Page 1: Electron therapy Class 2: Review questions - UT-H GSBS … … ·  · 2015-02-052 Raphex Question: T63, 2002 • In what situation is electron backscatter likely to be a problem?

1

Electron therapyElectron therapyClass 2: Review questionsClass 2: Review questions

Page 2: Electron therapy Class 2: Review questions - UT-H GSBS … … ·  · 2015-02-052 Raphex Question: T63, 2002 • In what situation is electron backscatter likely to be a problem?

2

Raphex Question: T63, 2002

• In what situation is electron backscatter likely to be a problem?

A. Using 1cm of tissue equivalent bolus on the skin.B. Using a lead intra-oral shield to protect the tongue

when treating a cheek.C. When electrons are incident at an angel greater than

30o.D. When electrons of different energies are matched on

the skin.

•• In what situation is electron backscatter likely to In what situation is electron backscatter likely to be a problem?be a problem?

A.A. Using 1cm of tissue equivalent bolus on the skin.Using 1cm of tissue equivalent bolus on the skin.B.B. Using a lead intraUsing a lead intra--oral shield to protect the tongue oral shield to protect the tongue

when treating a cheek.when treating a cheek.C.C. When electrons are incident at an angel greater than When electrons are incident at an angel greater than

3030oo..D.D. When electrons of different energies are matched on When electrons of different energies are matched on

the skin.the skin.

Page 3: Electron therapy Class 2: Review questions - UT-H GSBS … … ·  · 2015-02-052 Raphex Question: T63, 2002 • In what situation is electron backscatter likely to be a problem?

3

Raphex Question: T63, 2000

• The %DD at 5 cm for a 6MeV electron beam is approximately:

A. 100%B. 90%C. 80%D. 50%E. <5%

•• The %DD at 5 cm for a 6MeV electron beam is The %DD at 5 cm for a 6MeV electron beam is approximately:approximately:

A.A. 100%100%B.B. 90%90%C.C. 80%80%D.D. 50%50%E.E. <5%<5%

Page 4: Electron therapy Class 2: Review questions - UT-H GSBS … … ·  · 2015-02-052 Raphex Question: T63, 2002 • In what situation is electron backscatter likely to be a problem?

4

Raphex Question: T55, 1999

• The internal mammary nodes (IMN) can be included in wide tangents or treated with a direct beam. In the latter case, a combination of photons and electrons can be used. The reason for adding photons, instead of using electrons alone, is:

A. To reduce dose to underlying lung.B. To keep cord dose below tolerance.C. To reduce the skin dose.D. All of the above.

•• The internal mammary nodes (IMN) can be The internal mammary nodes (IMN) can be included in wide tangents or treated with a direct included in wide tangents or treated with a direct beam. In the latter case, a combination of photons beam. In the latter case, a combination of photons and electrons can be used. The reason for adding and electrons can be used. The reason for adding photons, instead of using electrons alone, is:photons, instead of using electrons alone, is:

A.A. To reduce dose to underlying lung.To reduce dose to underlying lung.B.B. To keep cord dose below tolerance.To keep cord dose below tolerance.C.C. To reduce the skin dose.To reduce the skin dose.D.D. All of the above.All of the above.

Page 5: Electron therapy Class 2: Review questions - UT-H GSBS … … ·  · 2015-02-052 Raphex Question: T63, 2002 • In what situation is electron backscatter likely to be a problem?

5

Raphex Question: T62, 2002

• A lateral neck tumor is treated with a single direct electron field. The max depth of the treatment volume is 1.5cm, and the minimum cord depth is 5.0cm. Two techniques are considered: 6MeV electrons 9MeV electrons with 0.5cm bolus

Which of the following is TRUE?A. The cord dose will be at least 50% of prescribed dose in either

case.B. The surface dose will be higher with the 6MeV Beam.C. The dose falloff beyond the tumor will be steeper with 6MeVD. The tumor would be underdosed with 9MeV

•• A lateral neck tumor is treated with a single direct electron A lateral neck tumor is treated with a single direct electron field. The max depth of the treatment volume is 1.5cm, field. The max depth of the treatment volume is 1.5cm, and the minimum cord depth is 5.0cm. Two techniques and the minimum cord depth is 5.0cm. Two techniques are considered:are considered: 6MeV electrons6MeV electrons 9MeV electrons with 0.5cm bolus9MeV electrons with 0.5cm bolus

Which of the following is TRUE?Which of the following is TRUE?A.A. The cord dose will be at least 50% of prescribed dose in either The cord dose will be at least 50% of prescribed dose in either

case.case.B.B. The surface dose will be higher with the 6MeV Beam.The surface dose will be higher with the 6MeV Beam.C.C. The dose falloff beyond the tumor will be steeper with 6MeVThe dose falloff beyond the tumor will be steeper with 6MeVD.D. The tumor would be The tumor would be underdosedunderdosed with 9MeVwith 9MeV

Page 6: Electron therapy Class 2: Review questions - UT-H GSBS … … ·  · 2015-02-052 Raphex Question: T63, 2002 • In what situation is electron backscatter likely to be a problem?

6

Raphex Question: T64, 2000

• Which of the following properties of electron beams are true?

1. The range in tissue in cm is about is about ½ the beam energy in MeV.

2. The distance between the 90% and 20% isodose levels on the axis increases with increasing energy.

3. The width of the 90% isodose decreases with depth.4. As energy increases, skin dose decreases.

A. 1,2,3B. 1,3C. 4 onlyD. All of the above

• Which of the following properties of electron beams are true?

1. The range in tissue in cm is about is about ½ the beam energy in MeV.

2. The distance between the 90% and 20% isodose levels on the axis increases with increasing energy.

3. The width of the 90% isodose decreases with depth.4. As energy increases, skin dose decreases.

A. 1,2,3B. 1,3C. 4 onlyD. All of the above

Page 7: Electron therapy Class 2: Review questions - UT-H GSBS … … ·  · 2015-02-052 Raphex Question: T63, 2002 • In what situation is electron backscatter likely to be a problem?

7

Raphex Question: T46, 2001

• Compared with 6MeV electrons, 16MeV electrons have:

1. A greater surface dose.2. A lower bremsstrahlung tail.3. A broader plateau region.4. A sharper fall-off between 80% and 20% isodose levels

A. 1,3B. 2,4C. 1,2,3D. 4 onlyE. 1,2,3,4

•• Compared with 6MeV electrons, 16MeV electrons have:Compared with 6MeV electrons, 16MeV electrons have:

1.1. A greater surface dose.A greater surface dose.2.2. A lower A lower bremsstrahlungbremsstrahlung tail.tail.3.3. A broader plateau region.A broader plateau region.4.4. A sharper fallA sharper fall--off between 80% and 20% off between 80% and 20% isodoseisodose levelslevels

A.A. 1,31,3B.B. 2,42,4C.C. 1,2,31,2,3D.D. 4 only4 onlyE.E. 1,2,3,41,2,3,4

Page 8: Electron therapy Class 2: Review questions - UT-H GSBS … … ·  · 2015-02-052 Raphex Question: T63, 2002 • In what situation is electron backscatter likely to be a problem?

8

Raphex Question: T57, 2002

• A 12 MeV electron beam has a range of____cm, and a 90% DD at approximately_____cm.

A. 6,4B. 12,4C. 12,6D. 4,3E. 9,6

•• A 12 A 12 MeVMeV electron beam has a range electron beam has a range of____cmof____cm, , and a 90% DD at and a 90% DD at approximately_____cmapproximately_____cm..

A.A. 6,46,4B.B. 12,412,4C.C. 12,612,6D.D. 4,34,3E.E. 9,69,6

Page 9: Electron therapy Class 2: Review questions - UT-H GSBS … … ·  · 2015-02-052 Raphex Question: T63, 2002 • In what situation is electron backscatter likely to be a problem?

9

Raphex Question: T69, 2003

• When a custom electron insert has dimensions smaller than the range of the electrons, all of the following are likely to occur except:

A. The output (cGy/MU) will be reducedB. Surface dose will decrease.C. PDD will decrease beyond dmax.D. Dmax will shift toward a shallower depth.

•• When a custom electron insert has dimensions When a custom electron insert has dimensions smaller than the range of the electrons, all of smaller than the range of the electrons, all of the following are likely to occur except:the following are likely to occur except:

A.A. The output (The output (cGycGy/MU) will be reduced/MU) will be reducedB.B. Surface dose will decrease.Surface dose will decrease.C.C. PDD will decrease beyond PDD will decrease beyond ddmaxmax..D.D. DmaxDmax will shift toward a shallower depth.will shift toward a shallower depth.

Page 10: Electron therapy Class 2: Review questions - UT-H GSBS … … ·  · 2015-02-052 Raphex Question: T63, 2002 • In what situation is electron backscatter likely to be a problem?

10

Electron therapyElectron therapyClass 2: Fundamentals Class 2: Fundamentals

continuedcontinued

Page 11: Electron therapy Class 2: Review questions - UT-H GSBS … … ·  · 2015-02-052 Raphex Question: T63, 2002 • In what situation is electron backscatter likely to be a problem?

11

Field flatness and symmetry

• Plane perpendicular to beam axis• Uniformity index (ICRU): Adose>90% /

Ageom

• Specifies dose<103% of CAX value• Depth: e.g. 95% beyond dmax (AAPM

TG#25)• Area: 2cm inside geometrid edge of fields

larger than 10x10cm• ± 5% (optimally be within ± 3%) TG#25• Symmetry: compares dose profile on one

side of central axis to the other. Max 2% variation (AAPM)

• Flatness and symmetry change: <1% of baseline

•• Plane perpendicular to beam axisPlane perpendicular to beam axis•• Uniformity index (ICRU): Uniformity index (ICRU): AAdosedose>90%>90% / /

AAgeomgeom

•• Specifies dose<103% of CAX valueSpecifies dose<103% of CAX value•• Depth: e.g. 95% beyond Depth: e.g. 95% beyond dmaxdmax (AAPM (AAPM

TG#25)TG#25)• Area: 2cm inside geometrid edge of fields

larger than 10x10cm• ± 5% (optimally be within ± 3%) TG#25• Symmetry: compares dose profile on one

side of central axis to the other. Max 2% variation (AAPM)

• Flatness and symmetry change: <1% of baseline

Page 12: Electron therapy Class 2: Review questions - UT-H GSBS … … ·  · 2015-02-052 Raphex Question: T63, 2002 • In what situation is electron backscatter likely to be a problem?

12

MU calculation

ODDMU

%

EOFSCSEOFFSCSEO cal ,,,,

D Prescribed dose%D Prescribed %O Dose output at R100

Output depends on energy, applicator, field size, SSD and skin collimation.

First consider different cones and field sizes:

Note: All at R100, which is dependent of field size and energy

EOFSCSEOFDDMU

cal

,,%

(more on MU calcs later)

Page 13: Electron therapy Class 2: Review questions - UT-H GSBS … … ·  · 2015-02-052 Raphex Question: T63, 2002 • In what situation is electron backscatter likely to be a problem?

13

Skin blocking

• Useful for:• Small fields• Critical structures close to field• Need to restore beam penumbra (e.g. extended SSD, arc therapy)

Page 14: Electron therapy Class 2: Review questions - UT-H GSBS … … ·  · 2015-02-052 Raphex Question: T63, 2002 • In what situation is electron backscatter likely to be a problem?

14

Effect of skin collimation on dosimetry

• PDD: – mostly scatter in patient

– Use field size determined by skin collimation

• Output factor:– Mostly scatter in air

– Use field size determined by secondary collimator (e.g. cerrobend cutout)

•• PDD: PDD: –– mostly scatter in patientmostly scatter in patient

–– Use field size determined by skin collimationUse field size determined by skin collimation

•• Output factor:Output factor:–– Mostly scatter in airMostly scatter in air

–– Use field size determined by secondary Use field size determined by secondary collimator (e.g. collimator (e.g. cerrobendcerrobend cutout)cutout)

Page 15: Electron therapy Class 2: Review questions - UT-H GSBS … … ·  · 2015-02-052 Raphex Question: T63, 2002 • In what situation is electron backscatter likely to be a problem?

15

Field equivalence

• For rectangular fields:

• (Note: neglects collimator scatter)

• Equivalent circular fields (field width=2a):

•• For rectangular fields:For rectangular fields:

•• (Note: neglects collimator scatter)(Note: neglects collimator scatter)

•• Equivalent circular fields (field width=2a):Equivalent circular fields (field width=2a):

Page 16: Electron therapy Class 2: Review questions - UT-H GSBS … … ·  · 2015-02-052 Raphex Question: T63, 2002 • In what situation is electron backscatter likely to be a problem?

16

WxWLxLLxW OFOFOF

Page 17: Electron therapy Class 2: Review questions - UT-H GSBS … … ·  · 2015-02-052 Raphex Question: T63, 2002 • In what situation is electron backscatter likely to be a problem?

17

Calculating PDD for non-square fields

• Depth dose can be determined from data from square fields:

• Need to renormalize for 100% at dmax for new PDD

• Can use same formula for output calculations

•• Depth dose can be determined from data Depth dose can be determined from data from square fields:from square fields:

•• Need to renormalize for 100% at Need to renormalize for 100% at dmaxdmax for for new PDDnew PDD

•• Can use same formula for output Can use same formula for output calculationscalculations

Page 18: Electron therapy Class 2: Review questions - UT-H GSBS … … ·  · 2015-02-052 Raphex Question: T63, 2002 • In what situation is electron backscatter likely to be a problem?

18

A possible scenario:

• Skin collimation 2x5

• Cutout in cone: 4x6

• Energy: 6MeV

• What determines output factor?

• What determines PDD?

•• Skin collimation 2x5Skin collimation 2x5

•• Cutout in cone: 4x6Cutout in cone: 4x6

•• Energy: 6MeVEnergy: 6MeV

•• What determines output factor?What determines output factor?

•• What determines PDD?What determines PDD?

Page 19: Electron therapy Class 2: Review questions - UT-H GSBS … … ·  · 2015-02-052 Raphex Question: T63, 2002 • In what situation is electron backscatter likely to be a problem?

19

Raphex Question: T61, 2002

•• An electron beam with a custom insert has a An electron beam with a custom insert has a measured OF measured OF ofof 0.954cGy/MU at 0.954cGy/MU at ddmaxmax. If 200cGy . If 200cGy are prescribed to the 90% are prescribed to the 90% isodoseisodose, the MU setting , the MU setting is ___.is ___.

A.A. 117117B.B. 210210C.C. 212212D.D. 222222E.E. 233233

Page 20: Electron therapy Class 2: Review questions - UT-H GSBS … … ·  · 2015-02-052 Raphex Question: T63, 2002 • In what situation is electron backscatter likely to be a problem?

20

Bonus question

•• The patient is to be treated to 200cGy, prescribed to the 80% The patient is to be treated to 200cGy, prescribed to the 80% isodoseisodose level.level.

•• The calibrated output for the 15 cone is 1cGy/MU at The calibrated output for the 15 cone is 1cGy/MU at ddmaxmax

•• The field size is 4x12cmThe field size is 4x12cm22 (15cone). (15cone). •• The output factor for a 4x4 field in the 15 cone is 0.954The output factor for a 4x4 field in the 15 cone is 0.954•• The output factor for a 12x12 field in the 15 cone is 0.997The output factor for a 12x12 field in the 15 cone is 0.997•• The MU setting is____.The MU setting is____.

A.A. 176176B.B. 184184C.C. 218218D.D. 250250E.E. 256256

Page 21: Electron therapy Class 2: Review questions - UT-H GSBS … … ·  · 2015-02-052 Raphex Question: T63, 2002 • In what situation is electron backscatter likely to be a problem?

21

Another bonus question

• We decide to add 1cm bolus to a 6MeV electron field. Which of the following is not true (if we do not change the MUs)?

a) The position of dmax moves upstream by 1cmb) The dose at 2cm depth in the tissue goes

downc) The skin dose increases to almost dmax

d) The maximum dose increases by 2%e) The maximum dose decreases by 2%

•• We decide to add 1cm bolus to a 6MeV We decide to add 1cm bolus to a 6MeV electron field. Which of the following is electron field. Which of the following is not true (if we do not change the not true (if we do not change the MUsMUs)?)?

a)a) The position of The position of dmaxdmax moves upstream by 1cmmoves upstream by 1cmb)b) The dose at 2cm depth in the tissue goes The dose at 2cm depth in the tissue goes

downdownc)c) The skin dose increases to almost The skin dose increases to almost ddmaxmax

d)d) The maximum dose increases by 2%The maximum dose increases by 2%e)e) The maximum dose decreases by 2%The maximum dose decreases by 2%

Page 22: Electron therapy Class 2: Review questions - UT-H GSBS … … ·  · 2015-02-052 Raphex Question: T63, 2002 • In what situation is electron backscatter likely to be a problem?

22

Raphex Question: T58, 1999

• Output for an electron cones depends on:

1. Cone size.2. Size of cut-out.3. Beam Energy.4. SSD

A. 1,2,3B. 1,3C. 2,4D. 4 onlyE. All of the above.

• Correct Answer: E. Output for electron beam depends on ALL of these factors.

•• Output for an electron cones depends on:Output for an electron cones depends on:

1.1. Cone size.Cone size.2.2. Size of cutSize of cut--out.out.3.3. Beam Energy.Beam Energy.4.4. SSDSSD

A.A. 1,2,31,2,3B.B. 1,31,3C.C. 2,42,4D.D. 4 only4 onlyE.E. All of the above.All of the above.

•• Correct Answer:Correct Answer: E. Output for electron beam depends E. Output for electron beam depends on ALL of these factors.on ALL of these factors.

Page 23: Electron therapy Class 2: Review questions - UT-H GSBS … … ·  · 2015-02-052 Raphex Question: T63, 2002 • In what situation is electron backscatter likely to be a problem?

23

Electron therapyClass 2: Algorithms

Acknowledgements• Many images, data and many slides are from Richard Popple, Ph.D., Department of Radiation Oncology, The University of Alabama at Birmingham, Birmingham, Alabama• Paul Yokoyama (Varian) was also very helpful

AcknowledgementsAcknowledgements•• Many images, data and many slides are from Richard Many images, data and many slides are from Richard PopplePopple, , Ph.D., Department of Radiation Oncology, The University of Ph.D., Department of Radiation Oncology, The University of Alabama at Birmingham, Birmingham, AlabamaAlabama at Birmingham, Birmingham, Alabama•• Paul Yokoyama (Varian) was also very helpfulPaul Yokoyama (Varian) was also very helpful

Laurence Court

Page 24: Electron therapy Class 2: Review questions - UT-H GSBS … … ·  · 2015-02-052 Raphex Question: T63, 2002 • In what situation is electron backscatter likely to be a problem?

24

Pencil beam algorithm

Page 25: Electron therapy Class 2: Review questions - UT-H GSBS … … ·  · 2015-02-052 Raphex Question: T63, 2002 • In what situation is electron backscatter likely to be a problem?

25

Schematic of Hogstrom algorithm

Page 26: Electron therapy Class 2: Review questions - UT-H GSBS … … ·  · 2015-02-052 Raphex Question: T63, 2002 • In what situation is electron backscatter likely to be a problem?

26

Heterogeneity correction in original Hogstrom pencil-beam algorithm

Slab approach to heterogeneity correction

Page 27: Electron therapy Class 2: Review questions - UT-H GSBS … … ·  · 2015-02-052 Raphex Question: T63, 2002 • In what situation is electron backscatter likely to be a problem?

27

Redefinition pencil-beam algorithm

Page 28: Electron therapy Class 2: Review questions - UT-H GSBS … … ·  · 2015-02-052 Raphex Question: T63, 2002 • In what situation is electron backscatter likely to be a problem?

28

Experimental evaluation

• Dose measured using watertank and diode• Dose measured using watertank and diode

Boyd et al, Med Phys 28, 950-958, 2001

Page 29: Electron therapy Class 2: Review questions - UT-H GSBS … … ·  · 2015-02-052 Raphex Question: T63, 2002 • In what situation is electron backscatter likely to be a problem?

29

20-MeV Horizontal Bone Slab

The University of Texas M. D. Anderson Cancer Center Department of Radiation Physics

Varian Clinac 2100, 15x15Varian Clinac 2100, 15x15--cmcm22 open applicator, 100 cm SSDopen applicator, 100 cm SSD

Off-Axis Position (cm)-10 -8 -6 -4 -2 0 2 4 6 8 10

Dep

th (c

m)

0123456789

101112

MeasuredPBRA

10%

100%

Criteria not met

Page 30: Electron therapy Class 2: Review questions - UT-H GSBS … … ·  · 2015-02-052 Raphex Question: T63, 2002 • In what situation is electron backscatter likely to be a problem?

30

Off-Axis Position (cm)-10 -8 -6 -4 -2 0 2 4 6 8 10

Dep

th (c

m)

0123456789

101112

MeasuredPBRA

10%

100%

Criteria not met

20-MeV Horizontal Air Slab

The University of Texas M. D. Anderson Cancer Center Department of Radiation Physics

Varian Clinac 2100, 15x15Varian Clinac 2100, 15x15--cmcm22 open applicator, 100 cm SSDopen applicator, 100 cm SSD

Page 31: Electron therapy Class 2: Review questions - UT-H GSBS … … ·  · 2015-02-052 Raphex Question: T63, 2002 • In what situation is electron backscatter likely to be a problem?

31

Parotid Gland - Transverse View

The University of Texas M. D. Anderson Cancer Center Department of Radiation Physics

Varian 2100, 16 MeV, 15x15Varian 2100, 16 MeV, 15x15--cmcm22 applicator, 100 cm SSDapplicator, 100 cm SSD

Page 32: Electron therapy Class 2: Review questions - UT-H GSBS … … ·  · 2015-02-052 Raphex Question: T63, 2002 • In what situation is electron backscatter likely to be a problem?

32

Parotid Gland - Transverse View

The University of Texas M. D. Anderson Cancer Center Department of Radiation Physics

Varian 2100, 16 MeV, 15x15Varian 2100, 16 MeV, 15x15--cmcm22 applicator, 100 cm SSDapplicator, 100 cm SSD

X Axis

-16 -14 -12 -10 -8 -6 -4 -2 0 2 4 6

Y A

xis

-4

-2

0

2

4

6

8

10

X Axis

-16 -14 -12 -10 -8 -6 -4 -2 0 2 4 6

Y A

xis

-4

-2

0

2

4

6

8

10

Monte CarloPBRA

10%

90%

AIR

LUNGTISSUE

BONE

Page 33: Electron therapy Class 2: Review questions - UT-H GSBS … … ·  · 2015-02-052 Raphex Question: T63, 2002 • In what situation is electron backscatter likely to be a problem?

33

IMC - Transverse Plane

Varian 2100, 16 MeV, 15x15-cm2 applicator, 105 cm SSDVarian 2100, 16 MeV, 15x15Varian 2100, 16 MeV, 15x15--cmcm22 applicator, 105 cm SSDapplicator, 105 cm SSD

The University of Texas M. D. Anderson Cancer Center Department of Radiation Physics

Page 34: Electron therapy Class 2: Review questions - UT-H GSBS … … ·  · 2015-02-052 Raphex Question: T63, 2002 • In what situation is electron backscatter likely to be a problem?

34

IMC - Transverse Plane

Varian 2100, 16 MeV, 15x15-cm2 applicator, 105 cm SSDVarian 2100, 16 MeV, 15x15Varian 2100, 16 MeV, 15x15--cmcm22 applicator, 105 cm SSDapplicator, 105 cm SSD

The University of Texas M. D. Anderson Cancer Center Department of Radiation Physics

AIR

LUNGTISSUE

BONE

X Axis

-4 -2 0 2 4 6 8 10

Y Ax

is

4

6

8

10

12

14

16

X Axis

-4 -2 0 2 4 6 8 10

Y Ax

is

4

6

8

10

12

14

16

Monte CarloPBRA

10%

50%

100%

Page 35: Electron therapy Class 2: Review questions - UT-H GSBS … … ·  · 2015-02-052 Raphex Question: T63, 2002 • In what situation is electron backscatter likely to be a problem?

35

Electron Monte Carlo dose calculations

• Accurate electron dose calculations (distributions) are very difficult

• Monte Carlo has potential to be most accurate method

• MC is very calculation intensive:– 2% SD at peak dose for 10x10cm field, 0.5cm

resolution requires 1,000,000 histories.

• Macro Monte Carlo (MMC) significantly reduces calculation time

•• Accurate electron dose calculations Accurate electron dose calculations (distributions) are very difficult(distributions) are very difficult

•• Monte Carlo has potential to be most Monte Carlo has potential to be most accurate methodaccurate method

•• MC is very calculation intensive:MC is very calculation intensive:–– 2% SD at peak dose for 10x10cm field, 0.5cm 2% SD at peak dose for 10x10cm field, 0.5cm

resolution requires 1,000,000 histories.resolution requires 1,000,000 histories.

•• Macro Monte Carlo (MMC) significantly Macro Monte Carlo (MMC) significantly reduces calculation timereduces calculation time

Page 36: Electron therapy Class 2: Review questions - UT-H GSBS … … ·  · 2015-02-052 Raphex Question: T63, 2002 • In what situation is electron backscatter likely to be a problem?

36

Primary Particle Transport – PDFs

‘Kugel’(!)

Precalculate PDFs for various Kugel sizes, materials

Page 37: Electron therapy Class 2: Review questions - UT-H GSBS … … ·  · 2015-02-052 Raphex Question: T63, 2002 • In what situation is electron backscatter likely to be a problem?

37

Primary Particle Transport

Page 38: Electron therapy Class 2: Review questions - UT-H GSBS … … ·  · 2015-02-052 Raphex Question: T63, 2002 • In what situation is electron backscatter likely to be a problem?

38

Eclipse implementation of MMC

1. Initial phase space model

2. Local simulation

3. Geometric pre-processing

4. Global simulation

1.1. Initial phase space modelInitial phase space model

2.2. Local simulationLocal simulation

3.3. Geometric preGeometric pre--processingprocessing

4.4. Global simulationGlobal simulation

Page 39: Electron therapy Class 2: Review questions - UT-H GSBS … … ·  · 2015-02-052 Raphex Question: T63, 2002 • In what situation is electron backscatter likely to be a problem?

39

(1) Initial Phase Space Model

• Four-source model

• Compute probability distributions of position, energy, direction of electrons and photons for each source, specificed at PSP

•• FourFour--source modelsource model

•• Compute probability Compute probability distributions of position, distributions of position, energy, direction of energy, direction of electrons and photons for electrons and photons for each source, each source, specificedspecificed at at PSPPSP

Page 40: Electron therapy Class 2: Review questions - UT-H GSBS … … ·  · 2015-02-052 Raphex Question: T63, 2002 • In what situation is electron backscatter likely to be a problem?

40

(2) Local Simulation

• Over 200,000 electrons transported through uniform spheres:– 0.05cm, 0.1cm, 0.15cm, 0.2cm, 0.3cm

– Air, lung, water, bone

– 0.2MeV – 25MeV (25 electron energies)

• Gives PDF representing direction and energy of exiting particles, and also secondary particules

•• Over 200,000 electrons transported through Over 200,000 electrons transported through uniform spheres:uniform spheres:–– 0.05cm, 0.1cm, 0.15cm, 0.2cm, 0.3cm0.05cm, 0.1cm, 0.15cm, 0.2cm, 0.3cm

–– Air, lung, water, boneAir, lung, water, bone

–– 0.2MeV 0.2MeV –– 25MeV (25 electron energies)25MeV (25 electron energies)

•• Gives PDF representing direction and Gives PDF representing direction and energy of exiting particles, and also energy of exiting particles, and also secondary secondary particulesparticules

Page 41: Electron therapy Class 2: Review questions - UT-H GSBS … … ·  · 2015-02-052 Raphex Question: T63, 2002 • In what situation is electron backscatter likely to be a problem?

41

(3) Geometric pre-processing

Pre-processing:

• CT# to density and material conversion (user-defined resolution)

• Identify heterogeneities, and assign ‘krugel index’

• Calculate mean ‘krugel density’

PrePre--processing:processing:

•• CT# to density and material CT# to density and material conversion (userconversion (user--defined defined resolution)resolution)

•• Identify heterogeneities, and Identify heterogeneities, and assign assign ‘‘krugelkrugel indexindex’’

•• Calculate mean Calculate mean ‘‘krugelkrugel densitydensity’’

Page 42: Electron therapy Class 2: Review questions - UT-H GSBS … … ·  · 2015-02-052 Raphex Question: T63, 2002 • In what situation is electron backscatter likely to be a problem?

42

Page 43: Electron therapy Class 2: Review questions - UT-H GSBS … … ·  · 2015-02-052 Raphex Question: T63, 2002 • In what situation is electron backscatter likely to be a problem?

43

(4) Global simulation

• Position, direction ,energy of particles exiting sphere (‘kugel’!) determined by random sampling of PDF for appropriate sphere (size, material, energy)• Dose deposited along straight line inside kugel

• Secondary particules forced to interact in sphere• Always transported with 20deg angle relative to incident electron

Page 44: Electron therapy Class 2: Review questions - UT-H GSBS … … ·  · 2015-02-052 Raphex Question: T63, 2002 • In what situation is electron backscatter likely to be a problem?

44

Calculation parameters – user interaction

Page 45: Electron therapy Class 2: Review questions - UT-H GSBS … … ·  · 2015-02-052 Raphex Question: T63, 2002 • In what situation is electron backscatter likely to be a problem?

45

Calculation parameters

Page 46: Electron therapy Class 2: Review questions - UT-H GSBS … … ·  · 2015-02-052 Raphex Question: T63, 2002 • In what situation is electron backscatter likely to be a problem?

46

Evaluation data set (dose distributions, impact of heterogeneities)

• R. A. Boyd, K. R. Hogstrom, J. A. Antolak, and A. S. Shiu, “A measured data set for evaluating electron beam dose algorithms.”Med. Phys. 28, 950–958 (2001).

• Set of measurements developed and published specifically for evaluating electron dose calculation algorithms.

•• R. A. Boyd, K. R. R. A. Boyd, K. R. HogstromHogstrom, J. A. , J. A. AntolakAntolak, , and A. S. and A. S. ShiuShiu, , ““A measured data set for A measured data set for evaluating electron beam dose algorithms.evaluating electron beam dose algorithms.””Med. Phys. 28, 950Med. Phys. 28, 950––958 (2001).958 (2001).

•• Set of measurements developed and Set of measurements developed and published specifically for evaluating electron published specifically for evaluating electron dose calculation algorithms.dose calculation algorithms.

Page 47: Electron therapy Class 2: Review questions - UT-H GSBS … … ·  · 2015-02-052 Raphex Question: T63, 2002 • In what situation is electron backscatter likely to be a problem?

47

Phantom geometries

Page 48: Electron therapy Class 2: Review questions - UT-H GSBS … … ·  · 2015-02-052 Raphex Question: T63, 2002 • In what situation is electron backscatter likely to be a problem?

48

Measurement configurations

Page 49: Electron therapy Class 2: Review questions - UT-H GSBS … … ·  · 2015-02-052 Raphex Question: T63, 2002 • In what situation is electron backscatter likely to be a problem?

49

Evaluation data set

• The measured data were normalized to the maximum dose on central axis in a water phantom (without inhomogeneity) at 100-cm SSD.

• All data were for a 15 x 15 cm2 field and were acquired using a single Varian Clinac 2100C accelerator (Varian Oncology Systems, Milpitas, CA) having a Series III electron foil/applicator set.

• 9MeV, 20MeV

•• The measured data were normalized to the The measured data were normalized to the maximum dose on central axis in a water phantom maximum dose on central axis in a water phantom (without (without inhomogeneityinhomogeneity) at 100) at 100--cm SSD.cm SSD.

•• All data were for a 15 x 15 cm2 field and were All data were for a 15 x 15 cm2 field and were acquired using a single Varian acquired using a single Varian ClinacClinac 2100C 2100C accelerator (Varian Oncology Systems, Milpitas, accelerator (Varian Oncology Systems, Milpitas, CA) having a Series III electron foil/applicator set.CA) having a Series III electron foil/applicator set.

•• 9MeV, 20MeV9MeV, 20MeV

Page 50: Electron therapy Class 2: Review questions - UT-H GSBS … … ·  · 2015-02-052 Raphex Question: T63, 2002 • In what situation is electron backscatter likely to be a problem?

50

Configuration / commissioning

• Data set of Boyd et al. not sufficient to configure Eclipse MMC engine.

• Depth dose data are present (configurations 1 and 3).

• Not present in data set:– CAX fractional depth dose without applicator

– In-air profile without applicator

– Absolute output

•• Data set of Boyd et al. not sufficient to Data set of Boyd et al. not sufficient to configure Eclipse MMC engine.configure Eclipse MMC engine.

•• Depth dose data are present (configurations Depth dose data are present (configurations 1 and 3).1 and 3).

•• Not present in data set:Not present in data set:–– CAX fractional depth dose without applicatorCAX fractional depth dose without applicator

–– InIn--air profile without applicatorair profile without applicator

–– Absolute outputAbsolute output

Page 51: Electron therapy Class 2: Review questions - UT-H GSBS … … ·  · 2015-02-052 Raphex Question: T63, 2002 • In what situation is electron backscatter likely to be a problem?

51

Evaluation method

• For each configuration/calculation parameter combination, the dose difference and distance-to-agreement were computed.

• The dose difference was determined by interpolating the measured dose at the calculation points using bilinear interpolation.

•• For each configuration/calculation For each configuration/calculation parameter combination, the dose difference parameter combination, the dose difference and distanceand distance--toto--agreement were computed.agreement were computed.

•• The dose difference was determined by The dose difference was determined by interpolating the measured dose at the interpolating the measured dose at the calculation points using bilinear calculation points using bilinear interpolation.interpolation.

Page 52: Electron therapy Class 2: Review questions - UT-H GSBS … … ·  · 2015-02-052 Raphex Question: T63, 2002 • In what situation is electron backscatter likely to be a problem?

52

Results - Accuracy

• Calculations performed for all possible values of accuracy parameter (1%, 2%, 3%, 5%, and 8%).

• 1 mm and 2.5 mm grid spacing for the 9 MeV and 20 MeV configurations, respectively.

• No smoothing.

•• Calculations performed for all possible Calculations performed for all possible values of accuracy parameter (1%, 2%, 3%, values of accuracy parameter (1%, 2%, 3%, 5%, and 8%).5%, and 8%).

•• 1 mm and 2.5 mm grid spacing for the 9 1 mm and 2.5 mm grid spacing for the 9 MeVMeV and 20 and 20 MeVMeV configurations, configurations, respectively.respectively.

•• No smoothing.No smoothing.

Page 53: Electron therapy Class 2: Review questions - UT-H GSBS … … ·  · 2015-02-052 Raphex Question: T63, 2002 • In what situation is electron backscatter likely to be a problem?

53

Results - Accuracy

52.6%65.8%8.9%8%33.9%48.8%5.2%5%17.7%29.8%3.2%3%9.9%19.7%2.4%2%3.1%9.8%1.9%1%2039.3%72.4%10.5%8%22.8%54.5%5.9%5%12.4%37.0%3.7%3%6.6%25.6%3.0%2%1.5%16.5%2.4%1%9

Fraction failing 3% difference and 3 mm distance-to-

agreement

Fraction failing 3% difference

RMS differenceAccuracyEnergy (MeV)

Page 54: Electron therapy Class 2: Review questions - UT-H GSBS … … ·  · 2015-02-052 Raphex Question: T63, 2002 • In what situation is electron backscatter likely to be a problem?

54

20 MeV, 100 cm SSD (Configuration 3)1% Accuracy

Isodose lines are 10%, 30%, 70%, 90%, and 95%. Calculation points violating 3% difference and 3 mm distance-to-agreement are shown in gray.

IsodoseIsodose lines are 10%, 30%, 70%, 90%, and 95%. Calculation lines are 10%, 30%, 70%, 90%, and 95%. Calculation points violating 3% difference and 3 mm distancepoints violating 3% difference and 3 mm distance--toto--agreement are shown in gray.agreement are shown in gray.

x (mm)

Dep

th (m

m)

-100 -50 0 50 100

0

20

40

60

80

100

120

Page 55: Electron therapy Class 2: Review questions - UT-H GSBS … … ·  · 2015-02-052 Raphex Question: T63, 2002 • In what situation is electron backscatter likely to be a problem?

55

20 MeV, 100 cm SSD (Configuration 3)3% Accuracy

x (mm)

Dep

th (m

m)

-100 -50 0 50 100

0

20

40

60

80

100

120

Page 56: Electron therapy Class 2: Review questions - UT-H GSBS … … ·  · 2015-02-052 Raphex Question: T63, 2002 • In what situation is electron backscatter likely to be a problem?

56

20 MeV, nose-shape (Configuration 14)3% Accuracy

x (mm)

Dep

th (m

m)

-100 -50 0 50 100

0

20

40

60

80

100

Page 57: Electron therapy Class 2: Review questions - UT-H GSBS … … ·  · 2015-02-052 Raphex Question: T63, 2002 • In what situation is electron backscatter likely to be a problem?

57

Results – Grid spacing

• Calculations performed for all possible values of grid spacing (1 mm, 1.5 mm, 2 mm, 2.5 mm, and 5 mm).

• 1% accuracy.

• No smoothing.

•• Calculations performed for all possible Calculations performed for all possible values of grid spacing (1 mm, 1.5 mm, 2 values of grid spacing (1 mm, 1.5 mm, 2 mm, 2.5 mm, and 5 mm).mm, 2.5 mm, and 5 mm).

•• 1% accuracy.1% accuracy.

•• No smoothing.No smoothing.

Page 58: Electron therapy Class 2: Review questions - UT-H GSBS … … ·  · 2015-02-052 Raphex Question: T63, 2002 • In what situation is electron backscatter likely to be a problem?

58

Results – Grid spacing

3.4%21.8%13.5%5.0

3.1%21.5%9.8%2.5

2.9%21.1%9.3%2.0

1.5%20.2%4.6%1.5

1.6%19.6%4.5%1.0206.6%9.3%47.6%5.0

2.9%6.1%23.7%2.5

1.9%5.8%18.3%2.0

2.2%6.2%19.1%1.5

1.5%4.9%16.5%1.09

Fraction failing 3% difference

and 3 mm distance-to-agreement

Fraction failing 3 mmDistance-to-agreement

Fraction failing 3% difference

Grid spacing (mm)Energy (MeV)

Page 59: Electron therapy Class 2: Review questions - UT-H GSBS … … ·  · 2015-02-052 Raphex Question: T63, 2002 • In what situation is electron backscatter likely to be a problem?

59

9 MeV, 2 cm stepped surface (Configuration 11)1 mm grid spacing

x (mm)

Dep

th (m

m)

-100 -50 0 50 100

0

50

Page 60: Electron therapy Class 2: Review questions - UT-H GSBS … … ·  · 2015-02-052 Raphex Question: T63, 2002 • In what situation is electron backscatter likely to be a problem?

60

9 MeV, 2 cm stepped surface (Configuration 11)5 mm grid spacing

x (mm)

Dep

th (m

m)

-100 -50 0 50 100

0

50

Page 61: Electron therapy Class 2: Review questions - UT-H GSBS … … ·  · 2015-02-052 Raphex Question: T63, 2002 • In what situation is electron backscatter likely to be a problem?

61

Results – smoothing

• Calculations performed for all possible values of 3-D smoothing (Low, Medium, and Strong).

• 1% accuracy.• Grid spacing

– 1 mm, 1.5 mm, 2 mm, and 2.5 mm for the 9 MeVconfigurations.

– 1.5 mm, 2 mm, 2.5 mm, and 5 mm for the 20 MeV configurations.

•• Calculations performed for all possible Calculations performed for all possible values of 3values of 3--D smoothing (Low, Medium, D smoothing (Low, Medium, and Strong).and Strong).

•• 1% accuracy.1% accuracy.•• Grid spacingGrid spacing

–– 1 mm, 1.5 mm, 2 mm, and 2.5 mm for the 9 1 mm, 1.5 mm, 2 mm, and 2.5 mm for the 9 MeVMeVconfigurations.configurations.

–– 1.5 mm, 2 mm, 2.5 mm, and 5 mm for the 20 1.5 mm, 2 mm, 2.5 mm, and 5 mm for the 20 MeVMeV configurations.configurations.

Page 62: Electron therapy Class 2: Review questions - UT-H GSBS … … ·  · 2015-02-052 Raphex Question: T63, 2002 • In what situation is electron backscatter likely to be a problem?

62

Results – smoothing

12.5%3.4%1.1%3.4%5.0

0.6%0.5%1.3%3.1%2.5

0.4%0.3%1.5%2.9%2.0

0.3%0.5%2.0%3.5%1.520

6.6%2.8%2.9%3.7%2.5

1.3%0.5%1.2%1.9%2.0

0.2%0.5%1.5%2.2%1.5

0.1%0.2%0.7%1.5%1.09

StrongMediumLowNoneGrid spacing

(mm)Energy

Fraction failing 3% difference and 3 mm distance-to-agreement

Page 63: Electron therapy Class 2: Review questions - UT-H GSBS … … ·  · 2015-02-052 Raphex Question: T63, 2002 • In what situation is electron backscatter likely to be a problem?

63

20-MeV beam incident on a flat surface with a horizontal bone heterogeneity (configuration 9) No smoothing

x (mm)

Dep

th (m

m)

-100 -50 0 50 100

0

20

40

60

80

100

Page 64: Electron therapy Class 2: Review questions - UT-H GSBS … … ·  · 2015-02-052 Raphex Question: T63, 2002 • In what situation is electron backscatter likely to be a problem?

64

20-MeV beam incident on a flat surface with a horizontal bone heterogeneity (configuration 9) 3D Medium smoothing

x (mm)

Dep

th (m

m)

-100 -50 0 50 100

0

20

40

60

80

100

Page 65: Electron therapy Class 2: Review questions - UT-H GSBS … … ·  · 2015-02-052 Raphex Question: T63, 2002 • In what situation is electron backscatter likely to be a problem?

65

Accuracy in presence of lung

• Experimental data from University of Bern (Switzerland) or Tampere University Hospital (Finland)

• 1cm perspex over lung-equivalent material

• 6MeV Varian beam

•• Experimental data Experimental data from University of Bern from University of Bern (Switzerland) or (Switzerland) or Tampere University Tampere University Hospital (Finland)Hospital (Finland)

•• 1cm 1cm perspexperspex over lungover lung--equivalent materialequivalent material

•• 6MeV Varian beam6MeV Varian beam

Page 66: Electron therapy Class 2: Review questions - UT-H GSBS … … ·  · 2015-02-052 Raphex Question: T63, 2002 • In what situation is electron backscatter likely to be a problem?

66

Page 67: Electron therapy Class 2: Review questions - UT-H GSBS … … ·  · 2015-02-052 Raphex Question: T63, 2002 • In what situation is electron backscatter likely to be a problem?

67

Conclusion (dose distributions)

• For judicious choices of parameters, dose calculations agree with expt. to better than 3% dose difference and 3-mm distance-to-agreement.

• Lung dose is less accurate.

• Note: Varian have also shown good results for oblique incidence

• The Eclipse MMC implementation is limited to– 2.5% and 2.5 mm for 9 MeV

– 3% and 3 mm for 20 MeV

•• For judicious choices of parameters, dose For judicious choices of parameters, dose calculations agree with expt. to better than 3% calculations agree with expt. to better than 3% dose difference and 3dose difference and 3--mm distancemm distance--toto--agreement.agreement.

•• Lung dose is less accurate.Lung dose is less accurate.

•• Note: Varian have also shown good results for Note: Varian have also shown good results for oblique incidenceoblique incidence

•• The Eclipse MMC implementation is limited toThe Eclipse MMC implementation is limited to–– 2.5% and 2.5 mm for 9 2.5% and 2.5 mm for 9 MeVMeV

–– 3% and 3 mm for 20 3% and 3 mm for 20 MeVMeV

Page 68: Electron therapy Class 2: Review questions - UT-H GSBS … … ·  · 2015-02-052 Raphex Question: T63, 2002 • In what situation is electron backscatter likely to be a problem?

68

Evaluation of output factors for range of clinical fields

• Data above only for simple field shapes •Comparison of measured and calculated outputs• (output includes shape and SSD differences)

Page 69: Electron therapy Class 2: Review questions - UT-H GSBS … … ·  · 2015-02-052 Raphex Question: T63, 2002 • In what situation is electron backscatter likely to be a problem?

69

Eclipse MMC Calculation Parameters

• Accuracy (mean statistical error in dose within the high dose volume) : 1%.

• Grid spacing:

• No smoothing

•• Accuracy (mean statistical error in dose within Accuracy (mean statistical error in dose within the high dose volume) : 1%.the high dose volume) : 1%.

•• Grid spacing:Grid spacing:

•• No smoothingNo smoothing2.5 mm2.5 mm18 18 MeVMeV

2.5 mm2.5 mm15 15 MeVMeV

2 mm2 mm12 12 MeVMeV

1.5 mm1.5 mm9 9 MeVMeV

1 mm1 mm6 6 MeVMeV

Grid spacingGrid spacingEnergyEnergy

Page 70: Electron therapy Class 2: Review questions - UT-H GSBS … … ·  · 2015-02-052 Raphex Question: T63, 2002 • In what situation is electron backscatter likely to be a problem?

70

Scatter plot of calculated output factor versus measured output factor

The solid line shows calculation equal to measurement, and the dashed lines show 3% deviation.

The solid line shows calculation equal to measurement, and the dThe solid line shows calculation equal to measurement, and the dashed lines ashed lines show 3% deviation.show 3% deviation.

0.8 0.85 0.9 0.95 1 1.05 1.10.8

0.85

0.9

0.95

1

1.05

1.1

Measured output factor

Ecl

ipse

MM

C o

utpu

t fac

tor

Page 71: Electron therapy Class 2: Review questions - UT-H GSBS … … ·  · 2015-02-052 Raphex Question: T63, 2002 • In what situation is electron backscatter likely to be a problem?

71

Histogram of difference distribution

-5 -4 -3 -2 -1 0 1 2 3 4 50

10

20

30

40Mean difference: -0.23%Std. deviation: 1.09%

Percent difference

Num

ber o

f ins

erts

Page 72: Electron therapy Class 2: Review questions - UT-H GSBS … … ·  · 2015-02-052 Raphex Question: T63, 2002 • In what situation is electron backscatter likely to be a problem?

72

Conclusion (output factors)

• The mean difference between the Eclipse MMC calculation and measurement was 0.2%. The standard deviation of the difference distribution was 1.1%.

• Of the 218 measurements,– 124 (56.9%) were within 1%. – 211 (96.8%) were within 2%. – 215 (98.6%) were within 3%.

• The largest difference was 3.4%.

•• The mean difference between the Eclipse The mean difference between the Eclipse MMC calculation and measurement was 0.2%. MMC calculation and measurement was 0.2%. The standard deviation of the difference The standard deviation of the difference distribution was 1.1%. distribution was 1.1%.

•• Of the 218 measurements,Of the 218 measurements,–– 124 (56.9%) were within 1%. 124 (56.9%) were within 1%. –– 211 (96.8%) were within 2%. 211 (96.8%) were within 2%. –– 215 (98.6%) were within 3%.215 (98.6%) were within 3%.

•• The largest difference was 3.4%.The largest difference was 3.4%.

Page 73: Electron therapy Class 2: Review questions - UT-H GSBS … … ·  · 2015-02-052 Raphex Question: T63, 2002 • In what situation is electron backscatter likely to be a problem?

73

Page 74: Electron therapy Class 2: Review questions - UT-H GSBS … … ·  · 2015-02-052 Raphex Question: T63, 2002 • In what situation is electron backscatter likely to be a problem?

74

Treatment planning algorithms: Summary

• Pencil beam

• Modified Monte Carlo

• Expect MU to be accurate within ~3%

• Relative isodoses should be accurate for most scenarios, but less reliable for lung

• Algorithms do not include skin collimation, customized bolus or electron arc therapy

•• Pencil beamPencil beam

•• Modified Monte CarloModified Monte Carlo

•• Expect MU to be accurate within ~3%Expect MU to be accurate within ~3%

•• Relative Relative isodosesisodoses should be accurate for should be accurate for most scenarios, but less reliable for lungmost scenarios, but less reliable for lung

•• Algorithms do not include skin collimation, Algorithms do not include skin collimation, customized bolus or electron arc therapycustomized bolus or electron arc therapy