105 ISSN: 2171-4703 © 2013 SECOIR Sociedad Española de Cirugía Ocular Implanto-Refractiva Ex-PRESS Glaucoma Filtration Device: eory, Technique and Results Iqbal Ike K. Ahmed, MD, FRCSC 1,2 ; Amandeep S. Rai MD 2 ABSTRACT: e Ex-PRESS glaucoma filtration device (Alcon Laboratories, Fort Worth, TX , USA) is designed to drain aqueous fluid from the anterior chamber to the subconjunctival space and form a filtration bleb, akin to a trabeculectomy. In this article we will review the surgical indications, preoperative evaluation, surgical techniques and postoperative considerations of this ophthalmic device. J Emmetropia. 2013; 4: 105-114 Description of Device e Ex-PRESS glaucoma filtration device (Alcon Laboratories, Fort Worth, TX, USA) is designed to drain aqueous fluid from the anterior chamber to the subconjunctival space and form a filtration bleb, akin to a trabeculectomy 1 . e Ex-PRESS glaucoma device is currently available in two models: an R-model and a P-model. e Ex PRESS R-model has a beveled tip, an external diameter of 400 microns (27-gauge), an internal lumen of 50 microns, a total device length of 2.96 mm and a uniform back plate. e Ex-PRESS P-model has a decreased bevel angle, an external diameter of 400 microns, a total device length of 2.64 mm and a vertical channel back plate. It is available in both a 50-micron and 200-micron internal lumen size (Figure 1) 1 . e Ex-PRESS glaucoma device is composed of biocompatible stainless steel with a spur to prevent extrusion of the device and a flat external backplate to prevent intrusion 2 . e backplate and spur are designed to conform to angle anatomy, and the distance between them approximates that of the scleral tract made by the device 3 . e P-model features vertical channels in the plate, designed to enhance posterior flow of aqueous and formation of a more diffuse posterior bleb. e devices also contain a relief port designed to provide an alternative pathway for aqueous should the axial lumen become occluded by iris, fibrin or blood 1 . Surgical indications e Ex-PRESS glaucoma device is indicated for use in open angle glaucoma. It is the authors’ experience that the device is effective in both wide open and narrow angles, as long as there is no peripheral anterior synechia in the area of intended implantation. e Ex-PRESS may be considered in patients who are candidates for a traditional trabeculectomy. Implantation of the Ex- PRESS shunt may be combined with lensectomy and intraocular lens (IOL) implantation. UPDATE/REVIEW 1 University of Utah, Salt Lake City, Utah, USA 2 University of Toronto, Toronto, Ontario, Canada Disclaimer: is article was previously published as Chapter 44 of the book Catarata y Glaucoma (ISBN: 9788493989804), coordinated by Dr Ramon Lorente. is book belongs to the series of annual monographs of the Spanish Society of Refractive Implant Ocular Surgery (SECOIR) and is reproduced with the consent of the copyright holders. Corresponding Author: Ike K. Ahmed, MD Credit Valley Eye Care, 3200 Erin Mills Parkway, Unit 1, Mississauga, Ontario, L5L 1W8, Canada. E-mail: [email protected] Figure 1. e Ex-PRESS R-model and P-model, with the dimensions and specifications of each.

Upload: trandang

Post on 01-May-2018

216 views

Category:

Documents


1 download

TRANSCRIPT

Page 1: Ex-PRESS Glaucoma Filtration Device: Theory, Technique … · Ex-PRESS Glaucoma Filtration Device: Theory, ... there were concerns with persistent hypotony, ... nylon sutures

105ISSN: 2171-4703© 2013 SECOIRSociedad Española de Cirugía Ocular Implanto-Refractiva

Ex-PRESS Glaucoma Filtration Device: Theory, Technique and Results

Iqbal Ike K. Ahmed, MD, FRCSC1,2; Amandeep S. Rai MD2

ABSTRACT: The Ex-PRESS glaucoma filtration device (Alcon Laboratories, Fort Worth, TX , USA) is designed to drain aqueous fluid from the anterior chamber to the subconjunctival space and form a filtration bleb, akin to a trabeculectomy. In this article we will review the surgical indications, preoperative evaluation, surgical techniques and postoperative considerations of this ophthalmic device.

J Emmetropia. 2013; 4: 105-114

Description of Device

The Ex-PRESS glaucoma filtration device (Alcon Laboratories, Fort Worth, TX, USA) is designed to drain aqueous fluid from the anterior chamber to the subconjunctival space and form a filtration bleb, akin to a trabeculectomy1. The Ex-PRESS glaucoma device is currently available in two models: an R-model and a P-model. The Ex PRESS R-model has a beveled tip, an external diameter of 400 microns (27-gauge), an internal lumen of 50 microns, a total device length of 2.96 mm and a uniform back plate. The Ex-PRESS P-model has a decreased bevel angle, an external diameter of 400 microns, a total device length of 2.64 mm and a vertical channel back plate. It is available in both a 50-micron and 200-micron internal lumen size (Figure 1)1.

The Ex-PRESS glaucoma device is composed of biocompatible stainless steel with a spur to prevent extrusion of the device and a flat external backplate to prevent intrusion2. The backplate and spur are designed to conform to angle anatomy, and the distance between them approximates that of the scleral tract made by the device3.

The P-model features vertical channels in the plate, designed to enhance posterior flow of aqueous and formation of a more diffuse posterior bleb. The devices also contain a relief port designed to provide an alternative pathway for aqueous should the axial lumen become occluded by iris, fibrin or blood1.

Surgical indications

The Ex-PRESS glaucoma device is indicated for use in open angle glaucoma. It is the authors’ experience that the device is effective in both wide open and narrow angles, as long as there is no peripheral anterior synechia in the area of intended implantation. The Ex-PRESS may be considered in patients who are candidates for a traditional trabeculectomy. Implantation of the Ex-PRESS shunt may be combined with lensectomy and intraocular lens (IOL) implantation.

UPDATE/REVIEW

1University of Utah, Salt Lake City, Utah, USA2University of Toronto, Toronto, Ontario, Canada

Disclaimer: This article was previously published as Chapter 44 of the book Catarata y Glaucoma (ISBN: 9788493989804), coordinated by Dr Ramon Lorente. This book belongs to the series of annual monographs of the Spanish Society of Refractive Implant Ocular Surgery (SECOIR) and is reproduced with the consent of the copyright holders.

Corresponding Author: Ike K. Ahmed, MDCredit Valley Eye Care, 3200 Erin Mills Parkway, Unit 1, Mississauga, Ontario, L5L 1W8, Canada.E-mail: [email protected]

Figure 1. The Ex-PRESS R-model and P-model, with the dimensions and specifications of each.

Page 2: Ex-PRESS Glaucoma Filtration Device: Theory, Technique … · Ex-PRESS Glaucoma Filtration Device: Theory, ... there were concerns with persistent hypotony, ... nylon sutures

106

JOURNAL OF EMMETROPIA - VOL 4, APRIL-JUNE

Preoperative evaluationIt is important to determine the preoperative status,

mobility, and health of the sclera and conjunctiva at the planned surgical site. Gonioscopy should be used to evaluate the anterior chamber angle for peripheral anterior synechiae at the planned insertion site. The crystalline lens should also be evaluated for the possibility of cataract, which can be treated with combined cataract-glaucoma surgery1.

Original surgical technique

The original surgical technique involved placement of the device under a conjunctival flap only. Under local or topical anesthesia, a 2-4 mm circumferential conjunctival opening was created 10-15 mm posterior to the limbus. The device was introduced using an injector, and slid under the conjunctiva and Tenon’s capsule. The device was implanted parallel to the iris plane, radial to the limbus. This was routinely combined with phacoemulsification and insertion of an intraocular lens. Particular advantages to this approach included reproducibility and simplicity: the procedure does not require a scleral flap and is easily taught.

This technique is no longer utilized due to high failure rates secondary to postoperative complications with flow control and positioning of the device. Specifically with regards to flow control, there were concerns with persistent hypotony, flat anterior chamber, choroidal detachment, and suprachoroidal hemorrhage. Specific positioning concerns included conjunctival erosion, shunt extrusion, conjunctival scarring and tube exposure1,4-8.

Current surgical technique

Under sterile field, the eye is first anesthetized with topical lidocaine. In combined surgery, the lensectomy and IOL implantation is performed prior to the Ex-PRESS implantation in order to ensure stability and deepen the anterior chamber, and to prevent excessive irrigation of fluid into the subconjunctival bleb. A temporal clear corneal or limbal incision is suggested for the cataract extraction to minimize manipulation and disturbance to the superior tissues where the bleb will ultimately form (Figure 2). Upon successful lensectomy and IOL insertion, viscoelastic should be left in the anterior chamber to maintain its stability. A suture is placed in the main incision to maintain watertight closure. The combination of viscoelastic and injected balanced saline solution ensures that intraocular pressure (IOP) remains adequately high to maintain globe and scleral rigidity.

A conjunctival peritomy is performed as the surgeon would normally do for a trabeculectomy, which is followed

Figure 2. In combined surgery, the phacoemulsification is completed with a temporal approach (2A) to ensure that the ExPRESS device can be implanted superiorly (2B).

Figure 3. After creation of a conjunctival peritomy, the scleral surface is prepared using light cautery.

by light cautery to prepare the scleral surface (Figure 3). The surgeon should plan for an adequately sized scleral flap that is positioned according to where the Ex-PRESS device will be implanted, in an approach to similar to a trabeculectomy scleral flap. The flap may need to be slightly larger in order to provide full scleral coverage over the footplate of the implant (Figure 4). A partial thickness scleral flap of one-half to two-thirds scleral thickness is created using a crescent blade just as the surgeon would normally do for a trabeculectomy. The shape of the flap is up to the surgeon’s discretion. The scleral dissection should be carried forward to the clear cornea to gain adequate exposure of the scleral spur, where the shunt will be implanted into the anterior chamber (Figure 5). Once the flap is prepared, an

EXPRESS GLAUCOMA FILTRATION DEVICE

Page 3: Ex-PRESS Glaucoma Filtration Device: Theory, Technique … · Ex-PRESS Glaucoma Filtration Device: Theory, ... there were concerns with persistent hypotony, ... nylon sutures

107

JOURNAL OF EMMETROPIA - VOL 4, APRIL-JUNE

Figure 4. (all measurements in millimeters): It is important to position and size the scleral flap in accordance with the anticipated site of implantation for the ExPRESS device. The surgical flap should be planned to provide adequate coverage lateral and posterior to the device in order to control aqueous flow.

Figure 5. Creation of a partial thickness scleral flap (about one-half to two-thirds scleral thickness). It is important to identify the scleral fibers and the scleral spur (the anticipated site of implantation), and continue the scleral dissection forward to the blue zone and clear cornea in order to ensure adequate exposure.

Figure 6. It is important to create the pilot hole at the anterior scleral spur (6A), and parallel to the iris plane, which can be facilitated by downward rotation of the eye (6B).

antimetabolite such as mitomycin-C (MMC) may be applied subconjunctivally and under the scleral flap, again in an approach similar to trabeculectomy.

After appropriate irrigation of the antimetabolite from the surgical field, a pilot hole is then fashioned using a sapphire blade (Alcon Laboratories, Fort Worth, TX, USA), a 26-gauge needle (if using a hypodermic needle) or a 25-gauge EdgePlus Trocar Blade (Alcon Laboratories, Fort Worth, TX, USA). The pilot hole is an opening into the anterior chamber and important for

successful placement of the implant. The eye should be positioned pointing down, and the needle should enter at the anterior spur at the level of the iris plane (Figure 6). Angling the needle appropriately is of the utmost importance: a posteriorly directed shunt may result in iris touch and subsequent occlusion of the shunt’s ostia, while an excessively anterior shunt may result in corneal touch and consequent endothelial trauma.

After creation of the pilot hole, the implant is ready for insertion (Figure 7). The Ex-PRESS implant comes preloaded on an injector system. The injector device is rotated 90 degrees prior to insertion, and is rotated to its final position once inside the eye. It is important to ensure that the back plate is flush with the scleral bed (Figure 8). The device is released at the tip of the injector system when the surgeon’s index finger depresses a portion of the injector’s shaft. An adequately planned scleral flap should now provide good lateral and posterior coverage of the Ex-PRESS device.

Good closure of the scleral flap is achieved using 10-0 nylon sutures. Two sutures are placed in each corner, with a slipknot that can be locked. Additional sutures can provide more flow control, if required (Figure 9). Similarly, the slipknots can be tightened or loosened to adjust the tension on the flap. The rate of flow can be assessed by continuously injecting balanced salt solution into either the anterior chamber or through a side port.

Firm conjunctival closure is achieved using 10-0 Vicryl sutures in a running horizontal mattress

EXPRESS GLAUCOMA FILTRATION DEVICE

Page 4: Ex-PRESS Glaucoma Filtration Device: Theory, Technique … · Ex-PRESS Glaucoma Filtration Device: Theory, ... there were concerns with persistent hypotony, ... nylon sutures

108

JOURNAL OF EMMETROPIA - VOL 4, APRIL-JUNE

technique. This technique allows a good conjunctival closure which prevents early leakage, while reducing the amount of tension on the bleb, which may help with bleb morphology.

Postoperative considerations

Despite the resistance provided by the fixed lumen size, the use of sutures to control flow, and intraoperative evaluation of aqueous egress, postoperative hypotony remains a potential complication of the Ex-PRESS implant. This may be caused by flow around the shunt through the pilot hole incision in which it is placed. Early postoperative hypotony warrants careful assessment for conjunctival leak at the site of wound closure as it may be due to a conjunctival leak. Early or late hypotony may necessitate pressurization or reformation of the anterior chamber using ophthalmic viscosurgical devices (OVDs). The smaller lumen of the Ex-PRESS device calls for a stepwise approach in

Figure 8. Final placement of the ExPRESS in the anterior chamber is slightly above and parallel to the iris (8A). The backplate of the device should be flush with the scleral bed (8B).

Figure 9. The scleral flap is closed with two sutures (10-0 nylon) in each corner, with a slipknot that can be locked (9A). Additional sutures can be used as required to control flow (9B).

Figure 7. The ExPRESS glaucoma filtration device comes preloaded on an injector device (7A). The injector is rotated 90 degrees prior to insertion (7B). The device enters the anterior chamber parallel to the iris (7C).

EXPRESS GLAUCOMA FILTRATION DEVICE

Page 5: Ex-PRESS Glaucoma Filtration Device: Theory, Technique … · Ex-PRESS Glaucoma Filtration Device: Theory, ... there were concerns with persistent hypotony, ... nylon sutures

109

JOURNAL OF EMMETROPIA - VOL 4, APRIL-JUNE

selecting OVDs starting with agents such as ProVisc and VisCoat (Alcon). If hypotony, choroidal effusions, or chamber instability persists, then the surgeon may select an agent such as Healon GV or Healon 5 (Abbott Medical Optics, Irvine, CA , USA). The surgeon should also titrate the volume of injections, starting with a small volume and increasing as necessary. Vigilant follow-up of these patients is mandatory.

Elevated postoperative IOP may indicate occlusion of the lumen of the device or episcleral fibrosis. If iris occlusion of the lumen is visualized, a 30-gauge needle can be inserted through the limbus at the slit lamp and used to mechanically sweep the iris away from the tip of the shunt. Occasionally, fibrin or a membrane may form at or in proximal tip of the device. In these cases, Nd:YAG laser can be used to remove the material, or if fibrin is felt to be the main issue, intracameral tPA (25ug/0.cc) may be injected into the anterior chamber to dissolve the fibrin. It is imperative that flow is reestablished in a timely fashion to prevent failure of bleb formation. Episcleral fibrosis with flattening of bleb is the more common reason for elevated postoperative IOP, and should be managed in a similar way as for trabeculectomy. In this scenario, subconjunctival injection of an antimetabolite like 5-fluorouracil or mitomcyin-C may be used, and/or needling of the bleb with lifting of the scleral flap can be performed. In cases of hyper-encapsulation or Tenon’s cyst formation, aqueous suppression should be used with a reduction of steroid usage, as per similar approach for a trabeculectomy.

The Ex-PRESS glaucoma device is subject to the short and long term complications known to exist with filtrating surgeries, including blebitis, endophthalmitis, bleb-related complications, corneal dellen, endothelial cell loss , and choroidal expansion. These are managed in a manner similar to trabeculectomy patients.

Efficacy

Several retrospective non-controlled case series have evaluated the IOP lowering effects of the Ex-PRESS glaucoma device implanted under a scleral flap. A study of 100 eyes demonstrated mean IOP lowering from 27.7 ± 9.2 mmHg to 14.0 ± 5.1mmHg (p < 0.0001), and glaucoma medications reduced from 2.7 ± 1.1 to 0.7 ± 1.1 (p < 0.0001)4. A study of 24 eyes demonstrated mean IOP lowering from 27.2 ± 7.1 mmHg to 14.5 ± 5.0 mmHg (p < 0.01)9. A report of 15 eyes with post-penetrating keratoplasty glaucoma demonstrated mean IOP lowering from 41.5 ± 14.3 mmHg to 12.1 ± 5.2 mmHg (p < 0.001)10. Another report of 37 eyes demonstrated mean IOP lowering from 27.6 ± 8.7 mmHg to 12.4 ± 3.4 mmHg (significance not reported)11.

Several retrospective non-randomized case-control studies have compared the IOP lowering effects of the Ex-PRESS glaucoma device under a scleral flap with conventional trabeculectomy. A study of 100 eyes found no difference in mean IOP or success rates between groups, but the Ex-PRESS group demonstrated statistically significant lower rates of early postoperative hypotony and choroidal effusion12. Similarly, a study of 93 eyes found no difference in mean IOP lowering or success rates, but the Ex-PRESS group had less early postoperative hypotony and choroidal effusion (although not statistically significant)13. A study of 70 eyes also reported no difference in success rates or mean IOP reductions, but higher IOP in the Ex-PRESS group at last follow-up (p = 0.008). The study also reported reduced rates of postoperative hypotony and hyphema in the Ex-PRESS group (although not statistically significant), faster visual recovery (p = 0.028), and less postoperative visits (p < 0.001). A study of 21 eyes found no difference in mean IOP but lower rates of shallow anterior chamber, hypotony, choroidal detachment (although not statistically significant), and hyphema in the Ex-PRESS group as well as improved visual recovery (p < 0.05)14.

A large retrospective non-randomized case-control study of 345 eyes found no difference in success rates when comparing Ex-PRESS alone with Ex-PRESS combined with phacoemulsification, both under scleral flaps. It is not possible to compare the postoperative IOPs directly between the groups because baseline IOPs were significantly different15.

A prospective randomized control trial of 80 eyes comparing the Ex-PRESS device under scleral flap with conventional trabeculectomy demonstrated higher rates of success (p = 0.023) and lower mean IOP (12.0 ± 2.7 mmHg vs. 13.9 ± 4.3; p = 0.02) in the Ex-PRESS group at 1-year follow-up with similar complication rates16. The same group reported 5-year follow-up data and found no difference in mean IOP or success rates between the two groups. 59% of Ex-PRESS patients and 46% of trabeculectomy patients demonstrated complete success (IOP < 18 mmHg without medications) (p = 0.25). 97% of Ex-PRESS patients and 100% of trabeculectomy patients demonstrated total success (IOP < 18 mmHg with or without medications)17.

From the data currently available on the Ex-PRESS glaucoma device implanted under a scleral flap (Table 1), the implant appears to provide an improved early postoperative safety profile without compromising IOP lowering efficacy when compared to trabeculectomy. Implantation of the Ex-PRESS device is a more predictable and reproducible technique, and less subject to the variability seen with trabeculectomy. In particular, there appears to be early postoperative advantages such as reduced rates of hypotony and choroidal effusion, faster visual recovery, and less postoperative visits.

Many of the studies are limited by their retrospective, non-randomized design.

EXPRESS GLAUCOMA FILTRATION DEVICE

Page 6: Ex-PRESS Glaucoma Filtration Device: Theory, Technique … · Ex-PRESS Glaucoma Filtration Device: Theory, ... there were concerns with persistent hypotony, ... nylon sutures

110

JOURNAL OF EMMETROPIA - VOL 4, APRIL-JUNE

Table 1. Data currently available on the Ex-PRESS glaucoma device implanted under a scleral flap

AUTHOR, YEAR Lankaranian 20114 Dahan 20059 Ates 201010 De Feo 200911 Seider 201113 Maris 200712 Good 201120 Sugiyama 201114 Kanner 200915 De Jong 200916 De Jong 201117

STUDY TYPE Retrospective non-control case series

Retrospective non-control case series

Retrospective non-control case series

Prospective non-control case series

Retrospective non-randomized case-control study

Retrospective non-randomized case-control study

Retrospective, non-randomized, case-control series

Retrospective, non-randomized, case-control series

Retrospective non-randomized, case control series

Prospective randomized control trial

Prospective 5 year follow-up (extension of previous trial listed)

PATIENTS 100 eyes of 100 patients, mean age 77.4y, mean F/U 27m ± 13.2

24 eyes of 23 patients, mean age 60.7y, mean F/U 71.0w (26-104)

15 eyes of 15 patients, post-penetrating keratoplasty glaucoma, mean age 37.4y, mean F/U 12.2m (8-19)

37 eyes of 35 patients, mean age 69y, mean F/U 18m (12-24)

93 eyes of 93 patients, mean age 71.0y, minimum F/U 6m

100 eyes of 97 patients, mean age 66y, mean F/U 10.8m (3.5-18)

70 eyes of 70 patients, mean age 69y, mean F/U 28m ± 3.2

21 eyes of 21 patients, mean age (EX: 64.2y, trab 71.3y), F/U 12m

345 eyes of 300 patients, mean age 72.4y(combined) and 63.9y (EX)*, mean F/U 21.9m ±12.5 (combined) 25.7m ± 11.1(EX)

80 eyes of 78 patients, mean age 62.3y (EX), 68.9y (trab)*, mean F/U 49.5w (EX), 51.0w (trab)

78 eyes of 78 patients, mean age 62.4y(EX), 68.6y (trab), mean F/U 262.4w(EX), 265.6(trab)

INTERVENTION Ex-PRESS under scleral flap

Ex-PRESS under scleral flap + MMC

Ex-PRESS under scleral flap + MMC

Ex-PRESS under scleral flap + MMC

Resident-performed Ex-PRESS under scleral flap (n=36) + MMC

Ex-PRESS under scleral flap + MMC

Ex-PRESS under scleral flap Ex-PRESS under scleral flap + MMC

Ex-PRESS under scleral flap + MMC + phacoemulsification

Ex-PRESS under scleral flap + MMC

Ex-PRESS under scleral flap + MMC

CONTROL N/A N/A N/A N/A Resident-performed trabeculectomy (n=57)

Trabeculectomy or phaco-trabeculectomy

Trabeculectomy Trabeculectomy + MMC

Ex-PRESS under scleral flap + MMC

Trabeculectomy + MMC

Trabeculectomy + MMC

OUTCOME MEASURES IOP success range: 5-21mmHg

IOP IOP, number of glaucoma medsIOP success range: 5-21mmHg

IOP,IOP success range: < 18mmHgStringent IOP success range: < 15mmHg

IOP success range: 6-21mmHg,medications

IOP success range: 5-21mmHg

IOP, bleb morphological features (Moorfields Bleb Grading System), medications, visual recovery, number of postoperative visits, IOP success range: 5-18mmHg

IOP, visual acuityIOP success range: 5-21mmHg

IOP success range: 5-21mmHg

IOP, medications, VA, IOP success range: 4-18mmHgStringent IOP success range: 4-15mmHg

IOP, medications,IOP success range: < 18mmHg

RESULTS Complete success in 60% at last F/U,Qualified success in 24% at last F/UMean IOP: 27.7±9.2 decreased to 14.0±5.1* Mean meds: 2.7±1.1 decreased to 0.7±1.1*

Mean IOP 27.2±7.1 decreased to 13.3±3.7 at 6m*, 14.5±5.0 at 12m*

Mean IOP 41.5±14.3 decreased to 12.1±5.2*.Complete success: 86.6% at last FU.Total success: 93.3% at last FUGlaucoma medications: 3.2(2-4) decreased to 0.3(0-3)*

IOP: 27.6±8.7 decreased to 12.4±3.4 at last FU (significance not reported)Complete success <18: 78.4%, qualified 5.4%Stringent complete success <15: 70.3%, qualified 10.8%

No difference in IOP or proportional IOP decrease between groups at all F/U points.Complete success higher in EX group at 3m (81% vs 61%, p=0.057), 6m (69% vs 47%, p=0.076)No difference at 12m (48% vs 45%, p=1.0)Total success similar at all time points.EX group used less medications at 3m, no difference at 6m or 12m

EX: IOP 26.2±10.5 decreased to 11.5±11.1Trab: IOP 25.52±9.9 decreased to 8.7± 9.9Mean IOP significantly lower in trab group up to 3m.No difference from 3-15m.No difference in success rates at last F/U (15m) (85.6% EX, 84.3% trab).

IOP: higher in EX group at 1y* and last F/U* (exact IOPs not reported). No difference in complete success: EX 77.1% vs trab 74.3% (p=1.00) or qualified success: EX 5.7% vs trab 8.6% (p=0.99)Bleb: no difference in bleb vascularity, height, diffuseness at last F/U. EX group returned to baseline visual acuity earlier*.EX group required fewer visits in first 3 postoperative months (6.05 v 8.23)*.No difference in medications at last F/U.

No difference in mean IOP or number of glaucoma medications at 1y.Improved visual recovery in Ex-PRESS group compared to trab*.Higher rate of qualified success (EX 100% vs trab 81.8%) (p=0.167)

Both groups demonstrated statistically significant decrease in IOP at last FUNo difference in success at 1y: Ex-PRESS alone (96.9%), combined (95.6%) (p=0.948)

EX had higher rate of complete success at 1y (84.6% vs 60%)*,stringent complete success at 1y (76.9% vs 50.0%)*.EX had lower IOP at 1y (12.0±2.7 vs 13.9±4.3*.No difference in glaucoma medications.

No difference in mean IOP at 5y: EX 11.5±2.9 vs. trab 11.3±1.9(p=0.73) or incomplete success at 5y: EX 59.0% vs. trab 46.2% (p=0.25) or in total success at 5y: EX 97.4% vs trab 100% (p=0.49)EX group used less glaucoma medications at 5y (significance not reported)

* statistical significance (p < 0.05); IOP, Intraocular pressure, measured in mmHg; F/U, Follow up; VA, visual acuity; AC, anterior chamber; w, weeks; m, months; y, years; EX, Ex-PRESS; Trab, trabeculectomyComplete success: IOP within the predefined IOP range, without any glaucoma medications; Qualified success: IOP within the predefined IOP range, with the use of glaucoma medications; Total success: IOP within the predefined IOP range, with or without the use of glaucoma medications

EXPRESS GLAUCOMA FILTRATION DEVICE

Page 7: Ex-PRESS Glaucoma Filtration Device: Theory, Technique … · Ex-PRESS Glaucoma Filtration Device: Theory, ... there were concerns with persistent hypotony, ... nylon sutures

111

JOURNAL OF EMMETROPIA - VOL 4, APRIL-JUNE

Table 1. Data currently available on the Ex-PRESS glaucoma device implanted under a scleral flap

AUTHOR, YEAR Lankaranian 20114 Dahan 20059 Ates 201010 De Feo 200911 Seider 201113 Maris 200712 Good 201120 Sugiyama 201114 Kanner 200915 De Jong 200916 De Jong 201117

STUDY TYPE Retrospective non-control case series

Retrospective non-control case series

Retrospective non-control case series

Prospective non-control case series

Retrospective non-randomized case-control study

Retrospective non-randomized case-control study

Retrospective, non-randomized, case-control series

Retrospective, non-randomized, case-control series

Retrospective non-randomized, case control series

Prospective randomized control trial

Prospective 5 year follow-up (extension of previous trial listed)

PATIENTS 100 eyes of 100 patients, mean age 77.4y, mean F/U 27m ± 13.2

24 eyes of 23 patients, mean age 60.7y, mean F/U 71.0w (26-104)

15 eyes of 15 patients, post-penetrating keratoplasty glaucoma, mean age 37.4y, mean F/U 12.2m (8-19)

37 eyes of 35 patients, mean age 69y, mean F/U 18m (12-24)

93 eyes of 93 patients, mean age 71.0y, minimum F/U 6m

100 eyes of 97 patients, mean age 66y, mean F/U 10.8m (3.5-18)

70 eyes of 70 patients, mean age 69y, mean F/U 28m ± 3.2

21 eyes of 21 patients, mean age (EX: 64.2y, trab 71.3y), F/U 12m

345 eyes of 300 patients, mean age 72.4y(combined) and 63.9y (EX)*, mean F/U 21.9m ±12.5 (combined) 25.7m ± 11.1(EX)

80 eyes of 78 patients, mean age 62.3y (EX), 68.9y (trab)*, mean F/U 49.5w (EX), 51.0w (trab)

78 eyes of 78 patients, mean age 62.4y(EX), 68.6y (trab), mean F/U 262.4w(EX), 265.6(trab)

INTERVENTION Ex-PRESS under scleral flap

Ex-PRESS under scleral flap + MMC

Ex-PRESS under scleral flap + MMC

Ex-PRESS under scleral flap + MMC

Resident-performed Ex-PRESS under scleral flap (n=36) + MMC

Ex-PRESS under scleral flap + MMC

Ex-PRESS under scleral flap Ex-PRESS under scleral flap + MMC

Ex-PRESS under scleral flap + MMC + phacoemulsification

Ex-PRESS under scleral flap + MMC

Ex-PRESS under scleral flap + MMC

CONTROL N/A N/A N/A N/A Resident-performed trabeculectomy (n=57)

Trabeculectomy or phaco-trabeculectomy

Trabeculectomy Trabeculectomy + MMC

Ex-PRESS under scleral flap + MMC

Trabeculectomy + MMC

Trabeculectomy + MMC

OUTCOME MEASURES IOP success range: 5-21mmHg

IOP IOP, number of glaucoma medsIOP success range: 5-21mmHg

IOP,IOP success range: < 18mmHgStringent IOP success range: < 15mmHg

IOP success range: 6-21mmHg,medications

IOP success range: 5-21mmHg

IOP, bleb morphological features (Moorfields Bleb Grading System), medications, visual recovery, number of postoperative visits, IOP success range: 5-18mmHg

IOP, visual acuityIOP success range: 5-21mmHg

IOP success range: 5-21mmHg

IOP, medications, VA, IOP success range: 4-18mmHgStringent IOP success range: 4-15mmHg

IOP, medications,IOP success range: < 18mmHg

RESULTS Complete success in 60% at last F/U,Qualified success in 24% at last F/UMean IOP: 27.7±9.2 decreased to 14.0±5.1* Mean meds: 2.7±1.1 decreased to 0.7±1.1*

Mean IOP 27.2±7.1 decreased to 13.3±3.7 at 6m*, 14.5±5.0 at 12m*

Mean IOP 41.5±14.3 decreased to 12.1±5.2*.Complete success: 86.6% at last FU.Total success: 93.3% at last FUGlaucoma medications: 3.2(2-4) decreased to 0.3(0-3)*

IOP: 27.6±8.7 decreased to 12.4±3.4 at last FU (significance not reported)Complete success <18: 78.4%, qualified 5.4%Stringent complete success <15: 70.3%, qualified 10.8%

No difference in IOP or proportional IOP decrease between groups at all F/U points.Complete success higher in EX group at 3m (81% vs 61%, p=0.057), 6m (69% vs 47%, p=0.076)No difference at 12m (48% vs 45%, p=1.0)Total success similar at all time points.EX group used less medications at 3m, no difference at 6m or 12m

EX: IOP 26.2±10.5 decreased to 11.5±11.1Trab: IOP 25.52±9.9 decreased to 8.7± 9.9Mean IOP significantly lower in trab group up to 3m.No difference from 3-15m.No difference in success rates at last F/U (15m) (85.6% EX, 84.3% trab).

IOP: higher in EX group at 1y* and last F/U* (exact IOPs not reported). No difference in complete success: EX 77.1% vs trab 74.3% (p=1.00) or qualified success: EX 5.7% vs trab 8.6% (p=0.99)Bleb: no difference in bleb vascularity, height, diffuseness at last F/U. EX group returned to baseline visual acuity earlier*.EX group required fewer visits in first 3 postoperative months (6.05 v 8.23)*.No difference in medications at last F/U.

No difference in mean IOP or number of glaucoma medications at 1y.Improved visual recovery in Ex-PRESS group compared to trab*.Higher rate of qualified success (EX 100% vs trab 81.8%) (p=0.167)

Both groups demonstrated statistically significant decrease in IOP at last FUNo difference in success at 1y: Ex-PRESS alone (96.9%), combined (95.6%) (p=0.948)

EX had higher rate of complete success at 1y (84.6% vs 60%)*,stringent complete success at 1y (76.9% vs 50.0%)*.EX had lower IOP at 1y (12.0±2.7 vs 13.9±4.3*.No difference in glaucoma medications.

No difference in mean IOP at 5y: EX 11.5±2.9 vs. trab 11.3±1.9(p=0.73) or incomplete success at 5y: EX 59.0% vs. trab 46.2% (p=0.25) or in total success at 5y: EX 97.4% vs trab 100% (p=0.49)EX group used less glaucoma medications at 5y (significance not reported)

* statistical significance (p < 0.05); IOP, Intraocular pressure, measured in mmHg; F/U, Follow up; VA, visual acuity; AC, anterior chamber; w, weeks; m, months; y, years; EX, Ex-PRESS; Trab, trabeculectomyComplete success: IOP within the predefined IOP range, without any glaucoma medications; Qualified success: IOP within the predefined IOP range, with the use of glaucoma medications; Total success: IOP within the predefined IOP range, with or without the use of glaucoma medications

EXPRESS GLAUCOMA FILTRATION DEVICE

Page 8: Ex-PRESS Glaucoma Filtration Device: Theory, Technique … · Ex-PRESS Glaucoma Filtration Device: Theory, ... there were concerns with persistent hypotony, ... nylon sutures

112

JOURNAL OF EMMETROPIA - VOL 4, APRIL-JUNE

ADVERSE EFFECTS Conjunctival leakage (6%), hyphema (2%), choroidal detachment (2%), wound gap (1%)

Transient hypotony spontaneously resolved (20.8%), device iris touch (12.5%), transient choroidal effusion spontaneously resolved (8.3%), hyphema (8.3%), shallow AC requiring reformation (4.1%)

After Ex-PRESS, all clear grafts remained clear; edematous grafts became clearer; opaque grafts showed no change.No change in central corneal graft thickness.Bleb needling to lower IOP (20%)

Serous choroidal detachment (24.3%, with 2.7% necessitating intervention), hypotony (21.6%, with 16.2% necessitating AC fill)

EX group had lower ratesof early postoperative hypotony, choroidal effusion, self-limited bleb leaks, but not statistically significant.Similar rates of prolonged postoperative hypotony, reoperation for glaucoma.EX group had higher rates of multiple laser suture lysis (LSL) procedures, and more LSL were done per patient*.Ex-PRESS group had higher rates of bleb leak requiring additional sutures, patients requiring AC reformation(not statistically significant)

Significantly lower rates of early postoperative hypotony (4% vs 32%)*, and choroidal effusion (8% vs 38%)* in EX group. No difference in flat AC, hypotony maculopathy, hyphema, bleb leak, bleb needling, endophthalmitis. More LSL in EX (2.2 vs 1.7)*.

EX group had lower rates of postoperative hypotony, and hyphema (not statistically significant).More laser suture lysis in EX group (51% vs 20%)*, and earlier (11.5d vs 15.8d)*.

Lower rates of shallow AC, hypotony, choroidal detachment, hyphema, fall in visual acuity in EX (not statistically significant).

No difference in blocked tube, early postoperative hypotony, flat AC, choroidal hemorrhage, aqueous misdirection, exposed implant, dislocated implant, dysthetic bleb, bleb leak

No difference in rates of shallow AC, flat AC, bleb leak, choroidal detachment, hyphema, IOP spikes, visual acuity

EX group had less needling, and development of cataract (significance not reported)

* statistical significance (p < 0.05); IOP, Intraocular pressure, measured in mmHg; F/U, Follow up; VA, visual acuity; AC, anterior chamber; w, weeks; m, months; y, years; EX, Ex-PRESS; Trab, trabeculectomyComplete success: IOP within the predefined IOP range, without any glaucoma medications; Qualified success: IOP within the predefined IOP range, with the use of glaucoma medications; Total success: IOP within the predefined IOP range, with or without the use of glaucoma medications

The devices showed minimal temperature changes (< 0.3 °C) under all protocols. The devices showed no induced torque, displacement, or rotation under the 1.5-T and 3.0-T MRI protocols. Under the 4.7-T MRI protocol, the device was displaced to the limit of the testing field, and rotational force could not be calculated.

Although the 316L stainless steel material is often described as non-magnetic, the Ex-PRESS glaucoma device does have ferromagnetic properties, as demonstrated under the 4.7-T MRI protocol. However, most MRI machines used in clinical settings worldwide use a magnetic field up to 3.0-T, and the device is safe under these conditions18.

Seven eyes of five patients with the Ex-PRESS device previously implanted who underwent axial T1-weighted, axial and coronal T2-weighted with fat saturation, and whole brain proton density (PD)/T2-weighted scans using a 1.5-T MRI were reviewed. Two (2) of the implants were under the conjunctiva, and 5 were implanted under a scleral flap, and mean time elapsed since surgery was 13 months (1-24 m). Interpretation of MRI scans of the orbit and brain are minimally affected by artifacts from the Ex-PRESS implant. Interpretation of optic nerve

With regards to adverse effects, many of the studies demonstrate an improved early postoperative safety profile with the Ex-PRESS device, particularly in regards to hypotony and choroidal effusion. However, given the relative rarity of the adverse outcomes and the limited sample sizes, many studies do not report statistical significance. Further trials, including prospective randomized control trials with sufficient numbers of patients and follow-up are warranted to strengthen the body of evidence.

Table 1. Data currently available on the Ex-PRESS glaucoma device implanted under a scleral flap (cont.)

AUTHOR, YEAR Lankaranian 20114 Dahan 20059 Ates 201010 De Feo 200911 Seider 201113 Maris 200712 Good 201120 Sugiyama 201114 Kanner 200915 De Jong 200916 De Jong 201117

EXPRESS GLAUCOMA FILTRATION DEVICE

Diagnostic Imaging

The American Society for Testing and Materials guidelines require magnetic resonance imaging (MRI) safety testing to evaluate materials for torque, displacement force, and radiofrequency heating under magnetic field. The Ex-PRESS glaucoma drainage device is composed of 316L stainless steel and was evaluated for movement with MRI using 1.5-Tesla (T), 3.0-T, and 4.7-T MRI scanning protocols. The process was performed with the Ex-PRESS device with both the P-50 and R-50 models18.

Page 9: Ex-PRESS Glaucoma Filtration Device: Theory, Technique … · Ex-PRESS Glaucoma Filtration Device: Theory, ... there were concerns with persistent hypotony, ... nylon sutures

113

JOURNAL OF EMMETROPIA - VOL 4, APRIL-JUNE

imaging may be affected, but was of sufficient quality for clinical neuroradiological evaluation. No change in IOP was observed in any eye in the first month following the MRI19. On plain radiographic images (X-ray) or computer tomography (CT) scans, the Ex-PRESS device appears as a radiopaque object1.

Biocompatibility Postoperative inflammation and scarring after

implantation of an implant stems from both the trauma of surgery and the biocompatibility of the implant. The biocompatibility of the Ex-PRESS device was established in a rabbit model. The device was implanted in the anterior chamber at the corneoscleral junction in the eyes of 8 white New Zealand rabbits while the other eye served as a control in each of the animals. Histopathological evaluation was carried out at 3 and 6 months postimplantation and found no evidence of active inflammation, and the lumen was patent in all samples at both timepoints. A thin, mature, fibrotic capsule of less than 0.04 mm thickness of fibroblastic nuclei without any inflammatory cells was noted in

most cases; it covered up to 25% of the implant’s outer surface area with the lumen remaining devoid of any obstruction in all cases. There was no evidence of pathological reaction at 3 or 6 months in any ocular tissue within the proximity of the device (cornea, limbus, sclera, iris, lens, vitreous body) nor in any ocular tissues distant from the device (choroid, retina). All testing was undertaken in rabbit models, which is suitable due to the greater fibrin formation and cellular proliferation in the rabbit eye compared to the human eye. As such, the low-grade reaction in this rabbit model bodes favourably for application to human eyes 3.

ConclusionThe Ex-PRESS glaucoma device provides an

alternative to conventional trabeculectomy and appears to have similar IOP lowering effects with a possible improved safety profile and improved intraoperative control and predictability. In patients who are being considered for trabeculectomy with or without phacoemulsification, consideration should be made for the Ex-PRESS shunt.

ADVERSE EFFECTS Conjunctival leakage (6%), hyphema (2%), choroidal detachment (2%), wound gap (1%)

Transient hypotony spontaneously resolved (20.8%), device iris touch (12.5%), transient choroidal effusion spontaneously resolved (8.3%), hyphema (8.3%), shallow AC requiring reformation (4.1%)

After Ex-PRESS, all clear grafts remained clear; edematous grafts became clearer; opaque grafts showed no change.No change in central corneal graft thickness.Bleb needling to lower IOP (20%)

Serous choroidal detachment (24.3%, with 2.7% necessitating intervention), hypotony (21.6%, with 16.2% necessitating AC fill)

EX group had lower ratesof early postoperative hypotony, choroidal effusion, self-limited bleb leaks, but not statistically significant.Similar rates of prolonged postoperative hypotony, reoperation for glaucoma.EX group had higher rates of multiple laser suture lysis (LSL) procedures, and more LSL were done per patient*.Ex-PRESS group had higher rates of bleb leak requiring additional sutures, patients requiring AC reformation(not statistically significant)

Significantly lower rates of early postoperative hypotony (4% vs 32%)*, and choroidal effusion (8% vs 38%)* in EX group. No difference in flat AC, hypotony maculopathy, hyphema, bleb leak, bleb needling, endophthalmitis. More LSL in EX (2.2 vs 1.7)*.

EX group had lower rates of postoperative hypotony, and hyphema (not statistically significant).More laser suture lysis in EX group (51% vs 20%)*, and earlier (11.5d vs 15.8d)*.

Lower rates of shallow AC, hypotony, choroidal detachment, hyphema, fall in visual acuity in EX (not statistically significant).

No difference in blocked tube, early postoperative hypotony, flat AC, choroidal hemorrhage, aqueous misdirection, exposed implant, dislocated implant, dysthetic bleb, bleb leak

No difference in rates of shallow AC, flat AC, bleb leak, choroidal detachment, hyphema, IOP spikes, visual acuity

EX group had less needling, and development of cataract (significance not reported)

* statistical significance (p < 0.05); IOP, Intraocular pressure, measured in mmHg; F/U, Follow up; VA, visual acuity; AC, anterior chamber; w, weeks; m, months; y, years; EX, Ex-PRESS; Trab, trabeculectomyComplete success: IOP within the predefined IOP range, without any glaucoma medications; Qualified success: IOP within the predefined IOP range, with the use of glaucoma medications; Total success: IOP within the predefined IOP range, with or without the use of glaucoma medications

Table 1. Data currently available on the Ex-PRESS glaucoma device implanted under a scleral flap (cont.)

AUTHOR, YEAR Lankaranian 20114 Dahan 20059 Ates 201010 De Feo 200911 Seider 201113 Maris 200712 Good 201120 Sugiyama 201114 Kanner 200915 De Jong 200916 De Jong 201117

EXPRESS GLAUCOMA FILTRATION DEVICE

Page 10: Ex-PRESS Glaucoma Filtration Device: Theory, Technique … · Ex-PRESS Glaucoma Filtration Device: Theory, ... there were concerns with persistent hypotony, ... nylon sutures

114

JOURNAL OF EMMETROPIA - VOL 4, APRIL-JUNE

REFERENCES1. Salim S. Ex-PRESS glaucoma filtration device-surgical

technique and outcomes. Int Ophthalmol Clin. 2011; 51:83-94.

2. Tam DY, Ahmed IIK. New Glaucoma Surgical Devices. In: Krieglstein GK & Weinreb RN. Essentials in Ophthalmology. 3rd ed. Heidelberg, Germany: Springer-Verlag Berlin Heidelberg. 2009:75-98.

3. Nyska A, Glovinsky Y, Belkin M, Epstein Y. Biocompatibility of the Ex-PRESS miniature glaucoma drainage implant. J Glaucoma. 2003;12:275-80.

4. Lankaranian D, Razeghinejad R, Prasad A, et al. Intermediate-term results of the Ex-PRESS miniature glaucoma implant under a scleral flap in previously operated eyes. Clin Exp Ophthalmol. 2011; 39:421-8.

5. Wamsley S, Moster MR, Rai S, et al. Results of the use of the Ex-PRESS miniature glaucoma implant in technically challenging, advanced glaucoma cases: a clinical pilot study. Am J Ophthalmol. 2004; 138:1049-51.

6. Stewart RM, Diamond JG, Ashmore ED, Ayyala RS. Complications following Ex-press glaucoma shunt implantation. Am J Ophthalmol. 2005; 140:340-1.

7. Rivier D, Roy S, Mermoud A. Ex-PRESS R-50 miniature glaucoma implant insertion under the conjunctiva combined with cataract extraction. J Cataract Refract Surg. 2007; 33:1946-52.

8. Traverso CE, De Feo F, Messas-Kaplan A, et al. Long term effect on IOP of a stainless steel glaucoma drainage implant (Ex-PRESS) in combined surgery with phacoemulsification. Brit J Ophthalmol. 2005; 89:425-9.

9. Dahan E, Carmichael TR. Implantation of a miniature glaucoma device under a scleral flap. J Glaucoma. 2005;14:98-102.

10. Ates H, Palamar M, Yagci A, Egrilmez S. Evaluation of Ex-PRESS mini glaucoma shunt implantation in refractory postpenetrating keratoplasty glaucoma. J Glaucoma 2010; 19:556-60.

11. De Feo F, Bagnis A, Bricola G, et al. Efficacy and safety of a steel drainage device implanted under a scleral flap. Can J Ophthalmol. 2009; 44:457-62.

12. Maris PJG, Ishida K, Netland PA. Comparison of trabeculectomy with Ex-PRESS miniature glaucoma device implanted under scleral flap. J Glaucoma 2007; 16:14-9.

13. Seider MI, Rofagha S, Lin SC, Stamper RL. Resident-performed Ex-PRESS Shunt Implantation Versus Trabeculectomy. J Glaucoma. 2012; 21:469-74

14. Sugiyama T, Shibata M, Kojima S, et al. The first report on intermediate-term outcome of Ex-PRESS glaucoma filtration device implanted under scleral flap in Japanese patients. Clin Ophthalmol. 2011; 5:1063-6.

15. Kanner EM, Netland PA, Sarkisian SR & Du H. Ex-PRESS miniature glaucoma device implanted under a scleral flap alone or combined with phacoemulsification cataract surgery. J Glaucoma. 2009; 18:488-91.

16. de Jong LA. The Ex-PRESS glaucoma shunt versus trabeculectomy in open-angle glaucoma: a prospective randomized study. Adv Ther. 2009; 26:336-45.

17. de Jong L, Lafuma A, Aguadé A, Berdeaux G. Five-year extension of a clinical trial comparing the EX-PRESS glaucoma filtration device and trabeculectomy in primary open-angle glaucoma. Clin Ophthalmol. 2011; 5:527-33.

18. Seibold LK, Rorrer RAL & Kahook MY. MRI of the Ex-PRESS stainless steel glaucoma drainage device. Brit J Ophthalmol. 2011; 95:251-4.

19. De Feo F, Roccatagliata L, Bonzano L, et al. Magnetic resonance imaging in patients implanted with Ex-PRESS stainless steel glaucoma drainage microdevice. Am J Ophthalmol. 2009; 147:907-11.

20. Good TJ, Kahook M.Y. Assessment of bleb morphologic features and postoperative outcomes after Ex-PRESS drainage device implantation versus trabeculectomy. Am J Ophthalmol. 2011; 151:507-13.

First author:

Ike K. Ahmed, MD

Credit Valley Eye Care, 3200 Erin Mills Parkway, Unit 1, Mississauga, Ontario, L5L 1W8, Canada.

EXPRESS GLAUCOMA FILTRATION DEVICE