Dispatch Date : 01-Jan-2018
Page 1
Hepatitis B serology
Distribution : 4202
Laboratory :
Operated by Public Health England NIS - Specialist Microbiology Services 133-155 Waterloo Road Wellington House London SE1 8UG
© Copyright. The data in UK NEQAS reports are confidential. Participants must consult the scheme organiser before quoting data from the scheme. UK NEQAS for Microbiology PO Box 63003 London NW9 1GH Published at 8:18 on Thursday 12 July 2018
Your Score Intended Result Your Report
Specimen 4138 HBsAg positive HBsAg positive 2 Anti-HBc positive Anti-HBc positive 2 Anti-HBc IgM negative Anti-HBc IgM negative 2 HBeAg negative HBeAg negative 2 Anti-HBe positive Anti-HBe positive 2
Specimen 4139 HBsAg positive HBsAg positive 2 Anti-HBc positive Anti-HBc positive 2 Anti-HBc IgM positive Anti-HBc IgM positive 2 HBeAg negative HBeAg negative 2 Anti-HBe positive Anti-HBe positive Not scored
Specimen 4140 HBsAg positive HBsAg positive 2 Anti-HBc positive Anti-HBc positive 2 Anti-HBc IgM negative Anti-HBc IgM negative 2 HBeAg negative HBeAg negative 2 Anti-HBe positive Anti-HBe positive 2
Specimen 4141 HBsAg negative HBsAg negative 2 Anti-HBc negative Anti-HBc negative 2 Anti-HBc IgM negative Anti-HBc IgM negative 2 HBeAg negative HBeAg negative 2 Anti-HBe negative Anti-HBe negative 2
Specimen 4142 HBsAg positive HBsAg positive 2 Anti-HBc positive Anti-HBc positive 2 Anti-HBc IgM negative Anti-HBc IgM negative 2 HBeAg negative HBeAg negative 2 Anti-HBe positive Anti-HBe positive 2
Specimen 4143 HBsAg negative HBsAg negative 2 Anti-HBc negative Anti-HBc negative 2 Anti-HBc IgM negative Anti-HBc IgM negative 2 HBeAg negative HBeAg negative 2 Anti-HBe negative Anti-HBe negative 2
of 14
Dispatch Date : 01-Jan-2018
Page 2
Hepatitis B serology
Distribution : 4202
Laboratory :
Operated by Public Health England NIS - Specialist Microbiology Services 133-155 Waterloo Road Wellington House London SE1 8UG
© Copyright. The data in UK NEQAS reports are confidential. Participants must consult the scheme organiser before quoting data from the scheme. UK NEQAS for Microbiology PO Box 63003 London NW9 1GH Published at 8:18 on Thursday 12 July 2018
Cumulative score information Total number of specimens sent to you for UK NEQAS for Hepatitis B serology over the last 3 distributions is 18 For these distributions specimen numbers 3789 3790 3791 3792 3793 3794 3956 3957 3958 3959 3960 3961 4138 4139 4140 4141 4142 4143 have been sent. Number of specimens reported too late for analysis (not scored) 0 Your cumulative score for the specimen/test combinations that you reported was 174 out of a possible total of 174 The mean score calculated from the reports returned by UK laboratories testing the specimen/test combinations you examined was 171.86 (with a standard error of 2.71). Performance rating Your performance rating for UK NEQAS for Hepatitis B serology (i.e. the number of standard errors by which your cumulative score lies above or below the mean for UK laboratories) is 0.79. A performance rating of more than 1.96 standard errors below the mean indicates possible poor performance. Your performance rating may change if other participants' results are amended. No score penalty is incurred for non return of reports. However non reporting of results may be used as a measure of poor performance.
Comments: Overall performance was excellent with 99.7% of participants reporting correctly for HBsAg, 99.9% for anti-HBc, 99.7% for anti-HBc IgM, 99.6% for HBeAg and 97.9% for anti-HBe. Comment on specimen 4139 Anti-HBe: 169/259 (65.3%) of participants reported a positive anti-HBe result for this specimen, 54/259 (20.8%) reported a negative result and 36/259 (13.9%) reported an equivocal result. Therefore consensus for this specimen was not reached and is not scored. Pre-distribution testing showed a positive anti-HBe result using the Abbott Architect and Biomerieux Vidas platforms with a negative anti-HBe result using the Diasorin ETI-EBK PLUS EIA. Cumulative performance information is presented on this page. In the histograms on page 3 and subsequent pages a maximum of 12 kit results are displayed: these include the most commonly used methods and the method(s) used in your laboratory is indicated by an arrow. The figures in the histograms and those in the overall results tables may differ (1) due to exclusion of kits displayed in the histograms resulting in apparently lower numbers of data sets in the histograms or (2) due to participants using more than one kit resulting in higher numbers of data sets in the histograms. HBsAg quantitative results : 45 participants reported quantitative results expressed in IU/mL using the Abbott Architect quantitative assay (n=14) and the DiaSorin Liaison assay (n=31). Median (lowest-highest) values by assay were as follows: Specimen 4138: 527 IU/mL (18-943) with Abbott Architect; 332 IU/mL (20-1000) with DiaSorin Liaison. Specimen 4139: 6 IU/mL (0-11) with Abbott Architect; 4 IU/mL (0-14) with DiaSorin Liaison. Specimen 4140: 519 IU/mL (16-946) with Abbott Architect; 340 IU/mL (22-1000) with DiaSorin Liaison. Specimen 4142: 4402 IU/mL (250-7751) with Abbott Architect; 3450 IU/mL (150-9643) with DiaSorin Liaison. Specimens 4141 and 4143 were negative for all markers. The quantitative results provided by the participants varied significantly to each other and were also different to the pre- distribution testing results that we have reported. If you use a quantitative assay please specify your method and report your results so we can include them in the analysis. Turn around time: The time taken to report your results was 4-days. This information is provided for your own use and does not form part of your performance assessment. Acknowledgements : We thank colleagues at PHE-NIS-Public Health Laboratory, Southampton; PHE-NIS-Virus Reference Division, Colindale; PHE-NIS-Public Health Laboratory, Manchester; Gloucestershire Royal Hospital; and NHSBT, Colindale for their kind assistance with pre-distribution testing. Enquiries: Repeat specimens can be ordered using the web form or by emailing [email protected]. stating your laboratory number, distribution number and type, and specimen number(s). Pre-distribution test results are available should you experience a technical failure and wish to discuss the results. Any technical queries about this distribution should be emailed to Habib Seyedzadeh at the above address. Report authorised by: Dr Sanjiv Rughooputh, Director
Per
form
ance
Rat
ing
Distribution
Your performance rating over the past 12 distributions Your current performance rating is 0.79
MICPERF2
3467 3517 3572 3677 3729 3787 3901 3954 4013 4086 4141 4202-4
-3
-2
-1
0
1
2A
llU
K
Tot
al s
core
Distribution
Total score you achieved for each of the last 12 distributionsYour total score for this distribution is 58
MICPERF2X
3467 3517 3572 3677 3729 3787 3901 3954 4013 4086 4141 42020
20
40
60
80
of 14
Dispatch Date : 01-Jan-2018
Page 3
Hepatitis B serology
Distribution : 4202
Laboratory :
Operated by Public Health England NIS - Specialist Microbiology Services 133-155 Waterloo Road Wellington House London SE1 8UG
© Copyright. The data in UK NEQAS reports are confidential. Participants must consult the scheme organiser before quoting data from the scheme. UK NEQAS for Microbiology PO Box 63003 London NW9 1GH Published at 8:18 on Thursday 12 July 2018
Six specimens were dispatched with the request to report on hepatitis B surface antigen (HBsAg) status and other hepatitis B markers. The specimens were derived from thrombinised human plasma (serum) donations. Specimens 4138, 4139, 4140 and 4142 were diluted 1:9.7, 1:4, 1:242, 1:6.9 respectively in negative thrombinised human plasma (serum). Specimens 4141 and 4143 were pools of negative thrombinised human plasma (serum).
Specimen : 4138 HBsAg positive (26 IU/mL)
UK (%) All (%)
8 (3.5) 9 (1.6)Biokit : Bioelisa60 (26.1) 76 (13.5)bioMerieux : Vidas49 (21.3) 178 (31.6)Architect HBsAg qual II
7 (3.0) 8 (1.4)Beckman : Access3 (1.3) 11 (2.0)Abbott : Architect5 (2.2) 14 (2.5)BioRad : Monolisa7 (3.0) 16 (2.8)Murex2 (0.9) 17 (3.0)Architect HBsAg quant
10 (4.3) 21 (3.7)Neutralisation (all kits)10 (4.3) 27 (4.8)DiaSorin : Liaison14 (6.1) 35 (6.2)Siemens:ADVIA Centaur47 (20.4) 108 (19.1)Roche
Number of Reports0 40 80 120 160 200
Positive Equivocal Negative
Your Report : HBsAg positive
HBsAg equivocal
140
HBsAg positive
0
Overall Results
388
-1
% Correct
00
100
------------------------------------------------------------UK
3880
Total
1HBsAg negative
140
All
100
------------------------------------------------------------
2
Score
Your Score : 2
Specimen : 4138 HBsAg positive (26 IU/mL) Anti-HBc positive
UK (%) All (%)
5 (3.3) 6 (1.5)Biokit : Bioelisa47 (30.7) 169 (42.5)Architect anti-HBc II
2 (1.3) 4 (1.0)DiaSorin0 (0.0) 4 (1.0)Abbott : Prism0 (0.0) 5 (1.3)OCD : VITROS6 (3.9) 7 (1.8)Beckman : Access1 (0.7) 8 (2.0)BioRad : Monolisa1 (0.7) 10 (2.5)Abbott : Architect7 (4.6) 19 (4.8)DiaSorin : Liaison9 (5.9) 28 (7.0)Siemens:ADVIA Centaur
34 (22.2) 44 (11.1)bioMerieux : Vidas39 (25.5) 80 (20.1)Roche
Number of Reports0 40 80 120 160 200
Positive Equivocal Negative
Your Report : Anti-HBc positive
Anti-HBc equivocal
129
Anti-HBc positive
0
Overall Results
347
-1
% Correct
00
100
------------------------------------------------------------UK
3470
Total
1Anti-HBc negative
129
All
100
------------------------------------------------------------
2
Score
Your Score : 2
Specimen : 4138 HBsAg positive (26 IU/mL) Anti-HBc IgM negative
UK (%) All (%)
30 (41.7) 121 (50.2)Abbott : Architect1 (1.4) 1 (0.4)BioRad : Monolisa0 (0.0) 1 (0.4)Beckman0 (0.0) 1 (0.4)DiaSorin0 (0.0) 2 (0.8)Dia.Pro1 (1.4) 2 (0.8)Beckman : Access0 (0.0) 2 (0.8)Unspecified0 (0.0) 6 (2.5)OCD : VITROS3 (4.2) 13 (5.4)DiaSorin : Liaison3 (4.2) 18 (7.5)Siemens:ADVIA Centaur
24 (33.3) 28 (11.6)bioMerieux : Vidas10 (13.9) 46 (19.1)Roche
Number of Reports0 30 60 90 120 150
Positive Equivocal Negative
Your Report : Anti-HBc IgM negative
Anti-HBc IgM equivocal
73
Anti-HBc IgM positive
244
Overall Results
244
2
% Correct
730
100
------------------------------------------------------------UK
00
Total
1Anti-HBc IgM negative
0
All
100
------------------------------------------------------------
-1
Score
Your Score : 2
Operated by Public Health England NIS - Specialist Microbiology Services 133-155 Waterloo Road Wellington House London SE1 8UG
© Copyright. The data in UK NEQAS reports are confidential. Participants must consult the scheme organiser before quoting data from the scheme. UK NEQAS for Microbiology PO Box 63003 London NW9 1GH Published at 8:18 on Thursday 12 July 2018
of 14
Dispatch Date : 01-Jan-2018
Page 4
Hepatitis B serology
Distribution : 4202
Laboratory :
Six specimens were dispatched with the request to report on hepatitis B surface antigen (HBsAg) status and other hepatitis B markers. The specimens were derived from thrombinised human plasma (serum) donations. Specimens 4138, 4139, 4140 and 4142 were diluted 1:9.7, 1:4, 1:242, 1:6.9 respectively in negative thrombinised human plasma (serum). Specimens 4141 and 4143 were pools of negative thrombinised human plasma (serum).
Specimen : 4138 HBsAg positive (26 IU/mL) HBeAg negative
UK (%) All (%)
24 (31.6) 28 (10.6)bioMerieux : Vidas31 (40.8) 137 (51.7)Abbott : Architect
0 (0.0) 1 (0.4)Dia.Pro0 (0.0) 2 (0.8)BioRad : Monolisa0 (0.0) 2 (0.8)Unspecified0 (0.0) 4 (1.5)OCD : VITROS1 (1.3) 4 (1.5)DiaSorin3 (3.9) 16 (6.0)DiaSorin : Liaison3 (3.9) 19 (7.2)Siemens:ADVIA Centaur
14 (18.4) 52 (19.6)Roche
Number of Reports0 30 60 90 120 150
Positive Equivocal Negative
Your Report : HBeAg negative
HBeAg equivocal
76
HBeAg positive
259
Overall Results
261
2
% Correct
751
99.2
------------------------------------------------------------UK
10
Total
1HBeAg negative
1
All
98.7
------------------------------------------------------------
-1
Score
Your Score : 2
Specimen : 4138 HBsAg positive (26 IU/mL) Anti-HBe positive
UK (%) All (%)
24 (31.6) 26 (10.0)bioMerieux : Vidas31 (40.8) 136 (52.1)Abbott : Architect
0 (0.0) 1 (0.4)Dia.Pro0 (0.0) 1 (0.4)Ortho0 (0.0) 1 (0.4)BioRad : Monolisa0 (0.0) 1 (0.4)Biokit : Bioelisa0 (0.0) 2 (0.8)Unspecified0 (0.0) 3 (1.1)OCD : VITROS1 (1.3) 4 (1.5)DiaSorin3 (3.9) 16 (6.1)DiaSorin : Liaison3 (3.9) 19 (7.3)Siemens:ADVIA Centaur
14 (18.4) 51 (19.5)Roche
Number of Reports0 30 60 90 120 150
Positive Equivocal Negative
Your Report : Anti-HBe positive
Anti-HBe equivocal
76
Anti-HBe positive
1
Overall Results
259
-1
% Correct
01
99.2
------------------------------------------------------------UK
2570
Total
1Anti-HBe negative
76
All
100
------------------------------------------------------------
2
Score
Your Score : 2
Operated by Public Health England NIS - Specialist Microbiology Services 133-155 Waterloo Road Wellington House London SE1 8UG
© Copyright. The data in UK NEQAS reports are confidential. Participants must consult the scheme organiser before quoting data from the scheme. UK NEQAS for Microbiology PO Box 63003 London NW9 1GH Published at 8:18 on Thursday 12 July 2018
of 14
Dispatch Date : 01-Jan-2018
Page 5
Hepatitis B serology
Distribution : 4202
Laboratory :
Six specimens were dispatched with the request to report on hepatitis B surface antigen (HBsAg) status and other hepatitis B markers. The specimens were derived from thrombinised human plasma (serum) donations. Specimens 4138, 4139, 4140 and 4142 were diluted 1:9.7, 1:4, 1:242, 1:6.9 respectively in negative thrombinised human plasma (serum). Specimens 4141 and 4143 were pools of negative thrombinised human plasma (serum).
Specimen : 4139 HBsAg positive (0.3 IU/mL)
UK (%) All (%)
8 (3.5) 9 (1.6)Biokit : Bioelisa60 (26.4) 77 (13.7)bioMerieux : Vidas49 (21.6) 179 (31.7)Architect HBsAg qual II
7 (3.1) 8 (1.4)Beckman : Access3 (1.3) 11 (2.0)Abbott : Architect5 (2.2) 14 (2.5)BioRad : Monolisa7 (3.1) 17 (3.0)Murex2 (0.9) 17 (3.0)Architect HBsAg quant9 (4.0) 20 (3.5)Neutralisation (all kits)
11 (4.8) 29 (5.1)DiaSorin : Liaison14 (6.2) 35 (6.2)Siemens:ADVIA Centaur46 (20.3) 107 (19.0)Roche
Number of Reports0 40 80 120 160 200
Positive Equivocal Negative
Your Report : HBsAg positive
HBsAg equivocal
140
HBsAg positive
0
Overall Results
388
-1
% Correct
02
99.5
------------------------------------------------------------UK
3861
Total
1HBsAg negative
139
All
99.3
------------------------------------------------------------
2
Score
Your Score : 2
Specimen : 4139 HBsAg positive (0.3 IU/mL) Anti-HBc positive
UK (%) All (%)
5 (3.2) 6 (1.5)Biokit : Bioelisa47 (30.3) 169 (42.0)Architect anti-HBc II
2 (1.3) 4 (1.0)DiaSorin0 (0.0) 4 (1.0)Abbott : Prism0 (0.0) 5 (1.2)OCD : VITROS6 (3.9) 7 (1.7)Beckman : Access1 (0.6) 8 (2.0)BioRad : Monolisa1 (0.6) 11 (2.7)Abbott : Architect7 (4.5) 19 (4.7)DiaSorin : Liaison9 (5.8) 28 (7.0)Siemens:ADVIA Centaur
35 (22.6) 45 (11.2)bioMerieux : Vidas40 (25.8) 81 (20.1)Roche
Number of Reports0 40 80 120 160 200
Positive Equivocal Negative
Your Report : Anti-HBc positive
Anti-HBc equivocal
130
Anti-HBc positive
0
Overall Results
348
-1
% Correct
00
100
------------------------------------------------------------UK
3480
Total
1Anti-HBc negative
130
All
100
------------------------------------------------------------
2
Score
Your Score : 2
Specimen : 4139 HBsAg positive (0.3 IU/mL) Anti-HBc IgM positive
UK (%) All (%)
30 (41.7) 121 (50.2)Abbott : Architect1 (1.4) 1 (0.4)BioRad : Monolisa0 (0.0) 1 (0.4)Beckman0 (0.0) 1 (0.4)DiaSorin0 (0.0) 2 (0.8)Dia.Pro1 (1.4) 2 (0.8)Beckman : Access0 (0.0) 2 (0.8)Unspecified0 (0.0) 6 (2.5)OCD : VITROS3 (4.2) 13 (5.4)DiaSorin : Liaison3 (4.2) 18 (7.5)Siemens:ADVIA Centaur
24 (33.3) 28 (11.6)bioMerieux : Vidas10 (13.9) 46 (19.1)Roche
Number of Reports0 30 60 90 120 150
Positive Equivocal Negative
Your Report : Anti-HBc IgM positive
Anti-HBc IgM equivocal
73
Anti-HBc IgM positive
1
Overall Results
244
-1
% Correct
10
99.6
------------------------------------------------------------UK
2430
Total
1Anti-HBc IgM negative
72
All
98.6
------------------------------------------------------------
2
Score
Your Score : 2
Operated by Public Health England NIS - Specialist Microbiology Services 133-155 Waterloo Road Wellington House London SE1 8UG
© Copyright. The data in UK NEQAS reports are confidential. Participants must consult the scheme organiser before quoting data from the scheme. UK NEQAS for Microbiology PO Box 63003 London NW9 1GH Published at 8:18 on Thursday 12 July 2018
of 14
Dispatch Date : 01-Jan-2018
Page 6
Hepatitis B serology
Distribution : 4202
Laboratory :
Six specimens were dispatched with the request to report on hepatitis B surface antigen (HBsAg) status and other hepatitis B markers. The specimens were derived from thrombinised human plasma (serum) donations. Specimens 4138, 4139, 4140 and 4142 were diluted 1:9.7, 1:4, 1:242, 1:6.9 respectively in negative thrombinised human plasma (serum). Specimens 4141 and 4143 were pools of negative thrombinised human plasma (serum).
Specimen : 4139 HBsAg positive (0.3 IU/mL) HBeAg negative
UK (%) All (%)
25 (32.5) 29 (10.9)bioMerieux : Vidas31 (40.3) 137 (51.5)Abbott : Architect
0 (0.0) 1 (0.4)Dia.Pro0 (0.0) 2 (0.8)BioRad : Monolisa0 (0.0) 2 (0.8)Unspecified0 (0.0) 4 (1.5)OCD : VITROS1 (1.3) 4 (1.5)DiaSorin3 (3.9) 16 (6.0)DiaSorin : Liaison3 (3.9) 19 (7.1)Siemens:ADVIA Centaur
14 (18.2) 52 (19.5)Roche
Number of Reports0 30 60 90 120 150
Positive Equivocal Negative
Your Report : HBeAg negative
HBeAg equivocal
76
HBeAg positive
258
Overall Results
261
2
% Correct
750
98.9
------------------------------------------------------------UK
30
Total
1HBeAg negative
1
All
98.7
------------------------------------------------------------
-1
Score
Your Score : 2
Specimen : 4139 HBsAg positive (0.3 IU/mL) Anti-HBe positive
UK (%) All (%)
25 (32.5) 27 (10.3)bioMerieux : Vidas31 (40.3) 136 (51.7)Abbott : Architect
0 (0.0) 1 (0.4)Dia.Pro0 (0.0) 1 (0.4)Ortho0 (0.0) 1 (0.4)BioRad : Monolisa0 (0.0) 1 (0.4)Biokit : Bioelisa0 (0.0) 3 (1.1)OCD : VITROS0 (0.0) 3 (1.1)Unspecified1 (1.3) 4 (1.5)DiaSorin3 (3.9) 16 (6.1)DiaSorin : Liaison3 (3.9) 19 (7.2)Siemens:ADVIA Centaur
14 (18.2) 51 (19.4)Roche
Number of Reports0 30 60 90 120 150
Positive Equivocal Negative
Your Report : Anti-HBe positive
Anti-HBe equivocal
76
Anti-HBe positive
54
Overall Results
259
NS
% Correct
1036
65.3
------------------------------------------------------------UK
16918
Total
NSAnti-HBe negative
48
All
63.2
------------------------------------------------------------
NS
Score
Your Score : Not scored
Operated by Public Health England NIS - Specialist Microbiology Services 133-155 Waterloo Road Wellington House London SE1 8UG
© Copyright. The data in UK NEQAS reports are confidential. Participants must consult the scheme organiser before quoting data from the scheme. UK NEQAS for Microbiology PO Box 63003 London NW9 1GH Published at 8:18 on Thursday 12 July 2018
of 14
Dispatch Date : 01-Jan-2018
Page 7
Hepatitis B serology
Distribution : 4202
Laboratory :
Six specimens were dispatched with the request to report on hepatitis B surface antigen (HBsAg) status and other hepatitis B markers. The specimens were derived from thrombinised human plasma (serum) donations. Specimens 4138, 4139, 4140 and 4142 were diluted 1:9.7, 1:4, 1:242, 1:6.9 respectively in negative thrombinised human plasma (serum). Specimens 4141 and 4143 were pools of negative thrombinised human plasma (serum).
Specimen : 4140 HBsAg positive (22 IU/mL)
UK (%) All (%)
8 (3.5) 9 (1.6)Biokit : Bioelisa60 (26.3) 76 (13.5)bioMerieux : Vidas49 (21.5) 178 (31.7)Architect HBsAg qual II
7 (3.1) 8 (1.4)Beckman : Access3 (1.3) 11 (2.0)Abbott : Architect5 (2.2) 14 (2.5)BioRad : Monolisa7 (3.1) 16 (2.8)Murex2 (0.9) 17 (3.0)Architect HBsAg quant
10 (4.4) 21 (3.7)Neutralisation (all kits)10 (4.4) 27 (4.8)DiaSorin : Liaison14 (6.1) 35 (6.2)Siemens:ADVIA Centaur47 (20.6) 108 (19.2)Roche
Number of Reports0 40 80 120 160 200
Positive Equivocal Negative
Your Report : HBsAg positive
HBsAg equivocal
140
HBsAg positive
4
Overall Results
388
-1
% Correct
00
99.0
------------------------------------------------------------UK
3840
Total
1HBsAg negative
140
All
100
------------------------------------------------------------
2
Score
Your Score : 2
Specimen : 4140 HBsAg positive (22 IU/mL) Anti-HBc positive
UK (%) All (%)
5 (3.2) 6 (1.5)Biokit : Bioelisa47 (30.3) 169 (42.3)Architect anti-HBc II
2 (1.3) 4 (1.0)DiaSorin0 (0.0) 4 (1.0)Abbott : Prism0 (0.0) 5 (1.3)OCD : VITROS6 (3.9) 7 (1.8)Beckman : Access1 (0.6) 8 (2.0)BioRad : Monolisa1 (0.6) 11 (2.8)Abbott : Architect7 (4.5) 18 (4.5)DiaSorin : Liaison9 (5.8) 28 (7.0)Siemens:ADVIA Centaur
35 (22.6) 44 (11.0)bioMerieux : Vidas40 (25.8) 81 (20.3)Roche
Number of Reports0 40 80 120 160 200
Positive Equivocal Negative
Your Report : Anti-HBc positive
Anti-HBc equivocal
130
Anti-HBc positive
0
Overall Results
347
-1
% Correct
00
100
------------------------------------------------------------UK
3470
Total
1Anti-HBc negative
130
All
100
------------------------------------------------------------
2
Score
Your Score : 2
Specimen : 4140 HBsAg positive (22 IU/mL) Anti-HBc IgM negative
UK (%) All (%)
30 (41.7) 121 (50.2)Abbott : Architect1 (1.4) 1 (0.4)BioRad : Monolisa0 (0.0) 1 (0.4)Beckman0 (0.0) 1 (0.4)DiaSorin0 (0.0) 2 (0.8)Dia.Pro1 (1.4) 2 (0.8)Beckman : Access0 (0.0) 2 (0.8)Unspecified0 (0.0) 6 (2.5)OCD : VITROS3 (4.2) 12 (5.0)DiaSorin : Liaison3 (4.2) 18 (7.5)Siemens:ADVIA Centaur
24 (33.3) 28 (11.6)bioMerieux : Vidas10 (13.9) 46 (19.1)Roche
Number of Reports0 30 60 90 120 150
Positive Equivocal Negative
Your Report : Anti-HBc IgM negative
Anti-HBc IgM equivocal
73
Anti-HBc IgM positive
243
Overall Results
245
2
% Correct
730
99.2
------------------------------------------------------------UK
20
Total
1Anti-HBc IgM negative
0
All
100
------------------------------------------------------------
-1
Score
Your Score : 2
Operated by Public Health England NIS - Specialist Microbiology Services 133-155 Waterloo Road Wellington House London SE1 8UG
© Copyright. The data in UK NEQAS reports are confidential. Participants must consult the scheme organiser before quoting data from the scheme. UK NEQAS for Microbiology PO Box 63003 London NW9 1GH Published at 8:18 on Thursday 12 July 2018
of 14
Dispatch Date : 01-Jan-2018
Page 8
Hepatitis B serology
Distribution : 4202
Laboratory :
Six specimens were dispatched with the request to report on hepatitis B surface antigen (HBsAg) status and other hepatitis B markers. The specimens were derived from thrombinised human plasma (serum) donations. Specimens 4138, 4139, 4140 and 4142 were diluted 1:9.7, 1:4, 1:242, 1:6.9 respectively in negative thrombinised human plasma (serum). Specimens 4141 and 4143 were pools of negative thrombinised human plasma (serum).
Specimen : 4140 HBsAg positive (22 IU/mL) HBeAg negative
UK (%) All (%)
24 (31.6) 28 (10.5)bioMerieux : Vidas31 (40.8) 137 (51.5)Abbott : Architect
0 (0.0) 1 (0.4)GBC EASE0 (0.0) 1 (0.4)Dia.Pro0 (0.0) 2 (0.8)BioRad : Monolisa0 (0.0) 2 (0.8)Unspecified0 (0.0) 4 (1.5)OCD : VITROS1 (1.3) 4 (1.5)DiaSorin3 (3.9) 16 (6.0)DiaSorin : Liaison3 (3.9) 19 (7.1)Siemens:ADVIA Centaur
14 (18.4) 52 (19.5)Roche
Number of Reports0 30 60 90 120 150
Positive Equivocal Negative
Your Report : HBeAg negative
HBeAg equivocal
76
HBeAg positive
261
Overall Results
261
2
% Correct
760
100
------------------------------------------------------------UK
00
Total
1HBeAg negative
0
All
100
------------------------------------------------------------
-1
Score
Your Score : 2
Specimen : 4140 HBsAg positive (22 IU/mL) Anti-HBe positive
UK (%) All (%)
24 (31.6) 27 (10.3)bioMerieux : Vidas31 (40.8) 136 (51.9)Abbott : Architect
0 (0.0) 1 (0.4)Dia.Pro0 (0.0) 1 (0.4)Ortho0 (0.0) 1 (0.4)BioRad : Monolisa0 (0.0) 1 (0.4)Biokit : Bioelisa0 (0.0) 2 (0.8)Unspecified0 (0.0) 3 (1.1)OCD : VITROS1 (1.3) 4 (1.5)DiaSorin3 (3.9) 16 (6.1)DiaSorin : Liaison3 (3.9) 19 (7.3)Siemens:ADVIA Centaur
14 (18.4) 51 (19.5)Roche
Number of Reports0 30 60 90 120 150
Positive Equivocal Negative
Your Report : Anti-HBe positive
Anti-HBe equivocal
76
Anti-HBe positive
20
Overall Results
259
-1
% Correct
15
90.3
------------------------------------------------------------UK
2342
Total
1Anti-HBe negative
73
All
96.1
------------------------------------------------------------
2
Score
Your Score : 2
Operated by Public Health England NIS - Specialist Microbiology Services 133-155 Waterloo Road Wellington House London SE1 8UG
© Copyright. The data in UK NEQAS reports are confidential. Participants must consult the scheme organiser before quoting data from the scheme. UK NEQAS for Microbiology PO Box 63003 London NW9 1GH Published at 8:18 on Thursday 12 July 2018
of 14
Dispatch Date : 01-Jan-2018
Page 9
Hepatitis B serology
Distribution : 4202
Laboratory :
Six specimens were dispatched with the request to report on hepatitis B surface antigen (HBsAg) status and other hepatitis B markers. The specimens were derived from thrombinised human plasma (serum) donations. Specimens 4138, 4139, 4140 and 4142 were diluted 1:9.7, 1:4, 1:242, 1:6.9 respectively in negative thrombinised human plasma (serum). Specimens 4141 and 4143 were pools of negative thrombinised human plasma (serum).
Specimen : 4141 HBsAg negative
UK (%) All (%)
8 (4.2) 9 (1.8)Biokit : Bioelisa40 (21.1) 54 (11.0)bioMerieux : Vidas48 (25.3) 173 (35.2)Architect HBsAg qual II
0 (0.0) 7 (1.4)OCD : VITROS7 (3.7) 8 (1.6)Beckman : Access2 (1.1) 9 (1.8)Abbott : Architect5 (2.6) 12 (2.4)Murex2 (1.1) 13 (2.6)Architect HBsAg quant5 (2.6) 14 (2.8)BioRad : Monolisa9 (4.7) 23 (4.7)DiaSorin : Liaison
13 (6.8) 34 (6.9)Siemens:ADVIA Centaur45 (23.7) 104 (21.1)Roche
Number of Reports0 40 80 120 160 200
Positive Equivocal Negative
Your Report : HBsAg negative
HBsAg equivocal
140
HBsAg positive
387
Overall Results
388
2
% Correct
1400
99.7
------------------------------------------------------------UK
10
Total
1HBsAg negative
0
All
100
------------------------------------------------------------
-1
Score
Your Score : 2
Specimen : 4141 HBsAg negative Anti-HBc negative
UK (%) All (%)
5 (3.7) 6 (1.6)Biokit : Bioelisa40 (29.6) 158 (42.2)Architect anti-HBc II
0 (0.0) 3 (0.8)Unspecified0 (0.0) 4 (1.1)Abbott : Prism0 (0.0) 5 (1.3)OCD : VITROS6 (4.4) 7 (1.9)Beckman : Access1 (0.7) 8 (2.1)BioRad : Monolisa1 (0.7) 9 (2.4)Abbott : Architect6 (4.4) 18 (4.8)DiaSorin : Liaison8 (5.9) 26 (7.0)Siemens:ADVIA Centaur
30 (22.2) 39 (10.4)bioMerieux : Vidas36 (26.7) 77 (20.6)Roche
Number of Reports0 40 80 120 160 200
Positive Equivocal Negative
Your Report : Anti-HBc negative
Anti-HBc equivocal
114
Anti-HBc positive
327
Overall Results
327
2
% Correct
1140
100
------------------------------------------------------------UK
00
Total
1Anti-HBc negative
0
All
100
------------------------------------------------------------
-1
Score
Your Score : 2
Specimen : 4141 HBsAg negative Anti-HBc IgM negative
UK (%) All (%)
16 (40.0) 94 (49.2)Abbott : Architect0 (0.0) 1 (0.5)Dia.Pro0 (0.0) 1 (0.5)Beckman0 (0.0) 1 (0.5)DiaSorin1 (2.5) 2 (1.0)Beckman : Access0 (0.0) 2 (1.0)Unspecified0 (0.0) 5 (2.6)OCD : VITROS1 (2.5) 11 (5.8)DiaSorin : Liaison2 (5.0) 15 (7.9)Siemens:ADVIA Centaur
14 (35.0) 17 (8.9)bioMerieux : Vidas6 (15.0) 42 (22.0)Roche
Number of Reports0 20 40 60 80 100
Positive Equivocal Negative
Your Report : Anti-HBc IgM negative
Anti-HBc IgM equivocal
41
Anti-HBc IgM positive
194
Overall Results
194
2
% Correct
410
100
------------------------------------------------------------UK
00
Total
1Anti-HBc IgM negative
0
All
100
------------------------------------------------------------
-1
Score
Your Score : 2
Operated by Public Health England NIS - Specialist Microbiology Services 133-155 Waterloo Road Wellington House London SE1 8UG
© Copyright. The data in UK NEQAS reports are confidential. Participants must consult the scheme organiser before quoting data from the scheme. UK NEQAS for Microbiology PO Box 63003 London NW9 1GH Published at 8:18 on Thursday 12 July 2018
of 14
Dispatch Date : 01-Jan-2018
Page 10
Hepatitis B serology
Distribution : 4202
Laboratory :
Six specimens were dispatched with the request to report on hepatitis B surface antigen (HBsAg) status and other hepatitis B markers. The specimens were derived from thrombinised human plasma (serum) donations. Specimens 4138, 4139, 4140 and 4142 were diluted 1:9.7, 1:4, 1:242, 1:6.9 respectively in negative thrombinised human plasma (serum). Specimens 4141 and 4143 were pools of negative thrombinised human plasma (serum).
Specimen : 4141 HBsAg negative HBeAg negative
UK (%) All (%)
14 (35.0) 16 (7.7)bioMerieux : Vidas14 (35.0) 104 (50.0)Abbott : Architect
0 (0.0) 1 (0.5)GBC EASE0 (0.0) 2 (1.0)BioRad : Monolisa0 (0.0) 2 (1.0)Unspecified0 (0.0) 3 (1.4)OCD : VITROS0 (0.0) 3 (1.4)DiaSorin2 (5.0) 15 (7.2)DiaSorin : Liaison2 (5.0) 16 (7.7)Siemens:ADVIA Centaur8 (20.0) 46 (22.1)Roche
Number of Reports0 30 60 90 120 150
Positive Equivocal Negative
Your Report : HBeAg negative
HBeAg equivocal
41
HBeAg positive
206
Overall Results
206
2
% Correct
410
100
------------------------------------------------------------UK
00
Total
1HBeAg negative
0
All
100
------------------------------------------------------------
-1
Score
Your Score : 2
Specimen : 4141 HBsAg negative Anti-HBe negative
UK (%) All (%)
13 (33.3) 14 (6.8)bioMerieux : Vidas14 (35.9) 105 (51.2)Abbott : Architect
0 (0.0) 1 (0.5)BioRad : Monolisa0 (0.0) 1 (0.5)Biokit : Bioelisa0 (0.0) 2 (1.0)Unspecified0 (0.0) 3 (1.5)OCD : VITROS0 (0.0) 3 (1.5)DiaSorin2 (5.1) 15 (7.3)DiaSorin : Liaison2 (5.1) 16 (7.8)Siemens:ADVIA Centaur8 (20.5) 45 (22.0)Roche
Number of Reports0 30 60 90 120 150
Positive Equivocal Negative
Your Report : Anti-HBe negative
Anti-HBe equivocal
40
Anti-HBe positive
204
Overall Results
204
2
% Correct
400
100
------------------------------------------------------------UK
00
Total
1Anti-HBe negative
0
All
100
------------------------------------------------------------
-1
Score
Your Score : 2
Operated by Public Health England NIS - Specialist Microbiology Services 133-155 Waterloo Road Wellington House London SE1 8UG
© Copyright. The data in UK NEQAS reports are confidential. Participants must consult the scheme organiser before quoting data from the scheme. UK NEQAS for Microbiology PO Box 63003 London NW9 1GH Published at 8:18 on Thursday 12 July 2018
of 14
Dispatch Date : 01-Jan-2018
Page 11
Hepatitis B serology
Distribution : 4202
Laboratory :
Six specimens were dispatched with the request to report on hepatitis B surface antigen (HBsAg) status and other hepatitis B markers. The specimens were derived from thrombinised human plasma (serum) donations. Specimens 4138, 4139, 4140 and 4142 were diluted 1:9.7, 1:4, 1:242, 1:6.9 respectively in negative thrombinised human plasma (serum). Specimens 4141 and 4143 were pools of negative thrombinised human plasma (serum).
Specimen : 4142 HBsAg positive (1798 IU/mL)
UK (%) All (%)
8 (3.5) 9 (1.6)Biokit : Bioelisa60 (26.5) 76 (13.6)bioMerieux : Vidas49 (21.7) 177 (31.7)Architect HBsAg qual II
7 (3.1) 8 (1.4)Beckman : Access3 (1.3) 11 (2.0)Abbott : Architect5 (2.2) 14 (2.5)BioRad : Monolisa6 (2.7) 15 (2.7)Murex2 (0.9) 17 (3.0)Architect HBsAg quant
10 (4.4) 21 (3.8)Neutralisation (all kits)9 (4.0) 26 (4.7)DiaSorin : Liaison
14 (6.2) 35 (6.3)Siemens:ADVIA Centaur47 (20.8) 108 (19.3)Roche
Number of Reports0 40 80 120 160 200
Positive Equivocal Negative
Your Report : HBsAg positive
HBsAg equivocal
140
HBsAg positive
0
Overall Results
387
-1
% Correct
00
100
------------------------------------------------------------UK
3870
Total
1HBsAg negative
140
All
100
------------------------------------------------------------
2
Score
Your Score : 2
Specimen : 4142 HBsAg positive (1798 IU/mL) Anti-HBc positive
UK (%) All (%)
5 (3.2) 6 (1.5)Biokit : Bioelisa47 (30.5) 169 (42.4)Architect anti-HBc II
2 (1.3) 4 (1.0)DiaSorin0 (0.0) 4 (1.0)Abbott : Prism0 (0.0) 5 (1.3)OCD : VITROS6 (3.9) 7 (1.8)Beckman : Access1 (0.6) 8 (2.0)BioRad : Monolisa1 (0.6) 11 (2.8)Abbott : Architect7 (4.5) 18 (4.5)DiaSorin : Liaison9 (5.8) 28 (7.0)Siemens:ADVIA Centaur
34 (22.1) 43 (10.8)bioMerieux : Vidas40 (26.0) 81 (20.3)Roche
Number of Reports0 40 80 120 160 200
Positive Equivocal Negative
Your Report : Anti-HBc positive
Anti-HBc equivocal
130
Anti-HBc positive
1
Overall Results
347
-1
% Correct
00
99.7
------------------------------------------------------------UK
3460
Total
1Anti-HBc negative
130
All
100
------------------------------------------------------------
2
Score
Your Score : 2
Specimen : 4142 HBsAg positive (1798 IU/mL) Anti-HBc IgM negative
UK (%) All (%)
30 (41.7) 120 (49.8)Abbott : Architect1 (1.4) 1 (0.4)BioRad : Monolisa0 (0.0) 1 (0.4)Beckman0 (0.0) 1 (0.4)DiaSorin0 (0.0) 2 (0.8)Dia.Pro1 (1.4) 2 (0.8)Beckman : Access0 (0.0) 2 (0.8)Unspecified0 (0.0) 6 (2.5)OCD : VITROS3 (4.2) 13 (5.4)DiaSorin : Liaison3 (4.2) 18 (7.5)Siemens:ADVIA Centaur
24 (33.3) 28 (11.6)bioMerieux : Vidas10 (13.9) 46 (19.1)Roche
Number of Reports0 30 60 90 120 150
Positive Equivocal Negative
Your Report : Anti-HBc IgM negative
Anti-HBc IgM equivocal
73
Anti-HBc IgM positive
243
Overall Results
245
2
% Correct
730
99.2
------------------------------------------------------------UK
20
Total
1Anti-HBc IgM negative
0
All
100
------------------------------------------------------------
-1
Score
Your Score : 2
Operated by Public Health England NIS - Specialist Microbiology Services 133-155 Waterloo Road Wellington House London SE1 8UG
© Copyright. The data in UK NEQAS reports are confidential. Participants must consult the scheme organiser before quoting data from the scheme. UK NEQAS for Microbiology PO Box 63003 London NW9 1GH Published at 8:18 on Thursday 12 July 2018
of 14
Dispatch Date : 01-Jan-2018
Page 12
Hepatitis B serology
Distribution : 4202
Laboratory :
Six specimens were dispatched with the request to report on hepatitis B surface antigen (HBsAg) status and other hepatitis B markers. The specimens were derived from thrombinised human plasma (serum) donations. Specimens 4138, 4139, 4140 and 4142 were diluted 1:9.7, 1:4, 1:242, 1:6.9 respectively in negative thrombinised human plasma (serum). Specimens 4141 and 4143 were pools of negative thrombinised human plasma (serum).
Specimen : 4142 HBsAg positive (1798 IU/mL) HBeAg negative
UK (%) All (%)
24 (31.6) 28 (10.6)bioMerieux : Vidas31 (40.8) 136 (51.3)Abbott : Architect
0 (0.0) 1 (0.4)GBC EASE0 (0.0) 1 (0.4)Dia.Pro0 (0.0) 2 (0.8)BioRad : Monolisa0 (0.0) 2 (0.8)Unspecified0 (0.0) 4 (1.5)OCD : VITROS1 (1.3) 4 (1.5)DiaSorin3 (3.9) 16 (6.0)DiaSorin : Liaison3 (3.9) 19 (7.2)Siemens:ADVIA Centaur
14 (18.4) 52 (19.6)Roche
Number of Reports0 30 60 90 120 150
Positive Equivocal Negative
Your Report : HBeAg negative
HBeAg equivocal
76
HBeAg positive
260
Overall Results
260
2
% Correct
760
100
------------------------------------------------------------UK
00
Total
1HBeAg negative
0
All
100
------------------------------------------------------------
-1
Score
Your Score : 2
Specimen : 4142 HBsAg positive (1798 IU/mL) Anti-HBe positive
UK (%) All (%)
24 (31.6) 26 (10.0)bioMerieux : Vidas31 (40.8) 136 (52.1)Abbott : Architect
0 (0.0) 1 (0.4)Dia.Pro0 (0.0) 1 (0.4)Ortho0 (0.0) 1 (0.4)BioRad : Monolisa0 (0.0) 1 (0.4)Biokit : Bioelisa0 (0.0) 2 (0.8)Unspecified0 (0.0) 3 (1.1)OCD : VITROS1 (1.3) 4 (1.5)DiaSorin3 (3.9) 16 (6.1)DiaSorin : Liaison3 (3.9) 19 (7.3)Siemens:ADVIA Centaur
14 (18.4) 51 (19.5)Roche
Number of Reports0 30 60 90 120 150
Positive Equivocal Negative
Your Report : Anti-HBe positive
Anti-HBe equivocal
76
Anti-HBe positive
0
Overall Results
259
-1
% Correct
00
100
------------------------------------------------------------UK
2590
Total
1Anti-HBe negative
76
All
100
------------------------------------------------------------
2
Score
Your Score : 2
Operated by Public Health England NIS - Specialist Microbiology Services 133-155 Waterloo Road Wellington House London SE1 8UG
© Copyright. The data in UK NEQAS reports are confidential. Participants must consult the scheme organiser before quoting data from the scheme. UK NEQAS for Microbiology PO Box 63003 London NW9 1GH Published at 8:18 on Thursday 12 July 2018
of 14
Dispatch Date : 01-Jan-2018
Page 13
Hepatitis B serology
Distribution : 4202
Laboratory :
Six specimens were dispatched with the request to report on hepatitis B surface antigen (HBsAg) status and other hepatitis B markers. The specimens were derived from thrombinised human plasma (serum) donations. Specimens 4138, 4139, 4140 and 4142 were diluted 1:9.7, 1:4, 1:242, 1:6.9 respectively in negative thrombinised human plasma (serum). Specimens 4141 and 4143 were pools of negative thrombinised human plasma (serum).
Specimen : 4143 HBsAg negative
UK (%) All (%)
7 (3.7) 8 (1.6)Biokit : Bioelisa39 (20.7) 53 (10.8)bioMerieux : Vidas48 (25.5) 172 (35.2)Architect HBsAg qual II
0 (0.0) 7 (1.4)OCD : VITROS7 (3.7) 8 (1.6)Beckman : Access2 (1.1) 9 (1.8)Abbott : Architect5 (2.7) 12 (2.5)Murex2 (1.1) 13 (2.7)Architect HBsAg quant5 (2.7) 14 (2.9)BioRad : Monolisa9 (4.8) 23 (4.7)DiaSorin : Liaison
13 (6.9) 34 (7.0)Siemens:ADVIA Centaur45 (23.9) 104 (21.3)Roche
Number of Reports0 40 80 120 160 200
Positive Equivocal Negative
Your Report : HBsAg negative
HBsAg equivocal
139
HBsAg positive
385
Overall Results
386
2
% Correct
1390
99.7
------------------------------------------------------------UK
10
Total
1HBsAg negative
0
All
100
------------------------------------------------------------
-1
Score
Your Score : 2
Specimen : 4143 HBsAg negative Anti-HBc negative
UK (%) All (%)
5 (3.7) 6 (1.6)Biokit : Bioelisa40 (29.4) 158 (42.1)Architect anti-HBc II
0 (0.0) 3 (0.8)Unspecified0 (0.0) 4 (1.1)Abbott : Prism0 (0.0) 5 (1.3)OCD : VITROS6 (4.4) 7 (1.9)Beckman : Access1 (0.7) 8 (2.1)BioRad : Monolisa1 (0.7) 10 (2.7)Abbott : Architect6 (4.4) 18 (4.8)DiaSorin : Liaison8 (5.9) 26 (6.9)Siemens:ADVIA Centaur
31 (22.8) 40 (10.7)bioMerieux : Vidas36 (26.5) 76 (20.3)Roche
Number of Reports0 40 80 120 160 200
Positive Equivocal Negative
Your Report : Anti-HBc negative
Anti-HBc equivocal
114
Anti-HBc positive
325
Overall Results
326
2
% Correct
1140
99.7
------------------------------------------------------------UK
10
Total
1Anti-HBc negative
0
All
100
------------------------------------------------------------
-1
Score
Your Score : 2
Specimen : 4143 HBsAg negative Anti-HBc IgM negative
UK (%) All (%)
16 (40.0) 91 (48.4)Abbott : Architect0 (0.0) 1 (0.5)Dia.Pro0 (0.0) 1 (0.5)Beckman0 (0.0) 1 (0.5)DiaSorin1 (2.5) 2 (1.1)Beckman : Access0 (0.0) 2 (1.1)Unspecified0 (0.0) 5 (2.7)OCD : VITROS1 (2.5) 11 (5.9)DiaSorin : Liaison2 (5.0) 15 (8.0)Siemens:ADVIA Centaur
14 (35.0) 17 (9.0)bioMerieux : Vidas6 (15.0) 42 (22.3)Roche
Number of Reports0 20 40 60 80 100
Positive Equivocal Negative
Your Report : Anti-HBc IgM negative
Anti-HBc IgM equivocal
41
Anti-HBc IgM positive
191
Overall Results
191
2
% Correct
410
100
------------------------------------------------------------UK
00
Total
1Anti-HBc IgM negative
0
All
100
------------------------------------------------------------
-1
Score
Your Score : 2
Operated by Public Health England NIS - Specialist Microbiology Services 133-155 Waterloo Road Wellington House London SE1 8UG
© Copyright. The data in UK NEQAS reports are confidential. Participants must consult the scheme organiser before quoting data from the scheme. UK NEQAS for Microbiology PO Box 63003 London NW9 1GH Published at 8:18 on Thursday 12 July 2018
of 14
Dispatch Date : 01-Jan-2018
Page 14
Hepatitis B serology
Distribution : 4202
Laboratory :
Six specimens were dispatched with the request to report on hepatitis B surface antigen (HBsAg) status and other hepatitis B markers. The specimens were derived from thrombinised human plasma (serum) donations. Specimens 4138, 4139, 4140 and 4142 were diluted 1:9.7, 1:4, 1:242, 1:6.9 respectively in negative thrombinised human plasma (serum). Specimens 4141 and 4143 were pools of negative thrombinised human plasma (serum).
Specimen : 4143 HBsAg negative HBeAg negative
UK (%) All (%)
14 (35.0) 16 (7.7)bioMerieux : Vidas14 (35.0) 104 (50.0)Abbott : Architect
0 (0.0) 1 (0.5)GBC EASE0 (0.0) 2 (1.0)BioRad : Monolisa0 (0.0) 2 (1.0)Unspecified0 (0.0) 3 (1.4)OCD : VITROS0 (0.0) 3 (1.4)DiaSorin2 (5.0) 15 (7.2)DiaSorin : Liaison2 (5.0) 16 (7.7)Siemens:ADVIA Centaur8 (20.0) 46 (22.1)Roche
Number of Reports0 30 60 90 120 150
Positive Equivocal Negative
Your Report : HBeAg negative
HBeAg equivocal
41
HBeAg positive
205
Overall Results
206
2
% Correct
410
99.5
------------------------------------------------------------UK
10
Total
1HBeAg negative
0
All
100
------------------------------------------------------------
-1
Score
Your Score : 2
Specimen : 4143 HBsAg negative Anti-HBe negative
UK (%) All (%)
14 (35.0) 15 (7.3)bioMerieux : Vidas14 (35.0) 105 (51.0)Abbott : Architect
0 (0.0) 1 (0.5)BioRad : Monolisa0 (0.0) 1 (0.5)Biokit : Bioelisa0 (0.0) 2 (1.0)Unspecified0 (0.0) 3 (1.5)OCD : VITROS0 (0.0) 3 (1.5)DiaSorin2 (5.0) 15 (7.3)DiaSorin : Liaison2 (5.0) 16 (7.8)Siemens:ADVIA Centaur8 (20.0) 45 (21.8)Roche
Number of Reports0 30 60 90 120 150
Positive Equivocal Negative
Your Report : Anti-HBe negative
Anti-HBe equivocal
41
Anti-HBe positive
205
Overall Results
205
2
% Correct
410
100
------------------------------------------------------------UK
00
Total
1Anti-HBe negative
0
All
100
------------------------------------------------------------
-1
Score
Your Score : 2
Operated by Public Health England NIS - Specialist Microbiology Services 133-155 Waterloo Road Wellington House London SE1 8UG
© Copyright. The data in UK NEQAS reports are confidential. Participants must consult the scheme organiser before quoting data from the scheme. UK NEQAS for Microbiology PO Box 63003 London NW9 1GH Published at 8:18 on Thursday 12 July 2018
of 14