eli 250 / eli 210 service manual - quince medical & surgical · 2020. 12. 23. · ref:...

111
REF: 9516-163-50-ENG Rev: B1 ELI 250 / ELI 210 Service Manual Manufactured by Mortara Instrument, Inc. Milwaukee U.S.A. CAUTION: Use of this service manual is intended for qualified service personnel only. CAUTION: Federal law restricts this device for sale to or on the order of a physician.

Upload: others

Post on 25-Jan-2021

8 views

Category:

Documents


0 download

TRANSCRIPT

  • REF: 9516-163-50-ENG Rev: B1

    ELI 250 / ELI 210 Service Manual

    Manufactured by Mortara Instrument, Inc. Milwaukee U.S.A.

    CAUTION: Use of this service manual is intended for qualified service personnel only.

    CAUTION: Federal law restricts this device for sale to or on the order of a physician.

  • Copyright© 2004 by Mortara Instrument, Inc.

    7865 N. 86th Street Milwaukee, Wisconsin 53224

    This document contains confidential information that belongs to Mortara Instrument, Inc. No part of this document may be transmitted,

    reproduced, used, or disclosed outside of the receiving organization without the express written consent of Mortara Instrument, Inc.

    Mortara is a registered trademark of Mortara Instrument, Inc

  • i

    TECHNICAL SUPPORT AND SERVICE

    Headquarters

    Mortara Instrument, Inc.

    7865 North 86th Street

    Milwaukee, WI 53224

    Tel: 414.354.1600

    Tel: 800.231.7437

    Fax: 414.354.4760

    Internet: http://www.mortara.com

    Europe Economic Community Representative

    Mortara Rangoni Europe, Srl

    (European Headquarters, Italy)

    Via Oradour 7

    40016 S. Giorgio di Piano, BO

    Tel: +39.051.6654311

    Fax: +39.051.6651012

    Service/Technical Support Group

    Mortara Instrument, Inc.

    7865 North 86th Street

    Milwaukee, WI 53224

    Tel: 414.354.1600

    Service: 888.MORTARA

    (888.667.8272)

    Fax: 414.354.4760

    E-mail: [email protected]

    24 Hour Technical Support

    Same Day Shipment of Replacement Parts

    Biomedical Training Classes

    Extended Warranties/Service Contracts

    Sales Support/ Supplies & Accessories

    Mortara Instrument, Inc.

    7865 North 86th Street

    Milwaukee, WI 53224

    Tel: 414.354.1600

    Fax: 414.354.4760

    E-mail: [email protected]

    Mortara Instrument GmbH

    (Germany)

    Henricistr. 124

    45136 Essen

    Tel: +49.201.268311

    Fax: +49.201.268313

    Mortara Instrument B.V.

    (The Netherlands)

    H. Dunantplein 6

    3731 CL De Bilt

    Postbus 131

    3720 AC Bilthoven

    Tel: +31.30.2205050

    Fax: +31.30.2201531

  • ii

    Notices Manufacturer’s Responsibility

    Mortara Instrument, Inc., is responsible for the effects on safety and performance only if:

    Assembly operations, extensions, readjustments, modifications or repairs are carried out only by persons authorized by Mortara Instrument,

    Inc.

    The device (ELI 250/210) is used in accordance with the instructions for use.

    Responsibility of the Customer The user of this product is responsible for ensuring the implementation of a satisfactory maintenance schedule. Failure to do so may cause

    undue failure and possible health hazards.

    Equipment Identification Mortara Instrument, Inc. equipment is identified by a serial and reference number on the back of the device. Care should be taken so that

    these numbers are not defaced.

    Copyright and Trademark Notices This document contains information that is protected by copyright. All rights are reserved. No part of this document may be photocopied,

    reproduced or translated to another language without prior written consent of Mortara Instrument, Inc.

    Other Important Information The information in this document is subject to change without notice. Mortara Instrument, Inc. makes no warranty of any kind with regard to this material including, but not limited to, implied warranties of

    merchantability and fitness for a particular purpose. Mortara Instrument, Inc. assumes no responsibility for any errors or omissions that may

    appear in this document. Mortara Instrument Inc. makes no commitment to update or to keep current the information contained in this

    document.

  • iii

    Warranty Information

    Your Mortara Warranty

    MORTARA INSTRUMENT, INC. (hereinafter referred to as “Mortara”) hereby warrants that Mortara products (hereinafter

    referred to as “Products”) shall be free from defects in material and workmanship under normal use, service and maintenance

    for the warranty period of such Product from Mortara or an authorized distributor or representative of Mortara. Normal use,

    service and maintenance means operation and maintenance in accordance with appropriate instructions and/or information

    guides. This Warranty does not apply to damage to the Products caused by any or all of the following circumstances or

    conditions:

    a) Freight damage;

    b) Parts and/or accessories of the Products not obtained from or approved by Mortara;

    c) Misapplication, misuse, abuse and failure to follow the Product instruction sheets and/or information guides;

    d) Accident, a disaster affecting the Products;

    e) Alterations or modifications to the Products not authorized by Mortara;

    f) Other events outside of Mortara’s reasonable control or not arising under normal operating conditions.

    THE REMEDY UNDER THIS WARRANTY IS LIMITED TO THE REPAIR OR REPLACEMENT WITHOUT CHARGE

    FOR LABOR OR MATERIALS, OR ANY PRODUCTS FOUND UPON EXAMINATION BY MORTARA TO HAVE

    BEEN DEFECTIVE. This remedy shall be conditioned upon receipt of notice by Mortara of any alleged defects promptly after

    discovery thereof within the warranty period. Mortara’s obligations under the foregoing warranty will further be conditioned

    upon the assumption by the purchaser of the Products (i) of all carrier charges with respect to any Products returned to

    Mortara’s principal place or any other place as specifically designated by Mortara or an authorized distributor or representative

    of Mortara, and (ii) all risk of loss in transit. It is expressly agreed that the liability of Mortara is limited and that Mortara does

    not function as an insurer. A purchaser of a Product, by its acceptance and purchase thereof, acknowledges and agrees that

    Mortara is not liable for loss, harm or damage due directly or indirectly to an occurrence or consequence therefrom relating to

    the Products. If Mortara should be found liable to anyone under any theory (except the expressed warranty set forth herein) for

    loss, harm or damage, the liability of Mortara shall be limited to the lesser of the actual loss, harm or damage, or the original

    purchase price of the Product when sold.

    EXCLUDED FROM THE LIMITED WARRANTY SET FORTH ABOVE ARE CONSUMABLE ITEMS SUCH AS PAPER,

    BATTERIES, ELECTRODES, PATIENT CABLES, LEAD WIRES AND MAGNETIC STORAGE MEDIUMS.

    EXCEPT AS SET FORTH HEREIN WITH RESPECT TO REIMBURSEMENT OF LABOR CHARGES, A PURCHASER’S

    SOLE EXCLUSIVE REMEDY AGAINST MORTARA FOR CLAIMS RELATING TO THE PRODUCTS FOR ANY AND

    ALL LOSSES AND DAMAGES RESULTING FROM ANY CAUSE SHALL BE THE REPAIR OR REPLACEMENT OF

    DEFECTIVE PRODUCTS TO THE EXTENT THAT THE DEFECT IS NOTICED AND MORTARA IS NOTIFIED WITHIN

    THE WARRANTY PERIOD. IN NO EVENT, INCLUDING THE CLAIM FOR NEGLIGENCE, SHALL MORTARA BE

    LIABLE FOR INCIDENTAL, SPECIAL OR CONSEQUENTIAL DAMAGES, OR FOR ANY OTHER LOSS, DAMAGE OR

    EXPENSE OF ANY KIND, INCLUDING LOSS OF PROFITS, WHETHER UNDER TORT, NEGLIGENCE OR STRICT

    LIABILITY THEORIES OF LAW, OR OTHERWISE. THIS WARRANTY IS EXPRESSLY IN LIEU OF ANY OTHER

    WARRANTIES, EXPRESS OR IMPLIED, INCLUDING, BUT NOT LIMITED TO THE IMPLIED WARRANTY OF

    MERCHANT ABILITY AND THE WARRANTY OF FITNESS FOR A PARTICULAR PURPOSE.

  • iv

    User Safety Information

    Warning:

    Means there is the possibility of personal injury to you or others.

    Caution:

    Means there is the possibility of damage to the equipment.

    Electrostatic Caution

    Means that there is the possibility of damage due to electrostatic discharge and appropriate precautions should be taken.

    Note: Provides information to further assist in the

    use of the device.

    Federal law restricts this device for sale to or on the order of a physician.

    Warning(s)

    • Device (electrocardiograph, Class I) captures and presents data reflecting a patient’s physiological condition that when reviewed by a trained physician or clinician can be useful in determining a diagnosis. However, the data should not be

    used as a sole means for determining a patient’s diagnosis.

    • To ensure that electrical safety is maintained during operation from AC (~) power, the device must be plugged into a Hospital Grade outlet.

    • To maintain designed operator and patient safety, peripheral equipment and accessories used that can come in direct patient contact, must be in compliance with UL 2601-1, IEC 601-1 and IEC 601-2-25.

    • Patient cables intended for use with the ELI 250/210 include series resistance (10 Kohm minimum) in each lead for defibrillation protection. Patient cables should be checked for cracks or breakage prior to use.

    • Conductive parts of the patient cable, electrodes and associated connections of Type CF applied parts, including the neutral conductor of the patient cable and electrode should not come into contact with other conductive parts, including earth

    ground.

    • To maintain designed operator and patient safety, only use parts and accessories supplied with the device and available through Mortara Instrument, Inc.

    • ECG electrodes could cause skin irritation and should be examined for signs of irritation or inflammation. • To prevent possible infection, single use components (e.g., electrodes) should be limited to one-time use only. • To avoid the possibility of serious injury or death during patient defibrillation, do not come into contact with device or

    patient cables. Additionally, proper placement of defibrillator paddles in relation to the electrodes is required to minimize

    harm to the patient.

    • To ensure the safety of both the patient and the device, 1.5 meters (5 feet) of open area should surround the patient. • A possible explosion hazard exists; do not use the device in the presence of flammable anesthetics.

  • v

    • Before attempting to use the device for clinical applications the operator must read and understand the contents of the manual and any documents accompanying the device.

    • Where the integrity of external protective earth conductor arrangement is in doubt, the ELI 250/210 shall be operated from its internal electrical power source.

    • All signal input and output (I/O) connectors are intended for connection of only devices complying with IEC 60601-1, or other IEC standards (e.g. IEC 60950), as appropriate to the device. Connecting additional devices to the ELI 250/210 may

    increase chassis and/or patient leakage currents. To maintain operator and patient safety, consideration should be given to

    the requirements of IEC 60601-1-1, and leakage currents should be measured to confirm no electric shock hazard exists.

    • The ELI 250/210 has not been designed for use with high-frequency (HF) surgical equipment and does not provide a protective means against hazards to the patient.

    • The quality of the signal produced by the electrocardiograph may be adversely affected by the use of other medical equipment, including but not limited to defibrillators and ultrasound machines.

    • The ELI 250/210 uses a lead-acid gel cell battery. Disposal of this battery must be in compliance with governmental regulations.

    • This unit can operate using AC power. Caution should be taken when working on this unit to limit the exposure to high voltages. The unit should not be connected to AC power and the battery fuse should be removed when replacing any

    assembly of this unit.

  • vi

    Caution(s)

    • To prevent possible damage to the keypad, do not use sharp or hard objects to depress keys, only use fingertips. • Do not attempt to clean the device or patient cables by submersing into a liquid, autoclaving, or steam cleaning. Wipe the

    exterior surfaces with a warm water and mild detergent solution and then dry with a clean cloth.

    • No user serviceable parts inside. Screw removal by qualified service personnel only. • The rechargeable internal battery is a sealed lead acid type and it is totally maintenance free. If the battery appears to

    become defective, refer to Mortara Instrument Service Department.

    • Do not pull or stretch patient cables as this could result in mechanical and/or electrical failures. Patient cables should be stored after forming them into a loose loop.

    Electrostatic Caution •

    Notes

    • Excessive patient movement could interfere with the operation of the device. • Proper patient preparation is important to proper application of ECG electrodes and operation of the device. • There is no known safety hazard if other equipment, such as pacemakers or other stimulators, are used simultaneously with

    the ELI 250/210; however, disturbance to the signal may occur.

    • If the ECG amplifier input is out of normal operating range, the display will indicate a lead fail for the lead(s) where this condition is present and if the signal is being printed, the respective lead(s) will print out as a square wave.

  • vii

    • As defined by IEC 60601-1 and IEC 60601-2-25, the device is classified as follows: • Class I equipment or internally powered • Type CF applied parts • Ordinary equipment • Not suitable for use in the presence of flammable anesthetics • Continuous operation

    NOTE: From a safety perspective, per IEC 60601-1 and derivative standards / norms, this unit is declared to be

    “Class I” and uses a three-prong inlet to ensure an earth connection is made along with mains. The ground terminal

    on the mains inlet is the only protective earth point in the unit. Exposed metal accessible during normal operation is

    double insulated from mains. Internal connections to earth ground are functional earth.

    • The ELI 250/210 will automatically turn off (blank screen) if the batteries have been severely discharged. • After operating the ELI 250/210 using battery power, always reconnect the power cord. This ensures that the batteries will

    be automatically recharged for the next time you use the ELI 250/210. A light will illuminate, next to the on/off switch,

    indicating that the unit is charging.

    • The ELI 250/210 is a UL Classified Device:

    WITH RESPECT TO ELECTRIC SHOCK, FIRE AND MECHANICAL HAZARDS ONLY IN ACCORDANCE WITH UL2601-1, IEC60601-1, CAN/CSA CC22.2 No. 601.1, AND IEC60601-2-25

  • viii

  • ix

    Table of Contents

    1 General

    Service Manual Purpose ......................................................................................................1-1 User Safety Information ........................................................................................................1-1

    Periodic Safety Inspections .....................................................................................1-1 Proper Power Cord ..................................................................................................1-1 Proper Fuse .............................................................................................................1-1 Do Not Operate in Explosive Atmospheres .............................................................1-1 Use Only Safe Methods of Interconnection .............................................................1-1 Do Not Mount Product above Patient ......................................................................1-1 Recommended Accessories ....................................................................................1-1 Sterilizing this Product .............................................................................................1-2 Liquid Spills .............................................................................................................1-2 System Information Log ...........................................................................................1-2 Product Information .................................................................................................1-2 Manual Symbol Delineation .....................................................................................1-2

    2 Maintenance and Cleaning

    Introduction (Maintenance and Cleaning) .............................................................................2-1 Recommended Cleaning Supplies .......................................................................................2-1 Equipment Required..............................................................................................................2-1 Cleaning and Inspection Techniques ...................................................................................2-2 Interior Cleaning ...................................................................................................................2-2 Magnetic Cleaning ................................................................................................................2-2 Printed Circuit Board Cleaning .............................................................................................2-2 Metallic and Plastic Parts Cleaning ......................................................................................2-3 Exterior Cleaning ..................................................................................................................2-3 Printhead Cleaning ...............................................................................................................2-3 Inspection of Writer Assembly Harness ...............................................................................2-3 Exterior Inspection ................................................................................................................2-4 Interior Visual Inspection ......................................................................................................2-4 Preventive Maintenance Schedule .......................................................................................2-4

    3 System Settings

    Access Configuration Menu for the ELI 250..........................................................................3-1 Summary of Configuration Menus.........................................................................................3-3 Configuration Page 1.............................................................................................................3-6 Software Version .........................................................................................................3-6 Cart Number ................................................................................................................3-6 Site Number.................................................................................................................3-6 Telephone Number ......................................................................................................3-6 Language.....................................................................................................................3-7 Volume.........................................................................................................................3-7 Battery Timeout ...........................................................................................................3-7 Configuration Page 2 ...........................................................................................................3-8 Flash Size ....................................................................................................................3-8 ID Format .....................................................................................................................3-8 AC Filter .......................................................................................................................3-8 Paper Speed................................................................................................................3-8 Filter .............................................................................................................................3-9 Height/Weight Units .....................................................................................................3-9

  • x

    Date Format .................................................................................................................3-9 Plot Format ..................................................................................................................3-9 Interpretation Option ....................................................................................................3.9 Reasons.......................................................................................................................3-10 Configuration Page 3 ...........................................................................................................3-10 Append to ECG............................................................................................................3-10 Number of Copies........................................................................................................3-10 ECG Retrieved.............................................................................................................3-10 Delete Rule ..................................................................................................................3-11 Storage Sensitivity .......................................................................................................3-11 Auto Save ECG............................................................................................................3-11 Auto Print ECG ............................................................................................................3-11 Serial Baud Rate..........................................................................................................3-11 Modem Baud Rate.......................................................................................................3-11 Use A4 Paper...............................................................................................................3-11 Caps Lock....................................................................................................................3-12 Configuration Page 4 ...........................................................................................................3-12 Rhythm Leads..............................................................................................................3-12 12-Lead Rhythm Printout.............................................................................................3-12

    Access Configuration Menu for the ELI 210..........................................................................3-13 Summary of Configuration Menus.........................................................................................3-16 Configuration Page 1.............................................................................................................3-18 Software Version .........................................................................................................3-18 Cart Number ................................................................................................................3-18 Site Number .................................................................................................................3-18 Telephone Number ......................................................................................................3-18 Language.....................................................................................................................3-18 Volume.........................................................................................................................3-18 Battery Timeout............................................................................................................3-19 AC Filter .......................................................................................................................3-19 Paper Speed................................................................................................................3-19 Filter .............................................................................................................................3-19 Height/Weight Units .....................................................................................................3-19 Configuration Page 2 ...........................................................................................................3-20 Date Format .................................................................................................................3-20 Plot Format ..................................................................................................................3-20 Interpretation Option ....................................................................................................3-21 Reasons.......................................................................................................................3-21 Append to ECG............................................................................................................3-21 Number of Copies........................................................................................................3-21 Use A4 Paper...............................................................................................................3-21 Configuration Page 3 ...........................................................................................................3-22 Rhythm Leads..............................................................................................................3-22 12-Lead Rhythm Printout.............................................................................................3-22

    4 Assembly/Disassembly of Unit

    Fuse Removal .......................................................................................................................4-1 Cover Assembly Removal .....................................................................................................4-2 Battery Removal ....................................................................................................................4-7 Keyboard/LCD Removal........................................................................................................4-7 Writer Removal / Installation..................................................................................................4-11 Printhead Removal / Installation............................................................................................4-13 Writer Motor Removal............................................................................................................4-15 Printed Circuit Board Assembly Removal / Installation .........................................................4-17 Front End Board Removal / Installation.................................................................................4-19

  • xi

    5 Technical Description

    Introduction............................................................................................................................5-1 ELI 250/210 Block Diagram...................................................................................................5-2 Main System Board ...............................................................................................................5-3 Front End Board ....................................................................................................................5-3 Input Connector Pin out.........................................................................................................5-4 ELI 250/210 Specifications....................................................................................................5-5 Parts List................................................................................................................................5-6

    6 Troubleshooting and Testing

    Introduction............................................................................................................................6-1 Troubleshooting.....................................................................................................................6-1 AC Power, Backlight on, Nothing on Display ..............................................................6-1 With AC On, Nothing on Display .................................................................................6-1 Unit Works on AC but not on Battery...........................................................................6-1 Unit turns on, No waveform on screen but Function Keys Displayed .........................6-2 No Keyboard Functions and Display Frozen...............................................................6-2 RS232 Transmit / Receive Problems ..........................................................................6-2 Optional Modem Transmit / Receive Problems...........................................................6-3 Testing ...................................................................................................................................6-4 Power Supply Test.......................................................................................................6-4 Keyboard test...............................................................................................................6-4 LCD Test......................................................................................................................6-4 Speaker Test ...............................................................................................................6-4 Date / Time Test ..........................................................................................................6-5 Writer Test ...................................................................................................................6-5 Writer Test Description ........................................................................................6-6 Adjusting the Writer Cue Sensor .................................................................................6-6 ECG Test .....................................................................................................................6-7 Noise Test....................................................................................................................6-7 Lead Fail Test ..............................................................................................................6-8 Transmitting Records...................................................................................................6-8 Testing Continued Serial Port Test ..................................................................................................6-10 Modem Test .......................................................................................................6-10 Modem Initialization ................................................................................6-10

    Receive ECGs ...................................................................................................6-12 WLAN Test.........................................................................................................6-13 Safety Test .........................................................................................................6-14

    Appendix A ELI 250 Connectivity

    Chapter Purpose ...................................................................................................................A-1 Transmitting records..............................................................................................................A-1 Direct Connection ..................................................................................................................A-3 Modem Connection ...............................................................................................................A-3 WLAN Transmission..............................................................................................................A-5 LAN Transmission .................................................................................................................A-6 Patient Request List ..............................................................................................................A-11 Custom ID Download.............................................................................................................A-13

  • xii

    Appendix B ELI 250 Connectivity Options

    Modem Options .....................................................................................................................B-1 XIRCOM Modem........................................................................................................B-1 Multitech Modem........................................................................................................B-3 LAN Option ............................................................................................................................B-5 Wireless LAN Option .............................................................................................................B-8

  • __________________________________________________________________________ Section 1

    1-1

    1 General

    Service Manual Purpose

    The purpose of this manual is to supply information to service personnel so they can maintain the ELI 250/210

    Interpretive and Non-Interpretive Electrocardiograph at the assembly and subassembly level. Although the manual

    includes parts lists, mechanical assembly parts, and printed circuit board information, it is intended to function

    primarily as a guide to preventative and corrective maintenance and electrical repairs considered field repairable.

    User Safety Information

    Periodic Safety Inspections

    Follow the recommended maintenance schedule. Inspect the power cord and transmission cables

    periodically for fraying or other damage and replace as needed. Broken or frayed wires may cause

    interference or loss of signal. Pay particular attention to points where wires enter connectors.

    Proper Power Cord

    Use only the power cord specified for the equipment. This product requires a three-wire, (18 gauge, SJT-

    grade) power cord, which is supplied with a three-terminal, polarized plug (hospital grade) for connection

    to the power source and protective ground. Use only a power outlet with a protective ground outlet. An

    interruption of the grounding connection could cause an electrical shock hazard.

    Proper Fuse

    Use only the fuse specified for the equipment (identical in type, voltage and current rating). Substituting a

    different fuse type could cause a fire hazard. Always make sure fuses have been installed before operating

    the unit.

    Do Not Operate in Explosive Atmospheres

    Do not operate the ELI 250/210 in the presence of flammable gasses or anesthetics; this environment could

    cause an explosion. Refer to Operator's Manual Safety Information: Warning(s) and Caution(s).

    Use Only Safe Methods of Interconnection

    To prevent electrical shock from the product when it is connected to other electrical equipment, proper

    grounding is essential. Refer to Operator's Manual Safety Information: Warning(s) Peripheral equipment.

    Do Not Mount Product above Patient

    Do not mount or place the product where it could fall on a patient or where it could be accidentally knocked

    off a shelf or other mounting arrangement.

    Recommended Accessories

    For the patient's safety and optimum equipment performance, use only the accessories specified by Mortara

    Instrument, Inc.

  • ELI 250/210 _________________________________________________________________________

    1-2

    Sterilizing this Product

    Do not sterilize this product or any accessories unless specifically directed by the manufacturer. Sterilization and

    sterilization environments can seriously damage many components and accessories.

    Liquid Spills

    Do not set beverages or other liquids on or near the ELI 250/210, and/or optional equipment.

    Product Information

    See Section 1 of Operator's Manual

    Manual Symbol Delineation

    Electrostatic sensitive devices

    Caution or Special Instructions

  • ____________________________________________________________________________Section 2

    2-1

    2 Maintaining and Cleaning

    CAUTION:

    Remove battery fuse before attempting any disassembly.

    Removal of the battery fuse will not cause a loss of ECGs or of the configuration. This

    information is stored in FLASH memory.

    Introduction This section provides servicing and maintenance instructions for the ELI 250/210 interpretive electrocardiograph.

    Subsequent parts of this section are disassembly, inspection techniques, cleaning techniques, and installation.

    Recommended Cleaning Supplies

    • Anti-static mat & wrist band, properly grounded • Clean, lint-free cloth • Cleaning solvent (isopropyl alcohol, 99% pure) • DRY, low pressure, compressed air (30 psi) • Electronic safe non-residue solvent, such as Flux-off • Masking tape • Non-metallic, soft-bristle brush • Naphtha • Vacuum cleaner, Static Safe

    Equipment Required

    • Screwdriver Philips #2. • Allen Wrench 0.05 inch or a 1.25 mm • Vibra-Tite or similar product to prevent screws from vibrating loose. • Digital Multi Meter (DMM)

    Note: The equipment and solvent mentioned above are standard shop commodities that are available from

    commercial sources. If in the performance of normal maintenance or repair, the PCB assembly, AC

    connector assembly, or writer assemblies are removed or replaced, a leakage test should be performed.

    See section on testing.

  • ELI 250/210__________________________________________________________________________

    2-2

    Cleaning and Inspecting Techniques This section contains instructions for periodic cleaning and inspection of the instrument as preventative maintenance

    measures. It also contains specific cleaning procedures to be conducted after disassembly. Parts having identical

    cleaning procedures are grouped under common headings. No special tools are required.

    Interior Cleaning

    WARNING

    Ventilate work area thoroughly when using solvents. Observe manufacturers warnings on

    solvent containers with regard to personnel safety and emergency first aid. Be sure that first

    aid equipment is available before using chemicals. Observe all shop safety and fire

    precautions.

    Magnetic Cleaning

    CAUTION

    Do not use solvents to clean transformers or inductors. The chemical action of solvents may

    remove the varnish from the wire coils, rendering the components useless. The solvent also

    neutralizes the adhesive of the cover tape, resulting in eventual tape separation from the

    windings.

    Clean transformers and inductors with a dry, non-metallic, soft bristle brush.

    Printed Circuit Board Cleaning The printed circuit board assemblies in the ELI 250/210 system contain static sensitive devices. Use special

    handling procedures to prevent damage due to ESD.

    Clean assembled parts with a vacuum cleaner or low pressure compressed air (30 psi). Take care when cleaning

    printed circuit boards that wires or component leads are not bent back and forth in such a manner as to weaken them

    and cause them to eventually break.

    Dry with low pressure compressed air.

  • ____________________________________________________________________________Section 2

    2-3

    Metallic and Plastic Parts Cleaning

    CAUTION

    Do not wipe over surfaces of nameplates or labels with abrasive cleaners or materials,

    as this will eventually wear away the nameplate information. Do not use solvents to

    clean plastic parts.

    Brush all surfaces and parts with a nonmetallic, soft bristle brush.

    Wipe metal surfaces with soft, nonabrasive cloth dampened with isopropyl alcohol.

    Dry surfaces with clean cloth.

    Wipe surfaces of nameplates and labels with dry cloth.

    Exterior Cleaning Use a damp cloth to clean external covers and the line cord. Do not use alcohol, solvents, or cleaning solutions.

    These cleaning agents may damage the surfaces of the instrument.

    Printhead Cleaning

    Open the writer cover as explained in the section on final assembly.

    Apply isopropyl alcohol to a clean cloth, and wipe the writer printhead until all foreign matter is removed.

    Dry the printhead thoroughly before use. Use a dry clean cloth for this.

    After cleaning is completed, inspect the unit using the techniques described previously.

    After the inspection is complete, install the cover as explained in the section on final assembly.

    Inspection of Writer Assembly Harness Visually inspect the wire harnesses for wear, maintenance damage, corrosion, deterioration, and damage resulting

    from dropping.

    If no defects are found, install the cover and restore the unit to service.

    If a defect is found in the cables, replace the damaged cable with a new cable.

  • ELI 250/210__________________________________________________________________________

    2-4

    Exterior Inspection

    Visually inspect the entire instrument for wear, maintenance damage, corrosion, deterioration, and damage resulting

    from dropping.

    Interior Visual Inspection

    Check all connectors for loose, bent or corroded contact points

    Check wire, harnesses and cables for signs of wear or deterioration.

    Inspect components and their leads for security of mounting, deterioration or leakage.

    Check terminals and connections for proper installation, failed soldering, loss or wear.

    Inspect PCB surfaces for charring, cracking or brittleness.

    Note: Some degree of discoloration of the PCB surface may be expected due to continued exposure to the

    operating temperatures of some of the components.

    Check the identification nameplate and other decals for legibility.

    Inspect chassis, covers, and brackets for warping, bending, surface damage or missing captive hardware.

    Check all screws and nuts for tightness or signs of stripped or crossed threads.

    Check for any other form of mechanical damage, which may indicate a failure.

    If, during the process of normal maintenance or repair, the PCB assembly, AC connector assembly, or writer

    assembly are removed and replaced, perform the leakage test as describe in the section on Testing and

    Troubleshooting.

    Preventative Maintenance Schedule:

    Maintenance to be Performed Period Notes

    Clean and inspect unit. 6 mo. Perform every 3 mo. if unit is in heavy use.

    Printhead cleaning 80 hrs Monthly cleaning recommended

    Leakage tests. - As required by facility or regulatory requirements.

    Table 1

  • ____________________________________________________________________________Section 3

    3-1

    3 System Settings

    Chapter Purpose

    This chapter is intended to provide the user with:

    • Instructions to access configuration menus • Summary of configuration options • Detailed descriptions of configuration settings Access Configuration Menus

    Four configuration pages define all ELI 250 operational conditions that do not change on a daily or patient-to-patient

    basis. Once you set these default conditions, you will rarely need to use the configuration screens again. When you

    apply power to the ELI 250, it will operate according to the settings you have selected. To access the configuration

    menus:

    Select More (F6) from the Real-Time ECG View.

  • ELI 250/210__________________________________________________________________________

    3-2

    Select Set Time/Date (number 3) from the application menu

    From the Set Time/Date screen, simultaneously press (SHIFT) + ALT + C. The first configuration screen will appear. Notice the page indicator in the upper right hand corner on

    each screen.

  • ____________________________________________________________________________Section 3

    3-3

    NOTES: Use Page (F4) to toggle through the configuration pages. Use ▲ (F1) and ▼ (F2) to move back and forth through each configuration option. Use ► (F3) to toggle through pre-programmed available settings per configuration field. Use Exit (F6) to return to the Real-Time ECG View. Any changes you have made will be saved. Use the BACKSPACE key to erase entry errors.

    The following chart summarizes the configuration screens and the available options for each parameter.

  • ELI 250/210__________________________________________________________________________

    3-4

    Summary of Configuration Menus

    Configuration Parameter

    Definition Configuration Screen

    Software Version The firmware version of the unit Screen One

    Cart Number User-defined (4 digits) Screen One

    Site Number User-defined (4 digits) Screen One

    Site Name User-defined (up to 30 alphanumeric characters) Screen One

    Telephone Number User-defined (up to 30 alphanumeric characters) Screen One

    Language Firmware language availability Screen One

    Volume Range from 0 – 8 Screen One

    Battery Timeout 10min, 30min, 60min Screen One

    Flash Size Normal memory or Expanded memory (optional) Screen Two

    ID Format Short, Long, Standard, Custom (optional) Screen Two

    AC Filter None, 60Hz, 50Hz Screen Two

    Paper Speed 25mm/sec or 50 mm/sec Screen Two

    Filter 40Hz, 150Hz, or 300Hz Screen Two

    Height/Weight Units lb/in or kg/cm Screen Two

    Date Format US (mm/dd/yyyy) or European (dd.mm.yyyy) Screen Two

    Plot Format Channels printed: 3+1, 6, 3+3, 12, 6+6; Cabrera or

    Standard Screen Two

    Interpretation Yes or No Screen Two

    Reasons Yes or No Screen Two

  • ____________________________________________________________________________Section 3

    3-5

    Append Unconfirmed Report, Reviewed by, Blank Screen Three

    #of copies 0 - 9 Screen Three

    #ECGs retrieved 0 - 9 Screen Three

  • ELI 250/210__________________________________________________________________________

    3-6

    Configuration Parameter

    Definition Configuration Screen

    Delete Rule Post plot, post transmit, post plot/xmt Screen Three

    Storage Sensitivity Normal or High Screen Three

    Auto-save ECG Yes or No Screen Three

    Auto-print ECG Yes or No Screen Three

    Baud Rate Serial Baud Rates: 9600, 19200, 38400, 57600, or

    115200 Screen Three

    Use A4 paper Yes or No Screen Three

    Caps Lock Yes or No Screen Three

    Rhythm format 3 channel, 6 channel or 12 channel Screen Four

    3 Rhythm Lead 1 V1-V6, I, II, III, aVR, aVL, aVF Screen Four

    3 Rhythm Lead 2 V1-V6, I, II, III, aVR, aVL, aVF Screen Four

    3 Rhythm Lead 3 V1-V6, I, II, III, aVR, aVL, aVF Screen Four

    6 Rhythm Lead 1 V1-V6, I, II, III, aVR, aVL, aVF Screen Four

    6 Rhythm Lead 2 V1-V6, I, II, III, aVR, aVL, aVF Screen Four

    6 Rhythm Lead 3 V1-V6, I, II, III, aVR, aVL, aVF Screen Four

    6 Rhythm Lead 4 V1-V6, I, II, III, aVR, aVL, aVF Screen Four

    6 Rhythm Lead 5 V1-V6, I, II, III, aVR, aVL, aVF Screen Four

    6 Rhythm Lead 6 V1-V6, I, II, III, aVR, aVL, aVF Screen Four

    3 + 1 Rhythm Lead V1-V6, I, II, III, aVR, aVL, aVF Screen Five

    3 + 3 Rhythm Lead 1 V1-V6, I, II, III, aVR, aVL, aVF Screen Five

  • ____________________________________________________________________________Section 3

    3-7

    3 + 3 Rhythm Lead 2 V1-V6, I, II, III, aVR, aVL, aVF Screen Five

    3 + 3 Rhythm Lead 3 V1-V6, I, II, III, aVR, aVL, aVF Screen Five

    Default XMT Media RS232, Modem, WLAN, LAN Screen Six

    DHCP (LAN only) Yes or No Screen Six

    IP Address XXX.XXX.XXX.XXX Screen Six

    Def Gateway XXX.XXX.XXX.XXX Screen Six

    Sub Net Mask XXX.XXX.XXX.XXX Screen Six

    Host IP XXX.XXX.XXX.XXX Screen Six

    Port Number XXX.XXX.XXX.XXX Screen Six

    SSID Yes or No Screen Six

    Channel Number 0, 1, 2, 3..... Screen Six

    Security WEP Yes or No Screen Seven

    WEP Key 0, 1, 2, 3..... Screen Seven

    WEP Key ID 26 digits in 13 sets of two digits Screen Seven

  • ELI 250/210__________________________________________________________________________

    3-8

    Configuration Page 1

    Software Version The displayed number identifies the firmware version of your electrocardiograph.

    Cart Number

    Cart numbers indicate which electrocardiograph transmitted a particular ECG.

    Site Number This option identifies the site of your ELI 250. Site numbers designate the hospital, clinic, or institution for ECG

    records stored in a Mortara Instrument, E-Scribe data management system and must be defined for transmitting and

    retrieving ECGs from the data management system. You can use up to four digits for the site number, numbers

    from 0 –4095 are supported.

    Site Name

    This option defines your clinic, hospital, or office name. You can enter up to 30 alphanumeric characters. The site

    name prints at the bottom, left edge of the ECG printout.

    Telephone Number

    This option specifies the telephone number for modem transmission (to another unit or to an E-Scribe data

    management system). Enter up to 30 alphanumeric characters, and use the BACKSPACE key to erase entry errors.

    To wait for an additional dial tone, use the letter W. For example, you may need to dial a 9 to get an outside line.

    EXAMPLE: 9W14145554321

    To insert a pause use a comma (,).

    To change tone dialing to pulse dialing, use the letter P.

    EXAMPLE: P14145554321

  • ____________________________________________________________________________Section 3

    3-9

    (If necessary, you can use both the letter W and the letter P in the same phone number.)

    TIP: Instead of entering the configuration menus, use a shortcut to quickly delete or modify a phone number. From

    the application screen, simultaneously press (SHIFT) + ALT + P. To edit an existing number, use the tab key.

    Language

    There are several languages available on the cardiograph.

    CAUTION: upon selecting a new language and exiting the configuration screen, function labels are

    immediately translated.

    If an unknown language is visible, use the following steps to revert to the language of your country:

    1. F6 from Real-Time ECG View 2. Select number 3 3. Simultaneously press (SHIFT) + ALT + C 4. Press F2 (▼) four times 5. Press F3 (►) until the desired language appears 6. F6 to return to Real-Time ECG View

    Volume This option defines the keyboard click loudness. Available settings rage from 0 (off) to 8 (loud).

    TIP: Use the BACKSPACE key to erase entry errors Battery Timeout

    This setting determines when the cardiograph will switch off in order to conserve the battery life of the unit. The

    battery timeout will only occur if the keypad has not been depressed for the time specified. The battery timeout

    setting is ignored if an active ECG signal is detected, during transmission, or while rhythm printing.

  • ELI 250/210__________________________________________________________________________

    3-10

    Configuration Page 2

    Flash Size Flash Size indicates the ECG storage capacity of your electrocardiograph. Normal Flash Size is the standard

    memory capacity. If the optional expanded memory has been installed, Expanded will be displayed.

    ID Format

    This option defines the format for the patient ID information prompts. There are three standard formats: short,

    standard, or long. A custom ID format is optional when downloaded from the E-Scribe Data Management System.

    The short format includes the patient's last name and first name, patient ID number, date of birth, which will

    automatically calculate the age, and gender field.

    The standard format includes the patient's last name, patient ID number, age, height, weight, gender, race,

    medication 1, medication 2, and a location field.

    The long format is identical to the standard format except that it includes the patient's first name, room and comment

    fields.

    The custom format, which is designed in the Mortara E-Scribe Data Management System, can be downloaded to the

    ELI 250. The custom ID format is uniquely designed to meet your facility’s needs. Please see Appendix A to

    download a Custom ID.

    AC Filter

    The ELI 250 removes 60Hz or 50Hz interference. The setting you select depends on the line frequency in your

    country. Always use the 60Hz setting in the United States. If the AC interference is present, check to see that the

    proper AC filter is selected.

    Paper Speed

    The writer speed can be configured at 25mm/s or 50mm/s for default ECG printouts. For rhythm printouts and

    display, speeds of 5mm/s or 10mm/s are also available. See Section 4 to change speeds for display or rhythm

    printing.

  • ____________________________________________________________________________Section 3

    3-11

    Filter The ECG plot frequency filter (or print filter) can be set to 40Hz, 150Hz, or 300Hz. The plot frequency filter does

    not filter the acquired digital record. A 40Hz plot filter setting will reduce the noise (40Hz and higher frequencies)

    on the printed ECG, a 150Hz plot filter setting will reduce the noise (150Hz and higher frequencies) on the printout

    and a 300Hz plot filter setting will not filter the printed ECG.

    Height/Weight Units

    This option defines the units of weight and height to either pounds/inches (lb/in) or kilograms/centimeters (kg/cm).

    Date Format This option defines the format for entering and displaying the patient’s date of birth in either the U.S. format or the

    European format.

    For example: July 16th

    , 2003 is displayed as follows: US Date Format: MM/DD/YYYY (07/16/2003) European Date Format: DD.MM.YYYY (16.07.2003)

    NOTE: the date format option does not modify the acquisition date printed on each ECG. Using the

    example above, the acquisition date would print as 16-Jul-2003.

    Plot Format This option defines the default for one of five available plot formats in either Standard or Cabrera presentation.

    Please note that regardless of the plot format selected, 10 seconds of 12-leads are always stored.

    The ECG plot options are:

    Format Option

    ECG Data

    3+1

    2.5 seconds of 12 leads in a three-channel format and the fourth channel is a 10 second rhythm strip of lead II (user defined).

    6

    5 seconds of 12-leads in a six-channel format

    3+3

    2.5 seconds of 12 leads in a three-channel format plus 10 second rhythm strip of leads II, V1 and V5 (user defined) in a three-channel format.

    12

    10 seconds of 12 leads in a 12 channel format

    6+6

    10 seconds of 12 leads in a 2 page printout (six channels per page)

    Interpretation Option

  • ELI 250/210__________________________________________________________________________

    3-12

    The ELI 250 automatically analyzes ECGs and prints the optional interpretation on the ECG printout. This option

    allows you to select or suppress the “interpretive” text on the ECG report.

    NOTE: A qualified physician should review the computer generated ECG interpretation before the

    treatment of any patient.

    Reasons

    The reason statements indicate why a particular interpretive statement was printed. Reason statements print in

    [square brackets] within the interpretive text (if the interpretation is turned on). Turning the reason statement

    function on or off does not affect the measurements performed or the interpretive statements selected by the analysis

    program.

    For Example:

    Anteroseptal Infarct [40+ MS Q WAVE IN V1-V4]

    Where “Anteroseptal Infarct” is the interpretive statement.

    And “40+ MS Q WAVE IN V1-V4” is the reason statement or explanation as to why the

    interpretive statement was printed.

    Configuration Page 3

    Append

    A status or statement phrase can be appended to the ECG and printed under the interpretive text printout. Either

    “unconfirmed report” or “reviewed by” can be selected. However, if you wish to have nothing appended to the

    ECG, “blank” may be selected.

    Number of Copies

    This option defines the number of printed copies when an ECG is taken. A zero (0) setting prints the original only;

    one (1) prints the original plus 1 copy; two (2) prints the original plus 2 copies, and so on. Up to 9 copies may be

    selected.

    ECGs Retrieved

  • ____________________________________________________________________________Section 3

    3-13

    This option defines the number of ECGs retrieved from an E-Scribe data management system. The ECGs are

    retrieved by ID number. A zero (0) setting retrieves the most current ECG for that ID number. Settings from one

    (1) to nine (9) retrieve the most current ECG plus “X” number of ECGs identified by the entered value.

    EXAMPLE: If you enter the number 5, you will retrieve the most current ECG plus the five preceding ECGs for

    that ID number. ECGs retrieved from the E-Scribe are only printed at the ELI250 and not saved.

    Delete Rule

    This option defines the ECG auto delete status in the patient directory. ECGs that are marked for deletion will be

    automatically removed or erased based on their record size to make room necessary for the new ECG record. ECGs

    are only erased from the directory when they are marked for deletion and if the directory becomes full. The delete

    rule selections are:

    Post Plot = ECG is automatically marked deleted after printing.

    Post Transmit = ECG is automatically marked deleted after transmission.

    Post Plot/Transmit = ECG is automatically marked deleted after transmission and printing.

    Storage Sensitivity The ECG storage sensitivity setting dictates the resolution of the stored record. The sensitivity settings are either

    Normal or High. If the configuration value is set to High, the stored ECG will have a high resolution. As a result,

    the records size will be large and will reduce the storage capacity in the ECG directory.

    Auto Save ECG

    This option defines whether or not the ELI 250 is to automatically save newly acquired ECGs in the directory, once

    they are acquired and printed. If the auto save configuration option is set to NO, and the record is printed, the ELI

    250 will prompt you to “Save ECG?” SAVE (F1) will store the ECG in the directory.

    NOTE: manually save ECGs by selecting More (F5) from the Acquired ECG view and Save (F5).

    Auto Print ECG

    This option defines whether or not the ELI 250 will automatically print the ECG after acquisition. If the selected

    configuration option is set to NO, a manual printout is possible after acquisition.

    Baud Rate

    The user-selectable baud rate determines the serial port’s data transmission rate in bits per second (bps). Set the

    baud rate to: 9600, 19200, 38400, 57600, or 115200 bps for direct data transmission between the ELI 250 and

    another Mortara cardiograph, and 38400 bps for a direct connection to the E-SCRIBE data management system.

    Use A4 Paper

    The ELI 250 accommodates use of Z-fold thermal paper in either the letter size (8.5 x 11 inches; 216 x 279 mm) or

    the “A4” size (8.27 x 11.69 inches; 210 x 297 mm). The provided paper tray spacer is required for use with A4 size

    paper. Enable this configuration option for A4 printing. The paper tray spacer should not be inserted if the letter

    size paper is used.

    Caps Lock All character entry is translated to upper case.

  • ELI 250/210__________________________________________________________________________

    3-14

    Configuration Page 4

    Rhythm Formats

    Configuration page four defines the default values for rhythm printing. It is possible to set a 3-channel, 6-channel or

    12-channel default rhythm format. Define rhythm leads one through three to customize a three-channel rhythm

    printout or define rhythm leads one through six to customize the six-channel rhythm printout. Use (F3) to toggle through each of the 12 leads.

    Configuration Page 5

    Rhythm Leads Configuration page five identifies the user-selectable 10-second rhythm lead printed in a 3 + 1 channel ECG printout

    and the three 10-second rhythm leads for the 3+3 channel ECG printout. Use (F3) to make your selection of any of the 12 leads.

    NOTE: rhythm acquisition is not stored in memory, only printed.

    NOTE: see section 4 to acquire a rhythm printout.

  • ____________________________________________________________________________Section 3

    3-15

    Configuration Page 6 (optional LAN connectivity)

    The ELI 250 connectivity options are RS232 (or direct), modem, LAN or WLAN. Configuration parameters

    associated with each type of transmission media will be presented in configuration pages six and seven only if those

    media are installed in the cardiograph.

    Default XMT Media

    Configuration page 6 identifies the default transmission media. Those connectivity options which have been

    optionally purchased and installed (RS232, modem, LAN or WLAN) in your ELI 250 will be available for default

    selection.

    DHCP

    This setting defines whether the Dynamic Host Communication Protocol (DHCP) will be used to obtain an IP

    address. If DHCP is Yes, the network will automatically and dynamically assign an IP address therefore, there is no

    need to manually enter an IP address. If DHCP is No, you must enter the IP address, Def Gateway and Sub Net

    Mask.

    NOTE: all parameters related to network connection must be entered under the direction of the IT

    Manager of the facility where the unit is installed.

    IP Address

    Define the fixed IP address used by the ELI 250 for network transmission (if DHCP is not selected).

    Def Gateway

    Enter the address of the default gateway (if DHCP is not selected).

    Sub Net Mask

    Enter the Sub Net address (if DHCP is not selected).

    Host IP

    Enter the IP address of the host server.

  • ELI 250/210__________________________________________________________________________

    3-16

    Note: addresses are always entered as 4 sets of 3 digits; therefore, an address of 192.168.0.7 must

    be entered on the ELI 250 as 192.168.000.007.

    Port Number

    Enter the port number used by the host server.

    Configuration Page 6, continued (optional WLAN connectivity)

    SSID

    Service Set Provider (SSID) is the name of the wireless network. All ELI 250 electrocardiographs that will transmit

    to the same network must have the same SSID name. This field is case sensitive.

    Channel Number

    This defines the channel used by the radio to search for the network. The exact channel number varies according to

    which type of wireless network is used and the country in which you install the WALN. When moving from one

    area to another, the ELI 250 WLAN card will automatically scan for the channel of the closest access point.

    Configuration Page 7(optional WLAN connectivity, continued)

  • ____________________________________________________________________________Section 3

    3-17

    Security (WEP)

    Wired Equivalent Privacy (WEP) is an encrypted security protocol, part of the 802.11 standard, and is used to

    protect the network from unauthorized users. The IT Manager will provide the 26 HEX digits that are the same to

    be used by all access points on the network. Therefore, only devices with the proper WEP will be able to login. The

    ELI 250 adopts a 128-bit encryption. Access points can have multiple WEP Keys stored. Each one of them is

    identified by a number (e.g., 0, 1, 2, 3).

    WEP Key

    Enter the WEP Key number (e.g., 0, 1, 2, 3).

    WEP Key ID

    Enter the 128-bit WEP Key ID value (26 digits in 13 sets of two digits).

  • ELI 250/210__________________________________________________________________________

    3-18

    Access Configuration Menus for the ELI 210

    Three configuration pages define all ELI 210 operational conditions that do not change on a daily or patient-to-

    patient basis. Once you set these default conditions, you will rarely need to use the configuration screens again.

    When you apply power to the ELI 210, it will operate according to the settings you have selected. To access the

    configuration menus:

    Select More (F6) from the Real-Time ECG View.

  • ____________________________________________________________________________Section 3

    3-19

    Select Set Time/Date (number 3) from the application menu.

    Simultaneously press ▲ (F1) and ▼ (F2). The first configuration screen will appear. There are a total of 3 configuration screens. Notice the page indicator in the upper right hand corner

    on each screen.

    NOTES: Use Page (F4) to toggle through the three configuration pages.

    Use ▲ (F1) and ▼ (F2) to move back and forth through each configuration option. Use ► (F3) to toggle through pre-programmed available settings per configuration field. Use Exit (F6) to return to the Real-Time ECG View. Any changes you have made will be saved.

    The following chart summarizes the configuration screens and the available options for each parameter.

  • ELI 250/210__________________________________________________________________________

    3-20

    Summary of Configuration Menus

    Configuration Parameter

    Definition Configuration Screen

    Software Version The firmware version of the unit Screen One

    Cart Number User-defined (4 digits) Screen One

    Site Number User-defined (4 digits) Screen One

    Language Firmware language availability Screen One

    Volume Range from 0 – 8 Screen One

    Battery Timeout 10min, 30min, 60min Screen One

    AC Filter None, 60Hz, 50Hz Screen One

    Paper Speed 25mm/sec or 50 mm/sec Screen One

    Filter 40Hz, 150Hz, or 300Hz Screen One

    Height/Weight Units lb/in or kg/cm Screen One

    Date Format US (mm/dd/yyyy) or European (dd.mm.yyyy) Screen Two

    Plot Format Channels printed: 3+1, 6, 3+3, 12, 6+6, Cabrera or

    Standard Screen Two

    Interpretation Yes or No Screen Two

    Reasons Yes or No Screen Two

    Append to ECG Unconfirmed, Reviewed by, Blank Screen Two

    # Copies 0 - 9 Screen Two

    Use A4 paper Yes or No Screen Two

  • ____________________________________________________________________________Section 3

    3-21

    Configuration Parameter

    Definition Configuration Screen

    Rhythm format 3 channel, 6 channel or 12 channel Screen Three

    3 Rhythm Lead 1 V1-V6, I, II, III, aVR, aVL, aVF Screen Three

    3 Rhythm Lead 2 V1-V6, I, II, III, aVR, aVL, aVF Screen Three

    3 Rhythm Lead 3 V1-V6, I, II, III, aVR, aVL, aVF Screen Three

    6 Rhythm Lead 1 V1-V6, I, II, III, aVR, aVL, aVF Screen Three

    6 Rhythm Lead 2 V1-V6, I, II, III, aVR, aVL, aVF Screen Three

    6 Rhythm Lead 3 V1-V6, I, II, III, aVR, aVL, aVF Screen Three

    6 Rhythm Lead 4 V1-V6, I, II, III, aVR, aVL, aVF Screen Three

    6 Rhythm Lead 5 V1-V6, I, II, III, aVR, aVL, aVF Screen Three

    6 Rhythm Lead 6 V1-V6, I, II, III, aVR, aVL, aVF Screen Three

    3 + 1 Rhythm Lead V1-V6, I, II, III, aVR, aVL, aVF Screen Four

    3 + 3 Rhythm Lead 1 V1-V6, I, II, III, aVR, aVL, aVF Screen Four

    3 + 3 Rhythm Lead 2 V1-V6, I, II, III, aVR, aVL, aVF Screen Four

    3 + 3 Rhythm Lead 3 V1-V6, I, II, III, aVR, aVL, aVF Screen Four

  • ELI 250/210__________________________________________________________________________

    3-22

    Configuration Page 1

    Software Version The displayed number identifies the firmware version of your electrocardiograph.

    Cart Number

    Cart numbers indicate which electrocardiograph transmitted a particular ECG. You can use up to four digits for the

    cart number.

    Site Number This option identifies the site of your ELI 210. You can use up to four digits for the site number.

    Language

    There are several languages available on the cardiograph.

    CAUTION: upon selecting a new language and exiting the configuration screen, function labels are

    immediately translated.

    If an unknown language is visible, use the following steps to revert to the language of your country:

    7. F6 from Real-Time ECG View 8. Select number 3 9. Simultaneously press ▲ (F1) and ▼ (F2). 10. Press F2 (▼) two times 11. Press F3 (►) until the desired language appears 12. F6 to return to Real-Time ECG View

    Volume This option defines the keyboard click loudness. Available settings rage from 0 (off) to 8 (loud).

    Battery Timeout

  • ____________________________________________________________________________Section 3

    3-23

    This setting determines when the cardiograph will switch off in order to conserve the battery life of the unit. The

    battery timeout will only occur if the keypad has not been depressed for the time specified. This setting will be

    ignored while the unit is transmitting, rhythm printing or detects an ECG input signal.

    AC Filter

    The ELI 210 removes 60Hz or 50Hz interference. The setting you select depends on the line frequency in your

    country. Always use the 60Hz setting in the United States. If the AC interference is present, check to see that the

    proper AC filter is selected.

    Paper Speed

    The writer speed can be configured at 25mm/s or 50mm/s for default ECG printouts. For rhythm printouts and

    display, speeds of 5mm/s or 10mm/s are also available. See Section 4 to change speeds for display or rhythm

    printing.

    Filter The ECG plot frequency filter (or print filter) can be set to 40Hz, 150Hz, or 300Hz. The plot frequency filter does

    not filter the acquired digital record. A 40Hz plot filter setting will reduce the noise (40Hz and higher frequencies)

    on the printed ECG, a 150Hz plot filter setting will reduce the noise (150Hz and higher frequencies) on the printout

    and a 300Hz plot filter setting will not filter the printed ECG.

    Height/Weight Units

    This option defines the units of weight and height to either pounds/inches (lb/in) or kilograms/centimeters (kg/cm).

  • ELI 250/210__________________________________________________________________________

    3-24

    Configuration Page 2

    Date Format This option defines the format for entering and displaying the patient’s date of birth in either the U.S. format or the

    European format.

    For example: July 16th

    , 2003 is displayed as follows: US Date Format: MM/DD/YYYY (07/16/2003) European Date Format: DD.MM.YYYY (16.07.2003)

    NOTE: the date format option does not modify the acquisition date printed on each ECG. Using the

    example above, the acquisition date would print as 16-Jul-2003.

    Plot Format This option defines the default for one of five available plot formats in either Standard or Cabrera presentation.

    Please note that regardless of the plot format selected, 10 seconds of 12-leads are always stored and printed.

    The ECG plot options are:

    Format Option

    ECG Data

    3+1

    2.5 seconds of 12 leads in a three-channel format and the fourth channel is a 10 second rhythm strip of lead II (user defined).

    6

    5 seconds of 12-leads in a six-channel format

    3+3

    2.5 seconds of 12 leads in a three-channel format plus 10 second rhythm strip of leads II, V1 and V5 (user defined) in a three-channel format.

  • ____________________________________________________________________________Section 3

    3-25

    12

    10 seconds of 12 leads in a 12 channel format

    6+6

    10 seconds of 12 leads in a 2 page printout (six channels per page)

    Interpretation Option

    The ELI 210 automatically analyzes ECGs and prints the optional interpretation on the ECG printout. This option

    allows you to select or suppress the “interpretive” text on the ECG report.

    NOTE: A qualified physician should review the computer generated ECG interpretation before the

    treatment of any patient.

    Reasons

    The reason statements indicate why a particular interpretive statement was printed. Reason statements print in

    [square brackets] within the interpretive text (if the interpretation is turned on). Turning the reason statement

    function on or off does not affect the measurements performed or the interpretive statements selected by the analysis

    program. See Appendix B for examples of the ECG reason statements on the sample printouts.

    For Example:

    Anteroseptal Infarct [40+ MS Q WAVE IN V1-V4]

    Where “Anteroseptal Infarct” is the interpretive statement.

    And “40+ MS Q WAVE IN V1-V4” is the reason statement or explanation as to why the

    interpretive statement was printed.

    Append to ECG

    A status or statement phrase can be appended to the ECG and printed under the interpretive text printout. Either

    “unconfirmed” or “reviewed by” can be selected. However, if you wish to have nothing appended to the ECG,

    blank may be selected.

    Number of Copies

    This option defines the number of printed copies when an ECG is taken. A zero (0) setting prints the original only;

    one (1) prints the original plus 1 copy; two (2) prints the original plus 2 copies, and so on. Up to 9 copies may be

    selected.

    Use A4 Paper

    The ELI 210 accommodates use of Z-fold thermal paper in either the letter size (8.5 x 11 inches; 216 x 279 mm) or

    the “A4” size (8.27 x 11.69 inches; 210 x 297 mm). The provided paper tray spacer is required for use with A4 size

    paper. Enable this configuration option and printing on A4 paper is possible. The paper tray spacer should not be

    inserted if the letter size paper is used.

    NOTE: Paper tray spacer must be inserted for use with A4 size paper.

  • ELI 250/210__________________________________________________________________________

    3-26

    Configuration Page 3

    Rhythm Formats

    Configuration page four defines the default values for rhythm printing. It is possible to set a 3-channel, 6-channel or

    12-channel default rhythm format. Define rhythm leads one through three to customize a three-channel rhythm

    printout or define rhythm leads one through six to customize the six-channel rhythm printout. Use (F3) to toggle through each of the 12 leads.

    Rhythm Leads Configuration page five identifies the user-selectable 10-second rhythm lead printed in a 3 + 1 channel ECG printout

    and the three 10-second rhythm leads for the 3+3 channel ECG printout. Use (F3) to make your selection of any of the 12 leads.

    NOTE: rhythm acquisition is not stored in memory, only printed

  • ___________________________________________________________________________ Section 4

    4-1

    4 Assembly/Disassembly of Unit

    CAUTION:

    Remove AC power and battery fuse before attempting any disassembly.

    Removal of the battery fuse will not cause a loss of ECG’s or of the configuration. This

    information is stored in FLASH memory.

    For illustration purposes the ELI 250 will be used to demonstrate the disassembly of the unit. The disassembly of

    the ELI 210 is identical except for the keyboard components. The parts list located on pages 5-6 and 5-7 of this

    manual.

    Fuse Removal

    Locate the fuse on the bottom of the unit.

    Insert a flat blade screwdriver into the slot of the fuse holder, push in slightly and rotate the screwdriver

    counterclockwise 1/4 turn. The fuse holder is spring loaded and will come out easily.

  • ELI 250/210 _________________________________________________________________________

    4-2

    Cover Assembly Removal / Installation

    From the standard operating position:

    Picture shown is ELI 250

  • ___________________________________________________________________________ Section 4

    4-3

    Flip unit over so that it is resting on the writer. Locate the mounting screws for the top cover. These are the screws

    around the perimeter of the unit and marked with arrows in the plastic.

    Loosen these screws, item number 14 in parts list, using a Philips head screwdriver.

    Note: When removing screws, loosen the screw and do not remove, the screw can remain in the screw hole

    by placing tape over the holes with the screw still in the hole. The tape will hold the screw so that it

    will not be lost during disassembly.

    Flip the unit upright so that it is in its normal position.

    CAUTION

    The unit will be in three pieces, the Top, the Bottom and the Connector Bulkhead. Care

    must be taken when flipping the unit over. Hold the top cover to the bottom of the unit

    securely when flipping unit over.

  • ELI 250/210 _________________________________________________________________________

    4-4

    Open the writer door approximately 4 to 5 inches.

  • ___________________________________________________________________________ Section 4

    4-5

    Lift up and rotate the right side of the top cover to left a small distance. Rest the cover on the top of the writer this

    will allow you to remove the Keyboard/LCD cable, item number 34, from the motherboard.

    Note: Slide your hand into the unit and disconnect the Keyboard/LCD, item 34, ribbon cable from the

    motherboard. This cable may be inserted into a quick-disconnect connector. Flip the latches of the

    connector and the cable will release. Not all motherboards will have the latches to secure the keyboard

    cable.

  • ELI 250/210 _________________________________________________________________________

    4-6

    Slide top cover off the unit, rotating as needed to clear the writer door.

    Set the cover to the side.

    Re-assemble in reverse order.

  • ___________________________________________________________________________ Section 4

    4-7

    Battery Removal Once the top cover is removed, the battery can be accessed for replacement if needed. Make sure that the battery

    fuse has been removed.

    Locate the battery, item number 11, near the front of the unit.

    Slide the cable connections off of the battery.

    The battery is attached to the unit using double-sided tape, using a flat blade screwdriver gently pry the battery

    loose from the bottom plastic chassis.

    Install the new battery in the same location using double sided tape, item number 49, and reconnect the battery

    cables.

    Replace the cover and install the battery fuse.

    Keyboard/LCD Removal With the top cover removed flip the cover over to access the inside of the top cover and locate the mounting screws

    for the keyboard and LCD.

    Note: 5 of the mounting screws are item number 13. The remaining screw, at the noted location is item 44

    and is shorter than the other screws. This becomes important when reinstalling the keyboard

    assembly.

    Remove the mounting screws from the underside of the cover.

    The keyboard assembly can then be removed from the top cover. The picture below shows the keyboard assembly

    removed from the top cover and the Keyboard/LCD cable, item 34, removed.

  • ELI 250/210 _________________________________________________________________________

    4-8

    Remove the LCD cable from the keyboard by 1) sliding the connector to the right then 2) lifting up on the

    connector.

  • ___________________________________________________________________________ Section 4

    4-9

    The keyboard circuit board, item number 3 for the ELI 250 and item number 4 for the ELI 210, can then be lifted

    off the assembly.

    The silicon key pad, item 9 for the ELI 250 and item 10 for the ELI 210, can now be removed from the keyboard

    bezel, item 22 for the ELI 250 and item 23 for the ELI 210, along with the LCD.

    The LCD is attached to the keyboard bezel by means of double-sided tape, item number 50, near the top of the

    LCD. Care should be taken when removing the LCD from the bezel.

  • ELI 250/210 _________________________________________________________________________

    4-10

    When replacing the LCD, remove the old double sided tape and cut a piece of double-sided tape approximately 3

    inches. Place this strip of double-sided tape, item number 50, onto mounting surface on Keyboard Bezel as shown.

    For the installation of a new LCD, item number 12, cut four pieces of foam tape, item number 51, to approximately

    2 inches. Place two strips of foam tape onto LCD as shown. Place the remaining two strips over the first two strips

    as shown. Leave the paper backing on the foam tape

    Reinstall Keyboard and LCD assembly in reverse order.

  • ___________________________________________________________________________ Section 4

    4-11

    Writer Removal / Installation

    Located on the perimeter of the motherboard are the cables for the writer. There are a total of 5 cables that need to

    be removed, 2 print head cables, 1 cue sensor cable, 1 print head ground cable and the writer motor cable.

    Remove the listed cables.

  • ELI 250/210 _________________________________________________________________________

    4-12

    Flip the unit over and remove the 4 screws, item 14, holding the writer to the bottom chassis.

    CAUTION

    Due to the weight of the battery, the bottom chassis of the unit may tilt off the writer while

    the unit is upside down. Make sure that the bottom chassis is supported while performing

    this procedure.

    Lift the bottom chassis straight off of the writer assembly.

    Re-assemble in reverse order.

  • ___________________________________________________________________________ Section 4

    4-13

    Printhead Removal / Installation

    CAUTION

    The printhead is a static sensitive part. Proper measures should be taken when

    handling this part.

    Note: The writer does not need to be removed from the unit for printhead to be replaced however removal of the

    writer from the chassis will make the job easier.

    With the top cover of the unit removed, close the writer door and remove the printhead cables from the

    motherboard. See section on removing the cover.

    Locate the O-Ring, on the underside of the printhead assembly, that secures the printhead assembly to the writer

    chassis and remove.

    CAUTION

    When the printhead spring bar is removed the printhead assembly will be loose. Care

    should be taken if this procedure is being performed with the writer assembly out of

    the unit and the writer assembly is upside down.

    View from bottom of writer assembly

    Open the writer door and remove the printhead spring bar.

  • ELI 250/210 _________________________________________________________________________

    4-14

    Carefully guide the printhead assembly out of the writer chassis making sure that no damage occurs to the cables

    attached to the printhead.

    Remove the 4 screws, item number52, that attach the printhead to the Printhead plastic support.

    Note: These screws are coated with Vibratite They are then installed tight and backed off one quarter turn to allow

    for expansion and contraction between the printhead and the plastic housing.

    Remove the cables attached to the printhead. Please note that on early units the cables are secured to the print head

    with adhesive. If the printhead or the cables require replacing the complete assembly should be replaced.

    Re-assemble in reverse order.

  • ___________________________________________________________________________ Section 4

    4-15

    Writer Motor Removal / Installation

    Flip the writer assembly over and locate the writer motor and gear assembly.

    Remove the three screws that hold this assembly to the writer assembly, item number 53.

  • ELI 250/210 _________________________________________________________________________

    4-16

    To remove the writer motor remove the two screws, item number 45, holding it to the gear assembly bracket. The

    writer motor will slide through the assembly.

    Use a 0.05 inch or a 1.25 mm size Allen wrench to remove the gear from the writer motor.

    Re-assemble in reverse order.

    Note: The screws holding the writer motor to the gear assembly bracket are coated with Vibra-tite to reduce the

    possibility of the screws vibrating loose. You may need to re-coat the screws to prevent vibration from loosening the

    screws.

  • ___________________________________________________________________________ Section 4

    4-17

    Printed Circuit Board Assembly Removal / Installation

    CAUTION

    The Motherboard is a static sensitive part. Proper measures should be taken when

    handling this part.

    When replacing the motherboard of the ELI 250/210 the configuration will need to be

    entered o