hid validation the 3500 series genetic analyzers · hid validation the 3500 series genetic...
TRANSCRIPT
The world leader in serving science Proprietary & Confidential
HID University Seminar Series 2010
HID Validation The 3500 Series Genetic Analyzers
2 Proprietary & Confidential
Outline
• Scope • Materials and Reagents
• Validation Elements • Selection of Validation Data
• Additional Sources of Validation Information • User Bulletin
3 Proprietary & Confidential
Validation Scope Hardware and Reagents
• Instruments • (3) 8 Capillary Instruments (3500) • (3) 24 Capillary Instruments (3500xL)
• Size Standard • GeneScan™ 600 LIZ™ Size Standard v2.0 • GeneScan™ 500 ROX™ Size Standard
• Profiler Plus™ / Cofiler™ do not use normalization
• Polymer and Array • POP-4™ and 36 cm Capillary Array
4 Proprietary & Confidential
Validation Scope Sample Summary
Test Name Samples Input Replicates
Genotype Concordance and Reproducibility
• 40 male and 42 female gDNA samples • 4 racial groups • Controls
1-2 ng total DNA • 1 replicate per sample • 2 injections
Sizing Precision and Accuracy
Allelic ladder 1 µL per well • See below
Sensitivity and Normalization
• 007 control DNA • 3 gDNA samples
• STR: 0.125ng, 0.5ng, 1ng, 6ng
• 5 replicates for DNA samples • 3 replicates for NTC • 4 injections • Minifiler: 0.125ng,
0.25ng, 0.5ng, .75ng
Mixture Analysis • 4 pairs of male and female gDNA • Mixture ratios • 1:0, 1:1, 1:3, 1:5, 1:9, and 0:1
• Recommended concentrations per kit
1 replicate per sample
Resolution • Allelic Ladder • Research material
• 1 µL per well 24 replicates
5 Proprietary & Confidential
Validation Scope AmpFℓSTR™ Kits
• Identifiler™
• Minifiler™
• SGM Plus™ • Yfiler™ • Sinofiler™ • SEfiler Plus™ • Profiler Plus™ • Cofiler™ • Identifiler™ Direct • Identifiler™ Plus • NGM™
The world leader in serving science Proprietary & Confidential
Precision and Accuracy
7 Proprietary & Confidential
Precision and Accuracy Example Configurations for Standard Deviation
Within Injection - 3500
8 data points per mean 12 injections per plate 12 columns 3 Instruments
12 data points per mean 4 injections per plate 8 rows 3 Instruments
Across Injections – 3500xL
8 Proprietary & Confidential
3500 Sizing Precision Study: Within an Injection of Identifiler® Allelic Ladder
9 Proprietary & Confidential
3500 Sizing Precision Study: Across Multiple Injections of Identifiler® Allelic Ladder Samples
96 Replicates of Identifiler® Allelic Ladder
3500 -12 Injections 3500xL - 4 Injections
The world leader in serving science Proprietary & Confidential
Introduction to Normalization
11 Proprietary & Confidential
Sources of CE Variation (3100)
• Numerous optical components, capillary array
• Electrophoresis, injection
• Sample and sample preparation
Raman Intensity, 3100file: INSTRUMENT DATA Mod2B.xls
y = -2.7202x + 6042.8R2 = 0.0059
0
1000
2000
3000
4000
5000
6000
7000
8000
9000
0 20 40 60 80 100 120 140Instrument
A/D
Units
Series1Linear (Series1)
median= 6026.50Stdev = 808.54Average = 6054.11Stdev/ average = 0.13
Almost 2-fold difference in signal from over 100 instruments from the factory floor
12 Proprietary & Confidential
Normalization: What is it? • This feature attenuates
signal variations associated with instrument, capillary array, sample salt load, and injection variability among capillaries and instruments
• Normalization can be applied during primary analysis of the data
Before normalization After normalization
3500-Series Normalization Feature
13 Proprietary & Confidential
• Normalization Target (NT) • The expected average RFU for eleven peaks in the GeneScan 600
LIZ® v2 size standard used for normalization
• Normalization Factor • Average peak height of the subset of peaks in the GeneScan 600
LIZ® v2 size standard used for normalization divided by the Normalization Target
• Normalization Factor Threshold • The passing range for Normalization Factor (default range is 0.3 to
3.0)
Normalization: Definitions
3500-Series normalization calculations
14 Proprietary & Confidential
No NormFactory only
Int Std onlyFactor+IntStd
Injection to Injection
Cap to Cap
Within
Instrument to Instrument
Total
39%
33%
25%
14%
31%
24% 24%
13%
20%19%
5%5%
12%11%
4%4%3%
3%
0%0%
0%
5%
10%
15%
20%
25%
30%
35%
40%
%CV
Levels of Normalization
Sources of Variation
Instrument to Instrument largest source of variation
Internal standard
Normalization greatest reduction in variation
Sources of variation by Normalization Method
The world leader in serving science Proprietary & Confidential
Sensitivity and GeneScan™ 600 LIZ™ Size Standard v2.0
16 Proprietary & Confidential
Sensitivity Study: Average Peak Height across a Range of DNA Input Amounts AmpFℓSTR™ MiniFilerTM PCR Amplification Kit
Normalization Method Factory Only Factory + Internal Std
0.75ng0.5ng0.25ng0.125ng
20000
15000
10000
5000
0
Conc. 0.75ng0.5ng0.25ng0.125ng
20000
15000
10000
5000
0
H1126
Ave
rage
Pea
k H
eigh
t
H1152
H1161 OO7
offon
Normalization
Boxplot of Average Peak Height
Panel variable: Sample
10000
5000
15000
0 0.125ng 0.25ng 0.50ng 0.75ng
10000
5000
15000
0
0.125ng 0.25ng 0.50ng 0.75ng Genomic DNA #2
Control DNA Standard 007
Genomic DNA #1
Genomic DNA #3
Peak
Hei
ght
17 Proprietary & Confidential
Normalization onoff
14000
12000
10000
8000
6000
4000
2000
0
Ave
rage
Pea
k H
eigh
t
Boxplot of Average Peak HeightConc. = 1ng
2000
4000
8000
10000
6000
12000
14000
Internal Standard Normalization
2000 2000
Disabled Enabled
Peak
Hei
ght
Sources of CE Variation (3100) Sources of CE Variation (3100) Sensitivity Study: Average Peak Height by Dye for 1ng DNA input AmpFℓSTR™ Identifiler™ PCR Amplification Kit
18 Proprietary & Confidential
Size Standard Peak Height Consistency Study: Five AmpFℓSTR™ PCR Amplification KitsGeneScan™ 600 LIZ™ Size Standard v2.0
6000
5000
4000
3000
2000
1000
0
Lot 1 Lot 2 Lot 3
GS600 Peak Height
Identifiler™ Identifiler™ Direct Identifiler™ Plus MinifilerTM NGMTM
7000
GS600 Lot Number
Peak
Hei
ght
19 Proprietary & Confidential
% CV
Size Standard Peak Height Consistency · 6 instruments · 3 lots of size standard
Size Standard Peak Height Consistency Study: %CV for 5 Different AmpFℓSTR® Kits GeneScan™ 600 LIZ® Size Standard v2.0
20 Proprietary & Confidential
Internal Standard Normalization: Injection to Injection Peak Height Consistency GeneScan™ 600 LIZ™ Size Standard v2.0
% CV
Normalization Method Factory Only Factory + Internal Std
40
30
20
10
0 Lot 1 Lot 2 Lot 3
Size Standard Peak Height Consistency · 6 instruments · 3 lots of size standard
The world leader in serving science Proprietary & Confidential
Resolution and Baseline Evaluation
22 Proprietary & Confidential
Resolution
• SEfiler Plus™ (SE33 triplet) and Identifiler™ (TH01 9.3/10) Allelic Ladders
• 160 injections per instrument on 6 instruments
• Research STR markers greater than 300 base pairs were also developed with single base pair differences (data not shown)
23 Proprietary & Confidential
Baseline Noise - Developmental Evaluation G5 Average Noise (10x SD)
0
20
40
60
80
100
120
140
160
180
200
FAM VIC NED PET LIZ TOTAL
10x
Ave
rage
Std
. Dev
.
Data generated during the Sensitivity experiment was then used to confirm the level of pull up and the peak amplitude threshold setting.
175 RFU
24 Proprietary & Confidential
Pull-up 360 Samples at 6 ng template
5.0%
20000
4000
8000
12000
16000
24000
The world leader in serving science Proprietary & Confidential
Additional Studies and Validation Documentation
26 Proprietary & Confidential
Identifiler™ Direct
• 82 gDNA, 25 cycles • Three 3500xl and three 3500 instruments. • Mother plate split into 6 daughter plates.
8000
8000
8 Capillary 3500
24 Capillary 3500xL
14000
14000
0
0
27 Proprietary & Confidential
Additional Studies
• Genotype Concordance • 100% Concordant with 3130xL
• GeneMapper™ ID-X v1.2 Evaluation • Normalization and RFID documentation functionality performed as
expected • Average Mx values obtained in the Mixture Study were similar to the
3130xl
28 Proprietary & Confidential
Validation Documentation
• AAFS and ISHI Posters • User Bulletin containing developmental validation • www.thermofisher.com/3500hid
29 Proprietary & Confidential
THANK YOU!!
Ariana Wheaton Senior Market Development Manager Thermo Fisher Scientific Email: [email protected]
© 2016 Thermo Fisher Scien3fic Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scien3fic and its subsidiaries unless otherwise specified. Windows is a trademark of MicrosoG Corpora3on. Products described herein are intended for forensic or paternity use only. Not for use in diagnos3cs procedures. The content provided herein may relate to products that have not been officially released and is subject to change without no3ce.