Download - Rapid electrochemical genosensor assay using a streptavidin carbon-polymer biocomposite electrode
Biosensors and Bioelectronics 19 (2003) 165�/175
www.elsevier.com/locate/bios
Rapid electrochemical genosensor assay using a streptavidin carbon-polymer biocomposite electrode
E. Williams 1, M.I. Pividori 2, A. Merkoci, R.J. Forster 1, S. Alegret *
Grup de Sensors i Biosensors, Departament de Quımica, Universitat Autonoma de Barcelona, 08193 Bellaterra, Catalonia, Spain
Received 9 May 2002; received in revised form 10 April 2003; accepted 23 April 2003
Abstract
A sensor capable of detecting a specific DNA sequence was designed by bulk modification of a graphite epoxy composite
electrode with streptavidin (2% w/w). Streptavidin is used to immobilise a biotinylated capture DNA probe to the surface of the
electrode. Simultaneous hybridisation occurs between the biotin DNA capture probe and the target-DNA and between the target-
DNA and a digoxigenin modified probe. The rapid binding kinetic of streptavidin�/biotin allows a one step immobilisation/
hybridisation procedure. Secondly, enzyme labelling of the DNA duplex occurs via an antigen�/antibody reaction between the Dig-
dsDNA and an anti-Dig-HRP. Finally, electrochemical detection is achieved through a suitable substrate (H2O2) for the enzyme-
labelled duplex. Optimisation of the sensor design, the modifier content and the immobilisation and hybridisation times was attained
using a simple nucleotide sequence. Regeneration of the surface is achieved with a simple polishing procedure that shows good
reproducibility. The generic use of a modified streptavidin carbon-polymer biocomposite electrode capable of surface regeneration
and a one step hybridisation/immobilisation procedure are the main advantages of this approach. In DNA analysis, this procedure,
if combined with the polymerase chain reaction, would represent certain advantages with respect to classical techniques, which prove
to be time consuming in situations where a simple and rapid detection is required. This innovative developed material may be used
for the detection of any analyte that can be coupled to the biotin�/streptavidin reaction, as is the case of immunoassays.
# 2003 Elsevier B.V. All rights reserved.
Keywords: Streptavidin�/biotin; Streptavidin carbon-polymer biocomposite; Horseradish peroxidase labelling; Amperometric genosensor; DNA
hybridisation; Digoxigenin; Anti-digoxigenin antibody
1. Introduction
‘World-wide infectious diseases account for approxi-
mately 40% of the total 50 million annual deaths, with
microbial diseases accounting for the majority of deaths in
developing countries ’ (World Health Organisation offi-
cial statistics 2000). Clearly, the development of a rapid,
sensitive, selective and highly accurate method for the
detection of DNA is one of the key health issues facing
our society today. Two competitive approaches exist for
DNA detection: direct sequencing and DNA hybridisa-
* Corresponding author. Tel.: �/34-93-581-1976; fax: �/34-93-581-
5972.
E-mail address: [email protected] (S. Alegret).1 Dublin City University, Glasnevin, Dublin 11, Ireland.2 Facultad de Bioquımica y Ciencias Biologicas, Universidad
Nacional del Litoral, Santa Fe, Argentina.
0956-5663/03/$ - see front matter # 2003 Elsevier B.V. All rights reserved.
doi:10.1016/S0956-5663(03)00171-4
tion. Although DNA sequencing has proved indispen-
sable for resolving the genetic make up of many
pathogenic species, as an analytical technique is too
time consuming (Turner et al., 1997). DNA hybridisa-
tion proves to be a simpler, quicker and more selective
method for the detection of previously known DNA
sequences.
The possibility of direct electrochemical detection of
DNA was initially proposed by Palecek (1958, 1960),
who recognised the capability of both DNA and RNA
to yield reduction and oxidation signals.Advancements in the areas of genetics with regards to
the elucidation of specific genes responsible for various
pathogenic diseases have encouraged research in the
area of genosensor development. This advancement is
shown by a large surge of publications relating to
electrochemical DNA detection as recently reviewed by
Pividori et al. (2000) and Palecek and Fojta (2001).
E. Williams et al. / Biosensors and Bioelectronics 19 (2003) 165�/175166
While many procedures still rely on the oxidation of the
guanine residues (Armistead and Thorp, 2001; Burrows
and Muller, 1998), much work has been done in
perfecting the immobilisation of DNA onto an electrodesurface along with the incorporation of electroactive
markers.
Extensive studies by Barton and co-workers (Erkkila
et al., 1999; Rajksi et al., 2000) have demonstrated the
capability of certain compounds to interact either by
intercalation, hydrophobic binding to the major and
minor grooves of DNA, or electrostatic interaction with
dsDNA, such as Ru(II) (Armistead and Thorp, 2000)Khellin (Radi, 1999) or Co(phen)3
3� (Li et al., 1997).
While, such a phenomena is of great interest in terms of
drug delivery, as in the case of daunomycin, it does not
lend itself easily for the development of genosensors due
to problems of non-specific adsorption onto the elec-
trode surface. Additionally, it has been reported (Erk-
kila et al., 1999) that the extent of intercalation depends
on the DNA sequence, favouring guanine rich DNAsequences. Such favouritism limits the range of its
application.
A more frequent use of electroactive markers involves
their attachment to the DNA duplex by direct covalent
attachment to ssDNA, utilisation of the biotin�/strepta-
vidin chemistry (Alfonta et al., 2001; Pividori et al.,
2001a,b) and the digoxigenin�/antidigoxigenin reaction
(Lumley-Woodyear et al., 1999). Ferrocene derivativesare commonly used as electroactive indicators by
immobilising them to a DNA probe (Xu et al., 2001;
Takenaka et al., 1994; Uto et al., 1997). More recent
studies by Kizek et al. (2002) have shown how using
osmium labelled ssDNA results in detection limits in the
region of 5�/10 ng/ml. Other studies show the labelling
of DNA with enzymes such as peroxidase, a known
redox enzyme which due to its slow heterogeneouselectron transfer kinetics require an additional electron
mediator to ascertain the response. Enzymes that have
been incorporated are horseradish peroxidase, (Pividori
et al., 2001a,b; Alfonta et al., 2001; Chen et al., 2000;
Lumley-Woodyear et al., 1999) and alkaline phospha-
tase (Palecek et al., 2002).
Various methods for the attachment of ssDNA
probes to the surface of the electrode exist. In brief,these methods may involve the use of an electrochemi-
cally pre-treated carbon electrode resulting in an in-
crease in the hydrophilicity of the transducer.
Subsequently, holding the electrode at positive poten-
tials results in the adsorption of the DNA probe through
hydrophilic and electrostatic attraction (Wang et al.,
1996). Covalent attachment of ssDNA to the surface of
the electrode can result in strong stable linkage with ahigh surface coverage, a key feature for the development
of sensors. Numerous methods for the covalent attach-
ment of biocomponents to the surface of electrodes are
reported in literature (Lee and Shim, 2001; Millan et al.,
1992; Millan and Mikkelsen, 1993), resulting in an
activated electrode surface which is capable of binding
with ssDNA. Covalent bonding of ssDNA to gold
electrodes has been achieved by attaching a mercapto-
hexyl group to the 5?-phosphate end of a DNA strand
(Hashimoto et al., 1994, 1998; Herne and Tarlov, 1997).
The aim of these efforts is to produce self assembled
monolayers similar to those described elsewhere (Nuzzo
and Allara, 1983).
In most of the aforementioned techniques, (Wang et
al., 1996, 1997a,b; Wang and Kawde, 2001; Marrazza et
al., 1999; Chiti et al., 2001) immobilisation of the DNA
probe depends on the chemical or electrochemical pre-
treatment step of the transducer. Many systems have
been developed involving the immobilisation of biotin
on the surface of electrodes in order to use biotin�/
avidin�/biotin sandwich techniques. One such immobi-
lisation procedures was proposed by Pantano et al.
(1991). However, a chemical pre-treatment step is again
required to immobilise the biotin onto the electrode
surface. Further advances and manipulation of the
biotin�/avidin�/biotin chemistry have been achieved
using UV light to initiate the attachment of a photo-
active biotin molecule to a transducer surface (Sha-
mansky et al., 2001).
An attractive feature when developing sensors is the
possibility of surface regeneration for repeated use. In
order to achieve these objectives in the aforementioned
references it is necessary to denature the DNA duplex.
Wang et al. (1998) have overcome the necessity of a
denaturation step by using a ssDNA probe dispersed
uniformly in the matrix of a polishable and renewable
carbon paste electrode. Here, the carbon paste acted as a
reservoir for ssDNA probes, such that each nucleic acid
assay was performed on a new surface using the same
electrode. Again however, the surface of the electrode
had to be activated.
This report describes the bulk modification of a
carbon-polymer composite electrode with streptavidin.
Exploitation of the strong interaction between biotin�/
streptavidin allows immobilisation of a biotin labelled
capture probe, containing the DNA sequence comple-
mentary to the target probe. Hybridisation of the target-
DNA to the capture probe and an antigen labelled DNA
sequence is achieved simultaneously. Electrochemical
detection is achieved when an enzyme-labelled antibody,
specific to the antigen (digoxigenin) labelled DNA is
added. One of the main advantages of this proposed
genosensor design is the capability of surface regenera-
tion through a simple polishing procedure along with a
one step immobilisation and hybridisation procedure.
To the best of the author’s knowledge the use of
immobilised streptavidin within a carbon composite in
addition to a one step immobilisation/hybridisation
procedure has not been reported yet.
E. Williams et al. / Biosensors and Bioelectronics 19 (2003) 165�/175 167
2. Materials and instrumentation
2.1. Chemicals and solutions
The composite electrodes were prepared using gra-
phite powder with a particle size of 50 mm (BDH, UK),
Epotek H77 (epoxy resin) and a hardener (Epoxy
Technology, USA). The DNA oligomers were obtained
from TIB MOLBIOL (Germany). The base sequences of
these oligomers can be seen below:
. Capture probe, biotin�/DNA(A): biotin-5?-AAA
AAA AAA AAA AAA AAA AA
. Target-DNA(B): 5?-AAA AAA AAA AAA AAA
AAA AAA AAA ATT TTT TTT TTT TTT TTT
TTT TTT TT-3?. Digoxigenin�/DNA(C): Dig-3?-TTT TTT TTT TTT
TTT TTT TT-5?
The hybridisation solution (10�/ SSC, 2�/ Den-
hardt’s, 200 mg/ml chloroform extracted salmon testes
DNA), bovine serum albumin, sodium dodecyl sulphate,
Tween 20, hydroquinone, and HRP�/biotin (peroxidase
biotinamidocaproyl) were purchased from Sigma. The
anti-digoxigenin horseradish peroxidase (anti-Dig-
HRP) and streptavidin were purchased from Roche
(Germany). Tris�/HCl, NaCl and hydrogen peroxide
were purchased from Merck (Germany). All other
reagents were of the highest grade available. All aqueous
solutions were prepared using double distilled water.
Fig. 1. Schematic representation of the DNA analysis based on an
electrochemical streptavidin carbon-polymer biocomposite. (A) Strep-
tavidin modified composite electrode. (B) One step immobilisation/
hybridisation procedure: target-DNA is hybridised with the biotiny-
lated capture probe and with the digoxigenin modified probe. The
complex biotin�/dsDNA-Dig is immobilised on the electrode surface
by linking the biotinylated capture probe with the streptavidin present
in the biocomposite. (C) Enzyme labelling based on the immunological
reaction between the immobilised dsDNA-Dig with anti-Dig-HRP.
(D) Electrochemical detection of the enzyme (HRP) labelled dsDNA.
2.2. Instrumentation and electrochemical measurements
Amperometric measurements were done with a LC-
4C Amperometric Controller (BAS Bioanalytical Sys-
tem, USA). Cyclic voltammetry measurements were
carried out using Autolab PGSTAT 20 (Eco-chemie,
The Netherlands). A platinum auxiliary electrode (Cri-
son 52�/67 1) and a double junction Ag/AgCl reference
electrode (Orion 900200 with 0.1 M KCl as an external
reference solution) were used. The carbon-polymer
composite electrode modified with streptavidin made
in our laboratory was used as the working electrode.
The incubations at controlled temperature were carried
out in the Eppendorf Thermomixer 5436.
Both cyclic voltammetry and chronoamperometry
were done in a 20 ml electrochemical cell. The electrolyte
in all cases was 0.1 M phosphate buffer, pH 7.4,
containing 0.1 M KCl and 1.81 mM hydroquinone.
For cyclic voltammetry the potential window ranged
between �/0.75 and 1.25 V (vs. Ag/AgCl). For chron-
oamperometry the applied potential was �/0.250 V (vs.
Ag/AgCl).
2.3. Construction of the streptavidin carbon-polymer
biocomposite electrodes
Graphite powder and epoxy resin were mixed in a 1:4(w/w) ratio. For every gram of graphite/epoxy mixture,
an additional 20 mg of streptavidin were added*/
resulting in a 2% (w/w) streptavidin carbon-polymer
biocomposite. This mixture was thoroughly hand mixed
to ensure the uniform dispersion of the streptavidin and
the carbon throughout the polymer. The resulting paste
was placed in a PVC cylindrical body (6 mm i.d.) that
had an electrical contact placed at a depth of 3 mm. Thecomposite was cured at 40 8C for 1 week (Santandreu et
al., 1997). Prior to each use the biocomposite was wetted
with doubly distilled water and then polished. Firstly
with abrasive paper and then with alumina paper
(polishing strips 301044-001, Orion) to give a smooth
mirror finish with a fresh renewable surface.
3. Procedure
3.1. Immobilisation and hybridisation
Fig. 1 illustrates the experimental procedure consist-
ing of three steps. Firstly, to an eppendorf containing a
volume of 120 ml of the hybridisation solution, pre-
Fig. 2. Cyclic voltammograms of (A) the streptavidin carbon-polymer
biocomposite, and (B) the carbon-polymer composite electrode. Scan
rate was 0.05 V/s in 0.1 M phosphate buffer pH 7.4, with 0.1 M KCl
and 1.81 mmol hydroquinone. All potential are recorded vs. Ag/AgCl
reference electrode.
E. Williams et al. / Biosensors and Bioelectronics 19 (2003) 165�/175168
viously equilibrated to 42 8C, 10 ml of each of the three
oligonucleotides, mentioned in Section 2.1, were added.
The final amount of all three DNA strands in the
hybridisation solution is 150 pmole. To this eppendorf,a freshly polished streptavidin carbon-polymer biocom-
posite electrode (Fig. 1A) was immersed and shaken
gently for 60 min at 42 8C. Immobilisation of DNA
occurs through the biotinylated capture probe. Simulta-
neous hybridisation occurs between the biotinylated
capture probe and the target-DNA and between the
target-DNA and a Dig-DNA as seen in Fig. 1B.
Removal of the electrode from the eppendorf is followedby a gentle washing in the thermomixer at 42 8C for 5
min using 0.01 M Tris�/HCl at pH 7.4, thereby removing
all unbound DNA.
3.2. Enzyme labelling
The next stage involves the attachment of the enzyme
label, anti-Dig-HRP to Dig-dsDNA, Fig. 1C. This
attachment is achieved by adding 10 ml of anti-Dig-HRP (12.45 mg/ml) to an eppendorf containing a volume
of 140 ml of blocking solution (1�/ PBS, 2% w/v BSA,
0.1% Tween 20, 5 mM EDTA) (Kricka, 1992), pre-
viously equilibrated to 37 8C. To this eppendorf, the
DNA modified streptavidin carbon-polymer biocompo-
site electrode was immersed and shaken gently for 40
min at 37 8C. The electrode is then washed with a post-
incubation solution (10 mM sodium phosphate pH 6.5,0.5 M NaCl, 0.05% w/v Tween 20, 0.1% w/v BSA, 1 mM
EDTA) (Kricka, 1992), with slight agitation at 37 8C for
5 min using the thermomixer. Once the immobilisation,
hybridisation and enzyme labelling have been achieved
the electrodes if not required immediately were stored in
Tris buffer pH 7.4 at 4 8C thereby allowing electro-
chemical measurements can be carried out later on.
3.3. Electrochemical enzyme activity determination
The amount of enzyme-labelled antibody bound to
the Dig-DNA duplex was determined using the follow-
ing electrochemical detection procedure (Fig. 1D).
Experiments were carried out at 229/2 8C using the
hydroquinone electrolyte solution mentioned in Section
2.2. Hydroquinone buffer was prepared daily because ofits sensitivity to photo-oxidation, by adding the correct
quantity of hydroquinone to a 0.1 M phosphate buffer
at pH 7.4 and 0.1 M KCl. This buffer was purged with
nitrogen prior to use for at least 1 h. Chronoampero-
metric experiments were then performed at a potential
of �/0.250 V vs. Ag/AgCl electrode. This potential of �/
0.250 V was chosen from cyclic voltammetry experi-
ments Fig. 2A. For the enzymatic activity determina-tions the freshly prepared carbon biocomposite
containing the immobilised enzyme-labelled dsDNA
was immersed in the hydroquinone electrolyte solution
and allowed to equilibrate for 10 min. Subsequently,
H2O2 was added to a final concentration of 176 mM,
which corresponds to the H2O2 concentration capable to
saturate the whole enzyme amount employed in the
labelling procedure. At Fig. 1D and Eq. (1) are
represented the reactions that occur at the genosensor
surface polarised at �/0.250 V (vs. Ag/AgCl) uponH2O2 addition in the presence of hydroquinone. After
addition of the substrate, the current was again mea-
sured for a further 10 min. This time is chosen to ensure
that a steady state current had been reached maximising
the catalytic activity of the enzyme horseradish perox-
idase. The electrode response is defined as the difference
in current before and after the addition of H2O2.
HRP(red)�H2O2 0 HRP(ox)�H2O
HRP(ox)�HQ(red) 0 HRP(red)�HQ(ox)HQ(ox)�ne� 0 HQ(red)
(1)
3.4. Variation of streptavidin content in the carbon-
polymer biocomposite. Stability study
The optimisation of the streptavidin content in the
biocomposite was done using electrodes with fourdifferent streptavidin contents (1, 1.5, 2 and 5%). These
electrodes were evaluated for 3 weeks storing them in an
upright position at 4 8C.
E. Williams et al. / Biosensors and Bioelectronics 19 (2003) 165�/175 169
The stability of the streptavidin biocomposite electro-
des was also studied over a 12-week period, during
which time the electrodes were also stored in an upright
position at 4 8C.In all cases four electrodes were used, and the same
polishing procedure described in Section 2.3 was fol-
lowed.
For these evaluations, the procedure involved immer-
sing the streptavidin modified electrodes in a 150 ml
biotin�/HRP in excess concentration of 50 mg/ml for 60
min at 37 8C with gentle agitation again using the
thermomixer. It is known that streptavidin forms acomplex with biotin at room temperature with an
association constant in the range of 1015/M (Gonzalez
et al., 1999). Electrochemical detection was carried out
as described in Section 3.3 above for the detection of
HRP labelled DNA duplex.
3.5. Non-specific adsorption evaluation. Study of the
mechanism of the one step immobilisation/hybridisation
procedure
As with all methods that use an enzyme label, the
problem of non-specific adsorption needs to be ad-
dressed. Furthermore, knowledge of the one step
immobilisation and hybridisation mechanism is useful
to understand how the genosensor works.
For these studies, (i) carbon-polymer compositeelectrodes, which contains no streptavidin, were used
and the same procedure described above was followed.
In the other three assays (ii, iii and iv), the streptavidin
carbon-polymer biocomposite electrodes prepared as
described above were used. The same procedure was
followed but omitting (ii) the target-DNA and (iii) the
biotinylated capture probe. Finally, assay (iv) was done
using all three oligonucleotide components to evaluatethe genosensor in capture format. In all cases (assays i,
ii, iii and iv) four electrodes were used following the
same polishing procedure described in Section 2.3.
3.6. Optimisation of genosensor response
3.6.1. Immobilisation as the limiting step
This assay was done using a two step procedure.
Immobilisation times were varied from 10 to 120 min atregular intervals of 10 min. For subsequent hybridisa-
tion, a time of 60 min was chosen in agreement with
literature (Gingeras et al., 1987). Detection of the
hybridisation was achieved using the electrochemical
procedure described above.
3.6.2. Hybridisation as the limiting step
This assay was also done using a two step procedure.In this instance, a constant immobilisation time of 60
min was used. Again, a variation of the hybridisation
time from 10 to 120 min at regular intervals of 10 min
was chosen. All other experimental conditions were the
same as above.
3.6.3. Variation of time for the one step hybridisation/
immobilisation procedure
This final assay used a one step procedure. A study of
the experimental time for the one step hybridisation/
immobilisation procedure vs. current response was
done. Again, a variation of the one step proceduretime between 10 and 120 min at regular intervals of 10
min was chosen. All other experimental conditions are
the same as above.
In all experiments, four different electrodes were used
and the electrode surface was renewed before each
assay.
4. Results and discussion
4.1. Cyclic voltammetry of streptavidin modified
electrodes
Cyclic voltammograms of the 2% (w/w) streptavidin
modified composite electrode with immobilised DNA in
a 1.81 mM hydroquinone phosphate buffer solution are
shown in Fig. 2A. The purpose of cyclic voltammetry is
to establish the location of the reduction potential ofbenzoquinone (HQox). In Fig. 2A*/for streptavidin
carbon-polymer biocomposite electrode*/the upper
curve represents the oxidation of hydroquinone to
benzoquinone with an Ep at 0.537 V, while the lower
curve represents the reverse reaction with an Ep at �/
0.230 V. Fig. 2B shows the voltammetry for a carbon-
polymer composite (without streptavidin), an electrode
used in the non-specific adsorption studies. In thisinstance the location of Ep for the oxidation and
reduction peak potentials is at 0.249 and �/0.127 V,
respectively. The difference in the peak potentials
between both composite electrodes may be attributed
to the presence of streptavidin, which may affect the
conductivity, surface properties and therefore the loca-
tion of the formal potential (Cespedes et al., 1996). It is
known that in the presence of a partially blockedelectrode the value for the peak-to-peak separation
will increase, with the oxidation peak moving in a
more positive direction and the reduction peak moving
in a more negative direction. However, since in this
instance only 2% of the total surface area of the
electrode contains streptavidin available for binding
with a biotin-probe, any shift in potential would be
negligible if compared with a genosensor design byHerne and Tarlov (1997) that use a self assembled
monolayer with the intention of achieving complete
surface coverage.
E. Williams et al. / Biosensors and Bioelectronics 19 (2003) 165�/175170
4.2. Variation of streptavidin content in the carbon-
polymer biocomposite
From a previously published report (Alegret, 1996) it
is known that varying the biocomponent content in a
carbon-polymer biocomposite can adversely affect the
overall electrochemical performance of a given sensor.
This loss of electrochemical performance is attributable
to the increased concentration of modified species
resulting in a separation of the graphite particles,
causing a reduction of the conductivity. For this reason
it was decided to optimise the streptavidin content in the
composition. This was achieved by constructing bio-
composites with 1, 1.5 and 2% w/w of streptavidin. The
experiments were carried out as described previously,
over a 3-week period during which the electrodes were
stored in an upright position at 4 8C. In all experiments,
four different electrodes were used and polished to a
mirror finish prior to immobilising biotin�/HRP on the
transducer surface. From Fig. 3 it can be seen that all
three types of biocomposite electrodes decrease their
current response with time, this is an issue to be
addressed in the following section. Additionally, there
is a decrease in current response for electrodes having a
lower w/w of streptavidin available for reaction with the
biotin�/HRP. This data suggests that increasing the
percentage of streptavidin from 1 to 2% in the biocom-
posite does not adversely affect the electrochemical
response due to loss of conductivity. Additionally, an
electrode with 5% streptavidin was also prepared. Upon
immersion into the electrolyte solution this electrode
appeared to ‘dissolve’ possibly because the high bio-
component content prevents sufficient crosslinking of
the polymer to provide a rigid matrix.
Fig. 3. Current response of carbon-polymer biocomposite electrodes
with different streptavidin content (SA). Experiments were done in 0.1
M phosphate buffer, 0.1 M KCl, pH 7.4 with 1.81 mmol of mediator
(hydroquinone). Substrate: H2O2, 176 mmol. Applied potential: �/
0.250 V (vs. Ag/AgCl). For each assay, n�/4.
4.3. Stability of streptavidin carbon-polymer
biocomposites
Due to the biological nature of these biocomposite
electrodes, another important parameter that needs to
be considered is their stability over time. As the
streptavidin is not chemically confined within the
composite electrode, there is a possibility of either loss
of streptavidin stability or the leaching of streptavidin
into either the hybridisation solution or the electrolyte
solution. Such phenomenon has been observed else-
where (Alegret et al., 1996) for other carbon-polymer
biocomposite electrodes. To study the stability of these
biocomposites, four electrodes were prepared and ex-
periments were carried on a weekly basis, over a 12-
week period, during which time the electrodes were
stored in an upright position at 4 8C. All other experi-
mental conditions were maintained constant. In all cases
the electrodes were polished prior to use exposing a fresh
surface with streptavidin that is available for binding,
avoiding the problems derived from breaking the
streptavidin/biotin bond by exposing it to high tempera-
tures or extreme conditions which may adversely affect
the stability of the biocomposite (Lo et al., 1999).As the concern here is the availability of streptavidin
for binding, it was decided to investigate the stability
using biotin�/HRP conjugate, where electrochemical
detection was carried out as described previously in
Section 3.3. The problem of non-specific adsorption
needs to be addressed. Therefore, all experiments were
done in parallel using a blank carbon composite
electrode. Fig. 4 shows that the contribution of non-
specific adsorption of the biotin�/HRP conjugate results
in a current response that is relatively consistent with
time and less than 150 nA in all cases. The results
obtained for the streptavidin biocomposite are also
shown in Fig. 4. After subtracting the response asso-
ciated with non-specific adsorption, these results show a
Fig. 4. Stability of the streptavidin carbon-polymer biocomposite
electrode, time (weeks) vs. current (nA). Non-specific adsorption of
biotin�/HRP on a carbon-polymer composite (') and current
response for streptavidin carbon-polymer biocomposite (j). For each
assay, n�/4. Other experimental details as in Fig. 3.
E. Williams et al. / Biosensors and Bioelectronics 19 (2003) 165�/175 171
variation over time ranging from a maximum value of
800 nA for week 1 to a minimum value of 185 nA at
week 11. Fig. 4 shows the mean value of a batch of four
electrodes, with a RSD of 15%. However, it is importantto remember that there is no uniform dispersion of
streptavidin within the carbon composite. Hence, one
electrode may expose more streptavidin than the other
and give rise to the large experimental error observed.
While, taking into account the variation of available
streptavidin binding sites, a general decrease in current
with time is observed for the freshly polished biocom-
posite surface suggesting denaturation of streptavidinwithin the biocomposite over long periods of time.
Over the 12-week period these electrodes were stu-
died, a quantitative result was obtained. However, if the
depth of the electrode was extended to 6 mm instead of 3
mm mentioned in Section 2.3 this would allow a greater
number of surface regeneration polishes, therefore,
allowing a greater number of uses.
4.4. Non-specific adsorption evaluation. Study of the
mechanism of the one step immobilisation/hybridisation
procedure
As with the more traditional enzyme immunoassays
the problem of non-specific adsorption continues to be
an issue in the development of genosensors based on
enzyme labelling. It is also necessary to study the one
step immobilisation/hybridisation mechanism. Toachieve this, a capture system with peculiar DNA
sequences (see Section 2.1) was chosen: the digoxigenin
modified probe (responsible for the analytical response)
matching with the target-DNA, but also with the
biotinylated capture probe.
Four assays were done for both evaluations. As
explained in Section 3.5, the first of these assays (i)
was done with an unmodified carbon composite elec-trode (containing no streptavidin), using all three
oligonucleotide components. Assays (ii), (iii) and (iv)
were carried out with streptavidin carbon-polymer
biocomposite electrodes. In assay (iv) (the genosensor
in capture format), all three oligonucleotides compo-
nents were used. In assays (ii) and (iii) the same
procedure as above was followed, but omitting (ii)
target-DNA, and (iii) biotinylated capture probe. Inall cases (assays i, ii, iii and iv) four electrodes were used,
and the same polishing procedure described in Section
2.3 was followed. Results are shown in Fig. 5. Fig. 5A
show the net amperometric signals while Fig. 5B shows
the signal percentage of assays (i), (ii) and (iii) with
respect to the genosensor signal in capture format (assay
iv).
It is useful to compare the results of assays (i), (iii) and(iv) to evaluate non-specific adsorption. By comparing
the results obtained in assays (i) (unmodified carbon-
polymer composite) and (iv) (the genosensor showing
the electrochemical response associated with the strep-
tavidin carbon-polymer biocomposite), the importance
of streptavidin in the carbon composite electrode is
evident from the resulting enhancement of the electro-chemical response. As seen in Fig. 5B, the contribution
of non-specific adsorption to the overall analytical
response was 8.95%.
The difference between the results obtained in assay
(i) and (iii) could be associated with non-specific
adsorption due to non-specific interactions with strep-
tavidin in the biocomposite (perhaps due to hydropho-
bic forces between streptavidin and anti-Dig-HRP). Asit is shown in Fig. 5B, the contribution of non-specific
adsorption onto streptavidin carbon-polymer biocom-
posite to the overall analytical response was 13.54%.
Finally, results in Fig. 5A (i) and (iii) show that such
non-specific adsorption does indeed occur. However, if
this response is compared with that of genosensor in
capture format (iv) it may be said that the overall
contribution is low with respect to the proposed assayand well within experimental error. For assay (i) this
error was 8.95% and for assay (iii) this error is 9.45%.
The comparison of results from assays (ii) and (iv)
with those of (i) and (iii) is useful to clarify the one step
immobilisation/hybridisation step mechanism.
Assay (ii) uses a streptavidin carbon-polymer bio-
composite, but in this instance the target-DNA is
omitted. This assay will indicate the contribution ofhybridisation between the digoxigenin modified probe
and the biotinylated capture probe, as the sequences are
complementary. As shown in Fig. 5B, the contribution
of the assay (ii) to the overall analytical response
obtained with the genosensor in capture format was
higher (18.45%) than the contribution of non-specific
adsorption (evaluated in assays (i) and (iii)). These
results indicate that hybridisation between the digox-igenin modified probe and the biotinylated capture
probe occurs to a lesser extent or that enzyme labelling
is favoured in assay (iv) with respect to assay (ii). In
assay (iv), the 50 bases of the target would act as a
spacer arm, exposing the digoxigenin to the solution to a
larger extent than in assay (ii), conferring flexibility and
favouring the reaction with the antibody anti-Dig-HRP.
Fig. 6 explains the possible reaction mechanism forassays (ii) and (iv).
In assay (ii), the biotinylated probe has two possibi-
lities (a) immobilisation, i.e. reaction with the biotin�/
streptavidin in the heterogeneous phase or (b) hybridi-
sation with the digoxigenin modified probe in solution.
Since the reaction biotin�/streptavidin is extremely fast
(Gonzalez et al., 1999), (a) is more likely to occur than
(b). As seen in Fig. 6, the hybridisation mechanismexpected between biotin-(dA)20 and digoxigenin-(dT)20
is in a heterogeneous phase (c) and not in solution (b).
Comparing the results shown in Fig. 5B it is evident that
hybridisation occurs in a heterogeneous phase (c)
Fig. 5. Evaluation of the non-specific adsorption. (A) (i) carbon-polymer composite (which contains no streptavidin), (ii) streptavidin carbon-
polymer biocomposite omitting target-DNA, (iii) streptavidin carbon-polymer biocomposite omitting biotinylated capture probe and (iv) genosensor
in capture format based on streptavidin carbon-polymer biocomposite. (B) Signal percentage of assay (i), (ii) and (iii) with respect to the genosensor
in capture format (assay iv). For each assay, n�/4. Other experimental details as in Fig. 3.
E. Williams et al. / Biosensors and Bioelectronics 19 (2003) 165�/175172
between biotin-(dA)20 and digoxigenin-(dT)20, repre-
senting 18.45% of the overall analytical response.
In assay (iv) (as shown in Fig. 6), the biotinylated
probe has three possible courses: immobilisation (a), i.e.
the biotin�/streptavidin reaction in heterogeneous phase
and hybridisation with the digoxigenin modified probe
(b) or with target-DNA (e), both in solution. As seen for
assay (ii), it is expected that the favoured reaction is
immobilisation implied by the biotin�/streptavidin reac-
tion. In this fashion the biotinylated probe would be less
available to hybridise in solution with the DNA-target
and with the digoxigenin-dT(20). Among the possible
hybridisation mechanisms, c, d and f (in Fig. 6, assay iv),
the hybridisation in solution (d) between the DNA
target and the digoxigenin-(dT)20 probe, is more likely
than the mechanism occurring in an heterogeneous
phase (c) (between biotin-(dA)20 and digoxigenin-
(dT)20), and (f) (between biotin-(dA)20 and the target-
DNA).
Fig. 6. Schematic representation of the reaction mechanism for the one step
carbon-polymer biocomposite omitting the target-DNA and for assay (iv) wit
polymer biocomposite. See the text for details.
Finally, hybridisation (g) occurs between the functio-
nalised duplex with digoxigenin and the biotinylated
probe immobilised on the composite.The proposed mechanism for assay (iv) is like a
pseudo two step procedure, occurring in one step. First,
immobilisation happens, followed by hybridisation. It is
in agreement with results shown in Fig. 7A and B for a
two step procedure assay, as explained above. The
amperometric signal produced in these assays was of
the same order confirming the mechanism proposed in
Fig. 6.
4.5. Optimisation of genosensor response
The novelty of this approach is in part attributed to
the simplicity of its design. However, combining the
hybridisation and the immobilisation of DNA in one
analytical step reduces the number of steps within theprotocol and, hence, reduces experimental time. There-
immobilisation/hybridisation procedure, for assay (ii) with streptavidin
h the genosensor in the capture format based on a streptavidin carbon-
Fig. 7. Experimental results for the optimisation of the streptavidin carbon-polymer genosensor, time (min) vs. current (nA). Two step procedures:
(A) variation of the immobilisation time with a constant hybridisation time of 60 min and (B) variation of the hybridisation time with a constant
immobilisation time of 60 min. And (C) variation of time for the one step hybridisation/immobilisation procedure. For each assay, n�/4. Other
experimental details as in Fig. 3.
E. Williams et al. / Biosensors and Bioelectronics 19 (2003) 165�/175 173
fore, as part of the optimisation procedure, it isimportant to ensure that by combining the immobilisa-
tion and hybridisation steps into a one step procedure
does not result in a compromise between sensitivity and
experimental time. For this reason, the following
optimisation experiments were carried out. In all
experiments, four different electrodes were used and
the electrode surface was renewed before each assay.
The first two assays used a two step procedure, while thefinal assay used a one step procedure. In Fig. 7A�/C the
average of the electrochemical response of these four
electrodes is shown and the bars show the standard
deviation of the replicates.
4.5.1. Immobilisation as the limiting step
A variation of the immobilisation time between 10and 120 min at regular intervals of 10 min was chosen.
For subsequent hybridisation, a time of 60 min was
chosen in agreement with literature (Gingeras et al.,
1987). Results of this study are presented in Fig. 7A
showing a clear relationship between immobilisation
time and current response, with a current of 790 nA
being observed at 60 min. However, for immobilisation
times greater than 60 min a current plateau is observed,suggesting a saturation of all available streptavidin
binding sites. Error analysis shows that in all cases
these electrodes show good reproducibility.
4.5.2. Hybridisation as the limiting step
In this instance, a constant immobilisation time of 60
min was used, as seen in Fig. 7A. Again, a variation ofthe hybridisation time between 10 and 120 min at
regular increments of 10 min was used. Results can be
seen in Fig. 7B. In agreement with the previous study it
is seen that experimental time and current are clearly
related. However, unlike the previous experiment that
showed a current plateau at 60 min, this time current
increases proportionally up to 80 min. Additionally, the
current plateau for this experiment is 10% lower thanthe one observed previously.
Results presented in Fig. 7A and B, suggest that the
hybridisation procedure is the limiting step (in agree-
ment with the proposed mechanism shown in Fig. 6).This result is not surprising considering the strong
binding affinity of biotin to streptavidin (Gonzalez et
al., 1999), with a known association constant in the
range of 1015/M. For such a strong binding reaction, one
would not expect the immobilisation of the capture
probe to be the limiting step in the overall procedure.
The experiments described establish the step that
limits the overall electrochemical response. They alsohelp estimate the optimum time for the one step
immobilisation/hybridisation procedure. However, it is
also possible that combining the immobilisation/hybri-
disation in one step will result in a loss of sensitivity.
Therefore a study of the experimental time vs. the
current response was done and results are shown in Fig.
7C. In agreement with Fig. 7A and B it is observed that
current increases with an increase of the immersion timefor times below 60 min. However, for times over 60 min
a current plateau is noticed. These results suggest that
all the available binding sites of the streptavidin
molecules exposed at the electrode surface have been
occupied within this time*/in agreement with results
shown in Fig. 7A.
Fig. 7A�/C shows that separation of hybridisation and
immobilisation in two individual steps does not increasethe amount of HRP available for detection. It should be
noted that optimisation of the enzyme labelling time of
the binding of anti-Dig-HRP to Dig-DNA has been
previously reported (Pividori, 2002) and the time of 40
min was chosen for the present study.
Based on the data presented in Fig. 7 it was decided to
work with the one step hybridisation procedure, as it
does not result in a loss of sensitivity due to a shortertime.
5. Conclusion
The proposed genosensor design has proven to besuccessful in using a simple bulk modification step,
hence, overcoming the complicated pre-treatment steps
associated with other genosensor designs. Also, the use
E. Williams et al. / Biosensors and Bioelectronics 19 (2003) 165�/175174
of a one step immobilisation and hybridisation proce-
dure reduces the cost of the experiment in addition to a
reduction in experimental time. Such a genosensor
design results a novel approach to DNA immobilisationbased on the well known biotin�/streptavidin interac-
tion. Attaching an enzyme-labelled marker to the DNA
duplex formed on the streptavidin carbon-polymer
biocomposite permits the detection of very small quan-
tities of DNA. An additional novel aspect of this work is
the capability of surface regeneration of the biocompo-
site electrodes allowing repeated analyses with the same
electrode without need to denature the dsDNA on theelectrode surface. Stability studies conducted demon-
strate the capability of the same electrode to be used for
a 12-week period.
The new material developed can be seen as a universal
electrochemical platform since it can be used to detect
any analyte that can be coupled to the biotin�/strepta-
vidin reaction as is the case of immunoassays in addition
to its use in DNA analysisFuture work will be focused on the possible applica-
tion of such method to screen printed electrodes,
allowing the development of disposable electrodes cap-
able of a quantitative DNA detection, in a simple, low
cost and easy to manipulate manner.
Acknowledgements
We thank the Ministerio de Educacion y Ciencia de
Madrid (Project BIO2000-0681-C02-01) for financial
assistance and Edna Williams thanks the Departamento
de Educacion y Ciencia and Higher Education Author-
ity for additional financial support.
References
Alegret, S., 1996. Rigid carbon-polymer biocomposites for electro-
chemical sensing. Rev. Analyst 121, 1751�/1758.
Alegret, S., Alonso, J., Bartrolı, J., Cespedes, F., Martınez-Fabregas,
E., del Valle, M., 1996. Amperometric biosensors based on bulk-
modified epoxy graphite biocomposites. Sensor Mater. 8 (3), 147�/
153.
Armistead, P.M., Thorp, H.H., 2000. Modification of indium tin oxide
electrodes with nucleic acids: detection of attomole quantities of
immobilised DNA by electrocatalysis. Anal. Chem. 72, 3764�/3770.
Armistead, P.H., Thorp, H.H., 2001. Oxidation kinetics of guanine in
DNA molecules adsorbed onto indium tin oxide electrodes. Anal.
Chem. 73, 558�/564.
Alfonta, L., Singh, A.K., Wollner, I., 2001. Liposomes labelled with
biotin and horseradish peroxidase: a probe for enhanced amplifica-
tion of antigen�/antibody or oligonucleotide*/DNA sensing pro-
cesses by precipitation of an insoluble product on electrodes. Anal.
Chem. 73, 91�/102.
Burrows, C.J., Muller, J.G., 1998. Metallodrug complexes that
mediate DNA and lipid damage via sulfide autoxidation: cop-
per(II) famotidine and iron(III) bis(salicylglycine). Inorg. Chim.
Acta 275�/276, 314�/319.
Chen, X., Ruan, C., Kong, J., Deng, J., 2000. Characterisation of
direct electron transfer and bioelectrocatalysis of horseradish
peroxidase in DNA films at pyrolytic graphite electrodes. Anal.
Chem. Acta 412, 89�/98.
Chiti, G., Marrazza, G., Mascini, M., 2001. Electrochemical DNA
biosensor for enviromental monitoring. Anal. Chim. Acta 427,
155�/164.
Cespedes, F., Martınez-Fabregas, E., Alegret, S., 1996. New materials
for electrochemical sensing. I. Rigid conducting composites. TrAC
15, 296�/303.
Erkkila, K.E., Odom, D.T., Barton, J.K., 1999. Recognition and
reaction of metallointercallators with DNA. Chem. Rev. 99, 2777�/
2795.
Gingeras, T.R., Kwoh, D.Y., Davis, G.R., 1987. Hybridization
properties of immobilized nucleic acids. Nucleic Acids Res. 15,
5373�/5390.
Gonzalez, M., Argarana, C.E., Fidelio, G.D., 1999. Extremely high
thermal stability of streptavidin and avidin upon biotin binding.
Biomol. Eng. 19, 67�/72.
Hashimoto, K., Ito, K., Ishimori, Y., 1994. Novel DNA sensors for
electrochemical gene detection. Anal. Chim. Acta 286, 219�/224.
Hashimoto, K., Ito, K., Ishimori, Y., 1998. Sequence-specific gene
detection with a gold electrode modified with DNA probes and an
electrochemically active dye. Anal. Chem. 66, 3830�/3833.
Herne, T.M., Tarlov, M.J., 1997. Characterization of DNA probes
immobilized onto gold surfaces. J. Am. Chem. Soc. 119, 8916�/
8920.
Kizek, R., Ludek, H., Frojta, M., Palecek, E., 2002. Determination of
nanogramquantities of osmium labelled single stranded DNA by
differential pulse stripping voltammetry. Bioelectrochemistry 55,
119�/121.
Kricka, L.J., 1992. Non-isotopic DNA Probe Techniques. Academic
Press, Inc, San Diego.
Lee, T.-Y., Shim, Y.-B., 2001. Direct DNA hybridisation detection
based on the oligonucleotide*/funtionalised conductive polymer.
Anal. Chem. 73, 5629�/5632.
Li, L., Wang, M., Dong, S., Wang, E., 1997. Electrochemical
investigation of DNA adsorbed on conducting polymer modified
electrodes. Anal. Sci. 13, 305�/310.
Lo, Y.-S., Huefner, N.D., Chan, W.S., Stevens, F., Harris, J.H.M.,
Beebe, T.P., 1999. Specific interactions between biotin and avidin
studied by atomic force spectroscopy using the Poisson statistical
analysis method. Langmuir 15, 1373�/1982.
Lumley-Woodyear, T., Campbell, C.N., Freeman, E., Freeman, A.,
Georgiou, G., Heller, A., 1999. Rapid amperometric verification of
PCR amplification of DNA. Anal. Chem. 71, 535�/538.
Marrazza, G., Chianella, I., Mascini, M., 1999. Disposable DNA
electrochemical biosensors for environmental monitoring. Anal.
Chim. Acta 387, 297�/307.
Millan, K.M., Mikkelsen, S.R., 1993. Sequence-selective biosensor for
DNA based on an electroactive hybridisation indicators. Anal.
Chem. 65, 2317�/2323.
Millan, K.M., Spurmanis, A.L., Mikkelsen, A.L., 1992. Covalent
immobilisation of DNA onto glassy carbon electrodes. Electro-
analysis 4, 929�/932.
Nuzzo, R.G., Allara, D.L., 1983. Adsorption of bifunctional organic
disulphides on gold surfaces. J. Am. Chem. Soc. 105, 4481�/4483.
Palecek, E., 1958. Oscillographic polarograhpy of nucleic acids and
their building blocks. Naturwiss 45, 186�/187.
Palecek, E., 1960. Oscillographic polarography of highly polymerized
deoxyribonucleic acid. Nature 188, 656�/657.
Palecek, E., Fojta, M., 2001. DNA hybridization and damage. Anal.
Chem., 75A�/83A.
Palecek, E., Kizek, R., Havran, L., Billova, S., Fojta, M., 2002.
Electrochemical enzyme linked immunoassay in a DNA hybridisa-
tion sensor. Anal. Chim. Acta 469, 73�/83.
E. Williams et al. / Biosensors and Bioelectronics 19 (2003) 165�/175 175
Pantano, P., Morton, T.H., Kuhr, W.G., 1991. Enzyme-modified
carbon-fiber microelectrodes with millisecond response times. J.
Am. Chem. Soc. 113, 1832�/1833.
Pividori, M.I., 2002. Ph.D. Thesis. Autonomous University of
Barcelona, Spain.
Pividori, M.I., Merkoci, A., Alegret, S., 2000. Electrochemical
genosensor design: immobilisation of oligonucleotides onto trans-
ducer surfaces and detection methods. Biosen. Bioelectron. 15,
291�/303.
Pividori, M.I., Merkoci, A., Alegret, S., 2001a. Classical dot blot
format implemented as an amperometric hybridisation genosensor.
Biosen. Bioelectron. 16, 1133�/1142.
Pividori, M.I., Merkoci, A., Alegret, S., 2001b. Dot-blot amperometric
genosensor for detecting a novel determinant of b-lactamase
resistance in Staphylococcus aureus . Analyst 126, 1551�/1557.
Radi, A., 1999. Volumetric study of Khellin at DNA-coated carbon
paste electrodes. Anal. Chim. Acta 386, 63�/68.
Rajksi, S.R., Jackson, B.A., Barton, J.K., 2000. DNA repair models
for damage and mismatch recognition. Mutat. Res. 447, 49�/72.
Shamansky, L.M., Brandon-Davis, C., Stuart, J.K., Kuhr, W.G.,
2001. Immobilisation and detection of DNA on microfludic chips.
Talanta 55, 909�/918.
Santandreu, M., Cespedes, F., Alegret, S., Martınez-Fabregas, E.,
1997. Amperometric immunosensors based on rigid conducting
immunocomposites. Anal. Chem. 69, 2080�/2085.
Takenaka, S., Uto, Y., Kondo, H., Ihara, T., Takagi, M., 1994.
Electrochemically active DNA probes. Detection of target DNA
sequences at femtomole level by High Performance Liquid
Chromatography with electrochemical detection. Anal. Biochem.
218, 436�/443.
Turner, S.J., Lewis, G.D., Bellamy, A.R., 1997. Detection of sewage-
derived Escherichia coli in a rural stream using multiplex PCR and
automated DNA detection. Water Sci. Technol. 35, 337�/342.
Uto, Y., Kondo, H., Abe, M., Suzuki, T., Takenaka, S., 1997.
Electrochemical analysis of DNA amplified by the polymerase
chain reaction with a ferrocenylated oligonucleotide. Anal. Bio-
chem. 250, 122�/124.
Wang, J., Cai, X., Jonsson, C., 1996. Adsorptive stripping potentio-
metry of DNA at electrochemically pre-treated carbon paste
electrodes. Electroanalysis 8, 20�/24.
Wang, J., Cai, X., Fernandes, J.R., Grant, D.H., Ozsoz, M., 1997a.
Electrochemical measurements of oligonucleotides in the presence
of chromosomal DNA using membrane-covered carbon electrodes.
Anal. Chem. 69, 4056�/4059.
Wang, J., Cai, X., Rivas, G., Shiraishi, H., Dontha, N., 1997b. Nucleic
acid immobilisation recognition and detection at chronoampero-
metric DNA chips. Biosen. Bioelectron. 12 (7), 587�/599.
Wang, J., Fernandes, J.R., Kubota, L.T., 1998. Polishable and
renewable DNA hybridisation biosensors. Anal. Chem. 70,
3699�/3702.
Wang, J., Kawde, A.-N., 2001. Pencil based renewable biosensors for
label free electrochemical detection of DNA hybridisation. Anal.
Chim. Acta 431, 219�/224.
Xu, C., Cai, H., He, P., 2001. Electrochemical detection of sequence
specific DNA using a DNA probe labelled with aminoferrocene
and chitosan modified electrode immobilised with ssDNA. Analyst
126, 62�/65.