![Page 1: Evaporative Light Scattering Detectors from Polymer ...schulzov/HPLC/Materialy pro studenty/ELSD... · Column: PL Hi-Plex Ca 9µm, 300x7.7mm Eluent: Water Flow Rate: 0.6ml/min Temp:](https://reader031.vdocuments.site/reader031/viewer/2022020120/5b7788297f8b9a4c438d53e8/html5/thumbnails/1.jpg)
Evaporative Light Scattering Detectorsfrom Polymer Laboratories
![Page 2: Evaporative Light Scattering Detectors from Polymer ...schulzov/HPLC/Materialy pro studenty/ELSD... · Column: PL Hi-Plex Ca 9µm, 300x7.7mm Eluent: Water Flow Rate: 0.6ml/min Temp:](https://reader031.vdocuments.site/reader031/viewer/2022020120/5b7788297f8b9a4c438d53e8/html5/thumbnails/2.jpg)
Introduction
Polymer Laboratories has over 15 years of experience in ELSD
ELSD can outperform traditional detectors when analysing non-chromophoricsamples by HPLC
Traditional HPLC detectors such as UV and RI have limited capabilities:
UV and RI are not compatible with a wide range of solvents
RI detection is not gradient compatible
Different analytes produce different UV responses depending on their extinction co-efficient
ELSDs can detect anything that is less volatile than the mobile phase
ELSD is universal and compatible with a wide range of solvents
![Page 3: Evaporative Light Scattering Detectors from Polymer ...schulzov/HPLC/Materialy pro studenty/ELSD... · Column: PL Hi-Plex Ca 9µm, 300x7.7mm Eluent: Water Flow Rate: 0.6ml/min Temp:](https://reader031.vdocuments.site/reader031/viewer/2022020120/5b7788297f8b9a4c438d53e8/html5/thumbnails/3.jpg)
Introduction
Polymer Laboratories presents 3 models of ELSD, each offering high sensitivity for a wide range of applications:
PL-ELS 2100
Regular HPLC with low temperature operation for volatile compounds
PL-ELS 1000
Routine HPLC and GPC using high boiling point solvents
PL-ELS 1000µ
For microbore and capillary LC
![Page 4: Evaporative Light Scattering Detectors from Polymer ...schulzov/HPLC/Materialy pro studenty/ELSD... · Column: PL Hi-Plex Ca 9µm, 300x7.7mm Eluent: Water Flow Rate: 0.6ml/min Temp:](https://reader031.vdocuments.site/reader031/viewer/2022020120/5b7788297f8b9a4c438d53e8/html5/thumbnails/4.jpg)
New PL-ELS 2100 Detector
![Page 5: Evaporative Light Scattering Detectors from Polymer ...schulzov/HPLC/Materialy pro studenty/ELSD... · Column: PL Hi-Plex Ca 9µm, 300x7.7mm Eluent: Water Flow Rate: 0.6ml/min Temp:](https://reader031.vdocuments.site/reader031/viewer/2022020120/5b7788297f8b9a4c438d53e8/html5/thumbnails/5.jpg)
Principles of Operation
The ELSD principle of operation employs three distinct stages:
Nebulisation
Evaporation
Detection
Eluent inlet
Gas
Liquid wasteLight Source
![Page 6: Evaporative Light Scattering Detectors from Polymer ...schulzov/HPLC/Materialy pro studenty/ELSD... · Column: PL Hi-Plex Ca 9µm, 300x7.7mm Eluent: Water Flow Rate: 0.6ml/min Temp:](https://reader031.vdocuments.site/reader031/viewer/2022020120/5b7788297f8b9a4c438d53e8/html5/thumbnails/6.jpg)
Nebulisation
Eluent flow mixed with N2 or Air
Concentric nebuliser
Efficient nebulisation: stable droplet plume, uniform droplet size
Temperature independently controlled Narrower cone of the plume (minimise band broadening)
Small nebuliser chamber(reduces band broadening)
![Page 7: Evaporative Light Scattering Detectors from Polymer ...schulzov/HPLC/Materialy pro studenty/ELSD... · Column: PL Hi-Plex Ca 9µm, 300x7.7mm Eluent: Water Flow Rate: 0.6ml/min Temp:](https://reader031.vdocuments.site/reader031/viewer/2022020120/5b7788297f8b9a4c438d53e8/html5/thumbnails/7.jpg)
Evaporation
Droplets pass through heated drift tube
30cm straight tube
Removes mobile phase to leave particulate form of analyte
Temperature set according to analytevolatility
Temperature controlled by user
Important to have laminar flow(reduces band broadening)
![Page 8: Evaporative Light Scattering Detectors from Polymer ...schulzov/HPLC/Materialy pro studenty/ELSD... · Column: PL Hi-Plex Ca 9µm, 300x7.7mm Eluent: Water Flow Rate: 0.6ml/min Temp:](https://reader031.vdocuments.site/reader031/viewer/2022020120/5b7788297f8b9a4c438d53e8/html5/thumbnails/8.jpg)
Detection
Particles irradiated with light source
LED (ca 400-500 nm)
Particles scatter light according to their size (mass sensitive)
Scattered light is detected by photomultiplier at fixed angle from incident light
Detection independent of optical properties of analyte
![Page 9: Evaporative Light Scattering Detectors from Polymer ...schulzov/HPLC/Materialy pro studenty/ELSD... · Column: PL Hi-Plex Ca 9µm, 300x7.7mm Eluent: Water Flow Rate: 0.6ml/min Temp:](https://reader031.vdocuments.site/reader031/viewer/2022020120/5b7788297f8b9a4c438d53e8/html5/thumbnails/9.jpg)
Advantages of Evaporative Light Scattering Detection
Universal - responds to all compounds in the mobile phase
Not dependent on spectroscopic properties of analyte
Produces more uniform detection sensitivity for analytes
Not susceptible to baseline drift during gradient elution, temperature or solvent pump fluctuations
ELSD compatible with a much wider range of solvents compared to Refractive Index
![Page 10: Evaporative Light Scattering Detectors from Polymer ...schulzov/HPLC/Materialy pro studenty/ELSD... · Column: PL Hi-Plex Ca 9µm, 300x7.7mm Eluent: Water Flow Rate: 0.6ml/min Temp:](https://reader031.vdocuments.site/reader031/viewer/2022020120/5b7788297f8b9a4c438d53e8/html5/thumbnails/10.jpg)
Removes the need for derivatisation steps (eg amino acids, toxins)Fast setup and equilibrationSensitivity in the order of 1-50ng (on column)(depends on eluent flow rate)No interference from solvent front peaks(enables fast analysis)Removal of mobile phase eluent allows rapid HPLC gradientsFlow rates up to up to 5ml/min can be achieved with no affect on baseline stabilityIdeal for High Throughput Screening
Advantages of Evaporative Light Scattering Detection
![Page 11: Evaporative Light Scattering Detectors from Polymer ...schulzov/HPLC/Materialy pro studenty/ELSD... · Column: PL Hi-Plex Ca 9µm, 300x7.7mm Eluent: Water Flow Rate: 0.6ml/min Temp:](https://reader031.vdocuments.site/reader031/viewer/2022020120/5b7788297f8b9a4c438d53e8/html5/thumbnails/11.jpg)
Column: PLRP-S 100Å 5µm, 50x4.6mmEluent A: Water + 0.05% TFAEluent B: ACN + 0.05% TFAGradient: 5-95% B in 1 minDetector: PL-ELS 2100
(neb=30°C, evap=30°C, gas=1.6 SLM)
Sample: Indapamide(IND), Dibutyl phthalate (DBP)
Flow rate increased from 2ml/min up to 5ml/min
Note: IND is non-volatile, DBP is relatively volatile
Fast Gradient, Fast Flow Rate Capability
![Page 12: Evaporative Light Scattering Detectors from Polymer ...schulzov/HPLC/Materialy pro studenty/ELSD... · Column: PL Hi-Plex Ca 9µm, 300x7.7mm Eluent: Water Flow Rate: 0.6ml/min Temp:](https://reader031.vdocuments.site/reader031/viewer/2022020120/5b7788297f8b9a4c438d53e8/html5/thumbnails/12.jpg)
0
6
2
8
4
0
0 1 2 3
2ml/min
3ml/min
4ml/min
5ml/min
INDDBP
Retention time (min)
Fast Gradient, Fast Flow Rate Capability
![Page 13: Evaporative Light Scattering Detectors from Polymer ...schulzov/HPLC/Materialy pro studenty/ELSD... · Column: PL Hi-Plex Ca 9µm, 300x7.7mm Eluent: Water Flow Rate: 0.6ml/min Temp:](https://reader031.vdocuments.site/reader031/viewer/2022020120/5b7788297f8b9a4c438d53e8/html5/thumbnails/13.jpg)
0.0 0.6 1.2 1.8 2.4 3.0
2ml/min
3ml/min
4ml/min
5ml/min
Stable baseline through the gradient
Fast Gradient, Fast Flow Rate Capability
![Page 14: Evaporative Light Scattering Detectors from Polymer ...schulzov/HPLC/Materialy pro studenty/ELSD... · Column: PL Hi-Plex Ca 9µm, 300x7.7mm Eluent: Water Flow Rate: 0.6ml/min Temp:](https://reader031.vdocuments.site/reader031/viewer/2022020120/5b7788297f8b9a4c438d53e8/html5/thumbnails/14.jpg)
Eluent A: 0.1% formic acid in water/ACN 95:5 v/vEluent B: 0.1% formic cid in water/ACN 5:95 v/vGradient: linear 0 - 100% B in 5 minFlow Rate: 0.9 mL/minSamples: 1 ng on columnDetector: ELS 1000
PL-ELS 2100 is transparent to DMSO at ambient temperature !!
High Throughput Screening
Nebuliser 30°C Evaporator30°C
(peak 2 DMSO )
![Page 15: Evaporative Light Scattering Detectors from Polymer ...schulzov/HPLC/Materialy pro studenty/ELSD... · Column: PL Hi-Plex Ca 9µm, 300x7.7mm Eluent: Water Flow Rate: 0.6ml/min Temp:](https://reader031.vdocuments.site/reader031/viewer/2022020120/5b7788297f8b9a4c438d53e8/html5/thumbnails/15.jpg)
Similar operating principles to LC-MS Volatile buffersFavours lower flow rates (ie 0.2-0.5ml/min)
Can develop LC methods on ELSD then transfer to LC-MS
ELSD can provide supporting information when used in tandem with LC-MS
Evaporative Light Scattering DetectionIdeal Complement to LC-MS
![Page 16: Evaporative Light Scattering Detectors from Polymer ...schulzov/HPLC/Materialy pro studenty/ELSD... · Column: PL Hi-Plex Ca 9µm, 300x7.7mm Eluent: Water Flow Rate: 0.6ml/min Temp:](https://reader031.vdocuments.site/reader031/viewer/2022020120/5b7788297f8b9a4c438d53e8/html5/thumbnails/16.jpg)
Ideal Complement to LC-MSSample Mixture of known 1:1 ratio
LC-MSresults show ratio to be
3:1
UV-Visresults show ratio to be
10:1
PL-ELS 2100results show ratio to be
1:1(Response independent
of optical properties)
![Page 17: Evaporative Light Scattering Detectors from Polymer ...schulzov/HPLC/Materialy pro studenty/ELSD... · Column: PL Hi-Plex Ca 9µm, 300x7.7mm Eluent: Water Flow Rate: 0.6ml/min Temp:](https://reader031.vdocuments.site/reader031/viewer/2022020120/5b7788297f8b9a4c438d53e8/html5/thumbnails/17.jpg)
Operation of the PL-ELS 2100
A new approach to ELSD applications
Nebuliser and evaporator temperatures determined by the volatility of the ANALYTE
Gas flow used to aid evaporation of the mobile phase
100% water @30°C requires high gas flow
100% Hexane @30°C requires no gas flow
PL-ELS 2100 can operate in 100% water @ 25°C
For Volatile & Semi-Volatile Solutes
![Page 18: Evaporative Light Scattering Detectors from Polymer ...schulzov/HPLC/Materialy pro studenty/ELSD... · Column: PL Hi-Plex Ca 9µm, 300x7.7mm Eluent: Water Flow Rate: 0.6ml/min Temp:](https://reader031.vdocuments.site/reader031/viewer/2022020120/5b7788297f8b9a4c438d53e8/html5/thumbnails/18.jpg)
Column: PL Hi-Plex Ca 9µm, 300x7.7mmEluent: WaterFlow Rate: 0.6ml/minTemp: 85°CInj Vol: 10µlDetector: PL-ELS 2100 (neb=30°C, evap=90°C, gas=1.6 SLM) Samples: 1. Fructose, 2. Glucose, 3. Sucrose I, 4. Lactose, 5. Stachyose
300 min
12
3
4
5
Non-Volatile Application: Sugars
![Page 19: Evaporative Light Scattering Detectors from Polymer ...schulzov/HPLC/Materialy pro studenty/ELSD... · Column: PL Hi-Plex Ca 9µm, 300x7.7mm Eluent: Water Flow Rate: 0.6ml/min Temp:](https://reader031.vdocuments.site/reader031/viewer/2022020120/5b7788297f8b9a4c438d53e8/html5/thumbnails/19.jpg)
Column: Lichrospher DIOL 5µm,150x2.1mmEluent A: IPA/Hexane/Water/Ammonia Hydroxide
57.8/40/2/0.2Eluent B: IPA/Hexane/Water/Ammonia Hydroxide
51.8/40/8/0.2 Gradient: 0-100% B in 7 mins (hold 8 mins)
100-0% B in 5 mins (hold 10mins)Flow Rate: 0.3ml/minDetector: PL-ELS 2100
(neb=30°C, evap=80°C, gas=1.0 SLM)
Samples: 1. Cholesterol2. Phosphatidylethanolamine3. Phosphatidylcholine4. Sphingomyelin5. Lysophosphatidylcholine
Non-Volatile Application: Phospholipid Separation
![Page 20: Evaporative Light Scattering Detectors from Polymer ...schulzov/HPLC/Materialy pro studenty/ELSD... · Column: PL Hi-Plex Ca 9µm, 300x7.7mm Eluent: Water Flow Rate: 0.6ml/min Temp:](https://reader031.vdocuments.site/reader031/viewer/2022020120/5b7788297f8b9a4c438d53e8/html5/thumbnails/20.jpg)
Column: Thermo-Hypersil ODS 5µm, 250x4.6mmEluent A: WaterEluent B: AcetonitrileGradient: 100% A in 5 mins hold, 0-40% B in 20 minsFlow Rate: 0.6ml/minInj Vol: 10µlDetector: PL-ELS 2100
(neb=50°C, evap=50°C, gas=1.6 SLM)
Semi-Volatile Application: Underivatised Amino Acids
![Page 21: Evaporative Light Scattering Detectors from Polymer ...schulzov/HPLC/Materialy pro studenty/ELSD... · Column: PL Hi-Plex Ca 9µm, 300x7.7mm Eluent: Water Flow Rate: 0.6ml/min Temp:](https://reader031.vdocuments.site/reader031/viewer/2022020120/5b7788297f8b9a4c438d53e8/html5/thumbnails/21.jpg)
0 30min
1. Serine2. Glutamic Acid3. Arginine4. Proline5. Valine6. Methionine7. Isoleucine8. Leucine9. Phenylanine10. Tryptophan
ELSD removes the need for derivatisation for applications such as amino acids
Semi-Volatile Application: Underivatised Amino Acids
![Page 22: Evaporative Light Scattering Detectors from Polymer ...schulzov/HPLC/Materialy pro studenty/ELSD... · Column: PL Hi-Plex Ca 9µm, 300x7.7mm Eluent: Water Flow Rate: 0.6ml/min Temp:](https://reader031.vdocuments.site/reader031/viewer/2022020120/5b7788297f8b9a4c438d53e8/html5/thumbnails/22.jpg)
Column: Adsorbosil C18 5µm, 150x4.6mmEluent A: Water + 0.1% TFAEluent B: ACN + 0.1% TFA Gradient: 60-90% B in 5 minsFlow Rate: 1.0ml/minDetector: PL-ELS 2100
(neb/evap=same temperature, gas=1.8 SLM)Samples: 1. Acetanilide
2. Indapamide3. Ibuprofen4. Dibutylphthalate
Note: Peak 2 is non-volatile Peaks 1, 3 & 4 are relatively volatile
Volatile Application: Effect of Changing Temperature
![Page 23: Evaporative Light Scattering Detectors from Polymer ...schulzov/HPLC/Materialy pro studenty/ELSD... · Column: PL Hi-Plex Ca 9µm, 300x7.7mm Eluent: Water Flow Rate: 0.6ml/min Temp:](https://reader031.vdocuments.site/reader031/viewer/2022020120/5b7788297f8b9a4c438d53e8/html5/thumbnails/23.jpg)
0Retention Time (min)
9
Ambient
35°C
50°C
1
2 3 4
• At high temperature, peak 2 has better S/N but peaks 1,3 & 4 are not detected• Running cold, all four peaks are detected
Volatile Application: Effect of Changing Temperature
![Page 24: Evaporative Light Scattering Detectors from Polymer ...schulzov/HPLC/Materialy pro studenty/ELSD... · Column: PL Hi-Plex Ca 9µm, 300x7.7mm Eluent: Water Flow Rate: 0.6ml/min Temp:](https://reader031.vdocuments.site/reader031/viewer/2022020120/5b7788297f8b9a4c438d53e8/html5/thumbnails/24.jpg)
Column: X-Terra C18 3.5µm, 30x2.1mmEluent A: Water Eluent B: ACN Gradient: 0-100% B in 3 mins (hold 0.7 min)Flow Rate: 0.5ml/minDetector: PL-ELS 2100
(neb=25°C, evap=25°C, gas=1.6 SLM)
Volatile Application: Phthalate SeparationSamples1. Diethylphthalate2. Dipropylphthalate3. Dibutylphthalate4. Dipentylphthalate5. Dioctylphthalate
![Page 25: Evaporative Light Scattering Detectors from Polymer ...schulzov/HPLC/Materialy pro studenty/ELSD... · Column: PL Hi-Plex Ca 9µm, 300x7.7mm Eluent: Water Flow Rate: 0.6ml/min Temp:](https://reader031.vdocuments.site/reader031/viewer/2022020120/5b7788297f8b9a4c438d53e8/html5/thumbnails/25.jpg)
Column: Adsorbosil C18 5µm, 150x4.6mmEluent A: Water Eluent B: ACN Gradient: 50-75% B in 7 minsFlow Rate: 1.0ml/minDetector: PL-ELS 2100
(Evap=30°C, Gas=1.0 SLM)
Nebuliser 30°C
Nebuliser 90°C
Volatile Application: Parabens SeparationSamples1. Methylparaben2. Ethylparaben3. Propylparaben
![Page 26: Evaporative Light Scattering Detectors from Polymer ...schulzov/HPLC/Materialy pro studenty/ELSD... · Column: PL Hi-Plex Ca 9µm, 300x7.7mm Eluent: Water Flow Rate: 0.6ml/min Temp:](https://reader031.vdocuments.site/reader031/viewer/2022020120/5b7788297f8b9a4c438d53e8/html5/thumbnails/26.jpg)
0 4min
Column: Adsorbosil C18 5µm, 150x4.6mmEluent: 60/40 Water/AcetonitrileFlow Rate: 1.0ml/minInj Vol: 10µlDetector: PL-ELS 2100
(neb=30°C, evap=30°C, gas=1.6 SLM) Sample: Urea 1mg/ml
PL-ELS 2100 can detect Urea,which has a MW 60
PL-ELS 2100Detection of Low Molecular Weight Compounds
![Page 27: Evaporative Light Scattering Detectors from Polymer ...schulzov/HPLC/Materialy pro studenty/ELSD... · Column: PL Hi-Plex Ca 9µm, 300x7.7mm Eluent: Water Flow Rate: 0.6ml/min Temp:](https://reader031.vdocuments.site/reader031/viewer/2022020120/5b7788297f8b9a4c438d53e8/html5/thumbnails/27.jpg)
Column: Spherisorb C18 5µm,150x4.6mmEluent A: 99% Water + 0.1% TFAEluent B: 1% ACN + 0.1% TFAFlow Rate: 0.5ml/minDetector: PL-ELS 2100
(neb=50°C, evap=70°C, gas=1.2 SLM)Sample: DL-Glutamic Acid
PL-ELS 2100Sensitivity: Limits of Detection
![Page 28: Evaporative Light Scattering Detectors from Polymer ...schulzov/HPLC/Materialy pro studenty/ELSD... · Column: PL Hi-Plex Ca 9µm, 300x7.7mm Eluent: Water Flow Rate: 0.6ml/min Temp:](https://reader031.vdocuments.site/reader031/viewer/2022020120/5b7788297f8b9a4c438d53e8/html5/thumbnails/28.jpg)
PL-ELS 2100Sensitivity: Limits of Detection
Column: ACE C8 3µm, 50x4.6mmEluent A: 100% Water Eluent B: 100% ACNIsocratic: 50/50 A/BFlow Rate: 1.0ml/minDetector: PL-ELS 2100
(neb=50°C, evap=50°C, gas=1.4 SLM)Sample: Caffeine
![Page 29: Evaporative Light Scattering Detectors from Polymer ...schulzov/HPLC/Materialy pro studenty/ELSD... · Column: PL Hi-Plex Ca 9µm, 300x7.7mm Eluent: Water Flow Rate: 0.6ml/min Temp:](https://reader031.vdocuments.site/reader031/viewer/2022020120/5b7788297f8b9a4c438d53e8/html5/thumbnails/29.jpg)
Low temperature operation even for 100% water
Extremely low dispersion for high resolution separations
High eluent flow rates, up to 5ml/min
Improved uniformity of response across a solvent gradient
Rapid equilibration
Extremely small footprint, stackable
Ergonomic design
Easy to use
Additional Benefits of the PL-ELS 2100
![Page 30: Evaporative Light Scattering Detectors from Polymer ...schulzov/HPLC/Materialy pro studenty/ELSD... · Column: PL Hi-Plex Ca 9µm, 300x7.7mm Eluent: Water Flow Rate: 0.6ml/min Temp:](https://reader031.vdocuments.site/reader031/viewer/2022020120/5b7788297f8b9a4c438d53e8/html5/thumbnails/30.jpg)
Les produits de nos partenaires Système LC ECD ALexys
![Page 31: Evaporative Light Scattering Detectors from Polymer ...schulzov/HPLC/Materialy pro studenty/ELSD... · Column: PL Hi-Plex Ca 9µm, 300x7.7mm Eluent: Water Flow Rate: 0.6ml/min Temp:](https://reader031.vdocuments.site/reader031/viewer/2022020120/5b7788297f8b9a4c438d53e8/html5/thumbnails/31.jpg)
Système LC ECD ALexys
Système intégré dédié à la détection électrochimique: Flexibilté : existe en version LC ou µLC
Sensibilité:
Ensemble du système optimisé pour la détection électrochimique
Compatibilité CFR 21 Par 11
Composants du système :•Pompe, amortisseur de pulsations•Dégazeur•Injecteur automatique analytique ou micro•DECADE II•Système de traitement de données
![Page 32: Evaporative Light Scattering Detectors from Polymer ...schulzov/HPLC/Materialy pro studenty/ELSD... · Column: PL Hi-Plex Ca 9µm, 300x7.7mm Eluent: Water Flow Rate: 0.6ml/min Temp:](https://reader031.vdocuments.site/reader031/viewer/2022020120/5b7788297f8b9a4c438d53e8/html5/thumbnails/32.jpg)
Decade II Véritable plate forme de travail
• Un four, Cage de faraday, abritant cellule de détection et colonne
• Pouvant contrôler une ou deux cellules de détection ( mode série ou parallèle)
• Concept de cellule original ( vaste choix d’électrodes de travail et de référence)
• Filtrage efficace du bruit de fond (ADF)
![Page 33: Evaporative Light Scattering Detectors from Polymer ...schulzov/HPLC/Materialy pro studenty/ELSD... · Column: PL Hi-Plex Ca 9µm, 300x7.7mm Eluent: Water Flow Rate: 0.6ml/min Temp:](https://reader031.vdocuments.site/reader031/viewer/2022020120/5b7788297f8b9a4c438d53e8/html5/thumbnails/33.jpg)
Cellule ampérométrique VT-03
• Cellule de type Wall Jet
• Stabilisation rapide,
• Faible volume(0,3µL à 10nL)
• Sensibilité
A: couche mince, B wall jet, C Poreuse
![Page 34: Evaporative Light Scattering Detectors from Polymer ...schulzov/HPLC/Materialy pro studenty/ELSD... · Column: PL Hi-Plex Ca 9µm, 300x7.7mm Eluent: Water Flow Rate: 0.6ml/min Temp:](https://reader031.vdocuments.site/reader031/viewer/2022020120/5b7788297f8b9a4c438d53e8/html5/thumbnails/34.jpg)
ADF avdanced digital filtrationFiltre conçu spécialement pour la détection
électrochimiqueNA - A - DHBA - D - 5HTRY - 5HT
-0.01
0
0.01
0.02
0.03
0.04
0.05
0.06
0 2 4 6 8 10 12 14 16 18 20t (min)
I (nA
)10Hz
0.005Hz
Les basses fréquences ( pics) passent
Les fréquences élevées (bruit) sont bloquées
![Page 35: Evaporative Light Scattering Detectors from Polymer ...schulzov/HPLC/Materialy pro studenty/ELSD... · Column: PL Hi-Plex Ca 9µm, 300x7.7mm Eluent: Water Flow Rate: 0.6ml/min Temp:](https://reader031.vdocuments.site/reader031/viewer/2022020120/5b7788297f8b9a4c438d53e8/html5/thumbnails/35.jpg)
Applications types
• Analyse de traces nécessitant: séléctivité et sensibilité
• Applications en Micro dialyse ( Off line ou On line)
![Page 36: Evaporative Light Scattering Detectors from Polymer ...schulzov/HPLC/Materialy pro studenty/ELSD... · Column: PL Hi-Plex Ca 9µm, 300x7.7mm Eluent: Water Flow Rate: 0.6ml/min Temp:](https://reader031.vdocuments.site/reader031/viewer/2022020120/5b7788297f8b9a4c438d53e8/html5/thumbnails/36.jpg)
ALEXYS system flow 1 ml/min; injection 25 ul (ul pick-up mode)
column 10 x 0.46 cm 3 um ODS-2; spiked with 10 fmol DA + 5HT/25ul (= 4 x 10-
10 M)
ADF 0.01 Hz up to 10 min, thereafter 0.002 Hz
Micro dialysate (frontal cortex + nucleus accumbens) settings