electric radiant panel testing: hardware maintenance parameters

9
ELECTRIC RADIANT PANEL TESTING: HARDWARE MAINTENANCE PARAMETERS KHANG D. TRAN, Ph.D. THE MEXMIL COMPANY SANTA ANA, CALIFORNIA, USA INTERNATIONAL AIRCRAFT MATERIALS FIRE TEST WORKING GROUP MEETING, 06/2003, HAMBURG, GERMANY

Upload: wyome

Post on 23-Feb-2016

54 views

Category:

Documents


0 download

DESCRIPTION

ELECTRIC RADIANT PANEL TESTING: HARDWARE MAINTENANCE PARAMETERS. KHANG D. TRAN, Ph.D. THE MEXMIL COMPANY SANTA ANA, CALIFORNIA, USA. INTERNATIONAL AIRCRAFT MATERIALS FIRE TEST WORKING GROUP MEETING, 06/2003, HAMBURG, GERMANY. Water supply. Exhaust. Control panel. Radiant panel. - PowerPoint PPT Presentation

TRANSCRIPT

ELECTRIC RADIANT PANEL TESTING: HARDWARE MAINTENANCE PARAMETERS KHANG D. TRAN, Ph.D.

THE MEXMIL COMPANYSANTA ANA, CALIFORNIA, USA

INTERNATIONAL AIRCRAFT MATERIALS FIRE TEST WORKING GROUP MEETING, 06/2003, HAMBURG, GERMANY

MEXMIL RADIANT PANEL TESTING SYSTEM

Control panel

Water supply

Radiant panel

Exhaust

Burner

RADIANT PANEL TESTING RESULTS

EQUIPMENT SETUP PARAMETERS

OPERATIONAL PARAMETERS

MATERIALS/SPECIMEN PARAMETERS

HARDWARE MAINTENANCE PARAMETERS

PARAMETERS THAT COULD AFFECT RADIANT PANEL TEST RESULTS

This presentation describes how the hard-ware maintenance parameters affect results of a radiant panel test

POWER CONTROLLER CIRCUIT BREAKER FAILURE

ELECTRIC RADIANT PANEL CORROSION RESISTANCE OF HEATING ELEMENTS BURNER TIP CLOGGING

• A Watlow electric radiant panel including 6 stainless steel radiant emitter strips are used in the system.• Maximum operating temperature: 1300 ºF• Strip resistance: 11.2 /16 Amp each phase

OBSERVATION: TEMPERATURE AND HEAT FLUX READINGS LEVEL OFF BEFORE REACHING SETPOINTS

VERIFICATION: Temperature of heating strips is not equal. Strips are not uniformly red hot.

1. Check the fuses (using an Ohm-meter or a Continuity Checker) 2. Check the strip resistance (using an Ohm-meter) Each strip should read 11.2 (Watlow strips)

3. Detect any corrosion signs: High temperature corrosion occurs due to the interaction at high temperatures of gases, or solids with materials.

4. Check the current intensity for each phase using a clamp-on ammeter (qualified electrician is required)

R

UNPLUG THE POWER CORD TO PERFORMMEASUREMENTS

A typical clamp-on ammeter

The electronic circuit of the strips

BURNER TIP MAINTENANCE

• Flue way/Air inlets clogged • Flame turns yellow, gets weaker,• Burner orifice dirty and curves up.• Gases are burning off • The inner blue cone becomes blurred

SPECIMEN

THE BURNER TIP SHOULD BE CLEANED USING ULTRA-SONIC BATCH CLEANING

A clean burner tip produces blue flame with bright inner cone

A BLUE FLAME MEANS A complete combustion A higher flame temperature A higher heat transfer rate

WHEREAS A YELLOW FLAME EXHIBITS An incomplete combustion A lower flame temperature A lower heat transfer rate

TEST RESULTS ARE UNPREDICTABLE

Surface burning occurs instead of flame penetration into supporting layers.This could make the process of flame propagation (on the surface of the specimen)unpredictable.