Analysis of Beverages using Large Volume Static Headspace / GCMS

Page 2: Hit “esc” to quit hit any key to continue Analysis of Beverages using Large Volume Static Headspace / GCMS

hit “esc” to quit hit any key to continue

Presentation Overview• Introduction to Large Volume Static

Headspace (LVSH) Technique

• Advantages of LVSH/GC(MS)

• Analytical Procedure

• Examples of Beverages Analyzed

• Remote Headspace Analysis

• Summary

Page 3: Hit “esc” to quit hit any key to continue Analysis of Beverages using Large Volume Static Headspace / GCMS

hit “esc” to quit hit any key to continue

GC/MS Detection Limits vs Headspace Sample Volume

0.001 0.01 0.1 1.0 10 100 1000

1000

100

10

1

0.1

0.01

10-3

10-4

10-5

PPM

LOD

MINIMUM VOLUME REQUIRED (CC)

Vol vs Conc.

Loop Preconcentration

Page 4: Hit “esc” to quit hit any key to continue Analysis of Beverages using Large Volume Static Headspace / GCMS

hit “esc” to quit hit any key to continue

Large Volume Headspace Requires Preconcentration

HP 5973 GCMS Volume Reduction Large Volume

Inlet or Autosampler

Page 5: Hit “esc” to quit hit any key to continue Analysis of Beverages using Large Volume Static Headspace / GCMS

hit “esc” to quit hit any key to continue

Large Volume Static Headspace Autosampler

• Introduces 10-200cc of headspace to the 7100 Preconcentrator

• Accommodates solids and liquids

• Detection limits 100x lower than conventional loop injection headspace

• “See what you can smell”

Page 6: Hit “esc” to quit hit any key to continue Analysis of Beverages using Large Volume Static Headspace / GCMS

hit “esc” to quit hit any key to continue

LVSH Sample Platforms and Enclosures

• Samples are placed in disposable vials, reducing potential for contamination of sample platforms.

• Sample Platforms use quick connects, simplifying removal for oven bakeout.

Page 7: Hit “esc” to quit hit any key to continue Analysis of Beverages using Large Volume Static Headspace / GCMS

hit “esc” to quit hit any key to continue

Advantages of Large Volume Static Headspace (LVSH)

• Cleaner. No aerosols or foaming to contaminate sample transfer lines

• Larger Sample Loading• Low detection limits without sample heating.

– Less chance of producing artifacts.– Sample remains in natural state.

• Ideal for kinetic / aging studies– Outgassing rates– Effects of Atmosphere and Storage Temperature

• Better recovery of reactive compounds– Strong carbon based adsorbents are avoided– Inert, Silonite tubing used throughout

• No sample fractionation (SPME)

Page 8: Hit “esc” to quit hit any key to continue Analysis of Beverages using Large Volume Static Headspace / GCMS

hit “esc” to quit hit any key to continue

Food and Beverage Analysis by Large Volume Static Headspace (LVSH) / GCMS Analysis

HP 5973 GCMS 7100 7032L 4600

7100 3-Stage Preconcentrator

7032L 21-Position LVSH/MiniCan Autosampler

4600 Multi-Channel Standards Diluter

Page 9: Hit “esc” to quit hit any key to continue Analysis of Beverages using Large Volume Static Headspace / GCMS

hit “esc” to quit hit any key to continue

7032-LVSH / 7100 /5973 Large Volume Static Headspace Analyzer

GCMS

6

7100 7032-LVSH

SL I/O SL I/O

M1

M2

M3

Page 10: Hit “esc” to quit hit any key to continue Analysis of Beverages using Large Volume Static Headspace / GCMS

hit “esc” to quit hit any key to continue

PCB Integrated wiring for improved reliability

Modular design.

Easy replacement of traps

MFC downstream of traps

7100 Open Architecture

Page 11: Hit “esc” to quit hit any key to continue Analysis of Beverages using Large Volume Static Headspace / GCMS

hit “esc” to quit hit any key to continue

• Silonite tubing used throughout.

•Complete heating of sample flow path. Easily accessed for trap and tubing replacement.

•External Heaters outside each cryotrap

7100 Heated Flow Path

7100 Preconcentrator

Page 12: Hit “esc” to quit hit any key to continue Analysis of Beverages using Large Volume Static Headspace / GCMS

hit “esc” to quit hit any key to continue

SiloniteTM Fused Silica Lined Tubing

Page 13: Hit “esc” to quit hit any key to continue Analysis of Beverages using Large Volume Static Headspace / GCMS

hit “esc” to quit hit any key to continue

Silonite Tubing VOC Adsorption/Absorption Study

• Less adsorption of Diethylbenzene on Silonite tubing. No deactivation needed.

• Faster sample equilibration with tubing surface.

• Less potential for carryover.

50

100

150

200

250

300

350

1 2 3 4

Silonite

SilcoSteel

Deact.SilcoSteel

R

E

S

P

O

N

S

E

Flush Time (Min.)

Silonite

Brand B

Deact. Brand B

Page 14: Hit “esc” to quit hit any key to continue Analysis of Beverages using Large Volume Static Headspace / GCMS

hit “esc” to quit hit any key to continue

InternalStandard

CalibrationStandard

Glass Beads Tenax

Cryogen in

CO2

Focuser

ToGC

VOCs

7100 Water and CO2 Management Techniques

H2O

Helium

PUMPMFC

Helium Carrier

Removing Water and CO2

Sample

Page 15: Hit “esc” to quit hit any key to continue Analysis of Beverages using Large Volume Static Headspace / GCMS

hit “esc” to quit hit any key to continue

Foods,Packaging

MP&T Glass Beads (-150C)

Tenax (-30C)

High CO2Beverages

ECTD Empty (-150C)

Tenax (-50C)

High CO2/EtOHBeverages(Heavy-Ends)

MP&T Tenax (20C) Tenax (-30C)

High CO2/EtOHBeverages(Light-Ends)

ECTD Tenax (-10C) Tenax (-60C)

3-Stage Flexibility for GCMS Analysis of Multiple Sample Types and Matrices

Sample Type Mode Trap1 Trap2

Trap 3 - On-column focuser for capillary GC/MS

Page 16: Hit “esc” to quit hit any key to continue Analysis of Beverages using Large Volume Static Headspace / GCMS

hit “esc” to quit hit any key to continue

Improper Water Management

400cc VOC STD, Full Scan 15-270 amu

Page 17: Hit “esc” to quit hit any key to continue Analysis of Beverages using Large Volume Static Headspace / GCMS

hit “esc” to quit hit any key to continue

Vinyl Chloride

y = 285937x - 13186

R2 = 1

-500000

500000

1500000

2500000

3500000

4500000

5500000

6500000

7500000

0 5 10 15 20 25

ppbv

MS

D A

bu

nd

an

ce Vinyl Chloride 100x Calibration Curve usingVariable Volume / Single Standard Calibration

Page 18: Hit “esc” to quit hit any key to continue Analysis of Beverages using Large Volume Static Headspace / GCMS

hit “esc” to quit hit any key to continue

3 .5 0 3 .55 3 .6 0 3 . 6 5 3 .70 3 .7 5 3 . 80 3 .5 8 3 . 9 0 3 . 95

7 5 0 0 02 6 0 0 0

2 5 0 0 0

2 4 0 0 0

2 3 0 0 0

2 2 0 0 0

2 1 0 0 0

2 0 0 0 0

1 9 0 0 0

1 8 0 0 0

1 7 0 0 0

1 6 0 0 0

1 5 0 0 0

1 4 0 0 0

1 3 0 0 0

1 2 0 0 0

1 1 0 0 0

1 0 0 0 0

9 0 0 0

8 0 0 0

7 0 0 0

6 0 0 0

5 0 0 0

4 0 0 0

3 0 0 0

2 0 00

1 0 0 0

0

7 0 0 0 0

4 5 0 0 0

6 5 0 0 0

5 0 0 0 0

4 0 0 0 0

3 0 0 0 0

2 5 0 0 0

3 5 0 0 0

2 0 0 0 0

1 5 0 0 0

1 0 0 0 0

5 0 0 0

6 0 0 0 0

5 5 0 0 0

E than o lm /z 4 54 .35 m in0 .17 ppb .

M ethan o lm /z 3 13 .67 m in .0 .49 ppb

Fast Injection and Proper Water Management Minimizes Tailing of Polar VOCs

Column: HP1, 60m, 0.32mm ID, 1um film.

Flow rate: 1.5 ccm

Carrier: He

Page 19: Hit “esc” to quit hit any key to continue Analysis of Beverages using Large Volume Static Headspace / GCMS

hit “esc” to quit hit any key to continue

Advantages of 7100 Multi-Stage Preconcentration

• Multiple Preconcentration Options Depending on Target Compounds and Matrix (CO2, ETOH)

• Reduced Temperature Trapping Minimizes Sample Stress, Maximizing Recovery of Thermally Labile Flavor and Odor Compounds

• Silonite Tubing Creates an Extremely Inert Sample Flow Path to Further Improve Recovery

• Demonstrated Recovery of Sulfur and Nitrogen Compounds

Page 20: Hit “esc” to quit hit any key to continue Analysis of Beverages using Large Volume Static Headspace / GCMS

hit “esc” to quit hit any key to continue

LVSH Analysis of Yuban Coffee - 100cc

Page 21: Hit “esc” to quit hit any key to continue Analysis of Beverages using Large Volume Static Headspace / GCMS

hit “esc” to quit hit any key to continue

100cc LVSH - Coke/Pepsi Challenge

Coke

Pepsi

Page 22: Hit “esc” to quit hit any key to continue Analysis of Beverages using Large Volume Static Headspace / GCMS

hit “esc” to quit hit any key to continue

TreeTop Apple Juice100cc, LVSH60m, HP1, 0.32mm ID, 1um7032LVSH/7100/5973

Apple Juice by LVSH / GCMS

Page 23: Hit “esc” to quit hit any key to continue Analysis of Beverages using Large Volume Static Headspace / GCMS

hit “esc” to quit hit any key to continue

Beer by LVSH/GCMS

Beer, ETOH Removed200cc, LVSH60m, HP1, 0.32mm ID, 1um7032LVSH/7100/5973

Page 24: Hit “esc” to quit hit any key to continue Analysis of Beverages using Large Volume Static Headspace / GCMS

hit “esc” to quit hit any key to continue

Beer by LVSH/GCMSBeer, ETOH RemovedFront End, Post ETOH200cc, LVSH60m, HP1, 0.32mm ID, 1um7032LVSH/7100/5973

1

2

3

4

1 Acetone2. Ethyl Formate3. Dimethyl Sulfide4. Propanol

Page 25: Hit “esc” to quit hit any key to continue Analysis of Beverages using Large Volume Static Headspace / GCMS

hit “esc” to quit hit any key to continue

Beer Light Ends by LVSH/GCMS

Beer, ETOH M1 RetainedCold Trap DehydrationLight Ends200cc, LVSH60m, HP1, 0.32mm ID, 1um7032LVSH/7100/5973

Page 26: Hit “esc” to quit hit any key to continue Analysis of Beverages using Large Volume Static Headspace / GCMS

hit “esc” to quit hit any key to continue

Beer Light Ends by LVSH/GCMS

Beer, ETOH M1 RetainedCold Trap DehydrationLight Ends200cc, LVSH60m, HP1, 0.32mm ID, 1um7032LVSH/7100/5973

Acetaldehyde

COS

Page 27: Hit “esc” to quit hit any key to continue Analysis of Beverages using Large Volume Static Headspace / GCMS

hit “esc” to quit hit any key to continue

2% Milk by LVSH/GCMS

2% Milk100cc, LVSH60m, HP1, 0.32mm ID, 1um7032LVSH/7100/5973

Page 28: Hit “esc” to quit hit any key to continue Analysis of Beverages using Large Volume Static Headspace / GCMS

hit “esc” to quit hit any key to continue

2% Milk100cc, LVSH60m, HP1, 0.32mm ID, 1um7032LVSH/7100/5973

Dimethyl Sulfide

2% Milk by LVSH/GCMS

Page 29: Hit “esc” to quit hit any key to continue Analysis of Beverages using Large Volume Static Headspace / GCMS

hit “esc” to quit hit any key to continue

MiniCan Samplers Allow Direct Sampling of Gas Phase Flavors and Odor Compounds

• Collection of Remote Headspace Samples Improves Monitoring Effeciency.

• MiniCan Samplers are SiloniteTM Coated to reduce surface losses.

• Monitor Process Streams, Holding Tanks, Reactors, etc.

Page 30: Hit “esc” to quit hit any key to continue Analysis of Beverages using Large Volume Static Headspace / GCMS

hit “esc” to quit hit any key to continue

Four Models of MiniCans

MC400 MC400L MC400S MC400V

Page 31: Hit “esc” to quit hit any key to continue Analysis of Beverages using Large Volume Static Headspace / GCMS

hit “esc” to quit hit any key to continue

7032 Autosampler

Gas Phase Analysis Using MiniCan Silonite Coated Samplers

Surrogate Spiking

Page 32: Hit “esc” to quit hit any key to continue Analysis of Beverages using Large Volume Static Headspace / GCMS

hit “esc” to quit hit any key to continue

5973/7100/7032 200cc Injection10 PPB Sulfur Std., ECTD

1

2

3 4

6

5

1. CO2

2. H2S

3. COS

4. MeSH

5. DMS

6. CS2

Page 33: Hit “esc” to quit hit any key to continue Analysis of Beverages using Large Volume Static Headspace / GCMS

hit “esc” to quit hit any key to continue

Si l coCAN

0

20

40

60

80

100

120

140

160

0 20 40 60 80 100

Rel at i ve Humi di ty (%)

Recovery (%)

Ethyl mercaptan

Propyl mercaptan

s- butyl mercaptan

t- Butyl mercaptan

SUMMA CANI STER

0

20

40

60

80

100

120

140

160

0 20 40 60 80 100

Rel at i ve Humi di ty (%)

Recovery (%)

Ethyl mercaptan

Propyl mercaptan

s- butyl mercaptan

t- Butyl mercaptan

2-Day Stability Study of Sulfur Compounds in

Fused Silica Lined vs Electropolished Canisters

7000 / HP5973 Data

Page 34: Hit “esc” to quit hit any key to continue Analysis of Beverages using Large Volume Static Headspace / GCMS

hit “esc” to quit hit any key to continue

5-PointSulfur Gas

Calibrations

H2S

MeSH CS2

DMS

Page 35: Hit “esc” to quit hit any key to continue Analysis of Beverages using Large Volume Static Headspace / GCMS

hit “esc” to quit hit any key to continue

Support for R&D and QA

• Fully Characterize Products down to Olfactory Detection Limits

• Identify Good and Bad “Markers” in product or packaging

• Monitoring on-line (Sequence Looping)

• Monitoring at-line (Sampling Ports)– Solids, Liquids by LVSH– Gases by MiniCan Samplers

Page 36: Hit “esc” to quit hit any key to continue Analysis of Beverages using Large Volume Static Headspace / GCMS

hit “esc” to quit hit any key to continue

Conclusion• Large Volume Static Headspace (LVSH) is a new analytical

tool available to food and beverage Chemists which enhances the detection of headspace components using GC and GCMS analysis.

• Maintaining the sample in a more natural state allows the analysis of “normally occurring” odor and flavor compounds.

• Inert flow paths and reduced temperature trapping increases the range of analytes that can be recovered, including oxygen, nitrogen, and sulfur containing compounds.