![Page 1: Using Hematology Instrument Data to Troubleshoot](https://reader033.vdocuments.site/reader033/viewer/2022061302/54e4c1734a79597b7b8b4a2b/html5/thumbnails/1.jpg)
MEDICAL MYSTERIESMEDICAL MYSTERIESUsing Hematology Instrument Data to
Troubleshoot
DR PETER JOHN LOGA, PhD; MS; MSc; BSc; SDMLT; DMLT; FIBMS; FZIMLS
![Page 2: Using Hematology Instrument Data to Troubleshoot](https://reader033.vdocuments.site/reader033/viewer/2022061302/54e4c1734a79597b7b8b4a2b/html5/thumbnails/2.jpg)
OBJECTIVESOBJECTIVES
Review instrument technology; compare & contrast normal vs abnormal
Apply this technology / knowledge to a variety of cases
“SOLVE” the medical mystery
![Page 3: Using Hematology Instrument Data to Troubleshoot](https://reader033.vdocuments.site/reader033/viewer/2022061302/54e4c1734a79597b7b8b4a2b/html5/thumbnails/3.jpg)
![Page 4: Using Hematology Instrument Data to Troubleshoot](https://reader033.vdocuments.site/reader033/viewer/2022061302/54e4c1734a79597b7b8b4a2b/html5/thumbnails/4.jpg)
PROVENPROVEN BECKMAN COULTERBECKMAN COULTER
TECHNOLOGIESTECHNOLOGIES
![Page 5: Using Hematology Instrument Data to Troubleshoot](https://reader033.vdocuments.site/reader033/viewer/2022061302/54e4c1734a79597b7b8b4a2b/html5/thumbnails/5.jpg)
The Coulter PrincipleThe Coulter Principle
External Electrode
Vacuum
External Housing(Aperture Bath)
Detail of Aperture
Internal Electrode
Suspension of Cells
Aperture Current Pathway
Aperture
Aperture Housing
![Page 6: Using Hematology Instrument Data to Troubleshoot](https://reader033.vdocuments.site/reader033/viewer/2022061302/54e4c1734a79597b7b8b4a2b/html5/thumbnails/6.jpg)
RBC HISTOGRAMRBC HISTOGRAM
NORMALNORMAL
![Page 7: Using Hematology Instrument Data to Troubleshoot](https://reader033.vdocuments.site/reader033/viewer/2022061302/54e4c1734a79597b7b8b4a2b/html5/thumbnails/7.jpg)
RBC HISTOGRAMRBC HISTOGRAM
RBC FRAGMENTS, MICROCYTIC RBCs, Giant PLTRBC FRAGMENTS, MICROCYTIC RBCs, Giant PLT DI RBCsDI RBCs
MACROCYTIC, TARGET CELLS, DI RBCMACROCYTIC, TARGET CELLS, DI RBCCOLD AGGLUTININCOLD AGGLUTININ
Post TransfusionPost Transfusion
![Page 8: Using Hematology Instrument Data to Troubleshoot](https://reader033.vdocuments.site/reader033/viewer/2022061302/54e4c1734a79597b7b8b4a2b/html5/thumbnails/8.jpg)
PLT HISTOGRAMSPLT HISTOGRAMS
NORMALNORMAL
![Page 9: Using Hematology Instrument Data to Troubleshoot](https://reader033.vdocuments.site/reader033/viewer/2022061302/54e4c1734a79597b7b8b4a2b/html5/thumbnails/9.jpg)
PLT Curve FittingPLT Curve FittingThe Curve Fitting Process Allows
More Accurate Counts When Platelets of Larger Than 20 fL Are Present
![Page 10: Using Hematology Instrument Data to Troubleshoot](https://reader033.vdocuments.site/reader033/viewer/2022061302/54e4c1734a79597b7b8b4a2b/html5/thumbnails/10.jpg)
PLT Counting & SizingPLT Counting & Sizing
Coulter impedance counting has a PATENTED CURVE FITTING process that is used in conjunction with WBC histogram review for platelet clump and giant platelet flags
![Page 11: Using Hematology Instrument Data to Troubleshoot](https://reader033.vdocuments.site/reader033/viewer/2022061302/54e4c1734a79597b7b8b4a2b/html5/thumbnails/11.jpg)
PLT HISTOGRAMSPLT HISTOGRAMS
Giant Platelets
Small Platelets
![Page 12: Using Hematology Instrument Data to Troubleshoot](https://reader033.vdocuments.site/reader033/viewer/2022061302/54e4c1734a79597b7b8b4a2b/html5/thumbnails/12.jpg)
Sensing Zone
Neutrophil
Oscilloscope
The Coulter PrincipleThe Coulter Principle
A white cell passes through WBC aperture
![Page 13: Using Hematology Instrument Data to Troubleshoot](https://reader033.vdocuments.site/reader033/viewer/2022061302/54e4c1734a79597b7b8b4a2b/html5/thumbnails/13.jpg)
Eos
Baso
Coulter WBC HistogramCoulter WBC Histogram
Lymphs
50 – 90 fL
Monos
90 -160 fL
Neuts
160 - 450 fL
![Page 14: Using Hematology Instrument Data to Troubleshoot](https://reader033.vdocuments.site/reader033/viewer/2022061302/54e4c1734a79597b7b8b4a2b/html5/thumbnails/14.jpg)
WBC HISTOGRAMSWBC HISTOGRAMS
ImmNE2 Eosinophilia
ImmNE1 & ImmNE2
Blasts
Lymphocytosis Variant Lymph
![Page 15: Using Hematology Instrument Data to Troubleshoot](https://reader033.vdocuments.site/reader033/viewer/2022061302/54e4c1734a79597b7b8b4a2b/html5/thumbnails/15.jpg)
WBC InterferenceWBC Interference
Percentage of interference analyzed for statistical significance
Flagging based on all three histograms instead of one
Histogram positional parameters used for further definition
Cellular Interference
![Page 16: Using Hematology Instrument Data to Troubleshoot](https://reader033.vdocuments.site/reader033/viewer/2022061302/54e4c1734a79597b7b8b4a2b/html5/thumbnails/16.jpg)
AccuCount TechnologyAccuCount Technology
– LH700 Series
WBC 0 – 400,000
RBC 0 – 8,000,000
HGB 0 - 25
PLT 0 – 3,000,000
AccuCount WBC and
AccuCount Plt Counts have
been “validated” by
Reference Flow
Cytometry
![Page 17: Using Hematology Instrument Data to Troubleshoot](https://reader033.vdocuments.site/reader033/viewer/2022061302/54e4c1734a79597b7b8b4a2b/html5/thumbnails/17.jpg)
VCS TECHNOLOGYVCS TECHNOLOGYAutomated Differential
Analysis
![Page 18: Using Hematology Instrument Data to Troubleshoot](https://reader033.vdocuments.site/reader033/viewer/2022061302/54e4c1734a79597b7b8b4a2b/html5/thumbnails/18.jpg)
Near Native WBC AnalysisNear Native WBC Analysis
Red Cells Removed From Sample Dilution Using a Lytic Process
Second Agent Prevents Alteration of the White Cells
Hydrodynamically Focused Flowcell
– Laminar Flow Ensures Single File Cell Passage
– Coincidence Effects Are Minimized
![Page 19: Using Hematology Instrument Data to Troubleshoot](https://reader033.vdocuments.site/reader033/viewer/2022061302/54e4c1734a79597b7b8b4a2b/html5/thumbnails/19.jpg)
BioPhysical Flow CytometryBioPhysical Flow Cytometry
SHEATH-FLUID IN
SHEATH-FLUID IN
SAMPLE IN
FLOW CELL
SAMPLE DILUTION
SHEATH FLUIDSENSING
AREA
Cells are hydrodynamically focused An electro-optical flow cytometer provides concurrent
electronic and optical measurements
![Page 20: Using Hematology Instrument Data to Troubleshoot](https://reader033.vdocuments.site/reader033/viewer/2022061302/54e4c1734a79597b7b8b4a2b/html5/thumbnails/20.jpg)
NUCLEAR SHAPE
AND COMPOSITION
CELL SIZECYTOPLASM
GRANULES
The 3-D VCS ScatterplotThe 3-D VCS Scatterplot
![Page 21: Using Hematology Instrument Data to Troubleshoot](https://reader033.vdocuments.site/reader033/viewer/2022061302/54e4c1734a79597b7b8b4a2b/html5/thumbnails/21.jpg)
COULTER COULTER VCSVCS TECHNOLOGYTECHNOLOGY
VOLUME = SIZE
CONDUCTIVITY = INTERNAL COMPOSITION
LIGHT SCATTER = CELL SHAPE / SURFACE
![Page 22: Using Hematology Instrument Data to Troubleshoot](https://reader033.vdocuments.site/reader033/viewer/2022061302/54e4c1734a79597b7b8b4a2b/html5/thumbnails/22.jpg)
VOLUMEVOLUME
DC Measures Total Cell Volume Using the Reference Method of Direct Current Impedance
Unaffected by cell orientation
![Page 23: Using Hematology Instrument Data to Troubleshoot](https://reader033.vdocuments.site/reader033/viewer/2022061302/54e4c1734a79597b7b8b4a2b/html5/thumbnails/23.jpg)
CONDUCTIVITYCONDUCTIVITY
RF Measures Internal Cell Structure Using Radiographic Imaging Similar to Ultrasound
Conductivity Is a Proprietary Technology
![Page 24: Using Hematology Instrument Data to Troubleshoot](https://reader033.vdocuments.site/reader033/viewer/2022061302/54e4c1734a79597b7b8b4a2b/html5/thumbnails/24.jpg)
LASER LIGHT SCATTERLASER LIGHT SCATTER
Light Scatter Measures Cell Surface Granularity Using a Broad Range of Angles. Over 60 angles of light scatter are analyzed.
![Page 25: Using Hematology Instrument Data to Troubleshoot](https://reader033.vdocuments.site/reader033/viewer/2022061302/54e4c1734a79597b7b8b4a2b/html5/thumbnails/25.jpg)
Lymphs
Monos
Basos
Neuts
Eos
3-D Cellular Analysis - VCS3-D Cellular Analysis - VCS
VOLUME (Y) CONDUCTIVITY (Z)
LIGHT SCATTER (X)
The 3 probes (DC, RF and Scatter) interrogateeach of the 8192 cells simultaneously.
Every cell is treated in the same manner and each cell is given an X, Y, and Z coordinate onthe dataplot; with 16 million points in the matrix.
ALL cell populations are DIRECTLYmeasured
![Page 26: Using Hematology Instrument Data to Troubleshoot](https://reader033.vdocuments.site/reader033/viewer/2022061302/54e4c1734a79597b7b8b4a2b/html5/thumbnails/26.jpg)
Better Better Abnormal Cell DetectionAbnormal Cell Detection
1 Mono-Blasts 2 Myelo-Blasts 3 Immature Granulocytes 4 Band Neutrophils 5 Lympho-Blasts 6 Variant Lymphocytes 7 Low Volume Lymphocytes 7a NRBCs 8 PLT Clumps 9 Giant Platelets10 RBC Parasites (Malaria,
etc)
1
2
34
56
7
8 9 10
7a
![Page 27: Using Hematology Instrument Data to Troubleshoot](https://reader033.vdocuments.site/reader033/viewer/2022061302/54e4c1734a79597b7b8b4a2b/html5/thumbnails/27.jpg)
DIFF TECHNOLOGYDIFF TECHNOLOGY
![Page 28: Using Hematology Instrument Data to Troubleshoot](https://reader033.vdocuments.site/reader033/viewer/2022061302/54e4c1734a79597b7b8b4a2b/html5/thumbnails/28.jpg)
NORMALNORMAL
![Page 29: Using Hematology Instrument Data to Troubleshoot](https://reader033.vdocuments.site/reader033/viewer/2022061302/54e4c1734a79597b7b8b4a2b/html5/thumbnails/29.jpg)
NORMALNORMAL
![Page 30: Using Hematology Instrument Data to Troubleshoot](https://reader033.vdocuments.site/reader033/viewer/2022061302/54e4c1734a79597b7b8b4a2b/html5/thumbnails/30.jpg)
NORMAL DATAPLOTNORMAL DATAPLOT
MONOCYTESNEUTROPHILS
EOSINOPHILS
LYMPHOCYTES
BASOPHILS
NRBC, PLT CLUMPS, GIANT PLT, MALARIAL PARASITES, DEBRIS, ETC
C O N
D U
C T
I V I
T Y
VOLUME
S C A T T E R
![Page 31: Using Hematology Instrument Data to Troubleshoot](https://reader033.vdocuments.site/reader033/viewer/2022061302/54e4c1734a79597b7b8b4a2b/html5/thumbnails/31.jpg)
CONDUCTIVITY
SCATTER
VO
LU
ME
CONDUCTIV
ITY
CUBE ROTATIONCUBE ROTATION
RED = VOLUME = SIZEGREEN = SCATTER = SURFACEBLUE = CONDUCTIVITY = INTERNAL
RED = VOLUME = SIZEGREEN = SCATTER = SURFACEBLUE = CONDUCTIVITY = INTERNAL
SCATTER
VO
LU
ME
![Page 32: Using Hematology Instrument Data to Troubleshoot](https://reader033.vdocuments.site/reader033/viewer/2022061302/54e4c1734a79597b7b8b4a2b/html5/thumbnails/32.jpg)
LH 700 SeriesLH 700 SeriesThe “6-Part Diff”The “6-Part Diff”
Fully automated
– No reflex or repeat testing required
– No additional reagent packs required
WBC count
automatically corrected
NRBC enumeration automatic with differential
![Page 33: Using Hematology Instrument Data to Troubleshoot](https://reader033.vdocuments.site/reader033/viewer/2022061302/54e4c1734a79597b7b8b4a2b/html5/thumbnails/33.jpg)
Decision RulesDecision Rules
4 Rule Types
Message-Action To Be Taken
And/Or Joins
Automatically Make Slide
UNLIMITEDRULES!
![Page 34: Using Hematology Instrument Data to Troubleshoot](https://reader033.vdocuments.site/reader033/viewer/2022061302/54e4c1734a79597b7b8b4a2b/html5/thumbnails/34.jpg)
Research Population Data (RPD)Research Population Data (RPD)
When VCS 3D Dataplot is optimized;
There is a change in the WBC Research Population Data
This appears to correlate with the presence of abnormal cells in previously undiagnosed patients
![Page 35: Using Hematology Instrument Data to Troubleshoot](https://reader033.vdocuments.site/reader033/viewer/2022061302/54e4c1734a79597b7b8b4a2b/html5/thumbnails/35.jpg)
The increasing SD correspondsto a more immature populationof cells
NE1
Research Population Data
![Page 36: Using Hematology Instrument Data to Troubleshoot](https://reader033.vdocuments.site/reader033/viewer/2022061302/54e4c1734a79597b7b8b4a2b/html5/thumbnails/36.jpg)
Research Population Data Research Population Data (RPD)(RPD)
WBC Research Population Data has been studied in the following clinical cases:– CLL– Left Shift– Malaria– Lymphoproliferative Disorders– Myelodysplasia– Sepsis
![Page 37: Using Hematology Instrument Data to Troubleshoot](https://reader033.vdocuments.site/reader033/viewer/2022061302/54e4c1734a79597b7b8b4a2b/html5/thumbnails/37.jpg)
Steve MarionneauxLaboratory ManagerThe Saint Vincent’s Comprehensive Cancer CenterNew York, New York
CLINICALCLINICAL APPLICATION APPLICATION
![Page 38: Using Hematology Instrument Data to Troubleshoot](https://reader033.vdocuments.site/reader033/viewer/2022061302/54e4c1734a79597b7b8b4a2b/html5/thumbnails/38.jpg)
MYSTERY #1MYSTERY #1
![Page 39: Using Hematology Instrument Data to Troubleshoot](https://reader033.vdocuments.site/reader033/viewer/2022061302/54e4c1734a79597b7b8b4a2b/html5/thumbnails/39.jpg)
CBC ResultsCBC Results8 Year Old Female8 Year Old Female
![Page 40: Using Hematology Instrument Data to Troubleshoot](https://reader033.vdocuments.site/reader033/viewer/2022061302/54e4c1734a79597b7b8b4a2b/html5/thumbnails/40.jpg)
DataPlot DataPlot ResultsResults
![Page 41: Using Hematology Instrument Data to Troubleshoot](https://reader033.vdocuments.site/reader033/viewer/2022061302/54e4c1734a79597b7b8b4a2b/html5/thumbnails/41.jpg)
MANUAL DIFF RESULTSMANUAL DIFF RESULTS
MANUAL DIFFSeg = 20Band = 2Lymph = 51Blast = 27
![Page 42: Using Hematology Instrument Data to Troubleshoot](https://reader033.vdocuments.site/reader033/viewer/2022061302/54e4c1734a79597b7b8b4a2b/html5/thumbnails/42.jpg)
Diff Cube RotationDiff Cube Rotation
CONDUCTIVITY
SCA
TT
ER
VO
LU
ME
SCATTER
VO
LU
ME
CONDUCTIV
ITY
PRE-B CELL Acute Lymphoblastic Leukemia
PRE-B CELL Acute Lymphoblastic Leukemia
![Page 43: Using Hematology Instrument Data to Troubleshoot](https://reader033.vdocuments.site/reader033/viewer/2022061302/54e4c1734a79597b7b8b4a2b/html5/thumbnails/43.jpg)
PRECURSORPRECURSOR B-CELL ALL B-CELL ALL
Low WBC, neutropenia Anemic Mononuclear population with smooth
chromatin CD34+, TdT+ population
![Page 44: Using Hematology Instrument Data to Troubleshoot](https://reader033.vdocuments.site/reader033/viewer/2022061302/54e4c1734a79597b7b8b4a2b/html5/thumbnails/44.jpg)
MYSTERY #2MYSTERY #2
![Page 45: Using Hematology Instrument Data to Troubleshoot](https://reader033.vdocuments.site/reader033/viewer/2022061302/54e4c1734a79597b7b8b4a2b/html5/thumbnails/45.jpg)
WBC &PLTWBC &PLTHISTOGRAMSHISTOGRAMS
AUTODIFF RESULTSAUTODIFF RESULTS
![Page 46: Using Hematology Instrument Data to Troubleshoot](https://reader033.vdocuments.site/reader033/viewer/2022061302/54e4c1734a79597b7b8b4a2b/html5/thumbnails/46.jpg)
RBC HISTOGRAMRBC HISTOGRAM
![Page 47: Using Hematology Instrument Data to Troubleshoot](https://reader033.vdocuments.site/reader033/viewer/2022061302/54e4c1734a79597b7b8b4a2b/html5/thumbnails/47.jpg)
CBC / RBC RESULTSCBC / RBC RESULTS
![Page 48: Using Hematology Instrument Data to Troubleshoot](https://reader033.vdocuments.site/reader033/viewer/2022061302/54e4c1734a79597b7b8b4a2b/html5/thumbnails/48.jpg)
MYSTERY #3MYSTERY #3
![Page 49: Using Hematology Instrument Data to Troubleshoot](https://reader033.vdocuments.site/reader033/viewer/2022061302/54e4c1734a79597b7b8b4a2b/html5/thumbnails/49.jpg)
46 Year Old Female46 Year Old Female
![Page 50: Using Hematology Instrument Data to Troubleshoot](https://reader033.vdocuments.site/reader033/viewer/2022061302/54e4c1734a79597b7b8b4a2b/html5/thumbnails/50.jpg)
46 / Female46 / Female
![Page 51: Using Hematology Instrument Data to Troubleshoot](https://reader033.vdocuments.site/reader033/viewer/2022061302/54e4c1734a79597b7b8b4a2b/html5/thumbnails/51.jpg)
Acute Promyelocytic Leukemia (Microgranular)
CONDUCTIVITY SCATTER
MANUAL DIFFLymph = 2Mono = 1Blast = 97
SCATTERCONDUCTIVIT
Y
VO
LU
ME
![Page 52: Using Hematology Instrument Data to Troubleshoot](https://reader033.vdocuments.site/reader033/viewer/2022061302/54e4c1734a79597b7b8b4a2b/html5/thumbnails/52.jpg)
MYSTERY #4MYSTERY #4
![Page 53: Using Hematology Instrument Data to Troubleshoot](https://reader033.vdocuments.site/reader033/viewer/2022061302/54e4c1734a79597b7b8b4a2b/html5/thumbnails/53.jpg)
Medical Mystery #4Medical Mystery #4
![Page 54: Using Hematology Instrument Data to Troubleshoot](https://reader033.vdocuments.site/reader033/viewer/2022061302/54e4c1734a79597b7b8b4a2b/html5/thumbnails/54.jpg)
Chronic Chronic LymphocyticLymphocytic
LeukemiaLeukemia
MANUAL DIFFSeg = 6Lymph = 92Mono = 2
![Page 55: Using Hematology Instrument Data to Troubleshoot](https://reader033.vdocuments.site/reader033/viewer/2022061302/54e4c1734a79597b7b8b4a2b/html5/thumbnails/55.jpg)
CLL
WITH
SMUDGE
CELLS
![Page 56: Using Hematology Instrument Data to Troubleshoot](https://reader033.vdocuments.site/reader033/viewer/2022061302/54e4c1734a79597b7b8b4a2b/html5/thumbnails/56.jpg)
CHRONIC CHRONIC LYMPHOCYTIC LYMPHOCYTIC
LEUKEMIALEUKEMIA Typically >60 years of age Initially asymptomatic Increased WBC Increased % of small, normal
lymphs (as disease progresses, more ‘immature’ lymphs appear
Smudge cells
![Page 57: Using Hematology Instrument Data to Troubleshoot](https://reader033.vdocuments.site/reader033/viewer/2022061302/54e4c1734a79597b7b8b4a2b/html5/thumbnails/57.jpg)
MYSTERY #5MYSTERY #5
![Page 58: Using Hematology Instrument Data to Troubleshoot](https://reader033.vdocuments.site/reader033/viewer/2022061302/54e4c1734a79597b7b8b4a2b/html5/thumbnails/58.jpg)
12 Month Old Male12 Month Old Male
HGB = 7.0
![Page 59: Using Hematology Instrument Data to Troubleshoot](https://reader033.vdocuments.site/reader033/viewer/2022061302/54e4c1734a79597b7b8b4a2b/html5/thumbnails/59.jpg)
12 Month Old Male12 Month Old Male
RBC Morphology2+ Poik 3+ Aniso 4+ Hypo4+ Micro 1+ Target 2+ Ellipto1+ Teardrop 1+ Poly
![Page 60: Using Hematology Instrument Data to Troubleshoot](https://reader033.vdocuments.site/reader033/viewer/2022061302/54e4c1734a79597b7b8b4a2b/html5/thumbnails/60.jpg)
MANUAL DIFFSeg = 42Lymph = 46Mono = 5Eo = 5Baso = 2NRBC = 1
![Page 61: Using Hematology Instrument Data to Troubleshoot](https://reader033.vdocuments.site/reader033/viewer/2022061302/54e4c1734a79597b7b8b4a2b/html5/thumbnails/61.jpg)
??????????
Iron Deficiency
Thalassemia
Iron Deficiency
Thalassemia
![Page 62: Using Hematology Instrument Data to Troubleshoot](https://reader033.vdocuments.site/reader033/viewer/2022061302/54e4c1734a79597b7b8b4a2b/html5/thumbnails/62.jpg)
RETIC RETIC RESEARCH POPULATIONSRESEARCH POPULATIONS
SickleThalassemiaLow Volume Lymphs =CLL
![Page 63: Using Hematology Instrument Data to Troubleshoot](https://reader033.vdocuments.site/reader033/viewer/2022061302/54e4c1734a79597b7b8b4a2b/html5/thumbnails/63.jpg)
MYSTERY #6MYSTERY #6
![Page 64: Using Hematology Instrument Data to Troubleshoot](https://reader033.vdocuments.site/reader033/viewer/2022061302/54e4c1734a79597b7b8b4a2b/html5/thumbnails/64.jpg)
Case Study Case Study HistoryHistory
74 year old female20lb unexplained weight lossFeverMalaiseSore throatMuscle aches
2 weeks duration
![Page 65: Using Hematology Instrument Data to Troubleshoot](https://reader033.vdocuments.site/reader033/viewer/2022061302/54e4c1734a79597b7b8b4a2b/html5/thumbnails/65.jpg)
Blasts or large Blasts or large lymphslymphs
HISTOGRAMDATA
![Page 66: Using Hematology Instrument Data to Troubleshoot](https://reader033.vdocuments.site/reader033/viewer/2022061302/54e4c1734a79597b7b8b4a2b/html5/thumbnails/66.jpg)
???????
AUER ROD
Auer rods aredefined as acoalescence ofthe azurophilicgranules and areonly seen in non-lymphocyticleukemias
![Page 67: Using Hematology Instrument Data to Troubleshoot](https://reader033.vdocuments.site/reader033/viewer/2022061302/54e4c1734a79597b7b8b4a2b/html5/thumbnails/67.jpg)
Manual DifferentialManual Differential
Seg = 4 Band = 1 Lymph = 17 Mono = 3 BLAST = 75 w/ occ
Auer rod
![Page 68: Using Hematology Instrument Data to Troubleshoot](https://reader033.vdocuments.site/reader033/viewer/2022061302/54e4c1734a79597b7b8b4a2b/html5/thumbnails/68.jpg)
ACUTE MYELOCYTIC ACUTE MYELOCYTIC LEUKEMIALEUKEMIA
Sudden onset Anemic Variable WBC Decreased PLT count >10% Blasts in peripheral blood Special Stains & Flow markers
+ for myelogenous cell lines
![Page 69: Using Hematology Instrument Data to Troubleshoot](https://reader033.vdocuments.site/reader033/viewer/2022061302/54e4c1734a79597b7b8b4a2b/html5/thumbnails/69.jpg)
MYSTERY #7MYSTERY #7
![Page 70: Using Hematology Instrument Data to Troubleshoot](https://reader033.vdocuments.site/reader033/viewer/2022061302/54e4c1734a79597b7b8b4a2b/html5/thumbnails/70.jpg)
Case StudyCase StudyHistoryHistory
83 year old maleUnexplained weight lossMalaiseNight sweatsSlight
hepatosplenomegaly
![Page 71: Using Hematology Instrument Data to Troubleshoot](https://reader033.vdocuments.site/reader033/viewer/2022061302/54e4c1734a79597b7b8b4a2b/html5/thumbnails/71.jpg)
![Page 72: Using Hematology Instrument Data to Troubleshoot](https://reader033.vdocuments.site/reader033/viewer/2022061302/54e4c1734a79597b7b8b4a2b/html5/thumbnails/72.jpg)
![Page 73: Using Hematology Instrument Data to Troubleshoot](https://reader033.vdocuments.site/reader033/viewer/2022061302/54e4c1734a79597b7b8b4a2b/html5/thumbnails/73.jpg)
LAB RESULTSLAB RESULTS
Manual Diff:Seg = 33Band = 15Lymph = 19Mono = 6Meta = 11Myelo = 10Blast = 6
![Page 74: Using Hematology Instrument Data to Troubleshoot](https://reader033.vdocuments.site/reader033/viewer/2022061302/54e4c1734a79597b7b8b4a2b/html5/thumbnails/74.jpg)
???
![Page 75: Using Hematology Instrument Data to Troubleshoot](https://reader033.vdocuments.site/reader033/viewer/2022061302/54e4c1734a79597b7b8b4a2b/html5/thumbnails/75.jpg)
Neutrophil Series
•Neutrophils
•Bands
•Metas
•Myelos
•Pros
•Ne Blasts
VCS3-D Data Plot
![Page 76: Using Hematology Instrument Data to Troubleshoot](https://reader033.vdocuments.site/reader033/viewer/2022061302/54e4c1734a79597b7b8b4a2b/html5/thumbnails/76.jpg)
MonocytesMonoblasts
VCS3-D Data Plot
![Page 77: Using Hematology Instrument Data to Troubleshoot](https://reader033.vdocuments.site/reader033/viewer/2022061302/54e4c1734a79597b7b8b4a2b/html5/thumbnails/77.jpg)
FLOW CYTOMETRYFLOW CYTOMETRYPATHOLOGIST PATHOLOGIST
INTERPRETATIONINTERPRETATIONThe immunophenotypic findings reveal increased
monocytes (26%) and 52% granulocytes with a shift toward immaturity and diminished side scatter. There is no evidence of increased blasts, a monoclonal B cell or aberrant T cell process.
The immunophenotypic findings are suggestive of a myeloproliferative process. Acute monocytic leukemia cannot be entirely excluded. Clinical pathologic correlation is required for final diagnosis.
![Page 78: Using Hematology Instrument Data to Troubleshoot](https://reader033.vdocuments.site/reader033/viewer/2022061302/54e4c1734a79597b7b8b4a2b/html5/thumbnails/78.jpg)
MYELOPROLIFERATIVE MYELOPROLIFERATIVE DISORDERSDISORDERS
Defined as a hypercellular bone marrow with increased quantities of one or more of the cells lines: erythrocytes, leukocytes or platelets in the peripheral blood.
It is thought to be a neoplastic, clonal proliferation of a single multipotential stem cell w/ one cell line predominating and often transforming into another.
![Page 79: Using Hematology Instrument Data to Troubleshoot](https://reader033.vdocuments.site/reader033/viewer/2022061302/54e4c1734a79597b7b8b4a2b/html5/thumbnails/79.jpg)
SUMMARYSUMMARY
Look at ALL the information provided by the instrument:– CBC parameters– WBC Histograms– RBC Histograms– PLT Histograms– Dataplots– Suspect Flags– Research Parameter
![Page 80: Using Hematology Instrument Data to Troubleshoot](https://reader033.vdocuments.site/reader033/viewer/2022061302/54e4c1734a79597b7b8b4a2b/html5/thumbnails/80.jpg)
SUMMARYSUMMARY
Combine this information with what you see at the microscope
Ask for a “second opinion” from a peer Create an “abnormal file”
SAVED LIST FOLDER
![Page 81: Using Hematology Instrument Data to Troubleshoot](https://reader033.vdocuments.site/reader033/viewer/2022061302/54e4c1734a79597b7b8b4a2b/html5/thumbnails/81.jpg)
Questions ???????Questions ???????
![Page 82: Using Hematology Instrument Data to Troubleshoot](https://reader033.vdocuments.site/reader033/viewer/2022061302/54e4c1734a79597b7b8b4a2b/html5/thumbnails/82.jpg)
ANY QUESTIONSANY QUESTIONS