fluorescence detection in liquid chromatography (agilent technologies)

Upload: leandregg

Post on 14-Apr-2018

220 views

Category:

Documents


0 download

TRANSCRIPT

  • 7/27/2019 Fluorescence Detection in Liquid Chromatography (Agilent Technologies)

    1/54

    Fluorescence detection in

    liquid chromatography

    A new approach tolower limits of detection andeasy spectral analysis

  • 7/27/2019 Fluorescence Detection in Liquid Chromatography (Agilent Technologies)

    2/54

    2

  • 7/27/2019 Fluorescence Detection in Liquid Chromatography (Agilent Technologies)

    3/54

    3

    A primer

    A new approach to

    lower limits of detectionand easy spectral analysis

    Applications offluorescence detectionin liquid chromatography

    Rainer Schuster andHelmut Schulenberg-Schell

  • 7/27/2019 Fluorescence Detection in Liquid Chromatography (Agilent Technologies)

    4/54

    4

    Copyright 2000 Agilent Technologies

    All rights reserved. Reproduction, adaption,

    or translation without proir written permission

    is prohibited, except as allowed under the

    copyright laws.

    Printed in Germany 01/00

    Publication number

    5968-9346E

  • 7/27/2019 Fluorescence Detection in Liquid Chromatography (Agilent Technologies)

    5/54

    5

    Acknowledgment The authors would like to thank our Agilent Technologiescolleagues Thomas Drr, Angelika Gratzfeld-Huesgen and

    Ludwig Huber for fruitful discussions and review of the

    manuscript. Udo Huber and Angelika Gratzfeld-Huesgen

    contributed to the applications.

  • 7/27/2019 Fluorescence Detection in Liquid Chromatography (Agilent Technologies)

    6/54

    6

    Preface High performance liquid chromatography (HPLC) is themethod of choice for separation and quantitation of polar

    and nonvolatile compounds. Retention time is the main tool

    for identification of analytes. The complexity of real world

    analytical problems requires the confirmation of peak

    identity through additional qualitative information. This can

    be provided from several detector types: diode array, mass

    spectrometric or fluorescence detectors.

    In 1982 HPs chemical analysis group (now part of Agilent

    Technologies) introduced the first commercially available

    diode array detector (DAD) for HPLC. This detector addeda third dimensionwavelengthto the chromatogram in

    addition to retention time and signal intensity. The diode

    array detector records UV/Visible absorption spectra in

    milliseconds while compounds are eluting from a column.

    A single chromatographic run provides quantitative results

    from the signal intensity and qualitative spectral information

    for peak confirmation. The improvements in detector

    design and the lower price have made it a sensitive and

    cost-effective tool that is about to replace single wavelength

    UV-detectors, even those performing routine analyses.

    Since the mid 1990s, mass spectrometric detectors for liquid

    chromatography (LC-MSD) have been available to provide

    analysts with both molecular weight and structural informa-

    tion for peak confirmation. The ruggedness and easy use of

    the instrumentation make this technique available for the

    chromatographer. From research and development, the

    technique will make its way into the cost-sensitive routine

    QA/QC laboratories.

    The fluorescence detector (FLD) is one of the most sensitive

    detectors in liquid chromatography. Both excitation and

    emission fluorescence spectra help to characterize indi-

    vidual compounds. While excitation spectra are identical

    to UV/Visible absorption spectra, emission spectra can giveadditional information. Until recently, however, FLDs have

    been built to provide single-wavelength information.

  • 7/27/2019 Fluorescence Detection in Liquid Chromatography (Agilent Technologies)

    7/54

    7

    Fluorescence spectra had to be acquired under stop-flow

    conditions and data analysis was time-consuming and

    cumbersome compared to data analysis of UV/Visible

    spectra.

    A completely new approach to fluorescence detector design

    has overcome these drawbacks and even improved the

    sensitivity. The Agilent 1100 Series fluorescence detector

    acquires spectra online simultaneously with the detector

    signal. During method development, one or two chromato-

    graphic runs can be sufficient to optimize wavelength

    settings for a series of analytes. This replaces tedious stop-flow experiments for each individual compound. Fluores-

    cence spectral data are analyzed with the same easy-to-use

    software tools as diode array spectra. In addition, up to four

    wavelengths can be recorded simultaneously to replace

    timetable-based wavelength switching. This ensures

    maximum sensitivity and selectivity without sacrificing the

    reliability of the analytical method in routine analysis of

    real-world samples.

    This primer introduces the new fluorescence detector

    technology and describes new strategies for rapid method

    development. A selection of applications demonstrates the

    wealth of information available with this new approach to

    fluorescence detection.

  • 7/27/2019 Fluorescence Detection in Liquid Chromatography (Agilent Technologies)

    8/54

    8

    Tableofco

    ntent

    Chapter 1Fluorescence detector technology ..................................................................................... 9How the fluorescence detector works .......................... .............. .............. .............. .............. ........ 11

    Data handling ................................................................................................................................... 12

    How to measure limits of detection .............................................................................................. 14

    Chapter 2Strategies for rapid method development ....................................................................... 15Step 1: Check the HPLC system for impurities............. .............. .............. .............. .............. ....... 17

    Step 2: Optimize limits of detection and selectivity ................................................................... 18

    Procedure I Take a fluorescence scan............... .............. ............... ............... .............. ........... 19

    Procedure II Take two HPLC runs with FLD ......................................................................... 21Procedure III Make a single run with the Agilent 1100 Series DAD/FLD combination ... 23

    Step 3: Set up routine methods ...................................................................................................... 25

    Multi wavelength detection ........................................................................................................ 25

    Fluorescence spectral libraries for peak confirmation.............. .............. ............. .............. .... 26

    Chapter 3The applications ................................................................................................................... 29Environmental:

    Polynuclear aromatic hydrocarbons ......................................................................................... 30

    Carbamates ................................................................................................................................... 33

    Glyphosate .................................................................................................................................... 37

    Food:

    Mycotoxins ................................................................................................................................... 40

    Aflatoxins B1/B2 and G1/G2 ............... .............. .............. .............. .............. .............. ............... ... 40

    Ochratoxin A ................................................................................................................................ 42

    Vitamins B2 and B6............. ............... .............. ............... .............. .............. ............... .............. .... 43

    Pharmaceutical:

    Quinidine....................................................................................................................................... 46

    Warfarin......................................................................................................................................... 48

    Amino acids ............ .............. .............. .............. .............. .............. .............. .............. ................ .... 50

    References........................................................................................................................................ 52

    Index ................................................................................................................................................. 53

  • 7/27/2019 Fluorescence Detection in Liquid Chromatography (Agilent Technologies)

    9/54

    9

    Chapter 1

    Fluorescencedetector

    technology

  • 7/27/2019 Fluorescence Detection in Liquid Chromatography (Agilent Technologies)

    10/54

    10

    Fluorescenc

    edetectortechnology

    Fluorescence detectors offer high selectivity combined

    with superior limits of detection (LOD) compared to UV

    detectors. Only about 10 percent of organic molecules

    have fluorophore structures, which enable the molecules

    to absorb light over a range of wavelengths. This takes

    electrons in the molecule to an excitation level as with any

    other molecule containing a chromophore. The fluorescent

    molecule has the ability to emit the absorbed energy at

    longer wavelengths (ref. 1).

    If the fluorescent light intensities are recorded while the

    excitation wavelength is changed and the emission wave-

    length is fixed, an excitation spectrum can be obtained.Or the excitation wavelength may be fixed and the emission

    wavelength changed. This procedure provides an emission

    spectrum of the analyte. Compounds without this native

    fluorescence may be derivatized to attach a fluorescent

    marker molecule in a pre- or post-column reaction (ref. 2).

    Fluorescence detectors offer limits of detection down to the

    ppt level. The signal intensities are very low compared to UV

    absorption and they are measured ideally versus a very low

    background noise level. This is inherently more sensitive

    than comparing two relatively large signals from a blank and

    a sample as done in UV absorption spectroscopy. However,

    the sensitivity of fluorescence detection is dependent on

    both the fluorophore properties and the detector design and

    settings. The response of a fluorophor is characterized by

    molar absorptivity and quantum yield at the applied experi-

    mental conditions. The sensitivity of the fluorescence

    detector depends on several factors: source intensity,

    efficiency of the optical system, bandwidth and so forth.

  • 7/27/2019 Fluorescence Detection in Liquid Chromatography (Agilent Technologies)

    11/54

    11

    How the fluorescencedetector works

    Figure 1Optical design of the Agilent 1100 Seriesfluorescence detector

    Fluorescenc

    edetectortechn

    ology

    Xenon

    flash lamp

    Excitation

    monochromator

    Emission

    monochromator

    Sample

    Photo-

    multiplier

    Lens

    Mirror

    Photodiode

    Lens

    Previous fluorescence detectors were equipped with motor-

    driven gratings for programmable excitation and emission

    wavelengths. The detector layout enabled positioning for a

    single wavelength at a time and required stop-flow condi-

    tions for acquisition of spectra. Thus, acquiring fluores-

    cence spectra was time-consuming and optimizing excita-

    tion and emission wavelengths was not possible without

    disturbing chromatographic separations.

    The new optical design of the Agilent 1100 Series fluores-

    cence detector is illustrated in figure 1.A Xenon flash lamp

    is used to offer the highest light intensities for excitation inthe UV range. The flash lamp ignites only for microseconds

    to provide light energy. Each flash causes fluorescence in

    the flow cell and generates an individual data point for the

    chromatogram. Since the lamp is not powered on during

    most of the detector operating time, it offers a lifetime of

    several thousand hours. No warmup time is needed to get a

    stable baseline. A holographic grating is used as a mono-

    chromator to disperse the polychromatic light of the Xenon

    lamp. The desired wavelength is then focused on the flow

    cell for optimum excitation. To minimize stray light from

    the excitation side of the detector, the optics are configured

    such that the emitted light is recorded at a 90 degree angle

    to the incident light beam. Another holographic grating is

    used as the emission monochromator. Both monochroma-

    tors have optimized light throughput in the visible range.

    A photomultiplier tube is the optimum choice to measure

    the low light intensity of the emitted fluorescence light.

    Since flash lamps have inherent fluctuations with respect to

    flash-to-flash intensity, a reference system based on a

    photodiode measures the intensity of the excitation and

    triggers a compensation of the detector signal.

    Since the vast majority of emission maxima are above

    280 nm, a cut-off filter (not shown) prevents stray light

    below this wavelength to enter the light path to the emis-sion monochromator. The fixed cut-off filter and bandwidth

    (20 nm) avoid the hardware checks and documentation

  • 7/27/2019 Fluorescence Detection in Liquid Chromatography (Agilent Technologies)

    12/54

    12

    Data handling

    Fluorescenc

    edetectortechn

    ology

    work that is involved with an instrument that has exchange-

    able filters and slits.

    The excitation and emission monochromators can switch

    between signal and spectral mode. In signal mode they are

    moved to specific positions that encode for the desired

    wavelengths, as with a traditional detector. This mode offers

    the lowest limits of detection since all data points are

    generated at a single excitation and emission wavelength.

    The spectral mode is used to obtain multi-signal or spectral

    information. The ignition of the flash lamp is synchronized

    with the rotation of the gratings, either the excitation or

    emission monochromator. The motor technology for the

    gratings is a long-life design as commonly used in high-

    speed PC disk drive hardware. Whenever the grating has

    reached the correct position during a revolution, the Xenon

    lamp is ignited to send a flash. The flash duration is below

    two microseconds while the revolution of the grating takes

    less than 14 milliseconds. Because of the rotating mono-

    chromators, the loss in sensitivity in the spectral mode is

    much lower compared to conventional dual-wavelength

    detection with UV detectors.

    In addition to its use as a detector for liquid chromatogra-

    phy, the Agilent 1100 Series FLD offers capabilities for off-

    line measurements in a refillable cuvette to obtain excita-

    tion and emission spectra in a single task for a pure com-

    pound. The result of this fluorescence scan can be viewed in

    a three-dimensional plot showing excitation wavelength,

    emission wavelength and fluorescence intensity on the axis.

    The sequence of data handling is shown schematically in

    figure 2. The example describes the collection of excitation

    spectra and a signal with Ex 250 nm/Em 350 nm. Alternating

    flashes are used either for chromatographic signals or tocontribute to spectra acquisition. About every 14 millisec-

  • 7/27/2019 Fluorescence Detection in Liquid Chromatography (Agilent Technologies)

    13/54

    13

    Figure 2Schematic representation of data flowin spectral modeOne signal at Ex 250 nm/Em 350 nm isrecorded and excitation spectra are takenstarting at Ex 230 nm.

    Spectrum

    Chromatogram

    Wavelength

    Time

    Flash# Time Ex Em Intensity

    [msec] [nm] [nm]

    1* 0 250 350 5

    2 14 230 350 4

    3 28 250 350 8

    4 42 240 350 15

    5 56 250 350 15

    6 70 250 350 30

    7 84 250 350 30

    8 98 260 350 25

    *flash duration = 1 microsecondFluorescenc

    edetectortechn

    ology

    With chromatographic data in three dimensions (emission

    or excitation wavelengths, retention time and intensity),

    the analysis can be displayed either in a three-dimensional

    plot or in a two-dimensional isofluorescence plot. Differ-

    ent colors signify different fluorescence intensities.

    Spectra can be evaluated against spectra from customized

    spectral libraries to determine compound identity or to

    control peak purity within a peak. While the quality of

    fluorescence spectra has proven to be useful for several

    applications, it has to be emphasized that UV/Vis spectraobtained with a diode array detector are of superior quality,

    especially for trace analysis.

    onds, a data point is obtained. The spectrum starts at

    230 nm. Just as with a DAD system, the fluorescence spectra

    and signals can be watched online and both types of data

    are stored on the PC hard disk. Additional signals can be

    extracted from the spectral data set during post-run data

    analysis. Spectral data points are corrected automatically

    for intensity changes over time based on the signal data.

  • 7/27/2019 Fluorescence Detection in Liquid Chromatography (Agilent Technologies)

    14/54

    14

    Fluorescenc

    edetectortechnology

    Figure 3Raman signal-to-noise (S/N)measurementSignal and ASTM noise at 397 nm (water),excitation at 350 nm, PMT=12, responsetime 8 s

    How to measure limitsof detection

    Fluorescence spectrophotometers can be qualified by the

    signal-to-noise ratio (S/N) for the Raman band for water or

    as a limit of detection (LOD) for a specific fluorescent

    compound.(ref.3)

    The Raman band is a result of Raman light scattering and

    not due to fluorescence. It simulates the phenomenon of

    fluorescence as it involves an initial light signal at a specific

    wavelength that causes a signal to occur at a longer wave-

    length. The specification based on Raman stray light is given

    as the ratio of signal to noise with the excitation wavelength

    at 350 nm and emission at 397 nm. The short-term noise is

    measured at 397 nm (figure 3) according to the procedures

    described in ASTM method 1657/94. For both values, the

    dark value at 450 nm (where no stray light appears) is taken

    as a reference point for the scale. Raman values greater than

    400 are typical for fluorescence detectors in HPLC.

    If anthracene is used to measure detector specifications,

    limits of detection as low as 10 femtogram anthracene are

    possible.

    LU

    38

    36

    34

    32

    30

    28

    26

    24

    0 2.5 5 7.5 10 12.5 [min]2017.515

    Signal and ASTM noise at 350/397 nm

    Dark value at 350/450 nm

    Raman S/N =Signal (397 nm) dark value (450 nm)

    ASTM noise (397 nm) dark value (450 nm)

  • 7/27/2019 Fluorescence Detection in Liquid Chromatography (Agilent Technologies)

    15/54

    15

    Chapter 2

    Strategies forrapid method

    development

  • 7/27/2019 Fluorescence Detection in Liquid Chromatography (Agilent Technologies)

    16/54

    16

    Methoddevelop

    ment

    Fluorescence detectors are used in liquid chromatography

    when superior limits of detection and selectivity are re-

    quired. Thorough method development, including spectra

    acquisition, is fundamental to achieve good results. This

    chapter describes three different steps that can be taken

    with the Agilent 1100 Series fluorescence detector. Table 1

    gives an overview of how to benefit from the operation

    modes during these steps.

    Table 1Steps for thorough method development

    Step 1: Step 2: Step 3:Check system Optimize limits of Set up routine

    detection and methodsselectivity

    Fluorescence Find impurities Determinescan (for example, in simultaneously the

    solvents and excitation and emissionreagents) spectra of a pure

    compound

    Signal mode Perform wavelength Use for lowestswitching limits of detection

    Spectral mode/ Determine Ex/Em Collect onlinemulti-wave- spectra for all spectra, perform

    length separated compounds library search,detection in a single run determine peak

    purityActivate up to four Replacewavelengths wavelengthsimultaneously switching

  • 7/27/2019 Fluorescence Detection in Liquid Chromatography (Agilent Technologies)

    17/54

    17

    Step 1:Check the HPLC systemfor impurities

    A critical issue in trace level fluorescence detection is to

    have an HPLC system free of fluorescent contamination.

    Most contaminants derive from impure solvents. Taking a

    fluorescence scan is a convenient way to check the quality

    of the solvent in a few minutes. This can be done, for

    example, by filling the FLD cuvette directly with the

    solvent for an offline measurement even before the start of

    a chromatographic run. The result can be displayed as an

    isofluorescence plot or a three-dimensional plot. Different

    colors reflect different intensities.

    Figure 4 shows a sample of slightly impure water which wasplanned for use as mobile phase. The area where fluores-

    cence of the contaminated water sample can be seen is

    between the stray light areas: the first- and second-order

    Raleigh stray light and Raman stray light. Since excitation

    and emission wavelength are the same for Raleigh stray

    light, the area of first-order Raleigh stray light is visible in

    the left upper area of the diagram. The Raman bands of

    water are seen below the first-order Raleigh stray light.

    Since the cut-off filter cuts off light below 280 nm, the

    second-order Raleigh stray light starts above 560 nm.

    Stray light acts in the same way as impurities in that it

    simulates background noise. In both cases, a higher noise

    level and therefore a higher limit of detection are obtained.This indicates that high sensitivity measurements should be

    done away from wavelength settings that have a high stray

    light background.

    Figure 4Isofluorescence plot of a mobile phaseA pure water sample was put into theflow cell. Spectra were recorded at 5 nmstep sizes.

    Impurity 1. order Raman 2. order

    450 nm

    220 nm

    Emission

    Excitation

    600 nm

    Methoddevelop

    ment

  • 7/27/2019 Fluorescence Detection in Liquid Chromatography (Agilent Technologies)

    18/54

    18

    Methoddevelop

    ment

    Wavelength [nm]

    250 300 350 400 450 500 550

    Norm.

    0

    5

    10

    15

    20

    25

    30

    35

    40

    Excitation Emission

    600

    Figure 5Excitation and emission spectra ofquinidineExcitation spectrum with emission at440 nm, emission spectrum with excitation

    at 250 nm of 1 g/ml quinidine.Detector settings: step size 5 nm, PMT 12,Response time 4 s.

    Step 2:Optimize limits ofdetection andselectivity

    To achieve optimum limits of detection and selectivity,

    analysts must find out about the fluorescent properties of

    the compounds of interest. Excitation and emission wave-

    lengths can be selected for optimum limits of detection and

    best selectivity. In general, fluorescence spectra obtained

    with different instruments may show significant differences

    depending on the hardware and software used. (ref. 4)

    The traditional approach is to extract an appropriate

    excitation wavelength from the UV spectrum that is similar

    to the fluorescence excitation spectrum (see figure 5) and

    to record the emission spectrum. Then with an optimumemission wavelength determined, the excitation spectrum

    is acquired.

    These tasks have to be repeated for each compound using

    either a fluorescence spectrophotometer or stop-flow

    conditions in HPLC. Usually each compound requires a

    separate run. As a result, a set of excitation and emission

    spectrum is obtained (figure 5) for each compound. Since

    this is a tedious procedure, it is applicable only when there

    is a limited number of compounds of interest.

  • 7/27/2019 Fluorescence Detection in Liquid Chromatography (Agilent Technologies)

    19/54

    19

    Methoddevelop

    ment

    The Agilent 1100 Series HPLC offers three different

    ways to obtain complete information on a compounds

    fluorescence:

    Procedure I - Take a fluorescence scan offline for a single

    compound as described above for the mobile phase. This is

    done preferably with a manual FLD cuvette when pure

    compounds are available.

    Procedure II - Use two HPLC runs with the Agilent 1100

    Series FLD to separate the compound mix under known

    conditions and acquire emission and excitation spectraseparately.

    Procedure III - Use an Agilent 1100 Series FLD/DAD

    combination and acquire UV/Visible spectra (equivalent to

    excitation spectra) with the DAD and emission spectra with

    the FLDboth in a single run.

    Procedure I Take a fluorescence scan

    Because fluorescence spectra traditionally have not been

    easily available with previous HPLC fluorescence detectors,

    standard fluorescence spectrophotometers have been usedin the past to acquire spectral information for unknown

    compounds. Unfortunately this approach limits optimiza-

    tion, as there are differences expected in optical design

    between an HPLC detector and a dedicated fluorescence

    spectrophotometer, or even between detectors. These

    differences can lead to variations for the optimum excita-

    tion and emission wavelengths.

    The Agilent 1100 Series fluorescence detector offers a

    fluorescence scan that delivers all spectral information

    previously obtained with a standard fluorescence spectro-

  • 7/27/2019 Fluorescence Detection in Liquid Chromatography (Agilent Technologies)

    20/54

    20

    Methoddevelop

    ment

    Figure 6Characterization of a pure compoundfrom a fluorescence scanAll excitation and emission spectra ofQuinidine (1 g/ml) are shown in onegraphic. Fluorescence intensity is plottedvs. excitation and emission wavelengths.Figure 5 gives detector settings.

    straylight1. order 350 nm Ex 315 nm Ex 250 nm Ex

    Ex-axis Em-axis

    photometer, independent of the HPLC fluorescence detec-

    tor. Figure 6 shows the complete information for quinidine

    as obtained with the Agilent 1100 Series FLD and a manual

    cuvette in a single offline measurement. The optima for

    excitation and emission wavelengths can be extracted as

    coordinates of the maxima in the three dimensional plot.

    One of the three maxima in the center of the plot can be

    chosen to define the excitation wavelength. The selection

    depends on the additional compounds that are going to be

    analyzed in the chromatographic run and the background

    noise that may be different upon excitation at 250 nm,

    315 nm or 350 nm. The maximum of emission is observedat 440 nm.

    440 Em

  • 7/27/2019 Fluorescence Detection in Liquid Chromatography (Agilent Technologies)

    21/54

    21

    Procedure II Take two HPLC runs with theFLD

    The conditions for the separation of organic compounds

    such as polyaromatic nuclear hydrocarbons (PNAs) are well

    described in various standard methods, including commonly

    used EPA and DIN methods. Achieving the best detection

    levels requires checking for the optimum excitation and

    emission wavelengths for all compounds. Yet taking fluores-

    cence scans individually makes this a tedious process. A

    better approach is to acquire spectra online for all com-

    pounds during a run. This speeds up method development

    tremendously. Two runs are sufficient for optimization.

    During thefirst run, one wavelength is chosen in the low

    UV range for the excitation wavelength and a spectral range

    for the emission wavelength. Most fluorophores show strong

    absorption at these wavelengths. Excitation is sufficient for

    collecting emission spectra.

    Figure 7 contains all emission spectra obtained in a single

    run from a mix of 15 PNAs. This set of spectra is used to set

    up a timetable for optimum emission wavelengths for all

    compounds.

    The individual compound spectra in the isofluorescence

    plot show that at least three emission wavelengths are

    needed to detect all 15 PNAs properly:

    0 min: 350 nm for naphthalene to phenanthrene

    8.2 min: 420 nm for anthracene to benzo(g,h,I)perylene

    19.0 min: 500 nm for indeno(1,2,3-cd)pyrene

    In the second run, three setpoints for emission wavelengths

    are entered into the time-program and excitation spectra

    are recorded, as shown in figure 8. The area of high intensity

    (red) is caused by stray light when emission spectra overlapwith the excitation wavelength. This can be avoided by

    fitting the spectral range automatically. Excitation at

    260 nm is most appropriate for all PNAs.

    Methoddevelop

    ment

  • 7/27/2019 Fluorescence Detection in Liquid Chromatography (Agilent Technologies)

    22/54

    22

    Methoddevelop

    ment

    Conditions for figure 7 and 8Column Vydac, 2.1 200 nm,

    PNA, 5 mMobile phase A = water; B = acetonitrileGradient 3 min, 60 %B;

    14 min, 90 %B;22 min, 100 %B

    Flow rate 0.4 ml/minColumntemperature 18 CInjectionvolume 5 lFLD settings PMT 12,

    response time 4 s,

    step size 5 nm

    Figure 8Optimization of the time-program for theexcitation wavelength

    Figure 7Optimization of the time-program for theemission wavelengthThis shows the isofluorescence plot ofemission spectra for 15 PNAs (5 g/ml)with a fixed excitation wavelength (260 nm).

    LU

    10

    20

    30

    40

    50

    60

    220 nm

    400 nm

    350 nm 420 nm 500 nm Emissionswitching

    1

    2

    34

    5

    67

    8

    9

    10

    11

    12

    13

    14

    1 Naphthalene

    2 Acenaphthene

    3 Fluorene

    4 Phenanthrene

    5 Anthracene6 Fluoranthene

    7 Pyrene

    8 Benz(a)anthracene

    9 Chrysene

    10 Benzo(b)fluoranthene

    11 Benzo(k)fluoranthene

    12 Benz(a)pyrene13 Benzo(g,h,i)perylene

    14 Indeno(1,2,3-cd)pyrene

    Excitationspectra

    0 2.5 7.5 10 12.5 15 17.5 20 22.5Time [min]

    5

    LU

    10

    20

    30

    40

    50

    60

    300 nm

    600 nm

    1

    2

    34

    5

    67

    8

    9

    10

    11

    12

    13

    14

    1 Naphthalene

    2 Acenaphthene

    3 Fluorene

    4 Phenanthrene

    5 Anthracene

    6 Fluoranthene

    7 Pyrene

    8 Benz(a)anthracene

    9 Chrysene

    10 Benzo(b)fluoranthene

    11 Benzo(k)fluoranthene

    12 Benz(a)pyrene

    13 Benzo(g,h,i)perylene

    14 Indeno(1,2,3-cd)pyrene

    Em-spectra

    fixed Ex

    0 2.5 5 7.5 10 12.5 15 17.5 20 22.5Time [min]

  • 7/27/2019 Fluorescence Detection in Liquid Chromatography (Agilent Technologies)

    23/54

    23

    Table 2Timetable for the analysis of 15 poly-nuclear aromatic hydrocarbonsThis timetable gives the conditions foroptimum detection based on the results

    of two chromatographic runs.

    Time Excitation Emission[min] wavelength wavelength

    [nm] [nm]

    0 260 3508.2 260 420

    19.0 260 500

    Methoddevelop

    ment

    The obtained data are combined to setup the time-table for

    for best limit of detection and selectivity. The optimized

    switching events for this example are summarized in

    table 2.

    Procedure III Make a single run with theAgilent 1100 Series DAD/FLD combination

    For most organic compounds, UV-spectra from diode array

    detectors are nearly identical to fluorescence excitation

    spectra. Spectral differences are caused by specific detectorcharacteristics such as spectral resolution or light sources.

    In practice, combining a diode array detector with a fluores-

    cence detector in series gives the full data set needed to

    achieve the optimum fluorescence excitation and emission

    wavelengths for a series of compounds in a single run. With

    the UV/Visible/excitation spectra available from the diode

    array detector, the fluorescence detector is set to acquire

    emission spectra with a fixed excitation wavelength in the

    low UV range.

    The example is taken from the quality control of carbam-

    ates. Samples are analyzed for the impurities 2,3-diamino-

    phenazine (DAP) and 2-amino-3-hydroxyphenazine (AHP).

    Reference samples of DAP and AHP were analyzed with

    diode array and fluorescence detection. Figure 9 shows the

    spectra obtained from both detectors for DAP. The excita-

    tion spectrum of DAP is very similar to the UV absorption

    spectrum from the diode array detector. Figure 10 shows

    the successful application of the method to a carbamate

    sample and a pure mixture of DAP and AHP for reference.

    The column was overloaded with the non-fluorescent

    carbamate (2-benzimidazole carbamic acid methylester/

    MBC) to see the known impurities, AHP and DAP.

  • 7/27/2019 Fluorescence Detection in Liquid Chromatography (Agilent Technologies)

    24/54

    24

    Methoddevelop

    ment

    Wavelength [nm]

    200 250 300 350 400 450 500 550

    0

    5

    10

    15

    20

    25

    30

    35

    Excitation Emission

    430 nm

    265 nm

    DAD-spectra

    UV

    Norm.

    Time [min]0 2 4 6 8 10 12

    LU

    0

    0.2

    0.4

    0.6

    0.8

    430/540 nm

    Standard

    Unknown2-amino-3-OH-phenazine

    2,3-diaminophenazine

    MBC

    265/540 nm

    Figure 9UV-spectrum and fluorescence spectra for 2,3-diaminophenazine (DAP)This is an impurity of carbamates. The excitation spectrum in a second runshows the similiarity of UV-spectra and fluorescence excitation spectra. Anexcitation wavelength at 265 nm was used for taking the emission spectrum andan emission wavelength at 540 nm was used for taking the excitation spectrum.

    ConditionsColumn Zorbax SB, 50 2 mm,

    PNA, 5 mMobile phase A = water; B = acetonitrileGradient 0 min, 5 %B;

    10 min, 15 %B;Flow rate 0.4 ml/minColumntemperature 35 CInjectionvolume 5 lFLD settings PMT 12,

    response time 4 s,step size 5 nm, Ex 265 nm

    and 430 nm, Em 540 nm

    Figure 10Qualitative analysis of MBC(2-benzimidazole carbamic acidmethylester) and impuritiesThe two upper traces are obtained usingtwo different excitation wavelengths.The lower trace is a pure standard of theknown impurities.

  • 7/27/2019 Fluorescence Detection in Liquid Chromatography (Agilent Technologies)

    25/54

    25

    Methoddevelop

    ment

    Step 3:Set up routine methods

    In routine analysis, sample matrices can have a significant

    influence on retention times. For reliable results, sample

    preparation must be thorough to avoid interferences or

    HPLC methods must be rugged enough. With difficult

    matrices, simultaneous multi-wavelength detection offers

    more reliability than timetable-controlled wavelength

    switching. The Agilent 1100 Series FLD can, in addition,

    acquire fluorescence spectra while it records the detector

    signals for quantitative analysis. Therefore qualitative data

    are available for peak confirmation and purity checks in

    routine analysis.

    Multi wavelength detection

    Time-programmed wavelength switching traditionally is

    used to achieve low limits of detection and high selectivity

    in routine quantitative analysis. Such switching is difficult

    if compounds elute closely and require a change in excita-

    tion or emission wavelength. Peaks can be distorted and

    quantitation made impossible if wavelength switching

    occurs during the elution of a compound. Very often this

    happens with complex matrices, influencing the retention

    of compounds.

    In spectral mode, the Agilent 1100 Series FLD can acquire

    up to four different signals simultaneously. All of them can

    be used for quantitative analysis. Apart from complex

    matrices, this is advantageous whenwatching for impurities

    at additional wavelengths. It is also advantageous for

    reaching low limits of detection or increasing selectivity

    through optimum wavelength settings at any time. The

    number of data points acquired per signal is reduced and

    thus limits of detection may be higher, depending on the

    detector settings compared to the signal mode.

    PNA analysis, for example, can be performed with simulta-

    neous multi wavelength detection instead of wavelength-

    switching. With four different wavelengths for emission, all

    15 PNAs can be monitored (figure 11).

  • 7/27/2019 Fluorescence Detection in Liquid Chromatography (Agilent Technologies)

    26/54

    26

    ConditionsColumn Vydac, 2.1 250 mm,

    PNA, 5 mMobile phase A = water; B = acetonitrileGradient 3 min, 60 %B;

    14.5 min, 90 %B;22.5 min, 95 %B

    Flow rate 0.4 ml/minColumntemperature 22 CInjectionvolume 2 lFLD settings PMT 12,

    response time 4 s

    1 excitation WL at 260 nm

    4 emission WL at 350, 420,

    440 and 500 nm

    Ex=260, Em=350

    Ex=260, Em=420

    Ex=260, Em=440

    Ex=260, Em=500

    Ex=275, Em=350, TTReferencechromatogramwith switching events

    0 5 10 15 20 25

    LU

    0

    20

    40

    60

    80

    100

    120

    140

    160

    180

    Time [min]

    12

    3

    5

    6

    7

    8

    9

    10

    11

    12

    13 14154

    1 Naphthalene

    2 Acenaphthene

    3 Fluorene

    4 Phenanthrene

    5 Anthracene

    6 Fluoranthene

    7 Pyrene

    8 Benz(a)anthracene

    9 Chrysene

    10 Benzo(b)fluoranthene

    11 Benzo(k)fluoranthene

    12 Benz(a)pyrene

    13 Dibenzo(a,h)anthracene

    14 Benzo(g,h,i)perylene

    15 Indeno(1,2,3-cd)pyrene

    Figure 11Simultaneous multi wavelengthdetection for PNA-analysisThe upper trace was received withtraditional wavelength switching.

    Fluorescence spectral libraries for peakconfirmation

    Previously, only diode array detectors and mass spectromet-

    ric detectors could deliver spectral information on-line to

    confirm peak identity as assigned by retention time.

    Now, fluorescence detectors provide an additional tool for

    automated peak confirmation and purity control. No

    additional run is necessary after the quantitative analysis.

    During method development, fluorescence excitation and

    emission spectra are collected from reference standards and

    entered into a libraryat the choice of the method devel-

    oper. All spectral data from unknown samples can then becompared automatically with library data. Table 3 illustrates

    Methoddevelop

    ment

    Ex/Em = 260/420 nm

    Ex/Em = 270/440 nm

    Ex/Em = 260/420 nm

    Ex/Em = 290/430 nm

    Ex/Em = 250/550 nm

  • 7/27/2019 Fluorescence Detection in Liquid Chromatography (Agilent Technologies)

    27/54

    27

    this principle using a PNA analysis. The match factor given

    in the report for each peak indicates the degree of similarity

    between the reference spectrum and the spectra from a

    peak. A match factor of 1,000 means identical spectra.

    In addition, the purity of a peak can be investigated by

    comparing spectra obtained within a single peak. When

    a peak is calculated to be within the user-defined purity

    limits, the purity factor is the mean purity value of all

    spectra that are within the purity limits.

    The reliability of the purity and the match factor dependson the quality of spectra recorded. Because of the lower

    number of data points available with the fluorescence

    detector in general, the match factors and purity data

    obtained show stronger deviations compared to data from

    the diode array detector, even if the compounds are identi-

    cal.

    Table 3Peak confirmation using a fluorescencespectral libraryThis shows an automated library searchbased on the emission spectra from a

    PNA reference sample.

    Meas. Library CalTblRetTime Sig Amount Purity Library Name

    [min] [min] [min] [ng] Factor # Match

    4.859 4.800 5.178 1 1.47986e-1 1 993 Naphthalene@em

    6.764 7.000 7.162 1 2.16156e-1 1 998 Acenaphthene@em

    7.137 7.100 7.544 1 1.14864e-1 1 995 Fluorene@em

    8.005 8.000 8.453 1 2.56635e-1 1 969 Phenanthrene@em

    8.841 8.800 9.328 1 1.76064e-1 1 993 Anthracene@em

    9.838 10.000 10.353 1 2.15360e-1 1 997 Fluoranthene@em

    10.439 10.400 10.988 1 8.00754e-2 1 1000 Pyrene@em

    12.826 12.800 13.469 1 1.40764e-1 1 998 Benz(a)anthracene@em

    13.340 13.300 14.022 1 1.14082e-1 1 999 Chrysene@em

    15.274 15.200 16.052 1 6.90434e-1 1 999 Benzo(b)fluoranthene@em

    16.187 16.200 17.052 1 5.61791e-1 1 998 Benzo(k)fluoranthene@em

    16.865 16.900 17.804 1 5.58070e-1 1 999 Benz(a)pyrene@em

    18.586 18.600 19.645 1 5.17430e-1 1 999 Dibenz(a,h)anthracene@em

    19.200 19.100 20.329 1 6.03334e-1 1 995 Benzo(g,h,i )perylene@em

    20.106 2 0.000 21.291 1 9.13648e-2 1 991 Indeno(1,2,3-cd)pyrene@em

    Methoddevelop

    ment

  • 7/27/2019 Fluorescence Detection in Liquid Chromatography (Agilent Technologies)

    28/54

    28

    Methoddevelop

    ment

  • 7/27/2019 Fluorescence Detection in Liquid Chromatography (Agilent Technologies)

    29/54

    29

    Chapter 3

    The applicationsEnvironmental

    Polynuclear aromatic hydrocarbons Carbamates

    Glyphosate

    Food

    Aflatoxins B1/B2 and G1/G2

    Ochratoxine A

    Vitamins B2 and B6

    Pharmaceutical

    Quinidine

    Warfarin

    Amino acids

  • 7/27/2019 Fluorescence Detection in Liquid Chromatography (Agilent Technologies)

    30/54

    30

    Environmentalapplica

    tions

    Polynuclear aromatic hydrocarbons (PNA) are formed

    during pyrolysis or incomplete combustion in industrial or

    private heaters, automobile exhaust fumes and tobacco

    smoke. These compounds are in many oil products such as

    diesel fuel, gasoline and bitumen. PNAs have become a

    ubiquitous class of compounds found in all environmental

    matrices, including air, soil and water.

    Many PNAs have been found to be carcinogenic or mu-

    tagenic. Because their structures differ, some are more

    carcinogenic than others. Benzo(e)pyrene, for example, is a

    weak carcinogen, while isomeric benzo(a)pyrene is a strongcarcinogen. For this reason, maximum concentration limits

    have been set for each individual PNA in air, soil and

    especially in water samples.

    Most environmental regulations in the U.S., many countries

    in Asia, and eastern Europe require PNA analysis according

    to U.S. Environmental Protection Agency (EPA) methods.

    These include analysis of 16 individual PNAs. In western

    Europe, most official methods describe the analysis of the

    same set of compounds as the EPA methodsexcept the

    European methods do not include acenaphthylene, a com-

    pound which does not show any fluorescence. A rapid

    method for analyzing only six PNAs using a fast isocratic

    HPLC method is described in the German standard method

    DIN 38 407 F8.

    Experiments and results

    With the rapid HPLC method, a PNA analysis is achieved in

    less then 10 min on a 25 cm microbore column (shown in

    figure 12). The lowest limits of detection for five of the six

    compounds is found at Ex 360 nm/Em 429 nm. For the last

    compound, the optimum is found at Ex 360/Em 490 nm and

    the compromise (Ex 360/Em 465 nm) gives medium limits

    Polynuclear aromatichydrocarbons

  • 7/27/2019 Fluorescence Detection in Liquid Chromatography (Agilent Technologies)

    31/54

    31

    ConditionsIsocratic methodColumn Vydac, 2.1 250 nm,

    PNA, 5 mMobile phase Water/acetonitrile = 10/90

    Flow rate 0.4 ml/minColumntemperature 18 CInjectionvolume 75 lFLD settings PMT 12,

    response time 4 s,step size 5 nm,Ex 360 nm, Em 465 nm

    LU1

    24

    5

    31 Fluoranthene

    2 Benzo(k)fluoranthene

    3 Benzo(b)fluoranthene

    4 Benz(a)pyrene

    5 Benzo(ghi)perylene

    6 Indeno(1,2,3-cd)pyrene

    2 4 6 8 10 12 14 16 18

    2

    4

    6

    8

    10

    Position-mode

    75ul ( 2ppb) = 150 pg

    LOD - 26ppt

    LOD - 7.6ppt

    LOD - 3.3ppt

    LOD - 4.9ppt

    6

    Time [min]

    of detection for all. The detection limit for five of the

    PNAs is down to 100 fg and 1 pg absolute for indeno(1,2,3-

    cd)pyrene (Ex 360 nm/Em 490 nm), the last eluting com-

    pound. Increasing the injection volume achieves even

    better detection limitsdown to low ppt levels.

    Figure 12Determination of PNAs according to DIN 38407 F8 down to a low ppt-levelA 2 ppb reference standard was analyzed at Ex 360 nm/Em 465 nm, with aninjection volume of 75 l. The limits of detection (LOD) are given at the S/N ratio=2.

    Most PNA standards contain the antioxidant trichlorophenol,

    which coelutes with chrysene. The antioxidant is so far

    observed only in the DAD analysis because of the UV

    absorption spectrum. With simultaneous multi wavelength

    detection, both chrysene at Ex 260/Em 420 nm and

    trichlorophenol at Ex 260/Em 310 nm can be quantitated

    selectively, as shown in figure 13. This is useful information

    when checking the stability of PNA standards.

    Environmentalapplica

    tions

  • 7/27/2019 Fluorescence Detection in Liquid Chromatography (Agilent Technologies)

    32/54

    32

    Environmentalapplica

    tions

    ConditionsGradient methodColumn Vydac, 2.1 250 nm,

    PNA, 5 mMobile phase A = water, B = acetonitrileGradient 3 min, 50 % B,

    14 min, 90 % B,22 min, 100 % B,

    Flow rate 0.4 ml/minColumntemperature 18 CInjectionvolume 5 lFLD settings PMT 12,

    response time 4 s,step size 5 nm, Ex=260 nm,Em=310 and 420 nm

    Figure 13Control of antioxidants in PNA reference standards with the Agilent 1100 Seriesfluorescence detectorWith multi wavelength detection, the antioxidant trichlorophenol and thePNAs could be analyzed in a single run, each at a specific wavelength.

    Time [min]

    LU

    14 16 18 20 22

    5

    10

    15

    20

    25

    30

    Chrysene Ex=260, Em=420

    Trichlorophenol Ex=260, Em=310

    Conclusions

    Fluorescence detection is the most sensitive HPLC

    detector for PNA analysis. Detection limits are in the

    low ppt range.

    Simultaneous multi wavelength detection can replace

    time-programmed wavelength switching.

    Spectral mode fluorescence detection makes PNA

    methods more reliable, as it can provide information on

    additives and impurities not seen in single-wavelength

    detection.

  • 7/27/2019 Fluorescence Detection in Liquid Chromatography (Agilent Technologies)

    33/54

    33

    Carbamates Pesticides are regarded as essential to protect the qualityof food during production, storage and distribution. The

    persistence of these chemicals requires monitoring of all

    major pesticides in crops as well as the environment.

    Carbamates are mainly used as insecticides on fruits and

    vegetables.

    According to EPA method 531.1, carbamates like the

    herbicide glyphosate need a postcolumn derivatization step.

    To meet the specified limits of detection, a fluorophor is

    attached to compounds separated on a dedicated column.

    Ortho-Phthalaldehyde is the reagent used most frequently.

    Experiments and results

    Carbamates are separated on dedicated columns. After

    hydrolysis of the compounds in the effluent, the derived

    methylamines react with an o-Phthalaldehyde/Thiofluor

    solution to the corresponding isoindole.

    Figure 14Excitation (Ex=450 nm) and emission spectra (Em=330 nm) of the OPA-derivativeof a carbamate

    ConditionsColumn Pickering Carbamate

    column, 150 4.6 nm,Mobile phase A = water, B = methanolGradient 0 min, 1 5 % B,

    29 min, 100 % B,Flow rate 1.0 ml/minColumntemperature 42 CInjectionvolume 1 lFLD settings PMT 12,

    response time 4 s,step size 5 nm,

    Post-column conditions in thePickering system:OPA reagent for derivatization, reactor forhydrolysis at 100 C, 23 sec dwell time,Derivatization: Ambient, 100 l, 4 s dwell

    timeWavelength [nm]

    Norm.

    250 300 350 400 450 500

    0

    0.5

    1.0

    1.5

    2.0

    2.5

    3.0330 nm

    450 nm

    Excitation Emission

    Environmentalapplica

    tions

  • 7/27/2019 Fluorescence Detection in Liquid Chromatography (Agilent Technologies)

    34/54

    34

    Environmentalapplica

    tions

    From the collected fluorescence spectra the lowest limit of

    detection is achievable when the detector is adjusted to

    Ex 230 nm/Em 450 nm or Ex 330 nm/Em 450 nm (figure 14).

    The individual carbamate residues do not shift the excita-

    tion and emission maxima significantly. The latter choice

    is slightly different from the literature where Ex 330 nm/

    Em 465 nm is proposed. (ref. 5) Figure 15 compares the two

    wavelength settings found with the Agilent 1100 Series FLD.

    Watching the ratio of the two signals (Ex 230 nm/Em 450 nm

    and Ex 330 nm/Em 450 nm) allows control of peak purity.

    In a post-column reaction, the reagent is constantly flowing

    through the detector. To measure selectively the fluores-

    cence of the derivatized carbamates, it is important to

    characterize the fluorescent properties of the reagent,

    impurities and the derivatized compounds completely.

    The fluorescence of the OPA-reagent in the mobile phase

    water/methanol is visible in a three-dimensional plot from afluorescence scan (shown in figure 16). Excitation at 330 nm

    and emission at 450 nm are the best choices to achieve the

    Figure 15Analysis of carbamates at two differentexcitation wavelengths, 230 and 330 nm

    ConditionsSee figure 14.

    12

    4

    53

    1 Aldicarb sulfonate

    2 Aldicarb sulfoxide

    3 Oxamyl

    4 Methomyl

    5 3-OH-Carbofuran

    6 Aldicarb

    6

    78

    9

    10

    7 Propoxur

    8 Carbofuran

    9 1-Naphthol

    10 Carbaryl

    11 Methiocarb

    12 BDMC

    0 5 10 15 20 25 30 35 40

    5

    10

    15

    20

    25

    30

    35

    40

    230/450 nm

    330/450 nm

    2.5 ng

    LU

    Time [min]

    1211

    45

  • 7/27/2019 Fluorescence Detection in Liquid Chromatography (Agilent Technologies)

    35/54

    35

    ConditionsSee figure 14.

    Figure 16Characterization of the mobile phase,including OPA reagent for carbamatesThe wavelength pair (Ex 330 nm/Em 450 nm)for quantitation of carbamates is separa-ted from the reagents fluorescence.Chromatographic conditions wereestablished and the fluorescence scanwas taken under stop-flow conditions.

    Food samples are complex matrices that need some sample

    preparation steps. In contrast, water can be clean enough to

    be injected directly. With high injection volumes, detection

    limits down to 100 ppt (S/N > 2) can be achieved (figure 17).

    Figure 17

    Analysis of carbamate standard (250 ppt)in waterWith an injection volume of 800 to 900 l, adetection limit of 100 ppt can be achieved(S/N > 2).

    Environmentalapplica

    tions

    lowest limits of detection when monitoring selectively

    derivatized compounds independent of the background from

    the reagent.

    LU

    Time [min]

    10 15 20 25 30 35

    14

    15

    16

    17

    18

    19

    330/450 nm

    900 l

    800 l

    1

    24

    5

    3

    1 Aldicarb sulfonate2 Oxamyl3 Methomyl4 3-OH-Carbofuran5 Aldicarb

    6 7 8 9

    10

    6 Propoxur7 Carbofuran8 1-Naphthol9 Carbaryl10 Methiocarb

    390

    330

    220280 450 480

  • 7/27/2019 Fluorescence Detection in Liquid Chromatography (Agilent Technologies)

    36/54

    36

    Environmentalapplica

    tions

    Conclusions

    Optimized wavelength settings allow superior limits

    of detection. Changes in chromatographic conditions may

    change optimum wavelengths.

    Simultaneous multi wavelength detection enables

    chromatographers to control peak purity by watching

    the ratio of the two signals.

    Information on reagent properties and possible fluores-

    cent impurities ensures reliable quantitative results.

  • 7/27/2019 Fluorescence Detection in Liquid Chromatography (Agilent Technologies)

    37/54

    37

    Glyphosate

    Environmentalapplica

    tions

    One of the most often used non-selective and post-emer-

    gence herbicides today is glyphosate (commonly found

    under the retail names of Roundup and Basta). Because

    of an increasing number of plants that are genetically

    engineered to resist glyphosate, the quantities applied may

    increase. Monitoring of glyphosate in soil, food and water is

    therefore becoming mandatory.

    The method of choice for analyzing glyphosate and its meta-

    bolite aminomethyl phosphonic acid (AMPA) is postcolumn

    derivatization based on a two-step mechanismoxidation

    with hypochlorit and reaction with o-phthalaldehyde. (ref.6)Several standard methods are currently being optimized,

    including DIN 38 407 F22 in Germany.

    Experiments and results

    Similar to the carbamate analysis, the optimum wavelengths

    for derivatizing glyphosate and AMPA are 230 nm excitation

    and 450 nm emission, as shown in figure 18.

    Because of the very high fluorescence background from the

    hypochlorit solution and OPA, the excitation wavelength

    340 nm results in a signal-to noise ratio two times higherthan the excitation wavelength 230 nm (see figure 19),

    although Ex 230 nm gives a much higher intensity.

    The limit of detection can be lowered to 5 pg (S/N>2).

    Depending on the injection volume, this is equivalent to

    0.5 g/l (500 ppt) or even lower (see figure 20).

  • 7/27/2019 Fluorescence Detection in Liquid Chromatography (Agilent Technologies)

    38/54

    38

    Environmentalapplica

    tions

    Conditions for figures 18, 19 and 20Column Pickering Glyphosate

    column, 150 4.6 nm,Mobile phase A, 5 mM KH

    2PO

    4, pH 2.0,

    B, 100 % 5 mM potassiumhydroxide

    Gradient 015 min, 100 % A,1517 min, 100 % B,1725 min, 100 % A,

    Flow rate 0.4 ml/minColumntemperature 55 CFLD settings PMT 12,

    response time 4 s,step size 5 nm,

    Post-column conditions in thePickering system:0.3 ml/min Pickerings hypochloritereagent for oxidation, 0.3 ml/min ofPickerings OPA reagent for derivatization,reactor for oxidation at 36 C, 500 l,43 s dwell timeDerivatization: Ambient, 100 l,4 s dwell time

    Wavelength [nm]

    Norm.

    200 250 300 350 400 450

    0

    25

    50

    75

    100

    125

    150

    175

    200

    Excitation Emission340 nm

    450 nm230 nm

    Time [min]

    Norm.

    0 2.5 5 7.5 10 12.5 15 17.5

    200

    400

    600

    800

    1000

    230/450 nm

    340/450 nm

    Figure 18Fluorescence spectra of a glyphosate derivative10 ng/ml glyphosate dissolved in water).

    Figure 19Analysis of glyphosate and AMPA atdifferent wavelengthsExcitation at 340 nm offers a lower noisecompared to Ex 230 nm.

  • 7/27/2019 Fluorescence Detection in Liquid Chromatography (Agilent Technologies)

    39/54

    39

    Environmentalapplica

    tions

    Time [min]

    LUGlyphosate

    AMPA

    0 2.5 5 7.5 10 12.5 15 17.5 20

    43.5

    43.6

    43.7

    43.8

    43.9

    44.0

    44.1

    44.2

    340/450 nmFigure 20Trace level analysis of glyphosate andAMPAInjection of 500 ppt glyphosate andAMPA with a10-l injection volume(5 pg absolute)

    Conclusions

    Excitation of glyphosate can be achieved in the UV

    or visible range.

    Excitation at the 340 nm offers better selectivity because

    excitation at 230 nm gives high background noise andresults in higher limits of detection.

    Glyphosate and AMPA can be detected down to the

    ppt level.

  • 7/27/2019 Fluorescence Detection in Liquid Chromatography (Agilent Technologies)

    40/54

    40

    Foodapplica

    tions

    Aflatoxins and ochratoxin A belong to a large family of

    compounds produced by funghi. These mycotoxins are

    highly toxic compounds to protect themselves against other

    organisms. All kinds of plant tissue can be growth media for

    funghi, and therefore all types of food can be contaminated

    with mycotoxins. Storage conditions define the extent of

    fungal growth.

    Aflatoxins are known to cause degradation of fruits and

    vegetables. Ochratoxine is the prominent mycotoxin found

    in cereals, flour and figs. Because of the carcinogenic,

    teratogenic and mutagenic character of mycotoxins, foodsamples require careful control down to trace levels.

    A suitable clean-up procedure and optimized fluorescence

    or mass spectrometric detection are fundamental in achiev-

    ing the required limits of detection in the low parts per

    billion (ppb) range.

    Experiments and results

    Two FLD runs produce optimum excitation and emission

    wavelengths, as described in chapter 2. The fluorescence

    spectra shown in figure 21 illustrate the result. Both the

    B- and G-type aflatoxins show similar spectra. The optimum

    excitation wavelength for both is 365 nm. The optimum

    emission wavelengths are different, 455 nm for the G-type

    and 445 nm for the B-type. These wavelengths deviate from

    the literature (ref. 7), which may be due to differences in

    experimental conditions such as the pH, the eluent compo-

    sition or the instrumentation used.

    With the optimized conditions listed below, the limit of

    detection is down in the low nanogram range, as shown in

    figure 22. This means that a sample of only a few grams can

    be sufficient to detect at the low ppb level.

    Mycotoxins

    Aflatoxins B1/B2 andG1/G2

  • 7/27/2019 Fluorescence Detection in Liquid Chromatography (Agilent Technologies)

    41/54

    41

    Foodapplications

    Figure 21Fluorescence spectra for aflatoxins G2and B2Reference standards were dissolved inmethanol (step size 5 nm).

    Time [min]

    LU

    0 2 4 6 8 10 12 14 16 18

    2

    4

    6

    8

    10

    12

    365/445 nm

    365/460 nm

    0.3 ng G2

    1 ng G1

    0.3 ng B2

    1 ng B1

    ConditionsColumn Hypersil ODS 100*2.1 mm,

    3 mMobile phase H

    2O/MeOH/CH

    3CN =

    63/26/11Flow rate 0.3 ml/minColumntemperature 25 CInjectionvolume 5 lFLD settings PMT 12,

    response time 4 s,step size 5 nm,

    Figure 22Analysis of aflatoxins at two differentwavelengths

    Wavelength [nm]

    Norm.

    200 250 300 350 400 450 500

    0

    2

    4

    6

    8

    Excitation Emission

    365 nmAflatoxin B29

    445 nm

    Norm. 365 nmAflatoxin G2 455 nm

    200 250 300 350 400 450 500

    -1

    0

    1

    2

    3

    4

    5

    Excitation Emission

    Time [min]

  • 7/27/2019 Fluorescence Detection in Liquid Chromatography (Agilent Technologies)

    42/54

    42

    Foodapplica

    tions

    Ochratoxin A is found in degrading plant materials as a

    product of aspergillus or penicillium funghi. Through the

    food chain, this compound may become enriched in animal

    tissue and act as a cancerogenic. It can also be a substance

    that is directly toxic.

    Experiments and results

    The chromatogram in figure 23 shows a 125 pg of ochra-

    toxine A at two different excitation wavelengths, 230 nm

    and 333 nm, with lower limits of detection at 230 nm but

    better selectivity over the matrix background at 333 nm.

    Ochratoxin A

    ConditionsColumn Zorbax SB C18, 150 2 mm,

    3.5 mMobile phase Water/acetonitrile = 50/50Flow rate 0.4 ml/minColumntemperature 40 CInjectionvolume 5 lFLD settings PMT 12,

    response time 4 s,step size 5 nm

    Figure 23Simultaneous multi wavelength detectionfor ochratoxine analysis125 pg ochrotoxine was injected. Themost appropriate wavelength can bechosen for quantitation depending onmatrix conditions. Conclusions

    Optimization of wavelength settings is mandatory if any

    change in experimental setup or chromatographic

    conditions occurs.

    The ability to detect simultaneously at multiple wave-

    lengths helps to obtain the best limits of detection and

    selectivity for the different species of aflatoxins.

    Ochratoxine A may be excited in the UV or visible range

    depending on the need for a higher signal or a more stable

    baseline.

    Time [min]

    LU

    0 2 4 6 8 10 12 14

    0.5

    0.6

    0.7

    0.8

    0.9

    1.0

    1.1

    1.2

    333/460 nm

    230/460 nm

    Ochratoxin A

  • 7/27/2019 Fluorescence Detection in Liquid Chromatography (Agilent Technologies)

    43/54

    43

    Foodapplica

    tions

    Vitamins B2 and B6 Vitamins are essential to avoid human and animal malnutri-tion. Improper storage may rapidly damage the natural

    vitamin content of food products or supplemental pharma-

    ceutical formulations.

    Vitamins are classified as either water-soluble or fat-soluble

    compounds. Each is analyzed by different methods. Among

    the water-soluble vitamins, only B2 (riboflavin and phospho-

    rylated riboflavins) and B6 (pyridoxamine, pyridoxal and,

    pyridoxine) show fluorescence. All compounds are sepa-

    rated in a single run. With previous fluorescence detectors

    they were detected using time-programmed wavelengthswitching. (ref.8)

    Experiments and results

    The optimal wavelengths for B6 vitamins are Ex 270/Em

    400 nm and for vitamin B2 Ex 270/Em 530nm, as shown in

    figure 24. The detection limit for B2 is down to 20 pg; the

    detection limit for B6 is 200 pg (S/N > 2).

    Figure 24Excitation and emission spectra ofvitamins B2 and B6Spectra were extracted from two FLD runsfor excitation and emission spectra(1 g/ml dissolved in water, step size 5 nm).

    Norm.

    Wavelength [nm]

    200 250 300 350 400 450 500 550

    05

    1015202530

    35

    Excitation Emission

    450 nm 530 nm

    362 nm262 nm

    Norm.

    Wavelength [nm]

    200 250 300 350 400 450 500 550

    05

    101520

    2530

    35

    Excitation Emission

    400 nm280 nm

  • 7/27/2019 Fluorescence Detection in Liquid Chromatography (Agilent Technologies)

    44/54

    44

    Foodapplica

    tions

    ConditionsColumn Zorbax SB 50 2.0 mm,

    5 mMobile phase A = 0.005 mM KH

    2PO

    4,

    pH 2.5 - H2SO

    4,

    B = acetonitrileGradient 025 % B in 10 minFlow rate 0.5 ml/minColumntemperature 35 CInjectionvolume 5 lFLD settings Response time 4 s

    During metabolism as well as during sample preparation

    from complex matrices, vitamins are subject to various

    transformations. The goal is therefore to monitor precursors

    and modified species as well as the main compounds in a

    single run. Figure 25 demonstrates this for vitamins B2 and

    B6. The small impurities close to Riboflavine (seen in figure

    26) can be identified through excitation spectra as belonging

    to the vitamin B2 complex. Reference standards help to

    identify them as monophosphates and diphosphates.

    Time [min]

    0 2 4 6 8

    LU

    0

    20

    40

    60

    80

    270/530 nm

    270/400 nm

    10

    Riboflavine

    Pryridoxamine

    Pryridoxal

    Pryridoxine

    Figure 25

    Analysis of vitamins B2 and B6Multi wavelength detection is used todetect both vitamins selectively andsensitive.

  • 7/27/2019 Fluorescence Detection in Liquid Chromatography (Agilent Technologies)

    45/54

    45

    Foodapplica

    tions

    Time[min]

    0 2 4 6 8 10 12 14

    LU

    -10

    0

    10

    20

    30

    40

    270/530 nm

    Wavelength [nm]250 300 350 400 450

    Norm.

    0

    10

    20

    30

    40

    50

    60

    70

    80

    1 Pyridoxamine

    2 Pyridoxal

    3 Pyridoxine

    4 Riboflavine-diphosphate

    5 Riboflavine-5'monophosphate

    6 Riboflavine

    12 3

    4

    56

    ConditionsSee figure 25.

    Figure 26Confirmation of byproducts of vitamin B2FLD excitation spectra (the red line is5monophosphat, the blue line is riboflavin)show the similarity of spectra from thephosphorylated and unphosphorylatedriboflavin (20 g/ml dissolved in mobilephase A).

    Conclusions

    The fluorescent vitamins B2 and B6 can be analyzed

    selectively at the pg level.

    Because of the difference in excitation and emission

    maxima, simultaneous multi wavelength detection is

    essential to detect both vitamins and byproducts at trace

    levels.

    Online spectra ensure confirmation of minor compounds

    throughout the complete HPLC run.

  • 7/27/2019 Fluorescence Detection in Liquid Chromatography (Agilent Technologies)

    46/54

    46

    Pharm

    aceuticalapplica

    tions

    Quinidine occurs as two stereoisomers and has been used in

    anti-malaria and antiarrhythmic drugs since the beginning of

    the twentieth century. Before that, American Indians used

    plant material containing quinidine for its antipyretic activity.

    This compound has a complex chemical structure containing

    a fluorophor and groups (tert. amine) that can be protonated

    (illustrated in figure 27). An HPLC separation with well

    shaped peaks requires a buffered eluent as described in the

    literature. (ref. 9) With a new generation of more stable

    column materials (such as Zorbax Stablebond), this com-

    pound has to be analyzed under acidic conditions. As muchas retention times change with the type of stationary phase,

    a different pH of the mobile phase changes UV absorption

    and fluorescence behavior.

    Experiments and results

    As pH decreases, quinidine shows a bathochrome shift:

    The emission wavelength changes from 380 nm to 450 nm

    as the pH shifts from pH 7 to pH 2.5 (shown in figure 28).

    This is important in maintaining optimum limits of detection.

    If slight changes in pH occur, these can contribute to

    non-reproducible separation and quantitation results.Recording spectra can help to view these changes. Conse-

    quently, spectral data can be used in routine work to check

    separation conditionsa prerequisite for reproducing

    quantitative results.

    Quinidine

    Figure 27Chemical structure of quinidineThe arrows indicate the functionalgroups that can be protonated.

    N

    N

    HHO

    CH3O

  • 7/27/2019 Fluorescence Detection in Liquid Chromatography (Agilent Technologies)

    47/54

    47

    Pharm

    aceuticalapplications

    Conditions

    Method 1 pH 7Column Purosphere 125 4 mm,

    5 mMobile phase H

    2O pH 7 (H

    2SO

    4)/

    acetonitrileFlow rate 0.6 ml/minColumntemperature 40 CInjectionvolume 1 lFLD settings PMT 12,

    response time 4 s,

    step size 5 nm

    Method 2 pH 2.5Column BDS 100 2.1 mm,

    3 mMobile phase H

    2O pH 2.5 (H

    2SO

    4)/

    acetonitrileFlow rate 0.3 ml/minColumntemperature 40 CInjectionvolume 1 lFLD settings PMT 12,

    response time 4 s,step size 5 nm

    Norm.

    Wavelength [nm]

    350 400 450 500 550 600

    0

    5

    10

    15

    20

    25

    30

    Emission

    pH 2.5pH 7

    450 nm380 nm35

    Emission

    Figure 28Influence of pH on emission spectra for quinidine

    Conclusions

    Fluorescence spectra are an excellent indicator of

    pH changes in the eluent.

    Quantitation of acidic or basic type substances requires

    strict control of pH not only to control retention times

    but for reliable quantitation as well.

  • 7/27/2019 Fluorescence Detection in Liquid Chromatography (Agilent Technologies)

    48/54

    48

    Pharm

    aceuticalapplica

    tions

    Warfarin Warfarin is an anticoagulant drug that is used in post-surgerytreatment. The chemical structure derives from coumarine

    and has phenolic character.

    In the literature, warfarin is analyzed mainly at neutral condi-

    tions (phosphate buffer pH 7.5) with fluorescence measured

    at Ex 290 nm/Em 390 nm or Ex 310 nm/Em 370 nm. (ref.10)

    Experiments and results

    Warfarin can be analyzed on a Zorbax Stablebond columnat pH 2.5. Under these conditions, excitation and emission

    maxima shifted to 272 nm and 355 nm, respectively (shown

    in figure 29).

    The resulting comparison between the analysis based on

    literature and actual optimized fluorescence wavelengths

    is shown in figure 30. Actual optimized fluorescence wave-

    lengths show about three times better limits of detection.

    Compared to UV detection, the Agilent 1100 Series FLD has

    a limit of detection about 20 times lower.

    Wavelength [nm]

    200 250 300 350 400 450 500

    Norm.

    0

    0.25

    0.50

    0.75

    1.00

    1.25

    1.50

    1.75

    Excitation Emission

    272 nm 355 nmConditionsColumn Zorbax SB C18, 50 2.1 mm,

    3 mMobile phase 0.005 M KH

    2PO

    4,

    pH 3/acetonitrileGradient 20 % B to 80 % B in 10 minWash 80 % B to 20 % B in 2 minColumntemperature 25 CInjectionvolume 1 lFLD settings PMT 12,

    response time 4 s,step size 5 nm

    Figure 29

    Fluorescence spectra for warfarin underacidic conditions (pH=3)10 g/ml dissolved in phosphate buffer.

  • 7/27/2019 Fluorescence Detection in Liquid Chromatography (Agilent Technologies)

    49/54

    49

    Pharm

    aceuticalapplications

    Time [min]

    LU

    0 1 2 3 4 5 6 7 8 9

    0.5

    1.0

    1.5

    2.0

    2.5

    3.0

    3.5

    272/355 nm

    290/390 nm

    accord. Lit.

    10 ng Warfarin

    Figure 30Analysis of warfarinResponse under literature conditions(ref. 10) and actual optimized fluorescencewavelengths.

    ConditionsSee figure 29.

    Conclusions

    New types of columns that produce different retention

    behavior require a rework of chromatographic conditions.

    This may induce significant shifts in fluorescence spectra,

    which in turn can influence the limits of detection with

    the fluorescence detector.

    In routine analysis, fluorescence spectra can be taken

    automatically and reviewed to check for accidental

    changes in chromatographic conditions.

  • 7/27/2019 Fluorescence Detection in Liquid Chromatography (Agilent Technologies)

    50/54

    50

    Amino acids are essential components for biological

    systems and have been found already during early stages of

    life on earth. For human nutrition a set of essential amino

    acids is needed that cannot be formed in the body but must

    be taken up from the daily food. As building blocks of the

    proteins they form the vast majority of all enzymes in

    biochemical transformations. A wide range of decease and

    drug research is targeted on protein biochemistry to under-

    stand the role of amino acids, peptides and proteins.

    After acidic or enzymatic hydrolysis amino acid composi-

    tion can be determined using reversed phase columns which

    will give typically a better resolution than ion chromatogra-

    phy. For sensitive detection amino acids are derivatized

    in a two-step pre-column derivatization. A post-column

    derivatisation is suitable for higher selectivity as required

    for complex biological samples. Fluorescence detection is

    used for concentrations below 100 pmol/l. UV-detection

    is a choice for higher concentrations up to the nanogram

    range. A standard HPLC system can be used for this straight-

    forward and cost-effective approach to amino acid analysis.

    Experimental and resultsPre-column derivatisation of primary amino acids is

    achieved with ortho-phthalaldehyde (OPA) and 9-fluorenyl-

    methylchloroformate (FMOC) is used for secondary amino

    acids. Seventeen different amino acids that are all found in

    protein hydrolysates were analyzed as shown in figure 31.

    A wavelength switching program is used to detect Proline.

    The limits of detection for the amino acids are listed in

    table 4 based on a signal-to-noise > 2. This is close to 100

    times more sensitive than UV-detection at 338nm. Retention

    time precision is typically below 0.2 % and peak area

    precision is typically below 5 %.

    Pharm

    aceuticalapplica

    tions

    Amino Acids

    Compound LOD for FLD FLDFLD

    (pmol) (fmol)

    Asp 19 0.139 0.924

    Glu 18 0.155 0.576Ser 21 0.156 1.015His 29 0.155 1.778Gly 21 0.118 1.124Thr 21 0.113 0.739Ala 20 0.120 0.767Arg 17 0.094 0.905Tyr 19 0.062 1.614Cys-SS-Cys not measuredVal 17 0.058 0.919Met 16 0.045 1.236Phe 17 0.048 1.079Ile 16 0.050 0.759Leu 18 0.040 0.952Lys 57 0.060 5.107Pro 22 0.044 4.379

    Table 4LODs for fluorescenceand UV detection

  • 7/27/2019 Fluorescence Detection in Liquid Chromatography (Agilent Technologies)

    51/54

    51

    Pharm

    aceuticalapplica

    tions

    Conclusion

    Reversed phase chromatography combined with pre-

    column derivatization is an ideal tool for automated cost-

    effective amino acid analysis on a standard HPLC setup.

    With the use of OPA and FMOC and fluorescence

    detection it is possible to push the limits of detection

    signifantly below 100 femtomoles for most amino acids.

    This technique is worked out to offer precise results for

    protein hydrolysates in a routine laboratory.

    Time [min]0 2 4 6 8 10 12 14 16 18

    LU

    0

    25

    50

    75

    100

    125

    175Asp

    Glu

    Ser

    His

    Gly

    Thr A

    la

    Arg

    Try

    C

    ys-S

    S-Cys

    Val

    Met

    Phe

    Ile

    Leu

    Lys

    Pro

    150

    10 pmol standardConditionsColumn 200 2.1 mm AA column

    and guard columnMobile phase A = 20 mMol NaAc + 0.018%

    TEA adjusted to pH 7.2 with1-2 % acetic acid, B = 20 %of 100 mMol NaAc adjustedto pH 7.2 with 1-2 % aceticacid + 40 % ACN and 40 %MeOH

    Flow rate 0.45 ml/minGradient start with: 100 %A, at 17 min

    60 %B, at 18 min, 100 %B,at 18.1 min flow 0.45, at18.5 min flow 0.8, at 23.9 minflow 0.8, at 24 min 100%Band flow 0.45, at 25 min 0%B

    Oven temp. 40 CPost time 5 minInjector program1 Draw 5.0 l from vial 10borate buffer2 Draw 1.0 l from vial 11OPA reagent3 Draw 0.0 l from vial 12water4 Draw 1.0 l from sample5 Draw 0.0 l from vial 12water6 Mix 8 l in air, max speed, six times7 Draw 1.0 l from vial 14FMOC8 Draw 0.0 l from vial 12water9 Mix 9 l in air, max speed, 3 times10 InjectFLD settings Excitation 340 nm

    Emission 450 nmPTM gain 12

    at 14.5 minExcitation =266 nmEmission = 305 nmPTM Gain 11

    Figure 31Analysis of 10 pmol/l amino acids with fluorescence detection

  • 7/27/2019 Fluorescence Detection in Liquid Chromatography (Agilent Technologies)

    52/54

    52

    Refere

    nces

    1. Lakowicz, J.R., Principles of Fluorescence Spectros-

    copy,Plenum Press, New York, 1983.

    2. Schuster, R., A comparison of pre- and post-column

    derivatization for the analysis of glyphosate,

    Agilent Technologies Application Note 5091-3621E, 1991.

    3. Froehlich, P.,Internatl Laboratory, No. 10, 42-44,1989.

    4. Brownrigg, J.T. and Sullivan, M. J.,Spectroscopy, Vol.1,

    No.2, 1989.

    5. Pickering Laboratories,Product brochure, Publ. No.

    B-CA 4, 1992.

    6. Pickering Laboratories,Application Note, Publ. No.

    B-CA 5, 1993.

    7. Official Methods of Analysis, Food Compositions;

    Additives, Natural Contaminants, 15th edition: AOAC:

    Arlington, VA, Vol. 2.; AOAC Official method 980.20,

    aflatoxins in cotton seed products, 1990.

    8. Mc Calley, D.V., J. Chrom., 357, 221, 1986.

    9. Chu, Y-Q. and Wainer, I.W.,Pharm. Res. 5, 680, 1988.

    10. Gratzfeld-Huesgen, A. and Schuster, R., HPLC for Food

    Analysis,Agilent Technologies Application Primer,

    5968-9345E, 2000.

    11. Gratzfeld-Huesgen, A., Sensitive and Reliable Amino

    Acids Analysis in Protein Hydrolysates using the Agilent

    1100 Series HPLC,Agilent Technologies Technical Note

    5968-5658E, 1999.

  • 7/27/2019 Fluorescence Detection in Liquid Chromatography (Agilent Technologies)

    53/54

    53

    Index

    Aaflatoxins, 4041

    B1/B2, 4041G1/G2 4041

    amino acids, 50AMPA, 3739antioxidant, 31anthracene, 14, 22, 26applications, 29ASTM, 142-amino-3-hydroxyphenazine (AHP),23,24

    B

    bandwidth, 10112-benzimidazole carbamic acidmethylester (MBC), 23, 24

    Ccarbamates, 3336cut-off filter, 11, 17cuvette, 12, 17, 19, 20chromophore, 10

    Ddata handling, 1213data analysis, 13derivatization, 33detector

    diode array, 6, 13, 23, 26, 31mass spectrometric, 6, 26, 40

    2,3-diamino-phenazine (DAP), 23, 24DIN method, 30, 37

    EEPA methods, 30, 33emission

    monochromator, 1112spectra, 6, 1014, 1627, 33, 43, 46, 47wavelength, 1014, 1625, 40

    excitationmonochromator, 1112spectra, 6, 1014, 1626, 33, 43wavelength, 1014, 1625, 40,

    Ffluorescence spectra, 13, 19, 25, 34, 38,40, 41, 48, 49

    fluorescence scan, 12, 17, 21, 34fluorophore, 21, 33, 46

    Ggerman standard method DIN, 30, 37glyphosate, 3739grating, 1112

    Iimpurities, 17, 23, 25isofluorescence plot, 13, 17, 22

    Llamp, 11

    Xenon, 1112light sources, 23limit of detection (LOD), 12, 1618, 23, 25,

    30, 31, 33, 34, 35, 36, 37, 39, 40, 42, 43, 46, 48,

    Mmass spectrometric, 6, 26, 40match factor, 27method development, 16272-benzimidazole carbamic acidmethylester (MBC), 23, 24mobile phase, 17, 34, 35, 46monochromator, 1112multi wavelength detection, 16, 25, 3132,36, 42, 4445,mycotoxins, 40

    Oo-phthalaldehyde (OPA), 3335

    ochratoxin A, 42

    Ppeak, 2527

    purity, 13, 27, 34identity, 26

    pesticides, 33photo-multiplier, 11photodiode, 11Pickering, 38polynuclear aromatic hydrocarbons(PNA), 21, 25, 3032precolumn derivatization,postcolumn derivatization, 33, 37purity factor, 27pyridoxal, 43, 44pyridoxamine, 43, 44pyridoxine, 43, 44

    Qqualitative analysis, 24quantitative analysis, 2526quinidine, 18, 19, 4647

    RRaman, 14,17Raleigh, 17riboflavine, 4344

    Sselectivity, 7, 18, 23, 25specifications, 14spectral libraries, 13

    spectral mode, 67, 32spectral range, 21stray light, 1112

    Raman, 11, 17Raleigh, 17

    Tthree-dimensional plot, 12, 13, 17, 21, 34trichlorophenol, 31, 32

    UUV/Visible absorption spectra, 6

    Vvitamins, 4345

    B2/B6, 4345

    Wwarfarin, 48wavelength, 12

    XXenon flash lamp, 1112

    ZZorbax, 46, 48

  • 7/27/2019 Fluorescence Detection in Liquid Chromatography (Agilent Technologies)

    54/54

    Copyright 2000 Agilent Technologies

    The information in this publication is

    subject to change without notice.

    All Rights Reserved. Reproduction,

    adaptation or translation without prior

    written permission is prohibited, except

    as allowed under the copyright laws.

    Printed in Germany 01/00

    Publication Number 5968-9346E

    www.agilent.com/chem