![Page 1: Estimation of Pulse Arrival Time Using Impedance ......Illustration retrieved from O. G. Martinsen, S. Grimnes, Bioimpedance and Bioelectricity Basics, 5/14 2nd ed., London: Academic](https://reader033.vdocuments.site/reader033/viewer/2022060211/5f04e2b17e708231d4103296/html5/thumbnails/1.jpg)
Estimation of Pulse Arrival Time Using Impedance Plethysmogram from Body Composition Scales
Birutė Paliakaitė1, Saulius Daukantas2, Andrius Sakalauskas2,Vaidotas Marozas1,2
1Department of Electronics Engineering, Kaunas University of Technology2Biomedical Engineering Institute, Kaunas University of Technology
April 13-15, 2015
![Page 2: Estimation of Pulse Arrival Time Using Impedance ......Illustration retrieved from O. G. Martinsen, S. Grimnes, Bioimpedance and Bioelectricity Basics, 5/14 2nd ed., London: Academic](https://reader033.vdocuments.site/reader033/viewer/2022060211/5f04e2b17e708231d4103296/html5/thumbnails/2.jpg)
Motivation
Arterial stiffness leads to the development of cardiovascular morbidity and mortality1.
Central (aortic) stiffness:
– elderly subjects,
– end-stage renal disease,
– hypertension,
– impaired glucose tolerance.
Peripheral (lower-limbs) stiffness:
– peripheral artery disease,
– diabetic peripheral neuropathy.
1/141S. Laurent et al. Expert consensus document on arterial stiffness: methodological issues and clinical applications. Eur. Heart J., vol. 27, no. 21, 2006.
![Page 3: Estimation of Pulse Arrival Time Using Impedance ......Illustration retrieved from O. G. Martinsen, S. Grimnes, Bioimpedance and Bioelectricity Basics, 5/14 2nd ed., London: Academic](https://reader033.vdocuments.site/reader033/viewer/2022060211/5f04e2b17e708231d4103296/html5/thumbnails/3.jpg)
Motivation
Arterial stiffness leads to the development of cardiovascular morbidity and mortality.
Central (aortic) stiffness1:
– elderly subjects,
– end-stage renal disease,
– hypertension,
– impaired glucose tolerance.
Peripheral (lower-limbs) stiffness:
– peripheral artery disease,
– diabetic peripheral neuropathy.
1/141S. Laurent et al. Expert consensus document on arterial stiffness: methodological issues and clinical applications. Eur. Heart J., vol. 27, no. 21, 2006.
![Page 4: Estimation of Pulse Arrival Time Using Impedance ......Illustration retrieved from O. G. Martinsen, S. Grimnes, Bioimpedance and Bioelectricity Basics, 5/14 2nd ed., London: Academic](https://reader033.vdocuments.site/reader033/viewer/2022060211/5f04e2b17e708231d4103296/html5/thumbnails/4.jpg)
Motivation
Arterial stiffness leads to the development of cardiovascular morbidity and mortality.
Central (aortic) stiffness:
– elderly subjects,
– end-stage renal disease,
– hypertension,
– impaired glucose tolerance.
Peripheral (lower-limbs) stiffness:
– peripheral artery disease2,
– diabetic peripheral neuropathy3.
1/14
2H. Yokoyama et al. Pulse wave velocity in lower-limb arteries among diabetic patients with peripheral arterial disease. J. Atheroscler. Thromb., vol. 10, no. 4, 2003.3M. Edmonds et al. Blood flow in the diabetic neuropathic foot. Diabetologia, vol. 22, no. 1, 1982.
![Page 5: Estimation of Pulse Arrival Time Using Impedance ......Illustration retrieved from O. G. Martinsen, S. Grimnes, Bioimpedance and Bioelectricity Basics, 5/14 2nd ed., London: Academic](https://reader033.vdocuments.site/reader033/viewer/2022060211/5f04e2b17e708231d4103296/html5/thumbnails/5.jpg)
Background
Arterial stiffness can be characterized by the propagation of the pulse pressure wave (PPW) along the arterial tree.
2/14
![Page 6: Estimation of Pulse Arrival Time Using Impedance ......Illustration retrieved from O. G. Martinsen, S. Grimnes, Bioimpedance and Bioelectricity Basics, 5/14 2nd ed., London: Academic](https://reader033.vdocuments.site/reader033/viewer/2022060211/5f04e2b17e708231d4103296/html5/thumbnails/6.jpg)
Background
Arterial stiffness can be characterized by the propagation of the pulse pressure wave (PPW) along the arterial tree.
Pulse arrival time (PAT) – the time interval between the R-wave of the QRS complex and the particular point in the PPW.
2/14
![Page 7: Estimation of Pulse Arrival Time Using Impedance ......Illustration retrieved from O. G. Martinsen, S. Grimnes, Bioimpedance and Bioelectricity Basics, 5/14 2nd ed., London: Academic](https://reader033.vdocuments.site/reader033/viewer/2022060211/5f04e2b17e708231d4103296/html5/thumbnails/7.jpg)
Background
Arterial stiffness can be characterized by the propagation of the pulse pressure wave (PPW) along the arterial tree.
Pulse arrival time (PAT) – the time interval between the R-wave of the QRS complex and the particular point in the PPW.
2/14
![Page 8: Estimation of Pulse Arrival Time Using Impedance ......Illustration retrieved from O. G. Martinsen, S. Grimnes, Bioimpedance and Bioelectricity Basics, 5/14 2nd ed., London: Academic](https://reader033.vdocuments.site/reader033/viewer/2022060211/5f04e2b17e708231d4103296/html5/thumbnails/8.jpg)
Background
Arterial stiffness can be characterized by the propagation of the pulse pressure wave (PPW) along the arterial tree.
Pulse arrival time (PAT) – the time interval between the R-wave of the QRS complex and the particular point in the PPW.
2/14
![Page 9: Estimation of Pulse Arrival Time Using Impedance ......Illustration retrieved from O. G. Martinsen, S. Grimnes, Bioimpedance and Bioelectricity Basics, 5/14 2nd ed., London: Academic](https://reader033.vdocuments.site/reader033/viewer/2022060211/5f04e2b17e708231d4103296/html5/thumbnails/9.jpg)
Background
Arterial stiffness can be characterized by the propagation of the pulse pressure wave (PPW) along the arterial tree.
Pulse arrival time (PAT) – the time interval between the R-wave of the QRS complex and the particular point in the PPW.
2/14
PAT
Arterial stiffness
![Page 10: Estimation of Pulse Arrival Time Using Impedance ......Illustration retrieved from O. G. Martinsen, S. Grimnes, Bioimpedance and Bioelectricity Basics, 5/14 2nd ed., London: Academic](https://reader033.vdocuments.site/reader033/viewer/2022060211/5f04e2b17e708231d4103296/html5/thumbnails/10.jpg)
Problem
Long-term periodic monitoring needed
3/14
![Page 11: Estimation of Pulse Arrival Time Using Impedance ......Illustration retrieved from O. G. Martinsen, S. Grimnes, Bioimpedance and Bioelectricity Basics, 5/14 2nd ed., London: Academic](https://reader033.vdocuments.site/reader033/viewer/2022060211/5f04e2b17e708231d4103296/html5/thumbnails/11.jpg)
Problem
Long-term periodic monitoring needed
Available devices for PPW recording
3/14
![Page 12: Estimation of Pulse Arrival Time Using Impedance ......Illustration retrieved from O. G. Martinsen, S. Grimnes, Bioimpedance and Bioelectricity Basics, 5/14 2nd ed., London: Academic](https://reader033.vdocuments.site/reader033/viewer/2022060211/5f04e2b17e708231d4103296/html5/thumbnails/12.jpg)
Problem
Long-term periodic monitoring needed
Available devices for PPW recording
Illustration retrieved from http://www.atcormedical.com/3/14
![Page 13: Estimation of Pulse Arrival Time Using Impedance ......Illustration retrieved from O. G. Martinsen, S. Grimnes, Bioimpedance and Bioelectricity Basics, 5/14 2nd ed., London: Academic](https://reader033.vdocuments.site/reader033/viewer/2022060211/5f04e2b17e708231d4103296/html5/thumbnails/13.jpg)
Problem
Long-term periodic monitoring needed
Available devices for PPW recording
Illustrations retrieved from http://www.atcormedical.com/ 3/14
![Page 14: Estimation of Pulse Arrival Time Using Impedance ......Illustration retrieved from O. G. Martinsen, S. Grimnes, Bioimpedance and Bioelectricity Basics, 5/14 2nd ed., London: Academic](https://reader033.vdocuments.site/reader033/viewer/2022060211/5f04e2b17e708231d4103296/html5/thumbnails/14.jpg)
Problem
Long-term periodic monitoring needed
Available devices for PPW recording
Illustrations retrieved from http://www.atcormedical.com/ 3/14
![Page 15: Estimation of Pulse Arrival Time Using Impedance ......Illustration retrieved from O. G. Martinsen, S. Grimnes, Bioimpedance and Bioelectricity Basics, 5/14 2nd ed., London: Academic](https://reader033.vdocuments.site/reader033/viewer/2022060211/5f04e2b17e708231d4103296/html5/thumbnails/15.jpg)
Problem
Long-term periodic monitoring needed
Available devices for PPW recording
Illustrations retrieved from http://www.atcormedical.com/ and http://www.atcormedical.com/ 3/14
![Page 16: Estimation of Pulse Arrival Time Using Impedance ......Illustration retrieved from O. G. Martinsen, S. Grimnes, Bioimpedance and Bioelectricity Basics, 5/14 2nd ed., London: Academic](https://reader033.vdocuments.site/reader033/viewer/2022060211/5f04e2b17e708231d4103296/html5/thumbnails/16.jpg)
Problem
Long-term periodic monitoring needed
Available devices for PPW recording:
Illustrations retrieved from http://www.atcormedical.com/ and http://www.atcormedical.com/
Operator dependent
Results rely on the placement
3/14
![Page 17: Estimation of Pulse Arrival Time Using Impedance ......Illustration retrieved from O. G. Martinsen, S. Grimnes, Bioimpedance and Bioelectricity Basics, 5/14 2nd ed., London: Academic](https://reader033.vdocuments.site/reader033/viewer/2022060211/5f04e2b17e708231d4103296/html5/thumbnails/17.jpg)
New Approach (1)
Impedance plethysmography (IPG) to determine changing tissue volumes (e.g. blood)
4/14
∆𝑹 = 𝝆𝒍𝟐
∆𝒗
![Page 18: Estimation of Pulse Arrival Time Using Impedance ......Illustration retrieved from O. G. Martinsen, S. Grimnes, Bioimpedance and Bioelectricity Basics, 5/14 2nd ed., London: Academic](https://reader033.vdocuments.site/reader033/viewer/2022060211/5f04e2b17e708231d4103296/html5/thumbnails/18.jpg)
New Approach (1)
Impedance plethysmography (IPG) to determine changing tissue volumes (e.g. blood).
ECG and IPG electrodes integrated into unobtrusive devices (e.g. bathroom scales)
4/14
𝑹 = 𝝆𝒍
𝑨
Illustration retrieved from OMRON HBF-510 Instruction Manual
![Page 19: Estimation of Pulse Arrival Time Using Impedance ......Illustration retrieved from O. G. Martinsen, S. Grimnes, Bioimpedance and Bioelectricity Basics, 5/14 2nd ed., London: Academic](https://reader033.vdocuments.site/reader033/viewer/2022060211/5f04e2b17e708231d4103296/html5/thumbnails/19.jpg)
New Approach (2)
The lower-body IPG signals are the sum of the local impedances of all segments between the voltageelectrodes,
5/14
![Page 20: Estimation of Pulse Arrival Time Using Impedance ......Illustration retrieved from O. G. Martinsen, S. Grimnes, Bioimpedance and Bioelectricity Basics, 5/14 2nd ed., London: Academic](https://reader033.vdocuments.site/reader033/viewer/2022060211/5f04e2b17e708231d4103296/html5/thumbnails/20.jpg)
New Approach (2)
The lower-body IPG signals are the sum of the local impedances of all segments between the voltageelectrodes,
5/14Illustration retrieved from O. G. Martinsen, S. Grimnes, Bioimpedance and Bioelectricity Basics, 2nd ed., London: Academic Press, 2008.
but the influence of the lower parts is the greatest.
![Page 21: Estimation of Pulse Arrival Time Using Impedance ......Illustration retrieved from O. G. Martinsen, S. Grimnes, Bioimpedance and Bioelectricity Basics, 5/14 2nd ed., London: Academic](https://reader033.vdocuments.site/reader033/viewer/2022060211/5f04e2b17e708231d4103296/html5/thumbnails/21.jpg)
New Approach (2)
The lower-body IPG signals are the sum of the local impedances of all segments between the voltageelectrodes,
5/14Illustration retrieved from O. G. Martinsen, S. Grimnes, Bioimpedance and Bioelectricity Basics, 2nd ed., London: Academic Press, 2008.
but the influence of the lower parts is the greatest.
The goal of this study is to demonstrate that PAT from the heart to the foot can be estimated using ECG and IPG recorded on the bathroom scales.
![Page 22: Estimation of Pulse Arrival Time Using Impedance ......Illustration retrieved from O. G. Martinsen, S. Grimnes, Bioimpedance and Bioelectricity Basics, 5/14 2nd ed., London: Academic](https://reader033.vdocuments.site/reader033/viewer/2022060211/5f04e2b17e708231d4103296/html5/thumbnails/22.jpg)
Measurement Setup
Body composition scales (Omron)
ECG: wireless ECG transmitter (Biopac)
IPG: electrical bioimpedance unit (Biopac)
IPG: photoplethysmogram amplifier unit (Biopac)
Illustration retrieved from OMRON HBF-510 Instruction Manual 6/14
![Page 23: Estimation of Pulse Arrival Time Using Impedance ......Illustration retrieved from O. G. Martinsen, S. Grimnes, Bioimpedance and Bioelectricity Basics, 5/14 2nd ed., London: Academic](https://reader033.vdocuments.site/reader033/viewer/2022060211/5f04e2b17e708231d4103296/html5/thumbnails/23.jpg)
Measurement Setup
Body composition scales (Omron)
ECG: wireless ECG transmitter (Biopac)
IPG: electrical bioimpedance unit (Biopac)
IPG: photoplethysmogram amplifier unit (Biopac)
Illustration retrieved from OMRON HBF-510 Instruction Manual 6/14
![Page 24: Estimation of Pulse Arrival Time Using Impedance ......Illustration retrieved from O. G. Martinsen, S. Grimnes, Bioimpedance and Bioelectricity Basics, 5/14 2nd ed., London: Academic](https://reader033.vdocuments.site/reader033/viewer/2022060211/5f04e2b17e708231d4103296/html5/thumbnails/24.jpg)
Measurement Setup
Body composition scales (Omron)
ECG: wireless ECG transmitter (Biopac)
IPG: electrical bioimpedance unit (Biopac)
IPG: photoplethysmogram amplifier unit (Biopac)
Illustration retrieved from OMRON HBF-510 Instruction Manual 6/14
![Page 25: Estimation of Pulse Arrival Time Using Impedance ......Illustration retrieved from O. G. Martinsen, S. Grimnes, Bioimpedance and Bioelectricity Basics, 5/14 2nd ed., London: Academic](https://reader033.vdocuments.site/reader033/viewer/2022060211/5f04e2b17e708231d4103296/html5/thumbnails/25.jpg)
Measurement Setup
Body composition scales (Omron)
ECG: wireless ECG transmitter (Biopac)
IPG: electrical bioimpedance unit (Biopac)
PPG: photoplethysmogram amplifier unit (Biopac)
Illustration retrieved from OMRON HBF-510 Instruction Manual 6/14
![Page 26: Estimation of Pulse Arrival Time Using Impedance ......Illustration retrieved from O. G. Martinsen, S. Grimnes, Bioimpedance and Bioelectricity Basics, 5/14 2nd ed., London: Academic](https://reader033.vdocuments.site/reader033/viewer/2022060211/5f04e2b17e708231d4103296/html5/thumbnails/26.jpg)
Methods
Four healthy subjects (one woman)
Paced respiration (0.1 Hz) to cause hemodynamics changes
7/14
![Page 27: Estimation of Pulse Arrival Time Using Impedance ......Illustration retrieved from O. G. Martinsen, S. Grimnes, Bioimpedance and Bioelectricity Basics, 5/14 2nd ed., London: Academic](https://reader033.vdocuments.site/reader033/viewer/2022060211/5f04e2b17e708231d4103296/html5/thumbnails/27.jpg)
Methods
Four healthy subjects (one woman)
Paced respiration (0.1 Hz) to cause hemodynamics changes
7/14
![Page 28: Estimation of Pulse Arrival Time Using Impedance ......Illustration retrieved from O. G. Martinsen, S. Grimnes, Bioimpedance and Bioelectricity Basics, 5/14 2nd ed., London: Academic](https://reader033.vdocuments.site/reader033/viewer/2022060211/5f04e2b17e708231d4103296/html5/thumbnails/28.jpg)
50 kHz400 µArms
V
Methods
Four healthy subjects (one woman)
Paced respiration (0.1 Hz) to cause hemodynamics changes
foot-to-foot single-foot
Illustration retrieved from OMRON HBF-510 Instruction Manual
Measurement cases
7/14
![Page 29: Estimation of Pulse Arrival Time Using Impedance ......Illustration retrieved from O. G. Martinsen, S. Grimnes, Bioimpedance and Bioelectricity Basics, 5/14 2nd ed., London: Academic](https://reader033.vdocuments.site/reader033/viewer/2022060211/5f04e2b17e708231d4103296/html5/thumbnails/29.jpg)
50 kHz400 µArms
V
Methods
Four healthy subjects (one woman)
Paced respiration (0.1 Hz) to cause hemodynamics changes
foot-to-foot single-foot
Illustration retrieved from OMRON HBF-510 Instruction Manual
Measurement cases
7/14
![Page 30: Estimation of Pulse Arrival Time Using Impedance ......Illustration retrieved from O. G. Martinsen, S. Grimnes, Bioimpedance and Bioelectricity Basics, 5/14 2nd ed., London: Academic](https://reader033.vdocuments.site/reader033/viewer/2022060211/5f04e2b17e708231d4103296/html5/thumbnails/30.jpg)
50 kHz400 µArms
V
Methods
Four healthy subjects (one woman)
Paced respiration (0.1 Hz) to cause hemodynamics changes
foot-to-foot single-foot
Illustration retrieved from OMRON HBF-510 Instruction Manual
Measurement cases
7/14
![Page 31: Estimation of Pulse Arrival Time Using Impedance ......Illustration retrieved from O. G. Martinsen, S. Grimnes, Bioimpedance and Bioelectricity Basics, 5/14 2nd ed., London: Academic](https://reader033.vdocuments.site/reader033/viewer/2022060211/5f04e2b17e708231d4103296/html5/thumbnails/31.jpg)
Signal Processing
8/14
![Page 32: Estimation of Pulse Arrival Time Using Impedance ......Illustration retrieved from O. G. Martinsen, S. Grimnes, Bioimpedance and Bioelectricity Basics, 5/14 2nd ed., London: Academic](https://reader033.vdocuments.site/reader033/viewer/2022060211/5f04e2b17e708231d4103296/html5/thumbnails/32.jpg)
Signal Processing
8/14
![Page 33: Estimation of Pulse Arrival Time Using Impedance ......Illustration retrieved from O. G. Martinsen, S. Grimnes, Bioimpedance and Bioelectricity Basics, 5/14 2nd ed., London: Academic](https://reader033.vdocuments.site/reader033/viewer/2022060211/5f04e2b17e708231d4103296/html5/thumbnails/33.jpg)
Signal Processing
8/14
![Page 34: Estimation of Pulse Arrival Time Using Impedance ......Illustration retrieved from O. G. Martinsen, S. Grimnes, Bioimpedance and Bioelectricity Basics, 5/14 2nd ed., London: Academic](https://reader033.vdocuments.site/reader033/viewer/2022060211/5f04e2b17e708231d4103296/html5/thumbnails/34.jpg)
Signal Processing
8/14
![Page 35: Estimation of Pulse Arrival Time Using Impedance ......Illustration retrieved from O. G. Martinsen, S. Grimnes, Bioimpedance and Bioelectricity Basics, 5/14 2nd ed., London: Academic](https://reader033.vdocuments.site/reader033/viewer/2022060211/5f04e2b17e708231d4103296/html5/thumbnails/35.jpg)
Signal Processing
8/14
![Page 36: Estimation of Pulse Arrival Time Using Impedance ......Illustration retrieved from O. G. Martinsen, S. Grimnes, Bioimpedance and Bioelectricity Basics, 5/14 2nd ed., London: Academic](https://reader033.vdocuments.site/reader033/viewer/2022060211/5f04e2b17e708231d4103296/html5/thumbnails/36.jpg)
Signal Processing
8/14
![Page 37: Estimation of Pulse Arrival Time Using Impedance ......Illustration retrieved from O. G. Martinsen, S. Grimnes, Bioimpedance and Bioelectricity Basics, 5/14 2nd ed., London: Academic](https://reader033.vdocuments.site/reader033/viewer/2022060211/5f04e2b17e708231d4103296/html5/thumbnails/37.jpg)
Results: example of the signals
9/14
Foot-to-foot caseSingle-foot case
![Page 38: Estimation of Pulse Arrival Time Using Impedance ......Illustration retrieved from O. G. Martinsen, S. Grimnes, Bioimpedance and Bioelectricity Basics, 5/14 2nd ed., London: Academic](https://reader033.vdocuments.site/reader033/viewer/2022060211/5f04e2b17e708231d4103296/html5/thumbnails/38.jpg)
Results: example of the estimated PAT
10/14
Single-foot case
Foot-to-foot case
![Page 39: Estimation of Pulse Arrival Time Using Impedance ......Illustration retrieved from O. G. Martinsen, S. Grimnes, Bioimpedance and Bioelectricity Basics, 5/14 2nd ed., London: Academic](https://reader033.vdocuments.site/reader033/viewer/2022060211/5f04e2b17e708231d4103296/html5/thumbnails/39.jpg)
Results: example of estimated PAT
10/14
Single-foot case
Foot-to-foot case
![Page 40: Estimation of Pulse Arrival Time Using Impedance ......Illustration retrieved from O. G. Martinsen, S. Grimnes, Bioimpedance and Bioelectricity Basics, 5/14 2nd ed., London: Academic](https://reader033.vdocuments.site/reader033/viewer/2022060211/5f04e2b17e708231d4103296/html5/thumbnails/40.jpg)
Results: example of estimated PAT
10/14
Single-foot case
Foot-to-foot case
![Page 41: Estimation of Pulse Arrival Time Using Impedance ......Illustration retrieved from O. G. Martinsen, S. Grimnes, Bioimpedance and Bioelectricity Basics, 5/14 2nd ed., London: Academic](https://reader033.vdocuments.site/reader033/viewer/2022060211/5f04e2b17e708231d4103296/html5/thumbnails/41.jpg)
Results: example of estimated PAT
10/14
Single-foot case
Foot-to-foot case
![Page 42: Estimation of Pulse Arrival Time Using Impedance ......Illustration retrieved from O. G. Martinsen, S. Grimnes, Bioimpedance and Bioelectricity Basics, 5/14 2nd ed., London: Academic](https://reader033.vdocuments.site/reader033/viewer/2022060211/5f04e2b17e708231d4103296/html5/thumbnails/42.jpg)
Results: foot-to-foot vs. single-foot
Data represent mean±SD 11/14
![Page 43: Estimation of Pulse Arrival Time Using Impedance ......Illustration retrieved from O. G. Martinsen, S. Grimnes, Bioimpedance and Bioelectricity Basics, 5/14 2nd ed., London: Academic](https://reader033.vdocuments.site/reader033/viewer/2022060211/5f04e2b17e708231d4103296/html5/thumbnails/43.jpg)
Results: foot-to-foot vs. single-foot
Data represent mean±SD 11/14
![Page 44: Estimation of Pulse Arrival Time Using Impedance ......Illustration retrieved from O. G. Martinsen, S. Grimnes, Bioimpedance and Bioelectricity Basics, 5/14 2nd ed., London: Academic](https://reader033.vdocuments.site/reader033/viewer/2022060211/5f04e2b17e708231d4103296/html5/thumbnails/44.jpg)
Results: boxplot of the absolute valuesof PAT
12/14
![Page 45: Estimation of Pulse Arrival Time Using Impedance ......Illustration retrieved from O. G. Martinsen, S. Grimnes, Bioimpedance and Bioelectricity Basics, 5/14 2nd ed., London: Academic](https://reader033.vdocuments.site/reader033/viewer/2022060211/5f04e2b17e708231d4103296/html5/thumbnails/45.jpg)
Results: boxplot of the absolute valuesof PAT
12/14
![Page 46: Estimation of Pulse Arrival Time Using Impedance ......Illustration retrieved from O. G. Martinsen, S. Grimnes, Bioimpedance and Bioelectricity Basics, 5/14 2nd ed., London: Academic](https://reader033.vdocuments.site/reader033/viewer/2022060211/5f04e2b17e708231d4103296/html5/thumbnails/46.jpg)
Results: boxplot of the absolute valuesof PAT
12/14
![Page 47: Estimation of Pulse Arrival Time Using Impedance ......Illustration retrieved from O. G. Martinsen, S. Grimnes, Bioimpedance and Bioelectricity Basics, 5/14 2nd ed., London: Academic](https://reader033.vdocuments.site/reader033/viewer/2022060211/5f04e2b17e708231d4103296/html5/thumbnails/47.jpg)
Results: boxplot of the absolute valuesof PAT
12/14
![Page 48: Estimation of Pulse Arrival Time Using Impedance ......Illustration retrieved from O. G. Martinsen, S. Grimnes, Bioimpedance and Bioelectricity Basics, 5/14 2nd ed., London: Academic](https://reader033.vdocuments.site/reader033/viewer/2022060211/5f04e2b17e708231d4103296/html5/thumbnails/48.jpg)
System Implementation
Custom-made bioimpedance unit integrated into body composition scales
13/14
![Page 49: Estimation of Pulse Arrival Time Using Impedance ......Illustration retrieved from O. G. Martinsen, S. Grimnes, Bioimpedance and Bioelectricity Basics, 5/14 2nd ed., London: Academic](https://reader033.vdocuments.site/reader033/viewer/2022060211/5f04e2b17e708231d4103296/html5/thumbnails/49.jpg)
Conclusions and Future Directions
Conclusions
– PAT can be estimated by using IPG and ECG sensors, which are integrated into body composition scales;
– PAT evaluated by the method introduced in this study correlates with PPG-based PAT;
– single-foot and foot-to-foot PATIPG slightly differs.
Future directions
– testing of the custom-made system;
– development of the algorithm for the calculation of PAT;
– a wider group of subjects with different health status.
14/14
![Page 50: Estimation of Pulse Arrival Time Using Impedance ......Illustration retrieved from O. G. Martinsen, S. Grimnes, Bioimpedance and Bioelectricity Basics, 5/14 2nd ed., London: Academic](https://reader033.vdocuments.site/reader033/viewer/2022060211/5f04e2b17e708231d4103296/html5/thumbnails/50.jpg)
Acknowledgment
This work was partly supported by the projects “Promotion of Student Scientific Activities” (VP1-3.1-ŠMM-01-V-02-003) from the Research Council of Lithuania and CARRE (No.611140) funded by the European Community 7th Framework Programme.
![Page 51: Estimation of Pulse Arrival Time Using Impedance ......Illustration retrieved from O. G. Martinsen, S. Grimnes, Bioimpedance and Bioelectricity Basics, 5/14 2nd ed., London: Academic](https://reader033.vdocuments.site/reader033/viewer/2022060211/5f04e2b17e708231d4103296/html5/thumbnails/51.jpg)