dr 7 ndt - dÜrr ndt · dr 7 ndt is controlled with the module crnet if the module cr scanner is...

20
EN DR 7 NDT Installation and Operating Instructions 2122100011L41 1705V002

Upload: others

Post on 17-Mar-2020

8 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: DR 7 NDT - DÜRR NDT · DR 7 NDT is controlled with the module CRNet If the module CR scanner is not required, (e g for a HD-CR 35 NDT with USB connection up to year of construction

EN

DR 7 NDTInstallation and Operating Instructions

2122100011L41

1705V

002

Page 2: DR 7 NDT - DÜRR NDT · DR 7 NDT is controlled with the module CRNet If the module CR scanner is not required, (e g for a HD-CR 35 NDT with USB connection up to year of construction
Page 3: DR 7 NDT - DÜRR NDT · DR 7 NDT is controlled with the module CRNet If the module CR scanner is not required, (e g for a HD-CR 35 NDT with USB connection up to year of construction

2122100011L41 1705V002 1

Operation10 Operation � � � � � � � � � � � � � � � � � � � � � � � 11

10�1 Taking an X-ray image� � � � � � � � � � 1110�2 Perform Defect Pixel calibration � � � 14

11 Cleaning � � � � � � � � � � � � � � � � � � � � � � � � 15

12 Maintenance� � � � � � � � � � � � � � � � � � � � � 15

Troubleshooting13 Tips for operators and service

technicians � � � � � � � � � � � � � � � � � � � � � � 1613�1 Poor X-ray image � � � � � � � � � � � � � 1613�2 Software error � � � � � � � � � � � � � � � � 17

EN

Contents

Important information1 About this document � � � � � � � � � � � � � � � 2

1�1 Warnings and symbols � � � � � � � � � � 21�2 Copyright information � � � � � � � � � � � 2

2 Safety � � � � � � � � � � � � � � � � � � � � � � � � � � � 32�1 Intended use � � � � � � � � � � � � � � � � � � 32�2 Improper use � � � � � � � � � � � � � � � � � 32�3 General safety information � � � � � � � � 32�4 Guarantee claim � � � � � � � � � � � � � � � 32�5 Qualified personnel � � � � � � � � � � � � � 32�6 Protection from electric shock � � � � � 32�7 Only use genuine parts � � � � � � � � � � 32�8 Transport � � � � � � � � � � � � � � � � � � � � 32�9 Disposal � � � � � � � � � � � � � � � � � � � � � 4

Product description3 Overview� � � � � � � � � � � � � � � � � � � � � � � � � 5

3�1 Scope of delivery � � � � � � � � � � � � � � 53�2 Special accessories � � � � � � � � � � � � 5

4 Technical data � � � � � � � � � � � � � � � � � � � � 64�1 Unit � � � � � � � � � � � � � � � � � � � � � � � � 64�2 Ambient conditions � � � � � � � � � � � � � 64�3 Type plate � � � � � � � � � � � � � � � � � � � � 74�4 Conformity assessment � � � � � � � � � � 7

5 Operation � � � � � � � � � � � � � � � � � � � � � � � � 7

Installation6 Requirements � � � � � � � � � � � � � � � � � � � � � 8

6�1 System requirements � � � � � � � � � � � 8

7 Installation � � � � � � � � � � � � � � � � � � � � � � � 87�1 Electrical connections � � � � � � � � � � � 87�2 Install driver(s) � � � � � � � � � � � � � � � � � 87�3 Installing the calibration data � � � � � � 9

8 Configuration � � � � � � � � � � � � � � � � � � � � � 98�1 Deactivate modules not required

in D-Tect� � � � � � � � � � � � � � � � � � � � � 98�2 Configure the unit in CRConfig � � � 10

9 Testing the device � � � � � � � � � � � � � � � � 10

Page 4: DR 7 NDT - DÜRR NDT · DR 7 NDT is controlled with the module CRNet If the module CR scanner is not required, (e g for a HD-CR 35 NDT with USB connection up to year of construction

2 2122100011L41 1705V002

Other symbolsThese symbols are used in the document and on or in the unit:

Note, e�g� specific instructions regarding efficient and cost-effective use of the unit�

Observe the accompanying documenta-tion�

CE labelling

Manufacturer

Dispose of correctly in accordance with EU Directive 2012/19/EU (WEEE)�

1.2 Copyright informationAll names of circuits, processes, names, soft-ware programs and units used in this document are protected by copyright� The Installation and Operating Instructions must not be copied or reprinted, neither in full nor in part, without written authorisation from Dürr NDT�

1 About this documentThese installation and operating instructions represent part of the unit�

Failure to comply with the instructions and information in these installation and operating instructions means that Dürr NDT will not be able to offer any warranty or assume any liability for the safe opera-tion and the safe functioning of the unit�

1.1 Warnings and symbols

WarningsThe warnings in this document are intended to draw your attention to possible injury to persons or damage to machinery� The following warning symbols are used:

General warning symbol

The warnings are structured as follows:

SIGNAL WORDDescription of the type and source of dangerHere you will find the possible conse-quences of ignoring the warning i Follow these measures to avoid the danger�

The signal word differentiates between four levels of danger:

– DANGERImmediate danger of severe injury or death

– WARNINGPossible danger of severe injury or death

– CAUTIONRisk of minor injuries

– NOTICERisk of extensive material/property damage

Important information

EN

Page 5: DR 7 NDT - DÜRR NDT · DR 7 NDT is controlled with the module CRNet If the module CR scanner is not required, (e g for a HD-CR 35 NDT with USB connection up to year of construction

2122100011L41 1705V002 3

Important information

2.5 Qualified personnel

OperationPersons who operate the units must ensure safe and correct handling based on their training and knowledge� i Instruct or have every user instructed in han-dling the unit�

Installation and repairs i All installation, resetting, alteration, extension and repair work must be carried out either by DÜRR NDT personnel or by a suitably quali-fied person approved by DÜRR NDT�

2.6 Protection from electric shock i When working on the units observe all the rel-evant electrical safety regulations�

i Immediately replace any damaged cables or plugs�

2.7 Only use genuine parts i Use only those accessories and optional ac-cessories specified or approved by DÜRR NDT�

i Only use only original wear parts and replace-ment parts�

DÜRR NDT accepts no liability for dam-ages resulting from the use of non-ap-proved accessories or optional accesso-ries, or from the use of non-original wear parts or replacement parts�The use of non-approved accessories, optional accessories or non-genuine wear parts / replacement parts (e�g�  mains cable) can have a negative effect in terms of electrical safety and EMC�

2.8 TransportThe original packaging provides optimum pro-tection for the unit during transport�Original packing for this unit can be ordered from DÜRR NDT as required�

DÜRR NDT does not accept liability for any damage to the unit resulting from transport in unsuitable packaging, even during the guarantee period�

i Only transport the unit in its original packag-ing�

i Keep the packing materials out of the reach of children�

2 SafetyDürr NDT has designed and constructed this unit so that when used in accordance with its Intended use there is no danger to people or property� Nevertheless, residual risks can re-main� You should therefore observe the follow-ing notes�

2.1 Intended useThe sensor is designed exclusively for the gen-eration of X-ray images in an industrial context�The sensor is designed for use of up to 70 kV and a total dosage of up to 20 Gy� Any use above this value can result in deterioration of the image quality and its informative value� In such a case, the guarantee claim expires�

2.2 Improper useAny other usage or usage of the sensor beyond this scope, especially its application for medical purposes is deemed to be improper� The manu-facturer accepts no liability for damages result-ing from improper usage� In these cases the user/operator will bear the sole risk�

WARNINGDanger of explosion from the ignition of combustible materials i Do not operate the unit in any rooms containing flammable mixtures�

2.3 General safety information i When operating this device always observe all guidelines, laws, and other rules and regula-tions that are applicable at the site of opera-tion�

i Prior to each use, check condition of the de-vice and make sure it is in perfect working or-der�

i Do not convert or modify the units� i Observe the Installation and Operating In-structions�

i Make the Installation and Operating Instruc-tions available to the person operating the de-vice at all times�

2.4 Guarantee claimThe guarantee claim will expire if the surround-ing conditions are not maintained in operation or the sensor suffers damage as a result of exterior force or overly-high exposure values�

EN

Page 6: DR 7 NDT - DÜRR NDT · DR 7 NDT is controlled with the module CRNet If the module CR scanner is not required, (e g for a HD-CR 35 NDT with USB connection up to year of construction

4 2122100011L41 1705V002

Important information

2.9 Disposal

UnitThe unit must be disposed of properly� Within the European Union, the unit must be disposed of in accordance with EU Directive 2012/19/EU (WEEE)�

i If you have any questions concerning the cor-rect disposal of the product, please contact DÜRR NDT or your specialist supplier�

EN

Page 7: DR 7 NDT - DÜRR NDT · DR 7 NDT is controlled with the module CRNet If the module CR scanner is not required, (e g for a HD-CR 35 NDT with USB connection up to year of construction

2122100011L41 1705V002 5

3 Overview

1

2

1 Sensor2 Calibration CD

3.1 Scope of deliveryThe following items are included in the scope of delivery:

DR 7 NDT . . . . . . . . . . . . . . . . . . 2122100013 – Sensor – Calibration CD – Installation and operating instructions

3.2 Special accessoriesThe following optional items can be used with the device: USB cable extension 4�8 m � � � � � � 2106-155-63

Product description

EN

Page 8: DR 7 NDT - DÜRR NDT · DR 7 NDT is controlled with the module CRNet If the module CR scanner is not required, (e g for a HD-CR 35 NDT with USB connection up to year of construction

6 2122100011L41 1705V002

Product description

4 Technical data

4.1 Unit

Electrical dataNominal voltage V DC 5Nominal current mA 100

General technical dataDimensions W x H x D

mm33�1 x 44�7 x 6�331�5 x 50�0 x 8�3

Cable length sensor m 4�5Max USB cable extension

m4�8 (see 3�2 Special

accessories)Computer connection USB 2�0

Compatible with USB 3�0

Sensor characteristicsActive sensor size W x H mm 26 x 36Min� pixel size µm 19Max� no� pixels 1368 x 1896Theoretical resolution LP/mm 26�3Sensor type CMOSScintillator Structured CsJ scintillator on fi-

bre-opticsMaximum voltage in free beam path kV 70Maximum total dosage Gy 20

4.2 Ambient conditions

Ambient conditions during operationTemperature °C 10 - 35Relative humidity % < 80Air pressure hPa 750 - 1160Height above sea level m < 2000

Ambient conditions during storage and transportTemperature °C -20 to +60Relative humidity % < 95Air pressure hPa 750 - 1160Height above sea level m < 16000

EN

Page 9: DR 7 NDT - DÜRR NDT · DR 7 NDT is controlled with the module CRNet If the module CR scanner is not required, (e g for a HD-CR 35 NDT with USB connection up to year of construction

2122100011L41 1705V002 7

Product description

5 OperationThe X-ray sensor is connected to the computer via the sensor cable� The sensor is placed and activated via an imag-ing software (e� g� D-Tect)� During the X-ray procedure, the x-rays are re-ceived by the scintillator (luminescent material) installed on the sensor and are converted into beams of light� These light beams are recorded by the sensor as image information, digitalised and transferred to the computer� If an error oc-curs during the transfer, the image information is sent again� No further X-ray image is possible during data transmission�

4.3 Type plateThe type plate is located on the sensor cable and on the case�

Figure 1: Type plate on the sensor cable

Figure 2: Type plate on the case

REF Order number/type numberSN Serial number

4.4 Conformity assessmentThis device has been subjected to conformity acceptance testing in accordance with the cur-rent relevant European Union guidelines� This equipment conforms to all relevant require-ments�

EN

Page 10: DR 7 NDT - DÜRR NDT · DR 7 NDT is controlled with the module CRNet If the module CR scanner is not required, (e g for a HD-CR 35 NDT with USB connection up to year of construction

8 2122100011L41 1705V002

7 Installation

7.1 Electrical connections

Combining devices safelyTake care when connecting units together or to parts of other systems as there is always an ele-ment of risk (e�g� due to leakage currents)� i Only connect units when there is no danger to the operator or to the environment�

i Only connect units when it is safe to do so and when there is no risk of damage or harm to the surroundings�

i If it is not 100% clear from the unit data sheet that such connections can be safely made or if you are in any doubt, always get a suitably qualified person (e�g� the manufacturer) to ver-ify that the setup is safe�

i Only connect peripheral units (e�g� computer, monitor, printer) that conform at least to the requirements set out in IEC 60950-1 (EN 60950-1)�

Connecting the unit to the computer

NOTICEDamage to the sensorThe interior parts of the sensor can be damaged by mechanical forces� i Do not drop the sensor� i Do not subject the sensor to any pressure�

i Do not fold, clamp or crush the cable� i Do not carry the sensor by the cable�

i Connect the sensor USB connector to a free socket on the computer�

7.2 Install driver(s)The drivers for the DR 7 NDT are automatically installed with the installation of D-Tect (from ver-sion 9�5)� i Connect the sensor USB connector to a free socket on the computer�

The unit is initialised automatically�

6 Requirements i The unit may only be operated given compli-ance with the locally-applicable legal specifi-cations (e� g� Röntgenverordnung (German X-ray Directive))�

i The unit must not be operated in a damp or wet environment (closed room / outside) (see "4 Technical data")�

i Do not subject the unit to direct sunlight or heat�

i The setup room may not to contain any major interference fields (e� g� strong magnetic fields)� They could cause faults in the unit�

6.1 System requirements

System requirements for computer sys-tems, see information sheet (order num-ber 9000-608-02) or on the Internet at www�duerr-ndt�de�

Installation

EN

Page 11: DR 7 NDT - DÜRR NDT · DR 7 NDT is controlled with the module CRNet If the module CR scanner is not required, (e g for a HD-CR 35 NDT with USB connection up to year of construction

2122100011L41 1705V002 9

Installation

8 ConfigurationThe unit supports the following imaging programs:

– D-Tect – ScanX View – Third-party software on request

8.1 Deactivate modules not required in D-Tect

i Start D-Tect� i In the Options menu, select > Display Configuration�

i Click on the Modules  button�

DR 7 NDT is controlled with the module CRNet� If the module CR scanner is not required, (e�g� for a HD-CR 35 NDT with USB connection up to year of construction mid-2013), this module must also be inactive� i Double-click D-Tect� i Select the Options registration card� i Drag the CR scanner module with into the Inactive field�

i Restart D-Tect�

7.3 Installing the calibration data

Every sensor requires the individual cali-bration data stored on the calibration CD included in the scope of delivery�

i Insert the calibration CD included in the scope of delivery in the CD drive�

The CD starts automatically and the calibra-tion data are installed�

i If the CD does not play automatically, execute the file setup.exe �

EN

Page 12: DR 7 NDT - DÜRR NDT · DR 7 NDT is controlled with the module CRNet If the module CR scanner is not required, (e g for a HD-CR 35 NDT with USB connection up to year of construction

10 2122100011L41 1705V002

Installation

9 Testing the deviceA X-ray image can be taken to test whether the unit has been connected correctly� i Select the Test registration card�

i Select the unit from the Registered Units list� i Select the mode class and mode� i Click on Scan Image�

i Record an X-ray image�

8.2 Configure the unit in CRConfig i Select Start > All Programs > Duerr NDT > CRScan > CRNetConfig�orIn D-Tect on the registration page Config. > Double-click   on CRNet module�

i Activate the connected unit in the Registered column�

i Click on the Edit  button�

i Change the unit name (Designation) as re-quired�

i Click Apply to save the configuration�

EN

Page 13: DR 7 NDT - DÜRR NDT · DR 7 NDT is controlled with the module CRNet If the module CR scanner is not required, (e g for a HD-CR 35 NDT with USB connection up to year of construction

2122100011L41 1705V002 11

The acquisition modes listed are standard con-figurations� The acquisition modes can be adapted, depending on requirements and appli-cation�Depending on the X-ray source and the record-ing conditions, the basis spatial resolution can change�

10.1 Taking an X-ray imageX-ray images with the DR 7 NDT are generated using multiple integration� Multiple integration produces a low SNR (signal to noise ratio)�Check before every use:

– The sensor and cable are not damaged� – The sensor and cable have been cleaned�

NOTICESensor damage and deterioration of the image quality resulting from high exposure values in the free beam path i With exposure values > 70 kV, protect the sensor accordingly�

i Comply with the maximum total dos-age of 20 Gy�

i Place the sensor� i Start the computer� i Start the imaging software� i Select the project� i Change to X-ray module� i Select the desired acquisition mode�

10 Operation

NOTICEDamage to the sensorThe interior parts of the sensor can be damaged by mechanical forces� i Do not drop the sensor� i Do not subject the sensor to any pressure�

i Do not fold, clamp or crush the cable� i Do not carry the sensor by the cable�

CAUTIONA damaged sensor can result in the issue of materials which damage health i Check the sensor and cable for dam-age before every use�

i Do not use a damaged sensor�

NOTICEShort circuit due to the build up of condensationStrong temperature fluctuations can damage the unit� i Only commission the unit once it has warmed up to room temperature�

i Do not subject the unit to strong tem-perature differences (heating max� 3 °C/min)� Exception: short-term use of the sensor at a temperature of up to 45 °C� This could produced in-creased levels of inherent noise in the image�

If DR 7 NDT is operated with D-Tect, you can generate X-ray images in the following acquisi-tion modes:

– NDT high (19 µm) – NDT Standard (19 µm) – NDT fast/binning (38 µm)

The transmission time takes a few seconds de-pending on the setting�

Operation

EN

Page 14: DR 7 NDT - DÜRR NDT · DR 7 NDT is controlled with the module CRNet If the module CR scanner is not required, (e g for a HD-CR 35 NDT with USB connection up to year of construction

12 2122100011L41 1705V002

Operation

i Trigger the X-ray image with Start integration�

The image is automatically transferred to the imaging software� No further X-ray image is possible during data transmission�

Perform offset-gain calibrationThe offset-gain calibration minimizes the falsifi-cation of the image information (e� g� through an inhomogeneous illumination of the X-ray source)� The offset-gain calibration is performed in two steps:

1� Sensor calibration without an active X-ray source (offset)

2� Gain calibration of the sensor with an active X-ray source without objects in the beam path (Gain)

i Click � i Working in the Calibration range, select Off-set-Gain from the selection list�

Under Status:Idling is displayed�

i Click the Scan button�

The Image acquisition window opens�

If an acquisition mode is used for the first time, or if this has been changed, offset-gain calibration must first be performed (see "Perform offset-gain calibration")�

i Click �

The settings for multiple integration open�

i Adapt the abort settings to the integration (see "Adapt the abort settings to the integra-tion")�

i Activate the X-ray source�

EN

Page 15: DR 7 NDT - DÜRR NDT · DR 7 NDT is controlled with the module CRNet If the module CR scanner is not required, (e g for a HD-CR 35 NDT with USB connection up to year of construction

2122100011L41 1705V002 13

Operation

Adapt the abort settings to the integrationMultiple integration has the following abort pre-conditions:

SNR threshold level

Multiple integration is ended automatically after the SNR threshold entered has been reached�

Dosage threshold val-ue

Multiple integration is ended automatically after the dosage threshold value has been reached�

Integration time

Multiple integration is ended automatically after the integra-tion time entered has been reached�

None The X-ray imaging continues until it is ended by the opera-tor�

If the abort preconditions are not fulfilled, imaging continues until it is ended by the operator�

i Select the required abort precondition� i Enter the desired value�

The standard settings are restored with Set standard�SNR threshold level: 150Dosage threshold value: 40000 µGyIntegration time: 30 s

i Click Save� i Close window� i Working in the abort preconditions SNR threshold value and dosage threshold preview image, select the area in which the value is to be determined�Left click in the preview image to place an area of defined size on the mouse pointer position�Alternatively an area of any size can be dragged open (left click) with the mouse�

i Click Start [Starten]�The X-ray source must be inactive to this end�

Calibration in idling is started�The quantity of images (min�/max�) to be generated is shown under Number of imag-es�Calibration in idling can be ended manually at any time� Calibration is ended automati-cally once the maximum number has been reached�Status: shows Start X-ray!�

NOTICESensor damage and deterioration of the image quality resulting from high exposure values in the free beam path i With exposure values > 70 kV, protect the sensor accordingly�

i Comply with the maximum total dos-age of 20 Gy�

i Activate the X-ray source� i Working under Dosage level, check whether the dosage of the X-ray source lies within the specifications�Should the dosage level lie outside this speci-fication, adapt the X-ray source�

i Working in the preview image, check whether objects are in the beam path and remove them if necessary�

i Click Start [Starten]�

Calibration with an active X-ray source is started�The quantity of images (min�/max�) to be generated is shown under Number of imag-es�The calibration can be ended manually at any time� Calibration is ended automatically once the maximum number has been reached�The offset-gain calibration has been com-pleted�

EN

Page 16: DR 7 NDT - DÜRR NDT · DR 7 NDT is controlled with the module CRNet If the module CR scanner is not required, (e g for a HD-CR 35 NDT with USB connection up to year of construction

14 2122100011L41 1705V002

Operation

i Click �

The defect pixel calibration settings will open�

i Click Start [Starten]�The X-ray source must be inactive to this end�

i Place DR 7 NDT 30 cm parallel to the X-ray source�

i Remove all objects from the beam path� i Set the X-ray source in accordance with the specifications in the dialog window�

i Click Start [Starten]� i Trigger the x-ray source five times�

The dialogue indicates when calibration has been completed�

i Close window�

The X-ray is stored as a PNG file (Calibra-tionFile_[Kalibrierungsnummer].kor.png) here: C:\Duerr\devices\imda\401\131E0025\calibrationData\[Serial number of the sensor] and can be viewed� The X-ray image can be evaluated in accord-ance with the standard specifications�

10.2 Perform Defect Pixel calibrationThe ageing of the sensor can result in pixilation� This would result in artefacts in the X-ray image� These pixels are identified through the defect pixel calibration and deducted from all X-ray im-ages�The calibration data are included in the scope of delivery for delivery� The calibration data contain the defect pixel file at the point of delivery�The defect pixel calibration can be performed at any time�

We recommend that you perform defect pixel calibration with CRNetConfig under Start > All programmes > Duerr NDT > CRScan�With CRNetConfig, it is not necessary to select a project to start an X-ray image�

i Start CRNetConfig�

i Change to the Test tab� i Working under Registered units: select the unit DR 7�

i Select the module class all� i Select Defect Pixel mode� i Click on Scan Image�

EN

Page 17: DR 7 NDT - DÜRR NDT · DR 7 NDT is controlled with the module CRNet If the module CR scanner is not required, (e g for a HD-CR 35 NDT with USB connection up to year of construction

2122100011L41 1705V002 15

Operation

12 MaintenanceThe appliance is maintenance-free�

11 Cleaning

NOTICEEquipment damage from moisture in the plug i Do not spray the USB connector�

i Unplug the unit� i Remove any soiling with a soft, damp, lint-free cloth�

i Allow the unit to dry completely before using it�

EN

Page 18: DR 7 NDT - DÜRR NDT · DR 7 NDT is controlled with the module CRNet If the module CR scanner is not required, (e g for a HD-CR 35 NDT with USB connection up to year of construction

16 2122100011L41 1705V002

13 Tips for operators and service technicians

Any repairs above and beyond routine maintenance must only be carried out by suitably quali-fied personnel or by one of our service technicians�

13.1 Poor X-ray image

Fault Probable cause Solution

No image transmission, despite X-ray image hav-ing been triggered

X-ray dose too low i Correct the exposure values�

Artefacts in image The sensor is damaged from me-chanical force

i Replace the sensor�

The sensor has suffered damage from excess exposure values (see "4 Technical data")

i Replace the sensor�

X-ray image too dark/bright

Incorrect exposure time set i Correct the exposure time�

Troubleshooting

EN

Page 19: DR 7 NDT - DÜRR NDT · DR 7 NDT is controlled with the module CRNet If the module CR scanner is not required, (e g for a HD-CR 35 NDT with USB connection up to year of construction

2122100011L41 1705V002 17

Troubleshooting

13.2 Software error

Fault Probable cause Solution

The operating system does not recognise the unit

A different USB unit is blocking the USB port

i Remove the blocking USB unit�

Error code E-0012 The unit driver has not been in-stalled correctly

i Install the unit driver afresh (see installation and configuration manual)�

Error code E-0077 Software installation corrupt i Inform your Service Technician�

Error code E-1001 The unit is not connected i Connect the unit to the USB port�

Unit is defective i Inform your Service Technician�

Error code E-1002 Multiple sensors are connected to the computer at the same time

i Disconnect the sensors not re-quired� Only a single sensor may be connected to the computer at any one time�

Error code E-1008 Sensor connected with over-long or incorrect USB extension

i Only use the USB extension listed in the optional acces-sories�

The computer USB port is unsuit-able for the data rate

i Connect the sensor to a differ-ent USB port�

Error code E-0012 The unit driver has not been in-stalled correctly

i Install the unit driver afresh (see installation and configuration manual)�

Error code E-1020 The current state of the sensor does not permit image acquisition

i Disconnect and reconnect the USB connection cable�

Sensor defective i Inform your Service Technician�

Error code E-1026 Incorrect acquisition mode i Select a different acquisition mode

i Inform your Service Technician�

Error code E-2006 No acquisition mode selected i Select the desired acquisition mode�

Error code E-10014 Sensor calibration data not in-stalled

i Install the sensor calibration data (see installation and configura-tion manual)�

Error code E-10016 Sensor calibration data incorrect i Inform your Service Technician�

EN

Page 20: DR 7 NDT - DÜRR NDT · DR 7 NDT is controlled with the module CRNet If the module CR scanner is not required, (e g for a HD-CR 35 NDT with USB connection up to year of construction

DÜRR NDT GmbH & Co. KGHöpfigheimer Straße 2274321 Bietigheim-Bissingen, GermanyPhone: +49 7142 99381-0 · Fax: +49 7142 [email protected] · www.duerr-ndt.de