epicardial pacing in children 12 years

8
A 12-Year Experience of Bipolar Steroid-Eluting Epicardial Pacing Leads in Children Maren Tomaske, MD, Bart Gerritse, PhD, Leo Kretzers, MS, Rene Pretre, MD, Ali Dodge-Khatami, MD, PhD, Mariette Rahn, MD, and Urs Bauersfeld, MD Division of Pediatric Cardiology, University Children’s Hospital and Division of Congenital Cardiovascular Surgery, University Children’s Hospital, Zurich, Switzerland; and Medtronic Bakken Research Center, Maastricht, the Netherlands Background. Cardiovascular abnormalities and small vascular size may preclude transvenous pacing and ne- cessitate epicardial lead implantation. This study evalu- ates the performance of steroid-eluting, bipolar epicar- dial pacing leads. Methods. We prospectively enrolled 114 children with 239 atrial and ventricular bipolar epicardial leads (Medtronic CapSure 10366 or 4968, Minneapolis, MN), followed up to 12.2 years (median, 3.2). Lead data were obtained at implant and at semi-annual visits. Analysis was done for left or right atrial and ventricular leads. Results. Median atrial and ventricular pacing thresh- olds remained below 1.2 V at 0.5 ms. Thresholds did not differ between pacing sites: left atrial, 0.82V at 0.5 ms; right atrial, 0.74 V at 0.5 ms (p 0.85); and left ventricu- lar, 0.96V at 0.5 ms; right ventricular, 0.94 V at 0.5 ms (p 0.65). Sensing demonstrated no difference for atrial leads, at left atrial, 3.4 mV; and right atrial, 2.9 mV (p 0.12), but there was superiority of left over right ventricular leads (11.2 vs 7.7 mV, p 0.002). During follow-up, the 239 atrial and ventricular leads experienced 19 (8%) lead failures. Lead survival at 2 and 5 years was 99% and 94% for atrial leads and 96% and 85% for ventricular leads, respectively. Conclusions. Bipolar steroid-eluting epicardial leads demonstrate excellent sensing characteristics and persis- tent low median pacing thresholds below 1.2 V at 0.5 ms in children during up to 12 years follow-up. Considering growing and active patients with most having congenital heart disease, the lead survival of 85% to 94% at 5 years is favorable. Subanalysis shows superior sensing for left ventricular leads. Bipolar steroid-eluting leads provide an alternative approach for permanent pacing and may also be considered for left atrial and ventricular pacing, resynchronization, or defibrillator therapy. (Ann Thorac Surg 2008;85:1704 –11) © 2008 by The Society of Thoracic Surgeons A dvances in lead and device technology allow pace- maker system implantation in infants and even in neonates [1, 2]. Besides bradycardia pacing for sinus node disease or heart block, resynchronization therapy and implantation of cardioverter defibrillators (ICD) are progressively more often required in pediatric patients and adults with congenital heart disease (CHD) [3]. Specific problems in children or adults with complex CHD can complicate pacemaker therapy. Small vessel size, cardiovascular abnormalities, or the intention to preserve the venous access often preclude a transvenous approach and require epicardial pacing [4, 5]. Moreover, physical activity and somatic growth may affect lead longevity in young patients [5–7]. Studies in the pediatric population have indicated superior longevity of transvenous over unipolar screw-in epicardial leads, mainly due to high thresholds, exit blocks, and fractures of epicardial leads [8]. Initial expe- riences with bipolar steroid-eluting epicardial leads have shown low pacing thresholds up to 18 months of follow-up [9]. A main disadvantage of transvenous leads is attribut- able to right ventricular (RV) pacing leading to impaired left ventricular (LV) systolic function over time [10]. The advantage of epicardial leads facilitating LV pacing and sensing has promoted further epicardial lead implanta- tion for single-site LV pacing, resynchronization therapy, ICD therapy, or patients with a diseased RV [3]. The aim of this study was to evaluate pacing and sensing charac- teristics, and survival of bipolar steroid-eluting epicardial pacing leads in pediatric patients. Patients and Methods Study Patients and Leads All children who received permanent pacing systems consisting of bipolar steroid-eluting epicardial leads (Medtronic CapSure Epi 10366 or 4968, Medtronic, Inc, Accepted for publication Feb 6, 2008. Address correspondence to Dr Tomaske, Division of Pediatric Cardiology, University Children’s Hospital, Steinwiesstrasse 75, Zurich, 8032, Swit- zerland; e-mail: [email protected]. Dr Bauersfeld, Dr Gerritse, and Mr Kretzers disclose that they have a financial relationship with Medtronic Inc. © 2008 by The Society of Thoracic Surgeons 0003-4975/08/$34.00 Published by Elsevier Inc doi:10.1016/j.athoracsur.2008.02.016 PEDIATRIC CARDIAC

Upload: ali-dodge-khatami-md-phd

Post on 12-Jan-2017

4 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: epicardial pacing in children 12 years

AEMA

DC

vcad

2(fow

odrl0ab

Annaap(

Cspapl

sebr

A

AUz

©P

PEDIA

TR

ICC

AR

DIA

C

12-Year Experience of Bipolar Steroid-Elutingpicardial Pacing Leads in Childrenaren Tomaske, MD, Bart Gerritse, PhD, Leo Kretzers, MS, Rene Pretre, MD,li Dodge-Khatami, MD, PhD, Mariette Rahn, MD, and Urs Bauersfeld, MD

ivision of Pediatric Cardiology, University Children’s Hospital and Division of Congenital Cardiovascular Surgery, University

hildren’s Hospital, Zurich, Switzerland; and Medtronic Bakken Research Center, Maastricht, the Netherlands

l2ffr

dtighivaar

Background. Cardiovascular abnormalities and smallascular size may preclude transvenous pacing and ne-essitate epicardial lead implantation. This study evalu-tes the performance of steroid-eluting, bipolar epicar-ial pacing leads.Methods. We prospectively enrolled 114 children with

39 atrial and ventricular bipolar epicardial leadsMedtronic CapSure 10366 or 4968, Minneapolis, MN),ollowed up to 12.2 years (median, 3.2). Lead data werebtained at implant and at semi-annual visits. Analysisas done for left or right atrial and ventricular leads.Results. Median atrial and ventricular pacing thresh-

lds remained below 1.2 V at 0.5 ms. Thresholds did notiffer between pacing sites: left atrial, 0.82V at 0.5 ms;ight atrial, 0.74 V at 0.5 ms (p � 0.85); and left ventricu-ar, 0.96V at 0.5 ms; right ventricular, 0.94 V at 0.5 ms (p �.65). Sensing demonstrated no difference for atrial leads,t left atrial, 3.4 mV; and right atrial, 2.9 mV (p � 0.12),

ut there was superiority of left over right ventricular

hf

alastIotp

P

SAc(

niversity Children’s Hospital, Steinwiesstrasse 75, Zurich, 8032, Swit-erland; e-mail: [email protected].

2008 by The Society of Thoracic Surgeonsublished by Elsevier Inc

eads (11.2 vs 7.7 mV, p � 0.002). During follow-up, the39 atrial and ventricular leads experienced 19 (8%) leadailures. Lead survival at 2 and 5 years was 99% and 94%or atrial leads and 96% and 85% for ventricular leads,espectively.

Conclusions. Bipolar steroid-eluting epicardial leadsemonstrate excellent sensing characteristics and persis-

ent low median pacing thresholds below 1.2 V at 0.5 msn children during up to 12 years follow-up. Consideringrowing and active patients with most having congenitaleart disease, the lead survival of 85% to 94% at 5 years

s favorable. Subanalysis shows superior sensing for leftentricular leads. Bipolar steroid-eluting leads providen alternative approach for permanent pacing and maylso be considered for left atrial and ventricular pacing,esynchronization, or defibrillator therapy.

(Ann Thorac Surg 2008;85:1704–11)

© 2008 by The Society of Thoracic Surgeons

dvances in lead and device technology allow pace-maker system implantation in infants and even in

eonates [1, 2]. Besides bradycardia pacing for sinusode disease or heart block, resynchronization therapynd implantation of cardioverter defibrillators (ICD)re progressively more often required in pediatricatients and adults with congenital heart disease

CHD) [3].Specific problems in children or adults with complex

HD can complicate pacemaker therapy. Small vesselize, cardiovascular abnormalities, or the intention toreserve the venous access often preclude a transvenouspproach and require epicardial pacing [4, 5]. Moreover,hysical activity and somatic growth may affect lead

ongevity in young patients [5–7].Studies in the pediatric population have indicated

uperior longevity of transvenous over unipolar screw-inpicardial leads, mainly due to high thresholds, exitlocks, and fractures of epicardial leads [8]. Initial expe-iences with bipolar steroid-eluting epicardial leads

ccepted for publication Feb 6, 2008.

ddress correspondence to Dr Tomaske, Division of Pediatric Cardiology,

ave shown low pacing thresholds up to 18 months ofollow-up [9].

A main disadvantage of transvenous leads is attribut-ble to right ventricular (RV) pacing leading to impairedeft ventricular (LV) systolic function over time [10]. Thedvantage of epicardial leads facilitating LV pacing andensing has promoted further epicardial lead implanta-ion for single-site LV pacing, resynchronization therapy,CD therapy, or patients with a diseased RV [3]. The aimf this study was to evaluate pacing and sensing charac-eristics, and survival of bipolar steroid-eluting epicardialacing leads in pediatric patients.

atients and Methods

tudy Patients and Leadsll children who received permanent pacing systems

onsisting of bipolar steroid-eluting epicardial leadsMedtronic CapSure Epi 10366 or 4968, Medtronic, Inc,

Dr Bauersfeld, Dr Gerritse, and Mr Kretzers disclosethat they have a financial relationship with Medtronic

Inc.

0003-4975/08/$34.00doi:10.1016/j.athoracsur.2008.02.016

Page 2: epicardial pacing in children 12 years

Mtc1lbesasodl

pacptc

TE

C

DPS

AWMC

PI

S

Ai

ble

2.M

easu

red

Pace

mak

erTe

lem

etry

Dat

aat

Dis

char

gean

dat

Follo

w-U

p

dD

isch

arge

a3

Mon

thsa

1Y

eara

3Y

ears

a5

Yea

rsa

8Y

ears

a10

Yea

rsa

ial

lead

sea

ds,

No.

107

102

7447

2515

7-w

ave,

mV

3.0

(1.8

–4.0

)3.

5(2

.0–5

.0)

3.2

(2.0

–4.0

)3.

5(2

.6–4

.0)

3.4

(2.3

–4.0

)3.

0(2

.0–3

.8)

2.5

(2.2

–4.3

)m

ped

ance

,oh

m64

2(5

87–7

27)

604

(568

–668

)61

7(5

72–6

83)

652

(588

–729

)69

6(6

46–7

67)

645

(600

–723

)72

7(6

37–7

28)

hre

shol

d,V

at0.

5m

sb0.

71(0

.5–1

.5)

0.67

(0.4

–1.3

)0.

67(0

.4–1

.2)

0.89

(0.7

–1.3

)0.

89(0

.7–1

.3)

0.97

(0.8

–1.1

)0.

89(0

.9–2

.3)

hre

shol

d,�

Jat

0.5

msb

0.39

(0.2

–1.8

)0.

36(0

.2–1

.2)

0.35

(0.2

–1.4

)0.

59(0

.3–1

.1)

0.68

(0.3

–1.0

)0.

83(0

.5–1

.0)

0.55

(0.5

–3.7

)n

tric

ula

rle

ads

ead

s,N

o.13

212

710

063

3622

10-w

ave,

mV

10.0

(7.0

–12.

5)10

.0(7

.0–1

2.5)

10.0

(6.0

–12.

5)10

.6(6

.5–1

2.5)

11.2

(5.0

–12.

8)9.

3(5

.0–1

2.5)

8.0

(5.8

–10.

0)m

ped

ance

,oh

m78

6(6

76–8

98)

734

(633

–820

)73

0(6

45–8

28)

752

(637

–846

)75

2(6

62–8

47)

725

(672

–856

)74

1(6

38–8

39)

hre

shol

d,V

at0.

5m

s,b

0.89

(0.7

–1.7

)0.

89(0

.7–1

.4)

0.89

(0.7

–1.3

)1.

11(0

.7–1

.7)

1.18

(0.7

–1.5

)0.

84(0

.7–1

.3)

0.92

(0.7

–1.1

)h

resh

old

,�J

at0.

5m

s),b

0.58

(0.0

–1.9

)0.

50(0

.3–1

.4)

0.59

(0.3

–1.4

)0.

86(0

.3–1

.9)

0.87

(0.4

–2.1

)0.

50(0

.3–0

.8)

0.61

(0.3

–1.0

)

nle

ssot

her

wis

ein

dic

ated

,dat

aar

egi

ven

asm

edia

ns

(in

terq

uar

tile

ran

ge).

acin

gvo

ltag

ean

den

ergy

thre

shol

ds

calc

ula

ted

for

0.5-

ms

pu

lse

du

rati

on.

1705Ann Thorac Surg TOMASKE ET AL2008;85:1704–11 BIPOLAR EPICARDIAL LEADS IN CHILDREN

PED

IAT

RIC

CA

RD

IAC

inneapolis, MN) between 1994 and 2006 were prospec-ively enrolled into the study. The study populationomprised 114 patients with a maximum follow-up of2.2 years. A total of 107 atrial and 132 ventricular pacingeads were implanted, including lead replacements oriventricular pacing in some patients. The choice forpicardial pacing systems was determined by patientize, cardiovascular abnormality, indication for pacing,nd patient preference. Leads were connected to variousingle- and dual-chamber devices, biventricular devices,r ICDs. The study excluded 5 children: 3 died in-hospitalue to a non-pacing-related cause and 2 children with

eads not connected to a device.The bipolar steroid-eluting epicardial leads have

latinized, porous electrode surfaces with a surfacerea of 14 mm2 (anode) and 6 mm2 (cathode). Theyontain 1.0 mg or less of dexamethasone sodium phos-hate. This study protocol was performed with institu-

ional Ethical Committee approval and written informed

able 1. Baseline Characteristics of 114 Children at Firstpicardial Lead Implant

haracteristicsNo (%) or Median

(range)

emographicsatients, No. 114exMale 70Female 44ge, years 6.2 (0.0–18.5)eight, kg 18.0 (1.0–88.0)edian follow-up, years 3.2 (0.1–12.2)ardiac anatomyStructurally normal heart 36 (32)AVSD 10 (9)VSD 14 (12)D-transposition great arteries and VSD 7 (6)L-transposition of great arteries 4 (4)Complex single ventricle 25 (22)Other 18 (16)

rior/concomitant cardiac operation 74 (65)ndication for pacing

Post-op AV block 39 (34)Congenital AV block 22 (19)Acquired AV block 9 (8)Post-op sinus node disease 6 (5)Congenital sinus node disease 24 (21)Resynchronization therapy 6 (5)ICD therapy 8 (7)

urgical access, n(%)Left lateral thoracotomy 65 (57)Right lateral thoracotomy 1 (1)Midline sternotomy 37 (32)Subxiphoid 11 (10)

V � atrioventricular; AVSD � atrioventricular septal defect; ICD �mplantable cardioverter device; VSD � ventricular septal defect.

onsent.

Ta Lea

Atr L P I T T

Ve L R I T T

aU

bP

Page 3: epicardial pacing in children 12 years

IFFwtvwievtppta

1

2

iel

ao

SAiodtpo

nou

dwtt

SAD

Fost0n0tb

TS

PAARVV

a

1706 TOMASKE ET AL Ann Thorac SurgBIPOLAR EPICARDIAL LEADS IN CHILDREN 2008;85:1704–11

PEDIA

TR

ICC

AR

DIA

C

mplant Data, Pacemaker Telemetry Data, and Leadailuresor all leads, impedances, P- or R-wave amplitudes, asell as pacing voltage and energy thresholds were ob-

ained at implant, at discharge, and regular follow-upisits. The first follow-up after discharge was within 4 to 6eeks. Further follow-up was scheduled 3 months after

mplant and every 6 months thereafter. Pacing voltage andnergy of lead thresholds were calculated for a standardalue of 0.5-ms pulse duration (V at 0.5 ms and �J at 0.5 ms)o allow for comparison by using the energy formulaublished previously [11]. In those patients in whom im-edance measurements were unipolar, the difference be-

ween unipolar and bipolar measurements was noted oncend added to the unipolar measurements during follow-up.

Lead events were grouped in two categories:

. lead failure, which was defined as lead fracture,insulation defect, replacement due to unacceptablyhigh thresholds or sensing abnormalities, loss of cap-ture, lead dislodgement, or primary infection; and

. secondary lead replacement, which was defined asevents of secondary infection if leads were exposedto pocket infection, accidental lead damage at car-diopulmonary bypass surgery, or elective lead re-placement at the time of device exchange.

Lead lifetime was defined as time elapsed from leadmplant to the event of lead failure. From the date ofxchange, new leads were enrolled as new implants withead measurements starting at implant.

Performances of left (LA) and right atrial (RA) and LVnd RV pacing leads, as well as the effect of prior cardiacperation, were also analyzed and compared.

urgical Techniqueccess for lead implantation was either by a subxiphoid

ncision to reach the RV apex or by a left lateral thoracot-my to reach the LV free wall and corresponding atria, asescribed previously [12]. In those children with concomi-

ant cardiac procedures, the epicardial leads were im-lanted through a midline sternotomy at the time of cardiacperation, with preference of an LV implant site.Standard surgical implant techniques were used with

onabsorbable sutures. For the distal and proximal suturef the triangular electrode, a single-knot technique was

able 3. Spearman Correlation for Individual Regressionlopes of Measured Telemetry Data and Lead Age

Pairs, No. Correlation (�) p Value

-wave 46 �0.049 0.75a

trial threshold 49 �0.001 0.99a

trial impedance 49 �0.141 0.34a

-wave 63 �0.088 0.49a

entricular threshold 70 �0.184 0.1a

entricular impedance 70 �0.096 0.43a

Not significant.

sed. Sutures were placed perpendicularly to the epicar- T

ium to avoid tissue trauma near the electrode. The deviceas implanted in the abdominal rectus sheath in 46 pa-

ients, in a left thoracic muscular pocket in 66, or subpec-orally in 2.

tatistical Analysisfollow-up period of 8 years was statistically analyzed.

ata are presented as median and interquartile (IQR)

ig 1. Subanalysis of 25 ventricular pacing leads with a lead agelder than 5 years. Box and whisker plots of individual regressionlopes demonstrated a slight incline of ventricular pacing voltagehresholds (A) for measurements beyond 5 years (median, �0.017 vs.052, p � 0.131). Measurements of ventricular impedances (B) sig-ificantly declined beyond 5 years (median, 5.53 vs �11.20, p �.003). The asterisk (*) denotes significance. The horizontal line inhe middle of each box indicates the median; the top and bottomorders of the box mark the 75th and 25th interquartile, respectively.

he whiskers mark is 97.5th and 2.5th percentile.
Page 4: epicardial pacing in children 12 years

rs

9lutuidefccm

Sadfdts(

R

PDao

1707Ann Thorac Surg TOMASKE ET AL2008;85:1704–11 BIPOLAR EPICARDIAL LEADS IN CHILDREN

PED

IAT

RIC

CA

RD

IAC

ange. A value of p � 0.05 was considered statisticallyignificant.

Survival is reported with Kaplan-Meier estimates and5% confidence intervals (CI) and compared with aog-rank test. Electrical measurements were comparedsing linear regression models, with parameter estima-

ion, corrected for multiple observations per patient,sing generalized estimating equations. To determine

ndividual change, regression slope coefficients over in-ividual repeated measurements were calculated forach patient’s course. Changes per year were calculatedor each electrical parameter. To focus on the long-termhanges, data collection for calculating regression slopeoefficients was started 6 months after implant, with ainimum of 5 contributing measurements.

Correlations between variables were measured by thepearman correlation. Mann-Whitney U tests were used fornalyzing differences in continuous variables between in-ependent groups. The Wilcoxon signed rank test was used

or within-group changes of continuous variables betweenifferent time periods. The �2 test provided a comparison of

he pacing site as well as the effect of cardiac operation. Alltatistical analyses were performed using SAS 9.1 softwareSAS Institute Inc, Cary, NC).

esults

atient Dataemographic and surgical data as well as clinical char-

cteristics at implantation are reported in Table 1. In 65%f the children, pacemaker system implantation was

Fig 2. (A) Atrial and (B) ventricular pacingvoltage thresholds at 0.5-ms pulse durationover time. No differences were found for me-dian pacing voltages for (A) the right atrial(RA, diamonds) or left atrial (LA, squares)leads (LA, 0.82 V at 0.5 ms vs RA, 0.74 V at0.5 ms; p � 0.85, or for the (B) right ventric-ular (RV, diamonds) or left ventricular (LV,squares) leads (LV, 0.96 V at 0.5 ms vs RV,0.94 V at 0.5 ms; p � 0.65).

Page 5: epicardial pacing in children 12 years

ra2o51t

LFmm1yitlMt0(i[ma

dltbdcss

a

d(PHs1

r0mwaopa

LDrlp

fatCtar

C

LR

T

O

FL

A

V

S

1708 TOMASKE ET AL Ann Thorac SurgBIPOLAR EPICARDIAL LEADS IN CHILDREN 2008;85:1704–11

PEDIA

TR

ICC

AR

DIA

C

elated to CHD with prior or concomitant cardiac oper-tion. Six children were lost to follow-up at a median of.5 years. No deaths related to the pacemaker systemccurred. Kaplan-Meier estimates of patient survival at 2,, and 10 years after enrolment were 98% (95% CI, 95% to00%), 95% (95% CI, 90% to 100%), and 95% (95% CI, 83%o 100%), respectively.

ead Follow-Up and Performanceollow-up consisted of 1251 visits contributing to atrialeasurements and 1677 to ventricular lead measure-ents. Median follow-up time was 2.6 years (range, 0.1 to

1.7 years) for atrial leads and 2.8 years (range, 0.1 to 12.2ears) for ventricular leads. Lead characteristics are givenn detail up to 10 years (Table 2). Median pacing voltagehresholds for atrial and ventricular leads remained be-ow 1.2 V at 0.5 ms during the entire observation period.

edian individual regression slopes indicated stablerends during follow-up for sensing (atrial, 0.000 [p �.88]; ventricular, 0.000 [p � 0.22]), pacing thresholdsatrial, 0.000 [p � 0.98], ventricular, 0.022 [p � 0.98]), andmpedances (atrial, 8.135 [p � 0.39]; ventricular, 0.625p � 0.13]). Neither of the individual regression slopes for

easured electrical parameters correlated with lead aget study closure (Table 3).A separate analysis of individual regression slopes was

one for 13 atrial leads and 25 ventricular leads with aead age older than 5 years (median, 9.4 years; range, 6.1o 12.2 years). Comparison of individual regression slopesefore and after a lead age of 5 years revealed no significantifferences for atrial leads. For ventricular leads, a signifi-ant decline of impedances as well as a slight but notignificant incline of pacing voltage thresholds was ob-erved for measurements beyond 5 years (Fig 1).

Analysis of lead performances between right or left

able 4. Lead Failures and Secondary Lead Replacements Du

utcome Tota

ollow-up, median (range) yearsead positionAtrial leads 107Ventricular leads 132trial lead failure, No. (%) 5 (5)High thresholdLead fractureInsulation defect

entricular lead failure, No. (%) 14 (11High thresholdLoss of captureLead fractureInsulation defectLead dislodgement

econdary lead replacement, No. (%) 5 (4)Secondary infectionAccidental lead damageRoutine lead replacement

trial and ventricular implant sites demonstrated no s

ifference between pacing voltage thresholds over timeFig 2A, B). Moreover, no difference was seen in median-wave amplitudes (LA, 3.4 mV vs RA, 2.9 mV; p � 0.12).owever, a significantly superior course of ventricular

ensing was seen for LV compared with RV leads (LV,1.2 mV vs RV, 7.7 mV; p � 0.002).

Analysis of the effect of prior or no cardiac operationevealed no difference for atrial (0.87 vs 0.67 V at 0.5 ms; p �.56) or ventricular pacing thresholds (1.01 vs 0.83 V at 0.5s; p � 0.21). No difference of prior or no cardiac operationas observed for P-wave (3.0 vs 3.3 mV; p � 0.31) or R-wave

mplitudes (10.7 vs 9.0 mV; p � 0.57), respectively. More-ver, no significant differences for lead impedances oracing energy thresholds were observed for the pacing sitesnd children with prior or no cardiac operation.

ead Failuresuring follow-up, 239 atrial and ventricular leads expe-

ienced 19 lead failures (8%). Characteristics of observedead failures and secondary lead replacements are re-orted in Table 4.At 2 and 5 years after implantation, the estimated

reedom from lead failure was 99% (95% CI, 97% to 100%)nd 94% (95% CI, 85% to 100%) for atrial leads, respec-ively; and was 96% (95% CI, 92% to 100%) and 85% (95%I, 76% to 94%), respectively, for ventricular leads. Fur-

hermore, estimated lead survival was compared by sites,nd no statistical significance was found between left andight implant sites (Fig 3A, B).

omment

ead Performance and Survivalecent advances in pacing lead technology, such as thin

Follow-up, Categorized for Lead Position

Right-sided Left-sided

3.5 (0.3–12.2) 2.2 (0.1–11.0)

37 7047 85

3 (8) 2 (3)— 12 11 —

8 (17) 6 (7)2 31 —4 2

— 11 —

3 (4) 2 (2)1 11 11 —

ring

l

)

teroid-eluting transvenous leads, facilitate implantation

Page 6: epicardial pacing in children 12 years

eti

cspdp[hosh

Hie

tevasdl7[

1709Ann Thorac Surg TOMASKE ET AL2008;85:1704–11 BIPOLAR EPICARDIAL LEADS IN CHILDREN

PED

IAT

RIC

CA

RD

IAC

ven in neonates and infants, thus maintaining the con-roversy about the use of transvenous or epicardial pac-ng leads [13].

The main concerns for transvenous RV pacing leads inhildren are venous obstructions [13], lead failures due toomatic growth [5], lead infections [14], and the risk ofaradoxic embolism in the presence of residual intracar-iac defects [15]. Major drawbacks of older epicardialacing leads have been the occurrence of lead fractures

16], increasing pacing thresholds over time [8, 16], andigh pacing thresholds in children with prior cardiacperation [17]. The invention of steroid-eluting electrodeurfaces combined with suture-on electrode tip designs

as improved the epicardial lead performance [18, 19]. f

owever, the potential advantage of the epicardial pac-ng system has yet to be demonstrated by prolongedxperiences.The present study enrolled 114 children after implan-

ation of 107 atrial and 132 ventricular bipolar steroid-luting epicardial pacing leads. Stable chronic atrial andentricular sensing as well as pacing thresholds werechieved over a maximum follow-up of 12.2 years. Per-istent low thresholds of steroid-eluting leads have beenemonstrated for up to 6 years for unipolar epicardial

eads [19, 20] and up to 7 years for transvenous leads [5,]. Similar to previous results of unipolar epicardial leads19, 20], low median pacing thresholds of 1.2 V at 0.5 ms

Fig 3. Kaplan-Meier estimates for lead sur-vival of (A) 107 atrial and (B) 132 ventricularleads. (A) No difference was found betweenatrial pacing sites (p � 0.37). At 2 and 5years, lead survival for right atrial (RA,dashed line) leads was 97% (95% confidence[CI], 90% to 100%) and 91% (95% CI, 75%to 100%), respectively; for left atrial (LA,solid line) leads it was 100% (95% CI, 100%to 100%) and 94% (95% CI, 83% to 100%),respectively. (B) No difference was found be-tween ventricular pacing sites (p � 0.92). At 2and 5 years, lead survival for right ventricle(RV, dotted line) leads was 100% (95% CI,100% to 100%) and 84% (95% CI, 66% to99%), respectively; for left ventricle (LV, solidline) leads it was 93% (95% CI, 85% to100%) and 86% (95% CI, 75% to 97%),respectively.

or atrial as well as ventricular leads were observed

Page 7: epicardial pacing in children 12 years

dtob[ces

wcrtdc5atlodtm

potlculalurmElr

SSaeqbif[ewrdcoavNud

sotoeoc

LCdisL[rmsLwipp

PTLfvcetalsamestaewlca

SAddIcnst

1710 TOMASKE ET AL Ann Thorac SurgBIPOLAR EPICARDIAL LEADS IN CHILDREN 2008;85:1704–11

PEDIA

TR

ICC

AR

DIA

C

uring follow-up in our study cohort, independently ofhe implant site. Recently, long-term pacing thresholdsf transvenous leads in children have been reported toe slightly superior, ranging from 0.6 to 0.9 V at 0.5 ms

7, 8]. However, the number of children with priorardiac operation and thereby higher risk for a dis-ased myocardium was smaller compared with ourtudy cohort.

Epicardial lead failures occurred in 8% of the leads,ith lead fractures and high thresholds being the most

ommon complication. These findings are in line withesults of recent studies reporting predominantly su-ured-on epicardial leads [7, 19]. Comparably, the inci-ence of lead failures of modern transvenous leads inhildren is reported to be about 7% [7, 8]. The estimated-year lead survival seen in our cohort was 94% for atrialnd 85% for ventricular pacing leads, independently ofhe implant site. These encouraging results of improvedead longevity confirm previous published data reportingf an estimated 5-year survival of 71% to 85% for epicar-ial leads [6, 7, 18]. Moreover, they are comparable with

he survival of transvenous leads, with a reported esti-ated 5-year survival rate of 84% to 89% [6, 7].Our favorable results of consistently low sensing and

acing thresholds, and improved lead longevity, werebserved in a study designed with uniform use of a singleechnology and single manufacturer of epicardial pacingeads, followed up in a relatively large population of 114hildren. Studies in children with permanent pacingsually have a small sample size. Previous reports of

ong-term experiences with epicardial lead thresholdsnd performance in large pediatric cohorts bear theimitation of showing performances of a variety of man-facturers, models, and technologies [5–8, 16]. The cur-ent study indicates improved long-term lead perfor-

ance of bipolar steroid-eluting epicardial leads.specially when connected to threshold tracking devices,

ow pacing thresholds will allow low energy pacing,esulting in a marked battery service life extension [21].

teroid Elutionteroid elution helps to diminish the initial inflammationfter lead implant; however, a secondary intention is itsffect on reducing stimulation thresholds over time. Theuestion of how long steroid-eluting electrodes are capa-le of maintaining their steroid-eluting effect to prevent

nflammation and fibrosis at the electrode-tissue inter-ace is not satisfactorily answered. An experimental study22] analyzed the remaining steroid in 25 transvenousxplanted leads and demonstrated that 20% of the steroidas still present at 10 years. Remarkably, individual

egression slopes indicated stable trends for telemetryata during the whole observation period and did notorrelate with lead age. A separate analysis of those leadslder than 5 years revealed stable lead performances fortrial leads and a slight incline for ventricular pacingoltages for measurements beyond a lead age of 5 years.evertheless, there was a significant decline of ventric-lar impedances measured beyond 5 years. A fast con-

ucting interface between the electrode and epicardial r

urface due to a decrease of the steroid-eluting potencyf the electrode, such as local edema or a deterioration ofhe lead insulation, could be a possible cause for thebserved lower ventricular impedances over time. How-ver, the restricted number of patients with a lead agelder than 5 years limits the statistical validity of aonclusive statement.

ead Implant Sitelinical trials with permanent RV apex pacing haveemonstrated asynchronous ventricular activation, lead-

ng to LV dysfunction over time [10]. Alternate pacingites have been intensively studied, indicating a superiorV systolic function in children paced from the LV apex

23, 24]. The potential prevention of myocardial deterio-ation has led to a favored LV approach for lead place-ent in our study cohort. An important finding was a

uperior performance of ventricular sensing was seen forV leads in our study cohort, whereas atrial sensing asell as atrial and ventricular pacing thresholds were

ndependent of the implant site. Moreover, sensing andacing characteristics of the epicardial leads were inde-endent of prior cardiac operation.

otential Clinical Implicationshe LV systolic function is less adversely affected fromV pacing. Because epicardial pacing of the LV seems

easible and safe, it could be the preferred approach forentricular pacing lead insertions. Furthermore, resyn-hronization therapy could be achieved by means ofpicardial LV instead of transvenous pacing by way ofhe coronary sinus [25] in grown-ups with CHD. Inddition to the favored LV position of the ventricularead, a LA lead may be preferable after extensive RAurgery such as after Mustard, Fontan, or Ebstein anom-ly surgery. Even though epicardial pacing lead place-ent inevitably involves a thoracotomy, LV pacing can

asily be achieved by a minithoracotomy. A muscle-paring minileft axillary approach is used in our insti-ution, as published recently [12]. This technique wasssociated with a low complication rate and providesxcellent cosmetic and functional results. Moreover,ith the superior course for LV sensing, epicardial

eads can safely be used in patients with ICD systemsonnected to epicardial sensing and pacing leads, andsubpleural defibrillation electrode [26].

tudy Limitationsmain limitation of this study is the use of different

evices. In the clinical setting, we observed slightlyifferent threshold measurements between the devices.n addition, the variable of R-wave measurementshanged over time. The new generations of devices doot offer the possibility to accurately determine R-waveensing above 12.5 mV. This circumstance could, poten-ially, even have led to an underestimation of the supe-

iority of R-wave sensing of LV leads.
Page 8: epicardial pacing in children 12 years

R

1

1

1

1

1

1

1

1

1

1

2

2

2

2

2

2

2

1711Ann Thorac Surg TOMASKE ET AL2008;85:1704–11 BIPOLAR EPICARDIAL LEADS IN CHILDREN

PED

IAT

RIC

CA

RD

IAC

eferences1. Villain E, Martelli H, Bonnet D, Iserin L, Butera G, Kachaner

J. Characteristics and results of epicardial pacing in neonatesand infants. Pacing Clin Electrophysiol 2000;23:2052–6.

2. Aellig NC, Balmer C, Dodge-Khatami A, Rahn M, Prêtre R,Bauersfeld U. Long-term follow-up after pacemaker implan-tation in neonates and infants. Ann Thorac Surg 2007;83:1420–3.

3. Walsh EP, Cecchin F. Recent advances in pacemaker andimplantable defibrillator therapy for young patients. CurrOpin Cardiol 2004;19:91–6.

4. Dodge-Khatami A, Johnsrude CL, Backer CL, Deal BJ, Stras-berger J, Mavroudis C. A comparison of steroid-elutingepicardial versus transvenous pacing leads in children.J Card Surg 2000;15:323–9.

5. Silvetti MS, Drago F, Grutter G, De Santis A, Di Ciommo V,Rava L. Twenty years of paediatric cardiac pacing: 515pacemakers and 480 leads implanted in 292 patients. Eu-ropace 2006;8:530–6.

6. Fortescue EB, Berul CI, Cecchin F, Walsh EP, Triedman JK,Alexander ME. Patient, procedural, and hardware factorsassociated with pacemaker lead failure in paediatrics andcongenital heart disease. Heart Rhythm 2004;1:150–9.

7. Fortescue EB, Berul CI, Cecchin F, Walsh EP, Triedman JK,Alexander ME. Comparison of modern steroid-eluting epi-cardial and thin transvenous pacemaker leads in paediatricsand congenital heart disease patients. J Intervent CardiacElectrophysiol 2005;14:27–36.

8. Udink ten Cate F, Breur J, Boramanand N, et al. Endocardialand epicardial steroid lead pacing in the neonatal andpaediatric age group. Heart 2002;88:392–6.

9. Bauersfeld U, Nowak B, Molinari L, et al. Low energyepicardial pacing in children: the benefit of Autocapture.Ann Thorac Surg 1999;68:1380–3.

0. Tantengco MV, Thomas RL, Karpawich PP. Left ventriculardysfunction after long-term right ventricular apical pacing inthe young. J Am Coll Cardiol 2001;37:2093–100.

1. Hamilton RM, Chiu C, Gow RM, Williams WG. A compar-ison of two stab-on unipolar epicardial leads in children.Pacing Clin Electrophysiol 1997;20:631–6.

2. Dodge-Khatami A, Kadner A, Dave H, Rahn M, Prêtre R,Bauersfeld U. Left heart atrial and ventricular pacingthrough a left lateral thoracotomy in children: a safe ap-proach with excellent functional and cosmetic results. EurJ Cardio Thoracic Surg 2005;28:541–5.

3. Ayabakan C, Rosenthal E. Endocardial pacemaker implan-tation in neonates and infants. Indian Pacing Electrophysiol

J 2006;6:57–62.

4. Klug D, Vaksmann G, Jarwe M, et al. Pacemaker leadinfection in young patients. Pacing Clin Electrophysiol 2003;26:1489–93.

5. Khairy P, Landzberg MJ, Gatzoulis MA, et al. Epicardialversus endocardial pacing and thromboembolic events in-vestigators: Transvenous pacing leads and systemic throm-boemboli in patients with intracardiac shunts: a multicenterstudy. Circulation 2006;113:2391–7.

6. Cohen MI, Bush DM, Vetter VL, et al. Permanent epicardialpacing in paediatric patients: Seventeen Years of experienceand 1200 outpatient visits. Circulation 2001;29:2585–90.

7. Karpawich PP, Walter H, Hakimi M. Chronic performance ofa transvenous steroid pacing lead used as epi-myocardialelectrode. Pacing Clin Electrophysiol 1998;21:1486–8.

8. Ector B, Willems R, Heidbüchel H, et al. Epicardial pacing: asingle-centre study on 321 leads in 138 patients. Acta Cardiol2006;61:343–51.

9. Cutler NG, Karpawich PP, Cavitt D, Hakimi M, Walters HL.Steroid-eluting epicardial pacing electrodes: six-year expe-rience of pacing thresholds in a growing paediatric popula-tion. Pacing Clin Electrophysiol 1997;20:2943–8.

0. Horenstein MS, Hakimi M, Walters HL, Karpawich PP.Chronic performance of steroid-eluting epicardial leads in agrowing pediatric population: a 10-year comparison. PacingClin Electrophysiol 2003;26:1467–71.

1. Tomaske M, Harpes P, Pretre R, Dodge-Khatami A, Bauers-feld U. Long-term experience with AutoCapture-controlledepicardial pacing in children. Europace 2007;9:645–50.

2. Mond HG, Stokes KB. The steroid-eluting electrode. A 10-yearexperience. Pacing Clin Electrophysiol 1996;19:1016–20.

3. Vanagt WY, Verbeek XA, Delhaas T, Mertens L, Daenen WJ,Prinzen FW. The left ventricular apex is the optimal site forpaediatric pacing: correlation with animal experience. Pac-ing Clin Electrophysiol 2004;27:837–43.

4. Vanagt WY, Verbeek XA, Delhaas T, et al. Acute hemody-namic benefit of left ventricular apex pacing in children. AnnThorac Surg 2005;79:932–6.

5. Nothroff J, Norozi K, Arnhold JO, Wessel A, Ruschewski W,Buchhorn R. Pacemaker implantation as a risk factor forheart failure in young adults with congenital heart disease.Pacing Clin Electrophysiol 2006;29:386–92.

6. Bauersfeld U, Tomaske M, Dodge-Khatami A, Rahn M,Kellenberger C, Pretre R. Initial experience with ImplantableCardioverter Defibrillator Systems using epicardial andpleural electrodes in paediatric patients. Ann Thorac Surg

2007;84:303–5.