Korea-Australia Rheology Journal June 2009 Vol. 21, No. 2 119
Korea-Australia Rheology JournalVol. 21, No. 2, June 2009 pp. 119-126
Numerical investigation on the blood flow characteristics
considering the axial rotation in stenosed artery
Kun Hyuk Sung, Kyoung Chul Ro and Hong Sun Ryou*
School of Mechanical Engineering, Chung-Ang University, 221 HeukSuk-Dong, Dongjak-Gu,Seoul 156-756, Korea
(Received March 13, 2009; final revision received May 4, 2009)
Abstract
A numerical analysis is performed to investigate the effect of rotation on the blood flow characteristics withfour different angular velocities. The artery has a cylindrical shape with 50% stenosis rate symmetricallydistributed at the middle. Blood flow is considered a non-Newtonian fluid. Using the Carreau model, weapply the pulsatile velocity profile at the inlet boundary. The period of the heart beat is one second. In com-parison with no-rotation case, the flow recirculation zone (FRZ) contracts and its duration is reduced in axi-ally rotating artery. Also wall shear stress is larger after the FRZ disappears. Although the geometry ofartery is axisymmetry, the spiral wave and asymmetric flow occur clearly at the small rotation rate. It iscaused that the flow is influenced by the effects of the rotation and the stenosis at same time.
keywords : blood flow, axially rotating velocity, stenosis, pulsatile flow, non-Newtoninan
1. Introduction
Arteriosclerosis is one of the most widespread diseases in
human beings. It is a significant factor in the death rate
because it affects hemodynamics and reduces the flow rate
of blood to the heart and the brain. In particular, sudden
body movement causes the blood pressure fluctuations of a
person with an arterial diseases and an ischemic poverty of
blood to increase more than those of a healthy person. In a
serious case, when the blood flow rate decreases, the per-
son may fall because of vertigo and experience temporary
eyesight and hearing trouble. Thus, human body move-
ments can affect the blood flow characteristics of persons
with arterial diseases more than those of healthy persons.
In research on human acceleration, Burton et al. (1974)
studied eyesight trouble caused by a change of blood flow
in extreme gravity circumstances. Hooks et al. (1972) con-
ducted a clinical study of the side effects of body accel-
eration. In research on blood flow and acceleration, Misra
and Sahu (1988) developed a mathematical model to study
the blood flow through large arteries under the action of
periodic body acceleration. Belardinelli et al. (1989) per-
formed an experimental study of the effect of blood pressure
by shock acceleration, and Mandal et al. (2007) performed
a numerical study on the blood characteristics of a cylin-
drical blood vessel with periodic accelerations. Nakamura et
al. (1988) and Luo et al. (1992) carried out an investigation
of the blood flow characteristics of stenosed and bifurcated
blood vessels. Ro et al. (2008) performed a numerical study
on the blood characteristic of the carotid bifurcation artery
with periodic accelerations. However, none of these resear-
ches focused on the effect of body acceleration on the char-
acteristics of blood flow without rotation.
From the medical side, studies on the effects of cervical
or spinal rotation on hemodynamic in arteries have been
executed, but they concentrated on the velocity distribution
caused by the change in artery volume. There have been
only a few studies on how the rotational movement of the
human body affects blood flow characteristics.
In fluid dynamics, the flow characteristics of an axially
rotating pipe without stenosis have been studied. For
example, Imao et al. (1992) showed the flow instability
problem in axially rotating pipes at the critical ratio of the
circumferential velocity to the mean axial velocity.
Kikuyama et al. (1983) showed that the transition of flow
state from the laminar to the turbulent in axially rotating
pipe can occur at a low Reynolds number. Both of these
experiments showed that rotation caused destabilization of
the flow in an axially rotating pipe. However, these exper-
iments used water, so they did not address the characteristics
of blood, which has non-Newtonian viscosity. Generally, the
assumption of Newtonian behavior of blood is acceptable
for high shear-rate flow, but it is not valid when the shear
rate is low (0.1 s-1), as it is in small arteries or on the down-
stream side of stenosis (Chien et al., 1982). It has also been
pointed out that, in some diseased conditions e.g. patients
with severe myocardial infarction, cerebrovascular diseases*Corresponding author: [email protected]© 2009 by The Korean Society of Rheology
Kun Hyuk Sung, Kyoung Chul Ro and Hong Sun Ryou
120 Korea-Australia Rheology Journal
and hypertension blood exhibits remarkable non-Newtonian
behavior. Thus, non-Newtonian viscosity must be consid-
ered when analyzing the characteristics of flow in stenosed
arteries.
Young (1979) showed that head loss is a nonlinear func-
tion of stenosis and that pressure losses become significant
only for stenoses greater than 50~70%. Thus, the stenosis
affects significantly the pressure distribution in the artery
and puts the person with an arterial disease in jeopardy.
The experiment showed the effect of inlet velocity on the
flow characteristics at the downstream of stenosis in the
artery (Deplano et al., 1999).
Extremely, the angular velocity is about 8 revolutions a
second when the figure skater performs a standing spin.
Although that case is rare to the common people, they
undergo rotations of body through daily exercise. So rotation
of body can affect the characteristic of blood flow such as
pressure drop, wall shear stress, and flow recirculation zone.
However, in spite of the importance of rotation effect,
there is no research on the effect of rotation on the flow
characteristics considered pulsatile velocity profile in
stenosed arteries.
Hence, for the basic study on the rotation effect to the
blood flow, we select the common carotid artery because
the artery rotates axially when people do a standing spin.
Therefore the purpose of this paper is a numerical analysis
of the effect of rotation and an unsteady pulsatile velocity
profile in a stenosed artery.
2. Numerical details
2.1. Governing equationsIn order to simulate the blood flow characteristics, mass
and momentum conservation equations are required and the
non-Newtonian viscosity and pulsatile flow must be con-
sidered. In addition, the axially rotating, centrifugal force
must be added to the momentum equation as a source term.
(1)
(2)
To simulate a non-Newtonian fluid problem, a consti-
tution equation is required for blood rheology character-
istics described by the second invariant of shear rate tensor:
(3)
where η and are apparent viscosity and shear rate.
Shear rate is represented as:
(4)
We use the Carreau viscosity model because it is more
suitable for representing blood rheology characteristics
(Cho, 1985):
(5)
where η0 is the zero shear viscosity (0.056 Pa·s), is
the infinite shear viscosity (0.00345 Pa·s), λ is the time
constant (3.313 s) and n is the power law index (0.356).
2.2. Modeling of an artery with stenosisFig. 1 shows a schematic view and grid generation of a
stenosed blood vessel. The stenosis, where is located
between 1D upstream and 1D downstream from the center
of stenosis, is modeled by Young’s model (Young, 1968),
as shown in equation (6). The diameter of the blood vessel
is 8 mm, and the minimum diameter of the stenosis is half
the size of the blood vessel. The stenosis rate is 50% with
no eccentricity.
(6)
where, a is the stenosis rate, z0 is half the length of the
stenosis ,1D , and z1 is the axial position from the starting
point of the stenosis where is 1D upstream from the center
of stenosis. The grid independent test is performed with
four different number of grid cells which are 43,587,
130,720, 434,808 and 580,320 when the angular velocity is
6 rev/s. In Fig. 2(a) and (b), the averaged wall shear stress
(WSS) with 434,808 grid cells follows that with 580,320
grid cells at each 3D and 5D downstream from the center
of stenosis. The difference of averaged WSS is within 5%
for a period of pulsatile in both cases. Thus, 434,808 hexa-
hedral grid cells is selected for numerical analysis. The
computing time for each case was about 8 hours with 8
nodes, 2.0 GHz CPU.
2.3. Boundary and initial conditionFor the numerical simulation, the 3-D time-dependent
Navier-Stokes equations were solved by the ANSYS CFX
V11.0 based on the finite volume method with the pres-
sure-based coupled solver. Fixed time step, 0.002 s, was
used with the Second Order Backward Euler scheme for
transient term. The flow is assumed to be a laminar flow,
incompressible, non-Newtonian, and the wall of artery is
∂ρ
∂t------ ∇ ρν( )⋅+ 0=
∂
∂t---- ρν( ) ∇ ρν ν⊗( )⋅+ =
pδ– µ ν ν∇( )T+∇( )+( ) ρω ω r×( )×–∇
τ ηγ· =
γ·
γ·
γ·1
2--- γij
·γji·
j
∑i
∑=
η η∞ η0 η∞–( ) 1 λγ·( )2
+[ ]n 1–( ) 2⁄
+=
η∞
R z( ) R a R 1 π z z1–( ) z0⁄cos+[ ]⋅–[ ]=
Fig. 1. The schematic view and grid generation of stenosed blood
vessel.
Numerical investigation on the blood flow characteristics considering the axial rotation in stenosed artery
Korea-Australia Rheology Journal June 2009 Vol. 21, No. 2 121
rigid with no slip conditions.
For the study of the effect of the pulsatile flow on the
blood flow characteristics in an artery, the idealized pul-
satile velocity profile of the common carotid artery is used
as the inlet boundary condition (Gijsen et al., 1999). Fig. 3
shows the pulsatile velocity profile which is dimensionless
by the period of pulsatile cycle, tp, which is 1 s. The time-
averaged dynamic viscosity for a period of pulsatile is
0.007865 . The peak of velocity is 0.20 m/s at inlet in
pulsatile and the density of blood is 1100 kg/m3. The pres-
sure boundary condition is used in the outlet of the artery.
The initial velocity through whole domain is equal to the
start of systole, t/tp =0, in pulsatile. For the study on the
effect of rotation on the blood flow characteristics, we use
a MRF (Multiple Reference Frame) method for application
to the rotating effect of blood vessel (Luo et al., 1994).
Through the MRF method, the only wall is rotated.
3. Results and discussion
For the validation of our numerical method, the numer-
ical results are compared with the experiment in axially
rotating pipe (Imao et al., 1992). The angular velocity of an
axially rotating artery is varied as 1, 2, 4 and 6 revolutions
per a second (rev/s), and the results are compared with
those of the no-rotation case.
We compare the axial velocity profiles of Newtonian
fluid flow to that of non-Newtonian fluid flow for 6 rev/s.
Consequently, the Newtonian fluid flow is more unstable
than another due to the magnitude of viscosity. Overall, the
viscosity of non-Newtonian fluid is larger than that of
Newtonian fluid. For unsteady flow, the simulations are
executed over at least three cycles to achieve a periodic
solution. The velocity variation after two cycles is less than
1% at test points behind the stenosis and results are saved
for the final cycle. In this section, the blood flow char-
acteristics such as axial velocity profiles, pressure distri-
bution, flow recirculation zone (FRZ) and wall shear stress
(WSS) distribution are presented. The blood flow char-
acteristics are obtained for the entire flow domain at four
different instants (t / tp = 0.14, 0.16, 0.40, 0.78) in pulsatile.
3.1. ValidationsDue to the instability problem of flow in the experiment
Pa s⋅
Fig. 2. Averaged axial WSS on circumferential lines at two dif-
ferent locations from the stenosis in pulsatile for various
number of grid cells. ((a) At 3D downstream from the
stenosis, (b) At 5D downstream from the stenosis).
Fig. 3. The pulsatile inlet velocity profile.
Kun Hyuk Sung, Kyoung Chul Ro and Hong Sun Ryou
122 Korea-Australia Rheology Journal
(Imao et al., 1992), the spiral wave appears as the ratio of
circumferential velocity to the axial velocity increases and
it is clear when the rotation rate is 3 in axially rotating pipe.
Thus, we have been performed the numerical analysis
when the rotation rate is 3. In the experiment, the Reynolds
number is 500 and the ratio of circumferential velocity to
the axial velocity is 3. The circumferential velocity of a
pipe and the flow rate of water are given as constant. Fig.
4(a) shows the circumferential velocity which is dimen-
sionless by the wall velocity and Fig. 4(b) shows the axial
velocity which is dimensionless by the inlet velocity in axi-
ally rotating pipe.
The numerical results follow the experimental results
except the axial velocity profile at Z=120. The profile of
measured axial velocity tends to become the turbulent at
Z=120. This reason is that the flow state changes from
laminar to turbulent by the rotation effect in experiment.
However, the circumferential velocity profile approaches
the solid-body rotation gradually as the numerical simu-
lation predicts.
3.2. Results of numerical analysisThe axial velocity profiles are presented in a plane con-
taining the axis of artery because the geometry is axi-
symmetric. Fig. 5(a), (b), (c) and (d) show axial velocity
profiles and FRZs when the artery is not rotating at dif-
ferent time phases in pulsatile. The contour of zero velocity
indicates the boundary of FRZ, in which the flow is either
stagnant or reversed. The jet velocity profile like a piston
shape appears behind the stenosis in flow acceleration
phase. Due to the decrease of the area at the stenosis and
reverse flow near the vessel wall, axial velocity of the center
line of artery increases in order to satisfy the flow rate con-
servation law at 1D downstream from the stenosis. The FRZ
appears at all time in pulsatile because the stenosis causes
disturbance to the flow and its size varies due to the change
Fig. 4. Dimensionless velocity profile ((a) The dimesionless cir-
cumferential velocity, (b) The dimensionless axial velocity).
Fig. 5. Axial velocity profiles and flow recirculation zone at dif-
ferent time phase in pulsatile. The position of profile is
presented as the multiple of diameter distal to stenosis.
((a) No rotation model, t/tp=0.14, (b) No rotation model,
t/tp=0.16, (c) No rotation model, t/tp=0.40, (d) No rotation
model, t/tp=0.78, (e) 6 rev/s model, t/tp=0.40, (f) 6 rev/s
model, t/tp=0.78).
Numerical investigation on the blood flow characteristics considering the axial rotation in stenosed artery
Korea-Australia Rheology Journal June 2009 Vol. 21, No. 2 123
of magnitude of velocity in pulsatile. The FRZ grows and
expands to the downstream of stenosis as flow accelerates
and extends up to the whole downstream at t / tp=0.4. After
that time, FRZ contracts rapidly and its size becomes con-
stant because the gradient of velocity is zero in pulsatile.
Due to the centrifugal force by rotating of vessel, the pres-
sure along axial line decreases but the pressure near wall
increases at the same time as shown in Fig. 6(a). Conse-
quently, the iso-pressure lines become more convex as the
angular velocity increases like Fig. 6(b). This causes the
friction increases and this phenomenon suppresses the heart
due to the increase of pressure drop at the rotating artery.
Fig. 5(e) and (f) show axial velocity profiles and FRZs
when the angular velocity is 6 rev/s. In comparison with the
no-rotation case, the FRZs are reduced remarkably due to
the increase of radial velocity from the centrifugal force at
t/tp =0.4. Also, axial velocity profiles and FRZs are asym-
metry although the geometry of artery is axisymmetry.
Fig. 7(a) shows the axial velocity contours when the
angular velocity is 6 rev/s at four different time phase in
pulsatile. Asymmetric contour with three or four protru-
sions appears at t/tp=0.4 and possesses the eccentricity . In
Fig. 6, the protrusions constitute the spiral wave as the
angular velocity is 4 rev/s. Fig. 7(b) shows axial velocity
contours as variation on the angular velocity at t/tp=0.40.
As the fluid flows to the downsteram, the four protrusions
of contour become dim at 6 rev/s. On the other hand, the
shape of four protrusions is maintained at 4 rev/s. The
asymmetric coutours appear after 3D downstream from the
stenosis at the whole rotating cases and after that location,
3D, the shape of protrusions rotates slowly.
Imao et al. (1992) investigated the structure and char-
acteristic of a spiral wave with the flow visualization tech-
nique in axially rotating pipe. In experimental study, the
spiral wave occurs as the variation of rotation rate which is
the ratio of circumferential velocity to the axial velocity.
As the rotation rate is 3, the spiral wave is most amplified
and is reduced with a greater rotation rate.
Fig. 8 shows the Iso-contour of the axial velocity at four
different instants in pulsatile. As the angular velocity is
4 rev/s, the spiral wave occurs at t/tp=0.4. The spiral wave
appears and disappears as the variation of rotation rate due
to the pulsatile inlet velocity. The stenosis affects the sta-
bility of flow in the artery. Tang et al. (1999) showed the
asymmetric flow patterns occur and become unstable in an
axisymmetric geometry with the stenosis. Buchanan et al.
(1998) showed in their numerical study that two co-rotat-
ing vortices occurred in a 75% (area reduction) axisym-
metrical stenosed model when the flow started to decel-
erate in pulsatile. In our numerical analysis, the spiral wave
and asymmetric flow occur clearly although the rotation rate
is smaller than 3. It is caused by the flow is influenced by
Fig. 6. Pressure distribution at t/tp=0.40 ((a) Pressure distribution
along the axial line at t/tp=0.40, (b) Pressure contour at yz-
plane as x=0).
Fig. 7. The Axial velocity contours ((a) Axial velocity contours
when the angular velocity is 6 rev/s at four different time
phase in pulsatile, (b) Axial velocity contours as variation
on the angular velocity at t/tp=0.40).
Kun Hyuk Sung, Kyoung Chul Ro and Hong Sun Ryou
124 Korea-Australia Rheology Journal
the effects of the rotation and the stenosis at the same time.
Variations of averaged axial WSS on circumferential
lines are shown in Fig. 9 at four different downstream loca-
tions through the whole pulsatile. Considering for the
asymmetric flow due to the rotation and stenosis of artery,
WSS are averaged on circumferential line of each location.
The intersections of WSS curves and the horizontal axis
are two points at which WSS changes signs and hence cor-
respond to the flow separation point (WSS changes from
positive to negative) and reattachment point (WSS changes
from negative to positive). At 1D downstream from the
stenosis, there is no intersection point and the sign of WSS
is negative all the time. Those indicate FRZ occurs through
the whole cycle. After 3D downstream from the stenosis, it
is notable that the difference of negative WSS caused by
the reverse flow. The WSS of no-rotation case is about
40% and nearly twice smaller than that of rotating case due
to the stronger reverse flow at 5D and 7D downstream
from the center of stenosis, respectively. The centrifugal
force caused by rotation effect suppresses the blood flow
toward wall and this pheonomenon decreases the reverse
Fig. 8. ISO-contour of the axial velocity which is 0.065 m/s when
the angular velocity is 4 rev/s.
Fig. 9. Averaged axial WSS on circumferential lines at four different downstream locations in pulsatile ((a) 1D downstream from the
stenosis, (b) 3D downstream from the stenosis, (c) 5D downstream from the stenosis, (d) 7D downstream from the stenosis).
Numerical investigation on the blood flow characteristics considering the axial rotation in stenosed artery
Korea-Australia Rheology Journal June 2009 Vol. 21, No. 2 125
flow and the FRZs. But, the WSS of rotation is larger after
the FRZs disappear because the circumferential gradient of
velocity is producted by the rotation of artery. This causes
the new vascular disorder by injuring endothelium of artery
such as arteroscelrosis at the downstream of stenosis (Fry,
1972). Also, when the artery is rotating, the existence time
of FRZs shortens further from the stenosis. In comparison
with no-rotation case, the existence time is smaller about
12.5% at 7D downstream from the center of stenosis.
4. Conclusion
In this paper, the rotation effects has been studied numer-
ically on a stenosed blood vessel with the pulsatile inlet
velocity profile.
The FRZs occur through the whole pulsatile because the
stenosis causes disturbance to the flow and its size varies
in pulsatile. In comparison with the no-rotation case, the
size and existence time of FRZs are reduced remarkably
due to the increase of radial velocity from the centrifugal
force in axially rotating artery. But, the friction increases
due to the increase of pressure by the centrifugal force.
This phenomenon suppresses the heart due to the increase
of pressure drop at the rotating artery.
Also, the WSS of no-rotation case is about 40% and
nearly twice smaller than that of rotating case due to the
stronger reverse flow at 5D and 7D downstream from the
stenosis, respectively. But the WSS of rotation is larger after
the FRZs disappear because the circumferential gradient of
velocity is producted by the rotation of artery. This causes
the new vascular disorder by injuring endothelium of artery
such as arteroscelrosis at the downstream of stenosis.
Although the geometry of artery is axisymmetry, the spi-
ral wave and asymmetric flow occur clearly in spite of the
small rotation rate. The contour of axial velocity is asym-
metric with four or three protrusions at t/tp=0.4 and pos-
sesses the eccentricity after 3D downstream from the
stenosis at the whole rotating cases. And it is caused that
the flow is influenced by the effects of the rotation and the
stenosis at the same time.
Acknowledgement
This research was partially supported by the Chung-Ang
University Grant in 2009.
References
Belardinelli, E., M. Ursino and E. Iemmi, 1989, A preliminary
theoretical study of arterial pressure perturbations under shock
acceleration, J. Biomech. Eng.-Trans. ASME 111, 233-240.
Buchanan Jr, J. and C. Kleinstreuer, 1998, Simulation of particle-
hemodynamics in a partially occluded artery segment with
implications to the initiation of microemboli and secondary
stenoses, J. Biomech. Eng.-Trans. ASME 120, 446-454.
Burton, R. R., S. D. Leverett Jr and E. D. Michaelson, 1974, Man
at high sustained +G(z) acceleration: a review, AEROSPACE
MED. 45, 1115-1136.
Chien, S., 1982, Hemorheology in clinical medicine, Clin.
Hemorheol. 2, 137-142.
Cho, Y. I., L. H. Back and D. W. Crawford, 1985, Experimental
investigation of branch flow ratio, angle, and Reynolds number
effects on the pressure and flow fields in arterial branch mod-
els, J. Biomech. Eng.-Trans. ASME 107, 257-267.
Deplano, V. and M. Siouffi, 1999, Experimental and numerical
study of pulsatile flows through stenosis: Wall shear stress
analysis, J. Biomech. 32, 1081-1090.
Fry, D., 1972, Response of the arterial wall to certain physical
factors. Atherogenesis: Initiating factors, A Ciba Foundation
Symp., 40-43.
Gijsen, F. J. H., E. Allanic, F. N. Van De Vosse and J. D. Janssen,
1999, The influence of the non-Newtonian properties of blood
on the flow in large arteries: Unsteady flow in a 90°curved
tube, J. Biomech. 32, 705-713.
Hooks, L., R. Nerem and T. Benson, 1972, A momentum integral
solution for pulsatile flow in a rigid tube with and without lon-
gitudinal vibration, Int. J. Eng. Sci. 10, 989-1007.
Imao, S., M. Itoh, Y. Yamada and Q. Zhang, 1992, The char-
acteristics of spiral waves in an axially rotating pipe, Exp. Fluids
List of symbols
R [m] radius of blood vessel
D [m] the maximum artery diameter
t [s] simulation time
tp [s] a period of pulsatile
z0 [m] half the length of the entire stenosis region, 1D
z1 [m]the axial position from the starting point of the
stenosis region
a [-] the rate of stenosis
v [-] dimensionless circumferential velocity, v’/R ω
r [-] dimensionless radial distance, r’/R
w [-] dimensionless axial velocity
ρ [kg/m3] density
[m/s] blood velocity tensor
η [Pa s] apparent viscosity
[1/s] shear rate
[rev/s] angular velocity
Subscripts
0 zero shear, amplitude
infinite shear
Superscripts
( )’ Dimension value
ν
γ·
ω
∞
Kun Hyuk Sung, Kyoung Chul Ro and Hong Sun Ryou
126 Korea-Australia Rheology Journal
12, 277-285.
Kikuyama, K., M. Murakami, K. Nishibori and K. Maeda, 1983,
Flow in an Axially Rotating Pipe: A calculation of flow in the
saturated region, B. JSME. 26, 506-513.
Luo, J. Y., R. I. Issa and A. D. Gosman, 1994, Prediction of
impeller-induced flows in mixing vessels using multiple
frames of reference, IChemE Symp. Ser. 136, 549-556.
Luo, X. and Z. Kuang, 1992, Non-Newtonian flow patterns asso-
ciated with an arterial stenosis, J. Biomech. Eng.-Trans. ASME
114, 512-514.
Mandal, P. K., S. Chakravarty, A. Mandal and N. Amin, 2007,
Effect of body acceleration on unsteady pulsatile flow of non-
newtonian fluid through a stenosed artery, Appl. Math. Com-
put. 189, 766-779.
Misra, J. C. and B. K. Sahu, 1988, Flow through blood vessels
under the action of a periodic acceleration field. A mathe-
matical analysis, J. Comput. Appl. Math. 16, 993-1016.
Nakamura, M. and T. Sawada, 1988, Numerical study on the
flow of a non-Newtonian fluid through an axisymmetric steno-
sis, J. Biomech. Eng.-Trans. ASME 110, 137-143.
Ro, K. C., S. H. Lee, S. W. Cho and H. S. Ryou, 2008, Numerical
Study on Blood Flow Characteristics of the Stenosed Blood
Vessel with Periodic Acceleration and Rotating Effect,
Springer Proceedings in Physics Series 124, 77-83.
Tang, D., C. Yang and D. N. Ku, 1999, A 3-D thin-wall model
with fluid-structure interactions for blood flow in carotid arter-
ies with symmetric and asymmetric stenoses, Comput. Struct.
72, 357-377.
Young, D. F., 1968, Effect of a time dependent stenosis on flow
through a tube, J. Engng. Ind. Trans. ASME 90, 248-254.
Young, D. F., 1979, Fluid mechanics of arterial stenoses, J. Bio-
mech. Eng.-Trans. ASME 101, 157-175.