gc chromatography in the fast lane - agilent · chromatography in the fast lane mark sinnott...

65
Chromatography in the Fast Lane Mark Sinnott Application Engineer December 13, 2018 1 Fast GC

Upload: others

Post on 01-Aug-2020

6 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: GC Chromatography in the Fast Lane - Agilent · Chromatography in the Fast Lane Mark Sinnott Application Engineer December 13, 2018 1 Fast GC. Page 2 Questions to Ask ... Fast Analysis

Chromatography in the Fast Lane

Mark SinnottApplication EngineerDecember 13, 2018

1

Fast GC

Page 2: GC Chromatography in the Fast Lane - Agilent · Chromatography in the Fast Lane Mark Sinnott Application Engineer December 13, 2018 1 Fast GC. Page 2 Questions to Ask ... Fast Analysis

Page 2

Questions to Ask

What information do you need from your analysis?

Do you have more baseline than you need between your peaks?

Do you need to resolve all of the components?

Page 3: GC Chromatography in the Fast Lane - Agilent · Chromatography in the Fast Lane Mark Sinnott Application Engineer December 13, 2018 1 Fast GC. Page 2 Questions to Ask ... Fast Analysis

Page 3

•Stationary Phase

•Temperature Programming

•Carrier Gas: type and linear velocity

•Shorten Column Length

•Decrease Film Thickness

•Decrease Internal Diameter

Variables for Shortening Run Times

Page 4: GC Chromatography in the Fast Lane - Agilent · Chromatography in the Fast Lane Mark Sinnott Application Engineer December 13, 2018 1 Fast GC. Page 2 Questions to Ask ... Fast Analysis

Page 4

Resolution

RN k

ks =

4 1

1

+

Factors effecting Rs

Efficiency N = (gas, L, rc)

Retention k = (T, df, rc)

Selectivity = (T, phase)

Variables

L Length

rc radius

df film thickness

T Temperature

Page 5: GC Chromatography in the Fast Lane - Agilent · Chromatography in the Fast Lane Mark Sinnott Application Engineer December 13, 2018 1 Fast GC. Page 2 Questions to Ask ... Fast Analysis

Page 5

Resolution

RN k

ks =

4 1

1

+

Factors effecting Rs

Efficiency N = (gas, L, rc)

Retention k = (T, df, rc)

Selectivity = (T, phase)

Variables

L Length

rc radius

df film thickness

T Temperature

Page 6: GC Chromatography in the Fast Lane - Agilent · Chromatography in the Fast Lane Mark Sinnott Application Engineer December 13, 2018 1 Fast GC. Page 2 Questions to Ask ... Fast Analysis

Page 6

Stationary Phase - Common Types

Siloxane polymers

Poly(ethylene) glycols

Porous polymers

Page 7: GC Chromatography in the Fast Lane - Agilent · Chromatography in the Fast Lane Mark Sinnott Application Engineer December 13, 2018 1 Fast GC. Page 2 Questions to Ask ... Fast Analysis

Page 7

Selectivity

Relative spacing of the chromatographic peaks

The result of all non-polar, polarizable and polar interactions that cause a stationary phase to be more or less retentive to one analyte than another

Page 8: GC Chromatography in the Fast Lane - Agilent · Chromatography in the Fast Lane Mark Sinnott Application Engineer December 13, 2018 1 Fast GC. Page 2 Questions to Ask ... Fast Analysis

Page 8

Optimizing Selectivity

Match analyte polarity to stationary phase polarity

-like dissolves like(oil and water don’t mix)

Take advantage of unique interactions between analyte and stationary phase functional groups

Page 9: GC Chromatography in the Fast Lane - Agilent · Chromatography in the Fast Lane Mark Sinnott Application Engineer December 13, 2018 1 Fast GC. Page 2 Questions to Ask ... Fast Analysis

Page 9

Start with the Right Phase

Time (min.)

0 2 4 6 8 10 12 14

Time (min.)

DB-1

15m x 0.32mm, 0.25µm

Oven:

40°C for 2 min

40-120°C at 5°C/min

0 1 2 3 4 5

DB-Wax

15m, 0.32mm, 0.25µm

Oven:

80-190°C at 20°C/min

Page 10: GC Chromatography in the Fast Lane - Agilent · Chromatography in the Fast Lane Mark Sinnott Application Engineer December 13, 2018 1 Fast GC. Page 2 Questions to Ask ... Fast Analysis

Page 10

Resolution

RN k

ks =

4 1

1

+

Factors effecting Rs

Efficiency N = (gas, L, rc)

Retention k = (T, df, rc)

Selectivity = (T, phase)

Variables

L Length

rc radius

df film thickness

T Temperature

Page 11: GC Chromatography in the Fast Lane - Agilent · Chromatography in the Fast Lane Mark Sinnott Application Engineer December 13, 2018 1 Fast GC. Page 2 Questions to Ask ... Fast Analysis

Page 11

Column Length and Efficiency (Theoretical Plates)

0.25 mm IDn/m = 4630 (for k = 5)

Length (m) n

15 69,450

30 138,900

60 277,800

Page 12: GC Chromatography in the Fast Lane - Agilent · Chromatography in the Fast Lane Mark Sinnott Application Engineer December 13, 2018 1 Fast GC. Page 2 Questions to Ask ... Fast Analysis

Page 12

Column Length and Resolution

Length X 4 = Resolution X 2

R n L

t L

Page 13: GC Chromatography in the Fast Lane - Agilent · Chromatography in the Fast Lane Mark Sinnott Application Engineer December 13, 2018 1 Fast GC. Page 2 Questions to Ask ... Fast Analysis

Page 13

Column Length VS Resolution and Retention: Isothermal

Double the plates, double the time but not double the the resolution

15 m 60 m30 m

R=0.84

2.29 min

R=1.68

8.73 min

R=1.16

4.82 min

Page 14: GC Chromatography in the Fast Lane - Agilent · Chromatography in the Fast Lane Mark Sinnott Application Engineer December 13, 2018 1 Fast GC. Page 2 Questions to Ask ... Fast Analysis

Page 14

BTEX

Carrier: Helium, 36 cm/sec at 40°c

Oven : 40°C for 3 min, 5°/min to 100°C

1. Benzene

2. Toluene

3. Ethylbenzene

4. m,p-Xylene

5. o-Xylene

.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 5.0 5.5 6.0 6.5 7.0 7.5 8.0 8.5 9.00.50 1.00 1.50 2.00 2.50 3.00 3.50 4.00 4.50 5.00 5.50 6.00 6.50 7.00 7.50 8.00 8.50 9.00

1

2 3

4

5

DB-5

30 m

0.53 mm I.D., 0.5 µm

DB-5

15 m

0.53 mm I.D., 0.5 µm

0.20 0.40 0.60 0.80 1.00 1.20 1.40 1.60 1.80 2.00 2.20 2.40 2.60 2.80 3.00 3.20 3.40 3.60 3.80 4.00 4.20 4.40 4.60 4.80 5.00 5.20 5.40 5.60 5.801.0 2.0 3.0 4.0 5.0

1 2 3

4

5

6.0

DECREASE THE LENGTH

Page 15: GC Chromatography in the Fast Lane - Agilent · Chromatography in the Fast Lane Mark Sinnott Application Engineer December 13, 2018 1 Fast GC. Page 2 Questions to Ask ... Fast Analysis

Page 15

Column Length and Cost

15m30m

60m

$ $ $ $ $ $ $

Page 16: GC Chromatography in the Fast Lane - Agilent · Chromatography in the Fast Lane Mark Sinnott Application Engineer December 13, 2018 1 Fast GC. Page 2 Questions to Ask ... Fast Analysis

Page 16

Length Summary

If you Decrease Length:

Efficiency Decrease

Resolution Decrease

Analysis Time Decrease

Pressure Decrease

Cost Decrease

Page 17: GC Chromatography in the Fast Lane - Agilent · Chromatography in the Fast Lane Mark Sinnott Application Engineer December 13, 2018 1 Fast GC. Page 2 Questions to Ask ... Fast Analysis

Page 17

Resolution

RN k

ks =

4 1

1

+

Factors effecting Rs

Efficiency N = (gas, L, rc)

Retention k = (T, df, rc)

Selectivity = (T, phase)

Variables

L Length

rc radius

df film thickness

T Temperature

Page 18: GC Chromatography in the Fast Lane - Agilent · Chromatography in the Fast Lane Mark Sinnott Application Engineer December 13, 2018 1 Fast GC. Page 2 Questions to Ask ... Fast Analysis

Page 18

Column Diameter - Theoretical Efficiency

I.D. (mm)

0.05

0.10

0.20

0.25

0.32

0.45

0.53

0.18

n/m

23,160

11,580

5830

4630

3660

2840

2060

6,660

30 m

20 m

10 m

k = 5

5 m

N ~ 112,000

N ~ 112,000

N ~ 112,000

N ~ 112,000

Total Plates

Page 19: GC Chromatography in the Fast Lane - Agilent · Chromatography in the Fast Lane Mark Sinnott Application Engineer December 13, 2018 1 Fast GC. Page 2 Questions to Ask ... Fast Analysis

Page 19

Different Column I. D. Equal Phase Ratios

Time (min)0 5 10 15 20

Column : DB-62430 m, 0.32 mm, 1.8 m

Carrier:Oven: 65°CInjection: SplitDetector: FID

Helium, 40(cm/sec)

Column : DB-62430 m, 0.53 mm, 3 m

Page 20: GC Chromatography in the Fast Lane - Agilent · Chromatography in the Fast Lane Mark Sinnott Application Engineer December 13, 2018 1 Fast GC. Page 2 Questions to Ask ... Fast Analysis

Page 20

PHASE RATIO ()Film Thickness

Column Dimensions Phase Ratio β

30 m x .53 mm x 3.0 m 44

30 m x .32 mm x 1.8 m 44

KC = k β

r2df

β =

Page 21: GC Chromatography in the Fast Lane - Agilent · Chromatography in the Fast Lane Mark Sinnott Application Engineer December 13, 2018 1 Fast GC. Page 2 Questions to Ask ... Fast Analysis

Page 21

Column Diameter and Capacity

Like Polarity Phase/Solute0.25 µm film thickness

I.D. (mm) Capacity (ng)

0.05 1-2

0.18 25-55

0.20 35-70

0.25 80-160

0.32 110-220

0.53 1000-2000

0.45 600-800

0.10 6-13

Page 22: GC Chromatography in the Fast Lane - Agilent · Chromatography in the Fast Lane Mark Sinnott Application Engineer December 13, 2018 1 Fast GC. Page 2 Questions to Ask ... Fast Analysis

Page 22

Column Diameter - Inlet Head Pressures (Helium)

30 metersHydrogen pressures x 1/2

I.D (mm) Pressure (psig)

0.10 90-130

0.20 25-40

0.25 15-25

0.32 10-20

0.45 3-7

0.53 2-4

0.18 30-45

0.05 275-400

Page 23: GC Chromatography in the Fast Lane - Agilent · Chromatography in the Fast Lane Mark Sinnott Application Engineer December 13, 2018 1 Fast GC. Page 2 Questions to Ask ... Fast Analysis

Page 23

Column Diameter and Carrier Gas Flow

Lower flow rates: Smaller diameter columns

Higher flow rates: Larger diameter columns

Low flow rates : GC/MSHigh flow rates: Headspace, purge & trap

Page 24: GC Chromatography in the Fast Lane - Agilent · Chromatography in the Fast Lane Mark Sinnott Application Engineer December 13, 2018 1 Fast GC. Page 2 Questions to Ask ... Fast Analysis

Page 24

Diameter Summary

If you decrease the inside diameter:

Efficiency Increase

Resolution Increase

Pressure Increase

Capacity Decrease

Flow rate Decrease

Page 25: GC Chromatography in the Fast Lane - Agilent · Chromatography in the Fast Lane Mark Sinnott Application Engineer December 13, 2018 1 Fast GC. Page 2 Questions to Ask ... Fast Analysis

Page 25

Combining a change in Length with a change in Diameter

Eff

icie

nc

y

Analysis Time

Decrease Length

Decrease Diameter

Page 26: GC Chromatography in the Fast Lane - Agilent · Chromatography in the Fast Lane Mark Sinnott Application Engineer December 13, 2018 1 Fast GC. Page 2 Questions to Ask ... Fast Analysis

Page 26

Carrier Gas ConsiderationsBest velocity?

Optimal range of velocities

Too low or high results in loss of resolution

Balance resolution and analysis time

BUT... there is no LAW against going FASTER

Page 27: GC Chromatography in the Fast Lane - Agilent · Chromatography in the Fast Lane Mark Sinnott Application Engineer December 13, 2018 1 Fast GC. Page 2 Questions to Ask ... Fast Analysis

Page 27

60

Excessive Diffusion

Poor Mass Transfer

Carrier Gas Considerations

Van Deemter Curve

10 20 30 40 50 60

0.25

0.50

0.75

1.00

u (cm/sec)

H (

min

)

He

N2

H 2

uopt

Page 28: GC Chromatography in the Fast Lane - Agilent · Chromatography in the Fast Lane Mark Sinnott Application Engineer December 13, 2018 1 Fast GC. Page 2 Questions to Ask ... Fast Analysis

Page 28

Carrier Gas ConsiderationsUtilizing Computer Controlled Flow Ramping (EPC)

Lonely Late Eluters? = P

Decrease retention and overall run time by increasing pressure

(speed up the gas!)

Page 29: GC Chromatography in the Fast Lane - Agilent · Chromatography in the Fast Lane Mark Sinnott Application Engineer December 13, 2018 1 Fast GC. Page 2 Questions to Ask ... Fast Analysis

Fast Detector Electronics – 6890 FID

200 Hz

100 Hz

50 Hz

20 Hz

10 Hz

Proper peak quantitation typically requires a minimum of 10 data points per peak

10 Hz data rate: (peaks less than one second wide will have less than 10 data points)

200 Hz data rate: (peaks as narrow as 50 msec can be handled reliably)

Page 30: GC Chromatography in the Fast Lane - Agilent · Chromatography in the Fast Lane Mark Sinnott Application Engineer December 13, 2018 1 Fast GC. Page 2 Questions to Ask ... Fast Analysis

Page 30

Easy Options with Method Translation Software

Different Column Dimensions

Switch He to H2 Carrier Gas and Try Faster Velocities

Same Column & Gas Type but Faster Velocities

Combination of all of the above

Page 31: GC Chromatography in the Fast Lane - Agilent · Chromatography in the Fast Lane Mark Sinnott Application Engineer December 13, 2018 1 Fast GC. Page 2 Questions to Ask ... Fast Analysis

Temperature Programming

30m, 0.25mm ID

Temperature

10m, 0.1mm ID

Temperature

Temperature program must be modified to give same temperature of elution (i.e. faster ramps, shorter hold times)

Page 32: GC Chromatography in the Fast Lane - Agilent · Chromatography in the Fast Lane Mark Sinnott Application Engineer December 13, 2018 1 Fast GC. Page 2 Questions to Ask ... Fast Analysis

Page 32

2 . 0 0 4 . 0 0 6 . 0 0 8 . 0 0 1 0 . 0 0 1 2 . 0 0 1 4 . 0 0 1 6 . 0 0 1 8 . 0 0

1 e + 0 7

2 e + 0 7

3 e + 0 7

4 e + 0 7

5 e + 0 7

6 e + 0 7

7 e + 0 7

8 e + 0 7

9 e + 0 7

1 e + 0 8

1 . 1 e + 0 8

1 . 2 e + 0 8

1 . 3 e + 0 8

1 . 4 e + 0 8

1 . 5 e + 0 8

1 . 6 e + 0 8

1 . 7 e + 0 8

1 . 8 e + 0 8

1 . 9 e + 0 8

T i m e

R e s p o n s e _

G C 3 - 6 7 0 7 . D \ E C D 2 B

22

21

20

19

18

17

16

1514

13

12

11

10

9

8

7

5,6

4

3

2

1

Column: DB-XLB30m x 0.32mm i.d., 0.25µm

Carrier: He, constant flow, 38 cm/s at 120°CInjector: Pulsed Splittless, 220 °C

Pulse pressure & time: 35psi for 1.15min2µL, 50ppb

Oven: 120°C for 1.17min120°C to 160°C at 25°/min160°C to 260°C at 10°/min260°C to 300°C (4min) at 15°/min

Detector: µ-ECD, 320°CAr/CH4 (P5) makeup gas at 60mL/min

<16 minutes

CLP-Pesticides - Original “Improved” Method 0.32mm I.D., Helium Carrier Gas

Page 33: GC Chromatography in the Fast Lane - Agilent · Chromatography in the Fast Lane Mark Sinnott Application Engineer December 13, 2018 1 Fast GC. Page 2 Questions to Ask ... Fast Analysis

Page 33

Input Original Method Parameters

Page 34: GC Chromatography in the Fast Lane - Agilent · Chromatography in the Fast Lane Mark Sinnott Application Engineer December 13, 2018 1 Fast GC. Page 2 Questions to Ask ... Fast Analysis

Page 34

New Velocity

New Temp. Program

Same Column and Gas TypeFast Analysis (OPGV)

Transfer same dimensions

Page 35: GC Chromatography in the Fast Lane - Agilent · Chromatography in the Fast Lane Mark Sinnott Application Engineer December 13, 2018 1 Fast GC. Page 2 Questions to Ask ... Fast Analysis

Page 35

Only Time!

<14 minutes

No Real Changes in Resolution!

Same Column and Gas TypeFast Analysis (OPGV)

2 . 0 0 4 . 0 0 6 . 0 0 8 . 0 0 1 0 . 0 0 1 2 . 0 0 1 4 . 0 0 1 6 . 0 0 1 8 . 0 0

1 e + 0 7

2 e + 0 7

3 e + 0 7

4 e + 0 7

5 e + 0 7

6 e + 0 7

7 e + 0 7

8 e + 0 7

9 e + 0 7

1 e + 0 8

1 . 1 e + 0 8

1 . 2 e + 0 8

1 . 3 e + 0 8

1 . 4 e + 0 8

1 . 5 e + 0 8

1 . 6 e + 0 8

1 . 7 e + 0 8

1 . 8 e + 0 8

1 . 9 e + 0 8

T i m e

R e s p o n s e _

G C 3 - 6 7 0 7 . D \ E C D 2 B

Before

Page 36: GC Chromatography in the Fast Lane - Agilent · Chromatography in the Fast Lane Mark Sinnott Application Engineer December 13, 2018 1 Fast GC. Page 2 Questions to Ask ... Fast Analysis

Page 38

Same Column, Hydrogen Carrier GasTranslate Only

Page 37: GC Chromatography in the Fast Lane - Agilent · Chromatography in the Fast Lane Mark Sinnott Application Engineer December 13, 2018 1 Fast GC. Page 2 Questions to Ask ... Fast Analysis

Page 39

Same Column, H2 Carrier GasTranslate Only

33% faster!

11 minutes

No Real Changes in Resolution

2 . 0 0 4 . 0 0 6 . 0 0 8 . 0 0 1 0 . 0 0 1 2 . 0 0 1 4 . 0 0 1 6 . 0 0 1 8 . 0 0

1 e + 0 7

2 e + 0 7

3 e + 0 7

4 e + 0 7

5 e + 0 7

6 e + 0 7

7 e + 0 7

8 e + 0 7

9 e + 0 7

1 e + 0 8

1 . 1 e + 0 8

1 . 2 e + 0 8

1 . 3 e + 0 8

1 . 4 e + 0 8

1 . 5 e + 0 8

1 . 6 e + 0 8

1 . 7 e + 0 8

1 . 8 e + 0 8

1 . 9 e + 0 8

T i m e

R e s p o n s e _

G C 3 - 6 7 0 7 . D \ E C D 2 B

Before

Page 38: GC Chromatography in the Fast Lane - Agilent · Chromatography in the Fast Lane Mark Sinnott Application Engineer December 13, 2018 1 Fast GC. Page 2 Questions to Ask ... Fast Analysis

Page 42

New Column Dimensions, H2 Gas, Translate Only

Input NEW dimensions

New Velocity

New Temp. Program

Close enough (got lucky)

Page 39: GC Chromatography in the Fast Lane - Agilent · Chromatography in the Fast Lane Mark Sinnott Application Engineer December 13, 2018 1 Fast GC. Page 2 Questions to Ask ... Fast Analysis

Page 43

New Column Dimensions, H2 Gas, Translate Only

Page 40: GC Chromatography in the Fast Lane - Agilent · Chromatography in the Fast Lane Mark Sinnott Application Engineer December 13, 2018 1 Fast GC. Page 2 Questions to Ask ... Fast Analysis

Page 44

10

11

5

6

New Column Dimensions, H2 Gas, Translate Only

Page 41: GC Chromatography in the Fast Lane - Agilent · Chromatography in the Fast Lane Mark Sinnott Application Engineer December 13, 2018 1 Fast GC. Page 2 Questions to Ask ... Fast Analysis

Page 45

New Velocity

New Temp. Program

New Column Dimensions, H2 Gas, Fast Analysis

Page 42: GC Chromatography in the Fast Lane - Agilent · Chromatography in the Fast Lane Mark Sinnott Application Engineer December 13, 2018 1 Fast GC. Page 2 Questions to Ask ... Fast Analysis

Page 46

0 . 5 0 1 . 0 0 1 . 5 0 2 . 0 0 2 . 5 0 3 . 0 0 3 . 5 0 4 . 0 0 4 . 5 0 5 . 0 0 5 . 5 0 6 . 0 0 6 . 5 0 7 . 0 0 7 . 5 0

2 0 0 0 0 0 0

3 0 0 0 0 0 0

4 0 0 0 0 0 0

5 0 0 0 0 0 0

6 0 0 0 0 0 0

7 0 0 0 0 0 0

8 0 0 0 0 0 0

9 0 0 0 0 0 0

1 e + 0 7

1 . 1 e + 0 7

1 . 2 e + 0 7

1 . 3 e + 0 7

1 . 4 e + 0 7

1 . 5 e + 0 7

1 . 6 e + 0 7

1 . 7 e + 0 7

1 . 8 e + 0 7

1 . 9 e + 0 7

2 e + 0 7

2 . 1 e + 0 7

2 . 2 e + 0 7

2 . 3 e + 0 7

2 . 4 e + 0 7

2 . 5 e + 0 7

2 . 6 e + 0 7

T i m e

R e s p o n s e _

G C 8 - 3 3 4 5 . D \ E C D 2 BNew Column Dimensions, H2 Gas, Fast Analysis

Better Resolution and Faster Analysis!

Page 43: GC Chromatography in the Fast Lane - Agilent · Chromatography in the Fast Lane Mark Sinnott Application Engineer December 13, 2018 1 Fast GC. Page 2 Questions to Ask ... Fast Analysis

Page 47

New Column Dimensions, H2 Gas, Higher Velocities

Page 44: GC Chromatography in the Fast Lane - Agilent · Chromatography in the Fast Lane Mark Sinnott Application Engineer December 13, 2018 1 Fast GC. Page 2 Questions to Ask ... Fast Analysis

Page 48

105 cm/sec

95 cm/sec

85 cm/sec

<6 minutes

5.5 minutes!?

New Column Dimensions, H2 Gas, Higher Velocities

Page 45: GC Chromatography in the Fast Lane - Agilent · Chromatography in the Fast Lane Mark Sinnott Application Engineer December 13, 2018 1 Fast GC. Page 2 Questions to Ask ... Fast Analysis

Page 49

0 . 5 0 1 . 0 0 1 . 5 0 2 . 0 0 2 . 5 0 3 . 0 0 3 . 5 0 4 . 0 0 4 . 5 0 5 . 0 0 5 . 5 0 6 . 0 0 6 . 5 0 7 . 0 0

2 0 0 0 0 0 0

4 0 0 0 0 0 0

6 0 0 0 0 0 0

8 0 0 0 0 0 0

1 e + 0 7

1 . 2 e + 0 7

1 . 4 e + 0 7

1 . 6 e + 0 7

1 . 8 e + 0 7

2 e + 0 7

2 . 2 e + 0 7

2 . 4 e + 0 7

2 . 6 e + 0 7

2 . 8 e + 0 7

3 e + 0 7

3 . 2 e + 0 7

3 . 4 e + 0 7

3 . 6 e + 0 7

T i m e

R e s p o n s e _

G C 8 - 3 3 6 7 . D \ E C D 2 B

22

21

20

19

18

17

161514

13

12

11

109

8

7

6

5

4

3

2

1

Column: DB-XLB20m x 0.18mm i.d., 0.18µm

Carrier: H2, constant flow, 77.3cm/s at 120°CInjector: Pulsed Splittless, 220 °C

Pulse pressure & time: 35psi for 0.5minFlow ramp at 6.25min of 99mL/min2 to 3mL/min2mm i.d. liner0.5µL, 50ppb

Oven: 120°C for 0.49min120°C to 160°C at 59.4°/min160°C to 260°C at 23.7°/min260°C to 300°C (1.69min) at 35.6°/min

Detector: µ-ECD, 320°CAr/CH4 (P5) makeup gas at 60mL/min

Final Method Used at EPA

Page 46: GC Chromatography in the Fast Lane - Agilent · Chromatography in the Fast Lane Mark Sinnott Application Engineer December 13, 2018 1 Fast GC. Page 2 Questions to Ask ... Fast Analysis

Page 50

0 . 5 0 1 . 0 0 1 . 5 0 2 . 0 0 2 . 5 0 3 . 0 0 3 . 5 0 4 . 0 0 4 . 5 0 5 . 0 0 5 . 5 0 6 . 0 0 6 . 5 0 7 . 0 0

2 0 0 0 0 0 0

4 0 0 0 0 0 0

6 0 0 0 0 0 0

8 0 0 0 0 0 0

1 e + 0 7

1 . 2 e + 0 7

1 . 4 e + 0 7

1 . 6 e + 0 7

1 . 8 e + 0 7

2 e + 0 7

2 . 2 e + 0 7

2 . 4 e + 0 7

2 . 6 e + 0 7

2 . 8 e + 0 7

3 e + 0 7

3 . 2 e + 0 7

3 . 4 e + 0 7

3 . 6 e + 0 7

T i m e

R e s p o n s e _

G C 8 - 3 3 6 7 . D \ E C D 2 B

DB-XLB

0 . 5 0 1 . 0 0 1 . 5 0 2 . 0 0 2 . 5 0 3 . 0 0 3 . 5 0 4 . 0 0 4 . 5 0 5 . 0 0 5 . 5 0 6 . 0 0 6 . 5 0 7 . 0 0

2 0 0 0 0 0 0

4 0 0 0 0 0 0

6 0 0 0 0 0 0

8 0 0 0 0 0 0

1 e + 0 7

1 . 2 e + 0 7

1 . 4 e + 0 7

1 . 6 e + 0 7

1 . 8 e + 0 7

2 e + 0 7

2 . 2 e + 0 7

2 . 4 e + 0 7

2 . 6 e + 0 7

2 . 8 e + 0 7

3 e + 0 7

3 . 2 e + 0 7

3 . 4 e + 0 7

3 . 6 e + 0 7

T i m e

R e s p o n s e _

G C 8 - 3 3 6 7 . D \ E C D 1 A

V i e w M o d e : I n t e g r a t i o n

DB-17ms

22

21

20

19

18

17

161514

13

12

11

1098

7

6

5

4

3

2

1

22

21

20

19

1817

1615

14

1312

11

1098

7

6

5

4

32

1

Final Method Used at EPA

Page 47: GC Chromatography in the Fast Lane - Agilent · Chromatography in the Fast Lane Mark Sinnott Application Engineer December 13, 2018 1 Fast GC. Page 2 Questions to Ask ... Fast Analysis

Page 51

Food/Fragrance – Method translation

Page 48: GC Chromatography in the Fast Lane - Agilent · Chromatography in the Fast Lane Mark Sinnott Application Engineer December 13, 2018 1 Fast GC. Page 2 Questions to Ask ... Fast Analysis

Page 52

Food/Fragrance – Method translation

Page 49: GC Chromatography in the Fast Lane - Agilent · Chromatography in the Fast Lane Mark Sinnott Application Engineer December 13, 2018 1 Fast GC. Page 2 Questions to Ask ... Fast Analysis

Page 53

Spearmint Oil

Δ -9.7 min (0.18 mm, He Carrier)

Δ -16.8 min (0.18 mm, H2

Carrier)

10.6 min

17.7 min

3 8 13 18 23 28

(0.25 mm, He Carrier)

Time (min)

27.4 min

Page 50: GC Chromatography in the Fast Lane - Agilent · Chromatography in the Fast Lane Mark Sinnott Application Engineer December 13, 2018 1 Fast GC. Page 2 Questions to Ask ... Fast Analysis

Page 54

10.6 min

17.7 min

3 8 13 18 23 28

27.4 min

(0.25 mm, He Carrier)

Δ -9.7 min (0.18 mm, He carrier)

Δ -16.8 min (0.18 mm, H2 Carrier)

Spearmint Oil – Resolution Check

Page 51: GC Chromatography in the Fast Lane - Agilent · Chromatography in the Fast Lane Mark Sinnott Application Engineer December 13, 2018 1 Fast GC. Page 2 Questions to Ask ... Fast Analysis

Practical Fast GC Applications with Capillary GC ColumnsMay 19, 2009

Page 55

Resolution Maintained

Compound Resolution

Compounds 0.25 mm

Helium

0.18 mm

Helium

0.18 mm

Hydrogen

Sabinene

β-Pinene1.52 1.59 1.56

α-Terpinene

p-Cymene1.61 1.73 1.86

Speed Gain N/A 35% 61%

Page 52: GC Chromatography in the Fast Lane - Agilent · Chromatography in the Fast Lane Mark Sinnott Application Engineer December 13, 2018 1 Fast GC. Page 2 Questions to Ask ... Fast Analysis

Page 56

CONCLUSIONSStationary Phases – Chosen for optimized selectivity

Diameter – Smaller allows shorter length but has less capacity

Make Small Changes – Again think capacity

Carrier Gas – Hydrogen, high velocity, but can still go fast with He

Temperature Program – Scale properly to preserve elution pattern

Method Translation Software – FREE, reliable

Flow Ramp – Increase at end of run for late eluters (if necessary)

System Upgrades – LTM, Capillary Flow Devices (back flush)

Page 53: GC Chromatography in the Fast Lane - Agilent · Chromatography in the Fast Lane Mark Sinnott Application Engineer December 13, 2018 1 Fast GC. Page 2 Questions to Ask ... Fast Analysis

The New Agilent Intuvo 9000 GC System

Page 54: GC Chromatography in the Fast Lane - Agilent · Chromatography in the Fast Lane Mark Sinnott Application Engineer December 13, 2018 1 Fast GC. Page 2 Questions to Ask ... Fast Analysis

Common frustrations with GC

• Measuring column length correctly

• Cutting your column correctly

• How tight it too tight?

• Clipping columns to deal with active sites, then updating retention times

Page 55: GC Chromatography in the Fast Lane - Agilent · Chromatography in the Fast Lane Mark Sinnott Application Engineer December 13, 2018 1 Fast GC. Page 2 Questions to Ask ... Fast Analysis

Common Care and Maintenance Scheme for GC Columns

1. Cut off 6”-1ft of the inlet end of the column.

2. Bake out the column for no more than 2 hours.

3. Cut off more column. (repeat as necessary)

Guard Column

Intuvo…

Change the guard chip

Page 56: GC Chromatography in the Fast Lane - Agilent · Chromatography in the Fast Lane Mark Sinnott Application Engineer December 13, 2018 1 Fast GC. Page 2 Questions to Ask ... Fast Analysis

Innovating the GC Flow Path

Click-and-run direct

connections

To detector

To detector

Conventionalflow path

Intuvoflow path

Inlet Inlet

Guard chip

Intuvo planar column

Gold seal

Nuts and ferrules

Classic capillarycolumn

Flow chip

Page 57: GC Chromatography in the Fast Lane - Agilent · Chromatography in the Fast Lane Mark Sinnott Application Engineer December 13, 2018 1 Fast GC. Page 2 Questions to Ask ... Fast Analysis

Easier and faster maintenance with Intuvo

• No more ferrules

• Direct face seal connections

• Audible and tactile click lets

you know connection is made

• Less unplanned downtime

• Fewer batch reruns and

precious samples lost

Page 58: GC Chromatography in the Fast Lane - Agilent · Chromatography in the Fast Lane Mark Sinnott Application Engineer December 13, 2018 1 Fast GC. Page 2 Questions to Ask ... Fast Analysis

No More

- measuring

- over-tightening

- trimming

Page 59: GC Chromatography in the Fast Lane - Agilent · Chromatography in the Fast Lane Mark Sinnott Application Engineer December 13, 2018 1 Fast GC. Page 2 Questions to Ask ... Fast Analysis

A New Portfolio of GC Consumables

UI Inlet LinersGuard Chip

No-trimcolumn

Intuvo Flow Chips

Smart Keys

Tools & Accessories

Page 60: GC Chromatography in the Fast Lane - Agilent · Chromatography in the Fast Lane Mark Sinnott Application Engineer December 13, 2018 1 Fast GC. Page 2 Questions to Ask ... Fast Analysis

No More:

- Measuring

- Over-tightening

- Trimming

Page 61: GC Chromatography in the Fast Lane - Agilent · Chromatography in the Fast Lane Mark Sinnott Application Engineer December 13, 2018 1 Fast GC. Page 2 Questions to Ask ... Fast Analysis

Tips to Assure a Good Column Installation

1. Finger tighten until only 1 thread on each of the two nuts is showing.

If more than 1 thread is showing, wiggle or re-position the column into place to further finger tighten the nuts to 1 thread.

Page 62: GC Chromatography in the Fast Lane - Agilent · Chromatography in the Fast Lane Mark Sinnott Application Engineer December 13, 2018 1 Fast GC. Page 2 Questions to Ask ... Fast Analysis

Tips to Assure a Good Column Installation

2. Check that the small integrated column nuts on the column are in form fitted place on the heater in the instrument.

3. Click and Run!!

Page 63: GC Chromatography in the Fast Lane - Agilent · Chromatography in the Fast Lane Mark Sinnott Application Engineer December 13, 2018 1 Fast GC. Page 2 Questions to Ask ... Fast Analysis

Smart Key Technology

• Smart chip tells your Intuvo

what you have

• Sets temperature limits for

you

• Keep track of performance

with read/writeable Smart Key

Page 64: GC Chromatography in the Fast Lane - Agilent · Chromatography in the Fast Lane Mark Sinnott Application Engineer December 13, 2018 1 Fast GC. Page 2 Questions to Ask ... Fast Analysis

Where Does it Get Dirty?

Here

Here

Here

Here

Here

Here

Page 78

Not the Column!

Page 65: GC Chromatography in the Fast Lane - Agilent · Chromatography in the Fast Lane Mark Sinnott Application Engineer December 13, 2018 1 Fast GC. Page 2 Questions to Ask ... Fast Analysis

New Autonomous Leak Checking