external quality control lecture md general 2014 course, clin path ain shams university, egypt

42
Quality Management in Medical Laboratory External Quality Control Dr. Adel A. Elazab Associate Professor of Clinical Pathology Faculty of Medicine Ain Shams University, Cairo, Egypt

Upload: adel-elazab-elged

Post on 07-May-2015

469 views

Category:

Education


2 download

DESCRIPTION

Short description of the categories of proficiency testing programs, management of external quality control assays, and evaluation of results

TRANSCRIPT

Page 1: External Quality Control Lecture MD General 2014 Course, Clin Path Ain Shams University, Egypt

Quality Management in Medical Laboratory

External Quality ControlDr. Adel A. ElazabAssociate Professor of Clinical PathologyFaculty of MedicineAin Shams University, Cairo, Egypt

Page 2: External Quality Control Lecture MD General 2014 Course, Clin Path Ain Shams University, Egypt
Page 3: External Quality Control Lecture MD General 2014 Course, Clin Path Ain Shams University, Egypt

Inter-laboratory comparisons and other performance evaluations that may extend throughout all phases of the testing cycle, including interpretation of results; determination of individual and collective laboratory performance characteristics of examination procedures by means of interlaboratory comparison

External Quality Assessment (EQA)

Page 4: External Quality Control Lecture MD General 2014 Course, Clin Path Ain Shams University, Egypt

A program in which multiple samples are periodically sent to members of a group of laboratories for analysis and/or identification, in which each laboratory’s results are compared with those of other laboratories in the group and/or with an assigned value, and reported to the participating laboratory and others

Proficiency Testing (PT)

Page 5: External Quality Control Lecture MD General 2014 Course, Clin Path Ain Shams University, Egypt

Introduced into laboratory medicine more than 60 years ago as an educational tool to address observations that results for aliquots of the same sample were different when measured by different laboratories.

PT/EQA programs have evolved in scope and sophistication and are now an essential component of a laboratory’s quality management system.

PT/EQA is a component of laboratory accreditation requirements

PT/EQA programs

Page 6: External Quality Control Lecture MD General 2014 Course, Clin Path Ain Shams University, Egypt

Preparation Steps for Serum-Based PT/EQA

Page 7: External Quality Control Lecture MD General 2014 Course, Clin Path Ain Shams University, Egypt
Page 8: External Quality Control Lecture MD General 2014 Course, Clin Path Ain Shams University, Egypt

Ideal samples for a PT/EQA program would fulfill a range of criteria: ◦ Stable for the conditions under which they will be transported

and stored◦ Homogeneous across all the aliquots produced◦ Have analyte concentrations that include the expected clinical

range◦ Include appropriate sample types (e.g., urine, whole blood,

serum)◦ Available in sufficient volume◦ Inexpensive enough for cost not to be an impediment◦ Behave in clinical laboratory measurement procedures in◦ the same manner as patient samples◦ Samples from a single donor or pooled samples from

multiple donors can be used

Samples used for PT/EQA

Page 9: External Quality Control Lecture MD General 2014 Course, Clin Path Ain Shams University, Egypt

Single donor Vs Pooled Samples

Single-donor samples:◦ Interfering substance may

be present that influences 1 or more of the measurement procedures.

Pooled samples:◦ Dilute an interfering substance

◦ Limited by interactions of components such as serum proteins or urine complexes from different donors may cause aggregation or precipitation that necessitates further processing and potential modification of the matrix

Page 10: External Quality Control Lecture MD General 2014 Course, Clin Path Ain Shams University, Egypt

Samples have traceable reference values (when reference values are used)

Behave like patient samples (commutability) The laboratory may also consider cost Similarity of PT samples to patient samples Method compatibility with peer groups Size of peer groups Frequency of challenges Timeliness and usefulness of reports Educational content Customer service

Selection of PT Programs

Page 11: External Quality Control Lecture MD General 2014 Course, Clin Path Ain Shams University, Egypt

Categories of PT/ EQA Schemes

Page 12: External Quality Control Lecture MD General 2014 Course, Clin Path Ain Shams University, Egypt

category 1 programs are limited because of:• Technical aspects such as a lack of reference measurement

procedures, absence of certified reference materials, inability to prepare commutable samples;

• Practical considerations such as the difficulty of preparing samples covering the full measuring interval and the complicated logistics of preparation and distribution of fresh or frozen samples;

• Psychological limitations such as lack of awareness of the quality factors important in PT/EQA or unwillingness to adopt these;

• Economic concerns because distributing commutable samples in sufficient quantity and providing target values with reference measurement procedures is expensive

Page 13: External Quality Control Lecture MD General 2014 Course, Clin Path Ain Shams University, Egypt

PT samples should be tested in the same manner as patient samples, to the extent possible

Some laboratories may improperly test PT samples differently from patient samples, by repeat testing of PT samples when patient samples are tested only once, or by having a specific analyst test PT samples rather than rotating PT testing among all the personnel who perform patient testing.

There should be no attempt to produce “best” results by replicate analysis or testing immediately following internal QC or recalibration

No Special Treatment

Page 14: External Quality Control Lecture MD General 2014 Course, Clin Path Ain Shams University, Egypt

The core content of the result report should resemble as closely as possible the content of a routine clinical result report

If the usual report is deemed inappropriate for a PT report, it may equally be inappropriate for a clinical report

A copy of all PT reports should be retained within the laboratory in order to verify the information handling by the PT provider.

No Special Treatment (cont.)

Page 15: External Quality Control Lecture MD General 2014 Course, Clin Path Ain Shams University, Egypt

Sending a set of samples from an organizing body to a group of participating laboratories for measurement of 1 or more analytes present in the samples

Samples are intended to simulate the clinical samples usually measured

Laboratories are not informed of the analyte concentration or activity in a particular sample

Timely schedules for running and reporting results are included

Laboratory perform measurements in the same manner as for patient samples

Results for the samples are returned to the PT/EQA organizer for evaluation of conformance to the expected results

PT Procedure

Page 16: External Quality Control Lecture MD General 2014 Course, Clin Path Ain Shams University, Egypt

The organizer prepares a report that includes:◦ the results reported by a laboratory◦ the method used for the measurements◦ the target values expected for each analyte◦ evaluation of whether the individual laboratory’s results

met the performance requirements◦ Reports may also include evaluation of the performance

of the various measurement procedures used by the participants

The laboratory evaluates its performance according to the provider report

PT Procedure (cont.)

Page 17: External Quality Control Lecture MD General 2014 Course, Clin Path Ain Shams University, Egypt

Limits or quality standards around the target value are established against which performance can be assessed by:◦ Regulatory: wider like US CLIA, German Rili-BAeK◦ Statistical: ± 2-3 SD◦ Clinically-based: on a difference that may affect clinical decisions or on

biological variation

Total error limits including bias, imprecision, and analytical nonspecificity can contribute to the variation in a single result

Have different limits to separately assess bias and imprecision when replicate samples are included

PT/EQA limits are set as a minimum standard to identify results that indicate poor performance. Thus, meeting these standards may not indicate that performance is optimal nor that performance meets all clinical needs

ACCEPTANCE CRITERIA FOR PT/EQA RESULTS

Page 18: External Quality Control Lecture MD General 2014 Course, Clin Path Ain Shams University, Egypt

Determines the accuracy by comparing PT/EQA results to those from a reference measurement procedure or from a designated comparison method or to an all-participant (or all-method) mean/ median. This arrangement is now referred to as accuracy-based evaluation

Assess agreement with other measurement procedures and imprecision among all methods as well as within a method group

Peer group evaluation provides valuable information to assess quality, verifying that a laboratory is using a measurement procedure in conformance to the manufacturer’s specifications and to other laboratories using the same technology

Participant Evaluation of Their Own PT/EQA Results

Page 19: External Quality Control Lecture MD General 2014 Course, Clin Path Ain Shams University, Egypt
Page 20: External Quality Control Lecture MD General 2014 Course, Clin Path Ain Shams University, Egypt
Page 21: External Quality Control Lecture MD General 2014 Course, Clin Path Ain Shams University, Egypt
Page 22: External Quality Control Lecture MD General 2014 Course, Clin Path Ain Shams University, Egypt

PT/EQA result represents 1 point in time and will occasionally be a random error

Repeat the measurement using a stored aliquot of the PT/EQA sample (assuming the measurand was stable on storage) to confirm if the problem has persisted or to conclude that the problem no longer exists and the original unacceptable result was a random event, and therefore no corrective action is indicated. If the repeated result is still unacceptable, the laboratory conducts further investigation to identify the root cause, and then initiates corrective action

Handling of Unacceptable PT result

Page 23: External Quality Control Lecture MD General 2014 Course, Clin Path Ain Shams University, Egypt

• Gather data related to the testing event to include records of calibration, reagent use, QC results, and maintenance procedures;

• Obtain other data on assay performance, e.g., previous PT/EQA results and relevant patient data;

• Identify the root cause of the error;• Take corrective action and preventive action ifindicated;• Monitor the success of the corrective action;• Document the investigation and the corrective

action.

Handling of Unacceptable PT Result (cont)

Page 24: External Quality Control Lecture MD General 2014 Course, Clin Path Ain Shams University, Egypt

• Was the testing material received in satisfactory condition?

• Was the appropriate sample tested?• Were procedures for sample preparation followed?• Was the appropriate method used for analysis?• Was the method performed according to

documented procedures?• Were appropriate reagents and controls used?• Was equipment operated according to documented

procedures?

Gathering and Reviewing Data

Page 25: External Quality Control Lecture MD General 2014 Course, Clin Path Ain Shams University, Egypt

• Was equipment appropriately maintained?• Was QC acceptable at the time of testing PT

samples?• Were results interpreted appropriately?• Has this problem occurred previously with PT

samples? Are data consistent with previous PT distributions? Is there a trend leading to failure or is the current set completely unexpected?

• Did repeat testing on the properly stored residual sample produce similar results?

• Were patient results acceptable at the time of PT testing?

Gathering and Reviewing Data (cont.)

Page 26: External Quality Control Lecture MD General 2014 Course, Clin Path Ain Shams University, Egypt

1. Clerical error;2. Methodologic problem;3. Equipment problem;4. Technical problem;5. Problem with proficiency testing materials;6. Problem with evaluation of results; and7. No explanation after investigation.

Classifying the Problem

Page 27: External Quality Control Lecture MD General 2014 Course, Clin Path Ain Shams University, Egypt

• result was not correctly transcribed from the instrument tape or read-out to the report form;

• the PT sample was mislabeled;• an incorrect instrument or method was

reported on the form;• incorrect units were reported;• a decimal point was misplaced; or• an incorrect reporting code was selected on

the report form.

1. Clerical Errors

Page 28: External Quality Control Lecture MD General 2014 Course, Clin Path Ain Shams University, Egypt

• no written procedure (method) for staff to follow;• procedure steps inadequately, incompletely, or

incorrectly described;• problem in manufacture of reagents or reference

materials;• imprecision due to result being close to detection

limit of method;• imprecision due to variation between reagent

lots;• incorrect assignment of calibrator value;• calibration is unstable;

2. Methodologic Problems

Page 29: External Quality Control Lecture MD General 2014 Course, Clin Path Ain Shams University, Egypt

• inadequate quality control method; e.g.,− QC material not relevant to analyte concentration, or− inappropriate QC rules or limits;

• result not within measuring range (linearity) for instrument or reagent system;

• method is biased;• method lacks sensitivity;• method lacks specificity;• carry-over from previous sample(s);• inappropriate incubation conditions (time, temperature,

and/or atmosphere);• method used without validation;• inappropriate reference intervals applied

2. Methodologic Problems (cont.)

Page 30: External Quality Control Lecture MD General 2014 Course, Clin Path Ain Shams University, Egypt

• obstruction of instrument tubing/orifice by clot or protein;• misalignment of instrument probes;• problem with instrument data processing functions;• problem in manufacture of reagents or reference materials;• problem with instrument settings specified by manufacturer;• automatic pipettor not calibrated to acceptable precision

and accuracy;• equipment malfunction;• errors or omissions in programming of equipment software

applications; and• scheduled instrument maintenance not performed

appropriately.

3. Equipment Problems

Page 31: External Quality Control Lecture MD General 2014 Course, Clin Path Ain Shams University, Egypt

• failure to follow recommended instrument function checks (e.g., temperatures, blank readings, pressures);

• incorrect reconstitution or storage of reference materials or reagents;

• improper reconstitution, preparation, or storage of PT material;

• delay in testing after reconstitution of PT material, causing evaporation or deterioration;

• failure to follow written procedures;

4. Technical Problems

Page 32: External Quality Control Lecture MD General 2014 Course, Clin Path Ain Shams University, Egypt

• failure to follow PT instructions;• failure to act on QC results that indicate a

method problem;• pipetting or dilution error;• calculation error;• contamination of PT sample during

processing

4. Technical Problems (cont.)

Page 33: External Quality Control Lecture MD General 2014 Course, Clin Path Ain Shams University, Egypt

• differences between PT samples and patient samples;

• sample deteriorated in transit (if time- or temperature-sensitive)− bacterial contamination− hemolysis− nonhomogeneous sample;

• sample contains interfering factors (which may be method-specific).

5. Problem With Proficiency Testing Materials

Page 34: External Quality Control Lecture MD General 2014 Course, Clin Path Ain Shams University, Egypt

• inappropriate peer group;• inappropriate target value from

nonhomogeneous testing material or lingering [“masked”] outliers

• inappropriate evaluation interval with an extremely precise method, the acceptable range may be much narrower than needed for clinical usefulness

• incorrect data entry by PT provider.

6. Problem With Proficiency Testing Evaluation

Page 35: External Quality Control Lecture MD General 2014 Course, Clin Path Ain Shams University, Egypt

An investigation fails to reveal an explanation for an unacceptable PT result 19 to 24% of the time

7. No Explanation After Investigation

Page 36: External Quality Control Lecture MD General 2014 Course, Clin Path Ain Shams University, Egypt

Investigation Form for Discordant Proficiency Testing Findings in Clinical Chemistry

Page 37: External Quality Control Lecture MD General 2014 Course, Clin Path Ain Shams University, Egypt
Page 38: External Quality Control Lecture MD General 2014 Course, Clin Path Ain Shams University, Egypt

Sample Form for Documenting Unacceptable PT Investigations

Page 39: External Quality Control Lecture MD General 2014 Course, Clin Path Ain Shams University, Egypt

Sample Form for Documenting Unacceptable PT Investigations (cont.)

Page 40: External Quality Control Lecture MD General 2014 Course, Clin Path Ain Shams University, Egypt

Split-Sample With Another Laboratory Internal Split-Sample Procedures Audit-Sample Procedure Analysis of Manufacturer’s Product Calibrator or

Trueness Control Material Analysis of Interlaboratory Quality Control Data Averages of Patient Data

Assessment of Laboratory Tests When Proficiency Testing is Not Available

Page 41: External Quality Control Lecture MD General 2014 Course, Clin Path Ain Shams University, Egypt

Reference Intervals Reevaluation of Interpreted Results Direct Observation of Technique-Dependent Tests Clinical Correlation Studies Government and University Interlaboratory

Comparison Programs Analysis of Data From Qualitative Alternative

Assessment Procedures

Assessment of Laboratory Tests When Proficiency Testing is Not Available (cont.)

Page 42: External Quality Control Lecture MD General 2014 Course, Clin Path Ain Shams University, Egypt

YOUR VALUABLE QUESTIONS