geo-cape workshop 18-20 aug 08 a geo venture class ...€¦ · slide9 spectrometer objective ir...
TRANSCRIPT
GEO-CAPE Workshop 18-20 Aug 08
A GEO Venture Class tropospheric O3 and HCHO measurement concept based on
the TIMSJack Kumer, Roche, A. E.; Rairden, R.L.; and J.L. Mergenthaler, with Lockheed Martin Adv. Tech. Center
and R.B. Chatfield with NASA Ames Res. Center
The USA National Research Council Decadal Survey (NRCDS) Report for Earth Science : –describes requirements for improved atmospheric measurements to gain crucial understanding for air quality, climate change, and weather.
–We are developing compact multi-channel grating mapping spectrometers (GMS) to meet this challenge.
In this presentation we will describe:–measurements of atmospheric spectra that we have made with a proto-type GMS with λ/∆λ = 6000 in the
midwave infrared region. [we refer to the projected space application of this GMS as “TIMS”]–a potential application in a pathfinder mission for the NRCDS Geostationary Coastal and Air Pollution Events
(GEO-CAPE) Mission. It would provide:–measurements of column O3 and HCHO on the Continental USA (CONUS) with
• hourly refresh and on contiguous double sampled 5 km north-south by 10 km west-east footprints–Input to the grating mapping spectrometer would be provided by a
• front end consisting of a scan mirror and telescope with 4.5 cm aperture• The scan mirror would west to east pushbroom the slit on a CONUS wide swath in 5 km steps • Using 7s per step provides full CONUS coverage with 2-hour refresh• The O3 and HCHO would be retrieved from spectra measured on two regions centered near 3.28 and
3.56 µm, respectively. Slide 1
Based on the demonstrated measurement capability of our proto-type GMS
–retrieved O3 column should have precision 1 to 2 % depending on albedo and pollution amount
• but with the added benefit of sensitivity to boundary layer (BL) O3. • In principle, when combined with the UV measurements, it could provide a measure of the
BL O3.
▪ Options other than UV include the thermal infrared bands @ 9.6 & 4.7µm
» These options are more costly than UV & this is a major consideration for the proposed pathfinder mission
» Further more it remains to judge whether these options would be as effective as the UV
–The HCHO retrieval should have precision 3 to 15 % depending on albedo and pollution amount
Figure below from P. K. Bhartia, et al,http://www.knmi.nl/omi/research/project/meetings/ostm10/presentations.html
Slide 2
4ºx8º
Geo measurement approach & spectral regions
Conus from Geo
~ 2500 km x 5000 km
scanline
scandirection
▬ 512 pixels along the scan line & 2 pixel slit widthSmallest footprint (ELF) is 5 km NS x 10 km WE
▬ 1000 5 km steps provide complete CONUS scan in 2 hours using 7s integration per step
Below: schematic of geo deployed tropospheric infrared mapping spectrometer TIMS, and UV mapping spectrometer. A 4.5 cm aperture is required.
Plots below of S/N per sample on a 5 x 5 km footprint and the transmittance difference in units %
UV
IR
Two MirrorScanner
TelescopePrimaryMirror
TIMS & UVSpectrometersEarth
Scene
Secondary
Slide 3
0.0
0.8
1.6
2.4
3.2
2760 2780 2800 2820wave number (cm-1)
S/N
per
spe
ctra
l sam
ple
0.0
0.2
0.4
0.6
0.8
1.0
1.2
% tr
ansm
ittan
ce d
iffer
ence
high end albedo = 0.08low end albedo =0.02% transmittance difference
wavelength = 3.53 to 3.62µm
column HCOH rmss[S/N] = 17 for albedo =0.08 [potential precision = 4.8 %]
and rmss[S/N]~6 for albedo =0.02 [potential precision = 16.7 %]
0
5
10
15
20
3035 3040 3045 3050 3055 3060wave number (cm-1)
S/N
per
spe
ctra
l sam
ple
0
2
4
6
% tr
ansm
ittan
ce d
iffer
ence
high end albedo = 0.08low end albedo=0.02% transmittance difference
column O3 rmss[S/N] = 88 for albedo =0.08 [potential precision = 1.1 %]
and rmss[S/N]~28 for albedo =0.02 [potential precision = 3.5 %]
wavelength = 3.27 to 3.30 µm
O3 & HCHO, the clean models are from the on line GATS browserclean & 40 ppbv surface enhanced O3
0
5
10
20 45 70 95 120O3 mixing ratio in ppbv
altit
ude
km
40 ppbv enhanced surface O3"clean" O3 profile
clean & doubled HCHO
0
5
10
0 1 2 3 4 5HCHO mixing ratio in ppbv
altit
ude
km
doubled HCHO profile"clean" HCHO profile
Slide 4
rmss[S/N] for 40 ppbv surface enhanced O3 on large aggregated footprints
rmss(S/N) "precision' rmss(S/N) "precision'0.080 20.8 4.8 10.4 9.60.020 8.2 12.2 4.1 24.4
50 x 50 km AG-footprintalbedo 25 x 25 km AG-footprint
Slide 5
0.0
2.0
4.0
3035 3040 3045 3050 3055 3060wave number (cm-1)
S/N
per
spe
ctra
l sam
ple high endalbedo = 0.08
low end albedo =0.02
surface O3 sensitivity on 50 x 50 km aggregated footprints+ 40 ppbv O3 rmss[S/N] = 20.8 for albedo =0.08 [potential precision = 4.8 %]
& rmss[S/N]=8.2 for albedo = 0.02 [potential precision = 12.2 %]
wavelength = 3.27 to 3.30 µm
0
4
8
3035 3040 3045 3050 3055 3060wave number (cm-1)
S/N
per
spe
ctra
l sam
ple
0
1
2
3
% tr
ansm
ittan
ce d
iffer
ence
high end albedo = 0.08low end albedo =0.02% transmittance difference
column O3 rmss[S/N] = 34 for albedo =0.0800 [potential precision = 2.9 %]
and rmss[S/N]=13 for ALBEDO =0.02 [potential precision = 7.4 %]
wavelength = 3.27 to 3.30 µm
O3 total column on double sampled 5 km x 10 km footprints in wet atmospheres & at 3.6 µm
–From above for the 3.3 µm region and for a standard atmosphere the total O3 can be retrieved at 1 to 3 % precision, depending on albedo, here we investigate
• The effects in the 3.3 µm region of tripling the atmospheric H2O content –The total O3 column precision is degraded to 3 to 7%, depending on albedo
• The total O3 column precision obtained from the data in the 3.6 µm region that is relatively insensitive to atmospheric H2O content
–This turns out to be the order 2 to 6 % precision, depending on albedo–Data simultaneously obtained in the two regions will together provide the order 1.5 to 4.6 % precision
Slide 6
Result for 3 x H2O
0.0
3.0
6.0
9.0
12.0
2760 2780 2800 2820wave number (cm-1)
S/N
per
spe
ctra
l sam
ple
0.0
1.0
2.0
3.0
4.0
% tr
ansm
ittan
ce d
iffer
ence
optimistic albedo = 0.080pessimistic albedo =0.0088% transmittance difference
column O3 rmss[S/N] = 55 for albedo =0.0800 [potential precision = 1.8 %]
and rmss[S/N]=18 for (H2O reflectance)/π =0.0088 [pot. precision = 5.8 %]
wavelength = 3.53 to 3.62 µm
Summary and conclusionsFrom Geostationary deployment it is important to diurnally map O3 and HCHO column in the CONUS with
•footprints the order 5 to 10 km and with a two hour refreshIt is most important to sense O3 in the boundary layer
We have demonstrated a tropospheric infrared mapping spectrometer (TIMS) lab demonstration grating mapping spectrometer GMS that when operated in the reflective SWIR address these requirements and provide measurements of– O3 column the order 1.5 to 4.6 % precision on double sampled 5 km NS by 10
km WE footprints with 2 hour refresh and withS/N the order 5 to 12 % for 40 ppbv near surface enhanced O3 on 50 x 50 km footprints
– Together with O3 retrieved from UV, the two measurements can provide a measure of the BL O3
– The geo deployed TIMS would also provide HCHO measurements with precision the order 5 to 17 % and intrinsically better than in the UV
Next we discuss our laborotory demonstration (Lab Demo) grating mapping spectrometer that these performance estimated are based upon Slide 7
Lab Demo GMS description
Slide 8
LN2 Container
Grating Holder
78K Optical Bench
Input Lens
LN2-Cooled DemonstrationSpectrometer Structure
λc=4.65 µm∆ν=0.32 cm-1
22--D Array MountD Array Mount
Laboratory demonstration mapping spectrometerLaboratory demonstration mapping spectrometer
35 cm
HAWAII HAWAII HgCdTeHgCdTe1024x1024 Pixels1024x1024 Pixels18.5 18.5 µµm pitchm pitch
1.9 cm
HAWAII 4-Quadrant ChipArray at spectrometer focal plane on LN2-cooled base plate
Large Format 2-D Infrared Array
The Basic Technique
Spatial512 pixels
Spectral 1024 pixels
Spatial512 pixels
Spectral 1024 pixels
Column 1λ=4.56 µ2193. cm-1
Column 1024λ=4.73 µ2113 cm-1
Total Array; Dispersion 80 cm-1Dispersion per Pixel: .08 cm-1
1024 columns
Echelle Grating52.7g/mmBlaze Angle~ 65 degSiO2 with Al Coating7x8 cm ruled areaUsed in 7th Order 49.5 deg
IncidentRay
78KTemperature105Well Capacity (e)
~ 190Dark Current (e/s) at 78K, 0.5V Bias
<60 CDSRead Noise (e rms)
0.95 µShort Wave Cuton
Long Wave cutoff
~60%QE (78K)
18.5 µPixel Pitch
1024x1024Format
78KTemperature105Well Capacity (e)
~ 190Dark Current (e/s) at 78K, 0.5V Bias
<60 CDSRead Noise (e rms)
0.95 µShort Wave Cuton
5.1µm Long Wave cutoff
~60%QE (78K)
18.5 µPixel Pitch
1024x1024Format
HAWAII HgCdTe MBE Array
FPA
Column 1λ=4.56 µ2193. cm-1
Column 1024λ=4.73 µ2113 cm-1
Total Array; Dispersion 80 cm-1Dispersion per Pixel: .08 cm-1
1024 columns
Echelle Grating52.7g/mmBlaze Angle~ 65 degSiO2 with Al Coating7x8 cm ruled areaUsed in 7th Order 49.5 deg
IncidentRay
78KTemperature105Well Capacity (e)
~ 190Dark Current (e/s) at 78K, 0.5V Bias
<60 CDSRead Noise (e rms)
0.95 µShort Wave Cuton
Long Wave cutoff
~60%QE (78K)
18.5 µPixel Pitch
1024x1024Format
78KTemperature105Well Capacity (e)
~ 190Dark Current (e/s) at 78K, 0.5V Bias
<60 CDSRead Noise (e rms)
0.95 µShort Wave Cuton
5.1µm Long Wave cutoff
~60%QE (78K)
18.5 µPixel Pitch
1024x1024Format
HAWAII HgCdTe MBE Array
FPA
5.00 CM
3.81cm
Eff Focal Length11.43 cmf/no=3
Grating
Array
LittrowOpticsDewar Window
Objective Lens
78K Enclosure
Filter
9° Along Slit
0.003° Slit width
Field of View
Laboratory demonstration grating mapping spectrometer (GMS) Optical schematic
slit
Lab Demo GMS test setup
Slide9
SpectrometerObjective
IR Source
Collimator
Laboratory Set-Up
Temperature Variable Calibration Source
slide # 1
Lab Demo GMS performance, gas cell absorption & noise
Slide 10
2112124213421442154216421742184
1 512 1024Pixel Number
Wavenumber
Spectral Calibration Using Absorption Cell DataPixel Data
Model
Spectral Calibration Using CO Absorption Cell Data-200
0
200
400
600
800
216721682169217021712172217321742175
Wavenumber
FWHM = 0.32 cm−1
[~3.7 Pix]
Assessing Spectral Line Shape Using CO absorption lines
Dispersion = 0.087 cm-1/Pix
Design Prediction: FWHM ~ 0.28 cm-1 for 60 µm Slit
Sensitivity Performance
1770
1780
1790
1800
1810
1820
1830
1840
1850
0 5 10 15 20 25Sample #
Cou
nts
Viewing 293K Blackbody
NESR(1s): 6.1x10-9 w/cm2/sr/cm-1
Extrapolated to Flight design
for single pixel
NESR 1.1x10-9
Standard Deviation: 16.9 counts ~106 Noise Electrons
Measured Responsivity: 3.6x10-10 w/cm2/sr/count
Each of the 25 sequentially observed data points is the 4 sec integratded count on a single pixel
Extrapolated to Flight design
for 36 pix aggregated footrpintl
NESR 1.8x10-10
•This observed noise performance is consistence with our noise performance modelTherefore the model can be used with confidence to extrapolate to the SWIR conceptual flight designThe S/N estimates presented above on slides 3 to 7 are based on that extrapolation of these data
Lab Demo GMS performance sky emission spectra1024 Pixels
512Pix
0 100 200 300 400 500 600 700 800 900 1000
2-Quadrant Image of Zenith SKY IR Emission Spectra
Spectra at yellow line
A First-Order Fit to the Data
2144 cm-1
2114 2124 2134 2144 2154 2164 2174 2184 2194
H20
H20H20
H20
CO COCO
N20
CO2
H20
H20
CO
2114 cm-1 2164 cm-1 2194 cm-1
Identification of atmospheric emission spectral features in the lab demo measurement range from ~ 4.56 to 4.73 µm
Slide11
Slide12
Concept for geo SWIR application using the Lab Demo
3.27 to3.30 µm 3.53 to 3.62 µm
To achieve the measurements as discussed on slides 1 through 7 above the demo GMS would be used for the IR spectrometer component (slide 3)
–It would have two slits for the 3.27 – 3.30, and the 3.53 to 3.62 µm regions, respectively–These would be
•in the orders 9 and 8, respectively of the demo grating (slide 8)•directed to separated regions on the focal plane as shown schematically below
–Separate order sorting filters for each would either be mounted on the respective slits, or on the focal plane, TBD
The IIP VSWIR design is an attractive option to the Lab Demo design Separate Arrays
Dichroic Beam splitters inserted in collimatedBundle
Re Imaging Optics and Filters
Single Grating,Different orders
Band 1 Band 2
The IIP VSWIR design facilitates multiple channel measurements by utilizing • a beam spilitter in the re-imaging arm• Sorting different grating orders in each branch• A detector array per branch
Trade off advantages of the IIP design include• Utilization of entire [both arrays], rather than just ½ of HAWAII 1rg, as does the Demo based design
• Spectral resolution improved by a factor two, improved spatial resolution• Requires a factor two larger aperture, therefore S/N increases by a factor two
Trade off disadvantage of the IIP design• Increased complexity and therefore cost
The IIP VSWIR design has been demonstrated:• To the right, note the validation of calibrated VSWIR data against the Denver University FTIR data
0.E+00
1.E‐07
2.E‐07
3.E‐07
4.E‐07
5.E‐07
6.E‐07
4280 4285 4290 4295 4300
0
0.1
0.2
0.3
0.4
0.5
0.6
TIMS calibrated dataDenver U FTIR data @ 0.25 cm‐1
Slide13
Slide14
estimates for emissivity near 3.3 & 3.6 um (1 of 3)
Albedo~ 0.08
Albedo~ 0.02approximate that albedo = 1- emissivity, then
Slide15
estimates for emissivity near 3.3 & 3.6 um (2 of 3)
Albedo~ 0.08
Albedo~ 0.02approximate that albedo = 1- emissivity, then
Slide16
estimates for emissivity near 3.3 & 3.6 um (3 of 3)
▪ illustrates trend to smaller albedos at larger wavelengths▪ assuming this holds on going from 4.3 to 3.3 – 3.6 um, then the large majority of the albedo in the latter region will be > 0.02, and a significant fraction will be > 0.08
Albedo~ 0.08
Albedo~ 0.02approximate that albedo = 1- emissivity, then
MAS images VIS, 2.10, 3.35 and 11.04 um
Slide17