immediate loading implant systemtechnical reasons hc2 2.9 mmd is not available with expanded apical...
TRANSCRIPT
TP
GH
C2 • Secure anit-rotation through high-precision internal hexagon
• Apical expanded bone thread with No-Itis® surface
• Excellent stability in all bone qualities: double condensation
• Universal application for fixed and removable prosthodontics
• Surface blasted, etched and anodised
• Abutment alignment and 100% tightness by the taper
IMMEDIATE LOADING IMPLANT SYSTEM TPG®
TWO PART IMPLANT SYSTEM HC2
2
HC2 IMPLANTS AND PROSTHETIC COMPONENTS
1. Ask your laboratory to prepare a drill template with the determined bore-
holes for the pilot drills. To be on the safe side, you can ask the laboratory
to insert guide sleeves (REF BFH) into the boreholes, which specify the
exact drill direction. Please use a 2.0 / 2.2 mm Ø drill for the pilot drilling.
2. For the following drill sequences you can use drill stops, which
can be attached and tightened to the drill according to the length of drill-
ing channel. Mucous membrane thickness and template height are taken
into account as needed. Thanks to the extremely high cutting efficiency of
our drills, no ascending drilling sequences will usually be required. Recom-
mended RPM: 2000-5000. Apply sufficient cooling and allow the cooling
to reach the working blades of the drills.
Contra-angle handpiece 2.0 /mm Ø
Drillstop
Holes with guide sleeve
Template
Muscosa
Jaw bone
(Example)
PRELIMINIARY WORK FOR TEMPLATE APPLICATION
DFN 2.9 13 DFN 2.9 15
DS 2
(Pilot drill)
(DFN 3.4 ) / DFN 3.7 DFN 4.2-4.5 (C-Drill 4.2 - 4.5)
(Corticalis drill)
DS 2
(Pilot drill)
DFN 3.0 (DFN 3.4) (C-Drill 3.7)
(Corticalis drill)
1.1 Drill for contra-angle handpiece
Typical sequence of the drills:
Step drill
Bone
Bone
Boneor
The use of DHC is done without previous pilot drill.
Max. torque during insertion of HC 2.9: 35 Ncm.
DS 2
(Pilot drill)
DFN 3.7 DFN 5.5 (C-Drill 5.5)
(Corticalis drill)
Bone
DS 2
(Pilot drill)
DS 2.8
(Pilot drill)
Bone
(hard bone: DFN/DFLN 3.4)
Max. torque during insertion of HC 3.3: 40 Ncm.
General note - HC2 implants are used as compression screws. In order to acchieve a good bone condensation and implant stability, the drilling should be carried out thin-
ner than the core diameter of the implant. The minimal diameter of the drill depends on the bone density. It is therefore not possible to advise drill-sequences which fit all
bone-qualities. Typically in the soft maxillary bone only small drill-diameters are used (e.g. the usage of DOS1 only, for HC2 implants with 3.3 - 5.5 mm diameter), whereas in
the highly mineralized lower jaw a specific drill sequence with respect to the mineralisation of the bone is necessary. For insertion under pressure use the handgrip. Due to
technical reasons HC2 2.9 mmd is not available with expanded apical thread.
HC2 2.9
(head diameter 3.2 mm)
HC2 3.7
HC2 3.3
HC2 5.5 mm
HC2 5.5 mm, 6+2
HC2 4.5 mm
HC2 4.5 mm 6+2
3
1. SURGERY
1.2 Implant packaging
1. Open the lid. The implant is connected
to the lid with a break joint.
2. Remove the implant without touching
the inner walls of the tube.
Original packaging Open packaging using the flap
Remove the label and stick
it intoyour patient’s file.
1.7
Result: a correctly inserted
implant
HC2 implant
Jaw
1.4
After you have attached the insertion
tool, firmly hold the lid in your hand and break
the implant off the top along the break joint.
1.3
Attach the insertion tool to the implant by hol-
ding the top, to which the implant is secured,
with your other hand.
Alternative: Firmly attach the assembled con-
tra-angle handpiece instrument IT 2.5 M to the
implant. For ratchets ITL 2.5 can be used as well.
1.5
Using the ratchet screw the implant clockwise
into the cavity.
The enossal part of the implant must be com-
pletely covered by the bone.
After insertion the implant can be turned by a
1/4 rotation backwards in order to relieve the
bone and allow blood access to the implant site.
TW 2
Jaw
1.6
Disengage the insertion tool
from the implant.
Jaw
TW2
IT 1 or IT 2
Make sure that the
hexagon is in the correct
position
HC2 Implant
Break joint
Lid with
implant mount
Break jointjoint
Lid
1.8
Close the implant with the
suitable cover screw
HT 1.25
CSTI
HC2 implant
Jaw
1.9
After healing period:
Remove cover screw
HT 1.25
CSTI
HC2 implant
Jaw
4
Impression tray
Impression post HLT
HC2 implant
Jaw
Hex-instrument HT 1.25
Thightening of the
impression post HLT
HC2 implant
Jaw
Arrow: Top view.
Direction of rotation:
clockwise.
1.10
Usage of the handgrip and HC2
adapter
1.11
Break off the holder
1.12
Insertion of the implant with axial pressure while turning
IT HC
2.1 PICK-UP IMPRESSIONS
2.1.1 Impression taking using the pick up method
Impression taking using an individual impression tray
2.1.2 Prior to the impression
Impression taking with an A-silicone such as Safeprint by
Dr. Ihde Dental. The use of open and closed impression
tray is possible.
The impression post HLT must not necessarily be unscre-
wed from the implant in order to remove the impression
tray. It can be repositioned later as well.
1. SURGERY AND PICK-UP IMPRESSIONS
Handgrip
5
2.1.3 Take impression
Disengage HLT from the implant: HLT remains in the im-
pression
After the impression is taken, the implant is closed with a
healing abutment (Gingiva former - straight or anatomic)
and the impression is sent to the laboratory.
2.2.1 Impression taking with a closed impression tray
Impression taking using an individual impression tray
2.2.2 Prior to the impression
Impression taking with an A silicone such as Safeprint® IM
by Dr. Ihde Dental.
When applying the closed-tray method, the impression
post TS/TSL HC is not located in the impression tray after
the impression is removed, but in the implant.
2.2.3 Remove impression
When the closed-tray method is applied, the impression
post TS/TSL HC remains on the implant after the impres-
sion tray is removed.
After removal of the impression tray the impression post
will be unscrewed and repositioned in the impression.
After the impression is taken, the implant is closed with a
healing abutment (Gingiva former - straight or anatomic)
and the impression is sent to the laboratory.
Loosen screws with HT 1.25
Window in impression tray
HLT
HC2 implant
Jaw
Tighten the impression post
manualy with the top screw
TS/TSL HC
HC2 implant
Jaw
Arrow: Top view.
Direction of rotation:
clockwise.
Impression post
TS/TSL HC
HC2 implant
Jaw
TS/TSL HC
HC2 implant
Jaw
2.2 CLOSED TRAY IMPRESSION TAKING
2.2 CLOSED TRAY IMPRESSION TAKING
6
3. PROCEDURES IN THE LABORATORY
3.1 Preparation of the impression tray for the
model generation
Pick up method:
Screw analog or M-analog against the impression
post.
3.2 Closed-tray method:
Screw analog IA or M-analog IA HC M to the transfer
post TS HC. �
Subsequently the impression post is repositioned in
the impression. �
The impression can now be casted. In M analogs (IA
HC M) block the lower access to the lock screw out prior
to casting.
3.3
The impression is poured. Then the impression
posts (HLT or TS/TSL HC) are unscrewed from the
laboratory analog.
3.4
The laboratory analog is now in the proper position
and orientation in the gipsum.
3.5
Positioning of the screwable abutments
TLA15 HC, thereby the optimal position and
adequate angulation must be determined.
NOTE The hexagon must be completely inserted
into the analog.
Fasten the laboratory
analog in the impression
using HT 1.25
HLT
IA or M-analog
Tighten the impression post onto
the laboratory analog using the
knurled screw
TS HC
IA or IA HC M
�
�
Gipsum
HT 1.25
screw eindrehen
TLA 15Achten Sie auf die korrekte
Stellung des Hexagons
IA or IA HC M
Gipsum
IA or IA HC M
Laboratory analog
Pour gipsum in
Arrow: Top view.
Direction of rotation:
clockwise.
Arrow: Top view.
Direction of rotation:
counterclockwise.
Arrow: Top view.
Direction of rotation:
clockwise.
7
The problem:
Implants, which are inserted with directional
divergences, cause the technician increased
difficulties
Advantages at a glance:
3. PROCEDURES IN THE LABORATORY
TLA 15 HC
Pattern Resin
Bone
3.6
Ensure proper position of the abutment when
transferring into the mouth.
Tightening torque of the screw during fastening on
the implant: 20 Ncm
3.7
If more than one angled abutment is used, your
laboratory will prepare a detachable synthetic bar (e.g.
from Pattern Resin) in order to facilitate the correct
positioning in the mouth.
TLA 15 HC
Bone
Access to
safety screw
Gingival region
Retention area
Jacket
Safety
screw
APPLICATION IA HC M ANALOGUE
Please note:
For HC2 implants with internal dual-hex (12 corners, HC2) the implant analogs IA 12 (REF 418177) or IA12 M (REF 418178) are
used.
The solution:
IA HC M tapered analog with jacket and TS HC or TSL HC impression
post (right)
The lock screw is firmly tightened against the analog using the hexagon wrench HT
1.77. Thereby the jacket is affixed. Now the M analog can be secured in the impres-
sion. Depending on the implant type this is done by attaching or tightening of the
M analog.
M analogs for HC2: Code: IA HC M
Pack with 1 analog, 1 lock screw and 5 sleeves (REF 418115)
• each M analog can be removed from the model at any time.
• no saw-cut model is necessary any more
• the 8 degree cone angle compensates for angulations between the implants
• all M analogs are equipped with rotation protection.
• damaged analogs can be easily replaced on the same model.
• adjustments can be made directly on the analog. The analog can be easily removed
from the model
• disturbing analogs can be temporarily removed
• these analogs do not need to be “deflasked” from previous models. Leave the jacket,
use the analogs right again
• the analogs can be reused. You save money
8
The surface is sandblasted and acid etched (high temperature), the apical thread has a No-Itis® surface. HC2 implants feature an internal 6-edge, an internal
marginal taper and a US standard internal thread.
NEW HC2 implant is now available with aggressive apical
thread and No-Itis® surface
As a result of many years of clinical observation of products, Dr. Ihde Dental AG has revised the design of the
famous HC2 implant: the broadened apical thread is fully self-cutting and carries the yellow No-Itis® surface.
Thanks to the new thread portion, the implant is more stable even in weak bone and high insertion torque
can be reached.
If the implant is anchored in the 2nd cortical, it may be used in immediate load protocols.
Especially in the upper jaw the usage of the new handgrip (REF311431, with Adapter IT HC REF418196, see
page 15) for inserting the implant is advisable. This tool allows to apply vertical insertion forces and will en-
hance the anchorage.
The drill sequence remains unchanged compared to the former design of the HC2 implant. And of course all
abutments and tools remain the same also if the new implant is applied.
IMPLANTS
Dimensions HC2 4.5 + 13
a) thread Ø basal 4.3 mm
b) hight of the wide thread part
c) micro thread length 2.5 mm
d) nominal Ø 3.7 mm
HEXACONE® IMPLANTS
Description enossal Ø enossal length Code REF Price cat.
HC2 2.9 13 2.9 mm 13 mm HC2 2.9 13 13-413200 G
HC2 2.9 15 2.9 mm 15 mm HC2 2.9 15 13-413201 G
HC2 3.3 8 3.3 mm 8 mm HC2 3.3 8 13-413220 G
HC2 3.3 10 3.3 mm 10 mm HC2 3.3 10 13-413221 G
HC2 3.3 11.5 3.3 mm 11.5 mm HC2 3.3 11.5 13-413222 G
HC2 3.3 13 3.3 mm 13 mm HC2 3.3 13 13-413223 G
HC2 3.3 15 3.3 mm 15 mm HC2 3.3 15 13-413224 G
HC2 3.7 8 3.7 mm 8 mm HC2 3.7 8 13-413202 G
HC2 3.7 10 3.7 mm 10 mm HC2 3.7 10 13-413203 G
HC2 3.7 11.5 3.7 mm 11.5 mm HC2 3.7 11.5 13-413210 G
HC2 3.7 13 3.7 mm 13 mm HC2 3.7 13 13-413204 G
HC2 3.7 15 3.7 mm 15 mm HC2 3.7 15 13-413205 G
HC2 4.5 8 4.5 mm 8 mm HC2 4.5 8 13-413206 G
HC2 4.5 10 4.5 mm 10 mm HC2 4.5 10 13-413207 G
HC2 4.5 11.5 4.5 mm 11.5 mm HC2 4.5 11.5 13-413208 G
HC2 4.5 13 4.5 mm 13 mm HC2 4.5 13 13-413209 G
HC2 5.5 8 5.5 mm 8 mm HC2 5.5 8 13-413211 G
HC2 5.5 10 5.5 mm 10 mm HC2 5.5 10 13-413212 G
HC2 5.5 11.5 5.5 mm 11.5 mm HC2 5.5 11.5 13-413213 G
HC2 5.5 13 5.5 mm 13 mm HC2 5.5 13 13-413214 G
HC2 2.9, 3.3, 4.5 and 5.5 implants will be converted from Dualhex to Hex.
HC2 2.9 must only be used in the load reduced area and not as single tooth implant.
c
b
a
d
9
IMPLANTS AND SURGICAL ACCESSORIES
HC2 6+2 was developed specially for the area of the 1st and 2nd molars in the upper and lower jaw. It is possible and recommendable to use it as a com-
pression screw implant in the upper jaw. Enossal length 6 mm. The upper edge of the polished 75° reverse cone can end at bone level or slightly above
it. The surface is sandblasted and acid etched in a high-temperature process.
Application limitations: HC2 2.9 mm implants may not be placed in a heavily loaded area, especially not in the molar area. Likewise these implants may
not be used where diagonal loading occurs, i.e. not for upper anteriors. Under no circumstances may HC2 2.9 mm implants be used for work that involves
unsupported occlusal surfaces (consoles). If used in immediate load protocols, the prosthetic construction must be inserted on the 2nd postoperative
day, and it should not be removed within the first 6 months.
SURGICAL ACCESSORIES
Each implant is supplied with this surgical screw (CSTI - REF 418101).
HEXACONE® 6+2 IMPLANTS
Description enossal Ø enossal length Code REF Price cat.
HC2 4.5 6+2 4.5 mm 6-8 mm HC2 4.5 6+2 13-413217 G
HC2 5.5 6+2 5.5 mm 6-8 mm HC2 5.5 6+2 13-413218 G
HC2 4.5 6+2 and 5.5 6+2 implants will be converted from Dualhex to Hex.
HC2 6+2 can be countersinked in the bone.
75 ° conus
b
a
c
a) enossal Ø: 4.5 - 5.5 mm
b) enossal length: 6 - 8 mm
c) reverse conus: 2 mm
4.6 mm 4.6 mm
6 mm 6 mm
Description Code REF Price cat.
Gingivaformer
for 3 mm
mucous membrane heightHSI 3 13-418111 B
for 5 mm
mucous membrane heightHSI 5 13-418112 B
Wide
gingivaformer
for 3 mm
mucous membrane heightHSIW 3 13-418191 B
for 5 mm
mucous membrane heightHSIW 5 13-418192 B
Anatomic
gingivaformer
3 mm hight, 4.5 mm widthHSI 3-4.5
13-418268 B
3 mm hight, 5.5 mm widthHSI 3-5.5
13-418269 B
5 mm hight, 6.7 mm width HSI 5-6.7 13-418270 B
Gingivaformer 3 mm hight, 3.3 mm width HSIS 3-3.3 13-418277 B
Gingivaformer MU for HC MU GF MU 13-418302 B
No-Itis® The innovative implant surface
Implants with the new No-Itis®-surface are double-sandblasted and then etched under heat. As final step an electro-chemical staining will be applied
and this last step of the surface preparation smoothens small wells (up to around 5 Micrometers) in the surface. Smooth implants (e.g. systems BCS, BOI)
receive the final surface treatment without any prior sandblasting or etching.
This surface, which is new in dental implantology, is extremely clean and keeps bacteria off the the implant. This surface has been proven successful in
traumatology for a long time. The reduction of roughness on the surface helps preventing peri-implantitis.
During the insertion of the implants the smoother surface reduces heat and thereby less torques is necessary and less heat develops. All other parameters
like the drilling sequence etc. remain unchanged.
10
ABUTMENTS FOR CEMENTED PROSTHETICS
HC SET 1
Screw Description Code REF Price cat.
Screwable abutment without anti-rotation protection for cemented
bridges. Trimmable down and abradable is possible. Height above
implant 8.5 mm, screw in with HT 1.25 Hex key.
The impression is made directly on the TCA, with tool TZ HC.
TCA 13-418129 B
Screwable abutment without anti-rotation protection for cemented
bridges. Trimmable down and abradable is possible. Screw in with HT
1.25 Hex key. The impression is made directly on the TCA.
TCA W 13-418173 B
Superstructure with hex and screw, straight, for cemented bridges.
Can be shortened and trimmed.
Height above implant 8.5 mm. Fasten with HT 1.25.
TLA HC 13-418133 D
Abutment narrow to HC 2.9
Fasten with HT 1.25.TLAS 13-418134 D
Abutment TLA HC 2 (GH) TLA HC2 13-418170 D
Abutment TLA HC 4 (GH) TLA HC4 13-418171 D
Abutment TLA W TLA W 13-418193 D
TLA, 15º angled, 1 mm gingival height (GH), including screw TLA 15 HC1 13-418135 F
TLA, 15º angled, 2 mm gingival height (GH), including screw TLA 15 HC2 13-418136 F
TLA, 15º angled, 3 mm gingival height (GH), including screw TLA 15 HC3 13-418137 F
TLA, 25ºangled, 1 mm gingival height (GH), including screw TLA 25 HC1 13-418139 F
TLA, 25º angled, 2 mm gingival height (GH), including screw TLA 25 HC2 13-418140 F
TLA, 25º angled, 3 mm gingival height (GH), including screw TLA 25 HC3 13-418141 F
Anatomic Abutment ANAB 13-418276 E
Recommended insertion torque: 20 Ncm
SF 21
SF 21
GH
GH
Impression taking and laboratory accessories
Code/REF TLA HC REF 13-418133 TZ HC REF 13-418179 PA HC REF 13-418181 IA 12 REF 13-418177**
Straight abutment Impression utility Castable cap Implant analogue
with 12-edge
REF 13-418182 for the complete SET
Price category for the SET F
Description Code REF Price cat.
Temporary cap, for TCA W HC, TLA W HC TPB W 13-418194 A
TEMPORARY CAP
** Note: For the HC2-implant with 12-edge, the implant analogue IA12 (REF 13-418177) or IA12 M (REF 13-418178) is used.
Impression taking and laboratory accessories
Code
REF
TSL HC
13-418110
TS HC
13-418109
HLT
13-418108
HLTC
13-418107
SF HLT long
13-418185
IA HC / IA 12 HC 2
13-418113/13-418177**
IA HC M / IA 12 M HC 2
13-418115/13-418178**
PA HC
13-418181
PA TLA HC
13-418172
Impression
post for HC.
Hight:
15.5 mm
Impression
post
for HC,
Hight:
10.6 mm
Impression
post for TLA,
TLA 15,
TLA 25
Transfer
post. Sui-
table for
6-edge and
12-edge.
Click on.
Pickup-
screw for
418108/HLT
impression
post
418113 = Implant
analogue for Hex
418177 = for Dual Hex
Implant analogue
with 5 sleeves,
intern. Hex
Castable
abutment
for TCA, TLA
Castable
abutment
for TLA
HC2/4
Price cat. C C C C B B C A A
11
ABUTMENTS FOR SCREW-IN PROSTHETICS
** Note: For the HC2-implant with 12-edge, the implant analogue IA12 (REF 13-418177) or IA12 M (REF 13-418178) is used.
Impression taking and laboratory accessories
CodeREF
TST13-418147
SFL BTT13-418100
PSTR (grey)
13-418124 (pack of 5)
PSTA
13-418123 (pack of 5)
SF13-418151
Transfer post Long screw TCT-analogue
Castable
abutment,
12 mm high, round
inside
Castable
abutment,
12 mm high, edged
inside
Fastening
screw
Price cat. B A B B B B
In this approach the position of the TCT hex is assigned.
Description Code REF Price cat.
Screwable abutments for bridges and bars,
deliverable for mucogingival levels of 3, 4, 5, 6 mm.
Screw in with HT 1.77.
TSA 3 13-418143 B
TSA 4 13-418144 B
TSA 5 13-418145 B
TSA 6 13-418146 B
Description Code REF Price cat.
Screwable mesostructure for bridges and bars,
screw in with HT 1.77 hex key. For mucogingival
levels of 0.5 mm, 1.5 mm and 2.5 mm.
TCT HC 0.5 13-418130 B
TCT HC 1.5 13-418131 B
TCT HC 2.5 13-418132 B
Recommended insertion torque: 20 Ncm
Impression taking and laboratory accessories
CodeREF
TS13-418142
BTS13-418152
PSS (white) SF13-418151
Impression post TSA-analogue Castable abutment, 10.5 mm high
5 per pack
Fastening screw for
PSS on BTS/TSA
Price cat. B B B B
Impression taking and laboratory accessories
Code
REF
IA HC / IA 12 HC 2
13-418113 / 13-418177**
HLT
13-418108
PSTR
13-418124 (5 pieces)
Implant analogue for HC2�
IA HC = 6 -edges, IA 12 HC = 12-edges
Transfer post for
HC and HC 2, with Hex.
Alternative: Castable abutment,
round inside 12 mm high
Price cat. B C B
IA HC IA 12 HC
Description Code REF Price cat.
Screw for PSTA SF TCTL 418165 A
Castable abutment,12 mm high, edged inside PSTA 418123 B
Screw-on mesio structure for bridges and bases TCTL 0.5 13-418138 D
TCT Set (TCTL 0.5, PSTA, SF TCTL) 13-418263 F
Hex
TCT-SET
SF TCTL
PSTA
TCTL 0.5
12
ABUTMENTS FOR REMOVABLE PROSTHETICS
CASTABLE ABUTMENTS
Description Code REF Price cat.
Castable abutments with hex, incl. metal basis (Ti6Al4V)
and screw for crown 10 mm high, white (not for casting-on)PLAB 13-418120 B
Castable abutments with hex, incl. metal basis (CoCrMo)
and screw for crown 10 mm high, white (for direct casting-on)PLAB 2 13-418183 G
Castable abutment including screw for bridges
(UCLA technique) 10 mm, red PLA 13-418158 B
Recommended insertion torque: 20 Ncm
** Note: For the HC2-implant with 12-edge, the implant analogue IA12 (REF 13-418177) or IA12 M (REF 13-418178) is used.
ABUTMENTS
Description Material Code REF Price cat.
HEX-reverse-abutment,
incl. screw SF 275 (REF 418275)Ti6Al4V HRA HC 13-418273 D
Tempbase for HRA HC TPB E 13-418274 C
This abutment turns the internal hexagon of the HC2-implant into an external standard-hexagon. The prosthetic screw is
screwed through. It tightens the prosthetic and the abutment at the same time.
Ti6Al4V CoCrMo
SF 275
TPB E
HRA HC
Impression taking and laboratory accessories
Code HLT HLTC SF HLT long IA HC / IA 12 HC 2
REF 13-418108 13-418107 13-418185 13-418113 / 13-418177**
Transfer post for
HC and HC 2, with Hex.
Transfer post for snap-
ping onto the HC2 im-
plant. Suitable for 6-edge
and 12-edge. Click on.
Pickup-screw for
418108 / HLT
impression post
Implant analogue for
HC2�
IA HC = 6 -edges
IA 12 HC = 12-edges
Preiskat. C C B B
IA HC IA 12 HC
Description Code REF Price cat.
Localicer® for HC2 hight 2 mm LOC HC 2 13-418116 C
Localicer® for HC2 hight 4 mm LOC HC 4 13-418117 C
Adequate tool: HT 1.77. We recommend to use minimally 6 implants per jaw and to use a single denture as splinting.
LOCALICER®
ACCESSORIES FOR LOCALICER®
Description Code REF Price cat.
instrument for mounting and de-mounting Localicer Caps LOC Tool 13-462335 C
Analog + impression cap set AA LOC 13-462337 C
Set with 4 nylon caps + 1 housing NCS 13-462338 D
Set of two, yellow, with increased friction strength R-Cap 13-462336 B
Pull-off force: *black 125 g, red 350 g, blue 500 g, pink 1000 g // *Black cap is for temporary solutions up to one month.
13
ABUTMENTS FOR REMOVABLE PROSTHETICS AND ACCESSORIES
MULTI - UNIT ABUTMENTS
ACCESSORIES FOR MULTI - UNIT ABUTMENTS
Description Code REF Price cat.
Temporary base, Ti6Al4V (SF MU2 has to be ordered separately) TC MU2 13-418290 D
Transfer straight, Ti6Al4V, incl. screw SFL MU2 TS MU2 13-418291 C
Castable for multiunit, incl. screw PA MU2 13-418292 A
Screw for TC MU, Ti6Al4V SF MU2 13-418293 A
Lab analogue for multiunit, Ti6Al4V IA MU2 13-418295 B
Hex-instrument, long HT 1.25 13-425100 C
Hex-instrument, long HT 1.77 13-425103 C
Screw
Description Code REF Price cat.
MU2 17 HC, angled, Ti6Al4V, incl. SF 20 MU2 17 HC 13-418281 L
MU2 35 HC, angled, Ti6Al4V, incl. SF 20 MU2 35 HC 13-418282 L
MU2S 0.5 HC, straight, Ti6Al4V MU2S 0.5 HC 13-418283 G
MU2S 1.5 HC, straight, Ti6Al4V MU2S 1.5 HC 13-418284 G
MU2S 2.5 HC, straight, Ti6Al4V MU2S 2.5 HC 13-418285 G
GF MU2 gingivaformer, Ti6Al4V. Incl. SF MU2 H above abutment shoulder 6 mm 13-418286 C
MU2 localicer, Ti6Al4V. Incl. SF MU2 Hight above abutment shoulder 6.7 mm 13-418287 C
Prosthetic screw for MU2 SF 20 13-420943 A
Insertion of the angled MU2-abutments with HT 1.25 // Insertion of the straight MU2S-abutments with HT 1.77
h = 3 mm
h = 4 mm
6 mm
6.7 mm
Description Code REF Price cat.
Ball abutments for fitting prostheses, head diameter 2.5 mm, for fit-
ting to TSA 3 - 6 abutments, application on TSA 3 - 6 abutmentsSB 13-418153 B
ACCESSORIES
Impression taking and laboratory accessories
Pu
ll-o
ff f
orc
es
Nylon cap NC Nylon cap R-NC
with increased friction strength
(for usage with worn-out Localicer)
for poly-
merization
directly into
the denture
Code TSA 3 to 6 TS BTS NC NC 1 NC 2 R-NC R-NC 1 R-NC 2 H
REF 13-418143 / -46 13-418142 13-418152 13-465028 13-465029 13-465030 13-465034 13-465033 13-465032 13-465031
Hight above
implant
3-6 mm
Impression post
for TSA and TCA
TSA-
analogue
ca. 1200 g
white
ca. 800 g
pink
ca. 500 g
yellow
green, strong pink,
medium
orange,
soft
Metal sleeve
for
all NC
Price cat. B B B A A A A A A B
Description Hight h Code REF Price cat.
Ball abutment, head - 0 2.5mm
Use with HT 1.25
Use with
NC-caps
0.5 mm TB 0.5 13-418126 B
2.0 mm TB 2 13-418127 B
4.0 mm TB 4 13-418128 B
hh
h
Description Code REF Price cat.
Bur cylinder, rverse, 20°. Incl. screw SF 20 FZ 20 HC 13-418262 D
REVERSE CONE CYLINDER
14
INSTRUMENTS AND TOOLS
Description Type Code REF Price cat.
IT 2.5 Short, 8 mm, click-on, 6-edges IT 2.5 13-418174 B
ITL 2.5 Long, 22 mm, click-on, 6-edges ITL 2.5 13-418175 B
ITM 2.5 Medium, 20 mm, click-on, 6-edges ITM 2.5 13-418176 B
IT 2.5 M Insertion tool for contra-angle handp. IT 2.5 M 13-418150 B
IT 2.5 M 12Insertion tool contra-angle, 12-edges
Insertion torque with max. 40 NcmIT 2.5 M 12 13-418200 B
ITWH 2.5 M Insertion tool contra-angle for HC ITWH 2.5 M 13-418184 C
IT1 12 Insertion tool short,12-edges IT1 12 13-418197 B
IT2 12 Insertion tool long, 12-edges IT2 12 13-418198 B
IT3 12 Insertion tool medium,12-edges IT3 12 13-418199 B
INSERTION TOOLS
Ø working rangeMax. working
depth
Total
lengthColor Code REF Price cat.
0.1 mm - 1.5 mm 15 mm 31.7 mm yellow BCD 1 13-900240 C
0.1 mm - 1.5 mm 15 mm 42 mm yellow BCDX 1 13-900243 C
2.0 mm / 3.6 mm 13 mm 30 mm DFN 2.9 13 13-418102 E
2.0 mm / 3.6 mm 15 mm 32 mm DFN 2.9 15 13-418103 E
2.0 mm 17 mm 36.5 mm DS 2 13-425001 D
2.8 mm 17 mm 36.5 mm DS 2.8 13-425005 D
2.7 mm 18 mm 36 mm DFN 3.0 13-425030 E
3.0 mm 18 mm 36 mm DFN 3.4 13-425031 E
3.4 mm 18 mm 36 mm DFN 3.7 13-425032 E
4.05 mm 18 mm 36 mm DFN 4.2 - 4.5 13-425033 E
4.4 mm 18 mm 36 mm DFN 5.5 13-425034 E
2.7 mm 18 mm 39 mm DFLN 3.0 13-425035 E
3.0 mm 18 mm 39 mm DFLN 3.4 13-425036 E
3.4 mm 18 mm 39 mm DFLN 3.7 13-425037 E
4.05 mm 18 mm 39 mm DFLN 4.2 - 4.5 13-425038 E
3.4 mm 11.5 mm 30 mm DFSN 3.7 13-425039 D
3.9 mm 11.5 mm 30 mm DFSN 4.2 - 4.5 13-425040 D
max. 3.4 mm 5 mm 27 mm C Drill 3.7 13-425043 D
max. 4.05 mm 5 mm 27 mm C Drill 4.2 - 4.5 13-425044 D
max. 5.8 mm 5 mm 27 mm C Drill 5.5 13-425045 D
It has been scientifically proven that HEATLESS® drills generate 55% less heat than traditional bone drills from other
manufacturers. This makes it possible to use higher rotational speeds: between 3,000 and 5,000 rpm are recommended
with good external cooling and intermittent drill technique.
Namely for the implant system HC2
HEATLESS® - DRILL DFN / DFLN FOR IMPLANTS WITH CONICAL CORE
TITAN BASE
Description Type Material Code REF Price cat.
Titan base
incl. screwAbutment basefor zirkonium Ti6Al4V MB HC 13-418267 D
Titan base
incl. screwTi6Al4V MBR HC 13-418272 D
Base for TPB 3 HC
incl. screw SF 20For scanner to HC implants BioPEEK® TPB 3 HC 13-418280 B
15
TOOLS
TOOLS
Description Type Length Code REF Price cat.
Hex-Tool 1.25 long 21 mm HT 1.25 13-425100 C
Torx-Tool 1.25 long 21 mm TT 1.25 13-425105 C
Hex-Tool for contra-angle, extra-long 45 mm HTW 1.25 13-425111 B
Hex-Tool 1.25 kurz 14 mm HTS 1.25 13-425101 C
Hex-Tool 1.77 for all suprastructures 19 mm HT 1.77 13-425103 C
Hex-Tool 1.25 M for contra-angle, long 26.1 mm HT 1.25 M 13-425112 B
Hex-Tool 1.77 M for contra-angle, long 28.6 mm HT 1.77 M 13-425113 B
Hex-Tool, extra-long 45 mm HTX 1.25 13-425102 C
Hex-Tool for contra-angle, extra-long 45 mm HTX 1.77 13-425104 C
Punch, 4.9 mm 0 for contra-angle PUW 1 13-425404 C
Punch, 3.9 mm 0 for contra-angle PUW 2 13-425405 C
Punch, 5.2 mm 0 manual PU 13-425406 C
standardized
probe
Scale 1 mm for X-ray
measurements22 mm PDG 13-425400 A
DX2Drill extension contra angle
handpiece, extends by 19 mmDX2 13-500704 D
DX2 H
Drill extention contra angle handpiece, extends
by 19 mm, W&H-hexagon on shank and in front
section
DX2 H 13-500708 D
Guide sleeveTitanium, for pilot drill,
5 pieces / unit10 mm BFH 13-425401 A
X-ray measuring
sphere
Surgical steel
5 pieces / unit, 0.5 mm 0RM 13-425403 A
Adapter for handgrip Adapter IT HC 13-418196 C
Handgrip, self-locking ** 13-311431 V
Handgrip for adapter, screwable 13-311430 V
Adapter for all contra angle handpiece
instruments, compatible to handle110 mm 13-310530 C
Ratchet For all Hex-instruments and applicators RAT 2 13-425051 K
RatchetAngled, for all Hex-instruments and
applicatorsRAT 3 13-425052 S
TW 2 * Torque Wrench (heavy duty), 10 - 70 Ncm TW 2 13-425402 S
UAW***
Universal ratchet adapter for all contra angle
handpiece instruments, for TW2
and RAT2. Max. 30 Ncm
UAW 13-425107 E
* It is recommended to have the torque ratchets recalibrated by us once a year.
** for cleaning this instrument an ultrasonic cleaning device and a thermo-desinfector (i.e. Miele TD-series) are required. If these devices
are not available in the dental office the handle with REF311430 should be purchased instead.
*** Do not use UAW in combination with RAT2 with Heatless® bone drills. The maximum insertion torque for UAW and Heatless® bone drills
is 35 Ncm. UAW may therefore be used only with TW2.
TOOLS
HANDGRIP-TRAY (empty)
REF 13-60043
EUR 69,90
Size of closed tray: B = 90 mm / L = 195 mm / H = 45 mm
For safe storage and sterilisation
of handgrips (max. 3 pieces) and
adapters (max. 8 pieces). Plastic,
autoclavable up to 134°C, not sui-
table for dry heat sterilizers.
16HC216
Autoclavable up to 134 °C, not suitable for dry heat sterilizers
This surgical kit contains all drills and tools for first works with the HC2 system.
Material: autoclavable plastic
Description Code REF Price €
Insertion tool IT 2.5 13-418174
Insertion tool ITL 2.5 13-418175
Insertion tool
for contra angle handpieceIT 2.5M 13-418150
Hex-instrument, long HT 1.25 13-425100
Spiraldrill DS 2 13-425001
Formdrill DFN 3.0 13-425030
Formdrill DFN 3.4 13-425031
Formdrill DFN 3.7 13-425032
Formdrill DFN 4.2 - 4.5 13-425033
Corticalis drill 3.7 C Drill 3.7 13-425043
Torque wrench TW2 13-425402
Starter-Tray for HC2 (with content)* 13-S60021-K 678,80
* tray without content upon request
Cortic
rqu
art
y w
Cor
Tor
Sta
* tray
HC2 STARTER - TRAY
17TP
GGGIMMEDIATE LOADING IMPLANT SYSTEM TPG®
18
enossal
length
1.8 mm
4.8 mm
Implant with Strauman® compatible
internal cone connection.
TPG® implants provide a sharp and cutting crestal thread, which
engages in to the corticals of the extraction sockets. The apical
compression thread provides perfect stability both in com-
pressed spongious and in cortical bone. TPG® implants may
be used both in extraction sockets and in healed bone areals.
TPG® implants are used for circular bridges and segments with
at least 3 implants. Under adequate loading conditions and if
enough implants are splinted, the treatment will be performed
in an immediate load protocol. Due to their polished surface
with No-Itis®-protection, TPG® implants are extremely resistant
agains bacterial colonialization and peri-implantitis.
TPG® implants may be inserted into fresh
extraction sockets with lateral and bi-corti-
cal engagement.
View from top: TPG® implants may be in-
serted into fresh extraction. The voids in the
extraction sockes may receive augmentati-
on or fill with woven bone.
TPG® implant engaged in the distal maxilla
and the pterygoid plate of the shenoid
bone. In this indication this implant may be
used for circular bridges and segments with
at least 3 implants.
TPG® IMPLANTS
EXAMPLES FOR THE USE OF TPG® IMPLANTS
Description enossal Ø enossal length H above bone Surface Material REF Pricecat.
TPG 4.1 8 4.8 mm 8 mm 1.8 mm Polished, No-Itis® Ti6Al4V 13-420106 H
TPG 4.1 10 4.8 mm 10 mm 1.8 mm Polished, No-Itis® Ti6Al4V 13-420107 H
TPG 4.1 12 4.8 mm 12 mm 1.8 mm Polished, No-Itis® Ti6Al4V 13-420104 H
TPG 4.1 15 4.8 mm 15 mm 1.8 mm Polished, No-Itis® Ti6Al4V 13-420100 H
TPG 4.1 17 4.8 mm 17 mm 1.8 mm Polished, No-Itis® Ti6Al4V 13-420101 H
TPG 4.1 19 4.8 mm 19 mm 1.8 mm Polished, No-Itis® Ti6Al4V 13-420102 H
TPG 4.1 21 4.8 mm 21 mm 1.8 mm Polished, No-Itis® Ti6Al4V 13-420103 H
TPG 4.1 23 4.8 mm 23 mm 1.8 mm Polished, No-Itis® Ti6Al4V 13-420105 H
Minimal differences to the displayed No-Itis® implant-colour has no influence on the quality of the surface.
All TPG® implants are being delivered incl. OF5 STI (REF 13-420426)
TPG 4.1 8
TPG 4.1 10
19
Description Neck Code REF Price cat.
Plattform offset abutment. Tool: IT K (REF 462320) * 1.5 mm OF 4 STI 13-420424 * E
Plattform offset abutment, large head. 1.3 mm OF 5 STI 13-420426 E
Plattform offset abutment, large head. 15° bendable. 4 mm OF 5L STI 13-420427 E
Insertion tool, length 20 mm. Tool: RAT, RAT2, TW2 IT K 13-462320 C
Insertion tool for large head IT2 BCS 13-900030 E
* Laboratory accessories can be found in the Simpladent KOC® BECES® system application (REF13-0003-09)
Description Code REF Pricecat.
Straight, screw-in abutment with pre-assembled screw for
cemented crowns or bridges. Fasten with TT 1.25.TLA STI 13-425110 D
15°angled, screw-in abutment with pre-assembled screw for
cemented crowns or bridges. Fasten with TT 1.25. TLA 15 STI 13-425111 E
20°angled, screw-in abutment with pre-assembled
screw for cemented crowns or bridges. Fasten with TT 1.25.TLA 20 STI 13-425112 E
25°angled, screw-in abutment with pre-assembled
screw for cemented crowns or bridges. Fasten with TT 1.25. TLA 25 STI 13-425113 E
One-piece abutment with solid thread, for cemented crowns or bridges,
height above implant 6.5 mm, 8°incline. Can be shortened or trimmed.
Flat on one side. Direct impression or transfer to lab
analogue IA STI. Fasten with HT 1.25.
TCA STI 13-420313 D
Single-piece and anti-rotational octagonal screw-in abutment;
for crowns, bridges and bar connectors. Height above implant: 1.5 mm.
Fasten with HT 1.77 / HTX 1.77
OSA STI 13-420338 D
CoCrMo bar coping for all S implants with SF 350 or SF 365,
height above implant: 5.5 mm. Screw directly onto implant.SKC STI 13-420366 E
With conus (without anti-rotation)
Description Code REF Pricecat.
Castable abutment incl. SF 215 replacement screw, 10.5 mm high.
For ULCA. Fasten with HT 1.25.PA STI 13-420358 C
Use with impression post TSX STI (REF13-420411) and analogue IA STI (REF13-420347)
TPG® ABUTMENTS
ABUTMENTS FOR CEMENTED PROSTHETICS
ABUTMENTS FOR SCREW-IN PROSTHETICS
PLATTFORM OFFSET ABUTMENTS
External
octagon
20
TPG® DRILLS AND TOOLS
Description Code REF Price cat.
Adapter TPG, Engages in the internal tri-lobe inside the conus. Adapter TPG 13-425106 D
Adapter short, for large head AA5 15° 13-462052 C
Adapter short, for large head AA5 25° 13-462053 C
Handgrip, not dismountable. Cleaning in ultrasonic bath at 45°,
with alkaline cleansing medium. For adapters, self-locking.Handgrip 13-311431 V
Handgrip, dismountable, can be cleaned manually.
For adapters, can be disassembled.Handgrip 13-311430 V
* must be used with max. 60 Ncm only. If the bone is harder, IT1-3 TPG has to be used with RAT 2 or TW2.
AA5 15°
AA5 25°
°
255°
Surgical steel, color coded, depth coded and autoclavable. The drill is marked with laser depth markings.
HEATLESS® DRILLS «DOS» FOR TPG® IMPLANTS
TOOLS FOR MANUAL INSERTION
Description Code REF Pricecat.
Insertion tool, short 12 mm IT1 TPG 13-425116 C
Insertion tool, medium 16 mm IT2 TPG 13-425117 C
Insertion tool, long 20 mm IT3 TPG 13-425118 C
Insertion tool for contra-angle, 30 mm ITM TPG 13-425119 B
Ratchet, for all Hex-instruments and insertion tools 13-RAT 2 K
Description max. working length Code REF Price cat.
BCD 1 15 mm 13-900240 C
BCDX 1 15 mm 13-900243 C
DOS 2, black 17 mm DOS 2 13-455312 D
DOS 4, blue 21 mm DOS 4 13-455314 D
DOS 5, green 17 mm DOS 5 13-455315 D
Twist Drill 2.0 / 21, Ø 2.0 mm 21 mm 13-90022 D
Countersink head, for implant-Ø 4.1 mm DRA 4.1 CS 13-425047 D
D
INSERTION TOOLS FOR RATCHET
21TPG21
TPG® ACCESSORIES
Description Code REF Pricecat.
Lab analogue, with cone IA STI 13-420347 B
Transfer post, engages on the outer cone, 9 mm high
Tighten with HT 1.25TS STI 13-420345 B
Transfer post, engages on the outer cone, 13 mm high
Tighten with HT 1.25TSX STI 13-420411 B
Hex-instrument. Length 21 mm, Ø 1.25 mm HT 1.25 13-HT 1.25 C
Hex-instrument extralong. Length 45 mm, Ø 1.25 mm HTX 1.25 13-425102 C
Torx-instrument. Length 21 mm, Ø 1.25 mm TT 1.25 13-TT 1.25 C
Synthetic base castable, 7.4 mm high. For TLA STI and TCA STI PS 13-420354 A
Synthetic base castable, 4 mm high. For TLA15 STI PS 15 13-420355 A
Synthetic base castable, 4 mm high. For TLA20 STI PS 20 13-420356 A
Synthetic base castable, 4 mm high. For TLA25 STI PS 25 13-420357 A
IMPRESSION AND LABORATORY ACCESSORIES
22
TPG® SMALL TRAY
Description Code REF Price € / cat.
Insertion tool IT2 BECES 13-900030
Insertion tool IT K 13-462320
Hex instrument, long HT 1.25 13-425100
Pathfinder drill BCD 1 13-900240
Pathfinder drill BCDX 1 13-900243
Heatless drill for implants with conical core DOS 2 13-455312
Heatless drill for implants with conical core DOS 4 13-455314
Heatless drill for implants with conical core DOS 5 13-455315
Countersink heads DRA 4.1 CS 13-425047
Standardized probe PDG 13-425400
Twist drill Twist Drill 2.0 / 21 13-90022
Insertion tool, short 12 mm IT1 TPG 13-425116
Insertion tool, medium 16 mm IT2 TPG 13-425117
Insertion tool, long 20 mm IT3 TPG 13-425118
Insertion tool for contra-angle, 30 mm ITM TPG 13-425119
Ratchet for all Hex-instruments and insertion tools RAT 2 13-425051
Small tray with content 13-S60047-K 729,90
Small tray empty 13-60047-K 149,90
optional Adapter TPG 13-425106 D
optional Handgrip for adapters, self-locking 13-311431 V
23
IMPLANT SURFACE AND MATERIAL
NO-ITIS® - THE INNOVATIVE IMPLANT SURFACE
Implants with the new No-Itis®-surface are double-sandblasted and then etched ander heat. As final step
an electro-chemical staining will be applied and this last step of the surface preparation smoothens small
wells (up to aroand 5 Micrometers) in the surface. Smooth implants (e.g. systems BECES®, COI) receive the
final surface treatment without any prior sandblasting or etching. This surface, which is new in dental
implantology, is extremely clean and keeps bacteria off the the implant. This surface has been proven
successful in traumatology for a long time. The reduction of roughness on the surface helps preventing
peri-implantitis. During the insertion of the implants the smoother surface reduces heat and thereby less
torques is necessary and less heat develops. All other parameters like the drilling sequence etc. remain
unchanged.
Ti6AL4V ELI
Ti6AL4V ELI also called „Grade 5“, is a purified version of the conventional 6/4 Ti-Alloy. It is used for more
than 50% of all human metallic implants. The material features superb mechanical and chemical pro-
perties, and it is very resistent against corrosion. Also regarding biocompatibility and the support of cell-
growth this specific titanium-alloy shows advantages compared to pure titanium. For these reasons the
material is used so widespread. Ti6AL4V ELI provides a 25% higher stability compared to pure titanium.
0029REF 13-0005-09_TPG-HC2_SAP_EN_V001_05/15
Simpladent GmbH Phone +41 (0)55 293 23 70
Dorfplatz 11 [email protected]
CH - 8737 Gommiswald / SG www.simpladent-implants.com
PLEASE CHECK THE VALID PRICE LIST OF YOUR LOCAL COORDINATOR
Symbols on the packs
Production
No.
Sterilized by
gamma radiation
Sterilization
Ethylene oxide
Single-
use
product
Instruction
for use
Expiry date Store
in a dry
place
Temperature
range from
-5 °C to 35 °C
Store tightly
keep closed
Do not use if
packing is
damaged
Do not
resterilize
Manu-
facturer
-5 °C
35 °C
Rx ONLYNon-
sterile
Intended for use by
dentists or surgeons
only
Production
date
We are certified DIN EN ISO 13485, and annex II of EEC Directive 93/42 EWG (2007).
Due to technical reasons the product dimensions shown in this brochure might deviate from reality. HC2 and TPG® are registered trademarks.
HC2 implants are patent-protected.
In case that implants would be reprocessed (cleaned, resterilized) infections could occur, because no validated procedures for reprocessing are available.
dent GmbH
tz 11
37 Gommiswa
CHECK THE VA
hone +41 (0)55 2
SG w
Sim
Do
CH
PLE
23 70
ent.ch
s.com
ATOR