i-CHROMATM T4 Test
Thyroxine (T4) is one of two major hormones produced by the thyroid gland (the other is called triiodothyronin,
or T3). T4 and T3 are regulated by a sensitive feedback system involving the hypothalamus and pituitary gland.
The hypothalamus releases the thyrotropin releasing hormone (TRH), which stimulates the pituitary to release the
thyroid stimulating hormone (TSH). This causes the thyroid to release T3 and T4 and these in turn regulate the
release of TRH and TSH via a feedback control mechanism. Normally, increased blood levels of T4 and T3 act to
decrease the amount of TSH secreted, thereby reducing the production and release of T4 and T3. Greater than
99% of T4 is reversibly bound to three plasma proteins in blood; thyroxine binding globulin (TBG) binds 70%,
thyroxine binding pre-albumin (TBPA) binds 20%, and albumin binds 10%. Approximately 0.03% of T4 is in the
free, unbound state in blood at any one time.
T4 is a useful marker for the diagnosis of hypothyroidism and hyperthyroidism. The level of T4 is decreased in
hypothyroidism, myxedema, chronic thyroiditis (Hashimoto’s disease). Increased levels of T4 have been found in
hyperthyroidism due to Grave’s disease and Plummer’s disease.
General Information
Key Features of i-CHROMATM T4 Test
• Quantitative Test Result
• Sample Type: Serum, and Plasma.
• Sample volume: 10 µl
• Detection Limit: 20 nmol/L
• Working range: 20-300 nmol/L
• Cut Off(normal range): 60-160 nmol/L
• Precision: CV% 2-10 in working range.
• Long Shelf Life: 20 months
• Fast Test Result: 10 minutes
• User Friendliness: 10 µl, 75 µl Pipette
• Comparable Test Result with Full Automatic Analyzer
• No Calibration Needed
Principle
The test uses a competitive immunodetection method, such that the detector antibody in buffer binds to T4 in
blood sample and antigen-antibody complexes competes with T4 covalently coupled to BSA that has been
immobilized on test strip as sample mixture migrates nitrocelluose matrix. Thus the presence of more T4 antigen
in blood would lead to accumulation of less antigen-antibody complexes on the test strip. Signal intensity of
fluorescence on detector antibody reflects amount of antigen captured and is processed from i-CHROMATM
Reader to show T4 concentration in specimen. The working range of i-CHROMATM T4 test is 20~300 nmol/L.
Comparison Data with Access2
1. Set a Test Device on a dust-free clean surface.
2. Check and insert ID Chip onto the instrument. Make sure that the Test Device lot # matches ID Chip lot #.
3. Take out a tube containing Detection Buffer from refrigerator and leave it at room temperature for 20
minutes.
4. Draw 10 µl of serum/plasma or control with a transfer pipette and add it to the tube containing Detection
Buffer.
5. Mix the specimen and Detection Buffer by tapping or inverting the tube
6. Take 75 µl of sample mixture and load it onto the well of disposable Test Device.
7. Leave the Test Device at room temperature for 10 minutes before inserting the device into the holder.
8. Insert Test Device onto the holder of i-CHROMATM Reader. And press “SELECT” button.
Checking the direction of Test Device, push the device back all the way.
9. Read the results on the display screen of i-CHROMATM Reader.
10. Test results will be shown on the display screen of i-CHROMATM Reader.
Test Procedure
Horm
on
e
i-CHROMATM T4 Test
BMP
ID chip
LOT:i-CHROMA T4
BMP
ID chip
LOT:i-CHROMA T4
BMP
ID chip
LOT:i-CHROMA T4
T4
T4
EXP :LOT :
for in vitro diagnostic use only
25Test
Detection Buffer Pouch 25 tubes of Detection Buffer
i-CHROMA Contents
Contents Box 25 Test Device Pouches(1 Test Device per Pouch)
Test Device 1 Mannual1 ID Chip
Select
123456
21 443Check the contents :1 ID chip, 25 Test Device,25 tubes of Detection buffer
Check and make sure thatthe ID chip lot number matchesTest Device lot number.
Insert the ID chip into the instrument.
Press "Select"
875 6Add the sample(10ul of serum/plasma) into the tube containing Detection buffer.
Collect 75ul of the mixture
10 times
Shake the tube up and down10 times or more : the mixture of buffer and thesample has to be used within 30 seconds.
T4
T4
D-Dm
er
1210 11
Transfer the mixture ontothe sample well of the Test Device.
10 minutes
Wait 10 minutes. Place the Test Device on theTest Device holder of the instrument.
000
D-Dm
er
Read the result on thedisplay screen.
Select
PSA
Press "Select" 13
T4
T4
same lot number
Detection Buffer
T4
TM