fluorescent glucose biosensors

Upload: mihai-oprea

Post on 03-Jun-2018

215 views

Category:

Documents


0 download

TRANSCRIPT

  • 8/12/2019 Fluorescent Glucose Biosensors

    1/2

    Fluorescent glucose biosensors

    Fluorescent glucose biosensors are devices that measure the concentration of glucose

    in diabetic patients by means of sensitive protein that relays the concentration by means of

    fluorescence, an alternative to amperometric sension of glucose. No device has yet entered the

    medical market, but, due to the prevalence of diabetes, it is the prime drive in the construction

    of fluorescent biosensors.

    Fluorescence is a property present in certain molecules, called fluorophores, in which

    they emit a photon shortly after absorbing one with a higher energy wavelength

    Blood glucose monitoring is a way of testing the concentration of glucose in the blood

    (glycemia). Particularly important in the care of diabetes mellitus, a blood glucose test is

    performed by piercing the skin (typically, on the finger) to draw blood, then applying the

    blood to a chemically active disposable 'test-strip'. Different manufacturers use different

    technology, but most systems measure an electrical characteristic, and use this to determine

    the glucose level in the blood. The test is usually referred to as capillary blood glucose and

    sometimes incorrectly called BM Stix (after one of the companies that makes the test kit).

    Healthcare professionals advise patients with diabetes on the appropriate monitoring

    regime for their condition. Most people with Type 2 diabetes test at least once per day.

    Diabetics who use insulin (all Type 1 diabetes and many Type 2s) usually test their blood

    sugar more often (3 to 10 times per day), both to assess the effectiveness of their prior insulin

    dose and to help determine their next insulin dose.

    The monitoring systems currently in general use have the drawback of below optimal

    number of readings, due to their reliance on a drop of fresh blood. Some continuous glucose

    monitors are commercially available, but suffer from the severe drawback of a short working

    life of the probe. The majority of these work amperometrically. As a result, there is an effort

    to create a sensor that relies on a different mechanism, such as via external infrared

    spectroscopy or via fluorescent biosensors . Various strategies exist to detect glucose levels

    using fluorescence, the first and most common being a Fret competition assay between

    glucose and a labelled glucose polymer for the binding site of Concanavalin A. Over the

    years, using a combination of rational design and screening approaches, many possible

    combinations of fluorescent sensor for glucose have been studied with varying degrees ofsuccess: In most approaches, the glucose concentration is translated into a change in

    fluorescence either by using a Fret pair or by using environment sensitive (solvatochromic)

    dyes.

    Another pair of papers suggest that the natural fluorescence (autofluorescence) of

    tissues may be harnessed to track glucose concentrations. These studies took advantage of the

    fact that NAD(P)H, in its reduced form, is autofluorescent, and that metabolites such as

    glucose cause a predictable increase in NAD(P)H reduction

  • 8/12/2019 Fluorescent Glucose Biosensors

    2/2

    Ultrasound can also measure levels of glucose because sound waves are transmitted

    through the tissue in the earlobe at different rates, according to how much glucose is present.

    Heat also travels at a different rate through tissue when blood glucose levels are high

    or low.

    The ear clip device also fires low-level electromagnetic waves into the tissue - these

    are transmitted differently depending on blood sugar.

    Though none of these three methods is accurate enough on its own to replace blood

    tests, by averaging out scores GlucoTrack can come up with a valid reading.