evolved gas analysis: tga/ftir - ta instruments...wire will do – an electric guitar string works...
TRANSCRIPT
TAINSTRUMENTS.COMTAINSTRUMENTS.COM
Evolved Gas Analysis: TGA/FTIR
Michael Bradley, Ph.D, M.B.A. ; Senior Manager,
Product Applications – ThermoFisher Scientific
James Browne, Applications Scientist
TA Instruments
TAINSTRUMENTS.COMTAINSTRUMENTS.COM
EGA Analysis: TGA/FTIR
•Evolved Gas Analysis is a powerful tandem for
today’s analytical chemist
•Sometimes referred to as ‘hyphenated
techniques’, the two most common
combinations with TGA are FTIR and Mass
Spectrometry
•In today’s discussion, we present the basics of
TGA/FTIR analyses.
TAINSTRUMENTS.COMTAINSTRUMENTS.COM
EGA Analysis: TGA / FTIR; General
Experimental Considerations
•No special sample preparation is necessary – Simply run a TGA experiment.
•Normal ramp rates work well (10 °C / min).
•Dynamic heating generally results in better data than an isothermal hold depending on the volatility of the analyte.
TAINSTRUMENTS.COMTAINSTRUMENTS.COM
EGA ANALYSIS: TGA / FTIR
General Tips for Obtaining Good Data
TAINSTRUMENTS.COMTAINSTRUMENTS.COM
EGA Analysis – TGA/FTIR; General
Experimental Considerations
•Be sure the TGA is clean
•Periodically disconnect the spectrometer transfer line and run an isothermal in air or oxygen at 1000 °C.
•This is especially important if you are running olefinic or oily samples.
•Clean the adapter port, make sure it is not clogged. Any stiff, thin wire will do – an electric guitar string works well. The thin ‘E’ string is good for the EGA adapter on the Discovery a wound ‘D’ string is good for the FTIR transfer line.
•Be careful cleaning the EGA adapter – it is very hot and will give a serious burn. Also be careful running a cleaning wire into the transfer line – it will come out very hot.
• If necessary, follow the procedures in the help menu for more thorough cleaning of the furnace.
•Disconnect or re-route the flow from the adapter when running samples that are not TGA/FTIR samples.
TAINSTRUMENTS.COMTAINSTRUMENTS.COM
EGA Analysis: TGA/FTIR – Know Your
Background
•Begin each experiment by obtaining a single-beam spectrum and a 100% line
�N.B. – A 100% line is simply a background and absorbance spectrum scan. You should see no evidence of obvious absorbances. Positive absorbances probably mean that residual sample remains, negative absorbances mean that the background is changing and probably improving.
•Check with the spectrometer manufacturer for any information regarding expected signal to noise performance, etc.
•Check the single beam spectrum between runs, be certain there is no residual sample remaining. Run a 100% line if you are not sure.
•Check with the spectrometer manufacturer for cleaning procedures for the gas cell, windows, etc.
TAINSTRUMENTS.COMTAINSTRUMENTS.COM
TGA/FTIR – Single Beam Spectrum
TAINSTRUMENTS.COMTAINSTRUMENTS.COM
TGA/FTIR – 100% Line
TAINSTRUMENTS.COMTAINSTRUMENTS.COM
TGA/FTIR – 100% Line / Single Beam Overlay
TAINSTRUMENTS.COMTAINSTRUMENTS.COM
TGA/FTIR – Use Spectral Tools to Your
Advantage – i.e. Subtract Interferences
•One of the most common by products of the TGA experiment is water vapor.
•Because of the ubiquitous nature of water vapor, and its spectral properties, it often interferes with making positive identifications of evolved gases.
•Fortunately, there is a simple solution:
�Collect and store a background spectrum of water vapor
�Obtain a purged background spectrum and open the furnace and collect the lab atmosphere.
�Store it and use it to subtract from your sample spectra and also correct for over-subtractions of water that can occur occasionally.
�FTIR software packages provide for spectral subtractions –check the directions for your particular software.
TAINSTRUMENTS.COMTAINSTRUMENTS.COM
EGA Analysis: TGA/FTIR – Water Vapor and
CO2
TAINSTRUMENTS.COMTAINSTRUMENTS.COM
TGA/FTIR – Water Vapor Minus CO2 – Asymmetric Stretching
and Wagging Regions Set to a Straight Line.
TAINSTRUMENTS.COMTAINSTRUMENTS.COM
Example: Unknown Pharmaceutical
TAINSTRUMENTS.COMTAINSTRUMENTS.COM
Example: Unknown Pharmaceutical
TAINSTRUMENTS.COMTAINSTRUMENTS.COM
Example: Unknown Pharmaceutical – Higher
Temperature
TAINSTRUMENTS.COMTAINSTRUMENTS.COM
Example: Unknown Pharmaceutical – Higher
Temperature
TAINSTRUMENTS.COMTAINSTRUMENTS.COM
Example – Unknown Decomposition Product
-0.008
-0.006
-0.004
-0.002
0.000
0.002
0.004
0.006
0.008
0.010
0.012
0.014
0.016
0.018
0.020
Ab
so
rba
nc
e
500 1000 1500 2000 2500 3000 3500 4000
Wavenumbers (cm-1)
NIST Reference of Cyclopentanone
Unknown
Unknown minus water
TAINSTRUMENTS.COMTAINSTRUMENTS.COM
Example: Analysis of Polyphenylene Oxide
•Engineering polymer
�Heat resistant
�Good tensile properties
�High surface energy –
paintable
TAINSTRUMENTS.COMTAINSTRUMENTS.COM
Example: Analysis of Polyphenylene Oxide
TAINSTRUMENTS.COMTAINSTRUMENTS.COM
Example: Analysis of Polyphenylene Oxide
Gram-Schmidt
Reconstruction
will often
resemble
derivative of
the mass loss
with respect to
temperature
TAINSTRUMENTS.COMTAINSTRUMENTS.COM
473°C
Example: Analysis of Polyphenylene Oxide
Peak intensity
at 473 °C
TAINSTRUMENTS.COMTAINSTRUMENTS.COM
0.005
0.010
0.015
0.020
0.025
0.030
0.035
0.040
0.045
0.050
0.055
0.060
0.065
0.070
0.075
0.080
0.085
Ab
so
rba
nc
e
500 1000 1500 2000 2500 3000 3500 4000
Wavenumbers (cm-1)
Example: Analysis of Polyphenylene Oxide
Peak intensity
at 473 °C
TAINSTRUMENTS.COMTAINSTRUMENTS.COM
0.005
0.010
0.015
0.020
0.025
0.030
0.035
0.040
0.045
0.050
0.055
0.060
0.065
0.070
0.075
0.080
0.085
Ab
so
rba
nc
e
500 1000 1500 2000 2500 3000 3500 4000
Wavenumbers (cm-1)
Example: Analysis of Polyphenylene Oxide
Peak intensity
at 473 °C
TAINSTRUMENTS.COMTAINSTRUMENTS.COM
-0.03
-0.02
-0.01
0.00
0.01
0.02
0.03
0.04
0.05
0.06
0.07
0.08
0.09
0.10
0.11
0.12
0.13
0.14
Ab
so
rba
nc
e
1000 1500 2000 2500 3000 3500 4000
Wavenumbers (cm-1)
Example: Analysis of Polyphenylene Oxide
Reference Spectrum in
Red
TAINSTRUMENTS.COMTAINSTRUMENTS.COM
0.001
0.002
0.003
0.004
0.005
0.006
0.007
0.008
0.009
0.010
Ab
sorb
ance
500 1000 1500 2000 2500 3000 3500 4000
Wavenumbers (cm-1)
Example: Analysis of Polyphenylene Oxide
Spectrum at 650 °C
TAINSTRUMENTS.COMTAINSTRUMENTS.COM
-0.002
-0.001
0.000
0.001
0.002
0.003
0.004
0.005
0.006
0.007
0.008
0.009
0.010
0.011
Ab
so
rba
nc
e
500 1000 1500 2000 2500 3000 3500 4000
Wavenumbers (cm-1)
Example: Analysis of Polyphenylene Oxide
Spectrum at 650 °C
Over-subtraction of
background corrected
TAINSTRUMENTS.COMTAINSTRUMENTS.COM
EGA Analysis: TGA/FTIR - Summary
•TGA combined with FTIR is a very powerful analytical tool.
�Plastics
�Pharmaceuticals
�Fuels
�Biomass
�Hazardous Decomposition Products
•The search capabilities of ThermoFisher’s Mercury and
Spectra software make FTIR spectroscopy very accessible
to all analytical chemists.
•Simple sample preparation
TAINSTRUMENTS.COMTAINSTRUMENTS.COM
Thank You
The World Leader in Thermal Analysis,
Rheology, and Microcalorimetry