User'sManual
Yokogawa Electric Corporation
IM 11G02N01-01E
IM 11G02N01-01E3rd Edition
IR400Infrared Gas Analyzer
iIM 11G02N01-01E
We are grateful for your purchase of Yokogawa’s Infrared Gas Analyzer, Model:IR400.
• First read this instruction manual carefully until an adequate understanding is acquired, andthen proceed to installation, operation and maintenance of the analyzer. Wrong handlingmay cause an accident or injury.
• The specifications of this analyzer are subject to change without prior notice for furtherproduct improvement.
• Modification of this analyzer is strictly prohibited unless a written approval is obtainedfrom the manufacturer. Yokogawa will not bear any responsibility for a trouble caused bysuch a modification.
• This instruction manual shall be stored by the person who actually uses the analyzer.
• After reading the manual, be sure to store it at a place easier to access.
• This instruction manual should be delivered to the end user without fail.
Manufacturer: Yokogawa Electric Corporation
Type: Described in Yokogawa’s company nameplate on main frame
Date of manufacture: Described in Yokogawa’s company nameplate on main frame
Product nationality: Japan
PREFACE
Request
• It is prohibited to transfer part or all of this manual withoutYokogawa’s permission in written format.
• Description in this manual is subject to change without priornotice for further improvement.
ii IM 11G02N01-01E
First of all, read this “Caution on safety” carefully, and then use the analyzer in the correct way.
• The cautionary descriptions listed here contain important information about safety, so they shouldalways be observed. Those safety precautions are ranked in 3 levels, “DANGER”, “CAUTION” and“PROHIBITION”.
Improper handling may cause dangerous situations that mayresult in death or serious injury.
Improper handling may cause dangerous situations that mayresult in medium-level troubles, minor injury, or propertydamage.
Items which must not be done are noted.
Caution on installation and transport of gas analyzer
• The unit is not of explosion-proof specifications. Do not use itin an atmosphere of explosive gases. Otherwise seriousaccidents such as explosion or fire may result.
• Install the analyzer, observing the rules provided in thismanual, in a place that endures the weight of the analyzer.Installation in an inadequate place may cause turnover or fall,resulting in injury.
• Be sure to wear protective gloves when lifting the analyzer.Lifting it with bare hands may result in injury.
• Be sure to fix the casing before transporting the analyzer.Transportation in unstable state may result in injury.
• The gas analyzer is heavy. Two or more persons should carryit, while exercising due care. Otherwise unexpected harm toyour body or injury may result.
• Take care not to let cable chips and other foreign objects enterthe unit during installation work. Otherwise fire, failure, ormalfunction may result.
DANGER
CAUTION
DANGER
CAUTION
PROHIBITION
CAUTION ON SAFETY
iiiIM 11G02N01-01E
Caution on use
• Be sure to read the instruction manual for reference gases beforehandling reference gases such as calibration gas to use themproperly.
• Leaving the analyzer unused for a long time or restarting it afterlong-term suspension requires procedures different from normaloperation or suspension procedures. Be sure to follow theinstructions in each instruction manual. Otherwise, intendedperformance may not be achieved, or accidents or injury mayresult.
• Do not operate the analyzer for a long time with its door leftopen. Otherwise, dust, foreign matter, etc. may stick on internalwalls, thereby causing faults.
Caution on wiring
• Be sure to turn off the power before installing wiring.Otherwise electric shock may result.
• Be sure to perform class D grounding work. Otherwise,electric shock or failure may result.
• Select a proper wiring material that satisfies the ratings of theinstrument. Otherwise, electric shock or fire may result.
• Be sure to connect a power supply of correct rating.Otherwise, fire may result.
Caution on piping
Be sure to observe the following precautions while installingpiping. Improper piping may result in gas leakage.
If the leaking gas contains a toxic component, seriousaccidents may result. If it contains combustible gases,explosion or fire may result.
• Connect pipes correctly referring to the instruction manual.
• Discharge the exhaust gas outdoors to prevent it fromremaining within the sampling device or indoors.
• Relieve the exhaust gas from the analyzer to the atmosphericpressure to prevent buildup of undesirable pressure to theanalyzer. Otherwise piping within the analyzer may bedisconnected, resulting in gas leakage.
• Use pipes and pressure reducing valves to which no oil/greaseis attached for piping. Otherwise, fire may result.
CAUTION
DANGER
DANGER
CAUTION
iv IM 11G02N01-01E
Caution on maintenance and check
• Before performing work with the cover of the analyzer keptopen for maintenance and check, be sure to purge completelynot only within the analyzer but also measuring gas lines withnitrogen or air. Otherwise, poisoning, fire, or explosion mayresult due to gas leakage.
Be sure to observe the following to perform work safely,avoiding electric shock or injury.
• Remove the watch and other metallic objects before work.
• Do not touch the instrument wet-handed.
• If the fuse is blown, eliminate the cause and replace it with theone of the same capacity and type. Otherwise, electric shockor accidents may result.
• Do not use replacement parts other than those specified by themanufacturer. Otherwise, intended performance may not beachieved, or accidents or failures may result.
• Dispose replacement parts such as maintenance parts asincombustibles according to the local waste disposalregulations.
PROHIBITION
DANGER
CAUTION
Others
• If the cause of a fault cannot be identified by referring to theinstruction manual, be sure to contact your dealer or Yokogawatechnician in charge of adjustment. Disassembling theinstrument carelessly may result in electric shock or injury.
CAUTION
Caution on use
• Do not touch the input/output terminals with metal or finger.Otherwise, electric shock or injury may result.
• Do not smoke or use flames near the analyzer. Otherwise, firemay result.
• Do not allow water to enter the analyzer. Otherwise, electricshock or internal fire may result.
vIM 11G02N01-01E
r After - Sales Warranty
d Do not modify the product.
d During the warranty period, for repair under warranty carry or send the product tothe local sales representative or service office. Yokogawa will replace or repair anydamaged parts and return the product to you.
d Before returning a product for repair under warranty, provide us with the modelname and serial number and a description of the problem. Any diagrams or dataexplaining the problem would also be appreciated.
d If we replace the product with a new one, we won’t provide you with a repair report.
d Yokogawa warrants the product for the period stated in the pre-purchase quotation.Yokogawa shall conduct defined warranty service based on its standard. When thecustomer site is located outside of the service area, a fee for dispatching themaintenance engineer will be charged to the customer.
d In the following cases, customer will be charged repair fee regardless of warrantyperiod.
• Failure of components which are out of scope of warranty stated in instructionmanual.
• Failure caused by usage of software, hardware or auxiliary equipment, whichYokogawa did not supply.
• Failure due to improper or insufficient maintenance by user.
• Failure due to misoperation, misuse or modification which Yokogawa does notauthorize.
• Failure due to power supply (voltage, frequency) being outside specifications orabnormal.
• Failure caused by any usage out of scope of recommended usage• Any damage from fire, earthquake, a storm and flood, lightning, disturbance, riot,
warfare, radiation and other natural changes.
d Yokogawa does not warrant conformance with the specific application at the usersite. Yokogawa will not bear direct/indirect responsibility for damage due to aspecific application.
d Yokogawa will not bear responsibility when the user configures the product intosystems or resells the product.
d Maintenance service and supplying repair parts will be covered for five years afterthe production ends. For repair this product, please contact the nearest sales officedescribed in this instruction manual.
≈
Blank Page
CONTENTS
PREFACE.............................................................................................................................. i
Caution on safety ................................................................................................................. ii
r After - Sales Warranty ................................................................................................... v
Contents ............................................................................................................................... vi
1. OVERVIEW ................................................................................................................. 1-1
2. NAME and description of eaCH UNIT ..................................................................... 2-1
2.1 Name and description of main unit ................................................................ 2-12.2 Input/Output terminal module ......................................................................... 2-2
3. INSTALLATION ......................................................................................................... 3-1
3.1 Installation conditions ..................................................................................... 3-13.2 Installation ....................................................................................................... 3-2
3.2.1 Installation of nanlyzer main frame ..............................................................3.2.2 Mounting input/output terminal module ................................................. 3-3
3.3 Piping............................................................................................................... 3-43.4 Sampling .......................................................................................................... 3-7
3.4.1 Conditions of sampling gas ..................................................................... 3-73.4.2 Sampling gas flow ................................................................................... 3-73.4.3 Preparation of standard gas ..................................................................... 3-73.4.4 Reduction of moisture interference ......................................................... 3-73.4.5 Purging of instrument inside ................................................................... 3-73.4.6 Pressure at sampling gas outlet ............................................................... 3-83.4.7 Example configuration of gas sampling system ..................................... 3-8
3.5 Wiring .............................................................................................................. 3-93.5.1 Power inlet ............................................................................................... 3-93.5.2 Input/output terminal module .................................................................. 3-9
4. Operation ...................................................................................................................... 4-1
4.1 Preparation for operation ................................................................................ 4-14.2 Warm-up operation and regular operation...................................................... 4-1
5. Description of display and operation panels ............................................................. 5-1
5.1 Name and description of operation panel ....................................................... 5-15.2 Overview of display and operation panels ..................................................... 5-25.3 Outline of display screen ................................................................................ 5-35.4 General operation ............................................................................................ 5-6
6. SETTING AND CALIBRATION .............................................................................. 6-1
6.1 Switch of range ............................................................................................... 6-16.1.1 Setting of range switch mode .................................................................. 6-16.1.2 Manual range switch ............................................................................... 6-2
6.2 Calibration setting ........................................................................................... 6-36.2.1 Setting of calibration concentration ........................................................ 6-36.2.2 Setting of manual zero calibration .......................................................... 6-56.2.3 Setting of calibration range ..................................................................... 6-7
6.2.4 Setting of auto calibration component/range .......................................... 6-96.3 Alarm setting ................................................................................................. 6-11
6.3.1 Setting of alarm values .......................................................................... 6-116.3.2 Hysteresis setting ................................................................................... 6-13
6.4 Setting of auto calibration ............................................................................. 6-146.4.1 Auto calibration ..................................................................................... 6-146.4.2 Forced run/stop of auto calibration ....................................................... 6-17
6.5 Setting of auto zero calibration ..................................................................... 6-206.5.1 Autozero calibration .............................................................................. 6-206.5.2 Forced run/stop of auto zero calibration ............................................... 6-22
6.6 Peak alarm setting ......................................................................................... 6-256.7 Parameter setting ........................................................................................... 6-276.8 Maintenance mode ........................................................................................ 6-346.9 Calibration ..................................................................................................... 6-40
6.9.1 Zero calibration ..................................................................................... 6-406.9.2 Span calibration ..................................................................................... 6-41
7. MAINTENANCE ......................................................................................................... 7-1
7.1 Daily check ...................................................................................................... 7-17.2 Daily check and maintenance procedures ...................................................... 7-17.3 Maintenance of analyzer unit .......................................................................... 7-2
7.3.1 Cleaning method for sample cell (pipe cell) ........................................... 7-27.3.2 Cleaning method for sample cell (block cell) ......................................... 7-47.3.3 Optical zero adjustment method (optical balance adjustment) ............... 7-67.3.4 Moisture interference compensation adjustment method ....................... 7-77.3.5 Replacement of fuse on analyzer unit ..................................................... 7-8
7.4 Inspection and maintenance of limited service-life components ................... 7-9
8. Trouble shooting for analyzer .................................................................................... 8-1
8.1 Error message .................................................................................................. 8-18.2 Troubleshooting............................................................................................... 8-3
9. SPECIFICATIONS...................................................................................................... 9-1
9.1 General specifications ..................................................................................... 9-19.2 Model and Suffix codes .................................................................................. 9-59.3 External Dimensions ..................................................................................... 9-10
Dedicated relay board (Option code: /R) .............................................. 9-14
Customer Maintenance Parts List ................................................ CMPL 11G02N01-01E
Revision Record .................................................................................................................... i
1 - 1IM 11G02N01-01E
1. OVERVIEW
This infrared gas analyzer (type: IR400) measures the concentration of NO, SO2, CO2, CO and CH4
contained in sampling gas on the principle that different atomic molecules have an absorption spectrum inthe wave band of infrared rays, and the intensity of absorption is determined by the Lambert-Beer law.
Since this instrument incorporates a compact paramagnetic O2 sensor, it allows measuring up to 5components simultaneously by using the built-in O2 sensor (up to 4 components if O2 sensor is excluded).
Furthermore, use of a microprocessor or large sized liquid crystal display realizes improvement ofoperability, accuracy and multi-functions.
This instrument is optimum for measuring combustible gas exhausted from boilers or incinerators, and itis effective for steel gas analysis (blast furnace, steel converter, thermal treatment furnace, sintering(Pellet equipment), coke furnace), storage and maturity of vegetable and fruit, biochemistry (microbe),[fermentation], air pollution [incinerator, exhaust gas desulfurization, denitration], automotive emission(excluding tester), protection against disasters [detection of explosive gas and toxic gas, combustion gasanalysis of new building material], growth of plants, chemical analysis [petroleum refinery plant,petroleum chemistry plant, gas generation plant], environment [landing concentration, tunnelconcentration, parking lot, building management] and various physical and chemical experiments.
≈
Blank Page
2 - 1IM 11G02N01-01E
Fig. 2-1
Name
(1) Handle
(2) Power switch
(3) Display/opera- tion
(4) Sampling gas inlet
(5) Sampling gas outlet
Description
Used for withdrawing the mainunit from the panel.
Used for ON/OFF the analyzer.
Liquid crysral diaplay and keysfor setting various functions
For connecting to the measuringgas tube
Connect to the exhaust line. (Apair of sampling gas inlet/outlet isprovided for each measuring unit.When ordered with purge, thepiping to measuring unit 2 is builtinside. In this case, the sample gasoutlet for measuring unit 2 is usedfor purge gas inlet.)
Name
(6) Sector motor
(7) Light source cover
(8) Input/output ter- minal connector
(9) Power inlet
(10) Protective cover
Description
For driving the rotation ofsector
Infrared light source isarranged in the cover.
For connecting to the externalinput/output terminal module
For connecting the power cable
Protective cover for the lightsource and the motor. May beremoved during operation.
POWER
(1) Handle
(2) Power switch
(5) Sampling gas outlet(4) Sampling gas inlet
(4) Sampling gas inlet
(3) Display/operation panel
For measuring unit 1
For measuring unit 2
(6) Sector motor
(7) Light source cover (9) Power inlet
(5) Sampling gas outlet
(8) Input/output terminal connector
(or purge gas inlet)
Front panel
Back panel
(10) Protective cover
2. NAME AND DESCRIPTION OF EACH UNIT
2.1 Name and description of main unit
2 - 2 IM 11G02N01-01E
2.2 Input/Output terminal moduleThis analyzer provides input/output of various signals from the supplied input/outpt terminalmodule by connecting the instrument to this module.
Fig. 2-2
TN1 TN2 TN3 TN4 TN5
Input/Output terminal module
(1) Mounting hole (φ4.5, 6 places)
(2) Input/Output terminal block (TN1 to TN5)
(6) Communication connector (CN2)
(3) Cable connection connector (CN1)(5) Calibration solenoid valve drive signal output connector (CN3)
(4) Analyzer main unit Input/Output terminal module connection cable (1m)
Name
(1) Mountinghole
(2) Input/outputterminalblock(TN 1 to TN5)
(3) Cableconnectionconnector
Description
Used for mounting input/outputterminal module. φ φ φ φ φ 4.5, 6 places
Input/output terminal forsignals of analog output, rangeidentification contact, alarmcontact output, etc.
Used for connecting the analyzermain unit and the input/outputterminal module (4).
Name
(4) Input/outputterminal moduleconnection cable
(5) Calibrationsolenoid valvedrive signaloutput connector
(6) Communicationconnector
Description
Used for connecting theanalyzer main unit to the input/output terminal module.
Cable connector for connectingthe analyzer to the relay boardfor automatic calibration.
Connect communication cable.*Please refer to another manual(IM 11G02P01-01E) aboutcommunication function.
3 - 1IM 11G02N01-01E
This unit is not explosion-proof type. Do not use it in a place with explosive gases to preventexplosion, fire or other serious accidents.
• Entrust the installation, movement or re-installation to a specialist or the supplier. A poorinstallation may cause accidental tipover, shock hazard, fire, injury, etc.
• The gas analyzer is heavy. It should be installed with utmost care. Otherwise, it may tipover or drop, for example, causing accident or injury.
• For lifting the gas analyzer, be sure to wear protective gloves. Bare hands may invite aninjury.
• This unit should be installed in a place which conforms to the conditions noted in theinstruction manual. Otherwise, it may cause electric shocks, fire or malfunction of the unit.
• During installation work, care should be taken to keep the unit free from entry of cable chipsor other foreign objects. Otherwise, it may cause fire, trouble or malfunction of the unit.
DANGER
CAUTION
3. INSTALLATION
3.1 Installation conditionsTo install the analyzer for optimum performance, select a location that meets the followingconditions;(1) This instrument is system built in type. This instrument should be used while embedded in a panel, locker, or enclosure of steel sheet.(2) Use this instrument indoors.(3) A vibration-free place(4) A place which is clean around the analyzer.(5) Power supply Rated voltage : 100 V to 240 V AC Operating voltage : 85 V to 264 V AC Rated frequency : 50/60 Hz Power consumption : 250 VA max. Inlet : Comformity to EN60320 class I type 3-pin inlet(6) Operation conditions Ambient temperature : -5˚ to 45˚C Ambient humidity : 90% RH or less, no condensation
3 - 2 IM 11G02N01-01E
3.2 Installation3.2.1 Installation of analyzer main frame
Installation methods for the analyzer main unit are divided into 2 types;
Note 1 Check and maintenance of the analyzer main unit may be carried out with the top cover detached.The guide rail method may be used if a space accessible for maintenance is provided at the top ofthe main unit. If maintenance space is not provided specially, it is recommended to use the sliderail method.
Recommended slide rail: Product No.: 305A-24 manufactured by Accuride International Co.
Note 2 For 19 inch rack mounting, the weight of the analyzer is supported with the bottom of the case(with the side of the case in case of slide rail method). For mounting dimensions of the slide rail,see “Item 9.3 External diagram”.
Don’t install the analyzer at a place which is exposed to direct sunlight.
The analyzer should be installed at a place where ambient temperature is within -5 to 45˚C, and
temperature fluctuation during use is minimum.
450 or more
465
101.
6
M6 Slide rail (Supports mass)
Slide rail (Supports mass)
Type External dimensions Mounting dimensions Mounting method
19 inch rack mounting guide rail method
19 inch rack mounting sliderail method
429
177
483
(Unit : mm)
3 - 3IM 11G02N01-01E
3.2.2 Mounting input/output terminal module
Mount the input/output terminal module on the panel; observing the following method.
(Note) To avoid the effect of noise generated from external units, mount the I/O terminal modulemounting plate on the panel for continuity at the mounting surface and connect the panelto the same ground as the analyzer main unit.
Note) How to ground analyzer main unit and I/O terminal module
* Don’t separate the analyzer and I/O terminal module, and be sure to ground them together.
Mount the analyzer and I/O terminal module on the same panel.
Bring I/O terminal module sheet metal and panel into continuity at ///// portion.
To avoid the effect of noises, etc. from external units, it is recommended to ground them by the procedure described below.
Bring analyzer main unit and panel into continuity at ///// portion or rail mounting portion.(No grounding is required at the power terminal).
Ground the panel casing.
External dimensions Mounting dimensions Mounting method
Terminal
Screwed to panel
Panel plate
To analyzer rear panel
Connection cable
4.5 mounting hole (× 6 places)M4 screw (× 6 places)
Rectangular hole
302
142
142
154
150 150
316
164
3 - 4 IM 11G02N01-01E
3.3 PipingPiping should be connected to the gas inlets and outlets of the front panel of the analyzer.• Use a corrosion resistant tube of Teflon, stainless or polyethylene to connect the instrument to
a sampling system. Even if there is a danger of corrosion, refrain from using a tube of rubberor soft vinyl. The instrument provides inaccurate indication due to gas absorption by pipingmaterials.
• Pipe connection port is Rc1/4 female thread (or 1/4 NPT). Piping should be cut as short aspossible for a quick response. About 4 mm inner diameter is recommended.
• Entry of dust into the instrument may result in defective operation. Use a clean piping orcoupling.
Connect the gas tube by the following method.
Sampling gas inlet: Attach the gas tube to introduce gas to be measured such as one that hascompleted dehumidification process and standard gases for zero and spancalibration to this inlet.Gas flow to be introduced should be constant within the range of 0.5 ± 0.2L/min.
Sampling gas outlet: Exhaust measured gas through the outlet. Attach the tube to exhaust mea-sured gas outdoors or to the atmosphere.
Purge gas inlet: It is used for purging the inside of the total gas analyzer . When the ana-lyzer must be purged, refer to Item 3.3.4 Purging of instrument inside.Use dry gas N
2 or instrumentation air for purge gas. (flow rate of 1L/min or
more should be used and no dust or mist is contained).
Sampling gas outlet
Sampling gas inlet
Sampling gas outlet(Purge gas inlet)
Sampling gas inlet
For measuring unit 1
For measuring unit 2
3 - 5IM 11G02N01-01E
Internal piping diagram
Measuring unit 2
Measuring unit 1
Sampling cellNote)2 cells may be used by combination of range.
OUTLET INLET OUTLET INLET
Sampling gas inlet
Sampling gas outlet
For measuring unit 1
Sampling gas inlet
Sampling gas outlet
For measuring unit 2
Built-inO2 sensor
Built-inO2 sensor
Note) When the purge gas inlet is provided, the piping to measuring unit 2 is built inside.
Correspondence of measured components and measuring units
Measuring components
1-component Analyzer for NO, SO2, CO2, CO and CH4
2-component Analyzer for NO-SO2 and CO2-CO
2-component Analyzer for NO-CO
3-component Analyzer for NO-SO2-CO
4-component Analyzer for NO-SO2-CO2-CO
Measuring unit 1
Each component NO-SO2CO2-CO
NO
NO-SO2
NO-SO2
Measuring unit 2
None
None
CO
CO
CO2-CO
Note) When there are two measuring units, the built-in O2 sensor must be connected to the measuring unit 2.
3 - 6 IM 11G02N01-01E
Example of connecting each measuring unit
NO2/NOconverter
Note) The NO2/NO converter is used when NO measurement is used for NOx measurement.
NO2/NOconverter
Note) The NO2/NO converter is used when NO measurement is used for NOx measurement.
? Two pair of gas inlet/outlet - (2)
IR400
? One pair of gas inlet/outlet
Filter
Flow meter0.5L/min
IR400
Exhaust
Release to atmosphere.
Sampling gas
Filter
Flow meter0.5L/min
Sampling gas inlet
Sampling gas outlet
Sampling gas inlet
Sampling gas outlet
IR400
Exhaust
Release to atmosphere.
Sampling gas inlet
Sampling gas outlet
Sampling gas inlet
Sampling gas outlet
Exhaust
Release to atmosphere.
Sampling gas
Filter
Flow meter 0.5L/min
Sampling gas
FilterFlow meter0.5L/min
? Two pair of gas inlet/outlet - (1)
Filter
For measuring unit 1
Sampling gas inlet
Sampling gas outlet
For measuring unit 2
For measuring unit 1
For measuring unit 2
3 - 7IM 11G02N01-01E
Analyzer without O2
measurement
N2 gas
Gas with concentra-tion of 90% or moreof full scale
Analyzer with external zirconia O2sensor
Dry air or atmospheric air (Note)
Gas with concentration of 90% ormore of full scale
1 to 2% O2
Zero gas
Span gas otherthan for O2
measurement
Span gas for O2measurement
3.4 Sampling3.4.1 Conditions of sampling gas
1. Dust contained in the sampling gas should be completely removed with a filter. For the finalstage filter, use a filter that allows removing dust particles of 0.3mm.
2. Dew point of sampling gas must be lower than the ambient temperature to avoid occurrenceof drain in the gas analyzer. If vapor is contained in the sampling gas, dew point should belowered to 0°C by using a dehumidifier.
3. If SO3 mist is contained in the sampling gas, use a mist filter or cooler to remove SO3 mist.Other mists should be removed by using a mist filter or cooler.
4. Corrosive gases such as Cl2, F2 and HCl, if they are contained in the sampling gas in consid-erable amounts, will shorten the life of instruments.
5. Temperature of sampling gas should be within 0 to 50°C. Provide a means that preventsentry of hot gas directly into the instrument.
3.4.2 Sampling gas flow
Flow of sampling gas should be 0.5 ± 0.2L/min.Avoid flow fluctuation during measurement.Observe the flow reading by a flowmeter provided as shown in the example of the samplingsystem configuration (Item 3.4.6).
3.4.3 Preparation of standard gas
Routine calibration is required by standard gas for keeping this instrument under normal operationcondition (once a week). Prepare a standard gas cylinder for zero calibration and span calibration.
3.4.4 Reduction of moisture interference
NO and SO2 measurement is subject to moisture interference.
As shown by the configuration example on the next page, provide a device for humidifying zerocalibration gas, thus controlling the moisture content at a constant level (moisture content insample gas should also be controlled here) in configuring a sampling system. That allows thesame moisture content as in the case of measurement to be contained in zero gas for calibration.
3.4.5 Purging of instrument inside
The inside of instrument need not be purged generally except for the following cases.
1. A combustible gas component is contained in sample gas.
2. Corrosive gas is contained in the atmospheric air at the installation site.3. The same gas as the sample gas component is contained in the atmospheric air at the installa-
tion site.
In such cases as above, the inside of analyzer should be purged with the air for instrumentationor N
2. Purging flow rate should be about 1L/min.
If dust or mist is contained in purging gas, it should be eliminated completely in advance.
Analyzer with built-in O2
sensor
N2 gas
Gas with concentration of90% or more of full scale
Gas with concentration of90% or more of full scale oratmospheric air (21%)
3 - 8 IM 11G02N01-01E
(1) Gas extractor
15
Drain Exhaust
2-way solenoid valve
Air
ZERO NO SO2 CO O2CO2
NO/N2
SO2/N2
CO2/N2
CO/N2
O2/N2
Infrared gas analyzermain unit CO,CO2 (O2)
IR400
or larger
(10)ZirconiaO2 senor
φ10/φ8Teflontube
(ZX8D)
(2) Mist filter
(3) Safety drain
trap
(4) Gas aspirator
Electronic gas cooler
(5)
(8) Flowmeter
(7) Membrane filter
(7) Membrane filter
(11)NO2/NOconverter
(6) Solenoid valve
Pressure reducing valve
(9) Standard gas
Infrared gas analyzermain unit NO, SO2
Description
Gas extractor with a heatingtype stainless steel filter of stan-dard mesh 40 m
Removes drain, mist, and dust.
The safety drain trap dividedinto two rooms for positive andnegative pressure. It monitorsand adjusts the sample gas pres-sure.
For aspiration of sample gas
Dries the moisture in samplegas to a dew point of approx.2˚C.
Used for introducing calibra-tion gas.
PTFE filter used to eliminatefine dust particles and permitmonitoring of dust adheringcondition on the front panel ofthe gas analyzer.
Name
(8) Flowmeter
(9) Standard gas
(10) Zirconia O2
sensor
(11) NO2/NO con-verter
Description
Adjusts and monitors the flowrate of sample gas.
Reference gas used for calibrat-ing zero and span of the ana-lyzer. Total 6 cylinders requiredfor zero gas air, span gas NO,SO2, CO, CO2 and O2.
External zirconia oxygensensor used for measuring theoxygen concentration insample gas.(This is not necessary in casewhen O2 sensor is built-in.)
Added to NOx analyzer.A special catalyst material forefficient conversion of NO2
gas to NO is used.
Name
(1) Gas extractor
(2) Mist filter
(3) Safety drain trap
(4) Gas aspirator
(5) Electronic gascooler
(6) Solenoid valve
(7) Membrane filter
3.4.6 Pressure at sampling gas outlet
Pressure at the sampling gas outlet should be adjusted to atmospheric pressure.
3.4.7 Example configuration of gas sampling system
The following illustrates a typical system configuration for five component gas measurement formonitoring combustion exhaust gas from boiler, refuse incinerator, etc.
Contact Yokogawa for system configuration matching the particular use or further information.
3 - 9IM 11G02N01-01E
3.5 Wiring3.5.1 Power inlet
The power inlet is provided at the rear panel.
Connect supplied power cable to this power inlet.
Input/output terminal connector
Power inlet
Rear panel
Grounding 2-pole plug
Exclusive cable (1 meter long)
Analyzer Input/output terminal module
Ferrite core (to analyzer side)
• Avoid installing this instrument near an electrical unit (high frequency furnace or electric welder)that generates much electrical noise. If using theinstrument near such a noise generating unit is
unavoidable, use a different power line to avoid noise.
• Mount a noise suppressor such as varister or spark killeras shown at right figure to the noise generating unit whennoise is generated from relays or solenoid valves.Mount the suppressor near the noise generating source,or it will have no effect.
3.5.2 Input/output terminal module
This analyzer should be connected to the input/output terminal module by supplied exclusivecable.
Plug this cable connector into the receptacle at the rear panel of the analyzer and the receptacle onthe PC board of the input/output module.
Connect the exclusive cable so that the ferrite core attached to the cable comes to the analyzerside.
Main unitpower supply
Varistor or spark killer
Install (connect)near the source.
Noise generating
source
When noise source is in the vicinity
3 - 10 IM 11G02N01-01E
(1) Analog output signal (AO): terminal block 1, 1 to 20, terminal block 2, 3 to 6.
Output signal: 4 to 20 mA DC or 0 to 1 V DC (selected when ordering)
Non-insulated output
Allowable load: 4 to 20 mA DC, 550Ω or less
0 to 1 V DC, 100kΩ or more
• Analog output is provided from each terminal corresponding to the channel displayed in themeasurement screen.
All of analog output signals for the instrument are not isolated. It is recommended to isolate signals individually to prevent interference from unnecessary signals or to prevent external interference, especially leading the cable of more than 30 meters or to outdoor.
(2) O2 sensor input: terminal block 2, 1 to 2.
Input signal:
External zirconia O2 analyzer: Zirconia O
2 sensor signal (ZX8Dp C output)
External O2 analyzer: 0 to 1 V DC (DC input resistor of 1MΩ or more)
• It is used when the external zirconia O2 analyzer or external O2 analyzer is specified as order.
• To connect to the output of the external Zirconia analyzer or external O2 analyzer preparedseparately.
• In case of an external O2 analyzer, input a signal of 0 to 1 V DC with respect to O2 full scaleof the analyzer.
• In case of built-in O2 analyzer, do not use the terminals.
O2 sensor input is not isolated. It is recommended to isolate when an external O2 analyzer is
installed apart from this analyzer. Zirconia O2 sensor Yokogawa make IR400 should be
installed at a location that is as close to this instrument as possible.
(3) Contact input (DI): terminal block 2, 1 to 20, terminal block 3, 5 to 10.
• It is for a contact input at no voltage. An input is provided when switching to short circuit(on) or open (off).
• No voltage is applied to the terminals.
(4) Contact output (DO): terminal block 3, 11 to 20, terminal block 4 and terminal block 5
• Contact rating: 250 V AC/2 A, load resistance
• An output is for a relay contact output. An output is provided when switching to conductive(on) or open (off).
Wiring of analog output signal, O2 sensor input and contact input should be fixed separatelyfrom the wiring of power supply and contact output.
Note) To avoid the effect of noise generated from external units, be sure to ground theanalyzer main unit. Continue between the I/O module mounting plate and the paneland connect the panel casing to the same ground as the analyzer.
3 - 11IM 11G02N01-01E
(5) List of terminal blocks
Terminal block 1<TN1>
11
12
13
14
15
16
17
18
19
20
1
2
3
4
5
6
7
8
9
10
(M3.5 screw)
Ch9 output(Ch9_OUT)
Ch10 output(Ch10_OUT)
Ch8 output(Ch8_OUT)
Ch7 output(Ch7_OUT)
Ch6 output(Ch6_OUT)
Ch1 output(Ch1_OUT)
Ch2 output(Ch2_OUT)
Ch3 output(Ch3_OUT)
Ch4 output(Ch4_OUT)
Ch5 output(Ch5_OUT) +
—+—
+—
+—
+—
+—
+—
+—
+—
Terminal block 2<TN2>
11
12
13
14
15
16
17
18
19
20
1
2
3
4
5
6
7
8
9
10
(M3.5 screw)
Ch4 remote rangeswitch input(R_RNG_Ch4)
Ch5 remote rangeswitch input(R_RNG_Ch5)
Ch3 remote rangeswitch input(R_RNG_Ch3)
Ch2 remote rangeswitch input(R_RNG_Ch2)
Ch1 remote rangeswitch input(R_RNG_Ch1)
Ch12 output(Ch12_OUT)
Ch11 output(Ch11_OUT)
Unassigned
Unassigned
O2 sensor input(O2_IN)
Note 1
Terminal block 3<TN3>
11
12
13
14
15
16
17
18
19
20
1
2
3
4
5
6
7
8
9
10
(M3.5 screw)
Ch4 range identificationcontact output (RNG_ID Ch4)
Ch5 range identificationcontact output (RNG_ID Ch5)
Ch3 range identificationcontact output (RNG_ID Ch3)
Ch2 range identificationcontact output (RNG_ID Ch2)
Ch1 range identificationcontact output (RNG_ID Ch1)
Auto calibration remote start
input (R_CAL)
Average value reset input (RESET)
Remote hold input (R_HOLD)
Unassigned
Unassigned
Terminal block 4<TN4>
11
12
13
14
15
16
17
18
19
20
1
2
3
4
5
6
7
8
9
10
(M3.5 screw)
Unassigned
Unassigned
Unassigned
Auto calibration status/contact output (CAL)
Calibration errorcontact output
(CAL_ALM)
Unassigned
Unassigned
Note 1 : For external O2 sensor input.
Instrument errorcontact output
(FAULT)
Pump ON/OFF contact output (PUMP)
Peak count alarm output (PEAK_ALM)
Connector<CN3>
Solenoid valve drive signal output for calibration(Transister output)
<D-sub 9P>
Contact output for zero calibration
Contact output forsample gas selection
Contact output for Ch1 span calibration
Contact output for Ch2 span calibration
Contact output for Ch3 span calibrationContact output for Ch4 span calibrationContact output for Ch5 span calibration
Power for relay drive5V DC
1
2
3
4
5
6
7
8
9
Terminal block 5<TN5>
11
12
13
14
15
16
17
18
19
20
1
2
3
4
5
6
7
8
9
10
(M3.5 screw)
Unassigned
Alarm 6 output (ALM_6)
Alarm 5 output (ALM_5)
Alarm 4 output (ALM_4)
Alarm 3 output (ALM_3)
Unassigned
Alarm 2 output (ALM_2)
Alarm 1 output (ALM_1)
3 - 12 IM 11G02N01-01E
(6) Description on terminal block
Terminal block 1<TN1>
11
12
13
14
15
16
17
18
19
20
1
2
3
4
5
6
7
8
9
10
(M3.5 screw)
Ch9 output(Ch9_OUT)
Ch10 output(Ch10_OUT)
Ch8 output(Ch8_OUT)
Ch7 output(Ch7_OUT)
Ch6 output(Ch6_OUT)
Ch1 output(Ch1_OUT)
Ch2 output(Ch2_OUT)
Ch3 output(Ch3_OUT)
Ch4 output(Ch4_OUT)
Ch5 output(Ch5_OUT) +
—
+—
+—
+—
+—
+—
+—
+—
+—
+—
+—
Terminal block 2<TN2>
11
12
13
14
15
16
17
18
19
20
1
2
3
4
5
6
7
8
9
10
(M3.5 screw)
Ch4 remote rangechangeover input(R_RNG_Ch4)
Ch5 remote rangechangeover input(R_RNG_Ch5)
Ch3 remote rangechangeover input(R_RNG_Ch3)
Ch2 remote rangechangeover input(R_RNG_Ch2)
Ch1 remote rangechangeover input(R_RNG_Ch1)
Ch12 output(Ch12_OUT)
Ch11 output(Ch11_OUT)
Unassigned
Unassigned
O2 sensor input(O2_IN)
Note 1
+—
+—
Note 1: For external O2 sensor input.
Terminal block 1 <TN1>
Terminal block for analog output (non-isolated output)
Between 1–2: Ch5 output
Between 3–4: Ch4 output
Between 5–6: Ch3output
Between 7–8: Ch2 output
Between 9–10: Ch1 output
Between 11–12: Ch10 output
Between 13–14: Ch9 output
Between 15–16: Ch8 output
Between 17–18: Ch7 output
Between 19–20: Ch6 output
Terminal block 2 <TN2>
Between 1–2: O2 sensor input
(For input of Fuji’s zirconiaoxygen sensor or externallyoxygen sensor. Must not be usedunless external O
2 sensor is
provided.)
Between 3–4: Ch12 output
Between 5–6: Ch11 output
Between 7–10 For internal connection. Must notbe wired. (Must not be used asjunction terminal).
Between 11–12: Ch5 remote range switch input
Between 13–14: Ch4 remote range switch input
Between 15–16: Ch3 remote range switch input
Between 17–18: Ch2 remote range switch input
Between 19–20: Ch1 remote range switch input
Action of remote range switchHigh range is selected when open.Low range is selected when short-circuited. For details of action, see“Item 6.1 Switch of range.”
3 - 13IM 11G02N01-01E
Terminal block 3 <TN3>
Between 1–4: For internal connection. Mustnot be wired. (Must not be usedas junction terminal.)
Between 5–6: Remote hold input. No holdwhen open. Output hold whenshort-circuited.
For details, refer to “Item 6.7Parameter setting, Output Hold”.
Between 7–8: Average value reset input. short-circuitting the contact input (forat 1.5 sec min.) resets O
2 average
and O2 corrected average simulta-
neously. Opening it restarts theaverage value.
For details, refer to “Item 6.7Parameter setting, Average ValueResetting”
Between 9–10: Automatic calibration remotestart input
After shorting for 1.5 sec. ormore, automatic calibration isstarted by the opening inputwhether the automatic calibrationsetting is ON/OFF.
For details, refer to “Item 6.4Setting of auto calibration”
Between 11–12: Ch5 range identification contactoutput
Between 13–14: Ch4 range identification contactoutput
Between 15–16 Ch3 range identification contactoutput
Between 17–18: Ch2 range identification contactoutput
Between 19–20: Ch1 range identification contactoutput
Action of range identification signalRange identification contact isconductive at low range and openat high range.
Terminal block 3<TN3>
11
12
13
14
15
16
17
18
19
20
1
2
3
4
5
6
7
8
9
10
(M3.5 screw)
Ch4 range identificationcontact output (RNG_IDCh4)
Ch5 range identificationcontact output (RNG_IDCh5)
Ch3 range identificationcontact output (RNG_IDCh3)
Ch2 range identificationcontact output (RNG_IDCh2)
Ch1 range identificationcontact output (RNG_IDCh1)
Auto calibration remote start
input (R_CAL)
Average value reset input (RESET)
Remote hold input (R_HOLD)
Unassigned
Unassigned
3 - 14 IM 11G02N01-01E
Terminal 4 <TN4>
Between 1–2: Peak count alarm contact output
It is conductive when peak countexceeds the setting time. Itremains open below the settingtime. For setting and operation,refer to “Item 6.6 Peak alarmsetting”.
Between 3–4: Contact output of auto calibrationstatus
When the auto calibration iscarried out and remote hold isON, it is conductive. Remainsopen otherwise.
Between 5–6: Pump ON/OFF contact output
Used when turning ON/OFF thepump. It is open during auto andmanual calibration status andconductive during measurement.
Between 7–8: Calibration error contact output
It is conductive when an erroroccurs during zero calibration orspan calibration. It is normallyopen.
Between 9–10: It is conductive when an erroroccurs to the analyzer unit. It isnormally open.
Between 11–20: For internal connection, wiring isnot allowed. (Do not use it asjunction terminal).
Terminal block 4<TN4>
11
12
13
14
15
16
17
18
19
20
1
2
3
4
5
6
7
8
9
10
(M3.5 screw)
Unassigned
Unassigned
Unassigned
Auto calibration status/contact output (CAL)
Calibration errorcontact output
(CAL_ALM)
Unassigned
Unassigned
Instrument errorcontact output
(FAULT)
Pump ON/OFF contact output (PUMP)
Peak count alarm output (PEAK_ALM)
3 - 15IM 11G02N01-01E
Terminal 5 <TN5>
Between 2, 3 and 4:
Ch3 alarm output
When the output exceeds the set value,it is conductive between 2 and 3, andopen between 3 and 4. Otherwise, it isopen between 2 and 3 and conductivebetween 3 and 4.
Between 5, 6 and 7:
Ch2 alarm output
When the output exceeds the set value,it is conductive between 5 and 6, andopen between 6 and 7. Otherwise, it isopen between 5 and 6, and conductivebetween 6 and 7.
Between 8, 9 and 10:
Ch1 alarm output
When the output exceeds the set value,it is conductive between 8 and 9, andopen between 9 and 10. Otherwise, itis open between 8 and 9.
Between 12, 13 and14:
Analyzer unit power OFF output
When the analyzer unit is turned ON,it is conductive between 12 and 13,and open between 13 and 14. Whenthe analyzer unit is turned OFF, it isopen between 12 and 13, and conduc-tive between 13 and 14.
Between 15, 16 and 17:
Ch5 alarm output
When the output exceeds the set value,it is conductive between 15 and 16,and open between 16 and 17. Other-wise, it is open between 15 and 16,and conductive between 16 and 17.
Between 18, 19 and 20:
Ch4 alarm output
When the output exceeds the set value,it is conductive between 18 and 19,and open between 19 and 20. Other-wise, it is open between 18 and 19,and conductive between 19 and 20.
For detailed action of the alarmcontact, refer to “Item 6.3 Alarmsetting”.
Terminal block 5<TN5>
11
12
13
14
15
16
17
18
19
20
1
2
3
4
5
6
7
8
9
10
(M3.5 screw)
Unassigned
Power disconnectionalarm output (POWER_OFF)
Ch5 alarm output (ALM_Ch5)
Ch4 alarm output(ALM_Ch4)
Ch3 alarm output (ALM_Ch3)
Ch2 alarm output (ALM_Ch2)
Ch1 alarm output (ALM_Ch1)
Unassigned
3 - 16 IM 11G02N01-01E
Connector <CN3>
Connector <CN3> provides outputs in combination with calibration action during auto calibrationand manual calibration.
An output is from a transistor (ratings: 5 V/50 mA).
A transistor is turned ON before starting each calibration.
Sample selection output is ON during measurement and OFF during calibration.
If calibration is not performed, the other transistors are OFF.
In case of auto calibration, sequential output is ON/OFF according to the setting.
Refer to “Item 6.4 Setting of auto calibration”.
Note) No. 9 pin is for solenoid valve ON/OFF relay drive power (5 V DC/0.5 A, max).Use No. 9 with reference to the diagram.
Connector<CN3>
(Transister output)
<D-sub 9P>
Contact output for zero calibration
Contact output forsample gas selection
Contact output for Ch1 span calibration
Contact output for Ch2 span calibration
Contact output for Ch3 span calibrationContact output for Ch4 span calibrationContact output for Ch5 span calibration
Power for relay drive 5V DC
1
2
3
4
5
6
7
8
9
Solenoid valve drive signal output for calibration
Cable
I/O terminal module Relay board
Contact relay
CN3
5V
SV1 to SV4: solenoid valves
Solenoid valve drive power
SV1
SV2
SV3
SV4
1
3
4
5
9
Standard gas for calibration
<Gas sampling system>
Refer to Item 3.4.7 Example configuration of gas sampling system .
<Electrical system>
Example of using solenoid valve drive signal output for calibration
Relay board and exclusive cable (D-sub 9p straight cable: 1.5 meters) are available on request.
SV4
Ch2span
SV3
Ch1span
SV2SV1
Zero
Membrane filter
Electronic gas cooler
Flowmeter
3 - 17IM 11G02N01-01E
¥ Zero calibration
¥ Span calibration
Zero calibrationoutput
Zero calibrationoutput
Pump ON/OFF contact
Pump ON/OFF contact
Sample selectionoutput
Ch1 to 5 span calibration output
Ch1 span calibration
Output hold function
Sample selectionoutput
(with hold ON setting)
ZERO ENTENT
Hold extension time.
Calibration gas flow
Calibration gas flow
Output signal hold
Output signal hold
Calibration end
Calibration end
on
offon
offon
off
off
Output hold function
(with hold ON setting)
SPAN ENTENT
Hold extemsion time.
Note)
When selecting Ch2 using and keys.
Ch2 span calibration
The hold extension time depends on the gas flow time of the automatic calibration settings.
(7) Timing of solenoid valve drive signal for calibration
1. Manual calibration (see “Item 6.9 Calibration”.)
3 - 18 IM 11G02N01-01E
Ch1 span calibration output
Ch1 span calibration
Ch3 span calibration
Ch2 span calibration
Ch4 span calibration
Ch5 span calibration
Ch2 span calibration output
Ch3 span calibration output
Ch4 span calibration output
Ch5 span calibration output
Pump ON/OFF contact
Sample selection output
Zero calibration output
Automatic calibration contact
Output hold function
Zero gas
350 s
Ch1 span gas350 s
Holdextension time.
Automaticcalibration
startZero calibration
(with hold ON setting)
2. In case of automatic calibration (example shown in Item 6.4.1, Automatic calibration settings)
4 - 1IM 11G02N01-01E
4.1 Preparation for operation
(1) Tube and wiring check
Double-check if tubes of the gas sampling and exhaust ports are correctly connected.
Double-check for proper wiring.
4.2 Warm-up operation and regular operation
(1) Operation procedure
1) Turn ON the power switch on the front panel of the analyzer unit.
The measurement screen appears on the front display panel in 1 or 2 seconds.
2) Wait for about 4 hours until the instrument is warmed up.
About 4 hours are required until the instrument allows accurate measurement.
Note) When in warm-up, the concentration reading may be beyond.
upper limit of range or
lower limit of range.
But, it is not an error.
3) Setting of various set values
Perform the various settings according to “Chapter 6. Setting and Calibration”.
4) Zero calibration and span calibration
Perform zero calibration and span calibration after warm-up operation.
Refer to “Chapter 6.9. Calibration”.
5) Introduction and measurement of measuring gas
Introduce the measuring gas into the analyzer unit before starting measurement.
4. OPERATION
≈
Blank Page
5 - 1IM 11G02N01-01E
Name
(1) MODE key
(2) SIDE key
(3) UP key
(4) DOWN key
Description
Used to switch the mode.
Used to change the selected item(by moving the cursor) andnumeral digit.
Used to change the selected item(by moving the cursor) and toincrease numeral value.
Used to change the selected item(by moving the cursor) and todecrease numeral value.
Name
(5) ESCAPE key
(6) ENTRY key
(7) ZERO key
(8) SPAN key
Description
Used to return to a previous screenor cancel the setting midway.
Used for confirmation of selecteditems or values, and for executionof calibration.
Used for zero calibration.
Used for span calibration.
This section describes the display unit and operation panel of the analyzer unit. It also explains the nameand description of function on the operation panel.
5.1 Name and description of operation panel
Fig. 5-1
MODE
POWER
ESC ZERO
ENT SPAN
5. DESCRIPTION OF DISPLAY AND OPERATION PANELS
5 - 2 IM 11G02N01-01E
Fig. 5-2
• Measurement mode
• Measurement mode
• Measurement mode
• User mode
• Switch Ranges
• CalibrationParameters
• Alarm Setting
• Setting of Auto Calibration
• Setting of Peak Alarm
• Parameter Setting
* 1) The panel configuration is changed depending on the display channel. (The measurement mode screen can be viewed by scrolling the arrow key up and down).
* 1
* 2) The peak alarm setting is added according to the code symbol when CO and O2 components exist.
ZERO Calibration
SPAN Calibration
• Selection of items Start Time Cycle Flow Time ON/OFF Auto zero Calibration Run / stop
• Setting of Auto Zero calibration
ESC
SPAN
ZEROMODE
ESCMODE
ESC
ESC
ESC
ESC
ESC
ESC
ESC
ESC
ESC
MODE
MODE
• User modeSwitch RangesCalibration ParametersAlarm SettingSetting of Auto CalibrationSetting of Auto Zero calibrationSetting of Peak AlarmParameter Setting
• Selection of itemsCalibration valueZero calibration Calibration range Auto calibration components / range
• Selection of itemsStart TimeCycleFlow TimeAuto calibration ON/OFFAuto Calibration Run / stop
• Selection of itemsPeak Alarm ON/OFFPeak ValuePeak CountHysteresis
• Selection of itemsCurrent time : Current time settingKey lock : Key lock ON/OFFOutput Hold : ON/OFFReset Av. Output : Average value
resettingResponse time : Response time (filter)Average Period : Average Period
settingBacklight Timer : ON/OFF, timeup timeTo Maintenance mode: Maintenance mode
(entry of password)
5.2 Overview of display and operation panels
5 - 3IM 11G02N01-01E
5.3 Outline of display screen
(1) Measurement mode screen (appears when the power is turned ON)
The measurement screen depends on the number of components. The following screen configura-tion as shown as an example is for NO, SO2, CO2, CO and O2 (output: 12 channel).
Fig. 5-3 Name and function of measurement mode screen
No.
(1)
(2)
(3)
(4)
Function
Displays component of instanta-neous value, corrected instanta-neous value, corrected averagevalue, etc.
Displays measured value ofconcentration.
Displays range values.
Displays unit with ppm andvol%.
Description
Displays peak alarm component.
Displays peak alarm concentra-tion display.(Upper limit value)
Displays the alarm timesexceeding the peak value.
Displays units of peak alarm withtimes/H.
No.
(5)
(6)
(7)
(8)
Name
Peak alarmcomponentdisplay
Peak alarmconcentrationdisplay
Peak alarmtimes
Peak alarmunit display
Name
Componentdisplay
Concentrationdisplay
Range display
Unit display
CV.AV.
* corrected instantaneous value ..... CV.
** corrected average value...............
* For outputs of more than 5 channels, scroll the arrow key or to view.
C h
C h
C h
C h
C h
(1) (2)
(3) (4)
(1) (2)
(4)(5) (6) (3) (7) (8)
C h
C h
C h
C h
5 - 4 IM 11G02N01-01E
(2) Setting/selection screen
The setting/selection screen is configured as shown below:
• In the status display area, the current status is displayed.
• In the message display area, messages associated with operation are displayed.
• In the setting item and selection item display area, items or values to be set are displayed, as required. To work on the area, move the cursor to any item by using , and keys.
Cursor
Status display area
• LCD screenMessage display area
Setting item/selection itemdisplay area
Fig. 5-4
• Instantaneous value and concentration value:
The concentration display of Ch (component) where sampling components such as “CO2”, “CO”
or “O2 are displayed in the component display, indicates current concentration values of the
measured components contained in gas that is now under measurement.
• O2 correction concentration values:
Ch components where “cv**” is displayed as “cv CO” in the component display are calculatedfrom the following equation, by setting sampling components, O
2 instantaneous/concentration
values and O2 correction reference value (see item 6.8).
Correction output=
The converted sampling components are NOX, SO
2 and CO only.
••••• O2 correction concentration average value:
In the Ch (component) and O2 average value where “ **” is displayed as “ CO” in thecomponent display, a value obtained by averaging O2 correction concentration value or O2
average value in a fixed time is output every 30 seconds.
Averaging time can be changed between 1 minute and 59 minutes or 1 hour and 4 hours accord-ing to the average time settings (See 6.7, Parameter setting).
(The set time is displayed as “1h” , for instance, in the range display.)
* The measurement ranges of O2 correction concentration value and O2 correction concentra-tion average value are the same as that of the measuring components. Also, the measure-ment range of O2 average value is the same as that of O2.
CVAV
CVAV
On:The value of the O2 correction referance value
(Value set by application)
Os: Oxygen concentration (%)
Cs: Concentration of relevant measured component.Note that Os does not exceed the O
2 limit value
21 - On21 - Os
× Cs
5 - 5IM 11G02N01-01E
(3)
Con
tent
s of
mea
sure
d ch
anne
l (C
h)
T
he fo
llow
ing
tabl
e gi
ves
mea
sure
men
t cha
nnel
s an
d th
eir
cont
ents
acc
ordi
ng to
the
sym
bols
.
Cor
rect
CO
O2
av.
Corre
ct NO
x av.
SO
2
CO
SO
2
CO O2
SO
2
SO
2
SO
2
O2
SO
2
O2
CO
SO
2
CO
COO2
O2
SO
2
CO
NO
x
NO
xN
Ox
NO
CO
2
NO
NO
Corre
ct NO
x av.
Corre
ct S
O2 a
v.
CO
CO
2
CO
CH
4
CO
NO
x
NO
SO
2
CH
4
NO
x
CO
2
CO
2
NO
NO
NO
CO
2C
OO2
O2
Corre
ct C
O a
v.
Corre
ct NO
x av.
Corre
ct C
O a
v.
Corre
ct S
O2 a
v.
Corre
ct NO
x av.
Cor
rect
NO
x
Corre
ct NO
x av.
Corre
ct NO
x av.
O2
av.
Cor
rect
SO
2
Cor
rect
NO
x
Cor
rect
NO
x
Cor
rect
CO
Corre
ct C
O a
v.
Cor
rect
SO
2
O2
av.
O2
av.
O2
anal
yzer
O2
CO
O2
av.
Cor
rect
NO
x
Cor
rect
CO
Cor
rect
SO
2
CO
2
CO
CO
O2
O2
CO O2
CO O2
O2
CO
2
O2
O2
av.
Cor
rect
SO
2
Cor
rect
CO
Corre
ct C
O a
v.
O2 a
v.1,
2, 3
1, 2
, 3
1, 2
, 3
1, 2
, 3
1, 2
, 3
1, 2
, 3
1, 2
, 3
1, 2
, 3
1, 2
, 3
1, 2
, 3
1, 2
, 3
1, 2
, 3
1, 2
, 3
1, 2
, 3
-K-B -G-D -H -J-F-A-K-H -J -L-G
N
-L -H-D -J -K
Ch1
0C
h9C
h11
Ch7
Ch6
CH
8
-F
O2 a
v.
Ch1
2
-B
Cod
e sy
mbo
l
Mea
sura
ble
com
pone
nt
-A -D-C
O2
corr
ectio
nC
h4C
h3C
h1C
h2C
h5
Out
put c
orre
spon
ding
to c
hann
els
Not
spe
cifie
d
/K/K/K /K /K/K/K
Not
spe
cifie
d
Not
spe
cifie
d
Not
spe
cifie
d
Not
spe
cifie
d
Not
spe
cifie
d
/K
exce
pt /K
exce
pt /K
exce
pt /K
/K
exce
pt /K
exce
pt /K
Not
spe
cifie
d
SO
2
Not
spe
cifie
d
Not
spe
cifie
d
Corre
ct C
O a
v.
Corre
ct S
O2 a
v.
NNN N NN
NO
Not
spe
cifie
dN N N
Corre
ct S
O2 a
v.
Corre
ct C
O a
v.
O2
av.
(not
e):
As
for
the
NO
met
er w
ithin
this
ran
ge, t
he d
ispl
ay o
n th
e in
dica
tor
beco
me
NO
x.T
he p
eak
coun
t ala
rm b
ecom
es a
con
tact
out
put.
The
"co
rrec
t" m
eans
O2
corr
ectio
n.T
he "
av."
mea
ns a
vara
ge v
alue
.
T08
.EP
S
5 - 6 IM 11G02N01-01E
5.4 General operation• Measurement mode
The measurement mode can be displayedup to 5 channels in a single screen. If 5channels or more are to be displayed in asingle screen, press the or key toscroll the channel one by one.
• User mode displays;
Changeover of Range
Setting about Calibration
Alarm Setting
Setting of Auto CalibrationSetting of Auto Zero Calibration
Setting of Peak Alarm
Parameter Setting.
For the setting contents, refer to“Chapter 6. Setting and calibration”.
Zero calibrationSee 6.9.1.
Span calibrationSee 6.9.2.
C h O2
0-25 vol%
C h CO0-100 ppm
C h CO20-10 vol%
C h SO20-100 ppm
C h NOX0-100 ppm0 0.
0 0.0 0 0
0 0.0 0 0.
.
C h O2
0-25 vol%
C h CO0-100 ppm
C h CO20-10 vol%
C h SO20-100 ppm
C h NOX0-100 ppm0 0.
0 0.0 0 0
0 0.0 0 0.
.
ZERO
SPAN
C hAV. O2
PEARK COUNTER
CO 500ppm 0 times/H
0-25 vol%
C h CO
0-100 ppm
C h SO2
0-100 ppm
C h NOX0-100 ppm
0 0 0.0 0.
0 0.0 0.
CV.AV.
CV.AV.
CV.AV.
ESCMODE
6 - 1IM 11G02N01-01E
MODE
( ) ENT
Range switch
or previous screen
6.1 Switch of range6.1.1 Setting of range switch mode
Set the range switch mode as follows.
(1) Press the MODE key in measurement mode to
display the User mode screen.
(2) Move the cursor to “Switch Ranges” andpress the ENT key.
(3) The “Channel Selection” screen appears.Move the cursor by pressing the orthe key on the channel selectionscreen that appears, and select Ch(component).
(4) Then press the ENT key.
(5) Selected range switch mode is highlighted.Press the or the key to select adesired switch mode.
(6) Then press the ENT key to confirm the
selection.If “MR” is selected, the cursor moves to“Range Switch.”
MR: Select a desired range on this screen.RR: Select a desired range according to
the remote range switch contact input.AR: Automatically switched from Range 1
to Range 2 when the measured concen-tration exceeds 90% of Range 1. Automatically switched from Range 2 to Range 1 when the measured concen-tration becomes smaller than 80% of Range 1.
* Operation set for each Ch only can be performed.
Description of setting
ENT
( ) ENT
6. SETTING AND CALIBRATION
6 - 2 IM 11G02N01-01E
6.1.2 Manual range switch
The range of the measured component can beswitched manually as follows.
(1) Select “MR” as range switch mode, and
then press the ENT key.
(2) Move the highlight of the cursor to rangeselection, and then select a desired range
by pressing the or the key. (The
mark indicates the currently selectedrange.)
(3) Then press the ENT key, and the
measurement is carried out in theselected range.
Note) If “RR” or “AR” is selected asrange switch mode, this operationcannot be performed.
The range for O2 correction value,
O2 correction average value, and O
2
average value is automaticallyswitched if correspondinginstantaneous value range isswitched.
ENT
( ) ENT
End of Range Switch
Press the key to end the setting of range switch mode or range switch operation or stop the operation in the middle, and the setting operation is made invalid and the previous screen appears.
To close the setting
ESC
The range identification contact output corresponding to each Ch (component) is conductive when Range 1 is selected, and open when Range 2 is selected, which is applicable to any of the range switch mode selected.Note that even if the range is switched during the hold of measurement value by remote hold contact input or the hold of measurement value at the time of calibration, the range identifica-tion contact output maintains the contact state immediately before the hold. After stop of the hold, the contact state of the current range is resumed.
Range identification contact operation
6 - 3IM 11G02N01-01E
6.2 Calibration settingThis mode is used to set calibration concentration and actions. The calibration setting involvescalibration concentration, zero calibration, calibration range and auto calibration component.
6.2.1 Setting of calibration concentration
It allows you to set concentrations of the standard gas (zero and span) of each channel used forcalibration.
MODE
( ) ENT
( ) ENT
( ) ENT
(1) During measurement, press the MODE key to
display the User mode.
(2) Point the cursor to “CalibrationParameters” by pressing the or key.
Press the ENT key.
(3) In the “Calibration Parameters” screenthat appears, point the cursor to “Calibra-
tion Value” by pressing the or key. Press the ENT key.
(4) In the “Calibration Concentration ChSelection” screen that appears, point thecursor to Ch you want to set by using the
or key. Press the ENT key.
6 - 4 IM 11G02N01-01E
ENT
End of Calibration
Concentration Setting
NOx, SO2, CO2, CO, CH4, external O2 measurement and buit-in paramagnetic O2 sensor
Span gas: 1 to 105% of full scale (Full scale (FS) is the same as each range value.)
External Zirconia O2 measurement
The setting cannot be performed beyond the range.
Zero gas: 5 to 25 vol%Span gas: 0.01 to 5 vol%
Setting range of values
Cursor for setting value
(5) In the “Calibration Concentration Selection”screen that appears, select any concentrationitem you want to set by pressing the , key.
Then press the ENT key, and the selected
value is highlighted.
(6) In the “Calibration Concentration ValueSetting” screen that appears, enter calibrationgas concentration values (zero and span).For value entry, press the or key, anda 1-digit value increases or decreases. By
pressing the , the digit moves.
After setting, save the entry by pressing theENT key. The saved value becomes valid
from the next calibration process.
Note) Enter settings that correspond to eachrange. If zirconia type is used as O
2
sensor, select 21.00 for the field of Zero(when air is used), and select theconcentration listed on the cylinder if theair contained in a cylinder is used.
To close the calibration concentration value setting process or cancel this mode midway, press the key. A previous screen will return.
To close the setting
ESC
ENT
6 - 5IM 11G02N01-01E
( ) ENT
( ) ENT
MODE
( ) ENT
(1) During measurement, press the MODE keyto display the User mode.
(2) Point the cursor to “Calibration Param-eters” by pressing the or key. Pressthe ENT key.
(3) In the “Calibration Parameters” screen thatappears, point the cursor to “About ZEROCalibration” by pressing the or key.Press the ENT key.
(4) In the “Manual ZERO Calibration ChSelection” screen that appears, point thecursor to Ch (component) you want to setby using the or key. Press the ENT
key.
6.2.2 Setting of manual zero calibration
When zero calibration is made manually, set either all measurement components should becalibrated simultaneously or each component should be calibrated while selecting one by one.
6 - 6 IM 11G02N01-01E
( ) ENT
To close the manual zero calibration setting or to
cancel this mode midway, press the ESC key.
A previous screen will return.
To close the setting
ESC
• When setting all components to “each”:
A single cursor will appear.
Manual Calibration screen
• When setting all components to “at once”:
Cursors will appear at all components where “at once” is set.
End of
Manual Zero Calibration Setting
Whether “each” or “at once” can be determined for each Ch (component).
•Setting “each” Select the Ch (component) on the manual zero calibration screen and then perform zero calibration.•Setting “at once”At a manual zero calibration, zero of Ch (components) for which “at once” was selected can simultaneously be calibrated.* When the cylinder air or atmospheric air is used for the zero gas, select “At once.”
Example
(5) In the “Manual ZERO CalibrationSelection” screen that appears, select “atonce” or “each” by pressing the or
key. When selecting “at once”, theCh (components) to be set can be zero-calibrated at the same time. Whenselecting “each”, either of the Ch (com-ponents) to be selected is zero-calibrated.After setting, press the ENT key.
6 - 7IM 11G02N01-01E
( ) ENT
( ) ENT
MODE
( ) ENT
(1) During measurement, press the MODE key todisplay the User mode.
(2) Point the cursor to “CalibrationParameters” by pressing the or key. Press the ENT key.
(3) In the “Calibration Parameters” screenthat appears, point the cursor to “AboutCalibration Range” by pressing the or
key. Press the ENT key.
(4) In the “Calibration Range Ch Selection”screen that appears, point the cursor to theCh you want to set by pressing the or
key. Press the ENT key.
6.2.3 Setting of calibration range
This mode is used to set if the range of each CH (component) at the calibration (manual calibrationor auto calibration) should be calibrated with a single range or 2 ranges.
6 - 8 IM 11G02N01-01E
(5) On the “calibration range selection”screen that appears, select “both” or“current” by pressing the or the key.
• If “both” is selected, zero or spancalibration is performed with Range 1and Range 2 of the selected Chinterlocked.
• If “current” is selected, zero or spancalibration is performed only for therange displayed when calibration ofselected Ch is performed.
Press the ENT key after the selection, andthe specified calibration is performed.
To perform calibration for “both,” set the same calibration gas concentration for both ranges.
Note
When setting NOx and CO to “both”
Two cursors will appear in both ranges (Ch1 and Ch4).
Manual Calibration screen
( ) ENT
End of Calibtation Range Setting
Ch1NOx
Ch2SO2
Range 1: 0 to 100 ppmRange 2: 0 to 2000 ppm
Range 1: 0 to 100 ppmRange 2: 0 to 2000 ppm
both
current
Ch1: Range 1 and Range 2 are calibrated together with zero and span calibration.Ch2: Only currently displayed range is calibrated with zero and span calibration.
Example
To close “Setting of Calibration Range” or
to cancel this mode midway, press the
ESC key. A previous screen will return.
To close “Setting of Calibration Range”
ESC
6 - 9IM 11G02N01-01E
( ) ENT
( ) ENT
MODE
( ) ENT
6.2.4 Setting of auto calibration component/range
Select the Ch (component) and the range with which auto calibration is to be performed. The Chfor which “AR” has been selected as range switch mode is calibrated in the range set here evenwhen auto calibration is performed.
(1) During measurement, press the MODE keyto display the User mode.
(2) Point the cursor to “Calibration Param-eters” by pressing the or key.Press the ENT key.
(3) In the “Calibration Parameters” screenthat appears, point the cursor to “AutoCalibration Components / Range” bypressing the or key. Press the
ENT key.
(4) In the “Auto Calibration Components /Range” selection screen that appears,point the cursor to the Ch you want to setby pressing the or key. Press the
ENT key.
6 - 10 IM 11G02N01-01E
(5) The cursor next to the range of theselected Ch (component) is highlighted.Select the range to be calibrated mainlyby pressing the or the key.
(6) Then press the ENT key, and calibration isperformed in the selected range.
Auto calibration and the manual calibration
of the component with which “AR” has
been selected as range switch mode are
performed in the range selected here. In this
case, once the calibration is started, the range
is automatically switched, and on completion
of the calibration, the original range is
resumed.
The range identification contact is inter-
locked with the range after the switch.
However, if the hold setting is set to “ON,”
the contact status before calibration is main-
tained.
To close "Auto Calibration Component/range" setting
(7) Press the key in the state described in
(5), and the highlight is switched between“enable” and “disable” auto calibration.
(8) Select “enable” of “disable” by pressingthe or the key.
(9) Then press the ENT key.
Press the key to exit automatic calibra-
tion component/range setting, and the pre-
vious screen appears.
To close the setting
ESC
Operation by setting
Auto calibration is performed under the following rules.
1. Zero calibration is performed at the same time, for the Ch (component) with which “enable”
is selected at the time of auto calibration and auto zero calibration.
2. Span calibration is performed in the order from smallest Ch No., for the Ch (component)
with which “enable” is selected at the time of auto calibration.
ZERO calibration on auto calibration and auto zero calibration of the component with which “enable” is selected are performed in batch irrespective of the description in “6.2.2 Setting of manual zero calibration.”
Note
( ) ENT
End of Auto Calibtation
Range Setting
( ) ENT
End of Auto Calibtation
Component Setting
6 - 11IM 11G02N01-01E
( ) ENT
MODE
( ) ENT
Note
Set the values so that H-limit value > L-limit value and that (H-limit value - L-limit value) > hysteresis.
6.3 Alarm setting6.3.1 Setting of alarm values
The High/Low limit alarm output setting for the measured concentration and power off alarm(alarm 6 only) setting can be made during measurement. Arbitrary 6 alarm contact outputs can beused.
To change alarm setting, set the alarm ON/OFF setting to OFF, and then change the value.
(1) During measurement, press the MODE keyto display the User mode.
(2) Point the cursor to “Alarm Setting” bypressing the or key. Press the ENT
key.
(3) After the alarm No. selection screen hasappeared, point the cursor to the AlarmNo. you want to set by pressing or
key .
Press the ENT key.
(4) After the alarm item selection screen hasappeared, operate the or key untilthe cursor is aligned with a desired itemand press the ENT key.
( ) ENT
6 - 12 IM 11G02N01-01E
Cursor for setting value
ENT
End of Alarm Setting
Note
For 10 minutes after turning on power, the alarm judgment is inactive.
To close the “Alarm Setting” or to cancel this
mode midway, press the key.
A previous screen will return.
To close the "Alarm Setting"
ESC
0% to 100% FS (Settable in each range).
Setting range
Description of setting items
The alarm contact assigned the same number as the alarm is operated accordingly.Channel: Channel setting targeted for issuance of alarm (Power off alarm can be
selected for alarm 6.) One Ch No. can be selected for multiple alarms.H-Limit value: Sets the high limit value (concentration) of alarm.L-Limit value: Sets the low limit value (concentration) of alarm.Kind of Alarm: Selects one of High limit alarm, Low limit alarm, and High limit or Low
limit alarm, HH limit alarm, and LL limit alarm.High, HH ... Alarm contact closes when above H-limit alarm.Low, LL ... Alarm contact closes when below L-limit alarm.High or Low ... Alarm contact closes when above H-limit value or
below lower limit value.If “Power” is selected for Channel, the contact is closed at all times while the power is on irrespective of the setting made here. (Alarm-6 only)ON/OFF: Enables the alarm function if set at ON, or disables the alarm function if set at OFF.* The H-limit value cannot be set below the L-limit value, and the L-limit value cannot be set
above the H-limit value.If it is desired to set the H-limit value below the L-limit value already stored in the memory, reduce the L-limit value beforehand, and vice versa.
When an H-limit alarm occurs, the “H-alarm” message comes on in the field of relevant Ch (component). (“L-alarm” for L-limit alarm, “HH-alarm” for HH limit alarm, and “LL-alarm” for LL limit alarm)
Typical on-screen display when an alarm occurs
C h O20-25 vol%
C h CO0-100 ppm
C h CO20-10 vol%
C h SO20-100 ppm
C h NOX0-100 ppm
0 0.0 0 0 3.
0 0.2 1 0 0.
- - - -H-alarm
(5) After setting, the alarm setting is nowcompleted by pressing the ENT key.
6 - 13IM 11G02N01-01E
Hysteresis (In case of upper limit alarm)
An alarm output is turned ON if measurement value exceeds the upper limit value as shownbelow. Once the alarm output has been turned ON, it is not turned OFF as long as theindication does not fall below the hysteresis width from the upper limit value.
0 to 20% of full scale[% full scale (FS)] represents the percentage with the width of the range of each compo-nent regarded as 100%.
Setting range
Upper limit value
ON (conductive)
OFF (open)
Hysteresis width
AlarmContact output
( ) ENT
ENT
End of Hysteresis Setting
NoteThe hysteresis is common to all alarms(components).
To close the “Hysteresis Setting” or cancel
the mode midway, press the ESC key.
A previous screen will return.
To close "Hysteresis Setting"
ESC
(1) In the “Alarm No. Selection” screen thatappears, point the cursor to “Hysteresis”by pressing the or key. Press the
ENT key.
(2) In the “Hysteresis Value Setting” screenthat appears, enter hysteresis values.
For the value entry, 1-digit value isincreased or decreased by pressing the
or key, and pressing the key
moves the digit. After setting, press theENT key.
6.3.2 Hysteresis setting
To prevent chattering of an alarm output nearthe alarm setting values, set the valueof hysteresis.
6 - 14 IM 11G02N01-01E
6.4 Setting of auto calibration6.4.1 Auto calibration
Auto calibration is automatically carried out at the time when zero calibration and span calibrationare set.
Before changing the setting of auto calibration, set the ON/OFF to OFF.
1. During measurement, press the MODE keyto display the User mode.
2. Point the cursor to “Setting of AutoCalibration” by pressing the or key. Press the ENT key.
3. In the “Setting of Auto Calibration”screen that appears, point the cursor toany item you want to set by pressing the
or key. Press the ENT key.
4. In the “Auto Calibration ParameterSetting” screen that appears, perform thevalue entry or the setting. For the valueentry or setting change, use the or
key. To change the setting, use the key to move the cursor to the right.
After setting, press the ENT key, andsetting of auto calibration is carried out .
( ) ENT
MODE
( ) ENT
ENT
End of Aoto Calibration Setting
To close the "Setting of Auto calibration" or cancel
this mode midway, press the ESC key.
A previous screen will return.
To close "Setting of Auto calibration"
ESC
Description of setting items• Start Time : Setting at the first calibration
(day of the week, hour, minute)• Cycle : A period between the start time of one
calibration and another (unit : hour/day)
• Flow Time : The time required for replacement by calibration gasTime required for replacement of sample gas after the calibration is completed (Set by calibration gas. See the next page.)
• ON/OFF : ON/OFF of auto calibration
6 - 15IM 11G02N01-01E
<Gas flow time> setting
(1) Press the ENT key in a state where thecursor is placed next to “Flow Time,” andthe flow time setting screen shown at rightappears.
(2) On the flow time setting screen thatappears, move the cursor to the gas youwant to change the setting by pressing the
or the key, and then press the ENT
key.
(3) The highlighted value can be changed.Change the value by pressing the orthe key, and then move the cursor tothe right by pressing the key.
(4) After changing the value, press the ENT
key.
(5) Press the ESC key to return to theautomatic calibration setting screen.
Note) Only the Chs used are displayed onthis screen. The Ex. time is theoutput signal hold extension timeafter the completion of calibration.It is valid only when the hold settingis set to “ON.” The Ex. time sethere is also the hold extension timeat the time of manual calibration.
MODE
( ) ENT
ENT
End of Gas Flow Time Setting
6 - 16 IM 11G02N01-01E
Remote start
Whether the auto calibration is set at ON or OFF, an auto calibration is available by keeping theremote start input closed for at least 1.5 seconds.
Remote start input Open
Closed (keep at least 1.5 sec.)
Start TimeCycleFlow Time
ON/OFF
SUN1ZeroCh1 SpanCh2 Span Ch3 Span Ch4 Span Ch5 Span EX. timeON
12:00day350 sec350 sec350 sec350 sec300 sec300 sec300 sec
(An example of “Ch1: through Ch5: enable”, as given in Item 6.2.4 “Auto Calibration Components/range”)
Flow time
: Auto calibration
Sunday12:00
Cycle
Zero calibration
350sec 350sec
Ch1Span calibration
350sec
Ch2Span calibration
Ch3Span calibration
Ch4Span calibration
Monday12:00
Tuesday12:00
Example
In case where auto calibration is carried out at the above setting.
Ch5Span calibration
350sec 300sec 300sec
Replace-ment time
300sec
Cycle : 1 to 99 hours or 1 to 40 days (initial value 7days)Flow time : 60 to 900 sec (initial value 300sec)
Setting range
When an auto calibration starts, the measurement screen automatically appears.Any key operation other than key lock ON/OFF and “Stop Auto Calibration” (see Item 6.4.2) is not permitted during auto calibration. “Stop Auto Calibration” cannot be performed with the key lock to ON. To cancel auto calibration forcedly, set the key lock to OFF and then execute “Stop Auto Calibration”.Turn on the power again after it is turned off (including the case of power failure) at the time set as the next start time in auto calibration, and then repeat it in the set cycle.
Caution••
•
Auto calibration status contact output is closed during auto calibration (including Ex. time), and is open inother cases.
6 - 17IM 11G02N01-01E
( ) ENT
( ) ENT
6.4.2 Forced run/stop of auto calibration
Auto calibration can be performed just once or forcibly stopped while the calibration isperformed.
6.4.2.1 Execution of auto calibration (only once)
(1) Display the User mode screen. Move thecursor to “Setting of Auto Calibration”by pressing the or the key, andthen press the ENT key.
(2) In the “Setting of Auto Calibration” itemselection screen that appears, point thecursor to “Auto Calibration Run” bypressing the or key. Press the ENT
key.
(3) “Run” is highlighted, displaying a mes-sage to confirm the execution of autocalibration. Press the ENT key to executethe auto calibration forcibly, and pressthe ESC key to cancel.
6 - 18 IM 11G02N01-01E
6.4.2.2 Forced stop of auto calibration
This mode is used to stop the auto calibration forcedly.
(1) In the User mode that is displayed, pointthe cursor to “Setting of AutoCalibration” by pressing the or key. Press the ENT key.
(2) In the “Setting of Auto Calibration” itemselection screen that appears, point thecursor to “Auto Calibration Stop” bypressing the or key. Press the ENT
key.
(“Auto Calibration Stop” appears whenthe screen is selected while auto calibration is performed.)
(3) “Stop” is highlighted, displaying amessage to confirm the stop of autocalibration. Press the ENT key to stop theauto calibration forcibly, and press the
ESC key to cancel (not stopped).
( ) ENT
( ) ENT
6 - 19IM 11G02N01-01E
During auto calibration, any key operation is not permitted other than operations such as key lock ON/OFF and “Stop Auto Calibration.”When the key lock is set at ON, even the “Auto Calibration Stop” cannot be used. To stop “Auto Calibration” forcedly, set the key lock to OFF and then execute “Auto Calibration Stop.”
Caution
• Zero calibration
ExampleIn case where setting the auto calibration components (see Item 6.2.4) to “Ch1: enable” and “Ch2: enable”
A message, “Zero cal.” blinks at Ch1 and Ch2.
• Ch1 span calibration
A message, “Span cal.” blinks at Ch1.
• Ch2 span calibration
A message, “Span cal.” blinks at Ch2.
“Auto Calibration” screen
C h O20-25 vol%
C h CO0-100 ppm
C h CO20-10 vol%
C h SO20-100 ppm
C h NOX0-100 ppm9 0 8
0 0.0 0 0.
.0 0.
0 0 0
C h O20-25 vol%
C h CO0-100 ppm
C h CO20-10 vol%
C h SO20-100 ppm
C h NOX0-100 ppm
9 5 0.0 0
0 0 0.
.0 0.
0 0 0
C h O20-25 vol%
C h CO0-100 ppm
C h CO20-10 vol%
C h SO20-100 ppm
C h NOX0-zero100 ppm0 5
0 3..
0 0 0 0.0 0.
2 1 0 2.
.
.
ZERO cal.
ZERO cal.
SPAN cal.
SPAN cal.
6 - 20 IM 11G02N01-01E
6.5 Setting of auto zero calibration6.5.1 Autozero calibration
Auto zero calibration is automatically carriedout at the time when zero calibration is set.Components for which a calibration is to bemade are determined by setting of auto cali-bration component in Item 6.2.4.
Before changing the setting of auto zerocalibration, set the ON/OFF to OFF.
(1) During measurement, press the MODE keyto display the User mode.
(2) Point the cursor to “Setting of Auto ZeroCalibration” by pressing the or key. Press the ENT key.
(3) In the “Setting of Auto Zero Calibration”screen that appears, point the cursor toany item you want to set by pressing the
or key. Press the ENT key.
(4) In the “Auto Zero Calibration ParameterSetting” screen that appears, perform thevalue entry or the setting. For the valueentry or setting change, use the or
key. To change the setting, use the key to move the cursor to the right.
After setting, press the ENT key, and autozero calibration is carried out by theentered setting value.
( ) ENT
MODE
( ) ENT
ENT
End of Auto Zero Calibration Setting
Description of setting items• Start Time : Setting at the first calibration
(day of the week, hour, minute)• Cycle : A period between the start time of one
calibration and another (unit : hour/day)
• Flow Time : The time required for the calibration gas to be replaced in the cell
• ON/OFF : ON/OFF of auto zero calibration
To close the “Setting of Auto Zero Calibration” or cancel this mode midway, press the key. A previous screen will return.
To close "setting of Auto Zero Calibration"
ESC
6 - 21IM 11G02N01-01E
Start timeCycleFlow timeON/OFF
SUN12300ON
12:00hoursec
(An example of “Ch1: through Ch5: enable,” as given in Item 6.2.4 “Auto Calibration Components/range”)
Flow time
: Auto zero calibration
Sunday12:00
Cycle
Zero calibration
300sec
Monday0:00
Monday12:00
Example
In case where auto zero calibration is carried out at the above setting.
Gas replacement time after calibration is the same as the flow time.
Replace-menttime
300sec
Cycle : 1 to 99 hours or 1 to 40 days (initial value 7days)Flow time : 60 to 900 sec (initial value 300sec)
Setting range
When an auto zero calibration starts, the measurement screen automatically appears.Any key operation other than key lock ON/OFF and "Stop Auto Zero Calibration" (see Item 6.5.2) is notpermitted during auto zero calibration. “Stop Auto Zero Calib-ration” cannot be performed with the key lock to ON. To cancel auto zero calibration forcedly, set the key lock to OFF and then execute “Stop Auto Zero Calibration.”If the auto calibration period and auto zero calibration period have overlapped, the auto calibration is retained, ignoring the auto zero calibrationof that period.When the hold setting is set to ON, the hold time of auto calibration contact and measurement value output signal is extended after calibration for gasreplacement time.Turn on the power again after it is turned off (including the case of power failure) at the time set as the next start time in auto zero calibration, and then repeat it in the set cycle.
Caution••
•
•
•
Auto calibration status contact output is closed during auto zero calibration, and is open in other cases.
6 - 22 IM 11G02N01-01E
( ) ENT
( ) ENT
6.5.2 Forced run/stop of auto zero calibration
Auto zero calibration can be performed just once, or auto zero calibration can be forcibly stoppedduring calibration.
6.5.2.1 Execution of auto zero calibration (just once)
(1) Move the cursor to “Setting of Auto ZeroCalibration” by pressing the or the
key on the user mode screen, andthen press the ENT key.
(2) In the “Setting of Auto Zero Calibration”item selection screen that appears, pointthe cursor to “Auto Zero CalibrationRun” by pressing the or key.Press the ENT key.
(3) “Run” is highlighted, displaying a mes-sage to confirm execution of auto zerocalibration. Press the ENT key to executethe calibration forcibly, and press the
ESC key to cancel.
6 - 23IM 11G02N01-01E
6.5.2.2 Forced stop of auto zero calibration
This mode is used to cancel the auto zero calibration forcedly.
(1) In the User mode that is displayed, pointthe cursor to “Setting of Auto ZeroCalibration” by pressing the or key. Press the ENT key.
(2) In the “Setting of Auto Zero Calibration”item selection screen that appears, pointthe cursor to “Auto Zero CalibrationStop” by pressing the or key.Press the ENT key.(“Auto Zero Calibration Stop” appearswhen the screen is selected while autozero calibration is performed.)
(3) “Stop” is inverted. A message appears,prompting you to verify that you want tostop auto zero calibration. Press the ENT
key to stop the auto zoro calibrationforcibly, and the ESC key to cancel (notstopped).
( ) ENT
( ) ENT
6 - 24 IM 11G02N01-01E
• Zero calibration
ExampleIn case where setting the auto calibration components (see Item 6.2.4) to “Ch1: enable” and “Ch2: enable”
A message, “Zero cal.” blinks at Ch1 and Ch2.
“Auto Zero Calibration” screen
ZERO cal.
ZERO cal.
. C h
0-25 vol%
C h
0-100 ppm
C h
0-10 vol%
C h
0-100 ppm
C h
0-100 ppm0 50 3.
.
0 0 0.0 0.
2 1 0 2.
ZERO cal.
ZERO cal.
During auto zero calibration, any key operation is not permitted other than operations such as key lock ON/OFF and “Stop Auto Zero Calibration.”When the key lock is set at ON, even the “Stop Auto Zero Calibration” cannot be used. To stop “auto zero calibration” forcedly, set the key lock to OFF and then execute “Auto Zero Calibration Stop.”
Caution
6 - 25IM 11G02N01-01E
6.6 Peak alarm settingWhen the peak number of times CO concentration exceeds the upper limit value during measure-ment exceeds the set number, an alarm is provided.
The peak alarm and this setting screen appear only when an option is added.
1. Press the MODE key in the Measurementmode, and the User mode appears.
2. Point the cursor to “Setting of PeakAlarm” by pressing the or key.Press the ENT key.
3. In the “Peak Alarm Setting” item selec-tion screen that appears, point the cursorto any item you want to set by pressingthe or key. Press the ENT key.
4. Then, enter numeric values and performthe setting.
Entering the numeric values or setting theitems should be carried out by using the or key.
After setting, press the ENT key, and the setvalues are saved.
( ) ENT
End of Peak Alarm Setting
MODE
( ) ENT
• Peak Alarm : ON/OFF of peak alarm • Alarm Value : If measuring value exceeds the
set alarm value, a peak counter counts1 time.
• Alarm Count : When a peak in excess of the setting time occurs, a peak count alarm output
is provided.• Hysteresis : To prevent possible chattering
when the measuring value may exceed the set peak concentration by only 1 time, the peak count has an allowance in the hysteresis width.
Description of setting items
( ) ENT
6 - 26 IM 11G02N01-01E
If CO concentration exceeds the alarm value, counting will begin. If the number of peaks is overthe set times per hour, a peak alarm contact output becomes closed (ON). If it is less than the settimes per hour, it is open (OFF). Since 5 times of peaks /hour is marked at (1) section fromthe above graph, the peak count alarm is turned ON. Since peaks of more than 5 times per 1 houroccur at the interval between (1) and (2) , the peak count alarm remains ON. Since at (2),peaks are reduced to 4 times per hour, it is turned OFF.
Like the hysteresis of the alarm setting , the hysteresis prevents possible chattering whenmeasured gas is fluctuated near the alarm value.
* For 10 minutes after the power is turned ON, a peak alarm counting is not carried out.
Releasing peak count alarm
To release the peak count alarm, set the peak alarm to OFF.Turning on the peak alarm initiates counting from 0.
Action of peak alarm
Example
[% full scale] represents the percentage with the CO range regarded as 100%.
Setting range
Alarm value : 10 to 1000 ppm 5 ppm step (initial value: 500 ppm)Alarm count : 1 to 99 times (initial value: 5 times)Hysteresis : 0 to 20 % of full scale (initial value: 0% of full scale)
6 - 27IM 11G02N01-01E
6.7 Parameter settingIt allows you to carry out the parameter setting such as time, key lock, etc., as required. Items to beset are as follows:
MODE
( ) ENT
• Current Time : Current year, month, date, day of the week, hour, and minute setting (The display appears in this order.)Note: The clock backup time is 2 days. If power is turned on after it is kept off
for 2 days or longer, make the time setting again.• Key Lock : Sets with ON/OFF so that any key operation except the key lock OFF cannot be
performed.• Output Hold : Sets whether Calibration Output is held or not, and the holding value setting.• Reset Av. Output : Resets the average value.• Response time : Sets the response time of electrical system.• Average Period : Sets the moving average time.• Backlight Timer : Sets automatic OFF of the backlight of display unit and the time until backlight out.• Maintenance mode : Enters passwords to switch to the Maintenance mode.
Description of setting items
* For the maintenace mode, see Item 6.8.
(1) To display the User mode, press the MODE
key in the measurement mode.
(2) Point the cursor to “Parameter Setting”by pressing the or key. Press the
ENT key.
(3) In the “Parameter Setting” screen thatappears, point the cursor to any item youwant to set by pressing the or key. Press the ENT key.
( ) ENT
6 - 28 IM 11G02N01-01E
4. In the Parameter Setting screen thatappears, enter the numeric values and setthe items. Entering the numeric values orsetting the items should be carried out byusing the or key. To move thecursor to the right, press the key.After setting, press the ENT key, that theparameter setting is carried out with thevalue you set.
To close the “Parameter Setting” screen or cancel this mode midway, press the key. A previous screen will return.
To close Parameter Setting screen
ESC
ENT
End of Parameter Setting
Calibration
Output hold
Press the key to flow gas. To execute calibration, press the ENT key.
Hold extending time.
Time set to gas flow time (See Item 6.4 Auto Calibration.)
or
ENT
ENT
ZERO
SPAN
• Hold setting : 0 to 100% FS• Response time : 1 to 60sec. (initial value: 15 sec)• Average period : 1 to 59 min or 1 to 4 hours (initial value: 1 hour)
When setting the unit of 1 to 59 minutes is terms of minute or 1 to 4 hours with hour
• Backlight Timer : 1 to 60 min (initial value: OFF)• Maintenance mode : 0000 to 9999 (initial value: 0000)
Setting Range
Output Hold
By setting an output hold to ON, an output signal of each channel are held during the calibration(manual calibration and auto calibration) and for the gas flow time (refer to Item 6.4, Setting ofAuto Calibration). Regardless of Hold ON/OFF setting, an output signal can be held via anexternal input.
(1) Manual calibration
6 - 29IM 11G02N01-01E
(2) Auto calibration
(3) External hold
Calibration
Output holdHold extending time.
Auto calibration start Auto calibration end
Time hold set to the gas flow time
Remote hold Input
Output hold
Close
Open
(4) Screen display during Holding
The “on Hold” message blinks on the measuring screen.
Since the screen displays the process of calibration is displayed during the manual calibration,“on Hold” is not displayed even if the output signal is held, but the screen is displayed with thehold extending time.
(5) If calibration is cancelled after the calibration gas is supplied regardless of during manualcalibration or auto calibration, the holding extending time will be performed.
6 - 30 IM 11G02N01-01E
(6) You can select the value for hold from the value immediately before entering output hold,“current,” and arbitrary value, “setting.”
Follow the procedures shown below to make the setting.
1) Press the ENT key in a state where thecursor is placed next to Hold.
2) “ON” or “OFF” is highlighted. Press the or the key to select ON or OFF.
Press the ENT key to return to (1).
3) Press the key in a state ON/OFF ishighlighted, and “Current” or “Setting” ishighlighted. Select “Current” or“Setting” by pressing the or the key.
4) Press the ENT key while “Current” isselected to return to (1). Press the ENT
key while “Setting” is selected to go tothe setting entering screen.“Current”: Holds the value immediatelybefore the hold.“Setting”: Holds the value arbitrarily set.
ENT
( ) ENT
6 - 31IM 11G02N01-01E
5) On the parameter hold screen thatappears, move the cursor next to the Ch(component) you want to make thesetting by pressing the or the key,and then press the ENT key.
6) The value is highlighted, indicating thatthe value can be changed. Change thevalue by pressing the or the key,and then move the cursor to the right bypressing the key.
7) After the value is changed, press the ENT
key.
8) Press the ESC key to return to theparameter setting screen.
The setting is expressed in % against the range for both ranges.When 0 to 1000 ppm is selected as the range, for example, if 10% FS is selected as hold setting, the output equivalent to 100 ppm is output and held irrespective of the measure-ment value at that time.
Meaning of setting
• Instantaneous value display of the measurement cannot be held. (Output only can be held.)• If set value is selected for hold, instantaneous O2 correction value is calculated and held
based on the set value.• Range identification contact output cannot be switched even if the range is switched during
the hold.
Description of setting
ESC
End of Hold Setting
Parameter Setting screen
ENT
( ) ENT
( ) ENT
6 - 32 IM 11G02N01-01E
Reset input
So long as close, resetting lasts.At the edge of changing from closing to opening, the average action restarts.
Close (hold at least 1.5 sec.)
Open
Average value reset
This mode is used to clear all average values O2 correction average and O2 average, and restartsaveraging. All average values are reset at a time. The indication value and output value is 0 ppm,vol% or so at the time of the reset input (Refer to the average peripd).
Response time
The response time of the electrical system canbe changed.
Setting is available by components.
Note) It does not provide exact seconds forthe setting time, but it gives a guideof the setting time.
The setting value can be modified as requested by the customer.
Average period
It allows you to set an average period of theaverage value of O2 correction and O2 average.
It enables you to set an average time of 1 to 59minutes (1-minute step) or 1 to 4 hours (1-hourstep).Changing the setting resets the average valueof O2 correction and O2 average value.(Pressing the ENT validates the resetting onlyfor components whose setting was changed.)
6 - 33IM 11G02N01-01E
In case the average period was set to 1 hour.
• Sampling occurs every 30 seconds.
• Every 30 seconds, the average for last 1 hour (time setting) is output.
• At the instant of resetting, zero is assumed for all past values. It means that the average value
will not be correct for 1 hour after resetting.
Example of average action
Average value
Time
Reset
1 Time
Backlight Timer
Automatic OFF setting of the backlight of the LCD unit can be made.
When the specified time elapses from when the measurement screen is resumed, the backlight isautomatically turned off. Press any key to reset backlight OFF.
Only when ON is selected, the time untilauto OFF is displayed. Press the key inthis state, and the time setting can bechanged by pressing the or the key.Press the ENT key to confirm the selection.
If OFF is selected, the backlight is not turnedoff.
Maintenance mode
Enter the password and then press the ENT key to enter the maintenance mode. The password canbe set by the password setting in maintenance mode. Default password setting at the time ofdelivery from the factory is “0000.” You can enter the maintenance mode with the value before itis changed.
6 - 34 IM 11G02N01-01E
• Error Log screen
6.8 Maintenance modeThis mode is used for check of sensor input values, display of error log files or setting of pass-words, etc. First, enter a password and then use it from the next operation. This mode is displayedby selecting the Maintenance Mode from “Item 6.7 Parameter Setting”.
Each “Maintenance” screen
( ) ENT
• NOx M : NOx sensor input value• NOx C : NOx interference compensation
sensor input value• SO2 M : SO2 sensor input value• SO2 C : SO2 interference compensation
sensor input value• CO2 M : CO2 sensor input value• CO2 C : CO2 interference compensation
sensor input value• CO M : CO sensor input value• CO C : CO interference compensation
sensor input value• Temperature: temperature sensor input value• O2 : O2 sensor input value
Description of Sensor Input Value screen
(1) Select the Maintenance mode from theParameter Setting screen to display thePassword Setting screen.
(2) Enter the password, and the MaintenanceMode item selection screen will bedisplayed. Point the cursor to the itemyou want to set by pressing the or key and press the ENT key.
(3) Next, each Maintenance screen isdisplayed.
Note) “To Factory Mode” is used for ourservice engineers only. Refrain fromusing this mode.
(4) Press the ESC key to return to theMaintenance Mode item selection screenfrom each screen.
Error history. Fourteen newest errors are logged.For error number, date and time (year, month, day, period) of occurrence, channel and other details of error, refer to Item 8.1 Error message.Select Clear Error Log and press the key, and the error log is cleared completely.
Description of Error Log screen
ENT
6 - 35IM 11G02N01-01E
• Calibration Log screen
Past calibration history.Sensor input value, concentration value, and the date when zero/span calibration is performed are logged. The 10 newest calibration data is logged by each component.Move the cursor to Clear Calibration Log and press the key, and the calibration log is cleared completely.
Z1 : Zero calibration (Z) of Range 1S1 : Span calibration (S) of Range 1M : Value of measuring detector at the time
of calibrationC : Value of the interference compensation
detector at the time of calibrationCon : Concentration value displayed before
calibration
Description of Calibration Log screen
ENT
If the following operation is maladjusted, the measurement may be adversely and excessively affected. Carry out the operation with utmost attention.
Caution
. Optical adjustment screen
For details of this item, refer to “Item 7.3.3Optical zero adjustment method”.
Press ENT key and turn ON the solenoidvalve signal for each calibration gasby using the or key.
ENT
6 - 36 IM 11G02N01-01E
• Moisture interference adjustment screen
For details of this item, refer to “Item 7.3.4Moisture interference adjustment method.”
Move the cursor to a desired Ch (compo-nent) by pressing the or the key, andthen press the ENT key, and the selected valueat right is highlighted.
Check that the gas for moisture interferencecompensation is flowing, change the mois-ture interference compensation coefficient
using the or the key, adjust the value
at left so that it becomes near zero, and then
press the ENT key to log moisture interfer-
ence compensation value.
In values on the left side of screen, the moisture interference for each component is already offset.The figures at right are interference compensation coefficients.
Description of moisture interference adjustment screen
CautionSince an interference compensation detector is not provided if the 1st range is beyond 0 to 10 vol%, no interference adjustment can be performed (no need).
ENT
6 - 37IM 11G02N01-01E
• Output adjustment screen
Analog output adjustment screen.Connect the digital multi meter to the output terminal corresponding to the number of OUT to be adjusted, and adjust the value so that 4mA or 0V is output at zero and 20mA or 1V is output at span.
Description of output adjustment screen
Move the cursor using the , , or the key to the output (OUT No. and zero/
span) to be adjusted, and then press theENT key.
The selected value is highlighted. Adjustthe value, while watching the output, bypressing the or the key. Press the
key to select the next digit.
On completion of the adjustment, pressthe ENT key.
ENT
6 - 38 IM 11G02N01-01E
• Other parameter
Password Set : Set the password used to move from the parameter setting screen to the maintenance mode. Arbitrary 4-digit number can be selected.
O2 ref. Value: Set the oxygen concentration reference value at the time of oxygen correction calculation. Settable in the range from 00 to 19%.
Limit : Set the oxygen concentration limit at the time of oxygen correction calculation. Settable in the range from 01 to 20%.
* Refer to the O2 correction concentration value in “5.3 Outline of display screen” for oxygen correction calculation procedure.
Station No. : Set the station No. for MODBUS communication. Settable in the range from 00 to 32.
Range setting : Moves to the screen on which measuring range is changed.
Description of each setting screen
Press the or the key to move thecursor to the item whose setting is to bechanged.
The values for password, oxygencorrection, limit, and station No. arehighlighted.
Press the or the key to change thevalue to desired one, and then press the
ENT key.
Note: Pay attention not to forget thepassword. Otherwise you cannotenter the maintenance mode.
6 - 39IM 11G02N01-01E
<How to set/change the range>
The measuring range can be arbitrarily selected inthe minimum and the maximum range specified atthe time of purchase. The range to be used can beselected 1 or 2.
(1) Move the cursor to the item to be set by press-ing the or the key, and then press the
ENT key.
(2) Move the cursor to the Ch (component) whosesetting is to be changed by pressing the orthe key, and then press the ENT key.
(3) Move the cursor to the item whose setting is tobe changed by pressing the or the key,and then press the ENT key.
(4) Press the or the key to change thevalue. Press the key to select the nextdigit. In a state where the decimal point ishighlighted, press the or the key, andthe decimal point position can be changed.
(5) When necessary change is made, press the ENT
key.
The value for range 1 and range 2 must fall within the range from the MIN and the MAX range (including the MIN and the MAX range), and at the same time range 1 must be smaller than range 2.The number of ranges is 1 or 2.
Settable range
ENT
( ) ENT
( ) ENT
ENT
End of Set/Change the Range
6 - 40 IM 11G02N01-01E
ENT
To Measurement screen after
executing Manual Zero Calibration
ZERO
6.9.1 Zero calibration
It is used for zero point adjustment. For zerocalibration, gas suited for an applicationshould be used according to “(3) Standard gasin Item 3.3 Sampling.”
(1) Press the ZERO key on the Measurementscreen to display the Manual Zero Cali-bration screen.
(2) Select the Ch (component) to becalibrated by pressing the or key.After selection, press the ENT key, andzero gas will be supplied.
(3) Wait until the indication is stabilizedwith the zero gas supplied. After theindication has been stabilized, press the
ENT key. Zero calibration in rangeselected by the cursor is carried out.
Note: For the Ch (component) forwhich “AR” is selected in “6.1.1Setting range switch mode,”the cursor automatically movesto the range selected in “Settingof auto calibration component/range” (6.2.4), and calibration iscarried out within that range.
For the Ch (components) that is set to “both” in the “Zero Calibration” of the Calibration Setting mode, zero calibration is also carried out at the same time.
Caution
To close the “Zero Calibration” or
cancel this mode midway, press the ESC
key. A previous screen will return.
To close "Zero Calibration"
ESC
( ) ENT
6.9 Calibration
6 - 41IM 11G02N01-01E
When “both” from “Calibration Range” of the Calibration Setting mode is set, span calibra-tion is performed together with 2 Ranges.
Caution
To close the “Span Calibration” or cancel this mode midway, press the key. A previous screen will return.
To close "Span Calibration"
ESC
(1) Press the SPAN key on the Measurementscreen to display the Manual Span Calibra-tion screen.
(2) Select CH (component) to be calibrated bypressing the or key and press theENT
key. The calibration gas is supplied.
To Measurement screen after
executing Manual Span Calibration
ENT
SPAN
(3) Wait until the indication is stabilized in thestate where the calibration gas is supplied.After the indication has been stabilized,press the ENT key. Span calibration ofRange selected by the cursor is performed.
Note: For the Ch (component) forwhich“AR” is selected in “6.1.1Setting range switch mode,” thecursor automatically moves to therangeselected in “Setting of autocalibration component/range”(6.2.4), and calibration is carriedout within that range.
( ) ENT
6.9.2 Span calibration
It is used to perform a span point adjustment. Supply calibration gas with concentration set to thespan value to perform the span calibration. For the span calibration gas for the NOX, SO2, CO2,CO measurement, use the standard gas with a concentration of 90% or more of the range value.For the span calibration gas for the O2 measurement, use the standard gas with a concentration of90% or more of the range value when measuring with the built-in O2 sensor, and use the standardgas of 1to 2 vol% when measuring with an external zirconia O2 sensor.
≈
Blank Page
7 - 1IM 11G02N01-01E
7.1 Daily check
(1) Zero calibration and span calibration
1. Perform zero calibration. For the calibration procedures, refer to “Item 6.9.1 Zero calibra-tion”.
2. Then, perform span calibration. For the calibration procedures, refer to “Item 6.9.2 Spancalibration”.
3. Zero calibration and span calibration should be carried out once a week, as required.
(2) Flow rate check
1. Sampling gas flow and purge gas flow are as follows:
• Sampling gas flow: 0.5 ± 0.2L/min
• Purge gas flow: About 1L/min
2. Check and maintenance should be carried out every day, as required.
7.2 Daily check and maintenance proceduresTable 7.1 Maintenance and check table
7. MAINTENANCE
7- 2 IM 11G02N01-01E
(1) Turn off the power switch, stop thesample gas, and allow the zero gas toflow for several minutes to purge thecell interior.
Loosen the setscrew (2 pieces) fromthe top cover and remove it.
(2) Remove the internal gas inlet tube.
(3) Loosen both right and left screws forcell holding plate.
• Remove the sample cell only.
(4) Turn to the left the sample cell windowand remove it from the sample cell (seeFig. 7-1).
(5) For cleaning the window and cellinside surface, first eliminate coarsedust by soft brush or the like and thenwipe them by soft rag.
The window is easy to get scratched.Pay utmost attention so as not todamage it.
(6) After the end of sample cell cleaning,mount the cell in place and proceed torunning.After cleaning sample cell, be sure toperform optical zero adjustment (seeItem 7.3.3) and moisture interferencecompensation adjustment (see Item7.3.4).
Gas outletGas inlet
Reference cell
Cell holdingplate
Cell holdingplate
Sample cell Screw Screw
Caution
Top cover
F10.12E.eps
If the window or the cell interior is very dirty, use a soft cloth moistened with absolute alcohol.A slightly corroded infrared transmission window or sample cell can be remedied by gently rubbingwith chromium oxide powder on cleaning cloth but an excessively corroded one must be replaced.When cleaning, do not exert an excessive stress.
7.3 Maintenance of analyzer unit7.3.1 Cleaning method for sample cell (pipe cell)
This section is strictly factory adjusted. Handle it with utmost attention.
If it is absolutely required, contact us.
7 - 3IM 11G02N01-01E
Infraredtransmission window
Window holder
O ring
Sample gas inlet
Cell
Bonded together
Sample gas outlet
Fig. 7-1 Structure of sample cell (pipe cell)
7- 4 IM 11G02N01-01E
(1) Turn off the power switch, stop thesample gas, and allow the zero gas toflow for several minutes to purge thecell interior.
Loosen the setscrew (2 pieces) fromthe top cover and remove it.
(2) Remove the internal gas inlet tube.
(3) Loosen the 2 detector set bolts.
Note) The distribution cell, block celland detector are fastened by thesame bolts.
(4) Using the furnished cell mounting tool,turn the window fixture to the left andremove it from the cell. (See thestructure of sample cell (block cell) inFig. 7-2.)
(5) For cleaning the infrared transmissionwindow and cell inside surface, firsteliminate coarse dust by soft brush orthe like and then wipe them by soft rag.
The window is easy to get scratched.Pay utmost attention so as not todamage it.
(6) After the end of sample cell cleaning,mount the cell in place and proceed torunning.
After cleaning sample cell, be sure toperform optical zero adjustment (seeItem 7.3.3) and moisture interferencecompensation adjustment (see Item7.3.4).
Caution
Infrared transmission window
Inner O ringOuter O ring
O ring holder
Window fixture
Top cover
F10.12E.eps
Gas inlet tube set cap nutDetector set bolt
If the window or the cell interior is very dirty, use a soft cloth moistened with absolute alcohol.A slightly corroded infrared transmission window or sample cell can be remedied by gently rubbingwith chromium oxide powder on cleaning cloth but an excessively corroded one must be replaced.When cleaning, do not exert an excessive stress.
7.3.2 Cleaning method for sample cell (block cell)
7 - 5IM 11G02N01-01E
Cell mounting tool
Window fixture
O ring holder
Outer O ring
Reference cell unit
Hole for bolt for fixing together withdistribution cell and detector
Inner O ring
Infrared transmission window
Sample cell unit
Sample gas outlet Sample gas inlet
Structure of sample cell (of 32, 16, 8, 4, 2 mm long) (sample cell and reference cell are integrated)
Note) Use the dedicated cell mounting tool (furnished).
Fig. 7-2 Structure of sample cell (block cell)
7- 6 IM 11G02N01-01E
7.3.3 Optical zero adjustment method (optical balance adjustment)
The adjustment is performed at reassembly after removing the sample cell, etc. for cleaning, etc.
If the following operation is maladjusted, the measurement may adversely be affected. If you arenot trained for adjustment, do not carry out this operation but contact the distributor or our service-man.
Caution
No. of componentsto be measured
1-component meter
3-component meter
4-component meter
2-co
mpo
nent
met
er
a
Main
NO-SO2
CO2-CO
NO-CO
CO2Main
NOMain
NO-SO2-CO
* O2 is excluded from the number of components.* Main is signal input value from the main detector of each component. Comp is signal input value from interference compensation detector of each component. If low range exceeds the range of 0 to 10vol%, detector signal of comp is not usable.
NO-SO2-CO2-CO
b
Comp
CO2Comp
NOComp
c
—
COMain
SO2Main
d
—
COComp
SO2Comp
e
—
—
—
f
—
—
—
g
—
—
—
h
—
—
—
NOMain
NOComp
— —CO
MainCO
Comp— —
NOMain
NOComp
— —CO
MainCO
CompSO2Main
SO2Comp
NOMain
NOComp
CO2Main
COMain
COComp
SO2Main
SO2Comp
—
Sensor values of which are not included in measuring components should be ignored.
<Correspondence between measurement detector and indicated position>
(1) Remove the top cover. Allow dry N2
or air to flow through the analyzer unitsample gas inlet until the readingstabilizes. The sample gas is introduceddirectly to the INLET of analyzer unitthrough the gas cylinder.
(2) Proceed to an optical adjustment in themaintenance mode. The display on theoperation panel of the main unit is asillustrated on the right. Balance adjust-ment is not required if the display fallswithin ±100.
a
b
c
d
e
f
g
h
Piping at optical balance adjustment
Short
Open
Introduce directly from cylinder
(not required for measuring unit 1 only)
Optical system 2 (measuring unit 2)
Optical system 1(measuring unit 1)
7 - 7IM 11G02N01-01E
7.3.4 Moisture interference compensation adjustment method
Proceed to an adjustment if excessively (beyond ± 2%FS) affected by moisture inteference.
After the end of optical balance adjustment, be sure tocarry out moisture inteference compensation adjust-ment.
(1) After warm-up, select the low range, allow dry gas(N
2, air) to flow at 0.5 L/min and carry out zero
calibration.
If the following operation is maladjusted, the measurement may adversely be affected. If you arenot trained for adjustment, do not carry out this operation but contact the distributor or our service-man.
Caution
N2 or AirFlow meter
ElectricalCooling Unit
Water Bubbler
Drain
Optical zero adjustment knob
Optical system 1
Optical system 2
(3) Carry out the adjustment in the procedure in (4) and subsequent.
• Adjust on the primary side of the optical system so that the values for a to d in 1-1 and1-2 become as close to 0 as possible within ±100 range.
• Adjust on the secondary side of the optical system so that the values for e to h in 2-1 and2-2 become as close to 0 as possiblewithin ±100 range.
(9) After the optical balance adjustment, mount the top cover of the analyzer unit, then carry outa moisture interference compensation adjustment, and perform zero and span calibrations.
* Before moving the dimmer plate, loosen the detector set bolts (just enough to makethe plate movable for snug adjustment).
(4) Operate the optical zero adjustmentknob to change the value displayed ata (or e ).
(5) Move the dimmer plate sideview tochange the value displayed at b (or
f ).
(6) Move the dimmer plate sidewise tochange the value displayed at c (or
g ).
(7) Move the dimmer plate sidewise tochange the value displayed at d (or
h ).
(8) Repeat the procedures in (4) to (7) tomake all the displayed values comeclose to 0 as possible within ±100range.
* Adjust the dimmer plate which isthe nearest to the zero adjustmentknob first, and sequentially.
Dimmer plate
Detector set bolt
7- 8 IM 11G02N01-01E
7.3.5 Replacement of fuse on analyzer unit
Note) Before the following work, be sure to remove the cause of blown fuse (short, etc) andrepair.
(1) Turn OFF the main power supply SW of
the analyzer.
(2) Loosen setscrews (4 pcs.) from the top of
the analyzer and remove the top cover.
(3) Turn the fuse holder (shown in Fig. 7-3)
counterclockwise and put it out, and the cap
will be removed. Remove the fuse
(250 V AC/3 A) from the holder. Replace it
with a new one .
(4) Mount the fuse holder cap and cover for
analyzer indication unit in the reverse
procedure. Turn ON the power supply SW.
The work will be completed if the analyzer
is normally worked.
(2) Display the moisture interferencecompensation screen of the analyzer unit(see “6.8 Maintenance mode”). Set thedew point to 28C by using an electroniccooler, and introduce bubbled N2 or airgas to the analyzer (shown on the figure).
(3) On the screen, select a desired Ch(component) by pressing the ENT key,adjust the value at right by pressing the
or the key so that the value at leftfalls within ±10 (make it as close to 0 aspossible), and then press the ENT key tomemorize the value. (Exiting by “ESC ”cancels the adjustment.)
Or, selecting the “ALL” and pressing the“ ENT ” key, zeroes all componentsintegrally.
(First, adjust all components by selectingALL and then perform fine adjustmentfor components one by one using and
keys.)
* If any components exceed the range of0 to 10vol%, no adjustment can beperformed (No interferencecompensation is required).
Front
Fuse holder
Fig. 7-3
( ) ENT
7 - 9IM 11G02N01-01E
7.4 Inspection and maintenance of limited service-life componentsThe analyzer uses limited-life components. The recommended replacement periods are listed in
the below table.
(1) Limited service-life components are those which wear out or for which failure is presumedwithin five years under normal operating or storage conditions. Components with more thanfive years of service life are the exception.
(2) The following table only involves the recommended periods for conducting preventive main- tenance for limited service-life components; these periods do not guarantee that accidental failures will not occur.
(3) The recommended replacement periods are tentative and depend on operating conditions.
(4) The recommended replacement periods may vary depending on the field date.
1. Light source Recommended replacement period: Every five years (*1)
2. Sector motor Recommended replacement period: Every two years (*1)
3. Detector without Recommended replacement period: Every five years (*1)
O2 sensor
4. O-ring Recommended replacement period: Every two years,
inspect it when cleaning of measuring cell. See Sec. 7.3
5. LCD panel Recommended replacement period: Every five years (*1)
6. Power supply Recommended replacement period: Every five years (*1)
7. Measuring cell Set up an appropriate maintenance period
( by refering to the check results)
8. Reference cell Recommended replacement period: Every five years (*1)
Checkpoint
Infrared gas analyzer
Check and maintenance items Procedure and criteria
Recommended check and maintenance periods
Every five years
Every tw
o years
Once a year
Every six m
onths
Every three m
onths
Once a m
onth
Once a w
eek
Routine
(*) In the check and maintenance columns, place a check mark ( s ) for check and confirmation work, a dark star ( w ) for replacement, and a white star ( q ) for parts preparation for preventive maintenance.(*1) These are seviceperson’s work, contact our sevicepersons.
ww
w
w
s
s
T7.5E.eps
q
d Precautions to be taken while checking
(1) When handling reference gas (during calibration), carefully read the reference-gas instruc-tion manual to use the gas correcrtly. In particular, special attention must be taken in han-dling carbon monoxide gases; otherwise, you may suffer from gas poisoning.
(2) During maintenance checks, be sure to keep the near fan on. If any gas leaks, you may suffer from gas poisoning.
(3) When replacing the analyzer gas filter or conducting maintenance service of the washer, completely shut the calibration-gas valve. Otherwise, you may suffer from gas poisoning.
≈
Blank Page
8 - 1IM 11G02N01-01E
8.1 Error messageIf errors occur, the following contents are displayed.
When errors No. 1 and No. 10 occur, analyzing block error contact output is closed.
When errors No. 4 to No. 9 occurs, calibration error contact output is closed.
Screen display and operation at the occurrence of error
In case of Error No. 1 to No. 4, No. 6, No. 8 to No. 10
Motor rotation detection signal faulty
Error No.1 • Motor rotation is faulty or stopped.• Motor rotation detector circuit is faulty.
Note) Sector motor is a consumption part.It is recommendable to exchange the motor once two years.
Measured values fluctuate too much during zero and span calibration.
Error No.8 • Calibration gas is not supplied.• Time for flowing calibration gas is short.
Calibration is abnormal during auto calibration.
Error No.9 • Error corresponding to No. 4 to No. 8 occurred during auto calibration.
Output cable connection is improper.
Error No.10 • Wiring is detached between analyzer and interface module.• Wiring is disconnected between analyzer and interface module.
Span calibration is not within the allowable range.
Error No.6
Amount of span calibration (difference between indication value and calibrated concentration) is over50% of full scale.
Error No.7
• Span gas is not supplied.• Calibrated concentration setting does not match cylinder concentration.• Zero calibration is not performed normally.• Span is deflected much due to dirty cell.• Detector sensitivity has deteriorated.
Zero calibration is not within. Error No.4
Amount of zero calibration (indication value) is over 50% of full scale.
Error No.5
• Zero gas is not supplied.• Zero is deflected much due to dirty cell.• Detector is faulty.• Optical balance is maladjusted.
Error display Error contents Probable causes
C H O20-25 vol%21 0 0.
C H CO0-100 ppm0 0
C H CO20-10 vol%0 0 0 0
C H SO20-100 ppm
C H NO20-25 ppm0 0 8
..
.
.Error No.9
1 3 6ENT
ESC
ESC
ESC
• Press the key to delete the error display. • If the key is pressed without removing the cause of an error, the error will be displayed again.
Measurement screen
• When more than one error occurs, pressing the key moves to another error display.
Display of error contents
8. TROUBLE SHOOTING FOR ANALYZER
8 - 2 IM 11G02N01-01E
In case of Error No. 5 and No. 7
Error log file
If error occurs, the history is saved in an error log file. The error log file exists in the mainte-nance mode.
Error log screen
* Up to 14 errors can be saved in the error history; the oldest error will be deleted one by one everytime a new occurs.
* If the power supply is turned OFF, the contents in the error log file will not be lost or damaged.
Deletion of error history
Press the ENT key on the above screen, and the “Error Log Clear” will be inverted. Furtherpressing the ENT key will clear the error history.
• Pressing delete the error display.
Calibration is continued. Unless another calibration error occurs, calibration is carried out to the end, the Measurement screen returns.
ENT
ENT ESC
ESC
Calibratedforcedly
C h
0 0 9.
C h
0 0
C h
0 0 0 0
C h
C h
9 0 8
..
.1 3 6
.
Errors that occurred
New
Date and time when an error occurred.
Component with which the error occurred.
Old
8 - 3IM 11G02N01-01E
8.2 TroubleshootingThe following table shows how to remedy other troubles, such as faulty readings.
• Check if the zero gas is supplied to the analyzer at the specified flow rate.• Check the optical balance. (See Subsection 7.3.3, "Optical zero adjustment method.")• Check the count value on the display panel. (Refer to the sensor input value by maintenance mode.)
• Check if the span gas is supplied to the analyzer at the specified flow rate.• Check if zero calibration can be performed properly.• Check the count value on the display panel. (Refer to the sensor input value by maintenance mode.)
• Take either of the above remedies depending of the source of the error: zero calibration or span calibration trouble.
• Check if the sample gas is supplied to the analyzer at the specified flow rate.• Check the optical system, e.g., the sample cell window, O-ring, detector window and inside of the cell for heavy dirt.
• Check for the influence of interference from moisture.• Check if the sample gas contains interfering components in large quantities or not.
• Check if the sample gas is supplied to the analyzer at the specified flow rate.• Check if the zero and span calibration is enabled.
Zero calibration fails.Error No. 4, 5, or 8 occurs.
Span calibration fails.Error No. 6, 7, or 8 occurs.
Error occurs during auto calibration.Error No. 9 occurs.
Drift.
Readings are abnormally high or low.
Readings do not increase.
• Locate and check gas leaking points, and take the proper remedy.• Adjust the optical balance.
• Locate and check gas leaking points, and take the proper remedy.• If zero calibration fails, first attempt remedies for zero calibration trouble.• Are there any changes from when zero gas is supplied?
• Locate and check gas leaking points, and take the proper remedy. (See Section 3.3.)• Clean the cell and window. Replace the part.(See Subsections 7.3.1 and 7.3.2.)
• Adjust interference from moisture.• Investigate the components of the sample gas and then contact our serviceperson. (See Sec. 9.1 item 4.)
• Locate and check gas leaking points, and take the proper remedy. (See Section 3.3.)• If enabled, sampling (check the points relating only to the sample gas and take the proper remedy.)• If not enabled, check the calibration related error items.
Symptom Checking Item Remedy, etc.
T8.2.eps
≈
Blank Page
9 - 1IM 11G02N01-01E
1. Standard specificationsMeasurable gas components and measuring range:
• Max. 5 components measurementincluding O2.
• Measuring range ratio ≤ 1:5 (O2)≤ 1:25(except for O2)
• For measurable components andpossible combinations of measuringranges, refer to Tables 1-(1) to (7).
Principle of measurement:NO, SO2, CO2, CO, CH4 ;
Non-dispersion infrared-ray absorptionmethodSingle light source and double beams(double-beam system)
O2 ; Paramagnetic O2 sensor (built in) or zir-conia O2 sensor (externally installed)
Measured value indication:Digital indication in 4 digits(LCD with back light)• Instantaneous value of each component• Instantaneous value after O2 correction
(only in NO, SO2, CO measurementwith O2)
• Average value after O2 correction(only in NO, SO2, CO measurementwith O2)
• O2 average valueAnalog output signals:
* Inputs/outputs of analog signals arepossible by combining with the input/output terminal module.
4 to 20mA DC or 0 to 1V DC,non-isolated output ; 12 points max.max.load 550Ω. for 4 to 20 mA DCmin.load 100kΩ. for 0 to 1V DC* Refer to Table 2, for the channel No. of
displayed values and analog outputsignals.
Analog input signal:For signal input from externally installedO2 sensor.Signal requirement;(1) Signal from Zirconia O2 sensor (when
ZX8DpC or ZX8DpD)(2) 0 to 1V DC from an O2 sensor Input
section is not isolated. This feature iseffective when an O2 sensor is notbuilt in.
(Depend on O2 input signal, measured concentration indication and O2 conver-sion.)
9. SPECIFICATIONS
9.1 General specifications
Minimum range Maximam range
CO2
COCH4
O2 (built in)
O2 (External Zirconia)
0 – 50ppm0 – 100vol%0 – 100vol%0 – 100vol%
0 – 25vol%
0 – 200ppm
0 – 5vol%
0 – 20ppm
0 – 25vol%0 – 5vol%
NOSO2 0 – 50ppm
0 – 5000ppm0 – 10vol%
0 – 50ppm
Relay Contact output:1a contact (250V AC/2A, resistive load)
Instrument error, calibration error,range identification, auto calibrationstatus, pump ON/OFF, peak alarm.
1c contact (250V AC/2A, resistive loadselectable 6 outputs)
High/Low limit alarm contact output.Power disconnection alarm.
* All relay contacts are isolated mutuallyand from the internal circuit.
Contact input: No-voltage contact (ON/0V, OFF/5VDC, 5mA flowing at ON)
Remote range switch, auto calibrationremote start, remote holding, averagevalue resetting, pump ON/OFF
Isolated from the internal circuit withphotocoupler. Contact inputs are not iso-lated from one another.
Transmission output:Solenoid valve drive signal for automaticcalibration.Transistor output (100mA or less)
Power supply: Voltage rating ; 100V to 240V ACAllowable range ; 85V to 264V ACFrequency ; 50Hz/60HzPower consumption; 250VA max.Inlet ; Conform to EN60320
Protection Class I typeOperation conditions:
Ambient temperature ; -5˚C to 45˚CAmbient humidity ; 90% RH max.,
non-condensingStorage conditions:
Ambient temperature ; -20˚C to 60˚CAmbient humidity ; 100% RH max.,
non-condensingDimensions (H ××××× W ××××× D):
Analyzer main unit;177 x 483 x 578mm (Excluding handles.)Input/output terminal module;164 x 318 x 55mm
Mass: Approx. 22 kg (only Analyzer)Finish color: Front panel ; Off-white (Munsell 10Y7.5/
0.5 or equivalent)Casing; Steel-blue (gray)
Enclosure: Steel casing, for indoor useMaterial of gas-contacting parts:
Gas inlet/outlet; SUS304Sample cell; SUS304,chloroprene rubberInfrared-ray transmitting window; CaF2
O2 sensor sample cell : SUS316Internal piping; Toaron, Teflon
Gas inlet/outlet: Rc1/4 or 1/4NPT internal threadPurge gas flow rate: 1L/min ( when required)
Installation altitude : 2000 m or less Safety and EMC conforming standards :
Safety: EN61010-1 Pollution degree 2 (Note)
Installation category II (Note) Note: Installation category, called over-voltage
category, specifies impulse withstanding voltage. Category II is for electrical equip-
ment. Pollution degree indicates the degree of
existence of solid, liquid, gas or other inclusions which may reduce dielectric strength. Degree 2 is the normal indoor
environment.EMC: EN61326
EN61000-3-2 EN61000-3-3
9 - 2 IM 11G02N01-01E
2. Standard FunctionsOutput signal holding:
Output signals are held during manual andauto calibrations by activation of holding(turning “ON” its setting).The values to be held are the ones just be-fore start calibration mode or setting value.It is selectable.Indication of instantaneous values will notbe held.
Remote output holding:Output signal is held at the latest value orsetting value by short-circuiting the remoteoutput holding input terminals.Holding is maintained while the terminalsare short-circuited.Indication of instanta-neous values will not be held.
Switch ranges:The switch ranges is available in manual,auto, and remote modes. Only presetswitch method is effective.
Manual: Allows range to switch by key operation.Auto: Allows range to switch from low to high
range when 90%FS or more is available inthe low range.Allows range to switch from high to lowrange when 80%FS or less is available inthe low range.
Remote: No-voltage contact input (for measurablecomponents)Allows range to switch via an external sig-nal when remote range switch input is re-ceived.When the contact input terminals for eachcomponent are short-circuited, the firstrange is selected, and it is switched to thesecond range when the terminals are open.
Range identification signal:The present measuring range is identifiedby a contact signal.The contact output terminals for each com-ponent are short-circuited when the firstrange is selected, and when the secondrange is selected, the terminals are open.
Auto calibration:Auto calibration is carried out periodically atthe preset cycle.When a standard gas cylinder for calibra-tion and a solenoid valve for opening/clos-ing the gas flow line are prepared externallyby the customer, calibration will be carriedout with the solenoid valve drive contactsfor zero calibration and each span calibra-tion turned on/off sequentially at the setauto calibration timing.
Auto calibration cycle setting:Auto calibration cycle is set.Setting is variable within 1 to 99 hours (inincrements of 1 hour) or 1 to 40 days (inincrements of 1 day).
Gas flow time setting:The time for flowing each calibration gas inauto calibration is set.Settable within 60 to 900 seconds (in in-crements of 1 second)
Auto calibration remote start:Auto calibration is carried out only onceaccording to an external input signal.Calibration sequence is settable in thesame way as the general auto calibration.Auto calibration is started by opening theauto calibration remote start input termi-nals after short-circuiting for 1.5 secondsor longer.
Auto zero calibration:Auto zero calibration is carried out periodi-cally at the preset cycle.This cycle is independent on “Autocalibradion” cycle.When zero calibration gas and solenoidvalve for opening/closing the calibrationgas flow line are prepared externally bythe customer, zero calibration will be car-ried out with the solenoid valve drive con-tact for zero calibration turned on/off at theset auto zero calibration timing.
Auto zero calibration cycle setting:Auto zero calibration cycle is set.Setting is variable within 1 to 99 hours (inincrements of 1 hour) or Setting is variablewithin 1 to 40 days (in increments of 1 day)
Gas flow time setting:The timing for flowing zero gas in autozero calibration is set.Settable 60 to 900 seconds (in incrementsof 1 second)
Upper/lower limit alarm:Alarm contact output turns on when mea-surement value reach to the preset high orlow limit alarm value.Contacts close when the channel value ofeach channel becomes larger than thehigh alarm limit value or smaller than thelow alarm limit value.
Instrument error contact output:Contacts close at occurrence of analyzererror No. 1, 3 or 10.
Calibration error contact output:Contacts close at occurrence of manual orauto calibration error (any of errors No. 4to 9).
Auto calibration status contact outputs:Contacts close during auto calibration.
Pump ON/OFF contact output:During measurement, this contact close.While calibration gas is flowing, this con-tact open. This contact is connected inpower supply of pump, and stop thesample gas while calibration gas flowing.
9 - 3IM 11G02N01-01E
3. Optional functionO2 correction: Correction of measured NO, SO2 and CO
gas concentrations into values at referenceO2 concentration
Correction formula:
C = –––––––– x Cs
C : Sample gas concentration after O2
correctionCs : Measured concentration of sample
gasOs : Measured O2 concentration (Limit set-
ting: 1 to 20% O2, default 17%)On : Reference O2 concentration (value changeable by setting.0 to 19%
O2, default 4%)Average value after O 2 correction and O 2 average value
calculation:The result of O2 correction or instanta-neous O2 value can be outputted as an av-erage value in the preset period of time.Used for averaging is the moving averagemethod in which sampling is carried out atintervals of 30 seconds.(Output is updated every 30 seconds. It isthe average value in the determined periodof time just before the latest updating.)Averaging time is settable within 1 to 59minutes (in increments of 1 minute) or 1 to 4hours (in increments of 1 hour).
Average value resetting:The above-mentioned output of averagevalue is started from the initial state byopening the average value resetting inputterminals after short-circuiting for 1.5 sec-onds or longer.Output is reset by short-circuiting and re-started by opening
CO concentration peak count alarm:
(added only for CO/O2 measurement)Alarm output turns on according to the pre-set concentration and count.Whenever the instantaneous value of COexceeds the preset concentration value,count increments. If the count exceeds thepreset value in one hour, the alarm contactsclose.
Communication function:RS-232C (9pins D-sub)Half-duplex bit serialStart-stop synchronizationModbusTM protcolContents : Read/Wright parameters
Read measurement concentra-tion and instrument status
Remark : When connecting via RS-485interface, a RS-232C RS-485 converter should be used.
21-On21-Os
4. PerformanceRepeatability : ±0.5% of full scaleLinearity : ±1% of full scaleZero drift : ±1% of full scale/week
(±2% of full scale/week; rangebetween 0 to 50ppm and 0 to200ppm)
(±2% of full scale/day; smallerthan 0 to 50ppm range)
Span drift : ±2% of full scale/week (±2% of full scale/day; smaller
than 0 to 50ppm range)Response time (for 90% FS response) :
Within 60 seconds including re-placement time of sampling gas(when gas flow rate is 0.5L/min)Gas replacement time dependson the number of measuringcomponents,and measuringrange
ffects of interfering gases
When sample gas contains gas components listed below, the measurement accuracy may suffer. Consult Yokogawa for countermeasures or effect on accuracy.
T03s.EPS
CO2 15% of CO2 is equivalent to 7 to 10 ppm of CO
15% of CO2 is equivalent to approx. 3 ppm of CH4
Analyzer
SO2 analyzer
CO analyzer
Interference gas Effect
NO2 50 ppm of NO2 is equivalent to -6 ppm of SO2
CH4 analyzer CO2
N2O 1000 ppm of N2O is equivalent to 80 ppm of CO
5. Standard Requirements for Sample GasFlow rate : 0.5 ±0.2L / minTemperature : 0 to 50˚CPressure : 10 kPa or less (Gas outlet side
should be open to the atmo-spheric air.)
Dust : 100µg/Nm3 in particle size of1µm or less
Mist : UnallowableMoisture : Below a level where saturation
occurs at 2˚C (condensation un-allowable).
Corrosive component: Hcl1 ppm or less
Standard gas for calibration:Zero gas ; Dry N2
Span gas ; Each sample gashaving concentra-tion 90 to 100% of itsmeasuring range(recommended).Gas beyond con-centration 100%FSis unusable.
In case a zirconia O2 analyzer isinstalled externally and calibra-tion is carried out on the samecalibration gas line:Zero gas ; Dry air or atmo-
spheric air (providedwithout CO2 sensor)
9 - 4 IM 11G02N01-01E
Span gas ; For other than O2
measurement, eachsample gas havingconcentration 90 to100% of its measur-ing range.For O2 measure-ment, O2 gas of 1 to 2vol%.
6. Installation Requirements• Indoor use. (Select a place where the equipment
does not receive direct sunshine, draft/rain or ra-diation from hot substances. If such a place can-not be found, a roof or cover should be preparedfor protection.)
• Avoide a place where receives heavy vibration.• Select a place where atmospheric air is clean.
9 - 5IM 11G02N01-01E
Without O2 analyzer External zirconia type O2 sensor (purchase separately: ZX8D ) External O2 analyzer (note 1)Built-in paramagnetic type O2 sensor
1st Component2nd Range(note 2)
2nd Component1st Range(note 2)
COCO2
SO2COCOSO2
SO2
NOSO2CO2COCH4NONOCO2NONO
ABCDEFGHJKLMPQRSTUN
N1
23
-A-B-C-D-F-G-H-J-K-L
050 ppm0100 ppm0200 ppm0250 ppm0500 ppm01000 ppm02000 ppm05000 ppm01%02%03%05%010%020%040%050%070%0100%Not available
CO
020 ppm (note 3)050 ppm0100 ppm0200 ppm0250 ppm0500 ppm01000 ppm02000 ppm05000 ppm01%02%03%05%010%020%040%050%070%0100%
050 ppm0100 ppm0200 ppm0250 ppm0500 ppm01000 ppm02000 ppm05000 ppm01%02%03%05%010%020%040%050%070%0100%Not available
IR400
ABCDEFGHJKLMPQRSTUN
VABCDEFGHJKLMPQRSTU
Model Suffix code Option code Description
Infrared gas analyzer 19-inch rack mounting type with slide rail
3rd1st 2nd 4thMeasurablecomponent(note 8)
O2 Analyzer
1st Component1st Range(note 2)
[Style:S3]
To be continued.
9.2 Model and Suffix codes
9 - 6 IM 11G02N01-01E
ABCDEFGHJKLMPQRSTUN
BCDEFGHJKLMPQRSTUN
BCDEFGHJKLMPQRSTUN
ABCDEFGHJKLMPQRSTUN
IR400
Model Suffix code Option code Description
2nd Component2nd Range(note 2)
3rd Component1st Range(note 2)
0100 ppm0200 ppm0250 ppm0500 ppm01000 ppm02000 ppm05000 ppm01%02%03%05%010%020%040%050%070%0100%Not Available
050 ppm0100 ppm0200 ppm0250 ppm0500 ppm01000 ppm02000 ppm05000 ppm01%02%03%05%010%020%040%050%070%0100%Not Available
0100 ppm0200 ppm0250 ppm0500 ppm01000 ppm02000 ppm05000 ppm01%02%03%05%010%020%040%050%070%0100%Not Available
4th Component1st Range(note 2)
050 ppm0100 ppm0200 ppm0250 ppm0500 ppm01000 ppm02000 ppm05000 ppm01%02%03%05%010%020%040%050%070%0100%Not Available
3rd Component2nd Range(note 2)
Infrared gas analyzer 19-inch rack mounting type with slide rail
To be continued.
Model and Suffix codes
9 - 7IM 11G02N01-01E
JEU
Rc 1/41/4 NPT
BCDEFGHJKLMPQRSTUN
123N
23N
-4-1
RT
IR400
Model Suffix code Option code Description
4th Component2nd Range(note 2)
0100 ppm0200 ppm0250 ppm0500 ppm01000 ppm02000 ppm05000 ppm01%02%03%05%010%020%040%050%070%0100%Not available
O2 Analyzer1st Range(note 2)
05%010%025%Not available
O2 Analyzer2nd Range(note 2)
010%025%Not available
Output 420 mA DC, non-isolation01 V DC, non-isolation
Piping
Indication, Power Cable (note 7) Japanese, Power Cable; rated voltage 125 V ACEnglish, Power Cable; rated voltage 125 V AC (UL)English, Power Cable; rated voltage 250 V AC (CEE)
/K
/A
/C
/P
/R
With O2 correction and O2 average value
With peak count alarm (note 4)
RS232C (note 9)
Analyzer internal purging
With dedicated cable
Option O2 correction and O2 avarage (note 4)
Communication
Internal purge (note 5)
Relay board (note 6)
Infrared gas analyzer 19-inch rack mounting type with slide rail
Model and Suffix codes
Footnotes:1: A signal from the external O2 analyzer should be 0-1 V DC linear to full scale.2: Possible combinations of ranges are specified in separate tables.3: Only available for CO2 measurement. Option code “/P,” Analyzer internal purging, must be specified.4: O2 correction is available only for NO, CO, and SO2. Both average value output after O2
correction and average O2 value output are provided at the same time. A peak count alarm can be provided only for CO measurement.5: When internal purging is specified with 3- or 4-component analyzers, only one set of gas inlet/outlet can be used and thus NO2/NO converter cannot be connected between two measurement sections.6: Should be specified when using a solenoid valve for automatic calibration.7: Suffix Codes “E” and “U” are power cables with different voltage rating and plug type. Select appropriate code according to the operating power supply voltage to be used in the field. Suffix Code “E” is of the North American plug type and “U” of the European type.8: For NOx measurement, a NO2/NO converter (P/N K9350LE or K9350LF) should be purchased separately.9: Should be specified when using ModbusTM communication.
9 - 8 IM 11G02N01-01E
Table 1. Measurable component and range - availability check table -
(1) Single-component analyzer (NO, SO 2, CO2, CO or CH 4)
(2) Double-component analyzer (NO and SO 2)
E
0500ppm
C 0200ppm
E
F
01000ppm
G
02000ppm
H
05000ppm
J
01%K
02%
0500ppm
F 01000ppm
G 02000ppm
D 0250ppm
H 05000ppm
J 01%
K 02%
L 03%
M 05%
P 010%
Q 020%
R 040%
S 050%
T 070%
U 0100%
A 050ppm
B 0100ppm
B
0100ppm
C
0200ppm
D
0250ppm
V
050ppm
A
020ppm
1st range
2nd range
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
C 0—200ppm
E
M0—5%
P0—10%
Q0—20%
S0—50%
T0—70%
0—500ppm
F 0—1000ppm
G 0—2000ppm
D 0—250ppm
H 0—5000ppm
J 0—1%
K 0—2%
L 0—3%
M 0—5%
P 0—10%
Q 0—20%
R 0—40%
S 0—50%
T 0—70%
U 0—100%
A 0—50ppm
B 0—100ppm
L0—3%
R0—40%
U0—100%
V 0—20ppm1st range
2nd range
:CO2 analyzer measurable range :CO analyzer measurable range :CH4 analyzer measurable range :SO2 analyzer measurable range :NO analyzer measurable range
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
SO2
B 0100ppmC 0200ppmD 0250ppmE 0500ppmF 01000ppmG
NO
02000ppm
A 050ppm
: Double components measurable range. 1st component ; NO, 2nd component ; SO2.
1st component(NO),1st range
2nd component (SO2), 1st range
H 05000ppm
E0500ppm
F01000ppm
G02000ppm
H05000ppm
B0100ppm
C0200ppm
D0250ppm
A050ppm
2
2222
2 2 2
* 1st range (low range) must meet the combination in above table. 2nd range, both NO and SO2 measurements are based on Table 1.
T04.EPS
9 - 9IM 11G02N01-01E
Table 3. Two-component analyzer (NO and CO)Both NO and CO analyzer must meet the range in Table 1; Single component analyzer.
Table 4. Two-component analyzer (CO 2 and CO)
CO
E F G HB C D
CO2
J01%
32.5
31
32
31
32
31
32
31
32
31
3235
3135
32.5
31
32.5
31
35
32
31
310
34
32
31
310
34
32
31 35
310
32
32
31
310
35
35
32.5
A
B
C
D
E
F
32
31
32310
31310
:Double components measurable. 1st component ; CO2, 2nd component ; CO.
2nd component (CO), 1st range
1st component(CO2),1st range
1st range (low range) must meet the combination in above table.2nd range, is specified;both CO2 and CO measurements are available up to 20 times of the 1st range.
is specified;CO measurement is available up to 20 times of the 1st range.CO2 measurement is available up to the ratio written after the mark.
is specified; both CO2 and CO measurements are available up to ratio The ratio, first value is for CO2, second value is for CO.
example: 3 2 3 5 means, 2nd range of CO2 is available up to double of 1st range, 2nd range of CO is 5 times of 1st range.
3 1 means only 1st range.
Note:
written after the mark.
A 050ppm
0100ppm
0200ppm
0250ppm
0500ppm
01000ppm
G
H
J
KL
MN
02000ppm
05000ppm
01%
02%
05%
010%
020%
050ppm 0100ppm 0200ppm 0250ppm 0500ppm 01000ppm 02000ppm 05000ppm
available up to
(For 0-200 ppm range, measurement is available up to 25 times.)
Table 5. Three-component analyzer (NO + SO 2 + CO)See Table 2 for NO + SO2 measurement of three-component analyzer (NO + SO2 + CO). See Table 1 for CO measurement.
Table 6. Four-component analyzer (NO + SO 2 + CO2 + CO)See Table 2 for NO + SO2 measurement and Table 4 for CO2 + CO measurement.
Table 7. O2 analyzer
2nd range1st range
1
20210%
30225%
025%
2 0210%
3 0225%
*O2 analyzer is selectable independently of combination with other components.
: Built-in O2 analyzer measurable range : External zirconia type O2 analyzer (in this case, Yokogawa s ZX8D Style C) measurable range
T07.EPS
9 - 10 IM 11G02N01-01E
Unit:mm
57
8 M
ax.
M4 (For mounting slide rail)
12
.7
12
3.8
28
8.9 31
5.8 4
66
.7
7.5
10
0.8
17
7
(22
)
26
.1
38
465
483
429
SPANENT
ZEROESCMODE
POWER
<Side View>
<Rear View>
<Top View>
INFRARED GAS ANALYZER
16
9
Power switch
Sample gas inlet Rc1/4 or 1/4NPT
Sample gas outlet Rc1/4 or 1/4NPT
(Purge gas inlet Rc1/4 or 1/4NPT)
<Front View>
<Rack mountimg hole>
M5
JIS : 100mmEIA : 101.6mm
Power inlet (100 to 240 V AC, 50/60 Hz)
Connector for imput/output terminal module
9.3 External Dimensions<Analyzer main unit> Unit : mm
9 - 11IM 11G02N01-01E
(140
)12
164
154
5
(54.
5) 29
316
6-[4.5 hole
150 150 8
(280)
1000630
8
Terminalblock
<TN1>
Terminalblock
<TN2>
Terminalblock
<TN3>
Terminalblock
<TN4>
Terminalblock
<TN5>
Connector <CN3>
Screw terminals M3.5
Input/output terminal module connector <CN1>
Cable for Connecting Input / Output Terminal: K9218SD (Accessory)
Dimensions for Mounting Input/Output Terminal Module
Mounting hole
Connector <CN2>
302
150 150
154
142
M4 screw
Cut M4 screw holes at 6 positions.Drill a rectangular hole of 302 3 142 mm or more in the center. F05.EPS
Unit: mm
Input/Output Terminal Module: K9218SC (Accessory)
9 - 12 IM 11G02N01-01E
35.3
60.
3
19.160.3 606.660.8
63263 Closed position flush or under
Closed
466.760.3
288.960.3
111.160.345460.3
565.160.3577.860.3
609.660.8
4.535.3 4.535.3123.860.3
25.460.3
15.960.5
12.760.3
12.760.3
(22.7) (9.5)
15.960.5
(24.
4)
Reinforcement plate
Cabinet member1.52t
Drawer member1.27t
Intermediatemember
Open
F06.EPS
EXTERNAL DEMENSIONS of ACCESSORY SLIDE RAILModel: 305A-24/Accuride International Inc.
19 inch rack mounting method:The mass of the instrument should be supported at the bottom of the unit (or the side of the unit when mounted with the slide rails).Also, for facilitate maintenance, a structure which allows extraction of the main unit by using the slide rail is recommendeg.
465
450 or more
101.
6or 10
0
465
450 or more
Slide rail mounted type
Rack size
Guide rail mounted type
Rack size
Mounting diagram
Slide rails
Mounting diagram
Guide rails
Guide railsFor the guide rail mounted type, a maintenance space (200 mm or more) should be provided on top of the main unit.
101.
6or 10
0
9 - 13IM 11G02N01-01E
STANDARD ACCESSORIES
T09.EPS
Note: Quantity in this is number of accessories supplied as standard. For instance, two K9218SE parts, i.e., two slide rails, are supplied as standard. When ordering separately, the required number of should be considered.
1
Name
1
QtyPart Number
1
K9218SD
2K9218SB
Cable
K9218SC
K9218SA
Input/outputterminal module
Power cable
Fuse
External terminal module
Connection cable between main unit and input/output terminal module (1 m)
Slide rail
standard inlet type (2.5 m )
Description
replacement fuse (250 V AC, 3 A, delay type) 31
2K9218SESlide rail
Dedicated Zirconia O 2 Sensor(to be purchased separately)For O2 correction, the IR400 can accept linealized 0 to 1 VDC signal coming from an analyzer calibrated to 0 to 25%O2 of full scale. Dedicated zirconia O2 sensor , ModelZX8D, is available from Yokogawa.
Measuring method:Zirconia systemMeasurable component and measuring range:
Measurable component Minimum range Maximum range
Oxygen (O2) 025 vol% 0225 vol%
T11.EPSRepeatability: Within ± 0.5% of full scaleLinearity: Within ± 1% of full scaleZero drift: Within ± 1% of full scale/weekSpan drift: Within ± 2% of full scale/weekResponse time:
Approx. 20 seconds (for 90% response)
Remark: The Zirconia system, due to its principle, may produce a measuring error due to relative concentration versus the com-bustible O2 gas concentration. Also, a corrosive gas (SO2 of 250 ppm or more, etc.) may affect the life of the sensor.
Gas inlet/outlet size: Rc1/4Power supply: 90 to 126 V AC or 200 to 240 V AC,
50/60 HzEnclosure: Steel casing, for indoor applicationIndication: Temperature indication (LED)Temperature alarm output:
Contact output 1a contact,Contact capacity 220 V AC, 1 A (resistiveload)
Safety and EMC conforming standards:Safety: EN61010-1
Pollution degree 2 (Note)Installation category II (Note)
Note: Installation category, called over-voltage category, specifies impulse withstanding voltage. Category II is for electrical equip-
ment. Pollution degree indicates the degree of
existence of solid, liquid, gas or other inclusions which may reduce dielectric strength. Degree 2 is the normal indoor
environment.EMC: EN61326
EN61000-3-2EN61000-3-3
Dimensions (H 3 W 3 D):140 3 170 3 190 mm
Weight: Approx. 3 kgFinish color: Munsell 5Y 7/1
ZX8D
pCpD
Powersupply
Model Suffix code Option code
2523
Style code
902126 V AC, 50/60 Hz2002240 V AC, 50/60 Hz
Dedicated zirconia O2 sensor
Description
Style C (Non-CE conformity) Style D (CE conformity)
T13.EPS
External Dimensions of ZX8D
+ -
POWER FUSE3A
OFF
1 2 3 4 5 6 7
AC E OUT ALM
800
7-M4
Gas inlet17 Gas outlet172.3
140
131
110
170
152
5
OXYGEN ANALYZER
TEMP
F09.EPS
External Connection Diagram
1 2 3 4 5 6 7
AC power50/60Hz
supply E Outputto IR400
Temperaturealarm output
+ -
F09-2.EPS
Measured gas flow rate:0.5 ± 0.25 L / min
Unit: mm
9 - 14 IM 11G02N01-01E
Dedicated relay board (Option code: /R)This relay board receives signals from connector CN3 ofthe IR400 I/O terminal module and activates the calibrationsolenoid valve directly.
• Relay contact :1 normally closed contactContact capacity; 250 V AC/2 A(resistive load)
Part Numbers
Item QtyPart No.
1
1K9218SG
K9218SFRelay board
Cable
For external contact point
Description
For relay boardT10.EPS
External Dimensions(Unit: mm)
CN1CN9
CN2 CN3 CN4 CN5
CN13
CN6 CN7 CN8
F10.EPS
7565
150
140
15 Max.
ø4.5Mounting hole
Connections
CN1CN9
Solenoid valvedriving power
Solenoid valve
CN2
CN13
CN3
Reserved
For switchingsample
CN3
For zero gas
Relay board
I/O terminal module
Dedicated cable (D-sub 9-pin straight cable 1.5 m)
For span gas
CN4 CN5 CN6 CN7 CN8
Ch1 Ch2 Ch3 Ch4 Ch5
F11.EPS
Contact action• During measurement: CN1; ON
Others; OFF• During calibration: CN1; OFF
Others; Contact corresponding to calibration timing is ON
Recommended Connector• CN1 to CN9: Housing; VHR-2N
(Japan Solderless Terminals (JST))Contact; SVH-21T-1.1
(Japan Solderless Terminals (JST))
9 - 15IM 11G02N01-01E
NO2/NO ConverterPart number: K9350LE (Non-CE conformity) K9350LF (CE conformity)Mounting: Indoor surface mountingTarget Gases: General boiler exhaust gas,
atmosphereCatalyst: Amount; 2 cm3
Replacement cycle; Approx. 12 months (at flow rate of 0.3 L/min with 5% O2, 10 ppm NO) Temperature setpoint; 210 6108C (Sensing tip: K thermocouple)
Wetted materials: Ceramic, Viton, glass filter, SUS31Conversion efficiency: 90% or higher, conforms to JISGas Flow Rate: 0.5 L/minAmbient Temperature: -5 to +458CPower Supply: 100 VAC , 50/60 Hz (K9350LE) 100 to 240 VAC , 50/60 Hz (K9350LF)
Power Consumption:Approx. 85 VA
Safety conforming standards:EN61010-1Pollution degree 2Installation category II
Weight: Approx. 1.1 kg (K9350LE)Approx. 1.2 kg (K9350LF)
Sample gas requirements:Dust/drain removed, gas temperature at1508C or less
T15.eps
Catalyst for NO2/NO converterGlass wool for NO2/NO converterFitting for NO2/NO converter
Part No.K9350LPK9350LQK9350LV
Qty222
One-year-Use Spare Parts
Item
External Dimensions
7 8
100V AC(Shared with internal wiring)
Wiring Diagram
Gas Inlet [9.5/[5.5
4-[5
135
(120
)10
5
(210
)
App
rox.
130
110
(150)
Gas Outlet [9.5/[5.5
Unit: mm
K9350LE
K9350LF
≈
Blank Page
CustomerMaintenanceParts List
All Rights Reserved, Copyright © 2002, Yokogawa Electric Corporation.Subject to change without notice.
Model IR400Infrared Gas Analyzer
CMPL 11G02N01-01E1/2
1st Edition : Aug. 2002 (YK)
1, 2
Item Parts No. Qty. Description
1 K9218SB 1 Fuse (Time Lag Fuse)
2 K9358QL 1 Fuse holder
2
CMPL 11G02M01-01E CMPL 11G02N01-01E2/2
1st Edition : Aug. 2002 (YK)
3
3
45
Item Parts No. Qty. Description
3 K9358SD 2 O-ring
4 K9358SQ 1 O-ring (Out-side)
5 K9358SR 1 O-ring (Out-side)
d Pipe Cell
d Block Cell
Revision Record
Manual Title : IR400 Infrared Gas Analyzer
Manual Number : IM 11G02N01-01E
Edition Date Remark (s)
1st Oct. 2002 Newly published
2nd Aug. 2006 Fully revised due to style change to S3
P. v., Added "After-Sales Warranty"
Section 2.1, "Name and description of main unit": Changed drawing partly
Section 3.4, "Piping": Changed drawing partly
Section 3.4.7, "Example configuration of gas sampling system": Changed drawing
Section 5.1, "Name and description of operation panel": Changed drawing partly
Section 9.1.1, "Standard specifications": Changed table "Measurable component and
measurement range"
Section 9.1.5, "Standard Requirements for Sample Gas": Changed description
Section 9.2, "Model and Suffix codes": Changed table partly
Section 9.2, "Measurable component and range 2 availability check table 2 ": (4),
"Double-component analyzer": Changed table partly
3rd Aug. 2007 Section 5.4, "General operation": Partially changed the figure.
Section 6.2, "Calibration setting": Partially changed the figure and descriptions in "Setting
range of values."
Section 6.4.1, "Auto calibration": Partially changed the figure and descriptions in Caution.
Section 6.4.2.1, "Execution of auto calibration": Partially changed descriptions in (3).
Section 6.4.2.2, "Forced stop of auto calibration": Partially changed descriptions in (3).
Section 6.5.1, "Auto zero calibration": Partially changed descriptions in Caution.
Section 6.5.2.1, "Forced run/stop of auto zero calibration": Partially changed descriptions in
(3).
Section 6.5.2.2, "Forced stop of auto zero calibration": Partially changed descriptions in (3).
Section 6.6, "Peak alarm setting": Partially changed the figure and descriptions.
Section 6.9.1, "Zero calibration": Partially changed the figure and descriptions in Caution.
Section 6.9.2, "Span calibration": Partially changed the figure.
Section 9.1, "General specifications," Safety and EMC conforming standards: Changed
descriptions and added Note.
Section 9.2, "Model and Suffix code": Deleted Footnote 8 and changed descriptions of
Footnote 8 (former 9).
Section 9.3, "External Dimensions": Changed power supply specification of NO2/NO
converter.
Section 9.3, "External Dimensions," NO2/NO Converter: Changed "Part number"
descriptions, added "Safety conforming standards" descriptions, deleted "CE Marking"
descriptions, and changed "External Dimensions" drawings,