Recent results

Marc & Qian

Page 2: Recent results

Outline

• 3D uniformity scan

• End cap TOF counter test result

• Report of moving laser box

Page 3: Recent results

3D Uniformity Scan

Page 4: Recent results

• We used USTC DAQ system to do the 13x13 3D uniformity scan = 171 total files, which covers a 3mm x 3mm square area. The reason for 171 is because we repeated the origin measurement three times (1st time is at initial origin, 2nd time is on the grid, 3rd time is final file.)

13

12

11

10

9

8

7

6

5

4

3

2

1

14

26

39

27

40

52

65

53

66

78/79

164

163

150

149

136

135

122

121

108

107

94

93

80

170

157

156

143

142

129

128

115

114

101

100

87

86

X

Y

(0.180,0.215)

Origin (0.240,0.275)

(0.300,0.335)

Page 5: Recent results

The percent variation over 2.5mm x 2.5mm ~ 13%.

Page 6: Recent results

Summary / Conclusion:

1. We did this test before we move the “Moon” Box.2. We know from our 2 D linear X,Y scans via O-Scope, that we are able to

achieve uniformity on the order of ~4%.3. We now conclude that the USTC/main DAQ interface requires more fine

tuning.4. We know from the ref-PMT data (reading from faster O-Scope) that the

laser is quite stable.5. (To be continued…)

Page 7: Recent results

End cap TOF counter test

Page 8: Recent results

Hole A

Hole B

We set the HV of EC and Barrel TOF separately to achieve both gain = 1E+5.

The Barrel TOF signal’s PH is 165.1 mV.

EC TOF Signal PH (mV) Wiggle fiber

Hole A 161.1 160~161 mV

Hole B 260.0 255~260 mV

Hole AHole B

For Hole B, if we use solenoid 3 with 63% transparency, the PH is 164 mV

Page 9: Recent results

Summary / Conclusion:

1. According the EC TOF Fiber installation method, the both holes are easy to achieve.

2. Using Hole A, we don’t need to use solenoid 3 or filter, the EC TOF and Barrel TOF signals’ PH are almost the same, and Hole A is more stable than Hole B when we wiggled the fiber.

Marc & Qian老善 刘倩

Page 10: Recent results

Moving Laser Box

Page 11: Recent results

Before we moved the Moon Box, we tested the Ref-PMT PH:

Origin PMT (mV) PH (mV)

CH1(east) (0.305,0.330) 460 202.0

CH2(west) (0.240,0.275) 486 202.0

Page 12: Recent results

We’ve moved the “Moon Box” into the rack.

Then we tested the PH:CH1 (460): 185.8 mVCH2 (486): 195.8 mV

Page 13: Recent results

Summary / Conclusion:

1. After we moved the moon box, we need to do the alignment again.2. The PH decreased about 8%.

Marc & Qian老善  刘倩