Built-in 5.7" TFT LCD Color Display Stand-alone or PC- connected operation 20 Analog Channels Standard, Expandable to 200 Input-to-output and channel-to-channel Isolation USB and Ethernet PC Interfaces GL800 midi Logger Features DATAQ Instruments, Inc. • 241 Springside Drive • Akron, Ohio 44333 • Tel: 330-668-1444 • Email: [email protected]m • www.dataq.com With its color monitor and internal memory the GL800 is a compact, lightweight, multi-channel data logger that provides 20 standard analog measurement channels, expandable to 200. The GL800 is equipped with a large internal flash memory to allow the direct capture of acquired data, and its built-in USB port may be used to connect any standard USB flash drive for incremental capacity. Alternatively, the USB or the integral Ethernet port may be connected to a PC to allow data upload in real time or from memory, as well as local or remote configuration and real time data acquisition. The Ethernet feature includes WEB and FTP server functions, which allows monitoring from a remote location as well as data transfer. Wide Voltage Measurement Range Each GL800 analog channel can measure from 20 mV to 50 VFS across eleven programmable measurement ranges. Full Electrical Isolation Per Channel Each analog GL800 channel is electrically isolated from all others and from instrument ground to allow accurate and safe measurements in industrial applications where ground potential differences are common. Voltage, Current, and Temperature Measurement Functions Use the GL800 to measure voltages, currents, 4-20 mA process currents, as well as thermocouple- and RTD-based temperatures. Four Unique ‘Pulse’ Inputs for Discrete Measurements The GL800 provides discrete input channels that can be used for counting and rotational speed measurement applications. Or program the discrete inputs as simple logic level input channels. Four Alarm Outputs Program the GL800 to trigger its open- collector outputs as a function of analog input signal level judgment, pulse judgment, or logic pattern. Wide Sample Interval Selections Sample intervals can be programmed to be one of sixteen values ranging from 100 ms to one hour. Bright TFT LCD Color Display The focal point of the GL800 is its built-in 5.7-inch color display that allows real time trending, data review, and complete instrument configuration. Engineering Units Scaling Each GL800 channel allows up to four break points to be programmed for accurate scaling into meaningful units like psi, grams, newtons, gallons per minute, etc. Flexible Triggering Options The GL800 allows data capture to be started or stopped based upon signal level, an external event, date/time, alarm, duration, or Boolean channel combinations. Analog signal triggers can be programmed based upon level and window tests: above threshold, below threshold, inside window, or outside window. Real Time and Post-recorded Calculations The GL800 may be programmed to calculate average value, peak value, minimum value, and rms. Flexible Power Requirements Power the GL800 from its provided international AC adaptor, from an optional built-in battery pack, or from any 9 to 24 VDC source using an optional cable. PC Connectivity via USB or Ethernet Allows data transfer to the PC either in real time or from the GL800’s memory. Also allows complete configuration of the GL800. PC Software Bundle Included The GL800 includes a Windows application for direct capture, measurement, and monitoring of GL800 data. In addition to waveform and data value capture and display, the application can export data to an Excel file for further analysis and report creation. The software includes built-in help for quick reference.

Upload: ngodang

Post on 13-Feb-2018

222 views

Category:

Documents


0 download

TRANSCRIPT

Built-in 5.7" TFT LCD Color Display

Stand-alone or PC-connected operation

20 Analog Channels Standard, Expandable to

200

Input-to-output and channel-to-channel

Isolation

USB and Ethernet PC Interfaces

GL800 midi Logger

Features

DATAQ Instruments, Inc. • 241 Springside Drive • Akron, Ohio 44333 • Tel: 330-668-1444 • Email: [email protected] • www.dataq.com

With its color monitor and internal memory the GL800 is a compact, lightweight, multi-channel data logger that provides 20 standard analog measurement channels, expandable to 200. The GL800 is equipped with a large internal flash memory to allow the direct capture of acquired data, and its built-in USB port may be used to connect any standard USB flash drive for incremental capacity. Alternatively, the USB or the integral Ethernet port may be connected to a PC to allow data upload in real time or from memory, as well as local or remote configuration and real time data acquisition. The Ethernet feature includes WEB and FTP server functions, which allows monitoring from a remote location as well as data transfer.

Wide Voltage Measurement RangeEach GL800 analog channel can measure from 20 mV to 50 VFS across eleven programmable measurement ranges.

Full Electrical Isolation Per ChannelEach analog GL800 channel is electrically isolated from all others and from instrument ground to allow accurate and safe measurements in industrial applications where ground potential differences are common.

Voltage, Current, and Temperature Measurement FunctionsUse the GL800 to measure voltages, currents, 4-20 mA process currents, as well as thermocouple- and RTD-based temperatures.

Four Unique ‘Pulse’ Inputs for Discrete MeasurementsThe GL800 provides discrete input channels that can be used for counting and rotational speed measurement applications. Or program the discrete inputs as simple logic level input channels.

Four Alarm OutputsProgram the GL800 to trigger its open-collector outputs as a function of analog input signal level judgment, pulse judgment, or logic pattern.

Wide Sample Interval SelectionsSample intervals can be programmed to be one of sixteen values ranging from 100 ms to one hour.

Bright TFT LCD Color DisplayThe focal point of the GL800 is its built-in 5.7-inch color display that allows real time trending, data review, and complete instrument configuration.

Engineering Units ScalingEach GL800 channel allows up to four break points to be programmed for accurate scaling into meaningful units like psi, grams, newtons, gallons per minute, etc.

Flexible Triggering OptionsThe GL800 allows data capture to be started or stopped based upon signal level, an external event, date/time, alarm, duration, or Boolean channel combinations. Analog signal triggers can be programmed based upon level and window tests: above threshold, below threshold, inside window, or outside window.

Real Time and Post-recorded CalculationsThe GL800 may be programmed to calculate average value, peak value, minimum value, and rms.

Flexible Power RequirementsPower the GL800 from its provided international AC adaptor, from an optional built-in battery pack, or from any 9 to 24 VDC source using an optional cable.

PC Connectivity via USB or EthernetAllows data transfer to the PC either in real time or from the GL800’s memory. Also allows complete configuration of the GL800.

PC Software Bundle IncludedThe GL800 includes a Windows application for direct capture, measurement, and monitoring of GL800 data. In addition to waveform and data value capture and display, the application can export data to an Excel file for further analysis and report creation. The software includes built-in help for quick reference.

GL800 Display, I/O, and Control Overview

330-668-1444 2 www.dataq.com

www.dataq.com 3 330-668-1444

GL800 Analog Input Circuit and Measurement RangesEach GL800 analog input channel features electrical isolation using a photo MOS relay switching method to maintain safe and accurate measurements in demanding industrial environments.

+

500kΩ

50Ω

0.05µF

-

0.05µF500kΩ

50Ω Channel Switching

Relay

Range Maximum SPAN Minimum SPAN Minimum Resolution20mV -22.000 to +22.000mV 0.200mV 0.001mV50mV -55.00 to +55.00mV 0.50mV 0.01mV100mV -110.00 to +110.00mV 1.00mV 0.01mV200mV -220.00 to +220.00mV 2.00mV 0.01mV500mV -550.0 to +550.0mV 5.0mV 0.1mV

1V -1.1000 to +1.1000V 0.0100V 0.0001V2V -2.2000 to +2.2000V 0.0200V 0.0001V5V -5.500 to +5.500V 0.050V 0.001V

10V -11.000 to +11.000V 0.100V 0.001V20V -22.000 to +22.000V 0.200V 0.001V50V -55.00 to +55.00V 0.50V 0.01V

Voltage Measurement Ranges per Channel

Range Maximum SPAN Minimum SPAN Minimum Resolution1-5 V -5.500 to +5.500V 0.050V 0.001V

Process Current Measurement (with external 250-ohm resistor)

Range Maximum SPAN Minimum SPAN Measurement Range Minimum ResolutionK

-270 to +2000°C 50°C

-200 to +1370°C

0.1°C

J -200 to +1100°CT -200 to +400°CR 0 to +1600°CE -200 to +900°CB +600 to +1920°CS 0 to +1760°CN 0 to +1300°CW 0 to +2315°C

PT100 -200 to +850°CJPT100 -200 to +500°CPT1000 -200 to +500°C

Temperature Measurement Ranges per Channel

Range Maximum SPAN Minimum SPAN Minimum Resolution0 to 100% 0 to +110% 1.0% 0.1%

Optional Humidity Measurement Range

330-668-1444 4 www.dataq.com

Typical GL800 Analog Signal ConnectionsTerminal Configuration

+ ......................................................... High -voltage terminal (terminal for high voltage signals)- ......................................................... Low-voltage terminal (terminal for low-voltage input signals)b ......................................................... Dedicated terminal when connection resistance temperature detector

Signal Types and Typical Connections

www.dataq.com 5 330-668-1444

Program the GL800 for Real-World Trigger ConditionsThe GL800 can adapt to just about any trigger condition you might encounter. Data recording can be stopped or started as a function of analog signal level, a definable alarm condition, an external event, or specific date and time. Beyond initiat-ing a data capture cycle, the GL800 can also be programmed to set a digital output to flag an external alarm condition. And after a trigger condition is executed you can program the GL800 to automatically rearm itself to wait for another trigger event, or stop entirely. You can even program the GL800 to detect and alarm on a thermocouple burnout condition. Here’s a summary of the GL800’s trigger and alarm features:

Setting Selections Available

Start side source setting

Off, Level, Alarm, External Input, DateLevel: Mode, Level, CombinationAlarm: Alarm port numberExternal input: noneDate: Date, Time

Stop side source setting

Off, Level, Alarm, External Input, Date, TimeLevel: Mode, Level, CombinationAlarm: Alarm port numberExternal input: noneDate: Date, TimeTime: Duration

Repeated capturing On, OffAlarm level settings Mode, Level, OutputAlarm hold On, OffSend burnout alarm On, Off

GL800 Trigger and Alarm Overview

GL800 Trigger ModesAbove Threshold

Outside Window

Inside Window

Below Threshold

CH1

Measurement starts (alarm generated)

CH1

Measurement starts (alarm generated)

CH1

Measurement starts (alarm generated)

CH1

Measurement starts (alarm generated)

330-668-1444 6 www.dataq.com

GL800 Logic, Pulse, Alarm, and External Trigger ConnectionsThe Logic Alarm Cable model B-513 provides access to the GL800’s discrete and pulse inputs, external trigger input, and alarm outputs. The cable is two meters in length, and is purchased separately.

Item DescriptionNumber of input channels 4Input voltage range 0 to +24V max (single-ended ground input)Threshold level Approx. +2.5VHysteresis Approx. 0.5V (+2.5 to +3 V)

Logic/Pulse Specifications

Item DescriptionNumber of input channels 1Input voltage range 0 to +24V max (single-ended ground input)Threshold level Approx. +2.5VHysteresis Approx. 0.5V (+2.5 to +3 V)

Trigger Input Specifications

Item DescriptionNumber of output channels 4Output format Open collector output

+5V, 10KΩ pull-up resistanceContact capacity 5 V to 24 V, 100mA or below

Alarm Output Specifications

Circuit Example of Relay Drive by Alarm OutputAlarm Output Circuit

GL800 +5V

10KΩ

+24VExternal Device

This relay turns ON when alarm is generated

www.dataq.com 7 330-668-1444

Flexible Computer Interfaces, and an Optional Battery PackThe GL800 provides the added benefit of PC connectivity to either a USB or Ethernet port. Both interfaces are standard, and the USB port doubles as a method to expand the GL-800’s internal 12 MB memory to a maximum 2 GB capacity us-ing standard USB flash memory. When either interface is connected to a PC you can upload measurement protocols to the GL800, monitor acquired data in real time, or download previously acquired data. The GL800’s Ethernet interface offers the advantage of allowing the GL800 to function as a Web server, allowing the GL800 to be manipulated from any web browser such as Firefox or Internet Explorer. Finally, an optional battery pack may be added to the GL800 to allow power-independent data recording whenever and wherever it’s required.

USB Flash Drive Memory Expansion USB-to-PC Interface Connection

Optional B-517 Battery PackEthernet Interface Connection

Ethernet Deployment ExamplesDirectly to a PC using a cross-over cable (100 meter maximum length)

Over a local area network (LAN) using hubs or switches(unlimited length)

330-668-1444 8 www.dataq.com

GL800 Analog Channel Expansion and Extension OptionsAdding more flexibility to where and how the GL800 can be deployed, the analog input screw terminal unit may be detached from the GL800 and located up to 23 inches away using the optional B-537 Extension Kit. If you need channel expansion beyond the 20 channels provided by the base GL800, then use one or more B-538 Expansion Kits (plus ONE B-537 Extension Kit) to expand your channel count in 20-channel increments to a maximum of two-hundred.

The Terminal Unit may be removed from the GL800...

…and Relocated up to 23 inches away using the B-537 Extension Kit

then use a B-538 Expansion kit to expand GL800 channel count to as many as 200

20 Channels 40 Channels 100 Channels 200 ChannelsGL-800 1 1 1 1

B-537 Extension Kit - 1 1 1B-538 Expansion Kit - 1 4 9

Expansion Ordering Chart

GL800 Display Quick-lookThe GL800’s keyboard and display are key components you’ll use for any typical data recording session. The display is a full color TFT LCD (thin-film transistor liquid crystal display), the same technology used in modern flat-panel televi-sions. The display measures 5.7 inches diagonally, and offers 320 x 240 pixels of bright, clear, high contrast resolution. The GL800’s keyboard allows full access to the instrument’s menu system as viewed through its display. Navigation is straightforward and intuitive using the keyboard’s navigation and ENTER keys that form the center of the array. Other keys support special operations that are clearly annotated.

GL800 Display Close-up and Modes

Keyboard Close-up

www.dataq.com 9 330-668-1444

GL800 Included and Optional AccessoriesIncluded PC Software

Item DescriptionCompatible OS Windows 2000, Windows XPFunctions Main unit control, real time data capture, data conversionMain unit settings Input, memory, alarm, triggerAllowed connection up to 10Number of channels per connection 200 ch maximumMaximum number of channels 500 ch maximumSettings AMP, data, trigger/alarm, report, othersCaptured data (CSV, Binary) Real time data, Memory data, USB memory dataDisplay Analog waveforms, logic waveforms, pulse waveforms, digital valuesDisplay modes Y-T, X-Y, Digital, Meter, ReportFile conversion Between cursors. All data.Monitor functions Alarm monitor enables sending of e-mail to the specified addressDual-screen function Displays the current data alongside past dataReport function Automatic creation of daily or monthly filesStatistic/History Displays max, min, and average values during measurement

Item DescriptionQuick Start Guide GL800-UM-8xxCD-ROM User’s manual, application softwareAC adapter 100 to 240 VAC, 50/60 Hz, power supply cord for each area

Included Accessories

Other Optional AccessoriesItem Option Description

DC power cable B-514 Bare tips (2 m)

Extension terminal base set B-537 Extension terminal base unit, cable

20 channel extension terminal set B-538 20 terminals, extension terminal base unit, connection plate, screws

Logic alarm cable B-513 2m, bare tips

Carrying Case B-536 Durable carrying case designed specifically for the GL800330-668-1444 10 www.dataq.com

Optional Battery Pack model B-517Item Description

Capacity 7.2V/2200mAhBattery type Lithium secondary batteryRunning time(up to 2 battery packscan be mounted)

<When LCD is ON> 1 battery pack: MAX brightness: approx. 2.5 hours; MIN brightness: approx. 3 hours 2 battery packs: MAX brightness: approx. 5 hours; MIN brightness: approx. 6 hours<When LCD is OFF> 1 battery pack: approx. 4.5 hours; 2 battery packs: approx. 9 hours

Charging method Mount in the main unitTime required for charging 1 battery pack: approx. 4 hours; 2 battery packs: approx. 8 hoursSwitchover in the case of power failure

Because the battery is used together with the AC adapter, the power supply will be switched automatically to the battery in the event of power failure. The AC adapter is the primary power source

Operating environment 15 to 40°C

Other functions When battery is running low, file is closed automatically (when captured to USB or internal memory).Remaining amount indicator.

Optional Humidity Sensor model B-530Item Description

Alowable temperature range -25 to 80°CAllowable Humidity Range 0 to 100%Relative humidity measurement accuracy ±3% RH (5 to 98% RH at 25°C)Response time 15 s (90% response when membrane filter installed)Sensor output 0 to 1 VDCSensor power source 5 to 16 VDCPower consumption approx. 4mAExternal dimensions 14mm × 80 mm (excluding cable)Cable length 3m

GL800 External Dimensions

330-668-1444 11 www.dataq.com

Analog Channel SpecificationsNumber of inputs: 20 channels (maximum 200 channels with exten-

sion unit)Input terminal type: M3 screw type terminals

Input method: Photo MOS relay scanning systemAll channels isolated, balanced input

Scan speed 0.1s/10 ch maximumMeasurement Ranges

Voltage: Voltage: 20, 50, 100, 200, 500 mV; 1, 2, 5, 10, 20, 50 V; 1-5 V F.S.

Temperature: Thermocouples: K, J, E, T, R, S, B, N, W (WRe5-26)Resistance temperature detector: Pt100, JPt100, Pt1000 (IEC751)

Humidity: Humidity: 0 to 100% (voltage 0 V to 1 V scaling conversion) *with B-530 (option)

Measurement accuracy*Voltage: 0.1% of Full Scale

Temperature(Thermocouple): TC Measurement Temperature

Range (°C)MeasurementAccuracy (°C)

R/S

0 ≤ Ts ≤ 100100 < Ts ≤ 300

R: 300 < Ts ≤ 1600S: 300 < Ts ≤ 1760

±5.2±3.0

±(0.05% of rdg +2.0)±(0.05% of rdg +2.0)

B 400 ≤ Ts ≤ 600600 < Ts ≤ 1820

±3.5±(0.05% of rdg +2.0)

K -200 ≤ Ts ≤ -100-100 < Ts ≤ 1370

±(0.05% of rdg +2.0)±(0.05% of rdg +1.0)

E -200 ≤ Ts ≤ -100-100 < Ts ≤ 800

±(0.05% of rdg +2.0)±(0.05% of rdg +1.0)

T -200 ≤ Ts ≤ -100-100 < Ts ≤ 400

±(0.1% of rdg +1.5)±(0.1% of rdg +0.5)

J-200 ≤ Ts ≤ -100-100 < Ts ≤ 100100 < Ts ≤ 1100

±2.7±1.7

±(0.05% of rdg +1.0)N 0 ≤ Ts ≤ 1300 ±(0.1% of rdg +1.0)W 0 ≤ Ts ≤ 2315 ±(0.1% of rdg +1.5)

Reference contact compensation accuracy: ±0.5°C

Resistance Temperature Detector: Type Measurement Tem-

perature Range (°C)Applied Current

MeasurementAccuracy (°C)

Pt100 -200 to 850 (FS=1050) 1mA ±1.0Jpt100 -200 to 500 (FS=700) 1mA ±0.8Pt1000 -200 to 500 (FS=700) 0.2mA ±0.8

* 23°C ±3°C when 30 minutes have elapsed after the power was switched on (filter On (10), 1 s/20 ch sampling, GND connected).

Reference contactcompensation accuracy:

Internal/External switching

A/D converter: 14 bitsTemperature coefficient: Gain: 0.01% of F.S./ °C

Input resistance: 1 MΩ ±5%Allowable signal source

resistance:Within 300Ω

Maximum permissible input voltage:

Between +/- terminals: Between input terminal/input terminal; Between input terminal/GND: 60 Vp-p

Withstand voltage: Between input terminal/input terminal : 1 min. at 350 Vp-pBetween input terminal/GND : 1 min. at 350 Vp-p

Insulation resistance: Between Input terminal/GND: At least 50 MΩ (at 500 VDC)Common mode rejection

ratio:At least 90 dB (50.60 Hz; signal source 300 Ω or less)

Noise: At least 48 dB (with +/- terminals shorted)Filter: Off, 2, 5, 10, 20, 40

Filter operation is on a moving average basis. The average value of the set sampling count is used.

Integral TFT LCD DisplayDisplay: 5.7-inch TFT color LCD (QVGA: 320 x 240 dots)

Displayed languages: Japanese, English, OthersBacklight life: 50,000 hrs (when brightness is down to 50%),

depends on operation environmentBacklight: Screensaver function (10, 30 sec., 1, 2, 5, 10, 30, 60 min.)

GL800 Specifications

330-668-1444 10 www.dataq.com

Overall SpecificationsNumber of analog inputs: 1 unit: 20 channels

Extension unit: Maximum 200 channelsExternal input/output: Trigger input, Logic input 4 channels or Pulse

input 4 channels, Alarm outputPC interface: Ethernet (10Base-T/100Base-TX), USB (high

speed supported) provided as standard features.Internal memory devices: Internal memory: approx. 12 MB

USB memory slot (FullSpeed supported) standardData backup functions: Setup conditions: EEPROM; Clock: lithium

secondary batteryOperating Environment: 0 to 45°C, 5 to 85% RH (15 to 40°C when using

batteries)Withstand voltage: Between each input channel and GND terminal: 1

minute at 350Vp-pBetween each input terminal: 1 minute at 350Vp-p

Power supply: AC adapter: 100 to 240 VAC, 50/60 HzDC input: 8.5 to 24 VDCBattery pack (option): 7.2 VDC (2200 mAh), 2 packs mountable

Power Consumption: AC Power consumption (when AC adapter is used)

Condition NormalConsumption

Consumption during battery recharge

LCD on 16 VA 28 VAScreensaver on 11 Va 22 VA

DC Power consumptionDC

Voltage Condition NormalConsumption

Consumption duringbattery recharge

+24V LCD on 0.3 VA 0.7 VA+24V Screensaver on 0.2 VA 0.6 VA+12V LCD on 0.55 VA Can’t Recharge+12V Screensaver on 0.3 VA Can’t Recharge+8.5V LCD on 0.8 VA Can’t Recharge+8.5V Screensaver on 0.45 VA Can’t Recharge

Note: normal status is when LCD brightness is set to MAXExternal Dimensions: 232 × 152 × 50 mm

Weight: 990g (excluding AC adapter and battery)Vibration-tested

conditions:Equivalent to automobile parts Type 1 Category A classification

Overall Functional SpecificationsDisplay Screen: Waveform screen + Digital screen, Expanded

Waveform screen, Digital screen + Calculation Display screen (can be key-toggled)

Sampling interval: 100 ms/10 ch maximum; 100, 200, 500 ms; 1, 2, 5, 10, 20, 30 s; 1, 2, 5, 10, 20, 30 min; 1 h

EU Scaling function: 4 points can be set for each channel.Review function: Data replay during data capture

Data save functions: Capture to internal memory, capture to USB memory, setup data can be saved, copy of data screen saved.

Calculation: Types of operation: Average value, peak value, max/min value, RMS.2 operations can be set simultaneouslyMethod: real time and between cursors specified (during data replay)Note: Calculation results are displayed in Digital screen + Calculation Display screen

Search functions: Function: Search the captured data for the required number of pointsSearch type: Channel pulse, logic, alarm search

Annotation inputfunction:

Function: A comment can be input for each channel.Inputtable characters: Alpha numericsNumber of characters: 11 (8 displayed)

GL800 Specifications (continued)Internal Memory Devices

Memory capacity: Internal memory: approx. 12MB Flash memoryUSB memory: Max 2 GB

Memory contents: Setup conditions, measured data, screen copy

Trigger Function SpecificationsRepeat trigger: Off, On

Trigger types: Start: Data capture starts when a trigger is generated.Stop: Data capture stops when a trigger is generated.

Trigger conditions: Start: Off, Level, External, Time, AlarmStop: Off, Level, External, Time, Duration, Alarm

Alarm judgment modes: Analog : H (↑), L (↓), Window In, Window OutLogic : H (↑), L (↓)Pulse : H (↑), L (↓), Window In, Window Out

Channel combination: OR, AND

PC InterfaceInterface types: Ethernet (10Base-T/100Base-TX

USB (HighSpeed)Software functions: Data transfer to PC (real time, memory)

PC control of the GL800Ethernet functions: Web server function: Displays GL800s screen

image on web broser, operation of GL800FTP server function : Transfers and deletes files from internal memory and USB memory

Real time data transfer speed:

100 msec/10 ch maximum

Discrete I/O SpecificationsInput/Output types: Trigger input (1 ch); Logic input (4 ch) or Pulse

input (4 ch); Alarm output (4 ch)Note: Switch between Logic and Pulse

Input specifications: Maximum input voltage: 0 to +24 V (single-ended ground input)Input threshold voltage: Approx. +2.5VHysteresis: Approx. 0.5 V (+2.5 to +3 V)

Alarm outputspecifications:

Output format: Open collector output (5 V, 10 KΩ pull-up resistance), Contact capacity 5 V to 24 V, 100 mA or belowOutput conditions: Level judgment, window judg-ment, logic pattern judgment, pulse judgment

Pulse inputRevolutions mode

(engines, etc):Function: Counts the number of pulses per second; enables them to be converted to rpms.Spans: 50, 500, 5000, 50 k, 500 k, 5 M, 50 M, 500 M PRM/F.S.

Counts mode(electric meters, etc.):

Function: Displays a count of the number of pulses for each sampling interval from the start of measurement.Spans: 50, 500, 5000, 50 k, 500 k, 5 M, 50 M, 500 M C/F.S.

Inst. mode: Function: Counts the number of pulses for each sampling interval. Resets the count value after each sampling interval.Spans: 50, 500, 5000, 50 k, 500 k, 5 M, 50 M, 500 M C/F.S.

Maximum number of pulse inputs:

Counts, Inst. modes: 50k/sampling intervalRevolution counts mode: 50k/s

DATAQ, the DATAQ logo and WinDaq are registered trademarks of DATAQ Instruments, Inc. All rights reserved. Copyright © 2009 DATAQ Instruments, Inc.The information on this data sheet is subject to change without notice.

DATAQ Instruments, Inc.241 Springside DriveAkron, Ohio 44333

Phone: 330-668-1444Fax: 330-666-5434

Data Acquisition Product Links(click on text to jump to page)

Data Acquisition | Data Logger | Chart Recorder | Thermocouple | Oscilloscope

Description Order No.GL800Compact, lightweight, multi-channel data logger that provides 20 standard analog measurement channels, expandable to 200.

GL800

Battery Pack7.2V/2200mAh Battery pack. B-517

DC power cable2-meter DC power cable, bare tips. B-514

Carrying CaseDurable case designed specifically for the GL800 B-536

Description Order No.

Extension Terminal Base SetExtension terminal base unit with cable. B-537

20 Channel Extension Set20 channel terminals, Extension terminal base unit, connection plate, screws.

B-538

Logic Alarm Cable2-meter, bare tips. B-513

Humidity Sensor3-meter with dedicated power connector. B-530

Ordering Guide