GS 1R1B30-E-H 4th edition Model RAKD Small Metal Rotameter General Specifications A float is guided concentrically in a conic metal tube. The position of this float is magnetically transmitted to the indicator. The short-tube Rotameter is used for measurement of low flow rates of liquids and gases. Its special application is in troubled, opaque or aggressive mediums and under high pressure. The instrument is mounted in a vertical pipeline with flow direction upwards. When the process conditions are changed the scale needs to be replaced by a new one of which the values should be calculated. Fig. 1a Indicator RAKD with tube without valve Fig. 1b Tube RAKD with valve FEATURES - Different process connections like internal threads and flanges - With valves (horizontal connection) and without valve (vertical connection) - Version with controller (only in combination with valve) - All wetted parts of stainless steel (1.4571) - Accuracy class 4 (according VDI/VDE 3513) - Round industrial standardized stainless steel housing of protection IP 65 (on request IP 67) - Light, guided floats resulting in low pressure loss and stable float movement - Maximum flow range 1-250 l/h water resp. 40-8000 l/h air, portioned in 13 flow ranges with a relation of 1:10 - Graduation of maximum flow ranges by factor 1,6 realised with 4 floats and 13 cone sizes - Fine controll valve with PTFE-gasket - Electric µP-controlled transmitter with linearized output - Electrical connection by fast connection technique (Quickon) - Inductive sensors as limit switches also in Fail-Safe-version can be fit - Pressure controller (normal up to 25 bar at 20°C) for a maximum flow of 100 l/h water resp. 3.250 l/h air - Connection of common transformer isolated barriers and transmitter power supplies possible - Alignment possibility of the electric transmitters with additional tuner (Service Box). - Also in Ex-version

Upload: others

Post on 26-Apr-2020

31 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: General Model RAKD Small Metal Rotameter Specifications ROTA model RAKD.pdf · Model RAKD Small Metal Rotameter General Specifications A float is guided concentrically in a conic

GS 1R1B30-E-H 4th edition

Model RAKD Small Metal Rotameter

General Specifications

A float is guided concentrically in a conic metal tube. The position of this float is magnetically transmitted to the indicator. The short-tube Rotameter is used for measurement of low flow rates of liquids and gases. Its special application is in troubled, opaque or aggressive mediums and under high pressure. The instrument is mounted in a vertical pipeline with flow direction upwards.

When the process conditions are changed the scale needs to be replaced by a new one of which the values should be calculated.

Fig. 1a Indicator RAKD with tube without valve

Fig. 1b Tube RAKD with valve

FEATURES - Different process connections like internal threads

and flanges - With valves (horizontal connection)

and without valve (vertical connection) - Version with controller (only in combination with valve) - All wetted parts of stainless steel (1.4571) - Accuracy class 4 (according VDI/VDE 3513) - Round industrial standardized stainless steel housing

of protection IP 65 (on request IP 67) - Light, guided floats resulting in low pressure loss and

stable float movement - Maximum flow range 1-250 l/h water resp.

40-8000 l/h air, portioned in 13 flow ranges with a relation of 1:10

- Graduation of maximum flow ranges by factor 1,6 realised with 4 floats and 13 cone sizes

- Fine controll valve with PTFE-gasket - Electric µP-controlled transmitter with linearized output - Electrical connection by fast connection technique

(Quickon) - Inductive sensors as limit switches also in

Fail-Safe-version can be fit - Pressure controller (normal up to 25 bar at 20°C) for a

maximum flow of 100 l/h water resp. 3.250 l/h air - Connection of common transformer isolated barriers

and transmitter power supplies possible - Alignment possibility of the electric transmitters with

additional tuner (Service Box). - Also in Ex-version

Page 2: General Model RAKD Small Metal Rotameter Specifications ROTA model RAKD.pdf · Model RAKD Small Metal Rotameter General Specifications A float is guided concentrically in a conic

GS 1R1B30-E-H

2

STANDARD SPECIFICATIONS The responsiblity with respect to the suitability and according application of our flowmeter is only situated by the customer.

Fluids to be measured : Liquid or gas

Measuring ranges : see flow table Measuring ranges ratio :10:1

Accuracy : class 4 acc. VDI/VDE 3513

Indication scale : Flow units or percent

Process connection – Inner thread : G 1/4; 1/4 NPT; G 3/8; 3/8 NPT – Cutting ring : 6 mm; 8 mm; 10 mm; 12 mm – Nozzle : 6 mm; 8 mm – Flange : - acc. EN 1092-1

DN15 and DN25 PN40; - acc. ASME B 16.5 ½” and 1” 150lbs, 300lbs

Materials : wetted parts 1.4571; other materials on request

Installation – Installation position : vertical – Flow direction : upwards – Face to face length : 125 mm (with flange 250 mm)

Housing – Material : CrNi-Steel (1.4301) – Protection : IP 65

Process temperature : without valve -25°C to 250°C with valve -25°C to 150°C See also fig. 6. Lower temperatures on request.

Weight : see table 9

ELECTRONIC TRANSMITTER (indicator type “E”) Temperature range : -25°C to + 65°C Power supply : 13.5-30 V DC Load resistance : (U-13.5V) /20mA Analog output : 4-20 mA Linearity : ≤ ± 0.25% f.s. Hysteresis : ≤ ± 0.15% f.s. Repeatability : ≤ ± 0.16% f.s. Influence of power supply : ≤ ± 0.1% f.s. Temp. coefficient of analog output

: ≤ ± 0.5% /10 K f.s. AC-part of analog output

: ≤ ± 0.15% f.s. Long time stability : ≤ ± 0.2% / Jahr Maximum output current : 21.5 mA Output current in case of failure

: ≤ 3.6 mA (NAMUR NE 43) Response time (99%) : appr. 1 s Electrical connection : QUICKON - Cable diameter : 4-6 mm - Max. cross section : 0.34 bis 0.75 mm2

OPTIONAL SPECIFICATIONS

PULSE OUTPUT ( OPTION /CP) Electronic switch with galvanic isolation acc. EN 60947-5-6 (NAMUR) Pulse length : 200 ms Max. frequency : 4 Hz Pulse rate : Qmax ≤ 1 → 0.0001

: 1 < Qmax ≤ 10 → 0.001 etc. e.g.. Qmax = 1 m3/h → 1 Puls = 0.0001 m3 = 0.1 l )

LIMIT SWITCH ES IN STANDARD VERSION (Option /K1 to /K3) Type : inductive proximity switch

SC 2-NO acc. DIN EN 60947-5-6 (NAMUR)

Temperature range : -25°C to + 100°C Nominal voltage : 8V DC Output signal : see table 1 Hysteresis : < 0.5 mm

LIMIT SWITCHES IN FAIL SAFE VERSION (Option /K6 to /K8) Type : inductive proximity switch

SJ 2-SN acc. DIN EN 60947-5-6 (NAMUR) Technical data on request.

Table 1

EXPLOSION PROOF VERSION, RAKD with ATEX- certificate (Option /XE) Intrinsic safe acc. EN 50020 Acc. certificate of conformity KEMA 00ATEX 1037X Explosion proof : EEx ia II T6; Gruppe II ; category 2G Entity parameter : Table 2

Function

MAX

MIN

Pointer

above GWbelow GW

above GWbelow GW

T1.EPS

SC 2-N0 SJ 2-SN

Switch Signal Switch Signal Fail safe

onoff

onoff

offon

offon

1mA3mA 3mA

3mA 3mA1mA

1mA

1mA1mA

1mA

GW = limit

Analog output Puls outputLimit switch

Type 2

Limit switch

Type 3

Ui [V] 30 16 16 16

Ii [mA] 100 20 25 52

Pi [mW] 750 64 64 169

Li [mH] 0.73 0 0.15 0.15

Ci [nF] 2.4 0 150 150

Tex1.EPS

Page 3: General Model RAKD Small Metal Rotameter Specifications ROTA model RAKD.pdf · Model RAKD Small Metal Rotameter General Specifications A float is guided concentrically in a conic

GS 1R1B30-E-H

3

F2.EPS

Housing Diaphragm Springs

R1 / R3 CrNi-Steel PTFE CrNi-Steel

R2 / R4 Brass Buna CrNi-Steel

T5.EPS

inlet pressure

l/h air at 20°C ; 1.013 bar abs.

Controller (Option /R1 to R4) Differential pressure controller for a constant flow at fluctuations of the process pressure. These are no valves to reduce the pressure. - Controller /R1 and /R2 for liquids with variable inlet or outlet pressure and for gases with variable inlet pressure and constan back pressure. - Controller /R3 and /R4 for gases with fluctuations of the back pressure.

Max. liquid flow : 100 l/h Max. gas flow : 3250 l/h Max. pressure : 25 bar Recommended differential pressure

: >400 mbar Temperature range : -25°C to + 80°C Materials

Temperature specification : Version 1: RAKD with indicator “T” and limit switch type 2 : Table 3

Version 2: RAKD with indicator “T” and limit switch type 3:: Table 4

Version 3 : RAKD with indicator “E” and with or without limit switch type 2 : Table 5

Version 4 : RAKD with indicator “E” with limit switch type 3: The smaller environmental temperature must be found according to the available temperature class and the maximum process temperature from table 4 and 5.

Temperature class T6 T5 T5 T4 T4

Max. ambient temperature 65°C 80°C 59°C 100°C 73°C

Max. process temperature 65°C 80°C 100°C 100°C 135°C

Tex2.EPS

Temperature class T6 T5 T5 T4 T4 T4

Max. ambient temperature 24°C 37°C 34°C 57°C 54°C 48°C

Max. process temperature 65°C 80°C 100°C 80°C 100°C 135°C

Tex3.EPS

Temperature class T6 T5 T5 T4

Max. ambient temperature 65°C 50°C 45°C 38°C

Max. process temperature 65°C 80°C 100°C 135°C

Tex4.EPS

EXPLOSION PROOF VERSION, RAKD “non incendive” (Option /XN) Type “n” (non incendive) acc. EN 50021. Explosion proof : EEx nL IIC T6 X Entity parameter : Table 6

Analog output Pulse output

/CP

Limit switch

SC2-N0, /K1...3

Limit switch

SJ2-SN, /K6...8

Ui [V] 30 16 16 16

Ii [mA] 100 20 25 25

Pi [mW] 750 64 64 64

Li [µH] 730 0 150 100

Ci [nF] 2,4 0 150 30

T12.EPS

POWER SUPPLY FOR LIMIT SWITCHES (Option /W_) Type : KFA6-SR2-Ex1.w (230 VAC)

KFD2-SR2-Ex1.w (24 VDC) Transmitter relay acc. DIN EN 50227 (NAMUR)

Power supply : 230 V AC + 10%/-15%, 50/60Hz : 24 V DC ± 20%

Relay output : 1 potential-free changeover contact : 2 potential-free changeover contacts

Switching capacity : max. 250 V AC, max. 2 A Control circuit : intrinsic safe [EEx ia] IIC; group II;

category (1) GD Entity parameter : see fig. 5

acc. certificate of conformity : PTB 00 ATEX 2081 (230 V AC) PTB 00 ATEX 2080 (24 V DC)

Fig. 2 Diagram controller characteristic

POWER SUPPLY FOR THE INTRINSIC SAFE ELECTRONIC TRANSMITTER (Option /U__) Type : Intrinsic safe power supply with

galvanic separated outputs and output (SINEAX B811) acc. certificate of conformity PTB 97 ATEX 2083

Poweer supply : 24 V to 60 V AC/DC 85 V to 230 V AC

Max. load resistance : 750 Ω Output current : 0/4 mA – 20 mA Control circuit : Intrinsic safe [EEx ia] IIC; group II,

category (1) G Entity parameter : see fig. 5

Page 4: General Model RAKD Small Metal Rotameter Specifications ROTA model RAKD.pdf · Model RAKD Small Metal Rotameter General Specifications A float is guided concentrically in a conic

GS 1R1B30-E-H

4

CONNECTION DIAGRAMS

i

i

ii

i

-

+

+

-

oo

o

oo

-

+-

+

~ ~

Limit switch type 2

I = 25mA

T; S = Connector QUICKON

P = 64mW

L = 0.15mH

C = 150nF

KEMA 00ATEX 1037X

EEx ia IIC T6...T5

U = 16V

EN 60947-5-6 (NAMUR)

Limit switch

RAKD - T80

PTB 00ATEX 2081

L = 97mH

C = 2320nF

I = 19.1mA

P = 51mW

Transmitter relay

KFA6-SR2-Ex2.W

U = 10.6V

EEx ia IIC

KFA6-SR2-Ex2.W

Power supply 230V AC

Limit switch MIN

Limit switch MAX

BU2

BN

BU

BN

1

2

1

10

12

6

4

3

1

11

7

8

9

14 15

MIN

F

MAX

S

T

U

Hazardous Area Safe Area

F3.EPS

Fig. 3 Ex-version of RAKD with 2 limit switches in combination with transmitter relay

Power supplyU = 13.5 bis 30V DC

T = Connector QUICKON

F

RAKD-E80

G

T

Output 4-20mA

U - 13.5V

20mAR <L

2

-

+

1

+ -

-

+

F4.EPS

Fig. 4 RAKD with electronic transmitterr

Page 5: General Model RAKD Small Metal Rotameter Specifications ROTA model RAKD.pdf · Model RAKD Small Metal Rotameter General Specifications A float is guided concentrically in a conic

GS 1R1B30-E-H

5

Limit switch / PulsLimit switch / Puls

T; S = Connector QUICKONT; S = Connector QUICKON

EEx ia IIC T6...T5EEx ia IIC T6...T5U = 16VU = 16V

P = 64mWP = 64mWI = 20mAI = 20mA

C = 0nFC = 0nFL = 0mHL = 0mH

KEMA 00ATEX 1037XKEMA 00ATEX 1037X

Puls outputPuls output

EEx ia IIC T6...T5EEx ia IIC T6...T5 U = 10.6VU = 10.6V

KEMA 00ATEX 1037XKEMA 00ATEX 1037X

L = 0.15mHL = 0.15mHC = 150nFC = 150nF

U = 16VU = 16VI = 25mAI = 25mAP = 64mWP = 64mW

PTB 00ATEX 2081PTB 00ATEX 2081

L = 97mHL = 97mHC = 2320nFC = 2320nF

I = 19.1mAI = 19.1mAP = 51mWP = 51mW

Limit switch type 2Limit switch type 2

EN 60947-5-6 (NAMUR)EN 60947-5-6 (NAMUR)

Limit switch orLimit switch orPulse output

KFA6-SR2-Ex1.WKFA6-SR2-Ex1.W

KFA6-SR2-Ex1.WKFA6-SR2-Ex1.WTransmitter relayTransmitter relay

EEx ia IICEEx ia IIC

4-20mA output

I = 100mAU = 30V

KEMA 00ATEX 1037X

L = 0.73mHC = 2.4nFP = 0.75W

EEx ia IIC T6...T5

RAKD-E80 Option /XESINEAX B 811

Option: /U2F

Transmitter power supply

PTB 97 ATEX 2083

Tamax = 55°C

C = 178nFL = 6.7mH

SINEAX B 811

U = 21V

P = 660mW

EEx ia IIC

I = 75mA

OutputOutput

4-20mA4-20mA

Power supplyPower supply

230V AC230V AC

i

i

i

i

i

o

i

i

i

i

i

o

o

o

o

-P

+P

-

-+

~ ~

ii

iii o

+

L-L-

L+L+

-

+

-

+

oo

oo

BUBU

1 K1 K

15 K15 K

2

1

2

3

1

1414

7

8

9

1515

BNBN 1

1

2

+/~+/~

1ININ

-/~-/~

5

6ININ

1010

OUTOUT4

OUTOUT9

S

S U

F

G

T T

Hazardous Area Safe Area

Fig. 5 Ex-version of RAKD with electronic transmitter with power supply and limit switch or pulse output in combination with transmitter relay

Page 6: General Model RAKD Small Metal Rotameter Specifications ROTA model RAKD.pdf · Model RAKD Small Metal Rotameter General Specifications A float is guided concentrically in a conic

GS 1R1B30-E-H

6

TEMPERATURE SPECIFICATION

50 55 60 65 70

75

100

200

Illustration of maximum allowable operation temperature depending onambient temperature for Rotameter type RAKD

Max

. op

erat

ion

tem

per

atu

re [°

C]

Ambient temperature [°C]

75 80

125

150

175

225

250

275

Version without valve Version with valve F6.EPS

MODEL SPECIFICATIONS Please make your decision in this order :

1. Option controller with controller without controller without controller without controller

2. Version with valve with valve without valve without valve

3. Max. Flow

Cone

1.0 - 100 l/h Water

40 - 3250 l/h Air

31 - 51

1.0 - 250 l/h Water

40 - 8000 l/h Air

31 - 53

1.0 - 100 l/h Water

40 - 3250 l/h Air

31 - 51

160 - 250 l/h Water

5000 - 8000 l/h Air

52 - 53

4. Process connection Inner thread

Cutting ring

Nozzle

Inner thread

Cutting ring

Nozzle

Inner thread

Cutting ring

Nozzle

Flange

Inner thread

Cutting ring

Flange

Specify the model code

according the mentioned page Page 7 Page 8 Page 9 Page 10

T10.EPSOrdering instructions Standard: a: Model, suffix and option code b: Flow conditions c: Temperature d: Pressure e: Viscosity f: Density For gases: cross reference of the scale Option/B: customer specification notes

Fig. 6

For option /XE (Ex-version) the maximum values for ambient and process temperature according to the respective temperature class mentioned in fig. 5 and tables 2 to 5 must be regarded. The minimum ambient temperature is -25°C. Lower temperatures on request.

Page 7: General Model RAKD Small Metal Rotameter Specifications ROTA model RAKD.pdf · Model RAKD Small Metal Rotameter General Specifications A float is guided concentrically in a conic

GS 1R1B30-E-H

7 R

AK

D w

ith

va

lve

an

d c

on

tro

lle

r 1

.0 –

10

0 l/h

wa

ter / 4

0 –

32

50

l/h

air

: P

ro

cess

co

nn

ecti

on

Inte

rn

al th

read

:

Rp 1

/4 P

N 2

5

41

R3

1/4

NP

T P

N 2

5

41

T3

Cu

ttin

g r

ing

:

∅ 6

PN

25

53

C3

∅ 8

PN

25

54

C3

∅ 1

0 P

N 2

5

55

C3

∅ 1

2 P

N 2

5

56

C3

No

zzle

:

∅ 6

PN

10

53

P1

∅ 8

PN

10

54

P1

Material: 1.4571 S

S

Ind

icato

r

Typ

e:

Local in

di c

ato

r T

Indic

at o

r w

ith

ele

ctr

ic o

ut p

ut

E

Ho

usin

g t

yp

e:

St a

i nl e

ss s

teel

80

Po

wer s

up

ply

:

none,

fo

r lo

cal in

di c

ato

r „T

NN

N

for

el e

ctr

ic t

ype „

E“,

24V

DC

, 2-w

ire,

4-2

0m

A

424

Sty

l e:

Vers

ion

D

Op

tio

ns

Ind

icato

r:

US

engin

eering u

nits

/A12

Mark:

V4A

-sta

inle

ss s

teel pla

te

(description a

ccord

ing c

usto

mer

dem

and)

/B1

Tag –

No. on s

cale

, or

oth

er

Rota

met e

r – S

pecifi c

ation (

max.

2x40 s

igns)

/ BG

Double

scal e

/ B

D

Pu

l se o

utp

ut:

( n

ot fo

r ty

pe T

and c

onta

ct )

Puls

e o

ut p

ut, p

ot e

nt ial off (

NA

MU

R E

N 5

0227)

/CP

Test:

Cert

ifi c

ate

oil a

nd g

rease f

ree f

or

oxygene u

se

/H1

Lim

it s

wi t

ch

es:

Mi n

- C

ont a

ct

/ K

1

Ma

x -

Cont a

ct

/K2

Min

- M

ax -

Conta

ct (o

nly

local in

di c

ato

r ty

pe: T

) / K

3

Min

- C

onta

ct

F

ail-S

af e

vers

ion

/K6

Ma

x –

Conta

ct F

ai l-S

afe

vers

i on

/K7

Min

- M

ax –

Conta

ct F

ail-S

afe

ve

rsi o

n

(

onl y

local in

dic

ato

r t y

pe: T

)

/K8

Intr

insic

all

y s

afe

:

CE

NE

LE

C A

TE

X

(only

for

ele

ct r

icle

out p

ut

and c

ont a

ct)

/X

E

„nL“

non incendi v

e

(only

for

ele

ct r

icle

outp

ut

and c

ont a

ct)

/ XN

Certi

ficate

s

Pre

ssure

test w

ith c

ert

ific

ate

/P

P

Cert

ific

ate

of com

pl iance 2

.1 a

cc. E

N 1

0204

/P2

Test re

port

2.2

acc. E

N 1

0204

/ P3

3. 1

B C

er t

ific

ate

acc. E

N 1

0204 o

nly

for

t ube,

connection h

eads, scre

w s

ealing p

lug

/P6

Reg

ler:

P

re p

ressu

re c

ontr

ol ler

1.4

57

1 (

onl y

with v

alv

e in

inl e

t; f

or

gas

wi th v

aria

ble

pre

pr e

ssu

re a

nd

liq

uid

s w

i th v

aria

ble

pre

and

back p

r essure

)

/R1

Pre

pre

ssu

re c

ont r

ol ler

bra

ss (

only

with

valv

e in inle

t; fo

r g

as

with v

aria

ble

pr e

pr e

ssu

re a

nd

li q

uid

s w

ith v

aria

ble

pr e

and

back p

r essure

)

/R2

Back p

r essure

contr

oller

1. 4

57

1 (

only

with v

al v

e in o

ut le

t ; f

or

gas w

i th v

ari

able

back p

ressu

re)

/R3

Back p

ressur e

contr

ol ler

bra

ss (

only

wi th v

alv

e in

outlet;

fo

r gas

with v

ar ia

ble

ba

ck p

r essure

) /R

4

Ad

dit

ion

al in

str

um

en

ts

T

ransm

itt e

r po

wer

sup

pl y

, S

INA

X B

81

1- 1

4, 2

30V

/11

5V

AC

with

ga

lva

nic

isol a

t ion,

outp

ut

0/ 4

- 20m

A for

Ex-v

ers

ion

/U2F

Tr a

nsm

itte

r po

wer

sup

ply

, S

INA

X B

81

1- 1

3, 2

4V

AC

/DC

with

gal v

ani c

isola

tion,

outp

ut

0/4

- 20m

A for

Ex-v

ers

ion

/U3F

Tra

nsfo

r me

r is

ola

ted b

arr

iers

KF

A6-S

R2-E

x1

. W/2

30 V

AC

; 1-c

han

nel, 1

cha

ng

eover

con

tact

/W2

A

Tra

nsfo

rme

r i s

ola

ted b

arr

iers

KF

A6-S

R2-E

x2

.W/2

30 V

AC

; 2-c

han

nel s

, 2 c

hang

eover

co

nt a

ct

/W2B

Tr a

nsfo

r me

r is

ola

ted b

arr

ier s

KF

A6-S

H- E

x1.W

/ 230 V

AC

; 1-c

han

nels

, 1 c

hang

eover

co

nta

ct

/W2E

Tra

nsfo

rme

r is

ola

t ed b

ar r

iers

KF

D2

-SR

2-E

x1

. W/2

4 V

DC

; 1-c

han

nel, 1

cha

ng

eover

con

tact

/W4

A

Tra

nsfo

rme

r i s

ola

ted b

arr

iers

KF

D2

-SR

2-E

x2

.W/2

4 V

DC

; 2-c

han

nel s

, 2 c

hang

eover

co

nt a

ct

/W4B

Tr a

nsfo

r me

r is

ola

ted b

arr

ier s

KF

D2

-SH

- Ex1.W

/ 24 V

DC

; 1-c

han

nels

, 1 c

hang

eover

co

nta

ct

/W4E

Versi o

n

Wit

h v

alv

e i

n in

l et

Gasket

Valv

e s

eat

PT

FE

S

ilver

VS

E

PT

FE

P

CT

FE

V

PE

Wit

h v

alv

e i

n o

utl

et

Gasket

Val v

e s

eat

PT

FE

S

ilber

VS

A

PT

FE

P

CT

FE

V

PA

Max. F

low

C

on

e

Wate

r

(l/h

) A

i r

(l/h

)

1

40

31

1.6

60

32

2.5

100

33

4.0

150

34

6. 0

200

37

10

325

41

16

500

42

25

800

43

40

1400

44

60

2000

47

100

3250

51

Cone

∆p m

bar

31-3

7

6

41-4

3

8

44-5

1

11

R

KD

A

S

S

80

D

*

/

For

calc

ula

tion t

o o

ther

fluid

s / p

rocess c

ond

itio

ns p

lease u

se o

ur

siz

ing p

rogra

m D

ure

p_

V.

T6.EPS

Page 8: General Model RAKD Small Metal Rotameter Specifications ROTA model RAKD.pdf · Model RAKD Small Metal Rotameter General Specifications A float is guided concentrically in a conic

GS 1R1B30-E-H

8 R

AK

D w

ith

va

lve

1

.0 –

250

l/h

wa

ter

/ 40

– 8

000

l/h

air

: P

rocess

co

nn

ecti

on

Inte

rnal th

read

:

G 1

/4 P

N 4

0

41

G4

G 1

/4 P

N 1

00

41

G6

1/4

NP

T P

N 4

0

41

T4

1/4

NP

T P

N 1

00

41

T6

Cu

ttin

g r

ing

:

∅ 6

PN

40

53

C4

∅ 6

PN

100

53

C6

∅ 8

PN

40

54

C4

∅ 8

PN

100

54

C6

∅ 1

0 P

N 4

0

55

C4

∅ 1

0 P

N 1

00

55

C6

∅ 1

2 P

N 4

0

56

C4

∅ 1

2 P

N 1

00

56

C6

No

zzle

:

∅ 6

PN

10

53

P1

∅ 8

PN

10

54

P1

Materila: 1.4571 S

S

In

dic

ato

r

Typ

e:

Local in

dic

ato

r T

Indic

ato

r w

ith

ele

ctr

ic o

utp

ut

E

Ho

usin

g t

yp

e:

Sta

inle

ss s

teel

80

Po

wer

su

pp

l y:

none,

for

local in

dic

ato

r „T

NN

N

for

ele

ctr

ic t

ype „

E“,

24V

DC

, 2-w

ire,

4-2

0m

A

424

Sty

le:

Vers

ion

D

Vers

ion

Wit

h v

alv

e i

n in

pu

t

Gasket

Valv

e s

eat

PT

FE

S

ilver

VS

E

PT

FE

P

CT

FE

V

PE

Wit

h v

alv

e i

n o

utp

ut

Gasket

Valv

e s

eat

PT

FE

S

ilver

VS

A

PT

FE

P

CT

FE

V

PA

Max. F

low

C

on

e

Wate

r (l

/h)

Air

(l

/h)

1

40

31

1.6

60

32

2.5

100

33

4.0

150

34

6.0

200

37

10

325

41

16

500

42

25

800

43

40

1400

44

60

2000

47

100

3250

51

160

5000

52

250

8000

53

Konus

∆p m

bar

31-3

7

6

41-4

3

8

44-5

1

11

52-5

3

13

R

KD

A

S

S

80

D

*

/

Fo

r cal c

ula

tion t

o o

ther

fluid

s / p

rocess c

ond

itio

ns p

l ease u

se o

ur

siz

ing p

rogra

m D

ure

p_

V.

Op

tio

ns

Ind

icato

r:

US

engin

eering u

nits

/A12

Mark

:

V4A

-sta

inle

ss s

teel pla

te

(description a

ccord

ing c

usto

mer

dem

and)

/B1

Tag –

No.

on s

cale

, or

oth

er

Rota

mete

r – S

pecific

ation (

max.

2x40 s

igns)

/BG

Double

scale

/B

D

Pu

lse o

utp

ut:

( n

ot fo

r ty

pe T

and c

onta

ct)

Puls

e o

utp

ut, p

ote

ntial off (

NA

MU

R E

N 5

0227)

/CP

Test:

Cert

ific

ate

oil

and g

rease f

ree f

or

oxygene u

se

/H1

Lim

it s

wit

ch

es:

Min

- C

onta

ct

/K

1

Ma

x -

Conta

ct

/K2

Min

- M

ax -

Conta

ct (o

nly

local in

dic

ato

r ty

pe: T

) /K

3

Min

- C

onta

ct

F

ail-S

afe

vers

ion

/K6

Ma

x -

Conta

ct F

ail-S

afe

vers

ion

/K7

Min

- M

ax -

Conta

ct F

ail-S

afe

vers

ion

(only

local in

dic

ato

r ty

pe: T

)

/K8

Intr

insic

all

y s

afe

:

CE

NE

LE

C A

TE

X

(only

for

ele

ctr

ic o

utp

ut

and c

onta

ct)

/XE

„nL“

non incendiv

e

(only

for

ele

ctr

ic o

utp

ut

and c

onta

ct)

/ XN

Cert

ific

ate

s

Pre

ssure

test w

ith c

ert

ific

ate

/P

P

Cert

ific

ate

of com

plia

nce w

ith 2

.1acc. E

N 1

0204

/P2

Test re

port

2.2

acc. E

N 1

0204

/ P3

Inspection c

ert

ific

ate

3.1

B a

cc. E

N 1

0204 o

nly

for

tube,

connection h

eads, scre

w s

ealing p

lug

/P6

Ad

dit

ion

al i n

str

um

en

ts

T

ransm

itte

r po

wer

sup

ply

, S

INA

X B

81

1-1

4, 2

30V

/ 11

5V

AC

with

galv

anic

isola

tion,

outp

ut

0/4

-20m

A for

Ex-v

ers

ion

/U2F

Tra

nsm

itte

r po

wer

sup

ply

, S

INA

X B

81

1-1

3, 2

4V

AC

/DC

with

galv

anic

isola

tion,

outp

ut

0/4

-20m

A for

Ex-v

ers

ion

/U3F

Tra

nsfo

rme

r is

ola

ted b

arr

iers

KF

A6-S

R2-E

x1

.W/2

30 V

AC

; 1-c

han

nel, 1

cha

ng

eover

con

tact

/W2

A

Tra

nsfo

rme

r is

ola

ted b

arr

iers

KF

A6-S

R2-E

x2

.W/ 2

30 V

AC

; 2-c

han

nels

, 2 c

hang

eover

co

nta

ct

/W2B

Tra

nsfo

rme

r is

ola

ted b

arr

iers

KF

A6-S

H-E

x1.W

/230 V

AC

; 1-c

han

nels

, 1 c

hang

eover

co

nta

ct

/W2E

Tra

nsfo

rme

r is

ola

t ed b

arr

iers

KF

D2

-SR

2-E

x1

.W/2

4 V

DC

; 1-c

han

nel, 1

cha

ng

eover

con

tact

/W4

A

Tra

nsfo

rme

r is

ola

ted b

arr

iers

KF

D2

-SR

2-E

x2

.W/ 2

4 V

DC

; 2-c

han

nels

, 2 c

hang

eover

co

nta

ct

/W4B

Tra

nsfo

rme

r is

ola

ted b

arr

iers

KF

D2

- SH

-Ex1.W

/24 V

DC

; 1-c

han

nels

, 1 c

hang

eover

co

nta

ct

/W4E

T7.EPS

Page 9: General Model RAKD Small Metal Rotameter Specifications ROTA model RAKD.pdf · Model RAKD Small Metal Rotameter General Specifications A float is guided concentrically in a conic

GS 1R1B30-E-H

9 R

AK

D w

ith

ou

t va

lve

1

.0 –

10

0 l/h

wa

ter

/ 40

– 3

250

l/h

air

:

Max. F

low

C

on

e

Wate

r (l

/h)

Air

(l

/h)

1

40

31

1.6

60

32

2.5

100

33

4.0

150

34

6.0

200

37

10

325

41

16

500

42

25

800

43

40

1400

44

60

2000

47

100

3250

51

Cone

∆p m

bar

31-3

7

6

41-4

3

8

44-5

1

11

Pro

cess

co

nn

ecti

on

Fla

ng

e:

DN

15 P

N 4

0

01

D4

DN

25 P

N 4

0

02

D4

AN

SI ½

150 lbs

01

A1

AN

SI 1 1

50 lbs

02

A1

AN

SI ½

300 lbs

01

A2

AN

SI 1 3

00 lbs

02

A2

Inte

rnal th

read

:

G 1

/4 P

N 1

00

41

G6

G 1

/4 P

N 1

60

41

G7

1/4

NP

T P

N 1

00

41

T6

1/4

NP

T P

N 1

60

41

T7

Cu

ttin

g r

ing

:

∅ 6

PN

100

53

C6

∅ 6

PN

160

53

C7

∅ 8

PN

100

54

C6

∅ 8

PN

160

54

C7

∅ 1

0 P

N 1

00

55

C6

∅ 1

0 P

N 1

60

55

C7

∅ 1

2 P

N 1

00

56

C6

∅ 1

2 P

N 1

60

56

C7

No

zzle

:

∅ 6

PN

10

53

P1

∅ 8

PN

10

54

P1

Material: 1.4571 S

S

Vers

ion

Without

valv

e

NN

N

Ind

icato

r

Typ

e:

Local in

dic

ato

r T

Indic

ato

r w

i th

ele

ctr

ic o

utp

ut

E

Ho

usin

g t

yp

e:

Sta

inle

ss s

teel

80

Po

wer

su

pp

ly:

none,

fo

r lo

cal in

di c

ato

r „T

“ N

NN

for

ele

ctr

ic t

ype „

E“,

24V

DC

, 2-w

i re,

4-2

0m

A

424

Sty

le:

Vers

ion

D

R

KD

A

S

S

N

NN

80

D

*

/

For

calc

ula

tion t

o o

ther

fluid

s / p

rocess c

ond

itio

ns p

lease u

se o

ur

siz

ing p

rogra

m D

ure

p_V

.

Op

tio

ns

Ind

icato

r:

US

engin

eering u

nits

/A12

Mark

:

V4A

-sta

inle

ss s

teel pla

te

(description a

ccord

ing c

usto

mer

dem

and)

/B1

Tag –

No. on s

cale

, or

oth

er

Rota

mete

r – S

pecifi c

ation (

max.

2x40 s

igns)

/BG

Double

scale

/B

D

Pu

lse o

utp

ut:

( n

ot fo

r ty

pe T

and c

onta

ct)

Puls

e o

utp

ut, p

ote

ntial off (

NA

MU

R E

N 5

0227)

/CP

Test:

Cert

ific

ate

oil

and g

rease f

ree f

or

oxygene u

se

/H1

Lim

it s

wit

ch

es:

Min

- C

onta

ct

/K

1

Ma

x -

Conta

ct

/K2

Min

- M

ax -

Conta

ct (o

nly

local in

dic

ato

r ty

pe: T

) /K

3

Min

- C

onta

ct

F

ail-S

afe

vers

ion

/K6

Ma

x -

Conta

ct F

ai l-S

afe

vers

ion

/K7

Min

- M

ax -

Conta

ct F

ai l-S

afe

vers

ion

(only

local in

dic

ato

r ty

pe: T

)

/K8

Intr

insic

all

y s

afe

:

CE

NE

LE

C A

TE

X

(onl y

for

ele

ctr

ic o

utp

ut

and c

onta

ct)

/X

E

„nL“

non incendiv

e

(only

for

ele

ctr

ic o

utp

ut

and c

onta

ct)

/X

N

Cert

ific

ate

s

Pre

ssure

test w

ith c

ert

ifi c

ate

/P

P

Cert

ific

ate

of com

plia

nce w

ith 2

.1acc. E

N 1

0204

/P2

Te

st re

port

2.2

acc. E

N 1

0204

/P3

Inspection c

ert

ific

ate

3.1

B a

cc. E

N 1

0204 o

nly

for

tube,

connection h

eads, scre

w s

ealing p

lug

/P6

Ad

dit

ion

al in

str

um

en

ts

T

ransm

itte

r po

wer

sup

ply

, S

INA

X B

81

1-1

4, 2

30V

/ 11

5V

AC

with

galv

anic

isola

tion,

outp

ut

0/4

-20m

A for

Ex-v

ers

ion

/U2F

Tra

nsm

itte

r po

wer

sup

ply

, S

INA

X B

81

1-1

3, 2

4V

AC

/DC

with

galv

anic

isola

tion,

outp

ut

0/4

-20m

A for

Ex-v

ers

ion

/U3F

Tra

nsfo

rme

r is

ola

ted b

arr

iers

KF

A6-S

R2-E

x1

.W/2

30 V

AC

; 1-c

han

nel, 1

cha

ng

eover

con

tact

/W2

A

Tra

nsfo

rme

r is

ola

ted b

arr

iers

KF

A6-S

R2-E

x2

.W/ 2

30 V

AC

; 2-c

han

nels

, 2 c

hang

eover

co

nta

ct

/W2B

Tra

nsfo

rme

r is

ola

ted b

arr

iers

KF

A6-S

H-E

x1.W

/230 V

AC

; 1-c

han

nels

, 1 c

hang

eover

co

nta

ct

/W2E

Tra

nsfo

rme

r is

ola

ted b

arr

iers

KF

D2

-SR

2-E

x1

.W/2

4 V

DC

; 1-c

han

nel, 1

cha

ng

eover

con

tact

/W4

A

Tra

nsfo

rme

r is

ola

ted b

arr

iers

KF

D2

-SR

2-E

x2

.W/ 2

4 V

DC

; 2-c

han

nels

, 2 c

hang

eover

co

nta

ct

/W4B

Tra

nsfo

rme

r is

ola

ted b

arr

iers

KF

D2

-SH

-Ex1.W

/24 V

DC

; 1-c

han

nels

, 1 c

hang

eover

co

nta

ct

/W4E

T8.EPS

Page 10: General Model RAKD Small Metal Rotameter Specifications ROTA model RAKD.pdf · Model RAKD Small Metal Rotameter General Specifications A float is guided concentrically in a conic

GS 1R1B30-E-H

10 R

AK

D w

ith

ou

t va

lve

1

60

– 2

50

l/h

wa

ter

/ 50

00

– 8

00

0 l

/h a

ir:

Max. F

low

C

on

e

Wate

r (l

/h)

Air

(l

/h)

160

5000

52

250

8000

53

Cone

∆p m

bar

52-5

3

13

Pro

cess

co

nn

ecti

on

Fla

ng

e:

DN

15 P

N 4

0

01

D4

DN

25 P

N 4

0

02

D4

AN

SI ½

150 lbs

01

A1

AN

SI 1 1

50 lbs

02

A1

AN

SI ½

300 lbs

01

A2

AN

SI 1 3

00 lbs

02

A2

Inte

rnal th

read

:

G 3

/8 P

N 1

00

42

G6

G 3

/8 P

N 1

60

42

G7

3/8

NP

T P

N 1

00

42

T6

3/8

NP

T P

N 1

60

42

T7

Cu

ttin

g r

ing

:

∅ 1

2 P

N 1

00

56

C6

∅ 1

2 P

N 1

60

56

C7

Material: 1.4571 S

S

V

ers

ion

Without

valv

e

NN

N

Ind

icato

r

Typ

e:

Local in

dic

ato

r T

Indic

ato

r w

ith

ele

ctr

ic o

utp

ut

E

Ho

usin

g t

yp

e:

Sta

inle

ss s

teel

80

Po

wer

su

pp

ly:

none,

fo

r lo

cal in

dic

ato

r „T

NN

N

for

ele

ctr

ic t

ype „

E“,

24V

DC

, 2-w

ire,

4-2

0m

A

424

Sty

le:

Vers

ion

D

Op

tio

ns

Ind

icato

r:

US

engin

eering u

nits

/A12

Mark

:

V4A

-sta

inle

ss s

teel pla

te

(description a

ccord

ing c

usto

mer

dem

and)

/B1

Tag –

No. on s

cale

, or

oth

er

Rota

mete

r – S

pecific

ation (

max.

2x40 s

igns)

/BG

Double

scale

/B

D

Pu

lse o

utp

ut:

( n

ot fo

r ty

pe T

and c

onta

ct)

Puls

e o

utp

ut, p

ote

ntial off (

NA

MU

R E

N 5

0227)

/CP

Test:

Cert

ific

ate

oil

and g

rease f

ree f

or

oxygene u

se

/H1

Lim

it s

wit

ch

es:

Min

- C

onta

ct

/K

1

Ma

x -

Conta

ct

/K2

Min

- M

ax -

Conta

ct (o

nly

local in

dic

ato

r ty

pe: T

) /K

3

Min

- C

onta

ct

F

ail-S

afe

vers

ion

/K6

Ma

x -

Conta

ct F

ail-S

afe

vers

ion

/K7

Min

- M

ax -

Conta

ct F

ail-S

afe

vers

ion

(only

local in

dic

ato

r ty

pe: T

)

/K8

Intr

insic

all

y s

afe

:

CE

NE

LE

C A

TE

X

(only

for

ele

ctr

ic o

utp

ut

and c

onta

ct)

/XE

„nL“

non incendiv

e

(only

for

ele

ctr

ic o

utp

ut

and c

onta

ct)

/XN

Cert

ific

ate

s

Pre

ssure

test w

ith c

ert

ific

ate

/P

P

Cert

ific

ate

of com

plia

nce w

ith 2

.1acc. E

N 1

0204

/P2

Test re

port

2.2

acc. E

N 1

0204

/P3

Inspection c

ert

ific

ate

3.1

B a

cc. E

N 1

0204 o

nly

for

tube,

connection h

eads, scre

w s

ealing p

lug

/P6

Ad

dit

ion

al in

str

um

en

ts

T

ransm

itte

r po

wer

sup

ply

, S

INA

X B

81

1-1

4, 2

30V

/11

5V

AC

with

galv

anic

isola

tion,

outp

ut

0/4

-20m

A for

Ex-v

ers

ion

/U2F

Tra

nsm

itte

r po

wer

sup

ply

, S

INA

X B

81

1-1

3, 2

4V

AC

/DC

with

ga

lva

nic

isola

tion,

outp

ut

0/4

-20m

A for

Ex-v

ers

ion

/U3F

Tra

nsfo

rme

r is

ola

ted b

arr

iers

KF

A6-S

R2-E

x1

.W/2

30 V

AC

; 1-c

han

nel, 1

cha

ng

eover

con

tact

/W2

A

Tra

nsfo

rme

r is

ola

ted b

arr

iers

KF

A6-S

R2-E

x2

.W/2

30 V

AC

; 2-c

han

nels

, 2 c

hang

eover

co

nta

ct

/W2B

Tra

nsfo

rme

r is

ola

ted b

arr

iers

KF

A6-S

H-E

x1.W

/230 V

AC

; 1-c

han

nels

, 1 c

hang

eover

co

nta

ct

/W2E

Tra

nsfo

rme

r is

ola

ted b

arr

iers

KF

D2

-SR

2-E

x1

.W/2

4 V

DC

; 1-c

han

nel, 1

cha

ng

eover

con

tact

/W4

A

Tra

nsfo

rme

r is

ola

ted b

arr

iers

KF

D2

-SR

2-E

x2

.W/2

4 V

DC

; 2-c

han

nels

, 2 c

hang

eover

co

nta

ct

/W4B

Tra

nsfo

rme

r is

ola

ted b

arr

iers

KF

D2

-SH

-Ex1.W

/24 V

DC

; 1-c

han

nels

, 1 c

hang

eover

co

nta

ct

/W4E

R

KD

A

S

S

N

NN

80

D

*

/

Fo

r calc

ula

tion t

o o

ther

fluid

s / p

r ocess c

ond

itio

ns p

lease u

se o

ur

siz

ing p

rogra

m D

ure

p_

V.

T9.EPS

Page 11: General Model RAKD Small Metal Rotameter Specifications ROTA model RAKD.pdf · Model RAKD Small Metal Rotameter General Specifications A float is guided concentrically in a conic

GS 1R1B30-E-H

11

DIMENSIONS Note : The dimensions a ; b ; c ; L1 ; L2 ; L3 are listed in table 7 and 8.

F7.EPS

F8.EPS

Fig. 7 Version without valve Fig. 8 Version with flange connections

F13.EPS

Fig. 9 Back view with mounting

Page 12: General Model RAKD Small Metal Rotameter Specifications ROTA model RAKD.pdf · Model RAKD Small Metal Rotameter General Specifications A float is guided concentrically in a conic

GS 1R1B30-E-H

12

F9.EPS

F12.EPS

Fig. 10 Version with inlet valve

Fig. 11 Version with outlet valve

Page 13: General Model RAKD Small Metal Rotameter Specifications ROTA model RAKD.pdf · Model RAKD Small Metal Rotameter General Specifications A float is guided concentrically in a conic

GS 1R1B30-E-H

13

F10.EPS

F11.EPS

Fig. 12 Version with inlet valve and inlet controller

Fig. 13 Version with outlet valve and back pressure controller

Page 14: General Model RAKD Small Metal Rotameter Specifications ROTA model RAKD.pdf · Model RAKD Small Metal Rotameter General Specifications A float is guided concentrically in a conic

GS 1R1B30-E-H

14

CONNECTION TYPES TABLE 7

a b c

Cone 31-51 Cone 52-53 Cone 31-53 Cone 31-51

Size

G 1/4 G 3/8 G 1/4 Rp 1/4

1/4 NPT 3/8 NPT 1/4 NPT 1/4 NPT

T2.EPS

Thread

INSTALLATION LENGTHS DEPENDING ON CONNECTION TYPE AND SIZE TABLE 8

L1 L2 L3

Process connection NW Cone 31-51 Cone 52-53 Cone 31-53 Cone 31-51

Cutting ring

6 mm 178 mm ---- 54.5 mm 142.5 mm

8 mm 172 mm ---- 51.5 mm 139.5 mm

10 mm 174 mm ---- 52.5 mm 140.5 mm

12 mm 174 mm 177 mm 52.5 mm 140.5 mm

Nozzle6 mm 182 mm ---- 56.5 mm 144.5 mm

8 mm 182 mm ---- 56.5 mm 144.5 mm

T3.EPS

Size

WEIGHTS TABLE 9

without valve with valve with controller

Weight approx. 600g approx. 1000g approx. 1800g

T4.EPS

Planning hints - The real working pressure has to be less than the specified pressure limit of the Rotameter. - Make sure that the wetted material is resistent to the medium. - Ambient and operation temperature has to be less than the specified maximum value. - If dirt accumulation is to be expected we recommend to install a bypass pipe. - To avoid float bouncing in case of gas application notice the recommandations of VDI/VDE 3513 Sheet 3. - To avoid mutual magnetic influence in case of a parallel design of several Rotameters take care that the distance

between the tube middle axes is not less than 120 mm. The distance to other ferrictic materials should not be less than 60 mm. - The strength of external magnetic fields close by the Rotameter should be approximately 0mT.

Page 15: General Model RAKD Small Metal Rotameter Specifications ROTA model RAKD.pdf · Model RAKD Small Metal Rotameter General Specifications A float is guided concentrically in a conic

GS 1R1B30-E-H

15

Page 16: General Model RAKD Small Metal Rotameter Specifications ROTA model RAKD.pdf · Model RAKD Small Metal Rotameter General Specifications A float is guided concentrically in a conic

GS 1R1B30-E-H

16