Page 1: Dr. Serena Psoroulas :: Technology development :: Centre for … · Beam width in air at iso-center as a Broadening of the beam width function of beam energy in air due to scattering

WIR SCHAFFEN WISSEN – HEUTE FÜR MORGEN

Motivation and challenges of fast dose delivery

Dr. Serena Psoroulas :: Technology development :: Centre for Proton therapy

EUCARD Workshop on Innovative Delivery Systems in Particle Therapy

Torino, 23.02.2017

Page 2: Dr. Serena Psoroulas :: Technology development :: Centre for … · Beam width in air at iso-center as a Broadening of the beam width function of beam energy in air due to scattering

Page 2

Treatment times vs room occupancy times in the PSI Centre for Proton Therapy

Room occupancy affected by many factors

Treatment times of order of 1-2 Gy/L/mincommon in clinical facilitiesFast room switching advantageous for multi-room facilities

23.2.2017 S. Psoroulas, Motivation and challenges of fast scanning

Gantry 2

Positioning time

Evtl. on-board imaging

Treatment time -

field on time

Act

ual

pat

ien

tti

me

so

n 2

2.0

2.2

01

7

Why faster

scanning?

Challenges?

Page 3: Dr. Serena Psoroulas :: Technology development :: Centre for … · Beam width in air at iso-center as a Broadening of the beam width function of beam energy in air due to scattering

Page 3

Motivation: moving tumours

23.2.2017 S. Psoroulas, Motivation and challenges of fast scanning

Page 4: Dr. Serena Psoroulas :: Technology development :: Centre for … · Beam width in air at iso-center as a Broadening of the beam width function of beam energy in air due to scattering

Page 4

Interplay effect with pencil beam scanning

Free breathing:

• Interference between motion

of target and motion of beam

across the different spots leads

to hot/cold spots (interplay

effect)

• Dose blurring at the edges

100 % = Target dose

Overdosage (clinically

not acceptable)

Underdosage (clinically

not acceptable)

Page 5: Dr. Serena Psoroulas :: Technology development :: Centre for … · Beam width in air at iso-center as a Broadening of the beam width function of beam energy in air due to scattering

Results of informal survey from A. Knopf (UMCG):Scanned beams preferred to scattering even for motion cases; but high need for • Fast scanning• Appropriate motion mitigation technique

What technique to treat moving targets?

Page 523.2.2017 S. Psoroulas, Motivation and challenges of fast scanning

Report on 11 centers (9 proton centres, 1 carbon centre, 1 proton + carbon

centre):

A. K

no

pf, 4

DT

P w

orksh

op

20

15

Page 6: Dr. Serena Psoroulas :: Technology development :: Centre for … · Beam width in air at iso-center as a Broadening of the beam width function of beam energy in air due to scattering

Page 6

Rescanning and machine performance

Goal: average hot-cold spots by re-applying the same field several times

Time 1 layer << breathing period

Affected by delivery performance as:• Energy switching time• Beam stability after energy switching

Results in increase of treatment timeChoice of rescanning strategy: likelymachine-dependent!

23.2.2017 S. Psoroulas, Motivation and challenges of fast scanning

From Seco et al 2009

Phys. Med. Biol. 54 N283

Volumetric Randomised

Layered Time-delayTime-delayed

Breath-sampling

Page 7: Dr. Serena Psoroulas :: Technology development :: Centre for … · Beam width in air at iso-center as a Broadening of the beam width function of beam energy in air due to scattering

Goal: reduce amplitude of motionTrigger delivery by monitoring in realtime tumour motion (or its surrogate)

Residual motion interplay to be suppressed using additional mitigation(eg rescanning) or appropriate planning parameters

Affected by: • Low efficiency in beam application,

particularly for synchrotrons (unlessspecial excitation cycles are used)

Results in increase of treatment time

Page 7

Gating and machine performance

23.2.2017 S. Psoroulas, Motivation and challenges of fast scanning

A. S

chae

tti, P

SI

G. Fattori et al, Testing of optical tracking

gating system at PSI Gantry 2 (in preparation)

Page 8: Dr. Serena Psoroulas :: Technology development :: Centre for … · Beam width in air at iso-center as a Broadening of the beam width function of beam energy in air due to scattering

Page 8

Breath hold and machine performance

Goal: quasi-static irradiation of tumoursin thorax/upper abdomen achievedwith patient’s cooperation

Affected by:• Total scanning time through volumeAs well as:• Reproducibility between breath-holds• Patient’s breath-hold length (Breath-

hold windows ~20 s)

Requires:• Fast scanning in all three dimensions• Precise patient monitoringTechniques such as apnea might relax the timing requirements

23.2.2017 S. Psoroulas, Motivation and challenges of fast scanning

Radiat Oncol J. 2014 Jun;32(2):84-94

Page 9: Dr. Serena Psoroulas :: Technology development :: Centre for … · Beam width in air at iso-center as a Broadening of the beam width function of beam energy in air due to scattering

Goal: compensate for CTV motion

Affected by:• Changes of WEPL along beam axis• Target deformation• Reaction time of tracking (and

adaptation) system• Reproducibility of motion (particularly

in inhomogeneous tissues)

Requires:• Real time monitoring of the internal

anatomy• Fast monitoring of changes in WEPL• Fast adaptation of beam configuration

(most challenging: range)• Appropriate treatment planning

parameters

Page 9

Tracking and machine performance

23.2.2017 S. Psoroulas, Motivation and challenges of fast scanning

From N. Saito et al 2009

Phys. Med. Biol. 54 4849

Page 10: Dr. Serena Psoroulas :: Technology development :: Centre for … · Beam width in air at iso-center as a Broadening of the beam width function of beam energy in air due to scattering

Page 10

How about immobilisation?

Immobilisation is a primary mean of motion/interplay mitigation:• Abdominal compression (liver cases)• Rectal balloon (prostate cases)

BUT:• Residual motion often not completely

negligible (eg compression)• Static configuration achieved for few

minutes, then changes in the anatomycan occur (eg drifts)

Immobilisation will work best with fast scanning (and if needed, combined with motion mitigation)

23.2.2017 S. Psoroulas, Motivation and challenges of fast scanning

Eccles, C. L., et al (2011),

IJROB 79(2), 602–608.

Page 11: Dr. Serena Psoroulas :: Technology development :: Centre for … · Beam width in air at iso-center as a Broadening of the beam width function of beam energy in air due to scattering

Fast energyswitching

Efficientdelivery

Fast transversescanning

Fast scanning to make motion mitigation possible

Page 1123.2.2017 S. Psoroulas, Motivation and challenges of fast scanning

Page 12: Dr. Serena Psoroulas :: Technology development :: Centre for … · Beam width in air at iso-center as a Broadening of the beam width function of beam energy in air due to scattering

Page 12

Fast scanning in depth: energy switching

strategy and beam stability

23.2.2017 S. Psoroulas, Motivation and challenges of fast scanning

Page 13: Dr. Serena Psoroulas :: Technology development :: Centre for … · Beam width in air at iso-center as a Broadening of the beam width function of beam energy in air due to scattering

Cyclotrons provide single energy: need

for energy degradation

• Upstream: energy degradation after

cyclotron

Need to adapt all magnets in the

beam line: settling time is the

bottleneck!

• Downstream: range shifter or ridge

filter before patient

Can reach fast switching (50 ms), but

at the price of beam size

Page 13

Energy switching strategies for cyclotrons

23.2.2017 S. Psoroulas, Motivation and challenges of fast scanning

Page 14: Dr. Serena Psoroulas :: Technology development :: Centre for … · Beam width in air at iso-center as a Broadening of the beam width function of beam energy in air due to scattering

Impact of energy switching strategyon beam size (PSI gantries)

1423.2.2017 S. Psoroulas, Motivation and challenges of fast scanning

70 100 130 170 200 2300

2

4Spot size Gantry 2, 18 cm air gap

FW

HM

[cm

]Energy [MeV]

FW

HM

(cm

)

Gantry 2

Beam width in air at iso-center as a function of beam energyBroadening of the beam width

in air due to scattering of the range shifter plates (0, 10, 20, 30: number of range shifter plates) as a function of air gap. From Pedroni et al., 2005

Gantry 1

105 MeV

68 MeV

134 MeV

D. Meer, PSI

Cyclotron facilities nowadays mostly rely

on upstream beam changes

Exception: Mevion

Page 15: Dr. Serena Psoroulas :: Technology development :: Centre for … · Beam width in air at iso-center as a Broadening of the beam width function of beam energy in air due to scattering

Impact of energy switching strategy on beamsize in synchrotrons (HIMAC simulation)

1523.2.2017 S. Psoroulas, Motivation and challenges of fast scanning

Upstream changes benefit

synchrotrons, but from timing point

of view a range shifter is still a

strong advantage

Under investigation: extended

flattop with multiple energies

Inaniwa, et al. 2012, Med. Phys., 39: 2820–2825.

downstream upstream

Page 16: Dr. Serena Psoroulas :: Technology development :: Centre for … · Beam width in air at iso-center as a Broadening of the beam width function of beam energy in air due to scattering

• Energy changes = changes in magnetic field = Eddy currents

• Lamination helps suppressing the effect – up to a limited degree

Impact of fast energy changeson bending magnets

1623.2.2017 S. Psoroulas, Motivation and challenges of fast scanning

Lamination of yoke perpendicular to orbit

Lamination parallel to particle orbit reduces

eddy currents in entrance/exit region

AMF3 dipole field map

Components parallel to

lamination on magnet edges

Gabard, A., et al. (2010), IEEE Transactions on Applied Superconductivity, 20.

Page 17: Dr. Serena Psoroulas :: Technology development :: Centre for … · Beam width in air at iso-center as a Broadening of the beam width function of beam energy in air due to scattering

0 20 40 60 80 100 120 140 160 180 2000

2000

4000

6000

water equivalent depth in MLIC [mm]

AD

C c

ou

nts

/cha

nn

el

0 0.5 1 1.5 2 2.5 3 3.5 4−0.05

0

0.05

Elapsed time since first spot [s]

rela

tive

devia

tio

n f

rom

mea

nm

easu

red 8

0%

fall−

off [

mm

]

200 MeV

150 MeV

120 MeV

Experimental validation of constancy of rangewith fast energy changes

Page 1723.2.2017 S. Psoroulas, Motivation and challenges of fast scanning

Measurements of range following a 100 ms energy change

20 ms beam delivery/spot

Range constancy better than 0.05 mm Psoroulas et al, manuscript in

preparation

Page 18: Dr. Serena Psoroulas :: Technology development :: Centre for … · Beam width in air at iso-center as a Broadening of the beam width function of beam energy in air due to scattering

Spot position drift at isocentre

1823.2.2017 S. Psoroulas, Motivation and challenges of fast scanning

AMA1 and upstream magnets:

no impact on spot position

AMF1 – AMF2: exponential spot drift,

decay within 1 s, effect < 0.5 mm

AMF3: slow decaying spot drift,

amplitude: > 2 mm, depending

on Energy change

Effect of the three magnets is partly compensating

(beam optics) – but residual drift visible: need online

correction (implemented in G2 control system at PSI)

Dis

per

sive

dir

ecti

on

ON

LY!!

Psoroulas et al,

manuscript in

preparation

Page 19: Dr. Serena Psoroulas :: Technology development :: Centre for … · Beam width in air at iso-center as a Broadening of the beam width function of beam energy in air due to scattering

• Strong impact of energy switching time, affecting the preferred strategy (e.g. Bernatowicz et al, PMB 58(22), 2013)

Rescanning

• For upstream energy switching systems: need low switching times!

• For synchrotrons: need synchronisation with accelerator (see later)

Gating and breath hold

• Need: 5 mm range compensation within 10 ms(Saito, et al., PMB 54(15), 2009)

• Current systems cannot match such requirementsTracking

Consequences for motion mitigation techniques

Page 1923.2.2017 S. Psoroulas, Motivation and challenges of fast scanning

Page 20: Dr. Serena Psoroulas :: Technology development :: Centre for … · Beam width in air at iso-center as a Broadening of the beam width function of beam energy in air due to scattering

Page 20

Fast scanning in transverse direction

23.2.2017 S. Psoroulas, Motivation and challenges of fast scanning

Page 21: Dr. Serena Psoroulas :: Technology development :: Centre for … · Beam width in air at iso-center as a Broadening of the beam width function of beam energy in air due to scattering

• Scanning magnets capable of

reaching up to 2 cm/ms

• Ferrite core: no hysteresis/eddy

currents effects at the expense of

field size

• Future: air-core magnets? (in

combination with laser-pulsed

acceleration)

21

Fast scanning: scanning magnets

23.2.2017 S. Psoroulas, Motivation and challenges of fast scanning

Dispersive direction:

Max 0.4 T

Transversedirection

Max 0.2 T

Ferrite core to avoid eddy currents

Pictures: PSI magnet section

Page 22: Dr. Serena Psoroulas :: Technology development :: Centre for … · Beam width in air at iso-center as a Broadening of the beam width function of beam energy in air due to scattering

Step-and-shoot approach

pioneered at PSI and GSI and

commonly used in all centres:

• Beam intensity constant (evtl.

fixed by the smallest spot length)

• Position constant

• Dose-driven: once all protons for

one position have been delivered,

move to the next

Very robust technique, from a time

when beam intensity was not very

stable...

Page 22

Current pencil beam scanning technique:dose-driven

23.2.2017 S. Psoroulas, Motivation and challenges of fast scanning

Image courtesy of Varian Medical Systems, Inc. All rights reserved.

Page 23: Dr. Serena Psoroulas :: Technology development :: Centre for … · Beam width in air at iso-center as a Broadening of the beam width function of beam energy in air due to scattering

Page 23

Alternative: time driven

All delivery techniques can be implemented also in time-driven mode:Dose at patients is defined as instantaneous beam current and scanning speed

Implemented by Sumitomo (independently also at PSI) to improve timing performance• Continuous speed modulation• Continuous beam delivery

Requires:• Stable beam currents• Closed-loop control of quantities (to react

to instabilities)• High-rate monitoring signals (to

eventually trigger interlocks)

23.2.2017 S. Psoroulas, Motivation and challenges of fast scanning

timet1 t2 t3 tn

Line from

measurement

with beam,

2015

G. Klimpki et al., submitted; talk tomorrow

Page 24: Dr. Serena Psoroulas :: Technology development :: Centre for … · Beam width in air at iso-center as a Broadening of the beam width function of beam energy in air due to scattering

• Higher scanning speeds (for the same beam current) means more rescanning

• Possible issue: minimum number of monitor unitsRescanning:

• Higher scanning speeds without increase in beam current might not win enough time…

Gating/breath hold

• Speeds > 0.5 cm/ms are already enough for corrections!Tracking

Consequences for motion mitigation techniques

Page 2423.2.2017 S. Psoroulas, Motivation and challenges of fast scanning

Page 25: Dr. Serena Psoroulas :: Technology development :: Centre for … · Beam width in air at iso-center as a Broadening of the beam width function of beam energy in air due to scattering

Page 25

Dose rate and dose modulation

23.2.2017 S. Psoroulas, Motivation and challenges of fast scanning

Page 26: Dr. Serena Psoroulas :: Technology development :: Centre for … · Beam width in air at iso-center as a Broadening of the beam width function of beam energy in air due to scattering

Maximum beam current is the decisive factor for reducing treatment time

Affected by:

• Shielding considerations

Beam losses in accelerator, along the beam line etc

• Patient safety system reaction time

Dose delivered in case of an incident constrained by safety standards

• Injection/extraction

For synchrotrons external source: limitation in injector, space charge, ring

emittance acceptance

For cyclotrons with internal source: source characteristics, speed of

modulation, extraction efficiency

• Beam losses/inefficiency after accelerator

Maximum beam current and treatment time

Page 2623.2.2017 S. Psoroulas, Motivation and challenges of fast scanning

Page 27: Dr. Serena Psoroulas :: Technology development :: Centre for … · Beam width in air at iso-center as a Broadening of the beam width function of beam energy in air due to scattering

Degrader highly impacts beam

divergence

Additional collimation to match

following beam line causes large

energy dependent transmission losses

At PSI: intensity compensation

strategies to flatten the transmission

curve

• Use current from the accelerator

• Use focussing/defocussing elements

in the beam line

• Different degrader materials (under

investigation...)

Page 27

Beam losses in transportwith upstream energy selection system

23.2.2017 S. Psoroulas, Motivation and challenges of fast scanning

van Goethem M J et al., Phys. Med. Biol. 54 5831–46

Page 28: Dr. Serena Psoroulas :: Technology development :: Centre for … · Beam width in air at iso-center as a Broadening of the beam width function of beam energy in air due to scattering

• Dose-driven approaches: can rely on

slow (~s) regulations loop

• Time-driven scanning approaches

rely on stability of beam current

In G2 conditions, current should

settle within ~150 μs

Need closed-loop controller, fast

detectors, and fast power supplies

Page 28

Fast scanning: beam intensity modulation

23.2.2017 S. Psoroulas, Motivation and challenges of fast scanning

Psoroulas et al, CYC2016

Page 29: Dr. Serena Psoroulas :: Technology development :: Centre for … · Beam width in air at iso-center as a Broadening of the beam width function of beam energy in air due to scattering

Phase controlled rescanning:

layered rescanning adapting beam

current to deliver one layer in one

gating window

• Requires control of the beam

intensity

• In combination with extended flat

tops and range shifter (to avoid

dead times due to acceleration

cycle)

• Beam intensity primarily will

depend on number of rescans –

and it might be too low to

deliver…

Page 29

Phase controlled rescanning and gating

23.2.2017 S. Psoroulas, Motivation and challenges of fast scanning

Furukawa et al., 2007, Med.

Phys., 34: 1085–1097

Page 30: Dr. Serena Psoroulas :: Technology development :: Centre for … · Beam width in air at iso-center as a Broadening of the beam width function of beam energy in air due to scattering

When rescanning low-weighted spots/layers, the minimum number of MU

deliverable by the system can be easily reached – what to do?

a) Skip those spots

number of skipped spots will depend on the number of rescans

b) Limit the maximum number of rescans

the spots contributing less to the total dose decide the number of rescans

c) Increase the minimum MU

will depend on the number of rescans

d) ‘Smart rescanning’: apply a different number

of rescans per each spot, compatible with MU

constraint (implemented at PSI)

Rescanning and deliverable spots

Page 3023.2.2017 S. Psoroulas, Motivation and challenges of fast scanning

Actis

et al, PTC

OG

55

Page 31: Dr. Serena Psoroulas :: Technology development :: Centre for … · Beam width in air at iso-center as a Broadening of the beam width function of beam energy in air due to scattering

• Higher flexibility in beam current might help – but not always…

• Minimum number of MU affects max number of rescansRescanning

• More affected by beam current limitations

• cyclotrons: extraction efficency, ESS losses

• synchrotrons: extraction technique, number of particles in ring

• Pulsed beams (synchrocyclotrons)?

Gating/Breath hold

• Beam modulation to correct for deformations/rotations of the target (Luechtenborg et al, IFMBE proc., 2009)

• Complex predictive algorithm required to account for off-axis dose of other raster positions

Tracking

Consequences for motion mitigation techniques

Page 3123.2.2017 S. Psoroulas, Motivation and challenges of fast scanning

Page 32: Dr. Serena Psoroulas :: Technology development :: Centre for … · Beam width in air at iso-center as a Broadening of the beam width function of beam energy in air due to scattering

Fast energyswitching

Efficientdelivery

Fast transversescanning

Fast scanning to make motion mitigation possible

Page 3223.2.2017 S. Psoroulas, Motivation and challenges of fast scanning

Tracking?

As low aspossible! Will benefit alltechniques

Will mainly benefit gatingand breath hold

Higher speeds could help with rescanning and with higher dose rates

Page 33: Dr. Serena Psoroulas :: Technology development :: Centre for … · Beam width in air at iso-center as a Broadening of the beam width function of beam energy in air due to scattering

Improving scanning is still substantial work!

Page 333/4/2017 S. Psoroulas, Centre for Proton Therapy

Thank you for

your attention!

PSI Gantry 2 scanning system