Fat and water separation techniques in NMR imaging NMR imaging Noura Azzabou 30 April 2013

Upload: others

Post on 27-Jun-2020

2 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Fat water Separation - lirmm.frstrauss/PageImage3/Fat_water_Separation.pdf · Fat and water separationtechniques : 2 point Dixon Acquisition : Echos time selection TEi= 1/447 s =

Fat and water separation techniques in

NMR imagingNMR imaging

Noura Azzabou

30 April 2013

Page 2: Fat water Separation - lirmm.frstrauss/PageImage3/Fat_water_Separation.pdf · Fat and water separationtechniques : 2 point Dixon Acquisition : Echos time selection TEi= 1/447 s =

A vector representation of the complex signal S for a a given

pixel with two spectral components, water (W) and fat (F).

Fat and water separation techniques : chemical shift principle

The phase angle of fat relative to

that of water due to their chemical

shift difference

The error phase due to the magnetic

field in homogeneity

another error phase due to other system imperfections

such as the spatial dependence of RF penetration and

different signal delay in the receiver chains.

Page 3: Fat water Separation - lirmm.frstrauss/PageImage3/Fat_water_Separation.pdf · Fat and water separationtechniques : 2 point Dixon Acquisition : Echos time selection TEi= 1/447 s =

A vector representation of the complex signal S for a a given

pixel with two spectral components, water (W) and fat (F).

Fat and water separation techniques : chemical shift principle

The phase angle of fat relative to

that of water due to their chemical

shift difference

The error phase due to the magnetic

field in homogeneity

another error phase due to other system imperfections

such as the spatial dependence of RF penetration and

different signal delay in the receiver chains.

Page 4: Fat water Separation - lirmm.frstrauss/PageImage3/Fat_water_Separation.pdf · Fat and water separationtechniques : 2 point Dixon Acquisition : Echos time selection TEi= 1/447 s =

Assuming that water-fat chemical shift (σσσσ) is 3.5 ppm, and B0 = 3T

3.5 ppm @ 3T = σ B0 γ = 3.5 x 3 x 42.58 MHz/T= 447 Hz

relative phase = = TE x 447 x 2π

w

f

w

f

Fat and water separation techniques : 2 point Dixon

Acquisition : Echos time selection

TEi= 1/447 s = 2.24 ms

TEi= 2/447 s = 4.48 ms

(“in phase” condition)

TEo= 1.5/447 s = 3.36 ms

TEo= 3/447 s = 6.71 ms

(“opposed phase” condition)

f

Page 5: Fat water Separation - lirmm.frstrauss/PageImage3/Fat_water_Separation.pdf · Fat and water separationtechniques : 2 point Dixon Acquisition : Echos time selection TEi= 1/447 s =

Assuming a prior knowledge about fat water content and that for example there is

Fat and water separation techniques : 2 point Dixon

Water and Fat signal reconsutruction

Assuming a prior knowledge about fat water content and that for example there is

more fat than water in the pixel we can write

In the opposite case we have

Page 6: Fat water Separation - lirmm.frstrauss/PageImage3/Fat_water_Separation.pdf · Fat and water separationtechniques : 2 point Dixon Acquisition : Echos time selection TEi= 1/447 s =

Fat and water separation techniques : 2 point Dixon

Reconstruction results

Fat ratio map, for the thigh of healthy volunteer.

Page 7: Fat water Separation - lirmm.frstrauss/PageImage3/Fat_water_Separation.pdf · Fat and water separationtechniques : 2 point Dixon Acquisition : Echos time selection TEi= 1/447 s =

Fat and water separation techniques : 2 point Dixon

Reconstruction results

Fat ratio map, for the thigh of healthy volunteer. For the one leg water and

fat signal are swapped

Page 8: Fat water Separation - lirmm.frstrauss/PageImage3/Fat_water_Separation.pdf · Fat and water separationtechniques : 2 point Dixon Acquisition : Echos time selection TEi= 1/447 s =

Fat and water separation techniques : 3 point Dixon

Reconstruction technique

After computing B0 inhomogenieties, fat and water signal are computed through

linear equation resolving.

Page 9: Fat water Separation - lirmm.frstrauss/PageImage3/Fat_water_Separation.pdf · Fat and water separationtechniques : 2 point Dixon Acquisition : Echos time selection TEi= 1/447 s =

Fat and water separation techniques : 3 point Dixon

Phase wrapping problem

Phase of the first in-phase image S1

Phase of the second in-phase image S3

Difference between phases.

Page 10: Fat water Separation - lirmm.frstrauss/PageImage3/Fat_water_Separation.pdf · Fat and water separationtechniques : 2 point Dixon Acquisition : Echos time selection TEi= 1/447 s =

Fat and water separation techniques : 3 point Dixon

Phase wrapping problem

10

Page 11: Fat water Separation - lirmm.frstrauss/PageImage3/Fat_water_Separation.pdf · Fat and water separationtechniques : 2 point Dixon Acquisition : Echos time selection TEi= 1/447 s =

Fat and water separation techniques : 3 point Dixon

Phase wrapping problem

Page 12: Fat water Separation - lirmm.frstrauss/PageImage3/Fat_water_Separation.pdf · Fat and water separationtechniques : 2 point Dixon Acquisition : Echos time selection TEi= 1/447 s =

Fat and water separation techniques : 3 point Dixon

Example of reconstructed images

in-phase opposed-phase F W

Page 13: Fat water Separation - lirmm.frstrauss/PageImage3/Fat_water_Separation.pdf · Fat and water separationtechniques : 2 point Dixon Acquisition : Echos time selection TEi= 1/447 s =

Fat and water separation techniques : 3 point Dixon

Quantitative analysis

Reference : Hines, C. D. G., Yu, H., Shimakawa, A., McKenzie, C. a, Brittain, J. H., & Reeder, S. B. (2009). T1 independent, T2* corrected

MRI with accurate spectral modeling for quantification of fat: validation in a fat-water-SPIO phantom. Journal of magnetic resonance

imaging: JMRI, 30(5), 1215-22.

Page 14: Fat water Separation - lirmm.frstrauss/PageImage3/Fat_water_Separation.pdf · Fat and water separationtechniques : 2 point Dixon Acquisition : Echos time selection TEi= 1/447 s =

Fat and water separation techniques : 3 point Dixon

Quantitative analysis

Reference : Hines, C. D. G., Yu, H., Shimakawa, A., McKenzie, C. a, Brittain, J. H., & Reeder, S. B. (2009). T1 independent, T2* corrected

MRI with accurate spectral modeling for quantification of fat: validation in a fat-water-SPIO phantom. Journal of magnetic resonance

imaging: JMRI, 30(5), 1215-22.

Page 15: Fat water Separation - lirmm.frstrauss/PageImage3/Fat_water_Separation.pdf · Fat and water separationtechniques : 2 point Dixon Acquisition : Echos time selection TEi= 1/447 s =

Fat and water separation techniques : 3 point Dixon

Quantitative analysis

Reference : Hines, C. D. G., Yu, H., Shimakawa, A., McKenzie, C. a, Brittain, J. H., & Reeder, S. B. (2009). T1 independent, T2* corrected

MRI with accurate spectral modeling for quantification of fat: validation in a fat-water-SPIO phantom. Journal of magnetic resonance

imaging: JMRI, 30(5), 1215-22.

Page 16: Fat water Separation - lirmm.frstrauss/PageImage3/Fat_water_Separation.pdf · Fat and water separationtechniques : 2 point Dixon Acquisition : Echos time selection TEi= 1/447 s =

voxelDixon technique is based on the chemical shift difference between water protons and lipid –CH2 protons.

Fat and water separation techniques : 3 point Dixon

Quantitative analysis

~3.5 ppm ~ 450 Hz @ 3T

Fat (-CH)

Page 17: Fat water Separation - lirmm.frstrauss/PageImage3/Fat_water_Separation.pdf · Fat and water separationtechniques : 2 point Dixon Acquisition : Echos time selection TEi= 1/447 s =

Fat and water separation techniques : Fat spectrum

Representative localized in vivo proton spectrum (PRESS: TR = 1.8 s, TE = 20 ms, voxel size = 2 mm × 2 mm × 2 mm, NEX = 500) of adipose tissue of a mouse.

Strobel K et al. J. Lipid Res. 2008;49:473-480

Page 18: Fat water Separation - lirmm.frstrauss/PageImage3/Fat_water_Separation.pdf · Fat and water separationtechniques : 2 point Dixon Acquisition : Echos time selection TEi= 1/447 s =

Fat and water separation techniques : Multi-peak Model

Page 19: Fat water Separation - lirmm.frstrauss/PageImage3/Fat_water_Separation.pdf · Fat and water separationtechniques : 2 point Dixon Acquisition : Echos time selection TEi= 1/447 s =

Fat and water separation techniques : Multi-peak Model

Page 20: Fat water Separation - lirmm.frstrauss/PageImage3/Fat_water_Separation.pdf · Fat and water separationtechniques : 2 point Dixon Acquisition : Echos time selection TEi= 1/447 s =

IDEAL - ALGORITHM

We can formulate this as a linear system with respect to fat and water signal.

In order to resolve this system we have to acquire at least M+2 images (with M total

number of fat peaks).

Page 21: Fat water Separation - lirmm.frstrauss/PageImage3/Fat_water_Separation.pdf · Fat and water separationtechniques : 2 point Dixon Acquisition : Echos time selection TEi= 1/447 s =

IDEAL - ALGORITHM

Page 22: Fat water Separation - lirmm.frstrauss/PageImage3/Fat_water_Separation.pdf · Fat and water separationtechniques : 2 point Dixon Acquisition : Echos time selection TEi= 1/447 s =

Fat and water separation techniques : IDEAL technique

Iterative reconstruction process

Page 23: Fat water Separation - lirmm.frstrauss/PageImage3/Fat_water_Separation.pdf · Fat and water separationtechniques : 2 point Dixon Acquisition : Echos time selection TEi= 1/447 s =

Fat and water separation techniques : IDEAL technique

Iterative reconstruction process : field map error estimation

Page 24: Fat water Separation - lirmm.frstrauss/PageImage3/Fat_water_Separation.pdf · Fat and water separationtechniques : 2 point Dixon Acquisition : Echos time selection TEi= 1/447 s =

Fat and water separation techniques : IDEAL technique

Iterative reconstruction process : field map error estimation

Page 25: Fat water Separation - lirmm.frstrauss/PageImage3/Fat_water_Separation.pdf · Fat and water separationtechniques : 2 point Dixon Acquisition : Echos time selection TEi= 1/447 s =

Fat and water separation techniques : IDEAL technique

Signal Decay

The decay is due to the spin-spin relaxation (T2) and an additional component due to B0

inhomogeniey. The over all decay is called T2* (R2*=1/T2*)

Page 26: Fat water Separation - lirmm.frstrauss/PageImage3/Fat_water_Separation.pdf · Fat and water separationtechniques : 2 point Dixon Acquisition : Echos time selection TEi= 1/447 s =

Fat and water separation techniques : IDEAL technique

Signal Decay

The decay is due to the spin-spin relaxation (T2) and an additional component due to B0

inhomogeniey. The over all decay is called T2* (R2*=1/T2*)

Page 27: Fat water Separation - lirmm.frstrauss/PageImage3/Fat_water_Separation.pdf · Fat and water separationtechniques : 2 point Dixon Acquisition : Echos time selection TEi= 1/447 s =
Page 28: Fat water Separation - lirmm.frstrauss/PageImage3/Fat_water_Separation.pdf · Fat and water separationtechniques : 2 point Dixon Acquisition : Echos time selection TEi= 1/447 s =

Fat and water separation techniques : IDEAL technique

Signal Model with R2* decrease

The same iterative algorithm is applied with the difference that the unknow field

map is a complex number.

Page 29: Fat water Separation - lirmm.frstrauss/PageImage3/Fat_water_Separation.pdf · Fat and water separationtechniques : 2 point Dixon Acquisition : Echos time selection TEi= 1/447 s =

Fat and water separation techniques : IDEAL technique

Page 30: Fat water Separation - lirmm.frstrauss/PageImage3/Fat_water_Separation.pdf · Fat and water separationtechniques : 2 point Dixon Acquisition : Echos time selection TEi= 1/447 s =

Fat and water separation techniques : IDEAL technique

Page 31: Fat water Separation - lirmm.frstrauss/PageImage3/Fat_water_Separation.pdf · Fat and water separationtechniques : 2 point Dixon Acquisition : Echos time selection TEi= 1/447 s =
Page 32: Fat water Separation - lirmm.frstrauss/PageImage3/Fat_water_Separation.pdf · Fat and water separationtechniques : 2 point Dixon Acquisition : Echos time selection TEi= 1/447 s =

Fat and water separation techniques : IDEAL technique

Page 33: Fat water Separation - lirmm.frstrauss/PageImage3/Fat_water_Separation.pdf · Fat and water separationtechniques : 2 point Dixon Acquisition : Echos time selection TEi= 1/447 s =

� IDEAL technique assumes that the initial guess of R2* and B0 field inohomogeniety

are close to real values.

� The reconstruction technique can lead to local minimums which can result in false

estimation of water and fat signal.

� The algorithm, is a voxel by voxel resolution, no regularization term.

Fat and water separation techniques : IDEAL technique

Iterative reconstruction process : limitations

Results of IDEAL technique on mayonnaise samples with different fat ratio : from left to

right : field map, water map and fat map.

Page 34: Fat water Separation - lirmm.frstrauss/PageImage3/Fat_water_Separation.pdf · Fat and water separationtechniques : 2 point Dixon Acquisition : Echos time selection TEi= 1/447 s =

The main objective is to

address the problem of

regularization as well as

local minimum problems local minimum problems

related to the IDEAL

resolution.

Page 35: Fat water Separation - lirmm.frstrauss/PageImage3/Fat_water_Separation.pdf · Fat and water separationtechniques : 2 point Dixon Acquisition : Echos time selection TEi= 1/447 s =

Fat and water separation techniques : VARPRO technique

10

12

14x 10

4 Error of the fit as a function of the field map value

VARPRO: reconstruction amounts to 1D function minimization

-1 -0.8 -0.6 -0.4 -0.2 0 0.2 0.4 0.6 0.8 10

2

4

6

8

dB0

err

or

Page 36: Fat water Separation - lirmm.frstrauss/PageImage3/Fat_water_Separation.pdf · Fat and water separationtechniques : 2 point Dixon Acquisition : Echos time selection TEi= 1/447 s =

Fat and water separation techniques : VARPRO technique

VARPRO: Algorithm

The algorithm is composed of two main step

1- Minimization

2- Regularization

Page 37: Fat water Separation - lirmm.frstrauss/PageImage3/Fat_water_Separation.pdf · Fat and water separationtechniques : 2 point Dixon Acquisition : Echos time selection TEi= 1/447 s =

VARPRO: comparaison with IDEAL

Fat and water separation techniques : VARPRO technique

Page 38: Fat water Separation - lirmm.frstrauss/PageImage3/Fat_water_Separation.pdf · Fat and water separationtechniques : 2 point Dixon Acquisition : Echos time selection TEi= 1/447 s =

An extension of the VARPRO approach, it performs minimization and regularization at

the same step

Page 39: Fat water Separation - lirmm.frstrauss/PageImage3/Fat_water_Separation.pdf · Fat and water separationtechniques : 2 point Dixon Acquisition : Echos time selection TEi= 1/447 s =

Fat and water separation techniques : VARPRO + Graph cut

Discrete optimization technique like graph

cut or other approaches

Page 40: Fat water Separation - lirmm.frstrauss/PageImage3/Fat_water_Separation.pdf · Fat and water separationtechniques : 2 point Dixon Acquisition : Echos time selection TEi= 1/447 s =

Fat and water separation techniques : VARPRO + Graph cut

Page 41: Fat water Separation - lirmm.frstrauss/PageImage3/Fat_water_Separation.pdf · Fat and water separationtechniques : 2 point Dixon Acquisition : Echos time selection TEi= 1/447 s =

Fat and water separation techniques : Summary

Two element are crucial in the estimation of fat and water signal in the tissue :

1- Accurate signal model : R2* decay and multi-peak fat spectrum

2- Efficient global optimization approach, on complex data.

Page 42: Fat water Separation - lirmm.frstrauss/PageImage3/Fat_water_Separation.pdf · Fat and water separationtechniques : 2 point Dixon Acquisition : Echos time selection TEi= 1/447 s =

Fat and water separation techniques : IDEAL technique

Page 43: Fat water Separation - lirmm.frstrauss/PageImage3/Fat_water_Separation.pdf · Fat and water separationtechniques : 2 point Dixon Acquisition : Echos time selection TEi= 1/447 s =

Fat and water separation techniques : Impact of fat spectrum

Taking into account M peaks, requires the acquisition of M+2 images

� Exam duration is increased

Pre-calibration of fat spectrum :

Page 44: Fat water Separation - lirmm.frstrauss/PageImage3/Fat_water_Separation.pdf · Fat and water separationtechniques : 2 point Dixon Acquisition : Echos time selection TEi= 1/447 s =

Fat and water separation techniques : Impact of fat spectrum

Page 45: Fat water Separation - lirmm.frstrauss/PageImage3/Fat_water_Separation.pdf · Fat and water separationtechniques : 2 point Dixon Acquisition : Echos time selection TEi= 1/447 s =

Fat and water separation techniques : Impact of fat spectrum

Page 46: Fat water Separation - lirmm.frstrauss/PageImage3/Fat_water_Separation.pdf · Fat and water separationtechniques : 2 point Dixon Acquisition : Echos time selection TEi= 1/447 s =

Fat and water separation techniques : Echo time selection

The variance of the estimator verifies

Page 47: Fat water Separation - lirmm.frstrauss/PageImage3/Fat_water_Separation.pdf · Fat and water separationtechniques : 2 point Dixon Acquisition : Echos time selection TEi= 1/447 s =
Page 48: Fat water Separation - lirmm.frstrauss/PageImage3/Fat_water_Separation.pdf · Fat and water separationtechniques : 2 point Dixon Acquisition : Echos time selection TEi= 1/447 s =
Page 49: Fat water Separation - lirmm.frstrauss/PageImage3/Fat_water_Separation.pdf · Fat and water separationtechniques : 2 point Dixon Acquisition : Echos time selection TEi= 1/447 s =

T1-weighted spin-echo MR images (TR/TE, 400/10)

obtained in 10-year-old patient with functional score

of 3.

1 = vastus lateralis, 2 = biceps femoris, 3 =

semitendinosus, 4 = vastus medialis, 5 = rectus

femoris, 6 = vastus intermedius.

Page 50: Fat water Separation - lirmm.frstrauss/PageImage3/Fat_water_Separation.pdf · Fat and water separationtechniques : 2 point Dixon Acquisition : Echos time selection TEi= 1/447 s =

Functional grade was strongly inversely correlated with fat percentage

measured with the Dixon technique (p < 0.05)

Fig. 3—Graphs show Dixon MRI measurements of intramuscular fat percentage increased significantly with

decreasing functional level in all muscles examined.