Product Information and User InstructionsModels CH5000S and CH5000
Caution – This device is only to be used by or on the prescription of a physician.
.......................................................................................................................DRIVE PURE OXYGEN CONCENTRATOR MODELS CH5000S and CH5000.......................................................................................................................2
Table of Contents
Introduction to Your Pure Oxygen Concentrator Page 2
How Your Pure Oxygen Concentrator Operates Page 2
Descriptions of Parts and Controls Page 3
Warnings and Cautions Page 4
Setup and Operation Page 5
Cleaning and Maintenance Page 6
Models, Parts, and Accessories Page 6
Alarms and Troubleshooting Page 7-8
Specifications Page 9
Separation Distances & Symbol Key Page 10
Warranty Page 11
Intended UseThe intended use of the Pure Oxygen Concentrator is to provide supplemental oxygen to
those individuals needing supplemental oxygen therapy. The device is not intended to be life
supporting or life sustaining.
Introduction to Your Pure Oxygen ConcentratorThe Pure Oxygen Concentrator is designed as a stationary supplemental oxygen source. A
physician has prescribed supplemental oxygen because you are not getting enough oxygen
from room air alone. The specific liter flow setting to meet your needs, as prescribed by your
physician, will increase the amount of oxygen your body receives. DO NOT change the flow
settings unless your health care professional instructs you to do so.
How Your Pure Oxygen Concentrator Operates The Pure is an electrically operated unit which produces concentrated oxygen from room air
providing supplemental oxygen flow of 0.5 to 5 liters per minute (LPM) through an oxygen
outlet and cannula. Your equipment provider will instruct you on proper and safe operation of
this unit. Please read and understand this entire manual before operating your concentrator.
.......................................................................................................................DRIVE PURE OXYGEN CONCENTRATOR MODELS CH5000S and CH5000.......................................................................................................................
.......................................................................................................................DRIVE PURE OXYGEN CONCENTRATOR MODELS CH5000S and CH5000....................................................................................................................... 3
Front View [Figure A]
1) LED Operation Indicator Lights
Oxygen Sensing CH5000S Model [Inset 1]
• Green Operation Normal, Power On Oxygen Purity Levels at 82% - 96%
• Yellow Operation Warning, Read Manual Oxygen Purity Levels at 72% - 81%
• Red Operation Alarm, Service Required Oxygen Purity Levels below 72% or
other system fault is present.
Non Oxygen Sensing CH5000 Model [Inset 2]
• Green Operation Normal, Power On
• Red Operation Alarm, Service Required
NOTE: For both the CH5000S and the CH5000 models, the Green LED will blink for 6-10
minutes during start up then become solid Green during normal operation.
2) Power Switch: I/0
• I = On
• 0 = Off
3) Flow Meter Knob
• Adjusts Liter Per Minute (LPM) output
4) Flow Meter
• Displays the Liter Per Minute (LPM) output
5) Oxygen Outlet Barb
• Cannula connection to dispense oxygen from concentrator
6) Hour Meter
• Measures the total operating hours of the concentrator
Rear View [Figure B]
7) Ventilation Slots
8) Humidifier bottle holder
9) Connection Adapter for humidifier bottle
10) Humidifier Bottle (not included)
11) Filter Door
NORMALOXYGEN
ALARMSERVICE REQUIRED
ONLY
IP21
Pure Oxygen Concentrator
CH5000
120V ~ 60Hz 3.0A
2Max O : 93% +/- 3%
Inovo, Inc401 Leonard Blvd N.Lehigh Acres, FL 33971888-446-6862
AL-00353/13/A
lement
I/O
Do Not Disassemble. Refer to Authorized Service Personnel.
Use No Oil Or Grease
MADE IN USA
REF.
SN.
Descriptions of Parts and ControlsPlease be sure the following items are included:
• User Manual
• Oxygen Concentrator
• Connection Adapter for humidifier bottle
Figure A Figure B
ONLY
IP21
Pure Oxygen Concentrator
CH5000S
120V ~ 60Hz 3.0A
2Max O : 93% +/- 3%
Inovo, Inc401 Leonard Blvd N.Lehigh Acres, FL 33971888-446-6862
AL-00352/13/A
lement
I/O
Do Not Disassemble. Refer to Authorized Service Personnel.
Use No Oil Or Grease
MADE IN USA
NORMALOXYGEN
WARNINGLOW OXYGEN
ALARMSERVICE REQUIRED
!
REF.
SN.
7
8
11
14
15 12
13
12) Foam Intake Filter
13) HEPA Intake Filter
14) Power Cord
15) Transfill Connection Port
9
15
10
Insert 1
Insert 21
2
3
45
6
.......................................................................................................................DRIVE PURE OXYGEN CONCENTRATOR MODELS CH5000S and CH5000.......................................................................................................................4
Warnings and Cautions • Under certain circumstances, oxygen therapy can be hazardous. Seeking medical advice
before using an oxygen concentrator is advisable.
• Proper operation of this device requires unobstructed ventilation of the rear filters and
ventilation slots. Keep the device a minimum of 6 inches (15 cm) from walls, furniture,
curtains and/or any other object which may impede airflow and ventilation slots.
• The oxygen concentrator should be located so as to avoid pollutants and fumes.
• Do not place the concentrator in a small closed space such as a closet.
• Oxygen accelerates combustion. Do not operate the concentrator within 5 feet
(1.5 meters) of hot, sparking objects, a heat source, or an open source flame. Not suitable
for use in the presence of a flammable anesthetic mixture with air, or with oxygen or
nitrous oxide.
• Do not smoke or permit others to smoke in the vicinity of the concentrator.
• Do not disassemble. Contact Drive Medical or an authorized Drive Medical provider for
assistance.
• If you experience discomfort during use of the concentrator, or if an equipment alarm is
triggered, contact your equipment provider or healthcare professional immediately.
• This concentrator generates supplemental oxygen for use on order of a physician or
healthcare provider. Oxygen supplied by this device should not be considered life
supporting or life sustaining.
• For those individuals who a healthcare professional has determined interruption in
oxygen therapy may have serious consequences, an alternate source of oxygen should
be available for immediate use.
• Do not use oil or grease on the concentrator or its components. When combined with
oxygen, these substances may greatly increase the potential of fire hazard or
personal injury.
• Do not use the concentrator if the power cord is damaged. Do not use extension cords
or electrical adapters with the concentrator.
• Do not clean the concentrator while it is plugged into an electrical outlet.
• Your equipment provider is responsible for appropriate preventive maintenance at
intervals recommended by Drive Medical.
• Do not place liquids on the concentrator.
• If liquid is spilled on the device, turn the power off and unplug the unit from the electrical
outlet prior to cleaning. Completely dry the unit prior to plugging in the power cord and
turning the power back on. If the device does not operate properly, contact your
equipment provider for assistance.
• Device operation outside of the voltage, LPM, temperature, humidity, and/or altitude
specifications (page 9) may result in decreased unit performance and oxygen purity
levels.
NOTE: The use of certain humidifiers and administration of accessories not specified for use
with this oxygen concentrator may impair its performance.
.......................................................................................................................DRIVE PURE OXYGEN CONCENTRATOR MODELS CH5000S and CH5000.......................................................................................................................
.......................................................................................................................DRIVE PURE OXYGEN CONCENTRATOR MODELS CH5000S and CH5000....................................................................................................................... 5
Setup and Operation 1) Position the concentrator in a location which will permit unrestricted air intake, be near
an electrical outlet, and be in a room in which you spend most of your time. Make sure
the device is at least 6 inches (15 cm) from walls, furniture and/or curtains, is not in a
closed confined space such as a closet, and is at least 5 feet (1.5 meters) from hot
sparking objects, a heat source, or an open flame.
2) Verify the air filter, located on the back of your unit, is clean and properly installed [See
Cleaning and Maintenance on Page 6].
3) After reading and understanding this manual, plug the power cord directly into an
electrical outlet.
Oxygen Cannula Connection without Humidification Bottle
4a) Connect the oxygen cannula to the Oxygen Outlet Barb
[Figure C]. Skip to Step 9.
Oxygen Cannula Connection with Humidification Bottle
4b) Connect the Connection Adapter to the Oxygen Outlet Barb
[Figure D]
5) Fill the humidifier bottle with distilled water.
6) Using the strap, firmly secure the humidifier bottle to the
concentrator [Figure E].
7) Run the Connection Adapter under the concentrator handle
and connect to the humidifier bottle according to the bottle
manufacturer’s instructions [Figure E].
8) Connect the cannula to the humidifier bottle per the bottle
manufacturer’s instructions.
9) Press the Power Switch to the ON position [Figure F]. During
start up, the Green LED Operations Indicator Light will flash.
Full oxygen purity will be reached within 6-10 minutes.
10) Check the Flow Meter to ensure, that at eye level, the center
of the ball is in line with the prescribed LPM flow rate
[Figure G].
a. To increase the flow rate, turn the Flow Meter Knob
counter-clockwise.
b. To decrease the flow rate, turn the
Flow Meter Knob clockwise.
11) Check that oxygen is flowing through the
cannula.
a. Refer to the Trouble Shooting Guide on
page 8 if there is no oxygen flow.
12) Put on the cannula as directed by your
equipment provider.
13) When not in use, turn off the concentrator.
Figure C
Figure D
Figure E
Figure F Figure G
.......................................................................................................................DRIVE PURE OXYGEN CONCENTRATOR MODELS CH5000S and CH5000.......................................................................................................................6
Cleaning and Maintenance Unplug the concentrator before cleaning. Do not use oil or grease.
Concentrator
Periodically, with a damp cloth, wipe the outside of the concentrator using a solution of hot
water and dishwashing detergent. If any medical cleaning solutions are used, be sure to
follow the manufacturer’s directions.
Filters
The Foam Intake filter is a washable filter. Rinse and wash as needed.
1) Remove the filter cover.
2) Remove both the Foam Intake filter and the HEPA filter.
3) Separate the Foam Intake filter from the HEPA filter.
4) Wash in a solution of warm water and dishwashing detergent.
5) Thoroughly rinse and dry.
6) Replace Foam Intake filter, HEPA filter, and Filter Cover.
The HEPA Intake Filter is a replaceable filter. Replacement is
recommended approximately every 12 months, or as needed.
Please contact your equipment provider for a replacement.
Cannula, Tubing, and Humidifier Bottles
Clean and replace as recommended and directed by the
manufacturer or your equipment provider.
Model, Parts and Accessories
Model Description
CH5000S Pure Oxygen Concentrator, with Oxygen Sensor
CH5000 Pure Oxygen Concentrator, without Oxygen Sensor
Parts Description
PL-2001844 Hooded Swivel Caster 1/each
PL-2001853 HEPA Intake Filter 1/each
PL-2001854 Reusable Foam Intake Filter 1/each
PL-3000706 Filter Intake Door 1/each
The following humidifier bottle is recommended for use with the Pure Oxygen Concentrator.
Accessories Description
HUM 001 Humidifier Bottle 50/case
HUM-TUB Humidifier Adapter Tubing 50/case
CON 550 Ridged Oxygen Tubing Connector 50/case
SOFT 225 Non-kinking Cozy Nasal Cannula 25' 25/case
SOFT 250 Non-kinking Cozy Nasal Cannula 50' 20/case
TUB NK 25 Non-kinking Oxygen Tubing 25' 25/case
TUB NK 50 Non-kinking Oxygen Tubing 50' 20/case
Filter Door
HEPA Foam Intake
.......................................................................................................................DRIVE PURE OXYGEN CONCENTRATOR MODELS CH5000S and CH5000.......................................................................................................................
.......................................................................................................................DRIVE PURE OXYGEN CONCENTRATOR MODELS CH5000S and CH5000....................................................................................................................... 7
Alarms and Troubleshooting
Light Indication Sounds Cause
Blinking Green None Normal Operation - Unit in Start-up Phase
Solid Green None Normal Operation
Solid Yellow None Oxygen Purity Between 72% and 81%. Device can still be used but service provider should be contacted.
Solid Red Solid Audible Tone
Oxygen Purity Below 71% and/or Low/No Flow (0 to 0.3 LPM). Service Required - Call Provider.
Solid Red with Solid Green
Solid Audible Tone
Oxygen Purity Above 82% and Low/No Flow (0 to 0.3 LPM). See section B of the troubleshooting section of this manual.
Solid Red with Solid Yellow
Solid Audible Tone
Oxygen Purity Between 72% and 81% and Low/No Flow (< 0.5 LPM). See section B of the trouble-shooting section of this manual.
None Intermittent Beep
The unit has lost the power connection and the switch is in the on position. See section A of the troubleshooting section of this manual.
Intermittent Beep
No power to Unit. See section A of the trouble-shooting section of this manual.
Light Indication Sounds Cause
Blinking Green None Normal Operation - Unit in Start-up Phase
Solid Green None Normal Operation
Solid Red Solid Audible Tone
Low/No Flow (0 to 0.3 LPM). Ensure tubing is not blocked. If tubing is not blocked and alarm contin-ues, contact service provider.
Solid Red with Solid Green
Solid Audible Tone
Low/No Flow (0 to 0.3 LPM). See section B of the troubleshooting section of this manual.
None Intermittent Beep
The unit has lost the power connection and the switch is in the on position. See section A of the troubleshooting section of this manual.
Intermittent Beep
No power to Unit. See section A of the trouble-shooting section of this manual.
Alarm Table – Non-Oxygen Sensor Model Number CH5000
Alarm Table – Oxygen Sensor Model Number CH5000S
.......................................................................................................................DRIVE PURE OXYGEN CONCENTRATOR MODELS CH5000S and CH5000.......................................................................................................................8
Symptom Probable Cause Potential Action
A. Unit does not operate. No lights are seen on the front panel when the power switch is “On”. An audible alarm is heard (intermittent beep).
1. Power cord not properly inserted into wall outlet.
Check power connection at the wall outlet.
2. Issue with power supplied to wall outlet.
Check your home circuit breaker and reset if necessary. Try using a different wall outlet if the situation occurs again.
B. Unit is running and either the red and green light OR red and yellow light is lit when the pow-er switch is in the “On” position. An audible alarm is heard.
1. Air filter is blocked.
Check the air filters. If dirty, clean/replace the filter(s) per the instructions in this manual (see "Cleaning and Maintenance" section of this manual). If unit ceases to alarm after a minimum of two minutes and proper flow appears to have returned, this was the issue.
2. Vents in back are blocked.
Check the rear cooling vents are not blocked or restricted. If unit ceases to alarm after a minimum of two minutes and proper flow appears to have returned, this was the issue.
3. Blocked tubing
Detach cannula, catheter, or face mask from oxygen outlet. If unit ceases to alarm after a minimum of two minutes and proper flow appears to have returned, this was the issue. Check for obstructions or kinks in external cannula, catheter, face mask, or tubing. Clean or replace, if necessary.
4. Blocked / defective humid-ifier bottle.
Detach the humidifier from the oxygen outlet. If unit ceases to alarm after a minimum of two minutes and proper flow appears to have returned, this was the issue.
5. Flow meter set too low.
Turn flow meter dial counter-clockwise to a minimum of 0.5 LPM or greater. If unit ceas-es to alarm after a minimum of two minutes and proper flow appears to have returned, this was the issue.
C. If any other problems occur with your oxygen concentrator or none of the solutions above work
Turn your unit “OFF”, switch to your reserve / alternate oxygen system, and contact your service provider immediately.
Basic Troubleshooting
.......................................................................................................................DRIVE PURE OXYGEN CONCENTRATOR MODELS CH5000S and CH5000.......................................................................................................................
.......................................................................................................................DRIVE PURE OXYGEN CONCENTRATOR MODELS CH5000S and CH5000....................................................................................................................... 9
Specifications
Electrical Requirements 120 VAC +/- 10%, 60 HZ
Rated Current 3A
Power 360 W @ 3 LPM
Typical Sound Level 45 dBA
Altitude 13,123 ft (4000 m) above sea level - concentration level at 93%
Operating Temperature 41°F - 104°F (5°C - 40°C)
Operating Humidity 20 - 60% RH, Non-condensing
Storage Temperature -13°F to 158°F (-25°C to 70°C) up to 93% RH
Oxygen Output Purity 87% to 96% for flows ranging 0.5 to 5 LPM
Outlet Pressure 5.5 PSI +/- 0.5 PSI
Flow Range 0.5 to 5 LPM
Low Flow Alarm < 0.5 LPM
Pressure Relief 35 PSI +/- 5 PSI (247kPa +/- 35 kPa)
Change in Maximum Recommended Flow when Back Pressure of 7kPa is Applied
0.5 LPM
Filters Compressor Inlet, Foam Intake, HEPA Intake
Safety Current Overload
High Pressure Relief
Power Loss
Dimensions 16.75" W x 10.25" D x 24.5" H
Weight 35lbs
Cabinet UL94V-O Flame Retardant Rated Thermoplastic
Electrical Double Insulated Cord - Do Not Use an Extension Cord
Duty Cycle Continuous Use
Warranty Limited 3 Year
Low Flow Audible Alarm
High Temperature Compressor
Oxygen Purity (OCI Model only)
.......................................................................................................................DRIVE PURE OXYGEN CONCENTRATOR MODELS CH5000S and CH5000.......................................................................................................................10
Symbol Key Description
Warning, Consult Accompanying Documents
Follow Instructions for Use
Manufacturer
Degree of Protection Against Electrical Shock: Type BF Unit
Class II Equipment
Do Not Dispose of in Household Waste
No Smoking
No Oil or Grease
Prescription Only
Do Not Disassemble
Serial Number
Drip Proof Equipment
SN
ONLY
Separation Distances
Recommended separation distances between portable and mobile RF Communications equipment and the Pure Oxygen Concentrator
The Pure Oxygen Concentrator is intended for use in the electromagnetic environment in
which radiated RF disturbances are controlled. The customer or user of the Pure Oxygen
Concentrator can help prevent electromagnetic interference by maintaining a minimum
distance between portable and mobile RF communications (transmitters) and the Pure
Oxygen Concentrator as recommended below. According to the maximum output power of the communication equipment.
Separation distance according to transmission frequency in meters
Max Output Power
150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2.5 GHz
(Watts)
D = (3.5/V1)(PP) D=(3.5/V
1)(PP) D=(7/V
1)(PP)
0.01 0.12 0.12 0.23
0.1 0.37 0.37 0.74
1 1.17 1.17 2.33
10 3.69 3.69 7.38
100 11.67 11.67 23.33
This equipment has been tested and is in compliance with the EMC limits specified in IEC/EN 60601-1-2. These limits are designed to provide a reasonable protection against harmful interference in a residential installation. If you do experience interference from other equip-ment you may try to correct or reduce the interference by one of the following measures: · Reposition, relocate, or increase the separation distance between the equipment · Connect the equipment into an outlet on a circuit that is different from that to which the other device is connected
IP21
.......................................................................................................................DRIVE PURE OXYGEN CONCENTRATOR MODELS CH5000S and CH5000.......................................................................................................................
.......................................................................................................................DRIVE PURE OXYGEN CONCENTRATOR MODELS CH5000S and CH5000....................................................................................................................... 11
Warranty
The Pure Oxygen Concentrator has been carefully manufactured and inspected in our
Lehigh Acres, FL USA facility and is warranted to be free from defects in workmanship
and materials. Under this warranty, Drive Medical and its subsidiaries obligation shall be
limited to the replacement or repair of any such units or parts that prove, by Drive Medical’s
inspection to be defective within three years from original date of purchase. This warranty is
not transferable or assignable to any subsequent purchaser or owner. Any abuse, operation
other than the intended use of the product as outlined in the manual, negligence, accident
or repair by someone other than a Drive Medical Authorized Service Professional, shall
immediately void this warranty.
Drive Medical will not accept damages or charges for labor, parts, or expenses incurred in
making field repairs, except upon written authorization prior to such action.
The foregoing warranty is exclusive and in lieu of all other expressed warranties. Implied
warranties, if any, including but not limited to the implied warranties of merchantability and
fitness for a particular purpose, shall not extend beyond the duration of the express warranty
provided herein. In no event shall Drive Medical or its subsidiaries be liable for loss of use or
profit or other collateral, special or consequential damages.
Drive Medical
99 Seaview Blvd
Port Washington, NY 11050
Toll-free 877-224-0946
www.drivemedical.com
Inovo, Inc, a Drive Medical Company
401 Leonard Blvd N.
Lehigh Acres, FL 33971
Toll Free 877-224-0946
Printed in the USA
PM-00076/15/C
Drive Medical99 Seaview Blvd
Port Washington, NY 11050
Toll-free 877-224-0946
www.drivemedical.com