glast lat project calibration & analysis meeting - august 29, 2005 benoît lott gamma-ray large...

27
GLAST LAT Project Calibration & Analysis Meeting - August 29, 2005 Benoît Lott Gamma-ray Large Gamma-ray Large Area Space Area Space Telescope Telescope Response of the GLAST LAT Calorimeter to Relativistic Heavy http://www.cenbg.in2p3.fr/ftp/astropart/glast/NIM_quenching.pdf

Post on 19-Dec-2015

214 views

Category:

Documents


0 download

TRANSCRIPT

GLAST LAT Project Calibration & Analysis Meeting - August 29, 2005

Benoît Lott

Gamma-ray Large Gamma-ray Large Area Space Area Space TelescopeTelescope

Response of the GLAST LAT Calorimeter to Relativistic Heavy Ions

http://www.cenbg.in2p3.fr/ftp/astropart/glast/NIM_quenching.pdf

GLAST LAT Project Calibration & Analysis Meeting - August 29, 2005

Benoît Lott

Energy calibration and response monitoring of the calorimeter: energy deposits of cosmic-ray heavy ions (C,N,O,Si,Fe)

Introduction

magnetic cutoff + “logarithmic-rise” domain:well-defined peaks in deposited energy per log

Fe: peak at 7.7 GeV <Emax> =25 GeV for a 300 GeV ray

GLAST LAT Project Calibration & Analysis Meeting - August 29, 2005

Benoît Lott

Motivation: “quenching”

GLAST LAT Project Calibration & Analysis Meeting - August 29, 2005

Benoît Lott

Temporal dependence of light ouput

p: fast+slow comp.ions: fast comp. only

low energy

GLAST LAT Project Calibration & Analysis Meeting - August 29, 2005

Benoît Lott

Quenching factor: definition

Quenching Factor= Emeas/Edep (normalized to 1 for MIPs)

GLAST LAT Project Calibration & Analysis Meeting - August 29, 2005

Benoît Lott

EM ASIC includes preamp, shaping amplifier (3.5 s) , 12-bit ADCAmplitude sampled through a Track&Hold stage at fixed delay time after the trigger signal

minical: 8 layers of 3 log each, with discrete electronics similar as EM, but with a 12-bit peak-sensing ADC

Fragment Separator (FRS) 1.7 GeV/nucleon Ni on Be target:4 dipole magnets: selection of particles according to magnetic rigidity (p/Q) “Cocktail” beam of A/Z=2 ionsDetection system includes MWPCs for positionMUSIC (ionisation chamber, Z measurement), plastic scintillation detectors ( A measurement via TOF, trigger)

Three data acquisition systems, with a common trigger signal.

Detector characteristics

GLAST LAT Project Calibration & Analysis Meeting - August 29, 2005

Benoît Lott

MUSIC calibration

GLAST LAT Project Calibration & Analysis Meeting - August 29, 2005

Benoît Lott

EM calibration

- non-linearity correction- pedestal subtraction- LEX8 absolute calibration- LEX1, HEX8, HEX1: cross calibration of adjacent ranges

GLAST LAT Project Calibration & Analysis Meeting - August 29, 2005

Benoît Lott

E (MeV)

Counts Independent calibration performed by Frederic Pironusing Gaussian+Landaufits

Perfect consistency

GLAST LAT Project Calibration & Analysis Meeting - August 29, 2005

Benoît Lott

minical calibration

Same as for EM.

Proton calibration checked against that obtained with:- 22Na g-rays- 20 GeV muons at CERN

In both cases, excellent consistency was found.

GLAST LAT Project Calibration & Analysis Meeting - August 29, 2005

Benoît Lott

blue: data red: Geant4

deposited energy (MeV)deposited energy (MeV)

50 GeV e+ 1.5 X080 GeV e+ 1.5 X0

CERN data: deposited energy distributions

GLAST LAT Project Calibration & Analysis Meeting - August 29, 2005

Benoît Lott

Quenching determination

Ion selection with MUSIC(ionization chamber)

Gaussian fits of ionization peaks independently for two log ends

<Emeas>= (<E+>+ <E->)/2

Quenching factor = <Emeas>/<EGEANT>

GLAST LAT Project Calibration & Analysis Meeting - August 29, 2005

Benoît Lott

Longitudinal dispersion by the spectrometer

GLAST LAT Project Calibration & Analysis Meeting - August 29, 2005

Benoît Lott

Calculated deposited energy

GLAST LAT Project Calibration & Analysis Meeting - August 29, 2005

Benoît Lott

Quenching factor for C: 1.23 “antiquenching”

Quenching factors

Sasha found a similar valuein an independent analysis

GLAST LAT Project Calibration & Analysis Meeting - August 29, 2005

Benoît Lott

Quenching factors

GLAST LAT Project Calibration & Analysis Meeting - August 29, 2005

Benoît Lott

-particle measurement

GLAST LAT Project Calibration & Analysis Meeting - August 29, 2005

Benoît Lott

Energy resolution

Another difference between MIPs and heavy ions

0.6 MeV/12 MeV = 5% extra width for MIPs

much less (< 1%) for heavy ions

GLAST LAT Project Calibration & Analysis Meeting - August 29, 2005

Benoît Lott

Compilation of quenching factors

GLAST LAT Project Calibration & Analysis Meeting - August 29, 2005

Benoît Lott

Possible interpretation of the antiquenching effect, based on an extrapolation on what is known at low-energy For MIPs, part of the energy goes into a slow scintillation component, that could be absent for heavy ions. This slow component is essentially filtered out by our electronics, resulting in the measured energy being lower for MIPs than for heavy ions.

GLAST LAT Project Calibration & Analysis Meeting - August 29, 2005

Benoît Lott

Scintillation components

measured with a standard phototube

GLAST LAT Project Calibration & Analysis Meeting - August 29, 2005

Benoît Lott

Hamamatsu S3590

GLAST LAT Project Calibration & Analysis Meeting - August 29, 2005

Benoît Lott

Scintillation components

GLAST LAT Project Calibration & Analysis Meeting - August 29, 2005

Benoît Lott

Ion selection with MUSIC(ionization chamber)

The ionization peaks correspondto non-interacting ions.

Loss to reactions

GLAST LAT Project Calibration & Analysis Meeting - August 29, 2005

Benoît Lott

Loss to reactions

GLAST LAT Project Calibration & Analysis Meeting - August 29, 2005

Benoît Lott

Charge-changing cross sections

GLAST LAT Project Calibration & Analysis Meeting - August 29, 2005

Benoît Lott

Conclusion

The CsI response to relativistic ions has been measured.

The “antiquenching” effect was unexpected, but is compatible with the trend observed at lower energy.

Data measured up to 1.7 GeV/nucleon: 90% of CR ions will have E<10 GeV/nucleon.Negligible change of quenching factors between 1 GeV and 1.7 GeV/nucleon (dE/dx changes by 7%).The assumption that they remain constant up to 10 GeV/nucleonis reasonable. Overall uncertainty ~ 3%.

The scintillation components observed with a standard phototube are similar for MIPs and Carbon ions (worth repeating with a red-sensitive phototube).