HPLC DETECTORS

BY:-VARUN GIRME

Page 2: HPLC Detectors

PAD:-DIODE ARRAY DETECTOR OR DAD OR HPLC PAD DETECTOR.

ELSD:-EVAPORATIVE LIGHT SCATTERING DETECTOR.

Page 3: HPLC Detectors

PAD:-DIODE ARRAY DETECTOR OR DAD OR HPLC PAD DETECTOR.

Page 4: HPLC Detectors
Page 5: HPLC Detectors

Diode array detector also referred to as a DAD detector or more specifically HPLC PAD detector.

PAD are used for obtaining spectral profiles from molecular mixtures or chromatographically separated samples.

Whether used to analyze molecules in a flow cell, static solution or solids they provide ultra-fast low noise spectral analysis.

Resolution determined by the number of diodes deployed over a specific wavelength range.

An HPLC photo diode array detector is coupled to elute of separation devices by molecular weight, hydrophobicity or ionic charge.

Page 6: HPLC Detectors

Certain HPLC PAD detectors handle the entire spectrum from the uv at 190nm to the near 1R at 1micron.

Other are specifically targeted to the uv visual or near 1R which up to 2048 elements provided for ultimate flexibility in resolution and multiplexing.

These mechanics provide high sampling speed of up to 190Hz and provide thermoelectric temp. control of sensor to limit noise.

Page 7: HPLC Detectors

ELSD:-EVAPORATIVE LIGHT SCATTERING DETECTOR.

Page 8: HPLC Detectors
Page 9: HPLC Detectors

Nebulization.The eluent from the HPLC system is forced

through a narrow orifice with a stream of gas flowingthrough a Venturi tube that sheaths the mobile phase toform small droplets that can be easily evaporated. Aflow of N2 at a pressure of 2–4 bar is typically used.

Page 10: HPLC Detectors

Vaporization. The mobile phase is removed from the nebulizer

droplets that contain the compound(s) of interest. The evaporation is performed by passing the droplets through a heated tube.

Page 11: HPLC Detectors

DetectionThe stream of solute particles that exits the

vaporization chamber enters the optical cell where the amount of light scattering, which is related to the mass of the compounds of interest, is measured. The sensitivity of the detector is ultimately determined by the number and shape of particles that are detected by the light scattering chamber in a given period

Page 12: HPLC Detectors

APPLICATION:-

Extremely small footprint to maximize availability bench space.

Wide gradient compatibility, no limitation on choice of solvent.

High sensitivity extremely low limits of detection.

Wide application range to cover very low to high molecular weight compounds.

Low dispersion minimum peak broadening for maximum resolution.

Rapid equilibrium fast set-up and high productivity.

Advanced easy to use instrument controls with built-in-safety.

Page 13: HPLC Detectors