epr spectrometer & dosimetry
Post on 02-Jun-2018
234 views
TRANSCRIPT
-
8/10/2019 EPR Spectrometer & Dosimetry
1/43
EPR spectrometer & dosimetry
in Hokkaido University
()
2014.09.16Kwon Choi
-
8/10/2019 EPR Spectrometer & Dosimetry
2/43
Lets say about EPR
Which exist around nucleus
Unpaired electrons
Electron Paramagnetic Resonance
-
8/10/2019 EPR Spectrometer & Dosimetry
3/43
Lets say about EPR
Paramagnetismisa form of magnetism whereby certain materialsare attractedby an externally applied magnetic field,and form internal, induced magnetic fieldsin the direction ofthe applied magnetic field. () Paramagnetic properties aredue to the presence of some unpaired electrons, (.)
(from Wikipedia )
Electron Paramagnetic ResonanceWhich you may not be familiar with
-
8/10/2019 EPR Spectrometer & Dosimetry
4/43
Lets say about EPR
In other word, align their directioninto the direction of the extenalmagnetic field.
Electron Paramagnetic ResonanceWhich you may not be familiar with
-
8/10/2019 EPR Spectrometer & Dosimetry
5/43
Lets say about EPR
Electron Paramagnetic ResonanceWhich you may not be familiar with
Form induced magnetic field
Up spin and down spin electrons have different energy states.ThisEnergy Splittingis determined by the strength of theexternal magnetic field.
=geBB0
ge: g factor, 2.00231930436153 (a.u.)for electron
B: Bohr magneton, 9.274009681024 (a.u.)
-
8/10/2019 EPR Spectrometer & Dosimetry
6/43
Lets say about EPR
Electron Paramagnetic ResonanceResonance absorption
Absorption of electromagnetic energyat a certain frequencyin a static external magnetic fieldby a quantum excitation of the material
-
8/10/2019 EPR Spectrometer & Dosimetry
7/43
Lets say about EPR
Electron Paramagnetic ResonanceResonance absorption
Energy
Absorption of theelectromagneticenergy
E =
Release of theenergy (heat, light)
Excitation of electrons Relaxation
The electromagnetic energywith the energy equal to the energy splitting of electrons,= E = geBB0 makes the resonance
-
8/10/2019 EPR Spectrometer & Dosimetry
8/43
2. Population distribution of electrons between two energy levels :
(Boltzmanns equation)
T : temperaturek : Boltzmann constant
1. Under the resonance :
The population of the lower level is greaterthan thatof the upper level.
Net energy absorption
http://encyclopedia2.thefreedictionary.com/Electron+Paramagnetic+Resonance+EPR
Energyemission
Energyabsorption
Both transitions with
same probability
=#
# <
How energy absorption occur
E = -1
0
E =1
0
-
8/10/2019 EPR Spectrometer & Dosimetry
9/43
-
8/10/2019 EPR Spectrometer & Dosimetry
10/43
The frequency of EM wave is determined by the utility.
Frequency => Energy splitting (Better signal)Dielectric loss (Heat loss)
Skin depth (Lower penetration)
-
8/10/2019 EPR Spectrometer & Dosimetry
11/43
= ( 20
)1/
: Resistivity of material0: dielectric constant: Frequency
Skin depth,
How deep the electromagnetic wave can penetrate
The frequency of EM wave is determined by the use.
Frequency => Energy splitting (Better signal)Dielectric loss (Heat loss)
Skin depth (Lower penetration)
Dielectric lossIn water
-
8/10/2019 EPR Spectrometer & Dosimetry
12/43
= ( 20
)1/
: Resistivity of material0: dielectric constant: Frequency
Skin depth,
How deep the electromagnetic wave can penetrate
At 1 GHz
(L-band)
Resistivity
(m)
Skin depth
(cm)Muscle 0.8 1.5
Liver 1 1.7
Fat 7-12 4-6
Bone 10-25 5-8
Brain 1 1.7
The frequency of EM wave is determined by the use.
Frequency => Energy splitting (Better signal)Dielectric loss (Heat loss)
Skin depth (Lower penetration)
Dielectric lossIn water
-
8/10/2019 EPR Spectrometer & Dosimetry
13/43
= ( 20
)1/
: Resistivity of material0: dielectric constant: Frequency
Skin depth,
How deep the electromagnetic wave can penetrate
At 1 GHz
(L-band)
Resistivity
(m)
Skin depth
(cm)Muscle 0.8 1.5
Liver 1 1.7
Fat 7-12 4-6
Bone 10-25 5-8
Brain 1 1.7
The frequency of EM wave is determined by the use.
Frequency => Energy splitting (Better signal)Dielectric loss (Heat loss)
Skin depth (Lower penetration)
Dielectric lossIn water
Under L-band (1-2 GHz) isappropriate for the use in vivo.
-
8/10/2019 EPR Spectrometer & Dosimetry
14/43
=geBB0
External magnetic field / mT
Because of the magneticmodulation
-> Basic form of an EPR signal
For
Microwave
Variant frequency of EM wave isused by the use.
External magnetic field / mT
-
8/10/2019 EPR Spectrometer & Dosimetry
15/43
=geBB0
External magnetic field / mT
Because of the magneticmodulation
-> Basic form of an EPR signal
Microwave
Variant frequency of EM wave isused by the use.
-
8/10/2019 EPR Spectrometer & Dosimetry
16/43
What can we observe with EPR?
Flaw of the
optic fiber
New material
(Ion liquid)
Process ofchemical
reaction fromUV irradiation
Tooth enamel ofthe irradiatedhuman
In vivo brain
/ tumor
How did radicals form in observed objects?
Elapse of timeMix the stableradical into thenew material
Unstable radicalsare formed bythe break of
chemical bonds
Stable radicals areformed in the
tooth enamel
Quantity ofradicals,relationship withthe amount ofradiation
Quantity ofradicals,distribution,relationship withtumor
Inject stable radicalsinto the vessel orregion of interest
What information do you want?
Quantity,distribution,location andquantity of flaw,
Interactionbetween stableradical and newmaterial
Dynamics(Change ofstructure,quantity,characteristicswith the timeelapse
Enamel
-
8/10/2019 EPR Spectrometer & Dosimetry
17/43
What can we observe with EPR?
Flaw of the
optic fiber
New material
(Ion liquid)
Process ofchemical
reaction fromUV irradiation
Tooth enamel ofthe irradiatedhuman
In vivo brain
/ tumor
How did radicals form in observed objects?
Elapse of timeMix the stableradical into thenew material
Unstable radicalsare formed bythe break of
chemical bonds
Stable radicals areformed in the
tooth enamel
Quantity ofradicals,relationship withthe amount ofradiation
Quantity ofradicals,distribution,relationship withtumor
Inject stable radicalsinto the vessel orregion of interest
-
8/10/2019 EPR Spectrometer & Dosimetry
18/43
Radiation dosimetry using tooth enamel
Requirement
Possibility of radiologic emergency situation(terrorist attack using a nuclear weapon, accident of nuclear power plants)
Need of patients triage (distinguish who may be irradiated)- Who exposed to over certain amount of radiation need rapid care.- Victims with doses starting at about 2 Gy can benefit from urgent
medical treatment.
General publication does not have the proper dosimeter.
EPR tooth dosimetry can be used for triagein situ.
Conventional blood test can be done at remote place that takes time.
-
8/10/2019 EPR Spectrometer & Dosimetry
19/43
Radiation dosimetry using tooth enamel
Hydroxyapatite contained in teeth and bones-> Suitable probe for dose reconstruction.
- High content of hydroxyapatite- Completely formed in childhood- Once formed never remodelled.
1. Carbonate impurities in hydroxyapatite crystals
Radiationexposure
radicals
(EPR sensitive stable radical as probe)
2. Tooth enamel is preferred for the retrospective dosimetry
-> Co2--radicals are accumulated over 107years
-
8/10/2019 EPR Spectrometer & Dosimetry
20/43
"Boys, be ambitious!Be ambitious not for money or for
selfish aggrandizement, not for thatevanescent thing which men call fame. Beambitious for that attainment of all that aman ought to be."
Regrettably, today in Korea this goodmaxim is generally used asBoys, be ambitious for fame and money!
, !, ..
William S. Clark Ph.D.
-
8/10/2019 EPR Spectrometer & Dosimetry
21/43
Farm (Part of the school of agriculture)
-
8/10/2019 EPR Spectrometer & Dosimetry
22/43
Farm (Part of the school of agriculture)
-
8/10/2019 EPR Spectrometer & Dosimetry
23/43
-
8/10/2019 EPR Spectrometer & Dosimetry
24/43
-
8/10/2019 EPR Spectrometer & Dosimetry
25/43
A little history about thediscovery of
electron paramagnetic resonance
-
8/10/2019 EPR Spectrometer & Dosimetry
26/43
-
8/10/2019 EPR Spectrometer & Dosimetry
27/43
Phenomenon called ElectronParamagnetic Resonance (EPR)
Isior Issac Rabi
Nobel Prize in Physics (1944)
Discovery of Nuclear Magnetic Resonance (1938)
Zavoisky was impressed at Rabis discovery, NMRand tried to detect NMR in liquids and solids.
-
8/10/2019 EPR Spectrometer & Dosimetry
28/43
Tooth dosimetry in Hokkaido University
-
8/10/2019 EPR Spectrometer & Dosimetry
29/43
-
8/10/2019 EPR Spectrometer & Dosimetry
30/43
Surface coil resonator for in vivo use
-
8/10/2019 EPR Spectrometer & Dosimetry
31/43
Surface coil resonator construction
-
8/10/2019 EPR Spectrometer & Dosimetry
32/43
Surface coil resonator construction
Home-made coaxial tube(Copper / Teflon)
Home made surface coil (Silver)
Sugawara et al, 2014,IEEE
-
8/10/2019 EPR Spectrometer & Dosimetry
33/43
Data Acquisition Program
In-house program made using Labview
Show the signal of each scan (upper) andintegration of number of those signals (lower)
SNR histogram (right)
-
8/10/2019 EPR Spectrometer & Dosimetry
34/43
Dosimetery
Synthetic Hydroxyapatite as dosimeter
- Used as a dental graft / implant
- Good substitute of human tooth enamel
Used electronic scale for the normalization of hydroxyapatitesamples
- Hydroxyapatite samples are irradiated different doses.
Irradiated dose
(cGy)
Weight
(g)0 0.254210 0.125530 0.123150 0.1292100 0.1338200 0.127300 0.1253600 0.0845
-
8/10/2019 EPR Spectrometer & Dosimetry
35/43
-
8/10/2019 EPR Spectrometer & Dosimetry
36/43
The quantity of radiation-induced radical can be determined relativelyby the peak-to-peak intensity of the spectrum.
External magnetic field / mTpeak-to-peak intensity
-
8/10/2019 EPR Spectrometer & Dosimetry
37/43
-0.06
-0.04
-0.02
0
0.02
0.04
0.06
0.08
39 39.5 40 40.5 41 41.5 42 42.5 43
The quantity of radiation-induced radical can be determined relativelyby the peak-to-peak intensity of the spectrum.
External magnetic field / mTpeak-to-peak intensity
Too noisy to determine the exactpeak-to-peak intensity
-
8/10/2019 EPR Spectrometer & Dosimetry
38/43
(mT)
-0.06
-0.04
-0.02
0
0.02
0.04
0.06
0.08
39 39.5 40 40.5 41 41.5 42 42.5 43
The quantity of radiation-induced radical can be determined relativelyby the peak-to-peak intensity of the spectrum.
External magnetic field / mTpeak-to-peak intensity
Too noisy to determine the exactpeak-to-peak intensity
Need of curve fitting
-
8/10/2019 EPR Spectrometer & Dosimetry
39/43
Curve fitting
Robinson et al., 1998, Journal of Magnetic Resonance
-0.08
-0.06
-0.04
-0.02
0
0.02
0.04
0.06
39 39.5 40 40.5 41 41.5 42 42.5 43
Including center frequency, linewidth, modulation frequency, Zeeman modulation
Also resulted signal intensity determined with the correction usingnumber of scans, weight of the sample
To determine the signal intensity
-
8/10/2019 EPR Spectrometer & Dosimetry
40/43
Results
Irradiated dose(cGy)
signal intensity
0 0.01981
10 0.04742
30 0.05276
50 0.05455100 0.06130
200 0.06347
300 0.08646
600 0.112680
0.02
0.04
0.06
0.08
0.1
0.12
0 100 200 300 400 500 600 700
(cGy)
(a.u.)
200 cGy sample didnt show good signal intensity.
As expected, doses under 100 cGy did not show
-
8/10/2019 EPR Spectrometer & Dosimetry
41/43
Discussion
1. Noise make the spectroscopy difficult.
Contamination of the surface coil is one reason.
Sugawara et al.Improvement of the baseline stability by HCl cleaning of the resonator
-
8/10/2019 EPR Spectrometer & Dosimetry
42/43
Discussion
2. Too small and various size of hydroxyapatite samples
-> Different amount of effect from the magnetic field of the surface coil
3. Only one sample per each radiation dose
-
8/10/2019 EPR Spectrometer & Dosimetry
43/43