epr spectrometer & dosimetry

Post on 02-Jun-2018

234 views

Category:

Documents


0 download

TRANSCRIPT

  • 8/10/2019 EPR Spectrometer & Dosimetry

    1/43

    EPR spectrometer & dosimetry

    in Hokkaido University

    ()

    2014.09.16Kwon Choi

  • 8/10/2019 EPR Spectrometer & Dosimetry

    2/43

    Lets say about EPR

    Which exist around nucleus

    Unpaired electrons

    Electron Paramagnetic Resonance

  • 8/10/2019 EPR Spectrometer & Dosimetry

    3/43

    Lets say about EPR

    Paramagnetismisa form of magnetism whereby certain materialsare attractedby an externally applied magnetic field,and form internal, induced magnetic fieldsin the direction ofthe applied magnetic field. () Paramagnetic properties aredue to the presence of some unpaired electrons, (.)

    (from Wikipedia )

    Electron Paramagnetic ResonanceWhich you may not be familiar with

  • 8/10/2019 EPR Spectrometer & Dosimetry

    4/43

    Lets say about EPR

    In other word, align their directioninto the direction of the extenalmagnetic field.

    Electron Paramagnetic ResonanceWhich you may not be familiar with

  • 8/10/2019 EPR Spectrometer & Dosimetry

    5/43

    Lets say about EPR

    Electron Paramagnetic ResonanceWhich you may not be familiar with

    Form induced magnetic field

    Up spin and down spin electrons have different energy states.ThisEnergy Splittingis determined by the strength of theexternal magnetic field.

    =geBB0

    ge: g factor, 2.00231930436153 (a.u.)for electron

    B: Bohr magneton, 9.274009681024 (a.u.)

  • 8/10/2019 EPR Spectrometer & Dosimetry

    6/43

    Lets say about EPR

    Electron Paramagnetic ResonanceResonance absorption

    Absorption of electromagnetic energyat a certain frequencyin a static external magnetic fieldby a quantum excitation of the material

  • 8/10/2019 EPR Spectrometer & Dosimetry

    7/43

    Lets say about EPR

    Electron Paramagnetic ResonanceResonance absorption

    Energy

    Absorption of theelectromagneticenergy

    E =

    Release of theenergy (heat, light)

    Excitation of electrons Relaxation

    The electromagnetic energywith the energy equal to the energy splitting of electrons,= E = geBB0 makes the resonance

  • 8/10/2019 EPR Spectrometer & Dosimetry

    8/43

    2. Population distribution of electrons between two energy levels :

    (Boltzmanns equation)

    T : temperaturek : Boltzmann constant

    1. Under the resonance :

    The population of the lower level is greaterthan thatof the upper level.

    Net energy absorption

    http://encyclopedia2.thefreedictionary.com/Electron+Paramagnetic+Resonance+EPR

    Energyemission

    Energyabsorption

    Both transitions with

    same probability

    =#

    # <

    How energy absorption occur

    E = -1

    0

    E =1

    0

  • 8/10/2019 EPR Spectrometer & Dosimetry

    9/43

  • 8/10/2019 EPR Spectrometer & Dosimetry

    10/43

    The frequency of EM wave is determined by the utility.

    Frequency => Energy splitting (Better signal)Dielectric loss (Heat loss)

    Skin depth (Lower penetration)

  • 8/10/2019 EPR Spectrometer & Dosimetry

    11/43

    = ( 20

    )1/

    : Resistivity of material0: dielectric constant: Frequency

    Skin depth,

    How deep the electromagnetic wave can penetrate

    The frequency of EM wave is determined by the use.

    Frequency => Energy splitting (Better signal)Dielectric loss (Heat loss)

    Skin depth (Lower penetration)

    Dielectric lossIn water

  • 8/10/2019 EPR Spectrometer & Dosimetry

    12/43

    = ( 20

    )1/

    : Resistivity of material0: dielectric constant: Frequency

    Skin depth,

    How deep the electromagnetic wave can penetrate

    At 1 GHz

    (L-band)

    Resistivity

    (m)

    Skin depth

    (cm)Muscle 0.8 1.5

    Liver 1 1.7

    Fat 7-12 4-6

    Bone 10-25 5-8

    Brain 1 1.7

    The frequency of EM wave is determined by the use.

    Frequency => Energy splitting (Better signal)Dielectric loss (Heat loss)

    Skin depth (Lower penetration)

    Dielectric lossIn water

  • 8/10/2019 EPR Spectrometer & Dosimetry

    13/43

    = ( 20

    )1/

    : Resistivity of material0: dielectric constant: Frequency

    Skin depth,

    How deep the electromagnetic wave can penetrate

    At 1 GHz

    (L-band)

    Resistivity

    (m)

    Skin depth

    (cm)Muscle 0.8 1.5

    Liver 1 1.7

    Fat 7-12 4-6

    Bone 10-25 5-8

    Brain 1 1.7

    The frequency of EM wave is determined by the use.

    Frequency => Energy splitting (Better signal)Dielectric loss (Heat loss)

    Skin depth (Lower penetration)

    Dielectric lossIn water

    Under L-band (1-2 GHz) isappropriate for the use in vivo.

  • 8/10/2019 EPR Spectrometer & Dosimetry

    14/43

    =geBB0

    External magnetic field / mT

    Because of the magneticmodulation

    -> Basic form of an EPR signal

    For

    Microwave

    Variant frequency of EM wave isused by the use.

    External magnetic field / mT

  • 8/10/2019 EPR Spectrometer & Dosimetry

    15/43

    =geBB0

    External magnetic field / mT

    Because of the magneticmodulation

    -> Basic form of an EPR signal

    Microwave

    Variant frequency of EM wave isused by the use.

  • 8/10/2019 EPR Spectrometer & Dosimetry

    16/43

    What can we observe with EPR?

    Flaw of the

    optic fiber

    New material

    (Ion liquid)

    Process ofchemical

    reaction fromUV irradiation

    Tooth enamel ofthe irradiatedhuman

    In vivo brain

    / tumor

    How did radicals form in observed objects?

    Elapse of timeMix the stableradical into thenew material

    Unstable radicalsare formed bythe break of

    chemical bonds

    Stable radicals areformed in the

    tooth enamel

    Quantity ofradicals,relationship withthe amount ofradiation

    Quantity ofradicals,distribution,relationship withtumor

    Inject stable radicalsinto the vessel orregion of interest

    What information do you want?

    Quantity,distribution,location andquantity of flaw,

    Interactionbetween stableradical and newmaterial

    Dynamics(Change ofstructure,quantity,characteristicswith the timeelapse

    Enamel

  • 8/10/2019 EPR Spectrometer & Dosimetry

    17/43

    What can we observe with EPR?

    Flaw of the

    optic fiber

    New material

    (Ion liquid)

    Process ofchemical

    reaction fromUV irradiation

    Tooth enamel ofthe irradiatedhuman

    In vivo brain

    / tumor

    How did radicals form in observed objects?

    Elapse of timeMix the stableradical into thenew material

    Unstable radicalsare formed bythe break of

    chemical bonds

    Stable radicals areformed in the

    tooth enamel

    Quantity ofradicals,relationship withthe amount ofradiation

    Quantity ofradicals,distribution,relationship withtumor

    Inject stable radicalsinto the vessel orregion of interest

  • 8/10/2019 EPR Spectrometer & Dosimetry

    18/43

    Radiation dosimetry using tooth enamel

    Requirement

    Possibility of radiologic emergency situation(terrorist attack using a nuclear weapon, accident of nuclear power plants)

    Need of patients triage (distinguish who may be irradiated)- Who exposed to over certain amount of radiation need rapid care.- Victims with doses starting at about 2 Gy can benefit from urgent

    medical treatment.

    General publication does not have the proper dosimeter.

    EPR tooth dosimetry can be used for triagein situ.

    Conventional blood test can be done at remote place that takes time.

  • 8/10/2019 EPR Spectrometer & Dosimetry

    19/43

    Radiation dosimetry using tooth enamel

    Hydroxyapatite contained in teeth and bones-> Suitable probe for dose reconstruction.

    - High content of hydroxyapatite- Completely formed in childhood- Once formed never remodelled.

    1. Carbonate impurities in hydroxyapatite crystals

    Radiationexposure

    radicals

    (EPR sensitive stable radical as probe)

    2. Tooth enamel is preferred for the retrospective dosimetry

    -> Co2--radicals are accumulated over 107years

  • 8/10/2019 EPR Spectrometer & Dosimetry

    20/43

    "Boys, be ambitious!Be ambitious not for money or for

    selfish aggrandizement, not for thatevanescent thing which men call fame. Beambitious for that attainment of all that aman ought to be."

    Regrettably, today in Korea this goodmaxim is generally used asBoys, be ambitious for fame and money!

    , !, ..

    William S. Clark Ph.D.

  • 8/10/2019 EPR Spectrometer & Dosimetry

    21/43

    Farm (Part of the school of agriculture)

  • 8/10/2019 EPR Spectrometer & Dosimetry

    22/43

    Farm (Part of the school of agriculture)

  • 8/10/2019 EPR Spectrometer & Dosimetry

    23/43

  • 8/10/2019 EPR Spectrometer & Dosimetry

    24/43

  • 8/10/2019 EPR Spectrometer & Dosimetry

    25/43

    A little history about thediscovery of

    electron paramagnetic resonance

  • 8/10/2019 EPR Spectrometer & Dosimetry

    26/43

  • 8/10/2019 EPR Spectrometer & Dosimetry

    27/43

    Phenomenon called ElectronParamagnetic Resonance (EPR)

    Isior Issac Rabi

    Nobel Prize in Physics (1944)

    Discovery of Nuclear Magnetic Resonance (1938)

    Zavoisky was impressed at Rabis discovery, NMRand tried to detect NMR in liquids and solids.

  • 8/10/2019 EPR Spectrometer & Dosimetry

    28/43

    Tooth dosimetry in Hokkaido University

  • 8/10/2019 EPR Spectrometer & Dosimetry

    29/43

  • 8/10/2019 EPR Spectrometer & Dosimetry

    30/43

    Surface coil resonator for in vivo use

  • 8/10/2019 EPR Spectrometer & Dosimetry

    31/43

    Surface coil resonator construction

  • 8/10/2019 EPR Spectrometer & Dosimetry

    32/43

    Surface coil resonator construction

    Home-made coaxial tube(Copper / Teflon)

    Home made surface coil (Silver)

    Sugawara et al, 2014,IEEE

  • 8/10/2019 EPR Spectrometer & Dosimetry

    33/43

    Data Acquisition Program

    In-house program made using Labview

    Show the signal of each scan (upper) andintegration of number of those signals (lower)

    SNR histogram (right)

  • 8/10/2019 EPR Spectrometer & Dosimetry

    34/43

    Dosimetery

    Synthetic Hydroxyapatite as dosimeter

    - Used as a dental graft / implant

    - Good substitute of human tooth enamel

    Used electronic scale for the normalization of hydroxyapatitesamples

    - Hydroxyapatite samples are irradiated different doses.

    Irradiated dose

    (cGy)

    Weight

    (g)0 0.254210 0.125530 0.123150 0.1292100 0.1338200 0.127300 0.1253600 0.0845

  • 8/10/2019 EPR Spectrometer & Dosimetry

    35/43

  • 8/10/2019 EPR Spectrometer & Dosimetry

    36/43

    The quantity of radiation-induced radical can be determined relativelyby the peak-to-peak intensity of the spectrum.

    External magnetic field / mTpeak-to-peak intensity

  • 8/10/2019 EPR Spectrometer & Dosimetry

    37/43

    -0.06

    -0.04

    -0.02

    0

    0.02

    0.04

    0.06

    0.08

    39 39.5 40 40.5 41 41.5 42 42.5 43

    The quantity of radiation-induced radical can be determined relativelyby the peak-to-peak intensity of the spectrum.

    External magnetic field / mTpeak-to-peak intensity

    Too noisy to determine the exactpeak-to-peak intensity

  • 8/10/2019 EPR Spectrometer & Dosimetry

    38/43

    (mT)

    -0.06

    -0.04

    -0.02

    0

    0.02

    0.04

    0.06

    0.08

    39 39.5 40 40.5 41 41.5 42 42.5 43

    The quantity of radiation-induced radical can be determined relativelyby the peak-to-peak intensity of the spectrum.

    External magnetic field / mTpeak-to-peak intensity

    Too noisy to determine the exactpeak-to-peak intensity

    Need of curve fitting

  • 8/10/2019 EPR Spectrometer & Dosimetry

    39/43

    Curve fitting

    Robinson et al., 1998, Journal of Magnetic Resonance

    -0.08

    -0.06

    -0.04

    -0.02

    0

    0.02

    0.04

    0.06

    39 39.5 40 40.5 41 41.5 42 42.5 43

    Including center frequency, linewidth, modulation frequency, Zeeman modulation

    Also resulted signal intensity determined with the correction usingnumber of scans, weight of the sample

    To determine the signal intensity

  • 8/10/2019 EPR Spectrometer & Dosimetry

    40/43

    Results

    Irradiated dose(cGy)

    signal intensity

    0 0.01981

    10 0.04742

    30 0.05276

    50 0.05455100 0.06130

    200 0.06347

    300 0.08646

    600 0.112680

    0.02

    0.04

    0.06

    0.08

    0.1

    0.12

    0 100 200 300 400 500 600 700

    (cGy)

    (a.u.)

    200 cGy sample didnt show good signal intensity.

    As expected, doses under 100 cGy did not show

  • 8/10/2019 EPR Spectrometer & Dosimetry

    41/43

    Discussion

    1. Noise make the spectroscopy difficult.

    Contamination of the surface coil is one reason.

    Sugawara et al.Improvement of the baseline stability by HCl cleaning of the resonator

  • 8/10/2019 EPR Spectrometer & Dosimetry

    42/43

    Discussion

    2. Too small and various size of hydroxyapatite samples

    -> Different amount of effect from the magnetic field of the surface coil

    3. Only one sample per each radiation dose

  • 8/10/2019 EPR Spectrometer & Dosimetry

    43/43