evaluation of the hybrid cochlear implant system: clinical ......if available in both ears: better...

20
NYU Cochlear Implant Center New York, New York Evaluation of the Hybrid Cochlear Implant System: Clinical Trial Results William H. Shapiro ACIA- Nashville 2014

Upload: others

Post on 28-Sep-2020

1 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Evaluation of the Hybrid Cochlear Implant System: Clinical ......If available in both ears: Better localization and Improved hearing for speech in the presence of spatially-separated

NYU Cochlear Implant Center New York, New York

Evaluation of the Hybrid Cochlear Implant System: Clinical Trial Results William H. Shapiro

ACIA- Nashville 2014

Page 2: Evaluation of the Hybrid Cochlear Implant System: Clinical ......If available in both ears: Better localization and Improved hearing for speech in the presence of spatially-separated

Cochlear™ Nucleus® Hybrid™ Implant System Components

Nucleus Hybrid L24 Implant

Remote Assistant Options

Nucleus 6 Sound Processor

Electric Component

Acoustic Component

Custom Sound Fitting Software – Version 4.1

Page 3: Evaluation of the Hybrid Cochlear Implant System: Clinical ......If available in both ears: Better localization and Improved hearing for speech in the presence of spatially-separated

Hybrid L24: Viable option for the “in-between” patient

Hybrid L24

Hybrid L24 Conventional, FLT, &

implantable HAs

Cochlear Implantation

Page 4: Evaluation of the Hybrid Cochlear Implant System: Clinical ......If available in both ears: Better localization and Improved hearing for speech in the presence of spatially-separated

Low Frequencies High Frequencies

Acoustic Hearing Electric Hearing (CI)

Why Hybrid Hearing?

Restores HF Sensitivity Good frequency

resolution

Maintain low-frequency acoustic hearing •  Easier to understand speech in noise

•  Better music appreciation

•  More natural sound quality

Acoustic

Restore high-frequency sensitivity •  Better speech perception

•  Completes the spectrum

Electric

If available in both ears: Better localization and Improved hearing for speech in the presence of spatially-separated noise

Two Ears

Page 5: Evaluation of the Hybrid Cochlear Implant System: Clinical ......If available in both ears: Better localization and Improved hearing for speech in the presence of spatially-separated

Cochlear® Nucleus™ Hybrid L24 Implant

Designed  to  restore  hearing  in  higher  frequencies  through  electrical  s3mula3on.  

Soft Tip for Minimal insertion trauma 1

2

3 Apical Diameter: .25mm 22 platinum electrode contacts

4

5

6

7

8

Tapered basal stiffener designed for a smooth, single motion insertion Basal Diameter: .4mm White Stopper at 16mm to indicate a full insertion Surgical Handle located opposite of the electrode contacts to assist with electrode orientation and atraumatic insertion

15 mm of active length

Page 6: Evaluation of the Hybrid Cochlear Implant System: Clinical ......If available in both ears: Better localization and Improved hearing for speech in the presence of spatially-separated

Candidacy criteria

Hybrid Aided word score between 10% and 60% in the ear to be implanted; 80% or less contralateral ear

Page 7: Evaluation of the Hybrid Cochlear Implant System: Clinical ......If available in both ears: Better localization and Improved hearing for speech in the presence of spatially-separated

HYBRID L24 PERFORMANCE OUTCOMES

7

Page 8: Evaluation of the Hybrid Cochlear Implant System: Clinical ......If available in both ears: Better localization and Improved hearing for speech in the presence of spatially-separated

Study Objectives

•  Primary (Hybrid mode): –  Mean scores for CNC (word and phoneme) measures, and AzBio

(sentences-in-noise) measure, at 6 months postactivation will be significantly improved over the preoperative measures (unilateral acoustic-only condition).

•  Secondary (Hybrid mode): –  On both CNC words and phonemes, most (> 75%) of the subjects

will score equal to or better than they did in the preoperative unilateral acoustic-only condition, and

–  On the AzBio sentences-in-noise, most (> 75%) of the subjects will score equal to or better than they did in the preoperative unilateral acoustic-only condition. Presentation Title Goes Here 8

Page 9: Evaluation of the Hybrid Cochlear Implant System: Clinical ......If available in both ears: Better localization and Improved hearing for speech in the presence of spatially-separated

Study Sites: Nucleus Hybrid L24 Implant

10 Implanting Centers • NYU (Lead Site) • Center for Hearing & Balance • Hearts for Hearing • Mayo Clinic, Rochester • Midwest Ear Institute • Northwestern University • The Ohio State University • Rocky Mtn Ear Center • University of Cincinnati • University of Iowa

Page 10: Evaluation of the Hybrid Cochlear Implant System: Clinical ......If available in both ears: Better localization and Improved hearing for speech in the presence of spatially-separated

Demographics Hybrid L24 Preoperative Audiograms

Frequency (Hz)125 250 500 750 1000 1500 2000 3000 4000 6000 8000

Hear

ing T

hres

hold

(dB

HL)

-10

0

10

20

30

40

50

60

70

80

90

100

110

120

130

N = 50

Hybrid L24 Preoperative AudiogramsFrequency (Hz)

125 250 500 750 1000 1500 2000 3000 4000 6000 8000

Hear

ing Th

resh

old (d

B HL

)

-10

0

10

20

30

40

50

60

70

80

90

100

110

120

130

N = 50

Ear  to  be  Implanted   Contralateral  Ear  

Characteris*c   Average  (S.D.)  

Age  at  Implanta3on   64.1    years  (±14.7)  

Dura3on  of  Hearing  Loss   28.1  years    (±14.9)  

Preopera3ve    Aided  CNC  Score  –  Implant  Ear   29.0%  (±14.6%)  

Preopera3ve    Aided  AzBio  Sentence  Score                                (+  5dB  SNR)  –  Implant  Ear   16.3%  (±14.4%)  

Page 11: Evaluation of the Hybrid Cochlear Implant System: Clinical ......If available in both ears: Better localization and Improved hearing for speech in the presence of spatially-separated

Frequency (Hz)

Soundfield Thresholds

125 250 500 750 1000 1500 2000 3000 4000

Hear

ing Le

vel (d

B)

0

10

2030

40

506070

80

90

100

110

120

Hearing Aid PreoperativeHybrid Initial Activation

Pre to Post Hearing Thresholds

Initial Activation

Low

Hearing sensitivity: RESTORED HIGH FREQUENCY AUDIBILITY

Page 12: Evaluation of the Hybrid Cochlear Implant System: Clinical ......If available in both ears: Better localization and Improved hearing for speech in the presence of spatially-separated

Speech Understanding 6 Months Post-Activation

CNC Words AzBio +5dB SNR

Perc

ent C

orre

ct

0

10

20

30

40

50

60

70

80

90

100Acoustic Alone PreHybrid Mode 6m

CNC Words AzBio +5dB SNR

Perc

ent C

orre

ct

0

10

20

30

40

50

60

70

80

90

100

Bilateral Acoustic PreCombined Mode 6m

34 points p<0.0001 33points

p<0.0001 36 points p<0.0001 33 points

p<0.0001

Speech perception significantly improved in both conditions

Page 13: Evaluation of the Hybrid Cochlear Implant System: Clinical ......If available in both ears: Better localization and Improved hearing for speech in the presence of spatially-separated

CNC Word Recognition: Different listening conditions at 6 months

Listening Conditions

Acoustic Bilateral Acoustic Electric Hybrid Combined

Perc

ent C

orre

ct

0

10

20

30

40

50

60

70

80

90

100

Preoperative Postoperative

N=38

28%

43%

18%

52%

70% 80%

Page 14: Evaluation of the Hybrid Cochlear Implant System: Clinical ......If available in both ears: Better localization and Improved hearing for speech in the presence of spatially-separated

Changing an Industry: Defining Preservation

Two Definitions - Two very different outcomes 1. Preservation of measurable acoustic hearing

•  Definition used by industry and in most publications •  Measure of detection/audibility •  Indicator of atraumatic surgery

2. Preservation of functional acoustic hearing •  A severe or better degree of hearing loss post-implantation

may provide a level of acoustic hearing that enhances the electrical stimulation provided by the cochlear implant1

1. Lenarz et. Al (2013) European Multi-center Study of the Nucleus Hybrid L24 cochlear implant. International Journal of Audiology. 1-11.

Functional = 5 frequency LFPTA (125 – 1000Hz)

Page 15: Evaluation of the Hybrid Cochlear Implant System: Clinical ......If available in both ears: Better localization and Improved hearing for speech in the presence of spatially-separated

Freedom Hybrid E Only Hybrid Mode

CNC Word Recognition6 Months Postactivation

Perc

ent C

orre

ct

0

10

20

30

40

50

60

70

80

90

100

MediansF = 54H E= 54.5HE = 82

Mean = 53.6

Mean = 76

Mean = 50

N=53 N=50 N=33

Why Hybrid Hearing?

Page 16: Evaluation of the Hybrid Cochlear Implant System: Clinical ......If available in both ears: Better localization and Improved hearing for speech in the presence of spatially-separated

Summary of Trial Results

• Audibility restored for the entire frequency range

• Outcomes on speech discrimination and self-assessment measures demonstrate statistically significant improvements

-Over 30% improvement on average in both quiet and noise conditions for the implanted ear

• At 6 months post-implant most individuals (90%) retain a level of measurable hearing and many (66%) utilize that hearing functionally in the implant ear

• Self assessment results corroborate speech perception results

Page 17: Evaluation of the Hybrid Cochlear Implant System: Clinical ......If available in both ears: Better localization and Improved hearing for speech in the presence of spatially-separated

Pending FDA approval

Thank you for listening

Page 18: Evaluation of the Hybrid Cochlear Implant System: Clinical ......If available in both ears: Better localization and Improved hearing for speech in the presence of spatially-separated

CLINICAL SIGNIFICANCE OF LOW FREQUENCY ACOUSTIC SENSTIVITY FOR THE IMPLANTED EAR

18

Page 19: Evaluation of the Hybrid Cochlear Implant System: Clinical ......If available in both ears: Better localization and Improved hearing for speech in the presence of spatially-separated

19

Improvement in Word Recognition and Sentence Recognition in Noise at 6 Months Hybrid (N=50)

Improvement in CNC Word Recognition(Percentage Points)

-40 -30 -20 -10 0 10 20 30 40 50 60 70 80 90 100

Impr

ovem

ent in

AzB

io Se

nten

ceRe

cogn

ition a

t + 5

dB S

NR

-40

-30

-20

-10

0

10

20

30

40

50

60

70

80

90

100

Group 1 (N=33)Group 2 (N=17)

Significant Decrementin Word Score

Significant Improvementin Word Score

Significant Decrementin AzBio

Significant Improvementin AzBio

Page 20: Evaluation of the Hybrid Cochlear Implant System: Clinical ......If available in both ears: Better localization and Improved hearing for speech in the presence of spatially-separated

R

R

R

Improvement in Word Recognition and Sentence Recognition in Noise at 6 Months Both Ears (N=50)

Improvement in CNC Word Recognition(Percentage Points)

-40 -30 -20 -10 0 10 20 30 40 50 60 70 80 90 100

Impr

ovem

ent in

AzB

io Se

nten

ceRe

cogn

ition a

t + 5

dB S

NR

-40

-30

-20

-10

0

10

20

30

40

50

60

70

80

90

100

Group 1 (N=33)Group 2 (N=17)

Significant Decrementin Word Score

Significant Improvementin Word Score

Significant Decrementin AzBio

Significant Improvementin AzBio