    ICARUSELECTRONIC GAS VOLUME CONVERTER

    1. INTRODUCTION 2. TECHNICAL DETAILS

    ICARUS is an electronic volume converter

    which, combined with a gas meter, provides

    the gas measurement in Sm3, calculated at

    base conditions of + 15 C and 1,01325 bar.

    ICARUS fully conforms to all current

    European standards and directives, such as

    EN 12405-1 and ATEX 94/9/EC, and is

    certified according to MID n 2004/22/EC.

    ICARUS is a type 1 device, manufactured

    with integrated pressure and temperature

    sensors.

    2.1 Pressure Measurement

    ICARUS is always fitted with an external

    piezo-resistive absolute pressure sensor,connected to the processing unit by a cable

    whose typical lenght is 3 mt.The sensor is enclosed in a stainless steel

    casing.

    The connection is G !, flat sealing.

    Available full scale ranges: 2,5; 6; 10; 30;

    81 bar A

    Accuracy: 0,1% of the full scale.

  • 7/24/2019 ICARUS EN

    2/6

    FIMIGAS S.p.A.Via C. Goldoni, 3 - 20063 CERNUSCO SUL NAVIGLIO (MI)Telefono n 02 9249234 - Fax n. 02 9249078E-mail: [email protected] http://www.fimigas.it

    2.2 Temperature Measurement

    ICARUS is always provided with a 4 wires PT

    1000 sensor compliant with European

    Standard EN IEC 60751 (its resistance value is

    1000" at 0 C). 4 wires connection grants ahigh accuracy measurement despite cable

    resistance.

    The sensor is connected to the processing unitby a cable whose typical lenght is 3 mt.

    The operating measuring range of the sensor is

    -30C70C.

    2.5 Outputs

    - N 1 optical serial port to communicate

    with a local PC (configuration setup, data

    download).

    - N 1 RS 232 serial port to connect to an

    external GSM/GPRS modem, necessary to

    achieve remote transmission of stored data

    for tele-reading applications.- N 1 RS 485 serial port, used or to

    communicate with a DCS by MODBUSRTU protocol, or to connect to an external

    GSM/GPRS modem, necessary to achieveremote transmission of stored data for tele-

    reading applications.

    2.3 Signal inputs

    ICARUS has an Open Collector high-

    frequency input and a low-frequency pulse

    input.When connected to a LF pulse emitter, gasmeter pulses are divided by a pulse weight

    factor [pulse/m3] which is set in the device.

    When Icarus is connected to a HF gas meter,

    its Open Collector HF input must be

    configured specifying:

    - The pulse weight factor.

    - The frequency value of the pulse train

    corresponding to the maximum flow

    rate.The maximum frequency that can be read

    by the input is 5Khz. HF input can operate

    only if ICARUS is powered by an external

    main DC voltage supply.

    - N 2 pulse outputs to transfer volume

    pulses associated to internal counters, or the

    presence of an alarm in the device.

    - N 1 PWM (pulse with modulated) output,

    by which its possible to have a 4-20 mAoutput associated to a parameter

    (temperature, pressure or flow-rate), thanks

    to an external converter that, located in safe

    area, converts the PWM signal into 4-20

    mA format.

    - N 1 LAN connection by ethernet port and

    n 4 4-20 mA outputs available on request.

    2.6 Flow-rate calculation

    ICARUS can calculate both measured flow-rate and base flow-rate (instantaneous or

    average flow-rate).

    The instantaneous flow-rate at measured gas

    conditions (Qm) is calculated by counting the

    number of read pulses and measuring the time

    elapsed since the last calculation (minimumtime 15 seconds).

    The instantaneous flow-rate at base gas

    conditions is calculated thanks to the

    following formula:

    Qb = Qm x C

    The average flow-rate Qb e Qm (stored every15 minutes) is calculated according to

    Standard UNI TS 11291-4.

    2.4 Conversion function

    Referring to ISO 12405 European Standard,

    ICARUS operates togheter with a gas meter to

    convert the measured gas volume into base

    conditions, according to the following

    formula:!"" #$ !"

    N 4 different algorithms for the calculation of

    compressibility:

    - SGERG 88

    - AGA8 Gross 1

    - AGA8 Gross 2

    - AGA NX19

  • 7/24/2019 ICARUS EN

    3/6

    FIMIGAS S.p.A.Via C. Goldoni, 3 - 20063 CERNUSCO SUL NAVIGLIO (MI)Telefono n 02 9249234 - Fax n. 02 9249078E-mail: [email protected] http://www.fimigas.it

    2.7 Keypad 2.9 Power

    The keypad is composed of five keys:

    - Four directional arrow shaped keys

    (Up, Down, Left, Right directions).

    - One confirmation central key (OK).

    Up/Down direction keys let the user:

    - Scroll up/down the items in a menu

    page.

    - Move the cursor of the display from

    one line to the upper/lower one.

    Left/Right direction keys let the user:

    -Return to the previous menu page.

    - Move the cursor of the display

    leftwards/rightwards.

    OK key is used:

    - To select a menu item.

    - To confirm a parameter setup.

    ICARUS can be powered by:

    - Internal lithium battery (3.6V)

    - External power (8Vdc)

    ICARUS is always provided with an internal

    battery satisfying ATEX requirement.The internal battery guarantees a life

    autonomy of 5 years at normal user

    conditions.

    In presence of the external power (compulsory

    to read HF incoming pulses or to activate the

    RS 485 serial port) the battery operates as

    backup power.

    Attention: external power is supplied by a

    specific model of power supply that is

    ATEX certified.It may only be used a battery, or a certified

    power supply, provided by Fimigas.

    3. INSTALLATION SITE

    ICARUS is generally installed in Zone 1

    hazardous area next to the gas pipeline. It canbe connected to a modem to remotely transfer

    recorded gas data. The modem shall always beinstalled in safe area and connected to

    ICARUS by an intrinsically safe barrier. The

    schemes reported below show the typical

    installation of the device.

    ICARUS enclosure is provided with 4 fixing

    threaded holes positioned on the back corners

    of the device. The internal diameter of the

    holes is 4 mm.

    By these holes ICARUS can be both wall-fixed and plate-fixed.

    2.8 Display

    The display of ICARUS lets the user:

    - Visualize real time gas data.

    - Visualize historical gas data.

    - Visualize active alarms on the device.

    - Visualize/modify setup.

    - Calibrate pressure and temperature

    sensors.

  • 7/24/2019 ICARUS EN

    4/6

    FIMIGAS S.p.A.Via C. Goldoni, 3 - 20063 CERNUSCO SUL NAVIGLIO (MI)Telefono n 02 9249234 - Fax n. 02 9249078E-mail: [email protected] http://www.fimigas.it

  • 7/24/2019 ICARUS EN

    5/6

    FIMIGAS S.p.A.Via C. Goldoni, 3 - 20063 CERNUSCO SUL NAVIGLIO (MI)Telefono n 02 9249234 - Fax n. 02 9249078E-mail: [email protected] http://www.fimigas.it

    4. DATA READING 9. REFERENCE DIRECTIVES AND

    STANDARDS

    Directives

    - European Directive 2004/22/EC related to

    measurement instruments (MID), MI-002.

    - European Directive 1994/9/EC related to

    equipment and protective systems

    intended to be used in potentially

    explosive atmospheres (ATEX).

    - European Directive 2006/95/EC related to

    low voltage devices;

    - European Directive 2004/108/EC related

    to electromagnetic compatibility (EMC);

    - European Directive 2002/95/EC related to

    the use of hazardous substances in

    electrical and electronic equipment(RoHs);

    - European Directive 2002/96/EC related to

    collection and recycling of electrical and

    electronic equipment (WEE).

    Standards

    - Italian Standard UNI 9167-2009 related to

    the design, manufacturing and test of

    natural gas measurements systems;

    - European Standard ISO 12405-1/A1, Gas

    meters Conversion instruments Part 1:

    Conversion of volume, 2005. related to

    the technical specifics of type 2 volume

    gas conversion devices. (this is the

    harmonized Norm of the cited MID

    Directive);

    - EN-IEC 60079-0, Explosive atmospheres

    Part 0: General requirements, 2007.- EN-IEC 60079-11, Explosive

    atmospheres Part 11: Equipment

    Protection by Intrinsic Safety, 2007.

    ICARUS has ample memory space for data

    storage.

    All the values in storage can be read in two

    different methods:

    By remote reading of the device viaexternal GSM/GPRS.

    By local connection to a PC, viaoptical or RS232/485 serial port.

    ICARUS meets all the requirements of

    the UNI 9167 Standard and is already

    equipped with the CTE communication

    protocol (upgradeable with future updates),

    according to UNI TS 11291:2009.

    5. ENVIRONMENTAL CONDITIONS

    -Ambient Temperature: -25C 55C

    - Storage Temperature: -40C 80C

    - Relative Humidity: 0% 100%

    6. MID CERTIFICATE

    ICARUS received by International Institute of

    Metrology the Certificate of Approval inaccordance with Directive MID (2004/22/CE):

    MID Certificate: CH-MI002-12009-00

    7. ATEX CERTIFICATE

    ICARUS is an intrinsically safe device and it

    has two ATEX markings:

    ATEX Certificate: TUV 11 ATEX 556245 X

    o Ex II 1G Ex ia IIC T4 Ga

    o Ex II (1)G [Ex ia] IIC

    8. DIMENSIONS AND WEIGHT

    IP 66 plastic case.

    Dimensions: 210 x 150 x 95 mm

    Weight: 1,5 Kg.

  • 7/24/2019 ICARUS EN

    6/6

    FIMIGAS S.p.A.Via C. Goldoni, 3 - 20063 CERNUSCO SUL NAVIGLIO (MI)Telefono n 02 9249234 - Fax n. 02 9249078E-mail: [email protected] http://www.fimigas.it

    - IEC 60079-14: Explosive atmospheres

    Part 14: Electrical installations design,

    selection and erection

    - IEC 60079-17: Explosive atmospheres

    Part 17: Inspection & Maintenance Of

    Installations In Hazardous Areas

    -

    EN-IEC 60079-28, Explosiveatmospheres Part 28: Protection of

    Equipment and transmission Systems

    using Optical radiation, 2006.

    - EN-IEC 60079-26, Explosive

    atmospheres Part 26: Equipment with

    Equipment Protection Level (EPL) GA,

    2007.

    - EN-IEC 62056-21, Electricity metering

    Data exchange for meter reading, tariff

    and load control Part 21: Direct local

    data exchange, 2002.

    - European Standard ISO 12213-1/2/3

    related to the calculation of gas

    compressibility factor (Z factor);

    - AGANX19 Standard related to the

    calculation of Z gas compressibility

    factor;

    - European Standard ISO 6976 related to

    the calculation of gas density;- International Standard IEC 60529,

    Degrees of Protection Provided by

    Enclosures (IP code), 2001.