ftr3008l6 - semspub.epa.gov

228
DflTft UntlDftTXOH D.HQ REPORT RPPBQUftL PROCESS o Data Ualidationi Both the analysts and the nanagers of itach laboratory section review the data generated i* ii»eir laboratory to insure that it neet* all of the applicable analytical and quality control requirement*. Before the data is released to the Project Hanager and Oocunent Control staff, th«: nanager*s review includes the following itensi e Check that the analysis was performed as specified by the client. o Check that all of the required samples were analyzed. o Check that no contaminants were present in blanks at greater than the quality acceptance limits. o Check that the required nunber of blanks, spikes, duplicates, st-nnOarcat etc. were a.nalyzed. o Check that all surrogate and spike recoveries are within quality acceptance limits. o Verify that instrufwot calibration ic acceptable. o Uerify that instrument tuning is acceptable. o Uerify that internal standard responses are acceptable. o Check/werify that any nethod/client specific analytical or quality control par&neters are acceptable. o Docunent any problems or exceptions to the above criteria <ie. Quality Assurance Notices). ftR3008L6 29

Upload: others

Post on 27-Apr-2022

52 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: ftR3008L6 - semspub.epa.gov

DflTft UntlDftTXOH D.HQ REPORT RPPBQUftL PROCESS

o Data Ualidationi

Both the analysts and the nanagers of itach laboratory sectionreview the data generated i* ii»eir laboratory to insure that itneet* all of the applicable analytical and quality controlrequirement*. Before the data is released to the ProjectHanager and Oocunent Control staff, th«: nanager*s reviewincludes the following itensi

e Check that the analysis was performed as specified by theclient.

o Check that all of the required samples were analyzed.

o Check that no contaminants were present in blanks atgreater than the quality acceptance limits.

o Check that the required nunber of blanks, spikes,duplicates, st-nnOarcat etc. were a.nalyzed.

o Check that all surrogate and spike recoveries are withinquality acceptance limits.

o Verify that instrufwot calibration ic acceptable.

o Uerify that instrument tuning is acceptable.

o Uerify that internal standard responses are acceptable.

o Check/werify that any nethod/client specific analytical orquality control par&neters are acceptable.

o Docunent any problems or exceptions to the above criteria<ie. Quality Assurance Notices).

ftR3008L629

Page 2: ftR3008L6 - semspub.epa.gov

ccRTxrxcflTXOits RHO PEEFW&IRHCE

CKDUCST regularly analyze* a 4*a.ri«ty of internally and externallyprovided QC sarrplev . -The *ubra**i«n, analysis, *nd reviewing ofinternal QC «e**ples is nonitored by the Quality Assurance Officera>nd the Uicr Prtgittent of Technitwl Services. The submission Andanalysis of •*xt«rnal QC *anple* is also nonitored by the Qualityftssurance Officer and tHe 'Dice-President of Technical Services,but the results are reviewed by the submitting party. These•anples cone fron a variety of sources, which include, tut ar* notlinitad by, the following exanplvs.

o Internal QC Sanplest

o Katrix spike^Xnatrix spike duplicatus

o Blank spikes

o Blind spikes

o External QC San&lesi

o CPB QC artpul**

o CPR blind eon-tract spikes*

o KPA erX«dic ssorrtract *udit samples

o State QC sanplas

o ConpuChfen QC anpl-S:*

o Outside vendor QC samples <ie. Environmental Research

o Client specific QC *anpl*s

o Client specific blind QC sanplcs

flR.30d8l720

Page 3: ftR3008L6 - semspub.epa.gov

ChemWest Analytical LaboratoriesAddendum A

Organizational Charts

RR300818

Page 4: ftR3008L6 - semspub.epa.gov

ORGANIZATIONAL CHARTS

ftR3Q0819CHEMWEST A n a l y t i c a l L a b o r a t o r i e s . Inc.

Page 5: ftR3008L6 - semspub.epa.gov

CO

tr co< i-x -PU CO

o IH <

< -PNI CM CD

< 5O DJrr coO c

BK>aL,

s

+> fc.c -^ oB U

O•3

__

__

V U C- - Ifc. *- *

B « B* * S i>X X C*» *> i c

I z *O O w

C — l-«- B B

£ 5 ?a c - cC B a B-• « • £*> * cB £ -aJC C C1. X k *-B C B 4>I • 4J e

<*- §S JCIfe V k.* * 1

S B CJC D

C C M« a CC 1C JIL B•s. OBC B Bf I- C

B £S 4 J 4 J

- 1 •£^ S -o£ O* BL.

B OV •

t = gs • —n B <p§ B

BS r i

1 ^11- - £

_ —

£

C ft

"o CLL.

I !uI

"~111

L

"*>

I I*> ».c a5 e» *-O •C *J

o otc

_

M X ?tl i. k- 0 «. •e *> • oi B aO I- B •t- 0 C KO £ mI 5 s 2

iX

1L, t, 5O B ££ B

-j c e1 1 ^. U)

t SO U*J C 0

k a

1 ? •»L.

JC

I.IB *> B^ B a «

i- *•O.

U.

i 1 1* i

IRSofti '. .

20

Page 6: ftR3008L6 - semspub.epa.gov

ORGANIZATIONAL CHART

Sample Control

October

1989

B "pa t

B C

E }*CD •n

WMstlBr

n

i-

Sample Control

TachnfctaTi

ToMver

B

U

i SamplB Control

TechntclBn

«J^

s

flR30082

Page 7: ftR3008L6 - semspub.epa.gov

AR300822

Page 8: ftR3008L6 - semspub.epa.gov

H

XU-1<2OMsNlMz<o£TO

T

. £^ 0L.

5 1CO SWUf Ut. o

X£aCOL.O>O•PCO

C X• i.u OQ *> V.£ « v .L. B

9 i•» i cf 5*£ 8

L-£O«COC5

^^«c|a

1u44CBUC>

*»B

8 51

*1«

»!E0]

X«£i."3XOkK

*>B

M 5S utff>

flR3'd0823

Page 9: ftR3008L6 - semspub.epa.gov

L.O•PCOL.

5CO a .

L.

s

Xa k.c 01.O tt *S B£ I « e ——— '

c • *^ j *u

BB5dB

B

i

B

^ •

8 f

W

§Dc£

•cL

£UC

jB

I "O O

E.en

L.

BC

sO•3

Bt^

B

trj

Page 10: ftR3008L6 - semspub.epa.gov

Joyce H. Y

oahloka

Inorganic

Specialist

5 o< 4Jx cou_J< COZ -J8 « v

N CM COZ O)< S-C Otr cO H

k B XB O <- 4.-9 — O B

C 44 OB K Btt o t c

V *J1- OBB — •»-

I

E U B1 1!• M tn

•3

B *>B D B

a T *. oc a o i*> B e. 6a» » »i asO

RR300825

Page 11: ftR3008L6 - semspub.epa.gov

OS*4

|

O

H o~CC

f} !•"•O

-J U „ _*t 4^ • £Z C x £ fe00 ^ 3 |M O t B ir^™ B •< CO 1 °M 4J aH C C*vz o ' —— —— 'of±Ja:o

cE»

k.^

5

O 4>b. B

I*O

c1L.h-

£0u

~--^ A

S B

0

5 £• no

t.BO

io.

•Jk.B

»-0 *>i. B*> fC v-

"p5 Im tr>o

BL.

3•O

X(DB«i

£ «*» —A "*

"o4J Q• no AR3QQ826

Page 12: ftR3008L6 - semspub.epa.gov

C£ CD< t)X C ,————————. gU CO j———————. 5J 3 - * ~ ' 5< (0z too <wf- X< -PM i-Z CO< 3o oo

cT3«^

O

B>B*Jn

BUC

3 l-B BB D

cX •iJ I..

B

6

«c6f-pj

J

Ccm

cuBt. *J

B f! «-< B

x "o2 £*- CO

1

flR30082?

Page 13: ftR3008L6 - semspub.epa.gov

H Oct c< CO5 £

tv o) Ov

< C f

a i ———*-i

*>

* cB a +JJC • —C C £ L.B B P) *0 1 "• « ________ _J "- „„.,-..o « CC B - . 4 1B C B C. ^ J 6o £ .£

K 4>0 •Ji. -•.5 1

1 1

5 .^ >* -.3 *» 43F « ct. •f4* *>

" •i* 'c *C -; «

in <

,_ -P g : : -- 5 gH co< LIS! 4H CflZ T-< C

Page 14: ftR3008L6 - semspub.epa.gov

•• -

0}r4_£o

c ^-• -• B+> BO*6 0 C• ~ " •B O> Xf Be ocX C -• .L C *»B B B«- ** JC«- B i.

J§ £ 1

J£B

£u•-»1

izIBL.O1.1

BU B^ iL. 4)B B

— *1 *6 L*3 c.

i *

B^*»B

S4JC

*- B5 EBb.&

H<0 i.

O) ? 5_J r ^ - • £—«• fc *> B o fc< t-

< CON 2M

o:o

flR300829

Page 15: ftR3008L6 - semspub.epa.gov

ChemWest Analytical LaboratoriesAddendum B

Staff Responsibilities and Resumes

SR30Q830

Page 16: ftR3008L6 - semspub.epa.gov

STAFF RESPONSIBILITIESAND RESUMES

AR30083ICHEHWEST A n a l y t i c a l Laborator1•• , Inc.

Page 17: ftR3008L6 - semspub.epa.gov

CHEMWEBTANALYTICAL LABORATORIES, INC.

CHEMWEST EMPLOYEES October 1988

____ ENVIRONMENTAL OPERATIONS_____

Offleers/Administration

Joel C. Bird President

Jill B. Henes, Ph.D. Vice President of Technical Services

Randall D. Smith Vice President of General Services [In-House Service Specialist]

Clarence Banks Business Manager

Michael P, Smith Controller

Rosanne Marguette . Administrative Assistant

Michele Fox Receptionistr

Quality Assurance _ ___ _ ...._.._.,__.._

Steven C. Madden Quality Assurance Manager [QualityAssurance Officer]

Jane I." LaTona Quality Assurance Specialist

Sample Control " . .. ._...

Bill McBenge Sample Control Manager [Sample coCustodian] co

OMichelle Toliver ~ ~ Sample Control Technician C5

coThomas S. Whistler Sample Control Technician &r

Sample Preparation Laboratory

F. Thomas Kwoka Sample Preparation Manager [SamplePreparation Supervisor; DioxinExtraction Specii

Charles F. Taylor ' Associate Chemist600W North Market Boulevard • Sacramento, CA 95834 • Phone [916] 923-0540

A Subsidiary of CompuChem Corporation

Page 18: ftR3008L6 - semspub.epa.gov

CHEMWEST EMPLOYEES October 1988

ENVIRONMENTAL OPERATIONS (Continued)

Sample Preparation Laboratory (continued)

James Kitchen Associate Chemist [Extractions__ Specialist - Extraction/Concentration

Expert]

Helen Smpardos Associate Chemist

Scott W. Pieters Laboratory Technician

GC Laboratory

Kirk J. Pocan ... . .. GC Manager [GC Laboratory Supervisor]

Troy J. Galloway GC Specialist [Pesticide ResidueAnalysis Expert]

Vincent C. Paul GC Specialist

-~s Brucie Estill GC Specialist

6C/MS Laboratory

Elaine Wong GC/MS Laboratory Manager [GC/MSSupervisor; Mass SpectralInterpretation Specialist]

John Medina, Jr. GC/MS Specialist [Dioxin Specialist;GC/MS Operator]

Richard M. Henson GC/MS Specialist

Annette D. Oleski GC/MS Specialist

Inorganics Laboratory

David R. Jarvis Inorganic Manager [InorganicSupervisor]

Jana Malcolm-Brown Inorganic Specialist

Linda C. Schneider Inorganic Specialist nnoO'3

CKEMWEST A n a l y t i c a l LatoeratoMaa,, Inc.

CHEMWEST "ANALYTICAL LABORATORIES. INC = , - "

Page 19: ftR3008L6 - semspub.epa.gov

CHEMWEST EMPLOYEES October 1988

ENVIRONMENTAL OPERATIONS (Continued)

Inorganics Laboratory (continued)

Janet C. Harlan - Inorganic Specialist

Allan Wbng Inorganic Specialist

Joyce H. Yoshioka Inorganic Specialist

Stephen C. Crisp- Inorganic SpecialistGrieser

Data Control

Robert T. Hart Data Control Manager

Denise Syljuberget Data Control Specialist

M Peter J. Gieger Data Control Assistant [Dioxin DataReporting a.nd Delivery Specialist]

Carol Trimble Data Control Specialist [Organic DataReporting and Delivery Officer]

Barbara Wiggins Data Control Specialist

Marketing

Jeffery Hesseltine Western regional Marketing Manager

Mar.jorie M. Namba Sales Representative

Toni C. Weeks Customer Service Representative

Note: Titles in brackets ([]) represent EPA functional Titles.

CHEMWEST A n a l y t i c a l La b o r a t o r 1 •• , Inc.

CHEMIST-ANALYTICAL LABORATORIES. INC

Page 20: ftR3008L6 - semspub.epa.gov

JOEL C. BIRD-president

CHEMWEST Analytical Laboratories, Inc.

Responsibilities: Mr. Bird is president and is responsible for• - — • - • • • - . - the day to day operation of CHEMWEST

.. _ Analytical Laboratories, Inc.

Education: Mr. Bird received his Bachelor of Sciencedegree in Biology from California StateUniversity, Fresno and his Master of BusinessAdministration from California StateUniversity, Sacramento.

Job Experience: Prior to joining CHEMWEST, Mr. Bird wasemployed for 2 1/2 years as Supervisor ofGC/MS, Priority Pollutants, for CaliforniaAnalytical Laboratories - Enseco. He wasemployed at CAL for ..over .5 1/2 years,involved with analysis of volatiles,semivolatiles, and dioxins by GC/MS, both asa manager and operator. Mr. Bird has 8 yearsof mass spectral data interpretationexperience..

CHEMWEST.ANALYTICAL LABORATORIES. INC

Page 21: ftR3008L6 - semspub.epa.gov

JILL B. HENES, PH.D.Vice President, Technical Services

CHEMWEST Analytical Laboratories, Inc.

Responsibilities: Dr. Henes is the Vice President of TechnicalServices responsible for managing thelaboratory staff in the accurate and timelyanalysis of environmental samples.

Education: Dr. Henes has an undergraduate degree inChemistry from the University of Vermont,Ph.D. in Chemistry from Case Western ReserveUniversity, and a Masters of BusinessAdministration from Duke University.

Job Experience: Prior to joining CHEMWEST, Dr. Henes wasemployed for four years at CompuChemLaboratories where she held the positions ofdirector of GC/Dioxin Programs, Manager ofGC/Dioxin Programs and Assistant Manager ofthe GC Laboratory.

Prior to joining CompuChem, Dr. Henes wasemployed as Technical Director withIndustrial and Environmental Analyst, Inc.,where she supervised the management ofcontracts to perform groundwater monitoringand sample analysis of trace organicchemicals. Dr.Henes also conductedindependent research in the field of enzymekinetics as a Post-Doctoral Researchassociate at Yale University.

Through the various laboratory positions Dr."Henes has held, she has gained 1 1/2 years ofexperience in the operation of the GC/MS/DSon environmental samples and 7 years ofexperience in the interpretation of massspectra gathered in the GC/MS analysis.Additionally, she has gained 8 years ofexperience in the preparation of extractsfrom environmental or hazardous waste samplesand 1 1/2 years of experience using the purgeand trap technique for .volatile organics.Dr. Henes has 4 years of experience inorganochlorine pesticide and PCB analysis,including clean-up procedures such as columnChromatography on environmental samples.

0

CHEMWEST ANALYTICAL LABORATORIES. INC

Page 22: ftR3008L6 - semspub.epa.gov

RANDALL D. SMITHVice President of General Services

CHEMWEST Analytical Laboratories, Inc.

Responsibilities: Mr. Smith is the Vice President of GeneralServices and is responsible for planning,coordinating and directing the operations ofCHEMWEST facilities and laboratoryinstrumentation, including specializedelectronic support of instrumentation for useby GC/MS and GC Laboratories. Mr. Smith isresponsible for daily interface with majorCHEMWEST vendors.

Education: Mr. Smith-xeceived his A.A. Degree in...Computer Technology from Control Datainstitute in Los Angeles, California. Hereceived extensive training in GC/MSelectronic maintenance from Finnigan.Institute in Cincinnati, Ohio, and fromExtrel Corporation in Pittsburgh,Pennsylvania.

Job. Experience:. .PrApr. 0 joining CHEMWEST, Mr. Smith wasemployed as a GC/MS Hardware Specialist withCalifornia Analytical Laboratories - ENSECO.He was responsible for all maintenance andrepair of 16 GC/MS/DS and MS/DS systems.Previous to his employment with CAL, he was asystems engineer with the FinniganCorporation for five years. Hisresponsibilities required specialized skillsin electronic circuitry, and the applicationof laboratory instrumentation. Prior to hisemployment with Finnigan Corporation, Mr.Smith was production Supervisor in electricalsub-assemblies for Me Donne11 Douglas-Commercial Aircraft Division. Hisresponsibilities included supervising 20persons in the assembly of electrical powercenters and radio racks for the DC-9aircraft.

AR300837

CHEMWEST ANALYTICAL LABORATORIES, INC

Page 23: ftR3008L6 - semspub.epa.gov

STEVEN C, MADDEN - -Quality Assurance Manager

CHEMWEST Analytical Laboratories, Inc.

Responsibilities: Mr. Madden is responsible for data review,training,and overall Quality Assurancemonitoring for CHEMWEST AnalyticalLaboratories, Inc.

Education: Mr. Madden received a Bachelor of Sciencedegree in Chemistry from the University ofCalifornia, Davis. Also, Mr. Madden has morethan 4 years of post graduate education in

~, • Agricultural Chemistry (EnvironmentalToxicology), at the University of California,Davis.

Job Experience: Prior to joining CHEMWESTf MR. Madden was theGC Laboratory Supervisor for 2 years,atRadian Corporation, in Sacramento. He wasresponsible for managing all operations ofthe GC laboratory including GC maintenance.Mr. Madden previously held positions as theManager of the Environmental Data Laboratory,Aerojet Strategic Propulsion Company and asGC Department Head, Carborundum AnalyticalLaboratory.

Through the variety of laboratory positionshe has held, Mr. Madden has gained 6 yearsof experience in the preparation of extractsfrom environmental and hazardous wastesamples, and 9 1/2 years of experience in theanalysis of organochlorine pesticide residueand PCS analysis including clean-upprocedures such as column Chromatography onenvironmental samples. In addition, Mr.Madden has 7 1/2 years of experience in usingthe purge-and -trap technique for theanalysis of volatile organics and 6 years ofexperience in the .analysis of otherpesticides and priority pollutant chemicals.

PCHE MWEST ANALYTICAL LABORATORIES. INC.

Page 24: ftR3008L6 - semspub.epa.gov

MICHAEL P. SMITHController

CHEMWEST AnalyticalJ^aboratpries, Inc.

Responsibilities: Mr. Smith is responsible for thefinancial/accounting and personneladministration functions of the laboratory.

Education: Mr. Smith received his MBA and BS in, Accounting from California State University,

Sacramento. He earned his California CPAcredential in 1986.

Job Experience:. _Prior t_p__joining CHEMWEST, Mr. Smith wascontroller for Software Dimensions, Inc. - amicro computer business software developer.

Previously, Mr. Smith held positions asDivision Controller with Northern telecon,Inc. - a PBX telephone systems firm andaccounting manager with Varian Associates - aproducer of laboratory analytical equipmentand accessories.

AR300839

CHEMWEST ANALYTICAL LABORATORIES. INC

Page 25: ftR3008L6 - semspub.epa.gov

BILL MCBENGESample Control Manager

CHEMWEST Analytical Laboratories, Inc.

Responsibilities: Mr. McBenge is responsible for the control ofall samples; both incoming and outgoing. Heis also responsible for all laboratorysupplies.

Education: Two Years of college.

Job Experience: Prior to joining CHEMWEST, Mr. McBenge wasemployed at California AnalyticalLaboratories - ENSECO as Sample CustodianSupervisor for 4 1/2 years. At CAL, Mr.McBenge designed and organized SampleControl, controlled all incoming and outgoingsamples (EPA and Non-EPA), completed Chain-of-Custody forms, Traffic Reports, SpecialAnalytical service forms, and Packing Lists,etc. Mr. McBenge was also responsible forproper disposal of the samples afteranalysis. He has 6 years of sample control

_ ., .;. ... -...,- -.experience.

Prior to that Mr. McBenge was employed atCaterpillar Tractor Company in Peoria,Illinois. Before being employed atCaterpillar, he was Quality Assurance—Inspector and Lab Assistant for Lehn & FinkCorporation, a subsidiary of Sterling Drug,where he performed water analysis, andinspection of all finished products andpromoted to Plant Supervisor 3rd shift.

CHEMWESl ANALYTICAL LABORATORIES. INC.

Page 26: ftR3008L6 - semspub.epa.gov

ft)

F.; THOMAS KWQKASample Preparation Laboratory ManagerCHEMWEST Analytical Laboratories, Inc.

Responsibilities: Mr. Kwoka is the manager of the OrganicExtractions Laboratory. 3He is responsible"for the supervision of all activities forthese laboratories including scheduling ofwork, preparation of all standards andensuring that high-quality work be performedin a timely and efficient'manner.

Education: Mr. Kwoka earned his Bachelor of ScienceDegree in Chemistry from California StateUniversity, Sacramento in 1983.

Job Experience: Prior to joining CHEMWEST, Mr. Kwoka workedfor three years for California AnalyticalLaboratories - EKSECO. Mr. Kwoka was thesupervisor of CAL's Dioxin Laboratory for thepast two years. Mr. Kwoka has 4 1/2 years ofexperience in the. .preparation of standardsand in the extraction and columnChromatography clean-up techniques ofpolychlorinated dibenzo-p-dioxins andpolychlprinated dibenzo. furans. Mr. Kwokahas 1 1/2 years experience using purge-and- -trap technique for volatile organics, TCA/TCEand gasoline analysis by GC. Mr. Kwoka has 5yea'rs of experience in the preparation ofextracts for semivolatiles and organochlorinepesticides/PCB's including columnChromatography clean-up procedures.

AR3008lfl

.CHEMWEST ANALYTICAL LABORATORIES. INC.

Page 27: ftR3008L6 - semspub.epa.gov

JAMES KITCHENAssociate Chemist

.... CHEMWEST Analytical Laboratories, Inc.

Responsibilities: Mr. Kitchen is. responsible for the extractionand concentration of water and soil samplesfor organic analysis, including prioritypollutants. Mr. Kitchen's duties alsoinclude the analysis of total petroleumhydrocarbons by IR and data reductionutilizing a program on an Epson Equity IIcomputer system.

Education: Mr. Kitchen earned his Bachelor of ScienceDegree in Biology from Dickinson StateUniversity in Dickinson, North Dakota in1987.

, Job Experience: While at CHSMWEST, Mr. Kitchen has continuedto demonstrate his proficiency in preparingsoil and water samples for analysis by EPA-CLP methods 8080, 8270, 8240, and TPH for GC

• •-- ———— -'- -analysis. MR. Kitchen also has gainedvaluable experience in column Chromatographyclean-up procedures, and extracting organiclead for ICAP analysis.

AR30081J2

CHtMWEST ANALYTICAL LABORATORIES.

Page 28: ftR3008L6 - semspub.epa.gov

CHARLES..F, TAYLOR ._. ... ...Associate Chemist

CHEMWEST Analytical "Laboratories, Inc.

Responsibilities: Mr. Taylor is responsible for the extractionand concentration of water and soil samplesfor organic analysis, including prioritypollutants, dioxins, and furans. He is alsoresponsible for column Chromatography cleanupprocedures. Mr. Taylor's duties also includesome inorganic analyses, such asnitrates/nitrites and cyanides, as well asthe analysis of total petroleum hydrocarbonsby IR and data reduction utilizing a programon an Epson Equity III.computer system.

Education: Mr. Taylor earned his Bachelor of ScienceDegree in Chemistry from California StateUniversity, Sacramento in 1987.

Job Experience: Prior to joining CHEMWEST, Mr. Taylor workedfor six months with the California Departmentof Food and Agriculture at their MeadowviewRoad Laboratory in the Pesticide Residuesection. He has ten months of experience inpreparing soil and water samples for analysisby EPA-CLP methods 8080, 8270, 8240 includingcolumn Chromatography cleanup procedures andTPH for GC analysis. Mr. Taylor has fourmonths experience in the extraction andcolumn Chromatography cleanup techniques ofpolychlorinated dibenzo-p-dioxins andpolychlorinated dibenzo furans. Mr. Tayloralso has experience analyzing samples forpesticides by GC.

AR3008U3

CHcNWEST. ANALYTICAL LABORATORIES. INC,

Page 29: ftR3008L6 - semspub.epa.gov

ftI ty

KIRK J. POCANGC Laboratory Manager

CHEMWEST Analytical Laboratories;, Inc.

Responsibilities: Mr. Pocan is responsible for the scheduling,data review, training, instrument maintenanceand the overall management of the GCLaboratory.

Education: Mr. Pocan received a Bachelor of Sciencedegree in Environmental Science from theUniversity of Nevada, Reno. Mr. Pocan alsohas two years of post graduate education.

Job Experience: Prior to. joining CHEMWEST, Mr. Pocan was GCLaboratory Supervisor for CaliforniaAnalytical Laboratories-ENSECO. Prior to hispromotion, Mr. Pocan was a staff chemistspecializing in pesticide and herbicideextractions and analysis. Mr. Pocan has over8 years of GC experience. Mr. Pocan has acomplete understanding of EPA 500, 600, 8000series methodology as they relate toChromatography analysis. .Mr. Pocan has used•packed capillary, and megabore columns and isfully familiar with the following detectors:FID, EC, TSD, HALL, Coulson, ITD, PID, UV,and fluorescence.

CHEMWESl ANALYTICAL LABORATORIES. tNC,

Page 30: ftR3008L6 - semspub.epa.gov

TROY J. GALLOWAYGC Specialist.

CHEMWEST Analytical Laboratories, Inc.

Responsibilities: Mr. Galloway is responsible for thedevelopment and application of GC methods forsamples requiring analysis using ECD, FID,PID, NPD, FPD, and HECD detectors.

Education: Mr. Galloway received a Bachelor of Sciencedegree in Chemistry from California StateUniversity, Sacramento.

Job Experience: -Prior to joining CHEMWEST, Mr. Galloway wasemployed by Canonie Environmental ServicesCorporation as a supervisor of GC and LCprojects in pesticide residue analysis. Heperformed GC and LC analysis on dust, foodtypes, environmental and hazardous wastes.

Mr. Galloway has 1 year of experience in theanalyses of metals using 3SPA flame andgraphite furnace methods and 4 years of GC

- - and LC experience in purgeable volatileorganics analyses, organochlorine pesticideresidue and PCS analysis.

/IR3008U5

ES' ANALYTICAL LABORATORIES. INC

Page 31: ftR3008L6 - semspub.epa.gov

ELAINE WONGGC/MS Manager

CHEMWEST Analytical Laboratories, Inc.

Responsibilities: Ms. Wong is responsible for scheduling,__._._ ,. working and reviewing all priority

pollutant data, she is also responsiblefor overseeing the operation of GC/MSsystems, spectral interpretation andquantitative data analysis.

Education: Ms. Wong received a Bachelor of Sciencedegree in Biology/Chemistry fromCalifornia State University, Sacramento.

Experience: Prior to joining CHEMWEST, Ms. Wongworked as a GC/MS operator for fouryears doing volatile and semivolatileand 3 years of mass spectral datainterpretating at California AnalyticalLaboratories ENSECO. Ms. Wong has alsodone six months of dioxin data reviewand interpretation. Before becoming a

" GC/MS operator, Ms. Wong was supervisorof Inorganic Wet Chemistry for one yearat California Analytical Laboratories.Ms. Wong was responsible for assigningwork, reviewing all wet chemistryanalyses, and training of all newpersonnel in this section.

Ms. Wong also worked, for the CaliforniaState Department of Food and agriculturefor three years doing pesticide,herbicide and fungicide extractions.

----- Ms. Wong also has six months of GC experience

flR3008U6

CHEMIST ANALYTICAL LABORATORIES. INC

Page 32: ftR3008L6 - semspub.epa.gov

.... .... .... -JOHN MEDINA Jr.GC/MS Specialist

CHEMWEST Analytical Laboratories, Inc.

Responsibilities: Mr. Medina is responsible for operatingand maintaining GC/MS instrumentation,mass spectral interpretation, datareduction, and review for volatile,semivolitile, and dioxin analysis.

Education: Mr. Medina earned his Bachelor ofScience Degree in Chemistry from theUniversity of California, Davis in Juneof 1984.

Job Experience: Prior to joining CHEMWEST, Mr. Medinaworked for California AnalyticalLaboratories-ENSECO (July 1984 toOctober 1986). Mr. Medina worked forthe Organic Extractions group doingsemivolatile and pesticide extractions,metals digest, dioxin extractions, andclean-up. In June of 1985, Mr. Medina

- ". -.-—-—:-~-.--..:- -:::T"joined ±he GC/MS group at CAL. Mr.Medina vorlced in the Volatile Organicsgroup from June of 1985 until joiningthe Dioxin group in May of 1986. Mr.Medina has 3 years experience in theGC/MS operations and in mass spectralinterpretation, and has attendedFinnigan Institute for courses in SuperIncos Data Systems,

flR3008U7

CHEMWEST ANALYTICAL LABORATORIES. INC

Page 33: ftR3008L6 - semspub.epa.gov

DAVID R. JARVISInorganics Manager

CHEMWEST Analytical Laboratories, Inc.

Responsibilities: Mr. Jarvis is responsible for the dailyoperation of the Inorganics Department.These responsibilities include methodsdevelopment, laboratory Q.C. conformance,data review, report generation oversight andsample analysis.

Education: Mr. Jarvis received his Bachelor of ScienceDegree in Environmental Planning/Geology fromCentral Michigan University and is presentlyworking on his masters degree in WaterResources Management at California StateUniversity, Chico,

Job Experience: Mr. Jarvis has 8 1/2 years of AnalyticalLaboratory experience including 1 1/2 yearsas an EPA-CLP chemist for CaliforniaAnalytical Laboratories-E.NSECO, working in

_ the ICP/AA Laboratory.

Further, Mr. Jarvis has gained a broad rangeof Analytical experience as Research Analyst(Methods Development and Military Q.C.) forFairchild Corporation, Water AnalysisCoordinator for Balazas AnalyticalLaboratories, and as Laboratory Field ServiceCoordinator for Johnson & Anderson ConsultingEngineers.

flR3008Z*8

ANALYTICAL LABORATORIES. INC

Page 34: ftR3008L6 - semspub.epa.gov

ROBERT HARTData Control Manager

CHEMWEST Analytical Laboratories, Inc.

Responsibilities: _Mr.__Hart is responsible tor managing the...__. daily reporting requirements of CHEMWEST. He

is responsible for the report generation andquality assurance review.

Education: Mr. Hart received a Bachelor of ScienceDegree in Biological Science from CaliforniaState University, Stanaislaus.

Job Experience: Prior to joining CHEMWEST,, Mr. Hart wasemployed at California AnalyticalLaboratories-ENSECO as Dioxin DataSupervisor. He was responsible for allDioxin reporting, production control and wasthe EPA liaison for the Contract LaboratoryProgram. Mr. Hart was responsible forproduction control for all dioxin data. Mr.Hart has also worked as a Microbiologist forvarious Central Valley companies.

CHENWEST ANALYTICAL LABORATORIES. INC

Page 35: ftR3008L6 - semspub.epa.gov

PETER GIENGERData Control Personnel

CHEMWEST Analytical Laboratories, Inc.

Responsibilities: Mr. Gienger is responsible for the reportingof Dioxin and Commericial reports. Heinsures the quality and the reportingrequirements of the reports.

Education: Mr. Gienger is continuing his educationtowards a Bachelor of Science degree inChemistry.

Job Experience: Prior to joining CHEMWEST, Mr. Gienger wasemployed at California AnalyticalLaboratories - ENSECO as a commerical datasupervisor. He was responsible for allcommericail reporting and production controlfor all commercial data. Mr. Gienger hasalso worked as a Laboratory technician forSemiconductor .Companies in the bay area.

CHENWEST ANALYTICAL LABORATORIES. INC

Page 36: ftR3008L6 - semspub.epa.gov

CAROL L. TRIMBLEData Specialist

CHEMWEST Analytical Laboratories, Inc.

Responsibilities: .Ms. Trimble is r_es_ponsible for thepreparation of the Organic EPA data, whichinvolves contact"with the USEPA SampleManagement Office as well as tracking and 'review of the data. She also preparescommercial data packages.

Education: Ms. Trimble received a Bachlors of ScienceDegree in Biochemistry form the University ofCalifornia, Davis, compleated in June of1988. The courses required for this degreeincluded a wide variety of chemistry as wellas biochemistry courses.

Job Experience.: _Ms_,_Tri_mble_ ha_s h.ad various positionsthroughout her college career, consisting ofclerical and word processing duties. She iswell versed in Word Perfect, WordStar,Microsoft Word, and MacWrite. Before comingto CHEMWEST she was employed as a medicaltranscriber, word processor and technicalwrite.

AR30085I

CHEMWEST ANALYTICAL LABORATORIES. INC

Page 37: ftR3008L6 - semspub.epa.gov

JEFFERY C. HESSELTINEEnvironmental Account Executive

CHEMWEST Analytical Laboratories, Inc.

Responsibilities: Mr. Hesseltine is responsible for themarketing and sales of all commercialclients in the Western United Statescomposed of the following states; CA,AZ, OR, WA, AL, ID, TJT, KM, CO, WY, MTand HA.

Education: Mr. Hesseltine received his bachelorsdegree in chemistry from CaliforniaState University, Northridge.

Job Experience: Previously Mr. Hesseltine had worked fortwo years as a field salesrepresentative for Sargent-WelchScientific Co. which isa laboratory supply distributor. Inthis position he was in charge of allsales and marketing for the NorthernCalifornia area.

AR300852

CHEMWEST ANALYTICAL LABORATORIES, INC

Page 38: ftR3008L6 - semspub.epa.gov

ChernWest Analytical LaboratoriesAddendum C

Certifications and Contracts

Page 39: ftR3008L6 - semspub.epa.gov

CERTIFICATIONS AND CONTRACTS

CHEMWEST ANALYTICAL LABORATORIES. INC

Page 40: ftR3008L6 - semspub.epa.gov

CALIFORNIA STATE DEPRATMENT OF

.HEALTH SERVICES.

Hazardous Materials Laboratory Section

AR300855

CHEMWEST ANALYTICAL LABORATORIES, INC.

Page 41: ftR3008L6 - semspub.epa.gov

ANALYTICAL LABORATORIES, INC.

CERTIFICATIONS AND CONTRACTS

May 1988

CALIFORNIA STATE DEPARTMENT _OF HEALTH SERVICES

CHEMWEST Analytical Laboratories, Inc._holds the following activecertificates from both the Hazardous Materials Laboratory Sectionand the Sanitation and Radiation Laboratory Section of theCalifornia State Department of Health Services.

o State of California Department of Health Services HazardousWaste-Testing Laboratory Certificate (Certificate #185,October 10, 1987 to October 9, 1989).

A copy of the certificate and three letters indicating thecategories of tests covered are attached.

CHEMWEST is awaiting notification from DOHS-HMLS concerningcertification for three other test categories: HalogenatedVolatile Organics (EPA Method 8010), Organic Lead, andTotal Petroleum Hydrocarbons.

o State of California Department of Health Services ApprovedWater^Laboratory Certificate (Certificate #0956, March 10,1988 to June 30, 1991).

A copy of the certificate is attached.

CHEMWEST is awaiting notification from DOHS-SRL concerningfull organic, full inorganic, and coliform certification.

UNITED STATES ENVIRONMENTAL PROTECTION AGENCY

CHEMWEST Analytical Laboratories, Inc. currently has an activeRoutine Analytical Services (RAS) contract with the US-EPA ContractLab Program (CLP). A copy of first two pages of Contract #68~W8-0010 is attached.CHEMWEST also performs Special Analytical Services (SAS) work forthe EPA-CLP. These contracts are on a case by case basis, and havebeen performed for both Dioxin and non-Dioxin work.

fiR30Q856

600W North Market Boulevard • Sacramento, CA 95634 • Phone (916) 923-0840 ' FAX (916) 923-1938

A Subsidiary of CompuChem Corporation -; ','

Page 42: ftR3008L6 - semspub.epa.gov

STATE OF CALIFORNIA—rtEAlTH AND WElfARE AGENCY______________________ GEORGE DfUKMEJIAN.

DEPARTMENT OF HEALTH SERVICESMRKElfY WAY

ir, CA W704 - - -- - -(415) 540-3105

May IB, 1986

Joel C. BirdChemwest Analytical Laboratories, Inc.600 K North Market Blvd.Sacramento, CA 95834

Dear Mr. Bird:

We are pleased to inform you that your application for HazardousWaste Testing Laboratory Certification has been approved. Thecategories certified are listed in the enclosed Hazardous WasteTesting Laboratory Certification List. Thisse additional testcategories will be added to our next update list.

- Sincerely,

Robert D. Stephens, Ph.D., ChiefHazardous Materials Laboratory

Fred Seto, Ph.D.Public Health Chemist

AR300857

Page 43: ftR3008L6 - semspub.epa.gov

HAZARDOUS WST_E TESTING LABORATORY CERTIFICATION LISTHazardous Materials Laboratory Section;"California Department of Health Services, 215! Berkeley fey, Btrkelty, CA

f..••west Analytical Laboratories, Inc. " PHONE: (816)923-0840 LABORATORY CATEGORY: Coaeerclali Korth Market Boulevard CERTIFICATE KUKSER: 185

CA 55634

ORGANIC CHEMICAL TESTING METHOD NUMBER (DATE CERTIFIED)

1.1 Halogenated Volatile Organic* ———————————————————————1.2 r Non-Halogenated Volatile Organics - ——————————— - ————————I.I Aromatic Volatile Organic* —————————————————————————1.4 Acrolein, Acrylonitrlli, Acetonltrile ——————————————————1.5 Phenols ————————————— ———————————————————1.5 Phthalati Esters ————————— ' ————————————————————1.1 OrganochloMne Pesticides ————————————————————————— •1.8 Polychlorfnated Biphenyls (PC8s) ——————————————————— — —1.9 Kitroaromatlcs and Cyclic Ketones —————————————————————1.10 Polynudear Aromatic Hydrocarbons - —————————————————————1.11 Chlorinated Hydrocarbons ———————————————————————————1.12 Organophosporus Pesticides • —————————————————— —— ———1.13 Chlorinated Herbicides - ————— = ——————————————————— —1.14 Carbawtes — — — — — --——.. .. .......-.-—..-,,.,..•- .———,.-...,-.1.15 BC/KS Method for Volatile Organics ————————————————————1 U ftT/lK Mftfknd fcvf Q»n> * uM » * 11 * ftfMiaMM; _— —— . ———— __———_™™ __,__—_

(f « CERTIFIED; X « MOT CERTIFIED)

———————————————— 8010(05-16-18) ——

———————————————— 8020(65-03-87) ——

———————————————— 8080(06-31-87) —————— . ——————————— 8080(08-31-87) ——

———————————————— 12*0(08-23-87) ——~-.,-...— — —_ —— . ......-tiififna.fi-a.eTt _.

—— y—— K—— Y- — -Xo—— N—— N—— Y—— Y—— N... M—— N—— 4!—— X

ijn—— Y_ _ _ _ v

INORGANIC CHEMICAL TESTING METHOD NUMBER (DATE CERTIFIED) (Y « CERTIFIED; N « NOT CERTIFIED)

-7041(08-04-87)————Y2.2 Arsenic ——————————————————————-———————————————————70SO(OB-04-87)————————————Y2.3 Bariut —————————————————————————6010(06-04-87)——————————————————————————————Y2.4 BerylHui ————————————————————————6010(08-04-87)————————————————————————————————Y2.5 Cad«iu« ——————————————————————————6010(06-04-87)———————————————————————————————Y2.6 Chroiiui(VI) ——————————————————————————————————————:———7186(06-31-87)—————————————Y2.7 Chroalui(total) ——————————————:——;——60'10(06-04-87)——————————————————————————————Y2.e Cobalt ——————————————————=——————6010(06-04-87)———————————————————————————————Y2j Copper ——————————————————————————5010(06-04-87)———————————————————————————————Y2.10 Lead ———————————————————————————6010(06-04-87)———————————————————1421(06-04-87)————Y2.11 Mercury———————•——————————————————————————————————————————————7471(08-04-87)————Y2.12 Molybdenum ————————————————————————6010(06-04-87)———————————————————————————————Y2.13 Nickel ——————————————————————————6010(06-04-87)———————————————————————————————Y2.14 Seleniui ————————————————————————————————————————————7740(08-04-87)—————————————Y2.15 Silver ———————————————————————————————————————————7760(08-04-87)————————————Y2.16 Thalliut —————————————————————————6010(06-04-87)———————————————————1841(06-04-87)————Y2.17 VanadiiM ————————————————————————5010(08-04-87)—————————————————————————————Y2.18 Zinc ———————————————————————————6010(06-04-87)———————————————————————————————Y2,19 Cyanide ———————————————————————————————•———————————————————8010(06-31-17)————Y2.20 Ruorida —————————————340.2(06-04-87)——————————————————————————————————————————Y2.21 Sulflde —————————————————————————————————————————————————————•«""" - — —Y

OTHER

« California Haste Extraction TestPhysical Property Testing ———

'-.« Aquatic Toxiclty Testing ————6.0 Bulk Asbestos Testing —————1.8 Total Organic Lead «1.0 Total Petroliwi Hydrocarbons

Page 44: ftR3008L6 - semspub.epa.gov

II

IIIIIIIIliII

IIII

en

£Z

C/3

o

t ^t , ••' ow i ^ T *• 3- U tf" f\ tc "J

u. LU wp j: <£

5 2 co Js 5i- tu LLJ g

UJoIIIIII _

oII §t K4

«< . fiRSQOf

II flR300859

(9

Page 45: ftR3008L6 - semspub.epa.gov

CALIFORNIA STATE DEFRAGMENT OF

HEALTH SERVICES

I)Sanitation and Radiation Laboratory'Section

CHEMWEST ANALYTICAL LABORATORIES. INC

Page 46: ftR3008L6 - semspub.epa.gov

IBJ

•n gv _ ..__ _• ••••

ll I IU ^^ eH P•H H

fe 15SJ •• Q** w O

>•5 °

8 § <O

^ OS

*

g;

.2 £«*• ^

IS T3g §O ^\ tJ3 O wes co *jC *~ -£^

•. c -p

**j ^ j j J

^3 t* Q 5J Q

S> g to P. eu>• 53 £4e i » «

1^ V — _*^ ws K »—* "o J

« ^ JJl 1l| I1 si 1-V

c Boo 5 -. •**5" 1 <£II £ .

•«• •*! j.t- B5 S

flR30086l

Page 47: ftR3008L6 - semspub.epa.gov

.\ o si

4-> -» £ *' (0 f-t n < iO • fK « -P R ti -rt ID Cue £ *»w in £ 4J tt) £ 5-rt C*i 0 S E » i4J g O 0 «wo J5 c — - ** *«> z 5 — 5 * IU •"- T3 0) *•

fe .. C H O«5 w 10 -H -H

wo S SH O M

O O 10 O5= m «H 4J*2 « o

J3 < .H« *H re in

> c/i « tt «"< _, •£ « Tl ££C S •< S K «-!, p -H 4J C^ K |*J j ^ o CO' g 4J l-t 0)£" W Z ^Q E"1 **-', K ^ "3 ^

w,r^^ * * U K «or i P* ^ %to H CJ (C «-, «*; ;w 'EzSQ 5^< « S - F k °<fc3O5 3W S « *K "3HSSS"^ O Jtf C P U C J - H«H<2: UJ2 £P.^K O > <I B o g

JT S -c<s - d£•0 i ' '£ ® i— *u tJla c

.1- 18• 1 I-ti 'D »

00 -S | t^/0 frj C E V**wi -e *-rt ft) tow .I I*K 3 £

W '-<&)co

RR300862

Page 48: ftR3008L6 - semspub.epa.gov

UNITED STATES'ENVIRONMENTAL PROTECTIONAGENCY - fXwrgar"T LABORATORY

? /

CHEMWES7 ANALYTICAL LABORATORIES, INC

Page 49: ftR3008L6 - semspub.epa.gov

Page 1WAS7K236

SOLICITATION, OFFER AND AWARDCertified for National Defense under BDSA Reg. 2 and/or CMS Reg. 1 .

RATING:TITLE: -Chemical Analytical Services for Hazardous Watt*

Field Samples for Qrganics by GC/MS Techniques

CONTRACT NO. 3 SOLICITATION NO. ! 4. TYPE OF SOLICITATIONWAB7K236 J Sealed Bid UF3>

DATE ISSUEDJJL 1 7 1987

6. REQUISITION/PURCHASE NO.

ISSUED BY<Hand~Carried/Courier Address)Environmental Protection AgencyBID/PROPOSAL ROOM/ 3rd Floor499 South Capitol Street, SWWashington/ DC 20003

8. ADDRESS OFFER TO <If other than Item(U S. Mail only)

Environmental Protection AgencyBID/PROPOSAL ROOM CPM214F)401 n St. SWWashington, DC 20460

IOTE: In sealed bid solicitations, "offer and offerer" mean "bid and bidder".

SOLICITATION'. Sealed offers in original and 4 copies for furnishing the supplies .orservices in the Schedule will be received at the place specified in Item S. or

ndcarried' in the depository listed in Item 7, until 1:00 P.M. local timeAL/G f Q 'C37 CAUTION-LATE Submissions, Modifications, and Withdrawals'

'wee Section L* Provision No. 32. 21*-7 or 52. 213-IO. All offers are subject toall terms and conditions contained in this solicitation.

"ol FOR INFORMATION CALL A. NAME Teresa LauB. TELEPHONE NO. (No Collect Calls) <202) 382-3239

jj_ TABL£ OF CONTENTSPART/SECTION DESCRIPTION

PART i - THE SCHEDULEA SOLICITATION/CONTRACT FORMB SUPPLIES OR SERVICES AND PRICES/COSTSC DESCRIPTION/SPECIFICATIONS/WORK STATEMENTD PACKAGING AND MARKINGE INSPECTION AND ACCEPTANCEF DELIVERIES OR PERFORMANCEC CONTRACT ADMINISTRATION DATAH SPECIAL CONTRACT REQUIREMENTS

PART II - CONTRACT CLAUSESI CONTRACT CLAUSES

PART III - LIST CF DOCUMENTS* EXHIBITS AND OTHER ATTACHMENTSJ LIST OF ATTACHMENTS

PART iv - REPRESENTATIONS AND INSTRUCTIONSK REPRESENTATIONS, CERTIFICATIONS AND OTHER STATEMENTS OF OFFERCR5L INSTRUCTIONS. CONDITIONS. AND NOTICES TO OFFERORSH EVALUATION FACTORS FOR AWARD AR30086J4

APPROVED BY OIRM 3/S4- FAR <4S CFR 53 2 l 4 C e > > EXCEPTION TO STANDARD FORM 33

Page 50: ftR3008L6 - semspub.epa.gov

Fagt 2WAB7K236 _ _ _

SOLICITATION, OFFER AND AWARDOFFER (Must be fully completed by offerer)

._ Item 12 does not apply if the solicitation includes the provisions at

.214-16. Minimum Bid Acceptance Period.

12. In compliance with the above, the undersigned agrees* if this effer isi cepted within 180 calendar days (120 days calendar days unless a differentp riod i* inserted by the offerer) from the date for receipt of offers specificabove* to furnish any or all items upon which prices ore offered at the pricet-t opposite each item, delivered at the designated pcint(s). within the timei ecified in the schedule.

13.0DISCOUNTCalendar

__ .%

FORday

PROMPT• J20•....__

PAYMENTCalendar__ %

(See Seeldays

Eion30

I* Clauso NO.Calendar ([ays

__ %

32.t_1J

232-8)_ Calendar

__ Xdays

1*. ACKNOWLEDGEMENT OF AMENDMENTS (The offerer acknowledges receipt of4 endments to the SOLICITATION for offerers and related documents numbered andsated:

AMENDMENT NO I DATE I AMENDMENT NO I DATE

i ONE CD ; 8/6/8713A. NAME AND ADDRESS OF OFFERORC] 'de: Facility:1 CHEMV7EST Analytical Laboratories, Inc.600W Korth Market Blvd.

»cramento, CA 95834) TELEPHONE NO. (Include Area Code)

fj? t 3 CH£CK IF REMITTANCE ADDRESS ISIN SCHEDULE ..

: '. S1CNATURE: T . ^

i :I 16. NAME AN0 TITLE OF PERSON: AUTHORIZED TO SIGN OFFER

Joel C. BirdPresident(Type or Print)

(916) 923-OEMODIFFERENT FROM ABOVE ENTER SUCH ADDRESS

US. OFFER DATE:i 8/17/87

(Joel C. B4#d) AWAR^D (To be completed fey the Government)ACCEPTED#S TO ITEMS NUMBERED !2V.. AMOUNT !2l. ACCOUNTING AND APPROPRIATION

I ! .-• •. *. AUTHORITY FOR USING OTHER THAN FULL AND OPEN COMPETITITION

C 3 10 U. S. C. £304Cc)( ) C 3 41 U. S. C. 2S3COC >

3. SUBMIT(4 cep

INVOICES TOies unless

ADDRESSotherwise

SHOWNepeci

IM ITEM t§3r<2~S" :fied >

'4. ADMINISTERED BY (If other than Item 7>: 23. PAYMENT WILL BE MADE BY! Environmental Protection AgencyI Financial Hgmt. Div. (MD32)! Attn: Account! Payable Br.! Research TriangU Pk. NC 27711

____ ______________________________ ; _________________ 126 NAME OF CONTRACTING OFFICER

MARIAN BERND(Tup, or Pi&Bttetins Officer ___________ _ _

F.. ORTANT - Award will be made on this Form or on Strtndard^Fbrm'126/ e>~ b"y otheruthorized official written notice.

APPROVED BY OIRM 3/64, FAR (48 CFR 33. 2i4Ce» EXCEPTION TO S \

. .. ...___...._._„__ ... ...... SR3Q0865" j

Page 51: ftR3008L6 - semspub.epa.gov

Chem West Analytical LaboratoriesAddendum D

Examples of Custody Forms

AR300866

Page 52: ftR3008L6 - semspub.epa.gov

Standard Operating Procedures (SOPs)

AR300867

ANALYTICAL LABORATORIES. INC

Page 53: ftR3008L6 - semspub.epa.gov

Sample Case Flow-Chart

- When samples are received at CHEMWEST they arechecked by the sample custodian for proper chain-of-custody from the shipper*

- The samples are issued CHEMWEST ID numbers during logIn procedures, their picture is taken and a samplefile is prepared.

- A complete set of copies are made and given to thePresident, Vice-president of Marketing, Technical LabDirector, and all Managers associated with therequested analyses.

- Extraction sheets are prepared by the project managerand passed on to the laboratory chemists.

- The chemist is responsible for checking the samplesin and out of sample control by signing the "SampleControl Log Sheet".

- When the chemist has finished preparing the samplesfor analysis, copies of the extractionist notebookand sample extraction sheet are passed on to theappropriate chemist, _who will be analyzing thesamples. The sample extracts are stored in a securedrefrigerator until time of analysis.

- When the samples have been analyzed and allappropriate data collected, the data is reviewed byanother chemist of manager and data sheets areinitialed upon approval, (See the CHEMWEST QualityAssurance Program Manual, page 29 for Data ReviewProcess).

- Then all of the sample data is passed on to DataControl where the information is compiled into asample data package according to the deliverablesindex listed in the contract. A book prepared as aCLP case is used as the SOP for assembling the datapackage.

- The new sample data package is returned to theoriginal operator who acquired the data or theDepartment Manager for approval and initialing of thedata sheets.

AR300868

LYTICAL LABORATORIES. INC

Page 54: ftR3008L6 - semspub.epa.gov

The case file is then reviewed by the project managerand finally by the CHEMWEST Vice-President ofTechnical Services. _

The case narrative for the case is signed by theVice-president of Technical Services or anotherqualified person (i.e. Project Manager, orEnvironmental Operations Manger).

The case is then copied and nailed to Vicr + co.,Sample Management Office, the appropriate EPA Regionand EKSL-Las Vegas using the same mode oftransportation for each.

The Data Control personnel are responsible for theorderly storage of sample documents in secured areas.EPA cases are kept in boxes having the EPA casenumber on them. Commercial samples are boxed inproduction, QC blank and QC spike/duplicatecontainers. The data specialist keep separate logsfor EPA case storage and commercial storage.

RR300869

CHEMWEST ANALYTICAL LABORATORIES. tNC

Page 55: ftR3008L6 - semspub.epa.gov

LOGGING IN SAMPLES

A) EPA (lows, mediums, 4 high hazard)

B) Non-EPA (Standard & Rush)

RR300870

Page 56: ftR3008L6 - semspub.epa.gov

EPA SAMPLE- LOG-IN

1) Check the ice chest-box for any EPA seals before opening.If no seals check condition of ice chest. Tape Intact.

2) Open the ieer chest in the hood and note any breakage(take pictures of any broken samples) - put on disposablegloves and gown. Take out and organize samples accordingto EPA numbers. Open samples in hood (uake sure hood ison). Empty remaining vermiculite, styrofoam in 30 gallontrash can in front of hood checking for any remaining samples

3) Examine traffic reports (fig 1) - see if you have receivedall component parts of the sample. Check bottle ID'Sagainst chain of custody ID numbers. Also examine chain-of-custody reports - see if all samples are included.Label samples according to contract. Colored tape I.D.(depending on contract) with black felt tip pen.

4) Check pH

5) Log in samples on lab ticket, (fig. 2,3,4,5,) include thefollowing information:

a) put CHEMWEST I.D, numbers in appropriate log booksb) client - inthis case US EPA and the appropriate contractc) analysis VMr, TCL (Target Compound List), Dioxins, etc.d) Case_number _ _e) CW I.D., EP A TVb., S"DG* and sample type (ext. or VDA)f) amount of sample received and in what type of con-

tainers, sizeg) VOA - any bubles? Septum up-side down?h) Sample tag information and numbers - or note if there

is no sample tagi) describe sample completely - e.g. any sediment, any

seals, any unusual problems at all - e.g. EPA numberon bottle or only on tape on bottle

j) Chain-of-custody - note on ticket if there is anychain-of-custody or not received (first ticket of series)

6) Till out traffic report - noting any problems on report e.g.seals intact or broken, chain-of-custody received or not re-ceived, samples broken, bubbles in VOA, numerical discrep-ancies, etc.

7) Sign chaing-of-custody (fig. 6), Note seal condition- makesure we've received those samples listed - sign in placesaying received by laboratory. Include date and time received.

B) Log samples in Sample Control Log.

SR300871

Page 57: ftR3008L6 - semspub.epa.gov

9) All information -_chain-of-custody, airbill, Traffic Reports,Tags, etc. placed in colored folders, (according to contract-color code) and turned in to Data Control, including pictures,if taken.

10) Medium and high hazard samples are placed In hood or glovebox. Open samples, only after sample custodian is gowned,gloved, and masked. Log in procedure same es others* Allmedium samples placed in ziplock plastic bags with red tapeto I.D, medium sample.

11) All EPA ice chests must be returned within two weeks.

AR3QQ872

Page 58: ftR3008L6 - semspub.epa.gov

I Order No JL

' CHEMWEST Analytical Laboratories, Inc. DiteRec/d_600 North Market Blvd.Sacramento, CA 95834(916)9230640 FAX (916) 923-1938 .

i

'i . Notebook______

U3.EPA-INORCANJCS TOTAL METALS)

i CONTRACTNO.

| ' ANALYSIS: TOTAL METALS

I CASE NO.

CHEMWEST I.D. _ EPA ID TAG NO SIZE ANALYSIS:

No. of containers.' Sample

Bottle seated: yes________No________,tape__________^OfikU! sealUbeE on bottle (other than preprinted label)

Packing Information (first lab ticket only)Chain of custody record rec'd: yes_____No.Official seaf on ice chest: yes_____ No_____; intact: yes_____No,.In lieu of scat-tape intact: yes_____No____^_EPA ID and tag numbers agree on all documents: yes______No——————Other Information:

573

Page 59: ftR3008L6 - semspub.epa.gov

. Order No

CHEMWEST Analytical Laboratories, Inc. Date Rec'd-.600 North Market Blvd. Compt. Date.

. Sacramento, CA 95834(916)923-0640 FAX (916) 923-1938

Notebook______

U3.EPA-VIAR AND COMPANY

COlstfRACTNO-SAS

ANALYSIS:

CASE NO.

.CHEMWEm.D. EPA ID TAG NO SIZE . ANALYSIS:.No. of containers_____________;____Volume.Sample Description:_____________________

Bottle sealed: yes_________No________tape___________Official sealLabel on bottie (other than preprinted label)

Packing Information (first lab ticket only)Chain of custody record rec'd: yes_____No______Official s«at on ice chest: yes______ No______; intact: yesIn lieu of seakape intact: yes_____No_____EPA ID and tag numbers agree on all documents: yes_____NoOiher Information:

Page 60: ftR3008L6 - semspub.epa.gov

I Order No_*.

CHEMWEST Analytical Laboratories, Inc. DateRecjd_600 North Market Blvd. Compl. Date.

* • Sacramento, CA 95834(916)9230640 FAX (916) 923-1938 " ———

I (Notebook_______

I U.S.EPA-ORGANICS

• CONTRAaNO.

IANALYSIS: HSL

I CASE NO.i , . .. . .!J CHEMWESTI.D. EPA ID TAG NO SIZE ANALYSIS:

\•No. of containers_________________Vofume.Sample Description:_______________________

Bottle sealed: yes________No________-tape__________Official sealLabel on bottle (other than preprinted label)

Packing Information (first Ub ticket only)Chain of custody record rec'd: yes_____No.Official seaf on fee chest: yes______No______• Intact: yes——————No.In lieu of seal-tape intact: yes______No______EPA ID and tag numbers agree on all documents: yes—————No—————Other Information:

RR300075

Page 61: ftR3008L6 - semspub.epa.gov

.NON-EPA SAMPLES

CHEMWEST Standard Ticket1} Check Ice chest, cardboard box, or wooden box for custody

seals, and check to sea that tape is intact.

2) Put disposable gown and gloves on and open ice chest,box, etc. in hood, organize samples; take picture of anybroken samples.

3) Check for chain-of-custody.

4} Label samples (white tape-orange pen), take Polaroid picture,

5) Take pH

£) Log in samples:a) Date and time received

b) P.O. No.

c) Contact person

d) Phone number

e) Analysis - write up - type~of containers, number,matrix, under chain-of-custody.

f) Location "~

g) How received - right corner, OTC, UPS, FE» etc.

h) If price supplied by client, put on right bottomof ticket.

i) Log sample I.D. nimber, Client and Date in SampleControl Log.

flR300876

Page 62: ftR3008L6 - semspub.epa.gov

CHEM WEST ANALYTICAL LABORATORIES,INC600 West North Market Blvd.Sacramento, California 95834(916) 923-0840 HOC (916) 923-1938

Order No._Date Rec'd..Compl. Date.Section;___

CLIENT- _________= ____- ___________ Project Name:._______________________________________ Project No._._________________________ pn NQ __

_______________________________________ Contact___________________________________________ Phone____ANALYSIS:

AR30U877

T n v\ nAvs TTXH r<; T nwnpT? TTMF is ARRANGED

Page 63: ftR3008L6 - semspub.epa.gov

RUSH SAMPLES

CHEMWEST Standard Ticket

Same procedure as other non-EPA

Stamp RUSH on ticket for the following analyses;;

Coliform, fecal & Total 6 Hrs.BOD 48 Hrs.Chlorine ImmediatelyChrome VI 24 Hrs.Color : . _ _ 48 Hrs.Odor 24 Hrs.Nitrate - 48 Hrs.Nitrite 48 Hrs.Ortho-phosphate & phosphorus 48 Hrs.Iodide 24 Hrs.Sulfite ImmediatelypH ImmediatelySettable Matter . ., . ..,.48 Hrs.Temperature . . . . . . " . _ __.." ImmediatelyTurbidity 48 Hrs.Methylene Blue Active Substances 48 Hrs.

RR300878

Page 64: ftR3008L6 - semspub.epa.gov

SAMPLE DISTRIBUTION

Internal Chain-of-custody - Log BooksAll samples removed from Sample Control will be monitored inlog books.

OrganicsInorganicsDioxins

flR300879

Page 65: ftR3008L6 - semspub.epa.gov

28 Federal Register / Vol. 49. No. 209 / Friday. October 2S. 1&&4 / Rules and Regulation!

r _ _ _ . _ _ _ ioTatla It Any perum may ao»unt »uch variunur duvi nut adwrauly ManUurinx and Support Laboratory ipply for* variance Irom the prescribed afted the inu-yrily of UK n» tuple. Such Regional Axlmmtiilrulorinuy gn.nl ftreservation techniques, container data will be forwarded by the Regional variance applicable to the specific /;jj| £*K!tatenaii, and maximum holding time* Administrator lo the Director of the dii-charte to the applicant, A deciut* •>•pplicable to samples taken trom a Environmental Monitoring and Support approve or deny a variance will bepecific discharge. Applications for Laboratory in Cincinnati, Ohio tor «uide within 90 days oTraceipt of ti*variances may be made by letters to the technical review and recommendations application by the ~Regional Administrator to the Region In lor action on the variance application, Adninifttntor.which the discharge will occur. Upon receipt of the recommendations • • • •Sufficient da.ta should be provided to from the Director of the Environment*!

TABLE u.—AEOUME& COMTAMEKC, PnecenvATiOM TECHMCuei. AMO HUAMQ Tut*

4.**fctl I

U.Ownc«

17.ft.

•i HfSav -r tfH)———————.Si. «a. MUMV trm my** Mtopvi.

I, M 10 It tt tl. «5 22. K tl & »-3< * S7. «i *t. Si, It S*-aO. 12, S3, t-Tl 14. n h*NMk me*t MOR«M VI •

«1 Ountf r«**i.

*- »«—*P*»|U <

«T.

aou.f* "nn •.frrarr" ———,at BMOUI fterttMraa* in*.M»T•l fabto tew«6 k«i

•?.•*•at

U. 1»-JC B. >«-», »*-*7, t»-4X «-<?. H. W, M. M. •*-«. ST.

a. P. K Pute eM ret,«, Aflrawr arc wpc

n-?4

P.O.P.O.

P.O-P.G.

P.O.P.C.P.G.P.O.P, G.P.G-P.G,

P,0-P.O.P.G.

P.O.P.G-P.O.

P.G.P.G.

G,.————————P.G———————P.G———————G am trtetaf.G««r.P.O.P.O.P.O.P.G.P.O.P.G.

P.G.P.G.

P.G-

P.G!P.O.

G.T

omrc.C-a.4-C,HJO.»||HCI.CM. re———_——CM J*

** -

Caot «t——————————___________.Caa.4t,twO»<topM>U.UeaKneaM*_»*O» » pM<L KJO. »fH<t.

Cea. «Xs iMO. to |H<1

Coa. __ __.Mlwwv

Cetf _ _______._toCoO. «'C. HCI v H O. * -H<1.

Coo. tt

Coot*t,

c-xrcwot*

CM. 4"C. UM% MAOi

CMtrc.

*•—' J**——-T-_**ffr- *fc1L*'

Cat rc. «0t% pto.Coo. «t ooorvnfc

HNO, k pM<S

•trnn.

'*»*

Ca&».

IP3S0880 i

TaMtSIMM*IB)

Page 66: ftR3008L6 - semspub.epa.gov

SAMPLE PRESERVATION

Complete and unequivocal preservation of samples, either domestic sewage, industrial wastes, ornatural waters, U a practical impossibility. Regardless of the nature of the sample, complete stabilityfor every constituent can never be achieved. At best, preservation techniques can only retard thechemical and biological changes that inevitably continue after the sample is removed from the parentsource. The changes that take place in s sample are either chemical or biological. In the former case,certain changes occur in the chemical structure of the constituents tthat are a function of physicalconditions. Metaf cations may precipitate as hydroxides or form complexes with other constituents;cations or unions may change valence sutes under certain reducing or oxidizing conditions; otherconstituents may dissolve or volatilize with the passage of time. Metal cations may also adsorb ontosurfaces (glass, plastic, quartz, etc.). such as, iron and lead. Biological changes taking place in asample may change the valence of an element or ft radical to a different valence. Soluble constituentsmay be convened to organically bound materials in cell structures, or cell lysis may result in releaseof cellular material into solution. The well known nitrogen and phosphorus cycles are example* ofbiological influence on sample composition. Therefore, as a general rule, it is best to analyze thesamples as soon as possible after collection. This is especially true whim the analyte concentration isexpected to be in the low ug/1 range.

Methods of preservation are relatively limited and are intended generally to (1) retard biologicalaction, (2) retard hydrolysis of chemical compounds and complexes, (3) reduce volatility ofconstituents, and (4) reduce absorption effects. Preservation methods are generally limited to pKcontrol, chemical addition, refrigeration, and freezing.

The recommended preservative for various constituents is given in Table 1. These choices are basedOn the accompanying references and on information supplied by various Quality AssuranceCoordinators. As more data become available, these recommended ho'!ding times will be adjusted toreflect new information. Other information provided in the table is an estimation of the volume ofcample required for the analysis, the suggested type of container, and the maximum recommendedholding times for samples properly preserved.

xv

Page 67: ftR3008L6 - semspub.epa.gov

TABLE 1

RECOMMENDATION FOR SAMPLING AND PRESERVATIONOF SAMPLES ACCORDING TO MEASUREMENT0'

Vol.Req. Holding

Measurement (ml) Container* Preservative*'4 Time8

100 Physical Properties• • - - . . .

Color 50 P,C Cool 4'C 48 Hit.

CofHirtmrf 100 P.G Cool 4'C £8 Days

Hardness 100 F,G HNOj to pH<2 £ Mot

Odor 200 G only Cool, 4'C 24 Hrs.

pH . 25 P,G None &rq. AnalyzeImmediately

* Residue

Filterable 100 P.G Cool 4*C 7 Days

Non- ._......Filterable 100 PrG Cool 4T 7 Days

Toul 100 P.G - Cool 4*C 7 Days

VoUUle 100 P.G Cool 4'C 7 Days

• Seuleable Miner 1000 : P.G Cool.-l*C 48 Hrs.

Temperature 1000 P.G Nooe.IUq. AnalyzeImmediately

Turfaidiry 100 P.G Cool 4X: 48 Hrs,

300 Metals

Dioolvcd to PtG Filter oo cite 6 Mas.HNO, to pH<2

Sufpe&ded 200 Filler oo aite

Total 100 P.G KNOjtopK<2

xvi

Page 68: ftR3008L6 - semspub.epa.gov

TABLE I (CONTD

Vol.., ' HoldingMeasurement (ml) Container' Preservative*"4 Time5

Chromium** 100 p,G Cool. 4"C 24 Hn.MercuryDinolved 100 P.G Film 28 Days

HNO, lo pH<2

Tool 100 P.G HNOj to pH<2 28 Dayt

300 loorEaaics. Kon-MetaHitt

Acidity 100 P,G Cool, 4'C 14 Days

Alkalinity _ 100 P.G Cool 4'C 14Da'y»

Bromide 100 P,G None JU<J. 28 Days

Chloride 50 P.G None Risq. 28 Days

Chlorine 200 P.G None Rrq. AnalyzeImmediately

Cyanides 500 P.G Cool 4*C H Days7NaOHwpH>120.6g auDrbic acid*

Fluoride 300 P.G None Roq. 28 Days

Iodide 100 P.G Cool *C 24 Hn.

Nitrogen»

Amrooma 400 P.G Cool4*C 26 Daysto pH<2

Total 500 P(O Cool 4t: 28 Daysto pH<2

Nitrate pJut Nitrite 100 P.G Cool 4"C 28 DaysH>SO4 tc>pH<2

Nitrate4 100 P.G Cool fC . <8 Hr»-

Nitrite 50 P.G Cool 4X1 48 Hn.

Page 69: ftR3008L6 - semspub.epa.gov

TABLE 1 (COND

Vol.Req. Holding

Measurement (ml) Container* Preservative*4 Time6

Dissolved OxygenProbe 300 G boule and top None Req,, Anatyie

ImmediatelyWinkJcr 300 O bottle and top Fix on site * I Hours

and storePbocpboms in darkOrtho-phosphate, 50 P.G Filter on tite 48 Hrs.Dissolved Cool 4'C

Hydrolyzable 50 P.G *Cool, 4'C 28 DaysH,S04tO!pH<2

Total 50 P.G Cool. 4'C 28 DaysH2SO4 to J»H<2

Total, 50 P.G Filter on idle 24 Hn.Dissolved Cool 4'C

H3SO4topH<2

Silica 50 P only Cool 4*C 28 Days f

Sulfate 50 P.G Cool 4'C 28 Days

Sulfide 500 P.G Cool, 4*C 7 Daysadd 2 ml linearrtuir plus NaOHto pH >9

Sulfite 50 P.G None Req. AnalyzeImmediately

400 Organic*

BOD 1000 P.G Cool 4X 48 Hr».

COD 50 P.G Cool. 4'C «DaysHjSO4 to pH<2

Oil & Grease 1000 G only Cool, 4*C 28 DaysH*SO4 to pH<2

Organic carbon 25 P.G Cool 4'C *& Days* H,SO4 or HQ to pH<2

Pbenolia 500 G cm!y ^ Cool.

AR30088^xviii

Page 70: ftR3008L6 - semspub.epa.gov

TABLE I (CONI)

VoLReq. Holding

Measurement (mJ) Container8 Preservative*14 Time6

•K

MBAS 250 P,G Cool 4T 48 Hn.

NTA 50 P.G Cool. 41C 24 Hn,

I. More specific instructions for preservation and sampling ait found with each procedure ASdetailed in this manual, A general discussion on sampling water and industrial wastewater maybe found in ASTM, Part 31, p. 72-82 (1976) Method D-3370.

2. Plastic (P) or Class (G). For metals, polyethylene with a polypropylene cap (no liner) ispreferred,

S. Sample preservation should be performed immediately upon sample collection. Forcomposite samples each aliquot should be preserved at the time of collection. When use ofan automated sampler makes it impossible to preserve each aliquot, then samples may bepreserved by maintaining at 4°C until compositing and lamplt splitting is completed.

4. When any sample is 10 be shipped by common carrier or sent through the United StatesMails, it must comply with the Department of Transportation Hazardous MaterialsRegulations (49 CFR Part 172). The person offering such material for transportation isresponsible for ensuring such compliance. For the preservation requirements of Table 1,the Office of Hazardous Materials, Materials Transportation Bureau, Department ofTransportation has determined that the Hazardous Materials Regulations do not apply 10the following materials: Hydrochloric acid (HCIJ in water solutions at concentrations of0.fM% by weight or less (pH about 1.96 or greater); Nitric acid (HNOj) in water solutionsatconcentrations of 0.15% by weight or less (pH about 1.62 or greater); Sulfuricacid (H*SO4>in water solutions at concentrations of 0.35% by weight or less (pH about 1.15 or greater);Sodium hydroxide (NaOH) in water solutions at concentrations of 0.080% by weight orless (pH about 1240 or less).

5. Samples should be analyzed as soon as possible after collection. The times listed are themaximum limes that samples may be held before analysis and still considered valid.Samples may be held for longer period* only if the permittee, or monitoring laboratory.has data on file to show that the specific types of cample under study are stable lor thelonger time, and has received a variance from the Regional Administrator. Some samplesmay not be stable for ihe maximum time period given in the table. A permittee, ormonitoring laboratory, is obligated to hold ihe sample for a shorter time if knowledgeexists to show this is necessary to maintain sample stability.

6. Should only be used in the presence of residual chlorine.

AR'300885

XIX

Page 71: ftR3008L6 - semspub.epa.gov

7. Maximum holding time is 24 hours when sulfide is present. Optionally, all samples may /"be tested with lead acetate paper before the pH adjustment in order todeiermme if sulfideis present. If sulfide is present, it can be removed by ihe addition of cadmium nitratepowder until a negative spot test is obtained. The sample is filtered and then NaOH isadded to pH 12.

8. Samples should be filtered immediately on-site before adding preservative for dissolved Imetals.

9. For samples from non-chlorinated drinking water supplies cone. HjSC>4 should be addedto lower sample pH to l«s than 2. The sample should be analyzed before 14 days.

xx

Page 72: ftR3008L6 - semspub.epa.gov

SAMPLE STORAGE LOCATIONS

A) RefrigeratorB) FreezerC) Metal CabinetsD) Dioxin Storage Room

SR3QQ887

Page 73: ftR3008L6 - semspub.epa.gov

SAMPLE STORAGE

Sample Control must be locked at all times when no one present.All cabinets, refrigerators, freezers locked except when per-sonnel present.

2. All samples are located by EPA & NQN-EPA Dump Slips.

3. Keys to cabinets, refrigerators, freezer are kept In a lock boxin Sample Control,

After sample log in, the samples are directly relinquished to alocked refrigerator, freezer or cabinet. Samples that need re-frigeration (all organics and some in-organics) ere kept in re-frigerator. Standards and Extracts are kept in the freezer orrefrigerator. All samples and extracts are checked in and outof their appropriate storage places in Sample Control Log Book.

Refrigerator R-l

Freezer -p-1

Cabinets C-lC-2CO

Dioxin Storage Room DSR

AR300888

Page 74: ftR3008L6 - semspub.epa.gov

COMPLETING APPROPRIATE FORMS

Dump slips - EPA & Non-EPA

Sample return form

Resubmittal Form

Intra-Lab Sub Sheet

Shipping Hazardous Samples

Bottle Request Form

CHEMWEST Chain of Custody

AR3Q0889

Page 75: ftR3008L6 - semspub.epa.gov

CHEMWEST I.D.

Category f Sample*

-----

•Date Reported

*Matrix Containers

Status;Already Cone.Discard ImmediatelyMold 30/60 Day*Bold Until notifiedReturn to Client

Va«e: - • • • • • • -

Fir»; ..............

IHVOICING / DISPOSAL FORM

MGR. • DATE RECEIVED

._..._.. Unit A&alyaisAnalysis Price I fcush Price Codes

. .... ..._ .___.. ._.._. ... _ ._... . .. ._ ..

• •

.

Total Price Invoice fMarket Sector Initials

»w*

Location

. . -... ,.._ . . . _ . . . _

. . . . . . . . . . . . . ....

Discard:Dispose/Move to;

Sink _, __Dump*t*r u ___ ,Acid V«itte Drum . ___Vast* Srilvan* Dru» _ ——

.............. . . K*d Tag"* -"-•——— t ~ Freccer fln^rinfion....... . : . . . Refrir.erator fill JOOoSU........ . Ambient

..... ....... Color Code (Vaste) __ __•

Phonej____________________________ Date Duuped

Page 76: ftR3008L6 - semspub.epa.gov

•EPA DUMP SLIP

SE »0:_______ ____ ANALYSIS:^ 'TPKEWT S»:

*

fit

rt KU-i.* TE DUE;

MANAGER:

CW Tf_

1

IVii);

'

EPA ID MATRIX CCWENTS" "" " BUMP

'

HAZ.VASTE. HPM H

REDTAG

••••MBl

SKIPPING PROBUXS:

ORGAKIC/lWDRGANIC:DCKP 66 DAYS AFTER DATA RHORTING: a n «-, r• HHOC

OXIN:DOKP 6 H3NTHS AFTES SAMPLE RECEIPT:)089!

UATE DUKPIB/IHITMLS /

tOCATJONCHEMVEST ANALYTICAL LABORATORIES

Page 77: ftR3008L6 - semspub.epa.gov

LU

G

£1COCO

CCLJUJE:O.E=30

J_,

E -* Sj. Ci E

HRIQQ892

Page 78: ftR3008L6 - semspub.epa.gov

ANALYTICAL LABORATORIES, INC.

6AHPLE RETURN FORM

CLIENT_________ ___ DATE

ATTN:

ANALYSIS OF THE ENCLOSED SAMPLE(S) HAS BEEN COMPLETED

CW ID'S:

SHIPPED VIAi

SHIPPED BYi

AR3GQ893

600W Nortn Market Boulevard * Sosromento, CA 95834 • Phone f?16) 923-0840 • FAX [916) 923-1938

A Sua*icfcary at Comp<-'Cf»»rn

Page 79: ftR3008L6 - semspub.epa.gov

ChexnWest Analytical Laboratories

NEW I.D. K U K B E R " "'

RESUBMITTAL FORM

CLIENT; _________________ PROJECT

ADDRESS: PROJECT NO:

PHONE: _________________ P.O. NO:

CONTACT:

OLD CHEMWEST I.D. SAMPLE I.D. KATRIX ANALYSIS

REQUESTED BY*_____________ DATE HEEDEDx————————

DATEs

Page 80: ftR3008L6 - semspub.epa.gov

ANALYTICAL LABORATORIES, INC.

IKTRA-LAB SUB-SHEET

CHEMtfEST I.D.

PROJECT DUE DATE_

MANAGER

SECTIOK PARAMETERS DATE DUE

COMMENTS

PROJECT MANAGED

6OWf4orth Ktartcef Boulevard •- Socram&nto. CA9S834 • Pbor* (916) 923 540 • FAX [916]

A Subsidiary & CornpuCf»m Corporotton

Page 81: ftR3008L6 - semspub.epa.gov

CHEMWESTANALYTICAL LABORATORIES, INC.

SHIPPING HAZARDOUS SAMPLES

Tape lid on Jar with electrical tape to prevent leakage.

Please put sample (802. jar) in cip-lok plastic bag.

Place jar in metal can with vermiculite (packing material).

Seal metal can with clips.

Place netal cans in ice chect packed in vermiculite.

Seal ice chest - fiber tape completely around two tinea.

Put necessary labels on ice chest:

FLAMMABLE SOLIDS . FLAMMABLE LIQUIDS

Flamnable Solid N.O.S. (Top) Flammable liquid H.O.S. (Top)UN 1325 (Top 6 Front) UK 1993 (Top t Front)Danger (Top) DangerFlammable Solid (Top & Front) Flammable Liquid (Top & Front)Co. Name, Address, ETC. Co. Name, Addreits, ETC.This End Up This End Dp

Put Chain of Custody in zip-lok plastic bag.

Tape plastic bag inside to lid of ice chest.

Ship Federal Express.

896

60CNV North Market Boutevard * Sacramento, CA 95834 • Phone £916) 923-OS40 • FAX [VTOJ

A Scfc*idtary of ConxxiChwn Corporahon

Page 82: ftR3008L6 - semspub.epa.gov

• 5CDreD.

•nuff " 1^^ Q

•LTa iCp

js:D

i^Stf 0^ CD

Ljf,1

REMA

RKS]

Sa3NIViNOOdOU39WnN

ccLU

5

COo"Z.occo

GO

oo

oo

UJ

Z

oLU

Occn.

N3HO 13M

S1V13N

810 - nooaoi-8'z4c's

(U38-S29OfrS8-frS9

s.aod0808/8090308/309

0108/109HJJ7X318NiX3/Hdi

PROJECT NO.

SAMPLERS

(SIGNATURE)

PRINTED NA

ME \

o(-og

UJ2'1-

UJ

a

6zLit_Ja.

i j .„.,,

TOTAL NUMBER

OF CONTAINERS

g

OoLU

3oUJcc

u

CO

QD

j]

L*

S

L

<s

maLU

C/)

iLUCC

e3acS>en

UJ

a

oUJ

UJOUJtrUJ

O

maUJT

1UJCC

5

o2cen

PMENT HANDLING

: REQUIREMENTS

wO

11E

ID

B)

C1. o

JJEp . -: :

<D

ffl

1C

a.

aa1

jjEh—

a

Ea.

IR

o£ O

Hal

11

3

1ao

L_^

a

REC'D BY (LABORATORY)

imaUJi

a_iUJcc

i

mnUJ

UJoUJCC

UJ

o50

|

o

B

§•CO

£recin

39

ictn

aZ

a.

uE

1BJ

.Ca.

jj

aEDzIC'CaLU£P

7

(CZ

a

.

csOO

1o.1

aao

600W

North Market Boulevard

• Sacramento, CA 95834 •

Phone (916)923-0840

Page 83: ftR3008L6 - semspub.epa.gov

ANALYTICAL LABORATORIES, INC.

To expedite the correct analysis and identificationef your sample*, please complete this forn. If you are"unsure of the analyses required, do not complete the"Analysis Required* part of the form. Please indicate acontact person and phone number which ChemWest staff can useto verify the appropriate analytical requirements.

Co. Names________________ Project Mcyr._____________

Project Name/*___________________Telephone

Your Sample I.D. Matrix Analysis Reauired

Commentsi

flR300898ChemWest contact person is:

60CW North Morkef Boulevard • Socromento, CA 95534 • Phone |[916] 923-0840 •A SuCwicKrY of CompuCr«rn Cor pcarotiori

Page 84: ftR3008L6 - semspub.epa.gov

CONTAINER REQUEST FORM

WATER/LIQUID

____ 40-ml VOA vial ____ 6-oz Amber glass

'• 40-ml VOA vial + HCL ____ 8-02 Amber glass + H..SO,———— 2 4

____ 1-gt Amber glass ____ 8-oz Amber glass + KaOH

____ 1-qt Amber glass + H2S04 ____ 8-02 Amber glass + HNO-

____ 1-gt Clear glass ____ 1-gt plastic

____ 1-gt Plastic + H2S04 ____ l~Pt Plastic + NaOH

____ 4-oz plastic sterile ____ 1-pt plastic + HN03

SOIL/SLUDGE : .._.__...=_._..,

____ 2-oz VOA jar

____ B-oz vide mouth clear glass jar

OTHER ...... ___... ..„. _ . . . . . .... COMMENTS: •_________

____ Chain of -Custody Forms _______________________

____ Custody Seals __ __ ___.____________________

CONTAINERS __ Ice Chest.__ Box

Blue Ice

CLIENT: ___. _:_ CONTACT;

PHONE: ( )

SHIPS ___ UPS ___ FED EX ___ GREYHOUND ___ KAIL ___ O.T.C. ___ OTHER

REQUESTED BYt __________' . ... . . _ DATE REQUESTED: ____________.

DATE WANTED:

.SHIPPED BY:

DATE: AR300899

LYTICAL lASORATCftlES. INC.

Page 85: ftR3008L6 - semspub.epa.gov

CHEMWESTANALYTICAL LABORATORIES, INC.

Attention:; Field Staplers*

The following sample containers are used for sample collection atCHEKfflSST. You may want to use the folowing list as a guide insample collection.

LIQUIDS

analysis (examples) Container(s)

volstiles (601, 602, 624) duplicate 40 ml VOA vials

nost organic extractables duplicate one-quart amber glass(658, 625, pesticides)

petroleum hydrocarbons (gas, one-quart amber glassdiesel)

EDB one-quart amber glass

oil £ grease one-quart clear glass t^SO^ preserved

TOC, COD 8-ounce amber glass, H2&04 preserved

penta and tetrachlorophenol 8-ounce amber glass, NaOK preserved

TOX ..._.- -8-ounce amber glass, HN03 preserved

most tret chemistry (general one-quart plastic bottlemineral)

total phenols one-quart glass, 1*2804 preserved

total cyanides one-quart plastic, NaOH preserved

colifotm bacteria 4-ounce plastic sterilized container

metals one-pint plastic, HN03 preserved

CAOTIOH: PRESERVED COHTAIKERS MAY COSTAIK ACID

SOLIDS

soil or solid sample analysis 8-ounce wide »outh glass jars

Vote that you may have-requested a specific anlaysis not listedabove. When in doubt, fill all the containers vhlcK we have shippedDnder no circumstances should you rinse these containers prior to fillingOnly the 40 ml VOA vials need to be completely *ullflDon " alothers should be left with a small headspace. flitJUU9QQ

Contact Person at CHEKWEST600W North Mo*et Bouh-word • Sacramento. CA 95SW • Phooe [°16) 923O&40 • RAX (916] 923-1938

Page 86: ftR3008L6 - semspub.epa.gov

CHEMWESTANALYTICAL LABORATORIES, INC.

CHEHWEST COURIER SERVICE

Date request received - ____ Receiving individual

Request received via phone_ mail _ other _______

Contact

phone (__)

Company _mAddress

Date pick-up needed _____________ afternoon __ morning

Type of samples to be picked up

Containers needed: yes _ no __ see attached

COMMENTS: " _____

r^ Marke* toijlevavl " ' V- ""r r .' -;C^9C83^ - Phone (916)923-0640 • FAX (9 16)923-1938

of CompuCt>em C rporolion

Page 87: ftR3008L6 - semspub.epa.gov

DUMPING OF SAMPLES

Whenever possible all remaining samples are returned to the client

Remaining samples are packed into 55 gallon drums.

Liquid drums, solvents, PCS oils, acids

Open face drums - all soils, including hazardous soils and liquidspacked in vermiculite

Color Code will be implemented 4 added to S.D.P™ as soon as in-formation is received from American Environmental Waste Disposal

&R300902

Page 88: ftR3008L6 - semspub.epa.gov

SHIPPING OF SAMPLES & ICE CHES5TS

A) Non-hazardous Samples

B) Hazardous Samples

C) Correct labeling of Hazardous Materials

AR300903

Page 89: ftR3008L6 - semspub.epa.gov

NON-HAZARDOUS SAMPLES

Samples shipped to other locations:

Asbestos, Check samples ancf other non-hazardous samples:

Place tape around caps and lidsPlace in ziplock bagsPut in box or ice chest IPack in vermiculite or styrofoam with ice, if requiredComplete necessary paper work, C of C, Hemo

Page 90: ftR3008L6 - semspub.epa.gov

SHIPPING HAZARDOUS SAMPLES

| Pack samples in ziplock bags end place in metal cans withvermiculite and seal with metal clips.

* Pack cans or containers in vermiculite in ice chest with ice, if necessary.

Seal ice chest with custody seals.

Complete necessary paper work:

Chain of CustodyI Seals on ice chests* Hazardous labelstffc See attached sheets (figures 1, 2, 3)"•

3R3Q0905

Page 91: ftR3008L6 - semspub.epa.gov

SOP FOR DISPOSAL OF LEFTOVER SAMPLES FOR VOA

These are the unused portions of samples received for VOA andthe portion remaining in the VOA vial after an albquat hasbeen taken for analysis.

- Combine the solutions in a 5 gallon carboy;- When fulli mix well and have an aloquat analyzed byGC/MS for VOA.

- Have the results reviewed by the CC/MS lab manager.- If within tolerance, dump down the drain followed by

excess water.- Submit the reults and a completed waste disposal form

(Figure 1) to the Safety Officer for filing.

AR3QQ9Q6

Page 92: ftR3008L6 - semspub.epa.gov

JMAS-TE DISPOSAL FORM

Check OneLEFTOVER VOA SAMPLES

SOLUTION FROM METAL DIGESTS

DILUTED DISTILLATE FROM CN ANALYSIS

Date Reviewed by _______________ ______"' ,._ Date

Neutralized by ______________ ______Date

Volume dumped in sink _______;____ ______Date

ATTACH RAW DATA AND CALCULATIONS

3*30090

Page 93: ftR3008L6 - semspub.epa.gov

PERSONAL ITEMS

Ice Chests:

Ice chests are returned to clients ASAP

All ice chests are cleaned, labels torn off ormarked out

Tape and returned per contract specified carrier,e.g. UPS-EPA

Copies of UPS-EPA shipments sent to Viar 4 Co.

Page 94: ftR3008L6 - semspub.epa.gov

SECURITY OF SAMPLE CONTROL ROOM

Sample Control must be locked at all times except when authorizedpersonnel are present.

The back door to Sample Control must be locked at all times ex-cept when receiving samples.

Page 95: ftR3008L6 - semspub.epa.gov

SANITATION

Sample Control must be kept clean at all times. This includesfloors, refrigerators, cabinets and hood.

Refrigerators - clean quarterlyFreezers - defrost annuallyCabinets - quarterlyJlpod - after cach^use ._....Floors - vaccuum dally

""3009)0

Page 96: ftR3008L6 - semspub.epa.gov

PHONE LOG

Kept at Sample Control deskLog all in-coming callsInclude the following information:

ClientPerson CallingTime & DateNature of Business

AR3009I

Page 97: ftR3008L6 - semspub.epa.gov

SAMPLE CONTAINER SOP

40 ML VOA VIAL

SOURCE! I-CHEM RESEARCH, XKC.1-800-443-16891-415-782-3905

PREPARATION; THERE IS NO PREPARATION FOR THESEBOTTLES. THEY COME PRECLEANED FROMI-CHEM. ATTACH A 2 X 3 1/2 CHEMWESTID LABEL.

PRESERVATIVE; NONE

40 ML VOA/HCL

SOURCE; I'CHEM RESEARCH, INC.1-800-443-16891-415-782-3905

PREPARATION: THERE IS NO PREPARATION FOR THESEBOTTLES. THEY COME PRECLEANED FROMI-CHEM. ATTACH A 2 X 3 1/2 CHEMWESTID LABEL.

PRESERVATIVE: PLACE 0.3 HLS OF 1 + 1 HCL INTO A VOAVIAL. GET THE SOLUTION PROM GC GROUP.

1-QT. AMBER GLASS BOTTLE

SOURCE; I-CHEM RESEARCH, INC.1-800-443-16891-415-782-3905

PREPARATION: THERE IS NO PREPARATION FOR THESEBOTTLES. THEY COME PRECLEANED FROMI-CHEM. ATTACH A 2 X 3 1/2 CHEKWESTID LABEL.

PRESERVATIVE: NONE AR3009 I 2

8-02. CLASS/H2S04

SOURCE: AMERICAN SCIENTIFIC PRODUCTS1-800-762-8610CATALOG NUMBER B7465-13

Page 98: ftR3008L6 - semspub.epa.gov

PREPARATION: RINSE WITH TWO 30 ML PORTIONS OFDISTILLED WATER. THE DISTILLED WATER ISFROM THE INORGANICS GROUP, ATTACH A 2 X3 1/2 CHEMWEST ID LABEL.

PRESERVATIVE: NONE

1-QT PLASTIC/H2S04_

SOURCE: AMERICAN SCIENTIFIC PRODUCTS1-800-672-8610CATALOG NUMBER B7510-32

VWR SCIENTIFIC1-800-841-0617CATALOG NUMBER 16058-145

PREPARATION; RINSE WITH TWO 30 ML PORTIONS OFDISTILLED WATER. THE DISTILLED WATER ISFROM THE INORGANICS GROUP. ATTACH A 2 X3 1/2 CHEMWEST ID LABEL AND A PRESERVEDWITH SULFURIC ACID LABEL.

— — - PLACE 2 MLS OF SULFURIC ACID (H2S04J"INTO BOTTLE. THE SULFURIC ACID ISREAGENT GRADE OFF THE SHELF.

1-PT PLASTIC

SOURCE: AMERICAN SCIENTIFIC PRODUCTS1-800-672-8610CATALOG NUMBER B?510-16_.......

VWR SCIENTIFIC1-800-841-0617CATALOG NUMBER 16058-123

PREPARATION: RINSE WITH TWO 30 ML PORTIONS OFDISTILLED WATER. THE DISTILLED WATER ISFROM THE INORGANICS GROUP. ATTACH A 2 X3 1/2 CHEMWEST ID LABEL.

PRESERVATIVE: NONE

flR3009I3

Page 99: ftR3008L6 - semspub.epa.gov

1-PT PLASTIC/HN03

SOURCE: AMERICAN SCIENTIFIC PRODUCTS1-800-672-8610CATALOG NUMBER B7510-16

VWR SCIENTIFIC1-800-841-0617CATALOG NUMBER 16058-123

PREPARATION: RINSE WITH TWO 30 ML PORTIONS OFDISTILLED WATER. THE DISTILLED WATER ISFROM THE INORGANICS GROUP. ATTACH A 2 X3 1/2 CHEMWEST ID LABEL AND A PRESERVEDWITH NITRIC ACID LABEL.

PRESERVATIVE: PLACE A 2 MLS OF NITRIC ACID (HN03) INTOBOTTLE. THE NITRIC ACID IS REAGENTGRADE OFF THE SHELF.

AR3009U

Page 100: ftR3008L6 - semspub.epa.gov

CONTROL SAFETY RULE'S

Safety glasses are highly recommended for anyone in the vicinity or*some one handling dispo pipets or other glassware that J&ight Jbe thrown ikthe trash can. k

£ab coats are highly recommended /or All*Cloves are mandatory /or all operations that could conceivjfabiy

allow the hands o/ someone to come in contact with acids, base's, solventsamples, glassrwool, and any similar aaterials.

person should wash their hands as o/tezi as is possible and as coo;a/ter hand J in g1 acids, bases, solvents, etc. as is possible.

Throw dispo latex gloves into the trash as jroon as they are taken o//.There is no smoking permitted in Sample Control.There is to be no eating or drinking in Sample Control.Cloves are reguired when sample* or received materials may come in

with the employees hands.Sa/ety glasses are aandatory when glass containers are being thrown

disosal container or trash can.£jcployees should wash their hands as often as possible to prevent the ;

o/ any contamination that Bay have occured due to shipping or sampling ttechniques, this includes paperwork, sample containers, packing materialLift only those items that are easy to' lift. Get assistance to lift or«ove heavy items. Lift with your legs not your bacJc.Company vehicles flay be operated for company business only. Buckle up-its the law. State and .company—— drivers and passengers.For the disposal of wastes, see company safety xianual /or wastespecific methods.

JCnow the location of all waste containers and theirrespective contents*

JCnow the location and use of all First Aid materials.JCnow the location and use of all emergency ©giiipmcnt; spill Jcits,

showers eye wash, /ire extinguishers--——Clean up spills immediately.Practice good house keeping and neatness in all wcrfc area.*;.Always add acid to water — «ot the reverse.If acid or base comes in contact with your ffcin, wash area with

running water for atleast ISmin. Notify Supervisor o/fl$gi§$i§.| 5Koti/y Supervisor of any and all accidents, dangerous situations,

incidents, hazards —— -— Do not assume that the Supervisor/manageralready Juiows .

/AR300915

Page 101: ftR3008L6 - semspub.epa.gov

Ko opened toed foot wear will.,be allowed.Be sure you Jtnow how to use a particular piece of apparatus or

how to perform a particular method BEFORE you undertake a tasJc. Askand expect instruction*

Do not use any unlabeled materials.Always label containers that you have transferred to.JCnow the location -and use of all emergency *guipment in lab.

iJCnow the location'and use of all First Aid materials.JCnow the location's/ waste container and their respective

contents.Do not dumpY acidic, basic, or organic wastes done theDo not perform task assignments in which you have not been

trained. Ask for and expect instructions and training.Jtead labels*Never use the contents of an unlabeled container.Always label the container to which you have trans/erred some-

thing to.Kash you hands and exposed areas as often as possible through

out the wor* day and especially after handling reagents and/orsamples. i

- - - : . - - . . . - . . _

Be sure ventilation devices are wording, zioti/y Supervisorof any suspected mal/unction.

AR3009I6

Page 102: ftR3008L6 - semspub.epa.gov

Cheia West Analytical LaboratoriesAddendum E

Facilities and Equipment Inventories

&R300917

Page 103: ftR3008L6 - semspub.epa.gov

FACILITY ANDEQUIPMENT INVENTORIES

ftR30Q918CHEMWEST A n a l y t i c a l L » b o r « t c r t « » , Inc.

Page 104: ftR3008L6 - semspub.epa.gov

ANALYTICAL LABORATORIES, INC.

THE CHEMWEST FACILITY

SPACE ALLOCATION October 19B8

Total Facility = 14,600 square feet

Individual Areas Sguare Feet

1. Sample Receiving, including: 1,050

Walk-In Refrigerator [60]Walk-In Freezer [60]

2. Organic Sample Prep. Lab 520

3. High Hazard Sample Prep. Lab 320

4. GC and GC/MS Lab, including: 1,360

VOA Room [350]

5. GC and GC/MS Refrigerator Room 120

6. ICP/AA (Metals) Laboratory 300

7. Wet Chemistry/Inorganics Sample 650Digestion Lab

8. Data Control 480

9. Computer Room and Data Storage . ._. 540

10. Electronics Repair Shop 12011. Office Areas 2,340

12. Clinical (Separate from 1,150Enviromental Operations)

13. Miscellanous 3,850

14. Warehouse/Storage 1,800

AR3009I9Total .. .

600W North Marke! Boulevard • - Sacramento, CA 95834 • Prrone ivioj y i-utwu * FAX (916] 923-1938

A Subsidiary of CompuCfvem Corporation

Page 105: ftR3008L6 - semspub.epa.gov

flR300920CHEMWEST A n a l y t i c a l ta berator1•> > Inc.

Page 106: ftR3008L6 - semspub.epa.gov

ANALYTICAL LABORATORIES; INC.

CHEMWEST GC/MS/DS ANALYTICAL INSTRUMENTATION

AND EQUIPMENT»

October 1988

2 - Finnigan 5100SP GC/MS Systems with Turbomolecular Pumps,and Incos Ill/Nova 4X Data Systems, including the NBS 42000compound Mass Spectral Library

1 - Finnigan 5100EF GC/MS Systems with Turbomolecular Pumps,and Incos Ill/Nova 4X Data Systems, including the NBS 42000compound Mass Spectral Library

1 - Extrel ELQ-4001 GC/MS .System with Diffusion Pump, and anIncos Ill/Nova 4X Data System, including the NBS 42000compound Mass Spectral Library

2 - Finnigan Stand Alone Incos Ill/Nova 4X Data System,including the NBS 42000 compound Mass Spectral Library

1 - Tekmar LSC-2 Purge and Trap Unit with Automatic Sampler

1 - Epson Equity III System (IBM-AT Compatible) with Epson DotMatrix Printer and Interface to Finnigan GC/MS DataSystems and Finnigan Formaster Softvrare for CLP FormsGeneration

AR30092

600W North Morket Boulevard • Sacramento, CA 95834 • Phone [916] 923-05

A Subsidiary of CompuChem Corporation

Page 107: ftR3008L6 - semspub.epa.gov

too£ O. 4Ju oc o

I

o

8s coM CO

• Co4J n* 0) 0)fH *J 4JrH (0 fQS ° *SB CDC <~

Co-HCD-H

0}OS*s»"t0)•o£

•aE 0)0) EGO M^ O»— t <W(0 l-tC Q)rt &

t-lQ}(-tP4-1O10*wsca£

4-JC0)

{44-JEC

.CO"x0)

Pktooo*in

S5

4.

C•HXO Pu-H t4Q U

C(yth

•HCc4ft*

SOCD0SO

in1 CT>SfO

rH5S ^o a KC5 M W

t-cox01

X

N , njs OB 2t) MC M

1-101 COaoE) GCO M

(QjS3

•Ha•3 M

0)C C(0 Q)•HC 10C 4-1-H (Cfe Q

EHO 1

^to ^104J «tO CkQ H

6

t*-CO*"«N

•tt•O

f>3fSrgN*

t£K

COCQ2

fr(0

CQ A aCQ H *O* a c

rH Ox to OwM M ES3 4J O\ U U

U CO *-"

(*.COxM

&COoorHin

(X.t-3u«.2CO

gIQF•HCc•H

COO1«H

mfS ml enSCTi

rH£U G B5U t-t CO

r*03>ivP-J

X

N^ gg*t O

1Q %U t-H13 wM W

(P OTao13 QtQ M

*(55

•H* f•B t-l0}

IS C10 0)I?*C3He e•H 0IM Q

6 csa> i'm um- .n-U «t(C O

(s»CO*s»OJ

•CO*5&uOSN01O3

*"

ifljsc

CQ035

&10

CQ A COO-H-O10 J B

S3 (0 QJSi^T*V. U L>< 0)

COCOxorH

Et*UOOrHin

§*.C•HMO•HQ

Cid-HCc-Hfe

*CO

rH

COm r*i f>

*t ftup. as

COCOor"i

X

w (05 oto SO MC MM Wt-l0) CO9 QCO M

*g£S5

rH10

•5 M0)

C C10 01cnoJC CC 4JH 63fej ci

01 1E U^

22*"-*n?.e oat - - - - o M

CO*>

CO

uXc

flR300922

Page 108: ftR3008L6 - semspub.epa.gov

01

EO•rl *4J CO0 0) O«-* +J 4J^H 0 00 G*0jj Q<CO DC ^ ^

^ ^M

cO-HCO•H^0)"XfctH01•oo£

•oCO QJOJ Eto &^ Or-t <M0 (-)E 0)< CM

HQJrlJj4JCJ0(M3B0£

4JC-OJig63C-

~"

ffcCO>v''ff

•CO*5&4g01010101^

rtjas

COCQ

5

t?0

a A coCO -HT3(0 tJ C£ 3

r-t Qffi (0 CU

Si o^.o u< 0)

pCO

• "x^Oi

ux^ofcorH COO0 E«* 01 -H

M <0w >

Cu3O•i

1

f«l0)4JK

X55M^»••*

If1

to O£U QO M

f*tCO0i

X

v (0$ oto asO WC MM rHtt0) COcxo53 0CQ H

<4as

r-H(0•a i-i

0)C E10 QJ-HE 0E 4J•H (0 ,&4 Q

g *0) 14-* £c

">(0

10 GG w

p«*CO*x01

•CO•O&a010103OS•*

rtjas

COCQ

5

&0M 01

10 «O *OCO H E10 (4 S£ o

r-t &K JO EM i-< oas 4J u< Q) %cu aw w «B

t*.COA

u*4(Bg

*

OJ1Uto

gss

4mg50?*

•ap, 1£te oM CtH CO•8 03 \D

01 co tnty>M *lt 4t53 Q asCU i-c CO

fK

COID

CQO Mt) HHcrH CO1 OM Gi,*"13 >s.CQ Z

ga

•H(j•B l-l

0>C C(D 0)•HB aC 4-*H 10ftu £

(PEO

E 16Q)jj >OCO ECO wf0

C Go.w

COCOxCO

COOMO t-lEH CO1 OM G0) WaCO 5

rtjJC

c•0 M

0)E Be Q)cnuHB 10C4J1 83r^O

OBOr-te <e

0)4J *Oa E

0

«QG M

[••.COxoarH

XJ

MHH rlrH 0)

4-)rN ID4J 0

IIITOWfc

*cas

E•2J 0tnC-H0 ECO Ed-HWfa

£0)4-)

0

« GQ H

o**

Zur

Page 109: ftR3008L6 - semspub.epa.gov

&HEMWESTANALYTICAL LABORATORIES, INC.

CHEMWEST GAS CHROMATOGRAPHIC ANALYTICAL

INSTRUMENTATION AND EQUIPMENT

October 1988

9 - Varian 3400 Gas Chromatographs

1 - Varian 3300 Gas Chromatograph

5 - Varian Electron Capture Detectors (BCD)6 - Varian Flame lonization Detectors (FID)

4 - Tracor Photoionization Detectors (PID)

5 - 0 1 Electrolytic Conductivity Detectors(ELCD - Hall type)

2 - Varian Thermionic Specific Detectors (TSD)

1 - Varian Flame Photometric Detector (FPD)w )^ 3 - Varian Split/Splitless Capillary Injectors

3 - Tekmar LSC-2 Purge and Trap Systems with 10-Port AutomaticSamplers

1 - Tekmar LSC-2000 Purge and Trap System with 16-PortAutomatic Sampler

1 - Tekmar LSC-4000 Purge and Trap System with 10-Port^Automatic Sampler ; -

4 - Varian 8035 Automatic Liguid Samplers

2 - NEC APC IV Computer Systems (IBM-AT Compatible) includingPanasonic Dot Matrix Printers

1 - Hewlett-Packard 3357 Laboratory Data Aguisition System

.AR30092I4

600W North Market Boulevard • Sacramento, CA 95834 • Phone (916) %o-utwu * FAX (916)923-1938

A Subsidiary of CompuCtem Corporation

Page 110: ftR3008L6 - semspub.epa.gov

a s"ftV, to SJ~ g I

CQcE3r-!OL>

EO-*H *-*.*J 030 QJ Or-(4J 4Jr-l 0 100 G*d4J aa »B *^M

BO•HCO-H>0)C«-l0)•ci•o

CQ 0)a im CNOI-H IHc nB 0)< PH

t-l0)M34->O0<H3B0£

4->EOJS^4->toBH

•O0>Xu0PH

P*CO"X.tO

GUUr-l0S CO*O J•03\Sin

fno oO CDrr*> «8

1P-*J

EHOCOw -PH CQ EH

PS CQU W WO tC PH

E0t-tW0>

inmOrfoo

rH «*m«t *t

U G &U M CO

&0•HrH•HOk0U

tOGO^01rH

^GL>U

G *BMfe -

• CO" • 3* c <

•Ho inO &4fO" (C Om O co

nP*b3£«>£E-*

B0•HU0>

*•c iin01loo

N ro

«t *tU Q 55U W CQ

N0) Mfc *Or-|J3fH« nHtPOj0) 0£0

p*COto

•-tcD•D

>O G0 UTf &(n w

<

^

B0•HU0£

\Dinrofnloon ro

*t *tu G aC W CQ

•OQJXU0PH

P*CO^rH

b.GUtJ GH w

&H•X.>«BO »0 Q•«• Hn p4

<gr*

B0-HH0>

^rH*«•fO1Oo*? m

«t *tU Q B5U w CO

•d01Xu0PH

P*CO*to

«G%K G

rH* .fc*•BO -0 Gr» Hcn p.

EEH*.<gr*

B0•HUIQf

*fmo**loo

VI fOn»t tc

U G SSU H CO

•o -QJxtl0PH

1CCOoarH

«OuurH«80S *•O -•«-»w^ E i-3>•*<o - ino cu<n•00m o eo

I&4

*>3EHUCOU -PH « EH

CC COO U WOffiPH

B0•HM0>

inrHrmloo

1C m«t »t

U G fcUHlfi

o) £rH 0OrHjQrH0 Titna0) «£U

toCO*01rH

*

^

GCOEH Q

0 ~S -*•C • CO

T-lrH>C43£ -H 0

rHO • -H0 fX(C^ B >ID U 0

EHCOECPHPHO

B0*-lM0>

flRii0310925

300925

Page 111: ftR3008L6 - semspub.epa.gov

•aOJ3E•rt4-1

OuCOCOOt

o4J

COI•HOu

EO•H — *4J CQ0 QJ 0)•H4J 4JfH 0 00 Q -C4J QCO DB ^M

BO-rt03•H^Q>UrHOJ1£•o

C Q)QJ ED EHNO— ,J u_i• ^ *p0 MB 0)< PH

MOJH34JU0*H3B0£

4JEQJ,

£44-JCQBH

0) MH 0

0 -HtnaQ) B£ U

COCO"SrH

GM(M

rH03 CQ*D "s mo o0 CO

ro«B

*OZ

PH*tc *s«&§

B0•Hf-t0

>

rHtoO1oo

CO *ft)

tt 4fcu GZC5rH CO

0) wUt (C£30-H

£ ft£U

COCOin

•8

OOu:> ,So0 QfO P-<

gEH•.

<*

1

B0j0

rHCOP-inlOoro

te <tOG ZO H W

QJ U»H 0

0 -HQJ tO£U

COCOf*.

•21Gwfc

xs

o0 GfO PH

*4§

C«•H

IQ

>

UCO

O\D1O

O 0rH <3-

pT)

ft XU G ZO H CO

Er0rHrH•rlcx0U

COCOOJ

• •»C^CU E

•-» 0 PH 0)0 U 4JP i-H 03G*-H B N0 3 CO

^3-0s-e c-4J

0 » - «CD Q t-iQCO U Ein ttt *H <1

EHCOPH Edi mPw CCJ OU CQ

1V hOJ 10S UOJ IQS PH

COrH P-

Err>QJ OQJ <kt-iu inWrH

OJ

<t %UG ZO fH CO

U*Hau

CD(0(SJ

«aM1^4 '*n

X.B

o -O% &4l> 0in u

55«B W

< a> COtPH ZJ CQU<

|4J *Oji tJOJ 0«"H XS oOJ 0E Cu

01

B0]0)1-tuCQ I

% *tU G ZU H CO

g

p*COrS

pinn

V0

0 J4J5 uO CrH UrH O0q* t _[

14J fj4J t-l0) 0•HXS U0) 0KPH

SE r iYJ01 O

4J < <(Q rH v3w 5

co i orH

0

§

P-CO*"SfcrH

COJrtj

v^s

01tCJCOJ

<Iz

t)0EXQJEH

) Ef)§C* IB ** W •

V<

*?fj *3i in

Q) fD

i '**-•*-

. /"Cw

J

0 G ZQMCO

^ _49

Uo nU)VZUJ

01

o

UDCO<T*CDOCOcc

Page 112: ftR3008L6 - semspub.epa.gov

•ea;3E•H4J

§

COCOenrH

O4J

RS

rHOu

EO-H —^4J CQB 01 Q)rH 4J 4JrH 0 00 Q*Cjj QJCQ SE «—w

BO•HCQ-H

OJ' K*XkrH01*£

TJCD QJ

10 5>*orH IH0 ME 0)< PH

H0)H34JU0*H3E0£

jjBQJ3H4-1BJ_CrH

,g*Z

p*CO"Xfcto

CQJS\s

OJ1CJCO**

<Jz

1-10IQJEH

COorH

0tH *8EH•3 01

1 010} 0]

& % %3 Q ZCM H CO

^Z

pCO>»fcto

CO(J*ssoooJ*1uCO14

*4<Z

(•)0

QJEH

a0H 1EH

•BI 1

0)tn % %£ M co

g

coCO*"m

COj<x^

oalCQ

<

i-l0

0)EH

aeij iEH

•8•C

1 1&1H 9t It3 G ZPH H CO

,4Z

COCO*«1orH

COJi£

Xsooo011uCO

*

KjJz

H0

50)

OH0l I

B

•81 1

OJtnM % 4tSG ZM CQ

*.o4*

O

_J

c<

u*ZutXo

flR300927

Page 113: ftR3008L6 - semspub.epa.gov

ICHEMWEBTANALYTICAL LABORATORIES, INC.

CHEMWEST INORGANIC ANALYTICAL INSTRUMENTATION

AND EQUIPMENT

October 1988

1 - Thermo Jarrel-Ash Model 61 ICP with Autosampler and IBMData System

2 - Perkin-Elmer 5000Z Graphite Furnace AAs with Autosamplersand IBM Data Systems (1st delivery scheduled Oct-Nov 1988;2nd delivery scheduled Dec-Jan 1988-1989)

1 - Perkin-Elmer 3030Z Graphite Furnace AA with Autosampler andData System

1 - Instrumentation Laboratories Video :i2E AA with GraphiteFurnace and Data System

1 - Instrumentation Laboratories Plasma 200 ICP withAutosampler and Data System

1 - Perkin-Elmer 5_OB Mercury Analyzer

1 - Varian VGA95 Hydride Generation Appartus

1 - Perkin-Elmer 1310 Infared Spectrophotometer

1 - . Dionex 4000i Ion Chromatograph with Autosampler and HitachiIntegrator

1 - Technicon TRAAC 800 Autoan.alyzer with IBM XT Data System _

1 - Milton-Roy Spectronic 21 UV/VIS Spectrophotometer

1 - Sequoia-Turner 390 UV/VIS Spectrophotometer

1 - Orion 940 lonanalyzer with RS232 Computer Interface andSpecific Ion Electrodes for pH, F-, NH3, C12 , C103, andOxygen (DO and BOD)

1 - Monitek 21 Nephelometer (Turbimeter)

1 - YSI 35 Conductivity Meter .. . .

1 - Labconco COD Digenstion Unit (plus Hach COD Kit)

1 - Penske-Martens Closed Cup Flashpoint Appara\"^UQ928

600W North Market Boulevard * Sacramento, CA 95B34 • Phone (916) 923-0840 -

A Subsidiary of CompuChem Corporation

Page 114: ftR3008L6 - semspub.epa.gov

CHEMWEST INORGANIC ANALYTICAL INSTRUMENTATION

AND EQUIPMENT

October 1988 (continued)

1 - Labconco Rapid Still II Auto Distillation/Digestion System(Kjeldahl)

1 - Blue M Fecal Coliform Incubator

1 - Barastead Nanopure II with Corning MegaPure MP-3A Still

1 - Bamstead Nanopure II with Barnsteadt Fistreen #2 Still

1 - Top Loadng Balance, Scientific Products TL1600

1 - Boekel Dessicator

2 - Blue M Water Baths

Miscellaneous Laboratory Glassware and Method Apparatus

- CN Distillation Appartus

Imhoff Cones

- Nessler Tubes

- Titration Apparatus

- Certified Thermometer (Fisher Scientific 14985-5E)

AR300929CHEMWEST A n a l y t i c a l L*bor*tor1«*. Inc.

ES7 ANWYTlCAL LABORATORIES. INC.

Page 115: ftR3008L6 - semspub.epa.gov

BO•H •"*jj n« QJ QJ•H JJ JJ•H 1C (0fl Q-O4J . PJto DC <~M

BO•rt09•H^OJ03*s•HQ)•3O£

•OID 0)OJ i03 M^s OiH M-t(0 l-tE QJ< CU

l_tOJUsJJEJID

PC10£

JJBOJ

JJDCw

r*CO04

^<cJJO 10o a<s•3

* B EE «C QJW JJ(C co niH J >s&< < CO

CO

SEHg

BO-HJJ (Qe 01JJ -HB MQJ O€ JJLj tj

JJ Oto JDB (0tH t-3

0,U

COCO

^

£OJ03

COCOJj (0l 4J

*0**s, Q5•3

H EVO 10

CQ1-2

EH&3SBO-HJJ 03« QJJJ -H* LjQJ O

E *°4J Om XlB toIH i4

(X' UM

t)QJ CO> CO•H *H HOJ H

OOH

|CQ tQ

< CO\ (0S JJ

1Coo *30 EIT) Q

O*-"CO " >tJ Jp fy

EH w wg^

MQJfHb21E-H

f-lQJ

g£c?

u c*OJ COs> *»**•H H•H 1OJ COO COxO fS

£OJ4J

CQ 0)

< CQ

**** (0^ JJ

(0CO Qoo -30 Eto 10

o*-'CQ ^< w£•4 ?6Q KlE* N^

MOJ

wtESOJ

1foO

COCOw

£OJ

CO COiJ >H< CQv (03* 4

(0ofj *30 E

O5s

CO " ^< £3EH w wS J

OJ41C•H

QJCH

S£&<C

l>IE

i-J

OJU £1C QJSJJa

? P IHfe CO

ts QJ eiH JJ J

•H IDO J= OQJ a•O 10*3•* E> u a:

IB

IHE

EO•H.p aid QJ.P-Hf* Lj

0) oSS

tl,p 003 -QE (0M tJ

itfi<

COCON

ino\<jO

QJCO

03**•>

CTiX

^

E(0-K1-1B^

OmjOJB0)C

!'•P4K

'300930

o*>mk.oA

Page 116: ftR3008L6 - semspub.epa.gov

BO•r! -~*4-> 030 Q) QJ«-H 4J JJrH RJ fl(0 Q TJjj aa c>B ^w

C5•HK•H>QJa;\tHQJT3O£

•303 QJa EB CXOrH twra inC QJ< CM

M?IHDJJU10<M=3'E5£

JJCQJP3£4jjCDEIH

r-CO

01

ffloin

toK

MQJErHUtB•HXMQJCM

rlQJN>irHIDsrt.hMDUfHQ)£

t-co\OJ

oHnH

U•5O*

ffiCMEH

rlQJErHW1E•HXMQJfe

UOaCO

•30)MrdMHEM

r*CO-*•*04

£OJ4J

V^ E> tN

COoO 10CD JJ

1CU Q

S £CC CQEH M

U»H«EH „M£MKOt-3OU

BOU•rlB.CUCJ£H

IHOJM>HrH(0EK

. OJJS<

CDCO^H

HOS

V•rtEOtHJJUQJato

oMCCE-it:£IHftOAou

JHoccfBOJjrH•H£

rtQJJJOJ

1 EO 0JJ 4JU Oa s:a aCO^D

t-CO*03

Oa*en

OMCCEH

SIHKOiJOu

1ra0 M•rl CJS BCF1 »-IQJ SCQ EH

MQJJJ

1 0)0 Erl 0JJ JJU OQJ naaCO

fe•-J

r*-CO

04

JJ »B nrH JMte n03 Om ••oa i *CO fe nK O* »rH'v.ffi U* a ^O*3 n*f C rHCA 10 O

L>IHbMU(CfrlCO

-35W 0&M

Bo•HMO

rlOJg.XrHrOEtoBOH^Xa

r-cc>•01

Hf4

b-tEHt-HanCQCCt3EH

J<iDJJ-Hao:E

.uOJ.P0)E•H•PUSIH

t-CO"s*08

inen

ew>wEHCJSQ35sLJ

IHCO>

IHOJjj0)£><JJ•H>•HJJt>3*3EOU

r-CD•.in

a9U

•3OJ0OrHD

EH55W0

XCQ»rfJ

COEa>jjIHCO£iOJXIDEQJCM

4JC•in ©HUj=CO«rHb

C

totu

1*00931

Page 117: ftR3008L6 - semspub.epa.gov

COCDOvH

rt

IJJCJo

BOiH ——JJ 03(0 QJ QJrH JJ JJrH ra re(0 OT3JJ QJ03 &B — *IH

BO03•H^QJKrHQJ•3O£

•3tt OJQJ E03 CSOrH U-i10 r)B QJ< CM

rtQJrt

JJOC(HS3B

2

BOJESJJtoE

COCD*sOH

•HOOo *5

K CQQJ t-3E<O-H\Q S

ljj

1* Lj

O JSIH CJ

xOJEO•HD

UHi

r<O

•H JJ£ (0U MJJ &10 OJJJ JJ•H E5; M

j>•Ea(0iHtno

,

O

*)X

AR300932

Page 118: ftR3008L6 - semspub.epa.gov

ANALYTICAL LABORATORIES, INC.CHEMWEST SAMPLE PREPARATION ANALYTICAL

INSTRUMENTATION AND EQUIPMENT

October 1988

1 - Hewlett-Packard 5880 Gas Chromatograph with dual ElectronCapture Detectors (ECD) and dual Spl&t/Splitless Injectors

1 - Hewlett-Packard 5790 Gas Chromatograph Flame lonizationDetector (FID) and Split/Splitless Injector

3 - Top Loading Balances, 2 Scientific Products TL1600 and 1Mettler 3200

1 - Analytical Balance, Scientific Products SP180

1 - pH Meter, Scientific Products

2 - Roto Evaporators, Haake Buchler

2 - Steam Baths, 10 Place, Custom

2 - Sonic Dismembrator, Artek Model 300

32 - Soxhlet Extraction Apparatus

20 - Kudema-Danish Concentration Apparatus

1 - Muffle Furnace, Thermolyne 1400

3 - Ovens, VWR Scientific 1305U

1 - 40 Place Orbital Shaker, Labline

4 - Nitrogen Slowdown Apparatus

1 - Dry Bath, Thermolyne

1 - Centrifuge ... . . .. ~_ _..... . . . . . . . . . .

Various Sizes of the Following Laboratory Glassware:

- Separatory Funnels

- Filtering Flasks - r

- Erlenmeyer Flasks _ '

" BeakerB AR300933- Chromatography Columns

600W North Mortcei Boulevard * Sacramento, CA 95834 * Phone (916) 923-OS40 • FAX (P 16] 923-193B

A Subsidiary of CompuChem Corporation

Page 119: ftR3008L6 - semspub.epa.gov

ANALYTICAL LABORATORIES, INC.

CHEMWEST ADDITIONAL COMPUTER EQUIPMENT

October 1988

1 - Epson Equity III Computer Systems (IBM-AT Compatibles)

B - NEC APC IV Computer Systems (IBM-AT Compatibles)

2 - CompuClub Computer System (IBM-AT Compatible)

1 - IBM XT Computer1 - Panasonic Senior Partner (IBM XT Compatible)

2 - Epson GQ-3500 Laser Printers

6 - Daisy Wheel Letter Quality Printers

10 - Dot Matrix Printers

3 - 1200 Baud Modems

AR3Q093I*

600W North Morket Boulevard * Sacramento, CA 95634 • Phone (916) 923-0840 • FAX [916} 923-1938

A Subsidiary of CompuCnem Corporation

Page 120: ftR3008L6 - semspub.epa.gov

Chem West Analytical LaboratoriesAddendum F

Sample Data Report for a non-CLP Analysis

flR300935

Page 121: ftR3008L6 - semspub.epa.gov

CHEMWESTANALYTICAL LABORATORIES, INC.

December 15, 1988

CompuChem Laboratories, Inc. - — ".3308 Chapel- Hill/Nelson" Hwy." "Research Triangle Park, NC 27709

Attention: M s . Angela Childress . . . "

Subject: Report,of. Data - Case Number 2816

Dear -Ms. Childress:" " " : -- - - "-'--'-

The technical staff at CHEM WEST, is pleased to provide our report for theanalysis you requested: Total,Petroleum Hydrocarbons - EPA Method 418.1.

Six samples (5 sgdls, 1-water)'for-Project. CompuChem, Project Number 88-196were -received December 5, 1388 in good condition. Results of the analysisare presented on the following pages. -

Thank you for choosing CHEMWEST Laboratories. Should you have questionsconcerning this data .report 'or..the analytical... methods employed, please do nothesitate to. contact Toni Weeks, our Technical Service Representative, or yourproject .manager. ;~We hope that,.you will gons.ig_er._CHEHWEST Laboratories foryour future analytical.support and service: requirements.

Sincerely/ '""

,.->

Jill.B. Henes, Ph.D. . . ... and Dave JarvisVice President of Technical Services . .-.. ..: Project Manager

DJ:bw . . - . .

cc: Joel Bird, President ' . . . . .File . ' = :. V-:. " "" '•'""" H ——-- --_- ~ __ - - _ ; ;

AR300936

600W North Market Boulevard • Sacramento, CA 95834'... • Phone (916) 923-0840 • FAX (916) 923-1938

A Subsidiary of CompuChem Corporation '''

Page 122: ftR3008L6 - semspub.epa.gov

CHEMWEST ANALYTICAL-LABORATORIESTOTAL PETROLEUM HYDROCARBONS

Date Extracted:" 12/12/88 . Case : 2816Date Analyzed : 12/14/88 Matrix: Water

AmountClient CHEMWEST - -" " ' Detected

ID -ID (mg/L)

Method Blank 2816-6MB '"""-. ;-. BRL233488 2816-6 " BRL233489 2816-7" . BRL

The reporting limit for Total Petroleum Hydrocarbons is 1.0 mg/L

BRL: Below~~Reporting Lim.it.

AR300937

Approved by: ?K\ - ; _ :_ .... , ~; .,--. ::;" "-•=-— -REV2:1.88

Page 123: ftR3008L6 - semspub.epa.gov

CHEMWEST ANALYTICAL LABORATORIESTOTAL PETROLEUM HYDROCARBONS

Date Extracted: 12/12/glT \r Case : 2816Date Analyzed : 12/14/88 Matrix: Soil

--- - - AmountClient ".' CHEMWEST - - Detected

ID ID . (mg/Kg)

Method Blank 2816-1MB . . ••---''- . " . - BRL233470 2816-1.- - -- - -•————; . BRL233476 2816-2" ; ... ^ -/" ~ , 12D233477 " J8I6-3'.".: .-'.:__:—. --- 4500231479 • • • • - - - : 2816-4- -• --.. ,. ' . - ":-.,. 1100233480 __-.- - - .2816-5! : - - - - - . - . . • 1 2 0

The reporting limit for Total Petroleum Hydrocarbons is 25 mg/Kg

BRL: Below Reporting Limit.. , .

AR300938

Approved by: Jrg . -- , - :..•.---.- •-• REV2:1.8"8

Page 124: ftR3008L6 - semspub.epa.gov

CHEMWEST ANALYTICAL LABORATORIESTPH by IR -

Date Extracted : 12/13/88 Case : 2832Date(s) Analyzed: 12/14/88 Matrix: Soil

AmountClient CHEMWEST DetectedID ID <mg/Kg)

Method Blank 2831-MB - BRL

The reporting limit for TPH by IR is 25 mg/Kg.

BRL: Below Reporting Limit.

SR3Q0339

Approved by: A"5 :_-, - - REV2:1.88

Page 125: ftR3008L6 - semspub.epa.gov

CHEMWEST ANALYTICAL LABORATORIESTPH by IR

Date Extracted : 12/13/88 Case : 2832pate(s) Analyzed: 12/14/88 Matrix: Soil

Spike AmountClient CHEMWEST cone. Detected %ID ID (mg/Kg} (mg/Kg) Recovery

233827/07BS31100301D 2831-10MS 146 120 82%

BRL: Below Reporting Limit.

3R3C03UO

Approved b y : A W . - - - - - " " " " . - """" " " R£V2:1.88

it-i'

Page 126: ftR3008L6 - semspub.epa.gov

CHEMWEST ANALYTICAL LABORATORIESTPH by IR

Date Extracted : 12/13/88 Case : 2832Date(s) Analyzed: 12/14/88 Matrix: Soil

Spike AmountClient - CHEMWEST cone. Detected %

ID ID (ing/Kg) (mg/Kg) Recovery

233827/07BS3-1100301D 2831-10MSD 147 131 89%

BRL: Below Reporting Limit.

fiR3QQ9UJ

Approved by: A^ - REV2:1.88

Page 127: ftR3008L6 - semspub.epa.gov

CHEMWEST ANALYTICAL LABORATORIESTPH by IR

Matrix-Spike - Matrix Spike DuplicateRelative Percent Difference

«Recovery Recovery

Compound 2832MS 2832MSD RPD

TPH 82% 89% 8%

Approved by: ,«^ - , .., . - ,- ,- -., ,, ppvb.i RR•——•—— - - - - - -- ----- - - - —-•-• -- - --- - - Kt> V i I I , 0 0

Page 128: ftR3008L6 - semspub.epa.gov

Regression Output:Constant - - _. - 0Std Err of Y.0.684917

- - 5-S.quared 0.998718No. of Observ 6Degrees .of Fr _ 5

X Coef 1.5942Std Er0.0l91 " . : .

w x L y y1absorb cone... conc.reading ;. .. .; -- -

0 0 01.4 2.5 ERR

~ 5 4.782809. I 10 . 9.565618JM 25 23.75461Jl.8 -50 50.69777

Samples . -• . Vi""J:."___Lr -,""," T" "" " ^: ""•"•' " '-"""" - . - :,Absorb .-,cw"| Cone f^P! : Final -Dilution ConeReading ' ' " C°nC ' ^^t Volume Factor in Sample

« _0 2816-6MB . .,... ,0 .; - ! 23.90 2R1 fi-fi . . . , . , . - , ,» . .. 1000 100 . ^» -£ 0 1 b - b - (7 -i « n rt ,„„ -1- 0.0e-> r> •>-,-—, - ±Bt5i3 100 T

2816-~7 ' : 0 Tflciw T"^ ! 0-015 25CC „- 23.9140!- ^ " J 0-00 2816-1 a " ?„ 100 1 23 9

13.1 2816-2 - 20788493 ' 17 \l - ±™ * e'.e12.3 2816-5 19.609513 -^.ll ^ \ "8-314.8 2 5 C C _ _ _ _ 23. 59519 ill {H } "4.7

5 2816-3,:: -- ' -' 77971348 J_ 7 fi7 i^ 1 23 • 61 . 1 - -2816-4 - 1.753696 " ' - ' 1 0 B 4511'2

15.2 25.CC .-5

Page 129: ftR3008L6 - semspub.epa.gov

CHEM WEST ANALYTICAL LABORATOR4NC600 West Nonh Market Blvd. -Sacramento, California 95834 ' -(916) 923-0840 FAX (916) 923-1938

1

2816Order No.DateCompl,J£ate.Section

rir i j n r . > PQ NO.v '- ^nn^HTVlJMt.^hori.

Phone

U A • -HU

4R3009W—-

SAMPLE WILL BE HELD 30 DAYS UNLESS LONGER TIME IS ARRANGED

Page 130: ftR3008L6 - semspub.epa.gov
Page 131: ftR3008L6 - semspub.epa.gov

•u-omzD

Page 132: ftR3008L6 - semspub.epa.gov

APPENDIX GNortheastern Analytical Corporation

Quality Assurance/Quality Control Plan

Page 133: ftR3008L6 - semspub.epa.gov

Northeastern Analytics! Corp.

QUALITY ASSURANCE/QUALITY CONTROL PLAN

STANDARD OPERATING PROCEDURES

Analysis; Sampling and Testing_ for the Environments! and Safety Professionalcorporate Ce^ie^STEestTSw)* Roac WB-iti Tfaev/Je e). 03053 [609) SBIa-BOOO FAX ("503) 985-B7DD

Page 134: ftR3008L6 - semspub.epa.gov

QA/QC STANDARD OPERATING PROCEDURES

Table of Contents

1.0 - Introduction and Objectives

2.0 Laboratory Responsibility and Organization

3.0 Quality Assurance Objectives

4.0 Sampling Procedures/Chain of Custody

5.0 Calibration Procedures

6.0 Analytical Procedures ..- . ._ .

7.0 Internal Quality Assurance Practices

8.0 . Data Reduction, Validation and Reporting

9.0 Performance Audits "

10.0 Procedures Used to Access Data Precision & Accuracy

11.0 Corrective Action

12-i 0 Analytical Instrumentation

13.0 Key Personnel Resumes . .

( f

Page 135: ftR3008L6 - semspub.epa.gov

1-0 INTRODUCTION"AND OBJECTIVES

1.1 Introduction ___....__._,..__.,.

NAC is a multidisciplined environmental testing organization.The main area of NAC's services are environmental analyses andasbestos testing. The environmental analyses are performedby NAC's Laboratory Division. NAC's management is committedto providing excellence in all of its' services. TheLaboratory Division produces data which can be relied upon byNAC's clientele. Management support is evidence through theemployment. of high caliber and experienced chemists andtechnicians, the construction of new laboratory facilities andthe acquisition .of modern instrumentation. The LaboratoryDivision traces its history back to 1971. Over seventeen (17)years of growth has resulted in a full range of environmentalanalytical capabilities offered today.

NAC is located in Marlton, New Jersey in a modern custom built27,000 square foot facility. The Laboratory Division providesanalytical support to clients from industry, consultinggroups, commercial and government. The Division specializesin the trace or -microanalysis of environmental contaminantsin variety of matrices. Much of the Divisions work is insupport of regulatory compliance, environmental investigationand the development of design data for environmental control.

The technical requirements and legal impact dictates strictadherence to approved and regulated procedures. The latestin analytical.instrumentation is incorporated into proceduresof "analysis. Personnel are trained in all phases of theanalytical scheme with emphasis on quality control and qualityassurance.techniques.

The Laboratory Division has its' foremost goal to provide highquality analytical data. Every phase of the Divisionoperation is monitored toward producing laboratory resultswhich have a high confidence level and are fully supportable.The Division is oriented to the needs of the clients to thedegree that, analytical data meeting the high standards can bemet. The Laboratory Division will not undertake analyticalprojects which will result in data failing to qualify understrict QA/QC standards.

NAC recognizes the importance of the reputation of theLaboratory Division and believes that continued success willcome from uncompromising quality goals. Over the past several-years NAC's management has demonstrated its commitments to theLaboratory Division and high quality analytical servicesthrough expenditures of millions of dollars and the attractionof experienced and knowledgeable personnel. This/fcfifflDpAtygni.-is built into NAC's future plans through planned expfen<nTiirfi?st?Utraining programs and the recruitment of specializedlaboratory and administrative personnel. .,

aR300950

Page 136: ftR3008L6 - semspub.epa.gov

1.2 Objectives and Policies _ . . . _ . .

The Quality Assurance/Quality Control QA/QC program of NAC'sLaboratory Division is designed to ensure the laboratory'sproduction of environmental analysis data in full compliancewith all regulatory procedures. The QA/QC program keeps allof NAC's procedures, SOPs and laboratory practices well withinthe guidelines and limits set by federal etnd state regulationsregarding accepted methods of analysis. The Program addressesthe Laboratory Division's organizational structure, thephysical facilities, the training and experience level oftechnical staff, operating procedures and reportingprocedures.

Operating procedures are detailed in the Laboratory AnalyticalMethods Standard Operating Procedures and Field SamplingStandard Operating Procedures. All of the Standard OperatingProcedures are readily available for examination by NAC'stechnical personnel. The proprietary nature of much of theinformation contained in NAC's QA/QC Program and SOPs requirespermission from the laboratory Division's Manager before theyare made available to non-laboratory personnel.

The SOPs form an integral part of the QA/QC Program.Procedures are defined for all aspects of sample collectionincluding preservation, decontamination of equipment, safetypractices and documentation. Sample holding times are definedfor parameters! of interest". .Procedures are spelled out indetail to facilitate analyses according to prescribed methods.Continuous control over samples is exercised from collectionto analysis, to reporting and finally retainage or disposal.This control is defined throughout Chain-of-Custodyprocedures. Analytical performance is monitored through theuse of'quality control techniques such as duplicate analysis,analysis of spiked samples, analysis of standards. The levelof QC samples exceeds all analytical guidelines published byregulatory authorities. _

NAC's QA/QC Program is utilized daily by the LaboratoryDivision's technical staff. The prograia including SOPs areupgraded _ continuously as new techniques and procedures areadopted by federal and state agencies.

Page 137: ftR3008L6 - semspub.epa.gov

2.0 LABORATORY RESPONSIBILITY AND ORGANIZATION

The general responsibilities of key personnel with respect toQA/QC are as follows: (See Figure 2-1)

2.1 Laboratory Manager

The Laboratory Manager's duties and responsibilitiesinclude the following:

Directs the laboratory's analytical programs.Executes laboratory administrative functions.Ensures " compliance with appropriate analyticalmethod and instrument performance specifications.

- Performs final review of test reports.

The Laboratory Manager reports to the Vice President,Operations.

2-2 Assistant Laboratory Manager

The. Assistant Laboratory Manger's duties andresponsibilities include the following:

Established and directs all activities relating toanalytical- QA/QC.

- Coordinates projects with .Marketing and scheduleswork, thereby not allowing work schedule overloads

- which would:affect quality.

The Assistant Laboratory Manager reports to theLaboratory Manager. . ,.

2.3 Laboratory Supervisors

The Laboratory Supervisors duties and responsibilitiesinclude the following:

. .:. ._..•=_——Ensure --proper-, application and performance ofspecified quality assurance practices.

- Perform first line review of the results of qualityassurance samples.

- Responsible for maintenance of quality assurancedocumentation.

All Laboratory Supervisors report to the AssistantLaboratory Manager.

AR30Q952

Page 138: ftR3008L6 - semspub.epa.gov

2.4 Sample Custodian

The Sample Custodian's duties and responsibilitiesinclude the following:

Ensures that all submitted samples are properlyaccepted into the laboratory in accordance withdocumented sample acceptance procedures.Ensures that associated sample acceptance data isentered into the laboratory sample tracking systems.Arranges proper secure sample storage.

The Sample Custodian reports to the Assistant LaboratoryManager. :

2.5 Analysts ._._.. ......: . _ _ _ _ . _ _._.. = ___..,...

Analyst's duties and responsibilities include thefollowing:

Equipment.Maintenance.- Equipment Calibration (See Section 5.0).

Sample preparation and analysis.Raw data manipulation and reporting.Inclusion of appropriate QC samples andconsiderations into all laboratory operations.

Analysts report to their respective LaboratorySupervisors. _ .

2-6 Administrative Technical Review ,

The administrative section duties and responsibilitiesinclude the following:

Initial assembly of test reports.Second review of quality assurance data.

M300953

Page 139: ftR3008L6 - semspub.epa.gov

-

CQco*H4J«Jt-t -P0) r-la a>O XI

si0)« --H Un0) CrH (QCU H

0}O-H>

rH0)CP(C r-(C 0)m toE in-H

£o c•P 4=(0 O*H raoXI3

.. . ,,=-__ _

^

O

rH 0)O r-i JJ>Q Q) CSO^*H

fd tc4-> (0CS* ^^(0 34J (C10 CU-rHto10

'

-

cMI

stration/

•HC-H

TJ

3 Nfl OH j- i

> 0)

>i U3-H

r-t>1 Q)kr^4J4J O CQ) 01 -H -—.•H -H (C &'e > cu c0) M -H£ ftJ r-l 4>U &3 O3 (0 <

4J W fc ~OS

>iao0 rHm o>o >fc tH (0-POO)U W rHQ)-H Ua>W rH T3

O rHo a fcc en uIB -H

17 KocH

>iaoU >iO JH rHO O rHrH 10 A3•P-H W0 >Q> iH TJa O rHco a roS J3a CA u

ffl -H(0 K£

>iX3a10rH 10tT* rJ IHO O 0)-P W XI«3 •*•( Ee > *cD L .e

<M(4H«JJCO

(U<M(04-JCO

IH«Hra4Jw

(0JJ

W

OH

o

Page 140: ftR3008L6 - semspub.epa.gov

Table 3-1

ACCURACY AND PRECISION OBJECTIVESFOR MEASUREMENT DATA

VOLATILES

Parameter Accuracy % Precision %GC GC/MS GC GC/MS

Acrolein 88-118 HA 20 NAAcrylonitrile ..._ 71-135 _ NA 20 NABenzene ' 39-150 37-151 21 26BromocUchloromethane 42-172 35-155 23 21Bromoform 13-159 45-169 21 18Bromomethane D-144 D-242 36 58Carbon Tetrachloride 43-143 70-140 20 11Chlorobenzene 38-150 37-160 17 28Chloroetharte 46-137 14-230 17 312-Chloroethyl vinyl ether 14-186 D-305 35 84Chloroform 49-133 51-138 19 18Chloromethane D-193 D-273 52 58Dibromochloromethane 24-191 53-149 24 17Dichlorobromomethane 42-172 35-l!55 20 211,1-Dichloroethane 47-132 59-1J55 14 161,1-Dichloroethene 28-167 D-234 29 431,2-Dichloroethane 51-147 49-1355 15 21t-l,2-Dichloroethene 38-155 54-156 17 191,2-Dichloropropane 44-156 D-210 23 45c-l,3-Dichloropropene 22-178 D-227 32 52t-l,3-Dichloropropene 22-178 17-183 32 34Ethylbenzene 32-160 37-162 26 24Methylene Chloride 25-162 D-221 21 361,1,2,2-Tetrachloroethane 8-184 46-157 26 20Tetrachloroethene 26-162 64-148 20 16Toluene 46-148 47-150 18 201,1,1-Trichloroethane 41-138 52 .62 20 211,1,2-Trichloroethane 39-136 52-150 20 18Trichloroethene , 35-146 71-157 23 13Trichlorofluoromethane 21-156 17-181 27 34Vinyl Chloride 28-163 D-251 27 651,2-Dichlorobenzene 37-154 NA 22 NA1,3-Dichlorobenzene 50-141 NA 19 NA1,4-DichlorobenKene 42-143 NA 20 NAM: Not analyzed by this wthod

Page 141: ftR3008L6 - semspub.epa.gov

Table 3-1ACCURACY AND PRECISION OBJECTIVESFOR MEASUREMENT DATA fContinued!

PBSTXCXDBS/PCB's

• Parameter Accuracy % Precision %

Aldrin 42-122 20a-BHC 37-134 23b-BHC 17-147 33g-BHC (Lindane) ' 32-127 25d-BHC 19-1*° 22Chlordane 45-119 194,4'-DDD 31-141 264,4'-DDE _ _ _ _ _ _ _ __36-145 2 84,4'-DDT 28-160 31Dieldrin 36-146 16Endosulfan I 45-153 18Endosulfan II D-202 45Endosulfan Sulfate 26-144 24Endrin 30-147 24Heptachlor 34-111 16Heptachlor Epoxide , 37-142 25Methoxychlor __ 5_6_ 137 23Toxaphene 41-126 2 0PCB-1016 50-114 15PCB-1221 - 15-178 35PCB-1232 10-215 31PCB-1242 39-150 21PCB-1248 38-158 25PCB-1254 29-131 17PCB-1260 8-127 35

flR'300957

Page 142: ftR3008L6 - semspub.epa.gov

Table 3-1

ACCURACY AND PRECISION OBJECTIVESFOR MEASUREMENT DATA (Continued)

BASE/NEUTRAL EXTRACTABLE ORGANICS

Parameter Accuracy % Precision %GC/MS GC/MS

Acenaphthene 47-147 21Acenaphthylene . 33-145 26Anthracene 27-133 27Benzo(a)anthracene 33-143 26Benzo(b)fluoranthene 24-159 30Benzo(k)fluoranthene 11-162 35Benzo(ghi)perylene D-219 51Benzo(a)pyrene 17-163 33Bis(2-chloroethoxy)nethane 33-184 28Bis(2-chloroethyl)ether 12-158 35Bis(2-chloroisopropyl)ether 36-166 26Bis(2-ethylhexyl)phthalate 8-158 364-Bromophenyl phenyl ether 53-127 17Butyl benzyl phthalate D-152 542-Chloronaphthalene 60-118 134-Chlorophenyl phenyl ether 25-158 30Chrysene 17-168 33Dibenzc(a,h)anthracene D-227 59Di-n-butyl phthalate 1-118 391,2-Dichlorobenzene 32-129 241,3-Dichlorobenzene D-172 411,4-Dichlorobenzene 20-124 293,3'-Dichlorobenzene D-262 50Diethyl phthalate D-114 52Dimethyl phthalate D-112 1042,4-Dinitrotoluene 39-139 222,6-Dinitrotoluene 50-158 19Di-n-octyl phthalate 4-146 38Fluoranthene ^ 26-137 28Fluorene 59-121 13Hexachlorobenzene D-152 42Hexachlorobutadiene 24-116 26Hexachloroethane 40-113 18Indeno(l,2,3cd)pyrene D-171 50Isophorone 21-196 33Naphthalene 21-133 29

AR30Q958

Page 143: ftR3008L6 - semspub.epa.gov

Table 3-1

ACCURACY AND PRECISION OBJECTIVESFOR MEASUREMENT DATA (Continued)

BASE/NEUTRAL EXTRACTABLE ORGANICS (Continued)

Parameter Accuracy % PrecisionGC/MS GC/MS

Nitrobenzene 35-180 27N-Nitroso-di-n-propylamine D-230 44Phenanthrene 54-120 15Pyrene ' 52-115 151,2,4-Trichlorobenzene 44-142 21

ACID EZTRACTABLE ORGANICS

Parameter Accuracy % Precision %GC/MS GC/MS

4-Chloro-3-methylphenol 22-147 302-Chlorophenol 23-134 292,4-Dichlophenol 39-135 222,4-Dimethylphenol 32-119 232,4-Dinitrophenol D-191 682-Methyl-4,6-dinitrophenol D-181 492-Nitrophenol 29-182 294-Nitrophenol D-132 47Pentachlorophenol 14-176 34Phenol 5-112 352,4,6-Trichlorophenol 37-144 23

Page 144: ftR3008L6 - semspub.epa.gov

Table 3-1

ACCURACY AND PRECISION OBJECTIVESFOR MEASUREMENT DATA (Continued*

METALS

Parameter Accuracy % Precision %

Aluminum 58-123 12Antimony 43-147 18Arsenic 68-127 10Barium 54-123 ' 13Beryllium ' -.. . _. 62-122 11Cadmium _ 80-108 5.0Chromium 61-102 8.2Copper 81-109 4.9Iron 31-166 23Lead 80-154 11Manganese 60-122 11Mercury 56-125 13Nickel 82-114 5.4Selenium 49-136 16Silver , ...__. 90-102 2.1Thallium 45-130 16Zinc , 76-127 8.4

AR300960

Page 145: ftR3008L6 - semspub.epa.gov

Table 3-1

ACCURACY AND PRECISION OBJECTIVESFOR MEASUREMENT DATA (Continued)

INORGANIC PARAMETERS

Parameter Accuracy % Precision %

Ammonia Nitrogen 69-113 8.2Biochemical Oxygen Demand 35-103 16Chloride 102-115 2.0Chemical Oxygen Demand 70-151 12Cyanide 80-113 5.7Fluoride - 91-117 4.2Hardness 84-115 5.2Nitrate Nitrogen 71-139 11Oil and Grease 78-104 4.8Petroleum Hydrocarbon 31-113 19Phenols 79-112 5.8Phospho rus 66-132 11Sulfate 65-136 12Total Organic Carbon 93-111 2.9Total Organic Halogen 65-117 9.5Hydrogen Sulfide 78-112 6.0Total Kjeldahl Nitrogen - 71-135 10

Note 1: Any analysis which is not performed by KAC shall beperformed by a certified subcontreictor.

AR30096

Page 146: ftR3008L6 - semspub.epa.gov

4.0 SAMPLING .PROCEDURES AND CHAIN OF CUSTODY

4.1 Sample Collection Procedures

KAC will .utilize approved sample collection methodsand equipment outlined in the following technicalpublications.

^ EPA/SW-846, "Test Methods for Evaluation Solid WastePhysical/Chemical Methods," Third Edition, USEPA,1986. - . . _ - . . _

NJDEP Field Sampling Procedures Manual, July 19B6.

EPA/Sw-611, "Procedures Manual for GroundwaterMonitoring at Solid Waste Disposal Facilities1'.

EPA 600/4-82-029, "Handbook for Sampling and SamplePreservation of Water and Wastewater, "September19S2. _ - ...

Specific: guidelines for a project for sample siteselection, selection of sampling equipment, typesof" "samples to be collected, standard samplecollection procedures, specific maintenance andcalibration procedures for sampling equipment, andother considerations are based upon site-specificrequirements. ..__ _ _ .... . :

4.1,1 Selection and Preparation of Sampling Equipment

The material of which sampling equipment isconstructed can affect analytical results. Thematerial selected for sampling certain parametersmust not contaminate or alter the sample beingcollected, and must be easily cleaned so thatsamples are not cross-contaminated. Also, theaction of the sampling device must not alter thesample bey collected (e.g., using an air-lift pumpfor volatiles in water may purge the compounds ofinterest).

4.1.2 Sample Containers, Preservatives and Holding Times

KAC recognizes that proper containers andappropriate preservatives are necessary for thecollection of valid ..samples. In addition, thesample must ~~ be analyzed within a prescribed timeframe for each parameter. The Laboratory SamplePreservation Summary details permissible samplecontainers, preservatives, holding times and thevolume of sample needed. The requirements ,pg fflf-ft n c otable correspond with the guidelines w.n vJ U u Iry udocuments listed in section 4.1.

Page 147: ftR3008L6 - semspub.epa.gov

4.1.3 Sample Documentation

KAC utilizes proper sample documentation measures torecord pertinent field data and ensure the legal validityof" all collected samples. These sample documentationmeasures will provide a detailed, legal record of allsampling activities including: sample collection,preservation, chain-of-custodypossession, transportationand laboratory submittal. Key component sampledocumentation measures include utilization of thefollowing record materials: Laboratory Field SampleLabels, Sample Seals, Chain-of-Custody Forms andLaboratory Sample Log.

4.1,4 Sample Custody . . . . . . .

Evidentiary documentation procedures continue with thelaboratory. The designated Laboratory Sample Custodianwill "receive and document all sample subinittals into thelaboratory. The Laboratory Sample Custodian will examinethe condition, preservation, and accompanyingdocumentation of all submitted sample prior to approvaland formal acceptance into the laboratory. Any sample,preservation, or documentation discrepancies (i.e. brokensample container, improper preservations, inadequatesample volume, poor documentation, etc.) will beresolved before the. .sample is approved and actually

*. accepted for~analyses. All required acceptance data willthen "be feco:rded an documented in.temperature/preservation checklist (Figure 4-2), labelledand placed in the secure sample storage area for

i. distribution to the appropriate analyst(s).

4.1.5 Standard Operating Procedures ..„...,_'(

The following pages are standard operating procedures for- groundwater sampling, trip and field blanks, holdingtimes, chain-of-custody", and sample "log in.

AR300963

Page 148: ftR3008L6 - semspub.epa.gov

METHOD: Ground Water Sampling Procedure

METHOD NO.: 013

"REFERENCE: NJDEP Field Sampling Procedure Manual,July 1986 Section XVIII

DATE: February 1987Revised June, 1988

1.0 Scop.e_.and Application ----.- —

This. .procedure describes the method for obtaining ground.water _ samples for organic and inorganic parameters. It isapplicable to wells such as found in fresh water supplyfields and monitoring wells in landfills, etc.

2.0 Summary of Method ___ _ ... _. . ,: . „ .__ _ ._.._,,

Prior -to sampling, the well -is jpurged to evacuate thestagnant - water. : Following the purging, samples arecollected" using bailers and transferred to approved samplingcontainers. " Every effort "is made to maintain sampleintegrity and prevent cross contamination via the use ofapproved materials and decontamination procedures. Trip andfield blanks are. .employed to -identify sources of potentialcontamination.

| 3 • 0 Apparatus and""Material's .

3,. 1 Field notebook

i 3,.2 Chain of Custody record\ 'i _ . . . . . . .. _

3.3 Keys for locked wells

3.4 Water level indicator

3 .5 "calculator"" :

3- 6 Submersible .pump, 1-12 gal/min capacity, with footvalve . . .

Page 149: ftR3008L6 - semspub.epa.gov

Method: Ground Water Sampling ProcedurePage 2-of 7

3.0 Apparatus and Materials (continued) ...

3., 7 Polypropylene cord

Note: Material Preference for well sampling:

1. Teflon ; 6. PVC2. Stain St. 316 .7. "Viton3. Stain St. 304 8. Conventional Polyethylene4. Polypropylene5. Polyethylene - _-

3.8 Tygon tubing or polythylene hose

3.9. Portable generator, suitable to power submersiblepump. . . . ' . : .

3.10- Stainless- steel.well bailer with Teflon ball valve and.stainless industrial steel cable or Teflon coatedcord. . . . . - . - ^ -

3.11- -Sample bottles, various

3.12 Filter apparatus

3".13 Vacuum pump

3..14 _.__0_.45 micron filters .. . .

3.15 Coolers, as required . -.— -

3.16 Ice br'~ice packs - - - - - _ - _ _

3.17 Washing brushes, various

3.18 DI water" dispensers -——

3.19 Sprayers (2)

3.20 Large plastic wash tub (3)

3.21 Poly sheet , AR300965

Page 150: ftR3008L6 - semspub.epa.gov

Method. Ground Water Sampling ProcedurePage 3 of 7 _ .....:....... ...... ....__..

3.0 Apparatus and Materials (continued)

3.22 pH meter and probe ...

3.23 Thermometers, Celsius

3 ,,24 Specific conductivity meter =.

3.25 Surgical gloves

3.26 Extension cords ... _ - .._....

4.0 Reagents ____„_______;..._......_._;.-.. . - . . - _ - • _ .. . . .

4.1 pH buffers, 4.0, 7.0, 10.0

- 4.2 DI—water, appropriate" volume ~'

4.3 Acetone, pesticide grade, appropriate volume

4.4 Non-phosphate detergent, appropriate volume

4.5 "Tap water, appropriate volume

5'° Interf erenc'es '" .'.'~_ ^ ;.;;. -..T" . - - - - _ — - - - ; ; . - . ;

Laboratory analysis/- regardless of how sophisticated, isonly as good as the quality of the sample. There, is nosubstitute .for paying close attention to factors that mayinfluence the quality of the sample. The proper apparatus,cleaned thoroughly and used carefully is the best means foracquiring quality laboratory data.

flR300966

Page 151: ftR3008L6 - semspub.epa.gov

Method: Ground Water Sampling ProcedurePage 4 of 7 ; ------- _ ----- —.-- .". :

•6.0 Procedure _; _ ^ _:^

6.1 Begin the sampling by verifying the well identity andrecording it accurately in the field notebook andinitiating the Chain of Custody (COC) record (seeprocedure NAC009).

6.2 Before unlocking and removing the well cap, clean anydebris off that may enter the well.

NOTE: • ANY EQUIPMENT THAT ENTERS THE WELL MUSTBE SUBJECTED TO DECONTAMINATIONPROCEDURES -PRIOR TO ITS USE, USINGAPPROVED METHODOLOGIES.

6,3 Using the water level indicator, determine the totaldepth of the . well to the top of the casing (TOC) .

— -. Record the depth in feet to two decimal places.

6.4 Determine the depth of -the..water by measuring from thetop of^ the casing to the water level. (DOW) . Recordthe depth in feet to two decimal places. •

6.5 --"Calculate the linear feet of water in the well by theheight of"the water column.

6.6 - Determine the number .of gallons of water in the wellusing the following table:

Casing Diameter .,. Gallons of water per linear feet

2" 0.164" 0.656" 1.47S" . . . 2.61

AR300967

Page 152: ftR3008L6 - semspub.epa.gov

Method: Ground Water Sampling ProcedurePage 5 of 7 _" ~~"".."_". ." =;":__"."____. . "" .".

6.6 (continued)

—-Calculate as follows:

TOC ft. - DOW ft. x Gallons/ft. = Gallons of water inwell .... : : ..; ;"= ~:... :Alternatively:

3.14 x R£- x h = Gallons of water in well231 .... . - ... .

where R = radius of well in inchesh = depth of water column in inches

Record all calculations in the field notebook.

6.7 "Determine the total number of gallons to be purged asfollows: . .

Purge volume -gallons = number of gallons of waterin well x 3

This- number represents the minimum amount of gallonsof-water to be purged from the well to represent threewell . volumes. " Three to- six well volumes, isrecommended. Record the minimum purge volume.

6.8 Lower the submersible pump into the well, using newpolypropylene rope and polyethylene hose. Locate theportable generator not less than 10 feet from the wellland begin purging. Measure the purge rate using the2 gallon bucket and a stopwatch. Determine the purge,time as follows: ...... : : .

Purge time, minutes = purge volume, galspurge rate, gal/min

Record the purge time, -needed for a minimum of 3volumes. .Record the start and stop time for the purgeto ensure the minimum purge volume is attained.

AR300968

Page 153: ftR3008L6 - semspub.epa.gov

Method: Ground Water Sampling ProcedurePage .6 of 7" "" " " : •-•--• --- -- ~-

6.9 "Locate the water inlet less 'than 6 ft. below thedynamic water level during purging. If the wellrecharges slowly, evacuate the well to near dryness

- - __-__-_. ___—-—and allow to recharge. Record the intermittent purgevolumes to ensure that the minimum purge volume isattained. -

i6.10 After the minimum purge volume has been pumped from

the well, remove the submersible pump and lay on theI poly sheet prior to decontamination. Cut off the rope

and hose.and store for disposal..

I 6.11 : Lower the appropriate .size bailer into the well,S-- .-____-.._=.=.._ minimizing the contact with the side walls and with as

little agitation of the water as possible. Fill the| required bottles and recap the well. If volatile? . vials are to filled, fill these first. Take the

sample not more than two hours .after purging. Ifdissolved metals are to be analyzed, filter the samplethrough a 0.45 micron filter prior to preservation.

*»- - - - -PRESERVE IMMEDIATELY AFTER FILTRATION IS COMPLETE.

t ' 6.12- Prior .to entry into the next well, the bailer,i,...-. . . . . . . . submersible pump, cable, electrical wire, etc., must

be -decontaminated. Do not reiuse the pump hose orrope. Between two well samplings, a field blankshould be prepared as described in NAC Procedure 007.

7.0 "Quality Assurance ' . . .

7.1 :Use only the equipment listed in section 3.0 to enterthe well.- Never take a sample using the submersiblepump. Do not reuse the rope or tubing.

7.2 Decontaminate the equipment prior to use and betweeneach well. Equipment decontaminated in the laboratoryis to be wrapped in aluminum foil, cleaned andautoclaved for transport.

flR30G969

Page 154: ftR3008L6 - semspub.epa.gov

Method; Ground Water Sampling ProcedurePage 7 of .7 ; ;

7., 3 Use the following decontamination steps on thebailers:

7.3.1 Scrub the equipment in a washtub using thenon-phosphate detergent solution andappropriate brushes to get inside the

— --- bailers.

7.3.2 Rinse thoroughly with tap water in a tub .

7.3.3 Rinse in a 10% HNO3 solution in a tube ifsample, is to be analyzed for metals.

7.3.4 Rinse: with DI water. _ .

7.3.5 Rinse with acetone if sample is to beanalyzed for semi-volatile organics. Do notuse if VOA are to be conducted.

,7.3.6 Rinse again with DI water.

7., 4 For the submersible-- pump, clean as above', making surethe scrub step gets into the small crevices, etc.Rinse _-the inside of the pump by flushing with potablewater..— . _ . ._ :: -— _ .____.-..:____ ._-__._ r. .

7.5 Where required, collect field and/or trip blanks usingprocedures described in NAC_ Procedure 007.

8.0 Safety

8.1 The safety of the personnel conducting the sampling isparamount. Most well samplings do not involve acutelyhazardous .conditions such as explosives or flammableconditions. However," the chemical composition of thesample is frequently unknown and must be assumed to beharmful. Contact with skin, eyes and breathing vaporsis to be avoided. In preparation of a sampling event,consult your supervisor - on the level of protectionrequired. . . - . ..

flR30Q970

Page 155: ftR3008L6 - semspub.epa.gov

Method;. Trip and Field Blanks

Method Number: 007

Reference: rNJDEPJField Sampling Procedures Manual, July, 1986

Date: December, 1986

1.0 Scope and Application . .__„_ .. ..... ,... .

1.1 To assess sampling device decontamination proceduresand the overall sample handling process, field and tripblanks are employed in. .certain instances. This methoddescribes how to prepare aqueous trip and field blanks"for a variety of parameters.

2.0 Summary of Methods ..,. __- _ . . _ _ , . . - _ . . . / ,,-. , _

2.1 Trip Blank -.a set of sample containers are filled withlaboratory pure water and travel with the samplecollector." 'They are analyzed in the same manner as the

— --=—— -samples."

2.2 Field Blank - an empty set of ssample containers arecarried with the sampler. In the field, laboratorypure water is passed through the s;ampling equipment andcaptured in _the_ empty set of containers. They areanalyzed in the,same manner as the samples.

3.0 Apparatus and Materials _ _..-.-. .... ..._._;_. ... _

3.1 Decontaminated Bottles and Field Equipment

3.1.1 500 ml amber bottles, Teflon lined caps

3.1.2 950 ml amber bottles, Teflon lined caps

3.1.3 -40 ml screw cap vials, Teflon lined septa

3.1.4 500 ml nalgene bottles

3.1.5 Field sampling equipment, as required

3.2 Glass or Nalgene carboy, 10 liters or more

3.3 pH paper _: . : ..,-.,- ../ ", ~ ~,- ;

AR30097I

Page 156: ftR3008L6 - semspub.epa.gov

Trip and Field BlanksPage 2 of 3

4.0 Reagents

4.1 Laboratory pure water, demonstrated to be free of theanalytes of interest

4.2 -Preservative reagents

4.2.1 HNO3, reagent grade, 6N4.2.2 H2 SO4 , reagent grade, 6N4.7.3 NaOH, reagent grade, 6N

5.0 .Procedure

5.1 Obtain a list of the analyses for the samples to betaken. See your supervisor. - -

5.2 Use the. following list to determine what and how manysample bottles to fill for each set of trip blanks forthe parameters. :

Parameter. .. -Quantity . Type _ . Preservative

VOA 3 40 ml vials Cool, 4°CSemi-volatiles - 2 -- -950 ml amber Cool, 4°CPesticide, PCBs 1 950 ml amber Cool, 4°CMetals l .500 ml nalgene HNO3, pH<2Cyanide 1 .... :,_.- -500 ml nalgene NaOH, pH>12Phenolics ....: l 500 ml amber " H2 SO4 pH<2Wet Chemistry * * - - - - *

* Consult ""your supervisor for the quantity, type ofcontainer, and apropriate-preservative.

5.3 Prepare 2 sets of bottles, one labeled "TripBlank", the other "Field Blank".

5.4 Place the set of bottles labeled "Field Blank" inthe coolers.

5.5 Using laboratory pure water, fill the set ofbottles labeled "Trip Blank" in the same manneryou would an aqueous.;sample. Be sure to check theVPA. vials for_air bubbles.__ Add. the preservativesand check .with pH paper. Place in th&r&&Q3.z$+.

/ £

Page 157: ftR3008L6 - semspub.epa.gov

Trip and Field BlanksPage. 3 of :.3

5;6 In the field, pass the laboratory pure water over— — -= and/or"through the decontaminated sampling

equipment and collect it in the set of bottlesmarked "Field Blank". Preserve appropriately.

5.7 If field decontamination procedures are used,collect the field blank after decontamination,prior to the next sample. Decontaminate theequipment again.

5.8 Keep the trip and field blanks with the samplesand return to the laboratory for analysis.

5.9 Log in the samples for analysis using agueoustechniques and holding times.

5.10 Your supervisor will review the data from thetrip and field blanks for indications thatchanges are needed in the sample handlingprocedures.r

300973

Page 158: ftR3008L6 - semspub.epa.gov

Method: Adhering to Holding Times

Method No.: 010

Reference: " Feoleral" Registry, 40 CFR, Part 136

Date: January 1987

1.0 Scope, .and Application .. . . . .

1.1..This is a procedure describing how the laboratoryassures that analyses are conducted within theirrequired holding times. - It is applicable to allparameters with EPA or NJDEP documented holding timesless than 30 days.

2 . 0 'Summary o f Method . . . . . " . ' . , . . .

2.1 When the samples are received, the sample custodianrecords the date that .each parameter must be analyzedby in order to be completed within proper holdingtimes. Each day, the .log of samples whose holdingtimes are about to expire is examined by the samplecustodian. The dates are checked against the jobfolder to determine if the analysis is complete. Ifthe . analysis has not been performed the LaboratoryManager is notified and the analysis is immediatelyscheduled so that the holding times are not violated.

3 . 0 "Definitions . . ........ . ; .. ... . . . _ . . " » . .

3.1 The EPA and NJDEP have established time limits withinwhich analyses must be performed'. If the analysis, isnot completed within the prescribed time limit the datawill probably be considered invalid. Holding timeshave been established for all regulated waterparameters and many soil parameters. Where no holdingtime is noted for soil parameters, NAC will adhere tothe - aqueous holding time. Sometimes the EPA holdingtimes are different from the NJDEP holding times. Itis essential to )cnow which regulatory agency haspreference. in most 'cases for NAC, the NJDEP holdin§times are prevalent. Note that holding times are legallimits that may not always reflect knowledge aboutchemical degradation or the lack thereof. It is thelegal .implications that make holding times soimportant. Holding times start as follo

3.1.1 -When the sampling date and time : are known, theholding time starts with the documented date andtime :.of sampling from the- Chain of -d.*-*"- 1'. (COC), bottle label or other document. - ;

r. " •;^ L-' _c

Page 159: ftR3008L6 - semspub.epa.gov

Adhering to Holding Times ™ -Page 2 of 5 . . . . . . . . ...-._ __. .

3 - ° Definitions , ._.. .. ... . __ ^__ __. __ . _ _ _ _ _•_.. ^

3.1.2 When only the date is known, the holding timestarts with the date of sampling.

3.1.3 If the date and time of sampling are not knownthe holding time starts with the date of samplereceipt. This practice is discouraged, andefforts should be made to .find out at least thedate of .sampling.

Note.: The above discussion assumes that properpreservation .techniques have been employed.

3.2 NAC will use the following schedule of holding times:

Holding Time..____...._... .___ ... . ........ . ... ... . .. . .Solid..: .._.

Parameter. . . -.-. .-: - - —rXcrueou^ Matrix . - Matrix

GC orr:GC/MS_Tests. „ ,.. .. .. _... _.,. _„ ..... _ . ___ ._..... _ ... ..

Volatile Organics 14 days " 7 days(EPA 601, 602, -7 days (ECRA)603, 624)

Semi-Volatile Organics 7 days (extraction) same(EPA 612, 625) " 7:40 days (analysis of extract) same

Pesticides- and PCB's ; 7 days (extraction) same(EPA 608) ... 40 days (analysis of extract) same

Wet Tests „...„__._.__ ,., .._:.._._.. ._i;:;._ - : .' "~I_ _:. ; ...._..

Alkalinity Immediately N/AAmmonia - - .-- - -.--28 days - - - " - Same*BOD .- "-- 48 hours -- . - ." N/ABromide --^--••-.... -28 days __ Same*COD ' - .: : -----28 "days ":."".._ :.. Same*Chloride., . -. . I.-, ...•::..•.:"•- "ki-: ~lr .: ::" "=--"l" : : : ,Chlorine -...-...-_.- ^ immediately . N/AColbr .1 :: :. " "=i"-48 hours " " N/ACyanide --": -....-•- --14 days (24 hrs. if S =

present) Same*Fluoride ..-- -V--28 days - - - - - - - «aiae4 „ _Hardness -. - 6 months H r^VAJ U 3 I JpH . . . . :-_ - _.::=. . -.lininecliately . - - - ys;fc

Page 160: ftR3008L6 - semspub.epa.gov

Adhering to Holding TimesPage 3 of 5 V ~. ".::: r:!_."_""~~'".' . "" ""-* - :~"."" ""-

,3.0 Definitions (Cont'd) . . . _ . . ...

Holding TimeSolid

Parameter ._._. ,__ -Aqueous Matrix Matrix

Wet Tests... cent'dl ._„„_.......„.. , - . ....

TKN 28 days - Same*Nitrate-N 48 hours •. Same*Nitrate-Nitrate-N 28 days Same*Nitrate-N 48 hours _: Same*Oil and Grease V " 2B days Same*Petroleum Hydrocarbons 24 hours 7 days (ECRA - gas)Ortho-P04 '~~~ - 48 hours -28 days (all others)Dissolved Oxygen - . • " . -_-...(Probe) " Immediately N/A

Dissolved-Oxygen -Wirikleiri:.:- ".•"..: •" " "" ; 8 hours " ' N/A

Phenols : .. - ~ -"- "28 days " Same*PhOSphorOUS . : •: •:•:.— • - ——

(elemental) ..... .._.__ 4_8. hours..... .... ...... ... .. Same*Phosphorous . ... .. .(total) : ...._._. . ."_ _28 "days"5 Same*

Total Solids 7 days "'" ' Same*T S S . ..i-- -—-•"j-. " " " ^r-^7_days" . . . . ^ N / ATDS - -- —— -"-: :_. "..Ji.- u-i-iVT S-jiqurs """ "" : , N/ATVS -"' .-' """- • ~7 days .-.-". - ' '.' - Same*Silica .. •":". - - 28 days Same*Specific Conductance..'.-immediately .. . . .._ N/ASulfate . _ ~^._.""" r/.V._." 7 7" days Same*Sulfide--_^_j^__i.-- -._-_^-7 days , ~~. - Same*Sulfite .-. _..._._. ,"_:..--Immediately _ Same*Surfactants "(MBAS) z"48 hours" - ----- --- N^ATemperature ._ __..._ ;_ ,.,-;,.; Immediately (on-site) N/ATurbidity 48 hours - :

Metals _. ..... .._....____. .: •_, . : . . . . . . . . . . . . . . . - • . . . . . . .

Chromium IV 2-4 hours N/A"Mercury . 28 days - .._. Same*All "other metals :6 months" . "" Same*

* No holding time established AR300976N/A: Not applicable .. .-... -.-,-,- ---

Page 161: ftR3008L6 - semspub.epa.gov

Adhering to Holding TimesPage 4 of 5 '" - - - - - - - - - - - — ----- -

4.0 Materials

4.1 Log-in form

4.2 Holding Time Record Book

4.2.1 The Holding Time Log Book is a bound laboratorynotebook kept in the log-in area.

4.272 Eactf'p'age "of the noteboolc has a date listed on__ .tG_e_top_pf the page.

4.3 Client folders, as needed

4.4 Work Out Log, as needed

4.5 Laboratory notebooks, as needed

5.0 "Procedure ___^'_~~'. _._^._'~_ .;,_7_"T1__ ""•".". -""" "_|.V_. : .. _. _ .

5.1 Recording the holding times as follows:

5.1.1:-Log in the samples using NAC Procedure 001.

- 5.1.2 Find out the date and time that the samples weretaken.

5.1.3 Using the date of the sample as a starting pointfind the date that each parameters' holding timeexpires. . -- •

5.1.4 On the date that the parameter holding timeexpires write the NAC number and parameter.

5,1.5 Repeat for" every"parameter with a holding timeof 30 days or less. If it falls on a Saturdayor Sunday; use the previous Friday as theexpira"tion date, but put the actual date inbrackets.

5.2 Checking on Adherence

5.2.1~-Every day at 8 am open the Holding Time RecordBook to the actual date.

Page 162: ftR3008L6 - semspub.epa.gov

Adhering to Holding Times , _ -Page 5 of 5 " " "._ . : . " " . . . . . . "

- 5 . 0 Procedure (Cont'd^ . . . . . .

5.2.2 Check for holding times about to expire asfollows:

5.2.2.1 Check the Work Out Log. If the reportis out, proceed no further.

5.2.2.2 If it is not logged out check theclient folders. If the data is listed,proceed no further.

5.2.2.3 If \the client file does not list the_ analysis as complete, refer this

immediately to the Laboratory Manager.He will refer you to the analyst tocheck his/her notebook. If theanalysis is not initiated, theLaboratory Manager must be notifiedimmediately.~* ^ I-M— _i

5.2 .2.4 The Laboratory Manager must place hisf initials next to any parameter not

initiated by the expiration date toprove that his attention was brought tothe matter.

5.3 The Laboratory Manager shall review all instanceswhere the analysis is not initiated until the dayor time of expiration with the Vice President-Operations.

5.4 The client must be notified immediately if anyanalysis is initiated after the day or time of

* expiration.

300978

Page 163: ftR3008L6 - semspub.epa.gov

Man_ Northeastern Analytical Corp

Method: Standard Chain of Custody

Method No.: 009 '

Reference: USEPA Handbook for Analytical Quality Control inWater and Wastewater Laboratories

Date: November, 1986Revised: October, 1987

1-0 Scope and Application

1.1 This is a procedure describing the proper use of chainof custody procedures. The primary objective is tocreate a written record which can be used to trace thepossession of a sample from its moment of collection toits disposal.

2.0 Summary of Method

2.1 A Chain of Custody (COC) document (Attachment) isfilled out by the field technician taking the sample.-Each change in the custody of the sample is recordeduntil its placement in a secure area of the laboratory.Its final disposal or disposition is noted on thedocument.

3.0 Definition

3.1 A sample is in someone's "custody1* if:

3.1.1 it. is in one's actual physical possession;

3.1.2 it is in one's view, after being in one'sphysical possession;

3.1.3 it is in one' s physical possession and thenlocked up so that no one can tamper with it;

3.1.4 it is kept in a secured area, restricted toauthorized personnel only.

3.2 This standard custody procedure does not extend tomovement of the sample within the laboratory. Once thelaboratory custodian has signed for the samples, thecustody chain remains as is except for finaldisposition or disposal.

AR300979

Page 164: ftR3008L6 - semspub.epa.gov

MOP. _ Northeastern Analytical Corp

Chain of 'CustodyPage 2 of 3 - .

4.0 Procedure . _ . . _ .__ ._

4.1 The Chain of Custody (COC) procedure begins at the timeof sample collection. The COC is kept with the samplesuntil custody is retained by the laboratory.

4.2 Referring to Attachment I,'fill out as follows for eachsample. Multiple samples/locations can go on one COC,but not multiple clients or sample sites.

4.2.1 "A" Project No: The four (4) or five (5) digitlaboratory log-in number assigned when thesamples are brought back to the lab.

4.2.2 "B" Project Name: Client name and location- - - - - - - - ..— _i.e., ABC Corporation, Bedford, NJ.

.4.2,3 "C" Name of the person(s) taking the samplesand retaining custody as defined in Section3.1 above.

4.2.4 "D" Sample No.: Number each separate sample,beginning with No. 1 to completion. If 8samples/locations are taken number thesamples 1 to 8.

4,2.5 "E" Date sample was taken.

4.2.6 "I"1 JTime sample was taken, using 24 hour clock.

4.2.7 "G" Note using a check mark if the sample issoil (or any solid) or aqueous.

4.2.8 "H" Sample Location: Description of samplelocation i.e., "outfall from 001" or drumII.

4.2.9 "I" The number of containers taken for thissample. A typical water sample taken formultiple parameters cam have 10 or moresample containers.

AR3;00980

Page 165: ftR3008L6 - semspub.epa.gov

MOP Nortneastem Analytical Corp

Chain of CustodyPage 3 of 3 .

4.0 Procedure fCont'dl

tf

4.2.10 "J" This area is to list the type of container,preservative or analysis. Typicaldescriptions are as follows:

4.2.10.1 250 ml plastic, sterile4.2.10.2 40 ml vials4.2.10.3 950 ml amber w/H2SO44.2.10.4 950 ml amber for organics4.2.10.5 500 ml clear, HNO34.2.10.6 500 ml clear, NaOH

4.2.11 "K" The quantity of each container type.

4.2.12 ML" .Note anything unusual.

4.2.2.. .13 "M" Signature of current sample _custodian"~' ""possessing custody to ne"xt" person as

defined in Section 3.1. Don't leave anyspaces-start signing here.

4.2.14 MN" Date and time of transfer.

4.2.15 "0" Signature of person assuming custody asdefined by Section 3.1. When shipping bycommon carrier, UPS, Federal Express, thiswill not apply. The signature of the

^ person""at the destination will go here.

4.3 Upon delivery at the laboratory, the samples will be- logged in according to Method No. 001, and a laboratorynumber assigned in location "A" on Attachment I.

4.4 The sample will be stored in the appropriate location,i.e., walk-in refrigerator and the custody signed overto the laboratory administrative assistant. The COCwill be kept in the "Active COC" file. IMPORTANT; Theperson assuming custody at any point is responsible fordetermining that the COC record is accurate with regardto • all information documented. Any deviation inquantity, descriptions, condition of sample, etc. mustbe noted and brought to. the attention of the Laborat^™Manager. -

4.5 When the final disposition is determined,to client or disposal, the COC will be signedisposal noted and the COC will be filed in the"Completed COC" file.

flR300981

Page 166: ftR3008L6 - semspub.epa.gov
Page 167: ftR3008L6 - semspub.epa.gov

Method: Log-in System .. . . ....

Method No.: 001

Date: August," 1986

1.0 Scope and Application _ . . .. .. . . .

1,1 . This is a procedure describing how samples are loggedinto the laboratory to initiate the analysis. All

. samples are given a unique numeric code to ensure. samples are not mixed up. The log-in system containsall of the information necessary for the samples to beanalyzed within the required holding time and for thecorrect parameters. " ~

2.0 Summary of Method .___„_.___..._..-,'-

A Customer Order (CO) (attached) is filled out for each setof : samples from each sampling location for each day.Information requested includes client, date of sampling,matrix, required analysis due date, etc. The CO isdistributed to the: laboratory departments who initiate the

-—analysis and to_ various other personnel who interact withthe system, v . . . ... --- -

3.0 "Procedure " .."" :. . ..; ..... . -'',.. ".._, .=.-=, =.r™ ;. .... . . . .

3.1 —Obtain a Customer Order (CO) form from the log-in area.

3.2 Unpack all of the samples and line up on a table in thelog-in area. Inspect for damage, air bubbles in vials,etc. Match against the chain of custody (COC) beforeyou sign if a custody change is to take place.

3.3 Obtain the next numerical Lab Order Number from the LabOrder -Logbook. .This is the master job number for thisset" of samples. Each sample within this set will besequentially numbered in addition, i.e., Job. No. 432.1may contain samples 4321-1, 4321--2, 4321-3 etc. Do notuse -1 if there is only one sample.

3.3.1 - -Organize your samples so that the log-insequence makes sense. Don't label sampleWell No." 1 as -"2 and Well No. 2 as -1.

flR30Q983

Page 168: ftR3008L6 - semspub.epa.gov

Log-in System " . . . . . .Page 2 of_.4 .... . . -.. ... .... ... ... ... ___ ___.. . . . _ ....... _

3 .-0 - Procedure (Cont' d^

3.3.2 Write the sample number on the bottles. In manycases, multiple sample containers are used for

- ——one sample, for different analyses on thatsample. Use the same number on each of thebottles taken for a sample.

3.3.3 Test the pH of all incoming aqueous samples thatare acid or base preserved. Record the resultsas acceptable or not "acceptable in the pHLogbook. Consult the Laboratory Manager in theevent of a problem.

3.3.4 Refrigerate or otherwise store the sample in".! _ .._-_.. . ." "accordance with EPA/NJDEP requirements

immediately upon completion of"3.3.3 above.

3.4 Fill out the required information in the Lab OrderLogbook.

3 .5 -"- Referring to Attachment I "

3.5.1 "A". ..—Insert "Lab Order Number

3.5.2 "B" Date sampled - if you do not know, as inthe... case of samples shipped . in, leaveblank.

3.5.3 "C" Insert date the sample actually arrived,not the date of log in, if different.

3.5.4 "D" Circle...appropriately

3.5.5 "E" -Date due - insert if known. This is veryimportant for priority turnaround samples.

3.5.6 "-F" Clients' full name.

3.5.7 "G" Clients' full address.including zip code.

3.5.8 "H" Principal .contact who authorized theanalyses, and his/her title.

3.5.9 "I" - Circle appropriately. If yes, inspect COCversus samples, call - any discrepancies tothe attention of the lab anaaer. File COCin "Active COC" file aftercustody to sample custodian.

HR3Q098U

Page 169: ftR3008L6 - semspub.epa.gov

Log-in SystemPage 3 of 4 . . . . _ " " 1

3.0 Procedure fCont'd)

3.5.10 "J11 Insert quantity of each sample as follows:

.,..=,i.e./ ••-• 2 Aqueous2/0/0 Solid/Sludge/Oil0/2/2 Air/Filter/Tube

3.5.11- "K" Check appropriately - This is extremelyimportant for the laboratory in determiningholding times, methodologies, etc.

3.5.12 "-L." .-Insert the sample number(s) and the sampledesignation

i.e., 4321-1 _ Well No. 14321-2 "Well No. 2

3.5.13 "M" Insert the required analytical tests, i.e.,if .the two above samples are to be analyzed

- - for:.EPA Method.601, 602 and 608 but 4321-2also is to be analyzed for Oil and Grease,fill .the CO out as follows:

Organic

A28022

601 . . . .602603

, 2 608624625 : : : - ' :. .PCBHERBIR

__ 1 O&G .TOCTOX127 _ _ ......FORM,

If all samples get a test, put the number oftests in the left column. If only certainsamples get a test, list the number of- testson the left, and the sample designation on

AR300985

Page 170: ftR3008L6 - semspub.epa.gov

Log-in SystemPage 4. of 4. ...._• -.-- -: • - ..-- -. - - - - -

3.0 Procedure fCont'd)

3.6" Attachment II shows a completely filled out CO for 4aqueous _and 4 solid samples. Use it as a guide.

3.7 A CO will be filled out for every job not currentlybeing handled by the "FHA" system.

3.8 The CO will be filled out by the NAC person who samplesthe j ob or who receives the j ob as it comes in thedoor. . . ..

3.9 - The CO must be completely and neatly filled in, with nomissing information. It will probably be necessary toconsult with the. Laboratory Manager or the MarketingDepartment to complete the Co for most jobs.

3.10 The complete CO must be distributed the same day thesamples come in if they arrive before noon. If thesample arrives, after noon, 10:00 AM the followingmorning is the deadline. ' _ - " ' _

3.11 A job folder is to be created with each CO. Theoriginal (8-1/2x14) of the CO goes to Marketing with 8-

_.__. -1/2x11.: copies :to the vice President of Operations andone for the job folder. The original and all" copiesare—to be in green. The original is 8-1/2x1-4 becausethe bottom section contains pricing information, thatonly marketing needs. The copies on 8-1/2x11 will cutthis, information off." """ _" " ..

flR300986

Page 171: ftR3008L6 - semspub.epa.gov

Ir*

.

ttxe (T}»ASDRZSS (

COtrrACT/TITLEj

NAC NUMBER

BATE SAWPLET

SAKTuZD BY:

T)^

,CLIENT DESIGNAT

> (B) DATE RECEIVED CO

KAC_G.IE T DATS

©COCj YES

QUMTTIT.:

tJ)-

ECRAl Y- - -— -- - -- - -

ION NAC NUMBER C.

DOT ©

HO

MATRIX

Soiid/sittd./cilAir/Filt«r/Tufi«-05)rs^^No- - - - -I SIT DESIGNATION

ro"— '

*;— — .:=. =-. -- ., t ....... :. . ._ _ ........ .....__,.

- .- ..

ORGANIC

>A2S3601602603606624

™" 625PCSSERBIF.O4GTOCTOX127FORM

Other T«atji

ANAL^T

KETALS

PPSDW1SDW2SbXjS*a«AJ.Br

" ' CdC4cr

• HEX* -"" Co

CuF*Pt)MqHnHq

' Sis«Xq

ICAL TESTS

wrr CHEMACIDAUCNE3PHBOD

• a ROM' ' C02

CODcr,CLDECLR£COLRCOfcTDCX

" DOr

""" HARDFPRE&CTKNO 3

"" H02ODOR

wrr CSEMHBASTSTSSTDSTVSSGSO4SO 3

EP TOXICITY

OLDKEVTMETPESTSi S3'VOAA££tf

BACTERIA

TCXF" FCXP ""

tfa PU£N FSME•mSnECZa,

OP04TP04

' siTURS

SPCMOLDFUHG_YZAS —— fiD^n

-£MARKS: ' ,..„_„„..,.,. .„«,.!' ' w w

Page 172: ftR3008L6 - semspub.epa.gov

• -

-•;- - •

TABLE

A-2

INSTRUCTIONS:

£

1

i.5X

j.-f 2.

Record actual t*I if

ont-.fti.te

WiTER DATA TO INDICATE COMPLIANCE WITH

1

r

6Q2

QJC

Qr-

fC

3. Record corrective action

in remarks.

4. Temperature:

Record actual tcjiperature

i11CNrHAl

VI

5.

y

o

%

j

1

1

|

P

i0iin

D8 CP

ECIAL INS5®pGnCWS/.WNCOMPIJO:

AND QA NOTATIONS

1_ZO*

1

30(39{38

,

Page 173: ftR3008L6 - semspub.epa.gov

5.0 CALIBRATION PROCEDURES ._..____. •___._/ _,

Laboratory Instruments

Northeastern Analytical Corporation (NAC) maintains specificprocedures for the operation and calibration of all analyticalinstruments. Along with proper maintenance, these practicesensure ..optimum instrument performance and accuracy. Theseprocedures include proper operator training and supervision;mandatory instrument performance specifications; andsystematic instrument calibration, verification, andmonitoring schedules.

The Laboratory utilizes mandatory instrument" performancespecifications to constantly ensure optimum instrumentalperformance. These performance specifications requireacceptable instrument response to specific performancestandards prior to initiating further instrument calibration

. and analyses. Acceptable instrument response criteria arebased upon the manufacture's or EPA's analytical methodspecifications.

Laboratory analysts record and document all instrumental runsin designated Laboratory Instrument Logbooks. These Logbooksidentify instrument operating parameters, settings, andperformance data associated with each instrumental run.

The Laboratory utilizes systematic instrument calibrationprocedures to constantly ensure analytical accuracy. Initialinstrument calibration curves are generated, systematicallyverified, and routinely monitored throughout the duration ofall instrumental analyses. Initial multipoint GC and GC/MScalibration curves are verified by injection of a knownworking standard every twelve (12) hours of instrumentoperations. An unacceptable working standard response (20%variation) requires construction of a new calibration curve.Other instrument calibration curves are constructedimmediately prior to analyses of each batch of samples. Allinitial instrument calibrations and/or verifications are thenroutinely monitored by systematic, repeat injections ofappropriate check standards throughout the duration ofanalyses. Additionally, all GC/MS analyses will utilizeinternal standards to ensure high quality data.

Specific calibration procedures for laboratory instrumentsincluding the frequency and standards used are listed in Table5-1 - CALIBRATION PROCEDURES. .

SR.3QQ989

Page 174: ftR3008L6 - semspub.epa.gov

TABLE 5-1 - CALIBRATION PROCEDURES

AA

Every 12 hours, the instrument shall be tuned to DFTPPor BFB to assure that instrument response meets EPAspecifications .

Generation of five (5) point calibration curve for allmethod compounds monthly, or more frequently if needed.

Analysis of standard after the analysis of twenty (20)samples to assure the consistency of instrument response.

Addition of internal standards to each sample.

Addition of surrogate spike compounds to every sample.. . . . . . . . _•._:_..::_ -,. = .... ;

Generation of three (3) point calibration curve for allanalyzed compounds monthly or prior to any sampleanalysis, as needed.. . _ . . _ . . . . _

Monitor .consistency of instrument response through theanalysis of a standard after every twenty (20) sampleanalyses. _._._.._

Demonstrate system cleanliness through the analysis ofa reagent blank prior to any sample analysis.

Maintain sample response within the limits of theresponse of the standards. ...

Construction of a three (3) point calibration curve foran element prior to the analysis-of any sample set.

Verification of the cleanliness of the system for eachelement through the analysis of a reagent blank with anyanalytical set." ~~ " "

Verification of instrumental consistency through theanalysis of spiked samples and controls after theanalysis of every twenty (20) samples.

Bracketing of sample response between the limits ofstandards response.

Analysis by method of standard additions when matrixinterference is suspected. , "•

&R300990

Page 175: ftR3008L6 - semspub.epa.gov

METALS

Barium

calciumChromic

MagnesiumManganese

Mercury

Nickel

potassium+ Selenium

1 flR30099l

Page 176: ftR3008L6 - semspub.epa.gov

TABLE 6-1 - LABORATORY ANALYTICAL METHODS SUMMARY

Parameter Method

Liquid " SolidMETALS _

Aluminum EPA 202.1, 202.2:' SW846 Method 7020Antimony EPA 204.1, 204.2 SW846 Method 7040,

__ 7041Arsenic EPA 206.2, 206.3 SW846 Method 7060,

7061Bariiom - - • • - . . . -.. EPA 208.1, 208.2 SW846 Method 7080,

7081Beryllium EPA 210.1, 210.2 SW846 Method 7090.

7091Boron ^ EPA 212.3 SW846 Method 7000Cadmium EPA 213.1, 213,2 SW846 Method 7130

7131 .Calcium EPA 215.1, 215,2 SW846 Method 7140Chromium EPA 218.1, 218.2 SW846 Method 7190

7191Chromium+6 EPA 218,4, 218.5 SW846 Method 7195Cobalt . EPA 219.1, 219.2 SW846 Method 720C

— - -:.--___ _ 7201Copper EPA 220.1, 220.2 SW846 Method 721C

~^" ' --: --"- :- •—--.; - --.,. ,. . . 7211Iron EPA~ 236.1, 236.2 SW846 Method 738i

" - - - - - - - - _ __ 7381Lead " • EPA 239.1, 239.2 SW846 Method 742

— - - -. _ 7421Magnesium - EPA 242.1 SW846 Method 745Manganese EPA 243.1, 243.2 SW846 Method 746

-""" ' ' ' -• .-- .- -.. .... 7461Mercury EPA 246.1, 246.2 SW846 Method 74'

7471Molybdenum EPA 246.1, 246.2 SW846 Method 74?

7481Nickel . EPA 249.1, 249.2 SW846 Method 75:

• - •-.--- :-.-. ... 7521Potassium EPA 258.1 SW846 Method 76Selenium - EPA 270.2, 270.3 SW846 Method 77

7741

AR30099?

Page 177: ftR3008L6 - semspub.epa.gov

BORATORY ANALYTICAL METHODS SUMMARY (Cont.)

Method

Solid . . . . . . . Liquid Solid

Method 7020 - -EPA 272.1, 272.2 SW846 Method 7760,

Meth° ' EPA 273.1, 273.2 SW846 Method 7770,Method 7060, EPA 279.1, 279.2 SW846 Method 7840,rn=^k-

wl^Method 7080, " SPA 282.1, 282.2•rnfti EPA 283.1, 283.2

Method 7090, IPA 286.1, 286.2 SW846 Method 7910,nc- * n

SwlU ethod 7380 " A 330.5

7841SW846 Method 7870SW846 Method 7000

•7O91 tyj-t.,-TTOA* MPthod 7000" """ : " 'PA 289.1, 289.2 SW846 Method 7950,SW8 4 o n«i . — . 7 Q mSW846 Method 7130, 79517131

SW846 Method 7140SW846 Method 7190

7195 ' 3A 305.1 - N/A-00 .A 310.1, 310.2 N/A

SW846 Method 7210. '325.3 :. Agronomy

N/A

7420,- * 335.1 SW846 Method 90107i ^ 335.2 SW846 Method 9010

SW846 Method 7450 "A 340.1 ...._. AgronomySWB46 Method 7460, ^^^ Agronomy

qW846 Method 7470, '' ~ . 351.3 Agronomy- 353.3 SW846 Method 9200

SW846 Method 7480, 353.3 . N/A7481 36°'2 N/ACW846 Method 7520, 365.2 Agronomy

- r 370'1 N/AMethod 7610 -" -:-• 375.4 SW846 Method 9035Method 7740, 376.1, 376.2 SW846 Method 9030

377-1 " - : N/A" 110.3 N/A

130.1, 130.2 N/A

'v

1 SR300993

Page 178: ftR3008L6 - semspub.epa.gov

TABLE 6-1 --LABORATORY ANALYTICAL METHODS SUMMARY fCont.1

Parameter __•__._..."._ .."_J:; _ r"."'."""" ^Me^hod

Liquid SolidMETALS : - . - . . - . : _Odor . -. EPA Method 140.~1 " N/ApH - -;EPA Method 150.1 SW846 Method 9040Residue . . . .Filterable EPA Method 160.1 N/ANon-Filterable EPA Method 160.2 N/ATotal .EPA Method 160.3 SM209Volatile . . . . . . _ .... EPA Method 160.4 SM209Settleable , EPA Method 160.5 N/A

Specific Conductance „.EPA Method 120.1 . N/ASpecific Gravity ASTM D-70 " " ~ jj/ATemperature - .EPA Method 170.1 EPA Method 170.1Turbidity EPA Method 180.1 N/ARCRA Ignitability SW846 Method 1010 ASTM D-56RCRA Reactivity SW846 Method 9010, SW846, Method

9030 9010,9030RCRA EP TOxicity SW846 Method 1310 . SW846 Method 1310

N/A: N o t Applicable . . . . . . .

REFERENCES:

EPA Methods - Methods for Organic Chemical Analysis of Municipal andIndustrial Wastewater, EPA-600/4-82-057, July 1982.

EPA Methods for Chemical Analysis of water and Wastes EPA 600/4-79-020, March 1983.

SW846 - Test/Methods for Evaluating Solid Waste, Physical/ChemicalMethods, Third .Edition, EPA, September, 1986.

ASTM - American Society for Testing and Materials.

SM - Standard Methods for the Examination of Water and Wastewater,American Public -Health Association, 16th Edition.

Agronomy - Methods of Soil Analysis, 2nd ed. American Society ofAgronomy

Page 179: ftR3008L6 - semspub.epa.gov

TABLE 6-2 - -LABORATORY' DETECTION LIMITS SUMMARY

Parameter _ _. , -H .; •„ ~ ,„_, : Water . ,__ _ Solids(and non-aqueous waste)

INORGANICS

METALS

Antimony " 0.005 0.5Arsenic - ... _ ........ .._- ._ .005 .0.5Barium 0.5 5Beryllium ..__ . . _ 0.01 1Cadmium 0.005 .5Chromium 0.03 3Copper _ :_ _- .01 1Lead .... __ _ ._ 0.005 0.5Mercury 0.005 0.05Nickel ' -,..•::" •'. :. v-;._ "O.i ._ 10Selenium . . . . . . .__.005 ,. . . 0.5Silver . ..-...- .--._.__-:::• ...:_.-___:..- - P-Q5_ __ _. 5.Thallium ~ - ~ • - - • - • • • ---.-,---,..- - - i 1 0Zinc . 0.05 . . . . . ' 5

Parameter ;-^,t,---- :-: r r - ... .._ j=.-- .... ._:.._-• jz - , .. ':_.••.:-•. . .- •

INORGANIC NOK KETALS

Acidity 3 N/AAlkalinity 3 N/ABiological Oxygen Demand (BOD) 4 N/AChemical Oxygen Demand (COD) 50 N/Achloride .... :: : '.:..".-"-•-T : : .soColor - - - - - - -- ^ - -——— ---5. N/AConductivity • 5 N/ACyanide, Free .": : :.::" "~~. .:":."-..:: :~0.01 1Cyanide, Total- - ' - ' 0.01 1Dissolved Carbon Dioxide . ..... __. - ... 1 . . N/AFluoride .... .... _.___._.._____ : _. .. O.L _.... . .... 5 _..........Hardness ^-—-- N/AIgnitability N/A N/AMethylene Blue Active Substances 0.05 N/ANitrogen:

Ammonia-Nitrogen " 1 - 5 0Total Kjeldahl Nitrogen 3 100Nitrate Nitrogen 1 20Nitrite Nitrogen 0.01 N/A

pH N/A N/APhenols, Total ... =. 0.2 40

Units: . .._•_--:_._._._:__= ..(sg/D (mg/k |fi300995

Page 180: ftR3008L6 - semspub.epa.gov

TABLE 6-2 - LABORATORY DETECTION LIMITS SUMMARY .Continued)

Parameter ,....,_......_...,_,._. Water . . Solids(and non-aqueous waste)

INORGANIC NON METALS

Phosphorous: . . . . .Ortho Phosphates .0.01 40Total Phosphorous '. 0.01 40

Silica, Dissolved _.._.. .... ... 1 N/A

Solids;; . . . . . .Total Dissolved Solids 5 N/ATotal Suspended Solids ... .__ 5 N/ATotal Volatile Solids .15 N/ATotal Solids 5 N/A

Specific Gravity N/A N/Asuifate . _ . ; •--;- . i . . 100Sulfide . . . 0.05 N/ASulfite _.. _ ; ; T 2 . N/ATotal Organic Carbon (TOG) 1 200Total Organic Halogen (TOX) 0.01 50Turbidity " .2 (NTU) N/A

Grease and Oils:Oil and Grease 5 200Total Petroleum Hydrocarbons 1 50

Units _ :_. -_:: - - .--.- ..-.__-- ._-(mg/l)K/A: Not applicable

ftR300996

Page 181: ftR3008L6 - semspub.epa.gov

TABLE 6-2 - LABORATORY DETECTION LIMITS SUMMARY (Continued^

VOLATILE ORGANICS BY GC

Solids (and non- Parameter _-._ ~r. " Aqueous aqueous wasted

Chloromethane 0.5 1.0Bromomethane 0.5 1.0Vinyl Chloride _. . ._. 0.5 . . 1.0Chloroethane 0.5 1.0Methylene Chloride . . 0.5 1.0Trichlorofluoromethane 0.5 1.01,1-Dichloroethene 0,5 1.01,1-Dichloroethane \ . 0.5 1.0trans-1,2-Dichloroethene 0.5 1.0Chloroform 0.5 1.01,2-Dichloroethane 0.5 1.01,1,1-T.richloroethane 0.5 1.0Carbon Tetrachloride ... .. .„_.. 0.5 1.0Bromodichloromethane 0.5 1.01,2-Dichloropropane 0.5 , 1.0trans-1,3-Dichloropropene 0.5 1.0Trichloroethene ... 0.5 1.0Dibromochloromethane . .0.5 . 1.01,1,2-Trichloroethane 0.5 1.0cis-1,3-Dichloropropene " _ 0.5 .._ 1.02-Chloroethylvinyl Ether 0.5 1.0Bromoform '-: T -- " 0.5 1.01,1,2,2-Tetrachloroethane 0.5 1.0Tetrachloroethene 0.5 1.0Chlorobenzene .0.5 1.01,3-Dichlorobenzene 0.5 1.01,2-Dichlorobenzene 0.5 1.01,4-Dichlorobenzene 0.5 1.0Benzene . , . . „ . . . , . 0 . 5 . . " " . . _ 1-0Toluene — .--0.5 1.0Ethylbenzene ... 0.5_. 1.0o-Xylene Q.5 1.0m-Xylene . . .. . . .. _0.5 _ 1.0p-Xylene . . 0.5 1-0Acrolein 10 " N/AAcrylonitrile . . 10 _ , N/A

Units . ._.-...-. . ; --...... Jug/1)

K/A: Not Applicable

Page 182: ftR3008L6 - semspub.epa.gov

TABLE 6-2 _^ LABORATORY DETECTION LIMITS SUMMARY fContinued-1

PESTICIDES AND POLYCHLORINATED BIPHENYLS

Solids (and non- Parameter .-.= ..._..... .^. :_ :__ . ,::: __. Aqueous aqueous waste)

Aldrin 0.03 0.03alpha BHC 0.03 0.03beta BHC 0.03 0.03gamma BHC . ,0.03 0.03delta BHC 0.03 0.03Dieldrin 0.03 0.03Chlordane = . , , . . . -,,0.01 0.01p,p'-DDE ' : -1 "7^ 0.03 0.03p,p'-DDT . . . . . . .. . . :._0..03" 0.03p,p'-DDD . • : -: :0.03 0.03Endosulfan I :. -.:••--• . :0.03 0.03Endosulfan II - _ 0.-05 . 0.05Endosulfan Sulfate " " " 0.2: 0.2Endrin " " 0.03 , 0.03Endrin Aldehyde , ._.-. .. 0.05 0.05Heptachlor ...- 0.03 0.03Heptachlor Epoxide . .....0.03 0.03Toxaphene 2 • 2Polychlorinated_Biphenyls: • . . .total as Aroclor-J.016 _. . 1. _ _ _. . . . . 1total as Aroclor-1221 ,. .. . _2 _...._. ... 2total as Aroclor.1232 1 . . 1total as Aroclor 1242 . - . . . 1. ^_.__ ... . 1total-as Arocl-or_1248 1 . . . . 1total as Aroclor- 1254 1 "~ 1total as Aroclor 1260 1 1

Units (ug/1)

RR300998

Page 183: ftR3008L6 - semspub.epa.gov

TABLE 6=2--- LABORATORY DETECTION LIMITS SUMMARY (Continued^

VOLATILE ORGANICS BY GC/MS

Solids (and nonparameter ... ..... =________ _._= Aqueous aqueous wasted

Chloromethane 10 10Bromomethane 10 . 10Vinyl Chloride -. ^ ....--.- JjO _. 10Chloroethane 10 10Methylene Chloride - .. ... ...5.0 5.01,1-Dichloroethene 5.0 5.01,1-Dichloroethane 5.0 5.0trans~l,2-Dichloroethene 5.0 5.0Chloroform ' .:.:-'- •*- : r"L - - - -.-- -'- -5.0 5.01,2-Dichloroethane 5.0 5.01,1,1-Trichloroethane . 5 . 0 . 5.0Carbon Tetrachloride. . : _ . -.. . 5.0 " _ . 5.0Bromodichloromethane 5.0 5.01,2-Dichloropropane 5.0 5.0trans-1,3-Dichloropropene. 5.0 " 5.0Trichloroetheneene . 5 . 0 5.0Dibromochloromethane 5.0 5.01,1,2-Trichloroethane 5.0 5.0Benzene - - . - . . 5.0 5.0cis-1,, 3-Dichlorophropene 5 -0 . 5.02-Chloroethylvinyl Ether - 10 10Bromoform 5.0 5.0Tetrachloroethene ..5.0 5.01,1,2,2-Tetrachloroethane 5.0 5.0Toluene . .. T ---.---.- 5.0 5.0Chlorobenzene 5.0 5.0Ethylbenzene . .. .5.0 5.0Total..Xylenes -". . - . ... 5.0 5.0Total Dichlorobenzenes 5.0 5.0

Units r.(ug/l) (mg/kg)

Page 184: ftR3008L6 - semspub.epa.gov

TABLE 6-2.- .-LABORATORY DETECTION LIMITS SUMMARY (Continued^

ACID EXTRACTABLES BY GC/MS

Solids (and nonParameter .. = _.;:_^._ .. - Aqueous = aqueous wasted

Phenol . 10 3302-Nitrophenol .10 3304-Nitrophenol 50 16702,4-Dinitrophenol 50 16704,6-Dinitro-o-cresol . 50 1670Pentachlorophenol .50 1670p-Chlorp-m-cresol """ '.".Y "~ 10 3302-Chlorophenol " " id " 3302,4-Dichlorophenol 10 3302,4,6-Trichlorophenol ". ' .10 " 3302,4-Dimethylphenol . ..... 1.0 330

BASE NEUTRAL EXTRACTABLES BY GC/KS

1 —-----•• ----- - Solids (and nonParameter .,_._ . -• —Aqueous aqueous wasted ...

1,2-Dichlorobenzene 10 3301,3-Dichlorobenzene : 10 3301,4-Dichlorobenzene 10 330Hexachloroethane . . 10 ... 330Hexachlorobutadiene 10.._..." ... 330Hexachlorobenzene :- • -- - - - 10._ _ . 3301,2,4-Trichlorobenzene .10 330bis (2-Chloroethoxy)Methane . . .... - _ .-,. , :, _,,?::10 _ .... , _ 330Naphthalene = .= ,._' ;10_._. _ _ 3302-Chloronaphthalene 10 . 330Nitrobenzene 10 330Isophorone 10 3302,4-Dinitrotoluene - - . 10 3302,6-Dinitrotoluene , 10 3304-Bromophenyl PhenylEther _ - - . . 10 330bis (2-Ethylhexyl)Phthalate ..,.-- , _ . , - . . . . 10_,_ . ... . , 330Di-n-octyl Phthalate . .,_.UQ. . . . .. 330Dimethyl Phthalate 1 0 . . . 3 3 0

Units .._... ... . _ (ug/1) (mg/kg)

AR3Q/000

Page 185: ftR3008L6 - semspub.epa.gov

TABLE 6-2 - LABORATORY DETECTION LIMITS SUMMARY (Continued,!

BASE NEUTRAL EXTRACTABLES BY GC/HS (Continuod)

Solid and NonParameter .. ... _ Aqueous Aqueous Waste

Di-n-butyl Phthalate 10 330Diethyl Phthalate 10 330Acenaphthy1ene 10 . 330Acenaphthene 10 330Butyl Benzyl Phthalate 10 330Fluorene 10 330Fluoranthene 10 330Chrysene 10 330Pyrene .10 330Phenanthrene - 10 ._ . 330Anthracene --.._- ....:. _, 10 __ 330Benzo (a) Anthracene 10 330Benzo. (b) Fluoranthene .. 10 •-•••- 330Benzo (k) Fluoranthene 10 _ 330Benz o (1) Pyrene 10 330Indeno (1,2,3-c, d) Pyrene 10 330Dibenzo (a,h) Anthracene .10 330Benzo (g,h,i) Perylene 10 3304-Chlorophenyl Phenyl Ether .10 , 330Benzidine - - - - ^ - - - - - - 2 0 6 6 03,3-Dichlorobenzidine ~~ 20 - --- - 660bis (2-Chloroethyl) Ether IP 330Hexachlorocyclopentadiene 10 330N-Nitrosodiphenylamine 10 330N-Nitrosodi-n-propylamine 10 330bis (2-Chloroisopropyl) Ether 10. .. Y 330

Units ' .:..-- .... (ug/1) (mg/kg)

Page 186: ftR3008L6 - semspub.epa.gov

7.0 Internal Quality Assurance Practices

Assurance/Quality Control Program is designed to ensure thescientific and legal validity of all analytical results. TheQA/QC Program consists primarily of a laboratory documentationnetwork in combination with systematic inclusion of variousanalytical quality control practicles into all componentlaboratory operations. These quality control practices willprovide constant, documented evaluation and surveillance ofacceptable analytical method performance.

7.1 Internal Quality Control Checks

Analytical quality control checks are performed in thelaboratory. These procedures are based upon USEPAanalytical methods guidance and generally acceptedstandards of good laboratory practice as outlined inTable 6-1. Key components of the Laboratory AnalyticalQuality Control Program include the fol1owing qualitycontrol practices and considerations:

- adherence to specified Laboratory Sample AcceptanceProcedures to ensure proper handling, processing,and storage of submitted samples (Section 4).

__.._._._. utilization of USEPA-approved Analytical Methods andInstrumentation (Section 6 and 12).

adherence to mandatory procedures for Operation,Calibration, and Maintenance of LaboratoryInstrumentation (Section 5).

utilization of proper Laboratory MeasuringEquipment, Glassware, Water, Chemical Reagents,Industrial Gases.

- constant surveillance and documentation ofacceptable analytical method accuracy and precisionthrough Initial Analytical Method PerformanceEvaluations and matrix spike/spike duplicateevaluations method blanks, field and trip blanks,bias control sample and EMSL's.

utilization of continuous Surrogate Spike RecoveryEvaluations where appropriate to ensure acceptablemethod performance.

- utilization of systematic Method Blank Evaluationsto identify analytical system interferences andbackground contamination levels.

AR301002

Page 187: ftR3008L6 - semspub.epa.gov

7.1 Internal Quality Control Checks (Continued)

adherence to proper Laboratory Documentationmeasures to ensure the complete integrity and legalvalidity of all Laboratory analyses.

utilization of Voluntary Intralaboratory PerformanceEvaluations to internally assess and evaluateanalytical performance.

participation in numerous Laboratory certifications,audits, and approval programs.

7.1.1 .Data Quality

principle criteria for validating data qualityis the continuous monitoring of analyticalaccuracy , precision , and overal 1 methodperformance through systematic analysis of qualitycontrol samples. KAC conducts both initial andcontinuous analytical - method performanceevaluations _to ensure that all generatedanalytical data meet acceptable quality controlmethod performance "criteria established by theUSEPA and the Laboratory. Each analytical methodcommonly used in the laboratory utilizes specificquality control procedures to continually monitoracceptable analytical method accuracy andprecision. These method quality controlprocedures primarily involve the mandatorysystematic insertion of quality control samplesinto at least 10% of all laboratory analyses, inaddition to strict adherence to instrumentperformance and calibration specifications. Thesespecific quality control procedures are thoroughlydetailed in the Analytical Methods SOP in thequality assurance section and. are based upon USEPAmethod guidance. (Table 6-1) .

7.1.1.1 Initial Method Performance Evaluation

Prior to the analysis of any samples, theLaboratory conducts Initial MethodPerformance Evaluations of each analyticalmethod as required commonly used in theLaboratory to demonstrate the ability toachieve acceptable method accuracy andprecision. Quality control data derivedfrom these evaluations is compared topublished USEPA method performancecriteria in order to determine #$£0 n i n n Oflnou I Uy g

Page 188: ftR3008L6 - semspub.epa.gov

7.1.1.1 Initial Method Performance Evaluation(Continued)

document acceptable analytical methodoperation. This Initial MethodPerformance Evaluation is conducted assummarized below.

A minimum of four (4) matrix spike samplesare prepared using the particularrepresentative sample matrix analyticalparameter(s) of interest. These matrixspike samples are prepared such that theparameter concentration(s) are within theworking range of the method and at leasttwo (2) times greater than the method'sbackground level.

The matrix spike samples are analyzed inaccordance with the method, and theaverage percent recovery (R) and thestandard deviation of the percentrecoveries (s) will be calculated from theanalytical results. The Laboratory valuesof R__and s jare compared to the published

••--•-- EPA method performance values of averagerecovery (X) and standard deviation (p)(if- applicable). The laboratory values

-- — are not acceptable if s>2p or X-R>2p.Unacceptable values require the Laboratoryto review potential analytical problemsand repeat the Initial Method PerformanceEvaluation until acceptable values areobtained.

The Laboratory does not analyze any actualsamples until the Initial MethodPerformance Evaluation demonstratesacceptable laboratory analytical methodcapability. Results from the InitialMethod Performance Evaluation are enteredinto the analytical method's InitialAccuracy and Performance Record maintainedin the Laboratory QC Manual.

7.1.1.2 Continuing Matrix Spike/spike DuplicateEvaluations

The laboratory conducts continuous methodperformance evaluations for eachanalytical method commonly us3f*DHYlV'V?nLaboratory to continuously demoneWraiy irtJQ Hdocument the maintenance of acceptablemethod accuracy and precision. Laboratory ,

Page 189: ftR3008L6 - semspub.epa.gov

7.1.1.2 -Continuing Matrix Spike/Spike DuplicateEvaluations (Continued)

quality control charts or tabularsummaries are constructed from this datain order to monitor and compare actuallaboratory quality control data withacceptable published USEPA or Laboratorymethod performance criteria.

Quality control samples consist of methodblanks, bias controls, and matrixspike/spike duplicate samples.

Mandatory matrix spike/spike duplicatesamples are systematically analyzed inorder to maintain continuous surveillanceof acceptable method performance.Approximately fifty percent (50%) of allquality control samples are matrix spikesamples. Percent Recovery determinations(R) from these matrix spike results aremonitored to provide a continuous measureof the overall accuracy and precision ofthe method in addition to determiningextraction efficiencies and sample matrixeffects.

7.1.1.3 Method Blank Evaluations

NAC systematically prepares and analyzesmethod blanks to continuously evaluateanalytical system interferences andbackground contamination levels. MethodBlank analyses include all components(glassware, chemical reagents,environment, etc.) of actual, routinemethod analyses, substituting reagentwater or other applicable clean matrix forthe actual sample. Approximately twenty-five percent (25%) of all quality controlsamples are method blanks. Analyses ofmethod blanks will provide a safeguardagainst interfering and/or contaminatedreagents, glassware, and laboratoryenvironments. Unfavorable method blankperformance will-render associated datasuspect and require corrective action (seeSection 11.0).

Page 190: ftR3008L6 - semspub.epa.gov

7.1.1.4 Bias Control Sample Evaluations

The laboratory performs analysis of bias' control samples for all analyses. Bias

control samples are made in the samemanner as matrix spikes, except thatanalyte is added 1to deionized waterinstead of a sample. This allows scrutiny

^ of any matrix effects that may beoperating. The bias control sample isrun when matrix spike samples are run.

k ._ .7.1.1.5 Continuous Surrogate Spike Recovery

Evaluation

3 NAC conducts continuous surrogate spikerecovery evaluations on all GC and GC/MSvolatile and semivolative analyses to

i . ensure acceptable method performance.Surrogate spikes consisting of similar,non-targeted compounds and/or methodcompound analogues are added to each GCand GC/MS volatile and semivolatile

'• fraction to continuously evaluate and1 document acceptable method performance,

without interfering and actual methodcompounds.- . _ . . . . . .

r^Surrogate spike recoveries must comparefavorably to published USEPA method orstatistically derived Laboratoryperformance limits in order for ananalysis to be acceptable. Unfavorablesurrogate spike recoveries renderassociated data suspect and requirecorrective action.

Page 191: ftR3008L6 - semspub.epa.gov

8.0 Data Reduction, Validation, and Reporting

8.1 Data Reduction

KAC currently utilizes a manual data management systemto record, document, and assimilate pertinent laboratorytechnical and administrative data. This system providesdata handling functions for a number of componentlaboratory activities including: Laboratory Sample login,Sample Analytical Results, Sample Tracking and AnalyticalQA/QC. This system is summarized below.

The data management system assigns an individualLaboratory Identification Number to each sample andcontains records pertinent technical and administrativesample data. Pertinent technical sample data includesthe client's -sample identification, sample physicaldescription, sampling data (if known), requiredanalytical parameters, and requested completion date.Pertinent administrative data is necessary for finalreporting and invoicing of results.

Laboratory worksheets are generated for distribution tothe appropriate analyst(s). These work sheets identifythe appropriate analytical parameters and associatedmethods necessary to complete the requested sampleanalyses along with the turnaround time requirements.These turnaround requirements not only specify requestedcompletion dates, but identify maximum allowable holdingtimes for samples and/ or extracts prior to analyses.

Laboratory personnel enter all completed sampleanalytical results and associated QC data onto theworksheets .

8.2 Data Validation Criteria

8.2.1- .Data Validation - During Analysis

principle criteria to be used to validate dataintegrity during analyses will be the following:

- Reagent blank results- Method preparation blank results- Calibration verification- Matrix spike/spike duplicate results

Quality Control check sample results

Page 192: ftR3008L6 - semspub.epa.gov

8.2.1 : Data Validation - During Analysis (Continued)

These measurements are to be wade by the analyst,using specific acceptance criteria. The analystwill either proceed with the analyses or takecorrective action (see Section 9) . All QC datais reviewed by the appropriate supervisor toensure that all QA procedures have been completed.

NAC utilizes complete laboratory documentationmeasures to ensure the integrity and legalvalidity of all sample analytical results. These

..._--. . .documentation measures encompass all analyticalactivities to create a traceable, legal historyof each sample and subsequent analysis. Alldocumented information is recorded in bound,consecutively-numbered analytical logbooks.Component analytical ..documentation measuresinclude:

8.2.2 Laboratory Sample Logbook

Samples submitted to KAC for laboratory analysisare recorded and documented in the Laboratory-Sample Logbook. Individual log entries includeclient, laboratory sample identification number,sample description, analytical requests, chain-of-custody possession statements, and additionalinformation."""

8.2.3 Laboratory Method Logbooks

All laboratory analyses are entered into variousLaboratory Method Logbooks which categoricallyrecord and document the raw data for eachanalytical parameter commonly determined. Eachanalytical parameter and/or activity is assigneda particular Laboratory Method Logbook whichrecords pertinent preparation, extraction, andinstrumental data for each sample. This includeslaboratory identification number, initial samplevolume or weight, extraction volumes, dilutionfactors, instrument values, and the name of theanalyst(s). These logbooks also systematicallyinclude quality control data. Method Logbooks aremaintained for seven years by KAC.

AR30I008

Page 193: ftR3008L6 - semspub.epa.gov

8.2.4 Laboratory Instrument Logbooks

All laboratory analyses requiring analyticalinstrumentation are recorded in various instrumentlogbooks which categorically record and documentanalytical instrument settings and performancedata . These__ logbooks record instrumentcalibration data,specific: sample volumes, instrument parameters,and corresponding performance data for each'sample.* Instrumental information has beenincluded and combined with Laboratory MethodLogbooks whenever possible to consolidate data.Laboratory Instrument Logbooks are retained forseven years.

8.2.5 Laboratory Equipment Calibration Logbooks. .- rr --

All Laboratory measuring equipment calibrationsare recorded in various Laboratory CalibrationLogbooks. These logbooks record the dates andprimary standard for calibration of variousLaboratory thermometers, balances, and glasswareitems.. ___ The Laboratory Calibration Logbooks areretained for five years.. ........ , _„.... r.. .. . . . . . .

8.2.6 Laboratory Chromatography Data File

All -Chromatography data generated by KAC iscategorically filed in the laboratoryChromatography data storage files. The filesinclude labeled, numbered chromatograms withcorresponding integrator print-outs and raw datasheets. This file is retained for seven years.

8.2.7 Laboratory GC/MS/DS Data File

All Chromatography data and correspondingquantitation lists generated by the LaboratoryGC/MS/DS Systems are categorically filed in theLaboratory GC/MS/DS Data File. Processed GC/MSData is filed on a daily basis in the appropriateLaboratory Processed .Data File (VOA, B/N/A. Inaddition, all GC/MS/DS Chromatography data,quantitation lists, and processed data arerecorded on nine track magnetic tape. Bothhardcopy data and magnetic computerized data areretained by the laboratory for seven years.

Page 194: ftR3008L6 - semspub.epa.gov

8.2.8 Analytical Quality Control Data Logs

The Analytical Quality Control Data Logs containall data generated by the Analytical QualityControl Program. These files record the raw dataand subsequent statistical calculations of variousLaboratory quality control components including:continuous method performance evaluations,surrogate spike recovery evaluations, and method"Hlank analyses. Laboratory QA/QC Data SummaryReports containing quality control datastatistical, .summaries and associated qualitycontrol charts are routinely generated from thisdata file.

8.3 Data Validation - Data Reporting

Subsequeht to data entry, the assistant laboratorymanager and the administrative group are responsible for-comparing analytical worksheets with actual data enteredonto the test report draft. The laboratory manager alsoreviews the quality control report that accompanies thefinal report. " .-..-=__._.._. -.

The associated project manager or account representativeis then responsible for reviewing the final reports priorto "release to the client. Reports are to be reviewedfor: — - - . . . . . . . _ _ . _ _ _ _ _ _ _ _ _ _ _ ......

Completeness of results for all parametersrequested: detection limits, units, dates, andsample descriptions are complete and correct.

Consistency - all parameters are viewed for internalconsistency (Cr VI Cr T, TKN NH3-N, TS TSS, etc.)

The final report is then filed after the above criteriahave been met.

8.4 Data Reporting Format

Laboratory Final Reports include appropriate introductorycomments, analytical methods summaries, sampling data,if applicable, results, and QA/QC data. The supportingmaterials previously mentioned are provided byadministrative personnel as appropriate for inclusion infinal reports. Additional technical narratives, alongwith supporting raw data may also be included aswarranted by special circumstances. Abridged analyticalresult summary final reports are provided for .smallprojects or those requiring rapid analytical^times. These abridged materials primar:"analytical results. Additional supporting materialbe provided .if., prearranged.

RR301010

Page 195: ftR3008L6 - semspub.epa.gov

9.0 Performance Audits " ....-- _ ._.

Participate in a number of performance audits, both internaland external, to monitor the capability and performance ofthe laboratory and its operations.

9.1 External Laboratory Certifications, Audits, and Approvals

KAC routinely participates in performance audits toveri fy the qua1ity of its measurement systems. Theseaudits are conducted on a regular basis as a part ofnormal laboratory operations. In addition, thelaboratory participates in a number of federal, state,and private Laboratory Certification, Audit and/orApproval Programs in order to demonstrate its analyticalcapabilities. Participation in these programs requiresthe laboratory to demonstrate acceptable laboratoryperformance through satisfactory completion ofperformance audits. As a part of its certification bythese various federal, state, and private agencies,submits to on-site external systems audits. Theinspection audits evaluate the adequacy of laboratorypersonnel, equipment, documentation, and QA/QC.Performance audits require satisfactory blind analysesof unknown interlaboratory performance evaluationsamples. A listing of Laboratory Certifications, Audits,and/or Approvals.'. currently maintained by the Laboratoryfollows:w

i-AGENCY .____..____ .._.. TYPE O'F CERTIFICATION

[ USEPA _ Laboratory No. NJ 101"=" ^ " KPDES DMR studies

r . . . . . . . Laboratory N o . C34006I (Atomospheric)i

State of New Jersey Water/wastewater, LaboratoryID No. 03117

AIHA - Metals, solvents, asbestos inair Laboratory ID No. (Asb) PATLaboratory No. 08003-001

. State of New York Water, wastewater, solidwaste,i air (pending)

Client Audits Various

ftRSOIOH

Page 196: ftR3008L6 - semspub.epa.gov

9.2 Internal Performance Evaluations

Representative analytical parameters are evaluated on aquarterly basis. Samples of known concentration ofvarious analytes (metals, organics, inorganics andphysical parameters) are made by the assistant laboratorymanager from analytes from EPA's Environmental Monitoringand Support Laboratory (EMSL). Compliance with respectto recovery and precision data is monitored.

\

flR3'OIOI2

Page 197: ftR3008L6 - semspub.epa.gov

10.0 Procedures Used to Access Data Precision and Accuracy

KAC uses specific routine procedures to assess the precisionand accuracy of its analytical data. These measures includethe validation and internal quality control proceduresdiscussed in section 8 and 9.

Specific procedures for assessing data accuracy and precisioninclude calculation of percent recoveries and relativepercent differences for all duplicate spike sample analyses.These calculations are summarized below, and apply to allanalytical parameters.

1. Accuracy = Percent Recovery = Observed Cone, x 100(R) Expected Cone.

2. Precision =- Relative Percent Difference = /C1-C2/ x 100(RPD) C1+C2

2

Where Cl and C2 are concentrations of duplicate spikes.

Analytical control limits are derived from statisticalmanipulation of each data category.. These limits areoutlined below.

Accuracy Precision

Upper Control Limit (UCL) ~" %R_ + 3S RPD + 3S

Lower Control Limit (LCL) %R - 3S RPD - 3S

. Where S is Standard Deviation

For each analytical... batch, the percent recovery RPD aremonitored and _documented in.a tabular ..and/or control chartformat. .Periodically, when not less than twenty such datapoints are obtained for,a. given analysis, the units for RPDand recovery are recalculated and the new clients areemployed to evaluate future data. The quality control chartsand tabular summaries will provide a continuous indicationof.method performance by visib comparing actual qualitycontrol recovery data to acceptable USEPA (if available)and/or laboratory method performance criteria.

0R30IGI3

Page 198: ftR3008L6 - semspub.epa.gov

10.0 Procedures Used to Access Data Precision and Accuracy

Laboratory method performance criteria are utilized toevaluate quality control data in the absence of publishedUSEPA __-_met_ho_d__performance criteria. Laboratory methodperformance criteria are statistically derived from duplicatematrix spike quality control data as follows:

Warning Limit (WL) = ±2sControl Limit (CL) «= ±3s

s = standard deviation of the method percent recoveries

A minimum of six (6) matrix spike sample percent recoverydeterminations are required to establish the above-statedperformance limits. These Laboratory method performancelimits are continually redefined, as quality controlinformation accumulates.

The mean value percent recovery (R) and standard deviation (s)determinations calculated from the matrix spike qualitycontrol results are used to generate Laboratory MethodAccuracy Statements for each analytical method commonly usedin the Laboratory. Method Accuracy Statements are definedas R ± 3s and are based upon a minimum of six (6) matrixspike sample determinations. These statements are updatedannually as calculated from quality control data accumulatedduring the previous year. All laboratory Method AccuracyStatements are maintained in the QC manual and should bewithin the recommended EPA criteria.

AR30IOU

Page 199: ftR3008L6 - semspub.epa.gov

11.0 Corrective Action —--

NAC routinely utilizes its quality control data to assess theneed for corrective action. Frequency review of data permitsrapid identification of the._source of analytical or samplingerror and implementation of corrective action.

11.1 Determination of the Need for Correction Action

Percent recovery determinations from the systematicmatrix spike quality control samples must comparefavorably to the published USEPA or laboratory methodperformance criteria outlined previously in order tovalidate and approve a corresponding batch of sampleanalyses. The method analyses are out of control and

( therefore unacceptable if:

1. One data point recovery value is outside of"i """ published USEPA performance limits or laboratory

control limits (CL).

2. Seven consecutive percent recovery values are on oneside of R line.

i 3. Percent recovery values increase or decrease overCJ - -- --- -—five consecutive analyses.

Unacceptable values will render the corresponding batchof sample analyses suspect, as will unacceptableresults for method blank analyses, until corrective

_T-_action demonstrates the return of acceptable methodperformance.

The analyst who reviews and complies raw data fromanalyses and associated matrix blank and matrix spikesamples must immediately notify the of manager ofdeviation from accepted standards. If any values areoutside of QC limits, corrective action will beinstituted at once.

Corrective action may also be implemented as a resultof .external . performance and systems audits,intralaboratory comparisons, QA project auditsconducted by concerned agencies, or other QA/QCactivities.

These corrective actions may involve any phase of theanalytical or: sampling method including: reagentquality, sample extraction, equipment cleaning,instrument calibration and/or performance,calculations, etc. •.**_._._..AR30IOI5

Page 200: ftR3008L6 - semspub.epa.gov

12.0

The following Table .12 is a current list of analyticalinstrumentation at the headquarters laboratory.

TABLE 12. ANALYTICAL -INSTRUMENTATION

OTY INSTRUMENTATION ..... KRNUFACT0RER

2 Gas Chromatograph/ Hewlett-PackardMass Spectrometer 5890 GC with

5970 MSD

6 Gas Chromatographs Varian 3700 ECD/FID/TCDTracor 560 PID/HECD

_ -Carle 311 FID/TCDTracor540 ECD/ECDTracor 540 PID/HECDHewlett-Packard 5980 FID

1 Chromatography Data System Nelson Analytical

4 Liquid Sample TEKMAR LSC--2Concentrators TEKMAR LSC-2

TEKMAR LSC-.2TEKMAR-LSC-2000

4 Integrator/plotters. 2 Hewlett-Packard 33902 Hewlett-Packard 3393

4 Autosamplers for Organic Hewlett-PackardAnalyses Y" . TEKMAR ALS-10

TEKMAR ALS 2016• Dynatech

1 TOX Analyzer .. : _ . . Dohrmann _MC3

1 TOC Analyzer ~" 01 Corp, Model 700with Solids Option

1 IR Spectrophotometer Perkin Elmer 1310

2 Atomic Absorption Perkin Elmer Model 5000Spectrophotometers withAutosamplers

1 Mercury Analyzer SpectroProducts HG-4

Various ovens, balances, pH meters, mV meters, hot plates, hoods,stir plates, etc. AR30IOI6

Page 201: ftR3008L6 - semspub.epa.gov

13.0 Resumes of Key Laboratory Personnel

&R3QIGI7

Page 202: ftR3008L6 - semspub.epa.gov

JOHN H. RISSELLABORATORY MANAGERSTATEMENT OF QUALIFICATIONS AND EXPERIENCE

EXPERIENCE

. Laboratory Manager for Northeastern AnalyticalCorporation (MAC), Mr. Rissel is responsible for theoverall, technical control of the Environmental LaboratoryDivision. Mr. Rissel supervises twenty-five (25)chemists, analysts and technicians in the routineanalysis of various matrices for contaminants ofinterest. He oversees the laboratory physical plant,including:

- Sample Reception/LoginInfr~a~-red Laboratory

- Gas Chromatography Laboratoryj _ wet Chemistry Laboratory! - Atomic Absorption Laboratory

- Bacteriologic Laboratoryf - Organic Preparation Laboratory1 - GC/MS LaboratoryC.; . . . . . . . _ . . _ _ .

Mr. Rissel has a diverse background with over • 13 yearsexperience in the conduct of environmental analyses invarious analytical matrices. These matrices include potablewater, wastewater, soil, sludge, and air pollution samples.Mr. Rissel has an "extensive working knowledge of EPAmethodologies including their requisite qualitycontrol/quality assurance requirements. He is familiar withEPA 500 and 600 series protocols for organics in water, aswell as EPA SW846 methods for solid matrices and disposalcriteria. He is familiar with the National Institute of

— Occupational Safety & Health analytical methods for airanalysis and also with various source analysis techniques.

Mr. Rissel's staff conducts the analysis proficiencysamples, submitted by the NJDEP, NIOSH, PADER & NYDOH, aswell as other regulatory agencies.

Laboratory Manager of Rossnagel & Associates. Mr. Risselconducted environmental-analyses of soil, water and sludgesusing standard methods. He supported the stack testinggroup by conducting analyses for specific air contaminantssuch as oxides of sulfur, oxides of nitrogen and varioussolvents. He supervised a laboratory staff and insured thequality pf_ results though strict adherence ,-jtp. .quaJLitycontrol- procedures and EPA quality assurancfl."_r«ittjr«iile.3ts.

.. -* * *—

RR3QIQ18

Page 203: ftR3008L6 - semspub.epa.gov

JOHN M. RISSELpage 2

EDUCATION

B.S. in Chemistry, Ursinus College, Pennsylvania

PROFESSIONAL AFFILIATIONS

American Chemical SocietyAmerican Water Works Association

AR30I019

Page 204: ftR3008L6 - semspub.epa.gov

PAUL P. PAINTER III . . . . . .ASSISTANT LABORATORY MANAGER AND CHEMISTSTATEMENT OF QUALIFICATIONS AND EXPERIENCE

EXPERIENCE ____._:__ ... ._._. . . . . . . . „- .

Assistant Laboratory Manager for Northeastern AnalyticalCorporation (NAC). Mr. Painter supervises the dailyoperations of the Environmental Laboratory Division,which includes the GC, GC/MS, Atomic Absorption and WetChemistry Laboratories. This encompasses diversifiedactivities including many managerial functions such asdelegating work assignments, sample tracking and custody,test report generation and the monitoring of laboratorysales. He performs trouble shooting of instruments andanalysis as well as the training of technicians. Mr.Painter frequently interacts with clients, explainingmethods of analysis, and interpreting results. He hasan extensive working knowledge of all USEPA protocols andNew Jersey Tier I & II formats. He is responsible fororganizing the quarterly Proficiency Analytical Testingprogram. ..... . ... ^

Atomic Absorption Technician at Century EnvironmentalTesting Labs, Inc. Mr. Painterfs laboratory dutiesincluded metal analysis by atomic .absorption and graphitefurnace using Perkin Elmer Models 303, 305, 360, 403, and5000 w/400 and 500 furnaces, and Baird Spectro-vac ICP.He also supervised a staff of 2employees in the daily operations of the metalsdepartment which included preparations, quality control,and reporting of data. He.supervised theMicroscopy Laboratory,and was responsible for itscomplete set-up, and obtaining certification in NewJersey and Pennsylvania. He supervised the dailyoperations.

EDUCATION

B.A. in Biological Science, Glassboro State College, NJA.A. in Liberal Arts, Muscatine Community College, IowaA.S. in Conservation, Muscatine Community College, IowaPerkin-Elmer Atomic Spectroscopy Training Course, Conn

PROFESSIONAL AFFILIATIONS

American Chemical Society

AR30I020

o r i r "

Page 205: ftR3008L6 - semspub.epa.gov

KEN CHAMBERSGAS CHROHOTOGRAPHY SUPERVISORSTATEMENT OF QUALIFICATIONS AND EXPERIENCE

EXPERIENCE

EDUCATION

Supervisor of Gas Chromatography (GC) Laboratory forNortheastern Analytical Corporation (NAC). Mr. Chambershas over 5 years experience operating the GasChromatography Laboratory and directing quality controlfor the Organic Preparations Laboratory. Mr. Chamberssupervises four (4) analysts in the daily operations ofGC analysis. Included in his scope of work is the reviewof analytical data and compliance with quality assuranceregulations. Mr. Chambers' capabilities include aworking knowledge of the analytical procedures of theUSEPA the National Institute of Occupational Safety &Health, and other governing agencies for the analysis oforganic contaminants of environmental matrices. He isfamiliar with all. preservation,_extraction and analysisprotocols for organic contaminants. Mr Chambers staffconducts the proficiency testing of organic samplessubmitted by New Jersey Department of EnvironmentalProtection in accordance with the Safe Drinking Water Actand the Clean Water Act, and various regulatory andcertifying agencies.

Laboratory Manager of Environmental Measurements andAnalysis, Inc. Mr. Chambers was responsible for theoperations of the entire laboratory division. Thesedivisions consisted of atomic absorption, gasChromatography and wet chemistry. He supervised seven(7) technicians and reviewed all analytical data andfinal reports. He continually assured that qualitycontrol, was being enforced, focusing on all aspects oflaboratory operations. Mr. Chambers is knowledgeableregarding USEPA protocols and is experienced in allphases of environmental laboratory analysis.

B.S. in Chemistry, Stockton State College, Pomona, NJ.Post Graduate Studies in Capillary Chromatography,American Chemical Society.Post Graduate Studies in Chromatography, WidnerUniversity.

PROFESSIONAL AFFILIATIONS

American Chemical Society AR3Q I 02 IChromatography Forum of the Delaware Valley c 'South Carolina Laboratory Management Society

Page 206: ftR3008L6 - semspub.epa.gov

RICHARD P. BARLEYGAS CHROMATOGRAPHY/KASS SPECTROKETRYSUPERVISORDIRECTOR OF QUALIFICATIONS AND EXPERIENCE

EXPERIENCE

EDUCATION

Gas Chromatography/Mass Spectronetry (GC/KB) Supervisorfor Northeastern Analytical Corporation (NAC). Mr.Earley is responsible for supervising analysis andquality control for the GC/MS department. He coordinatesall activities between the GC/MS and the organicextraction laboratories. He has extensive experiencewith GC/MS methodologies as regulated by the USEPA,including USEPA Methods 624, and 625, in addition tovarious methods documented in SW846, all of which dealwith volatile and semi-volatile compounds in water,wastewater and non-aqueous matrices. Mr. Earley hasextensive knowledge of New Jersey Tier I and Tier IIreporting formats and deliverables requirements, alongwith the USEPA Contract Lab Program (CLP) qualitycontrol/quality assurance protocols. He is experiencedalso with the organic preparations required for thesemethods.

Gas Chromatography Mass Spectrometry Chemist for NationalEnvironmental Testing, Inc. Mr. Earleyperformed environmental- priority pollutant analysisusing various USEPA methods, including Method 624 and625. Conducted analysis of proficiency samples for theNew Jersey Department of: Environmental Protection andthe State of New York. Mr. Earley was responsible forthe hardware maintenance and system trouble shooting forfour Hewlett-Packard Mass Spectrometers.

B.A. in,Biology, Mansfield State University,Pennsylvania. .GC/MS 5988/96 RTE Operator Course, offered by Hewlett -Packard.GC/MS Techniques and Interpretation Course, offered byHewlett -Packard.Purge and Trap Techniques Workshop, offered by ANKHEMLaboratory.

flR30IQ22

Page 207: ftR3008L6 - semspub.epa.gov

RICHARD C. CLEAVERATOMIC ABSORPTION LABORATORY SUPERVISORSTATEMENT OF QUALIFICATIONS AND EXPERIENCE

EXPERIENCE ..... ._.___.. . . .: . .: . .:_.

EDUCATIO_N

Atomic Absorption Laboratory Supervisor for NortheasternAnalytical Corporation (NAC). Mr. Cleaver is responsiblefor the operative functions of all metals analysis. Hesupervises four (4) technicians in the Atomic AbsorptionDepartment. He is skilled in various metal analysisprotocols. Mr. Cleaver conducts the mandatoryproficiency testing of trace metal samples for the NewJersey Department of Environmental Protection, and alsothe National Institute of Occupational Safety & Health.Mr. Cleaver's area of proficiency include:

Flame Atomic Absorption Spectroscopy.

- Furnace Atomic Absorption Spectroscopy.

- Cold Vapor Atomic Absorption Spectroscopy.

- Colorimetric Metals Analysis usingUltraviolet Spectroscopy.

Laboratory Supervisor for Environmental Measurements andAnalysis. Mr. Cleaver supervised a laboratorystaff of six (6) technicians in the limited chemistrydepartment. His duties included the supervision of thedaily operations of environmental analysis of soil, waterand sludge samples using standard methods. He was alsoresponsible for all trace metal analysis using AtomicAbsorption Spectroscopy. Mr. Cleaver wasdirectly involved in a project designed to computerizethe analysis QA/QC program. As a Laboratory Supervisor,Mr. Cleaver_was responsible for Project Management, whichincluded client interaction, and problem solving.

B.A. in Biology, 1984, Glassboro State College, Glassboro,N J . - — ' . _ . _ . .Post Graduate Studies: American Chemical.Society, Atomic- Absorption Spectroscopy.

PROFESSIONAL AFFILIATIONS _ ; __ _

South Carolina Laboratory Management Societ

Page 208: ftR3008L6 - semspub.epa.gov

Northeastern Analytical CorporationAddendum A

Sample Data Report for a non-CLP Analysis

Page 209: ftR3008L6 - semspub.epa.gov

02/15/89 16:17 -fyi_609~S85."9700 ' " N. E. ANALYTICAL -— CQMPUCHEM LABS @002

^^ Ixll _ II Northeastern Analytical Corp.• i 1

. . . _ . . - -.,._.,_...-_.._._,_...__...^..ANALYTICAL REPORT

for

COMPUCHEM LABORATORIESP.O. BOX 12652

3308 Chapel Hill/Nelson Hwy.Research Triangle Park, North Carolina 27709

Attention: MS. Angela Childress, Manager Planning and control

TEST REPORT NO. NAC89L-0095

Project No; 69-11

Client NAC DateDesignation Designation . Received Matrix

241220 89L-0095-1 1-19-89 Aqueous241222 89L-0095-2 1-19-89 Aqueous241225 39L-Q095-3 1-19-89 Aqueous241227 S9L-0095-4 1-19-B9 Aqueous241231 — 89L-0095-5 1-19-89 , Aqueous

Laboratory Name: Northeastern Analytical Corp.

Certification No: 03117*

Name: Jdhn M.

Title: ____ Laboratory Manaaer.

Date: January 31. .1989 ——

fl/R30!025

A ns lysis, Sampling and Testing for ihe Environmental sno a-BOOO PW< 1609) 985-970Q

Page 210: ftR3008L6 - semspub.epa.gov

02/15/S9 6H7 ©1 609 ..9.aa_-9700 ..____. :K, E-'\ ANALYTICAL —- COMPUCHEM LABS

NorLne.tSl.ernMOPCompuchem LaboratoriesTest Report No. 89L-0095January 31, 1989Page 2 of 9

Table of contents

I. Sampling Information . . . . . . - • • • • • • • 3

II* Chain of Custody Documentation . . . . . • • » * 4

III * Methodology. . . . . . . . . . . . - - - • • • • 6

IV. Laboratory Chronicle . . . . . . . . . * • • * • 7

V. Non-COTCpliance/QA Report . . . . . . . . . - • • 7

VI. Analytical Results . . . • . * . . . . . . - • • 6

VII. Quality Assurance Data . . . . . - • • • • • • • 9

1mFile: 9L\COMPUCHM\89L-0095

AR30I026

Page 211: ftR3008L6 - semspub.epa.gov

02725/89 16: 17 ' ©1 .603 985 "9700 ~':' " ' " N- E. ANALYTICAL ™ COMPUCHEM LABS ©<XK

^V| 1 _J I _" 1

_ Mortress-em Ana VL cai C.u ' i;

Coatpuchem LaboratoriesTest Report No. 69L-0095January 31, 1989Page 3 of 9

I. SAMPLING INFORMATION

Not submitted.

flR30!027

Page 212: ftR3008L6 - semspub.epa.gov

02/15/89 Y16T;_8Y~ '_J&\ 609" 985 970CT "~" N. E. ANALYTICAL —— COMPUCHEM LABS ©006

Ol

s

8

O nv »c>

\\

\

c\a

^

\

\

\

=T

\

rS

\\\

* U.

\ 1. ->i ———^ ^>» ^ ^ H

O

1Ii

i §

i

«£>

4

ifM.

1

J

Page 213: ftR3008L6 - semspub.epa.gov

02/-15/gHY;U6:18. . _©1 609 S.85" 3700" "X L~ANALYT"ICAL -""COMPUCHEM LABS U005.

MOP_ Northeastern Arslytica-' Uu1 1;

Compuchem LaboratoriesTest Report No. 89L-0095January 31, 1989Page 4 of 9

II. CHAIN OF CUSTODY DOCUMENTATION

1029

Page 214: ftR3008L6 - semspub.epa.gov

02/15/89 -16:19"^— ©1 609"_ 985 9700 YY.YY. N. E- ANALYTICAL —— COMPUCHEM LABS ©007

_ Nonneastern Ars,v:,cai Corp

Compuchem LaboratoriesTest Report No. 89L-0095January 31, 1989Page 6 of 9

METHODOLOGY

The samples were analyzed using methods as adopted from:

EPA Methods for the Chemical Analysis of Water andHastes, March, 1979.

Standard Methods for the Examination of, Water and_ Wastewater, 15th and/or 16th Edition.

SR30I

Page 215: ftR3008L6 - semspub.epa.gov

02/15/89 ""16:19 - ©1 609 '985 9700 ' "Y _. ..LTTEr ANALYTICAL —— COMPUCHEK LABS ©008 ~

NPC~ Northeastern Ano i ytioal C

Compuchem LaboratoriesTest Report Ho. 89L-0095January 31, 1989Page 7 of 9

IV. LABORATORY CHRONICLE

A- Date of Sampling: 1-17-89

B- Date of Receipt/Refrigeration: 1-19-89

C* Date of Analysis!

Parameter Date Analyzed

Total Organic Carbon 1-25-89Total organic Kalides 1-23-89Chromium Hexavalent 1-20-89Aaaonia-N 1-24-B9Biochemical Oxygen Demand 1-20-89Chemical Oxygen Demand 1-24-89Chloride 1-31-S9Nitrate-N 1-20-89Total Dissolved Solids 1-19-89Sulfate 1-23-89

NON-COHPLIANCE/QA REPORT

1) Hold time for hexavalent chromium was exceeded upon._ ,, receipt of samples.2) Samples for TOG, TOX, Ammonia-N and COD analyses were

received unpreserved.3) Samples for TOX analysis-were not received head space

free.

Supervisor Review and Approvais IV- lp \ fl R 30 I 0 3

Page 216: ftR3008L6 - semspub.epa.gov

02 15/89 3,6:20 "©1 609 965 .9700 N. E ANALYTICAL —— COMPUCHEM LABS- =' ©009

_ Northeastern Analytics! Corp

Compucheza LaboratoriesTest Report No. 89L-0095January 31, 1989Page 8 of 9

VI - ANALYTICAL RESULTS ........ _„_

Designation

S9L-O095-1 89L-0095-2 89L-0095-3Parameter ..__... _ 241220 241222 241225

Total Organic Carbon <l el <iTotal organic Halides 0.061 0.11 <0.0lChromium, hexavalent <0.05 <0.05 <0.05Aromonia-N 1.6 <l <iBiochemical Oxygen Demand,5 day total <4 <4 <4Chemical oxygen Demand <50 <50 <soChloride ,<1 6.7 <1Nitrate-K 1.4 2.8 <1Total Dissolved Solids 35 44 27Sulfdte 2.0 2.0 1.2

Units (mg/1) (mg/1) (ag/1)

e Designation

89L-0095-4 89L-009S-5.Paranetcr 241227 241231

Total Organic Carbon 2.8 <lTotal Organic Halides <0.01 <0.01Chromium, hexavalent <0.05 <0.05Ammonia-N <1 <1Biochemical Oxygen Demand,5 day total <4 <4Chemical Oxygen Demand 89 <50Chloride 6.7 <1Nitrate-N 1,8 <1Total Dissolved Solids 43 8.9Sulfate 1.2 1.2

Units (mg/1) (»«f/l)

Page 217: ftR3008L6 - semspub.epa.gov

02/15/89 -16:20: ' 1£1 609 _9S5 97.00 ;- -; --N. E,

„ - - - - - , - Corpa

S -ESS

Compuchem LaboratoriesTest Report No. 89L-0095January 31, 19893?aga 9 of 9

VII. QUALITY ASSURANCE DATA

Matrix_SpiKe and Matrix Spike Duplicate Recoveries

Sample Amount of Initial * DuplicateParameter Spiked Spike, ucr Recovery, j Recovery REP

TOC 0095-1 1,000 103 101 2.0TOX 0°95-4 10 91 . 110 19Chromium (Hex) control 25 93 —— - - -Aamonia-N 0060-6 5,000 94 93 1.1BOD 0096 1,000 57 61 6.8COD 0072-1 500 117 115 1.7Chloride 0095-2 5,000 107 107 0Nitrate-N 0067-10 5.0 97 96 1.0TDS 0082-3 Duplicate 100 ——Sulfate 0095-1 1f 000 129 129 0

Page 218: ftR3008L6 - semspub.epa.gov

Subcontractor: Chemwesf Laboratories . . . .-~ . ' .._:6'00'W. "North"Market "BlVd.Sacramento, California 95554 .^.. ','.'..".--- . :. -

Name/Date of last On-SIte- Evaluation: Bob MeTerer,' October^24, 1988

Date Approved: November, 1987 ' '" ""' """•"" "= = ;--;-—=•- - .

Certified Parameters: ...".. ™"7":!7" .". >^:_ T.. ,..'.. :'".:... t~:~"~" i~

Ammonia -- "total Pnqsphorus_ " .: :Chloride-"- -- :;:r.-.-" ..'_ ._" -Re&ct'fvlty' '"" '-'"--"•" ~"" """'Conductivity " Tofaf'pTssol ved SoHds ' ^Corros i v i ty """ -"'" Tot a r"Tus4pended SoTfds "" "*" V.::Cyanide.".:. -"" -.7-.-^ -J—l -: "Su'l fate _""_;•"•' -'• ""..."•".-: .: ""/ • jFiammability Tota i Petroleum HydrocarbonsFluoride . -.-. _ _ _ _ _ _ 0 i l & Grease^"" -- - " *" "-- - '-"Nitrates - -.- ~_Phenols .-"- *.~.' "~:- ~~%- Chlorine-Ph DioxinsCOD ~ •" . . . . . -..- ...._... : .Hardness""""-- "v: ~ TT— r— _ -_.s ^===--_-~;~; :Alkalinfty " . •-—•-.-.- .. -— —

Page 219: ftR3008L6 - semspub.epa.gov

Subcontractor .Name: Northeastern" Analytical Corporation ( NAC)734 Route "70 . "" " ' - - . -Medford. NJ OS055 = -

Name/Date~\ of On-Site Evaluation: Bob Meierer, January 30, 1987Bob Uhitehead. July 8. 1987

Da-te(s) Blind PE.Analyzed: March 2, 1387. flpril 20. 1S87. May 1. 1987

Date Apprdvco: June 8. 1987 " "

Certified Parameters: " ~ : " ."' "

Calcium ' Aluminum - SodiumMagnesium Antimony . . . ThalliumSodium " "" -~"- - — ----- —- - -ftrsenj_c ^ UenadiumPotassium Barium " " Zinc .Alkalinity . BerylliumSulfate. _... _." .._".. ."._._.___.._ __.±".~ CadmiumChloride . ... ___-.,.!-. J .. 111—1-. Chromium " "~"~"Total Dissloved Solids . . . CopperHardness .... ..__..==_..,. ,._.._— .. ,; - IronConduct ivity..__ .. LeadBiochemical Oxygen Demand ManganeseChemical Oxygen Demand : " MercuryTotal Phosphorus -——— - --- f^i^kel . _ "_..""Total Kje-lda.hl Nitrogen. . . _ _ .SeleniumOil & Grease -.. !____.....__._.."" _ ! - _ .SilverTSS" - " --- -"--"..:^^T^:==-.^-=r';. -»--——-_.— -^-~: : -:- ••

flR30!035

Page 220: ftR3008L6 - semspub.epa.gov

DAMES &MDORE71C1 WISCONSIN AVENUE. SLTTt 73?." BETHE;:rvC M.--R"V^AN? J .-i-iv :

June 19, 1989

M r . Randy Sturgeon ' - - - - - - -Remedial Enforcement Section (3HW16) ; :. JIIN'2 3U.S. Environmental Protection AgencyRegion III. ."- --._..-- : .:" " ---'." --~ - ..--,-. . — - - , ,., ., r..t841 Chestnut Building - - - - - fe^ U*.*;•Philadelphia, PA 19107 """ V-/..

Re: Modifications toHentokil QAPiP

Dear ,.Mr.' Sturgeon;

This letter is related'to "changes.which must be. made to theQuality Assurance Project Plan (QAPjP).. as -currently written. Thechanges are " necessitated: as "a. "result o'f corrections requested byour laboratory subcontractor. CompuChem Laboratories to informa-tion they supplied us which was subsequently discovered to be inerror. The changes are. minox and will have no impact on thescope of work .or quality assurance objectives for the project.

The., .following .required . changes should' be made to Table 8-2found on page .8-3 of "the QAPjPt "

o The TOC -analysis will be .performed by 'NortheasternAnalytical Corporation (NAG) rather than CompuChemLaboratories;

o The BOD analysis will be performed following EPA method#405.1, which is equivalent to SM #507 and for which anSOP "was provided in Appendix D.;._.

o The COD analysis will be performed following EPA method#410.4, the SOP of which was provided in Appendix D,rather than #410.1;

o The"TKN analysis-will be performed.following EPA method#351.3" ra'ther than #351". 2, and a replacement SOP isenclosed;

o The CLP-Organics SOW was."last revised on 2/88 and thismethodology will be followed;_and

o The CLP-Indrganics SOW was last revised on 7/88 andthis, methodology will be followed. a o r» n i

AR30 I 036

Page 221: ftR3008L6 - semspub.epa.gov

DAMES & MOOREThe fdll-owingr \change should be made to Section 6.3, which is

found on page 6-5: ,-.-- . - . ..- - - : : -

o The sentence "Samples will be-shipped directly from thefield to the laboratory performing the analysis inorder to minimize shipping times . and, thus, minimizethe chance of holding times being exceeded" should bechanged to read "Samples will be shipped from the fieldto CompuChem Laboratories where they will receive aCompuChem. identification number for sample trackingpurposes. Samples to be analyzed, by one of CompuChem'ssubcontracting laboratories (i.e., ChemWest AnalyticalLaboratories or Northeastern Analytical Corporation)will in general be repackaged and shipped to thesubcontractor laboratory by .CompuChem after havingreceived .CompuChem1s identification number. The onlyexception will "be samples for. analyses that havesensitive holding times (e.g., BOD, total phosphorus),which will be ." shipped..:directly to the laboratoryperforming :the .analysis so that•holding times are notexceeded."

Amended copies of page 6-5 and Table 8-2 containing thesechanges are enclosed for your review and the project record.

Dames & Moore assumes that EPA-will find these modificationsacceptable, since '_, they._ do not materially affect the datacollection eff6r~t"r"37Ple:ase__ca,ll. if this assumption is incorrect,or if .you "have questions or comments.

Sincerely,

DAMES & MOORE .

"John 0. OsgoodProject Manager

/cmm ........_.— .—_

cc: J. - PrzybylinskiD. Anderson

AR30IQ37

Page 222: ftR3008L6 - semspub.epa.gov

TABLE 8-2

Analytical Methods

LaboratoryPerforming

_____ Analytical Group " Methoda _______ Analysis

TCL volatiles : CLP --Organics SOW CompuChemTCL semivolatiies CLP - Organics SOW CompuChemTCL pesticides/PCBs -CLP - Organics SOW CompuChemTAL metals including, arsenic, CLP - Inorganics SOW CompuChemchromium, copper, and zinc — -- '-'-" " "~ -~Dioxins and furans EPA-SW-S46 - Method S2SO, ChemWest

Isomer SpecificNAC

BOD:--.. .,.._;„, ... -= -='=-EPA-600 05.1 NACCOD ". . - .'. <~~ — - --—— -:EPA-600 "#41 0.4" ChemWest

TKN EPA-600 #351,3 - NACTSS" ", _:::::::~r ":_—-:- :;- "-'-"T^pA-6o6"~#"i6d.2~ " NACTDS :- . .'."—" — -" .-'-"- ----- " TPA-60oTl60.I NAC

Hardness - -"- : --- - --v '^" ' "" -EPA-600 V/130.2 . ChemWestPhosphorus, Total " "EPA-600 #365.2, Procedure NAC

(00665) " . " . . 'Alkalinity " '" "'- """ " " " " "~ HPA-600 #310.1. ChemWestThiocyanate ."..,_.:T ". „. -J/ /.". -, "SM" ?/il2K" ' " " " " ChemWest

References: . "..:"" . - ": ;. "• ""-"" - - —-1. CLP SOW"- Organics ; -- - - :

EPA Contract Laboratory Program. Statement of Work for Organics Analysis,Multi-Media, (Multi-Concentration. 10/86. Revised 1/87, 2/87, 7/87, 8/87,

" 2/88". " - - - -

2. CLP SOW - Inorganics" . . . - . - . . - . -EPA Contract Laboratory Program. Statement of Work for InorganicsAnalysis, Multi-Media, Multi-Concentration. SOW No. 787. Revised 7/SS.

3. EPA-SW-S46 - Test .Methods for Evaluating Solid Waste, 3rd edition, Np^ i1986. .... - : ;:- ;:--:- flRSU I

4. EPA 600/4-79-200 - Methods for Chemical Analysis of Water and. "TC ',, :March 1979. ' '" "

5. SM "= Standard Methods for the Examination'of Water and Wedition. Approved by EPA in 40 CFR 136.3, Table IB.

8-3.

Page 223: ftR3008L6 - semspub.epa.gov

When the laboratory receives" the samples, the chain-of-custody forms will besigned and the date and time recorded by the laboratory sample custodian. Thesample custodian will ,.immediately inspect the shipment for damage andcompleteness, arid report any problems to the Dames & Moore sample shipper. Thelaboratory~sample custodian will then complete all the appropriate lab trackingsheets and logs.

6.3 LABORATORIES : . ,- ........

The laboratory sample custodian is responsible for the custody of samplesfrom the time of sample receipt to the time of sample discard. The contractedanalytical laboratories have established SOPs for sample handling, storage, anddispersemeht for analysis. These are detailed in laboratory SOPs and includeexamples of the Sample Receiving and Tracking Information Sheet, Sample LoggingSheets, and Laboratory Assignment Sheet, etc. Samples will be shipped from thefield to CompuChem Laboratories where they wilj receive a CompuChemidentification number for sample tracking purposes. Samples to be analyzed by one.of Compu'Chem's subcontracting laboratories (I.e., ChemWest AnalyticalLaboratoriesiMsnJSIp_rtheastern Analytical Corporation) will in general be repackagedand shipped to the subcontractor laboratory by CompuChem after having receivedCompuChem's identification number. The only exception will be samples foranalyses that have sensitive holding times (e.g., BOD, total phosphorus), which willbe shipped directly to the laboratory preforming the analysis so that holding timesare not exceeded. The chain-of-custody procedures employed by the laboratoriesare detailed in their respective Quality Assurance Plans.

0R30I039

6-5

Page 224: ftR3008L6 - semspub.epa.gov

6. Apparatus6.1 Digestion apparatus; A Kjeidahl digestion apparatus with 800 or 100 ml flasks and

suction takeoffto remove SO3 fumes and water.6.2 Distillation apparatus: The macro Kjeidahl flask is connected to a condenser and an

adaptor so that the distiUate can be collected!. Micro Kjeidahl steam distillationapparatus is commercially available.

6.3 Spectrophotometer for use at 400 to 425 nm with a light path of 1 cm or longer.7. Reagents

7.1 Distilled water should be free of ammonia. Such water is best prepared by the passage ofdistilled water through an ion exchange column containing a strongly acidic cationexchange resin mixed with a strongly basic anion exchange resin. Regeneration of thecolumn should be carried out according to the manufacturer's instructions.NOTE 1: Absolutions must be made with ammonia-free water.

7,2 Mercuric sulfate solution: Dissolve 8 g red mercuric oxide (HgO) in 50 ml of 1:4 sulfuricacid (10.0 ml cone. H2SG< : 40 ml distilled water) and dilute to 100 ml with distilledwater.. . . _ . : _ . . - - - - - - - - - - - — — " _'__:_ _ " "

7.3- Sulfuric acid-mercuric sulfate-potassium sulfate solution: Dissolve 267 g K2S04 in 1300ml distilled water and 400 ml cone. H2SO4. Add 50 ml mercuric sulfate solution (7.2) anddilute to 2 liters with distilled water.

7.4 Sodium hydroxide-sodium thiosulfate solution: Dissolve 500 g NaOH and 25 gNa2S2O3*5H2O in distilled water and dilute to 1 liter.

7.5 Mixed indicator: Mix 2 volumes of 0.2% methyl red in 95% ethanol with 1 volume of0.2% methylene blue in ethanol. Prepare fresh every 30 days.

7.6 Boric acid solution: Dissolve 20 g boric acid, H3BO3, in water and dilute to 1 liter withdistilled water, . ......_. _ __ '

7.7 Sulfuric acid, standard solution: (0.02 N) I" ml = 0.28 mg NH3-N. Prepare a stocksolution of approximately 0.1 N acid by diluting 3 ml of cone. H2SO4 (sp. gr. 1.84) to 1liter with CO2-free distilled water. Dilute 200 ml of this solution to 1 liter with CO2-freedistilled water. Standardize the approximately 0.02 N acid so prepared against 0.0200 NNa:COj solution. This last solution is prepared by dissolving 1.060 g anhydrous Na:CO3,oven-dried at 140°C, and diluting to 1 liter with CO2-free distilled water.NOTE 2: An alternate and perhaps preferable method is to standardize theapproximately 0.1 N H23O4 solution against a 0.100 N Na2CO3 solution. By properdilution the 0.02 N acid can the be prepared.

7.8 Ammonium chloride, stock solution: 1.0 ml = 1.0 mg NH3-N. Dissolve 3.819 g NH.C1in water and make up to 1 liter in a volumetric flask with distilled water.

7.9 Ammonium chloride, standard solution: 1.0ml = 0.01 mg NH3-N. Dilute 10.0 ml of thestock solution (7.8) with distilled water to 1 liter in a volumetric flask.

7.10 Nessler reagent: Dissolve 100 g of mercuric iodide and 70 g potassium iodide in a smallvolume of distilled water. Add this mixture slowly, with stirring, to a cooled solution of160 g of NaOH in 500 ml of distilled water. Dilute the mixture to 1 liter. The solution isstable for at least one year if stored in a pyrex bottle out of direct sunlight.

351.3-2

Page 225: ftR3008L6 - semspub.epa.gov

NOTE 3: Reagents 7.7, 7.8, 7.9, and 7.10 are identical to reagents 6.8, 6.2, 6.3, and 6.6described under Nitrogen, Ammonia (Colorimetric; Titrimetric; Potentiometric-Distillation Procedure, Method 350.2).

8. Procedure8.1 The distillation apparatus should be pre-steamed before use by distilling a 1:1 mixture of

distilled water and sodium hydroxide-sodium ttiosulfate solution (7.4) until the distillateis ammonia-free. This operation should be repeated each time the apparatus is out ofservice long enough to accumulate ammonia (usually 4 hours or more).

8.2 Macro Kjeidahl system8.2.1 Place a measured sample or the residue from the distillation in the ammonia

determination (for Organic Kjeidahl only) into an 800 ml Kjeidahl flask. Thesample size can be determined from the following table:

Kjeidahl Nitrogen Sample Sizein Sample, mg/1 ml

0-5 5005-10 " 25010-20. . .-.. . 10020-50 50.050-500 .; -- -,- =.- - - - 25.0

Dilute the sample, if required, to 500 ml with distilled water, and add 100 mlsulfuric acid-mercuric sulfate-potassium sulfate solution (7.3). Evaporate themixture in the Kjeidahl apparatus until S03 fumes are given off and the solutionturns colorless or pale yellow. Continue heating for 30 additional minutes. Cool theresidue and add 300 ml distilled water.

8.2.2 Make the digestate alkaline by careful addition of 100 mi of sodium hydroxide -thiosulfate solution (7.4) without mixing.NOTES: Slow addition of the heavy caustic solution down the tilted neck of thedigestion flask will cause, heavier solution to underlay the aqueous sulfuric acidsolution without loss of free-ammonia. Do not mix until the digestion flask hasbeen connected to the distillation apparatus.

8.2.3 Connect the Kjeidahl flask to the condenser with the tip of condenser or anextension of the condenser tip below the level of the boric acid solution (7.6) in thereceiving flask. .

8.2.4 Distill 300 ml at the rate of 6-10 ml/min., into 50 ml of 2%- boric acid (7.6)contained in a 500 ml Erlenmeyer flask.

8.2.5 Dilute the distillate to 500 ml in the flask. These flasks should be marked at the 350and the 500 ml volumes. With such marking, it is not necessary to transfer thedistillate to volumetric flasks. For concentrations above I mg/1, the ammonia canbe determined.titrimetrically. For concentrations below this value, it is determinedcolorimetrically. The potentiometric method is applicable to the range 0.05 to 1400mg/1.

flR30JQt_2. .351.3-3

Page 226: ftR3008L6 - semspub.epa.gov

8.3 Micro Kjeidahl system8.3.1 Place 50.0 ml of sample or an aliquot diluted to 50 ml in a 100 ml Kjeidahl flask

and add 10 nil sulfuric acid-mercuric sulfate-potassium sulfate solution (7.3).Evaporate the mixture in the Kjeidahl apparatus until SO3 fumes are given off andthe solution turns colorless or pale yellow. Then digest for an additional 30minutes. Cool the residue and add 30 ml distilled water.

8.3.2 Make the digestate alkaline by careful addition of 10 ml of sodium hydroxide-thiosulfate solution (7.4) without mixing. Do not mix until the digestion flask hasbeen connected to the distillation apparatus.

8.3.3 Connect the Kjeidahl flask to the condenser with the tip of condenser or anextension of the condenser tip below the level of the boric acid solution (7.6) in thereceiving flask or 50 ml short-form Nessler tube.

8.3.4 Steam distill 30 ml at the rate of 6-10 ml/'min., into 5 ml of 2% boric acid (7.6).8.3.5 Dilute the distillate to 50 ml. For concentrations above 1 mg/1 the ammonia can be

determined titrimetrically. For concentrations below this value, it is determinedcolorimetrically. The potentiometric method is applicable to the range 0,05 to 1400mg/1.

8.4 Determination of ammonia in distillate: Determine the ammonia content of the distillatetitrimetrically, colorimetrically, or potentiometrically, as described below.8.4.1 Titrimetric determinatic-h: Add 3 drops of the mixed indicator (7.5) to the distillate

and titrate the ammonia with the 0.02 N H3SO4 (7.7), matching the endpointagainst a blank containing the same volume of distilled water and H3BOj (7.6)solution...

8.4.2 Colorimetric determination: Prepare a series of Nessler tube standards as follows;

ml of Standard1.0 ml = 0.01 mg NH3-N mg_NH3-N/50.0 ml

0,0 ,- - ., . ... - , 0.00.5- . . . . . . . . - .-.- - 0.0051.0 -0.0102.0 _ -.. = .,..-.- 0.0204.0 - - . 0.0405.0 - -., - -- -- —.-0.0508.0 -" " 0;08010.0 0.10

Dilute each tube to 50 ml with ammonia free water, add 1 ml of Nessler Reagent(7.10) and mix. After 20 minutes read the absorbance at 425 nm against the blank.From the values obtained for the standards plot absorbance vs. mg NH3-N for thestandard curve. Develop color in the 50 ml diluted distillate in exactly the samemanner and read mg NH3-N from the standard curve.

8.4.3 Potentiometric determination: Consult the method entitled Nitrogen, Ammonia:Potentiometric, Ion Selective Electrode Method, (Method 350.3) in this manual.

8.4.4 It is not imperative that all standards be treated in the same manner as the samples.It is recommended that at least 2 standards (a high and low) be c!igest<ffi,tSiQififed, n J. On it vj u f U4O

351.3-4

Page 227: ftR3008L6 - semspub.epa.gov

and compared to similar values on the curve to insure that the digestion-distillationtechnique is reliable. If treated standards do not agree with untreated standards theoperator should find the cause of the apparent error before proceeding.

9. Calculation9.1 If the titrimetric procedure is used, calculate Total Kjeidahl Nitrogen, in mg/1, in the

original sample as follows:

TKN, mg/1 = <A - B)N X..F x 1,000

where: .. ... - ---- - - - - — — -- - - - -A = milliliters of standard 0.020 N H2SO4 solution used in titrating sample.B ==milliliters of standard 0.020 N H2SO4 solution usedin titrating blank.N = normality of sulfuric acid solution.F = milliequivalent weight of nitrogen (14 mg).S = milliliters of sample digested.

If thesulfuric acid is exactly 0.02 N the formula is shortened to:

TKN. mg/1 * (A - B' *

9.2- If the Nessler procedure is used, calculate the Total Kjeidahl Nitrogen, in mg/1, in theoriginal sample as follows; . .._ "-"--."- - *•

where:A = mg NHj-N read from curve.B = ml total distillate collected including the H3BO3.C = ml distillate taken for Nesslerization.D = ml of original sample taken. _

9.3 Calculate Organic Kjeidahl Nitrogen in mg/1, as follows:OrganicKjeldaJ.il Nitrogen = TKN-<NH3-N.)

351.3-5

Page 228: ftR3008L6 - semspub.epa.gov

9.4 Potentiometric determination: Calculate Total Kjeidahl Nitrogen, in mg/1, in theoriginal sample as follows: .--- -

TKN, mg/1 * • x A

where: ~_ J - ~ i_. - --A — mg NH3~N/1 from electrode method standard curve.B s= volume of diluted distillate in ml.D = ml of original sample taken.

10. Precision10.1. Thirty-one analysts in twenty laboratories analyzed natural water samples containing

exact increments of organic nitrogen, with the following results:

Increment as Precision as _____Accuracy as_____Nitrogen, Kjeidahl _ Standard Deviation fiias, Bias,

mg N/liter mg N/liter % mg N/liter

0.20 - 0.197 -f 15.54 +0,030,3 J 0.247 + 5.45 +0.024.10 1.056 + 1.03 +0.044.61 1.191 _ - 1.6_7 -0.08

(FWPCA Method Study 2, Nutrient Analyses)

Bibliography

1. Standard Methods for the Examination of Water and Wastewater, 14th Edition, p 437,Method 421(1975).

2. Schlueter, Albert, "Nitrate Interference In Total Kjeidahl Nitrogen Determinations and ItsRemoval by Anion Exchange Resins", EPA Report 600/7-77-017.

351.3-6