/2019 Immuno-monitoring using the Scepter 2.0 Cell Counter and Software Pro

    12/12

    EMD Millipore and the M logo are trademarks o Merck KGaA, Darmstadt, Germany. Guava, ViaCount andEmbryoMax are registered trademarks and Scepter, FlowCellect, guava easyCyte, and Milli-Mark are trademarkso Millipore Corporation. GentleMACS is a trademark o Miltenyi Biotec, Inc.Alexa Fluor is a registered trademark o Lie Technologies, Inc. Fc Block is a trademark o BD Biosciences.Lit No. AN3322EN00 BS-GEN-11-04565 Printed in the USA. 05/2011

    2011 Millipore Corporation. All rights reserved.

    www.millipore.com/CellularAnalysis

    To Place an Order or ReceiveTechnical Assistance

    In the U.S. and Canada, call toll-ree 1-800-645-5476

    For other countries across Europe and the world,

    please visit www.millipore.com/oices

    For Technical Service, please visit

    www.millipore.com/ofces

    Get Connected!

    Join EMD Millipore Bioscience on your avorite

    social media outlet or the latest updates, news,

    products, innovations, and contests!

    facebook.com/EMDMilliporeBioscience

    twitter.com/EMDMilliporeBio

    http://www.millipore.com/cellbiology/flx1/insightful_cell_analysis&open&cid=BIOS-S-EPDF-1043-1106-DShttp://www.millipore.com/offices/cp3/officeshome&open&cid=BIOS-S-EPDF-1043-1106-DShttp://www.millipore.com/offices/cp3/officeshome&open&cid=BIOS-S-EPDF-1043-1106-DShttp://www.millipore.com/offices/cp3/officeshome&open&cid=BIOS-S-EPDF-1043-1106-DShttp://www.millipore.com/techservices/techservices.do?open&cid=BIOS-S-EPDF-1043-1106-DShttp://www.millipore.com/techservices/techservices.do?open&cid=BIOS-S-EPDF-1043-1106-DShttp://www.facebook.com/EMDMilliporeBiosciencehttps://twitter.com/EMDMilliporeBiohttp://www.millipore.com/techservices/techservices.do?open&cid=BIOS-S-EPDF-1043-1106-DShttp://www.millipore.com/offices/cp3/officeshome&open&cid=BIOS-S-EPDF-1043-1106-DShttp://www.millipore.com/?cid=BIOS-S-EPDF-1043-1106-DShttps://twitter.com/EMDMilliporeBiohttp://www.facebook.com/EMDMilliporeBiosciencehttp://www.millipore.com/cellbiology/flx1/insightful_cell_analysis&open&cid=BIOS-S-EPDF-1043-1106-DS