1 of 4 L E S L I E D A N F A C U L T Y O F P H A R M A C Y U N I V E R S I T Y O F TORONTO ERWERKA D-6150 Hardness manual

Upload: others

Post on 01-Oct-2020

0 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: ERWERKA D-6150 Hardness manual - Pharmawiki.in · 3333 ooooffff 4444 12. On the menu bar located at the bottom left, click “Action” à “Calibration” à “Calibrate”. 13

1111 ooooffff 4444

L E S L I E D A N F A C U L T Y O F

P H A R M A C Y U N I V E R S I T Y O F

T O R O N T O

E R W E R K A D - 6 1 5 0

H a r d n e s s m a n u a l

Page 2: ERWERKA D-6150 Hardness manual - Pharmawiki.in · 3333 ooooffff 4444 12. On the menu bar located at the bottom left, click “Action” à “Calibration” à “Calibrate”. 13

2222 ooooffff 4444

P r e p a r i n g t h e

H a r d n e s s T e s t e r

1. Attach the Fujitsu Siemens Pocket PC to the hardness tester via the front panel. Slide the Pocket

PC in the rubber holder.

2. Make sure the SD start-up card is in the slot.

3. Plug in the hardness tester and turn it on using the switch on the back.

4. Follow the pocket PC prompts for set-up. Use the slide-out pen on the upper right side of the

Pocket PC.

5. At the home screen, remove the SD card and insert the “Siemens Update” SD card.

6. Click “Yes to All” if prompted to over-write files.

7. Click the “Start” button (upper left hand corner), then click “Programs” on the drop-down menu

8. Remove the “Siemens Update” SD card and replace it with the start-up card.

9. Click the “H-Test” icon.

10. Clean the tablet platform with the brush provided. Click “ok” to zero the jaw.

11. Enter the login account information:

• User : “admin”, Password: “abc123”, click “OK” to log in.

Page 3: ERWERKA D-6150 Hardness manual - Pharmawiki.in · 3333 ooooffff 4444 12. On the menu bar located at the bottom left, click “Action” à “Calibration” à “Calibrate”. 13

3333 ooooffff 4444

12. On the menu bar located at the bottom left, click “Action” à “Calibration” à “Calibrate”.

13. Dust of the stage with the brush and click the first “OK”.

14. Clean and insert the 10mm calibration block, polished finish facing the piston, and click the next

“OK”.

15. Clean and insert the 5mm calibration block, polished finish facing the piston, and click the next

“OK”.

16. Clean and insert the 10mm calibration block, polished finish facing the piston, and click the next

“OK”.

17. Clean and insert the 5 mm and 10mm calibration blocks together, polished finish facing the pis-

ton, and click the next “OK”.

18. Remove the calibration blocks from the stage.

C a l i b r a t i n g t h e

H a r d n e s s T e s t e r

Page 4: ERWERKA D-6150 Hardness manual - Pharmawiki.in · 3333 ooooffff 4444 12. On the menu bar located at the bottom left, click “Action” à “Calibration” à “Calibrate”. 13

4444 ooooffff 4444

19. Click “Action”, then “Measurement”, then “New”.

20. “Hardness” is the number of tablets to be tested. Set this to 10.

21. Leave the “Correction Factor” as 0.

22. Measure the diameter of your tablet with plastic calipers. Enter this value under “Diameter”.

23. Place a tablet on the stage, and click “Start”.

24. Click on “Inspection” to see the hardness values for each tablet.

25. The piston will squeeze the tablet until it breaks. While the piston pulls back, quickly sweep and

replace a new tablet on the stage for the next measurement until the last tablet is broken.

26. To review the data, click “Action” “Results” Select the most recent file (by date) “Show”

“Statistics”.

U s i n g t h e H a r d n e s s

T e s t e r