1 1 © 2010 PerkinElmer Fabio Mariconti Inorganic Line Leader Budapest, October 17th 2016 ICP-MS Technology, NexION 350

Upload: others

Post on 21-Jul-2020

4 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: ICP-MS Technology, NexION 350per-form.hu/wp-content/uploads/2016/10/Techniques... · Budapest, October 17th 2016 ICP-MS Technology, NexION 350 . 2 ICP-MS NexION 350... Environmental

1 1

© 2010 PerkinElmer

Fabio Mariconti

Inorganic Line Leader

Budapest, October 17th 2016

ICP-MS Technology,

NexION 350

Page 2: ICP-MS Technology, NexION 350per-form.hu/wp-content/uploads/2016/10/Techniques... · Budapest, October 17th 2016 ICP-MS Technology, NexION 350 . 2 ICP-MS NexION 350... Environmental

2 2

ICP-MS NexION 350...

Environmental38%

Chemical4%Bio/Clinical

15%

Petro2%

Teaching1%

Speciation1%

Semiconductor7%

Research9%

Nuclear3%

Geo/Metallurgical15%

Food5%

Page 3: ICP-MS Technology, NexION 350per-form.hu/wp-content/uploads/2016/10/Techniques... · Budapest, October 17th 2016 ICP-MS Technology, NexION 350 . 2 ICP-MS NexION 350... Environmental

3 3

1983 - ELAN 250 First commercially available ICP-MS

Dual Cone Interface

PlasmaLOK

Cryogenic pumping system

On-board computer

Manual operation

1987 - ELAN 500 Joint Venture between PerkinElmer + SCIEX formed 50% SCIEX (R&D / Manufacturing) - 50% PerkinElmer (Sales and Service)

Improved sensitivity

Improved resolution - new quadrupole and power supply

OmniRange - extended dynamic range

1990 - ELAN 5000 First turbo pumped ICP-MS

First free-running ICP RF generator

Improved interface

TotalQuant - First automated spectral interpretation and semi-quantitative analysis package

1994 - ELAN 6000 First simultaneous automatic extended dynamic range detection system – 9 orders of magnitude

First single lens system - AutoLens - simplicity with improved performance

1999 - ELAN 6100DRC Dynamic Reaction Cell (DRC)

First Cell ICP-MS capable of sub-ppt detection limits for all Period 4 elements (K-Se)

Wins PittCon Gold Award for best new product

Uses high temperature plasma conditions, minimal matrix effects, obsoletes cool plasma

2002 – 2005 ELAN 9000, DRCe, DRC II Improved low mass sensitivity

Dual Inlet Turbo pump

Cassette Torch

Chromera

2008 - Celebrating 25 years of success in ICP-MS

Over 3300 ELAN ICP-MS systems worldwide

Over 1600 DRC systems worldwide

NexION: Introducing the

Page 4: ICP-MS Technology, NexION 350per-form.hu/wp-content/uploads/2016/10/Techniques... · Budapest, October 17th 2016 ICP-MS Technology, NexION 350 . 2 ICP-MS NexION 350... Environmental

4 4

Introducing the

…the 7th generation ICP-MS from PerkinElmer

New RF Generator

New vacuum

system

New form factor

Better lab

facilities

utilization

New sample

introduction system

New torch

positioning

motorized XYZ

New torch mount

Improved cassette

New interface

Triple cone design

New ion optics

Quadrupole Ion

Deflector New detection system

New methods of

interference removal

Universal Cell Technology

New plasma viewport

color screen New

mass analyzer

NexION:

Page 5: ICP-MS Technology, NexION 350per-form.hu/wp-content/uploads/2016/10/Techniques... · Budapest, October 17th 2016 ICP-MS Technology, NexION 350 . 2 ICP-MS NexION 350... Environmental

5 5

NexION Features

New interface

Unique combination of Triple

Cone Interface and

Quadrupole Ion Deflector

(QID)

Why is this important?

New Methods of Interference Reduction (UCT™)

Q

3 Q

2

Q

1

Quadrupole

2

1

Cones

3

Page 6: ICP-MS Technology, NexION 350per-form.hu/wp-content/uploads/2016/10/Techniques... · Budapest, October 17th 2016 ICP-MS Technology, NexION 350 . 2 ICP-MS NexION 350... Environmental

6 6

NexION Features

Page 7: ICP-MS Technology, NexION 350per-form.hu/wp-content/uploads/2016/10/Techniques... · Budapest, October 17th 2016 ICP-MS Technology, NexION 350 . 2 ICP-MS NexION 350... Environmental

7 7

NexION Features

Page 8: ICP-MS Technology, NexION 350per-form.hu/wp-content/uploads/2016/10/Techniques... · Budapest, October 17th 2016 ICP-MS Technology, NexION 350 . 2 ICP-MS NexION 350... Environmental

8 8

Benefits of a Triple Cone Interface

Well confined beam (~2-3 mm)

Almost no deposition inside

mass spectrometer

NexION

Pressure drops in two

smaller steps:

• less beam divergence,

• dirt prevented from

entering mass spectrometer

– very little drift, reduced

maintenance

Sampler

cone Skimmer

cone

Hyper-

skimmer

cone

pre

ssure

distance

Large beam divergence

Lots of material deposits inside

mass spectrometer

Older ICP-MS systems –

Pressure drops in one large

step:

• large beam divergence,

• more dirt into mass

spectrometer – creates more

drift and required maintenance

Legend: Blue = beam from plasma (ions + neutrals + dirt);

Red = pressure drop per cone

Sampler

cone Skimmer

cone

pre

ssure

distance

Material that never

enters mass

spectrometer

Page 9: ICP-MS Technology, NexION 350per-form.hu/wp-content/uploads/2016/10/Techniques... · Budapest, October 17th 2016 ICP-MS Technology, NexION 350 . 2 ICP-MS NexION 350... Environmental

9 9

NexION Features

Incredible stability

Mass spectrometer (including cell) remains clean

when analyzing dirty samples

Ion Optics require very infrequent cleaning

More time running samples, less time cleaning or

recalibrating

Very low background noise

10x improvement in low mass sensitivity

Unique combination of Triple Cone Interface

and Quadrupole Ion Deflector provide:

Sampler

Skimmer

ICP

Hyper skimmer

QID

Photons, Neutrals and un-ionized particles

Analyte

ions

Page 10: ICP-MS Technology, NexION 350per-form.hu/wp-content/uploads/2016/10/Techniques... · Budapest, October 17th 2016 ICP-MS Technology, NexION 350 . 2 ICP-MS NexION 350... Environmental

10 10 10

QID: Excellent Mass Selectivity

.

100

61

25

67

100

57

31

68

100

0

20

40

60

80

100

120

Mg In U

% t

ran

sm

iss

ion

Autolens setting

DRCe Autolens bandpass

Mg

In

U

100

23

10

37

100

16 13

40

100

0

20

40

60

80

100

120

Mg In U

% t

ran

sm

iss

ion

Autolens setting

NexION 300X Quadrupole Ion Deflector bandpass

Mg

In

U

Page 11: ICP-MS Technology, NexION 350per-form.hu/wp-content/uploads/2016/10/Techniques... · Budapest, October 17th 2016 ICP-MS Technology, NexION 350 . 2 ICP-MS NexION 350... Environmental

11 11

15-hour stability using Rh as the internal standard. All results have been normalized to the

mean. Most analytes were +/- 3% over the 15 hour period. Sample matrix was TCLP

extract solution at 0.1% Sodium Acetate. The analytes were spiked at 100 ppb.

NexION Features

Page 12: ICP-MS Technology, NexION 350per-form.hu/wp-content/uploads/2016/10/Techniques... · Budapest, October 17th 2016 ICP-MS Technology, NexION 350 . 2 ICP-MS NexION 350... Environmental

12 12

The Spectral Interferences Problem in ICP-MS

.

Interferences on major isotopes inhibiting low-level determinations

27Al from 12C15N+

51V from 35Cl16O+

52Cr from 40Ar12C+

56Fe+ from 40Ar16O+, 40Ca16O+

58Ni from 58Fe+,40Ca18O+ (use 60Ni)

60Ni from 44Ca16O+

63Cu from 23Na40Ar+ (use 65Cu)

75As from 40Ar35Cl+

80Se from 40Ar40Ar+

Page 13: ICP-MS Technology, NexION 350per-form.hu/wp-content/uploads/2016/10/Techniques... · Budapest, October 17th 2016 ICP-MS Technology, NexION 350 . 2 ICP-MS NexION 350... Environmental

13 13

NexION Features

Universal Cell Technology - Tri-Mode cell instrument

Standard (STD) mode (uses no cell gases)

Unique vented cell provides true classical ICP-MS spectra

Collision mode (uses an inert cell gas, collision gas)

Separates analyte ions from polyatomic isobaric interferences using Kinetic

Energy Discrimination (KED)

Reaction mode (uses a reactive gas)

Separates analyte ions from interfering ions based on specific chemistry in the

Dynamic Reaction Cell (DRC)

The right mode for the application

Page 14: ICP-MS Technology, NexION 350per-form.hu/wp-content/uploads/2016/10/Techniques... · Budapest, October 17th 2016 ICP-MS Technology, NexION 350 . 2 ICP-MS NexION 350... Environmental

14 14

Universal Cell Technology (UCT™)

The STD mode is most suitable for:

Applications with few interferences on analytes of interest

Low mass or high mass elements

… BECs for interfered elements can be in the ppb range

Page 15: ICP-MS Technology, NexION 350per-form.hu/wp-content/uploads/2016/10/Techniques... · Budapest, October 17th 2016 ICP-MS Technology, NexION 350 . 2 ICP-MS NexION 350... Environmental

15 15

Important mid-mass elements suffer from common interferences

Ca (Ar)

K (ArH)

V (ClO)

Cr (ArC/ClO)

As (ArCl/ CoO)

Se (Ar2,Kr )

Cu (ArNa, NaCa)

Ni

(ArO/CaO/NaCl)

Limitations of STD mode

Cd (MoO)

Zn (SO2/ArMg/CaO)

Fe (ArO/CaO)

Mg (C2)

Al

(CN/CNH, N2 spread)

Page 16: ICP-MS Technology, NexION 350per-form.hu/wp-content/uploads/2016/10/Techniques... · Budapest, October 17th 2016 ICP-MS Technology, NexION 350 . 2 ICP-MS NexION 350... Environmental

16 16

ICP-MS : Interferences correction

Page 17: ICP-MS Technology, NexION 350per-form.hu/wp-content/uploads/2016/10/Techniques... · Budapest, October 17th 2016 ICP-MS Technology, NexION 350 . 2 ICP-MS NexION 350... Environmental

17 17

Universal Cell Technology (UCT™)

The Collision (KED) mode is most suitable for:

First row transition elements

Applications susceptible to common interferences at moderate levels

… BECs for interfered elements are in the ppt range

Page 18: ICP-MS Technology, NexION 350per-form.hu/wp-content/uploads/2016/10/Techniques... · Budapest, October 17th 2016 ICP-MS Technology, NexION 350 . 2 ICP-MS NexION 350... Environmental

Properties of KED mode operation

KED –

Dependent on collision cross sections, gas density, cell length and voltage barrier

Efficiency limited to 3-4 orders of magnitude interference reduction

Physical and electronic filtering process

Page 19: ICP-MS Technology, NexION 350per-form.hu/wp-content/uploads/2016/10/Techniques... · Budapest, October 17th 2016 ICP-MS Technology, NexION 350 . 2 ICP-MS NexION 350... Environmental

KED – Kinetic Energy Discrimination

Kinetic Energy Discrimination - The process of separating energy distributions by

exploiting differences in

In ICP-MS, it’s specifically the difference in collision cross section between elemental

ions and their polyatomic interfering isobars.

e.g. 56Fe & 56ArO, 78Se & 78Ar2, 51V & 51ClO

collision cross sections

Although polyatomic isobars share the same mass and energy

characteristics as the elemental ions, they can have significantly larger

collision cross sections (σ).

larger cross-section = more collisions = lower energy

Compare: V and ClO - K and ArH Se and Ar2 - Cd and MoO Fe and ArO - Au and TaO Δσ >> 0, good performance Δσ ~ 0, poor performance

Page 20: ICP-MS Technology, NexION 350per-form.hu/wp-content/uploads/2016/10/Techniques... · Budapest, October 17th 2016 ICP-MS Technology, NexION 350 . 2 ICP-MS NexION 350... Environmental

Energy distribution of ions inside the instrument

interface

= ion1 of m/z

= ion2 of same m/z

QID cell

entrance

cell

exit

KED

barrier

KED barrier repels ions that don’t have enough energy to enter the quad

Helium

σion1 >> σion2

Separation is based on difference Kinetic Energy

using appropriate HARDWARE (KED barrier, focousing multipole )

Page 21: ICP-MS Technology, NexION 350per-form.hu/wp-content/uploads/2016/10/Techniques... · Budapest, October 17th 2016 ICP-MS Technology, NexION 350 . 2 ICP-MS NexION 350... Environmental

21 21

Collisions /Reactions Cells

Take Away Here

Collision - reduction of energies of polyatomic ions

Page 22: ICP-MS Technology, NexION 350per-form.hu/wp-content/uploads/2016/10/Techniques... · Budapest, October 17th 2016 ICP-MS Technology, NexION 350 . 2 ICP-MS NexION 350... Environmental

22 22

Sensitivity Interferences

Elimination

Collision Cell KED (Energy Discrimination)

Page 23: ICP-MS Technology, NexION 350per-form.hu/wp-content/uploads/2016/10/Techniques... · Budapest, October 17th 2016 ICP-MS Technology, NexION 350 . 2 ICP-MS NexION 350... Environmental

Universal Cell Technology (UCT™)

The Reaction (DRC) mode is most suitable for:

Applications with the highest level of interferences that require the lowest BECs

… BECs for interfered elements are in the sub-ppt range

Page 24: ICP-MS Technology, NexION 350per-form.hu/wp-content/uploads/2016/10/Techniques... · Budapest, October 17th 2016 ICP-MS Technology, NexION 350 . 2 ICP-MS NexION 350... Environmental

Properties of DRC mode operation

DRC –

Uses electronics to drive the reaction to the products side

A + B ► C + D (bandpass forces C + D to dominate)

Dependent only on the reaction rates of the ions in the beam and the reaction gas

Faster reaction rates dominate over slower ones

Independent of mass

Can be extremely efficient at reducing interferences

Up to 9 orders of magnitude reductions are possible

Purely chemical filtering process between ions and molecules

Page 25: ICP-MS Technology, NexION 350per-form.hu/wp-content/uploads/2016/10/Techniques... · Budapest, October 17th 2016 ICP-MS Technology, NexION 350 . 2 ICP-MS NexION 350... Environmental

25 25

Low (RPq) and High (Rpa) mass cut-off applied

.

“q” parameter sets

“a” parameter sets

Ar +

CH 4

CH 3

+

CH 2 +

C H 2 3

+

C H 2 5

+

C H O 3 7

+

ArH +

C H n x+1

+

CH 4

+

C H n x-1

+

CH 4

CH 4

CH 4

CH 4

CH 4

C H n x

C H n x

C H O 3 6

C H O 3 6

C H O 3 6

C H O 3 6

AMU

B Mg Al Ca Fe Co Se + + + + + + +

low mass cut-off

high mass cut-off

Page 26: ICP-MS Technology, NexION 350per-form.hu/wp-content/uploads/2016/10/Techniques... · Budapest, October 17th 2016 ICP-MS Technology, NexION 350 . 2 ICP-MS NexION 350... Environmental

26 26

Now, let’s link the DRC

bandpass to the mass

analyzer ...

Ar +

CH 4

CH 3

+

CH 2 +

C H 2 3

+

C H 2 5

+

C H O 3 7

+

ArH +

C H n x+1

+

CH 4

+

C H n x-1

+

CH 4

CH 4

CH 4

CH 4

C H n x

C H n x

C H O 3 6

C H O 3 6

C H O 3 6

C H O 3 6

AMU

Mg +

Al +

Ca +

Fe +

Co +

Se +

B +

CH 4

This is the

DRC bandpass

This is the

quad bandpass

How DRC works…

Page 27: ICP-MS Technology, NexION 350per-form.hu/wp-content/uploads/2016/10/Techniques... · Budapest, October 17th 2016 ICP-MS Technology, NexION 350 . 2 ICP-MS NexION 350... Environmental

27 27

Ar +

CH 4

CH 3

+

CH 2 +

C H 2 3

+

C H 2 5

+

C H O 3 7

+

ArH +

C H n x+1

+

CH 4

+

C H n x-1

+

CH 4

CH 4

CH 4

CH 4

CH 4

C H n x

C H n x

C H O 3 6

C H O 3 6

C H O 3 6

C H O 3 6

AMU

B +

Mg +

Al +

Ca +

Fe +

Co +

Se +

… which provides for high

transmission of the analyte ions

...

How DRC works…

Page 28: ICP-MS Technology, NexION 350per-form.hu/wp-content/uploads/2016/10/Techniques... · Budapest, October 17th 2016 ICP-MS Technology, NexION 350 . 2 ICP-MS NexION 350... Environmental

28 28

Ar +

CH 4

CH 3

+

CH 2 +

C H 2 3

+

C H 2 5

+

C H O 3 7

+

ArH +

C H n x+1

+

CH 4

+

C H n x-1

+

CH 4

CH 4

CH 4

CH 4

CH 4

C H n x

C H n x

C H O 3 6

C H O 3 6

C H O 3 6

C H O 3 6

B +

Mg +

Al +

Ca +

Fe +

Co +

Se +

AMU

… while permitting reactive

isobars to be converted to

other masses ...

How DRC works…

Page 29: ICP-MS Technology, NexION 350per-form.hu/wp-content/uploads/2016/10/Techniques... · Budapest, October 17th 2016 ICP-MS Technology, NexION 350 . 2 ICP-MS NexION 350... Environmental

29 29

Ar +

CH 4

CH 3

+

CH 2 +

C H 2 3

+

C H 2 5

+

C H O 3 7

+

ArH +

C H n x+1

+

CH 4

+

C H n x-1

+

CH 4

CH 4

CH 4

CH 4

CH 4

C H n x

C H n x

C H O 3 6

C H O 3 6

C H O 3 6

C H O 3 6

B +

Mg +

Al +

Ca +

Fe +

Co +

Se +

AMU

… and simultaneously

suppressing the formation of

new interferences !

How DRC works…

Page 30: ICP-MS Technology, NexION 350per-form.hu/wp-content/uploads/2016/10/Techniques... · Budapest, October 17th 2016 ICP-MS Technology, NexION 350 . 2 ICP-MS NexION 350... Environmental

30 30

DRC Technology

Separation is based on different mass – charge ratio

using appropriate HARDWARE ( quadrupole to create a dynamic band pass )

“q” parameter sets

“a” parameter sets

low mass cut-off

high mass cut-off

Page 31: ICP-MS Technology, NexION 350per-form.hu/wp-content/uploads/2016/10/Techniques... · Budapest, October 17th 2016 ICP-MS Technology, NexION 350 . 2 ICP-MS NexION 350... Environmental

31 31

In a cell is the nature of the gas that we use to reduce/eliminate the interferences

that define if we have more reaction (CH4, NH3, O2, H2) or collision (He) events

Any supplier can introduce the gas that he wants in the cell but how to manage

what is happening in the cell depends from the hardware available in the different

instruments

THIS IS THE IMPORTANT DIFFERENCE AMONG THE INSTRUMENTS

AVAILABLE IN THE MARKET!

Interference Reduction (UCT™)

Page 32: ICP-MS Technology, NexION 350per-form.hu/wp-content/uploads/2016/10/Techniques... · Budapest, October 17th 2016 ICP-MS Technology, NexION 350 . 2 ICP-MS NexION 350... Environmental

32 32

In the past the customer had to chose

Not from Collision or Reaction cell !

Simply have everything ! All the technologies available nowadays to eliminate the

interferences in a Quadrupole ICP-MS !

EASY and POWERFUL !

Not from Collision or Reaction cell !

Interference Reduction (UCT™)

between KED (passive cell) approach or Band pass approach (active) cell

Page 33: ICP-MS Technology, NexION 350per-form.hu/wp-content/uploads/2016/10/Techniques... · Budapest, October 17th 2016 ICP-MS Technology, NexION 350 . 2 ICP-MS NexION 350... Environmental

33 33

NexION Features

Why is this important?

Improved Interference Reduction (UCT™)

Physics

1. Quadrupole can create a band pass (DRC)

2. Quadrupole can become a passive ion guide (KED)

3. Multi-pole can only be a passive ion guide

Axial Field

Electrode

Page 34: ICP-MS Technology, NexION 350per-form.hu/wp-content/uploads/2016/10/Techniques... · Budapest, October 17th 2016 ICP-MS Technology, NexION 350 . 2 ICP-MS NexION 350... Environmental

34 34

Extending Standard Working Range with EDR - Why is this important ?

EDR extends (not shifts) the upper linear range by 4 orders of magnitude

Selective attenuation of high signals through control of the bandpass parameters

If needed both low and high calibration ranges can run in a single method

Avoids dilutions

Avoids rerunning samples

All elements in the solution can be determined in a single

analytical run

Standard Calibration of Sodium – from DL to 10 mg/L Range Extended Calibration of Sodium – from DL to 500 mg/L Range

Linear Dynamic Ranges (LDR) were run for all major analytes: Aluminum, Calcium, Iron, Magnesium, Potassium, and

Sodium > 500 ppm

These analytes in environmental samples are usually high and frequently require additional dilutions to bring them

into range

Fewer dilutions for the user

Higher productivity for the lab