Page 1: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

URL: www.smatrix.com

Nordic Users Training – Helsinki – Sept. 8, 2011

Richard VerseputPresidentS-Matrix CorporationEureka, CA USA

Fusion AE – Quality by Design Software for LC Method Development

Page 2: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

Nordic Users Training – Helsinki – Sept. 2011

Thursday, September 8, 2011

15:00 – 15:30 Why QbD for LC Method Development

15:30 – 16:00 S-Matrix Quality by Design software – the complete solution for your LC Method Development workflow

16:00 – 16:30 What's new with Fusion AE Version 9.6

16:30 – 17:00 What does the future hold? (Interactive)

Page 3: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

LC Process Flow

Separation(Column)

AcceptableVariation

Raw MaterialComposition

(Mobile Phase)

HeatingChamber

(Column Oven)

Measurement(Detector)

X─API Resolution

= variation around setpoint

QbD for Method Development ? LC – a “Process in a Box

Method A - Small VariationMethod B - Large Variation 3

Page 4: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

Why QbD for Method Development ? ICH Q8 (R2)

Objective of the Guideline:

… The guideline also indicates areas where the demonstration of greaterunderstanding of pharmaceutical and manufacturing sciences can create a basis forflexible regulatory approaches. The degree of regulatory flexibility is predicated onthe level of relevant scientific knowledge provided.

Pharmaceutical Development:

… The information and knowledge gained from pharmaceutical developmentstudies and manufacturing experience provide scientific understanding to support theestablishment of the design space*, specifications, and manufacturing controls.

Page 5: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

Design Space:

The multidimensional combination and interaction of input variables (e.g., materialattributes) and process parameters that have been demonstrated to provide assuranceof quality. Working within the design space is not considered as a change. Movementout of the design space is considered to be a change and would normally initiate aregulatory post approval change process. Design space is proposed by the applicantand is subject to regulatory assessment and approval.

Formal Experimental Design:

A structured, organized method for determining the relationship between factorsaffecting a process and the output of that process. Also known as “Design ofExperiments”.

Fusion AE – Formal Experimental Design engines driving automated LC method development experimentation

Why Fusion AE for Method Development ? ICH Q8 (R2)

Page 6: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

Experiment run on HPLC in walk-away mode.

CDS generates chromatogram results.

6

Automated analysis, graphing, and reporting.

Report output formats: RTF, DOC, HTML, PDF.

Experiment Design

Ready-to-runmethods & sequences

File-less Data Exchanges

Steps 1 and 2 Step 3

Step 4

Step 5

Fusion AE – Automation Supported QbD Workflow

Page 7: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

• Select study variables

• Define study variable ranges

Define

Experimental

Region

• Build experimental design

• Run & Analyze results build equations

Develop

Knowledge

Space

• Define best method conditions

• Establish & verify robust design spaceEstablish

Design Space

A Quality-by-Design experimental approach consists of four distinct steps

1

2

3

• Define & Document Operating Space

• Establish Process SOP

Establish Operating

Space4

7

Page 8: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

FORMAL EXPERIMENTAL DESIGN

“A structured, organized method for determining the relationship between factors affecting a process and the output of that process. Also known as “Design of Experiments.”

[ICH Q8(R2) - Guidance for Industry, Pharmaceutical Development, August, 2009]

Simple, Template-driven Experiment Setup

8

Page 9: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

FORMAL EXPERIMENTAL DESIGN

“A structured, organized method for determining the relationship between factors affecting a process and the output of that process. Also known as “Design of Experiments.”

[ICH Q8(R2) - Guidance for Industry, Pharmaceutical Development, August, 2009]

Instant One-clickExperiment DesignGeneration

9

Page 10: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

Automation Reduces Risk

“Eliminate manual data transcription and checking via fully-automated, e-signature controlled, and fully automated data exchanges between regulatory compliant applications.”

Automated, Regulatory Compliant Data Exchanges

ChromatographyData Systems

Experiment Designs

ChromatogramResults

10

Page 11: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

Developing the Knowledge Space“The information and knowledge gained from pharmaceutical development studies and manufacturingexperience provide scientific understanding to support the establishment of the design space,specifications, and manufacturing controls.”

[ICH Q8(R2) - Guidance for Industry, Pharmaceutical Development, August, 2009]

pH

Oven Temp

Initial%

5.0

15.0

30.0 40.0

Rs = 9.3 + 4.2(X1) -5.4(X2)2 + 12.7(X4)2 + 1.3(X1*X3) + 1.6(X1)2X 2+ …

Linear Additive Effects Curvature Effects Higher-order Effects

11

Page 12: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

INITIAL DESIGN SPACE – MEAN PERFORMANCE

QbD Design Space - “The multidimensional combination and interaction of input variables (e.g., material attributes) and process parameters that have been demonstrated to provide assurance of quality.”

[ICH Q8(R2) - Guidance for Industry, Pharmaceutical Development, August, 2009]

Design Space (Un-shaded Region):

• Mean Performance Only

Edge of Failure –

Mean Performance

12

Page 13: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

FINAL DESIGN SPACE – MEAN PERFORMANCE + ROBUSTNESS

QbD Design Space - “The multidimensional combination and interaction of input variables (e.g., material attributes) and process parameters that have been demonstrated to provide assurance of quality.”

[ICH Q8(R2) - Guidance for Industry, Pharmaceutical Development, August, 2009]

Edges of Failure –Method RobustnessFinal Design Space:

• Mean Performance• Robustness

13

Page 14: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

[ICH Q8(R2) - Guidance for Industry, Pharmaceutical Development, August, 2009]

14[ICH Q8(R2) - Page 23

QbD “DESIGN SPACE”

QbD Design Space - “The multidimensional combination and interaction of input variables (e.g., material attributes) and process parameters that have been demonstrated to provide assurance of quality.”

Curves defining the acceptable performing regions, and the design space (region of overlap), are generated by equations (models)

Page 15: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

OPERATING SPACE – Design Space + Control Strategy

“CONTROL STRATEGY – the planned set of controls, derived from current product and process understanding thatensures process performance and product quality.”

VerificationRuns

[ICH Q8(R2) - Guidance for Industry, Pharmaceutical Development, August, 2009]

Operating Space

• Proven AcceptableRanges

15

Page 16: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

16

What’s New in Fusion AE – Version 9.5.0

Support for the Waters Acquity H-Class UPLC System and additional Waters Detectors

• Waters Acquity H-Class Column Manager (all stack configurations)

• Waters Acquity H-Class Direct Inject Sample Manager

• Waters Acquity H-Class Quaternary Solvent Manager

Page 17: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

17

What’s New in Fusion AE – Version 9.5.0

Implement new User Interface for the Pump Program

• Support for 2 – Step Gradient Studies

• Optional Step Inclusion/Exclusion Toggles

• Equilibration Time Integration

• Isocratic Pump Program Integration

• Gradient Slope / Gradient Curve Controls

• Pump Program Chart

• Coordination of Pump and Solvent Settings

Page 18: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

18

What’s New in Fusion AE – Version 9.5.0

Modifications to Solvents Settings Controls and Logic

• Strong Solvent Settings Dominance

• Support for More Than 1 Weak Solvent

• Solvent Type Labels based on the selected Chromatography type.

• Integration of the ‘Mobile Phase Blend Study’ factor

• Available Reservoirs Selection

Page 19: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

19

What’s New in Fusion AE – Version 9.5.0

Modifications to Aqueous Linked Factor System Settings Control

• Consolidated UI for Aqueous-linked Variables

• Metadata for pH and Buffer Strength

Page 20: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

20

What’s New in Fusion AE – Version 9.6.0

Advancements in Connectivity and Modeling

• Windows 7 and Windows Server 2008 Compatibility

• Empower 3 Compatibility

• Citrix ZenApp 5 and 6 Compatibility – Certified Citrix Partner

Page 21: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

21

What’s New in Fusion AE – Version 9.6.0

Advanced pH effects Modeling

Fairly Stable Region

1.00

2.00

9.0

pH

Rs

6.53.0

Desired Level of Characterization – the workable region for an optimization study.

Most Stable Region Least Stable Region

Page 22: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

22

What’s New in Fusion AE – Version 9.6.0

New Graphical Optimizer with Operating Space and Advanced Reporting

Page 23: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

23

What’s New in Fusion AE – Version 9.6.0

New Fusion Product Development Module

Tablet Coater Dissolution(Time Release)

Processing(Immediate Release)

Page 24: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

Nordic Users Training – Helsinki – Sept. 2011

Friday, September 9, 2011

07:00– 08:00 Breakfast08:00 – 09:30 Orientation to Fusion LC Method Development (FMD)

Design a column/solvent/pH screening experimentOrientation to the Screening Experiment chromatogram processing

protocol within Empower09:30 – 10:00 Analyze the Screening Experiment results – identify best pH range, column,

mobile phase, and initial gradient conditions to promote to optimization10:00 – 10:30 Coffee Break

10:30 – 12:00 Design a method optimization experimentOrientation to the Optimization Experiment chromatogram processing

protocol within EmpowerAnalyze the Optimization Experiment results – identify best pH, temperature, and

final gradient conditions in terms of mean method performance12:00 – 12:45 Use Robustness Simulator to enter (1) expected variation of the study parameters,

and (2) method robustness performance requirementsConnect mean performance and robustness requirements to identify FDA Design

Space and Safe Operating Ranges for method transfer12:45 – 13:00 Summary and end of training

Page 25: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

25

Design of Experiments – an equation building methodology.

Equations – are derived from real experimental data.

Knowledge – quantitative characterization of the study parameter effects on the critical performance characteristics.

FORMAL EXPERIMENTAL DESIGN

“A structured, organized method for determining the relationship between factors affecting a process and the output of that process. Also known as “Design of Experiments.”

[ICH Q8(R2) - Guidance for Industry, Pharmaceutical Development, August, 2009]

Page 26: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

26

Linear Equation (Model): Y = m(x) +b

1.00

2.00

40.0Oven Temp (°C)

Rs

y intercept = b0

0.0

∆y

30.0∆x

Rs = b0 + b1(x1)

1.80

1.30

∆y∆xSlope (b1) =

0.5010.00=

= 1.15 + 0.05(°C)

Page 27: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

)x(bbR 110s +=

Example Equation

27

● Linear Effect – One Factor

Page 28: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

)x(b)x(bbR 22110s ++=

Example Equation

28

● Linear Additive Effects – 2 Factors

(X1)

(X2)

Page 29: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

Example Equation

29

● Pairwise Interaction Effect )x(bbR 110s +=

)x(bbR 110s +=

pH = 4.0

pH = 6.0

Page 30: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

Example Equations

30

● Pairwise Interaction Effect

)xx(b)x(b)x(bbR 2*11222110s +++=

Page 31: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

31

2111110s )x(b)x(bbR ++=

● Curvature

Example Equation

Page 32: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

32

)x()x(b)x(b)x(b)x(b)x(bbR 22

11122

222222

111110s +++++=

● Curvature + Interaction = Complex Combined Effects

Example Equation

Page 33: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

33

● Equation Expressed Across the Combined Study Ranges

Response Surface Graph

Page 34: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

INITIAL DESIGN SPACE – MEAN PERFORMANCE

QbD Design Space - “The multidimensional combination and interaction of input variables (e.g., material attributes) and process parameters that have been demonstrated to provide assurance of quality.”

[ICH Q8(R2) - Guidance for Industry, Pharmaceutical Development, August, 2009]

Design Space (Un-shaded Region):

• Mean Performance Only

Edge of Failure –

Mean Performance

34

Page 35: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

35

Mean Performance Versus Robustness

Condition A – Good RobustnessCondition B – Poor Robustness

Conditions A and B – Identical Mean Performance ≠ identical robustness

Page 36: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

FINAL DESIGN SPACE – MEAN PERFORMANCE + ROBUSTNESS

QbD Design Space - “The multidimensional combination and interaction of input variables (e.g., material attributes) and process parameters that have been demonstrated to provide assurance of quality.”

[ICH Q8(R2) - Guidance for Industry, Pharmaceutical Development, August, 2009]

Edges of Failure –Method Robustness

Final Design Space:

• Mean Performance• Robustness

36

Page 37: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

OPERATING SPACE – Design Space + Control Strategy

“CONTROL STRATEGY – the planned set of controls, derived from current product and process understanding thatensures process performance and product quality.”

VerificationRuns

[ICH Q8(R2) - Guidance for Industry, Pharmaceutical Development, August, 2009]

Operating Space

• Proven AcceptableRanges

37

Page 38: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

38

“Difficult to Control” Parameters Next = LC Configuration change

• pH (manually switching reservoirs usually required)• Organic Solvent Type• Ion Pairing Agents (manually switching reservoirs usually required)• Column Type (manually switching columns required)

Traditional Approach to Method Development

Usually Done by Sequential Studies

“Easy to Control” Parameters 1st = Changeable in the method

• Gradient Time• Temperature

Even when studied together – traditional approachesdo NOT characterize interactions!}

Page 39: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

39

Phase 1 ─ Column/Solvent Screening = Major Effectors

• Gradient Time (constant slope - 5%-95%)• pH (wide range – automated solvent switching)• Column Type (multiple columns – automated column switching)• Solvent Type (w/wo blending)

Phase 2 ─ Method Optimization = Additional Effectors

• Pump Flow Rate• Gradient Slope (vary starting or end point % Organic) • Ion Pairing Agents (can be included)• pH (narrow range - robustness optimization)• Temperature (can be included in Phase 1)

QbD – a Risk-based Approach to Method Development

Capitalizes on automation!

Page 40: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

40

Column/Solvent/pH Screening

Experimental Design

Page 41: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

Screening Experiment – Design

41

Page 42: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

Experiment Variable Range or Level SettingsPump Flow Rate .5 mL/min

Injection Volume 1.0 uL

Gradient Time (min) 3.0 — 9.0 min Reasonable range – from Sample Workup

pH 2.0, 3.0, 4.0, 5.0, 6.0, 7.0 Wide range is screened – recommend 5-6 levelspKa of Primary Compound – 4.8

Column Temperature 30 C

Column Type Four Columns: Column screening – recommend wide selectivity rangeBEH C18 – all columns support pH rangeBEH Shield RP 18BEH PhenylBEH C8

Organic Solvents Mobile Phase A1-1: Aqueous Buffer, pH 2.0Mobile Phase A1-2: Aqueous Buffer, pH 3.0Mobile Phase A1-3: Aqueous Buffer, pH 4.0Mobile Phase A1-4: Aqueous Buffer, pH 5.0Mobile Phase A1-5: Aqueous Buffer, pH 6.0Mobile Phase A1-6: Aqueous Buffer, pH 7.0

Mobile Phase B1: AcetonitrileMobile Phase B2: Methanol

Experiment Setup

42

Page 43: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

Experiment Setup – LC Instrument Centric Template

43

Page 44: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

44

Experiment Setup – LC Instrument Centric Template

Page 45: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

45

Experiment Setup – LC Instrument Centric Template

Enter a pKa value only when the pKa is within the experimental range of pH.

Page 46: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

Generate Design

One Click:

Software maps the experimental design to the study factors.

46

C4

C3

C2

Gradient Time

pH

2.0

7.0

3.0 9.0

4.5

C1

Page 47: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

Generate Design – Statistical Efficiency

47

5 levels of Gradient Time

6 levels of pH

4 levels of Column Type

2 levels of Strong Solvent Type

5x6x4x2 = 240 possible combinations

Fusion Screening design = 46 runs

~ 5x efficiency.

Page 48: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

Export Wizard:

Software automatically reconstructs experimental design within the chromatography data software (CDS) as instrument methods and sample sets.

Export Design to the CDS

48

Page 49: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

49

Overnight Execution in Walk-away Mode

Solvent Selection Valve4-RelayPanel

LANCard

CustomerNetwork

Empower™

Page 50: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

• Column screening experiments, even those done by DOE, often have significant inherent data loss in critical results such as resolution.

• The data loss is due to both compound co-elution and also changes in compound elution order (peak exchange) between experiment trials.

These changes are often due to the major effects that pH and organic solvent type can have on column selectivity.

Complexity of Traditional Results Data

50

Page 51: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

Inherent Data Loss – Co-elution and Peak Exchange

LC Method Development Traditionally Involves Peak Tracking

But how do you track moving or disappearing peaks? 51

Page 52: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

2.27

5

2.86

2

3.22

13.

302

3.52

9

4.15

1

5.52

0 6.76

2

AU

0.00

0.50

1.00

1.50

2.00

Minutes0.00 1.00 2.00 3.00 4.00 5.00 6.00 7.00 8.00 9.00 10.00 11.00 12.00

1.81

0

2.01

5

2.50

42.

646

3.40

7

4.30

1

5.10

7

5.78

0

AU

0.00

0.50

1.00

1.50

2.00

2.50

Minutes0.00 0.50 1.00 1.50 2.00 2.50 3.00 3.50 4.00 4.50 5.00 5.50 6.00 6.50 7.00 7.50 8.00 8.50 9.00

Inherent Data Loss – Co-elution and Peak Exchange

Trial 37

52

Trial 41

Co-elution / Peak Exchange

Page 53: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

Run No. Gradient Time pH Column Type Imp E - USPResolution Imp F - USPResolution Imp G - USPResolution1.a.1.a 8.8 2 C18 1.4 2.32.a.1.a 6.3 2 Phenyl 1.783.a.1.a 10 2 C18 1.45 2.364.a.1.a 5 2 C18 1.155.a.1.a 10 2 Phenyl 1.06 2.256.a.1.a 5 2 Phenyl 1.667.a.1.a 10 2 RP8.a.1.a 5 2 RP9.a.1.a 7.5 2 RP 2.9510.a.1.a 7.5 4.5 C18 0.97 1.18 3.811.a.1.a 7.5 4.5 Phenyl 1.24 2.2712.a.1.a 7.5 4.5 RP 2.08 2.1513.a.1.a 5 4.5 RP 0.9814.a.1.a 7.5 4.5 C18 1 2.63 3.8515.a.1.a 7.5 4.5 Phenyl 1.29 2.2616.a.1.a 7.5 4.5 RP 2.0817.a.1.a 5 7 C18 2.3518.a.1.a 10 7 Phenyl 1.45 1.08 2.4519.a.1.a 5 7 Phenyl 2.0320.a.1.a 10 7 RP 3.0521.a.1.a 5 7 RP 1.0222.a.1.a 8.8 7 Phenyl 1.42 2.3423.a.1.a 6.3 7 RP 1.5424.a.1.a 10 7 C18 1.89 2.99 2.8225.a.1.a 10 7 C18 1.87 2.95 2.8126.a.1.a 5 7 C18

53

Peak exchange = inconsistent resolution data.

Co-elution = missing peaks = missing results.

Inherent Data Loss – Co-elution and Peak Exchange

Page 54: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

Inherent Data Loss Can Not Model Chromatography

Compound Name R2-Adj. Value

Impurity F 0.1785

Impurity H 0.9125

54

Inaccurate Predictions

Accurate Predictions

Problem – data often not accurately modeled.

Result – Phase 1 is reduced to a “Pick the Winner” Strategy.

Regression Model Statistics

On a scale of Zero to One (0 – 1), with 1.0000 = the perfect model:

Page 55: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

55

Trend Responses™

Page 56: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

56

QbD Requires Good Data.

Chromatogram Processing Protocol – Trend Responses

Page 57: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

57

Chromatogram Processing Protocol – Trend Responses

Page 58: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

Import Results from CDS

58

One Click:

Software automatically imports results data from the CDS.

Page 59: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

Import Wizard. Multiple Results Sets™

59

When the experiment is constructed as two or more sample sets due to instrument automation limitations, a single Import Results operation can bring in the results from all results sets linked to the experiment.

Page 60: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

Import Wizard. Auto-computed Trend Responses™

60

All CDS Responses can be utilized throughflexible Trend Response operators.

For example:

Max Peak # – can track API peak(s).

Tailing and/or Area can become additionalmetrics of method performance.

Page 61: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

Analysis Wizard. Automated Mode

61

Page 62: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

pH Level (X1)

Model Results

One Click: Software automatically builds an equation (model) from the CDS resultsfor each critical method performance characteristic (referred to in ICH Q8 as a Critical Quality Attribute, or CQA).

62

222222112

2111110s )x()x()x*x()x()x(R ββββββ +++++=

Gradient Time Level (X2)

Page 63: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

Visualize Results

63

One Click:

Software wizard generates graphical representations of each model – in this case the model of study factor effects on the resolution of a critical peak pair.

Note – effects are not always independent (linearly additive)

Page 64: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

Note: different Critical Method Attributes have different regions of good performance.

64

Trellis Visualization of Effects

Page 65: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

Automated Numerical Search for Optimum Method

65

Page 66: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

Numerical Solution Search – Best Conditions

66

For this design the software conducted 46 separate solution searches.

Answers are ordered from closest to furthest away in terms of simultaneously achieving all defined goals.

Page 67: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

An Overlay graph of all responses for which method performance goals are defined is automatically generated.

For this presentation we will “build” the overlay graph one response at a time.

Graphical Solution Search – Best Region Overall

67

Page 68: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

Fusion AE Overlay Graph.

Each color on the graph corresponds to a response for which goals have been defined.

A region shaded with a given color shows the study variable level setting combinations that will NOT meet the goals for the corresponding response.

Note: the un-shaded region corresponds to level setting combinations that meet all response goals.

Note: Shaded region indicates pHand Gradient Time combinations that do NOT meet performance requirements.

68

Graphical Optimizer

Page 69: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

Graphical Optimizer

Unshaded Region WithPredicted Best Settings:

~Gradient Time = 6.0 minpH = 3.0Mobile Phase = MeOHColumn = BEH Shield RP18

69

Page 70: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

Graphical Optimizer

Unshaded Region WithPredicted Best Settings:

~Gradient Time = 6.0 minpH = 3.0Mobile Phase = ACNColumn = BEH Shield RP18

70

Page 71: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

Point Predictions

71

Page 72: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

1.47

5

1.85

0

1.99

92.

105

2.21

2

2.50

5

3.16

5

3.80

4

4.37

0

AU

0.00

0.50

1.00

1.50

2.00

2.50

3.00

Minutes0.00 0.50 1.00 1.50 2.00 2.50 3.00 3.50 4.00 4.50 5.00 5.50 6.00 6.50 7.00 7.50 8.00 8.50 9.00

Analyze Results to Define Variable Levels for Optimization

72

1.72

1

1.91

4

2.03

0

2.16

3

2.26

6

2.43

7

2.85

6

3.23

5

3.53

4

AU

0.00

0.50

1.00

1.50

2.00

2.50

3.00

Minutes0.00 0.50 1.00 1.50 2.00 2.50 3.00 3.50 4.00 4.50 5.00 5.50 6.00

Methanol

Acetonitrile

There is still optimization work to do for 2 critical peak pairs.

Page 73: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

73

Method Optimization

Experimental Design

Page 74: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

74

Results: Optimize method for mean performance AND robustness.

DOE Optimization Design: Rapid Method Development Template 2Use column, pH, and solvent type defined in Phase 1.

Chromatographic Performance: Step 1. Model all factor effects on chromatographic characteristics.Step 2. Automatically compute robustness of each experiment method.Step 3. Model all factor effects on method robustness.

Note: no additional experimentsneeded to address robustness

Robustness Demonstration Experiment

Method Validation

MeetsALL PerformanceRequirements?

Yes

No

QbD Strategy

Phase 1 – Column/Solvent Screening

Phase 2 – Method Optimization

Page 75: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

75

Optimization Experiment – Planning

Page 76: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

Optimization Experiment – Design

76

Page 77: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

Experiment Variable Range or Level SettingsPump Flow Rate (mL/min) 0.5 mL/min

Injection Volume 1.0 uL

Gradient Slope(End Point % Organic)

Vary Starting point = 5.0 — 15.0 Vary Starting Point – Separation issues in1st half of chromatogram

End Point = 70.0%

Gradient Time (min) 6.0 minutes (Constant) Screening study result

pH 2.5, 3.0, 3.5 Recommend at least 3 levels

Oven Temperature 30.0 — 40.0 C Recommend at least 5.0 C

Column Type One ColumnBEH Shield RP 18 Screening study result

Organic Solvents Mobile Phase A1-1: Aqueous Buffer, pH 2.5Mobile Phase A1-2: Aqueous Buffer, pH 3.5Mobile Phase A1-3: Aqueous Buffer, pH 4.5Mobile Phase B1: Acetonitrile Screening study result

Optimization Experiment – Design

77

Page 78: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

Optimization Experiment – Design

78

Page 79: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

Optimization Experiment – Design

79

Page 80: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

80

Optimization Approach to pH

1.00

3.00

3.5

pH

Rs

3.02.5

2.00

Desired Level of Characterization – Optimum pH for Mean Performance and Robustness.

Page 81: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

Data Analysis

Method Optimization

81

Page 82: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

Import Results Wizard. Single Results Set.

82

Page 83: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

Import Wizard. Tracked-peak Responses

83

Optimization studies normally use identified (tracked) peak results data sets. All results computed from the CDS are available.

Page 84: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

Format Data for Analysis

84

Page 85: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

Analysis Wizard. Automated Mode

85

Page 86: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

Automated Numerical Search for Optimum Method

86

Page 87: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

Automated Numerical Search for Optimum Method

87

Page 88: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

Analyze Results to Define “Optimum” Method

Chromatogram = Predicted Best Conditions.

88All critical Peaks Well Resolved. Excellent Mean Performance.

Page 89: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

Graphical Optimizer – Mean Performance

89

Edges of Failure

Page 90: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

Graphical Optimizer – Mean Performance

90

Page 91: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

Integrating Robustness into Method Development

Page 92: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

ICH Q2A – Robustness

FDA – Reviewer Guidance: uses ICH Q2A

ICH Q2A:The robustness of an analytical procedure is a measure of its capacity to remain unaffected by small, but deliberate variations in method parameters and provides an indication of its reliability during normal usage.

In the case of liquid chromatography, examples of typical variations are:

• Influence of variations of pH in a mobile phase• Influence of variations in mobile phase composition• Different columns (different lots and/or suppliers)• Temperature• Flow rate

Note – the text “but deliberate” refer to the deliberate perturbation of critical instrument parameters about their method setpointsdone as part of a Validation-Robustness experiment.

92

Page 93: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

Method ScreeningSelect column & solventby visual inspection of

chromatograms

Method Development• Only evaluate mean

method performance

• Do NOT study parameterinteraction effects

Method ValidationFormal experiment to demonstratemethod robustness

Re-development• Method fails robustness

on validation testing

• Method fails robustness in the field

Traditional Method Development Approach to Robustness

93

Page 94: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

A Revealing Comment

Quality Implementation Working Group on Q8, Q9 and Q10 –Questions & Answers

Q: Does a set of proven acceptable ranges alone constitute a design space?

A: No, a combination of proven acceptable ranges (PARs) developed from univariate experimentation does not constitute a design space [see Q8(R1), Section 2.4.5.].

Proven acceptable ranges from only univariate experimentation may lack an understanding of interactions between the process parameters and/or material attributes. However, proven acceptable ranges continue to be acceptable from the regulatory perspective but are not considered a design space [see ICH Q8(R1) Section 2.4.5].

Quality Implementation Working Group on Q8, Q9 and Q10 -Questions & Answers, Current version dated June 10, 2009

94

Page 95: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

95

Setpoint Variation Envelope Around Target

Actual Nature of Setpoint Variation in Many LC Parameters

5 10 15 20

Method Run Time

Gra

dien

t % O

rgan

ic

Magnitude of shift from target varies from peak to peak. It is not a uniform bias of the chromatogram – puts OFAT approach at risk.

Page 96: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

FDA Reviewer Guidance –Validation of Chromatographic Methods

Methods validation should not be a one-time situation to fulfill Agency filingrequirements, but the methods should be validated and also designed by the developer or user to ensure ruggedness or robustness.

Regulatory Statements and Expectations

ICH Q2A – Text on Validation of Analytical ProceduresIX. ROBUSTNESS (8)The evaluation of robustness should be considered during the development phase and depends on the type of procedure under study. It should show the reliability of an analysis with respect to deliberate variations in method parameters.

96

Page 97: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

Method ScreeningSelect column & solvent

using quantitativeTrend Responses

Formal Method Development & Optimization

Method ValidationFormal experiment to demonstratemethod robustness

Characterize and model ALL study parameter effects on ALL critical method performance attributes

Method MeanPerformance Models

Method RobustnessModels

• Establish ICH Design Space

• Identify Optimal Method

• Establish Operating Space

Fusion QbD Approach to Robustness

97

Page 98: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

The quantitative metric of performance robustness that we will use is derived from traditional Process Capability studies.

The Robustness Metric

Considering the LC Instrument as aProcess-in-a-Box

98

Page 99: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

Process Flow

Separation(Column)

AcceptableVariation

Raw MaterialComposition

(Mobile Phase)

HeatingChamber

(Column Oven)

Measurement(Detector)

X─API Resolution

= variation around setpoint

Method Performance Variation – 100 Injections

Method A - Small VariationMethod B - Large Variation 99

Page 100: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

iationvar6LTLUTLCp σ

−=

Process Capability (Cp) – a direct, quantitative measure of process robustness used routinely in Statistical Process Control (SPC) applications. The classical SPC definition of “Inherent Process Capability” (Cp) is

UTL and LTL = Tolerance Limits (tolerance width).

6σ Variation = ±3σ process output variation.

Traditional Goal ≥ 1.33

- based on setting the UTL and LTL at ±4σ ofmethod performance variation.

- Note: a 6-sigma method would have a Cp = 2.00

Process (LC Method) Capability - Quantified

LTL UTL

API Resolution

6σVariation

X─

100

Page 101: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

Mean Value( X )

Process Capability - Quantified

±3σ

Rs Variation

LTL UTL

±3σ Variation = Tolerance Limit Interval00.166cp ==σσ

101

Page 102: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

Mean Value( X )

Process Capability - Quantified

±3σ

LTL UTL

±4σ

33.168cp ==σσ

Rs Variation

±3σ Variation = 75% of Tolerance Limit Interval

102

Page 103: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

Mean Value( X )

Process Capability - Quantified

±3σ

LTL UTL

±6σ

00.26

12cp ==σσ

Rs Variation

±3σ Variation = 50% of Tolerance Limit Interval

103

Page 104: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

Method Robustness

Class Exercise 8 – Method Robustness Optimization

Page 105: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

Previous Optimization – Mean Performance Only

Design and Operating Spaces – Mean Performance Goals Only

Not knowing the Edges of Failure for Robustness means the Operating Limits must be set “well within” the region of acceptable mean performance. 10

5

Page 106: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

Automated Stepwise Procedure for Generating Cp

S-Matrix Robustness Simulator™

Monte Carlo simulation using DOE-derived Models.

No additional experiment runs required.

We will demonstrate this 4-step process in the following slides.

106

Page 107: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

LCL UCL

Step 1 – Define Confidence Interval Around Setpoint for Each Variable

-0.1 +0.1

Maximum Expected Variation( 3σ Value) = ± 0.01

6σ ≈ 99.7%

Setpoint

E.g. – pH

NOTE - Use SOP for Buffer Preparation.

Expected 3σ Variation Limits

107

Page 108: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

LCL UCL

Maximum Expected Variation( 3σ Value) = ± 2.0

6σ ≈ 99.7%

Expected 3σ Variation Limits

Setpoint

Confidence Interval Around Setpoint

-2.0 +2.0

E.g. – Initial % Organic

NOTE - Use manufacturer’s specs for the ±3σ value or extend it based on the least-capable system which will be used on transfer. 10

8

Page 109: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

NOTE - Use manufacturer’s specs for the ±3σ value or extend it based on the least-capable system which will be used on transfer.

Confidence Intervals Around Setpoints

109

Page 110: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

-0.50 +0.50

USP Resolution

Step 2 – Define Maximum Tolerance Limits for each Response (CQA)

Mean Result Tolerance Limit Delta( distance) = ± 0.50X

IMPORTANT: the Tolerance Limit Delta values define the maximumacceptable limits on method performance variation.

This is normally your System Suitability Specification.

110

Page 111: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

Robustness is calculated for each key performance metric.

NOTE - the Tolerance Limit Delta values define the maximum acceptable limits on method performance variation.

This can be your System Suitability Specification.

Maximum Tolerance Limits Around CQAs

111

Page 112: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

Robustness Simulator* = Predicted Variation

* - U.S. Patent No. 7,606,685 B2

pH (X1)2.5

Initial %(X2)15

Step 3 – Predict Response Variation of each experiment run

E.g. Experiment Run 1 Method

RS

X —

RS = 9.3 + 4.2(X1) + 1.3(X1*X3) + 1.6(X1)2X2 - 5.4(X2)2 + 12.7(X4)2 …

112

Variation AroundSetpoint

Variation AroundSetpoint

Page 113: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

Tolerance Width Delta( distance) = ± 0.50

.I.C6LTLUTLCp σ

−=

0.70 1.20 1.70

Peak 4 – USPResolution

6σ C.I.

0.92 1.4957.000.1

92.049.170.070.1cp =

−−

=

Step 4 – Compute Robustness Cp for each experiment run

Cp = 1.75

E.g. Experiment Run 1 Method

LTL UTL

113

Page 114: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

Run No. Initial % Organic Oven Temperature pH Peak 4 - USPResolution Peak 4 - USPResolution - CpCondition Column - 1 5 30 2.501.a.1.a 5 30 2.50 1.20 1.752.a.1.a 15 30 2.50 2.43 2.963.a.1.a 5 30 2.50 1.20 1.754.a.1.a 15 30 2.50 2.42 2.96Condition Column - 2 10 30 3.005.a.1.a 10 30 3.00 2.77 1.39Condition Column - 3 15 30 3.506.a.1.a 15 30 3.50 5.12 0.627.a.1.a 5 30 3.50 5.54 0.41Condition Column - 4 10 35 2.508.a.1.a 10 35 2.50 2.03 1.91Condition Column - 5 10 35 3.009.a.1.a 10 35 3.00 2.85 1.5910.a.1.a 10 35 3.00 2.85 1.5911.a.1.a 10 35 3.00 2.85 1.5912.a.1.a 15 35 3.00 3.17 2.4813.a.1.a 5 35 3.00 2.36 1.18Condition Column - 6 10 35 3.5014.a.1.a 10 35 3.50 5.50 0.54Condition Column - 7 15 40 2.5015.a.1.a 15 40 2.50 2.56 6.3316.a.1.a 5 40 2.50 1.55 1.6817.a.1.a 5 40 2.50 1.57 1.68Condition Column - 8 10 40 3.0018.a.1.a 10 40 3.00 2.92 1.85Condition Column - 9 5 40 3.5019.a.1.a 5 40 3.50 5.52 0.4720.a.1.a 15 40 3.50 4.39 0.72

Computed ResolutionRobustness

ResolutionResults from CDS

Good mean performance

Poor Robustness

Both sets of results are modeled to identifythe best performing, most robust method.

Step 4 – Robustness Cp for All Experiment Runs

Good mean performance

Good Robustness

114

Page 115: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

Robustness prediction models are automatically derived.

Link the Mean Performance and Robustness equations via the

numerical and graphical optimizers to identify one or more

combinations of variable level settings that meet or exceed mean

performance and performance robustness goals for each response.

Step 5 – Generate a Robustness Cp Prediction Equation for Each CQA

115

Page 116: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

Design and Operating Space – Mean Performance Goals Only

116

Page 117: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

Final Design and Operating Space – Mean Performance + Robustness

117

Page 118: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

Step 3 – Generate the Point Predictions

3. Navigate to the Point Predictions activity, and enter the points into the prediction dialog.

118

Page 119: Fusion AE – Quality by Design Software for LC Method ... · flexible regulatory approaches. The degree of regulatory flexibility is predicated on. the level of relevant scientific

Step 4 – Export the Point Predictions to Empower and Run the Sample Set

4. The Point Predictions wizard will generate the predicted results for each method performance metric, including all robustness metrics. You then export these points to the CDS as a ready-to-run sample set and methods.

Documentation Notes

1. Maintain the verification run chromatograms in the Empower project.

2. Maintain the Fusion Method Development files in an archival system (e.g. NuGenesis).

Verification chromatogram –Operating Space Center Point

119